1 For Insulin Pump (Version 1.1) Written by: Morteza Yousef-Sanati Hamid Mohammad-Gholizadeh Assignment 2 Course: 703 Software Design Professor: Dr. Tom Maibaum March 2011
2 Table of Contents I INSULIN PUMP SPECIFICATIONS... 3 What is an insulin pump?... 3 How big is an insulin pump?... 3 How does an insulin pump work?... 3 How common is an insulin pump?... 3 II DESIGN DECISION... 5 II.1 SYSTEM DESIGN ARCHITECTURE... 5 II.1.1 GUI... 5 II.1.2 Controller... 6 II.1.3 Data Layer... 6 III OOM DIAGRAMS... 7 III.1 MODEL LEVEL DIAGRAMS... 7 III.1.1 Diagram PackageDiagram_ III.1.2 Diagram UseCaseDiagram_ III.2 PACKAGE DIAGRAMS CONTROLLER... 9 III.2.1 Diagram ACTButtonStateChart... 9 III.2.2 Diagram Controler ClassDiagram III.2.3 Diagram lockpump III.2.4 Diagram lockpumpcollaboration III.2.5 Diagram Pump State Chart III.2.6 Diagram reservoir_sate_chart III.2.7 Diagram resumepump III.2.8 Diagram resumepumpcollaboration III.2.9 Diagram setbasal III.2.10 Diagram setbasalcollaboration III.2.11 Diagram setbolus III.2.12 Diagram setboluscollaboration III.2.13 Diagram setdatetime III.2.14 Diagram setdatetimecollaboration III.2.15 Diagram showpumpstatus III.2.16 Diagram showpumpstatuscollaboration III.2.17 Diagram suspending PumpCollaboration III.2.18 Diagram SuspendingPump III.2.19 Diagram unlockpump III.2.20 Diagram unlockpumpcollaboration III.3 PACKAGE DIAGRAMS DATA III.3.1 Diagram DataLayerClassDiagram III.4 PACKAGE DIAGRAMS GUI III.4.1 Diagram GUI Class Diagram III.4.2 Diagram Pump Simulator User Interface Page 2 of 31
3 I Insulin Pump Specifications What is an insulin pump? The most recently available advance in insulin delivery is the insulin pump. In the United States, MiniMed and Disetronic market the insulin pump. An insulin pump is composed of a pump reservoir similar to that of an insulin cartridge, a battery-operated pump, and a computer chip that allows the user to control the exact amount of insulin being delivered. How big is an insulin pump? Currently, pumps on the market are about the size of a standard communications beeper. How does an insulin pump work? The pump is attached to a thin plastic tube (an infusion set) that has a soft cannula (or plastic needle) at the end through which insulin passes. This cannula is inserted under the skin, usually on the abdomen. The cannula is changed every two days. The tubing can be disconnected from the pump while showering or swimming. The pump is used for continuous insulin delivery, 24 hours a day. The amount of insulin is programmed and is administered at a constant rate (basal rate). Often, the amount of insulin needed over the course of 24 hours varies depending on factors like exercise, activity level, and sleep. The insulin pump allows the user to program many different basal rates to allow for variation in lifestyle. In addition, the user can program the pump to deliver a bolus (large dose of insulin) during meals to cover the excess demands of carbohydrate ingestion. How common is an insulin pump? Over 50,000 people with diabetes worldwide are using an insulin pump. This number is growing dramatically as these devices become smaller and more user-friendly. Insulin pumps allow for tight blood sugar control and lifestyle flexibility while minimizing the effects of low blood sugar (hypoglycemia). At present, the pump is the closest device on the market to an artificial pancreas. More recently, newer models of the pump have been developed that do not require a tubing, in fact - the insulin delivery device is placed directly on the skin and any adjustments needed for insulin delivery are made through a PDA like device that must be kept within a 6 foot range of the insulin delivery device, and can be worn in a pocket, kept in a purse, or on a tabletop when working. Page 3
4 Probably the most exciting innovation in pump technology is the ability to use the pump in tandem with newer glucose sensing technology. Glucose sensors have improved dramatically in the last few years, and are an option for patients to gain further insight into their patterns of glucose response to tailor a more individual treatment regimen. The newest generation of sensors allows for a real time glucose value to be given to the patient. The implantable sensor communicates wirelessly with a pager-sized device that has a screen. The device is kept in proximity to the sensor to allow for transfer of data, however, it can be a few feet away and still receive transmitted information. Depending on the model, the screen displays the blood glucose reading, a thread of readings over time, and a potential rate of change in the glucose values. The sensors can be programmed to produce a "beep" if blood sugars are in a range that is selected as too high or too low. Some can provide a warning beep if the drop in blood sugar is occurring too quickly. To take things one step further, there is one particular sensor that is new to the market that is designed to communicate directly with the insulin pump. While the pump does not yet respond directly to information from the sensor, it does "request" a response from the patient if there is a need for adjustments according to the patterns it is programmed to detect. The ultimate goal of this technology is to "close the loop" by continuously sensing what the body needs, and then responding by providing the appropriate dose of insulin. While this technology is a few more years in the making, the strides in this direction continue to grow. Page 4
