Laboratory Corrosion Testing of Medical Implants
|
|
|
- Cori Cross
- 9 years ago
- Views:
Transcription
1 Laboratory Corrosion Testing of Medical Implants Richard A. Corbett Corrosion Testing Laboratories, Inc. 60 Blue Hen Drive Newark, DE Copyright by ASM International
2 Laboratory Corrosion Testing of Medical Implants Richard A. Corbett Corrosion Testing Laboratories, Inc., Newark, Delaware, USA Abstract Performance evaluation of implantable devices is not new to medical device manufacturers, specifically implant manufacturers. A focus on corrosion resistance of implant devices has prompted members of the medical device industry, along with the United States Food and Drug Administration [USFDA] to set forth a guide for performance testing. Although actual corrosion resistance of a material can only be proven through long-term clinical trials, accelerated laboratory tests can be used to predict certain effects. Because corrosion of metals is an electrochemical process, accepted electrochemical testing techniques are used very effectively in the CPI (Chemical Process Industry) [namely ASTM G5, G61, and G71] have been adopted for use. Inconsistencies among laboratories, employing differing protocols, have resulted in varying interpretations by even the most experienced electrochemists and corrosion scientists. Some of the most widely used standard tests are also the most widely abused and misinterpreted, especially by persons who are less aware of materials properties. Since the USFDA provides no clear direction to testing, a modified electrochemical standard [ASTM F2129] has been adopted to address concerns regarding test environment, voltage scan rate, device configuration, and test protocol. The remaining issue is the establishment of an acceptance criterion. Introduction Implantable metal devices to treat diseases and repair the human body have been used for centuries. A gold plate was used to repair a cleft palate in In the 1600 s, iron, bronze and gold wire were used for sutures. 1 During the twentieth century Type 316 stainless steel was widely used as a prosthetic metal, and in recent years, the high-strength alloy MP35N, 2 and the shape-memory alloy Nitinol have been used in implantable device applications. 3 The most important requirement for an implantable material is that it must have corrosion resistance to physiological environments. All metallic implantable alloys are susceptible to corrosion to a certain degree, depending on the metallurgical condition, residual or service stresses, thermal history, and final surface treatment applied prior to implantation. It is the intent of this paper to prove our experience in the development and refining of laboratory tests to assist medical implant device manufacturers in evaluating their products for acceptance as being in the optimal corrosion resistant condition. There are a number of test tools available in achieving this goal. No one test is sufficient to provide all of the answers. Background Corrosion is defined as the deterioration of a material when exposed to an environment. For metals, this deterioration is an electrochemical process wherein a positive ion of the metal is released from the metal surface [identified as the anode] into the environment and subsequently an electron is released into the metal, where it combines with a positive dissolved ion [identified as the cathode]. Because of the flow of electrons [termed current] this process is an electrochemical reaction, and subsequently, electrical instruments, such as voltmeters and ammeters, can monitor this reaction. A metal achieves corrosion resistance by forming a protective oxide layer over its surface, somewhat akin to a barrier coating. This is termed passivation and is not synonymous with cleanliness. Cleaning a stainless steel in citric acid removes ferrous and ferric oxides but it does not passivate the surface, however exposing the stainless steel to nitric acid will rapidly form a protective chromium oxide layer that is passive. 4 This is addressed in ASTM F Testing a device to ASTM B will evaluate the presence of iron oxides and degree of cleanliness, but it will not determine the alloy s corrosion resistance. Development of ASTM Standards ASTM G 31 7 is an immersion test procedure that typically requires extended exposure periods, up to years, to evaluate the resistance of materials, which are employed in the implant device industry, to corrosion. As such, accelerating the testing by inducing changes on the material and monitoring the results has been an industry-accepted method.
3 The development of accelerated electrochemical tests to study corrosion began in earnest during the late 1960 s and into the 1980 s in the CPI. The original test standard, ASTM G 5, 8 was expanded from a stepped potentiostatic test to a potentiodynamic test as electronics developed, and subsequently to a cyclic experiment to examine susceptibility to localized pitting and crevice corrosion [ASTM G 61 9 ]. Galvanic studies between mixed metals in contact with each other were evaluated using the established ASTM G During the late 1970 s it was recognized that implantable alloys should not only be generally resistant to corrosion but also resistant to localized pitting and crevice attack. Research performed by Syrett 11 at the Stanford Research Institute identified a method of assessing the pitting resistance of alloys using passive-pitting-repassivation (PPR) curves. This method was modified by members of ASTM Subcommittee F04.02 to create the ASTM Standard F This test was designed to determine an inter-laboratory ranking of alloys, and not devices, to increasing resistance to pitting and crevice corrosion under specific conditions. The test conditions were intentionally designed to cause breakdown of Type 316L stainless steel, then considered acceptable for surgical implant use. The test results realized that those alloys that suffered localized attack during the test do not necessarily undergo localized corrosion when placed in the human body as an implant. The test consisted of placing a nonmetallic collar around a test specimen that was immersed in a 0.9% sodium chloride solution for one hour while measuring the potential (E 1 ) with respect to a saturated calomel electrode (SCE). After one hour, the potential of the specimen was shifted to +0.8 V (SCE) and held there for a specific period of time. Depending on the resulting current change over the course of time the resistance to localized attack could be assessed. Then the potential was shifted back to the one hour potential E1, and the current recorded for an additional period of time to determine the state of repassivation. The forming, shaping and finishing steps, including heat treatments, used in the manufacturing of implant devices can significantly alter the