Evaluation of painful total hip replacements modular metal taper junctions

Size: px
Start display at page:

Download "Evaluation of painful total hip replacements modular metal taper junctions"

Transcription

1 Orthopaedics Evaluation of painful total hip replacements R. Michael Meneghini, MD + Corresponding Author + Indiana University Health Physicians Department of Orthopaedic Surgery Indiana University School of Medicine Indianapolis, IN Intro & Background Benefits of modularity in total hip arthroplasty are numerous. Modular femoral heads have been the standard of care for many years as they allow for per-operative adjustments in leg length and offset. More recently, modern primary total hip arthroplasty femoral components have emerged to include those with modular necks, these provide the surgeon the unique advantage to more exactly reproduce or correct the patient s anatomy and hip biomechanics. The modular necks come in a variety of offsets, lengths and versions that allow the surgeon to create the appropriate hip anatomical geometry, which may optimize abductor tension and strength, equalize leg length and enhance hip stability. In addition, the modular neck can facilitate femoral component insertion in more technically challenging surgical approaches, such as the direct anterior approach, where femoral exposure and mobilization can be difficult. These modular neck prostheses therefore offer key benefits compared to standard components, especially in cases where deformity is present (e.g., dysplasia of the hip, post-traumatic arthritis, etc). These modular neck prostheses do also allow for the use of modular heads and so have two modular junctions as opposed to the single modular junction with conventional modular head prostheses. With the additional taper junction, the modular neck femoral components provide an additional interface that may, in rare situations, be a potential source for metal fretting and crevice corrosion. 1-5 Whenever two metal surfaces are in contact and there is the potential for relative motion, metal debris may be generated or corrosion can take place. There have been reports of fretting and/or corrosion between the stem and head of conventional modular head prostheses for many years. This has been reported with both titanium-titanium, CoCr-CoCr and CoCr-titanium taper junctions. In vivo production of metal debris from implants has been the subject of much scrutiny recently. Metal on metal bearings, for example, produce small (20-80nm diameter) metallic wear debris whenever motion is present and, in cases of edge loading, the load of metal debris can be highly elevated compared to well positioned metal on metal bearings. Furthermore, elevated blood serum metal ion levels (it is believed these metal ions are produced by corrosion of the wear debris in the case of metal on metal joints) and metal hypersensitivity resulting in an adverse local tissue reaction (ALTR) may occur with metal-metal articular surface bearings causing premature failure due to osteolysis, aseptic loosening and in some instances a devastating pseudotumor formation It has been shown that a similar reaction and potentially premature failure of the total hip arthroplasty may occur in rare cases of patients with both modular head 23 and modular-neck femoral components. 3 Therefore, the intent of this manuscript is to discuss the clinical presentation, evaluation and workup of patients who present with persistent pain and symptoms after successful total hip arthroplasty with a metal taper junction suspected of fretting and/or corrosion. It should be noted that the clinical presentation and symptoms for reactions to metal debris can be similar to other complications such as infection. The intent is to provide the clinician a basic guideline, only at our current level of understanding based on very few reported cases, for the potential clinical presentation, evaluation, pathology and treatment of such patients, including when to consider revision surgery.

2 Evaluation of painful total hip replacements Clinical Presentation Patients with persistent pain in a total hip arthroplasty with a modular neck or modular head primary femoral component who, after thorough evaluation, lacks a clinically evident etiology should be considered for a potential adverse local tissue reaction to fretting corrosion at the modular taper junction. The patients will typically present with pain which may be located in the anterior, lateral or posterior aspect of the hip. Patients may or may not have pain at rest, but more reliably have pain with weight-bearing. Patients will typically have pain with motion of their hip and with loading of the hip joint on physical examination. In the rare cases of pseudotumor formation, the patients may feel a large mass or fluid collection in the region of the hip and in those cases an investigation for significant local soft tissue destruction should be considered. Evaluation for Infection All patients with a painful total hip arthroplasty, regardless of the femoral component type used, should be evaluated for infection as soon as possible. There are both anecdotal cases and reported cases in the peer-reviewed literature documenting superimposed bacterial infection in the face of clinical ALTR from metal-metal articulations. 24 It is not clear if the infection led to the ALTR however there is also evidence that alteration of the local ph may create an environment, which facilitates metal corrosion which suggests that the ph alteration in occult periprosthetic infection may be a precursor to corrosion in metal taper junctions. 25 Further confounding the issue is the reported rare incidence of patients with an ALTR from metal hypersensitivity presenting with a clinical picture mimicking infection, including elevated serum inflammatory 2 markers. 18 Regardless, the initial workup for a patient with a painful total hip arthroplasty should be a screening ESR and CRP and if either or both of those tests are elevated, a hip aspiration should be performed. It is essential that infection be ruled out as a potential cause of pain in patients with a total hip arthroplasty. Serum Infection Markers and Hip Aspiration Hip aspiration is highly recommended in patients with an elevated ESR and/or CRP. It should be performed with a large bore 18 gauge needle under radiographic guidance. Fluid should be sent for culture and sensitivity, but also for cell count and differential. If a radiologist performs the aspiration, the fluid should be obtained without any radiopaque dye or contrast to confirm intra-articular needle placement, as this may result in a dilution of the cell count concentration, which is critical for accurate diagnosis of infection. Recently, evidence has emerged that demonstrate cutoff values for patients with elevated serum inflammatory markers that indicate infection is likely if synovial cell count and differential is elevated. Della Valle et al. reported that patients with a serum ESR > 30mm/hr and a C-Reactive protein > 10mg/dL who have a synovial fluid aspirate with a white cell count of greater than 3000 white blood cells per ml have a greater specificity, sensitivity and positive predictive value for the diagnosis of infection. 26 If patients have normal ESR and CRP, the data further supports infection is not present with 100% specificity. 26 The appearance of the aspirated fluid has not been definitely proven to be indicative of a tissue reaction or lack thereof. With patients with ALTR with metal on metal bearings the aspirated fluid has been described as brownish or grey in colour with a turbid appearance. Given the differences in mechanism of metallic debris production with modular junctions this fluid appearance may or may not be similar. In patients with a painful hip replacement with a modular taper junction, who do not have evidence for a diagnosis for infection, it is recommended the patient undergo a workup for the possibility of ALTR from metal junction fretting corrosion as well as other etiologies such as soft tissue impingement. Serum Metal Ion Levels When metallic debris is produced in vivo it can be corroded and the corrosion products can be detected in the blood as metal ions. Therefore, it is imperative that with those patients presenting with a painful THA and who have a modular metal taper junction the surgeon obtain whole blood or blood serum metal ion levels. This rationale is due to increased serum ion levels being associated with painful or poorly functioning metal-onmetal bearing total hip arthroplasties and have been shown to be valuable as a diagnostic tool in the workup of painful metal-on-metal hip replacements. 27 The relationship between blood/blood serum metal ion levels and fretting and/or corrosion of modular junctions has yet to be definitively established. However, De Smet et al demonstrated that chromium ion levels above 17μg/L and cobalt ion levels above 19μg/L were associated with high joint fluid ion levels and metallosis with metal on metal bearings. 27 These values may represent levels at which clinical significance might be considered for serum metal ion concentrations. In a recent study, Kwon et al. reported on a total of 201 metal-on-metal resurfacing arthroplasties evaluated at a mean follow-up of 61 months using ultrasound/magnetic resonance imaging and serum/hip aspirate cobalt and chromium measurements. 28 Pseudotumors found in 7 patients (4%)

