Transmitter for NH 3 CI 21 Operation Manual
|
|
- Lynn Simmons
- 7 years ago
- Views:
Transcription
1 Transmitter for NH 3 CI 21 Operation Manual
2 Content Page For your Safety 3 General Description 3 Detection Principle 3 Operational Notes 4 Design 4 Mounting Position 5 Mounting 5 Installation of Electrical Connections 6 Putting into Operation 6 Check of Electrical Zero Point 7 Check of Sensitivity (Calibration) / bump test 7 Check of Zero Point (Dead Band) 7 Trouble Shooting 8 Service 8 Service Address 8 Maintenance and Inspection 9 Spare Parts 9 Accessories 9 Connection Diagram CI 21 Built-in Sensor 10 Connection Diagram CI 21 External Sensor 11 Technical Data 12 2
3 TYPELABEL I NS I D E M E N U C H A N N E L O N F A UL T % For your Safety As with any piece of complex equipment, the GfG transmitter CI21 will do the job designed to do, only, if it is used and serviced in accordance with the manufacturer's instructions. Please protect yourself and your employees by following them. All individuals who have or will have the responsibility for using and servicing this product must carefully read this manual. The warranties made by GfG with respect to the product are voided, if the adjustment of functions or parameters is changed without the permission of GfG. They are also voided, if the product is not used and serviced in accordance with the instructions in this manual. Failures or false alarms caused by interfering gases or electrical signals are not part of the warranty obligation. The above does not alter any statements by GfG regarding warranties, conditions of sale and or delivery. General Description A fixed gas monitoring system consists of the transmitter (MWG) and a controller (GMA), which are connected by means of cable. The transmitter converts the unit for the gas concentration into an electrical signal and transmits it over the cable to the controller. Fixed Gas Detection System AMMONIA ou N A3 nm A2 A1 U A 3 A 2 A 1 FA UL T U TEST MENU QU T MENU Detection Evaluation Warning Fig. 1 The CI21 is an "intelligent" transmitter. The comprehensive electronics allows easy operation and maintenance and also increases the operational safety and accuracy. The following features characterize the CI21: Easy calibration at site. Compensation of temperature effects. Long sensor life. Detection Principle The CI21 is operated with a Charge Carrier Injection sensor. As long as there is no gas in the ambient air, the internal resistance of the sensor is rather high. As soon as NH 3 (Ammonia) is adsorbed at the sensor surface, the internal resistance is reduced. The change in resistance is the measure for the gas concentration and is converted into a standardized current signal ( ma or ma) by the integrated electronics. The Charge Carrier Injection sensor has a proven long-term stability, and selectivity. 3
4 Operational Notes The CI21 passes function and performance tests before shipment and is calibrated with suitable test gases. This does not overrule the need to perform another function test after putting the transmitter in operation. Design The design of the CI21 is shown in fig. 2. The sensor is mounted in the sensor support (pos. 5). The gas enters the sensor chamber through the diffusion inlet (pos. 6). The electronics convert the measurement signal into the output of ma or ma that is supplied to the controller by means of cable. CI21 Design 1. Casing top 2. Casing bottom 3. Type label 4. EMC cable gland (PG 11) 5. Sensor support 6. Diffusion inlet Fig. 2 Poti A Poti 0 Poti C Test Contacts Jumper Epoxy Resin Fuse Terminals EMC - PG 11 Sensor Fig. 3 / Schematic 4
5 Mounting Position It is essential to know the ambient conditions, which have to be taken into consideration before deciding on the mounting position. To achieve representative measurement results, take care of: the room ventilation, the gas density, and the danger highlights. Install the CI21 at a place where the gases pass the sensor even in case of bad ventilation. If necessary, use a smoke cartridge to check. Furthermore, take the following into consideration as well: Ambient temperature, Rain water, hose water, dripping water, condensation and Dust in the atmosphere. The CI21 is to a great extent protected against the entering of water and dust (IP 54). Special accessories are available to provide additional protection for very difficult conditions. Please contact GfG for detailed information.! Warranty may be voided, if the transmitter is exposed to ambient conditions that were unknown to GfG during planning, production or delivery. When deciding on the position for the transmitter, make sure that it is always accessible for service and maintenance. It is recommended to mount the CI21 with the sensor facing down to the floor. A different mounting position, however, does not affect the sensor accuracy. Mounting For connecting the CI21 refer to the connection diagram (also refer to the terminal diagram of your GMA controller). For mounting the transmitter remove the four lock screws and lift the casing top off. Fix the casing by means of two screws through the relevant borings.! The CI21 is not certified for use in hazardous areas 5
