PORTABLE RADIATION SURVEY INSTRUMENTS USE AND CALIBRATION
|
|
|
- Logan Douglas
- 9 years ago
- Views:
Transcription
1 PORTABLE RADIATION SURVEY INSTRUMENTS USE AND CALIBRATION PURPOSE This procedure provides basic criteria and methods for calibration of instruments used for radiation detection, measurements, or surveys. POLICY All instruments used for measuring exposure rates or determining the quantities of radioactivity present in samples or on surfaces (as contamination) are to be calibrated at least once a year. Calibrations are to be performed by individuals who meet the specified qualifications and using sources and procedures that assure compliance with federal and state regulations and license conditions. DEFINITIONS Calibration - the determination and correction of the response of an instrument relative to a series of known radiation values over the range of the instrument. Check Source (Reference Source) - a radioactive source (not necessarily calibrated) used to confirm the satisfactory operation of an instrument. Contamination Detection - the direct detection of radioactivity present on persons, objects, or surfaces; any numerical values obtained by direct detection of contamination must always be regarded as qualitative rather than quantitative. Directional Response - the variation in response of an instrument to a radiation field of constant intensity as a result of changes in orientation of the instrument with respect to the radiation source. Efficiency - a term used to describe (1) the intrinsic response of a specific detector, or (2) the overall response of a measurement system, to a standard source of radiation emissions (see the following specific definitions). Detector Intrinsic Efficiency - the fraction of nuclear particles or photons reaching a detector that interact with the detector in such a way as to deposit detectable energy. Measurement (Overall) Efficiency- the fraction of the particles or photons emitted by a standard source during a fixed time period that are detected by an instrument during the same time interval. The overall measurement efficiency includes the detector intrinsic efficiency, geometry factor, absorption corrections, scatter corrections, etc. (see "Yield" for more details). Point-Source Efficiency - the fraction of the particles or photons emitted by a discrete or point source detected by the survey instrument. A "point source" need not actually be a point, but a source with an active diameter much smaller than either the detector diameter or the source-to-detector distance. Energy Dependence - a change in instrument response with respect to the energy of radiation at a constant exposure or exposure rate. Exposure (Rate) Measurement - the determination of the intensity of a radiation field. Although the term "exposure" technically applies only to measurements of electromagnetic radiation by means of ionization in air, it is used here also in a general sense to refer to any measurement of field intensity, e.g. beta particles or neutrons, regardless of the detection mechanism. Geometry Factor - the fraction of all photons or nuclear particles emitted from a radiation source that would be intercepted by a detector if not scattered or absorbed. For a point source, it is the solid angle subtended by a detector external to the source divided by 4π. Response Check - the verification of proper operation of a survey instrument by observing its response to a check source. Standard Source - a radiation source exhibiting a disintegration, emission or exposure rate certified by or traceable to the U.S. National Institute of Standards and Technology (NIST). RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 1
2 Survey Instrument - a portable instrument used for direct detection and/or measurement of radioactivity or radiation field intensities. Transfer Instrument - an instrument or dosimeter exhibiting high precision which has been calibrated against a standard source. Transfer Standard - a radiation source whose emission or exposure rate has been determined and documented by comparison with one or more standard sources. "Yield", or overall efficiency, for a radiation-measuring instrument depends upon physical characteristics of the detector and of the source, the geometric relationship between the source and the detector, and the medium between the source and the detector, all of which may be dependent upon the energy of the radiation. Y = Y = I = F s = I F s B G F m F w E, where: yield, or "overall efficiency" (counts/disintegration) intensity of radiations of type to be detected, i.e. number of emissions per disintegration of a source nucleus fraction of emissions within the source that actually escape the source (always 1), i.e. 1 - fractional self-absorption within the source F b = fractional contribution from emissions backscattered from the material that supports the source (always 1), i.e. the total radiation directed toward the detector divided by the amount of radiation originally emitted in the direction of the detector G = F m = F w = geometry factor, or fraction of 4-pi solid angle subtended by the effective window of the detector (always 1) fraction of emissions leaving the source that pass through the intervening medium and reach the detector (always 1), i.e. 1 - fractional absorption in the medium fraction of the radiation reaching the window of the detector that passes through the detector to the sensitive volume (always 1), i.e. 1 - absorption in the window E = intrinsic efficiency of detection, i.e. the fraction of the radiation entering the sensitive volume of the detector that produces a measurable pulse (always 1) If a contamination survey instrument is calibrated with a source of the same nuclide as that which is to be detected, and mounted on a similar substrate, etc., most of the components of total counting yield are automatically included in the efficiency determined during calibration. However, it is important to always be aware of the factors that affect total yield when different sources or media are encountered. CONTAMINATION MEASUREMENTS Portable survey instruments are commonly used to detect and measure radioactive contamination on surfaces. The validity and accuracy of a direct measurement of a surface depends on a knowledge of the instrument's response to the contaminant nuclides and on the accessibility of the surface. The detector should be positioned approximately 1 cm from the surface. Measurement efficiency is then used to convert the instrument response to surface contamination density (dpm per 100 cm 2 ). Removable contamination on surfaces is evaluated by wiping areas of cm 2 with a filter paper or other absorbent material. In many situations, measurements of the activity on the wipe medium can be made with