Wavefront Ablation Profiles in Refractive Surgery: Description, Results, and Limitations
|
|
|
- Gordon Baldwin
- 10 years ago
- Views:
Transcription
1 REVIEW Wavefront Ablation Profiles in Refractive Surgery: Description, Results, and Limitations David Smadja, MD; Glauco Reggiani-Mello, MD; Marcony R. Santhiago, MD; Ronald R. Krueger, MD, MSE ABSTRACT PURPOSE: To provide an overview of the clinical results of different ablation profi les based on wavefront technology and discuss their characteristics and limitations. METHODS: Literature review of studies reporting results of ablation profi les based on wavefront technology in virgin healthy eyes. RESULTS: Over the past 10 years, a large number of studies comparing different treatment algorithms and newer excimer laser platforms have been published. Thirty-six clinical studies including 3637 eyes analyzing the clinical results obtained after wavefront-guided, wavefront-optimized, and Q-factor profi les have been reviewed. Although wavefront-driven profi les allowed reduction of the amount of induced optical aberrations with conventional ablations, thereby improving the quality of vision, it appears that no algorithm of treatment or excimer laser platform has demonstrated a clear superiority over another. Wound healing and unexpected biomechanical response to surgery affect the accuracy of customized treatments and produce variable results. In addition, it is diffi cult to rigorously analyze and compare fi ndings among different studies because of the diverse variety in which the data are reported. CONCLUSIONS: Despite several technological improvements over the years, wavefront ablation profi les have not consistently demonstrated superiority in terms of visual acuity and lower order aberrations compared to the standard procedure, although the induction of higher order aberrations has been reduced. The concept of an individualized eye model has emerged recently, based on the optical ray tracing algorithm, and could theoretically provide a higher level of customization, thus fulfi lling the promise of super vision. [J Refract Surg. 2012;28(3):xxx-xxx.] doi: / x Journal of Refractive Surgery Vol. xx, No. x, 2012 D uring the past decade, the field of refractive surgery has been flooded with dramatic advances introduced by new technologies. Aside from new applications of the femtosecond laser, further improvements were achieved with the development of new excimer laser platforms with higher repetition rates, faster eye trackers, and customized ablation profiles. Along with this rapid development, the primary goal of refractive surgery has evolved and today aims to provide super vision in terms of visual performance. The quest for super vision has generated a surge of interest, leading to the development of new algorithms of treatment and numerous technological improvements in the newer generations of excimer lasers. A large number of studies comparing different ablation profiles and different excimer laser platforms have been published. However, has an ideal algorithm of treatment emerged from these studies? Does one ablation profile clearly demonstrate superiority compared to the other in regards to visual acuity, contrast sensitivity, or night symptoms? After describing the technical properties of the various treatment algorithms using wavefront technology, we attempted to provide an overview of the clinical results available among several comparative studies. The factors tending to limit the success of these new procedures were also discussed along with future perspectives in the field. Studies reporting results of such profiles for retreatments, highly aberrated eyes, keratoconus, or transplant as well as results of topography-guided treatments were not included in this review. TECHNICAL CONSIDERATIONS OF ABLATION PROFILES The technical aspects of excimer laser ablation have evolved considerably since the introduction of the formula for standard From the Refractive Surgery Department, Cole Eye Institute, Cleveland Clinic Foundation, Cleveland, Ohio. Dr Krueger receives consulting and research funds from Alcon Laboratories Inc, Ft Worth, Texas. The remaining authors have no financial interest in the materials presented herein. Correspondence: David Smadja, MD, 1701 E 12th, W22Q, Cleveland, OH Tel: ; [email protected] Received: April 9, 2011; Accepted: December 6,
2 sphere-cylindrical correction described by Munnerlyn in Over the years, and with the surge of interest generated by wavefront technology, advanced laser treatment algorithms based on different concepts have been developed, aiming to correct not only lower but also higher order wavefront aberration errors. WAVEFRONT-GUIDED TREATMENT REQUIREMENTS Towards the end of the past decade, a new era began with the introduction of ablation profiles guided by the wavefront map to sculpt the cornea in a way to correct preexisting aberrations and reduce surgically induced aberrations. To achieve the precision required for this profile, advancements in excimer laser technology were needed. The broad beam cannot treat small irregularities in the cornea and most current devices use flying-spot technology, in which smaller beams (0.5 to 1.0 mm) are used for better and more accurate results in custom ablations and the correction of irregular astigmatism. Spot sizes 1 mm are required to adequately correct up to the fourth order terms whereas 0.6-mm beams are needed to treat up to the sixth order terms. 2 Some devices have a variable spot size (eg, VISX S4; Abbott Medical Optics Inc, Santa Ana, California). The shape of the beam is also important to produce a smoother surface, which is best achieved with an optimized spot overlap. 3 The energy is delivered in a Gaussian profile (greater energy density in the center of the beam than the periphery) with the Allegretto Wave Eye-Q laser (Alcon Laboratories Inc, Ft Worth, Texas), MEL 80 (Carl Zeiss Meditec AG, Jena, Germany), and NIDEK NAVEX (NIDEK Co Ltd, Gamagori, Japan). 4 The flying-spot laser must be delivered precisely for an accurate wavefront correction and requires highspeed eye-tracking systems because of the smaller spot size and risk of individual pulse decentration and misplacement compared to broad-beam lasers. Centration needs to be accurate, 5 as minimal misalignment or cyclotorsion that would normally cause only a small undercorrection in the spherical-cylindrical profile can induce a completely different aberration pattern. Theoretical modeling indicates that a decentration 0.10 mm can induce rather than reduce aberrations during myopic wavefront-guided treatments. 6 The repetition rate of laser pulses is another feature that varies among excimer lasers. Repetition rate correlates directly with treatment time. Longer periods cause drying of the corneal stroma and loss of fixation, which are unacceptable with customized profiles. In the first broad-beam lasers, the repetition rate was low (6 to 10 Hz). With scanning-spot technology, the rate needed to increase; otherwise, treatment times would be longer because more pulses would be needed for the same ablation. The scanning-spot frequency must not exceed a rate that can be followed by the tracking system. 6 In addition, excessive thermal damage in adjacent tissue can occur with repetition rates higher than 60 Hz. Modern devices, such as the Amaris excimer laser (SCHWIND eye-tech-solutions GmbH & Co KG, Kleinostheim, Germany), use higher repetition rates up to 500 Hz for faster treatments with scanning spots. In these devices, a repetition rate in a determined locale does not exceed 40 Hz because the pulses are being applied sequentially in different locations of the cornea to prevent thermal damage. 7 WAVEFRONT-OPTIMIZED PROFILE Controversy remains concerning the proper definition of the optimal ablation profile. Artal et al 8 found that eyes with smaller wavefront aberrations did not achieve the best visual quality. In 2004, Mrochen et al 9 designed an optimized aspheric profile to compensate only for the spherical aberration induced with the conventional treatment, preserving the preexisting optical aberration pattern of the eye. Since then, this algorithm of treatment has been introduced in several modern excimer laser platforms and is in theory less dependent on technical aspects that are critical in wavefront-guided treatment. This profile has been developed considering the loss in ablation efficacy due to the angle of incidence of the excimer laser pulses in the midperiphery (ie, the cosine effect, leading to an oval beam of reduced energy on the peripherally sloping cornea), which can lead to a decrease in the intended ablation depth in this region and an increase in spherical aberration. The optimized profiles compensate for this effect by increasing the pulse energy in the periphery. CUSTOM-Q FACTOR PROFILE Based on the concept that an ideal asphericity of the cornea could lead to a better visual performance, the Q-factor adjusted profile was developed to manipulate the asphericity by selecting the desired target of Q-value. Based on an aspheric eye model, Manns et al 10 suggested that a minimum of spherical aberration would be obtained at a target Q-factor of approximately 0.4. Some excimer laser platforms are equipped with this profile such as the Allegretto Wave Eye-Q laser with the adjustable Q software mode, allowing the surgeon to target the Q-value independently from the preexisting spherical aberration and the NAVEX system with the Optimized Prolate Ablation (OPA). The OPA profile uses both corneal topography and ocular 2 Copyright SLACK Incorporated
