Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:

Fundamental Autoclave Techniques

Guidance on safe use of Autoclaves

Effective Use of Autoclaves

Steam Sterilization Cycles for Lab Applications

UCCS Biosafety Management Policy

Decontamination and Waste Management

A GUIDELINE FOR THE SAFE USE OF AUTOCLAVES UNIVERSITY OF OTTAWA. Environmental Health and Safety Service

University of Colorado at Boulder

Autoclave Quality Control Policy

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

A Guide to the Handling and Disposal of Medical Waste Contents

JABSOM KAKA AKO STANDARDIZED AUTOCLAVE PROTOCOL

Laboratory Biosafty In Molecular Biology and its levels

Scott & White Institutional Biosafety Committee Compliance Program Biohazardous Material Spill Clean-Up Procedure Policy #IBC.002

Infectious Waste Management Plan

The procedures of this policy are consistent with applicable sections of the OSHA Bloodborne Pathogens Standard 29 CFR

Lab Safety and Standard Operating Procedures. Faculty of Dentistry And School of Biomedical Engineering

CHAPTER V: DISPOSAL OF WASTES CONTAMINATED WITH INFECTIOUS AGENTS

UNIVERSITY OF RICHMOND REGULATED MEDICAL WASTE MANAGEMENT GUIDELINES

Biosafety Spill Response Guide

BSL 1 Laboratory Biosafety Manual

Lab Exercise 3: Media, incubation, and aseptic technique

Appendix J IBC Biohazard Spill Management Plan

Biological Safety Program

THE UNIVERSITY OF NEWCASTLE- SCHOOL of BIOMEDICAL SCIENCES

IX. Decontamination and Spills

Each has the following responsibilities under this SOP:

or

North Carolina State University Emergency Facilities Closure Checklist- Part I

Policies. Prep Room Policies

Biohazardous Waste Disposal. Table of Contents

b) Exhaust Selector and Sterilization Times c) Unloading the Autoclave Appendix 1 - Autoclave Logbook...28

UNH SAFETY AUDIT Office of Environmental Health and Safety

3 Attest Biological Monitoring System

Biosafety Level 2 (BSL-2) Safety Guidelines

Laboratory Line. Vertical and Bench-Top Life Science Autoclaves

Appendix H IBC Managing Biohazardous Waste SOP

2.3. The management in each HCF shall be responsible for ensuring good waste management practices in their premises.

Appendix H Managing Biohazardous Waste SOP

BOSTON COLLEGE ENVIRONMENTAL HEALTH AND SAFETY LABORATORY GUIDE FOR RENOVATIONS, REMODELS, MOVES AND TERMINATIONS

Biohazardous Waste Management Plan

April Biohazardous Waste Disposal Plan

LABORATORY SAFETY INSPECTION CHECKLIST

THE UNIVERSITY OF MAINE BIOMEDICAL WASTE MANAGEMENT PLAN

Biosafety Guidelines. 5. Pipetting Mechanical pipetting devices must be available and used. Mouth pipetting is prohibited.

Accurate Convenient Affordable Results in 24 Hours. Mail-to-Lab Spore Testing. Mail-to-Lab Spore Testing

STANDARD OPERATING PROCEDURES SPILL RESPONSE AND CLEAN-UP OUTSIDE BIOSAFETY CABINET

Laboratory Safety and Environmental Health Assessment Program

Policy for the Disposal of Biological Waste

OESO Ergonomics Division ERGO(3746) Duke Police Corporate Risk Management

Carnegie Mellon University s Policy and Procedures for Recombinant and Synthetic Nucleic Acid Materials Spills

3. It is recommended that the water supply into the facility can be obtained from two (2) separate water lines if possible.

RADIATION CONTROL TECHNIQUE #2 INSTRUCTIONS FOR PREPARATION OF RADIOACTIVE WASTE FOR DISPOSAL

"ADOPTED STANDARDS FOR THE REGULATION OF MEDICAL WASTE" IN HEALTH CARE FACILITIES LICENSED BY THE MISSISSIPPI STATE DEPARTMENT OF HEALTH

TARLETON STATE UNIVERSITY Biohazardous Waste Program

The University of Texas at San Antonio Office of Environmental Health, Safety and Risk Management. Part A. Biological Waste Management Safety Plan

2. Laboratory equipment

Conservation of Momentum Greg Kifer

Laboratory Safety Inspection Checklist (Biology and Biochemistry)

Federal Wage System Job Grading Standard For Laboratory Working, Table of Contents

Archived. Gloves should be changed frequently during the analysis.