6 Installation of Electrical Connections An electrician obeying the applicable regulations should do procurement of cable and electrical connections. A shielded 18/3 cable should be used (Belden #8770 or equivalent) for distances less the 500 yards. Note: Shielding is effected over the EMC cable screwing. Terminals Resin Terminals Resin red red green yellow Signall Ground Voltage Supply EMC-PG 11 green brown white blue Signall Ground Voltage Supply EMC-PG 11 Sensor Sensor Sensor Connecting Wire, max. 2 m External Sensor Fig. 4 The supply voltage for the transmitter (terminal 4) is V DC. Once the installation is completed, close the casing top and fix it with the screws. Putting into Operation Once the CI21 is installed, a skilled person can put it into operation. Before shipment, the CI21 has been calibrated and tested after a specific warm-up time for several days. When connecting to a controller (power supply) you should allow approx. 60 minutes to warm-up the Charge Carrier Injection sensor before checking the sensitivity. Should you notice that a re-calibration is required, allow an extended warm up period of about 4 hours. In case you replace the sensor, adjust the jumper of the transmitter according to the specification on the new sensor. (Sensor replacement requires adjusting the dead band for the new sensor) The CI21 is calibrated and tested before shipment. Certain deviations, however, are possible due to transport, mounting and ambient conditions. Inside the CI21 there are two test contacts to check the electrical output signal (fig. 3). The voltage at the test contacts can be measured with a voltmeter. Warm-up time is approx. 60 minutes. Poti A Poti 0 Poti C Jumper Test Contacts Epoxy Resin Fuse Terminals EMC - PG 11 Sensor Fig 3 6
7 Check of Electrical Zero Point 1) Use a multimeter to check the output signal by measuring the voltage at the test contacts. 2) The reading should be in the range of: V Pot 0 Test Contacts V 3) In case of deviation it is possible to re-adjust the value by turning potentiometer O Sensor Check of Sensitivity (Calibration) / bump test 1) Use a multimeter to check the output signal by measuring the voltage at the test contacts. 2) Attach the calibration adapter to the diffusion inlet of the transmitter. Then select a suitable test gas (200, 1000 or 10,000ppm NH3 depending on the detection range) Pot C Test Contacts 1.00 V 3) Allow a flow of 0.5LPM or 1CFH for 3-5 minutes before reading and adjustments. 4) The reading should be in the range of 1.00V +/- 10% Sensor 5) In case of deviation you can correct the value by turning potentiometer C. When using a GMA controller signal checks and adjustments may be done at the GMA (please reference GMA manual).! Do not use a test gas with a balance of nitrogen. Calibration gas must be GfG approved. Check of Zero Point (Dead Band) Adjustment to the zero point (Pot A) should not be done without direct instruction from GfG. 1) Use a multimeter to check the output signal by measuring the voltage at the test contacts. 2) Attach the calibration adapter to the diffusion inlet of the transmitter. Then select a suitable test gas (20 or 30 ppm NH 3 depending on the detection range). 3) Allow a flow of 0.5LPM or 1CFH for 3-5 minutes before reading and adjustments. 4) The signal should be approx V. Pot A Sensor Test Contacts V 5) In case of deviation it is possible to re-adjust the value by turning potentiometer A. Potentiometer A is factory set and should not be adjusted without direct instruction or training from GfG. 7
8 Trouble Shooting CI 21 Troubleshooting Symptom Reason/s Solution 0 ma output current incorrect wiring verify wiring to CI 21 diagram defective fuse replace fuse cut cable/wire reconnect wiring dc voltage too high correct voltage to vdc defective electronics contact GFG MFG. for instructions no reaction to test gas improper gas use GFG approved test gas faulty sensor replace sensor Controller shows incorrect reading with test gas applied. controller not calibrated calibrate controller CI 21 out of calibration calibrate CI 21 High output current with no gas present. open ground wire reconnect ground wire possible interfering background gas identify background gas and eliminate Slow sensor response incorrect test gas use GFG approved test gas humidifier sponge too wet wring out sponge humidifier sponge too dry moisten sponge sensor defective replace sensor Service Service includes maintenance, inspection and repair of gas warning equipment and function checks. A function check includes: Check with alarm gas concentration Check of zero point and sensitivity (calibration) Check of response time Check of gas sampling and gas processing system (if any) Check of alarm signal activation Check of failure alarm The function check should be done before putting the system into operation, and it should be repeated at least once a year. Service Address For additional questions on the product or in case of failure and problems please contact: GfG Instrumentation, Inc Oak Valley Drive Suite 20 Ann Arbor, Mi Phone: Fax: info@gfg-inc.com Web: 8