corrosion resistance of the alloy from which the device was fabricated. During the selection of a material for use as an implantable device, certain tests, such as F 746, G 5, and G61, can identify corrosion resistance but this does not ensure that actual performance has not been modified because of component fabrication or assembly. Because corrosion of implant devices would have detrimental effects on their performance, it was necessary to determine their susceptibility to localized attack. Originally, ASTM G 61 was used by many to assess the susceptibility of localized corrosion, but this standard was too broad and the interpretation of repassivation was not addressed. Therefore, during the late 1990 s draft revisions and modifications were made to this standard. Thus, the ASTM Committee F04 on Medical and Surgical Materials and Devices adopted F as a modification of the ASTM Standard G 61. This standard assessed the corrosion resistance of small metallic devices or components using the G 61 technique but specific to implants. The devices, anticipated to be evaluated by this test method, included vascular stents, filters, support segments of endovascular grafts, cardiac occluders, aneurysm or ligation clips, staples, etc. The test specimens were to represent the device in its final form and finish, as it would be implanted, and they would be tested in their entirety. Summarizing the F 2129 test method, the device is immersed in a deaerated simulated physiological solution and the open circuit (rest) potential [E r ] is measured for one hour. Then potentiodynamic polarization scan is made from an initial potential of negative 100 mv from the one hour potential. The scan is in the positive, active to noble, direction at a rate of 0.16 mv per second. The scan is reversed when the current has reached two decades greater than that of the breakdown potential [E b ] (defined as a rapid increase of current per increment of applied potential). The reverse scan is stopped when the current becomes less than the current in the forward scan (defined as the protection potential, E p ) or the potential reaches the initial potential. The data is plotted on an x-y semi logarithmic diagram with the current density on the x-axis [logarithmic] in ma/cm2 and the potential versus a SCE on the linear y-axis. Appropriate references in their final form and finish are used as controls. Limitations on testing Nitrogen High purity nitrogen gas for purging was recommended in the original F 2129 standard although no purity specifications were given. In order to tighten the standard, maximum purities were specified in the current revision, such as % bottled gas or the vent gas coming from a Dewar of liquefied nitrogen, which serves the same purpose. Test solution The original F 2129 standard was non-mandatory in the selection of a test solution, although several different physiological environments were referenced. Phosphate buffered saline [PBS] was always our recommendation to clients who asked for our opinion. However, this was before the issuance of F 2129 in September 2001 when that document purposely excluded PBS. Since then, we have tested devices in Ringers and Hanks solutions. One of our major clients mandated that we use Hanks solution until it was discovered that the ph of the solution changed dramatically [ph 7.4 at the beginning of the test to ph 8.5 upon completion, Fig. 1]. Upon further investigation, it was discovered that calcium and magnesium precipitates covered the test cell as well as the test
4 specimen (Figs. 2a and b, energy dispersive x-ray spectrographic analysis, EDS). ph Hanks Ringers PBS Figure 1: Effect of deaeration with nitrogen on the ph of Hanks, Ringers and phosphate buffered solutions (a) Specimen holders Making an attachment to the device so as to minimize a crevice situation or to prevent a galvanic couple between mixed metals is always a challenge. The F 2129 standard states that wire or coil specimens can be mounted to the G 5 holder clamped between stainless steel nuts and partly immersed so as not to expose the hardware in the solution. This exposes the specimen to the air-liquid interface, which subjects it to localized crevice corrosion. An alternative is to coat the specimen at this interface; however, the coated area that is below the solution surface will also introduce a crevice. For stents or cylindrical devices, the F 2129 standard recommends using a special fixture consisting of a cylindrical mandrel that can accommodate the specimen holder described in G 5. The limitation of this device is coating the whole thing with a non-conductive resin, wherein any pinhole in the coating would lead to galvanic corrosion. Since it appears that no holder will completely eliminate crevices or a susceptibility to galvanic coupling, there has been success from using precision mini-clip test jumpers [RadioShack part number A]. These mini-clips are useful for small staples, laser cut tube, and wire mesh devices. The clips fit neatly into the glass electrode holder described in G 5. The clip provides a small contact point that minimizes coating area and galvanic exposure. Acceptance criteria An acceptance criterion is not addressed in any of the standards. Even the FDA does not specify a breakdown potential that must be met to consider a device acceptable. The burden is on the device manufacturer to justify the results they submit in their document package. (b) Figure 2a and b: (a) Clean platinum wire before exposure to Hanks solution: (b) after F 2129 polarization scan at 37 C. Note the presence of calcium, magnesium, oxygen, and increased carbon, indicative of carbonate precipitation In the current re-balloting of F 2129, PBS is the preferred test solution, although other physiological environments are listed as alternatives, including Ringers and Hanks. After evaluating several hundred devices as a third party laboratory, we have developed opinions as to what breakdown potentials should be acceptable, what are unacceptable, and what are marginal. It is our opinion that if a breakdown potential exceeds +600 mv [SCE] in PBS at 37 C, then the alloy is in its optimum corrosion resistant condition. If the breakdown potential is more electronegative than +300 mv, then the alloy is not in an optimal corrosion resistant condition. At breakdown potentials between +300 and +600 mv the alloy is marginal, and a high confidence level [i.e., a larger number of replicates] is needed for acceptance that the alloy will exhibit stable passive behavior. Replicates Electrochemical testing is dependant on the interaction between a metal and the environment. Subtle changes in either will result in minor or major shifts in the current densities when running a potential scan. This is nowhere better brought out than in the scatter bands depicted in G 5.
