Pulsed Discharge Detector Models D-3-I-HP and D-3-I-7890 Instruction Manual
|
|
|
- Randolph Logan Townsend
- 9 years ago
- Views:
Transcription
1 Valco Instruments Co. Inc. Pulsed Discharge Detector Models D-3-I-HP and D-3-I-7890 Instruction Manual Rev 1/15 North America, South America, and Australia/Oceania contact: Valco Instruments Co. Inc sales tech fax Europe, Asia, and Africa contact:: VICI AG International Schenkon, Switzerland Int Int [email protected] phone fax
2 This page intentionally left blank for printing purposes
3 Table of Contents Introduction Description and Operating Principles... 1 Safety Notes and Information... 2 Symbols Installation Category Safety Components of the Detector System... 3 System Requirements Components not Included with the Detector System... 4 System Purity... 4 Gas Specifications... 5 Installation General Precautions... 6 Checking the Firmware Version Agilent 6890 GC... 6 Agilent 7890 GC... 6 Mounting the Detector... 6 Installing and Configuring the FID Logic Board Agilent 6890 GC... 8 Agilent 7890 GC... 8 Gas Connections...11 Installing and Purging the Gas Regulator...11 Installing and Purging the Helium Purifier...12 Connecting the Discharge Gas to the Detector...12 Initial Power-Up...13 Capillary Column Connection...14 Packed Column Connection...14 Testing for Leaks...15 Mode Selection and Setup Helium Ionization Mode...16 Selective Photoionization Mode...16 Troubleshooting...18 Warranty...20 Detector Performance Log...21
4 This page intentionally left blank for printing purposes
5 Introduction 1 Description and Operating Principle The D3 is a non-radioactive pulsed discharge ionization detector (PDID), with models available for plug and play installation on the Agilent 6890 or 7890 GC. A schematic representation of the detector is shown in Figure 1. The D3 utilizes a stable, low power, pulsed DC discharge in helium as the ionization source. Elutants from the column, flowing counter to the flow of helium from the discharge zone, are ionized by photons from the helium discharge above. Resulting electrons are focused toward the collector electrode by the two bias electrodes. The principal mode of ionization is photoionization by radiation arising from the transition of diatomic helium He 2 (A 1 u + ) to the dissociative 2He(1S 1 ) ground state. This is the well-known Hopfield emission. The photon energy from the He 2 continuum is in the range of 13.5 ev to 17.7 ev. The D3 is essentially non-destructive ( % ionization) and highly sensitive. The response to organic compounds is linear over five orders of magnitude with minimum detectable quantities (MDQs) in the low picogram range. The response to fixed gases is positive (the standing current increases), with MDQs in the low ppb range. Detector response is universal except for neon, which has an ionization potential of ev. Since this potential is close to the energy of the He* metastable (19.8 ev) but greater than the photon energy from the He 2 continuum, neon exhibits a low ionization efficiency and low detector response. When a dopant is added to the discharge gas, the D3 also functions as a selective photoionization detector. (Suitable dopants include Ar for organic compounds, Kr for unsaturated compounds, or Xe for polynuclear aromatics.) GROUND PIN DISCHARGE REGION HELIUM INLET HIGH VOLTAGE SAPPHIRE INSULATORS CAPILLARY COLUMN VENT COLLECTOR ELECTRODE COLUMN INLET Figure 1: Schematic of the D-3 detector
6 Introduction 2 Safety Notes and Information Symbols ATTENTION Refer to the manual. Installation Category This equipment has been designed for installation category (overvoltage category) II, pollution degree 2. It has been approved for use only in heavy industrial environments and may not be used in the residential, commercial, or light-industrial environment. Safety This instrument left the factory in a safe condition. This instruction manual contains important information and warnings which must be followed by the user to insure safe operation and to retain the instrument in a safe condition. Use only with an approved mains supply cord having a rating of 2A, 250V, or greater. Do not use this equipment in a manner not specified herein. CAUTION: During normal operation, the detector produces ultraviolet energy (UVA, UVB), some of which may be emitted. Do not watch the arc without eye protection.
7 Introduction 3 Components of the Detector System Components of the detector system are listed in Table 1 below. Check the contents of the packages to verify that everything is present. Contact the factory if anything is missing or damaged. (NOTE: damaged shipments must remain with the original packaging for freight company inspection.) Description Quantity Product number Detector assembly (includes detector, heater, and electrometer): for Agilent PD-D3-I for Agilent PD-D3-I-7890 Pulser module 1 PD-M2 Pulser power supply 1 I Agilent FID board: for Agilent I-G for Agilent I-G Helium purifier 1 HP2 Miniature helium purifier 1 HPM Fittings kit 1 PD-KIT-D3 Includes: 1/32" polyimide ferrule 5 ZF.5V 1/16" gold-plated ferrule 5 ZF1GP mm polyimide column ferrule 5 FS mm polyimide column ferrule 5 FS.5 1/16" union 2 ZU1 1/16" tee 1 ZT1 1/8" external to 1/16" internal reducer 2 EZR21 1/8" to 1/16" reducing union 2 EZRU21 1/32" external nut 1 EN.5KN Packed column adapter 1 I D3 Restrictor, psi He 1 TGA-R-30F60P Instruction manual 1 MAN-PDD3 Table 1: Components of the D3 system
8 Introduction 4 System Requirements Components Not Included with the Detector System Helium (99.999% purity) and other support gases Ultra high purity grade gas pressure regulator with stainless steel diaphragm Any special adapters required for connection to the gas regulator SS tubing to go from gas supply to GC Flow measuring device System Purity Discharge/Carrier Gas Considerations The performance of the detector is adversely affected by the presence of any impurities in the gas streams (carrier, discharge, or dopant). We recommend that a quality grade of helium 5.0 (99.999% pure or better) be used at all times. Major gas suppliers offer research grade helium ( % pure) which is particularly low in fixed gas impurities and should give good results in a clean system, but even the highest quality carrier gas may contain some water vapor and fixed gas impurities; hence a helium purifier is included as part of the detector system. The discharge gas should always be run through the helium purifier. Whenever a new batch of discharge gas is received, we recommend performing a blank GC analysis of the gas in the PDHID mode to detect and identify the presence of any impurities. Gas purity requirements are specified on the next page. Tubing Standards of cleanliness that are suitable for many GC applications may be totally inadequate for the sensitive PDHID/PDPID work. All surfaces that contact the gas stream must be glass or stainless steel. Do not use copper tubing or brass fittings. All tubes must be thoroughly cleaned and baked before use. Flow Controllers The use of valves or flow controllers in which the gas stream is exposed to any polymer-based packing or lubricating material is to be particularly avoided. Pressure Regulators We recommend commercial ultra-pure grade regulators with stainless steel diaphragms. Regulators with diaphragms made of neoprene or other elastomers should never be used.