3 Evaluation of painful total hip replacements were associated with significantly higher cobalt and chromium levels and inferior functional scores. The median serum cobalt and chromium levels in female patients with bilateral arthroplasty in the pseudotumor group were 9.0μg/L and 12.0 μg/l, respectively. These metal ion levels were significantly greater than the median serum cobalt and chromium levels in female patients with bilateral arthroplasty without pseudotumor, which were 2.9μg/L and 3.2μg/L, respectively. The authors concluded that elevated levels of cobalt and chromium ions suggest that pseudotumors are associated with increased wear generated from metal-on-metal articulations. 28 The Medicines and Healthcare products Regulatory Agency (MHRA) in the United Kingdom has suggested a cut-off level of 7 parts per billion (ppb). However, the 7 ppb cut-off level for the maximum of cobalt or chromium has been shown to have a specificity of 89% and sensitivity 52% for detecting a pre-operative unexplained failed metal on metal hip replacement. 29 Furthermore, it was reported the optimal cut-off level for the maximum of cobalt or chromium was 4.97 ppb and had sensitivity 63% and specificity 86% and was concluded that blood metal ions have the ability to separate failed from wellfunctioning hip replacements. 29 These findings may or may not be relevant to hips with modular taper junctions as the process of metal debris formation is different in a taper junction compared to an articulation. However, it is may be that higher blood or blood serum ion levels are indicative of higher levels of metal debris production in the joint. In summary, while we are continuing to develop an understanding of serum metal ion levels and their correlation to adverse local tissue reactions in hip arthroplasty, the above values provide some context at our current understanding for evaluation of painful total hip arthroplasties suspected of an ALTR with modular taper junctions. Metal Hypersensitivity Nickel is the most common metal sensitizer in humans, followed by cobalt and chromium. Cell-mediated types of response with which orthopaedic implant-associated hypersensitivity reactions (metal sensitivity or metal allergy) are generally associated or are classified as type IV delayed-type hypersensitivity. 30 The incidence of metal sensitivity, as determined by patch testing among patients with both well and poorly functioning implants, is approximately 25%, roughly twice as high as that of the general population. The average incidence of metal sensitivity among patients with a failed implant is approximately 50% to 60%. 29 This increased prevalence of metal sensitivity among patients with a failed implant has prompted speculation that a metal allergy may contribute to implant loosening. However, there is a lack of a clear connection between incidence of metal sensitivity and implant duration, infection, reason for removal, or pain. At this time, it remains unclear what role, if any, metal sensitivity plays in implant loosening. Specific types of implants with greater propensity to release metal in vivo may, theoretically, be more prone to induce metal sensitivity (such as metal on-metal bearing surfaces or, to a lesser extent metal fretting corrosion of modular tapers). Although general patch testing and commercial kits exist for a variety of commonly antigenic substances there is mounting concern about the value of dermal testing to the study of immune responses to orthopaedic implants. The in vitro lymphocyte transformation test (LTT) involves measuring the proliferative response of lymphocytes (obtained from peripheral blood) following exposure with antigen. Reports indicate the LTT may be equally or better suited for the testing of implant-related sensitivity 30, 31 compared to dermal patch testing. Thomas et al. investigated 16 patients with a failed metal-metal bearing hip arthroplasty due to lymphocyte dominated periprosthetic inflammation. 32 In 13/16 patients (81%), systemic metal sensitivity was found based on PT and/or LTT. 32 In summary, as an additional piece of information in evaluating these challenging patients, diagnostic testing in the form of either patch testing or an in vitro lymphocyte transformation test should be considered in addition to serum ion levels to address the patient s level of metal ion exposure and their potential hypersensitivity. 30 Whilst these tests are not definitive, the additional information provided may help the diagnosing surgeon determine the appropriate treatment pathway. Radiographic Analysis Radiographic analysis of painful total hips should always include plain radiographs to evaluate for loosening, osteolysis and for assessment of implant position. Of particular concern in patients with demonstrated adverse local tissue reactions from metal debris or ions is early loosening or osteolysis. Particular attention should be paid to the medial calcar, trochanteric and peri-acetabular regions for cortical lucency or erosion. In many cases of ALTR from metal wear debris from any source, the plain radiographs are normal without any 3

4 Evaluation of painful total hip replacements Figure 2 Figure 1 evidence of bony erosion (Figure 1). In patients with persistent pain, who have elevated metal ion levels or positive metal hypersensitivity testing and normal plain radiographs, it is recommended that advanced imaging studies be considered. Ultrasound has been employed to evaluate solid or fluid soft-tissue collections and masses associated with pseudotumors or seen adjacent to the implants, however, more recently, CT and/or Magnetic Resonance Imaging (MRI) with metal artifact reduction techniques have been useful to delineate more subtle soft-tissue findings associated with ALTR around metal-metal bearings.33 MRI detects osteolysis and complications in the periprosthetic soft tissues such as wear-induced synovitis, periprosthetic collections, neurovascular compression, and quality of the muscle and tendons of the rotator cuff of the hip While there are local soft tissue abnormalities adjacent to all total hip replacements, those findings detailed above occurring 4 next to metal-metal bearings in greater frequency35 supports the notion that MRI is an appropriate modality to further investigate ALTR associated with a total hip arthroplasty affected by modular taper fretting corrosion. If one is considering an MRI evaluation of a painful THA suspected of metal wear debris associated ALTR, detailed dialogue with radiologists knowledgeable in the metal artifact reduction techniques are necessary to optimize the information available from these advanced imaging techniques.38, 39 MRI may also be a valuable tool in determining if the source of pain is in fact soft tissue impingement such as from the iliopsoas. Peri-Articular Histopathology Histology samples may be obtained during revision or possibly a biopsy in order to diagnose or confirm the diagnosis of a patient. Patients with an adverse local tissue reaction to local metal debris from fretting corrosion may demonstrate similar perivascular pathologic characteristics as those reported in patients with metal-metal bearings. The intra-operative hip fluid is typically described as brownish or grey in color. The characteristic histological features typically consist of diffuse and perivascular infiltrates of T and B lymphocytes and plasma cells, high endothelial venules, massive fibrin exudation, accumulation of macrophages with drop-like inclusions, and infiltrates of eosinophilic granulocytes and necrosis (Figure 2). Also, metal particles are uncommon and not a requisite for the diagnosis of an ALTR (also known as ALVAL or aseptic lymphocyte vasculitis associated lesion in some studies).6, 7, 14, 18, 22, 32 Recommended Surgical Treatment There is currently no specific clinical data to guide the surgeon in the treatment of patients who are found to have a painful total hip arthroplasty with modular taper junction fretting corrosion at either the head-neck or modular neck-stem junction. Based upon the current, limited information, certain tenets and patient care strategies should be considered. First, while difficult and stressful for the surgeon and patient, it is not recommended that observation without surgical intervention be used for any significant time period once the diagnosis of an adverse local tissue reaction has been confirmed. As has been reported with metal-metal bearings, pseudotumor formation with soft tissue destruction has been reported in patients with fretting corrosion of metal taper junctions albeit in only single case reports. In severe cases these pseudotumors can render abductors completely absent or via mass effect cause neurovascular compromise of surrounding nerves or vessels. It is generally believed that the longer the source of metal debris is present, the worse the soft tissue damage may be. Therefore, surgical

5 Evaluation of painful total hip replacements Figure 4 Figure 3 intervention is warranted once all othe sources of pain have been eliminated and an adverse local tissue reaction from metal fretting corrosion (or other source of metal debris) is strongly suspected. As always, a detailed discussion of the risks of observation versus operative intervention with the patient must take place prior to surgery. Once the decision to surgically intervene has been made, the method of revision reconstruction most likely should employ the principle of minimizing any modular metal taper junctions in this particular patient, as the presumption is that either their immune system plays a role in development of an ALTR via hypersensitivity or that their local biology may create an environment supportive of fretting corrosion for reasons that are as yet unclear. Acetabular components can typically be retained if well-fixed and in acceptable position. The acetabular liner should likely be composed of highly cross-linked polyethylene, although ceramic liners may be utilized as well. Furthermore, if a femoral component with a modular neck was used, this should be removed and replaced with a femoral component that does not utilize a modular neck. In addition, it is suggested that if a metal femoral head was utilized, a ceramic femoral head be used in the revision surgery with a new stem. It has been shown that ceramic-metal modular junctions have less fretting corrosion than metal-metal modular junctions,2 although only in an in vitro test using zirconia heads. The majority of studies that report surgical treatment of ALTR due to metal-metal bearings by conversion to an alternative bearing such as ceramic-ceramic or metalpolyethylene have reported resolution of pain and symptoms in the majority of patients.40 (Figure 3). Femoral stem tapers should not be re-used with new femoral heads or necks unless the new combination is specifically indicated in the manufacturer s product labelling. Intra-operative Findings Evidence of fretting corrosion of the modular taper junction (either at the neck-head or neck stem junction) is visualized by irregular black material on the surface of the metal contained within the junction (Figure 4). Further, the black material is typically associated with and in approximate location to surface irregularities on the metal Figure 5 taper surface in contact with the opposite metal surface, consistent with crevice corrosion.3 Adverse local tissue reactions may also create a substantial amount of intra-articular joint fluid, as is typically seen preoperatively on MRI, and is reported to be a brownish or grey color with a turbid consistency. While this fluid collection has been described, the intra-articular fluid characteristics have not been consistently characterized macroscopically or with respect to laboratory analysis and cell count at the time of this publication to our knowledge. Local soft-tissue necrosis has been seen in some cases of ALTR from metal debris and can extend as far as the peri-articular capsule, the abductor musculature and tendinous insertion onto the greater trochanter (Figure 5) with necrotic bone seen in 5