5 II Design Decision II.1 System Design Architecture Our design is based on 3-tier architecture which is reflected by three packages: GUI, Controller and Data. GUI includes classes related to GUI Design like, Frames, Buttons, Panels and etc. which is constructing our user interface. Controller includes core classes like classes which control the pump functionality and Data package contains a class which is responsible for storing and retrieving system data. GUI Controller Data System Design Architecture II.1.1 GUI Our GUI s completely reflects a real insulin pump interaction. Instead of using common GUI widgets like menus we decided to implement the menus just like the menus exist in a real insulin pump. This approach facilitate detaching the controller part of the software and embedding it in a real hardware with minimum effort. As you will see in the Class diagram we exploited the Command Pattern to control the complexity of different states of the buttons of the interface. As you will see in the state chart diagrams for the ACT button, its functionally changes according to the place the user interacting with. e.g while user is inside the menu the ACT button command differs Page 5
6 with the time the user setting the basal or bolus or even when the ser want to set the date and time. We also use singleton pattern for preventing some classes to have more than one instance throughout the life cycle of the system. e.g. ActBtn, UpBtn and other buttons in GUI. We should mention that we didn t indicate explicitly this pattern on these classes in class diagram. However it should be considered in implementation. We demonstrated each button as a child of a Button class setting their captions differently. It would be better to show these kinds of situations by using Object diagram and only change the value for that instances, But we avoided to use Object diagrams to avoid moe complexity. For changing the screen inside the frame, say the place we show information to the user, we will create one instance of a Panel class and will change according to each state of the pump. II.1.2 Controller We represented Battery and Reservoir with the corresponding classes in controller layer. These classes simulate the functionality of these hardwares like consuming the battery of decreasing the reservoir insulin. The pump class acts as a core of a controller system. it almost controls and synchronize the activity of the pumps. There are some enumerations in control layer which represents different states of the system like PumpState Enumeration. We used OCL constraint to specify the constraint of the system. these constraint are annotated in class diagrams elements. II.1.3 Data Layer Our data layer is responsible for storing system data. Actually there is one class in this layer doing all data related storing and retrieving tasks. For example we need to store the battery level and reservoir level when the pump is turned of (program closed) and then retrieve these data again when it is turned on (program started). Page 6
7 III OOM Diagrams III.1 Model level diagrams III.1.1 Diagram PackageDiagram_1 GUI Our Classesis organized in three packages, GUI: it includes classes related to GUI Design like, Frames, Buttons, Panels and... Controler: It includes core classes like classes wich control the pump functionality Data: This package contain a class which is responsible for storing and retriving system data Controller Data Page 7
8 III.1.2 Diagram UseCaseDiagram_1 Note: An insulin pump has five main buttons, including UP, Down, B, ACT and ESC/Back button and one indicator light set Alerts changereservoirlevel pump Initialize ShowPump Status changebatterylevel lock and unlock pump Set Bolus Patient <<include>> Access Bolus Menu set Date Time <<include>> Display Bolus History <<extend>> Review Basal Suspend & resume <<abstract>> Set Basal Navigating Menus to access Bolus Menu (This is a common Practice for other Usecases but I avoid to repeat it for simplicity) <<include>> Set Multiple Basal Rate Set Single Basal Rate Page 8
9 III.2 Package diagrams Controller III.2.1 Diagram ACTButtonStateChart SetBolusSelected ActBtn.Click [Bolus.amount > 0]/ StartDelivery ShowPumpStatusSelected ReturnBtn.Click ReturnBtn.Click ActBtn.Click [Pump is in Normal State] ActBtn.Click [Pump is in Normal State] DownBtn.Click SetBolusHighlighted DownBtn.Click SetBasalHighlighted ShowPumpStatusHighlighted ReturnBtn.Click UpBtn.Click ActBtn.Click [Pump is in Normal State] UpBtn.Click DownBtn.Click UpBtn.Click SetBasalSelected ReturnBtn.Click UpBtn.Click ShowBolusHistoryHighlighted ActBtn.Click [Pump is in Normal State] ReturnBtn.Click ShowBolusHistorySelected ActBtn.Click/ ShowNextDeliveryHistory ActBtn.Click [[Basal.Rate > 0 and basal. Time > 12A.M.]]/ NextBasalSetup() DownBtn.Click UpBtn.Click Suspend2 ReturnBtn.Click ActBtn.Click [Pump is in Normal State] SuspendSelected DownBtn.Click UpBtn.Click DownBtn.Click Resume2 ActBtn.Click [Pump is in Attention State] ReturnBtn.Click ResumeSelected Page 9
