Albuterol Delivery by 4 Different Nebulizers Placed in 4 Different Positions in a Pediatric Ventilator In Vitro Model
|
|
|
- Michael Gregory
- 9 years ago
- Views:
Transcription
1 Albuterol Delivery by 4 Different Nebulizers Placed in 4 Different Positions in a Pediatric Ventilator In Vitro Model Ariel Berlinski MD and J Randy Willis RRT-NPS BACKGROUND: The type of aerosol generator and the position in the ventilator circuit are crucial to determine aerosol delivery during mechanical ventilation. We compared lung deposition of albuterol aerosols generated by 4 different nebulizers placed in-line in 4 different positions in a pediatric ventilator model. METHODS: Two brands of continuously operated jet nebulizer (6 L/ min, oxygen), an ultrasonic, and a vibrating mesh were compared when placed at the ventilator, the humidifier, the Y-piece, and 30 cm before the Y-piece. The jet, ultrasonic, and vibrating mesh nebulizers were operated for 5, 15, and 15 min, respectively. The tested solutions contained 2.5 mg, 5.0 mg, and 7.5 mg of albuterol sulfate. The ventilator settings were: pressure-regulated volume control mode, tidal volume 200 ml, breathing frequency 20 breaths/min, PEEP 5 cm H 2 O, F IO2 0.4, inspiratory time 0.75 s, bias flow 2 L/min, and heater 37 C. The circuit was connected in series to a 5.5 mm cuffed endotracheal tube, a deposition filter, and a lung model. Albuterol was measured by spectrophotometry. RESULTS: Intra-device comparison: the jet and vibrating mesh nebulizers performed best at either the ventilator or humidifier, and worst at the Y-piece. The ultrasonic nebulizer performed best at the humidifier and worst at the Y-piece. Inter-device comparison: the vibrating mesh nebulizer outperformed both jet nebulizers at all tested positions, and the ultrasonic nebulizer when placed at either the ventilator or the humidifier. Lung deposition increased for the jet and ultrasonic nebulizers, but not for vibrating mesh nebulizer, when increasing the loading volume while maintaining the nominal dose. CONCLUSIONS: The vibrating mesh nebulizer was the most efficient device. The nebulizers were more efficient when placed at either the ventilator or the humidifier, and less efficient when placed at either the Y-piece or 30 cm from the Y-piece. These conclusions are valid for the tested conditions. Data regarding optimization of operating conditions should not be extrapolated among nebulizers of different operating principles. Key words: mechanical ventilation; albuterol; jet nebulizer; vibrating mesh nebulizer; ultrasonic nebulizer; position; pediatrics; aerosol; humidifier. [Respir Care 2013;58(7): Daedalus Enterprises] Introduction Pediatric patients receiving mechanical ventilation are often prescribed nebulized therapy. The ideal device and operating conditions are yet to be established. Although direct quantification of lung deposition would be the most accurate method to determine best device/operating conditions, ethical considerations regarding blood draws and The authors are affiliated with Arkansas Children s Hospital, Little Rock, Arkansas. Dr Berlinski is also affiliated with the Pulmonology Section, Department of Pediatrics, University of Arkansas for Medical Sciences, and with the Pediatric Aerosol Research Laboratory, Arkansas Children s Hospital Research Institute, Little Rock, Arkansas Mr Willis presented a version of this paper at the 56th AARC Congress, held December 6 9, 2010, in Las Vegas, Nevada. Dr Berlinski has disclosed relationships with Johnson & Johnson, MPEX Pharmaceutical, Gilead Sciences, Philips, and S&T Technologies. Mr Willis has disclosed no conflicts of interest. The Pediatric Aerosol Research Laboratory at Arkansas Children s Hospital Research Institute was partially established by, and receives partial support from, the George Endowment for Asthma. Correspondence: Ariel Berlinski MD, Pulmonary Medicine Division, Department of Pediatrics, University of Arkansas for Medical Sciences, 1 Children s Way, Slot , Little Rock AR [email protected]. DOI: /respcare RESPIRATORY CARE JULY 2013 VOL 58 NO 7
2 QUICK LOOK Fig. 1. Nebulizers tested. From right to left: Aerogen Solo, Maquet Ultrasonic model N E400E, Salter 8900, and Hudson Updraft II Opti-Neb. use of radio-labeled studies have limited their use. 1 Investigators have resorted to in vitro models, for which good in vivo/in vitro correlation has been established. 2,3 SEE THE RELATED EDITORIAL ON PAGE 1255 Current knowledge Aerosol lung deposition during pediatric mechanical ventilation is affected by the type of aerosol generator and its position in the circuit, endotracheal tube size, humidification, and ventilation mode. The optimal combination of each nebulizer type and position is unknown. What this paper contributes to our knowledge In an in-vitro pediatric ventilator model, a vibratingmesh nebulizer outperformed jet and ultrasonic nebulizers, irrespective of position in the ventilator circuit. Aerosol deposition was maximized with the nebulizer at the ventilator outlet or immediately after the humidifier. Deposition was less with the nebulizer at the Y- piece or 30 cm from the Y-piece. Delivery efficiency depends on many factors such as type of aerosol generator, position, endotracheal tube (ETT) size, and others. 4,5 The type of aerosol generator can determine a several-fold difference in drug delivery. 6 Currently jet nebulizers, ultrasonic nebulizers, and vibrating mesh nebulizers are available for use with ventilator circuits. Several authors have shown that the position of the aerosol generator can greatly influence drug delivery to the patient However, most of these studies were done using adult type settings and therefore are not applicable to pediatric scenarios In this study we compared drug delivery efficiency and absolute lung deposition of 2 jet nebulizers, an ultrasonic nebulizer, and a vibrating mesh nebulizer, placed in 4 different positions in a pediatric ventilator circuit model, using different albuterol doses and concentrations. Methods This study was performed at the Pediatric Aerosol Research Laboratory, Arkansas Children s Hospital Research Institute, and Arkansas s Children s Hospital, Little Rock, Arkansas. Nebulizers Four different brands of nebulizer, representing 3 different operating principles were tested (Fig. 1). Two continuously operated jet nebulizers were tested: Salter 8900 (Salter Labs, Arvin, California) and Hudson Updraft II Opti-Neb (Teleflex Medical, Research Triangle Park, North Carolina). Both nebulizers were connected to the circuit with a spring-loaded T-piece (Airlife Valved Tee adapter, Thayer Medical, Tucson, Arizona) and operated with central oxygen at 6 L/min for 5 min. The ventilator settings were adjusted during nebulization to compensate for the increase in flow. One ultrasonic nebulizer, model N E400E, Maquet, Solna, Sweden, and one vibrating mesh nebulizer, Solo, Aerogen, Galway, Ireland, were also tested. Both devices were connected to the ventilator circuit with a T-piece and were operated for 15 min. No ventilator adjustments were required to operate these 2 devices. The length of operation was decided after preliminary testing showed that these intervals allowed nebulization to dryness for all devices. Table 1 compares the different characteristics of the nebulizers, including the cost of delivering different amounts of albuterol sulfate for 3 different lengths of time. Solutions Several loading doses, volumes, and concentrations of albuterol sulfate nebulizer solution were tested. Initially, 2.5 mg/3 ml, 5.0 mg/3.5 ml, and 7.5 mg/4 ml were used. These solutions were made of single bullet units of albuterol sulfate 2.5 mg/3 ml (Nephron Pharmaceuticals, Orlando, Florida) and 0.5 mg/ml (Nephron Pharmaceuticals, Orlando, Florida). Previous research demonstrated that higher loading volumes determine higher drug output. 13 Therefore, we also tested 2 other solutions with the same nominal dose (2.5 mg) and different loading volumes (3.5 ml and 4 ml) to evaluate the effect of increasing the loading volume on lung dose. Positions of Aerosol Generators All nebulizers were tested in 4 different positions (Fig. 2). Position A was at the ventilator on the inspiratory side. RESPIRATORY CARE JULY 2013 VOL 58 NO
