Suction Catheter Size: An Assessment and Comparison of 3 Different Calculation Methods
|
|
|
- Vernon Garrett
- 9 years ago
- Views:
Transcription
1 Suction Catheter Size: An Assessment and Comparison of 3 Different Calculation Methods Christopher J Russian MEd RRT-NPS RPSGT RST, Joshua F Gonzales MHA RRT-NPS RRT-SDS, and Nicholas R Henry MSC RRT-NPS AE-C BACKGROUND: Current American Association for Respiratory Care (AARC) clinical practice guidelines recoend a suction catheter to endotracheal tube ratio (SC/ETT) based on the external diameter of the SC and the internal diameter of the ETT. An SC/ETT ratio of < 50% is consistent with the current recoendation. We theorized that a more satisfactory assessment of SC/ETT ratio could be accomplished using volume or area formulas and expansion of diameter recoendations. Some respiratory care texts recoend an SC/ETT ratio that exceeds the clinical practice guideline standard. METHODS: We calculated the internal volume and cross-sectional area of various ETT sizes, the external volume and cross-sectional area of various SC sizes, and the SC/ETT ratios. We also measured negative pressures created by suction in a lung model, during multiple suction maneuvers. RESULTS: Volume and area calculations provide an alternative method for determining the SC/ETT ratio. A volume or area ratio of 50% corresponds to a diameter ratio of 70%. Negative pressures during suctioning remain low at the new ratios, so a larger SC than current clinical practice guidelines still allows adequate air passage between the SC and ETT. CONCLUSIONS: Our results support an alternative SC/ETT ratio when pairing SCs and ETTs. Key words: suction catheter; endotracheal tube; endotracheal suctioning; airway clearance. [Respir Care 2014;59(1): Daedalus Enterprises] Introduction The authors are affiliated with the Department of Respiratory Care, Texas State University, San Marcos, Texas. Mr Russian presented a version of this paper at the OPEN FORUM of the 2011 AARC Congress, held November 5 8, 2011, in Tampa, Florida. This study was partly supported by a grant from Texas State University. The suction catheters were donated by Kimberly-Clark. Kimberly-Clark had no other involvement in the study. The authors have disclosed no conflicts of interest. Correspondence: Christopher J Russian MEd RRT-NPS RPSGT RST, Department of Respiratory Care, Texas State University, 601 University Drive, San Marcos TX [email protected]. DOI: /respcare In 1960 Rosen and Hillard published an extensive paper on the use of suctioning during clinical practice, and then a follow-up study in 1962 that focused specifically on negative pressure during tracheal suctioning. 1,2 They introduced a formula to determine the negative pressure generated within the lungs during suctioning. Based on their calculations they recoended that the suction catheter (SC) should be no more than half the diameter of the endotracheal tube (ETT). Those were the seminal publications on SC/ETT ratios. Regardless of widespread acceptance of an SC/ETT ratio based on tube diameter, there is evidence that clinicians and researchers use larger ratios than recoended. 3-6 The American Association for Respiratory Care (AARC) convened an expert panel to develop and periodically update clinical practice guidelines (CPG) on suctioning the artificial airway of the mechanically ventilated patient. 7,8 Section 2.0 of the AARC endotracheal suctioning CPG suggests selecting an SC based on the internal diameter (ID) of the artificial airway. 8 The expert panel cited 2 studies 9,10 to support this recoendation. However, Tiffin et al suggested using an SC/ETT ratio based on tube crosssectional area, rather than diameter. 9 Pedersen et al (a study not cited in the AARC CPG) suggested that the suction catheter should occlude less than half the internal lumen, rather than half the diameter of the ET-tube. 11 They considered lumen synonymous with volume, and 32 RESPIRATORY CARE JANUARY 2014 VOL 59 NO 1
2 provided a volume formula in their paper. 11 Collectively; there exist 3 different options for determining the relationship between SC and ETT size: diameter, volume, and cross-sectional area. We calculated the internal volumes and cross-sectional areas of multiple ETTs, calculated the external volumes and cross-sectional areas of multiple SCs, recorded the negative pressures generated during suctioning of an airway model, and compared the current AARC CPG recoendations with our calculated volumes, areas, and ratios. Methods This study was granted exemption status by the institutional review board at Texas State University-San Marcos. The study was performed in the instrumentation lab in the Department of Respiratory Care. We calculated the internal volumes and internal cross-sectional areas of various ETT sizes, the external volumes and external areas of various SC sizes, and the SC/ETT ratios. Standard geometry formulae for the volume and area calculations were used. We studied ETT sizes from 2.0 to 10.0 internal diameter (ID), and SC sizes from 4 to 16 French. The following millimeter and French conversions were used when necessary 12 : 1 3 French French QUICK LOOK Current knowledge The current American Association for Respiratory Care clinical practice guidelines recoend choosing suction catheter size based on the external diameter of the suction catheter and the internal diameter of the endotracheal tube: a ratio of 50% is recoended, to prevent suctioning-related complications, including hypoxemia. What this paper contributes to our knowledge A calculated suction-catheter-to-endotracheal-tube ratio of 50% of the endotracheal tube area or 50% of the endotracheal tube volume is equivalent to 70% of the endotracheal tube diameter. A ratio of 70% allows the clinician to determine the proper suction catheter size by doubling the endotracheal tube size. However, the appropriate suction catheter size should be judged by patient response to the procedure. diameter and ETT