adequate sensitivity, accuracy and precision with a portable survey instrument with no need for more quantitative laboratory analyses. For most beta-gamma emitting radionuclides, the recommended limits for removable contamination in unrestricted areas are 200 or 2000 dpm per 100 cm 2.. For alpha emitters it is 20 dpm per 100 cm 2. To improve sensitivity, wipes of 300 cm5 may be used if the wiped object has a surface area greater than 300 cm 2. A draft ANSI standard (1987) states that these limits are for the average over 1 m 2, and that any individual area of 100 cm 2 shall not exceed 3 times the limit. An NRC Regulatory Guide (1.86, 1974) applicable to nuclear reactors states that maximum values of 3 times the average limit apply only to the total contamination (fixed plus removable), implying that it does not apply specifically to removable contamination. Therefore, the activity on a wipe that must be detected with acceptable accuracy is in the range of 20 to 6000 dpm. RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 2
3 In general, a survey meter that is appropriate for making direct measurements of surface contamination is also appropriate for measuring removable activity on a wipe test. The question that must be addressed is: At a specified limit for removable contamination on a wipe, is the expected count rate on the survey instrument sufficiently above background to make an unambiguous decision that the activity is below the limit? In other words, what is the minimum detectable activity (MDA) for the measurement? Energy vs Absorption When radioactivity is to be measured on a filter, one factor to be considered is the fraction of the emitted radiation absorbed within the filter (F ABS ). For any alpha or beta emitter with sufficient energy to be detected by a thin-window GM tube, the absorbed fraction is rarely greater than 0.2; therefore, a conservative estimate of the unabsorbed fraction would be 80%. Therefore, the overall counting efficiency (yield) for measuring activity on a filter area that is no larger than the window of the detector would be 80% of the point-source efficiency, or 0.8 (Eff P ). Background Count Rate and Minimum Detectable Activity The variability of the background for a rate-meter instrument depends on the count rate and on the time constant of the meter. For measuring activity, the longest available time constant should be used and both background and sample readings should be long enough to allow the meter to stabilize. Observation of the meter for a few seconds allows one to estimate both the average and the upper limit of the meter readings. A typical rate meter with a thin-window GM detector might indicate a background rate in the range of 50 to 100 cpm in a low-radiation area. If the meter has a fairly long time constant, close observation should demonstrate that the reading rarely fluctuates to a value of more than 20 cpm above the average. This observation would be expected on the basis of radioactive decay statistics, since the standard deviation of the number of counts obtained in any one minute would be in the range of 7-10 cpm and 20 cpm would represent at least 2 standard deviations. Therefore, for typical GM survey meters, the minimum net response that is clearly above background would be approximately 20 cpm and the minimum detectable activity (MDA) on a filter would then be estimated as: MDA (dpm) = (Min. Net Rate)/(F ABS x Eff P ) = 20/(0.8 x Eff P ) Typical GM Survey Meter Response Typical point-source efficiencies for a thin-window GM survey meter are 4% for C-14, 22% for P-32 and 10% for alphas, for which the estimated MDAs on a filter would be: MDA (C-14) = 20 cpm/(0.8 x 0.04) = 625 dpm 1 RCL (C-14) = 1 nci x 0.04 x 0.8 x 2,220 dpm/nci = 70 cpm net MDA (P-32) = 20 cpm/(0.8 x 0.22) = 114 dpm 1 RCL (P-32) = 0.1 nci x 0.22 x 0.8 x 2,220 dpm/nci = 40 cpm net MDA (Alpha) = 20 cpm/(0.8 x 0.1 x 2,220 dpm/nci) = 0.1 nci 1 RCL (U) = 0.1 nci x 0.8 x 0.1 x 2,220 dpm/nci = 18 cpm net 1 RCL (Am) = 0.01 nci x 0.8 x 0.1 x 2,220 dpm/nci = 2 cpm net Obviously, a contamination survey instrument such as a TBM-3 is easily able to detect less than 1 RCL on a wipe filter for some nuclides, but not for others. Precision is of little importance since the intent is to rapidly identify areas that require decontamination. Leak Tests of Sealed Sources For all nuclides except radium-226, the leak test limit is 5 nci on a wipe of the source capsule or the accessible portions of the source housing. The appropriate instrument for measuring activity on a filter used for a leak test of a sealed source depends on the emissions from the source nuclide. For alpha emitters and medium-energy beta emitters, e.g. Co-60 and Cs-137, the TBM-3 point-source efficiency is about 0.1. The expected net count rate for the leak test limit of 5 nci on a wipe is: Net Rate = 5 nci x 0.1 x 0.8 x 2,220 dpm/nci = 888 cpm RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 3
4 To be on the safe side, round downward to 800 cpm as the acceptance criterion. For Sr-90, the TBM-3 point-source efficiency is about 0.2, resulting in a leak test acceptance criterion of 1,600 net cpm. Laboratory instruments should be used for counting wipes from gamma-only emitters such as Co-57 or Ba- 133 since portable instruments with calibrated efficiencies for these nuclides are usually not readily available. PERSONNEL QUALIFICATIONS FOR INSTRUMENT CALIBRATION Calibrations of instruments used for radiation protection purposes are to be performed by, or under the direct supervision of, individuals who have had specific training on the following subjects: 1. Principles and practices of radiation protection. 2. Radioactivity measurement standardization, monitoring techniques and instruments. 3. Mathematics and calculations pertinent to the use and measurement of radioactivity. 