3 Table 1. Features of Tested Nebulizers Hudson Updraft II Opti-Neb Salter 8900 Maquet Ultrasonic Aerogen Solo Vibrating Mesh Weight, g Dead volume External gas source required Yes Yes No No Provides additional flow to the circuit requiring adjustment Yes Yes No No of ventilator settings Effect on medication temperature Cost for 1 treatment* 2.5 mg/3 ml mg/3.5 ml mg/4 ml Cost for 28 treatments* 2.5 mg/3 ml mg/3.5 ml mg/4 ml Cost for 120 treatments* 2.5 mg/3 ml mg/3.5 ml mg/4 ml * Cost is expressed in United States dollars and was constructed with the following assumptions: nebulizer treatments to be given every 6 hours; jet nebulizer replaced every 48 hours, as per our hospital protocol; ultrasonic plastic cup used for all treatments; jet nebulizer run at 6 L/min of oxygen for 5 min; lifespan of controllers of vibrating mesh and ultrasonic is 2 years with 4 treatments/ day every day; spring-loaded T-piece and vibrating mesh circuit adapter used for all treatments; respiratory therapist time spent is similar among all devices. The ultrasonic nebulizer requires hardware and software to be installed on the ventilator. The additional cost was equated to the cost of the vibrating mesh controller. Fig. 2. Testing set-up. Position A: the ventilator on the inspiratory side. Position B: Between the humidifier and the inspiratory limb. Position C: Between the inspiratory limb and the Y-piece. Position D: In the inspiratory limb, 30 cm before the Y-piece. ETT endotracheal tube. Position B was between the humidifier and the inspiratory limb of the ventilator circuit. Position C was between the inspiratory limb of the ventilator circuit and the Y-piece. Position D was in the inspiratory limb of the ventilator circuit, at 30 cm before the Y-piece. The latter was achieved by cutting the circuit at 30 cm from the Y-piece, re-threading the heated wire, and placing a T-piece between the 2 pieces of circuit. Pediatric Ventilator Model A previously described pediatric ventilator model was used. 6 Briefly, a ventilator (Servo-i, Maquet, Solna, Sweden) operated in pressure-regulated volume control mode connected to a 1.83 m long and 22 mm inner diameter heated wire ventilator circuit (Evaqua, Fisher & Paykel, Auckland, New Zealand) was used. The settings were: tidal volume (V T ) 200 ml, breathing frequency 20 breaths/ min, PEEP 5 cm H 2 O, F IO2 0.4, inspiratory time 0.75 s, inspiratory rise 0.15 s, flow trigger 3, bias flow 2 L/min, and heater C. The circuit was connected to a 5.5 mm inner diameter cuffed ETT (Mallinckrodt Lo-Pro, Tyco/Mallinckrodt-Nellcor, Pleasanton, California). A lung model (SmartLung 600 ml, IMT Medical, Buchs, Switzerland) was connected in series, and a low-dead-space filter folder was interposed between them. The cuff was inflated to provide a closed circuit. Testing Procedure A new respiratory filter (Pari Respiratory Equipment, Midlothian, Virginia) was placed in the filter holder before each run. The filter holder was sealed with Teflon tape, and the efficiency of the seal was verified by checking the returned V T. The filter was removed upon completion of the run time, eluted with ultrapure water, and the washings were tested for albuterol with a spectrophotometer (Bio- Mate 3 UV-Vis Spectrophotometer, Thermo Fisher Scientific, Waltham, Massachusetts) at 276 nm. 14 The amount of drug captured in the filter was considered the lung dose. Five units of each aerosol generator were tested in each of the 4 positions, with 2.5 mg/3 ml, 5.0 mg/3.5 ml and 7.5 mg/4 ml. All positions except position D were tested with solutions of 2.5 mg/3.5 ml and 2.5 mg/4 ml RESPIRATORY CARE JULY 2013 VOL 58 NO 7
4 Statistical Analysis Data were compared as lung dose ( g of albuterol) and drug delivery efficiency (%). The latter was calculated as: Efficiency drug captured in the filter g / drug placed in the nebulizer g 100 Intra-device and inter-device comparisons were performed at all tested positions with each tested solution, using analysis of variance for repeated measures and analysis of variance, respectively, followed by the Tukey test when multiple comparison analysis was required. Differences within each device placed at each tested position between solutions of the same volume and different concentrations (2.5 mg/3.5 mg vs 5.0 mg/3.5 ml, 2.5 mg/ 4 ml vs 7.5 mg/4 ml, and 2.5 mg/3 ml vs 2.5 mg/4 ml) were tested using a 2-tailed paired t test. A statistical software package (MDAS 2.0, Eskay Software, Silver Spring, Maryland) was used to analyze the data. A P value.05 was considered statistically significant. Intra-device Comparison Results The vibrating mesh nebulizer performed best and worst when placed at either the humidifier or ventilator and Y- piece, respectively, for all tested solutions (P.01) (Fig. 3). These results were true for lung dose and delivery efficiency (Table 2). The difference in lung dose ranged from 3.2-fold with the 7.5 mg/4 ml solution to 4.3-fold with the 2.5 mg/3.5 ml solution. The Hudson nebulizer performed best and worse when placed at either the humidifier or ventilator and Y-piece, respectively, for all tested solutions (P.01) (see Fig. 3). These results were true for lung dose and delivery efficiency (see Table 2). The difference in lung dose ranged from 2.1-fold with the 7.5 mg/4 ml solution to 4-fold with the 2.5 mg/4 ml solution. The ultrasonic nebulizer performed best and worse when placed at the humidifier and Y-piece, respectively, for all tested solutions (P.01) (see Fig. 3). These results were true for lung dose and delivery efficiency (see Table 2). The difference in lung dose ranged from 2-fold with the 2.5 mg/3 ml solution to 5.7-fold with the 7.5 mg/4 ml solution. The Salter nebulizer performed best and worse when placed at either the humidifier or ventilator and Y-piece, respectively, for all tested solutions (P.01) (see Fig. 3). These results were true for lung dose and delivery efficiency (see Table 2). The difference in lung dose ranged from 1.6-fold with the 2.5 mg/3 ml solution to 2.9-fold with the 7.5 mg/4 ml solution. Inter-device Comparison When placed at the ventilator, the vibrating mesh nebulizer outperformed all other delivery devices, irrespective of solution tested (P.001) (Fig. 4). These results were true for lung dose and delivery efficiency (see Table 2). The ultrasonic nebulizer achieved higher lung dose than the jet nebulizers for some of the solutions tested. Lung dose differences among devices ranged from 5.7-fold with the 2.5 mg/3.5 ml solution to 9.3-fold with the 2.5 mg/ 3 ml solution. When placed at the humidifier, the vibrating mesh nebulizer outperformed all other delivery devices with all solutions tested, except for the ultrasonic nebulizer using the 7.5 mg/4 ml solution (P.001) (see Fig. 4). These results were true for lung dose and delivery efficiency (see Table 2). The ultrasonic nebulizer was also superior to both jet nebulizers (P.01). Lung dose differences among devices ranged from 7.1-fold with the 2.5 mg/4 ml solution to 8.2-fold with the 5.0 mg/3.5 ml solution. When placed at the Y-piece, the vibrating mesh nebulizer outperformed both jet nebulizers for all solutions tested (P.001) (see Fig. 4). The vibrating mesh nebulizer was only superior to the ultrasonic nebulizer when using the 7.5 mg/4 ml solution (P.01). These results were true for lung dose and delivery efficiency (see Table 2). Differences in lung dose among devices ranged from 3.7-fold with the 5.0 mg/4 ml solution to 11.2-fold with the 7.5 mg/ 4 ml solution. When placed 30 cm before the Y-piece, the vibrating mesh nebulizer outperformed both jet nebulizers, irrespective of solution tested (P.001) (see Fig. 4). The vibrating mesh nebulizer also performed better than the ultrasonic nebulizer with the 7.5 mg/4 ml solution. These results were true for lung dose and delivery efficiency (see Table 2). The ultrasonic nebulizer was also superior to both jet nebulizers (P.01). Lung dose differences among devices ranged from 3.5-fold with the 2.5 mg/3 ml solution to 5.4-fold with the 7.5 mg/4 ml solution. Evaluation of Changes in Loading Volume/Nominal Dose Effect on Lung Dose When albuterol solutions of 2.5 mg/3.5 ml and 5.0 mg/ 3.5 ml were compared, lung dose was higher with the latter in all positions and nebulizers tested (P.001), except for the ultrasonic nebulizer at the ventilator (P.29) (Table 3). RESPIRATORY CARE JULY 2013 VOL 58 NO