length were used to calculate ETT volume. We relied on the manufacturer labels for the ETT IDs. The 15 ETT connector was not included in the measurements. ETT length was measured from the distal end of the ETT 15 connector, while inserted in the ETT, to the proximal end of the Murphy eye. External diameter and length were used to calculate SC external volume. Length is crucial to volume measurements, so SC and ETT length were matched for the calculations. Original ETT length per the manufacturer was not altered, thus the ETT length determined SC length and the resulting volumes. The SC external diameter was measured with a digital caliper. Volume calculations were converted from cubic millimeters to milliliters. We calculated the ratio of the external volume of the SC to the internal volume of the ETT (SC/ETT volume ratio) for all measured sizes. ETT ID and SC external diameter were used to calculate crosssectional area. We also calculated the ratio of the external area of the SC to the internal area of the ETT (SC/ETT area ratio) for all measured sizes. We measured negative pressures during suctioning of an airway model consisting of corrugated and oxygen tubing. Cuffed ETTs ranging from 6 to 10 ID (SunMed, Largo, Florida) were inserted into corrugated tubing (diameter 21.0, length 30 cm). Cuffed ETTs of 4.0 and 5.0 ID (SunMed, Largo, Florida) were inserted into pediatric ventilator corrugated tubing (diameter 16.0, length 30 cm). A cuffed ETT of 3.0 ID (Wai Medical, Viroqua, Wisconsin) was inserted into neonatal ventilator corrugated tubing (diameter 10.6, length 23 cm). An uncuffed ETT of 2.0 ID (Sheridan/ Teleflex, Research Triangle Park, North Carolina) was inserted into oxygen tubing (diameter 4.0, length 19 cm). Half-size ETTs (eg, 5.5 ID) were not used for this portion of the study. A negative pressure manometer (Instrumentation Industries, Bethel Park, Pennsylvania) was connected to the distal end of each lung model and opposite the ETT point of insertion. The use of cuffed ETTs created a seal between the ETT and the airway model. This seal was crucial to our experiments, because it prevented air passage between the ETT and the airway model, thus reducing pressure attenuation from that pathway. Since the 2.0 ID ETT does not have a cuff, a 19-cm length of oxygen tubing was used as the airway model. Insertion of the ETT to the hub created a seal between the ETT and the oxygen tubing. Similar to the corrugated tubing models, a pressure manometer was attached to the distal end of the oxygen tubing, to measure lung pressure during suctioning. Pressures were recorded with the 2.0 ID ETT fully inserted into the oxygen tubing, to simulate a sealed ETT scenario. In-line SCs (Kimberly-Clark, Roswell, Georgia) were connected to each ETT, with the suction line connected to a vacuum source. The vacuum source was adjusted to RESPIRATORY CARE JANUARY 2014 VOL 59 NO 1 33
3 Table 1. ETT ID Suction Pressures Applied adhere to AARC CPG specifications, based on patient population (Table 1). 8,13 Each SC was inserted into the ETT until it reached the tip of the ETT. Suction was applied as the catheter was slowly withdrawn. The maximum negative pressure generated was recorded. The suctioning procedure was repeated to achieve 2 identical negative pressures. If the negative pressures differed by more than 2 cm H 2 O, the procedure was repeated. The ventilator connection of the in-line SC remained open during all suctioning maneuvers, so air passage was possible between the SC and ETT, depending on the SC/ETT ratio. Results Table 2 displays the external volume calculations for the SCs. A 14 French SC was compared to several ETTs of different lengths, so SC volume for each length is reported. Table 3 displays the external area calculations for the SCs. Table 4 displays the internal volumes and internal areas calculations for the ETTs. Table 5 displays an SC size comparison based on the AARC endotracheal suctioning CPG and our area and volume calculations. Due to the current AARC CPG language Table 5 has a column for 50% and a column for 50%. The area and volume recoendations are based on 50% of the ETT area and volume calculations, respectively. Our 50% area and volume calculations match the 70% diameter calculations. The SC sizes in Table 5 are limited to 14 French. A larger SC is possible with certain ETTs (eg, 10 ID). Table 6 displays the suctioning pressure results from our airway model. A curvilinear change in negative pressure occurs based on the SC/ETT ratio. Discussion Negative Suction Pressure ( Hg) ETT endotracheal tube ID internal diameter Our calculations allow comparison of SC/ETT area and volume ratios with diameter ratios per the current AARC CPG. Selection of an SC that is 50% of the internal area or Table 2. French Suction Catheter External s, Lengths, and External Volumes External Length Volume (ml) Table 3. French Suction Catheter Sizes and Calculated External Areas External Area ( 2 ) % of the internal volume of an ETT equates to an SC that is 70% of the diameter of an ETT. This relationship applies to adult, pediatric, and neonatal ETTs. Per the AARC CPG, only infants can be suctioned with an SC that is 70% of the ETT diameter. The calculations for area and volume require somewhat complicated formulae, versus using diameter calculations. Equally, ETT diameter calculations require a conversion from millimeter to 34 RESPIRATORY CARE JANUARY 2014 VOL 59 NO 1