4. Normal operating procedures and emergency procedures for all calibration sources to be used. Evidence of appropriate training may include satisfactory completion of formal academic courses on the subjects listed above, certification by the American Board of Health Physics, certification in radiological physics by the American College of Radiology, registration by the National Registry of Radiation Protection Technologists, or by specific training and examination administered by the RSO. CALIBRATION SOURCES Exposure rate: The 50-curie Cs-137 "Radiac" calibrator is used for essentially all photon exposure-rate calibrations. (The complete designation of the unit is the Radiac Calibrator AN/UDM-1A, also sometimes referred to as the "UDM".) The source provides three radiation field intensities by means of a rotatable source container and a movable aperture plug. The calibration room permits a range of calibration distances from approximately 0.5 to 5.0 meters. An adjustable calibration table rolls on a track to provide precise positioning of the instruments to be calibrated. Available exposure rates range from approximately 0.5 mr/hr to nearly 100 R/hr; these exposure rates have been verified many times with a variety of transfer instruments, resulting in a high degree of confidence in the radiation field intensity. A 1 mci Cs-137 source (Amersham CDC.701, X.7) is available for calibration of exposure rates below 0.5 mr/hr. Alpha activity: Alpha survey instruments are calibrated with a set of electrodeposited Pu-239 sources ranging from 1.1x10 3 to 1.44x10 6 alpha particles per minute emitted from the source surfaces (i.e. 2π emission). Beta activity: Calibration sources containing various beta-emitting nuclides are available for determining the energy response of survey instruments. Other sources containing various amounts of the same nuclide are available for determining the linearity of beta-detecting instruments over several operating ranges. For testing linearity of response, the calibration sources consist of sets of 4 planchets containing Tc-99 with activities covering 4 orders of magnitude. The sources are calibrated for 2π-emission rates, i.e. beta particles per minute emitted from the source surfacesfor testing energy dependence of the instrument response, sets of sources containing several nuclides, e.g. C-14, Cl-36, Sr-Y-90 and Pm-147, are available. Low-energy photon fluence: I-129 sources are available for calibration of instrument response to I-125. Although the photon energies are essentially identical, the yield of photons in the kev range is 77.5% for I-129 compared with 146.2% for I-125. This difference in yield (ratio = 1.89) must be accounted for in the calibration. Neutron fluence rates: Two neutron sources are available for checking the response of neutron survey meters: a. 10 mci of Am-Be ( 2x10 4 n/s) b. 2 Ci of Pu-Be ( 4x10 6 n/s) RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 4
5 CALIBRATION PROCEDURES Exposure Rate Instruments Before attempting to calibrate an instrument, verify that the instrument is not damaged, is not contaminated and that the batteries are in good condition, i.e. that the instrument responds satisfactorily to a check source. The mechanical zero of the meter should be checked with the instrument off and resting or held in a normal position. The mechanical adjustment screw on the meter movement is to be used only for setting the mechanical zero point of the meter and never for adjusting the electrical zero response point. Unless the meter has been heavily jarred or tampered with the mechanical zero should need no adjustment. Directional and energy response characteristics of survey instruments are normally taken from manufacturers' specifications. Calibrations are performed with the detector oriented toward the source in a manner comparable to that in which it is ordinarily used. Ionization chamber survey instruments with air-equivalent wall materials normally exhibit little directional dependence throughout the hemisphere directly facing the detector chamber. For calibration using a gamma source of moderate to high energy (e.g. Cs-137), the ion chamber is positioned with its axis perpendicular to the primary radiation beam. This allows greater precision in determining the distance from the source to the effective center of the chamber. Select calibration source configurations and distances to provide two exposure rates for each linear meter scale (or switch-selectable digital readout range) to be used. The two exposure rates should be within the lower 1/3 and the upper 1/3 of the range, as close as possible to 25% and 75% of full scale. Calibration points on ion chambers used for medical areas shall be separated by 50 percent of the scale reading. Select one exposure rate near the midpoint of each decade for an instrument with a logarithmic meter readout (or non-switchable digital readout). Set the calibration table at the appropriate distance and observe the background radiation reading of the instrument at that position. Be sure that the instrument reading is visible through the mirror to the source control position outside the calibration room. Leave the room, expose the source and observe the instrument reading. Return the source to the safe position. If the instrument response is within 10% of the true exposure rate at the calibration distance, record the results and proceed to the next exposure rate configuration. If the instrument response deviates by more than 10% from the true exposure rate, adjust the instrument according to the manufacturer's recommendations until it reads correctly. After any instrument adjustments, previous measurements may require recalibration as the adjustments may be interdependent. Repeat the above procedure until all calibration points have been recorded on the "EXPOSURE RATE METER CALIBRATION RECORD" (RPR 52A). For ionization chambers that are not hermetically sealed, corrections for air density (temperature and barometric pressure) can be made in two ways. For instruments to be used only on the University campus, or at the same elevation, the preferred method is to adjust the instrument to read exposure rate directly at this altitude. The instrument will then give correct readings without the use of correction factors; this fact should be noted on the calibration record (RPR 52A) and on the calibration sticker attached to the instrument. The second method is to adjust the instrument readings to those which would have been attained if the instrument had been calibrated at sea level. This method requires a correction for barometric pressure and temperature, both at the time of calibration and at the time of use. If the instrument response is within 20%, but not within 10%, of the true exposure rate on all scales, the instrument may be used provided it is accompanied by a calibration chart or graph that permits the correction of instrument readings to true exposure rates. If all readings are within 10%, a calibration graph is not required. Affix a green calibration label (RPR 52G) indicating the calibration date and results. Check Sources Immediately after the instrument has been calibrated, the regular user of the instrument should record its response to a convenient, long-lived check source. The instrument response to the check source should be verified each day the instrument is used and recorded at least quarterly. For instruments calibrated for low exposure rates (<0.5 mr/hr), a Coleman lantern mantle is adequate as a check source (see RPR 52E). Contamination and Wipe-Test Survey Instruments RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 5
