COMPARISON OF DYNAMIC CONTOUR TONOMETRY (PASCAL ) WITH PNEUMOTONOMETRY AND GOLDMANN TONOMETRY
|
|
|
- Brett Scott
- 10 years ago
- Views:
Transcription
1 ARCH SOC ESP OFTALMOL 2007; 82: ORIGINAL ARTICLE COMPARISON OF DYNAMIC CONTOUR TONOMETRY (PASCAL ) WITH PNEUMOTONOMETRY AND GOLDMANN TONOMETRY COMPARACIÓN DEL TONÓMETRO PASCAL CON EL NEUMOTONÓMETRO Y EL TONÓMETRO GOLDMANN HERAS-MULERO H 1, MORENO-MONTAÑÉS J 2, SÁDABA ECHARRI LM 2, MENDILUCE MARTÍN L 3 ABSTRACT Purpose: To compare the intraocular pressure measurements as defined by the Pascal tonometer, the Goldmann tonometer and the pneumotonometer. Methods: This was an observational clinical study, which included two hundred and five randomly selected subjects recruited from the Ophthalmology Department. The intraocular pressure measurements were performed with each tonometry technique in a randomized order. Results: The Pascal s intraocular pressure measurement was significantly higher than that measured by the other two tonometers (p<0.05). The quality data of Pascal was: optimum in 27.3% (56 of 205 patients), acceptable in 42% (86 of 205 patients) and unacceptable in 23.4% (48 of 205 patients). In 7.3% (15 of 205 patients) it was impossible to obtain any measurement using Pascal. A weak correlation coefficient between the Pascal and the Goldmann, and between Pascal and the pneumotonometer was found. The Bland-Altman method of measurement using these tonometers showed a high degree of discordance. Conclusion: As reported by others authors, the Pascal s intraocular pressure measurement is higher RESUMEN Objetivo: Comparar la medición de presión intraocular con el tonómetro de Pascal en córneas normales frente al tonómetro de Goldmann y al pneumotonómetro. Método: Estudio clínico observacional en el que se han incluido 205 pacientes consecutivos y randomizados que han acudido a la consulta de oftalmología. Se realiza una medición con cada tonómetro en orden aleatorio. Resultados: El Pascal, dio una medición de la presión intraocular mayor que la del resto de tonómetros. Se obtienen diferencias significativas del Pascal frente a los otros dos tonómetros (p<0,05). La fiabilidad del Pascal fue: fiables en el 27,3% (56 de 205 pacientes), aceptable en el 42% (86 de 205 pacientes) y en el 23,4% (48 de 205 pacientes) inaceptable. En el 7,3% (15 de 205 pacientes) restante no se obtuvo ninguna medición con el tonómetro de Pascal. Al comparar el tonómetro de Pascal frente al de Goldmann y al neumotonómetro se observa que el coeficiente de correlación es débil y presenta alta discordancia con el método de Bland- Altman. Received: 31/5/06. Accepted: 24/5/07. Department of Ophthalmology, University Clinic of Navarre. Pamplona (Navarre). Spain. 1 Graduate in Medicine. 2 Ph.D. in Medicine. 3 Optometrist optician. Correspondence: Henar Heras Mulero Clínica Universitaria de Navarra Departamento de Oftalmología C/. Pío XII, Pamplona Spain [email protected]
2 HERAS-MULERO H, et al. than that of the Goldmann tonometer. The measurement differs from 0.7 to 4.4 mmhg. In corneas with pathology, it is very difficult or even unacceptable to measure the intraocular pressure using the Pascal tonometer (Arch Soc Esp Oftalmol 2007; 82: ). Key words: Pascal, pneumotonometry, Goldmann tonometer, dynamic contour tonometry, intraocular pressure. Conclusión: Diversos autores han comparado el tonómetro de Pascal y el Goldmann coincidiendo todos en que el Pascal ofrece una presión intraocular mayor que el Goldmann con unas diferencias que van de 0,7 a 4,4 mmhg. En córneas patológicas es muy difícil realizar la medición y cuando ésto ocurre ésta suele ser informada como inaceptable. Palabras clave: Pascal, neumotonómetro, goldmann, tonómetro dinámico de contorno, presión intraocular. INTRODUCTION The Pascal dynamic contour tonometer is a digital contact tonometer that measures intraocular pressure between 5 and 200 mmhg with a precision of 0.2 mmhg (1-3). Together with the value of intraocular pressure, it notes the level of reliability and quality of the measurement with Q1 being reliable, Q2 and Q3 acceptable and Q4 and Q5 unacceptable, which requires repeating the measurement. The Pascal tonometer is independent of corneal thickness and biomechanical cornea properties such as elasticity, rigidity, level of corneal hydration or lamellar stroma configuration (1-3). The Goldmann tonometer is considered the golden standard to measure intraocular pressure. However, it presents a series of limitations and sources of error since the value it provides is not digital, heart beats make the semicircles change position, the amount of coloring varies the thickness of the semicircles, thus modifying the measurement and finally, corneal properties such as thickness and curvature can lead to a wrong measurement (1-3). The pneumatonometer (4) is an air pressure contact tonometer following the Mackay-Marg method. Compared to the Goldmann, it underestimates high pressures and overestimates low ones. This tonometer, as it is independent of cornea properties, is useful in edematous, irregular corneas or with leucomas. The purpose of this study was to analyze the reliability of the Pascal versus the Goldmann tonometer, the golden standard, and the pneumatonometer, which we know is not much influenced by the anatomical properties of the cornea. SUBJECTS, MATERIAL AND METHOD An observational clinical study was conducted which included 205 consecutive and randomized patients visiting the ophthalmology department. Of these 205 patients, 24 had glaucoma, 25 a cornea transplant and 156 patients had neither. Measurements were always carried out by the same person, only one eye per patient was included, and both the order of the tonometers and the eye selected for examination were random. Three measurements were conducted for the pneumatonometer (Model 30 Classic Pneumatonometer, Medtronic Ophthalmics, Jacksonville, USA) and the Goldmann (Applanation Tonometer AT 900 BQ, Haag-Streit, Köniz, Switzerland) noting down the mean. With Pascal (Pascal, SMT Swiss Microtechnology AG, CH-2562 Port, Switzerland) three measurements were carried out, choosing whichever was equal or lower than Q3. For those patients with whom an acceptable measurement level was not achieved after three measurements, the pressure provided with the first Pascal measurement was noted, together with its level of reliability. Analysis of methods was conducted by a nonparametric test using the Spearman coefficient for correlation and the Wilcoxon test to compare paired samples. Consistency between tonometers was achieved through the Bland-Altman method. We assessed the median and the interquantile range (25-75%). The statistical suite used was SPSS 11.0 (SPSS for Windows, SPSS Inc, Chicago, USA). 338 ARCH SOC ESP OFTALMOL 2007; 82:
3 Pascal Tonometer RESULTS The pneumatonometer offered measurements in 200 of the 205 patients, while the Goldmann tonometer did so in 198 patients and the Pascal tonometer was only capable of conducting acceptable measurements in 190 of the 205 patients (Q1 to Q3). Pascal provided a higher intraocular pressure measurement than that obtained with the other tonometers. When the measurements obtained with Pascal were compared to those of the other two tonometers, there were significant differences (p<0.001), the results of Pascal being different from those obtained with Goldmann or the pneumatonometer (table I). When the level of reliability was analyzed with the Pascal measurement, we found that in 27.3% (56 out of 205) of the patients, intraocular pressure figures were reported as reliable, in 42% (86 out of 205) acceptable and in 23.4% (48 out of 205) unacceptable, not reliable, and a new measurement was required; when a new measurement was conducted, a more reliable value was not obtained. In the remaining 7.3% (15 out of 205) it was not possible to obtain any measurements with the Pascal tonometer. When comparing the Pascal tonometer to the Goldmann, we saw the Spearman correlation coefficient was weak, only 0.48 (p=0.01) (fig. 1a) and inconsistent with the Bland-Altman method (6.42%) (fig. 1b). When Pascal was compared to the pneumatonometer, results were similar with a weak Spearman coefficient of 0.44 (p=0.01) (fig. 2a) and variance of 6.25% (fig. 2b). This might lead us to believe the results are due to the high percentage (23.4%, 48 out of 205 patients) of patients with low reliability with the Pascal tonometer, however, when these patients were extracted and the measurements reported with Pascal as acceptable were compared to the other two tonometers, Fig. 1a: Linear regression and correlation coefficient for Pascal and Goldmann. Fig. 1b: Bland-Altman for Pascal and Goldmann with a discordance coefficient of 6.42%. Table I. Summary table of data obtained with tonometers Pascal, Goldmann and pneumatonometer Tonometer Cases Median IR (25-75) p Wilcoxon* Pascal Goldmann <0.001 Pneumatonometer <0.001 * Comparisons conducted with Pascal versus Goldmann and Pascal versus pneumatonometer. Fig. 2a: Linear regression and correlation coefficient for Pascal and pneumatonometer. results did not vary much. So that the Spearman coefficient of Pascal versus that of Goldmann was 0.38 (p=0.01) (fig. 3a) and 0.47 (p=0.01) with the pneumatonometer (fig. 3b) and discordance between Pascal versus Goldmann and the pneumatonometer was significant, 5.88% in both cases (fig. 4). ARCH SOC ESP OFTALMOL 2007; 82:
4 HERAS-MULERO H, et al. Fig. 2b: Bland-Altman for Pascal and pneumatonometer with a discordance coefficient of 6.25%. Fig. 4a: Bland-Altman for Pascal and Goldmann with a discordance coefficient of 5.88%, only for values measured as acceptable for Pascal. Fig. 3a: Linear regression and correlation coefficient for Pascal and Goldmann only for values measured as acceptable for Pascal. Fig. 4b: Bland-Altman for Pascal and pneumatonometer with a variance coefficient of 5.88% only for values measured as acceptable for Pascal. Fig. 3b: Linear regression and correlation coefficient for Pascal and pneumatonometer only for values measured as acceptable for Pascal. DISCUSSION Several authors (1-3,5-7) have compared the Pascal tonometer to the Goldmann one and all agree that the Pascal provides greater intraocular pressure than Goldmann. Kotecha (2) has found the least differences, only an increase in favor of Pascal of 0.7 mmhg, compared to our study, which has found that Pascal overestimates intraocular pressure by 4.4 mmhg when compared to Goldmann. When we analyzed the existing correlation between Pascal and Goldmann we found very disparate data in literature, ranging from the weakest correlation by Kotecha (2) with a coefficient of 0.22, and the present study (spearman r=0.48 p=0.01) up to reaching the almost perfect correlation by Kamppeter (1) of Doyle (3) analyzed the existing relation between the Goldmann and the Pascal tonometers in various corneal thicknesses, noting that before a normal or thick cornea there were no significant differences in the intraocular pressure measurements, while when the cornea was thin there were differences (p=0.009) in the intraocular pressure measured with Goldmann and Pascal. 340 ARCH SOC ESP OFTALMOL 2007; 82:
5 Pascal Tonometer In literature there is only one paper comparing Pascal, Goldmann and the pneumatonometer and it was published by Kniestedt (4) in As all other authors have found, Pascal provided an average intraocular pressure of 18.3 mmhg, higher than that obtained with the pneumatonometer (17.1 mmhg) and Goldmann (16 mmhg). Also, he analyzed the correlation of each one of the three tonometers with the corneal thickness, finding there was only correlation with the Goldmann tonometer (r=0.24 p<0.01), the pneumatonometer (r=17.1 p=0.1) and the Pascal tonometer (r<0.1 p=0.80) being independent of corneal thickness. It has struck us that Pascal conducts, in many cases, not very reliable measurements or that it is not able to measure intraocular pressure. To conduct the measurement, this tonometer requires a round tear film contour not found in patients with lachrymal dryness, high astigmatism or keratoplasty. Although this tonometer can be very useful in healthy eyes, in patients with glaucoma under chronic treatment or elderly patients with ocular dryness, this low reliability can be clinically significant. Although Pascal has become popular as a system to measure intraocular pressure following LASIK, its use should be assessed with all type of patients as a tonometer not affected by corneal thickness. In this study, as in all other literature, the Pascal tonometer overestimates intraocular pressure. Measurements were not reliable in 23.4% (48 out of 205) of patients, and it was impossible to conduct measurements in 7.3% (15 out of 205). BIBLIOGRAFÍA 1. Kamppeter BA, Jonas JB. Dynamic contour tonometry for intraocular pressure measurement. Am J Ophthalmol 2005; 140: Kotecha A, White ET, Shewry JM, Garway-Heath DF. The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry. Br J Ophthalmol 2005; 89: Doyle A, Lachkar Y. Comparison of dynamic contour tonometry with goldmann applanation tonometry over a wide range of central corneal thickness. J Glaucoma 2005; 14: Kniestedt C, Lin S, Choe J, Bostrom A, Nee M, Stamper RL. Clinical comparison of contour and applanation tonometry and their relationship to pachymetry. Arch Ophthalmol 2005; 123: Schneider E, Grehn F. Intraocular pressure measurementcomparison of dynamic contour tonometry and goldmann applanation tonometry. J Glaucoma 2006; 15: Kaufmann C, Bachmann LM, Thiel MA. Comparison of dynamic contour tonometry with goldmann applanation tonometry. Invest Ophthalmol Vis Sci 2004; 45: Pache M, Wilmsmeyer S, Lautebach S, Funk J. Dynamic contour tonometry versus goldmann applanation tonometry: a comparative study. Graefes Arch Exp Ophthalmol 2005; 243: ARCH SOC ESP OFTALMOL 2007; 82:
How do you adjust for IOP in a LASIK patient?
How do you adjust for IOP in a LASIK patient? Sunita Radhakrishnan, M.D Glaucoma Center of San Francisco Glaucoma Research and Education Group Principle of applanation tonometry Imbert Fick law Applanation
Abstract. Email: [email protected]. Received 13 March 2014; revised 15 April 2014; accepted 6 May 2014
Open Journal of Ophthalmology, 2014, 4, 46-55 Published Online May 2014 in SciRes. http://www.scirp.org/journal/ojoph http://dx.doi.org/10.4236/ojoph.2014.42009 Dynamic Contour Tonometry, Tono-Pen XL,
Goldmann applanation tonometry (GAT) is the standard for
Intraocular Pressure Measurements Using Dynamic Contour Tonometry after Laser In Situ Keratomileusis Claude Kaufmann, 1 Lucas M. Bachmann, 2 and Michael A. Thiel 1 PURPOSE. Thinning of the corneal stroma
Make a more confident glaucoma risk assessment with Corneal Hysteresis, a superior predictor of glaucoma progression.
Make a more confident glaucoma risk assessment with Corneal Hysteresis, a superior predictor of glaucoma progression. Advancing Eye Care. American Innovation. Reichert Ocular Response Analyzer G3. Make
THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening
THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening GALILEI G4 Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software
Managing glaucoma is a. Glaucoma Management Edited by Kuldev Singh, MD, MPH, and Peter A. Netland, MD, PhD. Thickness and Hysteresis
Glaucoma Management Edited by Kuldev Singh, MD, MPH, and Peter A. Netland, MD, PhD Hysteresis: A Powerful Tool for Glaucoma Care Clinical data suggests that this measurement may be a valuable indicator
Underestimation of Intraocular Pressure in Eyes After Laser In Situ Keratomileusis
Underestimation of Intraocular Pressure in Eyes After Laser In Situ Keratomileusis Ako Arimoto*, Kimiya Shimizu, Nobuyuki Shoji, Kikuko Enomoto and Makio Kohara* *Department of Ophthalmology, Musashino
PROCEDIMIENTOPARALAGENERACIÓNDEMODELOS3DPARAMÉTRICOSA PARTIRDEMALLASOBTENIDASPORRELEVAMIENTOCONLÁSERESCÁNER
PROCEDIMIENTOPARALAGENERACIÓNDEMODELOS3DPARAMÉTRICOSA PARTIRDEMALLASOBTENIDASPORRELEVAMIENTOCONLÁSERESCÁNER Lopresti,LauraA.;Lara, Marianela;Gavino,Sergio;Fuertes,LauraL.;Defranco,GabrielH. UnidaddeInvestigación,DesaroloyTransferencia-GrupodeIngenieríaGráficaAplicada
LOCSU / WOPEC courses to support local service development
LOCSU / WOPEC courses to support local service development Excellence in Education for Eye Care Professionals LOCSU / WOPEC training courses are designed to support the delivery of enhanced eye care services
Using Biomechanics to Advance Glaucoma Care
1 FEBRUARY 2012 Using Biomechanics to Advance Glaucoma Care A panel of experts discusses how affordable new technology expands screening, diagnosis and monitoring of glaucoma. Sponsored by Ben Gaddie,
Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions
Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Document issued on: March 27, 2006 For questions regarding this document, contact Everette Beers, Ph.D. at 240-276-4200
Measure of Confidence. Glaucoma Module Premium Edition
Measure of Confidence Glaucoma Module Premium Edition The Changing Face of Glaucoma Practice Literature 1 Leske et al., Arch Ophthalmol 1997; 115:1051-1057 2 Doughty et al., Surv Ophthalmol 2000; 44:367-408
Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure
ORIGINAL ARTICLE Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure Joseph Frucht-Pery, MD; David Landau, MD; Frederik Raiskup, MD; Faik Orucov, MD;
Susana Sanduvete-Chaves, Salvador Chacón-Moscoso, Milagrosa Sánchez- Martín y José Antonio Pérez-Gil ( )
ACCIÓN PSICOLÓGICA, junio 2014, vol. 10, n. o 2, 3-20. ISSN: 1578-908X 19 THE REVISED OSTERLIND INDEX. A COMPARATIVE ANALYSIS IN CONTENT VALIDITY STUDIES 1 EL ÍNDICE DE OSTERLIND REVISADO. UN ANÁLISIS
Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery
Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery Farhad Hafezi, MD; Michael Mrochen, PhD; Franz Fankhauser II, MD; Theo Seiler, MD, PhD
Evaluation of the ICare Rebound Tonometer as a Home Intraocular Pressure Monitoring Device
ORIGINAL STUDY Evaluation of the ICare Rebound Tonometer as a Home Intraocular Pressure Monitoring Device Sanjay Asrani, MD,* Ashmita Chatterjee, BSc,w David K. Wallace, MD, MPH,* Cecilia Santiago-Turla,
New and Improved Femtosecond Laser Applications. Karl Stonecipher, MD Wavefront Congress 2008
New and Improved Femtosecond Laser Applications Karl Stonecipher, MD Wavefront Congress 2008 When birds don t fly, neither should you. When cows bunch together in a field, a storm is coming. When ants
UNDERSURFACE ABLATION OF THE CORNEAL FLAP FOR LASIK ENHANCEMENT
ARCH SOC ESP OFTALMOL 2007; 82: 697-704 ORIGINAL ARTICLE UNDERSURFACE ABLATION OF THE CORNEAL FLAP FOR LASIK ENHANCEMENT RETRATAMIENTOS EN CARA INTERNA DEL COLGAJO CORNEAL EN LASIK. ESTUDIO COMPARATIVO
Control of a variety of structures and idioms; occasional errors may occur, but
AP SPANISH LANGUAGE 2012 PRESENTATIONAL WRITING SCORING GUIDELINES SCORE DESCRIPTION TASK COMPLETION TOPIC DEVELOPMENT LANGUAGE USE 5 Demonstrates excellence 4 Demonstrates command 3 Demonstrates competence
Hospital San José, Bogotá, Colombia 1984-1985 Medical internship
BIOGRAPHICAL SKETCH NAME Salomon Esquenazi, M.D. era COMMONS USER NAME: sesque POSITION TITLE Research Assistant Professor of Ophthalmology and Neuroscience, Louisiana State University Health Sciences
How will this proposal affect clinical trials in Spain?
Barcelona, 16th October 2012 Cómo afectará esta propuesta p a los ensayos clínicos en España? How will this proposal affect clinical trials in Spain? Cristina Avendaño Solá Servicio de Farmacología Clínica
Comparison of Retinal Image Quality between SBK and PRK. Allen Boghossian, D.O. Durrie Vision Overland Park, KS
Comparison of Retinal Image Quality between and Allen Boghossian, D.O. Durrie Vision Overland Park, KS Study Design Purpose: To compare optical quality and intraocular scatter between and using double-pass
Tables of significant values of Jaccard's index of similarity
Miscel.lania Zooloqica 22.1 (1999) Tables of significant values of Jaccard's index of similarity R. Real Real, R., 1999. Tables of significant values of Jaccard's index of similarity. Misc. Zool., 22.1:
Ectasia Risk Scoring System (ERSS): Limitations and Variability in Predictive Value
Ectasia Risk Scoring System (ERSS): Limitations and Variability in Predictive Value Jay S. Pepose, M.D., Ph.D. Director, Pepose Vision Institute Professor of Clinical Ophthalmology Washington University
LINIO COLOMBIA. Starting-Up & Leading E-Commerce. www.linio.com.co. Luca Ranaldi, CEO. Pedro Freire, VP Marketing and Business Development
LINIO COLOMBIA Starting-Up & Leading E-Commerce Luca Ranaldi, CEO Pedro Freire, VP Marketing and Business Development 22 de Agosto 2013 www.linio.com.co QUÉ ES LINIO? Linio es la tienda online #1 en Colombia
We have witnessed the rapid development of
Experimental Results of Preparing Laser-shaped Stromal Implants for Laser-assisted Intrastromal Keratophakia in Extremely Complicated Laser in situ Keratomileusis Cases Mirko Jankov, MD; Michael C. Mrochen,
CORNEAL HYSTERESIS AND INTRAOCULAR PRESSURE MEASUREMENT
MEDICAL POLICY CORNEAL HYSTERESIS AND INTRAOCULAR PRESSURE MEASUREMENT Policy Number: CS026.D Effective Date: August 1, 2015 Table of Contents COVERAGE RATIONALE... APPLICABLE CODES.. DESCRIPTION OF SERVICES...
Resumen de Entrevista: Asociación de Agentes de Aduana del Puerto de Manzanillo
Resumen de Entrevista: Asociación de Agentes de Aduana del Puerto de Manzanillo 1. To your knowledge, to what extent do customs brokers run into operative inconveniences when it comes to collecting payment
CORNEAL HYSTERESIS AND INTRAOCULAR PRESSURE MEASUREMENT
CLINICAL POLICY CORNEAL HYSTERESIS AND INTRAOCULAR PRESSURE MEASUREMENT Policy Number: DIAGNOSTIC 048.10 T2 Effective Date: August 1, 2015 Table of Contents APPLICABLE LINES OF BUSINESS/PRODUCTS.. BENEFIT
BSM Connection elearning Course
BSM Connection elearning Course Scope of the Eye Care Practice 2008, BSM Consulting All Rights Reserved. Table of Contents OVERVIEW...1 THREE O S IN EYE CARE...1 ROUTINE VS. MEDICAL EXAMS...2 CONTACT LENSES/GLASSES...2
Memorial Health Care System Catholic Health Initiatives Financial Assistance Application Form
B Please note - Memorial Hospital may access external validation resources to assist in determining whether a full application for assistance is required. Financial Assistance Application 1) Patient Name
LOCSU Enhanced Services
LOCSU Enhanced Services Glaucoma Repeat Readings & OHT Monitoring Enhanced Service Pathway Issued by Local Optical Committee Support Unit May 2009 [Revised June 2012] Contents Page Executive Summary...
Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking
Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking Kay Lam, MD, Dan B. Rootman, MSc, Alejandro Lichtinger, and David S. Rootman, MD, FRCSC Author affiliations:
A comparative study of two models for the seismic analysis of buildings
INGENIERÍA E INVESTIGACIÓN VOL. No., DECEMBER 0 (-) A comparative study of two models for the seismic analysis of buildings Estudio comparativo de dos modelos para análisis sísmico de edificios A. Luévanos
Total Corneal Power Estimation: Ray Tracing Method versus Gaussian Optics Formula PATIENTS AND METHODS
Cornea Total Corneal Power Estimation: Ray Tracing Method versus Gaussian Optics Formula Li Wang, 1 Ashraf M. Mahmoud, 2 Betty Lise Anderson, 3 Douglas D. Koch, 1 and Cynthia J. Roberts 2 PURPOSE. To evaluate
LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth?
JKAU: Med. Sci., Vol. 18 No. 4, pp: 29-36 (2011 A.D. / 1432 A.H.) DOI: 10.4197/Med. 18-4.3 LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth? Ali M. El-Ghatit, MD,
Refractive errors are caused by an imperfectly shaped eyeball, cornea or lens, and are of three basic types:
Tips on Lasik Eye Surgery If you re tired of wearing glasses or contact lenses, you may be considering Lasik eye surgery one of the newest procedures to correct vision problems. Before you sign up for
Slit-lamp handling: Learning upgrade with webcam video recordings
ÓPTICA PURA Y APLICADA. www.sedoptica.es Type: Educational Paper / Tipo: Artículo Educativo Section: Optics Education / Sección: Enseñannza de la Óptica Slit-lamp handling: Learning upgrade with webcam
DIFFERENTIATIONS OF OBJECTS IN DIFFUSE DATABASES DIFERENCIACIONES DE OBJETOS EN BASES DE DATOS DIFUSAS
Recibido: 01 de agosto de 2012 Aceptado: 23 de octubre de 2012 DIFFERENTIATIONS OF OBJECTS IN DIFFUSE DATABASES DIFERENCIACIONES DE OBJETOS EN BASES DE DATOS DIFUSAS PhD. Amaury Caballero*; PhD. Gabriel
Ectasia after laser in-situ keratomileusis (LASIK)
Ectasia after laser in-situ keratomileusis (LASIK) 長 庚 紀 念 醫 院 眼 科 蕭 靜 熹 Post-LASIK ectasia A rare complication of LASIK Manhattan jury awarded a former investment banker a record $7.25 million for post-lasik
Verbos modales. In this class we look at modal verbs, which can be a tricky feature of English grammar.
Verbos modales In this class we look at modal verbs, which can be a tricky feature of English grammar. We use Modal verbs in English to show: Probability,Possibility, Capability, Permission, ObligaCon,
Central Islands After LASIK Detected by Corneal Topography
Korean J Ophthalmol Vol. 15:8-14, 2001 Central Islands After LASIK Detected by Corneal Topography Jin Seok Lee, MD, Choun-Ki Joo, MD Department of Ophthalmology, Kangnam St. Mary s Hospital, College of
Descemet s Stripping Endothelial Keratoplasty (DSEK)
Descemet s Stripping Endothelial Keratoplasty (DSEK) Your doctor has decided that you will benefit from a corneal transplant operation. This handout will explain your options to you. It explains the differences
A New Extension of the Exponential Distribution
Revista Colombiana de Estadística Junio 2014, volumen 37, no. 1, pp. 25 a 34 A New Extension of the Exponential Distribution Una nueva extensión de la distribución exponencial Yolanda M. Gómez 1,a, Heleno
Overview of Refractive Surgery
Overview of Refractive Surgery Michael N. Wiggins, MD Assistant Professor, College of Health Related Professions and College of Medicine, Department of Ophthalmology Jones Eye Institute University of Arkansas
Five-Year Outcome of LASIK for Myopia
Five-Year Outcome of LASIK for Myopia Naoko Kato, MD, PhD 1,2 Ikuko Toda, MD, PhD, 1,2 Yoshiko Hori-Komai, MD, PhD, 1,2 Chikako Sakai, BS, 1 Kazuo Tsubota, MD, PhD 2 Purpose: To investigate the efficacy
Descemet s Stripping Automated Endothelial Keratoplasty (DSAEK)
Descemet s Stripping Automated Endothelial Keratoplasty (DSAEK) John D. Goosey, MD Introduction DSAEK is a corneal transplant technique where the unhealthy, diseased, posterior portion of a patient s cornea
Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps
Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps Lingo Y. Lai, MD William G. Zeh, MD Clark L. Springs, MD The authors
Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes
Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes Xiangjun Chen 1,2 *, Aleksandar Stojanovic 1,2,3, Yanjun Hua 4,5, Jon Roger Eidet 6,DiHu 5, Jingting Wang 5,
El modelo estratégico de los canales de. contacto con el cliente dentro de los objetivos. de calidad y rentabilidad
El modelo estratégico de los canales de Click to edit Master text styles contacto con el cliente dentro de los objetivos Second level de calidad y rentabilidad Third level Fourth level Fifth level presentado
Oregon Eye Specialists, PC YOUR GUIDE CATARACT SURGERY. Improving VISION. Improving LIFE.
Oregon Eye Specialists, PC YOUR GUIDE TO CATARACT SURGERY Dinelli M. Monson, M.D. Comprehensive Ophthalmology Physician and Surgeon Tualatin Clinic: 19250 SW 65 th Ave, Ste 215 503.692.3630 Newberg Clinic:
GENERAL CODING INFORMATION. Use of Modifiers for Identifying Specific Eye
An Independent Licensee of the Blue Cross and Blue Shield Association. APPENDIX G OPTOMETRY & OPTICIAN GUIDELINES Acknowledgement: Current Procedural Terminology (CPT ) is copyright 2014 American Medical
Fecha: 04/12/2010. New standard treatment for breast cancer at early stages established
New standard treatment for breast cancer at early stages established London December 04, 2010 12:01:13 AM IST Spanish Oncology has established a new standard treatment for breast cancer at early stages,
INTELIGENCIA DE NEGOCIO CON SQL SERVER
INTELIGENCIA DE NEGOCIO CON SQL SERVER Este curso de Microsoft e-learning está orientado a preparar a los alumnos en el desarrollo de soluciones de Business Intelligence con SQL Server. El curso consta
Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD
Post LASIK Ectasia Gina M. Rogers, MD and Kenneth M. Goins, MD October 6, 2012 Chief Complaint: Decreasing vision after laser- assisted in- situ keratomileusis (LASIK) History of Present Illness: This
Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State
Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State University Columbus, Ohio Refractive Errors Myopia (nearsightedness)
ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography
ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography GALILEI G6 Clinical Applications High confidence premium IOL selection The GALILEI G6 offers a link to the ray-tracing software
Common Co-management Questions
Issue 037 efocus Innovation. Leadership. Passion for Perfection 415.922.9500 --- www.pacificvision.org Common Co-management Questions Top questions recently asked by optometrists co-managing refractive
Mathematical Model to Compare the Relative Tensile Strength of the Cornea After PRK, LASIK, and Small Incision Lenticule Extraction
ORIGINAL ARTICLE Mathematical Model to Compare the Relative Tensile Strength of the Cornea After PRK, LASIK, and Small Incision Lenticule Extraction Dan Z. Reinstein, MD, MA(Cantab), FRCOphth; Timothy
S canning laser polarimetry (SLP) of the retinal nerve fibre
627 SCIENTIFIC CORRESPONDENCE Influence of post-lasik corneal healing on scanning laser polarimetric measurement of the retinal nerve fibre layer thickness G Holló, Z Z Nagy, P Vargha, I Süveges... Br
INFORMED CONSENT TO HAVE LASIK
A Division of Scott & Christie and Associates INFORMED CONSENT TO HAVE LASIK This information is to help you make an informed decision about having Laser Assisted Intrastromal Keratomileusis (LASIK), an
Comparison to other anterior segment imaging devices
found helpful in a number of areas of ophthalmology. However, in the field of anterior segment tumors its usefulness is confined to a subset of tumors that are confined to the iris, hypo pigmented, and
Reduction in Intraocular Pressure after Cataract Extraction: The Ocular Hypertension Treatment Study
Reduction in Intraocular Pressure after Cataract Extraction: The Ocular Hypertension Treatment Study Steven L. Mansberger, MD, MPH, 1 Mae O. Gordon, PhD, 2 Henry Jampel, MD, MHS, 3 Anjali Bhorade, MD,
5/24/2013 ESOIRS 2013. Moderator: Alaa Ghaith, MD. Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD
ESOIRS 2013 Moderator: Alaa Ghaith, MD Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD 1 A systematic approach to the management of Keratoconus through the presentation of different
Scanning laser polarimetry and retinal thickness analysis before and after laser in situ k e r a t o m i l e u s i s
European Journal of Ophthalmology / Vol. 15 no. 4, 2005 / pp. 434-440 Scanning laser polarimetry and retinal thickness analysis before and after laser in situ k e r a t o m i l e u s i s E.M. HOFFMANN,
The «three-in-one» Microkeratome for traditional LASIK, EPI-LASIK, and anterior and posterior Lamellar Keratoplasty
The «three-in-one» Microkeratome for traditional LASIK, EPI-LASIK, and anterior and posterior Lamellar Keratoplasty Ziemer Group We develop and produce diagnostic and surgical products for the ophthalmic
Keratoconus and Riboflavin UV-Crosslinking of the Cornea
Keratoconus and Riboflavin UV-Crosslinking of the Cornea Institut für Refraktive und Ophthalmo-Chirurgie Dear patient, A keratoconus has been diagnosed on your cornea. In this disease, which affects 1
INFORMATIONAL NOTICE
Rod R. Blagojevich, Governor Barry S. Maram, Director 201 South Grand Avenue East Telephone: (217) 782-3303 Springfield, Illinois 62763-0002 TTY: (800) 526-5812 DATE: March 4, 2008 INFORMATIONAL NOTICE
Medicare and Corneal Surgery: Cosmetic versus Functional
Medicare and Corneal Surgery: Cosmetic versus Functional Riva Lee Asbell INTRODUCTION With the introduction of several new CPT (Current Procedural Terminology) codes for cornea, corneal coding is in the
ReLEx smile Minimally invasive vision correction Information for patients
ReLEx smile Minimally invasive vision correction Information for patients Seeing is living Our eyes are our most important sensory organ. The human brain obtains over 80 % of its information via the sense
PARALLEL TO THE DEVELOPMENTS OF SURGICAL
Comparison of Three Methods of Measuring Corneal Thickness and Anterior Chamber Depth WOLF BUEHL, MD, DANIJELA STOJANAC, MD, STEFAN SACU, MD, WOLFGANG DREXLER, MD, AND OLIVER FINDL, MD Accepted for publication