10 III.2.2 Diagram Controler ClassDiagram BolusHistory Bolus 0..* boluses Ordered - amount : int - amountdelivered : int start () stop () 0..1 addbolus (Bolus bolus) ClearHistory () exporttolist () <<enumeration>> PumpState NORMAL SPECIAL ATTENTION : int : int : int 0..1 currentbolus BasalRate - Time : Date - Rate : double - filepath : String DataProvider (Data) storebasal (Basal basal) retrivebasal () storebatterystate (Battery Parameter_1) retrivebatterystate () storereservoirstate () retrivereservoirstate ()... : Basal : Battery : Reservoir - - Pump state : int datetime : Date initialize () getstatus () alarm () suspend () resume () changebattery () changeresorvoir () finalize () lock () displaylocked... () : void : void : void : void : void : void : void : void Reservoir - tankcapacity : int - status : CapacityStatus consume (int amount) : void 0..1 Basal - basalratecounts : int Derived addbasalrate (BasalRate brate) deletebasalrate (BasalRate brate) clearbasalall () 1..* rates Ordered self<=100 and self>=0 charge>=0 and charge <=20 => status=low Battery - charge : int - status : CapacityStatus Derived getstatus () : CapacityStatus startconsumption () : void stopconsumption () : void tankcapacity>=0 and tankcapacity<=1000 BatteryStatus reservoirstatus <<enumeration>> CapacityStatus VERY_LOW LOW MID HIGH FULL : int : int : int : int : int Page 10
11 III.2.3 Diagram lockpump lockpump :ActBtn :Pump :UPBtn :DownBtn :BolusBtn :ReturnBtn patient2 1: doubleclick 2: lock( ) 3: create LockedCmd:Command 4: execute( ) 5: bindcmdstobtns( ) 6: setcmd(uplockedcmd) 7: setcmd(downlockedcmd) 8: setcmd(blockedcmd) 9: setcmd(rlockedcmd) 10: setcmd(actlockedcmd) 11: lock( ) Page 11
12 III.2.4 Diagram lockpumpcollaboration :Pump 2: lock( ) 3: create :BolusBtn :ActBtn 4: execute( ) 11: lock( ) 10: setcmd(actlockedcmd) 8: setcmd(blockedcmd) 1: doubleclick 5: bindcmdstobtns( ) LockedCmd:Command patient2 6: setcmd(uplockedcmd) 7: setcmd(downlockedcmd) 9: setcmd(rlockedcmd) :ReturnBtn :UPBtn :DownBtn Page 12
13 III.2.5 Diagram Pump State Chart Initial entry / Show Main Screen do / DoPumpInitialization [Initial Action Completed] Normal deliver idle delivery resume Attention entry / showattentionicon do / alertuser exit / beep suspend [self.reservoir.tankcapacity<10%] [self.battery.charge<10%] Special entry / showspecialicon do / alertuser lowbattery suspend lowbattandres lowreservoir Page 13
14 III.2.6 Diagram reservoir_sate_chart Reservoir of pump must be changed in order to get more capacity then there are not reverse transition from lower capacities to higher ones. consumereservoir [40%<self.tankCapacity<= 60%] high consumereservoir [60%<self.tankCapacity<= 80%] full [60%<self.tankCapacity<=80%] [80%<self.tankCapacity <= 100%] [40%<self.tankCapacity<=60%] very_low [self.tankcapacity<=10%] [10%<self.tankCapacity<=40%] middle consumereservoir [self.tankcapacity<=10%] consumereservoir [10%<self.tankCapacity<= 40%] low Page 14
15 III.2.7 Diagram resumepump resumepump patient7 actbtn2:actbtn upbtn2:upbtn downbtn2:downbtn pump2:pump 1: click( ) loop [ResumeMenu is Not Highlighted] alt 2: click( ) 3: click( ) 4: click( ) 5[Pump is in Attention State]: resume( ) Page 15
16 III.2.8 Diagram resumepumpcollaboration downbtn2:downbtn upbtn2:upbtn 2: click( ) 3: click( ) patient7 4: click( ) 1: click( ) 5[Pump is in Attention State]: resume( ) actbtn2:actbtn pump2:pump Page 16
17 III.2.9 Diagram setbasal setbasal patient5 setbasalmnu:menu setbasalcmd:command upbtn1:upbtn downbtn1:downbtn actbtn1:actbtn p5:pump 1: click( ) 2: execute( ) Show the Panel for setting Date and Time 3: bindcmdstobtns( ) user can use UP and Down key several times to get to the appropriat value loop alt [Condition] Condition 4: click( ) Condition 5: click( ) 6: click( ) finally user click Act button to confirm its desired value. 7: setdatetime( ) Page 17
18 III.2.10 Diagram setbasalcollaboration 3: bindcmdstobtns( ) setbasalcmd:command 2: execute( ) setbasalmnu:menu 1: click( ) patient5 5: click( ) downbtn1:downbtn 4: click( ) 6: click( ) upbtn1:upbtn 7: setdatetime( ) actbtn1:actbtn p5:pump Page 18
19 III.2.11 Diagram setbolus setbolus bolusbtn:bolusbtn actbtn:actbtn upbtn:upbtn dbtn:downbtn rbtn:returnbtn patient 1: click( ) 2: create boluscmd:boluscmd 3: execute( ) 4: create bolusscreen:bolusscreen 5: bindcmdstobtns( ) loop [True] alt upclicked 6: click( ) downclicked 7: click( ) break [actbtn or returnbtn Clicked] 8: click( ) 9: savebolus 10: showdefaultscreen 11: destroy 12: click( ) 13: showdefaultscreen 14: destroy Page 19
20 III.2.12 Diagram setboluscollaboration 10: showdefaultscreen 9: savebolus bolusscreen:bolusscreen 4: create 5: bindcmdstobtns( ) boluscmd:boluscmd 11: destroy actbtn:actbtn 14: destroy 13: showdefaultscreen 3: execute( ) 8: click( ) rbtn:returnbtn 2: create patient 12: click( ) 1: click( ) 6: click( ) bolusbtn:bolusbtn 7: click( ) upbtn:upbtn dbtn:downbtn Page 20
21 III.2.13 Diagram setdatetime setdatetime sdtmnu:menu DTcmd:Command p2:pump patient4 1: click( ) 2: execute( ) it bind new appropriat commands to every button including (UP, Down, Return, Bolus and Act) Show the Panel for setting Date and Time 3: bindcmdstobtns( ) 4: setdatetime( ) Page 21
22 III.2.14 Diagram setdatetimecollaboration sdtmnu:menu 2: execute( ) 1: click( ) 3: bindcmdstobtns( ) DTcmd:Command patient4 4: setdatetime( ) p2:pump Page 22
23 III.2.15 Diagram showpumpstatus showpumpstatus patient8 actbtn3:actbtn upbtn3:upbtn downbtn3:downbtn pump3:pump 1: click( ) loop [ShowPumpStatus is Not Highlighted] alt 2: click( ) 3: click( ) 4: click( ) 5: getstatus( ) Page 23
24 III.2.16 Diagram showpumpstatuscollaboration upbtn3:upbtn 2: click( ) patient8 3: click( ) downbtn3:downbtn 4: click( ) 1: click( ) 5: getstatus( ) actbtn3:actbtn pump3:pump Page 24
25 III.2.17 Diagram suspending PumpCollaboration upbtn:upbtn downbtn:downbtn 2: click( ) 3: click( ) patient6 4: click( ) 1: click( ) 5[Pump is in Normal State]: suspend( ) actbtn:actbtn pump:pump Page 25
26 III.2.18 Diagram SuspendingPump SuspendingPump patient6 actbtn:actbtn upbtn:upbtn downbtn:downbtn pump:pump 1: click( ) loop [SuspendMenu is Not Highlighted] alt 2: click( ) 3: click( ) 4: click( ) 5[Pump is in Normal State]: suspend( ) Page 26
