Development of an Automated Method for Monoclonal Antibody Purification and Analysis. Introduction. Gurmil Gendeh, 1 Wim Decrop, 2 and Remco Swart 2 1



Similar documents
A Complete Solution for Method Linearity in HPLC and UHPLC

Monoclonal Antibody Characterization Achieving Higher Throughput and Productivity

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide

Monitoring Protein PEGylation with Ion Exchange Chromatography

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Thermo Scientific Dionex AS-AP, AS-DV, and AS-HV Autosamplers. Automation, sample preparation. Autosamplers for Ion Chromatography

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary

What is Eluent Generation?

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Optimize Your Process Operations by Improving Process Monitoring

Affi-Prep Protein A Matrix Instruction Manual

Determination of Calcium and Magnesium in Brine

Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Continuous Chromatography for Monoclonal Antibody Purification from Cell Culture Supernatant

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)

Purification of reaction mixtures using flash chromatography.

INTRODUCTION. The principles of which are to:

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

How To Test For Contamination In Large Volume Water

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins

HiPer Ion Exchange Chromatography Teaching Kit

Concanavalin A Column for Analysis of Glycoproteins and Their Tryptic Glycopeptides

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry.

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Environmental Analysis Using the ICS-3000 Ion Chromatography System

Automated Nucleic Acid Extraction WorkStation Faster, cleaner,

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Thermo Scientific Dionex UltiMate 3000 Variable Wavelength Detectors

Using the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes

EQuan MAX Getting Connected Field Service Guide

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Structural Analysis of Labeled N-Glycans from Proteins by LC-MS/MS Separated Using a Novel Mixed-Mode Stationary Phase

Amino Acid Analyzer L-8900

Application Note. Determination of Amino acids by UHPLC with automated OPA- Derivatization by the Autosampler. Summary. Fig. 1.

Thermo Scientific Vanquish UHPLC System. Revolutionize Your. UHPLC Experience

POROS CaptureSelect affinity columns for rapid, small-scale purification and sample preparation of recombinant proteins

HiTrap Heparin HP, 1 ml and 5 ml

Smartline HPLC by design

Process-scale purification of monoclonal antibodies polishing using Capto Q

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Gel Filtration Standard

Efficient Multi-Well Protein Purification Strategies

Innovative vs Traditional

Data File. Sephadex G-25 media and pre-packed columns. Introduction. Sephadex G-25 Bead structure. Desalting/buffer exchange and gel filtration

Protocol: HPLC (amino acids)

Choose your optimal tools for protein studies

Selectivity and Detectability Optimizations in HPLC

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

Recent advances in the purification of IgM monoclonal antibodies

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

ab Protein Sumoylation Assay Ultra Kit

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

Earning Customer Loyalty

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

INSTRUCTIONS Edition AC

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software

ChromQuest 5.0 Chromatography Data System

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

3.3. NGC Chromatography Systems and ChromLab Software Instrument Guide Version 3.3

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

The Theory of HPLC. Gradient HPLC

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

FPLC standard operating procedure AKTA system (GE)

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

Transcription:

Development of an utomated Method for Monoclonal ntibody Purification and nalysis Gurmil Gendeh, 1 Wim Decrop, and Remco Swart 1 Thermo Fisher Scientific, Sunnyvale, C, US; Thermo Fisher Scientific, msterdam, The Netherlands Introduction During various stages in the development of biopharmaceuticals, purification and analytical characterization of the product are required. In the early phase of recombinant antibody process optimization, a large number of samples must be screened for titer, aggregation, and charge variants. Two variations of major importance occurring in monoclonal antibody (Mb) biopharmaceuticals are the presence of variants of the target molecule and their aggregation. These product-related substances can have different efficacies than the main product and may cause serious side effects, such as antidrugantibody formation. Protein aggregates are mostly the consequence of suboptimal production, purification, or handling conditions (e.g., temperature, ph). In the purification of antibodies, a protein-affinity separation is generally the first step. ffinity chromatography on protein or G columns typically yields a purity of more than 95% in a single step. lthough the purification on affinity columns yields information on the titer of the product, it is not selective with respect to related substances. To verify the purification efficiency, sample purity, or antibody quality, a technique such as ion-exchange chromataography (IEC) or size exclusion chromatography (SEC) is required. The SEC technique provides the necessary selectivity to identify agglomerates and size-based variations of the main component. Ion-exchange stationary phases provide good selectivity for separation of charge variants of the protein biopharmaceutical. The variations may be very subtle or small, and finding the optimal chromatographic conditions requires optimization. The work shown here discusses the development of an automated and unattended solution for purification and separation (by IEC and SEC) of monoclonal antibodies using a single HPLC system (Figure 1). In this twodimensional (D) process, multiple protein separation steps are performed automatically. The autosampler of the HPLC system was configured to perform the injection, fraction collection, and reinjection of the collected fractions. The intelligent program capabilities of the Thermo Scientific Dionex Chromeleon Chromatography Data System (CDS) software take care of all of the automation. FIGURE 1. () The Thermo Scientific Dionex UltiMate 3 Titanium solution with () the Thermo Scientific Dionex WPS-3T()FC bioinert autosampler/ fractionation module.

Instrumentation and D-LC Workflow HPLC experiments were carried out using a Dionex UltiMate 3 Titanium System equipped with: SRD-36 Membrane Degasser DGP-36M Dual-Gradient Pump System TCC-3SD Thermostatted Column Compartment with valve actuators and biocompatible pods WPS-3T()FC nalytical Dual-Valve Well Plate Sampler (Figure 1) VWD-34RS UV Detector equipped with a.5 µl flow cell The Chromeleon CDS extended fractionation software driver is added in the server configuration to control the advanced features of the Dionex WPS-3T()FC. This software extension is capable of managing timeand peak-based fractionation, including a wide range of settable parameters to adapt the fractionation to the specific needs of the analysis. Fractionation can be operated in continuous or pooled mode, as used in enrichment experiments. Further, this extension enables Chromeleon CDS postacquisition steps, which allows fractions to be reprocessed by another chromatographic method. The WPS-3T()FC This dual-valve well plate autosampler is capable of fractionating into various types of vials and well plates. It can operate at elevated flow rates by using the second valve as a diverter valve. The dual-needle design with a metal prepuncture needle allows fractionation in both capped and uncapped vials and well plates, as well as Eppendorf tubes. oth puncture depth and sample needle protrusion depth can be programmed for each sample container format. In addition, the WPS-3T() FC also offers sample handling and derivatization capabilities (e.g., in-well digestion, ph adjustment, dilution). It has a thermostatted sample compartment (4 45 C) and easy-to-adapt fluidics to suit the application needs, including full-loop injection volume ranges from 1 to 5 µl. It also accommodates multiple well plate types simultaneously. Finally, the module s excellent repeatability for injection and fractionation provides a robust platform for any analysis. FIGURE. Fluidic configuration of the automated offline D-LC system using the WPS-3T()FC well plate autosampler with fractionation. TCC Waste Pump 1 Protein Pump UV OUT IN The workflow and LC conditions for automated offline D-LC includes the following: Injection of 5 5 µl of an unpurified monoclonal/ polyclonal antibody sample first-dimension ( 1 D) affinity chromatography separation at a flow rate of. ml/min using the following steps: column wash/equilibration step of.75 min n elution step of 1 min utomated peak detection followed by fraction collection into a well plate in the autosampler The protein is reconditioned for the next analysis The total analysis time: approximately 3 min second-dimension ( D) separation (one of the following): Strong cation-exchange separation, applying a linear NaCl salt gradient ( mm MES buffer, ph 5.6) SEC separation ( mm sodium phosphate, 5 mm NaCl, ph 6.8) Prior to the D separation, add a neutralization buffer (Tris,.1 M) to each fraction to raise its ph, using the derivatization capabilities of the autosampler. SEC IEC 794 Development of an utomated Method for Monoclonal ntibody Purification and nalysis