9 Introduction 5 Gas Specifications Detector Mode PDHID Ar-PDPID Kr-PDPID Xe-PDPID Discharge gas Helium 2% Ar in He 1.5% Kr in He 0.8% Xe in He Carrier gas H elium * ** ** ** Any gas including He which has an ionization potential greater than 12 ev ** Any gas including He which has an ionization potential greater than 11 ev Purity Specifications Helium (discharge and carrier gas) must have a minimum purity of %, with < 20 ppm Ne impurity. For trace analysis of fixed gases, we strongly recomment % purity He with < 0.5 ppm Ne. Ar-PDPID mode: 2% ± 0.2% Ar in % He balance Kr-PDPID mode: 1.5% ± 0.1% Kr in % He balance Xe-PDPID mode: 0.8% ± 0.2% Xe in % He balance
10 6 Installation General Precautions Do not use plastic/polymer or copper tubes for gas handling and interconnectons. Use only stainless steel tubing with Valco gold-plated ferrules. Do not turn the discharge power on until the helium discharge gas is flowing through the detector. Do not shut off or disconnect the discharge gas when the detector is hot, even if the GC is turned off. Do not turn on the GC during detector installation. Checking the Firmware Version Agilent 6890 GC 1. Push <Options> and scroll down to select Diagnostics. Press <Enter>. 2. Select Instrument status, and press <Enter>. 3. Scroll down to read Version A.0x0x. Successful operation of the D3-I-HP requires either 6890 firmware version A or higher or the GC s FID pneumatic control (EPC). If the EPC is used, all it does is provide the detector title, which remains FID. Agilent 7890 GC How do they know if they have a problem related to firmware version? Will they get some message about incompatibility? When the GC is first powered on, the display will read: Agilent 7890 GC A.0x.0x Power on successful If the instrument is already on and displaying other information, push <Status> and then <Clear> to display the message shown above. The second line is the firmware version. Successful operation of the D3 requires 7890 firmware version A or higher. If you experience problems, contact Agilent technical support and request a firmware version compatible with Channel Partner detectors. If you are running ChemStation, please check with Agilent to determine that your version is compatible with Channel Partner detectors. A software upgrade may be required. Mounting the Detector 1. Place the D3 on the top of the GC in either the front or back detector position (Figure 2), and tighten the four captive Torx T-20 screws in the detector pallet. (Figure 3) 2. Connect the high voltage cable to the pulser module. The electrometer and heater cables will be connected after the FID logic board is installed.
11 Installation 7 MODEL D3 DETECTOR Agilent Technologies 7890A GC System Figure 2: A D3 detector in the Back position on a 7890A CAPTIVE SCREWS (4) DISCHARGE GAS INLET (30 ml/min@60 psi) VENT HEATER/SENSOR CABLE TO D3 PULSER MODULE (PD-M2) COLUMN INLET HIGH VOLTAGE CABLE ELECROMETER RIBBON CABLE CAPILLARY COLUMN ADAPTER (Packed column adapter shown on page10) Figure 3: Mounting and connecting the detector TO FID LOGIC BOARD
12 Installation 8 CAUTION: A D-3-I-HP will not work on a 7890 GC, and a D-3-I-7890 will not work on a If you are not sure which model you have, call the VICI detector department with the detector serial number. Installing and Configuring the FID Logic Board Agilent 6890 GC (units without EPC) The D3 uses the electrometer portion of the Agilent FID logic board, which must be configured to ignore the ignitor circuit. 1. Plug the FID logic board into the GC main board at the location corresponding to the detector mounting position (front or rear) which you intend for the D3. (Refer to the GC manual as required.) Tighten the detector card screw. (Figure 4) 2. Connect the heater/sensor cable to the main board and the electrometer cable to the interface board. 3. Push <Config> and scroll down to select Instrument. Press <Enter>. 4. Select F det: FID (for front position) or R det: FID (for rear position) and push <.><.><Mode/Type>. 5. Select Electrometer and press <Enter>. To switch back to standard FID mode, repeat the procedure and slect FID on the last step. Agilent 7890 GC Figure 4: 6890 FID board 1. Refer to the GC manual for instructions on removing access panels. Install the FID logic board as illustrated in Figure 5, at the location corresponding to the detector mounting position (front or back). Secure the thumb screw at the top right corner. 2. Connect the heater/sensor cable and the electrometer ribbon cable to the FID logic board as shown in Figure Connect the appropriate F-DET (front) or B-DET (back) cable to the logic board. The AUX DET1 cable is not be used with the VICI D3 detector. 4. Press <Options>, and use the scroll keys to select Keyboard & Display and press <Enter>. 5. Scroll to select Hard Configuration Lock. Set it to OFF by pressing the <Off/No> key. HEATER/SENSOR CABLE CONNECTOR FID BOARD SCREW FID LOGIC BOARD ELECTROMETER CABLE CONNECTOR
13 B-DET Installation 9 MODEL D3 IN BACK POSITION FRONT OF 7890 HEATER CABLE CONNECTIONS THUMBSCREW FID LOGIC BOARD FOR FRONT POSITION ELECTROMETER CABLE CONNECTIONS FID LOGIC BOARD FOR BACK POSITION F-DET CABLE CONNECTION B-DET CABLE CONNECTION F-DET Figure 5: 7890 FID logic board 6. Press the <Config> key, then press <Back det> or <Front Det>, corresponding to the location where the PDD will be installed. For this example we will select Back Det. The display will read: Configured (or Unconfigured) Ignore ready = False [B-DET] = (Signal) (FID) 7. Press <Mode/Type>. The display will read: CONFIGURE BACK DETECTOR Remove Module < (cursor is here) 8. Press <Enter> to clear the existing configuration. The display will read: CAUTION: Instrument power must be turned off and back on for set point to take effect. 9. Do no turn off the GC; instead press <Enter>. The display will read: CONFIGURE BACK DETECTOR Unconfigured: < (cursor is here) [B-Det] = (Signal) (FID) [EPC4] Not Found [A-Det2] Not Found 10. Press <Mode/Type>. The display will read: CONFIGURE BACK DETECTOR CPDET FID, No Htr, No EPC < (cursor is here) CPDET FID, Htr, No EPC
14 Installation Scroll to select the second line and press <Enter>. The display will read: CAUTION: Instrument power must be turned off and back on for set point to take effect. 12. Do not turn off the GC; instead press <Enter>. The display should read: CONFIGURE BACK DETECTOR Configured: Channel Partner Ignore Ready = False [B-det] = (Signal) (FID) 13. Turn off the GC and power it back up. Verify the configuration by pressing <Back Det>. The display should read: Back detector (CP) Temperature xx..x Output xxxx
15 Installation 11 Gas Connections Remember these three points discussed earlier: (1) all surfaces that contact the gas stream must be glass or stainless steel; (2) do not use copper tubing or brass fittings; and (3) all tubes must be thoroughly cleaned and baked before use. The installation instructions below assume that the detector discharge gas will be supplied from a nearby cylinder of helium of the proper purity. If your installation is different, you may need to modify the instructions appropriately. A number of Valco fittings have been supplied in the fittings kit to handle different situations. Since the distance from the helium supply to the GC varies from installation to installation, we do not supply tubing to go from that point to the GC. Figure 6 illustrates gas connections for the PD-D3-I detector system. Since the distance from the helium supply to the GC varies from installation to installation, we do not supply tubing for that purpose. VENT COLUMN INLET TGA-R-30F60P RESTRICTOR (30 ml/min minimum) DISCHARGE GAS INLET TEE (ZT1) EPC HPM HELIUM PURIFIER DISCHARGE GAS (99.999% He) ELECTRONIC PNEUMATIC CONTROL VICI MINIATURE HELIUM PURIFER COLUMN INJECTOR GAS CHROMATOGRAPH Figure 6: Gas connections for a PD-D3-I system Installing and Purging the Gas Regulator 1. Make sure the on/off valve on the helium cylinder is completely closed. Screw the CGA fitting nut of the regulator into the helium cylinder. Go beyond finger-tight, but do not tighten the nut all the way some leakage is required for the purging operation. 2. Turn the output pressure regulating knob completely counterclockwise. 3. Open the cylinder on/off valve slightly and quickly close it again. 4. Adjust the tightness of the regulator connecting nut to allow a pressure reduction of ~690 kpa/sec (100 psi/sec). With a new bottle, the gauge should start out at about 14 MPa (2000 psi). 5. When the pressure drops into the MPa ( psi) range, open the cylinder on/off valve slightly and quickly close it again. 6. Repeat Step 5 eight or ten times to be certain that all the air is purged. On the final purge, tighten the regulator connecting nut very securely as the pressure approaches the MPa ( psi) range.