27 III.2.19 Diagram unlockpump unlockpump b1:bolusbtn b2:upbtn p1:pump b4:actbtn b3:downbtn b5:returnbtn patient3 1: click( ) 2: click( ) 3: create unlockcmd:command 4: execute( ) 5: bindcmdstobtns( ) 6: setcmd( ) 7: setcmd( ) 8: setcmd( ) 9: setcmd( ) 10: setcmd( ) 11: unlock( ) Page 27
28 III.2.20 Diagram unlockpumpcollaboration b3:downbtn b5:returnbtn b1:bolusbtn 9: setcmd( ) 10: setcmd( ) 1: click( ) patient3 2: click( ) 8: setcmd( ) 7: setcmd( ) 4: execute( ) 3: create 5: bindcmdstobtns( ) unlockcmd:command 11: unlock( ) 6: setcmd( ) b2:upbtn p1:pump b4:actbtn Page 28
29 III.3 Package diagrams Data III.3.1 Diagram DataLayerClassDiagram - - charge status Battery (Controller) : int : CapacityStatus getstatus () startconsumption () stopconsumption () : CapacityStatus : void : void - - tankcapacity status Reservoir (Controller) : int : CapacityStatus consume (int amount) : void - filepath : String DataProvider storebasal (Basal basal) retrivebasal () storebatterystate (Battery Parameter_1) retrivebatterystate () storereservoirstate () retrivereservoirstate () storebolushistory (BolusHistory bolushis tory) retrivebolushistory () : Basal : Battery : Reservoir : BolusHistory Basal (Controller) - basalratecounts : int addbasalrate (BasalRate brate) deletebasalrate (BasalRate brate) clearbasalall () BolusHistory (Controller) addbolus (Bolus bolus) ClearHistory () exporttolist () : Bolus : void : void Page 29
30 III.4 Package diagrams GUI III.4.1 Diagram GUI Class Diagram MainScreen MainScreen represent the default screen when pump is started OR when it is not at any other especial screen PumpScreen {abstract} 1..* * submenu Ordered Menu - title : String click () 0..1 upmenu 0..1 Command {abstract} Command Pattern PumpFrame 0..1 execute () bindcmdstobtns () : void BolusScreen show () : void 0..1 cmd 1..* 0..* Singleton Pattern # # - caption icon action Button : String : String : int click () setcmd (Command cmd) {abstract} : void BolusCmd - showbolusscreen () self.caption="act" ActBtn {Rule_2} BolusBtn {Rule_3} DownBtn ReturnBtn self.caption="bolus" UPBtn self.caption="down" self.caption="return" self.caption="up" Page 30
31 III.4.2 Diagram Pump Simulator User Interface Frame Pump Simulator User Interface UP These are up and Down button Screen Down This is a Pump Panel, I may be represented by Panel Class in Java Bolus Esc(Return) ACT Light Pump Functional Buttons Page 31
Insulin pump therapy aiming for better blood glucose control in people with type 1 diabetes
Insulin pump therapy aiming for better blood glucose control in people with type 1 diabetes The Guy s and St Thomas NHS Foundation Trust insulin pump service is provided by a multidisciplinary team consisting
Get Primed on Pumps: A beginners guide to Insulin Pump Therapy
Get Primed on Pumps: A beginners guide to Insulin Pump Therapy Advantages of insulin pump therapy There are many advantages to using an insulin pump. Anyone can do it with the right training and support.
Insulin Pump Management and Continuous Glucose Monitoring Systems (CGMS)
Insulin Pump Management and Continuous Glucose Monitoring Systems (CGMS) Faith Daily, RN, BSN, CDE, CPT Certified Diabetes Educator/Insulin Pump Trainer August 16, 2014 Why Pump Therapy? Mimics normal
A Reference Guide for School Nurses. with the Medtronic MiniMed Insulin Pump
A Reference Guide for School Nurses with the Medtronic MiniMed Insulin Pump Table Of Contents I Table of Contents The Medtronic MiniMed Insulin Pump A Reference Guide for School Nurses For the 522, 722,
Insulin Delivery System. Reference Guide
Insulin Delivery System Reference Guide Reference Guide As an important member of our family, we would like you to know we understand the importance of managing your diabetes. We care about your health
ALVIN INDEPENDENT SCHOOL DISTRICT Diabetes Medical Management Plan
ALVIN INDEPENDENT SCHOOL DISTRICT Diabetes Medical Management Plan of Plan: School Year (must be current): This plan should be completed by the student s personal health care team and parents/guardian.
Diabetes Health Care Plan
The Public Schools of Brookline School Health Services of Plan: Diabetes Health Care Plan To be completed by the student s health care team and parents/guardian. Plan will be kept with the school nurse
Finally, the right fit for your insulin needs. www.tandemdiabetes.com
Finally, the right fit for your insulin needs. www.tandemdiabetes.com (Actual Size) For convenience and ease of use, the t:flex TM Insulin Pump just fits. It is designed to help make your day-to-day insulin
Taking Insulin Pumps to School. Rachel Calendo, MS, RN, CPNP, CDE, Martha Cuevas, RN, BSN, CPT
Taking Insulin Pumps to School Rachel Calendo, MS, RN, CPNP, CDE, Martha Cuevas, RN, BSN, CPT Insulin Pumps Today A micro-computer, about the size of a pager Programmed to deliver both a preset amount
The design of an insulin pump preliminary requirements (a technical note)
Pol J Med Phys Eng 2009;15(1):25-32. PL ISSN 1425-4689 doi: 10.2478/v10013-009-0003-y website: http://www.pjmpe.waw.pl Hubert J. Hawlas, Krzysztof Lewenstein The design of an insulin pump preliminary requirements
OFFICE OF CATHOLIC SCHOOLS DIOCESE OF ARLINGTON DIABETES MEDICAL MANAGEMENT PLAN Page 1 of 5 TO BE COMPLETED BY PARENT OR GUARDIAN
PART I OFFICE OF CATHOLIC SCHOOLS DIOCESE OF ARLINGTON DIABETES MEDICAL MANAGEMENT PLAN Page 1 of 5 TO BE COMPLETED BY PARENT OR GUARDIAN Student School Date of Birth Date of Diagnosis Grade/ Teacher Physical
Why is Insulin so Important?