5 Fig. 3. Intra-device comparison of albuterol captured at the end of the endotracheal tube (lung dose). With the Updraft II Opti-Neb jet nebulizer, lung dose with the 2.5 mg/3 ml solution when placed at the Y-piece was lower than all, and lung dose with the 5.0 mg/3.5 ml and 7.5 mg/4 ml solutions when positioned at the ventilator was similar to humidifier and was greater than at the Y-piece and 30 cm from the Y-piece. With the Salter 8900 jet nebulizer, lung dose with the 2.5 mg/3 ml solution when placed at the humidifier was greater than all other positions, and lung dose with the 5.0 mg/3.5 ml and 7.5 mg/4 ml solutions when positioned at the ventilator was similar to humidifier and was greater than Y-piece and 30 cm from the Y-piece. With the Maquet ultrasonic nebulizer, lung dose with the 2.5 mg/3 ml solution when placed at the humidifier was greater than all other positions, and also lung dose was greater at the ventilator than at the Y-piece; output with the 5.0 mg/3.5 ml solution when placed at the humidifier was greater than the Y-piece and 30 cm from the Y-piece and lung dose was greater at the ventilator than at the Y-piece; lung dose with the 7.5 mg/4 ml solution when placed at the humidifier was greater than all other positions, and lung dose was greater at the ventilator than at the Y-piece and 30 cm from the Y-piece. With the Aerogen Solo vibrating mesh nebulizer, lung dose with all solutions tested when positioned at the ventilator was similar to humidifier and was greater than Y-piece and 30 cm from the Y-piece. When albuterol solutions of 2.5 mg/4 ml and 7.5 mg/ 4 ml were compared, lung dose was higher with the latter in all positions and nebulizers tested (P.001), except for the Salter nebulizer at the Y-piece (P.09) (see Table 3). When albuterol solutions of 2.5 mg/3 ml and 2.5 mg/ 4 ml were compared, lung dose was higher with the latter except for the Hudson nebulizer at the Y-piece (P.08), the Salter nebulizer at the humidifier (P.39), the ultrasonic nebulizer at the ventilator (P.21) and the Y-piece (P.17). No differences were noted for the vibrating mesh nebulizer (P.54, P.41, and P.64, when placed at the ventilator, humidifier, or Y-piece, respectively) (see Table 3). Discussion Mechanically ventilated pediatric patients are often prescribed nebulized albuterol. Delivery device and position in the ventilator circuit play an important role in determining how much albuterol reaches the end of the ETT. We compared 4 different devices placed in 4 different 1128 RESPIRATORY CARE JULY 2013 VOL 58 NO 7
6 Table 2. Delivery Efficiency* Albuterol Nominal Dose/ Loading Volume Nebulizer Ventilator Humidifier Nebulizer Position Y-Piece 30 cm Before Y-piece 2.5 mg/3 ml Hudson Updraft II Opti-Neb Salter Maquet Ultrasonic Aerogen Solo vibrating mesh mg/3.5 ml Hudson Updraft II Opti-Neb Salter Maquet Ultrasonic Aerogen Solo vibrating mesh mg/4 ml Hudson Updraft II Opti-Neb Salter Maquet Ultrasonic Aerogen Solo vibrating mesh * Values are mean SD of 5 measurements, expressed as percentage and calculated as drug captured in the filter ( g)/drug placed in the nebulizer ( g) 100. Fig. 4. Inter-device comparison. * Deposition with the Aerogen Solo vibrating mesh nebulizer was greater than with all other nebulizers (P.01). Deposition with the Maquet ultrasonic nebulizer was greater than with the Salter 8900 nebulizer (P.05). Deposition with the Maquet ultrasonic nebulizer was greater than with the Hudson Updraft II Opti-Neb nebulizer (P.05). Deposition with the Aerogen Solo vibrating mesh nebulizer was greater than with the Hudson Updraft II Opti-Neb and Salter 8900 nebulizers (P.01). positions of the pediatric ventilator circuit model. The vibrating mesh device delivered the most albuterol to the end of the ETT. The Y-piece and either ventilator or humidifier were the worst and best positions to place the RESPIRATORY CARE JULY 2013 VOL 58 NO
7 Table 3. Evaluation of the Effect of Changes in Nominal Dose and Loading Volume on Lung Dose Nebulizer Albuterol Nominal Dose/Loading Volume Ventilator Nebulizer Position* Humidifier Y-Piece Hudson Updraft II Opti-Neb 2.5 mg/3 ml mg/4 ml mg/4 ml mg/3.5 ml mg/3.5 ml Salter mg/3 ml mg/4 ml mg/4 ml mg/3.5 ml mg/3.5 ml Maquet Ultrasonic 2.5 mg/3 ml mg/4 ml mg/4 ml 1, , mg/3.5 ml mg/3.5 ml Aerogen Solo vibrating mesh 2.5 mg/3 ml mg/4 ml mg/4 ml 2, , , mg/3.5 ml mg/3.5 ml 1, , * Values are mean SD lung dose. Drug output for 2.5 mg/4 ml was higher than for 2.5 mg/3 ml at this position. Drug output for 7.5 mg/4 ml was higher than for 2.5 mg/4 ml at all positions. Drug output for 5.0 mg/3.5 ml was higher than for 2.5 mg/3.5 ml at all positions. Drug output for 5.0 mg/3.5 ml was higher than for 2.5 mg/3.5 ml at this position. Drug output for 5.0 mg/3.5 ml was higher than for 2.5 mg/3.5 ml at this position. nebulizers. Increasing the loading volume without increasing the amount of drug does not increase lung dose when using vibrating mesh nebulizers. The problem of the effect of position has been previously looked at using adult type ventilator models In this study we provide lung dose data achieved by different aerosol generators located on 4 different places in a pediatric ventilator model. We found that both jet nebulizers performed better when placed at the ventilator or the humidifier. These data are in agreement with others Hughes et al and Quinn also found that lung dose increased when the generating device was placed closer to the ventilator (V T of 700 ml and 400 ml, respectively). 9,10 More recently, Ari et al compared lung dose of 3 different nebulizers (jet, ultrasonic, and vibrating mesh) placed in 3 positions (between the ETT and Y-piece, 15 cm from the Y-piece, and 15 cm from the ventilator). 11 They used an adult model of mechanical ventilation, without bias flow, and with/without humidification (V T 500 ml, breathing frequency 15 breaths/min, ETT 8 mm). They also found that lung dose was greater when jet nebulizers were placed at the ventilator. However, our current data are conflicting with other pediatric data using lower V T. 6,12 In a previous study, using a ventilator model similar to the one used in the present study, we compared lung dose achieved by a jet nebulizer (Salter 8900) and an intrapulmonary percussive ventilator device placed after the humidifier or before the Y-piece. 6 We found no difference in lung dose for the jet nebulizer when placed before the Y-piece or at the humidifier. In a follow-up study, Ari et al compared lung dose of a jet nebulizer placed before 15 cm before the Y-piece and 15 cm before the humidifier with a vibrating mesh nebulizer placed at the ETT and right before the humidifier. 12 They used humidified pediatric (V T 100 ml, breathing frequency 20 breaths/min, ETT 5 mm) and adult (V T 500 ml, breathing frequency 20 breaths/min, ETT 8 mm) ventilator models, with bias flow of either 2 or 5 L/min. They found no difference in lung dose for different positions for the jet nebulizer. One difference between their study and ours is that the jet nebulizers were operated at 2.5 L/min and 6 L/min in their study and our study, respectively. We speculate that the differences in lung dose seen between adult and pediatric models and 1130 RESPIRATORY CARE JULY 2013 VOL 58 NO 7