4 Table 4. Endotracheal Tube s, Lengths, Volumes, Calculated Maximum Volume Displacements, Cross-Sectional Areas, and Maximum Areas of Encroachment French Length Volume (ml) Maximum Volume Displacement (50% of calculated volume) (ml) Cross-Sectional Area ( 2 ) Maximum Area of Encroachment (50% of calculated area) ( 2 ) Table 5. Suction Catheter Size Recoendations Based on American Association for Respiratory Care Clinical Practice Guideline 2 and Area and Volume Calculations ETT Suction Catheter Size, French Based on American Association for Respiratory Care Clinical Practice Guideline 50% of the ETT 50% of the ETT 66% of the ETT 70% of the ETT Based on Area Calculation 50% of ETT Area Calculation Based on Volume Calculation 50% of ETT Volume Calculation * 14* 14* 14* * 14* 14* 14* * 14* 14* 14* * 14* 14* 14* * 14* 14* * Suction catheter size limited to 14 French. A larger size suction catheter is possible based on calculations. ETT endotracheal tube French, to match SC sizing. However, there is a quick method available to determine the SC/ETT ratio. Choosing an SC that is 70% of the ETT ID only requires that the clinician double the millimeter diameter size of the ETT to RESPIRATORY CARE JANUARY 2014 VOL 59 NO 1 35
5 Table 6. ETT Achieved Suction Pressure Versus ETT, Suction Catheter Size, and Set Suction Pressure Negative Pressure Achieved, cm H 2 O Suction Catheter Size, French Set Negative Suction Pressure ( Hg) 10.0* DNF DNF DNF DNF DNF DNF DNF * With the endotracheal tubes (ETTs) of 6 to 10 internal diameter we used large-bore corrugated tubing for the airway model. Pediatric ventilator circuit tubing was used for the airway model. Neonatal ventilator circuit tubing was used for the airway model. Oxygen tubing was used for the airway model. DNF suction catheter does not fit into ETT get the size of the SC in French. For example, a 7.0 ETT and a 7.5 ETT can accept a 14 French SC. This method to determine SC size applies for adult, pediatric, and neonatal ETTs. Furthermore, using an SC/ETT ratio exceeding the current AARC CPG recoendation is supported by 2 nationally recognized respiratory care textbooks. 14,15 Table 5 demonstrates the comparison between the 3 methods. It has been demonstrated that SCs that occlude half the area of the ETT do allow air to pass into the lungs during suctioning. 9 Figure 1 illustrates a comparison of an SC/ ETT ratio of 50% of the diameter versus 70% of the diameter. An SC that blocks 70% of the ETT ID still allows half of the ETT internal area for air passage, which is crucial for minimizing the negative pressure generated within the lungs and allowing ambient air to enter the SC. Our airway model experiment allowed the generation of negative pressure during suctioning, and, as expected, the pressure applied to the model lung became more negative as the SC/ETT ratio increased (Fig. 2). High suction pressure did not appear to occur until beyond the 70% SC/ETT ratio. Previous authors have calculated the negative pressure within the lungs when factoring in the applied suction pressure, the SC/ETT ratio, and the SC-ID/SC-externaldiameter ratio. 1,2 As Figure 3 demonstrates, an SC/ETT diameter ratio of 50% appears to generate a negative pressure on the flat or less steep portion of all 3 curves. Depending on the SC characteristics, the steep portion of the Fig. 1. Ratio of suction catheter outer diameter to endotracheal tube inner diameter. A: 50%. B: 70%. 36 RESPIRATORY CARE JANUARY 2014 VOL 59 NO 1
6 Fig. 2. Negative pressures generated in our airway model. The triangles indicate the recoended ratio of suction catheter outer diameter to endotracheal tube inner diameter, based on an area/ volume ratio of 50% and a diameter ratio of 70%. Since we did not study endotracheal tube (ETT) half-sizes (eg, 6.5 ), we make no recoendation regarding the 7 French and 5 French suction catheters. Fig. 4. Negative pressures calculated with the Rosen and Hillard formula. 1,2 The triangles indicate the recoendations based on an area/volume ratio of 50% and a diameter ratio of 70%. The American Association for Respiratory Care clinical practice guidelines suction pressures were converted to cm H 2 O and used in the calculations. Fig. 3. Calculated negative pressure developed in the lung plotted against the outside diameter of the suction catheter to the inside diameter of the airway. Each curve represents a catheter with a different wall thickness, expressed as the ratio of the catheter s inner diameter to outer diameter. (Data from reference 1.) curve does not occur until beyond an SC/ETT diameter ratio of 70% and suction pressure of 200 cm H 2 O. In addition, the suction pressure included in the original calculations (ie, 500 Hg) is beyond the negative pressure recoended in the AARC CPG (ie, in adults 150 Hg). 8,13 The formula developed by those authors was used to recalculate and plot the relationship between suction pressure, SC size, and ETT size. 1,2 Suction pressures were based on the current AARC CPG. Figure 4 demonstrates the curvilinear relationship between the SC and ETT sizes. The black triangles indicate the recoended SC and ETT combinations based on a volume/area ratio of 50% and a diameter ratio of 70%. As the SC/ETT ratio exceeds 70%, the calculated negative pressures approach the set suction pressure. The same curvilinear relationship was demonstrated with the airway model experiment. Limitations We did not assess the impact of a larger SC/ETT ratio on lung volumes. However, there is literature to address RESPIRATORY CARE JANUARY 2014 VOL 59 NO 1 37
7 this topic Additional research to determine the changes in lung volume during suctioning with various SC sizes is warranted. SC size was limited to 14 French, despite results indicating that a larger SC is possible. The reasons were that we found no literature supporting the use of a 16 French SC with an adult ETT, and there is literature suggesting that a 14 French SC is acceptable for adult patients. 