Insulin Therapy Why is Insulin so Important? If the glucose stays in your blood it doesn t do your cells (body) any good The glucose has to get inside the cells for the body to use it What Does Insulin
Basal and Bolus Insulin 7/16/2014. Jackie Aday RN, BSN, CDE Jeni Neighbors RN, BSN, CDE. BASAL: Small amount of insulin infused every few minutes
Jackie Aday RN, BSN, CDE Jeni Neighbors RN, BSN, CDE Insulin Pump Therapy Open looped system in which a small amount of insulin is continuously infused through a cannula or needle (basal rate) Larger doses
Design of an Insulin Pump. Purpose of an Insulin Pump:
Design of an Insulin Pump Purpose of an Insulin Pump: Insulin is a hormone central to regulating carbohydrate and fat metabolism in the body. It is secreted regularly within the body and aids in converting
Insulin Pumps. with Diabetes
Insulin Pumps with Diabetes Continuous subcutaneous insulin infusion (with an insulin pump) is an alternative to multiple daily injections for people with type 1 diabetes Streetwise What is an insulin
Pump Therapy MADE FOR YOU. Pump User Guide
Pump User Guide Pump Therapy MADE FOR YOU Dear ACCU-CHEK Spirit insulin pump owner, Thank you for choosing the ACCU-CHEK Spirit insulin pump. This guide will help you to get the most from your ACCU-CHEK
Debiotech will be introducing its new insulin JewelPUMP2 for Diabetes Type 2 during the 2014 ATTD conference in Vienna, Austria. (January 29 th, 2014)
Debiotech will be introducing its new insulin JewelPUMP2 for Diabetes Type 2 during the 2014 ATTD conference in Vienna, Austria. (January 29 th, 2014) Lausanne, Switzerland, January 29, 2014. The JewelPUMP2,
How to design an insulin pump
How to design an insulin pump Learn about the purpose of an insulin pump, its overall workings, and the requirements needed for its design as well as implementation. By Asha Ganesan Applications Engineer
Telephone: Home Work Cell E-mail Address Father/Guardian: Address:
SAMPLE Diabetes Medical Management Plan/Individualized Healthcare Plan Part A: Contact Information must be completed by the parent/guardian. Part B: Diabetes Medical Management Plan (DMMP) must be completed
Imagine a world... Believe in better control. MiniMed Veo Paradigm System Questions and Answers About Insulin Pumping
Imagine a world... Believe in better control MiniMed Veo Paradigm System Questions and Answers About Insulin Pumping 1 Imagine a world... Where you can exercise whenever you want and not have to carb load
SARASOTA MEMORIAL HOSPITAL
SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE MANAGEMENT OF PATIENT S OWN INSULIN PUMP/CONTINUOUS SUBCUTANEOUS INSULIN INFUSION PUMP (dia13) DATE: REVIEWED: PAGES: 08/84 10/15 1 of 7 PS1094 ISSUED
Diabetes Management and Treatment Plan for School (For the insulin pump student)
Lafayette School Corporation Health Services Diabetes Management and Treatment Plan for School (For the insulin pump student) Effective Dates: This plan should be complete by the student s personal health
MiniMed 640G System Intelligent for Better Control
Getting Started with the MiniMed 640G System Intelligent for Better Control Table Of Contents Introduction I MiniMed Care Introducing the MiniMed 640G System MiniMed Care...3 Welcome...4 The Delivery of
School Year 20 / 20. Diabetes Health Care Plan for Southgate Schools
School Year 20 / 20 Diabetes Health Care Plan for Southgate Schools Diabetes Medical Management Plan, Initialized Healthcare Plan and Physician Orders Part A: Contact Information must be completed by the
A pictorial guide to diabetes care, supplies, and devices
A pictorial guide to diabetes care, supplies, and devices Caution: This publication contains depictions of blood, needles and medical procedures related to diabetes care. This publication may be reproduced
Continuous Subcutaneous Insulin Infusion (CSII)
IMPORTANCE OF FOCUS CSII (Insulin pumps) have been used for more than 35 years. In the U.S. in 2005, the level of insulin pump penetration was estimated at 20 to 30% in patients with type 1 diabetes mellitus
Diabetes. New Trends Presented by Barbara Obst RN MS August 2008
Diabetes New Trends Presented by Barbara Obst RN MS August 2008 What is Diabetes Diabetes is a condition characterized by high levels of glucose. The glucose circulates in your blood and serves as the
The OmniPod Insulin Management System
Caregiver GUIDE The OmniPod Insulin Management System The OmniPod is an easy-to-use, two-part insulin delivery system. If you are a school nurse, daycare provider, or other secondary caregiver for someone
Contacts and Important Information
Doctor Name Address Contacts and Important Information Use the spaces below to record important health and product information. r Pharmacy Name Address Telephone and Fax Email Address Telephone and Fax
Anneli, Martina s daughter In better control with her pump since 2011 MY CHILD HAS TYPE 1 DIABETES
Anneli, Martina s daughter In better control with her pump since 2011 MY CHILD HAS TYPE 1 DIABETES Many parents whose child is diagnosed with Type 1 diabetes wonder: Why is this happening to my child?