8 between pediatric models of different V T might be because a certain V T is required for the differences in filter deposition to occur. The ultrasonic nebulizer performed best when placed at the humidifier. These data are conflicting with previously published data using an adult ventilator model. 12,15,16 Thomas et al, 15 using an adult model (ETT 9 mm, breathing frequency 15 breaths/min, inspiratory/expiratory ratio of 1:3, V T 900 ml), placed the devices in the inspiratory limb at 50 cm and right before the Y-piece, and found no differences in lung dose. 16 Ari et al, using an adult ventilator model, with no bias flow, found that placing the device 15 cm before the Y-piece provided the highest lung dose. The differences between their studies and ours could be due to several factors such as different ETT sizes, different ventilator settings, different devices, presence or absence of bias flow, and different placement of the devices. The vibrating mesh nebulizer performed best when placed at either the humidifier or at the ventilator. Our results confirm those of Ari et al using a pediatric model. 12 However, our delivery efficiency was 2-fold higher than the one reported by Ari et al. This difference could be due to a variety of factors such as the fact that we used a larger ventilator circuit (22 mm vs 15 mm inner diameter), a slightly larger ETT (5.5 mm vs 5.0 mm inner diameter), and a larger V T (200 ml vs 100 ml). In addition, the nebulizer was positioned in a slightly different way in our study: the nebulizer was connected immediately after the ventilator, while in their study a corrugated tube was interposed between the ventilator and the nebulizer that was connected to the humidifier. We also provide novel data on the placement of a vibrating mesh nebulizer in a pediatric ventilator model. However, these data are in conflict with the data Ari et al obtained using an adult model that showed lung dose was greater when the device was placed either before or at the Y-piece. 11 The differences are most likely due to the lack of bias flow in their study, since the same group showed a 2-fold increase in lung dose when moving the device from right before the Y-piece to before the humidifier when using a circuit with bias flow. 12 We found that in a pediatric ventilator model the 2 jet nebulizers tested and the vibrating mesh nebulizer delivered more albuterol to the end of the ETT when placed at either the ventilator or the humidifier. The ultrasonic nebulizer performed best at the humidifier. The worst performance was achieved by placing the nebulizer at the Y- piece. The position 30 cm before the Y-piece rendered similar results. These data confirm previous work showing that nebulizers are more efficient when placed farther away from the ETT. 9,10,12 We speculate that the ventilator circuit acts as a reservoir. In addition, the speed of the aerosol that reaches the Y-piece will decrease when the aerosol generator is placed away from the patient, therefore decreasing impaction of emitted aerosol. The influence of the type of device on lung dose in ventilator models and humans has also been previously studied. 6-8,11,12,16-18 However, there is paucity of data on the newly available technologies. In our study the vibrating mesh nebulizer outperformed 2 jet nebulizers in each position tested, and the ultrasonic nebulizer either at the ventilator or at the humidifier. O Doherty et al found no difference among a jet and 2 ultrasonic nebulizers when placed before the Y-piece and using a 3 ml nebulizer load. 8 Conversely, Harvey et al, in an in vivo study, demonstrated a 2-fold higher lung deposition with an ultrasonic nebulizer, compared to a jet nebulizer, placed before the Y-piece in an adult ventilator. 16 Similar results were obtained in our study. In a previous study we found that a jet nebulizer outperformed the intrapulmonary percussive ventilator when placed at humidifier. 6 Fok et al, in 2 in vivo studies (animal and neonatal), demonstrated a several-fold higher pulmonary deposition of ultrasonic nebulizers, compared to a jet nebulizer. 17,18 Our data are in agreement with their findings. Our data are also in partial agreement with results from Ari et al, who found in an adult mechanical ventilation model that the vibrating mesh outperformed the jet nebulizer but not the ultrasonic nebulizer when the devices were placed either at the Y-piece or 15 cm before the Y-piece. 11 They found no difference between devices when they were placed before the humidifier. In contrast, we found that the vibrating mesh nebulizer outperformed all other devices. These differences are most likely due to the fact that our circuit had a 2 L/min bias flow and theirs had none. The same group reported on a pediatric and an adult ventilator model that the vibrating mesh nebulizer had a 2-fold lung dose difference, when compared to the jet nebulizer. 12 Our data confirm their findings regarding the performance of vibrating mesh nebulizers. We speculate that the vibrating mesh device achieved a higher lung dose because of its design and operational characteristics. More specifically, it has a minimal dead space and it does not add extra flow to the ventilator circuit. Other authors have previously found that higher bias flow and higher nebulizer flow rates lead to lower lung doses. 12,19 We found that in a pediatric ventilator model the vibrating mesh nebulizer was superior to both the jet nebulizers at all tested positions, and to the ultrasonic nebulizer when placed at either the ventilator or the humidifier. When evaluating the effect of nebulizer loading volume/nominal dose on lung dose, we found that the increase in lung dose is not due only to an increase in loading volume, but also due to an increase in the nominal dose. This was reflected in the comparison between solutions of equal volume but different amount of albuterol (2.5 mg/3.5 ml vs 5.0 mg/3.5 ml, and 2.5 mg/4 ml vs 7.5 mg/4 ml). We also found that increasing the loading RESPIRATORY CARE JULY 2013 VOL 58 NO
9 volume increased lung dose for the jet and ultrasonic nebulizers but not for the vibrating mesh nebulizer. This was reflected in the comparison between 2 solutions with equal amounts of albuterol but different loading volumes. We speculate that these differences are due to the design characteristics of different nebulizer types, especially the amount of dead volume. These data are in agreement with previously published data on jet nebulizers not tested on a ventilator circuit. 13 These data are also in agreement with O Doherty et al, who found a significant increase in lung dose when increasing the loading volume of jet and ultrasonic nebulizers from 3 ml to 6 ml in an adult ventilator model. 