5 Therefore, to achieve statistical confidence in the results of testing, replicates need to be performed. How many replicates constitute acceptable confidence in the results depends on the number of variables, known and unknown, that are integrated into the manufacture of the device. Ideally, one wants to have a high confidence that a breakdown potential falls between a narrow range of potentials. In other words, it would be better to report that there is 95% confidence that the breakdown potential is between +600 and +800 mv than to be 99% confident that it is between +200 and +900 mv. As a third party laboratory, the number of replicates run is usually dictated by the budget of our clients and their experience with their material. Our guideline to clients is that five replicates falling within a potential range of 100 mv should have a very high confidence level, say 95%. However, without collecting data on a larger number of samples of that particular device, it is impossible to generate the statistical data necessary to enable the calculation of confidence levels for individual test programs. Based on our experience, an example of an acceptable potentiostatic scan with seven samples is shown in Fig. 3. The minimum breakdown potential is +400 mv and the highest is +700mV. Three other runs fall within this range, with two others exhibiting no breakdown. Figure 4 shows another scan with six samples all falling between +400 and +800 mv, and one exhibiting no breakdown. In contrast, Fig. 5 illustrates an unacceptable scan. There are seven scans with breakdown values falling between about 0 and +400 mv. The scatter, and hence the confidence in the testing is comparable to the two previous results, but, the absolute values of the breakdown potentials are too low. Furthermore, there is significantly more scatter in the rest potentials. Figure 6 illustrates a high level of confidence that the material is acceptable, wherein the rest potentials are within a tight band, and all seven curves do not exhibit breakdown. The test results show completely passive behavior up to +800 mv. Figure 7 illustrates both unacceptable material, with all breakdown potentials falling below +400 mv, and relatively low confidence in the test results because of the large amount of scatter in both the rest potential and the breakdown potential. Some of our clients are recognizing the need to establish a statistical basis for the reporting of both the rest potential and the breakdown potential. One of our clients has shown that for their implant about 17 replicates are required to establish the range for the breakdown potential within a 95% confidence. While this may not be typical, it stresses the importance of establishing a statistical database before assuming a few data points will accurately represent the performance of implants in service. Revisions to the ASTM F 2129 Standard The ASTM 2003 Annual Book of Standards Volume has just arrived, however, the Standard F 2129 has been revised and passed balloting in August 2003 but did not make this issue. Substantial changes have been made in order to improve this standard into a specific reference for all those involved in testing small implant devices. The more major changes include: Deleting the true corrosion potential (E corr ) and the corrosion current density (I corr ) because it was felt that the main purpose of F 2129 was to determine the susceptibility of localized corrosion by measurement of the breakdown and protection potentials. Defining the Rest Potential (E r ) as the working electrode voltage one hour after immersion into the test solution, and the Zero Current Potential (E zc ) as the voltage at which the current reaches a minimum during the forward potentiodynamic polarization scan. The test solution has been standardized on phosphate buffered saline (PBS), unless otherwise specified. The PBS is not to include calcium (Ca) or magnesium (Mg), as these species, most notably found in Ringers and Hanks Solutions, are not stable and precipitate out of solution which affects the ph. Deaeration is to be achieved with high purity nitrogen (minimum %), although nitrogen vented off a liquid nitrogen Dewar has been found satisfactory. The initiation voltage (E i ) for starting the forward potential scan is to be negative 100 mv of the rest potential, with the reverse (vertex) voltage (E v ) at two decades of current greater than the breakdown potential (E b ). Alternately, a reversing voltage of +1.0 volt may be used in the event that a breakdown potential is not realized. A warning was added to make the experimenter aware of imposed electromagnetic noise by heating baths used to control the temperature of the test cell that can shift potentials in excess of 100 mv. A method of testing for this is to monitor the rest potential of the test electrode with the heating system on, and then turn it off and monitor the potential for any changes. Summary And Conclusions As a third party, independent testing laboratory, confidentiality and non-disclosure agreements are to be respected, and specific information cannot be shared with other clients. Typically the laboratory is given the barest of information about the product to be test. Independent laboratories consider themselves lucky to receive what the
6 alloy is and its surface area, let alone surface finish, heat treatment, process conditions, etc. Many times clients do not even know which test to perform in order to complete their documentation package to the FDA. The independent laboratories are then put into a position to suggest the number of replicates, and even more demanding to recommend and support an acceptance criterion. Electrochemical testing, like any other test, is a single data point. It is one of several tests that can and should be performed to draw an ultimate decision about the resistance of a material to failure. By its nature, electrochemical testing provides a conservative, false positive result. This means that if there is a breakdown in the passive layer on the metal surface then the alloy is susceptible to attack. If there is no breakdown, then the alloy will not pit or crevice corrode in service, provided the variables do not change. Therefore, even if an alloy exhibits breakdown there is no guarantee that it will fail if implanted in the human body. It is the responsibility of the devise manufacturer to insure there is a statistical probability that the device will not fail. This includes running sufficient in vitro tests and selecting an acceptance breakdown criterion that is conservatively more electropositive than in vivo rest potential results. References 1. Mears, D.C., Metals in Medicine and Surgery, International Metals Reviews, pp (1977). 2. Escalas, F., J. Galante, W. Rostoker, and P.S. Coogan, MP35N: A Corrosion Resistant, High Strength Alloy for Orthopedic Surgical Implant: Bio-Assay Results, J. Biomedical Materials Research, vol. 9, p.303 (1975). 3. Rondelli, G. and B. Vicentini, Localized corrosion behaviour in simulated human body fluids of commercial Ni-Ti orthodontic wires, Biomaterials, vol. 20, p.785 (1999). 4. Merritt, K. and S.A. Brown, Studies on the Neutralization of Endotoxin (LPS) by Nitric and Citric Acid, A report to ASTM F on revision of F 86, ASTM International, Conshohocken, PA: ASTM International, 14 March ASTM F Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2003). 6. ASTM B Standard Practice for Operating Salt Spray (Fog) Apparatus. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2000). 7. ASTM G (1999). Standard Reference Test Method for Making Potentiostatic and Potentiodynamic Anodic Polarization Measurements. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2000). 8. ASTM G 5-94 (1999). Standard Reference Test Method for Making Potentiostatic and Potentiodynamic Anodic Polarization Measurements. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2000). 9. ASTM G (1998). Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements for Localized Corrosion Susceptibility of Iron-, Nickel-, or Cobalt-Based Alloys. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2000). 10. ASTM G (1999). Standard Guide for Conducting and Evaluating Galvanic Corrosion Tests in Electrolytes. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2000). 11. Syrett, B.C., PPR Curves A New Method of Assessing Pitting Corrosion Resistance, Corrosion, vol. 33, p. 221 (1977). 12. ASTM F (1999). Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2003). 13. ASTM F Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices. In: Annual book of ASTM Standards, vol , Conshohocken, PA: ASTM International (2003).