3 aberrometry to either maintain the natural corneal shape of the cornea after surgery or to target a specific corneal asphericity to improve the visual quality. 11 The MEL 80 platform is equipped with the CRS-MASTER Laser Blended Vision module, which allows manipulation of the corneal asphericity to increase the depth of focus to overcome presbyopia. This module has been used to create a non-linear aspheric ablation to provide micro-monovision. 12 CONVENTIONAL TREATMENT AND DEVELOPMENT OF CUSTOMIZED ABLATION PROFILES The main complaints reported by patients after conventional excimer laser refractive surgery are blurred vision under low light conditions, halos, and glare, which lead to decreased visual quality. 13 These complaints could occur despite achieving 20/20 visual acuity on the Snellen chart and are believed to be due to the surgically induced increase in high order aberrations (HOAs) after laser treatment. 14 By reshaping the cornea, the lasers modify the curvature and asphericity in a way that induces an increase in HOAs, especially spherical aberration, due to the optical effect of the transition zone between the treated and untreated cornea. The amount of induced optical aberrations is correlated with the level of attempted refractive correction. 15 It is well known that the effects of significant HOAs degrade the quality of the optical image on the fovea. Although some controversy exists, it has been suggested that the correction of HOAs could lead to an improvement in contrast sensitivity and visual acuity. 16,17 Therefore, the introduction of a customized correction based on the wavefront pattern of the eye was promising to overcome the main issues found in the conventional treatment. OVERVIEW OF THE CLINICAL RESULTS Over the past 10 years, a large number of studies comparing the different treatment algorithms and newer excimer laser platforms have been published. Despite efforts to show the superiority of wavefrontguided correction over the standard treatment, or a customized profile over another, it remains unclear whether an ablation profile has demonstrated its superiority over another in terms of visual performance and patient satisfaction. For clarity, and despite the lack of standardization concerning patient selection and treatment parameters surrounding the majority of these studies, we attempted to summarize the main results reported in the literature using the different available wavefront ablation profiles (wavefront-guided, wavefront-optimized, and custom-q) in virgin healthy eyes. Journal of Refractive Surgery Vol. xx, No. x, 2012 Potential advantages of using these ablation profiles in highly aberrated eyes such as retreatment for decentered ablation, small optical zone, irregular astigmatism, keratoconus, or transplant are not discussed in this review. INTRODUCTION OF WAVEFRONT ABLATION PROFILES Wavefront-guided Treatment. To avoid induction and reduce preexisting aberrations of the eye, the wavefront-guided profile was developed. The first results reported by Seiler et al 18 in 2000 after performing wavefront-guided LASIK on 35 eyes were encouraging. After 3 months, 93.5% of eyes had uncorrected distance visual acuity (UDVA) of 20/20 or better, and supernormal vision, defined as 20/10 or better, was achieved in 16%. Postoperative HOAs decreased in 22.5% of eyes. However, optical aberrations increased on average by a factor 1.44 with a 5-mm pupil. More recently, studies using different wavefront-guided platforms have supported these primary results, reporting improvements in visual acuity and contrast sensitivity postoperatively Although this ablation profile was developed with the goal of avoiding both induction of optical aberrations and reduction of preexisting aberrations, recent studies reported the limitations of such customized treatments. Bottos et al 22 retrospectively analyzed 177 myopic eyes and 32 hyperopic eyes that underwent wavefront-guided LASIK for correction ranging from 9.87 to 6.25 diopters (D). They reported a high correlation between the amount of preoperative refractive error and asphericity changes (r 2 =0.81, P.001) and induced corneal spherical aberrations (r 2 =0.34, P.001) and thus concluded that changes depended on the magnitude of the refractive correction. Additionally, they highlighted the tendency for the postoperative Q-value and spherical aberration to be more positive after myopic ablation and more negative after hyperopic ablation. Keir et al 23 reported similar results in a prospective study analyzing the impact of hyperopic wavefront-guided LASIK on HOAs and contrast sensitivity. A significant change was noted in postoperative spherical aberration, which had a reverse sign (became negative) and a high correlation was found between low contrast visual acuity and induced spherical aberration (r 2 =0.62, P.001). With the surge of interest generated by the wavefront-guided treatment, the need to improve outcomes to meet expectations led the current generation of lasers to undergo many technological improvements over the past 10 years. Following the company-driven technological developments, several studies comparing different wavefront-guided laser platforms have 3
4 been published despite the bias introduced such as the familiarization to the use of a new platform and the need for nomogram adjustment that is surgeon-dependent. Despite the introduction of different technological advancements in all platforms, superiority of one system over another has yet to be established in the peer-reviewed literature Wavefront-optimized Treatment. Good visual outcomes have also been reported with the wavefrontoptimized profile. George et al, 27 using the Allegretto Wave Eye-Q laser, found that 92% of myopes and 63% of hyperopes achieved UDVA of 20/20 after 3 months, and no statistically significant surgically induced spherical aberration was noted among myopic eyes. Using the AMARIS platform, Arbelaez et al 28 found similar results, with 84% of 50 myopic eyes achieving UDVA of 20/20 or better and 40% achieving UDVA of 20/16 or better after 6 months. No significant changes in aberration metrics were found postoperatively. Vongthongsri et al 29 also reported good results after myopic treatment using the optimized aspheric treatment zone (OATz) profile, which is available with the NIDEK NAVEX system. In their study, 94.2% of eyes were within 0.50 D of the intended refraction and 75% of eyes achieved UDVA of 20/20 or better after a mean follow-up of 3 months. RESULTS OF COMPARATIVE STUDIES BETWEEN CONVENTIONAL AND WAVEFRONT PROFILES Conventional vs Wavefront-guided Treatment. Numerous studies published in the literature have failed to demonstrate a clear advantage of wavefront-guided over conventional treatment for low to moderate myopia and myopic astigmatism. 