8 This study also demonstrated that increasing the nominal dose will not necessarily increase the lung dose. The vibrating mesh nebulizer showed an increase in lung dose irrespective of the position in the circuit. However, both jet nebulizers and the ultrasonic nebulizer failed to increase lung dose when placed at the Y-piece. This was also true for both jet nebulizers when positioned 30 cm from the Y-piece. We found that in a pediatric ventilator model the increase in loading volume without increasing the nominal dose increases lung dose in jet and ultrasonic nebulizers, but not in vibrating mesh ones. Our study has 2 limitations that are inherent to the in vitro nature of the study and the ventilator setup. The drug captured at the filter (lung dose) overestimates the amount of drug reaching the patient, because part of the exhaled aerosol is retained on it. This is critical for submicronic aerosols. However, this is a well established methodology that has been validated against human studies. Another limitation is that this study did not address the aerodynamic characteristics of the aerosols reaching the tip of the ETT. However, the mass median aerodynamic diameter of the aerosols generated by the nebulizers alone ranged from 3.1 m to4.4 m (data from our laboratory and manufacturer s information). In addition, Watterberg et al, in an in vitro study, found that, although a jet nebulizer and a submicronic ultrasonic nebulizer differed in particle size when measured at the nebulizer, those differences disappeared when measured at the tip of a 4.0 mm ETT. 20 Healthcare team members are now provided with delivery efficiency, lung dose, and cost data that should be used in conjunction with devices and patient characteristics, along with type of drug chosen for nebulization to personalize the care for a given patient. Conclusions This study provides a comparison of lung dose achieved by 4 devices placed in 4 different positions in the ventilator circuit, using different nominal doses of albuterol solution. In an in vitro pediatric ventilator model, a vibrating mesh nebulizer outperformed jet and an ultrasonic nebulizer, irrespective of their position in the ventilator circuit. The nebulizers were more efficient when placed at either the ventilator or the humidifier, and less efficient when placed at either the Y-piece or 30 cm from the Y- piece. The conclusions of this study are valid for the specific tested conditions. Conclusions regarding optimization of operating conditions should not be extrapolated among nebulizers of different operating principles. One important corollary of this work is that aerosol generators are part of a complex delivery device that also includes position of the device in the ventilator circuit, humidification, and type of ventilator circuit and ETT. REFERENCES 1. Everard ML. Ethical aspects of using radiolabelling in aerosol research. Arch Dis Child 2003;88(8): Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ. Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors. Am J Respir Crit Care Med 1999;159(1): Miller DD, Amin MM, Palmer LB, Shah AR, Smaldone GC. Aerosol delivery and modern mechanical ventilation: in vitro/in vivo evaluation. Am J Respir Crit Care Med 2003;168(10): Dhand R. Aerosol delivery during mechanical ventilation: from basic techniques to new devices. J Aerosol Med Pulm Drug Deliv 2008; 21(1): Fink JB. Aerosol delivery to ventilated infant and pediatric patients. Respir Care 2004;49(6): Berlinski A, Willis JR. Albuterol delivery by intrapulmonary percussive ventilator and jet nebulizer in a pediatric ventilator model. Respir Care 2010;55(12): O Riordan TG, Greco MJ, Perry RJ, Smaldone GC. Nebulizer function during mechanical ventilation. Am Rev Respir Dis 1992;145(5): O Doherty MJ, Thomas SH, Page CJ, Treacher DF, Nunan TO. Delivery of a nebulized aerosol to a lung model during mechanical ventilation. Effect of ventilator settings and nebulizer type, position, and volume of fill. Am Rev Respir Dis 1992;146(2): Hughes JM, Saez J. Effects of nebulizer mode and position in a mechanical ventilator circuit on dose efficiency. Respir Care 1987; 32(11): Quinn WW. Effect of a new nebulizer position on aerosol delivery during mechanical ventilation: a bench study. Respir Care 1992; 37(5): Ari A, Areabi H, Fink JB. Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified circuits during adult mechanical ventilation. Respir Care 2010;55(7): Ari A, alay OT, Harwood R, Sheard MM, Aljamhan EA, Fink JB. Influence of nebulizer type, position, and bias flow on aerosol drug delivery in simulated pediatric and adult lung models during mechanical ventilation. Respir Care 2010;55(7): Hess D, Fisher D, Williams P, Pooler S, Kacmarek RM. Medication nebulizer performance. Effects of diluent volume, nebulizer flow, and nebulizer brand. Chest 1996;110(2): Berlinski A, Waldrep JC. Four hours of continuous albuterol nebulization. Chest 1998;114(3): Thomas SHL, O Doherty MJ, Page CJ, Treacher DF, and Nunan T. Delivery of ultrasonic nebulized aerosols to a lung model during mechanical ventilation. Am Rev Respir Dis 1993;148: Harvey CJ, O Doherty MJ, Page CJ, Thomas SHL, Nunan TO, Treacher DF. Comparison of jet and ultrasonic nebulizer pulmonary 1132 RESPIRATORY CARE JULY 2013 VOL 58 NO 7
10 aerosol deposition during mechanical ventilation. Eur Respir J 1997; 10(4): Fok TF, Al-Essa M, Monkman S, Dolovich M, Girard L, Coates G, Kirpalani H. Pulmonary deposition of salbutamol aerosol delivered by metered dose inhaler, jet nebulizer, and ultrasonic nebulizer in mechanically ventilated rabbits. Pediatr Res 1997;42(5): Fok TF, Lam K, Ng PC, Leung TF, So HK, Cheung KL, Wong W. Delivery of salbutamol to nonventilated preterm infants by metereddose inhaler, jet nebulizer, and ultrasonic nebulizer. Eur Respir J 1998;12(1): Coleman DM, Kelly HW, McWilliams BC. Determinants of aerosolized albuterol delivery to mechanically ventilated infants. Chest 1996;109(6): Erratum in: Chest 1996;110(4): Watterberg KL, Clark AR, Kelly W, Murphy S. Delivery of aerosolized medication to intubated babies. Pediatr Pulmonol 1991;10(2): This article is approved for Continuing Respiratory Care Education credit. For information and to obtain your CRCE (free to AARC members) visit RESPIRATORY CARE JULY 2013 VOL 58 NO
Albuterol Delivery During Noninvasive Ventilation
Albuterol Delivery During Noninvasive Ventilation Matthew P Branconnier RRT and Dean R Hess PhD RRT FAARC HYPOTHESIS: Albuterol delivered during noninvasive positive-pressure ventilation is affected by
New Technology Offers New Opportunities: Continuous Bronchodilator Therapy. During Mechanical Ventilation. James B. Fink, MS, RRT, FAARC
New Technology Offers New Opportunities: Continuous Bronchodilator Therapy During Mechanical Ventilation James B. Fink, MS, RRT, FAARC James B. Fink, MS, RRT, FAARC Fellow Respiratory Science Key words:
Neurally Adjusted Ventilatory Assist: NAVA for Neonates
Neurally Adjusted Ventilatory Assist: NAVA for Neonates Robert L. Chatburn, MHHS, RRT-NPS, FAARC Research Manager Respiratory Institute Cleveland Clinic Professor Department of Medicine Lerner College
Ventilation Perfusion Relationships
Ventilation Perfusion Relationships VENTILATION PERFUSION RATIO Ideally, each alveolus in the lungs would receive the same amount of ventilation and pulmonary capillary blood flow (perfusion). In reality,
POLICIES & PROCEDURES. ID Number: 1115
POLICIES & PROCEDURES Title: OXYGEN ADMINISTRATION ID Number: 1115 Authorization [X] SHR Nursing Practice Committee Source: Nursing, Respiratory Therapy, Physiotherapy Date Revised: March 2015 Date of
MECHINICAL VENTILATION S. Kache, MD
MECHINICAL VENTILATION S. Kache, MD Spontaneous respiration vs. Mechanical ventilation Natural spontaneous ventilation occurs when the respiratory muscles, diaphragm and intercostal muscles pull on the
Suction Catheter Size: An Assessment and Comparison of 3 Different Calculation Methods
Suction Catheter Size: An Assessment and Comparison of 3 Different Calculation Methods Christopher J Russian MEd RRT-NPS RPSGT RST, Joshua F Gonzales MHA RRT-NPS RRT-SDS, and Nicholas R Henry MSC RRT-NPS
For every breath he takes. Trilogy200 ventilator s added sensitivity lets you breathe easier knowing your patients are where they belong home.
For every breath he takes Trilogy200 ventilator s added sensitivity lets you breathe easier knowing your patients are where they belong home. Sensitive to your patients needs Trilogy200, a portable life-support
VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE
VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE Critical Care SERVO-s 3 SERVO-s SIMPLY MAKES SENSE MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international provider of medical products
Oxygenation and Oxygen Therapy Michael Billow, D.O.
