Equipment used in connection with medical exposure
|
|
|
- Lorraine Logan
- 10 years ago
- Views:
Transcription
1 N Health (Third edition) This guidance is issued by the Health and Safety. Following the guidance is not compulsory and you are free to take other action. But if you do follow the guidance you will normally be doing enough to comply with the law. Health and safety inspectors seek to secure compliance with the law and may refer to this guidance as illustrating good practice. Equipment used in connection with medical exposure and Safety Guidance Notes are published under five subject headings: Medical Environmental Hygiene Chemical Safety Plant and Machinery General Introduction 1 The advice in this guidance note is for employers who have to any extent control of equipment used in connection with medical exposure to ionising radiation. This includes many NHS employers and private healthcare providers, medical x-ray services in industry, dentists, physiotherapists, osteopaths and chiropractors (this list may not be exhaustive). It may also include companies involved in providing facilities under Private Finance Initiative Schemes (PFI). The guidance will be useful to radiation protection advisers (RPAs) who have been consulted by such employers. 2 Advice is given on compliance with the Ionising Radiations Regulations (IRR99) in relation to equipment used in connection with medical exposure. In particular the guidance provides advice on the requirements of IRR99 Regulation 32(1), ie that the employer ensures equipment is of such design or construction and is so installed and maintained as to be capable of restricting so far as is reasonably practicable the exposure to ionising radiation of any person who is undergoing a medical exposure to the extent that it is compatible with the intended clinical purpose or research objective. The advice does not cover duties under the Ionising Radiation (Medical Exposure) Regulations 2000 (IR(ME)R) 2 but references are made to those Regulations where appropriate in the context. 3 Medical exposure is defined in IRR99 regulation 2(1) as exposure of a person to ionising radiation for the purpose of his medical or dental examination or treatment which is conducted under the direction of a suitably qualified person and includes any such examination for legal purposes and any such examination or treatment conducted for the purposes of research. For simplicity, all persons undergoing medical exposures are referred to as patients in this guidance. 4 This guidance supplements the general advice on compliance with IRR99 that can be found in Work with Ionising Radiation. Ionising Radiations Regulations 1999 Approved Code of Practice and Guidance L Scope 5 This guidance includes advice on: the selection, installation, maintenance, calibration and replacement of equipment; criteria of acceptability for both new and older equipment; quality assurance programmes, including adoption of suspension levels;* and the investigation of incidents involving a malfunction or defect in radiation equipment which results in an exposure much greater than intended. 1 of 18 pages
2 * Suspension Level is the level of non-compliance with quality assurance criteria at which an item of equipment should be removed from clinical use with immediate effect. For example, recommended levels for the suspension of diagnostic X-ray equipment are published in IPEM publication No Equipment used in connection with medical exposure includes all equipment whose design, construction, installation or maintenance (and any fault that might develop in it) can affect the magnitude or distribution of the absorbed dose received by the person undergoing a medical exposure. It includes: equipment intended to be used in connection with diagnostic or therapeutic procedures using ionising radiation and interventional radiology; radiation equipment (as defined by IRR99 regulation 32(8) and in paragraph 7); and ancillary equipment such as intensifying screens, computed radiography (CR) plates, CR readers, cassettes, digital fluorography systems, couches, anti-scatter grids, beam modifiers (eg filters and wedges), gamma cameras, computerised radiotherapy treatment planning systems and film processing units. 7 For the purpose of IRR99 regulation 32(5-8) radiation equipment means equipment which delivers ionising radiation to the person undergoing medical exposure and equipment which directly controls the extent of the exposure. Such equipment would include X-ray tubes, AEC devices (including image intensifiers), beam collimators, direct digital radiography systems, radiotherapy treatment machines and radionuclide dose calibrators. The section Exposures greater than intended relates solely to radiation equipment. 8 Employers should be aware that technology evolves and equipment design and specification may change significantly. Procedures and associated programmes should therefore be reviewed regularly to ensure they remain relevant to any new equipment and in compliance with IRR99 regulation This guidance applies to new and second-hand equipment and to equipment under loan (including for demonstration purposes, where that equipment is used in connection with medical exposures). General duties of employers under IRR99 10 The employers referred to in the introduction are defined as radiation employers in IRR99 regulation 2(1), as they either work or intend to work with ionising radiation. For simplicity, these are referred to as employers throughout the rest of this document. Such employers have a number of duties under IRR99 in relation to employees and members of the public. These are discussed in detail in L Employers must comply with aspects of IRR99 which apply to exposure to ionising radiation of employees and other persons (other than those undergoing medical exposure). Co-operation between employers 11 Employers who either share a workplace or an item of equipment used in connection with medical exposures, must co-operate with each other regarding the control of associated risks and compliance with IRR99 (and other relevant legislation). This means sharing information on the types of exposure for which the equipment is suitable, and the maintenance and quality assurance requirements needed to ensure its effectiveness at restricting exposure. If there are any modifications to the equipment, or new equipment is brought into service, the employer who authorises the modifications or new equipment should immediately inform other employers who use the equipment of the changes, and ensure that they are aware of anything that may affect the health and safety of the staff or the patient (see paragraphs of L121 3 ). 12 There may be situations in which employers are in control of premises where equipment is used in connection with medical exposures, but do not have specific responsibility for that equipment. Examples include companies owning premises where equipment that has been purchased or leased under publicprivate partnership schemes (or private finance initiative schemes), is used by another employer. In such situations it is essential that all employers cooperate to comply with IRR99 and to provide adequate information on associated risks. The detailed arrangements for co-operation and their periodic review should be agreed at the contract stage. It should be noted that the duties of IRR99 regulation 32 are placed on employers who have to any extent control over equipment used in medical exposures. Duties of manufacturers, suppliers, installers and erectors of articles for use in work with ionising radiation Design and construction 13 IRR99 regulation 31(1) extends the duties under Section 6 of the Health and Safety at Work etc Act (HSWA), so that there is a duty on manufacturers, importers and suppliers to ensure that any article for use with ionising radiation is so designed and constructed to restrict, so far as is reasonably practicable, the exposures of persons to ionising radiation arising out of its use. Although this requirement does not apply to the protection of persons undergoing medical exposures, it does apply to the exposure of other persons, including staff that 2 of 18 pages
3 carry out the exposure and members of the public. As discussed in the section Selection of equipment, IRR99 regulation 32(1) requires the employer to consider restriction of patient exposure when purchasing equipment. 14 The manufacturer is responsible for ensuring that their product complies with all the Essential Requirements of Annex 1 of the Medical Devices Directive (MDD) 93/42/EEC, 5 the Consumer Protection Act, 6 and the Medical Devices Regulations (MDR) 2002 (SI 2002 No 618). 7 The MDD is a single market measure designed to remove technical barriers to trade by harmonising safety and performance requirements for medical devices. The CE mark is applied to denote conformity, enabling manufacturers to market their products freely throughout the European Community without having to abide by any further national controls. HSE work closely with the Medicines & Healthcare products Regulatory Agency (MHRA), Scottish Health Department, Welsh Assembly Government, and Department of Health and Social Security (Northern Ireland), to ensure that manufacturers and employers comply with the relevant aspects of both IRR99 and MDR. 7 Critical examination 15 IRR99 regulation 31(2) requires a person who erects or installs an article for use in work with ionising radiation to: where appropriate, undertake a critical examination of the way in which the article was erected or installed; consult an RPA with regard to the nature and extent of the critical examination and its results; and provide the employer with adequate information about proper use, testing and maintenance of the article. This regulation does apply to the protection of persons undergoing a medical exposure. 16 A critical examination will be appropriate in cases where there might be radiation protection implications, for either staff or patients, associated with the incorrect installation of the equipment. Examples include the failure of the safety features or warning devices to operate correctly, poor location or inadequate shielding. The requirement applies to: installation of equipment (whether new, secondhand or refurbished); relocation of existing equipment (including relocation within the same premises); and in cases following major service or repair work where there may be radiation protection implications, eg following the fitting of: - a replacement x-ray tube or automatic exposure controls (AECs) (including to mobile or portable units); - a klystron on a linear accelerator; or - an ion chamber on a dose calibrator. 17 The critical examination should include the safety features, warning devices and protection from (unintended) exposure to ionising radiation provided for the protection of the patient. For example, AECs, exposure interlocks, back up timer, light/x-ray beam alignment and shielding from leakage radiation, may come within the scope of the critical examination. 