LC Conditions Protein ffinity Separation Used pplied iosystems Poros protein column ( µm, 4.6 mm i.d. 5 mm,.8 ml) for 1 D affinity-based separations with UV detection at 14 and 8 nm. pply between 5 and 5 µl of sample, making use of the large-volume injection kit and a user-defined program (UDP) to fill the sample loop. ffinity LC Conditions Mobile Phase : Mobile Phase : 1 mm Sodium phosphate, 15 mm NaCl, ph 7. 5 mm Glycine-HCl, 15 mm NaCl, ph.8 1 D Gradient: Wash and equilibration step for.75 min at 1%, followed by a 1 min elution step at 1% Flow Rate:. ml/min Temperature: 5 C Ion-Exchange Separation of Protein Variants Used Thermo Scientific MbPac SCX-1 column (4 mm i.d. 5 mm) for the D weak cation-exchange LC separation applying a linear salt gradient of MES buffer/ sodium-chloride solution, detection at 14 and 8 nm. Mobile Phase : mm MES buffer, ph 5.6 + 6 mm NaCl Mobile Phase : mm MES buffer, ph 5.5 + 3 mm NaCl D Gradient: From 15 to 45% in 3 min; 6 min wash step at 1% ; and 1 min equilibration time at 15% Flow Rate: 1. ml/min Temperature: 3 C Size-Exclusion Separation of Protein ggregates Used Thermo Scientific MbPac SEC-1 column for the aggregation analysis (4 mm i.d. 3 mm). The mobile phase is mm sodium phosphate, 5 mm NaCl, ph 6.8. Results Typically, the antibody is manually injected onto the affinity column. Here, the autosampler was used to introduce the sample to this column. fter an equilibration step with mobile phase, the antibody was eluted from the Protein column with mobile phase and fractionated into a 96-deep well plate using UV-based peak triggers (Figure 3). These triggers can be optimized easily to either collect or discard the breakthrough peak and to collect the antibody peak with a trigger based on retention time and signal properties. minimum fractionation/decision delay time of 5 s should be considered for peak-based fractionation. FIGURE 3. 1 D affinity chromatography separation of a Mb. The green labeled part of the chromatogram denotes the peak-based fractionation., 1. R1 - Mb - 1.547 % aff ph.8:. % 1 -.333. Flow: µl/min..5 1. 1.5..5 3. 795 Fractions can be visually and numerically reported in the Chromeleon CDS software. The fraction volume can be extracted and transferred automatically to the D sequences to enable numerous derivatizations. neutralizing buffer of 15% of the fraction volume was added to increase the fractions ph, using the derivatization capabilities of the autosampler (Figure 4). When a fraction is found that meets the requirements (e.g., titer/peak area), the Chromeleon CDS software can be used to initiate a postacquisition step for the D separation(s). Examples of D separations are: a strong cation-exchange-based separation yielding selectivity towards small charge differences in the antibody sample (Figure 4, IEC of a monoclonal IgG1 fraction); or an SEC to separate the antibody from its dimers or aggregates (Figure 4, SEC of a Mb). The ion-exchange chromatogram clearly shows the heavy chain C-terminal lysine heterogeneity variants of the Mb described by the three largest peaks. series of basic and acidic variants are also clearly visible. more detailed look into monitoring Mb heterogeneity and stability by IEC can be found in N 17 and N 18 1, and the MbPac column data sheet. 3 In the applied workflow all 1 D separations are carried out, followed by the automatically generated D separation sequence. The sample throughput can be further increased by accelerating the flow rate for the affinity and the SEC separations and shortening the different steps required in the affinity separation. The IEC separation also may be accelerated by modifying the gradient. 3