16 Installation Open the cylinder valve to pressurize the regulator once again. Close the valve and observe the needle of the high pressure gauge for 15 minutes. If it doesn t move, there is no critical leak on the high pressure side of the regulator. CAUTION: Never use leak detecting fluids on any part of this system. Installing and Purging the Helium Purifier 1. If the pressure regulator has a 1/8" male cone-type outlet port, install the Valco 1/8" external to 1/16" internal reducer (EZR21); if it has a 1/4" male cone-type outlet port, install the Valco 1/4" external to 1/16" internal reducer (EZR41). For other regulator outlet fittings, a wide variety of Valco adapters is available. 2. Remove the cap from the inlet tube of the Valco helium purifier and insert the tube fitting into the 1/16" reducer port. (Keep the outlet tube capped.) Use a 1/4" wrench to turn the nut one-quarter turn past the point where the ferrule first starts to grab the tubing. Do not remove the fitting. When made up properly, it should be leak-tight. 3. Turn the output pressure regulating knob clockwise until the gauge registers 345 KPA (50 psi). 4. Allow five minutes for equilibration, then turn the regulating knob all the way counterclockwise. 5. Observe the needle of the output pressure gauge for 15 minutes. There will be a slight initial drop, but if it doesn t move after that, consider that all the connections are tight. 6. If necessary, use an electronic leak detector to locate any leaks. If a leak detector is not available, tighten all the fittings (including the output pressure guage), and repressurize the system for another test. CAUTION: Never use leak detecting fluids on any part of this system. 7. Upcap the outlet tube of the purifier and purge the system for 15 to 30 minutes at ml/min to eliminate air from the purifier getter material. Connecting the Discharge Gas to the Detector 1. Connect the discharge gas source to restrictor TGA-R-30F60P, and connect the restrictor to the detector inlet. 2. Connect a flow measuring device to the vent and adjust the helium pressure to obtain a flow of ~30 ml/min.
17 Installation 13 Initial Power-Up CAUTION: Always make sure that discharge gas is flowing before heating and powering up the detector. 1. Before installing the column, set the gas flow to 30 ml/min (measured at the detector vent). Let it flow for 15 minutes so that all air is purged from the helium purifiers. 2. Plug in the helium purifiers and turn on the GC. 3. Set the detector temperature to 100 C and allow time for the detector and helium purifiers to reach the set temperature. 4. Plug the 24 VDC output from the power supply (I ) into the pulser module (PD-M2). 5. Plug the power supply cord to a VAC source, and look through the hole in the detector housing to see if the discharge is on. CAUTION: During normal operation, the detector produces ultraviolet energy (UVA, UVB), some of which may be emitted. Do not watch the arc without eye protection. 6. Check the standing/background current, and record it in the Detector Performance Log on the last page of this manual. Optimum range is pa at 100 C. Lower current indicates a clean, leak-free system. 7. The recommended detector temperature is 20 C above the column temperature, with a minimum of 100 C. Set the detector to the operating temperature required for the intended analysis. When the detector has reached the set temperature, read and record the standing current. 8. Install the column as illustrated on page 11, leaving the oven at ambient temperature. Start carrier flow, then read and record the standing current. The difference between this reading and the one previous is the ionization of the combined impurities in and eluting with the carrier gas. The smaller the difference, the better the quality of the gas exiting the column. 9. Set the column oven to the temperature required for the intended analysis. When the oven reaches the set temperature, read and record the standing current. The difference with the previous reading is the ionization of the column bleed. The smaller the difference, the better the column is conditioned. NOTE: Some stationary phases will have a higher bleed than others, but are still suitable for this detector. However, the lower the bleed, the lower the chances of contaminating the detector cell. From this point, the standing current should be observed and logged after any system change. In addition, logging the standing current (with and without the column) on a regular basis is an effective monitor of system integrity (leaktightness and cleanliness). We also recommend tracking the internal standard (quantity on column/area count) for sensitivity continuity.
