Engineering aspects of stents design and their translation into clinical practice
|
|
|
- Geoffrey Quinn
- 10 years ago
- Views:
Transcription
1 Ann Ist Super Sanità 2007 Vol. 43, No. 1: Engineering aspects of stents design and their translation into clinical practice Giuseppe Sangiorgi (a), Gloria Melzi (a), Pierfrancesco Agostoni (a), Clarissa Cola (b), Fabrizio Clementi (b), Paolo Romitelli (c), Renu Virmani (d) and Antonio Colombo (a) (a) Unità Operativa di Emodinamica, Ospedale San Raffaele, Milan, Italy (b) Dipartimento di Cardiologia, Università di Roma Tor Vergata, Rome, Italy (c) Divisione Vascolare, Medtronic Italia, Milan, Italy (d) CVPath Gaithersburg, Maryland, USA Summary. The implantation of coronary stents is a relevant part of interventional procedures for percutaneous revascularization. The wide acceptance of coronary stenting was based on the results of two highly significant trials which have shown the superiority of stenting over balloon angioplasty in terms of reduction of angiographic restenosis and need for repeated intervention in focal lesions and large coronary arteries. Since then, the growing use of stent market was impressive. A rapidly increasing number of different stent type with different material and designs has been introduced in the market both for bare metal stent and drug eluting stent. This review will summarize the different components of stent design that are important in term of biological response of the arterial wall and clinical outcome. In addition, new stent platforms, mainly represented by the biodegradable stent will be shortly reviewed since it may provide in the near future a more physiological answer to stent implantation, reducing vascular injury and accelerating vessel healing with consequent improving in clinical outcome. Key words: coronary artery disease, stent technology, drug-eluting stents, percutaneous intervention. Research and Methodologies/Ricerche e Metodologie Riassunto (Caratteristiche tecniche del disegno degli stent e risvolti correlati nella pratica clinica). L impianto di stent nell albero coronarico è una delle procedure più utilizzate nel campo della rivascolarizzazione meccanica percutanea della malattia aterosclerotica coronarica. L ampia accettazione di questi dispositivi è venuta dai risultati di due trials clinici che hanno mostrato come l utilizzo dello stent in arterie di grosso calibro ed in lesioni focali sia significativamente superiore alla sola angioplastica con palloncino in termini di riduzione della restenosi angiografica e di necessità di ripetere un ulteriore procedura a distanza. Da allora la crescita dello stent nel campo della cardiologia interventistica è stato esponenziale. Un numero sempre crescente di differenti tipi di stent con diverso disegno e materiali utilizzati per la costruzione sono stati introdotti sia per gli stent metallici tradizionali che più recentemente per gli stent medicati. Questo lavoro vuole riassumere dal punto di vista ingegneristico i differenti componenti e materiali che caratterizzano il disegno dello stent e dal punto di vista clinico come un disegno dello stent ideale possa influenzare la risposta biologica della parete vasale ed il risultato clinico. In ultimo saranno brevemente rivisti le nuove piattaforme di stent ed in particolari gli stent biodegradabili, che in un futuro ormai prossimo saranno introdotti nel mercato con la speranza che il ridotto danno vascolare e l aumentata risposta cicatriziale della parete che caratterizza questi stent di ultima generazione si possa tradurre in un maggiore beneficio clinico nei pazienti che si sottopongono alle procedure di rivascolarizzazione percutanea. Parole chiave: malattia aterosclerotica coronarica, aspetti tecnologici degli stent, stent medicati, interventistica percutanea. INTRODUCTION The implantation of coronary stents is an integral part of most interventional procedures for percutaneous revascularization. The wide acceptance of coronary stenting was based on the results of the BElgian NEtherlands STENT (BENESTENT) [1] and the STent REStenosis Study (STRESS) [2] trials, which showed the superiority of stenting over balloon angioplasty in terms of reduction of angiographic restenosis and need for repeated intervention in focal lesions and large coronary arteries. Since then, the growing use of stents in ever more complex lesions and patients [3] has stimulated the introduction of a rapidly increasing number of different stent designs. Among the reasons why different designs have been proposed, there are concepts of physiological mechanisms: indeed a primary concern of stent development is the need to reduce device pro- Indirizzo per la corrispondenza (Address for correspondence): Giuseppe Sangiorgi, Ospedale San Raffaele, Via Olgettina 64, Milan, Italy. [email protected].
2 90 Giuseppe Sangiorgi, Gloria Melzi, Pierfrancesco Agostoni, et al. files and to increase flexibility to facilitate safe delivery. A major caveat, almost completely resolved by double oral antiplatelet therapy, regards the possible occurrence of thrombotic episodes, hours to days after stent deployment, leading to relevant clinical events (sudden cardiac death or acute myocardial infarction). Other important issues are lesion coverage, to avoid plaque prolapse, and radial support, to prevent elastic recoil of the artery. Furthermore, the possibility to easily access side branches through the stent struts of a deployed stent in bifurcation lesions has progressively gained importance. Finally, radiological visibility is another important element to optimize the clinical benefits of a stent. In clinical practice, the operator must decide which stent is most appropriate for the patient, and even more importantly, for the lesion that is going to be treated. General characteristics pertaining to the ideal stent are reported in the following: - flexible; - trackable; - low unconstrained profile; - radio-opaque; - thromboresistant; - biocompatible; - reliably expandable; - high radial strength; - circumferential coverage; - low surface area; - hydrodynamic compatible. TYPES OF STENTS Stents can be classified according to several engineering variables [4] which influence stent characteristics, biocompatibility and outcome (Figure 1): Fig. 1 Stent design pyramid with different material and construction characteristics. - mechanism of expansion (self-expanding or balloon-expandable); - materials (stainless steel, cobalt-based alloy, tantalum, nitinol, inert coating, active coating, or biodegradable); - forms (sheet, wire or tube); - manufacturing methods (laser cut, water-jet cutting, photo-etching, etc.); - geometrical configurations/design (mesh structure, coil, slotted tube, ring, multi-design, or custom design); - addition to stent (grafts, radio-opaque markers, coatings, etc.). MECHANISM OF EXPANSION In general stents can be divided on the basis of the mechanism of expansion into two major types: selfexpanding stents and balloon-expandable stents. Balloon-expandable stents are made from materials that can be plastically deformed through the inflation of a balloon; after the balloon is deflated the stent remains in its expanded shape, except for a slight recoil caused by the elastic portion of the deformation. Self-expanding stents, on the other hand, are manufactured in the expanded shape, then compressed and constrained in a delivery system. Upon release from the delivery system they spring back, i.e. self-expand, to the preset diameter. MATERIALS Materials for metallic balloon-expandable or selfexpanding stents must exhibit excellent corrosion resistance and biocompatibility (Table 1); they should be adequately radio-opaque, and create minimal artifacts during MRI. For balloon-expandable stents the ideal material for construction should have a low yield stress (to make it deformable at manageable balloon pressures), high elastic modulus (for minimal recoil), and work hardened through expansion for high strength. Balloon-expandable stents are manufactured in the small diameter, i.e. deliverable configuration, pre-mounted over a balloon, and they are deployed by balloon inflation (that permits them to obtain their expanded shape) at the target site inside the vessel. On the contrary, the function of self-expanding stents is based on the elastic properties of the material used. Ideally, the material should have a low elastic modulus and a high yield stress for large elastic strains. Alternatively, the shape-memory effect of nitinol can be utilized. Here, large strains can be achieved either super-elastically, or via the thermal memory of the material. The most widely used material for stents is stainless steel, typically 316L, a particularly corrosion-resistant material with low carbon content and additions of molybdenum and niobium. In its fully annealed condition, stainless steel is easily deformable and, therefore, the standard material for balloon-expandable stents.