FIGURE 4. () Example of a D IEC separation (MbPac SCX-1, 5 4.6 mm i.d.) of a purified Mb fraction. () Example of a D SEC separation of a purified Mb fraction. FIGURE 5. () Example of a 15 min total analysis time D IEC separation (MbPac SCX-1, 15 4 mm i.d.) of a purified Mb fraction. () Example of a D SEC separation of a purified Mb fraction. (MbPac SEC-1, 15 4 mm i.d.) 1 K cidic Variants KK asic Variants 5 1 15 5 15 1 3 4 5 6 7 8 9 9 4 6 8 1 1 14 796 cceleration of nalysis Depending on the requirements for the separation, several options are available. Utilizing short columns allows high-throughput separations. Emerging techniques such as ph-gradient-based IEC are good alternatives to accelerate the screening process. 4,5 Examples of fast IEC as well as SEC are depicted in Figures 5 and 5. 1 3 4 5 6 7 8 8145 4 Development of an utomated Method for Monoclonal ntibody Purification and nalysis

Conclusions The dual-valve WPS-3T()FC autosampler enables high-flow fraction collection and reinjection of fractions as used in D workflows. Such workflows include automated offline D column coupling (e.g., affinity with IEC and SEC; RP with RP utilizing different ph levels; SPE with RPLC); as well as IEC with fractionation, trapping, and desalting prior to mass spectrometric detection. It is clear that this module can serve as a platform for coupling various chromatographic techniques for biopharmaceuticals and proteins analysis. The Chromeleon software package allows this high degree of automation with a wide range of flexibility to accommodate the most demanding applications. Sample handling and derivatization, purification, and HPLC analysis can be programmed both easily and in depth and gives the instrument unique capabilities in the field of sample purification, standard HPLC, and QC. References 1. Dionex Corporation. nalysis of Monoclonal ntibody Heterogeneity by Cation-Exchange Chromatography: Separation of C-Terminal Lysine Variants. N 17, LPN147-1, 9, Sunnyvale, C.. Dionex Corporation. Monitoring Monoclonal ntibody Stability by Cation-Exchange Chromatography. N 18, LPN15-1, 7, Sunnyvale, C. 3. Dionex Corporation.The MbPac SCX-1 Column for Monoclonal ntibody Variant nalysis. Datasheet, LPN567-3, 7, Sunnyvale, C. 4. Farnan, D.; Moreno, G. T. Multiproduct High-Resolution Monoclonal ntibody Charge Variant Separations by ph Gradient Ion-Exchange Chromatography. nal. Chem. 9, 81 (1), 8846 88575. 5. Rea, J. C.; Moreno, T.; Lou, Y; Farnan, D. Validation of a ph Gradient-ased Ion-Exchange Chromatography Method for High-Resolution Monoclonal ntibody Charge Variant Separations. J. Pharm. iomed. nal., 11, 54 (), 317-33. 11 Thermo Fisher Scientific, Inc. ll trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Dionex Products: 18 Titan Way, PO ox 363, Sunnyvale, C 9488-363, (48) 737-7 North merica: U.S./Canada (847) 95-75 South merica: razil (55) 11 3731 514 Europe: ustria (43) 616 51 5, enelux (31) 683 9768 (3) 3 353 494 Denmark (45) 36 36 9 9, France (33) 1 39 3 1 1, Germany (49) 6115 991 Ireland (353) 644 64, Italy (39) 51 6 167, Sweden (46) 8 473 338, Switzerland (41) 6 5 9966, United Kingdom (44) 176 6917 sia Pacific: ustralia (61) 94 533, China (85) 48 38, India (91) 764 735, Japan (81) 6885 113, Korea (8) 653 58, Singapore (65) 689 119, Taiwan (886) 875 6655 www.thermoscientific.com/dionex LPN 945-1 8/11