18 Where equipment components arrive on site ready-assembled, the person undertaking the critical examination on site may wish to request the records of assembly from the factory, and any critical examination associated with them. While mobile equipment that is delivered fully assembled does not formally require a critical examination (if erected in the European Union, it should have been critically examined by the manufacturer), employers are advised to check that the safety features are functioning prior to first clinical use, to demonstrate compliance with the requirements of regulation 32(1). 19 Portable equipment, eg that used for domiciliary radiography, should be tested appropriately and consideration given to the extent and suitability of shielding prior to any exposure taking place. The degree of testing, and what constitutes an acceptable result, should be agreed with the RPA prior to the domiciliary visit. It may be useful to use a written protocol for the purpose. 20 While the duty to carry out a critical examination rests with the installer, the RPA consulted may be either the RPA appointed by the installer or the employer's own RPA. The employer and the installer should establish at the contract stage who will carry out the critical examination, ie medical physics staff or the service engineer, and which RPA will take part. The RPA should have the relevant knowledge and experience in order to provide advice in relation to the critical examination. Following a satisfactory outcome to the critical examination, it is in the interests of both parties for the installer to prepare a report, endorsed by the RPA, to confirm that this is the case. It would be prudent to keep this report with the maintenance record of the equipment during its operational life. If the outcome of the critical examination is unsatisfactory, then the failure should be reported to the employer, remedial action taken and the examination repeated. In any event the employer should not bring any equipment into use unless, if appropriate, a critical examination has been satisfactorily completed. 3 of 18 pages
4 Requirements of IRR99 specific to equipment used in connection with medical exposures: regulation 32 Selection of equipment 21 An employer purchasing or leasing new, secondhand or transferred equipment should be satisfied that it is suitable and appropriate for the medical exposure for which it will be used, including the magnitude and distribution of patient dose. The employer should also consider the extent to which the equipment incorporates safety devices that inhibit operation in the event of a serious fault. At the selection stage the employer should consider the maintenance, servicing and quality assurance testing requirements. Maintenance of equipment is discussed in more detail in paragraph 34 and Appendix 1. The employer may find it helpful to consult the RPA and the medical physics expert (MPE) when selecting equipment to be used in connection with medical exposures. 22 When selecting new equipment (including ancillary equipment), the employer should consider any relevant reports or guidance provided by organisations such as the Radiation Protection Division of the Health Protection Agency (HPA(RPD)), the Institute of Physics and Engineering in Medicine (IPEM), British Institute of Radiology (BIR), National Health Service Breast Screening Programme (NHSBSP), British Nuclear Medicine Society (BNMS) and the Medicines & Healthcare products Regulatory Agency (MHRA) (this list is not exhaustive). For example, the MHRA has issued Medical Electrical Installation Guidance Notes (MEIGaN) 8 relating to the electrical installation of permanently installed medical devices (diagnostic imaging and radiotherapy installations). 23 Information can be obtained from the following sources: Evaluation reports on some new equipment are produced for the NHS Purchasing and Supply Agency (PASA). Evaluation reports published since 2002 are available electronically to download (as pdfs) from the Centre of Evidence-based Purchasing website at (use the contact details on the website to enquire about less recent evaluations). The KCARE programme SafeSpecs is a web application designed to assist in the process of preparing a technical specification which can be sent to X-ray equipment manufacturers (excluding those of CT scanners and X-ray Bone Densitometers). It will also aid in the interpretation of the data supplied, and provide general background information on X-ray equipment technology. SafeSpecs is available free to NHS users, and for a small fee for others, via the KCARE website ( For CT scanners the ImPACT group can provide equivalent specification, and further information is available on their website ( 24 All radiation equipment used for diagnosis installed for the first time after 1 January 2000 must, where practicable, be fitted with suitable means to inform the user of the quantity of radiation being produced (regulation 32(2)). Suitable means may include a device which shows the product of X-ray tube current and exposure time or a dose area product (DAP) meter. These devices may not be suitable or sufficient for certain types of equipment, such as dental, mammographic, CT and bone densitometry X-ray equipment, and alternatives should be considered. Examples include devices to display the following parameters: exposure time on fixed kv/ma dental X-ray units; kv, mas, target and filter (also compressed breast thickness where available) on mammographic units; CTDI on CT scanners; and scan time on bone densitometry units. Installation of equipment 25 The employer should find the installer s critical examination report (see paragraphs 15-20) useful in helping to ensure that the equipment has been properly installed. The employer should bear in mind that the critical examination is not a test of the initial integrity of the equipment but is a check to ensure that safety features and warning devices operate correctly, and that the radiation protection features are adequate. Further tests before use, including image quality, patient dose and electrical and mechanical safety will be required (see paragraphs 26-33). Quality assurance programmes 26 The employer is required to provide a suitable quality assurance (QA) programme for equipment used in connection with medical exposures (including ancillary equipment), IRR99 regulation 32(3,4). The QA programme must include the following: tests on equipment prior to its first clinical use (these tests are separate from those of the critical examination and will probably include acceptance tests (to confirm that the equipment is functioning as intended) and commissioning tests (to determine baseline results against which to compare future measurements and to determine appropriate clinical exposure settings)); adequate tests of equipment performance at appropriate intervals and after any major maintenance procedure; and where appropriate, measurements at suitable intervals to assess patient doses from radiation equipment. 4 of 18 pages
5 Testing may be carried out by the employer s own staff or by contractors, or both. Aspects of routine testing that may only be carried out by, or with the assistance of, the service engineer, should be agreed at the contract stage, together with the arrangements for test results to be transmitted to the employer without delay. It is the employer s responsibility to ensure that the tests are adequate and that appropriate remedial action is taken where necessary. 27 A suitable QA programme is one that clearly establishes planned and systematic actions necessary to ensure the equipment will satisfy the requirements of IRR99 regulation 32(1). The RPA should be consulted regarding QA programmes for equipment (including ancillary equipment). The extent of the programme will depend on the nature and range of equipment in use. The following should be considered when drawing up and implementing a suitable QA programme: advice from the manufacturer, particularly where the normal operating mode of the equipment (for example, some fluoroscopy/fluorography units) does not allow for non-interventional measurement of exposure factors such as operating potential; the workload of equipment. This will affect the required frequency of routine testing and maintenance. The frequency of testing should be clearly specified and written in QA schedules. As equipment ages or workload changes, the frequency of testing may need to be adjusted; results from commissioning tests (as defined in paragraph 26), which should be used as a baseline when considering the results of future tests. The magnitude of any deviation from the measured baseline value at which remedial action is required (remedial level), or immediate removal from clinical use is required (suspension level) should be clearly specified in QA schedules; and the complexity of equipment, which will affect the number of parameters to be measured in a QA programme and hence the time taken for tests to be carried out. Special consideration should be given to QA programmes for equipment which is used in connection with: medical exposures of children; a health screening programme; or high doses to the patient (such as interventional radiology, computed tomography, nuclear medicine or radiotherapy). Justification for these examinations or treatments (under IR(ME)R) 2 is different to those for general medical exposure of adults, as are the consequences of any exposure greater than intended. Considerations should include the amount, type and frequency of tests included in the QA programme. Special consideration should also be given to QA programmes for equipment incorporating new or developing technologies (for example computed radiography systems, direct digital radiography systems and intensity modulated radiotherapy systems). In such cases, advice should be sought from the manufacturer and any relevant organisations (such as MHRA, IPEM, BIR or HPA(RPD)) on appropriate parameters to measure, test methods and interpretation of results. Information regarding radiation dose issues with digital radiography systems is provided on the MHRA website at: GET_PAGE&useSecondary=true&ssDocName=CON Alternatively you can follow the route from the home page following these links: safety information/general safety information and advice/general and technical information/radiation dose. It may be useful to see how the requirements of other legislation will be met within the QA programme; such as that relating to mechanical or electrical safety. 28 The QA programme should include written procedures for recording: faults on equipment; whether the equipment was removed from clinical use; what action was necessary to correct the fault; and tests made before the equipment was returned to clinical use. Such fault logs provide an invaluable record of equipment performance and help identify problems that have the potential to cause future exposures much greater than intended (discussed in paragraphs 40-44). Equipment currently unfit for clinical use should be disabled (eg by isolating (and locking off) the unit from the mains supply or, in the case of a mobile unit, removing the main fuse) and posting a notice on the control panel. 29 The QA programme should clearly specify who has responsibility for organising the various elements, carrying out and recording tests or dose assessment, and for acting on adverse findings (such as suspending equipment from use). Results of tests carried out should be properly documented as part of the QA programme. 30 The employer should generally identify suitable test equipment as part of the QA programme, and will need to make arrangements to ensure that this is 5 of 18 pages