18 Installation 14 Capillary Column Connection 10.3 cm COLUMN INLET CAPILLARY COLUMN ADAPTER (IZERA1.5) COLUMN FERRULE NUT CAPILLARY COLUMN If the capillary column adapter is installed in the column inlet: 1. Make a mark on the column 10.3 cm from the end. 2. Remove the knurled nut and plug from the capillary column adapter in the column inlet at the bottom of the detector. (Figure 3) Slide the nut over the end of the column, followed by the appropriate column ferrule (FS.4 or FS.5, or ZF.5V for megabore). 3. Seat the ferrule in the detail of the column adapter and begin sliding the column through the capillary column adapter and into the column inlet. 4. Get the nut started on the threads and tighten it until you feel it contact the ferrule, then back off half a turn. Slide the column into the column inlet until the mark is flush with the surface of the knurled nut, and secure the column in the adapter by tightening the knurled nut finger tight only. If the capillary column adapter has been removed, reinstall it: 1. Unscrew the liner as far as it will go, then screw the fitting body into the column inlet fingertight. 2. While using a 1/8" wrench to prevent rotation of the liner (the part with the seat for the column ferrule), use a 1/4" wrench to tighten the body of the adapter until the ferrule has sealed. The liner will deform if it rotates. 3. Proceed to Step 1 above. Packed Column Connection COLUMN INLET To prevent detector contamination, we strongly recommend disconnecting the column from the detector during column bakeout procedures. 8.7 cm PACKED COLUMN PACKED COLUMN ADAPTER (I D3) The D3 is optimized for packed columns. The column tubing must be thoroughly cleaned and baked before the column is packed. Even when the best care is taken in column tubing cleaning and in the support and stationary phase selection, a new column will often bleed compounds, resulting in a considerable increase in the detector baseline. This initial bleed will usually be reduced to acceptable levels after the column is conditioned with clean carrier gas flow for several hours at the recommended bakeout temperature. 1. Loosen and remove the knurled nut and plug of the capillary column adapter, (or remove the column ferrule and the column if one has been installed). 2. Use a 1/8" wrench to hold the liner that part of the adapter in which the column ferrule sits. While the 1/8" wrench keeps the liner from rotating, use a 1/4" wrench on the fitting body to loosen the adapter 1/2 turn.
19 Installation Set aside the 1/8" wrench and completely remove the adapter from the column inlet. 4. Screw the packed column adapter into the column inlet by hand. Exercise caution, as the tip of the adapter is very fragile. Then tighten the adapter with a 1/4" wrench, using an additional wrench on the flats of the column inlet to support the detector. 5. Connect the 1/8" column to the packed column adapter with the EZRU21 reducing union supplied in the fittings kit. Testing for Leaks It is critical for the system to be leak-tight, and an additional check at this point can save many headaches later on. To test for leaks: 1. Cap the tube and pressurize the entire system with helium to 138 kpa (20 psi). 2. If the system does not hold pressure, check all the fittings with an electronic helium leak detector. DO NOT use leak detecting liquids. 3. Tighten fittings as required.
20 16 Mode Selection and Setup Helium Ionization Mode Since the PDHID mode provides a better indication of the cleanliness and the integrity (leak tightness) of the system, that mode is utilized for initial testing and startup. If the system is operating according to the parameters described thus far, it is ready for operation in the PDHID mode. Selective Photoionization Mode To maximize detector lifetime, turn off the discharge power when the GC is not actually analyzing samples. This is especially important in the Ar/Kr PID mode. Since the pulsed discharge detector is essentially a windowless helium photoionization detector, changing the discharge gas from pure helium to helium doped with argon, krypton, or xenon changes the discharge emission profile. This results in a change in the photon energy due to additional resonance atomic emissions and diatomic emissions from the rare gas added. Thus a single detector can be operated in any of the three photoionization detector (PID) modes: Ar-, Kr-, or Xe-PID. Doped helium is used rather than other pure gases in order to retain the benefits of the helium: namely, its transparency for Ar, Kr, an Xe resonance radiation and its efficient cooling of the electrodes. Any problems associated with the presence of a window between the photon source and the ionization chamber are eliminated. In most applications involving current commercial PIDs, analyte condensation and decomposition on the window attenuate the lamp energy, necessitating frequent cleaning and recalibration. Custom gas blends for the pulsed discharge detector are available from leading gas suppliers at special prices. Alternatively, they may be formulated on the spot by using appropriate fixed restrictors to mix appropriate amounts of pure helium and pure dopant through a tee. Since all gas streams must pass through a Valco purifier, the second option requires an additional purifier for each dopant. This may still be more cost effective than requesting a custom blend of the more expensive Kr or Xe; since the typical flow rate required for the pure dopant rare gas is about ml/min, a small lecture bottle can last for a long time. In either case, the total discharge gas flow rate should be the same as specified in Connecting the Discharge Gas to the Detector on page 12.
21 Mode Selection and Setup 17 Ar-PDPID Changing the discharge gas from helium to a mixture of 2% argon in helium changes the photon energy level from the ev range to the ev range. The argon emission consists of resonance radiation at 11.8 ev and 11.6 ev and the diatomic Ar 2 emission in the range of ev. Except for fixed gases and a few organic compounds like CH 4 (IP = 12.5 ev), CH 3 CN (IP = 12.2 ev) and some fluro-chloro hydrocarbons, the majority of organic compounds have ionization potentials lower than 11.8 ev. Thus the Ar-PDPID is nearly universal, like the flame ionization detector, but without the risks associated with the presence of an open flame and hydrogen. Kr-PDPID The recommended proportion is 1.4% Kr in He as the discharge gas. The krypton emission consists principally of resonance lines at 10.6 ev and 10.1 ev. The Kr-PDPID can detect compounds with IP < 10.6 ev, which includes unsaturated and cyclic hydrocarbons, alcohols, aldehydes, organic acids, esters, etc. Xe-PDPID The recommended proportion is 0.8% Xe in He as the discharge gas. The xenon emission consists principally of resonance lines at 9.6 ev and 8.4 ev, and can detect compounds with IP < 9.6, like aromatics, ethers, alcohols, aldehydes, etc. In addition to the specific compounds named in the three paragraphs above, certain important inorganic compounds like ammonia, hydrogen peroxide, arsenic trichloride, hydrogen sulfide, arsine, phosphine, nitric oxide, carbon disulfide etc. can be selectively detected using the appropriate photoionization mode. Each dopant gas requires an additional helium purifier, which must be purged and conditioned in the same manner as the purifier installed on the discharge gas supply. If you are using more than one dopant, we recommend use of a Valco multiposition stream selection valve so that no fittings have to be disconnected. Not only is this convenient, it keeps the system closed, minimizing chances of contamination. When changing from one dopant to another, allow at least one hour for the old gas to be purged from the system.
22 18 Troubleshooting High Background Current Check discharge color Pink Purple/ blue Disconnect the column and cap the column inlet Do a hydrogen leak test (described on next page) Baseline current falls to normal range (< 2000 pa) Baseline current is still high (> 2000 pa) Leaks in the fittings of the PDD plumbing Leaks inside the detector cell itself No detectable leaks. The detector isn t leaking. Check for column bleed and/or leaks in the GC setup. Tighten leaking fittings Consult Valco Possible bad carrier gas or empty He bottle. Install a new bottle. Baseline current falls to normal range (< 2000 pa) Baseline current is still high (> 2000 pa) Baseline current falls to normal range (< 2000 pa) Baseline current is still high (> 2000 pa) Baseline current falls to normal range (< 2000 pa) Baseline current is still high (> 2000 pa) Consult Valco Consult Valco Consult Valco Low Sensitivity Check background current with column removed and column inlet plugged Sensitivity is in acceptable range High background current (> 2000 pa) Refer to Troubleshooting: High Background Current Normal background current ( 2000 pa) Reinstall column and check background current Background current increases Background current remains in normal range Check column inlet for leaks. Tighten fittings. Check column position according to drawing on p. 14, and run a sample. Background current remains high Background current drops to normal range Sensitivity is in acceptable range Sensitivity is still too low Remove column and condition it, then run a sample Run a sample Remove column and bake detector at 300 C for four hours. Install column and run a sample. Sensitivity is still too low Sensitivity is in acceptable range Sensitivity is still too low Sensitivity is in acceptable range Sensitivity is still too low Consult Valco No Peaks If the background current is stable but there are no peaks: 1. Check column flow. 2. Check the column insertion position (10.3 cm for capillary columns, 8.7 cm for packed columns). 3. If there are still no peaks, check to make sure that the GC is actually making an injection. 4. If there are still no peaks, consult Valco.