3 Stents design 91 Table 1 Overview of materials used in balloon-expandable and self-expandable stents manufacture, different stent forms, stent fabrication, stent geometry and addictions Materials Form Fabrication Balloon expandable stents Self-expanding stents Wire Tube Sheet Ribbon Laser cutting Photochemical etching Braiding Knitting Vapor deposition Water jet - Stainless steel 316L (vast majority) - Tantalum - Martensitic nitinol - Platinum iridium - Polymers - Niobium alloy - Cobalt alloy - Superelsatic nickel-titanium nitinol (majority) - Cobalt alloy - Full hard (stainless steel) - Wallstent (cobalt alloy) - Bridge, S7, S660, (stainless steel, welded rings) - Angiostent (platinum iridium) - Strecker (tantalum) - Expander (nitinol) - Vast majority - NIR (steinless steel) - ZR1 (stainless steel) - GRII (stainless steel) - Endotex (nitinol) - Horizon prostatic (nitinol) - EndoCoil, esophacoil (nitinol) - Vast majority - NIR - Nitinol sheet - Coiled nitinol framework, eptfe covering - Wallstent (cobalt alloy) - Streaker (tantalum) - SCS, SCS-Z stent Helical spiral - Periodic peak to peak connections - No/minimal connections - Axial spine - Integral with graft Geometry Slotted Tube / Coil Woven Individual rings Sequential rings Open cells Closed cells - Braided - Knitted - Peak-peak connections - Peak-valley connections - Midstruts connections - Hybrids - Other - Regular peak-peak connection - Non-flex connector - Flex connector - Combined connector - Hybrid Coil Addictions Covering Radiopaque markers Radiopaque coating Biocompatibility coatings Drug eluting coating - WallGraft; coiled nitinol framework, eptfe covering - Tabs: (tantalum end, gold end, platinum within strut) - Sleeve:(gold, platinum) - Welded:(tantalum) - Gold, silicone carbide over gold - Tantalum coating, phosphorylcholine, carbon coating, silicone carbide - Rapamicyne, paclitaxel
4 92 Giuseppe Sangiorgi, Gloria Melzi, Pierfrancesco Agostoni, et al. In its full-hard condition, on the other hand, it exhibits enough elasticity for certain self-expanding stent designs. Alternative materials for balloonexpandable stents are tantalum, platinum alloys, niobium alloys and cobalt alloys. They are used for their better radio-opacity, higher strength, improved corrosion resistance, better MR compatibility or the combination of all these features. Better radio-opacity and higher strength allow the design of stents with smaller delivery profiles. As mentioned above, materials for self-expanding stents should exhibit large elastic strains. The most widely used material is nitinol, a nickel-titanium alloy that can recover elastic deformation of up to 10%. This unusually wide elastic range, commonly known as super-elasticity, is the result of a thermo-elastic martensitic transformation. The limited elastic range of more conventional materials, such as stainless steel or certain cobalt-based alloys, also limits design options. RAW MATERIAL FORMS Stents can be made from sheet, wire (round or flat) or tubing (Table 1). A large majority of balloon-expandable and self-expanding stents are made from wire or tubing. A few exceptions are made from sheet metal. Stents made from sheet metal have to be rolled up to a tubular configuration after the pattern has been created. FABRICATION METHODS The choice of fabrication method depends mainly on the raw material form used (Table 1). Wires can be formed into stents in various ways using conventional wire-forming techniques, such as coiling, braiding or knitting. The simplest shape for a wire stent is a coil. All coil stents marketed today are made from nitinol and are self-expanding. Welding at specific locations after wire-forming produces closedcell wire stents or increases longitudinal stability. The most common wire-based self-expanding stent is the WallStent (Boston Scientific), a braided design using multiple elgiloy (cobalt-based alloy) wires. This allows continuous production, i.e. the stents can be cut to a specific length from a long wire-mesh hose. Knitting allows the production of flexible balloon-expandable and self-expanding wire stents. The vast majority of coronary stents, and probably the majority of peripheral vascular stents, are produced by laser cutting from tubing. Typically, Nd: YAG lasers are used, allowing kerf widths of < 20 µm. Intricate patterns can be produced using tube sizes from 0.5 mm diameter. Balloon-expandable stents are cut in the crimped or near-crimped condition, and only require post-cutting deburring and surface treatment, typically electro-polishing. They are marketed as balloon-mounted, or -unmounted for hand-crimping. Self-expanding nitinol stents, on the other hand, can be cut either in the small configuration, requiring post-cutting expansion and shape-setting, or in the expanded condition. In either case, they have to be deburred and polished. Self-expanding stents have to be constrained in the delivery system and, therefore, are not available in an unmounted configuration. Laser cutting produces a heat-affected zone along the cut edge, which has to be removed for better performance. A cutting method that does not produce a heat-affected zone is water-jet cutting. A focused jet of water with some abrasive additives is used to cut the pattern instead of a laser beam. Another interesting manufacturing method is photochemical etching. Although this method is being used to produce stents from tubing, its real benefit is in sheet processing, when large numbers of parts can be processed in a single run. GEOMETRY Early designs were generally classified as either slotted tube geometries, such as the Palmaz stents, or coil geometries, such as the Gianturco-Roubin Flex stent. While slotted-tube type designs had excellent radial strength, they lacked flexibility. The opposite occurred for coil designs. The subsequent evolution of stent design yielded to the development of a rich variety of stent geometries, which can be classified into five main high-level categories: coil, helical spiral, woven, individual rings or sequential rings (Table 1): Coil Most common in non-vascular applications, as the coil design allows for retrievability after implantation. These designs are extremely flexible, but their strength is limited and their low expansion ratio results in high profile devices. Helical spiral These designs are generally promoted for their flexibility. With no or minimal internal connection points, they are very flexible, but also lack longitudinal support. As such, they can be subject to elongation or compression during delivery and deployment and, consequently, irregular cell size formation. With internal connection points, some flexibility is sacrificed in order to obtain a higher longitudinal stability and additional control over cell size. Woven This category includes a variety of designs constructed from one or more strands of wire. Braided designs are often used for self-expanding structures. While these designs offer excellent coverage, they typically shorten substantially during expansion. The radial strength of such a braided structure is also highly dependent on the axial fixation of its ends. Individual rings Single Z -shaped rings are commonly used to support grafts or similar prostheses; they can be in-
5 Stents design 93 dividually sutured or otherwise attached to the graft material during manufacture. These structures are not typically used alone as vascular stents. Sequential rings This category describes stents comprised of a series of expandable Z-shaped structural elements (known as struts ) joined by connecting elements (known as bridges, hinges, or nodes ). This type of construction accounts for the majority of commercially available stents. This category can be further refined according to the manner in which the structural elements are connected, and the nature of the resulting cells: - regular connection describes bridging elements that include connections to every inflection point around the circumference of a structural member; - periodic connection describes bridging elements that include connections to a subset of the inflection points around the circumference of a structural member. Connected inflection points alternate with unconnected inflection points in some defined pattern; - peak-peak connection or peak-valley connection are terms used to describe the locations where the bridging elements join adjacent structural members. Peak-peak bridging elements join the outer radii, and peak-valley bridging elements join outer radii to inner radii of the inflection points of adjacent structural members. Closed cell This describes sequential ring construction wherein all internal inflection points of the structural members are connected by bridging elements. Such a condition is typically only possible with regular peak-to-peak connections. Early slotted-tube type designs, such as the Palmaz stent, were strong, but inflexible. Later designs, such as the NIR stent, improved upon this concept by adding a flex-connector. These U-, V-, S-, or N-shaped elements plastically deform during bending, allowing adjacent structural members to separate or nest together, to more easily accommodate changes in shape. The primary advantages of closed-cell designs are optimal scaffolding and a uniform surface, regardless of the degree of bending. However, these advantages result in a structure that is typically less flexible than a similar open-cell design. Open cell This category describes construction wherein some or all the internal inflection points of the structural members are not connected by bridging elements. This allows periodic peak-to-peak connections, peak-to-valley connections, and mid-strut to midstrut connections, as well as innumerable hybrid combinations. In open-cell designs, the unconnected structural elements contribute to longitudinal flexibility. Periodically connected peak-to-peak designs are common among self-expanding stents, such as the SMART stent, as well as balloon-expandable stents, such as the AVE S7. The peak-to-valley connection of the ACS Multilink virtually eliminates foreshortening and assures that adjacent structural peaks are aligned peak-to-valley throughout the expansion range of the stent, optimizing scaffolding characteristics. However, the peak-to-valley connectors take up material that could otherwise be used for structural members. Consequently, structures with this type of peak-to-valley connection are generally not as strong as similar structures with peakto-peak connections. While these peak-to-peak and peak-to-valley connections are the most common, there are also examples of other variations, such as the BeStent, which feature mid-strut to mid-strut connectors. ADDITIONS Radio-opacity enhancements Stents made from stainless steel or nitinol are sometimes hard to see fluoroscopically, particularly if they are small and/or have thin and narrow struts. To improve X-ray visibility, markers are often attached to the stents. These additions are typically made from gold, platinum or tantalum, and can either be sleeves crimped around a strut, rivets coined into tabs at the end of the stent or integrated in a strut, or welded-on tabs (Table 1). Electroplating (with gold) is also being used to enhance X-ray visibility. Coatings Several active compounds have been used to cover stents in order to increase their biocompatibility and to possibly interact with different cells and molecules present in the blood and in the vessel wall. The rationale was to obtain a significant reduction in the activity of these cells and molecules, enhancing the safety and effectiveness of stents. Among the different compounds tested, heparin was one of the first. Its mode of action is to reduce the coagulation cascade (and thus possibly the thrombogenic risk) after the deployment of a stent. Other coatings, such as phosphorylcoline and silicon-carbide have been used in order to reduce platelet activation and interaction, thus possibly controlling their adhesion to the stent struts during the acute phase of stent re-endothelization. Passive coverage has been also shown to be useful. Indeed, covered stents have been created, in which a PTFE layer was put between two stents (Jostent graft, Jomed) or one stent was covered by a inner and an outer layer of PTFE (Symbiot, Boston Scientific) In both cases the coverage of the lesion is complete (i.e. the stent/artery ratio is 100%), thus the PTFE layer may entrap all the material possibly protruding from the stent struts. Drugs Optimized drug delivery stents require a combination of refined metal stent designs and drug delivery