6 provided, maintained and (if relevant) calibrated to an appropriate standard prior to use, and at suitable intervals. 31 Correct calibration of radiotherapy equipment is vital if the prescribed dose is to be delivered to each patient. Advice on calibration of such equipment is given in the Medical and Dental Guidance Notes (MDGN) 9 and IPEM Much equipment used in connection with medical exposures is supplied with programmable electronic control systems, ie computer-based systems which control the operation of the equipment. Hardware failures of such control systems, and/or errors in the software, can cause the equipment to malfunction and may lead to an exposure greater (or less) than intended to ionising radiation. This is a complex area and employers should be aware that the International Electrotechnical Commission (IEC) has produced comprehensive guidance on programmable electronic systems in safety related applications. 11 This document is the official IEC statement of safety philosophy for Programmable Electronic system Safety (PES) and where applicable, it will be adopted in the preparation of other IEC standards that have requirements for PES. Typically, this document will be used by industry sectors as the basis of sectorspecific and national standards 9,10,12,13 and HSE will promote the use of the standard in relevant industry sectors and stand-alone applications. 33 General advice is given in L121 3 paragraphs In addition, detailed information on quality assurance programmes for specific modalities (including criteria of acceptability, remedial and suspension levels and assessment of representative 10, 12, 25 patient doses) are available. Maintenance of equipment 34 Equipment used in connection with medical exposure should be under some form of maintenance/service contract, often with the manufacturer or installer. Alternatively, arrangements may be made with in-house engineers and technical staff. In this guidance, all reference to service engineers and service or maintenance includes such in-house arrangements. In any event, the employer has the legal responsibility for ensuring that equipment is properly maintained and this responsibility cannot be divested to a third party. 35 It is therefore recommended that employers consult their RPA when the contract is drawn up or reviewed, in order to ensure that the equipment maintenance arrangements are appropriate and assist the employer in meeting their legal duties with regard to IRR99 regulation 32. Appendix 1 details the issues to be considered by employers when arranging for maintenance of equipment used in connection with medical exposures. 36 The employer should establish clear procedures for acceptance of equipment back into clinical use following service or repair. The procedures may specify the information that will be provided by the engineer (eg test results), together with values of acceptable results of those tests (ie acceptable levels of performance). These procedures might be referenced in the local rules required by IRR99 regulation 17(1). No radiation equipment or ancillary equipment should be accepted back into service until a competent employer representative (such as a senior radiographer or medical physicist) has reviewed the service report/summary to confirm that the equipment has been left in a state fit for use, and that no alterations have been carried out which may significantly affect patient doses or radiation safety. If such alterations have been undertaken, the user should seek advice (for example from the RPA or MPE) before bringing the equipment back into use, as further tests may be needed to verify its performance. 37 Employers should keep adequate records of equipment maintenance. There should be sufficient information recorded to be able to identify, for example, any modification which may affect the patient dose or radiation safety (including software upgrades to electronic control systems). It may be convenient to keep the records of Quality Assurance testing alongside the maintenance record. Records should be maintained for the life of the equipment. When equipment reaches the end of its life the employer must decide whether or not the records should be maintained with respect to the likelihood of liabilities arising from its use. Note that this recommendation is in addition to the requirement for an equipment inventory under IR(ME)R. 2 Ageing equipment 38 If equipment has deteriorated in performance since it was installed, to a level which is significantly poorer than any currently acceptable level, it is unlikely to satisfy the requirements of IRR99 regulation 32(1). An example of such deterioration would be if the use of equipment resulted in the doses from medical exposures being significantly greater than local or national diagnostic reference levels (DRLs). The employer should therefore develop a programme for the progressive replacement of equipment before this situation is reached. In determining whether continued use of ageing equipment is justified, and in assigning priorities for equipment replacement, consideration should be given to the following: equipment performance (including comparison of recent performance test results with remedial and suspension levels); 6 of 18 pages
7 the magnitude of patient doses resulting from the use of that equipment; the frequency of use and number of patients likely to be affected by the continued use of the equipment; the range of patients likely to be affected by continued use (eg age, clinical condition). Such factors enable a determination of risk versus benefit for continued use of the equipment; and the cost of replacement. However, cost cannot normally be used as a legitimate reason for patients being subject to risks in excess of those normally encountered for the examinations undertaken. The employer should involve the RPA and MPE in the development of equipment replacement programmes. 39 The employer should consider the destination of redundant equipment, eg if the equipment is supplied to a veterinary practice for use in animal radiology, the employer has responsibilities under IRR99 regulation 31. Equipment that is going to be scrapped should be disabled, eg portable dental equipment should have the mains lead removed or circuit boards removed and separated from the generator. Any radioactive sources must be removed from redundant equipment, disposed of via an authorised route and properly accounted for. If it is not reasonably practicable to remove sources from equipment then the equipment itself must be disposed of via an authorised route. Exposures greater than intended: regulation 32(5), (6) and (7) Prevention and detection of exposures greater than intended 40 The employer must take all reasonably practicable steps to prevent failure of radiation equipment where such failure could result in exposure of patients to ionising radiation greater than intended. The employer must also implement steps to limit the consequence of such failure (IRR99 regulation 32(5)). 41 Advice on prevention of equipment failure is given in L121 3 paragraphs If the employer becomes aware of potential defects in equipment during use, they will need to assess, in consultation with the supplier (or the service agent/original manufacturer) and RPA, whether any further action is necessary. 42 If the failure of a single component can give rise to an unintended exposure of the patient, the employer may need to have additional controls in place. 9,15,21,24 Suitable controls include ensuring that the exposure is automatically terminated within an appropriate preset time, tube current-time product, or dose. Where this is not reasonably practicable, the employer may need to ensure that failure of equipment to terminate an exposure correctly is immediately detectable, so that the operator can take action. 43 In any case, the employer should have in place mechanisms to quickly detect malfunctions and/or defects in radiation equipment. Special care should be taken in cases where detection of exposures greater than intended is not straightforward; for example in the case of: some modern imaging technology (such as computed or direct digital radiography); during radiotherapy; or the use of radionuclide calibrators for nuclear medicine. Poor image quality cannot be relied upon to identify malfunctions in diagnostic X-ray or radionuclide imaging equipment. The employer should ensure that checks are made after each radiographic exposure and, in the case of radionuclide dose calibrators, on at least a daily basis prior to administrations of radioactive material, so that exposures greater than intended are prevented or quickly detected. 44 The employer will also need to provide a contingency plan for responding to equipment failures or malfunction. Following equipment failure and subsequent repair, the employer (in consultation with the RPA) should consider the likelihood and consequences of recurrence before returning the equipment to clinical use. Employers should consider the advice given in ICRP and ICRP Investigation and notification of patient exposures much greater than intended: IRR99 regulation 32(6) and (7) 45 When an employer suspects, or is informed, that an incident may have occurred in which a patient was exposed to ionising radiation to an extent much greater than intended as a result of malfunction or defect in any radiation equipment, an immediate investigation must be carried out. Unless the investigation shows beyond reasonable doubt that no such incident has occurred, the employer must immediately notify HSE. This notification should not be delayed pending further investigation. The employer is then required to make or arrange for a detailed investigation of the circumstances of the exposure, and an assessment of the dose received. 46 The employer must keep the initial investigation report for a period of at least two years from the date it was made. If a detailed investigation is made, the report must be kept for a period of at least 50 years from the date it was made (IRR99 regulation 32(7)). 7 of 18 pages