INFECTIOUS WASTE MANAGEMENT PLAN West Virginia School of Osteopathic Medicine, Fredric W. Smith Science Building

Blood Collection and Processing SOP

INFECTIOUS WASTE MANAGEMENT PLAN Marshall University Robert C. Byrd Biotechnology Science Center

OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE

A Guide to Managing Your Biological Waste at the University at Albany

CHEMICAL SPILL MANAGEMENT GUIDELINES

Microbiological Evaluation of the STI Series 2000 Medical Waste Treatment Process

MANAGEMENT AND DISPOSAL OF BIOLOGICAL WASTE TEXAS A&M UNIVERSITY. July 2003

Nu G Medical Waste System Technology (Pyrolysis / Thermal Decomposition)

GUIDELINES FOR THE MANGAGEMENT OF BIO- MEDICAL WASTE AT THE SCHOOL OF LIFE SCIENCES

Biohazardous Waste Disposal

UCLA Henry Samueli School of Engineering and Applied Science STANDARD OPERATING PROCEDURE MEDICAL WASTE (BIOHAZARD WASTE) MANAGEMENT

ENVIRONMENTAL HEALTH AND SAFETY STANDARD OPERATING PROCEDURES

Body Art Facility Infection Prevention And Control Plan Guideline

Required Biosafety Level Three (BSL-3) Practices, Procedures, Facilities, and Safety Equipment For BSL-3 and BSL3/2 Laboratories

Procedures for the Effective Use of Biological Safety Cabinets

Biohazardous, Medical & Biological Waste Guidance Chart

University of Hawai i Biohazardous Waste Disposal Guidelines

Because You Can t See Sterile.SM

Response to Biological Spills in the Laboratory (Intentional or Accidental)

Administrative Procedure

3053 Electrical Safety Training Program Course Outline

Spill Kits and Incidental Spill Response

TEXAS TECH UNIVERSITY HAZARDOUS WASTE PROGRAM DEPARTMENT OF ENVIRONMENTAL HEALTH AND SAFETY

Packing for Transportation

University of Otago. Laboratory Biohazard Waste Disposal. Guidelines

Bloodborne Pathogens Program Revised July,

Sterilization methods and equipment Lab 1-2

Effective Heat Sterilization in CO 2 Incubators

MANAGEMENT AND DISPOSAL OF BIOLOGICAL WASTE AT TEXAS A&M INTERNATIONAL UNIVERSITY

BIOLOGICAL WASTE PROGRAM

BIOLOGICAL SAFETY CABINET PROGRAM

Biomedical. Waste Management Guide

Transcription:

Please post or circulate Fact heet nvironment, ealth and afety Information for the Berkeley Campus No. 33 Revised 04/04/11 afe and ffective Use of Autoclaves Autoclaves are easy to use but can pose a safety hazard if operated incorrectly. This fact sheet was developed to help prevent injuries and ensure that biological wastes are effectively decontaminated prior to disposal. Please refer to the Fact heet Risk Group 1 Recombinant DNA Waste Disposal and your Biological Use Authorization (BUA) for detailed information on the types of biological wastes that must be autoclaved, as well as packaging and other waste handling requirements. What are autoclaves? Factors for effective sterilization Autoclaves use water, pressure and heat to create superheated steam that kills microorganisms and spores. They are used to decontaminate certain biological waste and sterilize media, instruments and lab ware. team sterilization is mainly a function of temperature, pressure and time. Temperature: ffective sterilization occurs when the steam temperature exceeds 250 F (121 C). Pressure: Autoclave pressurization should be at least 20 pounds per square inch (psi). Time: The amount of time needed to sterilize most organisms is dependent upon the temperature and pressure. At 250 F (121 C) in a vessel pressurized to 20 psi, bags require at least 30 minutes to sterilize. The autoclave s superheated steam must contact all areas of the load. Bottle caps should be loose during the autoclave cycle so vapor expanding during heating will not cause an explosion. Liquid materials such as media or saline in bottles should occupy no more than half of the autoclave chamber volume; this will permit steam to circulate freely around the load. Do not overfill autoclave bags and tie them loosely to allow steam to reach the center of bag contents. Larger loads and tightly packed materials will require more time. Dry heat cycle - when to use Autoclave safety considerations Office of nvironment, ealth afety University of California 317 University all #1150 Berkeley, CA 94720-1150 http://www.ehs.berkeley.edu (510) 642-3073 The autoclave may offer dry heat cycles. These can be utilized for sterilizing laboratory supplies, which can withstand high temperatures but would be affected by a steam cycle. The required exposure times for dry heat will depend on the material composition, load volume, and can require significantly more time than what would be required for steam sterilization at the same temperature. Because the required times for dry heat sterilization will vary on the material treated and the surface and content, it is best to vary initial loads to determine the most effective temperature and length of cycle. Never autoclave flammable or volatile liquids, because these liquids could off gas and explode in the superheated chamber. Firmly lock autoclave doors and their gaskets in place before operating the autoclave to prevent a sudden release of high-pressure steam. Most, but not all, autoclaves have safety interlocks that prevent the autoclave from running if the door isn t closed properly.

Fact heet: afe and ffective Use of Autoclaves Page 2 of 3 Autoclave safety considerations Wear personal protection equipment: o Lab coat o ye protection o Closed-toe shoes o eat-resistant gloves to remove items, especially hot glassware Wait for the pressure gauge to drop to zero with zero time remaining before opening the door. Never open an autoclave set for slow exhaust until the cycle is complete. uperheated liquids can boil over and damage the autoclave and injure the operator. Never use the fast exhaust cycle for liquids as they will evaporate. Open the door cautiously. tand behind the door and slowly open it. Allow 30 seconds to allow steam to escape before reaching inside. Let liquids stand another 10 20 minutes after the autoclave is opened to avoid any movement that could cause them to boil over, resulting in scalding. Remove all items carefully. Report any malfunctions or accidents immediately to your supervisor. Notify the facilities manager if PP-C or the vendor may be needed to repair any autoclave. Autoclave bags and equipment Autoclave trays made out of polypropylene or polycarbonate are used as a secondary container; waste bags touching the internal chamber may melt and cause costly repairs. tainless steel tubs are an alternative to polypropylene. It is a safety hazard to use trays or tubs that are cracked, warped or compromised. Clear, autoclavable bags are used for recombinant DNA and/or BL 1 waste. The bags cannot have biohazard symbols or wording. Once clear bags have gone through the cycle, they may be disposed of as regular trash. Red, autoclavable, Biohazard bags are used for BL3 waste. Autoclave all BL3 waste and place in the tubs provided by. Bottles - avoid autoclaving plastic bottles unless the plastic can withstand the autoclave cycle. Temperature indicators may be used to determine if the desired temperature was attained. Autoclave tape and temperature - indicating labels are attached to the outside of the bag and change color or appearance when the desired temperature is reached. These may be purchased from laboratory supply companies, although they are not required. To ensure the entire contents of bagged waste are decontaminated, a test of the autoclave must be conducted every month. ow to determine the effectiveness of an autoclave Autoclave users are responsible for conducting a monthly test on the autoclaves and reporting any problems to the department s facilities manager. Purchase a test kit from your laboratory supply company which uses Geobacillus stearothermophilus, one of the most heat-tolerant species of bacteria. If sterilization in an autoclave does not destroy the Geobacillus spores, the device is not working properly. (A good biological indicator for dry heat cycles Office of nvironment, ealth afety University of California 317 University all #1150 Berkeley, CA 94720-1150 http://www.ehs.berkeley.edu (510) 642-3073