6 94 Giuseppe Sangiorgi, Gloria Melzi, Pierfrancesco Agostoni, et al. technology. The combination of highly refined metallic-stent designs and polymer materials has been the standard approach in several drug-eluting stent (DES) initiatives. Stent-based drug delivery has been accomplished by 3 distinct mechanisms: - bio-absorbable polymeric stents can be loaded with a drug that is eluted slowly over time; - metal stents can have a drug bound to their surface or embedded within macroscopic fenestrations or microscopic nano-pores, thus providing more rapid drug delivery; - metal stents coated with an outer layer of polymer (bio-absorbable or non-bio-absorbable) can be drug-loaded, thus providing more controlled and sustained drug delivery, which might allow more effective drug-tissue interactions. Recent experimental data [5] suggest that stentstrut configuration directly determines the pattern and degree of drug delivery achieved by the stent. After deployment of even highly soluble and rapidly diffusing drugs, homogeneous drug delivery throughout the vessel with uniform concentration at various depths of the vessel wall is not achieved. Therefore stent designs that maintain regular strut spacing despite expansion in various anatomical circumstances will provide the most regular and predictable drug delivery. For drugs with wide toxic-to-therapeutic window, such as members of the sirolimus family, regularity of strut spacing might be less important and adequate doses can be applied to the stent surface so that, despite broad variability in the location of delivery, an adequate dose is uniformly released. On the other hand, drugs with narrower toxic-to-therapeutic ratios, such as paclitaxel, might suffer from inadequate dosing at sites where stent struts lie far apart and possibly from over-therapeutic or toxic dosing at sites where stent struts bunch together owing to vessel curvature or asymmetric expansion. Among the different DES currently approved in USA and Europe, the Cypher (sirolimuseluting) stent from CordisCorporation, Johnson & Johnson (Miami Lakes, FL, USA) and the Taxus (paclitaxel-eluting) stent from Boston Scientific (Natick, MA, USA) have a relatively similar platform: both have sequential ring design with inert and non-erodible polymeric coatings. The main difference between the two stents is the drug: sirolimus, along with its analogues, was primarily developed to suppress transplant rejection, whereas paclitaxel is a widely used cancer chemotherapeutic agent, whose window between efficacy and toxicity may be narrower in comparison to sirolimus. Several other next-generation DES are currently under investigation. The Endeavor stent from Medtronic (Medtronic Inc., Minneapolis, MN, USA), which uses the non-erodible polymer phosphorylcholine to release the sirolimus analogue ABT-578, will be the first DES to use a non-stainless steel alloy as the foundation for a polymer-coated DES, i.e. the thin-strut cobalt-chromium alloy (Driver). A novel approach is constituted by the XIENCE V stent from Abbott, Illinois, USA, which uses a highly deliverable metallic stent platform and a bio-stable acrylic coating for drug encapsulation and release, with the potential benefit of the restriction to the abluminal (outer) surface of each strut (so that drug and polymer are not exposed to the flowing blood in the arterial lumen) and the eventual degradation of the polymer to carbon dioxide and water, which are free of any toxic byproducts. Thus, the stent design guarantees the complete elimination of the drug from the stent over a finite period of time without drug retention, potential cause of late adverse events months to years after implantation. IMPACT OF STENT DESIGN ON CLINICAL OUTCOME Lesion-related (vessel diameter and length, ostial or bifurcational position, implantation techniques used, IVUS-guidance), and patient-related variables (diabetes, clinical presentation) may be all considered major determinants of acute, subacute and longterm clinical outcomes. Despite the fact that stent design has been shown to be relevant in influencing neointimal hyperplasia in different animal models, in humans the vascular wall response to different stent design does not seem to markedly influence the clinical outcome. On the other hand, in recent years active drug coating (i.e. with sirolimus or paclitaxel) has emerged as the new principal determinant of the reduction of angiographic restenosis and repeated revascularization in the target lesion. The choice of a particular type of stent design is mainly affected by the specific knowledge the operator has for one device or another, and by the a priori possible performance of that device in a specific lesion. Indeed, as different lesions behave in different ways after stent deployment, each type of lesion may be treated with a different stent. Acute clinical outcome Ideally, when a coronary lesion is treated by means of a percutaneous procedure, the endpoint is to dilate the stenosis and gain a lumen equal to the lumen of the non-diseased portion of the coronary artery, without residual dissections or images of haziness or plaque prolapse inside or around the stent deployed. Thus, this primary objective may be achieved after careful evaluation of the specific lesion characteristics and the stent properties. As interventionalists are progressively treating more challenging lesions in more complex situations, stents constructed for specific situations are always more important [6]. Tortuous lesions require particularly conformable and flexible stents, as the modification of the natural conformation of a vessel may alter flow dynamics and possibly increase the risk of events at follow-up. In case of ostial lesions, stents with strong radial support and good radiological visibility are needed
7 Stents design 95 Table 2 Stent versus stent randomized trials Trial Stent type No. of patients Early ST rate (%) RS rate (%) p-value RENEWAL [11] Lansky et al. [12] Thuesen et al. [13] ASCENT [14] Miketic et al.[16] Kastrati et al. [17] Kastrati et al. [18] Park et al. [19] Unverdorben et al. [20] DISTINCT [21] NIR Wallstent GR-II PS NIR Crossflex ML PS NIR Crown Inflow ML NIR PS PURA-A Gold inflow Steel inflow Gold NIR Steel NIR NIR Tenax DUET BiodivYsio PC * * 0.9 0* * * * 0 0 NA NA 0.6 0* 26 NS* * 17 26* * < 0.001* NS* 0.04* * 0.724* * * * < 0.05* 0.56* NS* * : p-values for comparison between stent types. GR: Gianturco-Roubin stent; ML: Multilink stent; PS: Palmaz-Schatz stent; NIR: NIR Stent; NA: not available; NS: not significant; RS: restenosis. ST: stent thrombosis. in order to avoid the elastic recoil of the artery and to permit a correct positioning (which is extremely important in this type of lesion) of the stent. Bifurcational lesions are approached with different stenting techniques, but in all of them, the possibility to rewire the side branch through the stent struts after stent deployment in the main branch is a major factor in determining a good result. Thus stents with large side openings that permit the passage of a balloon or a second stent should be preferred. Chronic total occlusions constitute a subset of lesions in which good lesion coverage and favorable radial support are important, due to the large amount of plaque present and to the possible formation of a false lumen with re-entry during the reopening of the occlusion with stiff guidewires. Small vessels require stents with good flexibility and very thin strut structure. Furthermore, as smaller vessels are usually situated in the distal coronary circulation, stent trackability (i.e. capacity to reach distal lesions) is of extreme importance in this situation. PTFE coated stents have been assessed for the treatment of degenerated saphenous vein grafts, in which extensive plaque burden associated with an elevated risk of distal embolization may be controlled by these covered self-expandable stents that constrain all the embolic material between the PTFE layer and the vessel wall. Other useful indications for PTFEcovered stents are coronary aneurysm exclusion and coronary perforation. As the stent is a foreign body not recognized by the blood, the most threatening acute complication of a stenting procedure is stent thrombosis. This event has been reduced to < 1% to 2% (with respect to 5-7% in the first trials), due to the introduction of high-pressure deployment of the device [7] and double anti-platelet therapy (aspirin and a thienopyridine) instead of aspirin and an oral anticoagulant [8, 9]. However, it has been demonstrated that there are substantial differences in haemodynamic and wall rheological characteristics of implanted stents of different designs, and the hydrodynamic compatibility of a stent, is now recognized as an important feature of ideal stent design. In this setting, Gurbel et al. [10] recently demonstrated that stent design can also affect the degree of platelet activation; stent thrombosis may thus be higher with coil than tubular stents. Long-term clinical outcome Stent configuration. Numerous randomized trials [11-21] comparing the mid-term clinical and angiographic outcomes of various stent designs have been published (Table 2). The wire mesh stent (like the self-expanding Wallstent, Boston Scientific) and the coil stent (like the Gianturco-Roubin Flex/GR-II Cook, Bloomington, Ind. USA; and the Wiktor