Objectives PERINATAL INSULIN PUMPS: BASICS FOR NURSES. Historical Perspective. Insulin Pumps in Pregnancy. Insulin Pumps in the US
Objectives PERINATAL INSULIN PUMPS: BASICS FOR NURSES Jo M. Kendrick, APN BC, CDE [email protected] Describe indications and contraindications for insulin pump use in hospitalized patients Differentiate
4/7/2015 CONFLICT OF INTEREST DISCLOSURE OBJECTIVES. Conflicts of Interest None Heather Rush. Heather M. Rush, APRN, CDE Louisville, KY
Heather M. Rush, APRN, CDE Louisville, KY CONFLICT OF INTEREST DISCLOSURE Conflicts of Interest None Heather Rush A conflict of interest exists when an individual is in a position to profit directly or
Is what you know about INSULIN PUMP THERAPY. This educational resource is provided by Medtronic MiniMed, Inc.
Is what you know about INSULIN PUMP THERAPY MYTH or? This educational resource is provided by Medtronic MiniMed, Inc. You have to know a lot about technology to learn to use a pump Interacting with your
Anneli, Martina s daughter In better control with her pump since 2011 MY CHILD HAS TYPE 1 DIABETES
Anneli, Martina s daughter In better control with her pump since 2011 MY CHILD HAS TYPE 1 DIABETES Many parents whose child is diagnosed with Type 1 diabetes wonder: Why is this happening to my child?
Dear Parent/Guardian and Physician of
DIABETIC MANAGEMENT PLAN Parent Consent and Physician Authorization POWAY UNIFIED SCHOOL DISTRICT HEALTH SERVICES 15250 Avenue of Science San Diego CA 92128 Dear Parent/Guardian and Physician of California
Insulin Pump Therapy and Continuous Glucose Sensor Use in the Management of Diabetes Mellitus
Insulin Pump Therapy and Continuous Glucose Sensor Use in the Management of Diabetes Mellitus Louis Haenel, IV, DO, FACOI, FACE Endocrinology Roper Hospital Charleston, SC Dr. Louis Haenel IV has disclosed
Introduction. We hope this guide will aide you and your staff in creating a safe and supportive environment for your students challenged by diabetes.
Introduction Diabetes is a chronic disease that affects the body s ability to metabolize food. The body converts much of the food we eat into glucose, the body s main source of energy. Glucose is carried
Diabetes Medical Management Plan (DMMP)
Diabetes Medical Management Plan (DMMP) This plan should be completed by the student s personal diabetes health care team, including the parents/guardian. It should be reviewed with relevant school staff
INSULIN PUMP. User Guide
INSULIN PUMP User Guide The Accu-Chek Spirit At a Glance Display Menu key Up key Down key Check key Infusion set Cartridge Adapter This user s guide is valid for version 2.XX of the Accu-Chek Spirit. Dear
B. Braun Space GlucoseControl (SGC)
B. Braun Space GlucoseControl (SGC) Quick Reference Guide Valid for software I9305_C Automated Infusion Systems Starting a new insulin therapy SpaceControl cannot be switched on and off by itself. It is
How to Attach Files in Blackboard Learn Email, Messages, Discussions, and Assignments
E-Mail 1. Launch the Email by clicking the icon from the Menu frame found on the left side of the screen. 2. From within the Send Email page, click the "Select Users" link. If you click any links above
Insulin Delivery Options: Inhale, Inject, and Infuse. Traci Evans, A.N.P., B.C.-A.D.M. Nurse Practitioner The Diabetes Center Ocean Springs, MS
Insulin Delivery Options: Inhale, Inject, and Infuse Traci Evans, A.N.P., B.C.-A.D.M. Nurse Practitioner The Diabetes Center Ocean Springs, MS Objectives Identify three ways patients are able to utilize
Insulin Infusion Pumps
Medical Coverage Policy Insulin Infusion Pumps EFFECTIVE DATE: 09/01/2004 POLICY LAST UPDATED: 08/06/2013 OVERVIEW The policy addresses insulin infusion pumps that are worn externally and those that are
ROYAL HOSPITAL FOR WOMEN
ROYAL HOSPITAL FOR WOMEN LOCAL OPERATING PROCEDURE CLINICAL POLICIES, PROCEDURES & GUIDELINES Approved by Quality & Patient Safety Committee 17 April 2014 INSULIN INFUSION PROTOCOL INSULIN DEXTROSE INFUSION
Table of Contents. Introduction to the model 508 insulin pump - - - - - - - - - - - - - - - - - - - - 1
ser s Guide This device is protected under one or more of the following.s. Patents: [S]4,562,751 [S]5,050,764 [S]5,376,070 [S]4,678,408 [S]5,080,653 [S]5,399,823 [S]4,685,903 [S]5,097,122 Other.S. and/or
Calculating Insulin Dose
Calculating Insulin Dose First, some basic things to know about insulin: Approximately 40-50% of the total daily insulin dose is to replace insulin overnight, when you are fasting and between meals. This
Alternative method for delivering insulin using continuous subcutaneous insulin infusion (CSII) open loop system
ISHP Spring CE Conference April 12, 2015 St. Luke s Medical Center Randi Lynn Griffiths, PharmD Clinical outpatient pharmacist Boise VA Medical Center Boise, Idaho Alternative method for delivering insulin
Insulin Pump Therapy
CHILDREN S SERVICES Insulin Pump Therapy These guidelines are not intended for starting a patient on an insulin pump. They are intended to give staff not part of the diabetic team information regarding