6 Evaluation of painful total hip replacements the most severe cases. In addition, there have been observations of medial femoral neck or calcar resorption, aseptic loosening and osteolysis associated with ALTR from metal debris and ion production and these may be encountered intra-operatively as well. Summary For patients with a persistently painful total hip arthroplasty that has a metal modular taper junction, it is suggested that a comprehensive workup for occult infection be undertaken. If no other mechanical, soft tissue impingement or infectious etiology for the persistent pain is identified, then a thorough evaluation should be performed for possible ALTR from metal debris via fretting and/or corrosion mechanisms at the modular taper junction. The evaluation should likely include serum metal ion levels, metal hypersensitivity testing and radiographic analysis including plain radiographs and preferably a metalartifact reducing MRI. While no single test is sensitive or specific enough to confirm the diagnosis of an adverse local tissue reaction from metal debris or ion production, the clinician should use the tests outlined above in conjunction with their clinical judgment and a thorough discussion with the patient to determine the appropriate treatment. If surgical intervention is deemed appropriate, attention to the techniques and principles of minimizing metal modular taper junctions should be employed as is feasible. In addition, patients who are suspected of having an ALTR from metal debris production should be followed closely to prevent the rare, but devastating progressive soft-tissue destruction that has been reported in this clinical setting. Modularity within total hip replacement allows for more precise restoration of the patient s anatomy during the operation. Increasing the modularity, such as by using modular necks as well as modular heads, increases the options available to the surgeon. Additional modular junctions are an additional interface, the incidence of fretting and/or corrosion of these junctions appears to be rare. There may be patient specific factors such as local chemistry influenced possibly by underlying infections however, there is little if any strong evidence in the literature. The most important thing to the surgeon and indeed the patient is the swift and unambiguous diagnosis of the cause of pain in a painful total hip arthroplasty so that in the rare cases where fretting and/or corrosion has occurred, there are no delays in providing the appropriate treatment. 6

7 Evaluation of painful total hip replacements References: 1. Brown SA, Flemming CA, Kawalec JS, et al. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. Spring 1995;6(1): Hallab NJ, Messina C, Skipor A, Jacobs JJ. Differences in the fretting corrosion of metalmetal and ceramic-metal modular junctions of total hip replacements. J Orthop Res. Mar 2004;22(2): Kop AM, Swarts E. Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty. Oct 2009;24(7): Reclaru L, Eschler PY, Lerf R, Blatter A. Electrochemical corrosion and metal ion release from Co-Cr-Mo prosthesis with titanium plasma spray coating. Biomaterials. Aug 2005;26(23): Vendittoli PA, Roy A, Mottard S, Girard J, Lusignan D, Lavigne M. Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations: a follow-up study. J Bone Joint Surg Br. Jan 2010;92(1): Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. May 2011;93(5): Browne JA, Bechtold CD, Berry DJ, Hanssen AD, Lewallen DG. Failed metal-on-metal hip arthroplasties: a spectrum of clinical presentations and operative findings. Clin Orthop Relat Res. Sep 2010;468(9): Donell ST, Darrah C, Nolan JF, et al. Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs. J Bone Joint Surg Br. Nov 2010;92(11): Ebramzadeh E, Campbell PA, Takamura KM, et al. Failure modes of 433 metal-on-metal hip implants: how, why, and wear. Orthop Clin North Am. Apr 2011;42(2): , ix. 10. Hasegawa M, Yoshida K, Wakabayashi H, Sudo A. Pseudotumor With Dominant B- Lymphocyte Infiltration After Metal-On-Metal Total Hip Arthroplasty With a Modular Cup. J Arthroplasty. Jun Hutt JR, Busch C, Hughes RA. Failure of a metal on metal hip prostheses presenting as a destructive soft tissue mass due to ALVAL. Rheumatol Int. Oct 2011;31(10): Iwamoto T, Ikari K, Momohara S. Pseudotumor from a Metal-on-Metal Hip. J Rheumatol. Oct 2011;38(10): Klapperich C, Graham J, Pruitt L, Ries MD. Failure of a metal-on-metal total hip arthroplasty from progressive osteolysis. J Arthroplasty. Oct 1999;14(7): Kwon YM, Thomas P, Summer B, et al. Lymphocyte proliferation responses in patients with pseudotumors following metal-on-metal hip resurfacing arthroplasty. J Orthop Res. Apr 2010;28(4): Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-onmetal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. J Bone Joint Surg Br. Jan 2010;92(1): Maezawa K, Nozawa M, Matsuda K, Yuasa T, Shitoto K, Kurosawa H. Early failure of modern metal-on-metal total hip arthroplasty using a Wagner standard cup. J Arthroplasty. Jun 2006;21(4): Maurer-Ertl W, Friesenbichler J, Liegl-Atzwanger B, Kuerzl G, Windhager R, Leithner A. Noninflammatory Pseudotumor Simulating Venous Thrombosis After Metal-on-Metal Hip Resurfacing. Orthopedics. 2011;34(10):e Mikhael MM, Hanssen AD, Sierra RJ. Failure of metal-on-metal total hip arthroplasty mimicking hip infection. A report of two cases. J Bone Joint Surg Am. Feb 2009;91(2): Ollivere B, Darrah C, Barker T, Nolan J, Porteous MJ. Early clinical failure of the Birmingham metal-on-metal hip resurfacing is associated with metallosis and soft-tissue necrosis. J Bone Joint Surg Br. Aug 2009;91(8): Park YS, Moon YW, Lim SJ, Yang JM, Ahn G, Choi YL. Early osteolysis following secondgeneration metal-on-metal hip replacement. J Bone Joint Surg Am. Jul 2005;87(7): Shahrdar C. Pseudotumor in large-diameter metal-on-metal total hip articulation. J Arthroplasty. Jun 2011;26(4):665 e Willert HG, Buchhorn GH, Fayyazi A, et al. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. Jan 2005;87(1): Kosukegawa I, Nagoya S, Kaya M, Sasaki K, Sasaki M, Yamashita T. Revision total hip arthroplasty due to pain from hypersensitivity to cobalt-chromium in total hip arthroplasty. J Arthroplasty. Sep 2011;26(6):978 e Watters TS, Eward WC, Hallows RK, Dodd LG, Wellman SS, Bolognesi MP. Pseudotumor with superimposed periprosthetic infection following metal-on-metal total hip arthroplasty: a case report. J Bone Joint Surg Am. Jul ;92(7): Merritt K, Brown SA. Effect of proteins and ph on fretting corrosion and metal ion release. J Biomed Mater Res. Feb 1988;22(2): Schinsky MF, Della Valle CJ, Sporer SM, Paprosky WG. Perioperative testing for joint infection in patients undergoing revision total hip arthroplasty. J Bone Joint Surg Am. Sep 2008;90(9): De Smet K, De Haan R, Calistri A, et al. Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Joint Surg Am. Nov 2008;90 Suppl 4: Kwon YM, Ostlere SJ, McLardy-Smith P, Athanasou NA, Gill HS, Murray DW. "Asymptomatic" pseudotumors after metal-on-metal hip resurfacing arthroplasty: prevalence and metal ion study. J Arthroplasty. Jun 2011;26(4): Hart AJ, Sabah SA, Bandi AS, et al. Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Joint Surg Br. Oct 2011;93(10): Jacobs JJ, Urban RM, Hallab NJ, Skipor AK, Fischer A, Wimmer MA. Metal-on-metal bearing surfaces. J Am Acad Orthop Surg. Feb 2009;17(2): Hallab NJ. Lymphocyte transformation testing for quantifying metal-implant-related hypersensitivity responses. Dermatitis. Jun 2004;15(2): Thomas P, Braathen LR, Dorig M, et al. Increased metal allergy in patients with failed metal-on-metal hip arthroplasty and peri-implant T-lymphocytic inflammation. Allergy. Aug 2009;64(8): Cahir JG, Toms AP, Marshall TJ, Wimhurst J, Nolan J. CT and MRI of hip arthroplasty. Clin Radiol. Dec 2007;62(12): ; discussion Hayter CL, Potter HG, Su EP. Imaging of metal-on-metal hip resurfacing. Orthop Clin North Am. Apr 2011;42(2): , viii. 35. Mistry A, Cahir J, Donell ST, Nolan J, Toms AP. MRI of asymptomatic patients with metalon-metal and polyethylene-on-metal total hip arthroplasties. Clin Radiol. Jun 2011;66(6): Sabah SA, Mitchell AW, Henckel J, Sandison A, Skinner JA, Hart AJ. Magnetic resonance imaging findings in painful metal-on-metal hips: a prospective study. J Arthroplasty. Jan 2011;26(1):71-76, 76 e Toms AP, Marshall TJ, Cahir J, et al. MRI of early symptomatic metal-on-metal total hip arthroplasty: a retrospective review of radiological findings in 20 hips. Clin Radiol. Jan 2008;63(1): Hayter CL, Koff MF, Shah P, Koch KM, Miller TT, Potter HG. MRI after arthroplasty: comparison of MAVRIC and conventional fast spin-echo techniques. AJR Am J Roentgenol. Sep 2011;197(3):W Toms AP, Smith-Bateman C, Malcolm PN, Cahir J, Graves M. Optimization of metal artefact reduction (MAR) sequences for MRI of total hip prostheses. Clin Radiol. Jun 2010;65(6): Rajpura A, Porter ML, Gambhir AK, Freemont AJ, Board TN. Clinical experience of revision of metal on metal hip arthroplasty for aseptic lymphocyte dominated vasculitis associated lesions (ALVAL). Hip Int. Jan-Mar 2011;21(1):