8 96 Giuseppe Sangiorgi, Gloria Melzi, Pierfrancesco Agostoni, et al. Stenosis (%) VISION Multi-Link DUET Fig. 2 Percentage stenosis due to neointimal formation with different Multi-Link family stents characterized by different strut thickness. Virmani R. et al. (personal communication). Medtronic) have been shown to have a high propensity for thrombosis and restenosis, thus leading to a higher major adverse cardiac event rate. These adverse outcomes were mainly determined by the high metal to surface area of the former and by the high degree of elastic recoil (associated with poor radial strength) and tissue prolapse of the latter. In fact, these stents are no longer used by cardiologists for coronary interventions. Other stent designs, such as the tubular stent (like the Palmaz-Schatz, Johnson & Johnson; NIR, Boston Scientific; and Crown, Johnson & Johnson) and the multicellular model (like the Multilink stent, Guidant), have been shown to attain better results than the wire mesh and coil stent. No significant imbalances between tubular and multicellular stents were noted in clinical trials, however multicellular stents, which have the same radial support property as the slotted tube stent type, but with less strut-strut intersections, appear to get the most favorable results. Indeed, in the large ASCENT trial [14], patients who were randomly assigned to receive the Multilink stent showed a trend towards a lower restenosis rate compared with that observed after tubular Palmaz-Schatz stent deployment. In another randomized trial evaluating five different stent models, the Multilink stent was associated with more favorable 6-month angiographic and 1-year clinical outcomes compared with the other four stent designs [17] (Figure 2). Strut thickness. Although the immediate stent performance may be improved by increasing strut thickness (which increases radiovisibility, radial strength and arterial wall support) excessive strut thickness, on the other hand, may impart more vascular injury, trigger more intimal hyperplasia, and engender a higher risk for restenosis than thinner struts. Clinical studies appear to confirm this direct relationship between strut thickness and arterial wall reaction. In the ISAR- STEREO study [18], in which two stent types of similar design with different strut thickness were randomly implanted in 651 patients with lesions in large coronary arteries (> 2.8 mm RVD), the 6-month binary restenosis rate and 1-year target vessel revascularization were higher following treatment with the ACS Multi-Link RX Duet stent (strut thickness of 0.14 mm) than with the ACS RX Multi-Link stent (strut thickness 0.05 mm); both stents were manufactured by the same company (Guidant, Advanced Cardiovascular Systems). Similar conclusions were demonstrated by the subsequent ISAR-STEREO-2 trial [22], in which two stent types with different design and different strut thickness were randomly implanted in 611 patients: the ACS RX Multi-Link stent (Guidant, Advanced Cardiovascular Systems) with thinner struts (strut thickness 0.05mm) elicited less angiographic and clinical restenosis than the thicker-strut (strut thickness of 0.14 mm) stent BX Velocity (Cordis Corporation, Johnson & Johnson). A similar finding for small vessels (< 3.0 mm RVD) stenting was observed in a retrospective analysis by Briguori et al [23] in which strut thickness was observed to be an independent predictor of in-stent restenosis (Figure 3). In an effort to further reduce strut thickness while maintaining adequate radiovisibility and radial strength, novel metallic materials such as cobalt-chromium alloy are being used for the production of stent. Stent coating. Active coating of stents with gold, a highly radiovisible and biocompatible material, has been demonstrated to be inferior to plain stainless steel stents in four randomized trials [19, 24-26]. A higher stent thrombosis and restenosis rate was observed with gold-coated stents compared with bare metal stents in all trials. Coating stents with silicon carbide, a potentially less thrombogenic and more compatible material than stainless steel, also did not result in any improvement in angiographic and clinical outcomes compared with bare metal stents in two recent randomized trials [20, 27]. Other randomized trials showed similar results with phosphorylcholine [21] and heparin-coating [28]. Indeed in all these studies there was no angiographic or clini-
9 Stents design 97 cal benefit compared to bare metal stents. A possible advantage of heparin-coated stents, i.e. the possible utilization of aspirin alone as the sole acute-phase antiplatelet regimen, was addressed in a non-randomized trial [29]. In this study, patients receiving heparin coated stents were treated with aspirin only and the rate of stent thrombosis was 1% favorably comparing with other historical controls. Passive coverage of stents with PTFE has been assessed as an appealing solution in the treatment of degenerated saphenous vein grafts containing a considerable amount of friable athero-thrombotic material. Covered stents may prevent distal embolization (and thus periprocedural myocardial infarctions) by entrapping this material against the vessel wall when deployed. These possible benefits were evident in the first registries [30], but two recent randomized trials did not fulfill the initial promise, showing a higher rate of periprocedural myocardial infarctions, compared to a similar restenosis rate [31]. Drug-elution. The breakthrough appearance of stents eluting anti-proliferative drugs with or without a carrier polymer has recently produced unparalleled results with an overall reduction in restenosis rate of between 70% to 85% and in major adverse cardiac events of about 60% compared with bare metal stents and an overall occurrence of restenosis and target lesion revascularization under 10% [32-42]. Sirolimus was used in several trials in the polymer-based system which provides a controlled drug-release over a 4 week period. In the RAVEL trial [32], in which 238 patients were randomized to receive either sirolimus-eluting stent (SES) or bare metal stent, mid-term neointimal hyerplasia was virtually absent in the SES group and correspondingly 6-month in-segment (in-stent and within 5 mm of the stent margins) restenosis was significantly lower in the SES group when compared to the bare metal stent group (0% versus 26.6% respectively, p < 0.001). In the SIRIUS trial [33] in which more real world patients with more complex lesions than RAVEL trial were enrolled, the superiority of SES was also evident, with a benefit in all patients and lesion subgroups, including those traditionally linked with heightened risk of in-stent restenosis such as small vessel size, long lesions and diabetics. The more recently completed European [34] and Canadian [35] SIRIUS trials and the Italian SES- SMART trial [36] confirmed and extended these findings by clearly demonstrating the efficacy of SES in smaller vessels without an increased risk of stent thrombosis. Paclitaxel has been evaluated in two different delivery systems, either by a carrier polymer or by direct impregnation onto the abluminal stent surface. The former, tested in the TAXUS trials, was shown to offer restenosis rates for paclitaxel-eluting stent (PES) significantly lower than those in the bare metal stent cohort. The results from the TAXUS I [37], TAXUS II [38] and TAXUS IV [39] trials were impressive and appeared comparable with those obtained with SES. Non-polymer-based PES, tested in two dose-finding trials, ELUTES [40] and ASPECT [41] offered promising results but the efficacy trial DELIVER [42] showed a restenosis rate of around 15%, definitely higher than that seen in the SES and TAXUS trials. Possible explanations for this failure come from to the fact that up to 40% of drug is lost during stent delivery and that non-polymer-based paclitaxel release is relatively rapid and complete within days to weeks, leaving the underlying bare metal stent exposed, while the polymer permits a Fig. 3 Restenosis rates in lesions treated with a stent with a strut thickness of < 0.10 mm (thin group; grey bars) and stent with a strut thickness 0.10 mm (thick group; white bars), according to the reference vessel diameter ( 2.50, 2,51 to 2,75 and 2.76 to 2.99 mm). Strut thickness significantly influences in-stent binary restenosis between 2.75 and 3.0 vessel size. Modified from Briguori C. et al. J Am Coll Cardiol 2002;40:403-9.
10 98 Giuseppe Sangiorgi, Gloria Melzi, Pierfrancesco Agostoni, et al. more gradual and controlled elution of the drug. Despite the initial enthusiasm, we now realize that while these agents are extremely effective they are not a panacea for restenosis. Indeed, some problems remain still open with DES: - increased costs for the patients and for the entire healthcare delivery system; - non completely satisfactory results in some subgroups such as bifurcation lesions [43] and diabetics [44]; - long-term safety, mainly in term of the occurrence of late stent thrombosis [45]; - role of basic design and structure of underlying stent platform. The ideal DES design may need to have a large surface area of contact with the vascular wall, minimal inter-filament gaps, robust radial support and symmetrical expansion to ensure uniform drug elution. At the same time, it would need to be slim, flexible and conformable to enable successful deployment in complex lesions. For drugs with a narrow toxic-totherapeutic index (the dosage window between beneficial biological effects i.e. inhibition of smooth muscle cells proliferation and toxic side effects i.e. endothelial and smooth muscle cells death), customized stent platforms may be required. The potential for long-term adverse effects of the synthetic polymers often used as carriers for anti-mitotic drugs is a major concern. Synthetic polymers may induce an enhanced inflammatory reaction and possibly a pro-thrombotic response [46]. Late stent thrombosis, late stent apposition and coronary aneurysm formation are thus real possibilities [47]. CONCLUSIONS Albeit stents are currently considered the gold standard for the treatment of narrowed coronary arteries, there is experimental and clinical evidence to indicate that a stent is not just a stent. Different stent models have different structural properties, with their own inherent advantages. The design, material composition and surface features of the stent, as well as the stent deployment technique, have an evident impact on its acute performance, risk of thrombosis, degree of vascular response and subsequent risk of in-stent restenosis. Tubular or corrugated stents are better than coil or mesh wire stents, in terms of a better acute and mid-term outcome. Stents with thinner struts and lower metal density yield a lower risk of restenosis than those with thicker struts and should be used for high-risk lesions such as those located in small vessels where the risk of restenosis is often magnified. The availability of new, highlybiocompatible and more radiovisible alloys with the same if not superior tensile strength than stainless steel will enable the production of low metal density stents that may further improve the anatomical and clinical outcomes of current stainless steel stents. Furthermore, stents coated with anti-proliferative agents, in particular sirolimus and paclitaxel, have recently opened a new era in interventional cardiology. They produce restenosis rates that are unrivalled by other bare metal stent models. However, several important questions regarding their cost-effectiveness, long-term safety and durability need to be addressed in order to clearly understand their potential impact in our daily practice. Moreover, as also these devices may be unsuccessful, the progressive understanding of the causes of these failures and of their different performance in various anatomical and biochemical settings becomes of pivotal importance. As scientists and companies are developing new types of stents with different anti-proliferative drugs, it is entirely foreseeable that most interventional procedures in the near future will involve DES, containing sirolimus, paclitaxel or even more effective drugs with both anti-mitotic and anti-thrombotic actions, impregnated onto highly-biocompatible carrier vehicles, and mounted onto a stent design with uniform expansion and with programmable, controllable drug-eluting capability. It is also possible that a coaction of different drugs, i.e. paclitaxel eluting stent and oral rapamycine given systemically, may further improve the clinical outcome in term of restenosis. Finally, new stent platforms, such as biodegradable stents or endothelial progenitor cell capturing stents may provide in the near future a more physiological answer to stent implantation, reducing vascular injury and accelerating vessel healing with consequent improving in clinical outcome. Received on 6 February Accepted on 5 March References 1. Serruys PW, de Jaegere P, Kiemeneij F, Macaya C, Rutsch W, Heyndrickx G, Emanuelsson H, Marco J, Legrand V, Materne P, et al. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent study group. N Engl J Med 1994;331: Fischman DL, Leon MB, Baim DS, Schatz RA, Savage MP, Penn I, Detre K, Veltri L, Ricci D, Nobuyoshi M, et al. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. Stent restenosis study investigators. N Engl J Med 1994;331: Wong P, Lau KW, Lim YL, Oesterle SN. Stent placement for non-stress/benestent lesions: a critical review. Catheter Cardiovasc Interv 2000;51: Stoeckel D, Bonsignore C, Duda S. A survey of stent designs. Min Invas Ther & Allied Technol 2002;11: Hwang CW, Wu D, Edelman ER. Physiological transport forces govern drug distribution for stent-based delivery. Circulation 2001;104: Colombo A, Stankovic G, Moses JW. Selection of coronary stents. J Am Coll Cardiol 2002;40: Colombo A, Hall P, Nakamura S, Almagor Y, Maiello