7 - - Figure 3: Acceptable breakdown potentials Figure 4: Acceptable breakdown potentials - - Figure 5: Unacceptable breakdown potentials Figure 6: Near prefect confidence with no breakdown - Figure 7: Unacceptable breakdown potentials, with wide scatter
Atomic Structure. Atoms consist of: Nucleus: Electrons Atom is electrically balanced equal electrons and protons. Protons Neutrons
Basics of Corrosion Performance Metals Sacrificial anode manufacturer Specialize in aluminum alloy anodes All products made in the USA (Berks county, PA) ISO9001/2001 Certified Quality System Also traditional
EXPERIMENT #9 CORROSION OF METALS
EXPERIMENT #9 CORROSION OF METALS Objective The objective of this experiment is to measure the corrosion rate of two different metals and to show the effectiveness of the use of inhibitors to protect metals
Dissimilar Metal Corrosion
PDHonline Course S118 (1 PDH) Dissimilar Metal Corrosion Instructor: D. Matthew Stuart, P.E., S.E., F.ASCE, F.SEI, SECB, MgtEng 2013 PDH Online PDH Center 5272 Meadow Estates Drive Fairfax, VA 22030-6658
The Electrical Control of Chemical Reactions E3-1
Experiment 3 The Electrical Control of Chemical Reactions E3-1 E3-2 The Task In this experiment you will explore the processes of oxidation and reduction, in which electrons flow between materials, and
Brochure & Price List. May 2012
Brochure & Price List May 2012 60 BLUE HEN DRIVE, NEWARK, DE 19713 TELEPHONE: 302-454-8200 FAX: 302-454-8204 ON THE WEB @ http://www.corrosionlab.com Email: [email protected] Corrosion Testing Laboratories,
Technical Data BLUE SHEET. Martensitic. stainless steels. Types 410, 420, 425 Mod, and 440A GENERAL PROPERTIES APPLICATIONS PRODUCT FORM
Technical Data BLUE SHEET Allegheny Ludlum Corporation Pittsburgh, PA Martensitic Stainless Steels Types 410, 420, 425 Mod, and 440A GENERAL PROPERTIES Allegheny Ludlum Types 410, 420, 425 Modified, and
Crevice Corrosion on Stainless Steel Propeller Shafts
Crevice Corrosion on Stainless Steel Propeller Shafts A Quick Summary: What is it? How to Prevent it. How to Repair it. Stainless steel propeller shafts and running gear are subject to pitting & crevice
THE ELECTROCHEMISTRY OF CORROSION Edited by Gareth Hinds from the original work of J G N Thomas
THE ELECTROCHEMISTRY OF CORROSION Edited by Gareth Hinds from the original work of J G N Thomas INTRODUCTION The surfaces of all metals (except for gold) in air are covered with oxide films. When such
Paper No. 4071 APPLICATION OF EQCM TO THE STUDY OF CO2 CORROSION
Paper No. 471 APPLICATION OF EQCM TO THE STUDY OF CO2 CORROSION Yang Yang, Bruce Brown and Srdjan Nešić Institute for Corrosion and Multiphase Technology, Department of Chemical and Biomolecular Engineering
AUTOCLAVE CORROSION INHIBITOR EVALUATION
AUTOCLAVE CORROSION INHIBITOR EVALUATION Cormetrics Job #: 12-123 Prepared for: ABC Company Lab: Bay 4-2280 39 th Avenue NE, Calgary, AB. T2E 6P7 Phone: 258-2853 Fax: 291-1423 ABC Company Page 1 of 9 1.
SALT SPRAY AND IMMERSION CORROSION TESTING OF PM STAINLESS STEEL MATERIALS. W. Brian James Hoeganaes Corporation. Cinnaminson, NJ 08077
SALT SPRAY AND IMMERSION CORROSION TESTING OF PM STAINLESS STEEL MATERIALS W. Brian James Hoeganaes Corporation Cinnaminson, NJ 08077 Leander F. Pease III Powder-Tech Associates Inc. Andover, MA 01845
Stainless Steel and Corrosion
Stainless Europe Stainless Steel and Corrosion What is corrosion? Metals, with the exception of the precious metals such as gold and platinum, that are found in their natural state are always extracted
POTENTIAL FOR HYDROGEN GENERATION AND EMBRITTLEMENT OF PRESTRESSING STEEL IN GALVANIZED PIPE VOIDED PILE
FDOT Research Report No. FL/DOT/SMO 04-477 POTENTIAL FOR HYDROGEN GENERATION AND EMBRITTLEMENT OF PRESTRESSING STEEL IN GALVANIZED PIPE VOIDED PILE submitted to Florida Department of Transportation State
North American Stainless
North American Stainless Flat Product Stainless Steel Grade Sheet 316 (S31600)/EN 1.4401 316L (S31603)/ EN 1.4404 INTRODUCTION NAS provides 316 and 316L SS, which are molybdenum-bearing austenitic stainless
CORROSION PROTECTION METHODS OF STRUCTURAL STEEL AGAINST ATMOSPHERIC CORROSION
CORROSION PROTECTION METHODS OF STRUCTURAL STEEL AGAINST ATMOSPHERIC CORROSION E. Daflou a, E. Rakanta b, *G. Batis c a Chemical Engineer, Chemical Engineering School, Section of Materials Science and
ATI 2205 ATI 2205. Technical Data Sheet. Duplex Stainless Steel GENERAL PROPERTIES. (UNS S31803 and S32205)
ATI 2205 Duplex Stainless Steel (UNS S31803 and S32205) GENERAL PROPERTIES ATI 2205 alloy (UNS S31803 and/or S32205) is a nitrogen-enhanced duplex stainless steel alloy. The nitrogen serves to significantly
CATHODIC PROTECTION SYSTEM DESIGN
CATHODIC PROTECTION SYSTEM DESIGN Presented By DENIS L ROSSI P.E. CORROSION ENGINEER New England C P Inc. Corrosion Fundamentals What is corrosion? It is defined as the degradation or deterioration of
Chapter 13: Electrochemistry. Electrochemistry. The study of the interchange of chemical and electrical energy.