14 Pulmonary secretions on the ETT wall decrease the ETT internal volume and cross-sectional area, and limit air passage during suctioning, which may increase the negative pressure within the lungs. We did not study the impact of secretions on ETT internal size and negative pressure. We also did not study distal negative pressure with a completely closed suctioning set-up. The proximal end of the in-line SC remained open to the atmosphere. The proximal end of the in-line SC is a source for air passage during suctioning. Adding positive-pressure ventilation to the proximal end of the in-line SC would impact the negative pressure generated during suctioning. Cuffed ETTs are not coonly used in neonatal and some pediatric patients. Without a cuff there is probably a leak between the ETT and the trachea, which would increase air passage during suctioning. We used cuffed ETTs to determine the negative pressure due to the SC/ETT ratio, so our suctioning results do not reflect clinical results that would occur with uncuffed ETTs. Conclusions A calculated SC/ETT ratio based on 50% of ETT area or 50% of ETT volume is equivalent to 70% of the ETT diameter. Using a diameter ratio of 70% would allow the clinician to quickly determine the proper SC size, by doubling the ETT size. This method for determining the SC/ETT ratio allows for a larger SC than is recoended by the current AARC CPG. 8 We stress, however, that our results establish the maximum SC size for each ETT. Bedside practitioners can select an SC that is smaller than the ratios provided above if the recoended size is unavailable or deemed inappropriate due to clinical outcomes. REFERENCES 1. Rosen M, Hillard EK. The use of suction in clinical medicine. Brit J Anaesth 1960;32(10): Rosen M, Hillard EK. The effects of negative pressure during tracheal suction. Anesth Analg 1962;41(1): Witmer MT, Hess DR, Sions M. An evaluation of the effectiveness of secretion removal with the ballard-circuit suction catheter. Respir Care 1991;36(8): Fernández MD, Piacentini E, Blanch L, Fernández R. Changes in lung volume with three systems of endotracheal suctioning with and without pre-oxygenation in patients with mild-to-moderate lung failure. Intensive Care Med 2004;30(12): DOI: /s Frengley RW, Closey DN, Sleigh JW, Torrance JM. The effect of closed system suction on airway pressures when using the Servo 300 ventilator. Crit Care Resuscitation 2001;3(4): Day T, Wainwright S, Wilson-Barnett J. An evaluation of a teaching intervention to improve the practice of endotracheal suctioning in intensive care units. J Clin Nurs 2001;10(5): American Association for Respiratory Care. AARC clinical practice guideline. Endotracheal suctioning of mechanically ventilated adults and children with artificial airways. Respir Care 1993;38(5): American Association for Respiratory Care. AARC clinical practice guideline. Endotracheal suctioning of mechanically ventilated patients with artificial airways: Respir Care 2010;55(6): Tiffin NH, Keim MR, Frewen TC. The effects of variations in flow through an insufflating catheter and endotracheal tube and suction catheter size on test lung pressures. Respir Care 1990;35(9): Vanner NH, Bick E. Tracheal pressures during open suctioning. Anaesthesia 2008;63(3): DOI: /j Pederson CM, Rosendahl-Nielsen M, Hjermind J, Egerod I. Endotracheal suctioning of the adult intubated patient-what is the evidence? Intensive Crit Care Nurs 2009;25(1): DOI: /j. iccn Iserson KV. J.F.B. Charriére: the man behind the French gauge. J Emerg Med 1987;5(6): American Association for Respiratory Care. AARC clinical practice guideline. Nasotracheal suctioning: 2004 revision and update. Respir Care 2004;49(9): Davies JD, May RA, Bortner PL. Airway management. In: Hess, DR, MacIntyre NR, Mishoe SC, Galvin WF, Adams AB, editors. Respiratory care: principles and practice, 2nd edition. Sudbury, MA: Jones & Bartlett Learning; 2012: Altobelli N. Airway management. In: Kacmarek RM, Stoller JK, Heuer AJ, editors. Egan s fundamentals of respiratory care, 10th editon. St Louis: Elsevier; 2013: Lindgren S, Almgren B, Hogman M, Lethvall S, Houltz E, Lundin S, Stenqvist O. Effectiveness and side effects of closed and open suctioning: an experimental evaluation. Intensive Care Med 2004;30(8): Morrow BM, Futter MJ, Argent AC. Endotracheal suctioning: from principles to practice. Intensive Care Med 2004;30(6): DOI: /s Tingay DG, Copnell B, Grant CA, Dargaville PA, Dunster KR, Schibler A. The effect of endotracheal suction on regional tidal ventilation and end-expiratory lung volume. Inten Care Med 2010; 36(5): DOI: /s x. This article is approved for Continuing Respiratory Care Education credit. For information and to obtain your CRCE (free to AARC members) visit 38 RESPIRATORY CARE JANUARY 2014 VOL 59 NO 1
TRACHEOSTOMY TUBE PARTS
Page1 NR 33 TRACHEOSTOMY CARE AND SUCTIONING Review ATI Basic skills videos: Tracheostomy care and Endotracheal suction using a closed suction set. TRACHEOSTOMY TUBE PARTS Match the numbers on the diagram
One Lung Ventilation Module (OLV)
1 One Lung Ventilation Module (OLV) A Thoracic Surgery Directors Association (TSDA) Cardiothoracic Surgery Resident Boot Camp Syllabus The ability to isolate one of the lungs is an essential skill set
endotracheal tube guide We provide Rusch quality.
endotracheal tube guide We provide Rusch quality. 2 RUSCH ENDOTRACHEAL TUBE GUIDE RUSCH ENDOTRACHEAL TUBE GUIDE 3 subglottic secretion removal a vap reduction strategy Ventilator-Associated Pneumonia (VAP)
Neonatal Intubation. Purpose. Scope. Indications. Equipment Cardiorespiratory monitor SaO 2 monitor. Anatomic Considerations.
Page 1 of 5 Purpose Scope Indications Neonatal Intubation To assure proper placement of endotracheal tubes for maximum ventilation using proper intubation procedures. The policy applies to all Respiratory
Nurses and Respiratory Therapists Working Together for Safe Alarm Systems Management
Nurses and Respiratory Therapists Working Together for Safe Alarm Systems Management May 11, 2015 9/25/2013 1 AAMI Foundation Vision: To drive the safe adoption and use of healthcare technology Visit our
NEONATAL AND PEDIATRIC CUFFED ENDOTRACHEAL TUBES: SAFETY AND PROPER USE
ENDOTRACHEAL TUBES: SAFETY AND PROPER USE by Esther Weathers RRT, RCP RC Educational Consulting Services, Inc. 16781 Van Buren Blvd, Suite B, Riverside, CA 92504-5798 (800) 441-LUNG / (877) 367-NURS www.rcecs.com
Ki m b e r l y-cl a r k* 72-Hour Closed-Suction Systems. Ba l l a r d* Tr a c h Ca r e* System. A unique design. A new standard in clean.