6 Instruments used to detect surface contamination are not calibrated in the same sense as are exposure rate survey instruments. Instead of adjusting the instrument response to a predetermined value, the response to any given source is simply recorded, as long as it is within the normal range for the instrument. Because the absorption correction factors cannot be quantitatively determined, direct surveys for surface contamination are always qualitative by nature. However, surface contamination measurements and wipe-test results must be reported as activity per 100 cm 2 and data must be provided for translating instrument readings into appropriate units, e.g. dpm per 100 cm 2. The detection efficiency will vary tremendously with the type of radiation and its energy. Low energy gamma, detectors, e.g. thin-window scintillation probes; thinwindow, gas-filled detectors, e.g. GM or proportional counters, usually exhibit approximately the same detection efficiency over the entire window area. The overall measurement efficiency should be expressed as a fraction, representing the ratio of instrument response to source activity, e.g. counts per disintegration or cpm per dpm. The point-source efficiency should be determined for each appropriate radiation beta-particle or gamma energy range. If the instrument is used exclusively for measuring a single nuclide, the calibration need be done only for that nuclide. Instruments used for a wide variety of nuclides should be calibrated for one nuclide in each beta-particle energy range, e.g C-14 (low), Cl-36 (medium) and Sr-Y-90 (high) and/or gamma energy, e.g. I-125 (low). Since each of these nuclides is not available in several sources with a range of activities, a single measurement is made for each nuclide, with the source at 1 cm from the center of the detector window is used to determine the point-source efficiency. Table 1 lists a few nuclides by energy groups and typical instrument measurement efficiencies for each energy group. Gamma counters, such as the TA=s TBS-6, possessing an external low-energy photon scintillator may be used for surveys and wipe-testing for I-125. When using calibration sources that do not emit the detected particle or photon with every disintegration, it is important to include the abundance of emissions in the calculation of efficiency. The efficiency should be calculated as the fraction of the appropriate emissions that are detected. To calculate the actual emission rate, multiply the source disintegration rate (e.g. dpm) by the fractional abundance of the detected emission. Note in Table 1 that the emission abundances for some sources are less than 1 and some are greater than 1. The linearity of response on all usable ranges of the instrument is determined only for Tc-99. The calibration data should be recorded on "CONTAMINATION METER CALIBRATION RECORD" (RPR 52B). Alpha scintillation probes are generally larger than thinwindow GM probes and may exhibit very significant differences in detection efficiency at various positions across the detector window. An average of at least 5 measurements distributed over the window area should be used as the measurement efficiency. The measurement efficiency for radionuclides most often encountered should be on the label attached to the instrument (RPR 52Y) or in some place of ready reference if estimates of quantitative surface contamination are needed. Immediately after the instrument has been calibrated, the regular user of the instrument should determine and record its response to a convenient, long-lived check source. The instrument response to the check source should be verified each day before being used to make measurements. Coleman lantern mantles are often used as check sources for contamination survey meters and low-range exposure-rate meters. For reproducibility and record keeping, a combined holder and record of checksource readings is available for routine use of these check sources (RPR 52E). Neutron Instruments If a neutron survey instrument is to be used to make dose-rate measurements, it must be calibrated by a qualified vendor, since the University has no calibrated neutron source. The response may be checked after calibration by a vendor, and before using the instrument with either the 10 mci Am-Be source or the 2 Ci Pu-Be source. REFERENCES American National Standards Institute, Radiation Protection Instrumentation Test and Calibration, ANSI N Surface Radioactivity Guides for Materials, equipment and Facilities to be Released for Uncontrolled Use, ANSI P/N13.12, 1987 (draft). RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 6
7 International Commission on Radiation Units and Measurements, - Determination of Dose Equivalents Resulting from External Radiation Sources, ICRU Report 39, National Council on Radiation Protection and Measurements, Instrumentation and Monitoring Methods for Radiation Protection, NCRP Report No. 57, Calibration of Survey Instruments for the Assessment of Ionizing Radiation Fields and Radioactive Surface Contamination, NCRP Report No. 112, U.S. Nuclear Regulatory Commission, 1980, Guide for the Preparation of Applications for Medical Programs, Reg. Guide 10.8, Rev. 1, Appendix D, Calibration of Instruments. - Termination of Operating Licenses for Nuclear Reactors, NRC Reg. guide 1.86, U.S. Nuclear Regulatory Commission, 1999, AInstrument Specifications and Model Survey Instrument and Air Sampler Calibration Program,@ NUREG-1556, Vol.11, Appendix O. RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 7
8 TABLE 1. TYPICAL INSTRUMENT RESPONSE Response to Calibration Source Nuclide, E (Average), Abundance Appropriate Survey Instrument Efficiency cpm/nci Very low-energy electron/beta emitters H-3, 6 kev, 100% Portable instruments are not applicable; use liquid scintillation Fe-55, 6 kev, 60% Ni-63, 17 kev, 100% Low-energy beta emitters C-14, 50 kev, 100% Thin-window GM, e.g.ta TBM-3 or TBS S-35, 50 kev, 100% Ca-45, 70 kev, 100% Tc-99, 101 kev, 100% Thin-window GM, e.g. TA TBM-3 or TBS Medium-energy beta emitters Cl-36, 279 kev, 98% Thin-window GM, e.g. TA TBM-3 or TBS High-energy beta emitters Sr-Y-90, 565 kev, 200% Thin-window GM, e.g. TA TBM-3 or TBS P-32, 695 kev, 100% Photon emitters with no particulate emissions: Low-energy photons I-125, kev, 147% Thin-crystal NaI scintillator, I-129, kev, 78% e.g. external probe on TA TBS-6 Medium-energy photons Co-57, 122 kev, 86% 136 kev, 11% Cr-51, 320 kev, 10% RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 8