8 47 Notification is not required for increased patient exposures due to ageing equipment whose performance has gradually deteriorated over time, provided the employer is fully aware of this deterioration in performance, and has made a positive decision to continue with its use. Performance specifications (remedial and suspension levels in the QA programme) will have been set for such equipment in consultation with the RPA and MPE. If equipment fails to meet the specification, action should be taken (see paragraphs 26-33). However, if the employer becomes aware of a sudden significant deterioration in the performance of the equipment, and a patient has received an exposure much greater than intended as a consequence, it will be necessary to treat this as an incident subject to notification to HSE under IRR99 regulation 32(7). 48 The requirement to notify HSE is solely in relation to a malfunction or defect in radiation equipment; it does not apply to incidents that occur as a result of errors by those persons performing medical exposures. Incidents caused by human error may need to be notified to the appropriate authority (Department of Health (DH), Scottish Health Department (SEHD), Welsh Assembly Government, or Department of Health Social Services and Public Safety (Northern Ireland) (DHSS&PS) under the requirements of IR(ME)R 2 )). 49 Sometimes a diagnostic medical exposure is repeated because of failure of ancillary equipment, eg a film processor or gamma camera and associated computer systems. Exposures that in total are higher than normal as a result of such failures are not reportable to HSE under IRR99 regulation 32(7). However, in such cases HSE would expect, as a matter of good practice, the employer to undertake an investigation of the causes and consequences of the incident and implement appropriate measures to limit the likelihood of recurrence. Such incidents may also involve an element of human error, eg failure to recognise that a problem has occurred before a number of repeat exposures have been carried out. In these cases the employer must consider whether notification to the appropriate authority DH under IR(ME)R 2 is required. 50 MDR 7 require manufacturers to report to the MHRA, UK competent Authority, any incident which led to, or might have led to, death or serious injury to patients, users and others. HSE therefore recommend users to report any such incidents to manufacturers. In addition, the MHRA, Scottish Healthcare Supplies and DHSS&PS operate a scheme for users to report any adverse incidents concerning medical devices. The employer is encouraged to report such incidents (even if they have not resulted in patient exposures much greater than intended) as this enables the Competent authority (MHRA) to take appropriate actions with the manufacturer. 51 IRR99 regulation 32 defines the requirement for the provision and maintenance of equipment to prevent exposures that are greater than intended. Exposures that are lower than intended can also have serious consequences, especially for therapy procedures (see Appendix 3). An investigation under IRR99 regulation 32(6) is not formally required if the malfunction or defect results in exposures less than intended. However, as a matter of good practice, the employer may wish to carry out their own investigations in such circumstances. Guidelines on notification 52 If the initial investigation shows beyond reasonable doubt that no incident has occurred or the patient has not received an exposure much greater than intended, the employer should make a simple report of the circumstances that led to the investigation and attach the accompanying evidence. The report is to be kept for two years, IRR99 regulation 32(7). 53 Where the initial investigation indicates that an exposure much greater than intended might have occurred, the employer must notify HSE immediately. Notification should not be delayed pending the outcome of a more detailed investigation. HSE needs to be notified of the basic details of the incident (type of radiation exposure, brief description of events, if possible an estimate of dose received and number of patients affected). The incident should be reported to the local HSE office, the address of which may be obtained from the HSE Infoline on or from the HSE website at A fax would suffice. The notification is made under IRR99 regulation 32(6) not RIDDOR The employer will need to decide whether or not the patient exposure was much greater than intended. This requires a professional judgement to be made and the employer is advised to consult the RPA and the MPE. However, to assist employers, the guidance in Table 1, Appendix 2 should be used for determining, under most circumstances, when incidents are likely to be notifiable. 55 To use the guidelines in Table 1, Appendix 2, the ratio of the suspected exposure to the intended exposure must be determined and compared with the appropriate factor in the table. If the suspected exposure was greater than the intended exposure by at least the factor shown, then HSE should be notified of the incident. 56 In some cases, the guidelines in Table 1, Appendix 2 will be of limited assistance, for example: 8 of 18 pages
9 if the incident involves diagnostic X-ray equipment which does not give post-exposure readout of tube loading factors and does not allow exposure factors to be selected; or if the remote after-loading equipment failed to achieve the correct source positioning. In these cases, it may be difficult to apply the guidelines to a single quantity and thus professional judgement will be particularly important in deciding whether or not to notify HSE of the incident. 57 There may be other incidents involving radiation equipment that the employer ought to investigate, but which fall outside the guidelines for HSE notification. These might include incidents involving a group of individuals, each of whom receives a relatively small additional exposure as a result of an equipment fault. 58 Table 1, Appendix 2 is only relevant to the requirement to notify HSE under IRR99 regulation 32(6); it is not intended to imply that incidents involving a lesser degree of overexposure are in some way acceptable. Regardless of whether or not a particular incident is notifiable, the employer will have responsibilities to minimise the occurrence of such incidents so far as reasonably practicable and to restrict the consequences of patient exposure. 59 The guidelines on notification refer to unintended exposures that may merit an investigation by HSE inspectors. Further explanation of the figures in Table 1 is given in Appendix An unintended exposure of a foetus during a medical exposure is not notifiable to HSE under IRR99, as a foetus is not a person independent of the mother. An accidental exposure of a foetus during a medical exposure is notifiable as part of the mother s exposure, if it results from malfunction of equipment used in connection with medical exposures. Detailed investigation of incidents Objectives of investigation 61 There are four main objectives in investigating incidents. These are: to establish what happened; to identify the defect or malfunction in the radiation equipment and to establish its causes; to decide upon and implement remedial action to prevent a recurrence; and to estimate the dose received by all persons involved in the incident. Appendix 3 gives advice on assessing patient doses as part of the investigation. It will probably be useful to analyse the immediate actions which were taken. The National Patient Safety Agency (NPSA) provides e-based training on root cause analysis, which may prove useful in such investigations, at: Persons to involve in the investigation 62 It is important to identify from the outset, or as early as possible, the persons who will be involved in the investigation, including those conducting the investigation and those whose evidence is to be considered. People who should always be involved include: the person in charge of the department where the incident took place; the person(s) acting as operator during the exposure (as defined by IR(ME)R 2 ); the service engineer who examined the equipment following the incident; the person who was responsible for quality assurance on the equipment; the RPA; and the RPS. Others who may be involved include: the person acting as practitioner for the exposure (as defined by IR(ME)R 2 ) as appropriate; the MPE; and the patient. The above list is not exhaustive. Sources of information 63 Sources of information that may prove helpful both in determination of what happened, and in the assessment of the dose received, include: the settings on the equipment, plus any equipment generated logfiles or performance data; recorded exposure parameters, such as for diagnostic radiology, including: - DAP readings; - operating potential (kv); - exposure time; - Focus skin distance (where available); - product of tube current and time; and for radiotherapy treatment machine monitor units: - beam time; - beam modifiers; and - geometric settings); other measures of exposure (such as for diagnostic radiology: film blackening, image parameters in computed radiography systems (eg CR exposure indicator); for nuclear medicine: count rate in a gamma camera image)). In addition, information may be gained from any 9 of 18 pages
10 patient TLD measurements undertaken as part of dose assessment programmes; QA records (including training records of those persons carrying out QA duties); training records of operators to eliminate human error as a major factor (such incidents fall under the requirements of IR(ME)R); 2 any fault reports and service records; any tests on equipment carried out for the purpose of the investigation; and an account of what happened by the person operating the equipment. Investigation report 64 It is recommended that the investigation report should include the following: the key facts concerning the incident; a record of the calculations and measurements that were made; and recommendations to avoid recurrence of the incident. 65 The report should be signed and dated by the person who prepared it. It should be copied to the manufacturer or maintenance contractor and retained by the employer for a period of 50 years. It is suggested that a record also be made of the actions taken to implement the report's recommendations. It may also be helpful to include the identity details of the patients involved (name, NHS number, date of birth etc) in the report. It should be noted that any such report should be given appropriate confidentiality, ie marked, stored and distributed accordingly (to comply with the Data Protection Act). 66 Nothing in this publication is intended to indicate whether or not patients should be informed of any incident resulting from malfunction or defect in equipment used in connection with medical exposure and the possible consequences of that exposure. Suitable guidance on this matter may be found in NHS publications: An organisation with a memory (2002) 27 and Building a safer NHS for patients (2001) of 18 pages