Ki m b e r l y-cl a r k* 72-Hour Closed-Suction Systems Ba l l a r d* Tr a c h Ca r e* System A unique design. A new standard in clean. Ki m b e r l y-cl a r k* 72-Hour Closed-Suction Systems Ba l l a
MECHINICAL VENTILATION S. Kache, MD
MECHINICAL VENTILATION S. Kache, MD Spontaneous respiration vs. Mechanical ventilation Natural spontaneous ventilation occurs when the respiratory muscles, diaphragm and intercostal muscles pull on the
Nocturnal Oxygenation Using a Pulsed-Dose Oxygen-Conserving Device Compared to Continuous Flow
Nocturnal Oxygenation Using a Pulsed-Dose Oxygen-Conserving Device Compared to Continuous Robert L Chatburn RRT-NPS FAARC, Joseph S Lewarski RRT FAARC, and Robert W McCoy RRT FAARC BACKGROUND: The pulsed-dose
Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context
Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context A subcommittee of the Canadian Neonatal Resuscitation Program (NRP) Steering Committee
High Impact Intervention Care bundle to reduce ventilation-association pneumonia
High Impact Intervention Care bundle to reduce ventilation-association pneumonia Aim To reduce the incidence of ventilation-associated pneumonia (VAP). Context The aim of the care bundle, as set out in
Albuterol Delivery by 4 Different Nebulizers Placed in 4 Different Positions in a Pediatric Ventilator In Vitro Model
Albuterol Delivery by 4 Different Nebulizers Placed in 4 Different Positions in a Pediatric Ventilator In Vitro Model Ariel Berlinski MD and J Randy Willis RRT-NPS BACKGROUND: The type of aerosol generator
Airways Resistance and Airflow through the Tracheobronchial Tree
Airways Resistance and Airflow through the Tracheobronchial Tree Lecturer: Sally Osborne, Ph.D. Department of Cellular & Physiological Sciences Email: [email protected] Useful links: www.sallyosborne.com
The Anesthesia Ventilator
The Anesthesia Ventilator Why is the piston replacing the bellows? For many decades, and millions of anesthetics, the bellows anesthesia ventilator has been a safe and effective clinical device. Indeed,
LESSON 4 ORAL, NASOPHARYNGEAL, AND NASOTRACHEAL SUCTIONING.
LESSON 4 ORAL, NASOPHARYNGEAL, AND NASOTRACHEAL SUCTIONING. 4-1. SUCTIONING a. Suctioning is a common nursing activity performed for the purpose of removing accumulated secretions from the patient's nose,
Management of cuffed endotracheal tubes
A comparison of endotracheal tube cuff pressures using estimation techniques and direct intracuff measurement Scott L. Stewart, CRNA, MSN Janet A. Secrest, RN, PhD Barbara R. Norwood, RN, EdD Richard Zachary,
Fuel Cell Characterization Lab
SJSU E 10 Introduction to Engineering Fuel Cell Characterization Lab What is a Fuel Cell? Before we answer that question, let s first review the process of electrolysis. In the process of electrolysis,
Battery Performance of 4 Intensive Care Ventilator Models
Battery Performance of 4 Intensive Care Ventilator Models Thomas C Blakeman RRT, Bryce RH Robinson MD, and Richard D Branson MSc RRT FAARC BACKGROUND: Hospital electrical power failure represents an important
Paediatric Intensive Care unit Nursing Procedure: Care of the ventilated child
Paediatric Intensive Care unit Nursing Procedure: Care of the ventilated child All nursing staff should read this policy to inform themselves of relevant areas prior to undertaking any aspect of care related
Pulmonary Function Testing: Coding and Billing Issues
Pulmonary Function Testing: Coding and Billing Issues Neil R MacIntyre MD FAARC and Catherine M Foss CPFT Introduction Accreditation Developing Coding/Billing Strategies Spirometry Lung Volume Testing
COURSE SYLLABUS RC 223 CLINICAL-3
COURSE SYLLABUS RC 223 CLINICAL-3 Class Hours: 0 Laboratory Hours: 24 Credit Hours: 4 Course Description: Entry Level Standards: This course will emphasize neonatal-pediatric intensive care, pulmonary
Department of Veterans Affairs VHA DIRECTIVE 2012-032 Veterans Health Administration Washington, DC 20420 October 26, 2012
Department of Veterans Affairs VHA DIRECTIVE 2012-032 Veterans Health Administration Washington, DC 20420 OUT OF OPERATING ROOM AIRWAY MANAGEMENT 1. PURPOSE: This Veterans Health Administration (VHA) Directive
The Determination of Carbon Dioxide in Exhaled Air A Calculator-Based Laboratory Experiment
The Determination of Carbon Dioxide in Exhaled Air A Calculator-Based Laboratory Experiment James Gordon * and Nathaniel Barbe Division of Science and Mathematics, Central Methodist University, Fayette,
Buoyant Force. Goals and Introduction
Buoyant Force Goals and Introduction When an object is placed in a fluid, it either floats or sinks. While the downward gravitational force, F g, still acts on the object, an object in a fluid is also
Evaluation of a Nasal Cannula in Noninvasive Ventilation Using a Lung Simulator
Evaluation of a Nasal in Noninvasive Ventilation Using a Lung Simulator Narayan P Iyer MD and Robert Chatburn MHHS RRT-NPS FAARC BACKGROUND: Nasal noninvasive ventilation (NIV) is a common form of noninvasive
AARC Clinical Practice Guidelines
AARC Clinical Practice Guidelines Endotracheal Suctioning of Mechanically Ventilated Patients With Artificial Airways 2010 An electronic literature search for articles published between January 1990 and
High-Frequency Oscillatory Ventilation
High-Frequency Oscillatory Ventilation Arthur Jones EdD, RRT Learning Objectives Describe the indications and rationale and monitoring for HFOV. Identify HFOV settings and describe the effects of their
GUIDELINES FOR THE MANAGEMENT OF OXYGEN THERAPY
SOUTH DURHAM HEALTH CARE NHS TRUST GUIDELINES FOR THE MANAGEMENT OF OXYGEN THERAPY AIM To supplement oxygen intake using the appropriate equipment in order to correct hypoxia and relieve breathlessness.
CHAPTER THREE COMMON DESCRIPTIVE STATISTICS COMMON DESCRIPTIVE STATISTICS / 13
COMMON DESCRIPTIVE STATISTICS / 13 CHAPTER THREE COMMON DESCRIPTIVE STATISTICS The analysis of data begins with descriptive statistics such as the mean, median, mode, range, standard deviation, variance,
Forced vital capacity: maximal volume of air exhaled with maximally forced effort from a maximal inspiration.