Chapter 13: Electrochemistry Redox Reactions Galvanic Cells Cell Potentials Cell Potentials and Equilbrium Batteries Electrolysis Electrolysis and Stoichiometry Corrosion Prevention Electrochemistry The
A SHORT INTRODUCTION TO CORROSION AND ITS CONTROL
A SHORT INTRODUCTION TO CORROSION AND ITS CONTROL CORROSION OF METALS AND ITS PREVENTION WHAT IS CORROSION Corrosion is the deterioration of materials by chemical interaction with their environment. The
Galvanic cell and Nernst equation
Galvanic cell and Nernst equation Galvanic cell Some times called Voltaic cell Spontaneous reaction redox reaction is used to provide a voltage and an electron flow through some electrical circuit When
Corrosion experiments in amine solutions
Corrosion experiments in amine solutions Andreas Grimstvedt Process technology SINTEF Materials and chemistry Wenle He Applied mechanics and corrosion SINTEF Materials and chemistry 1 Contents of presentation
The Potentiodynamic Polarization Scan. Technical Report 33
Technical Report 33 Technical Report 33 David G Enos Center for Electrochemical Science & Engineering Department of Materials Science & Engineering University of Virginia Charlottesville, VA Louie L Scribner
North American Stainless
North American Stainless Long Products Stainless Steel Grade Sheet AISI 316 UNS S31600 EN 1.4401 AISI 316L UNS S31630 EN 1.4404 INTRODUCTION NAS provides 316 and 316L SS, which are molybdenum-bearing austenitic
Metals and Non-metals. Comparison of physical properties of metals and non metals
Metals and Non-metals Comparison of physical properties of metals and non metals PHYSICAL PROPERTY METALS NON-METALS Physical State Metallic lustre (having a shining surface) Mostly solids (Liquid -mercury)
Experiment 9 Electrochemistry I Galvanic Cell
9-1 Experiment 9 Electrochemistry I Galvanic Cell Introduction: Chemical reactions involving the transfer of electrons from one reactant to another are called oxidation-reduction reactions or redox reactions.
SELECTIVE DISSOLUTION AND CORROSION FATIGUE BEHAVIORS OF 2205 DUPLEX STAINLESS STEEL
W.-T. Tsai, I.-H. Lo Department of Materials Science and Engineering National Cheng Kung University Tainan, Taiwan SELECTIVE DISSOLUTION AND CORROSION FATIGUE BEHAVIORS OF 2205 DUPLEX STAINLESS STEEL ABSTRACT
Question Bank Electrolysis
Question Bank Electrolysis 1. (a) What do you understand by the terms (i) electrolytes (ii) non-electrolytes? (b) Arrange electrolytes and non-electrolytes from the following substances (i) sugar solution
Galvanic Cells and the Nernst Equation
Exercise 7 Page 1 Illinois Central College CHEMISTRY 132 Laboratory Section: Galvanic Cells and the Nernst Equation Name: Equipment Voltage probe wires 0.1 M solutions of Pb(NO 3, Fe(NO 3 ) 3, and KNO
PRODUCT / APPLICATION INFORMATION
Abstract This paper will discuss performance life testing and stability testing used for self-regulating heating cables. It will show how the techniques were developed and evolved from the 970 s to the
CATHODIC PROTECTION P E FRANCIS
CATHODIC PROTECTION P E FRANCIS CONTENTS 1 INTRODUCTION...2 2 PRINCIPLES OF CATHODIC PROTECTION...3 3 METHODS OF APPLYING CATHODIC PROTECTION...5 3.1 IMPRESSED CURRENT...5 3.2 SACRIFICIAL ANODES...6 4
Lead & Magnet Wire Connection Methods Using the Tin Fusing Method Joyal A Division of AWE, Inc.
Lead & Magnet Wire Connection Methods Using the Tin Fusing Method Joyal A Division of AWE, Inc. Abstract The technology for connecting lead and magnet wires for electric motors and electro mechanical devices
GALVANIC MODEL FOR LOCALIZED CO2 CORROSION
Paper No. 687 GALVANIC MODEL FOR LOCALIZED CO CORROSION Jiabin Han, Srdjan Nešić and Bruce N. Brown Institute for Corrosion and Multiphase Technology Department of Chemical and Biomolecular Engineering
ALLOY 2205 DATA SHEET
ALLOY 2205 DATA SHEET UNS S32205, EN 1.4462 / UNS S31803 GENERAL PROPERTIES ////////////////////////////////////////////////////// //// 2205 (UNS designations S32205 / S31803) is a 22 % chromium, 3 % molybdenum,
reduction ore = metal oxides metal oxidation
Cathodic Protection and Interferences René Gregoor Madrid, June 18 th and 19 th 2009 1 Cathodic protection and interferences Corrosion Cathodic protection Protection criterion ON potential measurements
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS Sl. No. Contents Preamble 5.1 Aim 5.2 Introduction 5.2.1 Environmental Significance 5.3 Principle 5.4 Materials Required 5.4.1 Apparatus Required 5.4.2
Rajesh Swaminathan. March 13, 2005
Chemistry 12 IB Corrosion of Iron Rajesh Swaminathan March 13, 2005 1 Planning A 1.1 Aim The aim of the experiment is to investigate factors that affect the rate of corrosion of iron. More specifically,
North American Stainless
North American Stainless Flat Products Stainless Steel Grade Sheet 310S (S31008)/ EN 1.4845 Introduction: SS310 is a highly alloyed austenitic stainless steel designed for elevated-temperature service.