11 Stents design 99 L, Martini G, Gaglione A, Goldberg SL, Tobis JM. Intracoronary stenting without anticoagulation accomplished with intravascular ultrasound guidance. Circulation 1995;91: Schomig A, Neumann FJ, Kastrati A, Schuhlen H, Blasini R, Hadamitzky M, Walter H, Zitzmann-Roth EM, Richardt G, Alt E, Schmitt C, Ulm K. A randomized comparison of antiplatelet and anticoagulant therapy after the placement of coronary-artery stents. N Engl J Med 1996;334: Leon MB, Baim DS, Popma JJ, Gordon PC, Cutlip DE, Ho KK, Giambartolomei A, Diver DJ, Lasorda DM, Williams DO, Pocock SJ, Kuntz RE. A clinical trial comparing three antithrombotic-drug regimens after coronary-artery stenting. Stent anticoagulation restenosis study investigators. N Engl J Med 1998;339: Gurbel PA, Callahan KP, Malinin AI, Serebruany VL, Gillis J. Could stent design affect platelet activation? Results of the platelet activation in stenting (PAST) study. J Invasive Cardiol 2002;14: Nageh T, de Belder AJ, Thomas MR, Williams IL, Wainwright RJ. A randomised trial of endoluminal reconstruction comparing the NIR stent and the Wallstent in angioplasty of long segment coronary disease: results of the RENEWAL Study. Am Heart J 2001;141: Lansky AJ, Roubin GS, O Shaughnessy CD, Moore PB, Dean LS, Raizner AE, Safian RD, Zidar JP, Kerr JL, Popma JJ, Mehran R, Kuntz RE, Leon MB. randomized comparison of GR-II stent and Palmaz-Schatz stent for elective treatment of coronary stenoses. Circulation 2000;102: Thuesen L, Andersen HR, Krusell LR, Botker HE, Jorgensen E, Kelbaek H, Kristensen SD. Randomized comparison of the coil-design Crossflex and the tubular NIR stent. Catheter Cardiovasc Interv 2003;59: Baim DS, Cutlip DE, Midei M, Linnemeier TJ, Schreiber T, Cox D, Kereiakes D, Popma JJ, Robertson L, Prince R, Lansky AJ, Ho KK, Kuntz RE. Final results of a randomized trial comparing the MULTI-LINK stent with the Palmaz-Schatz stent for narrowings in native coronary arteries. Am J Cardiol 2001;87: Baim DS, Cutlip DE, O Shaughnessy CD, Hermiller JB, Kereiakes DJ, Giambartolomei A, Katz S, Lansky AJ, Fitzpatrick M, Popma JJ, Ho KK, Leon MB, Kuntz RE. Final results of a randomized trial comparing the NIR stent to the Palmaz-Schatz stent for narrowings in native coronary arteries. Am J Cardiol 2001;87: Miketic S, Carlsson J, Tebbe U. Randomized comparison of J&J Crown stent versus NIR stent after routine coronary angioplasty. Am Heart J 2001;142:E Kastrati A, Dirschinger J, Boekstegers P, Elezi S, Schuhlen H, Pache J, Steinbeck G, Schmitt C, Ulm K, Neumann FJ, Schomig A. Influence of stent design on 1-year outcome after coronary stent placement: a randomized comparison of five stent types in 1,147 unselected patients. Catheter Cardiovasc Interv 2000;50: Kastrati A, Mehilli J, Dirschinger J, Dotzer F, Schuhlen H, Neumann FJ, Fleckenstein M, Pfafferott C, Seyfarth M, Schomig A. Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR- STEREO) trial. Circulation 2001;103: Park SJ, Lee CW, Hong MK, Kim JJ, Park SW, Tahk SJ, Jang YS, Seung KB, Yang JY, Jeong MH. Comparison of goldcoated NIR stents with uncoated NIR stents in patients with coronary artery disease. Am J Cardiol 2002;89: Unverdorben M, Sippel B, Degenhardt R, Sattler K, Fries R, Abt B, Wagner E, Koehler H, Daemgen G, Scholz M, Ibrahim H, Tews KH, Hennen B, Berthold HK, Vallbracht C. Comparison of a silicon carbide-coated stent versus a noncoated stent in human beings: the Tenax versus Nir Stent Study s long-term outcome. Am Heart J 2003;145: Moses JW BC, Nukta ED, et al. The first clinical trial comparing a coated vs. a non-coated coronary stent: the Biocompatibles BiodivYsio stent in randomized control trial (DISTINCT) (abstr.). Circulation 2000;101:II Pache J, Kastrati A, Mehilli J, Schuhlen H, Dotzer F, Hausleiter J, Fleckenstein M, Neumann FJ, Sattelberger U, Schmitt C, Muller M, Dirschinger J, Schomig A. Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO-2) trial. J Am Coll Cardiol 2003;41: Briguori C, Sarais C, Pagnotta P, Liistro F, Montorfano M, Chieffo A, Sgura F, Corvaja N, Albiero R, Stankovic G, Toutoutzas C, Bonizzoni E, Di Mario C, Colombo A. Instent restenosis in small coronary arteries: impact of strut thickness. J Am Coll Cardiol 2002;40: Kastrati A, Schomig A, Dirschinger J, Mehilli J, von Welser N, Pache J, Schuhlen H, Schilling T, Schmitt C, Neumann FJ. Increased risk of restenosis after placement of gold-coated stents: results of a randomized trial comparing gold-coated with uncoated steel stents in patients with coronary artery disease. Circulation 2000;101: Reifart N, Morice MC, Silber S, Benit E, Hauptmann KE, de Sousa E, Webb J, Kaul U, Chan C, Thuesen L, Guagliumi G, Cobaugh M, Dawkins K. The NUGGET study: NIR ultra gold-gilded equivalency trial. Catheter Cardiovasc Interv 2004;62: Silber S. Final results of the NIRTOP trial (Personal Comunication). Euro percutaneous course of revascularization (PCR) symposium. Paris, France: May 14-18, Tanajura LF, Abizaid AA, Feres F, Pinto I, Mattos L, Staico R. Randomized intravascular ultrasound comparison between patients that underwent amorphous hydrogenated Silicon- Carbide coated stent deployment versus uncoated stents (abstr.). J Am Coll Cardiol 2003;41:58A. 28. Wohrle J, Al-Khayer E, Grotzinger U, Schindler C, Kochs M, Hombach V, Hoher M. Comparison of the heparin coated vs the uncoated Jostent-no influence on restenosis or clinical outcome. Eur Heart J 2001;22: Mehran R, Aymong ED, Ashby DT, Fischell T, Whitworth H, Jr., Siegel R, Thomas W, Wong SC, Narasimaiah R, Lansky AJ, Leon MB. Safety of an aspirin-alone regimen after intracoronary stenting with a heparin-coated stent: final results of the HOPE (HEPACOAT and an antithrombotic regimen of aspirin alone) study. Circulation 2003;108: Briguori C, De Gregorio J, Nishida T, Adamian M, Albiero R, Tucci G, Di Mario C, Colombo A. polytetrafluoroethylenecovered stent for the treatment of narrowings in aorticocoronary saphenous vein grafts. Am J Cardiol 2000;86: Schachinger V, Hamm CW, Munzel T, Haude M, Baldus S, Grube E, Bonzel T, Konorza T, Koster R, Arnold R, Haase J, Probst P, vom Dahl J, Neumann FJ, Mudra H, Hennen B, Thiele L, Zeiher AM. A randomized trial of polytetrafluoroethylene-membrane-covered stents compared with conventional stents in aortocoronary saphenous vein grafts. J Am Coll Cardiol 2003;42: Morice MC, Serruys PW, Sousa JE, Fajadet J, Ban Hayashi E, Perin M, Colombo A, Schuler G, Barragan P, Guagliumi G, Molnar F, Falotico R. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med 2002;346: Moses JW, Leon MB, Popma JJ, Fitzgerald PJ, Holmes DR, O Shaughnessy C, Caputo RP, Kereiakes DJ, Williams DO, Teirstein PS, Jaeger JL, Kuntz RE. Sirolimus-eluting stents
12 100 Giuseppe Sangiorgi, Gloria Melzi, Pierfrancesco Agostoni, et al. versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med 2003;349: Schofer J, Schluter M, Gershlick AH, Wijns W, Garcia E, Schampaert E, Breithardt G. Sirolimus-eluting stents for treatment of patients with long atherosclerotic lesions in small coronary arteries: double-blind, randomised controlled trial (E-SIRIUS). Lancet 2003;362: Schampaert E, Cohen EA, Schluter M, Reeves F, Traboulsi M, Title LM, Kuntz RE, Popma JJ. The Canadian study of the sirolimus-eluting stent in the treatment of patients with long de novo lesions in small native coronary arteries (C- SIRIUS). J Am Coll Cardiol 2004;43: Ardissino D, Cavallini C, Bramucci E, Indolfi C, Marzocchi A, Manari A, Angeloni G, Carosio G, Bonizzoni E, Colusso S, Repetto M, Merlini PA. Sirolimus-eluting vs uncoated stents for prevention of restenosis in small coronary arteries: a randomized trial. Jama 2004;292: Grube E, Silber S, Hauptmann KE, Mueller R, Buellesfeld L, Gerckens U, Russell ME. TAXUS I: six- and twelve-month results from a randomized, double-blind trial on a slow-release paclitaxel-eluting stent for de novo coronary lesions. Circulation 2003;107: Colombo A, Drzewiecki J, Banning A, Grube E, Hauptmann K, Silber S, Dudek D, Fort S, Schiele F, Zmudka K, Guagliumi G, Russell ME.. Randomized study to assess the effectiveness of slow- and moderate-release polymerbased paclitaxel-eluting stents for coronary artery lesions. Circulation 2003;108: Stone GW, Ellis SG, Cox DA, Hermiller J, O Shaughnessy C, Mann JT, Turco M, Caputo R, Bergin P, Greenberg J, Popma JJ, Russell ME. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N Engl J Med 2004;350: Gershlick A, De Scheerder I, Chevalier B, Stephens-Lloyd A, Camenzind E, Vrints C, Reifart N, Missault L, Goy JJ, Brinker JA, Raizner AE, Urban P, Heldman AW. Inhibition of restenosis with a paclitaxel-eluting, polymer-free coronary stent: the European evaluation of paclitaxel eluting stent (ELUTES) trial. Circulation 2004;109: Park SJ, Shim WH, Ho DS, Raizner AE, Park SW, Hong MK, Lee CW, Choi D, Jang Y, Lam R, Weissman NJ, Mintz GS. A paclitaxel-eluting stent for the prevention of coronary restenosis. N Engl J Med 2003;348: Lansky AJ, Costa RA, Mintz GS, Tsuchiya Y, Midei M, Cox DA, O Shaughnessy C, Applegate RA, Cannon LA, Mooney M, Farah A, Tannenbaum MA, Yakubov S, Kereiakes DJ, Wong SC, Kaplan B, Cristea E, Stone GW, Leon MB, Knopf WD, O Neill WW. Non-polymer-based paclitaxel-coated coronary stents for the treatment of patients with de novo coronary lesions: angiographic follow-up of the DELIVER clinical trial. Circulation 2004; 109: Colombo A, Moses JW, Morice MC, Ludwig J, Holmes DR, Jr., Spanos V, Louvard Y, Desmedt B, Di Mario C, Leon MB. Randomized study to evaluate sirolimus-eluting stents implanted at coronary bifurcation lesions. Circulation 2004;109: Moussa I, Leon MB, Baim DS, O Neill WW, Popma JJ, Buchbinder M, Midwall J, Simonton CA, Keim E, Wang P, Kuntz RE, Moses JW. Impact of sirolimus-eluting stents on outcome in diabetic patients: a SIRIUS (SIRolImUS-coated Bx Velocity balloon-expandable stent in the treatment of patients with de novo coronary artery lesions) substudy. Circulation 2004;109: McFadden EP, Stabile E, Regar E, Cheneau E, Ong AT, Kinnaird T, Suddath WO, Weissman NJ, Torguson R, Kent KM, Pichard AD, Satler LF, Waksman R, Serruys PW. Late thrombosis in drug-eluting coronary stents after discontinuation of antiplatelet therapy. Lancet 2004;364: Murphy JG, Schwartz RS, Edwards WD, Camrud AR, Vlietstra RE, Holmes DR Jr. Percutaneous polymeric stents in porcine coronary arteries. Initial experience with polyethylene terephthalate stents. Circulation 1992;86: Lau KW, Hung JS, Sigwart U. The current status of stent placement in small coronary arteries < 3.0 mm in diameter. J Invasive Cardiol 2004;8:411-6.