Oxygenation and Oxygen Therapy Michael Billow, D.O. The delivery of oxygen to all body tissues is the essence of critical care. Patients in respiratory distress/failure come easily to mind as the ones
5/30/2014 OBJECTIVES THE ROLE OF A RESPIRATORY THERAPIST IN THE DELIVERY ROOM. Disclosure
THE ROLE OF A RESPIRATORY THERAPIST IN THE DELIVERY ROOM Ona Fofah, MD FAAP Assistant Professor of Pediatrics Director, Division of Neonatology Department of Pediatrics Rutgers- NJMS, Newark OBJECTIVES
Clinical Foundations A Patient-focused Education Program for Respiratory Care Professionals
Clinical Foundations A Patient-focused Education Program for Respiratory Care Professionals Free Continuing Education for Respiratory Therapists (CRCE) and Nurses (CNE) See Page 12 Advisory Board Richard
Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003
Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003 Knowing the functions of the anesthetic delivery system
- Canine Chronic Bronchitis cannot be cured, but can be controlled
MD-096A-0211 1 2 - The symptoms of asthma can mimic other diseases such as heartworm, pneumonia and congestive heart failure (Padrid, Use of Inhaled Medications to Treat Respiratory Diseases in Dogs and
Artificial Ventilation Theory into practice
Artificial Ventilation Theory into practice Keith Simpson BVSc MRCVS MIET(Electronics) www.vetronic.co.uk [email protected] June 13 th 2014 Today we will discuss the administration of IPPV to anaesthetised
Nurses and Respiratory Therapists Working Together for Safe Alarm Systems Management
Nurses and Respiratory Therapists Working Together for Safe Alarm Systems Management May 11, 2015 9/25/2013 1 AAMI Foundation Vision: To drive the safe adoption and use of healthcare technology Visit our
Nocturnal Oxygenation Using a Pulsed-Dose Oxygen-Conserving Device Compared to Continuous Flow
Nocturnal Oxygenation Using a Pulsed-Dose Oxygen-Conserving Device Compared to Continuous Robert L Chatburn RRT-NPS FAARC, Joseph S Lewarski RRT FAARC, and Robert W McCoy RRT FAARC BACKGROUND: The pulsed-dose
The Anesthesia Ventilator
The Anesthesia Ventilator Why is the piston replacing the bellows? For many decades, and millions of anesthetics, the bellows anesthesia ventilator has been a safe and effective clinical device. Indeed,
Forced vital capacity: maximal volume of air exhaled with maximally forced effort from a maximal inspiration.
SOP Spirometry 1. General considerations Spirometry serves as a physiological test to quantify pulmonary disease severity and to assess clinical change in respiratory function over time. Standard spirometric
High Flow Through a Nasal Cannula and CPAP Effect in a Simulated Infant Model
High Flow Through a Nasal Cannula and CPAP Effect in a Simulated Infant Model Teresa A Volsko MHHS RRT FAARC, Kathy Fedor RRT-NPS, Jason Amadei RRT, and Robert L Chatburn MHHS RRT-NPS FAARC BACKGROUND:
Critical Care Therapy and Respiratory Care Section
Category: Section: Title: Critical Care Therapy and Respiratory Care Section Policy #: 03 Revised: 5/02 1.0 DESCRIPTION Clinical Ventilator Management Invasive and Noninvasive Positive Pressure Ventilation
Survey of Directors of Respiratory Therapy Departments Regarding the Future Education and Credentialing of Respiratory Care Students and Staff
Survey of Directors of Respiratory Therapy Departments Regarding the Future Education and Credentialing of Respiratory Care Students and Staff Robert M Kacmarek PhD RRT FAARC, Thomas A Barnes EdD RRT FAARC,
Common Mistakes with inhaler Technique
Common Mistakes with inhaler Technique Inhaling medication into the lungs can be very effective - but frequently it is not. The most common reasons for patients not responding to treatment are:- Being
SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / 8353. Instruction Manual. Page 1 of 9
SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / 8353 Instruction Manual Page 1 of 9 TABLE OF CONTENTS Introduction...3 Important Safeguards..3 System Components...5 Setup and Operation...5 Setup...5 Operation....6
Is Albuterol Sulfate Inhalation Aerosol A Steroid
Is Albuterol Sulfate Inhalation Aerosol A Steroid buy albuterol sulfate solution for nebulizer albuterol sulfate inhaler order online metered doses in combivent albuterol sulfate inhaler albuterol sulfate
Evaluation of a Nasal Cannula in Noninvasive Ventilation Using a Lung Simulator
Evaluation of a Nasal in Noninvasive Ventilation Using a Lung Simulator Narayan P Iyer MD and Robert Chatburn MHHS RRT-NPS FAARC BACKGROUND: Nasal noninvasive ventilation (NIV) is a common form of noninvasive
Coding Guidelines for Certain Respiratory Care Services July 2014
Coding Guidelines for Certain Respiratory Care Services Overview From time to time the AARC receives inquiries about respiratory-related coding and coverage issues through its Help Line or Coding Listserv.
Battery Performance of 4 Intensive Care Ventilator Models
Battery Performance of 4 Intensive Care Ventilator Models Thomas C Blakeman RRT, Bryce RH Robinson MD, and Richard D Branson MSc RRT FAARC BACKGROUND: Hospital electrical power failure represents an important
AHA Instructor Renewal
AHA Instructor Renewal Thank you for your continuing on as American Heart Association AHA Instructor! In this packet you will we have provided information on: A detailed outline of the process to renew
From AARC Protocol Committee; Subcommittee Adult Critical Care Version 1.0a (Sept., 2003), Subcommittee Chair, Susan P. Pilbeam
AARC - ADULT MECHANICAL VENTILATOR PROTOCOLS 1. Guidelines for Using Ventilator Protocols 2. Definition of Modes and Suggestions for Use of Modes 3. Adult Respiratory Ventilator Protocol - Guidelines for
Ki m b e r l y-cl a r k* 72-Hour Closed-Suction Systems. Ba l l a r d* Tr a c h Ca r e* System. A unique design. A new standard in clean.
Ki m b e r l y-cl a r k* 72-Hour Closed-Suction Systems Ba l l a r d* Tr a c h Ca r e* System A unique design. A new standard in clean. Ki m b e r l y-cl a r k* 72-Hour Closed-Suction Systems Ba l l a
INTRODUCING RESMED S. Home NIV Solutions. S9 VPAP ST-A with ivaps S9 VPAP ST. Why choose average when you can choose intelligent?
INTRODUCING RESMED S Home NIV Solutions S9 VPAP ST-A with ivaps S9 VPAP ST Why choose average when you can choose intelligent? Now you can provide intelligent air through ResMed s intelligent Volume-Assured
GRADE 11F: Biology 3. UNIT 11FB.3 9 hours. Human gas exchange system and health. Resources. About this unit. Previous learning.
GRADE 11F: Biology 3 Human gas exchange system and health UNIT 11FB.3 9 hours About this unit This unit is the third of six units on biology for Grade 11 foundation. The unit is designed to guide your
Respiratory Care. A Life and Breath Career for You!
Respiratory Care A Life and Breath Career for You! Respiratory Care Makes a Difference At 9:32 am, Lori Moreno brought a newborn baby struggling to breathe back to life What have you accomplished today?