9 RPR 52A. EXPOSURE RATE METER CALIBRATION RECORD Responsible User: Group #: Task #: Location: Phone: INSTRUMENT: Manufacturer: Model: Ser.#: Batteries: Replaced OK, not replaced Detector/Probe: Hermetically sealed? Yes No Temperature, T = o C Pressure, P = mm Hg Air Density Correction Factor = 760 (273 + T)/295 P = (should be 1.2) Detector Orientation to Radiation Field Perpendicular Parallel CALIBRATION SOURCE(S): 50 Ci Cs-l37, AN/UDM-1A for gamma exposure rates >0.5 mr/hr 1 mci Cs-137, Amersham CDC.701, X.7 for exposure rates <0.5 mr/hr 10 mci Am-Be or 2 Ci Pu-Be for neutron response check Other: RESULTS: Distance Exposure Rate Instrument Response (mr/hr, R/hr) Error as Left (cm) (mr/hr, R/hr) Scale As Found As Left Corrected 10% 20% If error on any scale to be used exceeds 20%, the instrument is unacceptable; affix RED LABEL. If error exceeds 10%, but is less than 20%, the instrument is acceptable ONLY IF a calibration chart must be attached to the instrument. If the error is within 10%, the instrument is acceptable without a calibration chart. If the instrument is acceptable, affix a GREEN LABEL. Calibrated by: Date: RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 9
10 RPR 52B. CONTAMINATION SURVEY METER EFFICIENCY CALIBRATION RECORD Responsible User: Group #: Task #: Location: Phone: INSTRUMENT: Manufacturer: Model: Ser.#: Batteries: Replaced OK, not replaced Detector/Probe: Window thickness: mg/cm 2 Window area: cm 2 CALIBRATION SOURCE(S): Eberline Pu-239 source set, ID#: for alpha activity Eberline Tc-99 source set (DNS-19), ID#: for beta linearity response N.E.N. multiple nuclide set (NES-269), ID#: for beta energy response N E N I-129 (NES-269), ID#. for I-125 energy response Other: POINT SOURCE CALIBRATION RESULTS: (Source at 1 cm from center of detector window) Detection Efficiency Instrument Response Cal. Point Scale Avg. Nuclide Activity Units Scale Units As Found As Left (cpm/dpm) (cpm/dpm) If detection efficiencies for the same nuclide on different scales are within 10% of each other, and the efficiencies for different energies are within the expected (nominal) range, affix a YELLOW LABEL to the instrument. Calibrated by: Date: RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 10
11 RPR 52D. SURVEY INSTRUMENT CALIBRATION LABELS Reduce to 13" X 22" (64%) before printing; permanent self-adhesive labels. CONTAMINATION SURVEY METERS - YELLOW LABEL, RPR 52Y: CONTAMINATION SURVEY METER C-14 P-32 I-125 Efficiency: Note: By: S/N: Date: Check Source Reading (if any): Date: Due By: Radiological Health Department ( ) RPR 52Y (12/2013) EXPOSURE RATE SURVEY METERS - GREEN LABEL, RPR 52G: EXPOSURE RATE SURVEY METER [ ] WITHIN 10% ON ALL USABLE SCALES. (Indicate uncalibrated scales) [ ] WITHIN 20%; USE CALIBRATION GRAPH. Cs-137 calibration: consult instrument manual for response at other energies. By: Date: S/N: Due By: Radiological Health Department ( ) RPR 52G (12/2013) OUT OF SERVICE METERS- FLUORESCENT RED LABEL, RPR 52FR: In the case where the instrument is Red Tagged and the laboratory(s) has access to the other survey meters, or the lab(s) does not want to repair the instrument, or the meter is unrepairable; then the instrument will be wrapped with the OUT OF SERVICE label. OUT OF SERVICE OUT OF SERVICE Date: Radiological Health Department ( ) RPR 52FR (12/2013) RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 11
12 FAILED CALIBRATION METERS RED CARDSTOCK, RPR 52R: Red Tag means to affix a notice to an instrument which has failed calibration and needs to either be sent back to the manufacturer for appropriate service, onsite repair/decontamination if available, or the instrument may require parts. If the instrument, during calibration at RHD, fails then it should be labeled with the RED TAG. FAILED CALIBRATION METER Meter Make Model S/N DO NOT USE UNTIL Meter Calibrated/Repaired/Decontaminated o Check Electronics o Check Audio o Check Detector o Calibrated o Repaired o Decontaminated Date: Name: Comments: Radiological Health Department ( ) RPR 52R (12/2013) BATTERY REPLACEMENT METERS FLUORESCENT LIME GREEN LABEL, RPR 52FLG: In some cases, lack of batteries in the instrument may cause a false reading. The instrument should be checked with the proper battery style; if the problem is related to the battery issue, lab(s) personnel may label it on their own. Use the proper Tag for this reason which, after correction, lab(s) personnel can remove it. DO NOT USE THIS METER UNTIL BATTERIES ARE REPLACED Date: Radiological Health Department ( ) RPR 52FLG (12/2013) RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 12
13 RPR 52E. CHECK SOURCE RESPONSE Instrument reading should be recorded immediately after calibration and at least quarterly thereafter. Manufacturer Model Serial # Date Reading Date Reading _ fold here _ Instrument Check Source Property of Instructions: Cut a 1" diameter hole For replacement, call: centered on the cross lines. Radiological Health Mount Coleman lantern mantle Ext inside the folded card. Take reading with the detector centered on the cross lines. RPR 52E (12/2013) RPR 52. INSTRUMENT CALIBRATION & USE (12/2013) - 13
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the
Quality Assurance Testing of AtomLab Dose Calibrators
(Manufacturer s* Instructions) Introduction The following procedures are the manufacturer s instructions for qualifying AtomLab 100, 100+, 200, 300, 400, and 500 dose calibrators in accordance with NRC
Main properties of atoms and nucleus
Main properties of atoms and nucleus. Atom Structure.... Structure of Nuclei... 3. Definition of Isotopes... 4. Energy Characteristics of Nuclei... 5. Laws of Radioactive Nuclei Transformation... 3. Atom
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
Emergencies Involving Radiation
Page 1-1 Section 1 Emergencies Involving Radiation Contents A. Emergency Phone Numbers...1-2 1. Campus... 1-2 a. Radiation Safety Office 543-0463... 1-2 b. University Police 9-911... 1-2 2. Off Campus...
Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material)
Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material) Table of Content Receiving a Radioactive Package Constancy Test Daily Survey Weekly Wipes Accuracy Test Linearity Geometric
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
Radiation Detection and Measurement
Radiation Detection and Measurement June 2008 Tom Lewellen [email protected] Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron
EMERGENCY PROCEDURES FOR RADIOACTIVE MATERIAL. 2. Prevent the spread of contamination by covering the spill with absorbent paper.
EMERGENCY PROCEDURES FOR RADIOACTIVE MATERIAL MINOR SPILLS: 1. Notify persons in the area that a spill has occurred. 2. Prevent the spread of contamination by covering the spill with absorbent paper. 3.
Introduction to Geiger Counters
Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists
HAZMAT Employee Training D.O.T. Radioactive Material Shipments #8 (02/12)
HAZMAT Employee Training D.O.T. Radioactive Material Shipments #8 (02/12) HAZMAT EMPLOYEE: TRAINING DATE: 1. Hazardous Materials Regulations (HMR) originate from the following government agency: a. U.S.
