A survey of stent designs
|
|
|
- Marion Preston
- 9 years ago
- Views:
Transcription
1 Min Invas Ther & Allied Technol 2002: 11(4) A survey of stent designs D. Stoeckel 1, C. Bonsignore 1 and S. Duda 2 1 Nitinol Devices and Components, Fremont, CA USA 2 Dept of Diagnostic Radiology, University of Tuebingen, Tuebingen, Germany Summary Over 100 different stent designs are currently being marketed or are in evaluation for vascular and non-vascular indications. This paper attempts to differentiate stent designs by engineering aspects. A stent design pyramid is presented, which breaks the differentiating aspects into materials, raw material forms, fabrication methods, geometrical features, and additions. The primary distinguishing factor in all groups is balloon-expansion versus self-expandability. Typical examples for each category of the pyramid are shown. Keywords stents, materials, fabrication methods, geometrical features Introduction The 2002 Handbook of Coronary Stents edited by Serruys and Rensing [1] lists 43 coronary stents or stent families, and Koronarstenting, published in 2001 by Machraoui, Grewe and Fischer [2], brings the number of stents tabulated to 59. Neither book claims to be complete; each focuses on cardiology, excluding stents specifically marketed for peripheral or non-vascular indications. It is therefore probably safe to assume that there are close to 100 different stents currently being marketed or in evaluation worldwide, with most of them available in Europe. As a result of strict FDA regulations, the number of approved stents in the USA is not quite as high, but is still substantial. These stents compete for a market that is estimated to be near $3 billion, and is expected to double with the advent of drug eluting devices. Most surveys differentiate stents by their clinical use, e.g. vascular or non-vascular, coronary or peripheral. This paper proposes classifications based on design and engineering characteristics of these structures, which are illustrated in the stent design pyramid in Figure 1. The chosen starting point is materials used, and distinguishes between balloonexpandable and self-expanding stents. From there, the classifications branch out into forms of materials used, such as sheet, wire or tube; manufacturing methods, such as laser-cutting, waterjet-cutting, photo-etching; and various wire-forming techniques. Next, the vast array of geometrical configurations that Figure 1. Stent design pyramid. Correspondence: D. Stoeckel, Nitinol Devices and Components, Westinghouse Drive, Fremont, CA 94539, USA Isis Medical Media Ltd 137 Martin Dunitz Taylor&Francis healthsciences
2 D. Stoeckel et al. have been explored in stent designs are considered. The classification ends with additions to stents, such as grafts, radiopaque markers and coatings. Throughout the text, each branch of a stent design map is presented that classifies nearly 100 commercialised stent designs. Designs included in this survey have been documented in texts [1,2], brochures, and company websites [3 25]. Like others, this review is probably not complete, and may describe stents that are not yet, or are no longer available. Materials Materials for metallic balloon-expandable or selfexpanding stents must exhibit excellent corrosion resistance and biocompatibility (Figure 2). They should be adequately radiopaque, and create minimal artifacts during MRI. Balloon-expandable stents are made from materials that can be plastically deformed through the inflation of a balloon. After the balloon is deflated the stent remains in its expanded shape, except for a slight recoil caused by the elastic portion of the deformation. The ideal material for these stents therefore has a low yield stress (to make it deformable at manageable balloon pressures), high elastic modulus (for minimal recoil), and is work hardened through expansion for high strength. Balloon-expandable stents are manufactured in the small diameter, i.e. deliverable configuration, and balloon-dilated to the expanded shape at the target site inside the vessel. Self-expanding stents, on the other hand, are manufactured in the expanded shape, then compressed and constrained in a delivery system. Upon release from the delivery system they spring back, i.e. self-expand, to the preset diameter. Their function, therefore, is based on the elastic properties of the material used. Ideally, the material should have a low elastic modulus and a high yield stress for large elastic strains. Alternatively, the shapememory effect of nitinol can be utilised. Here, large strains can be achieved either superelastically, or via the thermal memory of the material. The most widely used material for stents is stainless steel, typically 316L, a particularly corrosionresistant material with low carbon content and Figure 2. Overview of materials used in stent manufacture. 138
3 A survey of stent designs additions of molybdenum and niobium. In its fully annealed condition, stainless steel is easily deformable and, therefore, the standard material for balloon-expandable stents. In its full-hard condition, on the other hand, it exhibits enough elasticity for certain self-expanding stent designs. Alternative materials for balloon-expandable stents are tantalum [BSC Strecker (Figure 3), Cordis Crossflex, Medtronic Wiktor ], platinum alloys (AngioDynamics Angio Stent ), niobium alloys (Inflow Dynamics Lunar Starflex ) and cobalt alloys. They are used for their better radiopacity, higher strength, improved corrosion resistance, better MR compatibility or the combination of all these features. Better radiopacity and higher strength allow the design of stents with smaller delivery profiles. As mentioned above, materials for self-expanding stents should exhibit large elastic strains. The most widely used material is nitinol, a nickel titanium alloy that can recover elastic deformations of up to 10%. This unusually large elastic range, commonly known as superelasticity, is the result of a thermo-elastic martensitic transformation. The limited elastic range of more conventional materials, such as stainless steel (Cook Z Stent ) or certain cobalt-based alloys (BSC WallStent ), also limits design options. While the WallStent offers excellent wall coverage and flexibility, its shortcoming is its length change during deployment. The zig-zag configuration of the Z-Stent does not change length during deployment, but does not provide wall coverage in its bare configuration. Raw material form Stents can be made from sheet, wire (round or flat) or tubing (Figure 4). A big majority of the balloonexpandable and self-expanding stents are made from wire or tubing. A few exceptions are the BSC/Medinol NIR, the Navius ZR1, the EndoTex ratcheting stent and the Cook GRII (Figure 5), which are made from sheet metal. Stents made from sheet metal have to be rolled up to a tubular configuration after the pattern has been created. The NIR stent is then welded, while the ZR1 and the EndoTex use special mechanical locking features. Figure 3. Strecker stent made of knitted tantalum wire. Figure 5. Cook GRII, formed from stainless steel sheet, featuring an axial backbone with integral gold markers. Figure 4. Overview of stent forms 139