23 Troubleshooting 19 High Noise Level If the noise lievel is high: 1. See if it improves with the GC fan turned off. 2. Check the column insertion position (10.3 cm for capillary columns, 8.7 cm for packed columns.) 3. If the noise level is still high, do a hydrogen leak test, described below. 4. If there are no leaks found, or leaks are found and repaired and the noise level is still high, consult Valco. Performing a Hydrogen Leak Test A hydrogen leak test allows you to pinpoint loose fittings and leaks inside the detector. 1. Connect a length of plastic tubing to a regulated hydrogen bottle. Establish a flow of 5-10 ml/min. 2. Hold the hydrogen outlet tube at a fitting connection for ten seconds while monitoring the baseline. 3. Tighten the fitting and test it again, repeating as necessary until every connection has passed the ten second test. 4. Hold the hydrogen outlet close to the detector. (Do not insert the tube into any detector holes.) If a leak in the detector is indicated, contact Valco.
24 20 Warranty This Limited Warranty gives the Buyer specific legal rights, and a Buyer may also have other rights that vary from state to state. For a period of 365 calendar days from the date of shipment, Valco Instruments Company, Inc. (hereinafter Seller) warrants the goods to be free from defect in material and workmanship to the original purchaser. During the warranty period, Seller agrees to repair or replace defective and/or nonconforming goods or parts without charge for material or labor, or, at the Seller s option, demand return of the goods and tender repayment of the price. Buyer s exclusive remedy is repair or replacement of defective and nonconforming goods, or, at Seller s option, the repayment of the price. Seller excludes and disclaims any liability for lost profits, personal injury, interruption of service, or for consequential incidental or special damages arising out of, resuiting from, or relating in any manner to these goods This Limited Warranty does not cover defects, damage, or nonconformity resulting from abuse, misuse, neglect, lack of reasonable care, modification, or the attachment of improper devices to the goods. This Limited Warranty does not cover expendable items. This warranty is VOID when repairs are performed by a nonauthorized service center or representative. For information about authorized service centers or representatives, write Customer Repairs, Valco Instruments Company, Inc, P.O. Box 55603, Houston, Texas 77255, or phone (713) At Seller s option, repairs or replacements will be made on site or at the factory. If repairs or replacements are to be made at the factory, Buyer shall return the goods prepaid and bear all the risks of loss until delivered to the factory. If Seller returns the goods, they will be delivered prepaid and Seller will bear all risks of loss until delivery to Buyer. Buyer and Seller agree that this Limited Warranty shall be governed by and construed in accordance with the laws of the State of Texas. The warranties contained in this agreement are in lieu of all other warranties expressed or implied, including the warranties of merchantability and fitness for a particular purpose. This Limited Warranty supercedes all prior proposals or representations oral or written and constitutes the entire understanding regarding the warranties made by Seller to Buyer. This Limited Warranty may not be expanded or modified except in writing signed by the parties hereto.
25 Detector Performance Log 21 In addition to the occasions indicated in the Comments area of the table below (see Initial Power-Up, page 13), the standing current should be observed and logged after any system change. Logging the standing current (with and without the column) on a regular basis is also an effective monitor of system integrity (leaktightness and cleanliness). To check sensitivity continuity, we recommend tracking the internal standard (quantity on column/area count). Additional log pages can be downloaded from the support/manuals section of Detector Model: D-3-I- Serial Number: Date of purchase: Date Operator Comments Column temp ( C) Detector temp ( C) Noise level Sample Standing current (pa) Initial power-up (IPU) 100 IPU, detector to analysis temp IPU, column installed Ambient IPU, column to analysis temp
The Electron Capture Detector (ECD)
Theory of operation The ECD is based on the phenomenon that electronegative species can react with thermal electrons present to form negatively charged ions. The loss of such electrons is related to the
23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
RGC-IR Remote Gas Calibrator for IR400
Remote Gas Calibrator for IR400 The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by
Operator s Manual. Pressure Injection Cell Model PC8500. Congratulations! Contents
Operator s Manual Pressure Injection Cell Model PC8500 Congratulations! Congratulations on your purchase of a Next Advance Pressure Injection Cell. Please read this operator s manual which explains proper
USE &CARE GUIDE. Remote Faucet Pump System. See Important Safeguards on page 2
Remote Faucet Pump System USE &CARE GUIDE See Important Safeguards on page 2 An exclamation point within an equilateral triangle is intended to alert user to the presence of important operating and maintenance
Micro GC Fusion Pre-Installation Checklist
Site Preparation Proper site preparation prior to installing is very important. Contact INFICON (www.inficon.com) with any questions regarding site preparation. NOTE: Checkboxes ( ) identify hardware that
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
MODEL G300 BRAKE BLEEDER
MODEL G300 BRAKE BLEEDER Installation, Operation & Repair Parts Information Branick Industries, Inc. 4245 Main Avenue P.O. Box 1937 Fargo, North Dakota 58103 REV060616 P/N: 81-0035G 1 THIS PAGE INTENTIONALLY
STC 3S012-020-A Series Solenoid Valves
STC 3S012-020-A Series Solenoid Valves 3S012-020-A Series Solenoid Valve Specifications Valve Model 3S012-1/8-A 3S012-1/4-A 3S020-1/8-A 3S020-1/4-A Port Size (NPT) 1/8 1/4 1/8 1/4 Valve Type Action 3 Way,
Stainless Steel Single and Dual Circulation Kits
Instruction Sheet P/N 160780 01 Stainless Steel Single and Dual Circulation Kits Introduction The single and dual high-pressure circulation kits allow you to vary and control the circulation rate of coating
SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE
SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE Cable Processor Module overview WARNING! When tipping the Cable Processor Module back, (after removing the toggle arm pin), use extreme caution not to drop
Service Guide 12/27/03 TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models)
TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models) WARNING - HIGH VOLTAGE AC electrical circuits are connected to this heater. Do not attempt any service work on the
Name of Equipment Silver King Model SKMCD1P/C1. This equipment chapter is to be inserted in the appropriate section of the Equipment Manual.