Activities at temporary job site utilizing U.S. NRC Radioactive Material License #06-30556-01 Amendment 03 0303531(,
CABRERA SERVICES RADIOLOGICAL. ENGINEERING. REMEDIATION May 17,2011 Regional Administrator U.S. Nuclear Regulatory Commission Region I 475 Allendale Road King ofprussia, P A 19406 ATTN: Director, Division
PERSONNEL MONITORING AND DOSIMETRY POLICIES
PERSONNEL MONITORING AND DOSIMETRY POLICIES All individuals who are required to have their exposure to ionizing radiation monitored must be trained prior to using the source(s) of radiation. The radioactive
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)
Ionizing Radiation, Czech Republic, (Czech Metrology Institute) Calibration or Measurement RADIOACTIVITY 1.0E+00 1.0E+02 Bq cm -2 C-14 1.0E+01 1.0E+02 Bq cm -2 Co-60 1.0E+01 1.0E+02 Bq cm -2 Sr-90 1.0E+01
Radioactivity III: Measurement of Half Life.
PHY 192 Half Life 1 Radioactivity III: Measurement of Half Life. Introduction This experiment will once again use the apparatus of the first experiment, this time to measure radiation intensity as a function
INFORMATION NOTICE: 09-01-01 USE OF XRF DEVICES CONTAINING RADIOACTIVE MATERIAL North Dakota Department of Health Radiation Control Program (2/06)
INFORMATION NOTICE: 09-01-01 USE OF XRF DEVICES CONTAINING RADIOACTIVE MATERIAL North Dakota Department of Health Radiation Control Program (2/06) Radioactive material is regulated by the North Dakota
Radiation Strip Thickness Measurement Systems
Radiation Strip Thickness Measurement Systems During the past years we have increased our sales of radiometric Vollmer strip thickness measurement systems, i.e. X-ray or isotope gauges, dramatically. Now,
notes MERRTT IS-302 Radiological Survey Instruments and Dosimetry Devices
INTRODUCTION This module provides general information about radiological survey instruments. Basic functions and limitations are discussed. Because there are many models of survey instruments available,
RADIATION MONITORING SYSTEMS
RADIATION MONITORING SYSTEMS Area monitors Process monitors In-line, Adjacent-to-Line monitors Local/remote processors, Data Acquisition Systems NUPIC Certified, NRC Approved Vendor Wide Range Sensitivity
Radium-223 Therapy: Handling & Radiation Safety Issues
Nuclear Physics Enterprises Radium-223 Therapy: Handling & Radiation Safety Issues Jeffry A. Siegel, Ph.D. President & CEO 2014 Meeting Mid-Atlantic States Radiation Control Programs Malvern, PA NRC s
Selection, use and maintenance of portable monitoring instruments
HSE information sheet Selection, use and maintenance of portable monitoring instruments Ionising Radiation Protection Series No 7 Introduction The Ionising Radiations Regulations 1999 (IRR99) 1 require
Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry
Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry John Keightley NSUF 18 May 2010 Acknowledgement: Several discussions with Stefaan Pommé, IRMM, EC-JRC, Geel Some diagrams
Application Note. ipix A Gamma imager to support various applications
Application Note ipix A Gamma imager to support various applications Introduction ipix is a unique gamma imager that quickly locates low level radioactive sources from a distance and estimates the dose
P O L Y T E C R.M. S.R.L.
INFORMACIÓN FUENTES EXCEPTUADAS PARA ENSEÑANZA RSS-2 G-M Resolving Time Set This set is used for determining the resolving time of GM Counters. It consists of three half discs, two of which contain 5 microcuries
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics
COMPENDIUM OF EPA-APPROVED ANALYTICAL METHODS FOR MEASURING RADIONUCLIDES IN DRINKING WATER
COMPENDIUM OF EPA-APPROVED ANALYTICAL METHODS FOR MEASURING RADIONUCLIDES IN DRINKING WATER June 1998 Prepared by the Office of Environmental Policy and Assistance Air, Water and Radiation Division (EH-412)
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting. Method Number: ELAP method code 7037
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting Method Number: ELAP method code 7037 SOP Number: Revision Number: SOP Date: Personnel / Data
Chapter NP-5. Nuclear Physics. Nuclear Reactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 NUCLEAR REACTIONS 2.0 NEUTRON INTERACTIONS
Chapter NP-5 Nuclear Physics Nuclear Reactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 2.0 NEUTRON INTERACTIONS 2.1 ELASTIC SCATTERING 2.2 INELASTIC SCATTERING 2.3 RADIATIVE CAPTURE 2.4 PARTICLE
Nuclear Physics. Nuclear Physics comprises the study of:
Nuclear Physics Nuclear Physics comprises the study of: The general properties of nuclei The particles contained in the nucleus The interaction between these particles Radioactivity and nuclear reactions
GUIDE FOR THE PREPARATION OF RADIOACTIVE MATERIAL APPLICATIONS FOR WELL LOGGING OPERATIONS IN KENTUCKY
GUIDE FOR THE PREPARATION OF RADIOACTIVE MATERIAL APPLICATIONS FOR WELL LOGGING OPERATIONS IN KENTUCKY Radiation Control Cabinet for Human Resources 275 East Main Street Frankfort, Kentucky 40621 JANUARY
Radiological mapping and characterization at the Barsebäck nuclear power plant
Radiological mapping and characterization at the Barsebäck nuclear power plant Leif Spanier, Scandpower AB Lars Håkansson, Barsebäck Kraft AB Radiological Characterisation for Decommissioning Workshop
GAMMA-RAY SPECTRA REFERENCES
GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.