Fact heet: afe and ffective Use of Autoclaves Page 3 of 3 is Bacillus atrophaeus.) Read the manufacturers instructions prior to use. Check for expiration dates, storage, and other safety considerations. Procedures: 1. Place the spore strip or ampoule in the center of the load. A string or wire is used to retrieve the spore indicator after the run. ecure the strip or ampoule with autoclave-tape and a length of cotton string with the string extending out of the bag opening (some indicators are manufactured with an attached string). Remove the ampoule after the cycle is completed 2. Place the autoclaved test vial and a non-autoclaved control vial in an incubator, and incubate at 133 to 140 F (56 to 60 C). If you don t have an incubator, buy a spore test that you can ship to the manufacturer for incubation. pore strips typically require incubation for up to seven days after the run (follow manufacturer instructions). 3. Any bacterial growth indicates that the autoclave is not working properly. Do not continue to use an autoclave if it fails the spore test contact your facilities manager immediately. Disposal and Other Considerations. Used indicators with spore growth should be autoclaved again and disposed of as regular trash. Do not conduct tests in bags containing chemical disinfectants as these may kill the strip agent. Recordkeeping It is the responsibility of the autoclave users to establish a schedule for testing. The following autoclave records must be maintained for three years in the autoclave room, preferably in a binder accessible by or PP-C personnel: On-site maintenance records (remember that maintenance is based on hours of usage, rather than a chronological period) Autoclave use log (each load of material inactivated shall be logged as follows): * Date, time, and operator s name * Contact information: Laboratory, room number, phone number * Type of material sterilized * Temperature, pressure, and length of time the load is sterilized * The autoclave print-out, if the autoclave has a working printer or the cycle wheel when filled. The attached form Monthly Autoclave Test Log may be used to documenmaintenance and testing dates and results. Office of nvironment, ealth afety University of California 317 University all #1150 Berkeley, CA 94720-1150 http://www.ehs.berkeley.edu (510) 642-3073

Monthly Autoclave Test Log Building, Room # and Autoclave# Date Generator Name Time at 121C Max Tem p Max Press. Wt. in Lbs. Monthly sterilization Check Pass/Fail Comments Operator

Bacillus Ampoule Test Procedure, for Autoclave Monitoring The following testing will be conducted on a monthly basis and results will be recorded in an autoclave log. Log records to be kept for a minimum of three years. Initial Process: 1. Retrieve two spore ampoules (Bacillus tearothermophillus). 2. Identify the indicators (spore ampoules) by labeling them with the proper information (date, autoclave number, control, test, etc.). 3. Place the spore ampoules in a horizontal position with representative materials to be sterilized. Both ampules should be located in the part of the load that would be considered most questionable for sterilization. Remember to attach a piece of autoclave tape onto the load. 4. elect appropriate cycle to process the load (usually 30min for liquid or dry cycle). 5. Once the cycle has come to a finish allow autoclave to decompress and cool down (min. 10 min). 6. Remove load from autoclave and allow it to further cool down (10-15min). 7. Retrieve the spore test ampoules from the load. 8. It is common to see a color change from the spore ampules from purple to black; usually this distinguishes indicators that have been exposed to steam to those that have not. Incubation process: Use any incubator that is adjusted to a temperature between 55 0 C and 60 0 C. Take both indicator ampoules from the load that was run and gently squeeze the plastic crusher to break the glass ampoule and place those in the incubator. Take a third spore ampoule (label control ampoule) that was not run in the load proceed to break the glass ampoule and also incubate in the same incubator, allow incubation and verify all three spore ampoules after 24 hours. Interpretation: 1. xamine indicator ampoules for any color change after 24 hours. The presence of yellow color change indicates a fail test and positive bacterial growth. No color change indicates a passing test and proper sterilization of sample. The control ampoule should have a yellow color change. 2. If a fail test is confirmed run the test again and discontinue normal use of the autoclave for the time being. If the second test round fails again, inform all appropriate personnel as well as nvironmental ealth afety (Biosafety Officer ext. 3-6562). Discontinue full use of autoclave machine until approved for use. 3. If a passing test is confirmed from the first round or second round, you can proceed with regular use of the autoclave. 4. Record all results in your monthly autoclave log. 5. nsure to dispose of all used Bacillus ampoules by autoclaving first before discarding.