Name of Equipment Silver King Model SKMCD1P/C1 This equipment chapter is to be inserted in the appropriate section of the Equipment Manual. Manufactured exclusively for McDonald s By Silver King Refrigeration,
CABLE MODEM QUICK START
CABLE MODEM QUICK START This Quick Start describes how to connect your Zoom cable modem to a cable modem service. This lets your cable modem provide Internet access to a computer or other device connected
Oil and Coolant Circulating Heating System. Model - OCSM
Oil and Coolant Circulating Heating System Model - OCSM Installation & Operation Manual 216280-000 REV 2 Identifying Your System The HOTSTART heating system is designed to heat fluids for use in marine
430 Power/Electronics Replacement
Replacing the main board WARNING Before proceeding, turn off the main power switch and unplug the power cord. Caution Make sure you are properly grounded with an ESD strap before continuing. The main printed
OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608
Electric Water Heater OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608 If your water Heater is Damaged or you have questions
Micro Cryo-Trap Model 971
Micro Cryo-Trap Model 971 for Liquid CO2 Cooling Scientific Instrument Services, Inc. 1027 Old York Road Ringoes, NJ 08551 Phone (908) 788-5550 Part #951001M revised 10/ /98 Notice The information in this
PALLET JACK - 2.5 TON
PALLET JACK - 2.5 TON 39939 SET UP AND OPERATING INSTRUCTIONS Visit our website at: http://www.harborfreight.com Read this material before using this product. Failure to do so can result in serious injury.
Ceiling Mount Air Handler Manual
www.surna.com 303.993.5271 Ceiling Mount Air Handler Manual Models: CMAH12, CMAH18, CMAH24, CMAH30, CMAH36, CMAH48, CMAH60 Revised: September 2014 Table of Contents Warranty Information 4 Limited Warranty
Thermal Conductivity Detector
Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)
16/32 Channel 1U Rack Mount CCTV Power Supply
16/32 Channel 1U Rack Mount CCTV Power Supply Manual PH-A3224-GUQ Shown 16-Channel 32-Channel PTC PH-A1612-PUQ PH-A3224-PUQ Glass Fuse PH-A1612-GUQ PH-A3224-GUQ Industrial design 12 Amp 3 Amps per channel
Advantium 2 Plus Alarm
ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia
Installation and Troubleshooting Instructions for Electric Tankless Residential Water Heaters.
Model Number: Serial Number: Information Manual Installation and Troubleshooting Instructions for Electric Tankless Residential Water Heaters. ATTENTION: IF YOU ARE NOT A LICENSED PLUMBER OR A LICENSED
ATS Overhead Table Shelf System INSTRUCTION MANUAL
ATS Overhead Table Shelf System INSTRUCTION MANUAL ATS Overhead Table Shelf System Instruction Manual Warranty Newport Corporation warrants this product to be free of defects in material and workmanship
INSTANT HOT WATER RECIRCULATING SYSTEM
INSTANT HOT WATER RECIRCULATING SYSTEM INSTALLATION AND OPERATING INSTRUCTIONS Save manual for future reference MODEL 500800 Warning Please read carefully before proceeding with installation. Your failure
VK-250 WARRANTY REGISTRATION FORM
VK-250 WARRANTY REGISTRATION FORM Unit Serial Number: Customer Name: Address: Date of Purchase: Purchased From: Dealer Name: Address: IMPORTANT NOTE: In order to receive the full five year product warranty,
3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE
4 3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE A. Changing the Control Board B. Replacing a Heating Element C. Thermistor Replacement D. High Limit Switch Replacement E. Level Detector Replacement
MAC FAUCETS. Owner s Manual. Models: FA444
MAC FAUCETS Owner s Manual Models: FA444 Installation procedure: 1. Take a moment to view drawings and read special Installation note on page 3 of this manual before proceeding with installation. Connect
CHROMOCTANE O N LINE ANALYZER
GECIL Process GT instruments CHROMOCTANE O N LINE ANALYZER ASTM D6733 DHA and RON and MON on line GC for Reformer unit Automatic Chromatographic system for analysis of naphta feeds and reforming gasolines
CARING FOR YOUR WATER HEATER
http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html Water Heater Inspections CARING FOR YOUR WATER HEATER Venting System (Direct Vent Only) The venting system should be inspected
Installation Instructions 6028.801
DAZZLE Installation Instructions 08.80 Spread Lavatory Faucet with Speed Connect Drain* Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American
PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: AW 120 / AW 122 / AW 125 AWD 120 / AWD 121 / AWD 129
PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: The information included in this Splendide Repair Manual may change without notice. Please see our web site www.splendide.com/service/docs.html for updates,
Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.
1.0 IMPORTANT RECEIVING INSTRUCTIONS Visually inspect all components for shipping damage. Shipping Damage is not covered by warranty. If shipping damage is found, notify carrier at once. The carrier is
Ozone Generator Manual
ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 Ozone Generator Manual 80003-118 MCN-282, 05/14/04 Table of Contents 1.0 INTRODUCTION... 1 1.1 Unpack...1 1.2 Check Order...1 1.3 Serial Numbers...1
ROTOR LOADER OWNER S MANUAL
ROTOR LOADER OWNER S MANUAL ROTOR LOADER OWNER S MANUAL WARNING IMPORTANT SAFETY INSTRUCTIONS AND GUIDELINES. Misuse of paintball equipment may cause serious injury or death. QUICK SET-UP GUIDE BATTERY
Installation, Maintenance, & Repair Series 4000B/LF4000B
Installation, Maintenance, & Repair Series 4000B/LF4000B Reduced Pressure Zone Assemblies Sizes: 1 2" 2" (15 50mm) RP/IS-A-4000B 4000B/LF4000B Size: 3 /4"! WARNING Testing Read this Manual BEFORE using
Propane Conversion Kit PROPANE CONVERSION KIT INSTALLATION MANUAL
Propane Conversion Kit Supersedes: 145.25-IOM2 (708) Form 145.25-IOM2 (908) PROPANE CONVERSION KIT INSTALLATION MANUAL "LPKIT " - PROPANE CONVERSION KIT Kits are available for field conversion from natural
Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)
Service Manual Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com
Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.