A Survey of Stent Designs. Stoeckel, Bonsignore, Duda. Min Invas Ther & Allied Technol 11(4) pp. 137 147
We are Nitinol. ASurveyofStentDesigns Stoeckel,Bonsignore,Duda MinInvasTher&AlliedTechnol 11(4) pp.137 147 2002 www.nitinol.com 47533 Westinghouse Drive Fremont, California 94539 t 510.683.2000 f 510.683.2001
Clinical Research Intracoronary Stenting with Crushing in Coronary Artery Bifurcation Lesions: Initial Results and Medium-Term Follow Up
Hellenic J Cardiol 45: 379-383, 2004 Clinical Research Intracoronary Stenting with Crushing in Coronary Artery Bifurcation Lesions: Initial Results and Medium-Term Follow Up PETROS S. DARDAS, DIMITRIS
A survey of stent designs
Min Invas Ther & Allied Technol 2002: 11(4) 137 147 A survey of stent designs D. Stoeckel 1, C. Bonsignore 1 and S. Duda 2 1 Nitinol Devices and Components, Fremont, CA USA 2 Dept of Diagnostic Radiology,
TECHNOLOGICAL INNOVATION FOR CORONARY STENTS Francesco Migliavacca
LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it TECHNOLOGICAL INNOVATION FOR CORONARY STENTS Francesco Migliavacca Erice, 1 maggio 2015 International School of Cardiac Surgery Introduction
MEDICAL POLICY No. 91580-R1 DRUG-ELUTING STENTS FOR ISCHEMIC HEART DISEASE
DRUG-ELUTING STENTS FOR ISCHEMIC HEART DISEASE Effective Date: October 1, 2015 Review Dates: 10/11, 10/12, 10/13, 8/14, 8/15 Date Of Origin: October 12, 2011 Status: Current Summary of Changes Clarifications:
Yukon Choice DES + Translumina - The polymer-free DES solution. www.translumina.de. Coronary Stent System for Drug Application
Coronary Stent System for Drug Application Yukon Choice DES + new catheter polymer-free DES Translumina - The polymer-free DES solution www.translumina.de A First Class Stent The YUKON Choice DES+ is the
Different Properties of Coronary Stents György Ring 1, Eszter Bognár 1, Zsófia Bálint-Pataki 1, János Dobránszky 2
Different Properties of Coronary Stents György Ring 1, Eszter Bognár 1, Zsófia Bálint-Pataki 1, János Dobránszky 2 1 Budapest University of Technology and Economics, Department of Materials Science and
Advances in Stent Technology
Nurse/Technologist Symposium Advances in Stent Technology Antonietta Tolentino, CCRN, MSN, ANP-C Adult Nurse Practitioner, Cardiac Catheterization Laboratory The Zena and Michael A. Weiner Cardiovascular
Orsiro Hybrid Drug Eluting Stent Industry's first hybrid DES
Vascular Intervention Drug Eluting Stents Orsiro Orsiro Hybrid Drug Eluting Stent Industry's first hybrid DES Orsiro An ideal combination of passive and active components The Orsiro Hybrid Drug Eluting
Review Article CORONARY STENTS - PAST, PRESENT AND FUTURE.
Review Article CORONARY STENTS - PAST, PRESENT AND FUTURE. VV JAIN*, S YELWATKAR**, OP GUPTA*** Introduction The advent of period of angioplasty has revolutionized the practice of interventional cardiology.
The Minvasys Amazonia Pax & Nile Pax Polymer Free Paclitaxel Eluting Stent Program
The Minvasys Amazonia Pax & Nile Pax Polymer Free Paclitaxel Eluting Stent Program Jean Fajadet, MD, FESC Clinique Pasteur - Toulouse - France Disclosure statement Nothing to disclose Dedicated Delivery
USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT
USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT V. Astier, S. Bidal, V. Garitey, F. Mouret Abstract : A stent is a wire metal mesh tube placed into an artery or blood vessel to hold the structure
Duration of Dual Antiplatelet Therapy After Coronary Stenting
Duration of Dual Antiplatelet Therapy After Coronary Stenting C. DEAN KATSAMAKIS, DO, FACC, FSCAI INTERVENTIONAL CARDIOLOGIST ADVOCATE LUTHERAN GENERAL HOSPITAL INTRODUCTION Coronary artery stents are
Overview of Newer Stent Devices for Aneurysm Treatment
Overview of Newer Stent Devices for Aneurysm Treatment Randall C. Edgell, M.D. Associate Professor Vascular and Interventional Neurology Saint Louis University Disclosure Outcome adjudication for THERAPY,
Coronary Stents. What is a Coronary Artery stent?
What is a Coronary Artery stent? Coronary Stents A coronary stent is stainless tube with slots. It is mounted on a balloon catheter in a "crimped" or collapsed state. When the balloon of is inflated, the
Drug-Eluting Balloons. Klaus Bonaventura Department of Cardiology and Angiology Heart Thorax Vascular Center, Klinikum Ernst von Bergmann, Potsdam
Drug-Eluting Balloons Klaus Bonaventura Department of Cardiology and Angiology Heart Thorax Vascular Center, Klinikum Ernst von Bergmann, Potsdam Potential conflicts of interest Speaker s name: Klaus Bonaventura
Deployment of a Self-expanding Stent in an Artery
Deployment of a Self-expanding Stent in an Artery Santosh Prabhu*, Chris Feezor*, Andy Denison*, Nuno Rebelo** and Mouhsine Serrar** * Guidant Corporation 3200 Lakeside Drive Santa Clara, CA 95054 and
Status of Drug-Eluting Coronary Stents
Status of Drug-Eluting Coronary Stents EVOLUTION OF PERCUTANEOUS CORONARY INTERVENTIONAL TECHNOLOGIES The introduction of percutaneous transluminal coronary angioplasty (PTCA) in the late1970s provided
Understanding Coronary Artery Disease, Cardiac Catheterization, and Treatment Options. A Guide for Patients
Understanding Coronary Artery Disease, Cardiac Catheterization, and Treatment Options A Guide for Patients Coronary Artery Disease If you or a member of your family has been diagnosed with coronary artery
The Bioresorbable Vascular Stent Dr Albert Ko
The Bioresorbable Vascular Stent Dr Albert Ko Dr Albert Ko MB BS, FRACP, FCSANZ Interventional/General Cardiologist Ascot Cardiology Symposium 2013 Treatment Goals for Coronary Artery Disease Relieve of
GUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA - A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES -
EUROPEAN COMMISSION ENTERPRISE AND INDUSTRY DIRECTORATE-GENERAL Consumer goods Cosmetics and Medical Devices MEDDEV 2.7.1 Appendix 1 December 2008 GUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA
Antiplatelet and anticoagulation treatment of patients undergoing carotid and peripheral artery angioplasty
Round Table: Antithrombotic therapy beyond ACS Antiplatelet and anticoagulation treatment of patients undergoing carotid and peripheral artery angioplasty M. Matsagkas, MD, PhD, EBSQ-Vasc Associate Professor
Coronary Bifurcation Treatment: Update from the European Bifurcation Club. Remo Albiero, MD Ist. Clinico S. Rocco Brescia (Italy)
Coronary Bifurcation Treatment: Update from the European Bifurcation Club Remo Albiero, MD Ist. Clinico S. Rocco Brescia (Italy) Disclosure Statement of Financial Interest Within the past 12 months, I
Cilostazol versus Clopidogrel after Coronary Stenting
Cilostazol versus Clopidogrel after Coronary Stenting Seong-Wook Park, MD, PhD, FACC Division of Cardiology, Asan Medical Center University of Ulsan College of Medicine Seoul, Korea AMC, 2004 Background
Stent Design for a Percutaneous Heart Valve
Stent Design for a Percutaneous Heart Valve A Major Qualifying Project submitted to the Faculty of WORCESTER POLYTECHNIC INSTITUTE in partial fulfillment of the requirements for the Degree of Bachelor
Intracoronary Stenting and. Robert A. Byrne, Julinda Mehilli, Salvatore Cassese, Franz-Josef Neumann, Susanne Pinieck, Tomohisa Tada,
Prospective, Randomized Trial of Paclitaxel-Eluting Balloon versus Paclitaxel-Eluting Stent versus Balloon Angioplasty for Treatment of Coronary Restenosis in Limus- Eluting Stents Intracoronary Stenting
Ostial LAD: Single stent approach is the best. Antonio A. Pocoví, MD, FSCAI, MTSAC, Advisory Council Member, CACI
Ostial LAD: Single stent approach is the best Antonio A. Pocoví, MD, FSCAI, MTSAC, Advisory Council Member, CACI Chair, Interventional Cardiology Sanatorio San Lucas Instituto Alexander Fleming Buenos
LEADERS: 5-Year Follow-up
LEADERS: -Year Follow-up from a Prospective, Randomized Trial of Biolimus A9-eluting Stents with a Biodegradable Polymer vs. Sirolimus-eluting Stents with a Durable Polymer : Final Report of the LEADERS
A Post-market Study to Assess the STENTYS Self-exPanding COronary Stent In AcuTe myocardial InfarctiON in Real Life APPOSITION III
A Post-market Study to Assess the STENTYS Self-exPanding COronary Stent In AcuTe myocardial InfarctiON in Real Life APPOSITION III Gilles Montalescot, MD, PhD Pitié-Salpêtrière Hospital, Paris, France
Antonio Colombo MD on behalf of the SECURITY Investigators
Second Generation Drug-Eluting Stents Implantation Followed by Six Versus Twelve-Month - Dual Antiplatelet Therapy - The SECURITY Randomized Clinical Trial Antonio Colombo MD on behalf of the SECURITY
Majestic Trial 12 Month Results
Majestic Trial 12 Month Results S.Müller-Hülsbeck, MD, EBIR, FCIRSE, FICA ACADEMIC HOSPITALS Flensburg of Kiel University Ev.-Luth. Diakonissenanstalt zu Flensburg Knuthstraße 1, 24939 FLENSBURG Dept.