SOP Spirometry 1. General considerations Spirometry serves as a physiological test to quantify pulmonary disease severity and to assess clinical change in respiratory function over time. Standard spirometric
Survey of Directors of Respiratory Therapy Departments Regarding the Future Education and Credentialing of Respiratory Care Students and Staff
Survey of Directors of Respiratory Therapy Departments Regarding the Future Education and Credentialing of Respiratory Care Students and Staff Robert M Kacmarek PhD RRT FAARC, Thomas A Barnes EdD RRT FAARC,
Critical Care Therapy and Respiratory Care Section
Critical Care Therapy and Respiratory Care Section Category: Clinical Section: Special Procedures Title: Transport of Critically Ill Patients Policy #: 05 Revised: 04/00 1.0 DESCRIPTION 1.1 Definition:
Guiding Protocolized Patient Care through Branching Logic. By Cindy Sparkman, BS, RRT-NPS and Mickey Roach, BS, RRT
Guiding Protocolized Patient Care through Branching Logic By Cindy Sparkman, BS, RRT-NPS and Mickey Roach, BS, RRT University of Utah 2 University of Utah FY2012 Hospitals and Clinics; 4 Hospitals 10 Community
POCKET GUIDE. NAVA and NIV NAVA in neonatal settings
POCKET GUIDE NAVA and NIV NAVA in neonatal settings Table of contents EMPTY 1 2 3 4 Introduction and background facts Invasive ventilation with NAVA Non invasive ventilation with NAVA NAVA and NIV NAVA
Pulmonary Ventilation
Pulmonary Ventilation Graphics are used with permission of: Pearson Education Inc., publishing as Benjamin Cummings (http://www.aw-bc.com) Page 1. Introduction Pulmonary ventilation, or breathing, is the
Secrets of the Pediatric Intubated Patient:
Secrets of the Pediatric Intubated Patient: Evidence-Based Keys to Success Leslie Grant, RN, BSN, CFRN, CCRN Matt Payne, RN, CFRN, FP-C Ranked #1 in Virginia by U.S. News & World Report VCU Total Artificial
NHS FORTH VALLEY Chest Drain Insertion by Seldinger Method (Pigtail drains)
NHS FORTH VALLEY Chest Drain Insertion by Seldinger Method (Pigtail drains) Date of First Issue 12/12/2012 Approved 12/12/2012 Current Issue Date 12/12/2012 Review Date 12/12/2014 Version 1 EQIA Yes 14/12/2012
TY-CARE TM TY-CARE TM exel TY-CARE TM exel Neo/Ped Closed Suction Systems
COMPLETE TY-CARE TY-CARE product range 444SP01010 10 10 ea/box 444SP01012 12 10 ea/box 444SP01014 14 10 ea/box 444SP01016 16 10 ea/box A COMPLETE RANGE FOR ADULTS TO NEONATES TY-CARE TM TY-CARE TM exel
Edwards FloTrac Sensor & Edwards Vigileo Monitor. Understanding Stroke Volume Variation and Its Clinical Application
Edwards FloTrac Sensor & Edwards Vigileo Monitor Understanding Stroke Volume Variation and Its Clinical Application 1 Topics System Configuration Pulsus Paradoxes Reversed Pulsus Paradoxus What is Stroke
Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003
Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003 Knowing the functions of the anesthetic delivery system
CHAPTER TWELVE TABLES, CHARTS, AND GRAPHS
TABLES, CHARTS, AND GRAPHS / 75 CHAPTER TWELVE TABLES, CHARTS, AND GRAPHS Tables, charts, and graphs are frequently used in statistics to visually communicate data. Such illustrations are also a frequent
Vaporization of Liquid Nitrogen
Vaporization of Liquid Nitrogen Goals and Introduction As a system exchanges thermal energy with its surroundings, the temperature of the system will usually increase or decrease, depending on the direction
CUESTA COLLEGE REGISTERED NURSING PROGRAM CRITICAL ELEMENTS
CUESTA COLLEGE REGISTERED NURSING PROGRAM CRITICAL ELEMENTS LEVELS I through IV A. OVERRIDING CRITICAL ELEMENTS Violation of an overriding area will result in termination and failure of the particular
UC Davis Medical Center Emergency Medicine Airway Equipment: Contents, Cleaning and Stocking
UC Davis Medical Center Emergency Medicine Airway Equipment: Contents, Cleaning and Stocking Introduction: Due to the large volume of critically ill patients who present for care at the U.C. Davis Emergency
5/30/2014 OBJECTIVES THE ROLE OF A RESPIRATORY THERAPIST IN THE DELIVERY ROOM. Disclosure
THE ROLE OF A RESPIRATORY THERAPIST IN THE DELIVERY ROOM Ona Fofah, MD FAAP Assistant Professor of Pediatrics Director, Division of Neonatology Department of Pediatrics Rutgers- NJMS, Newark OBJECTIVES
Veterinary Clinics and Hospitals
Anesthetic Gases: Guidelines for Workplace Exposure Veterinary Clinics and Hospitals Inhalation anesthesia in veterinary hospitals is practiced in a manner similar to that in human hospitals. Generally,
Shiley Product Guide. Shiley TM XLT Extended-Length Tracheostomy Tubes. Shiley TM Disposable Inner Cannula
Shiley Product Guide Shiley TM XLT Extended-Length Shiley TM Disposable Inner Cannula Shiley TM Percutaneous Disposable Inner Cannula Shiley TM Accessories for Disposable Inner Cannula Tubes Shiley TM
Ventilator Application of the Passy-Muir Valve David A. Muir Course Outline Benefits Review of the Biased Closed Position No Leak Passy-Muir Valves
Ventilator Application of the Passy-Muir Valve Michael S. Harrell, B.S., RRT Director of Clinical Education-Respiratory [email protected] (949) 833-8255 David A. Muir 23 year-old ventilator dependent
Introduction Hypothesis Methods Results Conclusions Figure 11-1: Format for scientific abstract preparation
ABSTRACT AND MANUSCRIPT PREPARATION / 69 CHAPTER ELEVEN ABSTRACT AND MANUSCRIPT PREPARATION Once data analysis is complete, the natural progression of medical research is to publish the conclusions of
GE Healthcare. Welcome to the future of anesthesia Aisys * Carestation *
GE Healthcare Welcome to the future of anesthesia Aisys * Carestation * The way you practice today may be different from the past. Diverse patient types and co-morbidities make your world challenging.