9 Maintenance and Inspection With maintenance and inspection the proper and planned status of the gas detection system is controlled. The CI 21 does not require special maintenance. Depending on the ambient conditions, gas detection systems may show a different behavior. Therefore, it is important to perform a visual check every day, particularly during the first few days after putting into operation. Check of pump assembly and gas processing system (if any). Check of gas supply for soiling (for correct measurement the gas supply to the sensor must not be blocked). Sensors are subject to ageing and exhausting processes. Depending on their type and on their exposure to gas they are more or less used. A sensitivity check can only be done with a suitable test gas after certain periods of time. The sensitivity calibration is an expert calibration and is usually performed by GfG service or by authorized persons. Spare Parts Part No. Spare Sensor/built-in Spare Sensor/external In case you replace the sensor, adjust the jumper inside the transmitter according to the specification on the new sensor. (Sensor replacement requires adjusting the dead band for the new sensor) Accessories Part No. Calibration Adapter Calibration Kit w/o Gas (fixed flow regulator, tubing, inline humidifier & adapter) Calibration gas (200ppm NH3 balance air) Calibration gas (1000ppm NH3 balance air) Calibration gas (10,000ppm NH3 balance air) Calibration Kit with gas (10,000 ppm) Calibration Kit with gas (1,000 ppm) Calibration Kit with gas (200 ppm)
10 Connection Diagram CI 21 Built-in Sensor Built-in Sensor MK red red green yellow white green brown Connect shield to Cable Gland Signal ma GND 20V / 24V GMA GMA motherboard post 1997 (blue) GMA motherboard pre 1996 (no color) GMA GMA GMA81 (A)
11 Connection Diagram CI 21 External Sensor MK External Sensor green brown white blue white green brown Connect shield to Cable Gland Signal ma GND 20V / 24V GMA GMA motherboard post 1997 (blue) GMA motherboard pre 1996 (no color) GMA GMA GMA81 (A)
12 Technical Data Transmitter CI21 Type: Sensor Type: CI21 MK (external) or MK (built-in) Gas: Ammonia, NH 3 Standard range: 200ppm, 1,000ppm or 10,000ppm Gas Supply: Diffusion Detection Principle: Charge Carrier Injection (CI) Supply Voltage: V DC Output Current: mA (4kΩ) or mA (200Ω) Response Time t A : < 8 s Typical Sensor Life: > 2 years Climate Conditions Casing Temperature: C ( F) Humidity: 0 99 % r. h., non condensing Pressure: hpa Material: Aluminum Protection: IP 54 Cable Gland: ½ in. conduit Transmitter Cable: 3 core shielded cable AWG 18 for runs less than 500 yards Belden #8770 or equivalent Dimensions: 100 x 100 x 57 mm (WxHxD) Weight: approx. 370 g 12
13 Notes
14
15
16 1194 Oak Valley Drive, Suite 20 Ann Arbor, Michigan United States of America Phone: (800) or (734) Fax (734) Website: GfG reserves the right of modification JA 7004-C21
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
More informationCombustible Gas Transmitter 65-2400RK-05
Combustible Gas Transmitter 65-2400RK-05 Overview This detection insert describes the combustible gas transmitter. This insert also describes how to install, start up, maintain, and calibrate the transmitter.
More informationIntroduction. Refrigerant Leak Detecting System. Feature and Benefits
PSC European Refrigeration Controls Catalogue Catalog Section 9 Product Bulletin PD-GAS-E Refrigerant Leak Detecting System Introduction This range of refrigerant leak detecting systems is designed for
More informationPolyGard Carbon Monoxide CO Transmitter ADTX3 1110
PolyGard Carbon Monoxide CO Transmitter ADTX3 1110 DESCRIPTION CO transmitter including digital measurement value processing and temperature compensation for the continuous monitoring of the ambient air
More informationPART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
More informationCONTOIL DFM 8EDM. Table of contents. Mounting and operating instructions
Mounting and operating instructions CONTOIL DFM 8EDM Table of contents Safety instructions 2 - Designed use - Installation, commissioning and operation - Operational safety - Return of the instruments
More informationTIG INVERTER INSTRUCTION MANUAL
TIG INVERTER INSTRUCTION MANUAL Contents Warning General Description Block Diagram Main Parameters Circuit Diagram Installation and Operation Caution Maintenance Spare Parts List Troubleshooting 3 4 4
More informationPSDC 08128. PSDC 12V/8A/8x1A Power supply for up to 8 analog cameras PTC/TOPIC.
PSDC 08128 v.1.2 PSDC 12V/8A/8x1A Power supply for up to 8 analog cameras PTC/TOPIC. EN** Edition: 9 from 25.05.2015 Supersedes the edition: 8 from 12.12.2014 the 8x 1A/12V DC power output for powering
More information98371000 REV. REL 1302 WEST BEARDSLEY AVENUE P.O. BOX 1127 ELKHART IN 46515 (574) 295-8330 FAX (574) 293-9914 2011 ELKHART BRASS MFG. CO., INC.
81471067 Electric Monitor Motor Control Panel For use with Model 8394059 SPIT-FIRE Monitor Installation, Operation, and Maintenance Instructions FOR ATEX APPLICATIONS 1302 WEST BEARDSLEY AVENUE P.O. BOX
More informationEX-5175-EC Sensor/Transmitter Manual
PO Box 979 Ann Arbor, MI 48106-0979 www.enmet.com EX-5175-EC Sensor/Transmitter Manual Manual Part Number 80003-099 MCN-13-008, 11/06/13 Table of Contents 1.0 INTRODUCTION... 1 1.1 Unpack... 1 1.2 Check
More informationModel 201 Wiegand Touchpad Reader Installation Guide
Model 201 Wiegand Touchpad Reader Installation Guide P/N 460353001C 15AUG11 2011 UTC Fire & Security. All rights reserved. This document may not be copied in whole or in part or otherwise reproduced without
More informationACS-30-EU-EMDR-10-MOD
Multi-application heat tracing Control and Monitoring in commercial and residential buildings Roof & Gutter de-icing sensor module Technical Information Approvals Module IP rating Ambient operating temperature
More informationModel SRMD Setra Remote Monitoring Display
Model SRMD Setra Remote Monitoring Display 1.0 GENERAL INFORMATION Thank you for purchasing the Setra Remote Monitoring Display (SRMD). The SRMD is a digital panel meter with a bright 1 LED display for
More informationPower Supply and Indicator Unit Type 5024-1. Fig. 1 Type 5024-1. Mounting and Operating Instructions EB 9539 EN
Power Supply and Indicator Unit Type 5024-1 Fig. 1 Type 5024-1 Mounting and Operating Instructions EB 9539 EN Edition October 2003 Application 1 Application The Type 5024-1 Power Supply and Indicator Unit
More informationDuct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
More informationSensor for Ammonia. 11511 West 83 rd Terrace email sales@manningsystems.com. 1.800.444.9935 or 913.894.1185. fax 913.894.1296
EC-F2-NH 3 Sensor for Ammonia Instruction and Installation Manual Rev H MOUNT ENCLOSURE THIS END UP. DO NOT BLOCK PERFORATED VENT HOLES. SENSOR 11511 West 83 rd Terrace email sales@manningsystems.com Lenexa,
More informationPart I - Installation
400 Series Pressure and Differential Pressure Switches Types: H400, H402, H403, H400K, H402K, J400, J402, J403, J400K, J402K UNITED ELECTRIC CONTROLS Installation and Maintenance Instructions Please read
More informationVentostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Measurement & Control. Features:
GE Measurement & Control Ventostat Wall Mount Telaire Wall Mount, Humidity and Temperature Transmitters Features: Patented, Absorption Infrared Gas sensing engine provides high accuracy in a compact low
More informationContents. Document information
User Manual Contents Document information... 2 Introduction... 3 Warnings... 3 Manufacturer... 3 Description... Installation... Configuration... Troubleshooting...11 Technical data...12 Device Scope: PCB
More informationENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 ENG 97D STAND-ALONE MANUAL. Manual Part Number 80003-035 MCN-389, 01/21/08
PO Box 979 Ann Arbor, MI 48106-0979 ENG 97D STAND-ALONE MANUAL Manual Part Number 80003-035 MCN-389, 01/21/08 Table of Contents 1.0 Introduction...1 1.1 Unpack...1 1.2 Check Order...1 1.3 Serial Numbers...1
More informationSolar Home System. User Manual. AEH-SHS01-10W2L Solar Home System 2 Lamps
Solar Home System User Manual AEHSHS0110W2L Solar Home System 2 Lamps All rights reserved Specifications subject to change without prior notice 2 Dear Customer, Thank you for purchasing Schneider Electric
More informationOEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &
More informationAdvantium 2 Plus Alarm
ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia
More informationSeries 510 Submersible Level Transmitters
Series 510 Submersible Level Transmitters The Series 510 Submersible Level Transmitters are solid state instruments designed for direct submergence into many types of liquid for quick, accurate and reliable
More informationABB Drives. User s Manual. Pulse Encoder Interface Module RTAC-01
ABB Drives User s Manual Pulse Encoder Interface Module RTAC-0 Pulse Encoder Interface Module RTAC-0 User s Manual 3AFE 64486853 REV A EN EFFECTIVE:.5.00 00 ABB Oy. All Rights Reserved. Safety instructions
More informationGREISINGER electronic GmbH
T38.0.0X.6C-03 CO2 - Transmitter Operating Manual GT10 - CO2-1R GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 Tel.: +49 9402 / 9383-0, Fax: +49 9402 / 9383-33, email: info@greisinger.de
More informationI nstallation. M100Q Series Proportional Actuator with R81Q Controller Board for Thermistor Sensor Applications. Tools Needed.
FANs 268.1, 1628.3 Installation Bulletin M100Q Issue Date 1099 M100Q Series Proportional Actuator with R81Q Controller Board for Thermistor Sensor Applications I nstallation Parts Included M110QGA-1 and
More informationOPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
More informationViZion Installation Guide
ViZion Installation Guide v2.0 1 ViZion Installation Guide Table of Contents Inventory Hardware Setup Understanding Synchronization Cable Begin by taking an inventory of the required equipment DR Unit
More informationATL Fuel Level Sender Probes
T E C H N I C A L S P E C I F I C A T I O N The ATL EL-AD-151 (Resistance Output) and EL-AD-152 (Voltage Output) Fuel Level Senders are highly advanced sensors for continuously measuring the contents of
More informationManual for Fire Suppression & Methane Detection System
Manual for Fire Suppression & Methane Detection System Fogmaker North America Post address: 150 Gordon Dr Exton, PA 19341 Delivery address: 150 Gordon Dr Exton, PA 19341 Tel: 610-265-3610 Fax: 610-265-8327
More informationStreaming Potential System (SPT1000, SPD1000)
Streaming Potential System (SPT1000, SPD1000) User's Manual Sentrol Systems, Inc. 3949 Cotswold Dr. SW, Lilburn, GA 30047-2371 Tel: 770-564-1541, Fax: 770-564-8605 www.sentrolsystems.com Since the streaming
More informationEE-SPY801/802. Photomicrosensors for detecting wafer-carrier mounting. Wafer-carrier Mounting Photomicrosensors. Ordering Information.
Wafer-carrier Mounting Photomicrosensors EE-SPY1/ CSM_EE-SPY1 DS_E_3_ Photomicrosensors for detecting wafer-carrier mounting. The mounting position is set with a pedestal. The contact surface with the
More informationFIRERAY 2000 Installation Guide
FIRERAY 2000 Installation Guide Features Range 33ft to 330 ft. 24Vdc operation Selectable alarm thresholds Low current consumption Ground level electronics Manual or Automatic reset System Description
More information12 Volt 30 Amp Digital Solar Charge Controller
12 Volt 30 Amp Digital Solar Charge Controller User s Manual WARNING Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions
More informationAXIS T81B22 DC 30W Midspan
INSTALLATION GUIDE AXIS T81B22 DC 30W Midspan ENGLISH About this Document This document includes instructions for installing AXIS T81B22 on your network. Previous experience of networking will be beneficial
More informationWater Leak Detection System
Water Leak Detection System Installation and Operating Manual 505-334-5865 ph 505-334-5867 fax www.rodisystems.com email:info@rodisystems.com 936 Highway 516 Aztec, NM 87410-2828 Manual Revisions and Copyright
More informationSubminiature Load Cell Model 8417
w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...
More informationBEP FD-2 CONTOUR MATRIX FUME DETECTOR
BEP FD-2 CONTOUR MATRIX FUME DETECTOR INSTALLATION AND OPERATING INSTRUCTIONS Table of Contents 1 WARRANTY...3 2 IMPORTANT...3 3 CAUTIONS WHEN USING GAS SENSORS...3 4 FEATURES...4 5 SPECIFICATION...4 6
More informationY7230A CO 2 Alarm System
Y7230A CO 2 Alarm System FOR CO 2 GAS DETECTION PRODUCT DATA FEATURES APPLICATION Y7230A CO 2 alarm system is used to detect gas leaking from CO 2 beverage dispensing systems and posts an alarm to occupants
More informationi ChatterBox! Motorcycle Security
i Before you Start the Installation * Please read this manual to become familiar with the requirements necessary to complete the installation. * Use a high quality multi-meter to test all wires before
More informationRAEGuard EC FGM-13XX
RAEGuard EC FGM-13XX User s Guide P/N 033-4102-000 Rev B May 2007 - READ BEFORE OPERATING - This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining,
More information600-GD & 600-GDL 600 SERIES GAS DETECTOR
600-GD & 600-GDL 600 SERIES GAS DETECTOR INSTALLATION AND OPERATING INSTRUCTIONS This page has been deliberately left blank. Page 1 Table of Contents 1 WARRANTY...3 2 IMPORTANT...3 3 CAUTIONS WHEN USING
More informationPSU-U-24V/AC-6A/10/FTA-TR-MC PSAC 0624
PSU-U-24V/AC-6A/10/FTA-TR-MC PSAC 0624 v.1.0 AC/AC power supply unit Edition: 1 from 24 th October 2008 Supercedes edition: ---------- 10062009 CONTENTS: 1. Technical description. 1.1. General description
More informationPIEZOMETER AIRFLOW MEASURING RING
PIEZOMETER AIRFLOW MEASURING RING M.K. Plastics is pleased to introduce The Piezometer Airflow Measuring Ring, now available as an accessory for our steel Axijet fans. The system consists of a Piezometer
More informationSingle Station Remote Alarm
ADI 5106G Certified ISO 9001:2000 Single Station Remote Alarm 529 5106-01-120 529 5106-01-220 INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Controls Corporation
More informationKNX TH65-AP. Thermo-Hygrometer. Technical specifications and installation instructions Item number 70184
EN KNX TH65-AP Thermo-Hygrometer Technical specifications and installation instructions Item number 70184 Elsner Elektronik GmbH Control and Automation Engineering Herdweg 7 D 75391 Gechingen Phone +49
More information19 LED Tube Controller ORDERCODE 41003
19 LED Tube Controller ORDERCODE 41003 Congratulations! You have bought a great, innovative product from Showtec. The Showtec 19 LED Tube Controller brings excitement to any venue. Whether you want simple
More informationINSTALLATION INSTRUCTIONS
LIGHTING CONTROL PANELS 4 AND 8 RELAYS INSTALLATION INSTRUCTIONS INSTALLATION OVERVIEW The installation instructions contained in this document are provided as a guide for proper and reliable installation.
More informationCharge Regulator SCR 12 Marine
Charge Regulator SCR 12 Marine Manual Many thanks for purchasing a superwind product. The SCR 12 Marine is a charge regulator of highest quality and will perfectly and reliably charge your batteries for
More informationStructureScan HD Module. Installation Guide ENGLISH. www.bandg.com www.simrad-yachting.com www.lowrance.com
StructureScan HD Module Installation Guide ENGLISH www.bandg.com www.simrad-yachting.com www.lowrance.com Disclaimer As Navico is continuously improving this product, we retain the right to make changes
More information543-0032-00, 943-0032-00. User s Manual
543-0032-00, 943-0032-00 User s Manual 1 Comfort Alert Diagnostics Faster Service And Improved Accuracy The Comfort Alert diagnostics module is a breakthrough innovation for troubleshooting heat pump and
More informationUser Manual DONGGUAN SHENGYANG INDUSTRIAL CO., LTD. SY WMVC Smart Grid Tie Microinverter. Dongguan Shengyang Industrial Co., Ltd.
DONGGUAN SHENGYANG INDUSTRIAL CO., LTD SY WMVC Smart Grid Tie Microinverter User Manual Dongguan Shengyang Industrial Co., Ltd. ADD: No. 1 Shangxin Road, New Village, Xin An Community, Chang An Town, Dongguan
More informationCAUTION OPC-LM1-ID. Option Card for Frequency Divider. Fuji Electric FA Components & Systems Co., Ltd. Instruction Manual
Instruction Manual Option Card for Frequency Divider Deliver this instruction manual without fail to those who actually operate the equipment. Read this operation manual and understand the description
More informationDetector transparent with Color Inserts. FAA 500 TR P Trim Ring transparent with Color Inserts. FCA 500 / FCA 500 E Detector Bases
Detector Color Detector transparent with Color Inserts FAA 500 TR W Trim Ring FAA 500 TR P Trim Ring transparent with Color Inserts FAA 500 BB Ceiling Mount Back Box FCA 500 / FCA 500 E Detector Bases
More informationEzFlex Combustible Gas Leak Detector Model EZ40
User's Guide EzFlex Combustible Gas Leak Detector Model EZ40 Introduction Congratulations on your purchase of the Extech EZ40 Combustible Gas Leak Detector. The EZ40 detects the presence of combustible
More informationOzone Generator Manual
ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 Ozone Generator Manual 80003-118 MCN-282, 05/14/04 Table of Contents 1.0 INTRODUCTION... 1 1.1 Unpack...1 1.2 Check Order...1 1.3 Serial Numbers...1
More informationLED...A70..4-..Q Series Sealed High-Intensity
LED...A70..4-..Q Series Sealed High-Intensity Area Lights High-Power Lighting for use with PresencePLUS and Other Vision Systems Features Rugged, waterproof housing, rated IEC IP68 Compact area light for
More informationDigital inductive conductivity transmitter
Digital inductive conductivity transmitter Optimal solution for conductivity measurements in difficult fluids (polluted, dirty,...) PEEK/PPA version for CIP applications Large range of process connections
More informationTechnote. Frese OPTIMA Compact actuators DN10-DN32. Application. Features thermic actuators. Features motoric actuators. Approval. www.frese.
Page 1 of 6 Application Proportional 0-10V or 3-position modulating or On/ Off control of Frese OPTIMA Compact valves in heating, ventilating and air conditioning systems. The actuator can be mounted on
More informationSDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual January 2011 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
More informationMCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
More informationHow To Use A Ds340 (Dsp)
Ideal for monitoring dust levels in the exhaust gas of industrial combustion or air filtration processes. Innovative Dynamic Detection Principle (DDP) measurement technique Immune to gradual reductions
More informationPK-01. Standalone door control module. SATEL sp. z o.o. ul. Schuberta 79 80-172 Gdańsk POLAND tel. + 48 58 320 94 00
Standalone door control module PK-01 Firmware version 1.00 pk-01_en 06/12 SATEL sp. z o.o. ul. Schuberta 79 80-172 Gdańsk POLAND tel. + 48 58 320 94 00 info@satel.pl www.satel.eu WARNINGS Read carefully
More informationEPSOLAR LS0512R / LS0524R. Solar Light Controller INSTRUCTION MANUAL. Please read this manual carefully before using the product!
EPSOLAR LS0512R / LS0524R Solar Light Controller INSTRUCTION MANUAL Please read this manual carefully before using the product! LandStar LS0512R / LS0524R Solar Light Controller Nominal system voltage
More informationBLUE RIBBON CORP. BC001 Birdcage Installation Manual
Product Overview: This manual is applicable for the BC001 Birdcage Submersible Level Transmitters. If the product you have has a different prefix part number than listed above please contact the factory
More informationInstallation and Operation Guide for PD4100 Series Power Control Centers
Installation and Operation Guide for PD4100 Series Power Control Centers Extended warranties are available for purchase at www.progressivedyn.com Member Thank you for selecting Progressive Dynamics as
More informationSeries AMEPR30-AZ up to 2.5A AC-DC / DC-DC LED Driver / Converter
FEATURES: Click on Series name for product info on aimtec.com Models Single output Model Max Output Power (W) 1 AC-DC Constant Current or Constant Over temperature protection Voltage LED Driver Over current
More informationSeries 6000 Torque measured metal bellow coupling
Properties Free of float metal bellow coupling with integrated torque measurement Non-contact measurement system, high robustness High torsional stiffness Limited torque of inertia Performance Measurement
More informationFLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases
FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY RESULT IN INJURY TO YOU OR DAMAGE TO
More informationHow To Control A Power Supply With A Mini Transmitter And Switch (Power Supply)
transmitter (2 wire) / switch for continuous or On/Off Control Indication, monitoring, transmitting and continuous or On/Off control in one device Output signal 4...20 ma, 2-wire for continuous control
More informationWE-350 Series ¼ Turn Electric Actuator
WE-350 Series ¼ Turn Electric Actuator Operation and Installation Manual Pg 1 (Rev. 020113) Table of Contents 1.0 General 1.1 Pre-Installation Inspection 1.2 Storage 1.3 Features & General Information
More informationLMU-5000. Hardware and Installation Guide
LMU-5000 Hardware and Installation Guide Plan The Installation Verify Power, Ground and Ignition. Be sure to check each source (power, ground and ignition) to ensure that the proper signaling exists. This
More informationMTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter
Thermal Mass meter and Temperature Transmitter Description The MTI10 insertion thermal mass flowmeter and the MTL10 in-line thermal mass flowmeter provide accurate mass flow measurement of clean, dry gases
More informationModels SFP33 & SFP32 Pressure Transmitters
SFP33 Flow-Through Pressure Transmitter SFP32 Single-Ended Pressure Transmitter DS-PR-SFP-eng April, 2016 Data Sheet Models SFP33 & SFP32 Pressure Transmitters Pressure Brooks Seal Free Pressure Measurement
More informationATS Overhead Table Shelf System INSTRUCTION MANUAL
ATS Overhead Table Shelf System INSTRUCTION MANUAL ATS Overhead Table Shelf System Instruction Manual Warranty Newport Corporation warrants this product to be free of defects in material and workmanship
More informationThreaded Inline Air Heaters
Threaded Inline Air Heaters FOR SAFETY & LONG HEATER LIFE, CAREFULLY READ THIS MANUAL BEFORE USE. Safety SHOCK HAZARD! Only qualified individuals should install this heater and related controls. Follow
More informationManual. Solar Fountain Mobile Phone Charger
EN Manual Solar Fountain Mobile Phone Charger Copyrights 2008 Victron Energy B.V. All Rights Reserved This publication or parts thereof may not be reproduced in any form, by any method, for any purpose.
More informationPERF10 Rubidium Atomic Clock
Owner s Manual Audio Stanford Research Systems Revision 1.0 February, 2011 2 Certification Stanford Research Systems certifies that this product met its published specifications at the time of shipment.
More informationSignal Manager Installation and Operation Manual
Signal Manager Installation and Operation Manual Rev.11/28/2008 Part # 26550007 Table of Contents CHAPTER 1: SYSTEM DESCRIPTION... 3 FUNCTION AND THEORY... 3 CHAPTER 2: SYSTEM INSTALLATION... 4 HARDWARE
More informationTelefon (+45) 43 43 82 00 Telefax (+45) 43 43 74 75 mail@uni-valve.com www.uni-valve.com. Installation and operating manual
Uni-Valve A /S VENTILER & INSTRUMENTER Telefon (+45) 43 43 82 00 Telefax (+45) 43 43 74 75 mail@uni-valve.com www.uni-valve.com UNI-EL Electric actuator Installation and operating manual Remote position
More informationOperating instructions Diffuse reflection sensor with background suppression O1D101 O1D104 706114 / 00 01 / 2012
Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D104 706114 / 00 01 / 2012 Contents 1 Preliminary note 3 1.1 Symbols used 3 1.2 Warning signs used 3 2 Safety instructions
More informationCAUTION OPC-LM1-IL. Option Card for Encoder of Line Driver Output. Instruction Manual
Instruction Manual OPC-LM1-IL Option Card for Encoder of Line Driver Output CAUTION Deliver this instruction manual without fail to those who actually operate the equipment. Read this operation manual
More informationFMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters
FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters M-4270/0707, pg. 2 of 29 READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY
More informationTABLE OF CONTENTS. SECTION 1 General Information: Pg 2 Description of Operation
GTRDD SERIES HEATLESS DRYER INSTALLATION, OPERATION & MAINTENANCE MANUAL TABLE OF CONTENTS SECTION 1 General Information: Pg 2 Description of Operation SECTION 2 Safety Instructions: Pg 3 Installation
More informationLED Par 56 Short DMX ORDERCODE 42421
LED Par 56 Short DMX ORDERCODE 42421 Congratulations! You have bought a great, innovative product from Showtec. The Showtec LED Par 56 brings excitement to any venue. Whether you want simple plug-&-play
More informationVANGUARD A OPERATOR S MANUAL PORTER. FM-952 Rev. E 7/07
PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 VANGUARD A OPERATOR S MANUAL FM-952 Rev.
More informationOASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE
OASIS-PLUS 120V INFORMATION & OPERATING INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE 56041852 WARNING:
More informationVARIO PROX CARD READER INSTALLATION MANUAL
VARIO PROX CARD READER INSTALLATION MANUAL TABLE OF CONTENTS 1.0 INTRODUCTION...2 1.1 Legend...3 1.2 Terminology...4 2.0 MOUNTING...5 3.0 WIEGAND CONNECTION...6 4.0 AUX BUTTON INPUT...7 5.0 INSTALLER PROGRAMMING...8
More informationInstallation and Operation Guide for PD4000 Series Power Control Center
Extended warranties are available for purchase at www.progressivedyn.com Installation and Operation Guide for PD4000 Series Power Control Center Member Thank you for selecting Progressive Dynamics as your
More informationSOLAR POWER - THE NEW WAY TO RV!
IMPORTANT! - YOU'RE SOLAR READY SOLAR POWER - THE NEW WAY TO RV! Your RV is PRE-WIRED for SOLAR POWER! Benefits of Solar Zamp Solar Portable Solar Kit Maximize Battery Life Electrical Independence Green,
More informationDifferential Pressure Sensors for air and non-corrosive gases, with calibration certificate
1 919 1919P01 1919P03 QBM65/C, QBM75-1U/C Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate QBM751-1/C QBM65/C QBM75/C Highly accurate measurements with calibration
More informationHow To Power A Schen
Automation / Mini Type SCH32F Hollow Shaft Encoder - Ø 32 mm Hollow Bore: Ø 6 mm to Ø 3/8 inch Resolution up to 5000 ppr IP 65 (IP 50 for IDC connector option) Electrical Specifications Code: Resolution:
More informationDrayton Digistat +2RF/+3RF
/+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer
More informationDMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor
DMP Industrial Pressure Transmitter for Low Pressure Stainless Steel Sensor accuracy according to IEC 60770: standard: 0.5 % FSO option: 0.5 / 0. % FSO from 0... 00 mbar up to 0... 40 bar Output signals
More informationTMS TANK MANAGEMENT SYSTEM
TMS TANK MANAGEMENT SYSTEM Page 1 of 9 Operating Instructions GENERAL The Tank Management System is a bespoke design to control, monitor and accommodate efficient storage and dispensing of TMS. FUNCTIONS
More informationMaterial: Weight: Bearing Life: Shaft Speed: Starting Torque: Mass Moment of Inertia: Shaft Loads:
Automation / Mini Type SCH32F Hollow Shaft Encoder - Ø 32 mm Hollow Bore: Ø 6 mm to Ø 3/8 inch Resolution up to 5000 ppr IP 65 (IP 50 for IDC connector option) Electrical Specifications Code: Resolution:
More information12 Volt 30 Amp Digital Solar Charge Controller Installation & Operation Manual
12 Volt 30 Amp Digital Solar Charge Controller Installation & Operation Manual This 30Amp charge controller is designed to protect your 12Volt Lead-acid or Gel-cell battery from being overcharge by solar
More informationTechnical data sheet TDS0070
Technical data sheet TDS0070 ATEX Certified II 2 G Ex d IIC T4 Gb IECEx Certified Ex d IIC T4 Gb Ta = -20ºC to +60ºC FEATURES INCLUDE: - ATEX and IECEx Certified explosion proof Exd housing for Premier
More informationElectronic Pressure Switch EDS 300
Electronic Pressure Switch EDS 300 User manual Page 2 of 16 Content 1. Functions of the EDS 300...3 2. Mounting...3 3. Operating keys on the membrane keypad...4 4. Digital display...4 5. Output function...5
More information