Technical Paper. Corrosion Resistance of Bolt Coatings. Dmitry Zhmurkin, Ph.D. Tyco Electronics Harrisburg, PA. July 23, 2009
Technical Paper Corrosion Resistance of Bolt Coatings Dmitry Zhmurkin, Ph.D. Tyco Electronics Harrisburg, PA July 23, 2009 Hot-Dip Galvanizing Batch hot-dip galvanizing has been the most commonly used
Corrosion Control & Cathodic Protection for Water & Wastewater Systems
Corrosion Control & Cathodic Protection for Water & Wastewater Systems Presented By: James T Lary Corrpro Companies, Inc 1090 Enterprise Dr. Medina, OH 44256 Tel. 330-723 723-5082 (x1215) email: [email protected]
Durcomet 100 CD4MCuN. Bulletin A/7l
Durcomet 100 CD4MCuN Bulletin A/7l Durcomet 100 Introduction Durcomet 100 is a duplex stainless steel produced to ASTM specification A995 or A890, Grade CD4MCuN (1B). It is indicated by the Flowserve casting
THE INVESTIGATION OF CORROSION BEHAVIOR OF 316 Cb AUSTENITIC STAINLESS STEEL IN ACIDIC SOLUTIONS
316 Cb OSTENİTİK PASLANMAZ ÇELİĞİN ASİDİK ÇÖZELTİLERDEKİ KOROZİF DAVRANIŞININ İNCELENMESİ M. SONCU*, S. ZOR**, L.ÇAPAN* *Kocaeli Üniversitesi Makine Mühendisliği Bölümü, 41380, Kocaeli **Kocaeli Üniversitesi
GENERAL PROPERTIES //////////////////////////////////////////////////////
ALLOY 625 DATA SHEET //// Alloy 625 (UNS designation N06625) is a nickel-chromium-molybdenum alloy possessing excellent resistance to oxidation and corrosion over a broad range of corrosive conditions,
Name Electrochemical Cells Practice Exam Date:
Name Electrochemical Cells Practice Exam Date: 1. Which energy change occurs in an operating voltaic cell? 1) chemical to electrical 2) electrical to chemical 3) chemical to nuclear 4) nuclear to chemical
North American Stainless
North American Stainless Flat Products Stainless Steel Grade Sheet 304 (S30400)/ EN 1.4301 304L (S30403) / EN 1.4307 304H (S30409) Introduction: Types 304, 304L and 304H are the most versatile and widely
DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT
DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT BACKGROUND AND PRODUCT DESCRIPTIONS Fracture Diagnostics provides state-of-the-art test equipment, utilizing the Rising Step Load testing technique.
Practical Examples of Galvanic Cells
56 Practical Examples of Galvanic Cells There are many practical examples of galvanic cells in use in our everyday lives. We are familiar with batteries of all types. One of the most common is the lead-acid
CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001)
CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001) I. Introduction Calibration experiments or procedures are fairly common in laboratory work which involves any type of instrumentation. Calibration
Chapter 3 BASICS OF CORROSION MEASUREMENTS. Mixed-Potential Theory. The mixed potential-theory (1) consists of two simple
Chapter 3 BASICS OF CORROSION MEASUREMENTS Mixed-Potential Theory. The mixed potential-theory (1) consists of two simple hypothesis: (1) any electrochemical reaction can be divided into two or more partial
of the Corrosion Resistance of Conductive Elastomers:
CHO-TR19C September/1992 Quantitative Determination of the Corrosion Resistance of Conductive Elastomers: Part III. Materials Based on Silver-Plated Aluminum Fillers (Various Manufacturers) Prepared By:
Corrosion of Copper in Water
Corrosion of Copper in Water Digby D. Macdonald Center for Electrochemical Science and Technology Department of Materials Science and Engineering Pennsylvania State University 209 Steidle Bldg University
Utilizing of Sodium Nitrite as Inhibitor for Protection of Carbon Steel in Salt Solution
Int. J. Electrochem. Sci., 7 (2012) 6941-6950 International Journal of ELECTROCHEMICAL SCIENCE www.electrochemsci.org Utilizing of Sodium Nitrite as Inhibitor for Protection of Carbon Steel in Salt Solution
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope Rakesh Sidharthan 1 Gnanavel B K 2 Assistant professor Mechanical, Department Professor, Mechanical Department, Gojan engineering college,
North American Stainless
North American Stainless Long Products Stainless Steel Grade Sheet 2205 UNS S2205 EN 1.4462 2304 UNS S2304 EN 1.4362 INTRODUCTION Types 2205 and 2304 are duplex stainless steel grades with a microstructure,
Best Practice in Boiler Water Treatment
Best Practice in Boiler Water Treatment Boiler Water Treatment Part 2 Internal Treatment Objectives of Internal Water Treatment 1 To control the level of total dissolved solids (TDS) within the boiler
The Pennsylvania State University. The Graduate School. Engineering Science and Mechanics
The Pennsylvania State University The Graduate School Engineering Science and Mechanics DEVELOPMENT OF NEW MULTIFUNCTIONAL COATINGS FOR PROTECTION AGAINST EROSION AND CORROSION IN QATAR OIL AND GAS PRODUCTION
Anodes and Misc Equipment