8 325 Corporate Drive Mahwah, NJ t: f: A surgeon must always rely on his or her own professional clinical judgment when deciding whether to use a particular product when treating a particular patient. Stryker does not dispense medical advice and recommends that surgeons be trained in the use of any particular product before using it in surgery. The information presented is intended to demonstrate the breadth of Stryker product offerings. A surgeon must always refer to the package insert, product label and/or instructions for use before using any Stryker product. Products may not be available in all markets because product availability is subject to the regulatory and/or medical practices in individual markets. Please contact your Stryker representative if you have questions about the availability of Stryker products in your area. Stryker Corporation or its divisions or other corporate affiliated entities own, use or have applied for the following trademarks or service marks: Stryker. All other trademarks are trademarks of their respective owners or holders. NL12-NA-HP-433 MS/GS 04/12 Copyright 2012 Stryker Printed in USA

Clinical Testing for Metal-on-Metal Prosthetic Wear and Tear

Clinical Testing for Metal-on-Metal Prosthetic Wear and Tear Clinical Testing for Metal-on-Metal Prosthetic Wear and Tear Frederick G. Strathmann, PhD, DABCC (CC, TC) February 8 th, 2013 Disclosures None Objectives Compare heavy metal analysis in synovial fluid

More information

Current Concerns with Metal-on-Metal Hip Arthroplasty

Current Concerns with Metal-on-Metal Hip Arthroplasty Information Statement Current Concerns with Metal-on-Metal Hip Arthroplasty The American Academy of Orthopaedic Surgeons gratefully acknowledges the work of the Association of Hip & Knee Surgeons in the

More information

Cover Page. The handle http://hdl.handle.net/1887/25896 holds various files of this Leiden University dissertation

Cover Page. The handle http://hdl.handle.net/1887/25896 holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/25896 holds various files of this Leiden University dissertation Author: Weegen, Walter van der Title: Metal-on-metal hip arthroplasty : local tissue reactions

More information

Answers to commonly asked questions from patients with metal-on-metal hip replacements / resurfacings. Contents

Answers to commonly asked questions from patients with metal-on-metal hip replacements / resurfacings. Contents Answers to commonly asked questions from patients with metal-on-metal hip replacements / resurfacings John Skinner 1 and Alister Hart 2, Consultant Orthopaedic Surgeons and Directors of the London Implant

More information

Metal-on-Metal Total Hip Arthroplasty: Causes and High Incidence of Early Failure

Metal-on-Metal Total Hip Arthroplasty: Causes and High Incidence of Early Failure Metal-on-Metal Total Hip Arthroplasty: Causes and High Incidence of Early Failure David Fabi, MD; Brett Levine, MD; Wayne Paprosky, MD; Craig DellaValle, MD; Scott Sporer, MD; Gregg Klein, MD; Harlan Levine,

More information

Metal-on-Metal Hip Systems

Metal-on-Metal Hip Systems Meeting of the Orthopaedic and Rehabilitation Devices Advisory Panel June 27-28, 2012 Metal-on-Metal Hip Systems Center for Devices and Radiological Health U.S. Food and Drug Administration Meeting of

More information

Information for Our Patients Regarding Metal-on-Metal (MoM) Hip Replacements

Information for Our Patients Regarding Metal-on-Metal (MoM) Hip Replacements Information for Our Patients Regarding Metal-on-Metal (MoM) Hip Replacements The American Association of Hip and Knee Surgeons, The Hip Society and the American Academy of Orthopaedic Surgeons Concerns

More information

Y O U R S U R G E O N S. choice of. implants F O R Y O U R S U R G E R Y

Y O U R S U R G E O N S. choice of. implants F O R Y O U R S U R G E R Y Y O U R S U R G E O N S choice of implants F O R Y O U R S U R G E R Y Y O U R S U R G E O N S choice of implants F O R Y O U R S U R G E R Y Your Surgeon Has Chosen the C 2 a-taper Acetabular System The

More information

Running head: PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES 1 PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES: A LITERTURE REVIEW WITH CASE STUDIES

Running head: PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES 1 PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES: A LITERTURE REVIEW WITH CASE STUDIES Running head: PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES 1 PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES: A LITERTURE REVIEW WITH CASE STUDIES November 13, 2013 PSEUDOTUMORS IN ASSOCIATION WITH

More information

Metal-on-Metal A clinical overview

Metal-on-Metal A clinical overview Metal-on-Metal A clinical overview Adverse reactions to metal debris In response to the earlier medical device alert published by the MHRA on all metal-on-metal hip replacements, Corin aims to provide

More information

MRI Findings in Painful Metal-on- Metal Hip Arthroplasty

MRI Findings in Painful Metal-on- Metal Hip Arthroplasty Musculoskeletal Imaging Original Research Hayter et al. MRI of Painful Metal-on-Metal Hip Prostheses Musculoskeletal Imaging Original Research Catherine L. Hayter 1 Stephanie L. Gold 1 Matthew F. Koff

More information

on-metal Hips: Device Mechanics and Failure Modes

on-metal Hips: Device Mechanics and Failure Modes 1 Metal-on on-metal Hips: Device Mechanics and Failure Modes Steven M. Kurtz, Ph.D., and Richard Underwood, Ph.D. Exponent, Inc., and Drexel University NIH R01 AR47904 2 Contracts: DePuy Orthopaedics,