30,31 In a contralateral study comparing 10 eyes treated with conventional LASIK and 10 eyes with wavefront-guided LASIK, Phusitphoykai et al 30 did not find any statistically significant differences regarding visual outcomes and optical aberrations postoperatively. More recently, Dougherty and Bains, 31 in a series of 290 eyes, compared the results of conventional LASIK using the NI- DEK EC-5000 (NIDEK Co Ltd, Gamagori, Japan) with those obtained after using 2 different wavefront-guided laser platforms, the VISX S4 CustomVue (VISX, Santa Clara, California) and CustomCornea (Alcon Laboratories Inc). No statistically significant differences were observed regarding safety, efficacy, or predictability among groups. Other studies concluded that wavefront-guided ablation induces fewer aberrations than conventional treatment, but still increases total HOAs. 32,33 Kim et al, 32 in a contralateral eye comparative study including 24 patients treated with LASIK using the Technolas 217z excimer laser (Bausch & Lomb Surgical, Rochester, New York), reported a reduction in induced HOAs following wavefront-guided treatment. However, no clinical differences in terms of visual acuity, patient preference, and mesopic contrast sensitivity were observed between the two treatments. Another group of studies reported a significant superiority of wavefront-guided compared to conventional treatment in terms of night visual performance and glare symptoms. 20,34 Schallhorn et al 20 observed a significant improvement of night driving visual performance after wavefront-guided correction compared to conventional treatment. Conventional Treatment vs Optimized Profile. Although several authors reported good visual outcomes and only a small increment in surgically induced HOAs after treatment with aspheric profiles, 27,28 few comparative studies with conventional treatment are available in the peer-reviewed literature. Mastropasqua et al 35 compared the visual outcomes and wavefront errors of 50 myopic eyes treated with photorefractive keratectomy (PRK) using an aspheric profile with the MEL 80 laser platform to 24 myopic eyes treated with a standard treatment using the MEL 70 laser platform. They found that both treatments caused an increase in HOAs. Nevertheless, the total wavefront error and spherical aberration were significantly lower with the aspheric profile compared to conventional treatment. de Ortueta et al 36 compared 70 myopic eyes with conventional treatment with 70 eyes treated with an aspheric profile using the ESIRIS laser platform (SCHWIND eye-techsolutions GmbH & Co KG). Their results were similar to those reported by Mastropasqua et al, 35 with an increase of HOAs and spherical aberrations after ablation in both groups, but a smaller induction of HOAs after aspheric ablation than standard ablation. When analyzing these data, despite showing a small reduction in the induction of HOAs with the wavefront-optimized profile, it is not possible to conclude that it leads to superior visual performance compared to the standard treatment. RESULTS OF COMPARATIVE STUDIES BETWEEN DIFFERENT WAVEFRONT ABLATION PROFILES Wavefront-guided vs Wavefront-optimized. Both of these profiles have suggested a smaller increase in HOAs postoperatively compared to the conventional treatment. The wavefront-optimized profile was introduced to preserve the natural prolate shape of the cornea and to compensate for the so-called cosine effect. The idea is also to reduce the variables that can influence the final outcome, resulting in a more simple procedure. In wavefront-guided treatments, as an ex- 4 Copyright SLACK Incorporated
5 ample, the quality of the wavefront map and correct axis alignment tend to be more crucial than in wavefront-optimized treatments. To determine the place of this new algorithm in the strategy of treatment, Stonecipher and Kezirian, 37 in a United States Food and Drug Administration trial, compared 374 myopic eyes randomized to receive either the wavefront-optimized profile or wavefrontguided profile (which was approved for use in the United States in 2003) using the Allegretto Wave Eye- Q. Regarding the visual outcomes after 3 months, most of the results analyzed did not show any statistically significant differences between groups, except for the percentage of eyes that achieved uncorrected distance visual acuity of 20/16, which was slightly higher in the wavefront-optimized group (76% vs 64%). Regarding postoperative wavefront errors, no difference was found between groups in 83% of eyes that had 0.30 μm of preoperative HOA root-mean-square (RMS). In eyes that had 0.30 μm of preoperative HOA RMS, the wavefront-guided profile resulted in less HOAs postoperatively than the wavefront-optimized profile. In terms of these results, the authors concluded that wavefront-guided treatment did not have an advantage over wavefront-optimized treatment in the majority of cases. They suggested wavefront-guided treatment be considered if the magnitude of preoperative HOA RMS is 0.35 μm, which represents approximately 20% to 25% of eyes in the overall general population. Other recent studies have supported these findings. 38,39 Hantera 39 reported similar results with the NAVEX, comparing the OATz algorithm (equivalent to the wavefront-optimized profile) to the optimized path difference custom aspheric treatment (equivalent to the wavefront-guided profile). The author found a greater tendency for increasing HOAs in eyes with 0.30 μm HOA RMS preoperatively in the wavefrontguided group. The reverse tendency was observed in the wavefront-optimized group. Therefore, the author recommended using the wavefront-optimized profile for eyes with preoperative HOA RMS 0.30 μm. Recently, Feng et al 40 supported these findings in a metaanalysis including 7 studies and 930 eyes comparing wavefront-guided to wavefront-optimized LASIK for myopia. They reported no difference between profiles in terms of UDVA, corrected distance visual acuity (CDVA), and induction of HOAs for patients with preoperative HOA RMS 0.30 μm. However, in eyes with preoperative HOA RMS 0.30 μm, the wavefrontguided profile induced less postoperative HOAs than the wavefront-optimized profile. In a prospective, randomized, double-masked study of 200 eyes that underwent LASIK for myopia and Journal of Refractive Surgery Vol. xx, No. x, 2012 astigmatism, Yu et al 41 found no statistically significant difference between treatments after 6 months. The variables analyzed included visual acuity, total HOAs, and contrast sensitivity in dim light conditions. An interesting finding in this study was the absence of a significant difference in patient satisfaction level after both treatments, evaluating night driving symptoms; visual symptoms such glare, halos, or ghost images; fluctuation of vision; and dryness. Perez-Straziota et al, 42 comparing 66 eyes receiving either wavefrontguided or wavefront-optimized LASIK, respectively, found similar results with no difference between groups in terms of UDVA, CDVA, manifest refraction spherical equivalent, and HOAs after 3 months. Another interesting finding in this study was that 14 eyes initially scheduled for wavefront-guided treatment had to be converted to wavefront-optimized for poor limbal ring alignment (8 eyes), a Wavescan that was inconsistent with the manifest refraction (5 eyes), and no iris registration (1 eye), indicating the high level of requirements for wavefront-guided compared to wavefront-optimized treatment. George et al 27 retrospectively evaluated the results of 285 eyes treated with the wavefront-optimized profile using the WaveLight Allegretto laser and found excellent visual outcomes with a low postoperative increase in HOAs of 0.03 μm for myopic LASIK and μm spherical aberration induced after 3 months. Comparing these results to former outcomes obtained with the wavefront-guided treatment using the LADAR CustomCornea, they reported a greater increase in surgically induced spherical aberration with the wavefront-guided compared to the wavefront-optimized profile after myopic LASIK. These results, although in contrast with previous studies comparing both ablation profiles, are consistent with the primary aim of the wavefront-optimized profile, which is to preserve the natural shape of the cornea without inducing ocular aberrations. Custom-Q Profile. Koller et al 43 compared this treatment algorithm to wavefront-guided ablation in a contralateral eye study of 35 patients treated with LASIK for myopia and astigmatism. They reported equal safety and efficacy using both treatments. Although the corneal asphericity was less impaired in the Q-factor profile up to 5.00 D, no statistically significant difference was noted regarding visual or optical parameters up to 9.00 D, except after 1 month when the wavefront-guided group had significantly lower comalike aberration. The unexpected result was the similar induced spherical aberration in both groups. Ghoreishi et al, 44 in a contralateral eye study comprising 56 patients, found similar results comparing one eye treated 5