Ventilator Application of the Passy-Muir Valve David A. Muir Course Outline Benefits Review of the Biased Closed Position No Leak Passy-Muir Valves
Ventilator Application of the Passy-Muir Valve Michael S. Harrell, B.S., RRT Director of Clinical Education-Respiratory [email protected] (949) 833-8255 David A. Muir 23 year-old ventilator dependent
Airway Pressure Release Ventilation
Page: 1 Policy #: 25.01.153 Issued: 4-1-2006 Reviewed/ Revised: Section: 10-11-2006 Respiratory Care Airway Pressure Release Ventilation Description/Definition Airway Pressure Release Ventilation (APRV)
High-Frequency Oscillatory Ventilation
High-Frequency Oscillatory Ventilation Arthur Jones EdD, RRT Learning Objectives Describe the indications and rationale and monitoring for HFOV. Identify HFOV settings and describe the effects of their
Ventilators/Transport*
Ventilators/Transport* pneuton A (Pneumatic Transport Ventilator) pneuton (pronounced new ton ); a small, lightweight transport ventilator designed for use on patients from pediatric to adult in size.
Neonatal Reference Guide
Operated by REACH Air Medical Services Assessment Heart Rate (beats/min.) Age Rate
AHA/AAP Neonatal Resuscitation Guidelines 2010: Summary of Major Changes and Comment on its Utility in Resource-Limited Settings
AHA/AAP Neonatal Resuscitation Guidelines 2010: Summary of Major Changes and Comment on its Utility in Resource-Limited Settings Resuscitation step Recommendations (2005) Recommendations (2010) Comments/LOE
University of Kansas. Respiratory Care Education
University of Kansas Respiratory Care Education What is Respiratory Care? Respiratory Care is the health profession that specializes in the promotion of optimum cardiopulmonary function and health Respiratory
BY UV SPECTROPHOTOMETRY
1 of 9 1. PURPOSE This Standard Operating Procedure describes the method for measuring the concentration of viral particles using UV/Vis spectrophotometry at 260 and 280 nm. 2. SCOPE This procedure is
Common Ventilator Management Issues
Common Ventilator Management Issues William Janssen, M.D. Assistant Professor of Medicine National Jewish Health University of Colorado Denver Health Sciences Center You have just admitted a 28 year-old
Home Ventilator. 2009 Breath of Life Home Medical Equipment and Respiratory Services
Home Ventilator 2009 Breath of Life Home Medical Equipment and Respiratory Services 1 Your Prescription Your physician has prescribed a home ventilator system for your child. This book will help you to
POCKET GUIDE. NAVA and NIV NAVA in neonatal settings
POCKET GUIDE NAVA and NIV NAVA in neonatal settings Table of contents EMPTY 1 2 3 4 Introduction and background facts Invasive ventilation with NAVA Non invasive ventilation with NAVA NAVA and NIV NAVA
NOC: 3214. Occupation: Respiratory Therapist
NOC: 3214 Occupation: Respiratory Therapist Occupation Description: Responsibilities include assisting in the diagnosis, treatment and care of residents with respiratory and cardiopulmonary disorders.
«The compact and mobile testing device for ventilators.»
«The compact and mobile testing device for ventilators.» Developed for mobile use simple. compact. reliable. Simple operation CITREX is simple and intuitive to operate. The colour screen offers excellent
Boaray 700 Anesthesia Machine
Boaray 700 Anesthesia Machine Features CE certification. 15 inch colored LCD, touch screen setting, big font display, free changeable angles of views. Various ventilation modes - VCV, PCV, SIMV+VCV, SIMV+PCV.
1.40 Prevention of Nosocomial Pneumonia
1.40 Purpose Audience Policy Statement: The guideline is designed to reduce the incidence of pneumonia and other acute lower respiratory tract infections. All UTMB healthcare workers and contract healthcare
Respiratory Care Solutions Improving Quality of Life
Respiratory Care Solutions Improving Quality of Life Improving quality of life The need to rehabilitate patients effectively after respiratory disorders, is of extreme importance, not only to decrease
The Anesthesia Ventilator
The Anesthesia Ventilator Why is the piston replacing the bellows? For many decades, and millions of anesthetics, the bellows anesthesia ventilator has been a safe and effective clinical device. Indeed,
Sensors / Modules / Monitors. Breath Gas Sensing and Monitoring. For life supporting systems in intensive, emergency and ambulant care
Sensors / Modules / Monitors Breath Gas Sensing and Monitoring For life supporting systems in intensive, emergency and ambulant care Your competent partner for breath gas sensing and monitoring EnviteC
GUIDELINES FOR THE MANAGEMENT OF OXYGEN THERAPY
SOUTH DURHAM HEALTH CARE NHS TRUST GUIDELINES FOR THE MANAGEMENT OF OXYGEN THERAPY AIM To supplement oxygen intake using the appropriate equipment in order to correct hypoxia and relieve breathlessness.
Laryngeal Mask Airways (LMA), Indications and Use for the Pre-Hospital Provider. www.umke.org
Laryngeal Mask Airways (LMA), Indications and Use for the Pre-Hospital Provider Objectives: Identify the indications, contraindications and side effects of LMA use. Identify the equipment necessary for
PLEUR EVAC CHEST DRAINAGE SYSTEMS. Great Innovations to Enhance Chest Drainage
PLEUR EVAC CHEST DRAINAGE SYSTEMS Great Innovations to Enhance Chest Drainage 2 INTRODUCTION Teleflex Great Innovations to Enhance Chest Drainage Trusted brands make Teleflex a reliable and strong partner.
Nasal-Aire II Critical Care Single-use CPAP interface by InnoMed The comfortable, economical solution to non-invasive ventilation!
CPAP Nasal-Aire II Critical Care Single-use CPAP interface by InnoMed The comfortable, economical solution to non-invasive ventilation! Advantages:. Can be worn for any duration without discomfort-no uncomfortable
HUDSON RCI RESPIRATORY SOLUTIONS. Breathing, made easier.
HUDSON RCI RESPIRATORY SOLUTIONS Breathing, made easier. 1 BREATHING, MADE EASIER. As a leader in respiratory care for more than 65 years, Hudson RCI is dedicated to providing innovative products that
Heliox During Mechanical Ventilation
Heliox During Mechanical Ventilation Shekhar T Venkataraman MD Introduction Prediction of Physiologic Changes With Heliox During Mechanical Ventilation Lower-Airway Obstruction Noninvasive Ventilation
CHAPTER THREE COMMON DESCRIPTIVE STATISTICS COMMON DESCRIPTIVE STATISTICS / 13
COMMON DESCRIPTIVE STATISTICS / 13 CHAPTER THREE COMMON DESCRIPTIVE STATISTICS The analysis of data begins with descriptive statistics such as the mean, median, mode, range, standard deviation, variance,
Veterinary Clinics and Hospitals
Anesthetic Gases: Guidelines for Workplace Exposure Veterinary Clinics and Hospitals Inhalation anesthesia in veterinary hospitals is practiced in a manner similar to that in human hospitals. Generally,
Preventing Medication Errors in EMS
Preventing Medication Errors in EMS Examples and Case Reports John Gallagher MD Phoenix Fire Dept. 1 Six R s Right Medication Right Dose Right Time Right Route Right Patient Right Documentation Bryan Bledsoe
TRANSPORT OF CRITICALLY ILL PATIENTS
TRANSPORT OF CRITICALLY ILL PATIENTS Introduction Inter-hospital and intra-hospital transport of critically ill patients places the patient at risk of adverse events and increased morbidity and mortality.