More information

Total Hip Joint Replacement. A Patient s Guide

Total Hip Joint Replacement. A Patient s Guide Total Hip Joint Replacement A Patient s Guide Don t Let Hip Pain Slow You Down What is a Hip Joint? Your joints are involved in almost every activity you do. Simple movements such as walking, bending,

More information

The Right Choice. Exeter. total hip system

The Right Choice. Exeter. total hip system The Right Choice Exeter total hip system Exeter The Right Choice Anatomic Reconstruction Offset The objectives of total hip replacement are to: relieve pain increase mobility and function Achieving a correct

More information

BONE PRESERVATION STEM

BONE PRESERVATION STEM TRI-LOCK BONE PRESERVATION STEM Featuring GRIPTION Technology SURGICAL TECHNIQUE IMPLANT GEOMETRY Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was

More information

Appendix B: FDA Safety Communication (Website) for MoM Hip Systems

Appendix B: FDA Safety Communication (Website) for MoM Hip Systems Appendix B: FDA Safety Communication (Website) for MoM Hip Systems. 1 Metal on Metal Hip Implant Systems Home Page The Hip Joint Hip Implant Systems Metal-on-Metal Hip Implant Systems Total Hip Replacement

More information

Medical Device Alert. Device All metal-on-metal (MoM) hip replacements. Action by. CAS deadlines. Ref: MDA/2012/036 Issued: 25 June 2012 at 11:00

Medical Device Alert. Device All metal-on-metal (MoM) hip replacements. Action by. CAS deadlines. Ref: MDA/2012/036 Issued: 25 June 2012 at 11:00 Medical Device Alert Issued: 25 June 2012 at 11:00 Device All metal-on-metal (MoM) hip replacements Problem The MHRA is issuing updated information and advice about the follow-up of implanted with metal-on-metal

More information

Metal on Metal Hips. Total Hip Replacement. Total Hip Replacement

Metal on Metal Hips. Total Hip Replacement. Total Hip Replacement Metal on metal hips: Questions Metal on Metal Hips Jim Sullivan Why were metal on metal hips popular? Are all metal on metal hips the same? What are the problems with the ASR hip? What are the issues with

More information

MRI Predicts ALVAL and Tissue Damage in Metal-on-Metal Hip Arthroplasty

MRI Predicts ALVAL and Tissue Damage in Metal-on-Metal Hip Arthroplasty Clin Orthop Relat Res DOI 10.1007/s11999-013-2788-y Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons SYMPOSIUM: PAPERS PRESENTED AT THE ANNUAL MEETINGS

More information

Hip Resurfacing 2011 ORIGINAL ARTICLE. James W. Pritchett MD. Introduction. Abstract

Hip Resurfacing 2011 ORIGINAL ARTICLE. James W. Pritchett MD. Introduction. Abstract 1 ORIGINAL ARTICLE Hip Resurfacing 2011 James W. Pritchett MD Abstract In 1938 Marion Smith-Peterson placed a cobalt chromium cup on a reshaped femoral head to perform the first hip resurfacing. 15 Also,

More information

Exeter. Surgical Technique. V40 Stem Cement-in-Cement. Orthopaedics

Exeter. Surgical Technique. V40 Stem Cement-in-Cement. Orthopaedics Exeter Orthopaedics V40 Stem Cement-in-Cement Surgical Technique Exeter V40 Stem Cement-in-Cement Surgical Technique Table of Contents Indications and Contraindications...2 Warnings and Precautions...2

More information

Why an Exactech Hip is Right for You

Why an Exactech Hip is Right for You Why an Exactech Hip is Right for You Why do I need a total hip replacement? Which surgical approach is best for me? How long will it last? Which implant is right for me? Founded in 1985 by an orthopaedic

More information

Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability

Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability Robert Wetzel, MD, Lalit Puri, MD, S. David Stulberg, MD Disclosure In accordance with ACCME

More information

Randal S. Ford. ATTORNEY AT LAW (205) 752-4440 attorney@randalsford.com. DePuy ASR Hip Implant Recall Fact Sheet

Randal S. Ford. ATTORNEY AT LAW (205) 752-4440 attorney@randalsford.com. DePuy ASR Hip Implant Recall Fact Sheet DePuy ASR Hip Implant Recall Fact Sheet Recalled Devices: ASR XL Acetabular System on the market since 2004, and sold worldwide. Dates of Implant: July 2003 to 2010 ASR Hip Resurfacing System on the market

More information

The Journal of Arthroplasty

The Journal of Arthroplasty The Journal of Arthroplasty 29 (2014) 219 224 Contents lists available at ScienceDirect The Journal of Arthroplasty journal homepage: www.arthroplastyjournal.org One-Component Revision of Failed Hip Resurfacing

More information

Clinically Proven Fiber Metal Material. Power to choose advanced bearing technologies to match patient demands

Clinically Proven Fiber Metal Material. Power to choose advanced bearing technologies to match patient demands Trilogy IT Acetabular System Design Rationale 1 For orthopedic surgeons treating a wide range of patients, the Trilogy IT Acetabular System provides a clinically proven1-5 Fiber Metal Material and the

More information

Jenaro Fernández-Valencia, Xavier Gallart, Guillem Bori, Sebastián Garcia Ramiro, Andrés Combalía, and Josep Riba

Jenaro Fernández-Valencia, Xavier Gallart, Guillem Bori, Sebastián Garcia Ramiro, Andrés Combalía, and Josep Riba Advances in Orthopedics, Article ID 982523, 4 pages http://dx.doi.org/10.1155/2014/982523 Research Article Assessment of Patients with a DePuy ASR Metal-on-Metal Hip Replacement: Results of Applying the

More information

Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology

Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology Hips designed to fit the unique anatomies of men and women Independent control for a natural fit Simple, practical solutions for optimal restoration

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Health Technology Appraisal Total hip replacement and resurfacing arthroplasty for the treatment of pain or disability resulting from end stage arthritis

More information

IMPORTANT PRODUCT RECALL UPDATE RA2012-067 EXT_2

IMPORTANT PRODUCT RECALL UPDATE RA2012-067 EXT_2 325 Corporate Drive Mahwah, NJ 07430 t: 201 831 5000 Orthopaedics IMPORTANT PRODUCT RECALL UPDATE RA2012-067 EXT_2 Reference: RA 2012-067_EXT_2 (update of the action reference RA2012-067_EXT_1 dated on

More information

Clinical and Radiographic Outcomes of 139 Hips with Articular Surface Replacement Total Hip Arthroplasty*

Clinical and Radiographic Outcomes of 139 Hips with Articular Surface Replacement Total Hip Arthroplasty* Clinical and Radiographic Outcomes of 139 Hips with Articular Surface Replacement Total Hip Arthroplasty* Assaf Kadar MD 1, Ran Ankory MD 1, Haggay Sherman MD 1, Iris Eshed MD 2, adav Shasha MD 1, Aviram

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. Dall-Miles Cabling System Dall-Miles Recon and Trauma Cable System Trochanteric Reattachment Using the Trochanteric

More information

Hip Replacement Surgery Understanding the Risks

Hip Replacement Surgery Understanding the Risks Hip Replacement Surgery Understanding the Risks Understanding the Risks of Hip Replacement Surgery Introduction This booklet is designed to help your doctor talk to you about the most common risks you

More information

DePuy Orthopaedics ASR Recall Reference Guide for U.S. Healthcare Professionals

DePuy Orthopaedics ASR Recall Reference Guide for U.S. Healthcare Professionals DePuy Orthopaedics ASR Recall Reference Guide for U.S. Healthcare Professionals DPYUS 46 (Version 2) 1 October 18, 2010 Contents 1. Available Resources... 3 Surgeon-to-Patient Communications Support...

More information

A Magazine Dedicated to Mass Torts Law. The Dangers of Metal-on-Metal Hips. Details of DePuy ASR Settlement Announced. www.langdonemison.