LIST OF APPENDICES. PART 1 - MODEL PROCEDURES THAT APPLICANTS MAY USE TO PLAN RADIATION SAFETY PROGRAMS [See Appendix X for imact of Part 20 Revision]
LIST OF APPENDICES PART 1 - MODEL PROCEDURES THAT APPLICANTS MAY USE TO PLAN RADIATION SAFETY PROGRAMS [See Appendix X for imact of Part 20 Revision] Appendix A Model Training Program... A-1 B Model Procedure
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
Antoine Henri Becquerel was born in Paris on December 15, 1852
Discovery Antoine Henri Becquerel was born in Paris on December 15, 1852 Summit Environmental Technologies, Inc. Analytical Laboratories 3310 Win Street Cuyahoga Falls, Ohio 44223 Fax: 1-330-253-4489 Call
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
MASS DEFECT AND BINDING ENERGY
MASS DEFECT AND BINDING ENERGY The separate laws of Conservation of Mass and Conservation of Energy are not applied strictly on the nuclear level. It is possible to convert between mass and energy. Instead
DOE-HDBK-1122-99. Module 2.18 Air Sampling Equipment. Instructor s Guide
Course Title: Radiological Control Technician Module Title: Air Sampling Equipment Module Number: 2.18 Objectives: 2.18.01 Identify the factors that affect the operator's selection of a portable air sampler.
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
Qualitätsmanagement-Handbuch
U r Kalibrier- und Messmöglichkeiten Qualitätsmanagement-Handbuch Relative erweiterte Messunsicherheit, die sich aus der Standardmessunsicherheit durch Multiplikation mit dem Erweiterungsfaktor k = 2 ergibt.
RPR 54 RADIOACTIVE WASTE MANAGEMENT
RPR 54 RADIOACTIVE WASTE MANAGEMENT PURPOSE This procedure specifies how the Radiological Health Department collects, stores, processes, packages, and disposes of radioactive waste. POLICY All users of
Ramon 2.2 Radon-Monitor
Ramon 2.2 Radon-Monitor Applications Measurement Modes - Technology - Quality 1 1) Application of the Ramon 2.2: The Ramon2.2 Radon monitor is a unique device and it can be best described as an automatic,
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
Radiation Monitoring System RMS
Radiation Monitoring System RMS VF RMS radiation monitoring system is a modular and standardized system for the radiological monitoring. Variety of radiation monitors, display and alarm units can be connected
MATERIALS LICENSE. 1. Honeywell International, Inc. 3. License Number: SUB-526, Amendment 6-1
NRC FORM 374 U.S. NUCLEAR REGULATORY COMMISSION Page 1 of 5 Pursuant to the Atomic Energy Act of 1954, as amended, the Energy Reorganization Act of 1974 (Public Law 93-438), and Title 10, Code of Federal
Using the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
............... [2] At the time of purchase of a Strontium-90 source, the activity is 3.7 10 6 Bq.
1 Strontium-90 decays with the emission of a β-particle to form Yttrium-90. The reaction is represented by the equation 90 38 The decay constant is 0.025 year 1. 90 39 0 1 Sr Y + e + 0.55 MeV. (a) Suggest,
ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.
1 ATOMIC SPECTRA Objective: To measure the wavelengths of visible light emitted by atomic hydrogen and verify the measured wavelengths against those predicted by quantum theory. To identify an unknown
Objectives 404 CHAPTER 9 RADIATION
Objectives Explain the difference between isotopes of the same element. Describe the force that holds nucleons together. Explain the relationship between mass and energy according to Einstein s theory
Environmental Health and Safety Radiation Safety. Module 1. Radiation Safety Fundamentals
Environmental Health and Safety Radiation Safety Module 1 Radiation Safety Fundamentals Atomic Structure Atoms are composed of a variety of subatomic particles. The three of interest to Health Physics
MATERIALS LICENSE. In accordance with the letter dated January 15, 2009,
NRC FORM 374 U.S. NUCLEAR REGULATORY COMMISSION PAGE 1 OF 5 PAGES Pursuant to the Atomic Energy Act of 1954, as amended, the Energy Reorganization Act of 1974 (Public Law 93-438), and Title 10, Code of
Basics of Nuclear Physics and Fission
Basics of Nuclear Physics and Fission A basic background in nuclear physics for those who want to start at the beginning. Some of the terms used in this factsheet can be found in IEER s on-line glossary.
AP2 Magnetism. (c) Explain why the magnetic field does no work on the particle as it moves in its circular path.
A charged particle is projected from point P with velocity v at a right angle to a uniform magnetic field directed out of the plane of the page as shown. The particle moves along a circle of radius R.
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
ORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software
ORTEC DET-SW-UPG Latest Software Features Three Search Modes: Gamma/Neutron total count rate. SNM search mode. Sliding average "monitor" mode. (NEW) User choice of identification schemes: Classify mode
Measurement of Germanium Detector Efficiency
Measurement of Germanium Detector Efficiency Marcus H. Wiggs 2009 Notre Dame Physics REU Advisor: Dr. Philippe Collon Mentors: Matthew Bowers, Daniel Robertson, Chris Schmitt ABSTRACT: A possible discrepancy
Introduction. Chapter 15 Radiation Protection. Regulatory bodies. Dose Equivalent. Regulatory bodies. Main Principles of Radiation Protection
Introduction Chapter 15 Radiation Protection Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. F.M. Khan, The Physics of Radiation Therapy, 4th ed., Chapter
Part 5 of 9 Radiological Control Technician Training Site Academic Training Instructor s Guide Phase I
Part 5 of 9 Radiological Control Technician Training Site Academic Training Phase I Coordinated and Conducted for the Office of Health, Safety and Security U.S. Department of Energy Radiological Control
Chemistry 1000 Lecture 2: Nuclear reactions and radiation. Marc R. Roussel
Chemistry 1000 Lecture 2: Nuclear reactions and radiation Marc R. Roussel Nuclear reactions Ordinary chemical reactions do not involve the nuclei, so we can balance these reactions by making sure that
Copies of this document shall be classified either as Controlled or Uncontrolled, and shall be identified as such on the cover sheet.