Comparison of the Rate of Improvement in Gas Exchange between Two High Frequency Ventilators in a Newborn Piglet Lung Injury Model
Comparison of the Rate of Improvement in Gas Exchange between Two High Frequency Ventilators in a Newborn Piglet Lung Injury Model Kurt Gillette, MD, San Antonio Military Medical Center Background: High
Allergy Emergency Treatment Protocol
Allergy Emergency Treatment Protocol I. Initial evaluation of possible allergic reaction a. Cease administration of allergenic extracts b. Notify physician c. Record vital signs: blood pressure, pulse,
Continuous nebulization therapy for asthma with aerosols of 2 agonists
Continuous nebulization therapy for asthma with aerosols of 2 agonists Otto G Raabe, PhD; Timothy M Wong, BS; Garrett B Wong; James W Roxburgh, MBA, RRT; S David Piper; and James I C Lee, BS Background:
Pulmonary Function Testing: Coding and Billing Issues
Pulmonary Function Testing: Coding and Billing Issues Neil R MacIntyre MD FAARC and Catherine M Foss CPFT Introduction Accreditation Developing Coding/Billing Strategies Spirometry Lung Volume Testing
ST. ROSE HOSPITAL Job Description
ST. ROSE HOSPITAL Job Description JOB TITLE/CLASSIFICATION... REGISTERED RESPIRATORY THERAPIST (RRT) (163) DEPARTMENT... PULMONARY SERVICES STATUS... NON-EXEMPT RELATIONSHIPS Reports to... PULMONARY SUPERVISOR
Therapist Multiple-Choice Examination
Therapist Multiple-Choice Examination Effective: January 2015 Detailed Content Outline Items are linked to open cells. Each scored form will include 20-item pretests. I. PATIENT DATA EVALUATION AND RECOMMENDATIONS
PROP Acute Care/Rehab Discharge Planning Requirements 1. PROP Medical Criteria 2. PROP Prescription for Services 3
BC Association for Individualized Technology and Supports for People with Disabilities (BCITS) Proviinciiall Respiiratory Outreach Program ((PROP)) Discharge Planning Guide For ventilator dependent and
3100B Clinical Training Program. 3100B HFOV VIASYS Healthcare
3100B Clinical Training Program 3100B HFOV VIASYS Healthcare HFOV at Alveolar Level Nieman,, G, SUNY 1999 Who DO We Treat? Only Pathology studied to date has been ARDS Questions about management of adults
MEDUMAT Easy CPR The Guiding Emergency Ventilator
MEDUMAT Easy CPR The Guiding Emergency Ventilator MEDUMAT Easy CPR The Guiding Emergency Ventilator Safety is on Duty Although patient resuscitation is not part of your daily work, you are regularly confronted
Oxygen & Aerosol Therapy
Oxygen & Table of Contents Unomedical Oxygen Delivery System...3-7 Unomedical Oxygen Delivery System... 3 Low... 4 Nasal Oxygen Cannulas... 4 Oxygen Catheters... 4 Medium... 5 Medium Oxygen Masks... 5
81 First Responder Respiratory
81 First Responder Medical Scenarios Asthma Scenario: You are called to a local house for a woman with trouble breathing. You arrive to find a 67-year-old woman sitting upright in a chair. She states she
Australian/New Zealand Standard
AS/NZS 4236:1994 Australian/New Zealand Standard Respiratory therapy equipment Jet nebulizers and jet nebulizer air pumps AS/NZS 4236:1994 This Joint Australian/New Zealand Standard was prepared by Joint
RESPIRATORY THERAPIST CLASSIFICATION SERIES
STATE OF WISCONSIN CLASSIFICATION SPECIFICATION RESPIRATORY THERAPIST CLASSIFICATION SERIES Effective Date: June 16, 1985 Modified Effective: November 5, 2000 Modified Effective: May 6, 2001 I. INTRODUCTION
Trilogy accessory guide
Trilogy accessory guide 2 Trilogy accessory guide Light, versatile, and easy to use, Trilogy portable life-support ventilators are designed for use in the home and alternative care sites and provide noninvasive
Detach Patient's Instructions for Use from package insert and dispense with the product.
Attention Health Care Professional: Detach Patient's Instructions for Use from package insert and dispense with the product. PROVENTIL HFA (albuterol sulfate) Inhalation Aerosol FOR ORAL INHALATION ONLY
Evaluation copy. Figure 1
O 2 Extraction by the Lungs Computer 23 Oxygen is required for cell metabolism. During inhalation air is brought into the lungs, where oxygen is extracted. Oxygen passes into the bloodstream at the membrane
Gas Exchange. Graphics are used with permission of: Pearson Education Inc., publishing as Benjamin Cummings (http://www.aw-bc.com)
Gas Exchange Graphics are used with permission of: Pearson Education Inc., publishing as Benjamin Cummings (http://www.aw-bc.com) Page 1. Introduction Oxygen and carbon dioxide diffuse between the alveoli
ECMO AND LIFE SUPPORT SYSTEMS QUADROX PLS AND ROTAFLOW HARDWARE AND ACCESSORIES
ECMO AND LIFE SUPPORT SYSTEMS QUADROX PLS AND ROTAFLOW HARDWARE AND ACCESSORIES The Gold Standard Cardiovascular QUADROX PLS and ROTAFLOW 3 QUADROX PLS AND ROTAFLOW INNOVATIVE PERFORMANCE AND RELIABILITY
Flow Measurement with Respironics Flow Sensors
W H I T E P A P E R Flow Measurement with Respironics Flow Sensors ABSTRACT Flow measurement issues with mechanically ventilated patients are reviewed with particular emphasis on the Respironics flow sensors.
ACLS Provider Manual Comparison Sheet Based on 2010 AHA Guidelines for CPR and ECC. BLS Changes
ACLS Provider Manual Comparison Sheet Based on 2010 AHA Guidelines for CPR and ECC CPR Chest compressions, Airway, Breathing (C-A-B) BLS Changes New Old Rationale New science indicates the following order:
STATE OF NEBRASKA STATUTES RELATING TO RESPIRATORY CARE PRACTICE ACT
2012 STATE OF NEBRASKA STATUTES RELATING TO RESPIRATORY CARE PRACTICE ACT Department of Health and Human Services Division of Public Health Licensure Unit 301 Centennial Mall South, Third Floor PO Box
Enables MDA Medical Teams to categorize victims in mass casualty scenarios, in order to be able to triage and treat casualties
MDA Disposable ALS + BLS Medical Ambulance Equipment Prices shown in CDN. Funds Items Description Picture Mass Casualty ID tag 1000 units = $350 Enables MDA Medical Teams to categorize victims in mass
AMERICAN HEART ASSOCIATION 2010 ACLS GUIDELINES: WHAT EVERY CLINICIAN NEEDS TO KNOW
AMERICAN HEART ASSOCIATION 2010 ACLS GUIDELINES: WHAT EVERY CLINICIAN NEEDS TO KNOW Joseph Heidenreich, MD Texas A&M Health Science Center Scott & White Memorial Hospital Scott & White Memorial Hospital
EXPERIMENT 13: THE IDEAL GAS LAW AND THE MOLECULAR WEIGHT OF GASES
Name Section EXPERIMENT 13: THE IDEAL GAS LAW AND THE MOLECULAR WEIGHT OF GASES PRE-LABORATORY QUESTIONS The following preparatory questions should be answered before coming to lab. They are intended to
Gas Exchange Graphics are used with permission of: adam.com (http://www.adam.com/) Benjamin Cummings Publishing Co (http://www.awl.
Gas Exchange Graphics are used with permission of: adam.com (http://www.adam.com/) Benjamin Cummings Publishing Co (http://www.awl.com/bc) Page 1. Introduction Oxygen and carbon dioxide diffuse between
Michigan Adult Cardiac Protocols CARDIAC ARREST GENERAL. Date: May 31, 2012 Page 1 of 5
Date: May 31, 2012 Page 1 of 5 Cardiac Arrest General This protocol should be followed for all adult cardiac arrests. Medical cardiac arrest patients undergoing attempted resuscitation should not be transported
Talcum Powder and Cancer
Talcum Powder and Cancer What is talcum powder? Talcum powder is made from talc, a mineral made up mainly of the elements magnesium, silicon, and oxygen. As a powder, it absorbs moisture well and helps