A Magazine Dedicated to Mass Torts Law. The Dangers of Metal-on-Metal Hips. Details of DePuy ASR Settlement Announced. www.langdonemison. d3 Dangerous Drugs and Devices A Magazine Dedicated to Mass Torts Law The Dangers of Metal-on-Metal Hips Details of DePuy ASR Settlement Announced www.langdonemison.com Johnson & Johnson Agrees to Pay

More information

Anterior Hip Replacement

Anterior Hip Replacement Disclaimer This movie is an educational resource only and should not be used to manage Orthopaedic health. All decisions about the management of hip replacement and arthritis management must be made in

More information

Acetabular Wedge Augment System

Acetabular Wedge Augment System Orthopaedics Acetabular Wedge Augment System Stryker Hips. Implant with confidence. Restoration Acetabular Wedge Augment System Restoration Acetabular Wedge Augments, united with Tritanium, Trident, MDM

More information

Metal-on-Metal: Questions & Answers

Metal-on-Metal: Questions & Answers Metal-on-Metal: Questions & Answers What are the benefits of metal-on-metal hip replacement? Cobalt Chrome alloy has been used since the 1930 s for a multitude of orthopaedic applications. Due to its relatively

More information

Purpose. Causes. Causes of Early Hip Disease. Causes. Clinical Evaluation 1/7/2010. Causes of early hip disease Femoral Acetabular Impingement (FAI)

Purpose. Causes. Causes of Early Hip Disease. Causes. Clinical Evaluation 1/7/2010. Causes of early hip disease Femoral Acetabular Impingement (FAI) Purpose Henry R. Boucher, M.D. Union Memorial Hospital Baltimore, Maryland Causes of early hip disease Femoral Acetabular Impingement (FAI) Clinical and radiographic work up Treatment conservative and

More information

Metal-on-Metal Hip Arthroplasty: A Review of Adverse Reactions and Patient Management

Metal-on-Metal Hip Arthroplasty: A Review of Adverse Reactions and Patient Management J. Funct. Biomater. 2015, 6, 486-499; doi:10.3390/jfb6030486 Review OPEN ACCESS Journal of Functional Biomaterials ISSN 2079-4983 www.mdpi.com/journal/jfb Metal-on-Metal Hip Arthroplasty: A Review of Adverse

More information

Early failure of a Birmingham resurfacing hip replacement with lymphoreticular spread of metal debris: pre-operative diagnosis with MR

Early failure of a Birmingham resurfacing hip replacement with lymphoreticular spread of metal debris: pre-operative diagnosis with MR The British Journal of Radiology, 82 (2009), e87 e91 CASE REPORT Early failure of a Birmingham resurfacing hip replacement with lymphoreticular spread of metal debris: pre-operative diagnosis with MR 1

More information

PARTIES, VENUE AND JURISDICTION

PARTIES, VENUE AND JURISDICTION F. John Caldwell, Jr., Esquire NJ Bar #9951994 Altom M. Maglio, Esquire (Pro Hac Vice Pending) FL Bar #88005 Ilyas Sayeg, Esquire (Pro Hac Vice Pending) FL Bar #0099140 MAGLIO CHRISTOPHER & TO ALE P.A.

More information

Patient Labeling Information System Description

Patient Labeling Information System Description Patient Labeling Information System Description The Trident Ceramic Acetabular System is an artificial hip replacement device that features a new, state-of-the-art ceramic-on-ceramic bearing couple. The

More information

Summary notes from Metal Hip Replacements: Solving The Uncertainties

Summary notes from Metal Hip Replacements: Solving The Uncertainties These notes represent a summary of each presentation at the meeting, Metal on- Metal Hip Replacements: Solving The Uncertainties. The meeting was: sponsored by Orthopaedic Research UK; hosted by Alister

More information

frequently asked questions Knee and Hip Joint Replacement Technology

frequently asked questions Knee and Hip Joint Replacement Technology frequently asked questions Knee and Hip Joint Replacement Technology frequently asked questions Knee and Hip Joint Replacement Technology Recently, you may have seen advertisements from legal companies

More information

Understanding the biological outcomes and responses to orthopaedic surgical implants.

Understanding the biological outcomes and responses to orthopaedic surgical implants. Understanding the biological outcomes and responses to orthopaedic surgical implants. Dr Catherine Bladen, University of Leeds,UK A report for the Winston Churchill Memorial Trust, 2012 1 Acknowledgements:

More information

Hip Resurfacing (Re-Review)

Hip Resurfacing (Re-Review) 20, 2012 Health Technology Assessment Hip Resurfacing (Re-Review) Final Evidence Report October 14, 2013 Health Technology Assessment Program (HTA) Washington State Health Care Authority PO Box 42712 Olympia,

More information

Disclosures. Complications of Wear or Corrosion of Chrome-Cobalt Hip Implants. Where are we going? 21 st Century Cures Act.

Disclosures. Complications of Wear or Corrosion of Chrome-Cobalt Hip Implants. Where are we going? 21 st Century Cures Act. Complications of Wear or Corrosion of Chrome-Cobalt Hip Implants Stephen S. Tower, M.D. Affiliated Professor UAA/ WWAMI Tower Joint Replacement Clinic www.tjrclinic.com Disclosures Index Case of Arthroprosthetic

More information

SUBJECT: MANAGEMENT OF BREAST EFFECTIVE DATE: 12/16/99 IMPLANTS REVISED DATE:

SUBJECT: MANAGEMENT OF BREAST EFFECTIVE DATE: 12/16/99 IMPLANTS REVISED DATE: MEDICAL POLICY SUBJECT: MANAGEMENT OF BREAST PAGE: 1 OF: 5 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical policy

More information

VERILAST Technology for Hip Replacement Implants

VERILAST Technology for Hip Replacement Implants VERILAST Technology for Hip Replacement Implants Surgeon Name Clinic Name Clinic Address Clinic i Address Phone Number Web Address Total Hip Replacement What Is VERILAST Hip Technology? OXINIUM Oxidized

More information

Musculoskeletal Infection Care Process Model

Musculoskeletal Infection Care Process Model Musculoskeletal Infection Care Process Model Musculoskeletal infections are serious and potentially life-threatening. Musculoskeletal infections include necrotizing fasciitis, septic arthritis, osteomyelitis,

More information

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION Richard A. Sweet, M.D. Louisville Orthopaedic Clinic Louisville, KY TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION INTRODUCTION Total hip replacement surgery (THR) has

More information

Accolade. System. Brochure. Achieving Perfect Balance. Cemented and Cementless Femoral Hip System

Accolade. System. Brochure. Achieving Perfect Balance. Cemented and Cementless Femoral Hip System Accolade System Brochure Achieving Perfect Balance Cemented and Cementless Femoral Hip System Accolade Cemented Hip System Potential for Perfect Balance The Accolade C Femoral Component: A perfect balance

More information

Robotic-Arm Assisted Surgery

Robotic-Arm Assisted Surgery Mako TM Robotic-Arm Assisted Surgery for Total Hip Replacement A Patient s Guide Causes of Your Hip Pain Your joints are involved in almost every activity you do. Movements such as walking, bending and

More information

BIRMINGHAM HIP Resurfacing (BHR ) System PATIENT INFORMATION

BIRMINGHAM HIP Resurfacing (BHR ) System PATIENT INFORMATION BIRMINGHAM HIP Resurfacing (BHR ) System PATIENT INFORMATION Table of Contents 1.0 What is the BHR Device? 2.0 What is the Purpose of the BHR Device? 3.0 When Should the BHR Device Not Be Used? (Contraindications)

More information

FDA Executive Summary Memorandum Metal-on-Metal Hip Implant Systems

FDA Executive Summary Memorandum Metal-on-Metal Hip Implant Systems FDA Executive Summary Memorandum Metal-on-Metal Hip Implant Systems Prepared for the June 27-28, 2012 Meeting of the Orthopaedic and Rehabilitation Devices Advisory Panel Gaithersburg Hilton Gaithersburg,