Report reference guide. mmol/l. One. solution EMR. No fuss
Report reference guide mmol/l One solution No fuss 0000000000 00 0 0000 00 0 0 0 0 0 0 000000000 0 0 00 0 0 0 00 00 0 0 0 0 0 0 00 00000000000000000000 00 EMR About. Plug Transmitter into your outlet is
The MiniMed Paradigm Veo System A step-by-step guide
The MiniMed Paradigm Veo System A step-by-step guide www.medtronic-diabetes.com.au Your HCPs Contact Details Staying in touch with your Healthcare Professionals (HCPs) Your Endocrinologist: Phone number:
Insulin Pump Therapy in Older Adults
Insulin Pump Therapy in Older Adults Joshua Akers, PharmD Candidate, 2007 College of Pharmacy Washington State University Spokane, WA Phone: 509-432-3332 Email: [email protected] and Stephen M. Setter,
Basal Rate Testing Blood sugar is affected at any time by 1) basal insulin 2) food (carbohydrate) intake 3) bolus insulin (meal time and correction)
Basal Rate Testing Blood sugar is affected at any time by 1) basal insulin 2) food (carbohydrate) intake 3) bolus insulin (meal time and correction) 4) activity and 5) other factors such as stress and
The Diabetes Self Management Insulin Pump Therapy Program
The Diabetes Self Management Insulin Pump Therapy Program Is part of a Nationally recognized program taught by CPT s Offers classes during the daytime in a three part series: PRE-PUMP CLASS SALINE START
Abbott Diabetes Care Inc. www.abbottdiabetescare.com. ART20494 Rev. B 02/11
Abbott Diabetes Care Inc. www.abbottdiabetescare.com ART20494 Rev. B 02/11 Table of Contents Contents Preface...vi Overview of the CoPilot Health Management System...vi Intended Use and Application...vi
Diabetes Insulin Pump Health Care Plan District Nurse Phone: 262-560-2104 District Nurse Fax: 262-560-2106
EMPOW ERING A COMMUNITY OF LEARNERS AND LEADERS Diabetes Insulin Pump Health Care Plan District Nurse Phone: 262-560-2104 District Nurse Fax: 262-560-2106 Student DOB School Grade Doctor Phone School Year
My Sick Day Plan for Type 1 Diabetes on an Insulin Pump
My Sick Day Plan for Type 1 Diabetes on an Insulin Pump When you are sick, your blood sugar levels may be harder to keep under control. Your blood sugar may go too high or too low. Use this guide to help
Use of Continuous Subcutaneous Insulin Infusion (CSII) Pumps in Hospitalized Patients Policy and Procedure
Purpose: To ensure safe and accurate administration of insulin for patients using their own external continuous subcutaneous insulin infusion pump during hospitalization. Definitions: Insulin pump: An
Section 504 Plan (pg 1 of 8)
Section 504 Plan (pg 1 of 8) of Birth School Today s Section 504 Plan for: Disability: Diabetes School Year: Grade: Homeroom Teacher: Bus Number: Background Objectives The student has type diabetes. Diabetes
Diabetes and Technology. Disclosures Certified Insulin Pump Trainer for: Animas Medtronic Diabetes Omnipod. Rebecca Ray, MSN, APRN, FNP-C
Diabetes and Technology Rebecca Ray, MSN, APRN, FNP-C Insulin Pump Therapy and Continuous Glucose Monitoring In Patients with Type 2 Diabetes Page 1 Disclosures Certified Insulin Pump Trainer for: Animas
H-TRONplus. H-TRON plus Reference Manual INSULIN PUMP. Dear H-TRONplus pump user,
H-TRON plus Reference Manual H-TRONplus Dear H-TRONplus pump user, Congratulations on your new H-TRONplus Insulin Pump. You and your H-TRONplus Insulin Pump are unique. Therefore, it is important that
Inpatient Guidelines: Insulin Infusion Pump Management
Inpatient Guidelines: Insulin Infusion Pump Management Developed by the Statewide Diabetes Clinical Network Steering Committee July 2012 Clinical Access and Redesign Unit Table of Contents Purpose...4
DIABETES AND INSULIN PUMP THERAPY TOP TIPS FOR TRAVELLING
DIABETES AND INSULIN PUMP THERAPY TOP TIPS FOR TRAVELLING Information Leaflet Your Health. Our Priority. Page 2 of 6 People with diabetes who use an insulin pump can still enjoy foreign travel. However
WHAT CAN I DO TO REDUCE MY RISK OF DEVELOPING THE COMPLICATIONS OF TYPE 1 DIABETES?
Christian In better control with his pump since 2012 WHAT CAN I DO TO REDUCE MY RISK OF DEVELOPING THE COMPLICATIONS OF TYPE 1 DIABETES? Many people with Type 1 diabetes worry about potential long-term
Insulin Delivery Devices
My Doctor Says I Need Insulin to Control My Diabetes... How Should I Take It? BD Getting Started Insulin Delivery Devices Why is Insulin So Important? Insulin is a life-saving drug for people with diabetes.
Diabetes Medical Management Plan
Diabetes Medical Management Plan 1 School District: School: School Year: Grade: Student Name: DOB: Provider Name: Phone #: Fax #: Blood Glucose Monitoring at School Blood Glucose Target Range: - mg/dl
UW MEDICINE PATIENT EDUCATION. Using Insulin. Basic facts about insulin and self-injection. What is insulin? How does diabetes affect the body?
UW MEDICINE PATIENT EDUCATION Using Insulin Basic facts about insulin and self-injection This handout explains what insulin is, the different types of insulin, how to store it, how to give an injection
If you have type 1 diabetes, you will probably get symptoms quite suddenly, over a few days or weeks.