Anodes and Misc Equipment Application: Platinised Titanium Anodes Platinised titanium anodes are recommended for use in the following electrolytic processes:- Precious metal electroplating - e.g. Au, Pt,
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical
THE USE OF OZONE IN COOLING TOWERS
THE USE OF OZONE IN COOLING TOWERS Paul D. McNicholas Ozonia Ltd Duebendorf, Switzerland Abstract Ozone has been successfully applied to industrial cooling water systems with the resultant improvement
A NEW FRONTIER FOR DEPOSIT STRESS MEASUREMENTS
A NEW FRONTIER FOR DEPOSIT STRESS MEASUREMENTS By Frank H. Leaman Specialty Testing and Development Company, Inc. York, PA USA ABSTRACT Internal stress exists as an inherent force within electroplated
Electrochemical Half Cells and Reactions
Suggested reading: Chang text pages 81 89 Cautions Heavy metals, such as lead, and solutions of heavy metals may be toxic and an irritant. Purpose To determine the cell potential (E cell ) for various
THE STRAIN GAGE PRESSURE TRANSDUCER
THE STRAIN GAGE PRESSURE TRANSDUCER Pressure transducers use a variety of sensing devices to provide an electrical output proportional to applied pressure. The sensing device employed in the transducers
Implant materials. Learning outcomes. Implant materials in trauma. How to use this handout? Functions of implants. Types of materials
Implant materials How to use this handout? The left column is the information as given during the lecture. The column at the right gives you space to make personal notes. Learning outcomes At the end of
DETERMINING THE MASS OF A COPPER ATOM
DETERMINING THE MASS OF A COPPER ATOM LAB ADV.31 From Vernier Software & Technology, 2004 STANDARDS ADDRESSED 3.4.10 A Explains concepts about the structure and properties of matter. 3.4.12 A Apply concepts
ALLOY C276 DATA SHEET
ALLOY C276 DATA SHEET //// Alloy C276 (UNS designation N10276) is a nickel-molybdenum-chromium-iron-tungsten alloy known for its corrosion resistance in a wide range of aggressive media. It is one of the
Axial and Radial Leaded Multilayer Ceramic Capacitors for Automotive Applications Class 1 and Class 2, 50 V DC, 100 V DC and 200 V DC
Axial and Radial Leaded Multilayer Ceramic Capacitors for Automotive Applications Class 1 and Class 2, 5 V DC, 1 V DC and 2 V DC DESIGNING For more than 2 years Vitramon has supported the automotive industry
North American Stainless
Introduction: North American Stainless Flat Products Stainless Steel Grade Sheet 309S (S30908)/ EN1.4833 SS309 is a highly alloyed austenitic stainless steel used for its excellent oxidation resistance,
Discovering Electrochemical Cells
Discovering Electrochemical Cells Part I Electrolytic Cells Many important industrial processes PGCC CHM 102 Cell Construction e e power conductive medium What chemical species would be present in a vessel
Electrochemistry Voltaic Cells
Electrochemistry Voltaic Cells Many chemical reactions can be classified as oxidation-reduction or redox reactions. In these reactions one species loses electrons or is oxidized while another species gains
Application Note CORR-1
Application Note CORR-1 Subject: Basics of Corrosion Measurements POTENTIODYNAMIC POLARIZATION MEASUREMENTS Introduction Many corrosion phenomena can be explained in terms of electrochemical reactions.
Eveready Carbon Zinc (Zn/MnO ² ) Application Manual
Page 1 of 13 Eveready Carbon Zinc (Zn/MnO ² ) Application Manual Eveready carbon zinc batteries are marketed in two basic categories--classic and Super Heavy Duty. The Classic category, our least expensive
CHM1 Review Exam 12. Topics REDOX
CHM1 Review Exam 12 Topics REDOX REDOX Reactions Oxidation Reduction Oxidizing agent Reducing agent Galvanic (Voltaic) Cells Anode Cathode Salt bridge Electrolyte Half-reactions Voltage o Positive voltages
MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.
Chemistry 1C-Dr. Larson Chapter 20 Review Questions MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) is reduced in the following reaction: Cr2O7
ELECTROCHEMICAL CELLS LAB
ELECTROCHEMICAL CELLS LAB Purpose: The purpose of this lab is to demonstrate the ability of chemistry to make electric current using oxidation/reduction (REDOX) reactions, and to measure the electric current
A Potentiometric Analysis of Fluoride Ion in Toothpaste
CHEM 311L Quantitative Analysis Laboratory Revision 2.0 A Potentiometric Analysis of Fluoride Ion in Toothpaste In this laboratory exercise, we will analyze a toothpaste sample for its Fluoride Ion (F
o Electrons are written in half reactions but not in net ionic equations. Why? Well, let s see.
REDOX REACTION EQUATIONS AND APPLICATIONS Overview of Redox Reactions: o Change in Oxidation State: Loses Electrons = Oxidized (Oxidation number increases) Gains Electrons = Reduced (Oxidation Number Reduced)
Figure 1. A voltaic cell Cu,Cu 2+ Ag +, Ag. gas is, by convention, assigned a reduction potential of 0.00 V.