Neurally Adjusted Ventilatory Assist: NAVA for Neonates
Neurally Adjusted Ventilatory Assist: NAVA for Neonates Robert L. Chatburn, MHHS, RRT-NPS, FAARC Research Manager Respiratory Institute Cleveland Clinic Professor Department of Medicine Lerner College
INJECTION MOLDING COOLING TIME REDUCTION AND THERMAL STRESS ANALYSIS
INJECTION MOLDING COOLING TIME REDUCTION AND THERMAL STRESS ANALYSIS Tom Kimerling University of Massachusetts, Amherst MIE 605 Finite Element Analysis Spring 2002 ABSTRACT A FEA transient thermal structural
Certified Respiratory Therapist Exam Review Guide
55119_FMxx_5282.qxd 5/12/09 7:09 AM Page i Jones and Bartlett Publishers, LLC. NOT FOR SALE OR DISTRIBUTION Certified Respiratory Therapist Exam Review Guide Edited by CRAIG L. SCANLAN, EdD, RRT, FAARC
Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P2.6.3.2. Thermodynamic cycle Heat pump
Heat Thermodynamic cycle Heat pump LD Physics Leaflets P2.6.3.2 Investigating the function of the expansion valve of the heat pump Objects of the experiment g To study the operational components of the
Nurse Practitioners and Physician Assistants in the United States: Current Patterns of Distribution and Recent Trends. Preliminary Tables and Figures
Nurse Practitioners and Physician Assistants in the United States: Current Patterns of Distribution and Recent Trends Preliminary Tables and Figures Kevin M. Stange, PhD Assistant Professor Gerald R. Ford
Lab 1: Introduction to PSpice
Lab 1: Introduction to PSpice Objectives A primary purpose of this lab is for you to become familiar with the use of PSpice and to learn to use it to assist you in the analysis of circuits. The software
AEROSPACE MEDICAL SERVICE SPECIALTY INDEPENDENT DUTY MEDICAL TECHNICIAN EMERGENCY MEDICINE PROCEDURES
QTP4N0X1C-9 02 July 2015 AEROSPACE MEDICAL SERVICE SPECIALTY INDEPENDENT DUTY MEDICAL TECHNICIAN EMERGENCY MEDICINE PROCEDURES Volume 9 TOTAL FORCE, TOTAL CARE EVERYTIME, ANYWHERE 383d TRAINING SQUADRON/XUFB
How To Determine The Number Of Respiratory Therapists
AARC White Paper November 8, 2012 BEST PRACTICES IN RESPIRATORY CARE PRODUCTIVITY AND STAFFING This paper provides guidance and considerations in the application of the AARC Position Statement: Best Practices
X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1)
CORRELATION AND REGRESSION / 47 CHAPTER EIGHT CORRELATION AND REGRESSION Correlation and regression are statistical methods that are commonly used in the medical literature to compare two or more variables.
ANAESTHESIA AND GENERAL SURGERY PRODUCTS PHOENIX BRAND
ANAESTHESIA AND GENERAL SURGERY PRODUCTS PHOENIX BRAND PHOE Apr 11 ENDOBRONCHIAL TUBE, SINGLE USE Original Dr Frank Robertshaw design. D shaped lumina provides maximum cross sectional area, minimising
Pennsylvania System of School Assessment
Pennsylvania System of School Assessment The Assessment Anchors, as defined by the Eligible Content, are organized into cohesive blueprints, each structured with a common labeling system that can be read
Difficult Pre-hospital Airway Management
Difficult Pre-hospital Airway Management Frans L. Rutten, MD, FDSA Elisabeth Hospital Tilburg, The Netherlands In management of patients with an emergency, airway management and ventilatory care are the
The Anesthesia Ventilator
The Anesthesia Ventilator Why is the piston replacing the bellows? For many decades, and millions of anesthetics, the bellows anesthesia ventilator has been a safe and effective clinical device. Indeed,
NEONATAL RESUSCITATION PROVIDER (NRP) RECERTIFICATION TABLE OF CONTENTS
NEONATAL RESUSCITATION PROVIDER (NRP) RECERTIFICATION TABLE OF CONTENTS NEONATAL FLOW ALGORITHM.2 INTRODUCTION 3 ANTICIPATION OF RESUSCITATION 4 TEMPERATURE CONTROL.4 CLEARING THE AIRWAY OF MECONIUM 5
High Flow Through a Nasal Cannula and CPAP Effect in a Simulated Infant Model
High Flow Through a Nasal Cannula and CPAP Effect in a Simulated Infant Model Teresa A Volsko MHHS RRT FAARC, Kathy Fedor RRT-NPS, Jason Amadei RRT, and Robert L Chatburn MHHS RRT-NPS FAARC BACKGROUND:
Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.
The Physics of Flow Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK. Introduction Flow is defined as the quantity of fluid
Critical Care Therapy and Respiratory Care Section
Category: Section: Title: Critical Care Therapy and Respiratory Care Section Policy #: 03 Revised: 5/02 1.0 DESCRIPTION Clinical Ventilator Management Invasive and Noninvasive Positive Pressure Ventilation
Ventilation Perfusion Relationships
Ventilation Perfusion Relationships VENTILATION PERFUSION RATIO Ideally, each alveolus in the lungs would receive the same amount of ventilation and pulmonary capillary blood flow (perfusion). In reality,
Understanding patent claims (d) Double pipe
Understanding patent claims (d) Double pipe The invention The invention relates to a double pipe and a method of manufacturing it. The double pipe is preferably employed in a vehicle air conditioner for
E/M Experiment: Electrons in a Magnetic Field.
E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.