Patient information from the BMJ Group Diabetes type 1 Type 1 diabetes is a long-term condition. It can lead to serious health problems. But making changes to your lifestyle and taking insulin can help
Equipment and Supplies Checklist for Parents Student: DOB: School: Grade: Equipment and Supplies to be Provided by Parent. Parent Signature Date
Equipment and Supplies Checklist for Parents Student: DOB: School: Grade: Equipment and Supplies to be Provided by Parent Parent Signature Date Daily Snacks (for AM/PM snack times): Specify: Extra Snacks
MiniMed 530G System User Guide
MiniMed 530G System User Guide Copyright & trademarks 2012 Medtronic MiniMed, Inc. All rights reserved. This product is covered by U.S. Patents. Other U.S. and/or foreign patents may be pending. Please
Are you ready to pump?
The decision to use an insulin pump is not an easy one. There is a lot involved with managing diabetes using a pump. Take some time to make this decision. Talk about it with your family and your health
insulin & diabetes What is insulin? Why must it be injected? What if I have to go on to insulin? Are there different types of insulin?
Talking diabetes No.24 insulin & diabetes Insulin injections are required when the body produces little or no insulin, as with type 1 diabetes. They are also required for some people with type 2 diabetes
Insulin Pump Therapy during Pregnancy and Birth
Approvals: Specialist Group: Miss F Ashworth, Dr I Gallen, Dr J Ahmed Maternity Guidelines Group: V1 Dec 2012 Directorate Board: V1 Jan 2013 Clinical Guidelines Subgroup: July 2011 MSLC: V1 Nov 2012 Equality
An estimated 280 Australians develop diabetes every day. It is Australia s fastest-growing chronic disease.
Diabetes and insulin Summary Even with the help of your doctor and diabetes nurse educator, it may take a while to find the right insulin dose to reduce your blood glucose to your target levels. Insulin
Safe use of insulin e- learning module
Safe use of insulin e- learning module Page 1 Introduction Insulin is a hormone produced by the beta cells in the pancreas, it is released when blood glucose levels are raised for example after a meal.
A patient guide to the use of insulin for diabetes
A patient guide to the use of insulin for diabetes Delivering the best in care UHB is a no smoking Trust To see all of our current patient information leaflets please visit www.uhb.nhs.uk/patient-information-leaflets.htm
OmniPod. Extension. for the CoPilot Health Management System. Version 1.1. For use with PDM model UST200
OmniPod Extension for the CoPilot Health Management System Version 1.1 For use with PDM model UST200 OmniPod Extension for CoPilot Quick Start Guide 1 Quick Start Guide Use the OmniPod Extension to download
Type 1 Diabetes - Managing a Critical Ratio
Objectives Students will learn about how ratios and proportions are a key part of managing Type 1 Diabetes Students will learn about the regulation of blood glucose through insulin (hormone) binding with
INSULIN PUMP THERAPY
INSULIN PUMP THERAPY Information Leaflet Your Health. Our Priority. Page 2 of 5 Insulin management plan for pregnant women using insulin pump therapy If at any point, you are not able to control your blood
Insulin Pens & Improving Patient Adherence
Insulin Pens & Improving Patient Adherence Bonnie Pepon, RN, BSN, CDE Certified Diabetes Educator Conemaugh Diabetes Institute Kip Benko, MD FACEP Asst Clinical Professor University of Pittsburgh School
Continuous Subcutaneous Insulin Infusion (CSII) pump therapy
Continuous Subcutaneous Insulin Infusion (CSII) pump therapy GHPI1267_04_15 Department: Diabetes team Review due: April 2018 www.gloshospitals.nhs.uk 21 BETTER FOR YOU BETTER FOR YOU 1 Introduction This
Citrix EdgeSight for Load Testing User s Guide. Citrix EdgeSight for Load Testing 3.8
Citrix EdgeSight for Load Testing User s Guide Citrix EdgeSight for Load Testing 3.8 Copyright Use of the product documented in this guide is subject to your prior acceptance of the End User License Agreement.
Insulin. and diabetes. What is insulin? Who needs to inject insulin? Why must it be injected? What if I have to go on to insulin?
Insulin What is insulin? and diabetes Insulin is a hormone made by special cells, called beta cells, in the pancreas. When we eat, insulin is released into the blood stream where it helps to move glucose
BLOOD GLUCOSE METER. Getting Started Guide for Single Patient Use Only
BLOOD GLUCOSE METER Getting Started Guide for Single Patient Use Only Before You Start Testing About the meter and test strips Carefully read and follow the instructions in the Getting Started Guide,
Paradigm 512 and 712 Infusion Pumps
Paradigm 512 and 712 Infusion Pumps Models MMT-512 MMT-712 User Guide 2005, Medtronic. All rights reserved. This product is covered by U.S. Patent Nos. 6,551,276 and 6,554,798. Other U.S. and/or foreign
Insulin Dose Adjustment REAL-Time CGMS Guidelines for Subjects on Pump Therapy
Insulin Dose Adjustment REAL-Time CGMS Guidelines for Subjects on Pump Therapy In addition to using the blood sugar logs to adjust your insulin doses every week, you should also use your continuous glucose
sbchoi~kku.edu jeremy~sooil.com DANA Diabecare IlS Insulin Infusion pump Infusion pump Class II
510(k) Summary for DANA Diabecare IS SPONSOR SOOIL Development Co, Ltd. 196-1, Dogok-dong, Kangnam-gu Seoul 135-270 Korea 2 2007 Contact Person: Soo Bong Choi Telephone: 82-2-3463-0041 Fax: 82-2-3463-7077
Diabetes Education Information for Teachers and School staff
Diabetes Education Information for Teachers and School staff The information below is intended to educate school staff on the basics of diabetes care. If you have a student in your classroom or you will
SMS Database System Quick Start. [Version 1.0.3]
SMS Database System Quick Start [Version 1.0.3] Warning ICP DAS Inc., LTD. assumes no liability for damages consequent to the use of this product. ICP DAS Inc., LTD. reserves the right to change this manual