Voltaic Cells Introduction In this lab you will first prepare a set of simple standard half-cells and then measure the voltage between the half-cells with a voltmeter. From this data you will be able to
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical
RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE
Roulette, T., J. Roulette, and S. Pons. Results of ICARUS 9 Experiments Run at IMRA Europe. in Sixth International Conference on Cold Fusion, Progress in New Hydrogen Energy. 1996. Lake Toya, Hokkaido,
CATHODIC PROTECTION OF REINFORCED CONCRETE STRUCTURES
CATHODIC PROTECTION OF REINFORCED CONCRETE STRUCTURES AN OVERVIEW CLEARLY THE BEST PROTECTION FOR YOUR INVESTMENT Introduction Cathodic protection is applied to reinforced concrete structures to either
An Ammonium-free, Acid. Zinc/Nickel (12-15% Ni) Process
An Ammonium-free, Acid Zinc/Nickel (12-15% Ni) Process Requirements of an acid zinc/nickel process Technical requirements High plating speed (current efficiency up to 90 %) High thicknesses of deposits
Chair: David Kroon Vice-Chair: Chuck Lawrence
CONFERENCE CALL MINUTES TASK GROUP 284 CATHODIC PROTECTION, GALVANIC ANODE FOR INTERNAL SUBMERGED SURFACES OF STEEL WATER STORAGE TANKS: REVIEW OF NACE SP0196-2011 ASSIGNMENT: To review and update NACE
Determining Equivalent Weight by Copper Electrolysis
Purpose The purpose of this experiment is to determine the equivalent mass of copper based on change in the mass of a copper electrode and the volume of hydrogen gas generated during an electrolysis reaction.
Ch 20 Electrochemistry: the study of the relationships between electricity and chemical reactions.
Ch 20 Electrochemistry: the study of the relationships between electricity and chemical reactions. In electrochemical reactions, electrons are transferred from one species to another. Learning goals and
How Sensors Work. How Oxygen, Electrochemical Toxic, and Metal Oxide Semiconductor Sensors Work *
How Oxygen, Electrochemical Toxic, and Metal Oxide Semiconductor Sensors Work * 1. Oxygen sensor detection principle Most portable or survey instruments used for workplace evaluation of oxygen concentrations
Corrosion of Steel, Aluminum and Copper in Electrical Applications
Corrosion of Steel, Aluminum and Copper in Electrical Applications By: Alex Mak, P. Eng. - Senior Field Applications Engineer, General Cable Executive Summary Corrosion, typically defined as the deterioration
PASSIVATION TREATMENTS FOR CORROSION-RESISTANT STEEL
FEDERAL SPECIFICATION QQ-P-35C October 28, 1988 SUPERSEDING Fed. Spec. QQ-P-35B April 5, 1973 PASSIVATION TREATMENTS FOR CORROSION-RESISTANT STEEL This specification was approved by the Assistant Administrator
TECHNICAL SERVICE DEPARTMENT Technical Service Bulletin 1-800-432-8373. Anode Rods, Cathodic Protection and the Porcelain (glass) Lining
Corrosion can be defined as the destructive attack of a metal by an electrochemical reaction with its environment. Steel exposed to moisture and oxygen will rust and corrode. Corrosion is defined as the
DEGASSED CATION CONDUCTIVITY MEASUREMENT
(Presented at EPRI's 8th International Conference on Cycle Chemistry in Fossil and Combined Cycle Plants with Heat Recovery Steam Generators - June 20-23, 2006 Calgary, Alberta Canada) DEGASSED CATION
9707 Key West Avenue, Suite 100 Rockville, MD 20850 Phone: 301-740-1421 Fax: 301-990-9771 E-Mail: [email protected]
9707 Key West Avenue, Suite 100 Rockville, MD 20850 Phone: 301-740-1421 Fax: 301-990-9771 E-Mail: [email protected] Part of the recertification process is to obtain Continuing Education Units (CEUs). One way
EDEXCEL INTERNATIONAL GCSE CHEMISTRY EDEXCEL CERTIFICATE IN CHEMISTRY ANSWERS SECTION C
EDEXCEL INTERNATIONAL GCSE CHEMISTRY EDEXCEL CERTIFICATE IN CHEMISTRY ANSWERS SECTION C Chapter 16 1. Burn sulfur in air to give sulfur dioxide. S(s) + O 2 (g) ----> SO 2 (g) Pass this with more air over
CYCLE CHEMISTRY ph MEASUREMENT
CYCLE CHEMISTRY ph MEASUREMENT David M. Gray and Edward P. Santini Thornton Associates, Waltham MA Electric Utility Chemistry Workshop Champaign, Illinois May 12-14, 1998 Sponsored by University of Illinois
CONTENTS. ZVU Engineering a.s., Member of ZVU Group, WASTE HEAT BOILERS Page 2
WASTE HEAT BOILERS CONTENTS 1 INTRODUCTION... 3 2 CONCEPTION OF WASTE HEAT BOILERS... 4 2.1 Complex Solution...4 2.2 Kind of Heat Exchange...5 2.3 Heat Recovery Units and Their Usage...5 2.4 Materials
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this document
LASER CUTTING OF STAINLESS STEEL
LASER CUTTING OF STAINLESS STEEL Laser inert gas cutting is the most applicable process type used for cutting of stainless steel. Laser oxygen cutting is also applied in cases where the cut face oxidation
Comparison of the Corrosion Protection Effectiveness of Vapor Corrosion Inhibitor and Nitrogen Blanketing System
Comparison of the Corrosion Protection Effectiveness of Vapor Corrosion Inhibitor and Nitrogen Blanketing System Behzad Bavarian 1, Jia Zhang 1, Lisa Reiner 1 and Boris Miksic, FNACE 2 Dept. of Manufacturing
Chem 1721 Brief Notes: Chapter 19
Chem 1721 Brief Notes: Chapter 19 Chapter 19: Electrochemistry Consider the same redox reaction set up 2 different ways: Cu metal in a solution of AgNO 3 Cu Cu salt bridge electrically conducting wire
FEATURES AND BENEFITS OF DIFFERENT PLATINUM ALLOYS. Kris Vaithinathan and Richard Lanam Engelhard Corporation
FEATURES AND BENEFITS OF DIFFERENT PLATINUM ALLOYS Kris Vaithinathan and Richard Lanam Engelhard Corporation Introduction There has been a significant increase in the world wide use of platinum for jewelry