6 with custom-q PRK targeted to a Q-value of 0.2 and the fellow eye treated with wavefront-guided PRK. They found no differences in terms of safety and visual outcomes and a similar increase in spherical aberrations postoperatively. Stojanovic et al 45 compared this same algorithm of treatment with a target Q-value set to 0.5 or 0.6, depending on the patient s preoperative asphericity and age, to the wavefront-optimized profile. All eyes showed a tendency towards an oblate shift postoperatively; however, less shift was noted in the Custom-Q group than in the wavefront-optimized group. No differences were found regarding safety, refractive outcomes, low contrast visual acuity, and wavefront errors. Of the few studies available in the peer-reviewed literature comparing the custom-q profile to other ablation profiles, most failed to demonstrate any superiority of this algorithm of treatment over another. These findings also suggest that the custom-q profile was not able to deliver exactly the attempted asphericity with the current technology, probably affected also by an unexpected wound-healing response. In a multivariate regression analysis of 160 eyes that underwent wavefront-guided treatment, Tuan and Chernyak 46 found no significant correlation between the change in corneal shape and visual performance. The authors concluded that no standard optimal asphericity exists and the ideal shape needs to be customized on an individual basis. 37 LIMITING FACTORS PATIENT-RELATED LIMITATIONS Despite all technological improvements, final outcome is limited by patient-specific factors, such as wound healing and biomechanical response. Current excimer lasers are precise and can sculpt a plastic model with submicron precision. However, corneal wound-healing response tends to smooth features applied to the corneal stroma and potentially avoid the effect of the intended ablation to correct HOAs. As mentioned by Netto and Wilson, 47 hyperplasia of the epithelium within the ablated zone by even a single cell is likely to have a major affect on a customized corneal treatment, given that the diameter of an epithelial cell is larger than the ablation volume to correct 1 μm RMS of wavefront irregularity in a single point. Furthermore, regarding the biomechanical response, Dupps and Roberts 48 suggested that a central ablation, by releasing tension in the periphery, can lead to expansion and thickening in the midperiphery causing a secondary flattening of the center and thus a change in curvature. These postoperative issues are known to hinder the accuracy of customized treatments and produce variable results because the majority of these profiles do not take into account these unexpected responses. Wavefront-guided treatment aims to correct the preexisting aberrations and avoid the induction of significant amounts of aberrations to improve visual performance, but questions have been raised with recent publications showing that the total RMS of HOAs is not a reliable predictor of visual performance. It has been shown that visual acuity varied significantly depending on the combination of Zernike modes and the relative contribution of each mode. 49 From these findings, it remains to be seen whether the treatment of all HOAs is more beneficial for visual performance and if not, which specific combinations of aberrations must be treated or, on the contrary, maintained or unchanged. Technology, such as adaptive optics, might be a useful tool to combine with wavefront technology to reach a higher level of customization. Preoperative patient simulations with different combinations of aberrations might help in determining the specific amount and Zernike mode of aberration to target with the treatment. TECHNOLOGY-RELATED LIMITATIONS Wavefront sensors are precise and reproducible when dealing with low to moderate magnitudes of higher order aberrations. However, because the measurement of ocular aberrations is based on the position of a collection of spot images, if the eye is highly aberrated (decentered ablations, small optical zone, or irregular astigmatism), the spots overlap and the computer cannot determine the origin of an individual spot, reducing the precision of the measurement. In addition, data analysis by a wavefront sensor does not take into account the quality of individual spots formed by the lenslet array, which can cause false readings in opacities and disruption of the tear film. Furthermore, differences in the technology used by the different wavefront sensing devices such as Hartmann- Shack, Tscherning (position-based aberrometers), or dynamic skiascopy (time-based aberrometers) lead to different wavefront measurements, thereby further increasing the discrepancy in results. 50 FACTORS LIMITING INTERPRETATION AND ANALYSIS OF THE AVAILABLE RESULTS Due to the disparity in results among the different comparative studies, and although customized ablation seems to provide better outcomes than conventional ablation in terms of induced aberrations, it appears that no algorithm of treatment or excimer laser 6 Copyright SLACK Incorporated
7 TABLE 1 Overview of the Disparity in Study Design, Surgical Techniques, Excimer Laser Platforms, and Outcomes Analyzed Among Studies Comparing Different Ablation Profiles Design Technique Parameters analyzed Excimer Study (y) Profile Eyes (N) CL/NCL P/R PRK LASIK Platform VA WF CS Symptom Bottos et al 22 (2011) WFG 209 NCL R FS laser VISX S4 X X Keir et al 23 (2011) WFG 62 NCL P FS laser LADAR X X X Awwad et al 21 (2005) WFG 200 NCL R FS laser Multiple X X X X Arbelaez et al 28 (2009) WFO 50 NCL R FS laser Amaris X X George et al 27 (2010) WFO 285 NCL R FS laser Allegretto X X Schallhorn et al 20 (2009) WFG vs Cv 36 NCL R Keratome Multiple X X Dougherty et al 31 (2008) WFG vs Cv 290 NCL R Keratome Multiple X X Kim et al 32 (2004) WFG vs Cv 48 CL P Keratome Zyoptix X X X Mastropasqua et al 33 (2006) WFO vs Cv 74 CL P X MEL X X de Ortueta et al 36 (2009) WFO vs Cv 70 NCL R Keratome ESIRIS X X Stonecipher & Kezirian 37 (2008) WFG vs WFO 374 NCL P FS laser Allegretto X X Yu et al 41 (2008) WFG vs WFO 200 NCL P Keratome Allegretto X X X X Perez-Straziota et al 42 (2009) WFG vs WFO 134 NCL R Keratome Multiple X X Padmanabhan et al 28 (2008) WFG vs WFO 54 CL P Keratome Allegretto X X X Hantera 39 (2009) WFG vs WFO 23 NCL P Keratome CXIII X X Koller et al 43 (2006) WFG vs Q 70 CL P Keratome Allegretto X X X Ghoreishi et al 44 (2009) WFG vs Q 112 CL P X Allegretto X X X Stojanovic et al 45 (2008) WFO vs Q 46 NCL R X Allegretto X X X CL/NCL = contralateral/noncontralateral study, P/R = prospective/retrospective study, VA = visual acuity, WF = wavefront aberration error, CS = contrast sensitivity, WFG = wavefront-guided, WFO = wavefront-optimized, Cv = conventional, Q = adjusted Q profile, PRK = photorefractive keratectomy, FS = femtosecond VISX S4 (Abbott Medical Optics Inc, Santa Ana, California); LADAR, Allegretto (Alcon Laboratories Inc, Ft Worth, Texas); Amaris (SCHWIND eye-tech-solutions, Kleinostheim, Germany); Zyoptix (Bausch & Lomb, Rochester, New York); MEL (Carl Zeiss Meditec, Jena, Germany); ESIRIS, CXIII (NIDEK Co Ltd, Gamagori, Japan) Journal of Refractive Surgery Vol. xx, No. x,