TAXUS Express 2. Paclitaxel-Eluting Coronary Stent System. Patient Information Guide
TAXUS Express 2 Paclitaxel-Eluting Coronary Stent System Patient Information Guide 1 Table of Contents Coronary Artery Disease...2 Who Is at Risk?...3 Diagnosis of Coronary Artery Disease...3 Treatment
Coronary-Artery Stents
review article Drug Therapy Coronary-Artery Stents Patrick W. Serruys, M.D., Ph.D., Michael J.B. Kutryk, M.D., Ph.D., and Andrew T.L. Ong, M.B., B.S. T he use of percutaneously introduced prosthetic devices
Parameterization and Optimization of balloon expandable Stent
Parameterization and Optimization of balloon expandable Stent Karthik Shankaran, and Senthil Karrupaswamy Detroit Engineered Products (DEP) Inc Abstract: Vascular stents are deployed in the blocked arteries
Rotational Atherectomy for the Treatment of In-Stent Restenosis
Home SVCC Area: English - Español - Português Rotational Atherectomy for the Treatment of In-Stent Restenosis Steven L. Goldberg, MD Cardiac Catheterization Laboratory, University of Washington Medical
StentViz Enhanced Stent Visualization
GE Healthcare StentViz Enhanced Stent Visualization Dr. Morice, Dr. Lefèvre, Dr. Hovasse, Dr. Chevalier, Dr. Louvard Institut Cardiovasculaire Paris Sud, Massy, France Assessing the deployment of stents
Drug-Eluting Coronary Stents. Paul Montero PGY-II University of Colorado Health Sciences Center Surgical Grand Rounds Resident Debate
Drug-Eluting Coronary Stents Paul Montero PGY-II University of Colorado Health Sciences Center Surgical Grand Rounds Resident Debate Outline Coronary Artery Disease Evolution of Percutaneous Coronary Intervention
CORONARY STENTS: CLINICAL EXPERIENCE AND COST -EFFECTIVENESS
Canadian Coordinating Office for Health Technology Assessment CORONARY STENTS: CLINICAL EXPERIENCE AND COST -EFFECTIVENESS CCOHTA Report 1997: 1E Additional copies of Coronary Stents: Clinical Experience
La facilitazione alla PCI con statine
La facilitazione alla PCI con statine Bertinoro, 16 aprile 2010 Aldo Miola, M.D., Ph.D. Medical Manager Primary Care Pfizer Italy 1 Are statins beneficial in patients undergoing PCI? 2 3 Statin therapy
Angioplasty and Stent Education Guide
Angioplasty and Stent Education Guide Table of Contents Treating coronary artery disease...2 What is coronary artery disease...3 Coronary artery disease treatment options...4 What are coronary artery
2015 European Coronary Stent New Product Innovation Award
2015 European Coronary Stent New Product Innovation Award 2015 Contents Background and Company Performance... 3 Industry Challenges... 3 New Product Attributes and Customer Impact... 3 Conclusion... 4
Is Stenting or Coronary Artery By-pass Grafting the Better Treatment for This Patient?
Is Stenting or Coronary Artery By-pass Grafting the Better Treatment for This Patient? --- NIRS-IVUS TVC Imaging Adds Additional Information for the Heart Team Dr. Luis Tami Memorial Regional Hospital
Experience of Direct Coronary Stenting at National Institute of Cardiovascular Diseases
Experience of Direct Coronary Stenting at National Institute of Cardiovascular Diseases T. Masood,T. Sagheer,D. Jan,N. Qamar,A.M.A. Faruqui ( National Institute of Cardiovascular Diseases (NICVD), Karachi.
PATIENT INFORMATION BOOKLET
PATIENT INFORMATION BOOKLET Wingspan Stent System with Gateway PTA Balloon Catheter TABLE OF CONTENTS Definitions... 2 What is the Purpose of This Booklet?... 3 What is an Intracranial Lesion?... 3 Who
Tracheal Stent System
nt Tracheal Stent System Available sizes: 08mm x 20mm NEW 08mm x 30mm 08mmx 50 mm 08mm x 60mm 08mm x 70mm 08mm x 80mm 08mm x 90mm 10mm x 20mm NEW 10 mm x 30 mm 10 mm x 50 mm 10 mm x 60 mm 10 mm x 70 mm
ANESTHESIA FOR PATIENTS WITH CORONARY STENTS FOR NON CARDIAC SURGERY. Dr. Mahesh Vakamudi. Professor and Head
ANESTHESIA FOR PATIENTS WITH CORONARY STENTS FOR NON CARDIAC SURGERY Dr. Mahesh Vakamudi Professor and Head Department of Anesthesiology, Critical Care and Pain Medicine Sri Ramachandra University INTRODUCTION
Review of Intracoronary Radiation for In-Stent Restenosis
Review of Intracoronary Radiation for In-Stent Restenosis Robert Lew, MBBS, FRACP, Andrew Ajani, MD, Ron Waksman, MD, FACC Vascular Brachytherapy using beta and gamma radiation is the standard care for
A Patient s Guide to Primary and Secondary Prevention of Cardiovascular Disease Using Blood-Thinning (Anticoagulant) Drugs
A Patient s Guide to Primary and Secondary Prevention of PATIENT EDUCATION GUIDE What Is Cardiovascular Disease? Cardiovascular disease (CVD) is a broad term that covers any disease of the heart and circulatory
Technology Insight Report
Technology Insight Report Coronary Stents Cardiology has come a long way over the decades with huge advances having been made in areas such as invasive surgery and interventional cardiology. As a result
Early healing after treatment of coronary lesions by everolimus, or biolimus eluting bioresorbable polymer stents
Early healing after treatment of coronary lesions by everolimus, or biolimus eluting bioresorbable polymer stents One-month results of the study Ida Riise Balleby 1, Trine Krejberg Ørhøj 1, Christian Juhl
High incidence of thrombus formation at 18 months after paclitaxel-eluting stent implantation: Angioscopic comparison with sirolimus-eluting stent
High incidence of thrombus formation at 18 months after paclitaxel-eluting stent implantation: Angioscopic comparison with sirolimus-eluting stent Masahiko Hara, MD, Masami Nishino, MD, Masayuki Taniike,
California Health and Safety Code, Section 1256.01
California Health and Safety Code, Section 1256.01 1256.01. (a) The Elective Percutaneous Coronary Intervention (PCI) Pilot Program is hereby established in the department. The purpose of the pilot program
CORONARY ARTERY BYPASS GRAFTS, STENTS, AND EXTRACORONARY CARDIAC DZ. Charles White MD
CORONARY ARTERY BYPASS GRAFTS, STENTS, AND EXTRACORONARY CARDIAC DZ Charles White MD Director of Thoracic Imaging Department of Radiology University of Maryland CORONARY ARTERY BYPASS GRAFTS First performed
Clinical Programs. Medtronic Coronary Stent Systems. driver BMS
Clinical Programs Medtronic Coronary Stent Systems Endeavor DES driver BMS July 2010 Contents Overview of Clinical Programs... 2 Drug-Eluting Stents ENDEAVOR I*... 4 ENDEAVOR II*... 6 ENDEAVOR II Continued
SAMPLE. Asia-Pacific Interventional Cardiology Procedures Outlook to 2020. Reference Code: GDMECR0061PDB. Publication Date: May 2014
Asia-Pacific Interventional Cardiology Procedures Outlook to 2020 Reference Code: GDMECR0061PDB Publication Date: May 2014 Page 1 1 Table of Contents 1 Table of Contents... 2 1.1 List of Tables... 4 1.2
Modelling of Stent Deployment and Deformation in Diseased Arteries by Considering Vessel Anisotropy
Modelling of Stent Deployment and Deformation in Diseased Arteries by A. Schiavone, L.G. Zhao [email protected] Abstract In this paper, finite element simulation of stent deployment was carried out using
COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) DRAFT
European Medicines Agency London, 22 March 2007 Doc. Ref. EMEA/CHMP/EWP/110540/2007 COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) DRAFT GUIDELINE ON THE DEVELOPMENT OF MEDICINAL SUBSTANCES CONTAINED
Carotid Artery Stenting? Comparison of the outcome of octogenarian patients according to the stent type. Data from a multicentre registry
Closed- or Open-cell stents in Carotid Artery Stenting? Comparison of the outcome of octogenarian patients according to the stent type. Data from a multicentre registry Nikas D 1, Reimers B 2, Iakovou
JUL 2 2008. Ms. Kendra Basler Regulatory Affairs Associate Abbott Vascular Cardiac Therapies 3200 Lakeside Drive Santa Clara, CA 95054-2807
DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service JUL 2 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Kendra Basler Regulatory Affairs Associate Abbott Vascular
Evolu'on of Balloon Angioplasty From POBA to newer technologies NCVH May 26, 2015
Evolu'on of Balloon Angioplasty From POBA to newer technologies NCVH May 26, 2015 Cezar Staniloae, MD Heart and Vascular Ins7tute New York University Medical Center Overview 1. Physics of balloon angioplasty
PRECOMBAT Trial. Seung-Whan Lee, MD, PhD On behalf of the PRECOMBAT Investigators
Premier of Randomized Comparison of Bypass Surgery versus Angioplasty Using Sirolimus-Eluting Stent in Patients with Left Main Coronary Artery Disease PRECOMBAT Trial Seung-Whan Lee, MD, PhD On behalf
SCORM. For more patient education, please visit www.cypherusa.com
Attach Label Understanding Coronary Artery Disease, Cardiac Catheterization, and Treatment Options Stent Implant Card CYPHER Sirolimus-eluting Coronary Stent SCORM P.O. Box 025700 Miami, FL 33102-5700,
The Pantera Lux Paclitaxel DEB Device Description and Clinical Studies. Christoph Hehrlein, University Clinic Freiburg i.br.