11 Appendix 1: Maintenance of equipment used in connection with medical exposures To include any maintenance which may affect the radiation safety of equipment used in connection with medical exposure, or the capability of the equipment to restrict exposure to any person undergoing a medical exposure. 1. The employer and service contractor (or in-house provider) should agree the tests to be undertaken by the contractors and the pass/fail criteria to be applied to these tests. These should be consistent with the remedial and suspension levels chosen by the employer for the equipment. The criteria will depend on one of the following: advised to highlight in the service report any faults that have not been completely corrected or identified, if those faults could foreseeably lead to a patient exposure that is significantly greater than intended (see paragraphs 40-66). The employer (in consultation with the RPA and where appropriate, mechanical or electrical engineers) should then assess the risk of such occurrence and decide whether bringing the equipment back into service is justified. Following service or maintenance of radiotherapy equipment, the QA tests should identify any changes which might affect patient dose. in the case of new equipment, the original specification and/or results of acceptance or commissioning tests. The results of these tests should be used as a baseline with which to compare future performance tests; when original records are not available for equipment, tests should be undertaken to establish a baseline performance, the results of these tests can then be used as a basis for future comparisons; or for equipment which is no longer able to meet original specification, the minimum acceptable standards may be based on the advice given by the relevant professional organisation or EC criteria. 2. As with frequency of routine tests for the QA schedule, intervals between preventive maintenance inspections should reflect the workload and age of individual items of equipment. 3. Service contractors have an obligation to ensure that any of their employees who undertake maintenance of X-ray equipment are adequately trained and competent to do so (MHRA Device Bulletins DB9801 (Chapter 8) 29 and DB2000(02) 30 ). The employer must ensure that any third party service agent can provide the necessary level of expertise. Software diagnostic packages for X-ray generators may need to be licensed. 4. It would be advisable for the service report to refer to agreed tests and the pass/fail criteria applied to the equipment. It is important that any work carried out on the equipment that may significantly affect the patient doses (such as changes to X-ray output or factors relating to image quality) is noted within the report. The criteria should specify what constitutes a significant change in the radiation dose received by patients undergoing medical exposures (and this might be included in the QA programme). The engineer is 11 of 18 pages
12 Appendix 2 Table 1 Guidelines for notification of incidents involving radiation equipment used in connection with medical exposure Type of diagnostic examination Guideline multiplying factor applied to intended dose Interventional radiology, radiographic, 1.5 and fluoroscopic procedures involving contrast agents, nuclear medicine with intended dose >5mSv and computed tomography examinations. Mammography, nuclear medicine with 10 intended E)5mSv but >0.5mSv, all other radiographic examinations not referred to elsewhere in this table. Radiography of extremities, skull, 20 dentition, shoulder, chest, elbow, knee, and nuclear medicine with intended E)0.5mSv. Type of treatment Guideline multiplying factor applied to intended dose Beam therapy, brachytherapy Unsealed radionuclide therapy Explanation of figures in Table (whole course) or 1.2 (any fraction) 1.2 (any administration) Diagnostic procedures 1. For diagnostic procedures, the guideline factors reflect a level of performance that is significantly outside the normal specification of equipment used for medical exposures. 2. The notification guidelines in Table 1, Appendix 2 are broadly representative of patient exposure, ie effective dose or mean glandular dose. Suitable measurements for determining these quantities are: dose-area product; duration of exposure; product of tube current and time (mas); volume of tissue irradiated; and/or activity administered. 3. For diagnostic procedures in nuclear medicine, the radiation equipment likely to be involved in any unintended exposure would be the equipment used to determine the activity of the radiopharmaceutical administered. However, IRR99 regulation 32(6) is not relevant to equipment which cannot be used directly to control the extent of the exposure, for example if radionuclide calibrators are being used only to provide a rough check on volumetric measurements made for radionuclides, eg 51 Cr. 4. The guidelines for nuclear medicine are based on the presumption that the intended dose would be equivalent to the diagnostic reference level of dose (in terms of activity administered for a particular examination determined locally within the national framework provided by Administration of Radioactive Substances Advisory Committee (ARSAC) in their notes for guidance. 31 ) 5. The factors in Table 1 are intended to be applied to the entire patient exposure received during the particular procedure in which the faulty radiation equipment was used. 6. HSE should not be notified in cases where equipment failure leads to an exposure which is less than intended and a repeat is carried out, leading to a total exposure for the procedure which is 1.5 times greater than the intended exposure. However, as a matter of good practice, the employer may wish to carry out their own investigations in such circumstances. Therapeutic procedures 7. In the case of therapeutic procedures (including palliative treatments) for malignant and non-malignant conditions, guidelines are given for a single patient treatment fraction and for a whole course of treatment. The figures of 10% excess exposure for a course of treatment and 20% for a single fraction ensure that the majority of significant incidents would be notifiable. The figures take account of normal clinical tolerance levels of the final delivered dose, and the possibility of unexpected deterministic effects. The figures also reflect a level of performance that is significantly outside the normal specification of equipment used for medical exposure. 8. For radiotherapy and brachytherapy, the guideline factors can be applied to the prescribed dose to the target volume and/or the intended dose to any critical tissues. For therapeutic nuclear medicine, the guideline factor is intended to be applied to the activity of the radionuclide administered to the patient. General comments on multiplying factors in Table 1 9. The requirement to notify HSE about exposures much greater than intended (IRR99 regulation 32(7)) is to enable HSE to further investigate incidents that may have resulted from a failure to comply with IRR99 regulation 32. The guideline factors are not solely based on increased risk to the individual (a very 12 of 18 pages
13 complicated determination, which must take into account the range of patient ages, clinical conditions and actual magnitude of the exposure). The single factor of 50% excess (diagnostic) and factors of 10% and 20% (therapeutic) are considered to adequately reflect incidents which may require further investigation by HSE, while retaining a level of simplicity and transparency. 13 of 18 pages
14 Appendix 3: Assessment of doses to patients following exposures greater than intended Recommended quantities for assessing dose received 1. An assessment of the dose received, as required by IRR99 regulation 32(6), allows the clinician responsible for the patient to make an assessment of the potential harm caused by the unintended exposure. Methods for assessing dose 2. In order to provide uniformity in the assessment of doses, possible dosimetric approaches appropriate to seven general categories of procedure are summarised in Table 2. With the exception of radiotherapy and incidents involving only paediatric patients, the methods provide estimates of the effective dose to a standard phantom that is representative of an adult patient of average size and composition. It is strongly recommended that the dose assessment be carried out in consultation with the RPA. 3. It is recommended that estimates be made both of the dose actually delivered in the incident and the dose that would have been received if the incident had not occurred. This will allow a comparison to be made. An incident may involve one or more diagnostic or therapeutic exposures. Table 2 Summary of recommended strategies for dosimetric assessment of incidents arising from different categories of medical exposure Procedure Dose quantity a Method b Radiography E Estimates of ESD per film or DAP rad and normalised dose data c D max Appropriate summation of ESD per film Mammography D gland Estimates of entrance air KERMA per film and normalised dose data E c Estimates of doses to other organs Fluoroscopy & Fluorography E Estimates of DAP flouro or entrance surface dose and normalised dose data c D max Estimate of entrance surface dose rate and knowledge of procedure used CT E CTDI or DLP and normalised dose data c D max Measurements on head or body phantom Nuclear medicine E Administered activity and normalised dose data (31-34) Paediatrics c E, D max Use age-specific data when available Radiotherapy D crit Knowledge of dose to radiosensitive organs Notes (a) E = effective dose Dmax = maximum surface dose Dgland = dose to glandular tissue of the breast Dcrit = absorbed doses to the critical organs (b) ESD = entrance surface dose (with backscatter) DAPrad = dose-area product for radiographic exposure DAPflouro= dose-area product for fluoroscopic and/or fluorographic exposure CTDI = Axial CT dose in free air DLP = Dose Length Product for a CT examination (c) These quantities are to be calculated only as necessary when considering the following categories of medical exposure: Dmax = radiography, fluoroscopy, paediatrics and CT E = mammography 14 of 18 pages
15 References 1. Ionising Radiations Regulations 1999 SI 1999/3232 The Stationery Office 1999 ISBN Ionising Radiation (Medical Exposure) Regulations 2000 SI 2000/1059 The Stationery Office 2000 ISBN Work with ionising radiation. Ionising Radiations Regulations Approved Code of Practice and guidance L121 HSE Books 2000 ISBN Health and Safety at Work etc Act 1974 Ch37 The Stationery Office 1974 ISBN Medical Devices Directive Council Directive 93/42/EEC Official Journal L Consumer Protection Act 1987 The Stationery Office 1987 ISBN Medical Devices Regulations 2002 SI 2002/618 HMSO 2002 ISBN Medical Electrical Installation Guidance Notes MHRA 2005 Available via the MHRA website at: 9. Medical and Dental Guidance Notes: A good practice guide on all aspects of ionising radiation protection in the clinical environment Institute of Physics and Engineering in Medicine 2002 York ISBN IPEM Report 81: Physical aspects of quality control in radiotherapy Institute of Physics and Engineering in Medicine 1999 York ISBN X 11. IEC International Standard Functional safety of electrical/electronic/programmable electronic safety-related systems Parts 1-7 IEC 2005 International Electrotechnical Commission ISBN IPSM Report 54: Commissioning and quality assurance of linear accelerators (With a supplement on gamma units reproduction copy) Institute of Physical Science 1988 York ISBN IPEM Report 68: A guide to commissioning and quality control of treatment planning systems Institute of Physics and Engineering in Medicine 1996 York ISBN Guidance notes for dental practitioners on the safe use of X-ray equipment National Radiological Protection Board 2001 NRPB Chilton ISBN This document can be found on the Internet at: ns/dental_guidance_notes.htm 15. European Commission 1997 Radiation protection 91: Criteria for acceptability of radiological (including radiotherapy) and nuclear medicine installations Luxembourg: Office for Official Publications of the European Communities ISBN IPEM Report 77: Recommended standards for routine performance testing of diagnostic X-ray imaging systems Institute of Physics and Engineering in Medicine 1997 York ISBN Assurance of quality in the diagnostic imaging department British Institute of Radiology 2002 ISBN IPSM, NRPB, CoR National protocol for patient dose measurements in diagnostic radiology National Radiological Protection Board 1992 NRPB Chilton 19. Guidelines on radiology standards for primary dental care Report by the Royal College of Radiologists and the National Radiological Protection Board NRPB Documents 5 (3) National Radiological Protection Board 1994 NRPB Chilton ISBN IPSM Report 89: The commissioning and routine testing of mammographic x-ray systems Institute of Physics and Engineering in Medicine 2005 York ISBN Avoidance of radiation injuries from medical interventional procedures No. 85 ICRP Publication: Annals of the ICRP 30 (2) International Commission on Radiological Protection 2001 ISBN IPEM Report 65: Quality standards in nuclear medicine Institute of Physics and Engineering in Medicine 1992 York ISBN IPEM Report 66: Quality control of gamma cameras and associated computer systems Institute of Physics and Engineering in Medicine 1993 York ISBN Prevention of accidental exposures to patients undergoing radiation therapy No. 86 ICRP Publication: Annals of the ICRP 30 (3) International Commission for Radiological Protection of 18 pages