More information

Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumours

Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumours Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumours Y. -M. Kwon, S. Glyn-Jones, D. J. Simpson, A. Kamali, P. McLardy-Smith, H. S. Gill, D. W. Murray From

More information

Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner

Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner The Journal of Arthroplasty Vol. 24 No. 2 2009 Case Report Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner William J. Robertson, MD, Christopher J. Mattern, MD/MBA, John

More information

Rotator Cuff Pathophysiology. treatment program that will effectively treat it. The tricky part about the shoulder is that it is a ball and

Rotator Cuff Pathophysiology. treatment program that will effectively treat it. The tricky part about the shoulder is that it is a ball and Rotator Cuff Pathophysiology Shoulder injuries occur to most people at least once in their life. This highly mobile and versatile joint is one of the most common reasons people visit their health care

More information

Metal-on-metal (MoM) total hip arthroplasty

Metal-on-metal (MoM) total hip arthroplasty S77 Metal-on-Metal Total Hip Arthroplasty Five- to 11-Year Follow-Up Vassilios S. Nikolaou, M.D., Ph.D., Alain Petit, Ph.D., Kevin Debiparshad, M.D., Olga L. Huk, M.D., M.Sc., David J. Zukor, M.D., and

More information

Oftentimes, as implant surgeons, we are

Oftentimes, as implant surgeons, we are CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the

More information

Clinical performance of endoprosthetic and total hip replacement systems

Clinical performance of endoprosthetic and total hip replacement systems Veterans Administration Journal of Rehabilitation Research and Development Vol. 24 No. 3 Pages 49 56 Clinical performance of endoprosthetic and total hip replacement systems P. M. SANDBORN, B.S. ; S. D.

More information

Continuing Medical Education Article Imaging of Multiple Myeloma and Related Plasma Cell Dyscrasias JNM, July 2012, Volume 53, Number 7

Continuing Medical Education Article Imaging of Multiple Myeloma and Related Plasma Cell Dyscrasias JNM, July 2012, Volume 53, Number 7 Continuing Medical Education Article Imaging of Multiple Myeloma and Related Plasma Cell Dyscrasias JNM, July 2012, Volume 53, Number 7 Authors Ronald C. Walker 1,2, Tracy L. Brown 3, Laurie B. Jones-Jackson

More information

We have carefully reviewed your questions and the recent promotion for your program on the ABC website.

We have carefully reviewed your questions and the recent promotion for your program on the ABC website. 23 May 2014 Mr P Cronau Senior Producer, Four Corners, The ABC, GPO Box 9994, Sydney NSW 2001. Dear Mr Cronau, We have carefully reviewed your questions and the recent promotion for your program on the

More information

Zimmer M/L Taper Hip Prosthesis. Surgical Technique

Zimmer M/L Taper Hip Prosthesis. Surgical Technique Zimmer M/L Taper Hip Prosthesis Surgical Technique Zimmer M/L Taper Hip Prosthesis 1 Zimmer M/L Taper Hip Prosthesis Surgical Technique Table of Contents Preoperative Planning 2 Determination of Leg Length

More information

Joint Revision Surgery - When Do I Need It?

Joint Revision Surgery - When Do I Need It? Page 1 of 7 Joint Revision Surgery - When Do I Need It? Joint replacement surgery is undoubtedly one of the greatest medical advances of our time. Hip and knee replacements have been performed in millions

More information

In this brochure you will: Keep in mind:

In this brochure you will: Keep in mind: Hip Implant Options Bearing Basics Today s hip implant patients are fortunate. Doctors can now choose from advanced implant materials. One choice doctors make when selecting the implant for you is called

More information

.org. Ankle Fractures (Broken Ankle) Anatomy

.org. Ankle Fractures (Broken Ankle) Anatomy Ankle Fractures (Broken Ankle) Page ( 1 ) A broken ankle is also known as an ankle fracture. This means that one or more of the bones that make up the ankle joint are broken. A fractured ankle can range

More information

FEMORAL NECK FRACTURE FOLLOWING TOTAL KNEE REPLACEMENT

FEMORAL NECK FRACTURE FOLLOWING TOTAL KNEE REPLACEMENT 1 FEMORAL NECK FRACTURE FOLLOWING TOTAL KNEE REPLACEMENT László Sólyom ( ), András Vajda & József Lakatos Orthopaedic Department, Semmelweis University, Medical Faculty, Budapest, Hungary Correspondence:

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. X3 Knee Technology The Next Generation in Highly Crosslinked Knee Polyethylene Innovation Matters As a pioneer

More information

Minimally Invasive Hip Replacement through the Direct Lateral Approach

Minimally Invasive Hip Replacement through the Direct Lateral Approach Surgical Technique INNOVATIONS IN MINIMALLY INVASIVE JOINT SURGERY Minimally Invasive Hip Replacement through the Direct Lateral Approach *smith&nephew Introduction Prosthetic replacement of the hip joint

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. UHR Universal Head Bipolar System T he Stryker Orthopaedics UHR Universal Head is a leader among bipolar component

More information

Zimmer Longevity Highly Cross-linked Polyethylene

Zimmer Longevity Highly Cross-linked Polyethylene Zimmer Longevity Highly Cross-linked Polyethylene Clinical Value Dossier April 9, 2012 TOC Clinical Value Dossier for Zimmer Longevity Highly Cross-linked Polyethylene 1 Executive Summary 7 2 Burden of

More information

HIP RESURFACING ARTHROPLASTY

HIP RESURFACING ARTHROPLASTY HIP RESURFACING ARTHROPLASTY MEDICAL POLICY Policy Number: 2015T0503K Effective Date: December 1, 2015 Table of Contents BENEFIT CONSIDERATIONS COVERAGE RATIONALE APPLICABLE CODES.. DESCRIPTION OF SERVICES...

More information

UNITED STATES DISTRICT COURT DISTRICT OF MINNESOTA

UNITED STATES DISTRICT COURT DISTRICT OF MINNESOTA EXHIBIT A UNITED STATES DISTRICT COURT DISTRICT OF MINNESOTA IN RE: STRYKER REJUVENATE HIP STEM AND ABG II MODULAR HIP STEM LITIGATION MDL NO. 13-2441 (DWF-FLN) This Document Relates to All Actions. PLAINTIFFS,

More information

Cormet Hip Resurfacing System

Cormet Hip Resurfacing System Cormet Hip Resurfacing System Patient Product Information 325 Corporate Drive Mahwah, NJ 07430 t: 1-888-STRYKER www.aboutstryker.com The information presented in this brochure is for educational purposes

More information

DHCPL_RA2014-170-EXT Page 1 of 6

DHCPL_RA2014-170-EXT Page 1 of 6 September 11th 2015 URGENT: Field Safety Notice RA2014-170(EXT) FSCA identifier: Type of Action: Description: Product Field Action RA2014-170(EXT) Field Safety Corrective Action: Return to supplier. LFIT

More information

YOUR GUIDE TO TOTAL HIP REPLACEMENT

YOUR GUIDE TO TOTAL HIP REPLACEMENT A Partnership for Better Healthcare A Partnership for Better Healthcare YOUR GUIDE TO TOTAL HIP REPLACEMENT PEI Limited M50 Business Park Ballymount Road Upper Ballymount Dublin 12 Tel: 01-419 6900 Fax:

More information

Knotilus TM. Anchor Instability Repair. Technique Guide

Knotilus TM. Anchor Instability Repair. Technique Guide Knotilus TM Anchor Instability Repair Technique Guide Instability Repair Using the Knotilus TM Anchor Introduction While the shoulder has more mobility than any other joint in the body, it is also the

More information

Your Practice Online

Your Practice Online P R E S E N T S Your Practice Online Disclaimer This information is an educational resource only and should not be used to make a decision on Revision Hip Replacement or arthritis management. All decisions

More information

Diagnosing and investigating adverse reactions in metal on metal hip implants

Diagnosing and investigating adverse reactions in metal on metal hip implants Follow the link from the online version of this article to obtain certified continuing medical education credits PRACTICE POINTER Diagnosing and investigating adverse reactions in metal on metal hip implants

More information

Radiologic Assessment of Total Hip Arthroplasty: Loosening

Radiologic Assessment of Total Hip Arthroplasty: Loosening March 2004 Radiologic Assessment of Total Hip Arthroplasty: Loosening Grant Garrigues, Harvard Medical School Anatomy Liner Acetabular Component Calcar Stem Femoral Component 2 www.yoursurgery.com/proceduredetails.cfm?br=5&proc=27

More information

UNITED STATES DISTRICT COURT CENTRAL DISTRICT OF ILLINOIS PEORIA DIVISION

UNITED STATES DISTRICT COURT CENTRAL DISTRICT OF ILLINOIS PEORIA DIVISION UNITED STATES DISTRICT COURT CENTRAL DISTRICT OF ILLINOIS PEORIA DIVISION MARCIA OLSON, ) ) Plaintiff, ) ) Case No: v. ) ) HOWMEDICA OSTEONICS CORP. d/b/a ) STRYKER ORTHOPAEDICS, ) JURY TRIAL DEMANDED

More information

Health Benchmarks Program Clinical Quality Indicator Specification 2013

Health Benchmarks Program Clinical Quality Indicator Specification 2013 Health Benchmarks Program Clinical Quality Indicator Specification 2013 Measure Title USE OF IMAGING STUDIES FOR LOW BACK PAIN Disease State Musculoskeletal Indicator Classification Utilization Strength

More information

Metasul LDH Large Diameter Head

Metasul LDH Large Diameter Head Metasul LDH Large Diameter Head Surgical Technique Metasul LDH Large Diameter Head Surgical Technique Enhancing Stability and Increasing Range of Motion Metasul LDH Large Diameter Head Surgical Technique

More information

ReCap Product Rationale

ReCap Product Rationale FLH 155 01/09 Biomet UK Ltd Waterton Industrial Estate Bridgend, South Wales CF31 3XA, United Kingdom Tel. +44 (0)1656 655221 Fax. +44 (0)1656 645454 ReCap Product Rationale A complete resurfacing system

More information

IN THE FEDERAL COURT OF AUSTRALIA (FCA) NEW SOUTH WALES REGISTRY - FEDERAL COURT OF AUSTRALIA

IN THE FEDERAL COURT OF AUSTRALIA (FCA) NEW SOUTH WALES REGISTRY - FEDERAL COURT OF AUSTRALIA IN THE FEDERAL COURT OF AUSTRALIA (FCA) NEW SOUTH WALES REGISTRY - FEDERAL COURT OF AUSTRALIA GENERAL DIVISION No: NSD213/2011 NOTICE OF FILING This document was filed electronically in the FEDERAL COURT

More information

A Patient s Guide to Artificial Cervical Disc Replacement

A Patient s Guide to Artificial Cervical Disc Replacement A Patient s Guide to Artificial Cervical Disc Replacement Each year, hundreds of thousands of adults are diagnosed with Cervical Disc Degeneration, an upper spine condition that can cause pain and numbness

More information

High-Flex Solutions for the MIS Era. Zimmer Unicompartmental High Flex Knee System

High-Flex Solutions for the MIS Era. Zimmer Unicompartmental High Flex Knee System High-Flex Solutions for the MIS Era Zimmer Unicompartmental High Flex Knee System Zimmer Unicompartmental High Flex Knee Built On Success In today s health care environment, meeting patient demands means

More information

Oxford Partial Knee. A Definitive Implant. Tibial Component. Anatomical shape for optimal bone coverage

Oxford Partial Knee. A Definitive Implant. Tibial Component. Anatomical shape for optimal bone coverage Oxford Partial Knee Oxford Partial Knee A Definitive Implant With over 35 years clinical experience, the Oxford Partial Knee is the most widely used 1 and proven 2 partial knee system in the world. Tibial

More information

PRESENTATIONS/PAPERS INTERNATIONAL. RAISING STANDARDS IN JOINT ARTHROPLASTY Course Co-Chairman, The Great Debate - London, June 2013

PRESENTATIONS/PAPERS INTERNATIONAL. RAISING STANDARDS IN JOINT ARTHROPLASTY Course Co-Chairman, The Great Debate - London, June 2013 PRESENTATIONS/PAPERS INTERNATIONAL 1998 2013 RAISING STANDARDS IN JOINT ARTHROPLASTY Course Co-Chairman, The Great Debate - London, June 2013 HOW TO CHOOSE THE RIGHT STEM FOR THE RIGHT PATIENT Corin Symposium

More information

Simplified surgery. Personalized performance.

Simplified surgery. Personalized performance. Simplified surgery. Personalized performance. VISIONAIRE Patient Matched Technology Accuracy Advanced surgical precision, resulting in reproducible outcomes Efficiency Simplifying surgery, reducing costs,

More information

HEADER TOTAL HIP REPLACEMENT SURGERY FROM PREPARATION TO RECOVERY

HEADER TOTAL HIP REPLACEMENT SURGERY FROM PREPARATION TO RECOVERY HEADER TOTAL HIP REPLACEMENT SURGERY FROM PREPARATION TO RECOVERY ABOUT THE HIP JOINT The hip joint is a ball and socket joint that connects the body to the legs. The leg bone is called the femur. The

More information

TRI Product NewsFlash. December 2015

TRI Product NewsFlash. December 2015 TRI Product NewsFlash December 2015 Study Overview 2015 Dear Partners Year in, year out, we are screening all major scientific journals to ensure that our TRI Performance Concept still reflects the latest

More information

Options for Cervical Disc Degeneration A Guide to the M6-C. clinical study

Options for Cervical Disc Degeneration A Guide to the M6-C. clinical study Options for Cervical Disc Degeneration A Guide to the M6-C clinical study Each year, hundreds of thousands of adults are diagnosed with Cervical Disc Degeneration, an upper spine condition that can cause

More information

Disclosures. ! Consultant. ! Stryker. ! Smith and Nephew. Why Resurface? ! Save bone the day of surgery. ! No bone loss from stressshielding

Disclosures. ! Consultant. ! Stryker. ! Smith and Nephew. Why Resurface? ! Save bone the day of surgery. ! No bone loss from stressshielding Disclosures irmingham Hip Resurfacing Hip Resurfacing in 2015 Peter rooks MD, FRCS(C) Cleveland Clinic! Consultant! Stryker! Smith and Nephew HR Goals Why Resurface? Why Resurface?! Summarize benefits!

More information

COMPARISON OF DIFFERENT DIAGNOSTIC MODALITIES IN EVALUATION OF INFECTION OF TOTAL HIP ARTHROPLASTY. Summary

COMPARISON OF DIFFERENT DIAGNOSTIC MODALITIES IN EVALUATION OF INFECTION OF TOTAL HIP ARTHROPLASTY. Summary J Biomed Clin Res Volume 2 Number 1, 2009 Original Article COMPARISON OF DIFFERENT DIAGNOSTIC MODALITIES IN EVALUATION OF INFECTION OF TOTAL HIP ARTHROPLASTY Plamen S. Kinov, 1 Teodor Z. Petrov, 1 Dilyana

More information

.URGENT DEVICE CORRECTION

.URGENT DEVICE CORRECTION zimmer.urgent DEVICE CORRECTION Zimmer, Inc. P.O. Box 706 Warsaw, IN 46561-0706 574.267.6131 www.zimmer.com July 22, 2008 Dear Surgeon: Since we last wrote to you in May 2008 regarding the Durom Acetabular

More information

Case Series on Chronic Whiplash Related Neck Pain Treated with Intraarticular Zygapophysial Joint Regeneration Injection Therapy

Case Series on Chronic Whiplash Related Neck Pain Treated with Intraarticular Zygapophysial Joint Regeneration Injection Therapy Pain Physician 2007; 10:313-318 ISSN 1533-3159 Case Series Case Series on Chronic Whiplash Related Neck Pain Treated with Intraarticular Zygapophysial Joint Regeneration Injection Therapy R. Allen Hooper

More information