8 platform has demonstrated a clear superiority over another. The main reason lies in the difficulties in analyzing and comparing findings among those different studies. The Table summarizes the disparities in study design and parameters analyzed among several comparative studies. The lack of standardization in patient selection, range of attempted correction, excimer laser platform and ablation algorithm used, nomogram adjustments, and visual function tested (visual acuity, contrast sensitivity, wavefront errors, or subjective patient satisfaction and visual symptoms) lead to a high degree of bias and significant discrepancies in the results. Furthermore, over the past 10 years, numerous improvements in excimer laser platforms such as faster repetition rates, faster eye trackers, and wavefront analyzers (providing more accurate measurements), may have led to superior outcomes, inducing more variables and thus further confusing the critical analysis of the results. PERSPECTIVES Despite the technological improvements and improvements in refractive outcomes in terms of induction of higher order aberrations and reduction of night vision symptoms compared to the standard procedure, it is now a well known fact that wavefrontguided ablation profiles are not as effective as originally promised. It was demonstrated that wavefront-based ablation profiles could lead to incomplete correction of HOAs or induction of additional aberrations, especially when higher refractive corrections are required. The reason is believed to be due to an inaccurate transfer of a theoretically calculated ablation profile to the cornea by simplifying the effect of the multi-lens system of the eye. As suggested by Mrochen et al 51 only an individualized eye model, based on an optical ray tracing algorithm, can provide an ideal ablation profile by achieving the highest degree of customization. This eye model is obtained by taking into account the following parameters: wavefront of the entire eye, shape and curvature of the anterior and posterior surface of the cornea, corneal and lens thicknesses, anterior chamber depth, axial ocular length, and after an optimization process, the data related to the lens surface. Unpublished data collected in the first European multicenter clinical trial on the ray tracing ablation profile demonstrated excellent safety and efficacy visual outcomes. After 3 months, 75% of eyes gained one line of CDVA and 5% gained two lines, whereas no eyes lost lines. These results seem to be promising and if a clinically significant effect is confirmed, this ray tracing ablation profile may be able to fulfill the promise of super vision. AUTHOR CONTRIBUTIONS Study concept and design (D.S.); data collection (D.S.); analysis and interpretation of data (D.S., G.R.M., M.R.S., R.R.K.); drafting of the manuscript (D.S., G.R.M.); critical revision of the manuscript (M.R.S., R.R.K.); supervision (R.R.K.) REFERENCES 1. Munnerlyn C, Koons SJ, Marshall J. Photorefractive keratectomy: a technique for laser refractive surgery. J Cataract Refract Surg. 1988;14(1): Waheed S, Krueger RR. Update on customized excimer ablations: recent developments reported in Curr Opin Ophthalmol. 2002;14(4): Huang D, Arif M. Spot size and quality of scanning laser correction of higher-order wavefront aberrations. J Cataract Refract Surg. 2002;28(3): Lubatschowski H, Kermani O, Welling H, Ertmer W. A scanning and rotating slit arf excimer laser delivery system for refractive surgery. J Refract Surg. 1998;14(2 Suppl):S186-S Ciccio AE, Durrie DS, Stahl JE, Schwendeman F. Ocular cyclotorsion during customized laser ablation. J Refract Surg. 2005;21(6):S772-S Bueeler M. Maximum permissible lateral decentration in aberration-sensing and wavefront-guided corneal ablation. J Cataract Refract Surg. 2003;29(2): Iseli HP, Mrochen M, Hafezi F, Seiler T. Clinical photoablation with a 500-Hz scanning spot excimer laser. J Refract Surg. 2004;20(6): Artal P. What aberration pattern (if any) produces the best vision? Personal communication, 6th International Wavefront Congress; Feb 11-13, 2005; Athens, Greece. 9. Mrochen M, Donitzky C, Wüllner C, Löffler J. Wavefront-optimized ablation profiles: theoretical background. J Cataract Refract Surg. 2004;(30): Manns F, Ho A, Parel JM, Culbertson W. Ablation profiles for wavefront-guided correction of myopia and primary spherical aberration. J Cataract Refract Surg. 2002;28(5): Holladay JT, Bains HS. Optimized prolate ablations with the NIDEK CXII excimer laser. J Refract Surg. 2005;21(5 Suppl):S595-S Reinstein DZ, Archer TJ, Gobbe M. LASIK for myopic astigmatism and presbyopia using non-linear aspheric micro-monovision with the Carl Zeiss Meditec MEL 80 platform. J Refract Surg. 2011;27(1): Chalita MR, Chavala S, Xu M, Krueger RR. Wavefront analysis in post-lasik eyes and its correlation with visual symptoms, refraction, and topography. Ophthalmology. 2004;111(3): Marcos S, Barbero S, Llorente L, Merayo-Lloves J. Optical response to LASIK surgery for myopia from total and corneal aberration measurements. Invest Ophthalmol. 2001;42(13): Moreno-Barriuso E, Lloves JM, Marcos S, Navarro R, Llorente L, Barbero S. Ocular aberrations before and after myopic corneal refractive surgery: LASIK-induced changes measured with laser ray tracing. Invest Ophthalmol Vis Sci. 2001;(42): Williams D, Yoon GY, Porter J, Guirao A, Hofer H, Cox I. Visual benefit of correcting higher order aberrations of the eye. J Refract Surg. 2000;16(5):S554-S Rocha KM, Vabre L, Chateau N, Krueger RR. Enhanced visual acuity and image perception following correction of highly ab- 8 Copyright SLACK Incorporated
9 errated eyes using an adaptive optics visual simulator. J Refract Surg. 2010;26(1): Seiler T, Mrochen M, Kaemmerer M. Operative correction of ocular aberrations to improve visual acuity. J Refract Surg. 2000;16(5):S619-S Tuan K-mon A, Liang J. Improved contrast sensitivity and visual acuity after wavefront-guided laser in situ keratomileusis: in-depth statistical analysis. J Cataract Refract Surg. 2006;32(2): Schallhorn SC, Tanzer DJ, Kaupp SE, Brown M, Malady SE. Comparison of night driving performance after wavefront-guided and conventional LASIK for moderate myopia. Ophthalmology. 2009;116(4): Awwad ST, Haithcock KK, Oral D, Bowman RW, Cavanagh HD, McCulley JP. A comparison of induced astigmatism in conventional and wavefront-guided myopic LASIK using LADARVision4000 and VISX S4 platforms. J Refract Surg. 2005;21(6):S792-S Bottos KM, Leite MT, Aventura-Isidro M, et al. Corneal asphericity and spherical aberration after refractive surgery. J Cataract Refract Surg. 2011;37(6): Keir NJ, Simpson T, Hutchings N, Jones L, Fonn D. Outcomes of wavefront-guided laser in situ keratomileusis for hyperopia. J Cataract Refract Surg. 2011;37(5): Moshirfar M, Espandar L, Meyer JJ, Tanner JR, Holz HA. Prospective randomized trial of wavefront-guided laser in situ keratomileusis with the CustomCornea and CustomVue laser systems. J Cataract Refract Surg. 2007;33(10): Awwad S, Bowman RW, Cavanagh HD, McCulley JP. Wavefront-guided LASIK for myopia using the LADAR CustomCornea and the VISX CustomVue. J Refract Surg. 2007;23(1): Durrie DS, Stahl J. Randomized comparison of custom laser in situ keratomileusis with the Alcon CustomCornea and the Bausch & Lomb Zyoptix systems: one-month results. J Refract Surg. 2004;20(5):S614-S George MR, Shah RA, Hood C, Krueger RR. Transitioning to optimized correction with the WaveLight ALLEGRETTO WAVE: case distribution, visual outcomes, and wavefront aberrations. J Refract Surg. 2010;26(10):S806-S Arbelaez MC, Vidal C, Arba-Mosquera S. Excimer laser correction of moderate to high astigmatism with a non-wavefrontguided aberration-free ablation profile: six-month results. J Cataract Refract Surg. 2009;35(10): Vongthongsri A, Phusitphoykai N, Siriboonkoom J. Comparison of conventional and OATz ablation profiles for laser in situ keratomileusis to correct myopia up to diopters. J Refract Surg. 2004;20(5 Suppl):S674-S Phusitphoykai N, Tungsiripat T, Siriboonkoom J, Vongthongsri A. Comparison of conventional versus wavefront-guided laser in situ keratomileusis in the same patient. J Refract Surg. 2003;19(2 Suppl):S217-S Dougherty PJ, Bains HS. A retrospective comparison of LASIK outcomes for myopia and myopic astigmatism with conventional NIDEK versus wavefront-guided VISX and Alcon platforms. J Refract Surg. 2008;24(9): Kim T, Yang S, Tchah H. Bilateral comparison of wavefrontguided versus conventional laser in situ keratomileusis with Bausch and Lomb Zyoptix. J Refract Surg. 2004;20(5): Mastropasqua L, Nubile M, Ciancaglini M, Toto L, Ballone E. Prospective randomized comparison of wavefront-guided and conventional photorefractive keratectomy with the Meditec MEL 70 Laser. J Refract Surg. 2004;20(5): Lee HK, Choe CM, Ma KT, Kim EK. Measurement of contrast sensitivity and glare under mesopic and photopic conditions following wavefront-guided and conventional LASIK surgery. J Refract Surg. 2006;22(7): Mastropasqua L, Toto L, Zuppardi E, et al. Photorefractive keratectomy with aspheric profile of ablation versus conventional photorefractive keratectomy for myopia correction: six-month controlled clinical