16 ISBN European Commission 1999 European guidelines on quality criteria for computed tomography EUR EC Luxembourg ISBN This document can be found on the Internet at: Reporting of Injuries, Diseases and Dangerous Occurrences Regulations 1995 SI 1995/3163 The Stationery Office 1995 ISBN An organisation with a memory: Report of an expert Group on learning from adverse events in the NHS chaired by the Chief Medical Officer The Stationery Office 2000 ISBN This document can be found on the Internet at: Building a safer NHS for patients: Implementing an organisation with a memory Department of Health London 2001 This document can be found on the Internet at: Medical Devices Agency Medical devices and equipment management for hospital and community based organisations Device Bulletin MDA DB9801 Chapter 8 London MDA 1998 Available from the Medicines & Healthcare products Regulatory Agency at: [email protected] 30. Medical Devices Agency Medical devices and equipment management: Repair and maintenance provision Device Bulletin MDA DB2000(02) London MDA 2000 Available from the Medicines & Healthcare products Regulatory Agency at: [email protected] 31. Notes for guidance on the clinical administration of radiopharmaceuticals and use of sealed radioactive sources Administration of Radioactive Substances Advisory Committee 1998 Department of Health London This document can be found on the Internet at: Radiation dose to patients from radiopharmaceuticals No. 53 ICRP Publication: Annals of the ICRP 18 (1-4) International Commission on Radiological Protection 1988 ISBN ICRP 80 Radiation dose to patients from radiopharmaceuticals Addendum 2 to ICRP53: Annals of the ICRP 28 (3) International Commission on Radiological Protection 2000 ISBN Further reading 1. Safe use of work equipment. Provision and Use of Work Equipment Regulations Approved Code of Practice and guidance L22 (Second edition) HSE Books 1998 ISBN Electricity at Work Regulations 1989 SI 1989/635 The Stationery Office ISBN X 3. IPEM Report 79: The critical examination of x-ray generating equipment in diagnostic radiology by The Critical Examination Working Party of the IPEM 1999 Institute of Physics and Engineering in Medicine ISBN Recommendations of the International Commission on Radiological Protection No. 60 ICRP Publication: Annals of the ICRP 21 (1-3) International Commission on Radiological Protection 1991 ISBN Hart D, Wall B F, Shrimpton P C et al Reference doses and patient size in paediatric radiology National Radiological Protection Board R318 NRPB 2001 ISBN X 6. Tanner R J, Wall B F, Shrimpton P C et al Frequency of medical and dental X-ray examinations in the UK 1997/98 National Radiological Protection Board R320 NRPB 2000 ISBN Hart D and Wall B F Radiation exposure of the UK population from medical and dental x-ray examinations National Radiological Protection Board W4 NRPB 2002 ISBN McCollough C H and Schueler B A Calculation of effective dose Medical Physics (5) ICRP Addendum 1 to Publication 53 Radiation dose to patients from radiopharmaceuticals ICRP Publication 62 Radiological Protection in Biomedical Research Annals of the ICRP 22 (3) International Commission on Radiological Protection 1993 ISBN of 18 pages
17 Glossary MHRA Medicines & Healthcare products Regulatory Agency (formed by the AEC Automatic Exposure Control merger of the Medical Devices Agency (MDA) and Medicines Control BNMS British Nuclear Medicine Society Agency (MCA)) DGN Dental Guidance Notes MPE medical physics expert DHSS&PS(NI) Department of Health Social Services NHS National Health Service and Public Safety (Northern Ireland) NHSBSP National Health Service Breast DR Direct (digital) radiography Screening Programme DRL Diagnostic Reference Level NIAIC Northern Ireland Adverse Incident Centre (DHSS&PS(NI)) CR Computed radiography PACSnet Picture Archiving and Communication CT Computed tomography Systems National Evaluation Team CTDI Computed tomography dose index PASA NHS Purchasing and Supply Agency DH Department of Health PES Programmable electronic system safety HPA (RPD) Health Protection Agency Radiation Protection Division formerly known as QA Quality Assurance National Radiological Protection Board (NRPB) RCR Royal College of Radiologists HSWO78 Health & Safety at Work Order 1978 REMS Radiation and Environmental (Northern Ireland) Modelling Scotland (formerly NRPB Scotland) IEC International Electrotechnical Commission RPA radiation protection adviser ImPACT Imaging Performance Assessment of RPS radiation protection supervisor CT scanners SEHD Scottish Health IPE Institute of Physics and Engineering Department in Medicine SHS Scottish Healthcare Supplies IR(ME)R Ionising Radiation (Medical Exposure) Regulations 2000 IRR99 Ionising Radiations Regulations 1999 KCARE L121 MDGN King s Centre for the Assessment of Radiological Equipment Work with ionising radiation. Ionising Radiations Regulations Approved Code of Practice and guidance L121 Medical and Dental Guidance Notes MDR Medical Devices Regulations 2002 MEIGaN Medical Electrical Installation Guidance Notes 17 of 18 pages
18 Further information HSE produces a wide range of documents. Some are available as printed publications, both priced and free, and others are only accessible via the HSE website, HSE priced and free publications are available by mail order from HSE Books, PO Box 1999, Sudbury, Suffolk CO10 2WA Tel: Fax: Website: (HSE priced publications are also available from bookshops and free leaflets can be downloaded from HSE s website: For information about health and safety ring HSE's Infoline Tel: Fax: Textphone: [email protected] or write to HSE Information Services, Caerphilly Business Park, Caerphilly CF83 3GG. This document contains notes on good practice which are not compulsory but which you may find helpful in considering what you need to do. This document is available web only at: Crown copyright This publication may be freely reproduced, except for advertising, endorsement or commercial purposes. First published 03/06. Please acknowledge the source as HSE. Published by the Health and Safety PM77 (Third edition) 03/06 18 of 18 pages
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act
Safety and Radiation Protection Office Working with X-Ray Equipment
Safety and Radiation Protection Office Working with X-Ray Equipment Radiation Safety Rules May 2016 Radiation Safety Rules X-Ray Equipment General Rules for the protection of persons exposed to ionising
Guidance on the Safe Use of Dental Cone Beam CT (Computed Tomography) Equipment
HPA-CRCE-010 Guidance on the Safe Use of Dental Cone Beam CT (Computed Tomography) Equipment Prepared by the HPA Working Party on Dental Cone Beam CT Equipment Health Protection Agency Centre for Radiation,
HSE information sheet Industrial radiography - managing radiation risks Ionising Radiation Protection Series No. 1(rev 1)
HSE information sheet Industrial radiography - managing radiation risks Ionising Radiation Protection Series No. 1(rev 1) Introduction This information sheet is for clients and main or principal contractors
List of Contents Section 10, Nuclear Safety and Radiation Protection Legislation Page
List of Contents Section 10, Nuclear Safety and Radiation Protection Legislation Page Overview 1 The Radiation Protection (Medical Exposures) Directive 11 The Public Information (Radiological Emergency)
SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration
Implementation Page #: Page 1 of 10 Last Reviewed/Update 1. Purpose and Scope The purpose of this document is to describe the Medical Physics and Radiation Safety program at Boston University (BU) and
The Ionising Radiations Regulations 1999 (IRR'99): Guidance Booklet
The Ionising Radiations Regulations 1999 (IRR'99): Guidance Booklet Responsible person: Maria Murray Published: Sunday, July 1, 2012 ISBN: 9781-871101-88-3 Summary Radiographic staff play a pivotal and
Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
The practice of Clinical Technologists is divided into Clinical Physics and Clinical Engineering. They work in the following disciplines:
VRCT Voluntary Register of Clinical Technologists Regulating the Clinical Technologist Profession The Clinical Technologist: Scope of Practice Draft Document Clinical Technologists are Healthcare Scientists
PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS
Sec. B.1 - B.4 SSRCR Volume I - February 2009 PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS Sec. B.1 - Purpose and Scope. a. This
EU DIRECTIVE ON GOOD CLINICAL PRACTICE IN CLINICAL TRIALS DH & MHRA BRIEFING NOTE
EU DIRECTIVE ON GOOD CLINICAL PRACTICE IN CLINICAL TRIALS DH & MHRA BRIEFING NOTE Purpose 1. The Clinical Trials Directive 2001/20/EC heralds certain additional responsibilities for the Medicines and Healthcare
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
25 TEXAS ADMINISTRATIVE CODE 289.226. Registration of Radiation Machine Use and Services. Texas Regulations for Control of Radiation
25 TEXAS ADMINISTRATIVE CODE 289.226 Registration of Radiation Machine Use and Services Texas Regulations for Control of Radiation (revisions effective September 1, 2004 are shown as shaded text) 289.226(a)
Guidance Notes for Dental Practitioners on the Safe Use of X-Ray Equipment
Guidance Notes for Dental Practitioners on the Safe Use of X-Ray Equipment June 2001 Membership of the Working Party for producing the Guidance Notes for dental practitioners on the safe use of x-ray equipment
Staff Doses & Practical Radiation Protection in DEXA
Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:
Consultation. Review of the scopes of practice for registration in the profession of medical radiation technology
Consultation Review of the scopes of practice for registration in the profession of medical radiation technology CONTENTS Page Consultation Information... 3 Context of the Consultation... 4-7 Legislative
Veterinary applications of ionising radiation
Veterinary applications of ionising radiation HERCA Task Force on Veterinary Applications Main results of the Questionnaire National regulatory requirements with regard to veterinary medical applications
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.
USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY
GUIDE ST 5.3 / 4 MAY 2007 USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY 1 Ge n e r a l 3 2 Safety licences and exemption from licensing 3 3 Safety licence-exempt use of radiation 3
Study-wide considerations: identifies the areas of the criterion the lead reviewer should consider when conducting the review
UK STUDY-WIDE GOVERNANCE CRITERIA FOR R&D REVIEW INTRODUCTION The following document is intended to provide operational guidance to the person taking the UK lead in conducting the R&D Study-Wide review
5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ
5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ European Commission 5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21$1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/(;32685(6
318 DECREE. of the State Office for Nuclear Safety of 13 June 2202,
318 DECREE of the State Office for Nuclear Safety of 13 June 2202, On details for emergency preparedness assurance at nuclear installations and workplaces with ionizing radiation sources and on requirements
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
ADMINISTRATIVE RULES OF THE VERMONT BOARD OF RADIOLOGIC TECHNOLOGY
ADMINISTRATIVE RULES OF THE VERMONT BOARD OF RADIOLOGIC TECHNOLOGY Part 1. GENERAL INFORMATION 1.1 The Board's Purpose The State Board of Radiologic Technology (the Board) has been created and given powers
HEALTH PROFESSIONS ACT 56 OF 1974
HEALTH PROFESSIONS ACT 56 OF 1974 REGULATIONS DEFINING THE SCOPE OF THE PROFESSION OF RADIOGRAPHY Published under Government Notice R2326 in Government Gazette 5349 of 3 December 1976 and amended by GN
Information Security Policies. Version 6.1
Information Security Policies Version 6.1 Information Security Policies Contents: 1. Information Security page 3 2. Business Continuity page 5 3. Compliance page 6 4. Outsourcing and Third Party Access
GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES
GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound
WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE
WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 [email protected]
THE ROYAL COLLEGE OF RADIOLOGISTS
THE ROYAL COLLEGE OF RADIOLOGISTS TEL: 020-7636-4432 FAX: 020-7323-3100 38 PORTLAND PLACE LONDON W1B 1JQ BFCR(06)4 (updated February 2008) Retention and Storage of Images and Radiological Patient Data
How To Inspect A Blood Bank
Site visit inspection report on compliance with HTA minimum standards Belfast Cord Blood Bank HTA licensing number 11077 Licensed for the procurement, processing, testing, storage, distribution and import/export
Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994
Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical SUR-G0014-1 12 NOVEMBER 2013 This guide does not purport to be an interpretation of law and/or regulations
IPEM guidance on coding Medical Physics and Clinical Engineering staff within ESR for Healthcare Science Leads
IPEM guidance on coding Medical Physics and Clinical Engineering staff within ESR for Healthcare Science Leads Background Workforce Standards at the Health & Social Care Information Centre have recently
Personal Dosimetry in Dental Radiology
Personal Dosimetry in Dental Radiology Position Statement Personal Dosimetry in Dental Radiology Radiological Protection Institute of Ireland May 2011 3 Clonskeagh Square, Dublin 14 Background The RPII
Leaflet 14. Accident and Incident Investigation and Reporting
Leaflet 14 Accident and Incident Investigation and Reporting CONTENTS Paragraph 1 Scope 2 Statutory requirements Duties 3 Commanding Officer and Head of Establishment (CO/HoE) 4 Radiation Safety Officer
ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY
ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY B. Wall, (Co-ordinator) 1, D. Hart 1, H. Mol 2, A. Lecluyse 2, A. Aroua 3, P. Trueb 3, J. Griebel 4, E. Nekolla 4, P. Gron 5, H. Waltenburg 5, H. Beauvais-March
European Academy of DentoMaxilloFacial Radiology
European Academy of DentoMaxilloFacial Radiology Framework for Specialist Training in Dental and Maxillofacial Radiology Background The scope of DentoMaxilloFacial Radiology DMFR (Dental and Maxillofacial
Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC
Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC August 2013 Contents 1 Introduction...3 2 Scope of the directive...3 2.1 What is an in vitro diagnostic medical device?... 3 2.2 Specimen
Nuclear Safety Council Instruction number IS-19, of October 22 nd 2008, on the requirements of the nuclear facilities management system
Nuclear Safety Council Instruction number IS-19, of October 22 nd 2008, on the requirements of the nuclear facilities management system Published in the Official State Gazette (BOE) number 270 of November
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,
Policy for the Management of Asbestos in Buildings
in Buildings Health and Safety Services 1 A1. Introduction If asbestos fibres enter the lungs they can cause cancer or incurable disease. Consequently the University is committed to ensure that all reasonably
Guidance notes for the installation of electrical supply in mobile trailers for breast screening
NHS Breast Cancer Screening Programme Cancer Screening Programmes Guidance notes for the installation of electrical supply in mobile trailers for breast screening B Mansfield, CP Lawinski NHSBSP PUBLICATION
NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM
NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM Purpose: To provide licensees and registrants with a
Radiation Rule Guide. Veterinary Facilities. Hawaii Administrative Rules, Chapter 11-45, Radiation Control. Hawaii State Department of Health
Radiation Rule Guide Veterinary Facilities Hawaii Administrative Rules, Chapter 11-45, Radiation Control Hawaii State Department of Health MAY 2009 Radiation Rule Guide for Veterinary Facilities Hawaii
3rd International Symposium of the System of Radiological Protection Seoul, Korea, 20-22 October 2015
3rd International Symposium of the System of Radiological Protection Seoul, Korea, 20-22 October 2015 Dr. Pedro Ortiz López Member of ICRP Committee 3 on Radiological Protection in Medicine 2 1895 Discovery
Technical Notes NDT 001 Specific Requirements for Non- Destructive Testing Laboratories
ACCREDITATION SCHEME FOR LABORATORIES Technical Notes NDT 001 Specific Requirements for Non- Destructive Testing Laboratories Technical Notes NDT 001, Apr 14 The SAC Accreditation Programme is managed
Application for the Regulation of the Clinical Technologist Profession by the Health Professions Council
Application for the Regulation of the Clinical Technologist Profession by the Health Professions Council The Voluntary Register of Clinical Technologists Institute of Physics and Engineering in Medicine
Risk management a practical approach
Risk management a practical approach Introduction Preventing work related accidents and injuries is the primary concern for all those involved in health and safety. Work related accidents and injuries
Waste Management Policy
HEALTH AND SAFETY ISSUES Waste Management Policy Policy No: 39 Date of issue: October 1998 Review Dates: August 2001, December 2003, July 2008 Date to be Reviewed: July 2010 Page 1 of 13 Date of Issue:
EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information
Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment
Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Atomic Energy Regulatory Board Mumbai 400 094 1 DEFINITIONS 1. Quality Assurance (QA): Planned and systematic actions
Bonn Call-for-Action Joint Position Statement by the IAEA and WHO
Bonn Call-for-Action Joint Position Statement by the IAEA and WHO The International Atomic Energy Agency (IAEA) held the International Conference on Radiation Protection in Medicine: Setting the Scene
CODE OF PRACTICE. Safety Management. Occupational Safety and Health Branch Labour Department CODE OF PRACTICE ON SAFETY MANAGEMENT 1
CODE OF PRACTICE On Safety Management Occupational Safety and Health Branch Labour Department CODE OF PRACTICE ON SAFETY MANAGEMENT 1 This Code of Practice is prepared by the Occupational Safety and Health
How To Deal With Asbestos In Minster Of Minsteria
RELATED TO EXPOSURE TO ASBESTOS AT WORK [S.L.424.23 1 SUBSIDIARY LEGISLATION 424.23 RELATED TO EXPOSURE TO ASBESTOS AT WORK REGULATIONS LEGAL NOTICE 323 of 2006. 15th December, 2006 1. (1) The title of
APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS
APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS SAFETY REPORTS SERIES No. 39 APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL
New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01)
Safeguarding public health New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01) Tony Orme, Senior GDP Inspector Inspection, Enforcement and Standards Division
DESCRIPTION OF THE MEDICINES FOR HUMAN USE (CLINICAL TRIALS) REGULATIONS 2004
DESCRIPTION OF THE MEDICINES FOR HUMAN USE (CLINICAL TRIALS) REGULATIONS 2004 Page 1 of 17 TABLE OF CONTENTS INTRODUCTION 4 EXECUTIVE SUMMARY 4 PUBLIC HEALTH BENEFITS 5 Good clinical practice 5 Good manufacturing
GUIDELINES FOR FOOD IMPORT CONTROL SYSTEMS
GUIDELINES FOR FOOD IMPORT CONTROL SYSTEMS SECTION 1 SCOPE CAC/GL 47-2003 1. This document provides a framework for the development and operation of an import control system to protect consumers and facilitate
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
The Design of Diagnostic Medical Facilities where Ionising Radiation is used
RPII 09/01 The Design of Diagnostic Medical Facilities where Ionising Radiation is used A Code of Practice issued by the Radiological Protection Institute of Ireland June 2009 The Design of Diagnostic
SAFETY and HEALTH MANAGEMENT STANDARDS
SAFETY and HEALTH STANDARDS The Verve Energy Occupational Safety and Health Management Standards have been designed to: Meet the Recognised Industry Practices & Standards and AS/NZS 4801 Table of Contents
Guideline on good pharmacovigilance practices (GVP)
1 2 20 February 2012 EMA/541760/2011 3 4 Guideline on good pharmacovigilance practices (GVP) Module I Pharmacovigilance systems and their quality systems Draft finalised by the Agency in collaboration
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
Health and Safety Management Standards
Health and Safety Management Standards Health and Safety Curtin University APR 2012 PAGE LEFT INTENTIONALLY BLANK Page 2 of 15 CONTENTS 1. Introduction... 4 1.1 Hierarchy of Health and Safety Documents...