1.0 IMPORTANT RECEIVING INSTRUCTIONS Visually inspect all components for shipping damage. Shipping Damage is not covered by warranty. If shipping damage is found, notify carrier at once. The carrier is
WH/WIH/WIHD/WIHD&M/WIHL AND WIH 17-67 SERIES
WH/WIH/WIHD/WIHD&M/WIHL AND WIH 17-67 SERIES Service and Installation Manual Please read this manual completely before attempting to install or operate this equipment! Notify carrier of damage! Inspect
TABLE OF CONTENTS. SECTION 1 General Information: Pg 2 Description of Operation
GTRDD SERIES HEATLESS DRYER INSTALLATION, OPERATION & MAINTENANCE MANUAL TABLE OF CONTENTS SECTION 1 General Information: Pg 2 Description of Operation SECTION 2 Safety Instructions: Pg 3 Installation
Foodservice Equipment Specialists P.O. Box 880 Saco, ME. / U.S.A. 04072 877-854-8006 * FAX (207) 283-8080
Foodservice Equipment Specialists P.O. Box 880 Saco, ME. / U.S.A. 04072 877-854-8006 * FAX (207) 283-8080 FOR SERVICE ASSISTANCE U.S. AND CANADA CALL: 1-877-854-8006 24 HOURS/DAY 7 DAYS/WEEK TABLE OF CONTENTS
Service manual. Website: www.andico.com.au CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL.
Website: www.andico.com.au Service manual CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL. - ONLY FOR AUTHORISED SERVICE PERSONNEL. MODELS: MPK1-09CR-QB8 MPK1-12ER-QB6
Powers Controls TH 192 HC Heating/Cooling Room Thermostat
Powers Controls TH 192 HC Heating/Cooling Room Thermostat Technical Instructions Document No. 155-066P25 TH 192-2 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 192 HC thermostats are proportional
DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual
DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR Installation and Operation Manual 1 ENGLISH Safety Instructions 2 Features 2-3 Specifications 3 Installation 4-5 Wiring Diagrams 6-7 Warranty 8
Re-packing the SRI FID/Methanizer tube 7-10-2008
7-10-2008 The SRI FID/Methanizer is located on the right side of the GC s column oven. Remove the cover and insulation from the FID/Methanizer heater block using a screwdriver. Loosen the nut holding the
Indirect-Fired Storage Water Heater Models WH-30 through WH-80 INSTALLATION AND OPERATING INSTRUCTIONS
Indirect-Fired Storage Water Heater Models WH-30 through WH-80 INSTALLATION AND OPERATING INSTRUCTIONS Contents Page Ratings and Specifications..................... 2 Installation Requirements......................
Seaward Products OWNER S MANUAL WATER HEATERS. Serial Number:
Seaward Products WATER HEATERS OWNER S MANUAL Serial Number: IMPORTANT SAFETY INSTRUCTIONS WARNING When using electrical appliances, basic safety precautions to reduce the risk of fire, electrical shock,
VOYAGER 570G. 744A Sprayer Control
VOYAGER 570G 744A Sprayer Control U S E R M A N U A L U S E R M A N U A L Table of Contents CHAPTER 1 - INTRODUCTION...1 SYSTEM CONFIGURATIONS...1 KIT CONTENTS...3 CONTROL HOUSING ASSEMBLY...5 CHAPTER
USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model # 900172
Model # 900172: Page 1 USER MANUAL Bottom Loading Bottled Water Dispenser Model # 900172 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING
Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00
E Refrigerant Charging Unit ICOGD 020AH1000 Operating Manual FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00 Contents Technical Specifications... 20 Safety... 21 A/C System... 22 Components... 23 Control Panel...
Global Water Instrumentation, Inc.
Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 6 Tipping Bucket: RG200 02/05/10 01-343 - 1 - Congratulations on your
What you can expect from your Solar Hot Water System
Owner s Manual Your Solar Hot Water System Thank you for your investment in the FAFCO Solar Hot Water System. Since 1969 our goal has been to design, test, and produce solar heating systems that save you
Line to Refrigerator Ice/Water Dispenser
Standard 18 Line to Refrigerator Ice/Water Dispenser Pump Module Bottled Water How The System Works The FLOJET Bottled Water Dispensing System was designed to pump water from a commercially available 5-gallon
6 & 12 Volt Battery and Systems Tester with 100 Amp Load
6 & 12 Volt Battery and Systems Tester with 100 Amp Load Form No. 841-731 -000 DESCRIPTION This Load Tester tests 6 or 12 volt automotive-size lead-acid batteries under load. It will also test 6 or 12
SOLAR ELECTRIC MODULE ES-124 & ES-62T Owners Manual and Installation Guide
SOLAR ELECTRIC MODULE ES-124 & ES-62T Owners Manual and Installation Guide circuit. Reverse connection will damage the module and may result in fire. CAUTIONS Solar electric modules produce DC electricity
OPERATING and MAINTENANCE MANUAL SERIES 2700A CONTROL VALVE CONTENTS. INTRODUCTION...1 Scope...1 Description...1 Valve Identification...
OPERATING and MAINTENANCE MANUAL SERIES 2700A CONTROL VALVE CONTENTS INTRODUCTION...1 Scope...1 Description...1 Valve Identification...2 1.0 VALVE INSTALLATION...2 2.0 VALVE MAINTENANCE...3 Actuator Disassembly...3
Installing a Capillary Column into an ECD and a Micro ECD
Installing a Capillary Column into an ECD and a Micro ECD Procedure: Preparing capillary columns You must prepare your capillary column before installation. Proper preparation assures that the column end
Say Hello to Your Precision Cooker
Say Hello to Your Precision Cooker B A C D E Current Temperature WI-FI Set Temperature Start/Stop Numeric Scroll Wheel Removable Skirt Adjustable Ring Clamp Water Pump Cap F G H Setting up Your Precision
OASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE
OASIS-PLUS 120V INFORMATION & OPERATING INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE 56041852 WARNING:
Installation Operation Maintenance Troubleshooting
Installation Operation Maintenance Troubleshooting Version 3, 2.1, June/09 Apr/09 1416 1402 Table of Contents Important information 2 Weights & Dimensions 3 Operator s Instructions 4 Flow Diagram 5 Purging
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
USER MANUAL Stand Alone Power Supply PSQ 2909 / PSQ 3909 / PSQ 4909 PSQ 2920 / PSQ 3920 / PSQ 4920
USER MANUAL Stand Alone Power Supply PSQ 2909 / PSQ 3909 / PSQ 4909 PSQ 2920 / PSQ 3920 / PSQ 4920 [This page intentionally left blank] Warning for Your Protection 1. Read these instructions. 2. Keep these
AUTOMATIC CHANGEOVER MANIFOLD Series 915
AUTOMATIC CHANGEOVER MANIFOLD Series 915 FEATURES Fully automatic, simple, hassle-free operation No variance in delivery pressure Designed for ultra high purity gas service Built-in audio and visual alarm
Bottom Loading Water Dispenser
Bottom Loading Water Dispenser Model # 601000 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING DISPENSER. SAVE THIS MANUAL FOR FUTURE
GC-2010 Gas Chromatograph INSTRUCTION MANUAL
221-40407F Gas Chromatograph INSTRUCTION MANUAL Read this instruction manual thoroughly before using this instrument. Save the manual for future reference. Copyright Shimadzu Corporation 2000. All rights
FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters
FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters M-4270/0707, pg. 2 of 29 READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY
HP UPS R1500 Generation 3
HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications
EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board.
EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board. This Troubleshooting Guide covers the electric coolant valve and control
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CHCSCS
Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CHCSCS Voltage rating: 120V~60Hz Power rating : 1400W(Cooling) Power rating : 1350W(Heating) Customer Support : 1-800-474-21477
MP-4V Heavy Duty Riveter / 39048
MP-4V Heavy Duty Riveter / 39048 This newly designed heavy-duty air/hydraulic riveter is ergonomically designed with the professional in mind. The light weight 3.7 lbs. well balanced MP-4V includes a Vacuum
CEILING CASSETTE FAN COIL UNIT
CEILING CASSETTE FAN COIL UNIT Operation and Installation Manual MHCFC4W-04, 08, 12, 16 Four way ceiling cassette fan coil unit Please read this manual before using the fan coil unit. Installation and
36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve
Operator: Save these instructions for future use! FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL COULD CAUSE PERSONAL INJURY AND/OR PROPERTY DAMAGE. DESCRIPTION
MCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
Pre-Fab Easy Installation For Models: 006-CT-USK, 008-CT-USK Installation and Operating Instructions for Existing Homes with Standard Piping
Instruction Sheet 102-140 Pre-Fab Easy Installation For Models: 006-CT-USK, 008-CT-USK Installation and Operating Instructions for Existing Homes with Standard Piping Plant I.D. 001-1024 EFFECTIVE: November
Installation and Operating Instructions Installation Instructions for SS EPE-316L Series
INSTR3010 0406 Installation and Operating Instructions Installation Instructions for SS EPE-316L Series Congratulations on your purchase of this Aqua-Pure high flow, single housing filtration system. This
Installation Instructions
7.3L & 6.0L 5/8 FUEL TANK PICKUP KIT Fits 94-07 7.3L & 6.0L Powerstroke Diesel Trucks & Excursions Installation Instructions These instructions are intended simply to be a guide for the installation of
Your safety and the safety of others are very important.
NATURAL GAS TO PROPANE CONVERSION KIT 090 INSTALLATION INSTRUCTIONS FOR ALTITUDES 0 -,00 FT. ONLY PROPANE CONVERSION KIT SAFETY... INSTALLATION REQUIREMENTS... Tools and Parts... LP Gas Requirements...
Watts Pure Water Systems
Watts Pure Water Systems WM-120-PT Installation and Operation Manual Premier Water Systems Do not use with water that is microbiologically unsafe or of unknown quality without adequate disinfection before
PTU Series Ultrahigh-Purity Pressure Transducers
PTU Series UltrahighPurity Pressure Transducers User s Manual The Swagelok PTU series pressure transducer provides electronic monitoring of system pressure for ultrahighpurity applications. The PTU series
About this Manual. Support for Your Product
About this Manual We ve added this manual to the Agilent website in an effort to help you support your product. This manual is the best copy we could find; it may be incomplete or contain dated information.
Cryogenic Vacuum Pumps
Oxford Instruments Austin Cryogenic Vacuum Pumps Decontamination Procedure Only use helium gas with at least 99.999% purity for decontamination procedure described in this document. Decontamination procedures
7 High-Resolution Digital Photo Frame
TM 16-1003 User s Guide 7 High-Resolution Digital Photo Frame One demo photo included, as illustrated Please read this user s guide before using your new photo frame. Package contents Photo Frame AC Adapter
! WARNING. McDonnell & Miller Installation & Maintenance Instructions MM-217(I) Series 150S and 157S (Snap Switch, All Models except 157S-RB-P)
Series 150S and 157S (Snap Switch, All Models except 157S-RB-P) Low Water Cut-Off/Pump Controllers For Steam Boilers and Other Level Control Applications McDonnell & Miller Installation & Maintenance Instructions
Electric Water Heater
Electric Water Heater Operation and Installation Manual (LIMITED WARRANTY AND TANK REPLACEMENT POLICY) The following information should be noted at time of installation and retained for future reference.
Owner s Manual & Safety Instructions
Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W
MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries
INSTALLATION INSTRUCTIONS
INSTALLATION INSTRUCTIONS VGBQ SERIES CONVERSION KIT VIKING RANGE CORPORATION 111 Front Street Greenwood, Mississippi (MS) 38930 USA (662) 455-1200 IMPORTANT: PLEASE READ AND FOLLOW 1. Before beginning,
OPERATING MANUAL CONTRACTOR STYLE AXIAL FANS
Manual No. BLWR043 Rev. 4 September 2012 OPERATING MANUAL CONTRACTOR STYLE AXIAL FANS CVF-8AC CVF-8AC50 CVF-6ACAN CVF-15ACAN CVF-25ACAN CVF-8A15KIT CVF-8A25KIT MODELS: CVF-8DC CVF-6DCAN CVF-15DCAN CVF-25DCAN
Installation Instructions 4508 4508S
SYMPHONY Spread Lavatory Faucet with Speed Connect Drain Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American Standard faucets. Speed Connect
BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets
BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets Any Application. Any Environment. Code No: 0816407 Rev. 4 (07/12)
MP8000 Series Pneumatic Valve Actuators
14-1161- 9, Rev. A Technical Bulletin MP8000 Issue Date March 2016 MP8000 Series Pneumatic Valve Actuators Pre-Assembly Details Page 3 MP8000 Actuator Components 3 Tools Required to Assemble an MP8000
Gas Chromatograph GC Consumables. p/n: 220-94898-02
Gas Chromatograph GC Consumables p/n: 220-94898-02 Diag. # 1) Septum Nut 2) Needle Guide 3) Septum 4) Injection Port A Assy 5) O-ring for Glass Insert 6) Glass Insert 7) Gold Gasket 8) Capillary Adaptor
Installation Manual DIAPHRAGM WELL TANK
Installation Manual DIAPHRAGM WELL TANK IN-LINE SERIES: 2-5 & 7 GALLON VERTICAL SERIES: 14-20-25-32-36-52-65-86-96-119 GALLON HORIZONTAL SERIES: 7-14 & 20 GALLON NO LEAD NO LEAD: The weighted average of
Manual for Fire Suppression & Methane Detection System
Manual for Fire Suppression & Methane Detection System Fogmaker North America Post address: 150 Gordon Dr Exton, PA 19341 Delivery address: 150 Gordon Dr Exton, PA 19341 Tel: 610-265-3610 Fax: 610-265-8327
Installation Manual DIAPHRAGM WELL TANK
Installation Manual DIAPHRAGM WELL TANK Safety Instructions Installation Warranty IN-LINE SERIES: 2-5 & 7 GALLON HORIZONTAL SERIES: 7-14 & 20 GALLON VERTICAL SERIES: 14-20-32-36-52-86-96-119 GALLON KEEP