trial. J Cataract Refract Surg. 2006;32(1): de Ortueta D, Arba Mosquera S, Baatz H. Comparison of standard and aberration-neutral profiles for myopic LASIK with the SCHWIND ESIRIS platform. J Refract Surg. 2009;25(4): Stonecipher KG, Kezirian GM. Wavefront-optimized versus wavefront-guided LASIK for myopic astigmatism with the ALLEGRETTO WAVE: three-month results of a prospective FDA trial. J Refract Surg. 2008;24(4):S424-S Padmanabhan P, Mrochen M, Basuthkar S, Viswanathan D, Joseph R. Wavefront-guided versus wavefront-optimized laser in situ keratomileusis : contralateral comparative study. J Cataract Refract Surg. 2008;34(3): Hantera M. Comparison of postoperative wavefront aberrations after NIDEK CXIII optimized aspheric transition zone treatment and OPD-guided custom aspheric treatment. J Refract Surg. 2009;25(10 Suppl):S922-S Feng J, Yu J, Wang Q. Meta-analysis of wavefront-guided vs. wavefront-optimized LASIK for myopia. Optom Vis Sci. 2011;88(12): Yu J, Chen H, Wang F. Patient satisfaction and visual symptoms after wavefront-guided and wavefront-optimized LASIK with the WaveLight platform. J Refract Surg. 2008;24(5): Perez-Straziota CE, Randleman JB, Stulting RD. Visual acuity and higher-order aberrations with wavefront-guided and wavefront-optimized laser in situ keratomileusis. J Cataract Refract Surg. 2010;36(3): Koller T, Iseli HP, Hafezi F, Mrochen M, Seiler T. Q-factor customized ablation profile for the correction of myopic astigmatism. J Cataract Refract Surg. 2006;32(4): Ghoreishi SM, Naderibeni A, Peyman A, Rismanchian A, Eslami F. Aspheric profile versus wavefront-guided ablation photorefractive keratectomy for the correction of myopia using the Allegretto Eye Q. Eur J Ophthalmol. 2009;19(4): Stojanovic A, Wang L, Jankov MR, Nitter TA, Wang Q. Wavefront optimized versus custom-q treatments in surface ablation for myopic astigmatism with the WaveLight ALLEGRETTO laser. J Refract Surg. 2008;24(8): Tuan KM, Chernyak D. A. Corneal asphericity and visual function after wavefront-guided LASIK. Optom Vis Sci. 2006;83(8): Netto MV, Wilson SE. Corneal wound healing relevance to wavefront guided laser treatments. Ophthalmol Clin North Am. 2004;17(2): Dupps W, Roberts C. Effect of acute biomechanical changes on corneal curvature after photokeratectomy. J Refract Surg. 2001;17(6): Applegate RA, Marsack JD, Ramos R, Sarver EJ. Interaction between aberrations to improve or reduce visual performance. J Cataract Refract Surg. 2003;29(8): Kim DS, Narváez J, Krassin J, Bahjri K. Comparison of the VISX WaveScan and NIDEK OPD-Scan aberrometers. J Refract Surg. 2009;25(5): Mrochen M, Bueeler M, Donitzky C, Seiler T. Optical ray tracing for the calculation of optimized corneal ablation profiles in refractive treatment planning. J Refract Surg. 2008;24(4):S446- S451. Journal of Refractive Surgery Vol. xx, No. x,
10 AUTHOR QUERIES Your repeatedly mention that no algorithm has proven to be superior over another. Please remove all repetitive statements throughout the paper. Regarding reference 8, did Dr Artal present a talk at this meeting? Did the author speak during the talk or after? The following was changed. Okay as edited? Original: Decreased quality of vision reported by the patient as blurred vision under low light conditions, halos, and glare are the main complaints after conventional excimer laser refractive surgery. 13 Edited: The main complaints reported by patients after conventional excimer laser refractive surgery are blurred vision under low light conditions, halos, and glare, which lead to decreased visual quality. 13 Please clarify the following. Following the company-driven technological developments, several studies comparing different wavefront-guided laser platforms have been published despite the bias introduced such as the familiarization to the use of a new platform and the need for nomogram adjustment that is surgeon-dependent. Please clarify the following. These findings also suggest that the custom-q profile was not able to deliver exactly the attempted asphericity with the current technology, probably affected also by an unexpected wound healing response. Please clarify the following. However, corneal wound-healing response tends to smooth features applied to the corneal stroma and potentially avoid the effect of the intended ablation to correct HOAs. Please provide a source and year for the unpublished data in the last paragraph of the article. In unpublished data collected in the first European multicenter clinical trial on the ray tracing ablation profile, excellent safety and efficacy visual outcomes were reported. After 3 months, 75% of eyes have gained one line of CDVA and 5% gained two lines, whereas no eyes lost lines In the last sentence of the article, super normal vision was changed to super vision, as used in the Abstract and Introduction. Okay as edited? In the Table, Femto was changed to FS laser and MK was changed to keratome. Okay as edited? 10 Copyright SLACK Incorporated
Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser
Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser Sudhank Bharti, MD; Harkaran S. Bains ABSTRACT PURPOSE: To investigate the predictability
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System Jan Venter, MD ABSTRACT PURPOSE: To determine the predictability, effi cacy, safety, and stability of LASIK using custom ablation
When refractive surgeons began to understand
The Fundamentals of Customized Ablation Customized treatments are especially useful when aberrations are high and symptomatic. BY LAURA DE BENITO-LLOPIS, MRCOphth, MD, PhD When refractive surgeons began
Excimer Laser Instrumentation
Ch073-X0016.qxd 5/10/07 6:39 PM Page 1 CHAPTER 73 Excimer Laser Instrumentation Theo Seiler and Tobias Koller AU: Please supply Key Features. ELEMENTS OF AN EXCIMER LASER SYSTEM The excimer laser emits
Wavefront Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography
Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography Maria Regina Chalita, MD, 1 Sai Chavala, MD, 1 Meng Xu, MS, 2 Ronald R. Krueger, MD, MSE 1 Purpose: To evaluate
The pinnacle of refractive performance.
Introducing! The pinnacle of refractive performance. REFRACTIVE SURGERY sets a new standard in LASIK outcomes More than 98% of patients would choose it again. 1 It even outperformed glasses and contacts
Wavefront technology has been used in our
Wavefront Customized Ablations With the WASCA Asclepion Workstation Sophia I. Panagopoulou, BSc; Ioannis G. Pallikaris, MD ABSTRACT PURPOSE: WASCA (Wavefront Aberration Supported Cornea Ablation) is a
A Contralateral, Randomized Comparison of Optimized Prolate Ablation and Conventional LASIK for Myopia With the NIDEK Excimer Laser Platform
ORIGINAL ARTICLES A Contralateral, Randomized Comparison of Optimized Prolate Ablation and Conventional LASIK for Myopia With the NIDEK Excimer Laser Platform Alaa M. El Danasoury, MD, FRCS; Jack Holladay,
Laser refractive surgery is becoming increasingly
Correlation of Aberrations With Visual Symptoms Using Wavefront Analysis in Eyes After Laser in situ Keratomileusis Maria Regina Chalita, MD; Meng Xu, MS; Ronald R. Krueger, MD, MSE ABSTRACT PURPOSE: To
Customized corneal ablation and super vision. Customized Corneal Ablation and Super Vision
Customized Corneal Ablation and Super Vision Scott M. MacRae, MD; James Schwiegerling, PhD; Robert Snyder, MD, PhD ABSTRACT PURPOSE: To review the early development of new technologies that are becoming
Refractive Surgery. Evolution of Refractive Error Correction
Refractive Surgery Techniques that correct for refractive error in the eye have undergone dramatic evolution. The cornea is the easiest place to place a correction, so most techniques have focused on modifying
Laser refractive surgery has evolved full circle.
Clinical Pearls for Success in Femto-LASIK Refractive Surgery We have 3 months of experience using the Femto LDV in combination with the Amaris Total Tech Laser. BY MARIA CLARA ARBELAEZ, MD; AND SAMUEL
Richard S. Hoffman, MD. Clinical Associate Professor of Ophthalmology Oregon Health & Science University
Zeiss Mel 80 and Visumax Refractive Laser Systems Richard S. Hoffman, MD Clinical Associate Professor of Ophthalmology Oregon Health & Science University No Financial Interest ZEISS Workstation CRS-Master
The Evolution of the Optical Zone in Corneal Refractive Surgery. Bruce Drum, Ph.D.
The Evolution of the Optical Zone in Corneal Refractive Surgery. Bruce Drum, Ph.D. FDA, Division of Ophthalmic and ENT Devices, Rockville, MD Disclaimer This presentation represents the professional opinion
Common visual problems in older LASIK patients
丘 子 宏 LASIK 手 術 後 的 視 覺 Visual acuity:the measurement of high contrast Snellen acuity but not other functions under different condition Quality of vision: measure the visual functions in variable condition
VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment)
CustomVue Advantage Patient Information Sheet VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment) Statements
Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda
Slit Skiascopic-guided Ablation Using the Nidek Laser Scott MacRae, MD; Masanao Fujieda ABSTRACT PURPOSE: To present the approach of using a scanning slit refractometer (the ARK 10000) in conjunction with
Wavefront-Guided LASIK for the Correction of Primary Myopia and Astigmatism
Ophthalmic Technology Assessment Wavefront-Guided LASIK for the Correction of Primary Myopia and Astigmatism A Report by the American Academy of Ophthalmology Steven C. Schallhorn, MD, Ayad A. Farjo, MD,
Comparing Femtosecond Lenticule Extraction (FLEx) and Femtosecond Laser In-situ Keratomileusis (LASIK) for Myopia and Astigmatism
Original Article Philippine Journal of OPHTHALMOLOGY Comparing Femtosecond Lenticule Extraction (FLEx) and Femtosecond Laser In-situ Keratomileusis (LASIK) for Myopia and Astigmatism Tina Marie Saban-Roa,
OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment
Shinagawa LASIK Center OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment Minoru Tomita, MD, Ph.D 1) Executive Medical Director at Shinagawa LASIK Center, Tokyo, Japan 2) Clinical Professor
Comparison of higher-order aberrations after wavefront-guided laser in situ keratomileusis and laser-assisted subepithelial keratectomy
J CATARACT REFRACT SURG - VOL 32, MAY 2006 Comparison of higher-order aberrations after wavefront-guided laser in situ keratomileusis and laser-assisted subepithelial keratectomy So-Hyang Chung, MD, In
It has been 10 years since ophthalmology was captivated
Wavefront-Guided Treatments: Past, Present, and Future An overview of wavefront basics and a personal account of one surgeon s experience. BY A. JOHN KANELLOPOULOS, MD It has been 10 years since ophthalmology
Accelerated Refractive Performance
Accelerated Refractive Performance Get There at the Speed of WaveLight Designed to accommodate your refractive technology goals now and into the future, the WaveLight Workstation is a faster way to get
Consumer s Guide to LASIK
Consumer s Guide to LASIK A Community Service Project brought to you by Price Vision Group Your Guide To A Successful LASIK Procedure The purpose of this educational guide is to help prospective patients
Keratoconus Detection Using Corneal Topography
Keratoconus Detection Using Corneal Topography Jack T. Holladay, MD, MSEE, FACS ABSTRACT PURPOSE: To review the topographic patterns associated with keratoconus suspects and provide criteria for keratoconus
LASIK. Complications. Customized Ablations. Photorefractive Keratectomy. Femtosecond Keratome for LASIK. Cornea Resculpted
Refractive Surgery: Which Procedure for Which Patient? David R. Hardten, M.D. Minneapolis, Minnesota Have done research, consulting, or speaking for: Alcon, Allergan, AMO, Bausch & Lomb, Inspire, Medtronic,
Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY
Sponsored by Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Results of the T-CAT Phase III Clinical Trial TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Topography-Guided Custom Ablation Treatments (T-CAT) with
Cornea. Wavefront-guided Laser-assisted in situ Keratomileusis. Wavefront Sensing Methods and Devices. Preoperative Diagnostics.
Wavefront-guided Laser-assisted in situ Keratomileusis Steve Schallhorn, MD Professor of Ophthalmology, University of California, San Francisco, US; Chief Medical Director, Optical Express, UK; and Private
THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening
THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening GALILEI G4 Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software
Topography-guided laser refractive surgery
REVIEW C URRENT OPINION Topography-guided laser refractive surgery Simon Holland a,b, David T.C. Lin a, and Johnson C.H. Tan b,c Purpose of review Topography-guided laser refractive surgery regularizes
The future of laser refractive surgery is exciting
The Cornea is Not a Piece of Plastic Cynthia Roberts, PhD Editorial The future of laser refractive surgery is exciting with the potential for ever-improved postoperative visual performance. In the past,
Despite being a relatively safe and
APACRS Asia-Pacific The news magazine of the Asia-Pacific Association of Cataract & Refractive Surgeons Supplement to EyeWorld Asia-Pacific Winter 2015 Shaping the refractive laser landscape Sponsored
Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information
Rediscover quality of life thanks to vision correction with technology from Carl Zeiss Patient Information 5 2 It was really w Vision defects: Light that goes astray For clear vision the eyes, cornea and
Comparison of Night Driving Performance after Wavefront-Guided and Conventional LASIK for Moderate Myopia
Comparison of Night Driving Performance after Wavefront-Guided and Conventional LASIK for Moderate Myopia Steve C. Schallhorn, MD, David J. Tanzer, MD, Sandor E. Kaupp, MS, Mitch Brown, OD, Stephanie E.
FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US
FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US JON DISHLER, MD DENVER, COLORADO, USA INTRODUCTION AND STUDY OBJECTIVES This article summarizes the first US
Advanced personalized nomogram for myopic laser surgery: First 100 eyes
ARTICLE Advanced personalized nomogram for myopic laser surgery: First 1 eyes Ruth Lapid-Gortzak, MD, Jan Willem van der Linden, BOpt, Ivanka J.E. van der Meulen, MD, Carla P. Nieuwendaal, MD PURPOSE:
Q-factor customized ablation profile for the correction of myopic astigmatism
J CATARACT REFRACT SURG - VOL 32, APRIL 2006 Q-factor customized ablation profile for the correction of myopic astigmatism Tobias Koller, MD, Hans Peter Iseli, MD, Farhad Hafezi, MD, Michael Mrochen, PhD,
Supplement to September 2013 FEATURED AUTHORS:
Supplement to September 2013 SHAPING THE FUTURE With SCHWIND AMARIS 1050RS and PresbyMAX Hybrid FEATURED AUTHORS: Minoru Tomita, MD, PhD María Clara Arbelaez, MD Michiel H.A. Luger, MD Pierre Baudu, MD
SCHWIND CAM Perfect Planning wide range of applications
SCHWIND CAM Perfect Planning wide range of applications ORK-CAM PresbyMAX PALK-CAM PTK-CAM 2 SCHWIND CAM the system solution The latest version of the modular SCHWIND CAM represents an even more efficient
EDITORIAL Wavefront Guided Ablation
EDITORIAL Wavefront Guided Ablation SCOTT M. MACRAE, MD, AND DAVID R. WILLIAMS, PHD ADAPTIVE OPTICS WAS FIRST SUGGESTED IN 1953 BY astronomer Horace Babcock to remove the blurring effects of turbulence
OPTIMIZED. Excimer Laser Surgery,
Supplement to March 2009 Excimer Laser Surgery, OPTIMIZED Surgeons are adopting the ALLEGRETTO WAVE Eye-Q laser for the efficiency, safety, and undeniable results of its wavefront-optimized ablations.
MicroScan. Excimer laser system for all types of vision corrections OPTOSYSTEMS LTD.
MicroScan Excimer laser system for all types of vision corrections OPTOSYSTEMS LTD. МicroScan is a new generation excimer laser system for all types of vision corrections: myopia, hypermetropia, astigmatism.