SUCTION & TUBING YANKAUER, POOLE, FRAZIER AND SIGMOID INSTRUMENTS PREMIUM CONNECTORED AND ECONOMY ROLL TUBING
SUCTION INSTRUMENTS & TUBING YANKAUER, POOLE, FRAZIER AND SIGMOID INSTRUMENTS PREMIUM CONNECTORED AND ECONOMY ROLL TUBING A complete line of quality, disposable Suction Instruments and Tubing to meet your
Density Determinations and Various Methods to Measure
Density Determinations and Various Methods to Measure Volume GOAL AND OVERVIEW This lab provides an introduction to the concept and applications of density measurements. The densities of brass and aluminum
Children's Medical Services (CMS) Regional Perinatal Intensive Care Center (RPICC) Neonatal Extracorporeal Life Support (ECLS) Centers Questionnaire
Children's Medical Services (CMS) Regional Perinatal Intensive Care Center (RPICC) Neonatal Extracorporeal Life Support (ECLS) Centers Questionnaire Date: RPICC Facility: CMS use only Include the following
Reflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
1 of 9 2/9/2010 3:38 PM
1 of 9 2/9/2010 3:38 PM Chapter 23 Homework Due: 8:00am on Monday, February 8, 2010 Note: To understand how points are awarded, read your instructor's Grading Policy. [Return to Standard Assignment View]
Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials.
Lab 3 Tension Test Objectives Concepts Background Experimental Procedure Report Requirements Discussion Objectives Experimentally determine the yield strength, tensile strength, and modules of elasticity
Albuterol Delivery During Noninvasive Ventilation
Albuterol Delivery During Noninvasive Ventilation Matthew P Branconnier RRT and Dean R Hess PhD RRT FAARC HYPOTHESIS: Albuterol delivered during noninvasive positive-pressure ventilation is affected by
Enzyme Action: Testing Catalase Activity
Enzyme Action: Testing Catalase Activity DataQuest 12 Many organisms can decompose hydrogen peroxide (H 2 O 2 ) enzymatically. Enzymes are globular proteins, responsible for most of the chemical activities
Guidelines for Standards of Care for Patients with Acute Respiratory Failure on Mechanical Ventilatory Support
Guidelines for Standards of Care for Patients with Acute Respiratory Failure on Mechanical Ventilatory Support Copyright by the SOCIETY OF CRITICAL CARE MEDICINE These guidelines can also be found in the
Obtaining Valid Consent to Participate in Perinatal Research Where Consent is Time Critical
Obtaining Valid Consent to Participate in Perinatal Research Where Consent is Time Critical February 2016 Obtaining Valid Consent to Participate in Perinatal Research Where Consent is Time Critical This
Artificial Ventilation Theory into practice
Artificial Ventilation Theory into practice Keith Simpson BVSc MRCVS MIET(Electronics) www.vetronic.co.uk [email protected] June 13 th 2014 Today we will discuss the administration of IPPV to anaesthetised
RESISTANCE AND COMPLIANCE
by Kevin T. Martin BVE, RRT, RCP RC Educational Consulting Services, Inc. 16781 Van Buren Blvd, Suite B, Riverside, CA 92504-5798 (800) 441-LUNG / (877) 367-NURS www.rcecs.com BEHAVIORAL OBJECTIVES UPON
High-frequency oscillatory
High-frequency oscillatory ventilation in adults: Respiratory therapy issues Jason Higgins, BS, RRT; Bob Estetter, RRT; Dean Holland, RRT; Brian Smith, RRT; Stephen Derdak, DO Objective: To summarize clinical
ronc itrad "~' Set Automatic entry into the left main-stem bronchus Ordering Information Bronchitrac "L" Set
Bronchitrac "L" Set ronc itrad "~' Set Automatic entry into the left main-stem bronchus The only suction catheter specifically designed for left-bronchus access in an easy-ta-use tray. Simply advance the
Respiratory Care Technology (RC)
Cincinnati State 1 Cincinnati State offers a comprehensive program in Respiratory Technology. Students develop a wide range of clinical skills in traditional and nontraditional roles and gain proficiency
AHA/AAP Neonatal Resuscitation Guidelines 2010: Summary of Major Changes and Comment on its Utility in Resource-Limited Settings
AHA/AAP Neonatal Resuscitation Guidelines 2010: Summary of Major Changes and Comment on its Utility in Resource-Limited Settings Resuscitation step Recommendations (2005) Recommendations (2010) Comments/LOE
Tracheostomy Care at Home
Tracheostomy Care at Home Patient Education CARE AND TREATMENT This information will help you understand how to care for a person with a tracheostomy at home. It is a supplement to the teaching offered
FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER
FLOW MEASUREMENT 200 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER Dr D G Stewart, NEL Dr M Reader-Harris, NEL Dr R J W Peters, McCrometer Inc INTRODUCTION The V-Cone
DRG 475 Respiratory System Diagnosis with Ventilator Support. ICD-9-CM Coding Guidelines
DRG 475 Respiratory System Diagnosis with Ventilator Support ICD-9-CM Coding G The below listed g are not inclusive. The coder should refer to the applicable Coding Clinic g for additional information.
Designing and Drawing a Sprocket Visualizing ideas through the creation of CAD solid models is a key engineering skill.
05 Webster St. Hanover Massachusetts 0339 Tel. 78 878 5 Fax 78 878 6708 Designing and Drawing a Sprocket Visualizing ideas through the creation of CAD solid models is a key engineering skill. The following
DATA EXPRESSION AND ANALYSIS
NAME Lab Day DATA EXPRESSION AND ANALYSIS LABORATORY 1 OBJECTIVES Understand the basis of science and the scientific method. Understand exponents and the metric system. Understand the metric units of length,
MECHANICS OF SOLIDS - BEAMS TUTORIAL 2 SHEAR FORCE AND BENDING MOMENTS IN BEAMS
MECHANICS OF SOLIDS - BEAMS TUTORIAL 2 SHEAR FORCE AND BENDING MOMENTS IN BEAMS This is the second tutorial on bending of beams. You should judge your progress by completing the self assessment exercises.
Measurement of Length, Mass, Volume and Density
Measurement of Length, Mass, Volume and Density Experimental Objective The objective of this experiment is to acquaint you with basic scientific conventions for measuring physical quantities. You will