What is Radiology and Radiologic Technology?
What is Radiology and Radiologic Technology? Note: Prospective CSI Radiologic Technology students are encouraged to print this document out and study it thoroughly before applying for admission to the
STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT
2010 STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT Department of Health and Human Services Division of Public Health Licensure Unit 301 Centennial Mall South, Third Floor PO Box
Health & Safety Policy and Procedure MEDICAL DEVICES IN DENTAL POLICY. Version: v1. Date approved: pending
Health & Safety Policy and Procedure MEDICAL DEVICES IN DENTAL POLICY Version: v1 Date approved: pending Document Control Document Reference Title of document Authors name(s) Authors job title(s) Directorate(s)
Health and Safety (Sharp Instruments in Healthcare) Regulations 2013
Health and Safety (Sharp Instruments in Healthcare) Regulations 2013 Guidance for employers and employees HSE information sheet Health Services Information Sheet 7 This information sheet is for healthcare
X-RAY SAFETY MANUAL. WRHA X-ray Safety Committee
X-RAY SAFETY MANUAL A manual of guidance, policies and procedures specific to the use of x-rays in Diagnostic Imaging, Winnipeg Regional Health Authority. WRHA X-ray Safety Committee Edition: April 2013
GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES
Foreword GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES These guidance notes do not aim to be a definite interpretation of National Laws and/or regulations and are for guidance purpose only.
X-ray Safety Manual 2013
X-ray Safety Manual 2013 X-ray Machines and Other Ionizing Radiation Producing Devices The guidelines in this manual, when followed will ensure that appropriate protective and regulatory actions required
RC-17. Alejandro V. Nader National Regulatory Authority Montevideo - Uruguay
RC-17 Radiation Protection in Waste Management and Disposal Implementing the Joint Convention on the Safety of Spent Fuel Management and on the Safety of Radioactive Waste Management Alejandro V. Nader
OHS INCIDENT REPORTING & INVESTIGATION - SOP 104
1. Purpose The purpose of this SOP is to provide a process for OHS Incident Reporting and Investigation. 2. Approval Managing Director 3. Definitions Dangerous Occurrence means an occurrence as defined
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric
TITLE 32: ENERGY CHAPTER II: ILLINOIS EMERGENCY MANAGEMENT AGENCY SUBCHAPTER b: RADIATION PROTECTION
TITLE 32: ENERGY CHAPTER II: ILLINOIS EMERGENCY MANAGEMENT AGENCY SUBCHAPTER b: RADIATION PROTECTION PART 405 CERTIFICATION OF INDIVIDUALS TO PERFORM INDUSTRIAL RADIOGRAPHY Section 405.10 Purpose and Scope
Caedmon College Whitby
Caedmon College Whitby Data Protection and Information Security Policy College Governance Status This policy was re-issued in June 2014 and was adopted by the Governing Body on 26 June 2014. It will be
REGULATION (EEC) No 2309/93
REGULATION (EEC) No 2309/93 Council Regulation (EEC) No 2309/93 of 22 July 1993 laying down Community procedures for the authorization and supervision of medicinal products for human and veterinary use
Illinois. Accreditation. Requirements. Medical Radiation. Technology
Illinois Emergency Management Agency Division of Nuclear Safety A Guide for Illinois Accreditation Requirements in Medical Radiation Technology Pat Quinn Governor Jonathan Monken Director CONTENTS 1. Introduction...
(OJ L 169, 12.7.1993, p. 1)
1993L0042 EN 11.10.2007 005.001 1 This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents B COUNCIL DIRECTIVE 93/42/EEC of 14 June 1993 concerning
The Control of Major Accident Hazards Regulations (Northern Ireland) 2015
STATUTORY RULES OF NORTHERN IRELAND 2015 No. 325 HEALTH AND SAFETY The Control of Major Accident Hazards Regulations (Northern Ireland) 2015 Made - - - - 26th August 2015 Coming into operation - 28th September
The Nuclear Safety Council, in exercise of the functions conferred to
Nuclear Safety Council (BOE no. 43, of 19/02/2008) The Nuclear Safety Council s INSTRUCTION IS-17, of 30 th January, on the recognition of training courses and programmes for personnel that manage the
Example of a food company quality
Appendix A manual Example of a food company quality Contents Date: 13/03/95 RME-QLMN-OO Page 1 of 3 Section Title ISO 9001 reference 01 In trod uction 02 Purpose 03 Scope 04 Definitions 05 Management responsibility
Policy Document Control Page
Policy Document Control Page Title Title: Medical Devices Management Policy Version: 10 Reference Number: CO16 Supersedes Supersedes: Version 9 Description of Amendment(s): Originator Addition of 4.8 Sharps
OH&S Management Systems Audit Checklist (NAT, E3)
3.1.2 3.1.1 Introduction OH&S Management Systems Audit Checklist (NAT, E3) This audit checklist is based on Element 3 (Implementation) of the National Self-Insurers OHS Audit Tool. For a full copy of the
Professional Behaviour and Ethical Conduct. Code of Practice relating to:
Code of Practice relating to: Professional Behaviour and Ethical Conduct Promoting transparency and enhancing public confidence in the dental profession February 2012 Code of Practice relating to: Professional
Health and safety regulation... a short guide
...... a short guide...... a short guide Why this guide? The Health and Safety Commission (HSC) conducted a review of health and safety regulation in 1994. It found that people were confused about the
OCCUPATIONAL HEALTH AND SAFETY AUTHORITY ACT (CAP. 424) Protection of Workers from the Risks related to Exposure to Asbestos at Work Regulations, 2006
L.N. 323 of 2006 B 4649 OCCUPATIONAL HEALTH AND SAFETY AUTHORITY ACT (CAP. 424) Protection of Workers from the Risks related to Exposure to Asbestos at Work Regulations, 2006 IN EXERCISE of the powers
Drinking Water Quality Management Plan Review and Audit Guideline
Drinking Water Quality Management Plan Review and Audit Guideline This publication has been compiled by Queensland Water Supply Regulator, Department of Energy and Water Supply. State of Queensland, 2013.
A Short Guide to The Safety, Health and Welfare at Work Act, 2005
A Short Guide to The Safety, Health and Welfare at Work Act, 2005 3 A Short Guide to the Safety, Health and Welfare at Work Act, 2005 Published in August 2005 by the Health and Safety Authority, 10 Hogan
Patient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University