Laser in situ keratomileusis for mixed astigmatism using a modified formula for bitoric ablation
European Journal of Ophthalmology / Vol. 18 no. 6, 2008 / pp. 869-876 Laser in situ keratomileusis for mixed astigmatism using a modified formula for bitoric ablation D. DE ORTUETA, C. HAECKER Augenlaserzentrum
Assessment of Contrast Sensitivity and Aberrations After Photorefractive Keratectomy in Patients with Myopia Greater than 5 Diopters
ORIGINAL REPORT Assessment of Contrast Sensitivity and Aberrations After Photorefractive Keratectomy in Patients with Myopia Greater than 5 Diopters Alireza Fahim 1, Bijan Rezvan 1, and Hassan Hashemi
INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the
INTRACOR An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the 1 Dr Luis Ruiz Presbyopia treatment with INTRACOR Luis
Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay
Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay Minoru Tomita, MD, PhD 1,2 1) Shinagawa LASIK, Tokyo,
LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY
Thi-Qar Medical Journal (TQMJ): Vol(4) No(4):1(14-21) SUMMARY: LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY Dr. Ali Jawad AL- Gidis (M.B.Ch.B., D.O., F.I.C.O.)* Background: LASIK which
Detection of static cyclotorsion and compensation for dynamic cyclotorsion in laser in situ keratomileusis
ARTICLE Detection of static cyclotorsion and compensation for dynamic cyclotorsion in laser in situ keratomileusis Jean-Luc Febbraro, MD, Douglas D. Koch, MD, Hamza N. Khan, MD, MPH, FRCSC, Alain Saad,
Overview of Refractive Surgery
Overview of Refractive Surgery Michael N. Wiggins, MD Assistant Professor, College of Health Related Professions and College of Medicine, Department of Ophthalmology Jones Eye Institute University of Arkansas
Clinical Results of Topography-based Customized Ablations in Highly Aberrated Eyes and Keratoconus/Ectasia With Cross-linking
Clinical Results of Topography-based Customized Ablations in Highly Aberrated Eyes and Keratoconus/Ectasia With Cross-linking David T.C. Lin, MD, FRCSC; Simon Holland, MD, FRCSC; Johnson C.H. Tan, MBBS,
Cassie Schroeder Refractive Surgery Coordinator Boozman-Hof Regional Eye Clinic (479) 246-1820 (479) 531-3937
Thank you for your interest in refractive surgery here at Boozman-Hof Regional Eye Clinic. Enclosed is a bio on Dr. Cole and articles about LASEK. There is also information on our financing company that
Changes in higher order aberrations after wavefront-guided PRK for correction of low to moderate myopia and myopic astigmatism: Two-year follow-up
European Journal of Ophthalmology / Vol. 17 no. 4, 2007 / pp. 507-514 Changes in higher order aberrations after wavefront-guided PRK for correction of low to moderate myopia and myopic astigmatism: Two-year
LASER VISION C ORRECTION REFRACTIVE SURGERY CENTER
LASER VISION C ORRECTION REFRACTIVE SURGERY CENTER W e l c o m e Throughout our history, physicians at Mass. Eye and Ear have led clinical advances and research that have resulted in the discovery of disease-causing
What now, when Presbyopia sets in? Minoru Tomita, MD, PhD Executive Director Shinagawa LASIK Center, Tokyo, JAPAN
What now, when Presbyopia sets in? Minoru Tomita, MD, PhD Executive Director, Tokyo, JAPAN Financial disclosure Travel expenses for the ASCRS meeting were paid by Ziemer and AcuFocus, Inc. Presbyopia:
Alain Saad, MD, Alice Grise-Dulac, MD, Damien Gatinel, MD, PhD
CASE REPORT Bilateral loss in the quality of vision associated with anterior corneal protrusion after hyperopic LASIK followed by intrastromal femtolaser-assisted incisions Alain Saad, MD, Alice Grise-Dulac,
Intracorneal Ring Segments Implantation Followed by Same-day Topography-guided PRK and Corneal Collagen CXL in Low to Moderate Keratoconus
SURGICAL TECHNIQUE Intracorneal Ring Segments Implantation Followed by Same-day Topography-guided PRK and Corneal Collagen CXL in Low to Moderate Keratoconus Waleed Al-Tuwairqi, MD; Mazen M. Sinjab, MD,
IntraLase Corp. Clinical Studies Fact Sheet
Page 1 of 12 Contact: Liana Miller (949) 595-4320 [email protected] IntraLase Corp. Clinical Studies Fact Sheet Clinical data validates IntraLase as a superior technology for creating corneal flaps
Multi-Centre Evaluation of TransPRK outcomes with SCHWIND AMARIS using SmartPulse Technology
study 2015 Multi-Centre Evaluation of TransPRK outcomes with SCHWIND AMARIS using SmartPulse Technology David T. Lin, MD, Canada, Paolo Vinciguerra, MD, Italy, Maria C. Arbelaez, MD, Oman, Shady T. Awwad,
Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation
Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation Bettina B. Jendritza, MD; Michael C. Knorz, MD; Steve Morton ABSTRACT PURPOSE: To investigate the use of wavefront-guided
Surgical Management of Irregular Astigmatism
Peer-Reviewed Literature: Surgical Management of Irregular Astigmatism Editor: Ming Wang, MD, PhD, Clinical Associate Professor of Ophthalmology at the University of Tennessee and Director of the Wang
Our Commitment To You
SYSTEM SUPPORT Quality-crafted, the system boasts dependability with high efficiency and low gas usage. We provide responsive service and maintenance contract options, supported by our nationwide direct
Customized ablation algorithm for the treatment of steep central islands after refractive laser surgery
J CATARACT REFRACT SURG - VOL 32, MAY 2006 Customized ablation algorithm for the treatment of steep central islands after refractive laser surgery Farhad Hafezi, MD, Mirko Jankov, MD, Michael Mrochen,
Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia
Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia Alaa El Danasoury, FRCS Magrabi Hospitals & Centers Surgical options for the Correction of Presbyopia Monovision Reversal of Presbyopia:
Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps
Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps Lingo Y. Lai, MD William G. Zeh, MD Clark L. Springs, MD The authors
LASIK for Hyperopia With the WaveLight Excimer Laser
LASIK for Hyperopia With the WaveLight Excimer Laser A. John Kanellopoulos, MD; Joseph Conway, MD; Lawrence H. Pe, MD ABSTRACT PURPOSE: To evaluate the safety and effi cacy of the ALLEGRETTO WAVE excimer
Excellent outlook for your eyes: With SCHWIND AMARIS technology
Excellent outlook for your eyes: With SCHWIND AMARIS technology Laser surgery for the eyes Comfortable, fast and safe The SCHWIND laser eye procedure When performing laser procedures, eye surgeons apply
Save the Date. 16th SCHWIND User Meeting January 21-24, 2016, Singapore
Save the Date 16th SCHWIND User Meeting January 21-24, 2016, Singapore SCHWIND Programme Booklet ESCRS Barcelona September 5-8, 2015 SCHWIND eye-tech-solutions, Programme ESCRS 2015 3 ESCRS Congress 2015
Life Science Journal 2014;11(9) http://www.lifesciencesite.com. Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000)
Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000) Gamal Mostafa Abo El Maaty, Mohamed Elmoddather, Mahmoud Ibrahem Ghazy, Mohamed Al-Taher Ophthalmology Department, Faculty of
How To See With An Cl
Deciding on the vision correction procedure that s right for you is an important one. The table below provides a general comparison of the major differences between Visian ICL, LASIK and PRK. It is NOT
Faster recovery of visual acuity at all distances
news Special issue PresbyMAX Faster recovery of visual acuity at all distances PresbyMAX μ-monovision: The latest technique for treating presbyopia SCHWIND now offers PresbyMAX µ-monovision as a new option
Alexandria s Guide to LASIK
Alexandria s Guide to LASIK A Community Service Project sponsored by: Wallace Laser Center Your Guide To A Successful LASIK Procedure The word LASIK is actually an acronym for Laser Assisted In-Situ Keratomileusis.
CustomVue Treatments for Monovision in Presbyopic Patients with Low to Moderate Myopia and Myopic Astigmatism
CustomVue Treatments for Monovision in Presbyopic Patients with Low to Moderate and Myopic Introduction Pre-Operative Examination Surgical Technique 1 2 IMPORTANT INFORMATION CustomVue Monovision treatments
Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday?
Financial Disclosure Arturo Chayet, MD Tijuana, BC Mexico Perry S. Binder, MS, MD San Diego CA USA I have the following financial interests or relationships to disclose: AMO/IntraLase Corporation - C Acufocus
Dr. Booth received his medical degree from the University of California: San Diego and his bachelor of science from Stanford University.
We've developed this handbook to help our patients become better informed about the entire process of laser vision correction. We hope you find it helpful and informative. Dr. Booth received his medical