4 D. Stoeckel et al. Fabrication methods The choice of fabrication method depends mainly on the raw material form used (Figure 6). Wires can be formed into stents in various ways using conventional wire-forming techniques, such as coiling, braiding, or knitting. The simplest shape of a wire stent is a coil, e.g. the IntraTherapeutics IntraCoil. All coil stents marketed today are made from nitinol and are selfexpanding. Welding at specific locations after wireforming produces closed-cell wire stents [BSC Symphony (Figure 7), self-expanding nitinol stent] or increases longitudinal stability (Cordis Crossflex, balloon-expandable SS stent). The most common wire-based self-expanding stent is the WallStent (BSC), a braided design using multiple elgiloy (cobaltbased alloy) wires (Figure 8). This allows continuous production, i.e. the stents can be cut to length from a long wire-mesh hose. Knitting allows the production of flexible balloon-expandable and self-expanding wire stents. Examples are the BSC Strecker tantalum stent and the Cook ZA nitinol stent. The vast majority of coronary stents, and probably the majority of peripheral vascular stents, are produced by laser cutting from tubing Typically, Nd:YAG lasers are used, allowing kerf widths of, 20 µm. Intricate patterns can be produced using tube sizes from 0.5 mm diameter. Balloon-expandable stents are cut in the crimped or near-crimped condition, and only require post-cutting deburring and surface treatment typically electropolishing. They are marketed balloon-mounted, or unmounted for hand-crimping. Self-expanding nitinol stents, on the other hand, can be cut either in the small configuration, requiring post-cutting expansion and shape-setting, or in the expanded condition. In either case, they have to be deburred and polished. Selfexpanding stents have to be constrained in the delivery system and, therefore, are not available in an unmounted configuration. Figure 7. Symphony stent, nitinol wire welded to form a closed-cell structure. Figure 8. WallStent, braided stent fabricated from cobalt alloy wire. Figure 6. Overview of stent fabrication. 140
5 A survey of stent designs Laser cutting produces a heat-affected zone along the cut edge, which has to be removed for better performance. A cutting method that does not produce a heat-affected zone is waterjet cutting. A focussed jet of water with some abrasive additives is used to cut the pattern instead of a laser beam. Only one stent on the market has been produced by this method, the St Come SCS stainless-steel stent. Another interesting fabrication method is photochemical etching. Although this method is being used to produce stents from tubing (Interventional Technologies LP stainless-steel stent), its real benefit is in sheet processing, when large numbers of parts can be processed in a single run. Examples are the BSC/Medinol NIR stainless-steel stent and the Vascular Architects aspire nitinol stent frame (Figure 9). Geometry Early designs were generally classified as either slotted tube geometries, such as the Palmaz stents, or coil geometries, such as the Gianturco Roubin Flex stent. While slotted-tube type designs had excellent radial strength, they lackedflexibility. The opposite was true of coil designs. Conflicting design imperatives spawned a rich variety of stent geometries competing in a very crowded marketplace, each seeking an optimal balance of strength and flexibility. The course of this evolution has been documented in the geometry branch of the design map, illustrated in Figures We have chosen to classify stent geometries into five high-level categories: coil, helical spiral, woven, individual rings, or sequential rings, each of which is further refined into appropriate sub-categories as described below. Coil Most common in non-vascular applications, as the coil design allows for retrievability after implantation. These designs are extremely flexible, but their strength is limited and their low expansion ratio results in high profile devices. Figure 13 shows an example of the InStent Esophacoil device. Helical spiral These designs are generally promoted for their flexibility. With no or minimal internal connection points, they are very flexible, but also lack longitudinal support. As such, they can be subject to elongation or compression during delivery and deployment and, consequently, irregular cell size. With internal connection points, some flexibility is sacrificed in exchange for longitudinal stability and additional control over cell size. The Crossflex stent depicted in Figure 14 is an example of a minimally connected helical spiral geometry. Woven This category includes a variety of designs constructed from one or more strands of wire. Braided designs are often used for self-expanding structures, such as the WallStent, as shown in Figure 8. While these designs offer excellent coverage, they typically shorten substantially during expansion. The radial strength of such a braided structure is also highly dependent on axial fixation of its ends. The Strecker stent (Figure 3) is an example of a balloonexpandable knitted tantalum stent, while the Cook ZA (Figure 15) stent demonstrates a self-expanding knitted nitinol wire design. Individual rings Single Z -shaped rings are commonly used to support grafts or similar prostheses; they can be Figure 9. Vascular architect s aspire. Framework (left) is fabricated by photochemical etching of nitinol sheet and covered with eptfe material (right). 141
6 Figure 10. Stent geometry: helical spiral, woven, coil. individually sutured or otherwise attached to the graft material during manufacture. These structures are not typically used alone as vascular stents. Sequential rings This category describes stents comprised of a series of expandable Z-shaped structural elements (known as struts ) joined by connecting elements (known as bridges, hinges, or nodes ). This type of construction accounts for the majority of commercially available stents, and 70% of the designs included in this survey. This category can be further refined by describing the manner in which the structural elements are connected, and the nature of the resulting cells: n n Regular connection describes bridging elements that include connections to every inflection point around the circumference of a structural member. Periodic connection describes bridging elements that include connections to a subset of the inflection points around the circumference of a structural member. Connected inflection points alternate with unconnected inflection points in some defined pattern. n Peak peak connection or peak valley connection are terms used to describe the locations at which the bridging elements join adjacent structural members. Peak peak bridging elements join the outer radii, and peak valley bridging elements join outer radii to inner radii of the inflection points of adjacent structural members. Closed cell This describes sequential ring construction wherein all internal inflection points of the structural members are connected by bridging elements. Such a condition is typically only possible with regular peak-to-peak connections. Early slotted-tube type designs, such as the Palmaz stent (Figure 16), were strong, but inflexible. Later designs, such as the NIR stent (Figure 17), improved upon this concept by adding a flexconnector. These U-, V-, S-, or N-shaped elements plastically deform during bending, allowing adjacent 142
7 A survey of stent designs Figure 11. Geometry of stents manufactured as individual rings, sequential rings/closed cells. structural members to separate or nest together, to more easily accommodate changes in shape. The primary advantages of closed-cell designs are optimal scaffolding and a uniform surface, regardless of the degree of bending. However, these advantages result in a structure that is typically less flexible than a similar open-cell design. Open cell This category describes construction wherein some or all the internal inflection points of the structural members are not connected by bridging elements. This allows periodic peak-to-peak connections, peakto-valley connections, and mid-strut to mid-strut connections, as well as innumerable hybrid combinations. In open-cell designs, the unconnected structural elements contribute to longitudinal flexibility. Periodically connected peak-to-peak designs are common among self-expanding stents, such as the SMART stent (Figure 18), as well as balloonexpandable stents, such as the AVE S7 (Figure 19). The peak-to-valley connection of the ACS Multilink (Figure 20) virtually eliminates foreshortening and assures that adjacent structural peaks are aligned peak-to-valley throughout the expansion range of the stent, optimizing scaffolding characteristics. However, the peak-to-valley connectors take up material that could otherwise be used for structural members, consequently, structures with this type of peak-tovalley connection are generally not as strong as similar structures with peak-to-peak connections. While these peak-to-peak and peak-to-valley connections are most common, there are also examples of other variations, such as the BeStent 143
8 D. Stoeckel et al. Figure 12. Geometry of sequential rings/open cells. Figure 13. Esophacoil: coil stent fabricated from nitinol ribbon. Figure 14. Crossflex: a minimally connected helical spiral stent fabricated from stainless-steel wire. 144
9 A survey of stent designs Figure 15. Cook ZA: knitted nitinol wire design, featuring sleeve-type gold markers. Figure 19. AVE S7 stent: balloon expandable open-cell sequential ring design with periodic peak-to-peak non-flex connections. Figure 16. Palmaz Schatz stent: each half represents a closed-cell slotted tube structure. Figure 20. ACS Multilink: balloon expandable open-cell sequential ring design, with peak-to-valley connections. (Figure 21), which feature mid-strut to mid-strut connectors. Finally, the Navius ZR1 (Figure 22) is a unique ratcheting design that defies categorisation. Figure 17. NIR stent: a closed-cell structure featuring V flex-hinges. Figure 18. SMART stent: self-expanding open-cell sequential ring design with periodic peak-to-peak non-flex connections. Additions Although the final layer of the stent design pyramid and the branch in Figure 23 covers a range of modifications to stent designs, it would exceed the scope of this review to comment on all options. Therefore, we will only comment on radiopacity enhancements. Stents made from stainless steel or nitinol are sometimes hard to see fluoroscopically, particularly if they are small and/or have thin and narrow struts. To improve X-ray visibility, markers are often attached to the stents. These additions are typically made from gold, platinum or tantalum, and can either be sleeves crimped around a strut (Cook ZA nitinol stent with gold marker, BSC Symphony nitinol stent with platinum marker); rivets coined into tabs at the end of the stent [Cook Zilver nitinol stent with gold marker, Cordis SMARTeR nitinol stent with tantalum marker (Figure 24)] or integrated in a strut (Medtronik BeStent stainless steel with gold marker, Sorin Carbostent stainless steel with platinum marker); or welded-on tabs [Bard Luminexx Nitinol stent with tantalum tabs (Figure 25)]. Electroplating is also being used to enhance X-ray visibility. The Biotronik Tenax XR stainless steel stent 145
10 D. Stoeckel et al. has gold-plated end-segments, while the Inflow Dynamics Inflow Gold and the BSC/Medinol NIR Royal stainless steel stents were completely goldplated. Figure 21. BeStent: balloon expandable open-cell sequential ring design, with midstrut-to-midstrut connections and integral gold markers. Figure 22. Navius ZR1: ratcheting stent design fabricated from stainless-steel sheet. Conclusions In summary, we are confronted with more than 100 different stent designs. Why? This development has mainly been driven by patent and marketing issues rather than actual scientific considerations. Obviously, we do not need that many brands. Currently there are solid investigations underway testing biocompatibility, potentiodynamic polarization corrosion resistance, thrombogenicity, radiopacity, chronic fatigue behaviour in addition to more classic properties like flexibility, trackability and sheaf compatibility. An example for such investigations is the evaluation of gold coatings on stainless steel stents. The modification of stent surfaces using metallic, ceramic or polymer (not drug-eluting) coatings is considered the next step in the evolution of stent designs. However, adding surface layers does not always improve the performance of a stent. Experimental evidence had suggested that coating stents with a gold layer may have a beneficial influence. However, clinical trials were not able to Figure 23. Overview of additions to stents. 146
11 A survey of stent designs prove this hypothesis. For example, in a randomized trial by Kastrati et al. patients with symptomatic coronary artery disease were randomly assigned to receive either a gold-coated Inflow stent (n = 367) or an uncoated Inflow stainless steel stent (n = 364) of identical design [26]. Follow-up angiography was routinely performed at 6 thrombotic events observed during the first 30 days after intervention. However, the gold-coated stents were associated with a considerable increase in the risk of restenosis over the first year after stenting (49.7% in the gold-stent group and 38.1% in the steel-stent group; P = 0.003). The same finding was reported from a Korean trail in 216 patients [27]. Trials like these may allow us to conclude that stent design, material and surface preparation may have a significant impact on long-term clinical outcome. We also have to take the stent design into account when we interpret results from trials that used different stent designs. This will certainly help us Figure 24. SMARTeR tantalum radiopaque marker in nitinol tab. to discriminate actual achievements from me too products and will enable us to get closer to the ideal stent. References 1 Serruys PW, Rensing BJ editors. Handbook of coronary stents. 4th edn. London: Martin Dunitz, Machraoui A, Grewe P, Fischer A, editors. Koronarstenting. Darmstadt: Steinkopf, visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6h July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July visited on 6th July Kastrati A, Schomig A, Dirschinger J et al. Increased risk of restenosis after placement of gold-coated stents: results of a randomized trial comparing gold-coated with uncoated steel stents in patients with coronary artery disease. Circulation 2000; 101: Park SJ, Lee CW, Hang MK et al. Comparison of goldcoated NIR stents with uncoated NIR stents in patients with coronary artery disease. Am J Cardiol 2002; 89: Figure 25. Luminexx radiopaque tantalum marker welded onto nitinol. 147
A Survey of Stent Designs. Stoeckel, Bonsignore, Duda. Min Invas Ther & Allied Technol 11(4) pp. 137 147
We are Nitinol. ASurveyofStentDesigns Stoeckel,Bonsignore,Duda MinInvasTher&AlliedTechnol 11(4) pp.137 147 2002 www.nitinol.com 47533 Westinghouse Drive Fremont, California 94539 t 510.683.2000 f 510.683.2001
Stent Design for a Percutaneous Heart Valve
Stent Design for a Percutaneous Heart Valve A Major Qualifying Project submitted to the Faculty of WORCESTER POLYTECHNIC INSTITUTE in partial fulfillment of the requirements for the Degree of Bachelor
USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT
USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT V. Astier, S. Bidal, V. Garitey, F. Mouret Abstract : A stent is a wire metal mesh tube placed into an artery or blood vessel to hold the structure
Overview of Newer Stent Devices for Aneurysm Treatment
Overview of Newer Stent Devices for Aneurysm Treatment Randall C. Edgell, M.D. Associate Professor Vascular and Interventional Neurology Saint Louis University Disclosure Outcome adjudication for THERAPY,
Different Properties of Coronary Stents György Ring 1, Eszter Bognár 1, Zsófia Bálint-Pataki 1, János Dobránszky 2
Different Properties of Coronary Stents György Ring 1, Eszter Bognár 1, Zsófia Bálint-Pataki 1, János Dobránszky 2 1 Budapest University of Technology and Economics, Department of Materials Science and
Orsiro Hybrid Drug Eluting Stent Industry's first hybrid DES
Vascular Intervention Drug Eluting Stents Orsiro Orsiro Hybrid Drug Eluting Stent Industry's first hybrid DES Orsiro An ideal combination of passive and active components The Orsiro Hybrid Drug Eluting
Deployment of a Self-expanding Stent in an Artery
Deployment of a Self-expanding Stent in an Artery Santosh Prabhu*, Chris Feezor*, Andy Denison*, Nuno Rebelo** and Mouhsine Serrar** * Guidant Corporation 3200 Lakeside Drive Santa Clara, CA 95054 and
Tracheal Stent System
nt Tracheal Stent System Available sizes: 08mm x 20mm NEW 08mm x 30mm 08mmx 50 mm 08mm x 60mm 08mm x 70mm 08mm x 80mm 08mm x 90mm 10mm x 20mm NEW 10 mm x 30 mm 10 mm x 50 mm 10 mm x 60 mm 10 mm x 70 mm
TECHNOLOGICAL INNOVATION FOR CORONARY STENTS Francesco Migliavacca
LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it TECHNOLOGICAL INNOVATION FOR CORONARY STENTS Francesco Migliavacca Erice, 1 maggio 2015 International School of Cardiac Surgery Introduction
Yukon Choice DES + Translumina - The polymer-free DES solution. www.translumina.de. Coronary Stent System for Drug Application
Coronary Stent System for Drug Application Yukon Choice DES + new catheter polymer-free DES Translumina - The polymer-free DES solution www.translumina.de A First Class Stent The YUKON Choice DES+ is the
Advances in Stent Technology
Nurse/Technologist Symposium Advances in Stent Technology Antonietta Tolentino, CCRN, MSN, ANP-C Adult Nurse Practitioner, Cardiac Catheterization Laboratory The Zena and Michael A. Weiner Cardiovascular
Sheet metal operations - Bending and related processes
Sheet metal operations - Bending and related processes R. Chandramouli Associate Dean-Research SASTRA University, Thanjavur-613 401 Table of Contents 1.Quiz-Key... Error! Bookmark not defined. 1.Bending
Modelling of Stent Deployment and Deformation in Diseased Arteries by Considering Vessel Anisotropy
Modelling of Stent Deployment and Deformation in Diseased Arteries by A. Schiavone, L.G. Zhao [email protected] Abstract In this paper, finite element simulation of stent deployment was carried out using
Parameterization and Optimization of balloon expandable Stent
Parameterization and Optimization of balloon expandable Stent Karthik Shankaran, and Senthil Karrupaswamy Detroit Engineered Products (DEP) Inc Abstract: Vascular stents are deployed in the blocked arteries
North American Stainless
North American Stainless Long Products Stainless Steel Grade Sheet 2205 UNS S2205 EN 1.4462 2304 UNS S2304 EN 1.4362 INTRODUCTION Types 2205 and 2304 are duplex stainless steel grades with a microstructure,
North American Stainless
North American Stainless Flat Products Stainless Steel Grade Sheet 430 (S43000)/ EN 1.4016 Introduction: SS430 is a low-carbon plain chromium, ferritic stainless steel without any stabilization of carbon
Engineering aspects of stents design and their translation into clinical practice
Ann Ist Super Sanità 2007 Vol. 43, No. 1: 89-100 89 Engineering aspects of stents design and their translation into clinical practice Giuseppe Sangiorgi (a), Gloria Melzi (a), Pierfrancesco Agostoni (a),
Implant materials. Learning outcomes. Implant materials in trauma. How to use this handout? Functions of implants. Types of materials
Implant materials How to use this handout? The left column is the information as given during the lecture. The column at the right gives you space to make personal notes. Learning outcomes At the end of
PATIENT INFORMATION BOOKLET
PATIENT INFORMATION BOOKLET Wingspan Stent System with Gateway PTA Balloon Catheter TABLE OF CONTENTS Definitions... 2 What is the Purpose of This Booklet?... 3 What is an Intracranial Lesion?... 3 Who
EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR PRODUCTION ON THE PIERCED FLAT SHEET METAL USING LASER FORMING PROCESS
JOURNAL OF CURRENT RESEARCH IN SCIENCE (ISSN 2322-5009) CODEN (USA): JCRSDJ 2014, Vol. 2, No. 2, pp:277-284 Available at www.jcrs010.com ORIGINAL ARTICLE EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR
Flexible Printed Circuits Design Guide
www.tech-etch.com/flex Flexible Printed Circuits Design Guide Multilayer SMT Assembly Selective Plating of Gold & Tin-Lead Fine Line Microvias Cantilevered & Windowed Leads 1 MATERIALS CONDUCTOR Copper
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope Rakesh Sidharthan 1 Gnanavel B K 2 Assistant professor Mechanical, Department Professor, Mechanical Department, Gojan engineering college,
Modeling Stents Using Abaqus
Day 1 Lecture 1 Introduction Lecture 2 Geometry and Meshing Lecture 3 Element Selection and Mesh Convergence Workshop 1a Balloon-expanded stent geometry and meshing Workshop 1b Self-expanding stent geometry
North American Stainless
North American Stainless Flat Products Stainless Steel Sheet T409 INTRODUCTION NAS 409 is an 11% chromium, stabilized ferritic stainless steel. It is not as resistant to corrosion or high-temperature oxidation
TARIFF CODE and updates standard
TARIFF CODE and updates standard No HS CODE AHTN CODE PRODUCT DESCRIPTION PRODUCT TYPE STANDARDS IDENTIFIED 7207 Semi finished products of iron or non alloy steel Containing by weight less than 0.25% of
NEW SIX-SEALS CUTTING RING. INTERNATIONAL INDUSTRIAL PATENT Nr. 864061 of the 10/03/99 FLANKS AND DOES NOT REPLACE THE STANDARD RING CURRENTLY IN USE
B4 NEW SIX-SEALS CUTTING RING. INTERNATIONAL INDUSTRIAL PATENT Nr. 864061 of the 10/03/99 FLANKS AND DOES NOT REPLACE THE STANDARD RING CURRENTLY IN USE AVAILABLE IN CARBON AND STAINLESS STEEL 23 THEORY
Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied
Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied Stress and strain fracture or engineering point of view: allows to predict the
MANUFACTURING OF STENTS: OPTIMIZE THE STENT WITH NEW MANUFACTURING TECHNOLOGIES
12 MANUFACTURING OF STENTS: OPTIMIZE THE STENT WITH NEW MANUFACTURING TECHNOLOGIES Andreas SCHUESSLER, Ullrich BAYER, Gerd SIEKMEYER, Rainer STEEGMUELLER, Martin STROBEL, Admedes SCHUESSLER We proposed
Mechanical Properties of Nitinol Stents and Stent-grafts. Comparison of 6 mm Diameter Devices
Mechanical Properties of Nitinol Stents and Stent-grafts Comparison of 6 mm Diameter Devices P E R F O R M A N C E b y d e s i g n Purpose The association of target vessel restenosis with stent fractures
Simulation of Residual Stresses in an Induction Hardened Roll
2.6.4 Simulation of Residual Stresses in an Induction Hardened Roll Ludwig Hellenthal, Clemens Groth Walzen Irle GmbH, Netphen-Deuz, Germany CADFEM GmbH, Burgdorf/Hannover, Germany Summary A heat treatment
THE STRAIN GAGE PRESSURE TRANSDUCER
THE STRAIN GAGE PRESSURE TRANSDUCER Pressure transducers use a variety of sensing devices to provide an electrical output proportional to applied pressure. The sensing device employed in the transducers
Corrugated Tubular Heat Exchangers
Corrugated Tubular Heat Exchangers HEAT EXCHANGERS for the 21st CENTURY Corrugated Tubular Heat Exchangers (CTHE) Corrugated Tube Heat Exchangers are shell and tube heat exchangers which use corrugated
EDEXCEL NATIONAL CERTIFICATE/DIPLOMA MECHANICAL PRINCIPLES OUTCOME 2 ENGINEERING COMPONENTS TUTORIAL 1 STRUCTURAL MEMBERS
ENGINEERING COMPONENTS EDEXCEL NATIONAL CERTIFICATE/DIPLOMA MECHANICAL PRINCIPLES OUTCOME ENGINEERING COMPONENTS TUTORIAL 1 STRUCTURAL MEMBERS Structural members: struts and ties; direct stress and strain,
TURQUOISE PTCA BALLOON DILATATION CATHETER
TURQUOISE PTCA BALLOON DILATATION CATHETER TURQUOISE PTCA BallOOn Dilatation CATheter The right BAlance TO meet the clinical need TURQUOISE PTCA s advanced catheter technology offers the right balance
Material data sheet. EOS StainlessSteel GP1 for EOSINT M 270. Description, application
EOS StainlessSteel GP1 for EOSINT M 270 A number of different materials are available for use with EOSINT M systems, offering a broad range of e-manufacturing applications. EOS StainlessSteel GP1 is a
When Procedural Support really matters. Navien TM A+ Intracranial Support Catheter
When Procedural Support really matters Navien TM A+ Intracranial Support Catheter Coils / Liquid Embolics DURABLE ovalization resistance, Maintains patent lumen in tortuosity for smooth micro catheter
Coronary Stents. What is a Coronary Artery stent?
What is a Coronary Artery stent? Coronary Stents A coronary stent is stainless tube with slots. It is mounted on a balloon catheter in a "crimped" or collapsed state. When the balloon of is inflated, the
Chapter 5 - Aircraft Welding
Chapter 5 - Aircraft Welding Chapter 5 Section A Study Aid Questions Fill in the Blanks 1. There are 3 types of welding:, and, welding. 2. The oxy acetylene flame, with a temperature of Fahrenheit is produced
Majestic Trial 12 Month Results
Majestic Trial 12 Month Results S.Müller-Hülsbeck, MD, EBIR, FCIRSE, FICA ACADEMIC HOSPITALS Flensburg of Kiel University Ev.-Luth. Diakonissenanstalt zu Flensburg Knuthstraße 1, 24939 FLENSBURG Dept.
Biomaterials in tissue engineering
Biomaterials in tissue engineering S. Swaminathan Director Centre for Nanotechnology & Advanced Biomaterials School of Chemical & Biotechnology SASTRA University Thanjavur 613 401 Tamil Nadu Page 1 of
Profiles in stock Standard catalogue. Discover the possibilities with aluminium profiles from Sapa
Profiles in stock Standard catalogue Discover the possibilities with aluminium from Sapa You can create your own profile or you can choose from a wide selection of standard. Delivery: 1 2 weeks for standard
Metal Injection Molding (MIM) of components made of Titanium and its alloys
Metal Injection Molding (MIM) of components made of Titanium and its alloys 1 Presentation content Introduction to Metal Injection Molding (MIM) Technology - explaination Products - examples Company brief
COATED CARBIDE. TiN. Al 2 O 3
COATED CARBIDE GENERAL INFORMATION CVD = Chemical Vapour Deposition coated grades GC2015, GC2025, GC2135, GC235, GC3005, GC3015, GC3020, GC3025, GC3115, GC4015, GC4025, GC4035, S05F, and CD1810. PVD =
Evolu'on of Balloon Angioplasty From POBA to newer technologies NCVH May 26, 2015
Evolu'on of Balloon Angioplasty From POBA to newer technologies NCVH May 26, 2015 Cezar Staniloae, MD Heart and Vascular Ins7tute New York University Medical Center Overview 1. Physics of balloon angioplasty
FLAME CuTTIng te la P is utting B c
FLAME CUTTING Both Oxy-LPG and Oxy-acetylene processes are acceptable for sectioning all thicknesses of BISPLATE. With these processes, the following techniques are recommended: Gas pressure to be the
TruLaser Tube: Optimal tube. and. profile cutting. Machine Tools / Power Tools Laser Technology / Electronics Medical Technology
TruLaser Tube: Optimal tube and profile cutting Machine Tools / Power Tools Laser Technology / Electronics Medical Technology Giving you the edge in laser tube processing. The TRUMPF Group ranks among
Reliable Metal Bending
Reliable Metal Bending Since 1984 Welcome to Bendpro CNC Mandrel Bending Our commitment to you... P3 Accuracy Guaranteed Quick response quotes F.I.S. freight inclusive pricing Short lead-times 25+ years
ELEVATOR TRAVELING CABLE - DESIGN EVOLUTION
ELEVATOR TRAVELING CABLE - DESIGN EVOLUTION by Richard Laney Introduction Elevator traveling cable is a vital link between the elevator car and controller. In conventional elevators, all power and signal
Crimp Tooling Where Form Meets Function
Crimp Tooling Where Form Meets Function Quality, cost, and throughput are key attributes for any production process. The crimp termination process is no exception. Many variables contribute to the results.
Cable Construction Guide This article was kindly donated by George Cardas
Cable Construction Guide This article was kindly donated by George Cardas Recent years have seen ferment in every area of high-end audio. Nowhere has this produced more of a revolution (controversy to
INTRODUCTION TO BEAMS
CHAPTER Structural Steel Design LRFD Method INTRODUCTION TO BEAMS Third Edition A. J. Clark School of Engineering Department of Civil and Environmental Engineering Part II Structural Steel Design and Analysis
7.3 Fit selection. Inner ring: Rotating. Outer ring: Stationary. Inner ring: Stationary. Outer ring: Rotating. Inner ring: Stationary
7. Bearing Fits 7. Fitting For rolling s, inner and outer rings are fixed on the or in the housing so that relative movement does not occur between fitting surfaces during operation or under. This relative
November 20, 2014. Vascular Flow Technologies Ltd. Edwin Lindsay VP of QA/RA Prospect Business Centre, Gemini Cresent Dundee DD2 1TY United Kingdom
DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 20, 2014 Vascular
StentViz Enhanced Stent Visualization
GE Healthcare StentViz Enhanced Stent Visualization Dr. Morice, Dr. Lefèvre, Dr. Hovasse, Dr. Chevalier, Dr. Louvard Institut Cardiovasculaire Paris Sud, Massy, France Assessing the deployment of stents
Technical Data. 7. Bearing Fits. 7.1 Interference. 7.2 Calculation of interference F B LLLLLLLLL( A-54
Technical Data 7. Bearing Fits 7.1 Interference For rolling s the rings are fixed on the or in the housing so that slip or movement does not occur between the mated surface during operation or under. This
Mechanical Property Changes in Steel during the Pipe Making Process Brent Keil 1
381 Mechanical Property Changes in Steel during the Pipe Making Process Brent Keil 1 Abstract Welded Steel Pipe (WSP) is arguably the most widely utilized pipe material for the transmission of water throughout
CE APPROVALS FOR MORE THAN 70 DIFFERENT PRODUCTS COMPANY PROFILE
CONTENT 05 06 08 09 10 12 14 16 18 Company Profile Board of Directors Modified Generic Strategy Unique Technology Development Methodology Our 12 Step Manufacturing & Development Facilities & Products History
The Minvasys Amazonia Pax & Nile Pax Polymer Free Paclitaxel Eluting Stent Program
The Minvasys Amazonia Pax & Nile Pax Polymer Free Paclitaxel Eluting Stent Program Jean Fajadet, MD, FESC Clinique Pasteur - Toulouse - France Disclosure statement Nothing to disclose Dedicated Delivery
Ostial LAD: Single stent approach is the best. Antonio A. Pocoví, MD, FSCAI, MTSAC, Advisory Council Member, CACI
Ostial LAD: Single stent approach is the best Antonio A. Pocoví, MD, FSCAI, MTSAC, Advisory Council Member, CACI Chair, Interventional Cardiology Sanatorio San Lucas Instituto Alexander Fleming Buenos
An Introduction to Spiral Retaining Ring Installation and Removal
An Introduction to Spiral Retaining Ring Installation and Removal FACTS AND FIGURES OF SPIRAL RETAINING RINGS Retaining Rings act as a removable shoulder on a shaft or in a bore. When installed in a groove,
Structural Integrity Analysis
Structural Integrity Analysis 1. STRESS CONCENTRATION Igor Kokcharov 1.1 STRESSES AND CONCENTRATORS 1.1.1 Stress An applied external force F causes inner forces in the carrying structure. Inner forces
AUSTENITIC STAINLESS DAMASCENE STEEL
AUSTENITIC STAINLESS DAMASCENE STEEL Damasteel s austenitic stainless Damascene Steel is a mix between types 304L and 316L stainless steels which are variations of the 18 percent chromium 8 percent nickel
Bending, Forming and Flexing Printed Circuits
Bending, Forming and Flexing Printed Circuits John Coonrod Rogers Corporation Introduction: In the printed circuit board industry there are generally two main types of circuit boards; there are rigid printed
PS-SEAL PTFE based high performance seal
PS-SEAL PTFE based high performance seal Advancing the Science of Sealing Advancing the Science of Sealing Table of content PS-SEAL 3 PS-SEAL STANDARD Standard seal in stock 4 PS-SEAL NON-STANDARD Customized
Heat Pipe Selection Revision 12/04/2001
Heat Pipe Selection Revision 12/04/2001 Heat pipes are being used very often in particular applications when conventional cooling methods are not suitable. Once the need for heat pipe arises, the most
Anodes and Misc Equipment
Anodes and Misc Equipment Application: Platinised Titanium Anodes Platinised titanium anodes are recommended for use in the following electrolytic processes:- Precious metal electroplating - e.g. Au, Pt,
Technology Insight Report
Technology Insight Report Coronary Stents Cardiology has come a long way over the decades with huge advances having been made in areas such as invasive surgery and interventional cardiology. As a result
Local buckling of plates made of high strength steel
Local buckling of plates made of high strength steel Tapani Halmea, Lauri Huusko b,a, Gary Marquis a, Timo Björk a a Lappeenranta University of Technology, Faculty of Technology Engineering, Lappeenranta,
WWER Type Fuel Manufacture in China
WWER Type Fuel Manufacture in China Yang Xiaodong P.O. Box 273, CJNF, YiBin City, Sichuan, China, [Fax: (+86)8318279161] Abstract: At CJNF, a plan was established for implementation of technical introduction
FLEXIBLE COPPER & ALUMINIUM CONNECTIONS. Product Range 1 VGL - ALLIED CONNECTORS
Product Range ROUND STRANDED COPPER CABLE Electrolytic Copper / Oxygen Free Copper Conforms to DIN 46438 Standards Finish: Bare Copper / Tinned / Silver-plated ROUND BRAID / ROUND STRANDED CABLE WITH OVERALL
Weld Cracking. An Excerpt from The Fabricators' and Erectors' Guide to Welded Steel Construction. The James F. Lincoln Arc Welding Foundation
Weld Cracking An Excerpt from The Fabricators' and Erectors' Guide to Welded Steel Construction The James F. Lincoln Arc Welding Foundation Weld Cracking Several types of discontinuities may occur in welds
PAAM COMMERCIAL PRIVATE LIMITED
METAL PACKAGING We at Paam Group have been providing vast range of packaging products to our customers based on their preferences, needs and specifications. We provide all the materials required in metal
Angioplasty and Stent Education Guide
Angioplasty and Stent Education Guide Table of Contents Treating coronary artery disease...2 What is coronary artery disease...3 Coronary artery disease treatment options...4 What are coronary artery
Application Tooling Specification Sheet
Modular Crimp Head Order No. 63825-2170 Application Tooling Specification Sheet TYPE 4A Hand Crimp Tool Order No. 63825-2100 FEATURES % A full cycle ratcheting hand tool ensures complete crimps % Ergonomically
Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials.
Lab 3 Tension Test Objectives Concepts Background Experimental Procedure Report Requirements Discussion Objectives Experimentally determine the yield strength, tensile strength, and modules of elasticity
Popliteal artery: to stent or not to stent?
Popliteal artery: to stent or not to stent? Karl-Ludwig Schulte Vascular Center Berlin Ev. Hospital Königin Elisabeth St. Gertrauden Hospital University Hospital Charité, CC11 Humboldt-University Berlin
THE INFLUENCE OF STEEL GRADE AND STEEL HARDNESS ON TOOL LIFE WHEN MILLING IN HARDENED TOOL STEEL
THE INFLUENCE OF STEEL GRADE AND STEEL HARDNESS ON TOOL LIFE WHEN MILLING IN HARDENED TOOL STEEL S. Gunnarsson, B. Högman and L. G. Nordh Uddeholm Tooling AB Research and Development 683 85 Hagfors Sweden
Metal-O-rings and C-rings for extreme operating conditions
Metal-O-rings and C-rings for extreme operating conditions From ultrahigh vacuum up to 6800 bar (100000 psi) pressure and temperatures from -269 C cryogenic up to +980 C (-425 F up to 1800 F). Resistant
INVESTIGATION OF METALLIC SURFACE AREA OF CORONARY STENTS
INVESTIGATION OF METALLIC SURFACE AREA OF CORONARY STENTS Dóra Károly 1, Miksa Kovács 1, Andrew Attila Terdik 1, Eszter Bognár 1,2 1Department of Materials Science and Engineering, Faculty of Mechanical
Stability. Security. Integrity.
Stability. Security. Integrity. PN #MBHPT Foundation Supportworks provides quality helical pile systems for both new construction and retrofit applications. 288 Helical Pile System About Foundation Supportworks
Unit 6: EXTRUSION. Difficult to form metals like stainless steels, nickel based alloys and high temperature metals can also be extruded.
1 Unit 6: EXTRUSION Introduction: Extrusion is a metal working process in which cross section of metal is reduced by forcing the metal through a die orifice under high pressure. It is used to produce cylindrical
Engine Bearing Materials
Engine Bearing Materials Dr. Dmitri Kopeliovich (Research & Development Manager) The durable operation of an engine bearing is achieved if its materials combine high strength (load capacity, wear resistance,
Comparison of Second-Generation Stents for Application in the Superficial Femoral Artery: An In Vitro Evaluation Focusing on Stent Design
Comparison of Second-Generation Stents for Application in the Superficial Femoral Artery: An In Vitro Evaluation Focusing on Stent Design Müller-Hülsbeck S, Schäfer PJ, Charalambous N, Yagi H, Heller M,
MCE380: Measurements and Instrumentation Lab. Chapter 9: Force, Torque and Strain Measurements
MCE380: Measurements and Instrumentation Lab Chapter 9: Force, Torque and Strain Measurements Topics: Elastic Elements for Force Measurement Dynamometers and Brakes Resistance Strain Gages Holman, Ch.
21. - 22. 11. 2012, Plzeň, Czech Republic, EU
PRO-BEAM TECHNOLOGY - HEAT TREATMENT OF METALS Václav HOŠEK a, Filip VRÁBLÍK a, Uwe CLAUß b, Pavel STOLAŘ a a ECOSOND s.r.o., Čerčany, Czech Republic, EU, [email protected] b pro-beam AG & Co.KgaA, Neukirchen,
ROUNDO 3-Roll Plate Bending Machines Type PS
ROUNDO 3-Roll Plate Bending Machines Type PS Universal Machine used for all Plate Bending Double Pinch Design 1. Align the plate using the alignment grooves in the lower rolls or against the rear roll.
MASTER DEGREE PROJECT
MASTER DEGREE PROJECT Finite Element Analysis of a Washing Machine Cylinder Thesis in Applied Mechanics one year Master Degree Program Performed : Spring term, 2010 Level Author Supervisor s Examiner :
Handbook on the Ultrasonic Examination. Austenitic Welds
Handbook on the Ultrasonic Examination Austenitic Welds The International Institute of Welding Edition Handbook On the Ultrasonic Examination of Austenitic Welds Compiled by COMMISSION V Testing, Measurement,
Chapter 17: Springs. Fundamentals of Machine Elements, 3 rd ed. Schmid, Hamrock and Jacobson. 2014 CRC Press
Chapter 17: Springs It must be confessed that the inventors of the mechanical arts have been much more useful to men than the inventors of syllogisms. Voltaire A collection of helical compression springs.
High-strength and ultrahigh-strength. Cut sheet from hot-rolled steel strip and heavy plate. voestalpine Steel Division www.voestalpine.
High-strength and ultrahigh-strength TM steels Cut sheet from hot-rolled steel strip and heavy plate Josef Elmer, Key account manager voestalpine Steel Division www.voestalpine.com/steel Weight savings
ElastoFlake & ElastoTec
ElastoFlake & ElastoTec Relining Group 02 Revolutionally injection casting ElastoFlake, a new material from the Relining Group, is a polymer plastic mass intended for injection casting, which was developed
WallFlex Biliary RX Stent. Fully, Partially and Uncovered Self-Expanding Metal Stents
WallFlex Biliary RX Stent Fully, Partially and Uncovered Self-Expanding Metal Stents WallFlex Biliary RX Stent Fully, Partially and Uncovered Self-Expanding Metal Stents The WallFlex Biliary RX Stent is
Radial Insert Ball Bearings Housing Units. Our Expertise is Your Benefit. Schaeffler Group Industrial
Radial Insert Ball Bearings Housing Units Our Expertise is Your Benefit Schaeffler Group Industrial A p p l i c a t i o n s Quality for every eventuality: Radial insert ball bearings in industry Radial
Abrasive-Flow Machining
1 Polishing Using Magnetic Fields Figure 25.30 Schematic illustration of polishing of balls and rollers using magnetic fields. (a) Magnetic float polishing of ceramic balls. (b) Magnetic-field-assisted
Non- Carbon Fiber Electrical Heating Textile Introduction:
Non- Carbon Fiber Electrical Heating Introduction: This Heating is a new generation electrical heating system which is specially designed for mobile heating, energy saving and safety orientated requirements
Flex Circuit Design and Manufacture.
Flex Circuit Design and Manufacture. Hawarden Industrial Park, Manor Lane, Deeside, Flintshire, CH5 3QZ Tel 01244 520510 Fax 01244 520721 [email protected] www.merlincircuit.co.uk Flex Circuit