INTRODUCTION This document describes an inspection system broken down into four separate levels. When Inspection Level IV is invoked, this system meets the requirements of former MIL-I-45208A. The primary
Decay of Radioactivity. Radioactive decay. The Math of Radioactive decay. Robert Miyaoka, PhD. Radioactive decay:
Decay of Radioactivity Robert Miyaoka, PhD [email protected] Nuclear Medicine Basic Science Lectures https://www.rad.washington.edu/research/research/groups/irl/ed ucation/basic-science-resident-lectures/2011-12-basic-scienceresident-lectures
RADIATION SAFETY MANUAL
Approved by the University Radiation Safety Committee 2014 - September 29 Table of Contents i. Scope and Purpose...3 ii. Abbreviations...4 iii. Glossary of Terms / Definitions...5 iv. Concordia University
Thyroid Screening for Radioiodine
Regulatory Document Thyroid Screening for Radioiodine July 2008 CNSC REGULATORY DOCUMENTS The Canadian Nuclear Safety Commission (CNSC) develops regulatory documents under the authority of paragraphs 9(b)
NORM MANAGEMENT GUIDELINES & PRINCIPLES
Introducing NORM NORM MANAGEMENT Three is glad to present some basic hints about NORM that should be taken onto consideration whilst dealing with these matters The acronym NORMmeans all Naturally Occurring
WM2012 Conference, February 26-March 1, 2012, Phoenix, Arizona, USA
Radioactive Waste Characterization Strategies; Comparisons Between AK/PK, Dose to Curie Modeling, Gamma Spectroscopy, and Laboratory Analysis Methods- 12194 Steven J. Singledecker, Scotty W. Jones, Alison
Azeri, Chirag & Gunashli Full Field Development Phase 3 Environmental & Socio-economic Impact Assessment. A10.1 Introduction...
APPENDIX 10 Normally Occurring Radioactive Material (NORM) Contents A10.1 Introduction... A10/3 A10.2 Occurrence & Concentration of NORM in Oil and Gas Operations... A10/4 A10.2.1 Probability of NORM formation
Vacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials
Illicit Trafficking, Sub/cross-national threats, Poster presentation A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials Ronald M. Keyser, Timothy R. Twomey,
USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY
GUIDE ST 5.3 / 4 MAY 2007 USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY 1 Ge n e r a l 3 2 Safety licences and exemption from licensing 3 3 Safety licence-exempt use of radiation 3
PUMPED Nd:YAG LASER. Last Revision: August 21, 2007
PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
Principle of Thermal Imaging
Section 8 All materials, which are above 0 degrees Kelvin (-273 degrees C), emit infrared energy. The infrared energy emitted from the measured object is converted into an electrical signal by the imaging
Section 4 RADIATION SAFETY RULES AND REGULATIONS
Page RR- 1 Section 4 RADIATION SAFETY RULES AND REGULATIONS UW Environmental Health and Safety Page RR-3 A. Historical Perspective Standards and regulations for radiation protection have been evolving
C-10 Radiation Monitoring Network
C-10 Radiation Monitoring Network Table of Contents C-10 Foundation History C-10 Mission & Goals History of C-10 Radiological Monitoring Network C-10 Radiation Monitoring Network (CRMN) CRMN System Architecture
Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25
Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 CHAPTER 227. RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT, X-RAY GAUGING EQUIPMENT, ELECTRON MICROSCOPES AND X-RAY CALIBRATION SYSTEMS Sec. 227.1.
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
notes Assessing Package Integrity
INTRODUCTION The federal government regulates radioactive material packaging, labeling, and transport. The labels used on packages of radioactive material can be used to help you obtain information about
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
W. C. Reinig. Savannah River Laboratory E. I. du Pent de Nemours and Company Aiken, South Carolina 298o1
.*. *.-a /dp73j/3~ DP-MS-68-48 calforn1um-252: A NEW SOTOPC SOUR(!EFOR NEUTRON RADOGRAPHY by W. C. Reinig Savannah River Laboratory E.. du Pent de Nemours and Company Aiken, South Carolina 298o1. SRL7
Alpha/Beta Counting Manual Systems
Alpha/Beta Counting Manual Systems Ultra High Performance Alpha/Beta Counting Systems for Maximum Laboratory Productivity ORTEC offers a full suite of MANUAL alpha / beta counters incorporating either
Captus 3000 Thyroid Uptake And Well System
(201) 825-9500 / (800) 631-3826 THRYOID UPTAKE SYSTEMS 51 Captus 3000 Thyroid Uptake And Well System The Ultimate... User Friendly... Intuitive Performing Thyroid Uptake measurements has never been easier.
1. In the general symbol cleus, which of the three letters. 2. What is the mass number of an alpha particle?
1. In the general symbol cleus, which of the three letters Z A X for a nu represents the atomic number? 2. What is the mass number of an alpha particle? 3. What is the mass number of a beta particle? 4.
Passive Whole Body Monitoring Nuclear Power Industry Application Introduction to the Argos -5AB Zeus (Gamma Option) and the GEM
Application Note Passive Whole Body Monitoring Nuclear Power Industry Application Introduction to the Argos -5AB Zeus (Gamma Option) and the GEM -5 Gamma Exit Monitor CANBERRA s personnel contamination
Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies
Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.
Waste Management 04 Conference, February 29 - March 4, 2004, Tucson, AZ Copyright WM Symposia, Inc. All Rights Reserved. Reprinted with permission.
RADIOLOGICAL FALSE POSITIVES IN ENVIRONMENTAL SOIL AND GROUNDWATER DATA FROM COMMERCIAL LABORATORIES Walter Kubilius Westinghouse Savannah River Company Thomas Coffey, Paul Mark EXR, Inc. Allen Volesky
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy Upp, Daniel L.; Keyser, Ronald M.; Gedcke, Dale A.; Twomey, Timothy R.; and Bingham, Russell D. PerkinElmer Instruments, Inc. ORTEC,
Nuclear Science Merit Badge Workbook
Merit Badge Workbook This workbook can help you but you still need to read the merit badge pamphlet. The work space provided for each requirement should be used by the Scout to make notes for discussing
APPENDIX N. Data Validation Using Data Descriptors
APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection
RADIOACTIVE WASTE MANAGEMENT PLAN
RADIOACTIVE WASTE MANAGEMENT PLAN Copies of this plan will be made available to all members of the campus community with the potential to generate radioactive waste. Contents I. Purpose II. Regulatory
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality