Rethinking Risk and Benefit in Dental and Maxillofacial Imaging Dose matters
|
|
|
- Eugenia Robbins
- 9 years ago
- Views:
Transcription
1 Rethinking Risk and Benefit in Dental and Maxillofacial Imaging Dose matters John B Ludlow, DDS, MS, FDS RCSEd University of North Carolina, Chapel Hill, NC Background Radiation risk is frequently front page news recent example - Japan s Fukushima Nuclear Plant following an Earthquake & Tsunami - 3/11/2011 Our patients are concerned and often confused about risk associated with different exposures We can help by being well informed 1
2 Learning objectives: Identify the risks from ionizing radiation Describe options in CBCT units which affect dose Discuss importance of matching options to objectives of imaging Explore ways to reduce patient risk Explain how to talk about risks with patients What are the risks? Gamma bomb exposure Radioactive spider bite 2
3 Stochastic vs Deterministic Effects Stochastic effects A linear-no-threshold hypothesis of x ray risk fits most data for cancer development but is extrapolated to doses below 100 mgy (adult exposure) mg (fetal exposure) No expressions of germ cell mutations have been observed in human populations Deterministic effects (Non stochastic effects) Threshold for in-utero birth defects msv Cataracts 2-5 Gy radiation burns 3 Gy (reddening) radiation mucositis 30+ Gy (therapy typically Gy) Source contribution to total effective dose (6.2 msv) per capita in the US From NCRP REPORT No. 160, 2009 Source % µsv Ubiquitous background Medical Consumer
4 CT procedures per year - US Annual growth of >10% per year From NCRP REPORT No. 160, 2009 Background NEJM Computed Tomography An Increasing Source of Radiation Exposure David Brenner, et al. From 1.5% to 2% of all cancers in the United States may be attributable to the radiation from CT studies 4
5 Background: CT & CBCT Effective Doses (2007 ICRP) Large FOV CBCT scans µsv Medium FOV CBCT scans µsv Small FOV CBCT scans µsv Medium FOV MDCT scans µsv Ludlow JB, Ivanovic M. Comparative Dosimetry of Dental CBCT Devices and 64 row CT for Oral and Maxillofacial Radiology Oral Surg Oral Med Oral Pathol Oral Radiol Endodont 2008;106: Dose and Risk Estimation 1990 Recommendations of the International Commission on Radiological Protection Effective dose calculation (Sv) Summed doses to weighted organs & tissues known to be most susceptible to radiation damage Mathematical expression: E = w T x H T 5
6 ICRP 2007 Recommendations 2007 Recommendations of the ICRP Reassessment of Risk based on cancer incidence data from the Life Span Study of Japanese atomic bomb survivors Revision of list of tissues Adjustment of weights Effective dose: E = w T x H T Tissue weighting factors for calculation of Effective Dose Comparison of 1990 and 2007 ICRP Recommendations * Adrenals, brain, upper large intestine, small intestine, kidney, muscle, pancreas, spleen, thymus, uterus Adrenals, Extrathoracic region, Gall bladder, Heart, Kidneys, Lymphatic nodes, Muscle, Oral Mucosa, Pancreas, Prostate, Small Intestine, Spleen, Thymus, and Uterus/cervix Recommendations of the ICRP. Publication 60. Ann ICRP 1991; 21: Recommendations of the ICRP. Publication 103. Ann ICRP 2007; 37: Tissue w T w T Bone marrow Breast Colon Lung Stomach Bladder Esophagus Gonads Liver Thyroid Bone surface Brain remainder 0.01 Salivary glands Skin Remainder Tissues 0.05*
7 Summary of changes ICRP additional weighted tissues 10% increase in weight of tissues located in maxillofacial area 28% increase in weight adjusted for distribution of tissues in maxillofacial area 3 of the newly weighted tissues are entirely within the maxillofacial area: oral mucosa, extrathoracic region, and salivary glands Effective Dose and Detriment (Risk) Calculation Detriment includes the weighted probabilities fatal and non-fatal cancer relative length of life lost hereditary effects Cancer risk alone may be used for the 2007 ICRP risk estimates for dental radiography Cancer Risk = E(Sv) x Cancer risks for children are 2 or more times greater than for adults 7
8 How do we measure dose? Effective Dose calculation preferred Human phantom studies Expensive Simple acrylic phantoms CTDI VOL Easy but inaccurate Monte Carlo modeling Promising but model and software dependent Dose area Product x coefficient for head/neck exposure Easy but inaccurate RANDO Level 6 Levels
9 Fields of view Large Medium Small > 15 cm cm < 10 cm Dental CBCT units available in the US Small FOV Medium FOV Large FOV Unit Name Company Unit Name Company Unit Name Company Picasso Trio VaTech NewTom Vgi AFP Imaging NewTom 3G AFP Imaging Accuitomo 80 J. Morita Reve 3D VaTech Alphard 3030 Belmont / Asahi Veraview-epocs 3D J. Morita CB-500 Gendex Master 3DS VaTech D Kodak D med FOV Kodak i-cat Next Gen Imaging Sciences Int Skyview 6 FOV My-Ray Accuitomo 170 J. Morita Ilumina Imtech / 3M Promax CBCT Planmeca Skyview 9 FOV My-Ray D large FOV Kodak Orion Ritter Imaging Scanora 3D med FOV Soredex Promax 3D Max Planmeca Suni 3D Suni Galileos Comfort Sirona NewTom 9000* QR Verona Prexion TeraRecon Galileos Compact Sirona i-cat Classic* Imaging Sciences Int Auge Zio OP 300 3D Asahi Roentgen Intrumentarium CB Mercuray* Hitachi Medical * No longer manufactured Orthophos XG 3D Sirona 9300 Kodak 15 manufacturers 32 units since 2000 (as many as 40 unit variations world-wide) 9
10 Field of view Note that detector sizes need to be larger than the FOV due to image magnification Image Intensifiers produce spherical FOVs Flat panels produce cylendrical FOVs Cylinders typically provide larger useful FOVs than spheres. Cylinder height Sphere diameter Large FOV Units ex: Kodak kv 2-15 ma Pulsed Flat Panel large FOV version (18 cm x 20 cm) 10
11 Large FOV CBCT Dose Calculations (Based on ICRP 2007 Recommendations) Large FOV Techniques Effective Dose in µsv Dose as multiple of average Panoramic Dose Days of per capita background* Probability of x in a million fatal cancer NewTom3G Large FOV CB Mercuray Facial FOV (maximum quality) CB Mercuray Facial FOV (standard quality) i-cat Classic Extended Field Next Generation i-cat Portrait mode Kodak cm x 18 cm (medium adult) Iluma (standard) X Iluma (ultra) SCANORA 3D dual scan *3,000 µsv, NCRP Report No. 145, 2003 Average of 5 units dose in µsv x 5.5x10-2 Medium FOV units ex: Kodak kv 2-15 ma Pulsed Flat Panel medium FOV version (9 cm x 15 cm) 11
12 Medium FOV CBCT Dose Calculations (Based on ICRP 2007 Recommendations) Medium FOV Techniques Effective Dose in µsv Dose as multiple of average Panoramic Dose Days of per capita background* Probability of x in a million fatal cancer CB Mercuray Panoramic FOV Classic i-cat Standard scan Next Generation i-cat Landscape mode Galileos (default exposure) SCANORA 3D large FOV Newtom VG CB-500 extended diameter scan Kodak cm x 15 cm (medium adult) Somaton 64 MDCT Somaton 64 MDCT w/ CARE Dose 4D *3,000 µsv, NCRP Report No. 145, 2003 Average of 5 units dose in µsv x 5.5x X Small FOV Units ex: Kodak D Panoramic unit Sensor switches from pan to 3D electronically Volume size: 3.7 cm x 5 cm Voxel size 67 µm CMOS w/ optical fiber kv 2-15 ma Pulsed 16 bit 12
13 Small FOV CBCT Dose Calculations (Based on ICRP 2007 Recommendations) Small FOV Techniques CB Mercuray I FOV (maxillary) CB cm x 8 cm Standard 0.3 or 0.4 mm Orthophos XG 3D 8 cm x 8 cm (medium adult) Promax 3D 8 cm x 8 cm (medium adult) PreXion 3D 8 cm x 8 cm (standard exposure) OP300 8 cm x 6 cm FOV Standard dose & res SCANORA 3D 6 cm x 6 cm (avg sextant) Kodak cm x 5 cm (avg sextant) Effective Dose in µsv Dose as multiple of average Panoramic Dose Days of per capita background* Probability of x in a million fatal cancer X *3,000 µsv, NCRP Report No. 145, 2003; Average of 5 units; dose in µsv x 5.5x10-2 Kodak 9000 effective dose* * ICRP 2007 calculation Average of 5 units: Sirona - Orthophos XG, Planmeca - ProMax, Kodak , SCANORA 3D, Instrumentarium - OP 200 VT 13
14 Variable FOV units ex: Kodak FOVs + pan Voxel size µ Scan time sec Effective Dose Comparison of Carestream CS Cone Beam CT (CBCT) and Multi-Slice CT (MSCT) 2,3,4 Systems Carestream CS 9300 (CBCT) 1 and Multi-Slice CT (MSCT) 2,3,4 Effective Dose Comparison Sinus 25x12 cm Temporal 25x8 cm Sinus 25x12 cm Temporal 25x8 cm Sinus 17x13.5 cm Head 17x13.5 cm Sinus 17x11 cm Jaw 10x10 cm Temporal 17x6 cm Posterior maxillary 5x5 cm Temporal 5x5 cm TMJ 8x8 cm Full panoramic Standard MSCT Low-Dose MSCT CS 9300 CBCT msv 1 Ludlow JB. Effective doses of CS 9300 cone beam CT system conducted in June ICRP 2007 tissue weights used. Absorbed dose calculated for bone marrow, thyroid, esophagus, skin, bone surface, salivary glands, brain, lymphatic nodes, extrathoracic airway, muscle, oral mucosa. 2 Faccioli et al. Radiation dose saving through the use of cone beam CT in hearing impaired patients. Radiol Med. 114: , Niu et al. Radiation dose to the lens using different temporal bone CT scanning protocols. AJNR 31: , Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 106: ,
15 Panoramic Replacement? CBCT is inferior to conventional panoramic imaging for Caries interpretation False positive diagnoses Radiation dose Periodontal assessment CBCT detector Metal artifact effects Some CBCT units have a conventional panoramic option Panoramic detector Kodak D Conventional pan vs. CT panoramic layer conventional 20 mm 10 mm 15
16 Affect of layer thickness on CT panoramic image 5 mm 20 mm Effective dose distribution in maxillofacial imaging 40% of total E dose is from bone marrow & thyroid exposure 55% of total E dose is from salivary gland and remainder exposure 16
17 Effect of Field of View on dose distribution Dose proportion decreases with decreasing FOV Dose proportion increases with decreasing FOV Example of use of field restriction to reduce dose Orthophos XG 3D 17
18 Other dose associated technical factors Pulsed x-ray source Scintillator coating Cesium Iodide Gadolinium Oxy-bromide Detector design Image intensifier / Sphere Flat panel / Cylinder Resolution? Spatial Resolution and Dose NewTom VGi 0.3 vs 0.15 voxel size 8 cm x 8 cm Standard Resolution High Resolution 8 cm x 12 cm ICRP Effective Dose in µsv Average of 70% reduction in dose from use of the standard resolution option 18
19 Additional dose associated technical factors kvp Added filtration Kodak D Effect of added filtration and increased kv Effective dose ICRP 2007 µsv Large FOV (18 x 21 cm) pre-production configuration added filtration configuration Small adult 93 % reduction in dose Medium adult % Large adult % Medium FOV (9 x 15 cm) Small adult % Medium adult % Large adult 166 avg reduction 43% 19
20 The most significant dose associated factor patient selection criteria ADA/FDA Selection Criteria THE SELECTION OF PATIENTS FOR DENTAL RADIOGRAPHIC EXAMINATIONS Originally developed 1987 Most recently revised topics_radiography_examinations.pdf 20
21 USE OF CONE-BEAM COMPUTED TOMOGRAPHY IN ENDODONTICS Joint Position Statement of the American Association of Endodontists and the American Academy of Oral and Maxillofacial Radiology What patients are most likely to benefit? Difficult diagnosis Equivocal signs / symptoms Superimposed structure Internal / external resorbtion Unusual morphology Root or canal numbers Root curvature Intraoperative complication Refractory to conventional treatment Pre-surgical Proximity and relationship to nerve canal or sinus Pathology of non-endodontic origin suspected Joint Position Statement of the American Association of Orthodontists and the American Academy of Oral and Maxillofacial Radiology THE USE OF RADIOLOGY RADIOGRAPHIC EXAMINATIONS IN ORTHODONTICS currently under development children may be two to ten times or more sensitive to radiation carcinogenesis than mature adults* *Smith-Bindman R, et al. Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med
22 Risk differences due to age Children may be 2 to 10 times more sensitive to radiation than adults Brenner D NEJM Risk vs Benefit RISK Risks Dollars Dose ALARA Benefits? 22
23 Explaining Risk to patients Do Provide an estimate of cancer risk (this should be adjusted for children) Compare with Ubiquitous Background Dose Compare with alternative exam equivalence (pan or FMX) Compare with Commonly encountered risks of life Don t Say it s nothing, it s unimportant, or similar dismissive statements. Use analogy of a day at the beach Comparable Risk Table Situation of a one in million risk of dying Risk Quantity Nature Life Living in stone building 2 Months Natural Radioactivity Living in Denver, CO 2 weeks Cosmic Radiation Travel Canoe 6 minutes Accident Bicycle 10 miles Accident Car 300 miles Accident Airplane 1000 miles Accident Airplane 6000 miles Cosmic Radiation Work Typical Factory 10 days Accident Miscellaneous Smoking 1.4 cigarettes Cardiovascular Disease, Cancer Wine 500 cc Cirrhosis 23
24 Explaining Benefit to the patients and parents Accurate diagnosis = Reduced cost reduced time reduced discomfort Better outcomes Fewer complications Risk Example The effective dose from a Kodak 9000 medium adult panoramic scan is about 15 µsv The effective dose from average Kodak x 5 cm jaw sextant is about 21 µsv This dose from these combined examinations is equivalent to about 4 days of average naturally occurring background dose The added risk of cancer from this dose is about 2 in 1,000,000 exposures. Keep in mind the population risk of lifetime fatal cancer is 1 in 5. 24
25 In accordance with the AGD, I declare that I have received expense reimbursement and honoraria from Carestream Dental for dosimetry studies performed on Carestream CBCT units discussed in this presentation and have received expense reimbursement and an honorarium for this talk. Course Code CE unit Presentation slides available at: Approved PACE Program Provider FAGD/MAGD Credit 5/31/2010 to 5/31/
Cone Beam Imaging Why all the Fuss about the X-ray Dose? D.A. Miles BA, DDS, MS, FRCD(C) Dip. ABOM, Dip. ABOMR. Introduction
Cone Beam Imaging Why all the Fuss about the X-ray Dose? D.A. Miles BA, DDS, MS, FRCD(C) Dip. ABOM, Dip. ABOMR Introduction The newest imaging modality to become available to dentists is making its impact
3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?
Disclosures Understanding IAC CT Accreditation Serve as one of two AAPM representatives to IAC CT Board of Directors, Serve as IAC representative to AAPM for focus group on accreditations Provide 3-hour
What Parents Should Know about the Safety of Dental Radiology.
What Parents Should Know about the Safety of Dental Radiology. There are many different types of x-ray images (pictures) that can be taken of children in the dental office to assist in diagnosis. These
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
Practical exercise: Effective dose estimate in CT
Practical exercise: Effective dose estimate in CT TRAINING COURCE PROGRAM 19 20 May 2011, Sofia, Bulgaria Virginia Tsapaki Medical Physics Dpt Konstantopoulio General Hospital email: [email protected]
Effective dose from cone beam CT examinations in dentistry
The British Journal of Radiology, 82 (2009), 35 40 Effective dose from cone beam CT examinations in dentistry 1 J A ROBERTS, MSc, 2 N A DRAGE, BDS, FDS, DDRRCR, 3 J DAVIES, BDS, MFDS, RCS(Ed) and 1 D W
cbct feature September 2011 n orthotown.com
by Sean Carlson, DMD, MS; John Graham, DDS, MD; William E. Harrell Jr., DMD; Edward Y. Lin, DDS, MS; Aaron Molen, DDS, MS and Wm. Randol Womack, DDS A general consensus in dentistry exists, at the present
Assessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
DENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
Dolphin 3D Imaging 11.8
Dolphin 3D Imaging 11.8 The Dolphin 3D software module is a powerful tool that makes processing 3D data extremely simple, enabling dental specialists from a wide variety of disciplines to diagnose, plan
Digital Imaging & Communications in Dentistry: DICOM and Interoperability
2 nd International Congress, Budapest, 2005 Digital Imaging & Communications in Dentistry: DICOM and Interoperability Allan G. Farman ADA Representative to the DICOM Standard Committee Co-Chair Chair DICOM
Veraviewepocs 3D R100 & F40
Veraviewepocs 3D R100 & F40 Innovative 3D Reuleaux Full Arch FOV Thinking ahead. Focused on life. Veraviewepocs 3D R100 A New Frontier in X-ray Diagnostics Veraviewepocs 3D R100 has changed the shape of
I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit
I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
ENGLISH. Planmeca ProMax 3D Mid
ENGLISH Planmeca ProMax 3D Mid Genuine all-in-one unit Planmeca ProMax 3D Mid is a genuine all-in-one CBVT (Cone Beam Volumetric Tomography) unit including 3D imaging, digital panoramic, digital cephalometric
Oftentimes, as implant surgeons, we are
CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the
The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.
RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer
Veraviewepocs 3D R100 & F40
Veraviewepocs 3D R100 & F40 Innovative 3D Reuleaux Full Arch FOV Thinking ahead. Focused on life. Veraviewepocs 3D R100 A New Frontier in X-ray Diagnostics Veraviewepocs 3D R100 has changed the shape of
The wait is over CS 8100 3D. 3D imaging is now available for everyone
The wait is over CS 8100 3D 3D imaging is now available for everyone Complexity is no longer the standard We re removing the obstacles that traditionally come with moving to 2D/3D imaging Despite previous
CS 9000. 3D Panoramic Cephalometric. Innovation, in reach
CS 9000 3D Panoramic Cephalometric Innovation, in reach Innovation made simple We believe in innovation. We always have. In fact, our products have consistently distinguished themselves as groundbreaking
X-Rays Benefits and Risks. Techniques that use x-rays
X-Rays Benefits and Risks X-rays are a form of electromagnetic radiation, just like light waves and radiowaves. Because X-rays have higher energy than light waves, they can pass through the body. X-rays
Effective dose and risk assessment from film tomography used for dental implant diagnostics
Effective dose and risk assessment from film tomography used for dental implant diagnostics N.L. Frederiksen, B.W. Benson and T.W. Sokolowski* The Baylor College of Dentistry and "The Sammons Cancer Center
ENGLISH. Planmeca ProMax 3D concept
ENGLISH Planmeca ProMax 3D concept Profound understanding of anatomy Planmeca ProMax 3D concept is an intelligent and multipurpose X-ray unit series designed to obtain complete information on patient anatomy
For a new dimension of a practice s success: The 3D unit family.
CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS ORTHOPHOS XG 3D ORTHOPHOS XG 3D Ceph GALILEOS
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
X-ray (Radiography) - Abdomen
Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board
Radiation Safety Issues for Radiologic Technologists
Radiation Safety Issues for Radiologic Technologists Greg Sackett, M.S., CHP Medical Physicist Radiation Worker Risks? 1 Patient Risks? Acute Effects? Delayed Effects? Patient Questions? Radiation Dose
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
Multi-slice Helical CT Scanning of the Chest
Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.
Veraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life.
Veraview IC5 HD High definition, digital imaging excellence Thinking ahead. Focused on life. Thinking ahead. Focused on life. 2 3 IC5 HD Super High-Speed with High Definition Clarity For dental radiology
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
Users Group Meeting 2011 Northeastern Society of Orthodontists 89th Annual Meeting November 11-14, 2010 Hilton Montréal Bonaventure Montréal, Quebec, Canada 14 Lecture Schedule Saturday continued w Doctor
Endodontics. Colleagues for Excellence. in Endodontics. Published for the Dental Professional Community by the American Association of Endodontists
Endodontics Colleagues for Excellence Winter 2011 Summer 2011 Root Cone Canal Beam-Computed Irrigants and Tomography Disinfectants in Endodontics Published for the Dental Professional Community by the
Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral
Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come
Clinical Considerations for Cone Beam Imaging in Dentistry
Quality Resource MetLife is an ADA CERP Recognized Provider Accepted Program Provider FAGD/MAGD Credit 11/01/08 to 12/31/12 1.0 CE credit contact hour will be awarded for the review of this Quality Resource
Are children more sensitive to radiation than adults?
Are children more sensitive to radiation than adults? By Madan M. Rehani Director of Radiation Protection, European Society of Radiology There is a commonly held belief that children may be two to three
ORGAN SYSTEMS OF THE BODY
ORGAN SYSTEMS OF THE BODY DEFINITIONS AND CONCEPTS A. Organ a structure made up of two or more kinds of tissues organized in such a way that they can together perform a more complex function that can any
The Challenge of CT Dose Records
The Challenge of CT Dose Records Kimberly Applegate, MD, MS, FACR Professor of Radiology and Pediatrics Emory University *financial disclosures: -Springer Textbook contracts -AIM advisory board for patient
X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581
X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers
The professional for your diagnostics.
CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS DIGITAL PANORAMIC X-RAY MADE SIMPLE The professional for your diagnostics. System concept ORTHOPHOS XG 5 completes the next
Name Class Date Laboratory Investigation 24A Chapter 24A: Human Skin
Name Class Date Laboratory Investigation 24A Chapter 24A: Human Skin Human Anatomy & Physiology: Integumentary System You may refer to pages 386-394 in your textbook for a general discussion of the integumentary
Concepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
A Clinician s Guide to Understanding Cone Beam Volumetric Imaging (CBVI)
Earn 4 CE credits This course was written for dentists, dental hygienists, and assistants. A Clinician s Guide to Understanding Cone Beam Volumetric Imaging (CBVI) A Peer-Reviewed Publication Written by
5G - R15.1 - EN ENGLISH. QR srl - Via Silvestrini, 20-37135 Verona Italy Tel. +39 045 8202727-045 583500 [email protected] www.newtom.
5G - R15.1 - EN ENGLISH QR srl - Via Silvestrini, 20-37135 Verona Italy Tel. +39 045 8202727-045 583500 [email protected] www.newtom.it FIRST IN CONE BEAM, ACCURATE IN RESULTS. 360 degree imaging, reduced
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
Radiation safety in dental radiography
Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum
BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY
BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY N. GEERAERT a,b,c, R. KLAUSZ a, S. MULLER a, I. BLOCH c, H. BOSMANS b a GE Healthcare, Buc, France b Department
COMMUNICATING RADIATION RISKS IN PAEDIATRIC IMAGING
COMMUNICATING RADIATION RISKS IN PAEDIATRIC IMAGING Information to support healthcare discussions about benefit and risk Executive summary Executive summary Advances in technologies using ionizing radiation
3D Accuitomo FPD XYZ Slice View Tomograph. Super-High Resolution Images of Region of Interest
3D Accuitomo FPD XYZ Slice View Tomograph. Super-High Resolution Images of Region of Interest Thinking ahead. Focused on life. Thinking ahead. Focused on life. The new era of user-friendly, 3-dimensional
Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose!
Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Dale A. Miles BA, DDS, MS, FRCD(C) Diplomate, American Board of Oral and Maxillofacial Radiology It used to be that panoramic images
Environmental Radiation Risk Assessment
Environmental Radiation Risk Assessment Jerome Puskin, PhD Center for Science & Risk Assessment Radiation Protection Division Office of Radiation and Indoor Air (ORIA) 2 Outline 1. Ionizing radiation definitions,
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Radiation and Pregnancy
Radiation and Pregnancy General Special attention must be paid to the pregnant, potentially pregnant, or breast feeding patient. The developing embryo or fetus is particularly sensitive to radiation. If
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
University of Colorado Radiology Dose-Risk Smartcard
University of Colorado Radiology Dose-Risk Smartcard Estimates of cancer risk from the low-dose radiation exposures in this card come from ICRP-103 and BEIR-VII (see references). These studies assume a
PATIENT ORGANS DOSE CALCULATIONS IN NUCLEAR MEDICINE
www.arpapress.com/volumes/vol11issue1/ijrras_11_1_17.pdf PATIENT ORGANS DOSE CALCULATIONS IN NUCLEAR MEDICINE Nadia Helal Radiation Safety Dep. NCNSRC., Atomic Energy Authority, 3, Ahmed El Zomor St.,
SOURCES AND EFFECTS OF IONIZING RADIATION
SOURCES AND EFFECTS OF IONIZING RADIATION United Nations Scientific Committee on the Effects of Atomic Radiation UNSCEAR 2008 Report to the General Assembly with Scientific Annexes VOLUME I UNITED NATIONS
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination
Digital panoramic X-ray for practical diagnostics
ORTHOPHOS XG 3 instruments treatment centers Imaging systems cad/cam systems ORTHOPHOS XG 3 Digital panoramic X-ray for practical diagnostics System concept Standard panoramic X-ray with proven technology.
Multislice computed tomography (CT) provides
Comparison of the Accuracy of Cone Beam Computed Tomography and Medical Computed Tomography: Implications for Clinical Diagnostics with Guided Surgery Marcus Abboud, DMD, PhD 1 /José Luis Calvo Guirado,
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
Endoscope Optics. Chapter 8. 8.1 Introduction
Chapter 8 Endoscope Optics Endoscopes are used to observe otherwise inaccessible areas within the human body either noninvasively or minimally invasively. Endoscopes have unparalleled ability to visualize
Radiation Safety Characteristics of the NOMAD Portable X-ray System
Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality
Radiation Protection Update
Radiation Protection Update Pennsylvania Dental Association Revised December 2013 Reviewed by the Pennsylvania Department of Environmental Protection Radiation Protection Update Radiation in the Healing
Energy Employees Occupational Illness Compensation Program (EEOICPA)
Energy Employees Occupational Illness Compensation Program (EEOICPA) Joseph Falco, MD MPH Manager, Occupational Medicine Clinic BNL Point-of-Contact EEOICPA Program November 10, 2011 Energy Employees Occupational
Low-dose CT Imaging. Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital
Lung Cancer Screening with Low-dose CT Imaging Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital Despite recent declines in the incidence of lung cancer related to the
Monitoring Patient Radiation Dose in VA. Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration
Monitoring Patient Radiation Dose in VA Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration TWENTY ONE/NOVEMBER 2011 Computed Tomography (CT), Magnetic Resonance
What is Cone-Beam CT and How Does it Work?
Dent Clin N Am 52 (2008) 707 730 What is Cone-Beam CT and How Does it Work? William C. Scarfe, BDS, FRACDS, MS a, *, Allan G. Farman, BDS, PhD, DSc, MBA b a Department of Surgical/Hospital Dentistry, University
Cone Beam Implant Planning Manual
Cone Beam Implant Planning Manual (Everything you want to know about CBCT planning) Page 2 Overview & Guide Selection Criteria Page 3 Scan Appliance basics When are they required? Page 4 Scan Appliance
Patient Prep Information
Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center
Kidney Cancer OVERVIEW
Kidney Cancer OVERVIEW Kidney cancer is the third most common genitourinary cancer in adults. There are approximately 54,000 new cancer cases each year in the United States, and the incidence of kidney
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
Dose conversion factors for radon
Dose conversion factors for radon James Marsh, John Harrison, Margot Tirmarche, Dominique Laurier, Eric Blanchardon and François Paquet HEIR, New Mexico, May 2009 ICRP C1-task group 64 Headed by Margot
CT Dose to Patients. CT Dose Reporting Requirements of CA Senate Bill 1237. Sources of Ionizing Radiation Exposure (then) 5/3/2011
CT Dose to Patients CT Dose Reporting Requirements of CA Senate Bill 1237 In U.S., CT comprises only 11% of all exams but generates 67% of total diagnostic dose» Mettler 2000 Melissa C. Martin, M.S., FACMP,
Production of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
Patient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University
for breast cancer detection
Microwave imaging for breast cancer detection Sara Salvador Summary Epidemiology of breast cancer The early diagnosis: State of the art Microwave imaging: The signal processing algorithm 2D/3D models and
Introduction to Dental Anatomy
Introduction to Dental Anatomy Vickie P. Overman, RDH, MEd Continuing Education Units: N/A This continuing education course is intended for dental students and dental hygiene students. Maintaining the
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
SARCOIDOSIS. Signs and symptoms associated with specific organ involvement can include the following:
SARCOIDOSIS Sarcoidosis is a disease that occurs when areas of inflammation develop in different organs of the body. Very small clusters of inflammation, called granulomas, are seen with sarcoidosis. They
Purchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage
Treating Thyroid Cancer using I-131 Maximum Tolerable Dose Method
Treating Thyroid Cancer using I-131 Maximum Tolerable Dose Method Christopher Martel, M.Sc., CHP Lisa Thornhill,, NRRPT, RT(NM) Boston University Medical Center Thyroid Carcinoma New cases and deaths in
6 RADIATION DOSE, RADIATION PROTECTION AND THE IONISING RADIATIONS REGULATIONS
6 RADIATION DOSE, RADIATION PROTECTION AND THE IONISING RADIATIONS REGULATIONS Overview This chapter discusses the issues of radiation dose, radiation protection and the relevant aspects of United Kingdom
Radiation Protection Series
Radiation The Radiation is published by the Australian Radiation and Nuclear Safety Agency (ARPANSA) to promote practices which protect human health and the environment from the possible harmful effects
P R E S E N T S Dr. Mufa T. Ghadiali is skilled in all aspects of General Surgery. His General Surgery Services include: General Surgery Advanced Laparoscopic Surgery Surgical Oncology Gastrointestinal
IONISING RADIATION. X-rays: benefit and risk
IONISING RADIATION X-rays: benefit and risk Impress Federal Office for Radiation Protection Postfach 10 01 49 D - 38201 Salzgitter Telephone: + 49 (0) 30 18333-0 Fax: + 49 (0) 30 18333-1885 Website: www.bfs.de
Comparison of radiation dose from X-ray, CT, and PET/ CT in paediatric patients with neuroblastoma using a dose monitoring program
Comparison of radiation dose from X-ray, CT, and PET/ CT in paediatric patients with neuroblastoma using a dose monitoring program Poster No.: C-0591 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific
Diagnostic Exposure Tracking in the Medical Record
Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand
Radiation safety in dental radiography
Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.