The Pantera Lux Paclitaxel DEB Device Description and Clinical Studies Christoph Hehrlein, University Clinic Freiburg i.br. Germany Disclosure Statement of Financial Interest Within the past 12 months,
Coronary in-stent restenosis predictors, treatment and prevention
European Heart Journal (2000) 21, 1739 1749 doi:10.1053/euhj.2000.2153, available online at http://www.idealibrary.com on Review Coronary in-stent restenosis predictors, treatment and prevention R. Hoffmann
The largest clinical study of Bayer's Xarelto (rivaroxaban) Wednesday, 14 November 2012 07:38
Bayer HealthCare has announced the initiation of the COMPASS study, the largest clinical study of its oral anticoagulant Xarelto (rivaroxaban) to date, investigating the prevention of major adverse cardiac
A Patient Information Guide for the CYPHER Sirolimus-eluting Coronary Stent
bringing evidence to life Understanding Coronary Artery Disease, Cardiac Catheterization, and Treatment Options A Patient Information Guide for the CYPHER Sirolimus-eluting Coronary Stent Patient s Name
Presenter: Marco Valgimigli, MD PhD, FESC Erasmus MC, Thoraxcenter Rotterdam The Netherlands
Comparing zotarolimus-eluting and bare-metal stent efficacy in selected high bleeding risk patients treated with a short dual antiplatelet therapy duration. A pre-specified analysis from the The Zotarolimuseluting
Mechanical Properties of Nitinol Stents and Stent-grafts. Comparison of 6 mm Diameter Devices
Mechanical Properties of Nitinol Stents and Stent-grafts Comparison of 6 mm Diameter Devices P E R F O R M A N C E b y d e s i g n Purpose The association of target vessel restenosis with stent fractures
INVESTIGATION OF METALLIC SURFACE AREA OF CORONARY STENTS
INVESTIGATION OF METALLIC SURFACE AREA OF CORONARY STENTS Dóra Károly 1, Miksa Kovács 1, Andrew Attila Terdik 1, Eszter Bognár 1,2 1Department of Materials Science and Engineering, Faculty of Mechanical
Cardiovascular disease has become a dominant cause of
Abstract Immediate and Six-month Clinical Outcome of Percutaneous Coronary Intervention in a Tertiary Hospital in the Sultanate of Oman Panduranga Prashanth, Mohamed Mukhaini, Abdulla A. Riyami, Kadhim
TAXUS Express2 Paclitaxel-Eluting Coronary Stent System TAXUS Liberté Paclitaxel-Eluting Coronary Stent System. A Patient s Guide
TAXUS Express2 Paclitaxel-Eluting Coronary Stent System TAXUS Liberté Paclitaxel-Eluting Coronary Stent System A Patient s Guide Table of Contents Coronary Artery Disease... 2 Who Is at Risk?... 3 Diagnosis
Understanding your Renal Stent Procedure. A patient Guide (COVER PAGE) TABLE OF CONTENTS (inside front page)
Understanding your Renal Stent Procedure. A patient Guide (COVER PAGE) TABLE OF CONTENTS (inside front page) The Kidney and the Renal Arteries... 1 Renal Artery Disease... 2 Diagnosis of Renal.Artery Disease...
TURQUOISE PTCA BALLOON DILATATION CATHETER
TURQUOISE PTCA BALLOON DILATATION CATHETER TURQUOISE PTCA BallOOn Dilatation CATheter The right BAlance TO meet the clinical need TURQUOISE PTCA s advanced catheter technology offers the right balance
WallFlex Biliary RX Stent. Fully, Partially and Uncovered Self-Expanding Metal Stents
WallFlex Biliary RX Stent Fully, Partially and Uncovered Self-Expanding Metal Stents WallFlex Biliary RX Stent Fully, Partially and Uncovered Self-Expanding Metal Stents The WallFlex Biliary RX Stent is
Is There A LIfe for DES after discontinuation of Clopidogrel
Chicago 2014 Is There A LIfe for DES after discontinuation of Clopidogrel Six-month versus 24-month dual antiplatelet therapy after implantation of drug eluting stents in patients non-resistant to aspirin:
When Procedural Support really matters. Navien TM A+ Intracranial Support Catheter
When Procedural Support really matters Navien TM A+ Intracranial Support Catheter Coils / Liquid Embolics DURABLE ovalization resistance, Maintains patent lumen in tortuosity for smooth micro catheter
Clinical Study Synopsis
Clinical Study Synopsis This Clinical Study Synopsis is provided for patients and healthcare professionals to increase the transparency of Bayer's clinical research. This document is not intended to replace
Abbott Vascular Everolimus-Eluting Bioresorbable Vascular Scaffold Components
Abbott Vascular Everolimus-Eluting Bioresorbable Vascular Scaffold Components Bioresorbabl e Scaffold Bioresorbabl e Coating Everolimus XIENCE V Delivery System *Except for platinum markers All illustrations
NOVOSTE BETA-CATH SYSTEM
HOSPITAL INPATIENT AND OUTPATIENT BILLING GUIDE FOR THE NOVOSTE BETA-CATH SYSTEM INTRAVASCULAR BRACHYTHERAPY DEVICE This guide is intended solely for use as a tool to help hospital billing staff resolve
Do método diagnóstico ao terapêutico: a História da Hemodinâmica e da Cardiologia Intervencionista, em 15 minutos. Eulógio E Martinez
Do método diagnóstico ao terapêutico: a História da Hemodinâmica e da Cardiologia Intervencionista, em 15 minutos Eulógio E Martinez IVUS vs luminologia Positivo Negativo estenose TCE oclusão CD sístole
How To Make A Vascular Graft
LCL-GRAFT TM - Innovative Large and Small Caliber Vascular Grafts for Coronary Bypass and Peripheral Vascular Surgery. A. Perets 1, M. Li 1, P. Uttayarat 1, P. Pimton 1, A. Wu 2, R. J. Levy 3, R. J. Composto
Christopher M. Wright, MD, MBA Pioneer Cardiovascular Consultants Tempe, Arizona
Christopher M. Wright, MD, MBA Pioneer Cardiovascular Consultants Tempe, Arizona Areas to be covered Historical, current, and future treatments for various cardiovascular disease: Atherosclerosis (Coronary
North American Stainless
North American Stainless Long Products Stainless Steel Grade Sheet 2205 UNS S2205 EN 1.4462 2304 UNS S2304 EN 1.4362 INTRODUCTION Types 2205 and 2304 are duplex stainless steel grades with a microstructure,
DATE: 29 August 2012 CONTEXT AND POLICY ISSUES
TITLE: Dual Antiplatelet Therapy and Enoxaparin or Unfractionated Heparin for patients with ST-elevation Myocardial Infarction: A Review of the Clinical Evidence DATE: 29 August 2012 CONTEXT AND POLICY
Copenhagen University Hospital Rigshospitalet Aarhus University Hospital Skejby Denmark
Long-term outcome after drug-eluting versus bare-metal stent implantation in patients with ST-elevation myocardial infarction 3 year follow-up of the randomised trial Peter Clemmensen, Henning Kelbæk,
