Program Requirements
|
|
- Eric Leonard
- 7 years ago
- Views:
Transcription
1 METCAP Program Requirements METrology and Calibration Accreditation Program ANSI/NCSL Z subclause 5.3 Accreditation Bureau (L-A-B) - All Rights Reserved. August 31, No part of this document may be reproduced or utilized, in any form, or by any means electronic or mechanical, including photocopy or recording, or by any information storage and retrieval system, without written permission from the publisher, the Accreditation Bureau (L-A-B). LABPR Accreditation Bureau Coldwater Rd. Suite 101 Fort Wayne, IN Phone: (260) Fax: (260) LABPR Rev 3 5/27/11 Page 1 of 27
2 1. Introduction 1.1. The purpose of this document is to establish specific program requirements as defined by the Accreditation Bureau, hereafter known as L-A-B. This document is written to help laboratories develop and implement a laboratory management system that is in compliance with ANSI/NCSLI Z , Subclause 5.3 that is suitable for their size and workload, and that will meet their clients needs Laboratories that are contractually bound to control their measuring and test equipment or that of their customers, the requirements of this standard is applicable. All requirements of the standard can be superseded by the requests of the customer ANSI/NCSLI Z Subclause 5.3 Requirements for the calibration of measuring and test equipment Checklist sets out the criteria for laboratories wishing to demonstrate that they are technically competent, and are able to generate technically valid calibration and test results. The standard will form the basis for the accreditation of competence of laboratories by accreditation bodies. 2. L-A-B Specific Requirements 2.1. Applications Laboratories applying for the accreditation of ANSI/NCSL Z540.3 subclause 5.3 will be required to apply the requirements of subclause 5.3 to their own Measuring and Test Equipment (M&TE) and calculate the Probability of False Acceptance (PFA) for their own M&TE Laboratories that calibrate equipment internally and are not accredited to calibrate those particular items, the laboratories will be required to calculate PFA for those supporting calibrations Documents to be submitted At the time of the notification of the assessment, documentation will be sent to the laboratory describing what is needed for the review prior to the assessment including: Annex A and B of this document Procedures for the determination of Measuring and Test Equipment calibration requirements Procedure for Measurement Assurance L-A-B Form 001 Traceability Tracking Procedure for the estimation and expression of the uncertainty of measurement Uncertainty analysis Proposed scope of accreditation. Scope of Accreditation examples can be found Annex B of L-A-B AG s procedure(s) for calculating PFA Examples of the PFA calculations Procedure for establishing and adjusting calibration intervals Example of a calibration certificate that is issued to the customer complying with the requirements of ISO/IEC 17025, L-A-B Policy s and PR requirements Name(s) of personnel who are responsible for maintaining compliance to these requirements. LABPR Rev 3 5/27/11 Page 2 of 27
3 3. References 3.1. ANSI/NCSL Z Requirements for the calibration of measuring and test equipment 3.2. Handbook for the Application of ANSI/NCSL Z Requirements for the calibration of measuring and test equipment 3.3. ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories 3.4. International Vocabulary of Basic and General Terms in Metrology (VIM) 4. Definitions 4.1. Calibration - The set of operations that establish, under specified conditions, the relationship between values of quantities indicated by a measuring instrument or measuring system, or values represented by a material measure or a reference material, and the corresponding values realized by standards Calibration system, calibration program - The set of interrelated or interacting elements necessary to maintain the measurement performance of measuring and test equipment to defined requirements Guard band - The offset of a measurement decision value from a stated specification or tolerance Measurement assurance - The result of a process to provide adequate confidence that a measurement will satisfy stated requirements Measurement decision risk - The probability that an incorrect decision will result from a measurement Measurement process - The set of operations to determine the value of a measurement quantity Measurement quantity - An attribute of a phenomenon, artifact, or substance that may be distinguished qualitatively and determined quantitatively Measurement reliability - The probability that all the applicable measurement quantities of measuring and test equipment are within tolerance Measuring and test equipment - The measuring instrument, measurement standard, reference material or auxiliary apparatus, or a combination thereof, necessary to realize a measurement process M&TE- Abbreviation for Measuring and Test Equipment Organization - A group of people and facilities with an arrangement of responsibilities, authorities, and relationships PFA Abbreviation for Probability of False Acceptance Test uncertainty ratio - The ratio of the span of the tolerance of a measurement quantity subject to calibration, to twice the 95% expanded uncertainty of the measurement process used for calibration Tolerance - Extreme values of an error permitted by specifications, regulations, etc., for a given measuring instrument, test, or measurement application Validation - Confirmation, through the provision of objective evidence, that the requirements for a specific intended use or application have been fulfilled. General Requirements above and beyond ISO/IEC LABPR Rev 3 5/27/11 Page 3 of 27
4 5. Requirements of ANSI/NCSL Z540.3 subclause Calibration of measuring and test equipment - Calibration of measuring and test equipment shall be in accordance with the requirements of this National Standard. Calibration may be performed within or outside a designated calibration facility, e.g., in situ, on-site, or at a customer s facility, provided compliance with the requirements of this National Standard is maintained. The scope of the calibration capability shall be consistent with the calibration requirements and provide levels of measurement decision risk acceptable to both the customer and supplier. 5.3 a) Where calibrations provide for reporting measured values, the measurement uncertainty shall be acceptable to the customer and shall be documented. 5.3 b) Where calibrations provide for verification that measurement quantities are within specified tolerances, the probability that incorrect acceptance decisions (false accept) will result from calibration tests shall not exceed 2% and shall be documented. Where it is not practicable to estimate this probability, the test uncertainty ratio shall be equal to or greater than 4:1. Calibration-servicing components may be considered competent to provide calibration services when they have been accredited to meet ANS/ISO/IEC 17025, including the requirements of this sub-clause, or otherwise found to be in conformance by an authority acceptable to the customer. Competence shall be confirmed for the required calibration services and reflected in a scope of accreditation or in a similar listing of calibration capability and conformity in the non-accredited laboratory. All measuring and test equipment included in the calibration system, including measuring systems, calibration equipment, reference standards and material, and other inspection and monitoring equipment, shall be calibrated prior to use and recalibrated at predetermined intervals to ensure acceptable measurement uncertainty, traceability, and reliability. Intervals may be based on usage or time since last calibration. When there is doubt as to the suitability of an item for calibration, when an item does not conform to the description provided, or the calibration required is not specified in sufficient detail, the calibration-servicing component shall consult with the customer for further instructions before proceeding and shall record the discussion Calibration procedures - Calibrations shall be performed using calibration procedures that: address the measuring and test equipment performance requirements; are acceptable to the customer; are current and appropriate for the calibrations; and provide reasonable assurance that the calibration results are as described. All calibration procedures shall: a) contain sufficient information on requirements for the associated measurements and instructions to perform the calibrations. In addition, the number of different measurement quantities and values in a calibration procedure shall be sufficient to ensure conformity of the measuring and test equipment to determined requirements; b) provide for determining and recording the as-found performance of the measuring and test equipment being calibrated; and c) be suitable for use by the calibration staff to perform and reproduce a calibration to the established performance criteria. Deviation from calibration procedures shall occur only if the deviation has been documented, technically justified, authorized, and accepted by the customer. LABPR Rev 3 5/27/11 Page 4 of 27
5 When a calibration procedure proposed by the customer is considered to be inappropriate, the customer shall be notified Contents - calibration procedures shall include the following information: a) identification and document control information; b) scope and/or description of item to be calibrated; c) measurement quantities and ranges to be determined for the item to be calibrated and any associated tolerances; d) minimum performance requirements of the equipment to be used for calibration, including measurement and reference standards, and reference materials; e) environmental conditions required and any stabilization period needed; f) description of steps associated with the calibrations to be performed; g) criteria and/or requirements for calibration decisions, such as approval or rejection; and h) data to be recorded and method of analysis and presentation Validation - Calibration procedures and their modifications, shall be validated. The validation shall be as extensive as is necessary to meet the needs of the procedure s application Measurement assurance procedures - Measuring processes incorporating measurement assurance methods, such as statistical process control, shall use a measurement assurance procedure. This procedure shall be systematically applied and include stated measurement uncertainty or reliability goals, control criteria, and methodology to verify that the goals and criteria are being attained. The controls shall be adjusted when the results of the previous measurements indicate that such action is appropriate to maintain acceptable measurement uncertainty or reliability. Measurement assurance controls may be based on the use of calibrated check standards, usage, and/or time since the last performance. The measurement assurance procedure shall include mandatory instructions to preclude the use of the measuring process that exceeds its controls. The measurement assurance procedure and any associated measuring and test equipment shall be documented as a calibration procedure in accordance with the provisions of this National Standard Measurement uncertainty and traceability - The uncertainty and traceability of all measurement results associated with processes included in the calibration system shall meet the requirements of their applications. Measurement uncertainty components which have an influence on such measurement results shall be included in the estimates of measurement uncertainty Expression of measurement uncertainty - A documented procedure shall be used to estimate and express the uncertainty of measurement for all calibrations. As a minimum, the procedure shall address: a) sources of measurement uncertainty; b) estimation and combining of uncertainties; c) conditions and assumptions; d) documentation and reporting criteria; and e) bibliography. LABPR Rev 3 5/27/11 Page 5 of 27
6 Measurement traceability- The results of a calibration or measurement shall be traceable through a controlled, unbroken chain of competent calibrations to and through the National Institute of Standards and Technology to the SI units of measurement. This traceability to a national measurement institute other than the National Institute of Standards and Technology is acceptable when: a) a mutual recognition agreement, such as the Comité International des Poids et Mesures (CIPM) Mutual Recognition Arrangement (MRA), is in effect with the National Institute of Standards and Technology and sufficient equivalence of applicable calibration services exists; or b) when the calibration service of the National Institute of Standards and Technology is not available or does not meet the measurement performance requirements. Where traceability to SI units through national metrology institutes is not available, or SI units are not established, a consensus standard including a reference standard and related calibration procedures, which are clearly specified and mutually agreed upon by all parties concerned, shall be applied Calibration equipment - All measuring and test equipment required for the correct performance of calibrations and related measurements, including calibration and reference standards and reference material, shall be available to the calibrationservicing component. In those cases where the calibration-servicing component needs to use equipment outside its permanent control, it shall ensure that the requirements of this National Standard are met. Measuring and test equipment used for calibration and associated measurements shall be capable of achieving the performance required and shall comply with specifications relevant to the calibrations performed. Measuring and test equipment that may affect the results of the calibrations shall be calibrated and included in a calibration system meeting the requirements of this National Standard Documentation- Instructions on the use and operation of all relevant equipment shall be available to the calibration-servicing component where the absence of such instructions could jeopardize the results of calibrations Calibration personnel - Personnel who operate specific equipment, perform calibrations or measurements, evaluate results, give opinions and interpretations, prepare calibration records, and/or sign calibration reports shall be competent. When using personnel who are undergoing training, appropriate supervision shall be provided Authorization - Specific personnel shall be authorized: to perform particular types of calibrations; to issue calibration procedures and reports; to give opinions and interpretations; to prepare calibration records; and to operate particular types of equipment. Records of such authorization shall be maintained, readily available, and include the date of authorization Influence factors and conditions - All factors and conditions of the calibration area that adversely influence the calibration results shall be defined, monitored, recorded, and mitigated to meet calibration process requirements Calibration quality - Calibration quality shall be monitored to assure the validity of the calibration services. The monitoring process shall be documented and the resulting data recorded and evaluated. This monitoring process may include, but not be limited to, the following: LABPR Rev 3 5/27/11 Page 6 of 27
7 a) sampling of calibrated measuring and test equipment; b) calibrations using different procedures or methods; c) regular use of measurement assurance methods; and/or d) participation in measurement inter-comparison programs. When the results of any of the procedures indicate suspect calibrations, appropriate actions shall be taken to identify the root cause and take any corrective and preventive actions Reporting the results - The results of each calibration or series of calibrations shall be reported accurately, clearly, unambiguously, and objectively. The results reported shall include all information requested by the customer and necessary for the interpretation of the calibration results, and all information required by the calibration procedure used Calibration reporting - Where a calibration report is issued, the report shall include at least the following information: a) title; b) identification of the calibration-servicing component, and the location where the calibrations were carried out, if different; c) unique identification of the calibration report; d) identification of the customer; e) identification of the calibration procedure used; f) description of, the condition of, and unambiguous identification of the item calibrated; g) date of performance of the calibration; h) calibration results with, where appropriate, the units of measurement; i) calibration results before and after adjustment or repair, if available; j) person authorizing the calibration report; k) statement of measurement traceability; l) factors and conditions under which the calibrations were made that have an influence on the measurement results; and m) uncertainty of measurement, where appropriate, and/or a statement of conformance. Where a statement of conformance with specified calibration performance requirements and/or calibration procedure is provided, the extent of conformance shall be provided and included in the calibration records Opinions and interpretations - When opinions and interpretations are included, the basis upon which the opinions and interpretations have been made shall be documented. Opinions and interpretations shall be clearly marked as such in a calibration report Control and source - When it is necessary to amend or issue a complete new calibration report, it shall be uniquely identified and shall contain a reference to the original that it amends or replaces Calibration records - Records of the calibration process shall be completed to attest to the correctness of the results, as appropriate. These records shall be maintained and made available. Records of the calibration process shall demonstrate whether LABPR Rev 3 5/27/11 Page 7 of 27
8 measuring and test equipment have satisfied the performance requirements specified. The records should be sufficient to demonstrate the traceability of the measurements and reproduce the calibration results under conditions close to the original conditions. Records shall be maintained of each item of equipment and software where its use contributes to the results of the calibrations performed. The records shall include the following: a) description and unique identification of the measuring and test equipment (manufacturer, model, nomenclature, serial number, etc.); b) identification of the calibration-servicing component; c) identification of the calibration procedure used; d) date on which the calibration was completed; e) as-found measurement performance condition of the equipment; f) calibration actions taken (adjusted, repaired, new value assigned, limited, derated, modified, etc.); g) calibration results and the as-returned measurement performance condition of the equipment; h) limitations of use; i) assigned calibration interval; and j) identification of calibration reports. Generation, amendment, issuance, and deletion of records shall be authorized. In addition, the reason for an amendment or deletion of a record shall be documented. REVISION HISTORY Revision Revision Level Date Revised By Brief Description of Revision Original 08/31/09 Randy Long Rev 1 11/30/10 Ryan Fischer Original Issue, requirements section numbering matching that of Z540.3 Further identified the documents needed prior to the assessment in Approved: Date: 12/13/2010 LABPR Rev 3 5/27/11 Page 8 of 27
9 Annex A Documentation Checklist Name: Number of Lab Techs: Instructions: This documentation checklist and the documentation stated below are required to be sent to L-A-B at least 30 days prior to the scheduled Full Assessment. L-A-B does not maintain any laboratory quality documentation on file. Any delays in receiving the necessary documentation may result in a rescheduling of the assessment per L-A-B Policy 004. Electronic documentation must be sent to laboratorydocuments@l-a-b.com. All remaining hard copy documents must be sent to L-A-B operations office. Documentation Requirements for an Assessment Date Sent to L-A-B Electronic Hard Copy Proof of Purchase of ANSI/NCSL Z540.3 (Initial Assessment Only) Procedures for the determination of Measuring and Test Equipment calibration requirements. Procedure for Measurement Assurance L-A-B Form 001 Traceability Tracking Please provide this form which details the traceability of the calibrated items in the lab. Procedure for the estimation and expression of the uncertainty of measurement Proposed scope of accreditation. Scope of Accreditation example can be found Annex B of L-A-B AG s procedure(s) for calculating PFA Examples of the PFA calculations Procedure for establishing and adjusting calibration intervals Calibration Certificate or Test Report Example of a calibration certificate that is issued to the customer complying with the requirements of ISO/IEC 17025, L-A-B Policy s and PR requirements LABPR Rev 3 5/27/11 Page 9 of 27
10 Information Company Name Annex B Assessment Checklist Address (Include addresses of all sites covered by this assessment) Quality Manager Phone FAX Contact Form Completed By Date Form Completed Assessor Name Date of Assessment LABPR Rev 3 5/27/11 Page 10 of 27
11 Instructions Accreditation Bureau (L-A-B) and its applicants for laboratory accreditation will use this document to conduct an ANSI / NCSL Z , subclause 5.3 Quality System Review, and as an assessment checklist during the assessment process. The clauses of ISO/IEC are paraphrased in this document, and do not necessarily reflect the complete requirements. The laboratory is required to prove compliance with the full requirements as stated in ANSI / NCSL Z , subclause 5.3 in their entirety. Applicant The applicant shall complete the column Your Document by referencing its document that shows they are compliant with the specific requirement. This is the only column that is to be filled out by the applicant. This reference should indicate the document name, page and paragraph(s) in which the compliance can be confirmed. This checklist and all documentation referenced on the checklist will then be forwarded to L-A-B Operations office. Lead Assessor Document Review The lead assessor must confirm the documentation s compliance with the requirements of ANSI / NCSL Z , subclause 5.3, and indicate the laboratories compliance in the document compliance C and N columns of this form. The assessor should inform the client of any observations or requirements the laboratory may need to address prior to the assessment. Lead Assessor Pre-assessment Visit (Optional) The lead assessor will use this document to review each section and highlight areas which may be necessary for further review during the pre-assessment visit. Lead Assessor Assessment Visit: The lead assessor is responsible for completing this checklist by indicating the laboratory s compliance to its own policies and procedures, and the requirements of ANSI / NCSL Z , subclause 5.3. The laboratories compliance shall be indicated in the Assessment C and N columns, and comments on compliance, nonconformances, and observations shall be documented in the Assessment Notes column. Assessment Results Assessor shall complete Form 14 for Assessment results, and recommendations. LABPR Rev 3 5/27/11 Page 11 of 27
12 Assessment Summary LABPR Rev 3 5/27/11 Page 12 of 27
13 Comments on Calibration of Measuring and Test Equipment 5.3 Calibration or measuring and test equipment shall be IAW with ANSI/NCSL Z Calibration may be performed within or outside a designated calibration facility. The scope of the calibration activities shall be consistent with the calibration requirements and provide levels of measurement risk acceptable to customer and supplier. a) b) Where calibrations report measured values, the measurement uncertainty shall be documented and acceptable to the customer. Calibrations that provide for verification that measurement quantities are within specified tolerances, making statements of compliance: The probability that incorrect acceptance decisions will result from calibration tests shall not exceed 2% and shall be documented. If not practicable to estimate probability the test uncertainty ratio (TUR) shall be equal to or greater than 4:1 Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 13 of 27
14 Comments on Independent laboratories shall apply a 95 % Calibration Process Uncertainty Guard Band or TUR based Guard Band where reliability data of the UUT is not available Services provided to the organization supporting calibration services shall be considered competent when they have been at least: Accredited to Met the requirements of ANSI Z540.3 Found in conformance by an authority acceptable to the customer Competence for the calibration services shall be confirmed and reflected in a scope of accreditation or similar listing of calibration capability and conformity in the non-accredited laboratory. All measuring and test equipment included in the calibration system shall be calibrated prior to use and recalibrated at predetermined intervals. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 14 of 27
15 Comments on Where doubt to the suitability of an item for calibration or it does not conform to the description provided, or there is insufficient detail of the required services. The customer must be consulted and the discussion recorded. Calibration Procedures Calibration procedures shall assure: The measuring and test equipment performance requirements are addressed Are acceptable to the customer. Are current and appropriate for the calibrations. Provide reasonable assurance that the calibration results are as described. All calibration procedures shall: Contain sufficient information on requirements and instructions to perform the calibrations. a) Different measurement quantities and values in a calibration procedure shall ensure conformity of the measuring and test equipment to determined requirements. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 15 of 27
16 Comments on b) c) Include determining and recording the as-found performance of the M&TE being calibrated. Be suitable for use by the calibration staff to perform and reproduce a calibration to the established performance criteria. Deviation shall occur only if it has been documented, technically justified, authorized, and accepted by the customer. The customer shall be notified when a calibration procedure proposed by the customer is considered to be inappropriate Contents Calibration procedures shall include the following information: a) Identification and document control information. b) Scope and/or description of item to be calibrated. c) Measurement quantities and ranges to be determined for the item to be calibrated and any associated tolerances Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 16 of 27
17 Comments on d) e) f) Minimum performance requirements of the equipment to be used for calibration, including measurement and reference standards, and reference materials. Environmental conditions required and any stabilization period needed. Description of steps associated with the calibrations to be performed. Measurement Assurance Procedures A procedure shall be established for measuring processes incorporating measurement assurance methods Measurement uncertainty or reliability goals, control criteria, and methodology to verify that the goals and criteria are being attained shall be stated and systematically applied. Measurement assurance controls may be based on the use of calibrated check standards, usage, and/or time since the last performance. The controls shall be adjusted when the results of the previous measurements are needed to maintain acceptable measurement uncertainty or reliability. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 17 of 27
18 Comments on The measurement assurance procedure shall include mandatory instructions to preclude the use of the measuring process that exceeds its controls. The measurement assurance procedure and any associated measuring and test equipment shall be documented as a calibration procedure in accordance with the provisions of this National Standard Measurement Uncertainty and Traceability a) b) The uncertainty and traceability of all measurement results associated with processes included in the calibration system shall meet the requirements of their applications. Measurement uncertainty components which have an influence on such measurement results shall be included in the estimates of measurement uncertainty Expression of Measurement Uncertainty A documented procedure shall be used to estimate and express the uncertainty of measurement for all calibrations. The procedure shall address at a minimum: Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 18 of 27
19 Comments on a) Sources of measurement uncertainty b) Estimation and combining of uncertainties c) Conditions and assumptions d) Documentation and reporting criteria e) Bibliography Measurement Traceability The results of a calibration or measurement shall be traceable through a controlled, unbroken chain of competent calibrations to and through NIST to the SI units of measurement. This traceability to a national measurement institute other than the National Institute of Standards and Technology is acceptable when: a) A mutual recognition agreement is in effect with the NIST and sufficient equivalence of applicable calibration services exists. b) When the calibration service of the NIST is not available or does not meet the measurement performance requirements. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 19 of 27
20 Comments on Where traceability to SI units through national metrology institutes is not available, or SI units are not established, a consensus standard including a reference standard and related calibration procedures, which are clearly specified and mutually agreed upon by all parties concerned, shall be applied Calibration Equipment All M&TE required for the correct performance of calibrations and related measurements, including calibration and reference standards and reference material, shall be available to the calibration-servicing component. In those cases where the calibration-servicing component needs to use equipment outside its permanent control, it shall ensure that the requirements of this National Standard are met. Measuring and test equipment used for calibration and associated measurements shall be capable of achieving the performance required and shall comply with specifications relevant to the calibrations performed. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 20 of 27
21 Comments on Measuring and test equipment that may affect the results of the calibrations shall be calibrated and included in a calibration system meeting the requirements of this National Standard Documentation Instructions on the use and operation of all relevant equipment shall be available. Calibration Personnel Personnel who operate specific equipment, perform calibrations or measurements, evaluate results, give opinions and interpretations, prepare calibration records, and/or sign calibration reports shall be competent. Authorization Specific personnel shall be authorized: to perform particular types of calibrations, issue calibration procedures and reports, give opinions and interpretations, prepare calibration records and to operate particular types of equipment. Records of authorization shall be maintained, readily available, and include the date of authorization. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 21 of 27
22 Comments on Influence Factors and Conditions All factors and conditions of the calibration area that adversely influence the calibration results shall be defined, monitored, recorded, and mitigated to meet calibration process requirements. Calibration shall be stopped when the adverse effects of the influence factors and conditions jeopardize the results of the calibration. Calibration Quality a) b) c) Calibration quality shall be monitored to assure the validity of the calibration services. The monitoring shall be documented and the resulting data recorded and evaluated. This monitoring process may include, but not be limited to, the following: Sampling of calibrated measuring and test equipment. Calibrations using different procedures or methods. Regular use of measurement assurance methods. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 22 of 27
23 Comments on Participation in measurement inter-comparison programs. d) When the results of any of the procedures indicate suspect calibrations, appropriate actions shall be taken to identify the root cause and take any corrective and preventative actions. Reporting the results The results of each calibration or series of calibrations shall be reported accurately, clearly, unambiguously, and objectively. The results reported shall include all information requested by the customer and necessary for the interpretation of the calibration results, and all information required by the calibration procedure used. Calibration reporting Where a calibration report is issued, the report shall include at least the following information: a) Title Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 23 of 27
24 Comments on b) Identification of the calibration-servicing component, and the location where the calibrations were carried out, if different c) Unique identification of the calibration report d) Identification of the customer e) Identification of the calibration procedure used f) Description of, the condition of, and unambiguous identification of the item calibrated g) Date of performance of the calibration h) i) Calibration results with, where appropriate, the units of measurement Calibration results before and after adjustment or repair, if available j) Person authorizing the calibration report k) Statement of measurement traceability l) m) Factors and conditions under which the calibrations were made that have an influence on the measurement results. Uncertainty of measurement, where Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 24 of 27
25 Comments on appropriate, and/or a statement of conformance. Where a statement of conformance with specified calibration performance requirements and/or calibration procedure is provided, the extent of conformance shall be provided and included in the calibration records. Opinions and Interpretations When opinions and interpretations are included, the basis upon which the opinions and interpretations have been made shall be documented. Opinions and interpretations shall be clearly marked as such in a calibration report. Control and Source When it is necessary to amend or issue a complete new calibration report, it shall be uniquely identified and shall contain a reference to the original that it amends or replaces. Calibration Records Records of the calibration process shall be completed to attest to the correctness of the results. Shall be maintained and made available. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 25 of 27
26 Comments on Records of the calibration process shall demonstrate whether measuring and test equipment have satisfied the performance requirements specified. Records shall be maintained of each item of equipment and software where its use contributes to the results of the calibrations performed. The records shall include the following: a) Description and unique identification of the measuring and test equipment. b) Identification of the calibration-servicing component. c) Identification of the calibration procedure used. d) Date on which the calibration was completed. e) f) As-found measurement performance condition of the equipment. Calibration actions taken (adjusted, repaired, new value assigned, limited, derated, modified, etc.) Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 26 of 27
27 Comments on g) Calibration results and the as-returned measurement performance condition of the equipment. h) Limitations of use. i) Assigned calibration interval. Identification of calibration reports. j) Generation, amendment, issuance, and deletion of records shall be authorized. An amendment or deletion of a record shall be documented. Legend: C=Compliant, NC=Noncompliant, NA=Not Applicable, Document=laboratory s document where compliance to the LABPR Rev 1 11/30/10 Page 27 of 27
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
More informationNIST HANDBOOK 150 CHECKLIST
NIST HANDBOOK 150 CHECKLIST Instructions to the Assessor: This checklist addresses the general accreditation criteria prescribed in NIST Handbook 150, NVLAP Procedures and General Requirements (2006 edition).
More informationQuality Assurance Manual January 2008 Revision I Revision date of 01/17/13. Pi Tape Corporation. 344 North Vinewood St. Escondido CA 92029-1337 USA
Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13 Pi Tape Corporation 344 North Vinewood St. Escondido CA 92029-1337 USA Phone # 760 746-9830 Fax # 760 746-9196 In conformance
More informationQuality Management System Policy Manual
Quality Management System Burns Engineering, Inc. 10201 Bren Road East Minnetonka, MN 55343 4. 4.1 General Requirements The Quality Manual consists of the quality system policy documents and the quality
More informationProcedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
More informationISO/IEC 17025 QUALITY MANUAL
1800 NW 169 th Pl, Beaverton, OR 97006 Revision F Date: 9/18/06 PAGE 1 OF 18 TABLE OF CONTENTS Quality Manual Section Applicable ISO/IEC 17025:2005 clause(s) Page Quality Policy 4.2.2 3 Introduction 4
More informationOIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL DOCUMENT OIML D 18 Edition 2008 (E) The use of certified reference materials in fields covered by metrological control exercised by national services of legal metrology. Basic principles
More informationTITLE: Quality System Manual DOCUMENT NUMBER: QMS
1 of 19 BUSINESS PROCESS (BP) QUALITY MANAGEMENT SYS NAME SIGNATURE DATE PREPARED BY BILL ATKINSON SIGNATURE ON FILE 12/3/2014 APPROVALS DEPARTMENT NAME SIGNATURE DATE PRESIDENT DANNY ANGEL SIGNATURE ON
More informationNIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING
NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING Instructions to the Assessor: This checklist addresses specific accreditation criteria prescribed in NIST Handbook 150-5, Construction Materials
More informationNon-Controlled Copy. Quality Manual Revision Log. Rev # Date Description
Quality Manual Revision Log Rev # Date Description 1 Oct 1984 Original Document. 2 Aug 1986 Document Hillsboro Move. Complete Revision. 3 May 1988 Incorporate/Beaverton Move. Reformat to Word Perfect 4.0.
More informationProcedures and General Requirements
NIST HB 150-2016 National Voluntary Laboratory Accreditation Program Procedures and General Requirements Warren R. Merkel Vanda R. White This publication is available free of charge from: http://dx.doi.org/10.6028/nist.hb.150-2016
More informationCALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS 17025.
CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS 17025. Nicholas Mason Metrology Software Project Manager Calibration Group, Fluke Corporation 6920 Seaway Blvd. Everett,
More informationEagle Machining, Inc. Quality Management System
Eagle Machining, Inc. Quality Management System 1 of 10310 Antoine Drive Bldg D, Houston, Texas 77086 BUSINESS OPERATING MANUAL (QUALITY MANUAL) Revision Date: 08/01/2014 Approved By: Joseph Vu Date: 08/01/2014
More informationGap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing
Gap Analysis May 2013 Issued: May 2013 Gap Analysis of and in the field of Medical Testing Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
More informationTHERMAL INSULATION MATERIALS
NIST HANDBOOK 150-15 2006 Edition National Voluntary Laboratory Accreditation Program THERMAL INSULATION MATERIALS Lawrence I. Knab National Voluntary Laboratory Accreditation Program Standards Services
More informationOD-2017-Ed.1.5 CHECK LIST FOR TESTING AND CALIBRATION LABORATORIES
IECEE OD-2017-Ed.1.5 OPERATIONAL & RULING DOCUMENTS CHECK LIST FOR TESTING AND CALIBRATION LABORATORIES Rev 2006-09-22 OD-2017-Ed.1.5 IEC - IECEE 2007 - Copyright all rights reserved Except for IECEE members
More informationEA-7/01. EA Guidelines. on the application. Of EN 45012. Publication Reference PURPOSE
Publication Reference EA-7/01 EA Guidelines on the application Of EN 45012 PURPOSE The purpose of the document is to provide explanations with a view to harmonise the application of ISO/IEC Guide 62/EN
More informationCORPORATE QUALITY MANUAL
Corporate Quality Manual Preface The following Corporate Quality Manual is written within the framework of ISO 9001:2008 Quality System by the employees of CyberOptics. CyberOptics recognizes the importance
More informationGENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION
International Document OIML 1 st CD GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Organization of Legal Metrology Contents 01 Scope 2 12
More informationISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES
ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/01/01 DOCUMENT NUMBER: MA 2002 TABLE OF CONTENTS Foreword... 3 1. Application... 4 2.
More informationRTP s NUCLEAR QUALITY ASSURANCE PROGRAM
RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP operates under one quality program, whether you purchase products that are commercial grade, nuclear safety-related or industrial safety compliant (IEC 61508).
More informationGuidelines on the Reporting of Compliance with Specification
Guidelines on the Reporting of Compliance with Specification ILAC-G8:03/2009 Copyright ILAC 2009 ILAC encourages the authorized reproduction of this publication, or parts thereof, by organisations wishing
More informationcalibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper
calibrate with confidence White Paper Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Authored by Keith Bennett Director of Metrology Howard Zion Director of Technical Operations TRANSCAT
More informationNew Program Quality Requirements for Suppliers (NPQR) For Limited Manufacturing of Components and/or Processes for Engineering Certification
New Program Quality Requirements for Suppliers (NPQR) For Limited Manufacturing of Components and/or Processes for Engineering Certification Revision A New Program Quality Requirements for Suppliers Rev.
More informationISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi The Innovation Hub, Pretoria Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements
More informationAmerican Association for Laboratory Accreditation
Page 1 of 15 R104 GENERAL REQUIREMENTS: ACCREDITATION OF FIELD TESTING AND FIELD CALIBRATION LABORATORIES July 2015 2015 by A2LA All rights reserved. No part of this document may be reproduced in any form
More information14620 Henry Road Houston, Texas 77060 PH: 281-447-3980 FX: 281-447-3988. WEB: www.texasinternational.com QUALITY MANUAL
14620 Henry Road Houston, Texas 77060 PH: 281-447-3980 FX: 281-447-3988 WEB: www.texasinternational.com QUALITY MANUAL ISO 9001:2008 API Spec Q1, 9th Edition API Spec 8C 5 Th Edition MANUAL NUMBER: Electronic
More informationCalibration Service Selection Guide
Calibration Service Selection Guide Certification of Accreditation available for download at www.ricelake.com/accreditation. Answer Questions to determine the calibration service you need Need accredited*
More informationHow To Ensure Quality Assurance
1001 Robb Hill Road Oakdale, PA 15071-3200 412/923-2670 PROCUREMENT SPECIFICATION: SUPPLIER QUALITY MANAGEMENT SYSTEM REQUIREMENTS Specification Number: ATI-QA-703, Rev 4 Date: 11/12/13 Page: 1 of 7 Written
More informationRaad voor Accreditatie (Dutch Accreditation Council RvA) Assessment of Conformity Assessment Schemes
Raad voor Accreditatie (Dutch Accreditation Council RvA) Assessment of Conformity Assessment Schemes Document code: RvA-T033-UK Version 3, 28 februari 2014 A Rv A-Explanatory note describes the policy
More informationNABL NATIONAL ACCREDITATION
NABL 160 NABL NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES GUIDE for PREPARING A QUALITY MANUAL ISSUE NO. : 05 AMENDMENT NO : 00 ISSUE DATE: 27.06.2012 AMENDMENT DATE: -- Amendment
More informationLuminus Devices, Inc Luminus Testing Laboratory Quality Management Systems Manual
Luminus Devices, Inc Luminus Testing Laboratory Quality Management Systems Manual ISO/IEC 17025 NVLAP, NIST HANDBOOKS 150,150-1 This document belongs to Luminus Devices, Inc. It cannot be reproduced without
More informationIAF Informative Document for the Transition of Food Safety Management System Accreditation to ISO/TS 22003:201X from ISO/TS 22003:2007
IAF Informative Document IAF Informative Document for the Transition of Food Safety Management System Accreditation to ISO/TS 22003:201X from ISO/TS 22003:2007 (IAF ID X:201X) Page 2 of 6 The (IAF) details
More informationQuality Assurance QUALITY ASSURANCE PLAN
Revision 2 Page 1 of 40 QUALITY ASSURANCE PLAN PLAN APPROVALS: Jeff Shouse Signature on File DIRECTOR OF QUALITY ASSURANCE DIRECTOR OF QUALITY ASSURANCE (signature) DATE Rodney Baltzer Signature on File
More informationSouth Coast Air Quality Management District
South Coast Air Quality Management District Laboratory Approval Program Administrative, Operational and Technical Requirements November 2015 Direct Inquiries to: Laboratory Approval Program Coordinator
More informationEuropean cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING. The Expression of Uncertainty in Quantitative Testing
European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING Accreditation of Laboratories Publication Reference EAL-G23 The Expression of Uncertainty in Quantitative Testing
More informationCRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective
More informationChecklist. Standard for Medical Laboratory
Checklist Standard for Medical Laboratory Name of hospital..name of Laboratory..... Name. Position / Title...... DD/MM/YY.Revision... 1. Organization and Management 1. Laboratory shall have the organizational
More informationISO 9001:2000 Gap Analysis Checklist
ISO 9001:2000 Gap Analysis Checklist Type: Assessor: ISO 9001 REQUIREMENTS STATUS ACTION/COMMENTS 4 Quality Management System 4.1 General Requirements Processes needed for the quality management system
More informationGeneral and statistical principles for certification of RM ISO Guide 35 and Guide 34
General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in
More informationNATO STANDARD AQAP-2310 NATO QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR AVIATION, SPACE AND DEFENCE SUPPLIERS
NATO STANDARD AQAP-2310 NATO QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR AVIATION, SPACE AND DEFENCE SUPPLIERS Edition A Version 1 April 2013 NORTH ATLANTIC TREATY ORGANIZATION ALLIED ADMINISTRATIVE PUBLICATION
More informationISO 9001:2008 Audit Checklist
g GE Power & Water ISO 9001:2008 Audit Checklist Organization Auditor Date Page 1 Std. 4.1 General s a. Are processes identified b. Sequence & interaction of processes determined? c. Criteria for operation
More informationGLOBAL AEROSPACE PROCEDURE
1 of 13 1. PURPOSE: This procedure establishes the minimum PPG Aerospace requirements for a Supplier's Quality Plan. These requirements pertain to the documentation and inspection/testing necessary to
More informationNABL accreditation is a formal recognition of the technical competence of a medical testing laboratory
NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory Third party assessment NABL has Mutual Recognition Arrangement with APLAC and ILAC NABL accredited
More informationState Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist
This checklist is intended to help laboratories improve the quality of their testing programs with regard to meat and poultry testing samples, in order to ensure those laboratories are at least equal to
More informationQuality Assurance System Manual
The online version of this procedure is official. Therefore, all printed versions of this document are unofficial copies. St. Louis Public Schools Quality Assurance System Manual 801 North 11 th Street
More informationISO 9001 (2000) QUALITY MANAGEMENT SYSTEM ASSESSMENT REPORT SUPPLIER/ SUBCONTRACTOR
Page 1 of 20 ISO 9001 (2000) QUALITY MANAGEMENT SYSTEM ASSESSMENT REPORT SUPPLIER/ SUBCONTRACTOR SUPPLIER/ SUBCONTRACTOR NAME: ADDRESS: CITY AND STATE: ZIP CODE: SUPPLIER/MANUFACTURER NO PHONE: DIVISION:
More informationCHECKLIST ISO/IEC 17021:2011 Conformity Assessment Requirements for Bodies Providing Audit and Certification of Management Systems
Date(s) of Evaluation: CHECKLIST ISO/IEC 17021:2011 Conformity Assessment Requirements for Bodies Providing Audit and Certification of Management Systems Assessor(s) & Observer(s): Organization: Area/Field
More informationGeneral Rules for the Certification of Management Systems Code: RG
General Rules for the Certification of Management Systems Code: RG Drafted on: 1 April 2012 Effective from: 1 October 2012 TABLE OF CONTENTS CHAPTER TITLE PAGE CHAPTER 1 GENERAL 3 CHAPTER 2 REFERENCE STANDARD
More informationPROPOSED UPDATED TEXT FOR WHO GOOD MANUFACTURING PRACTICES FOR PHARMACEUTICAL PRODUCTS: MAIN PRINCIPLES (JANUARY 2013)
January 2013 RESTRICTED PROPOSED UPDATED TEXT FOR WHO GOOD MANUFACTURING PRACTICES FOR PHARMACEUTICAL PRODUCTS: MAIN PRINCIPLES (JANUARY 2013) DRAFT FOR COMMENTS Please address any comments on this proposal
More information1 For more information T: +1.847.664.1499 / E: DAP@ul.com / W: ul.com/dap
DATA ACCEPTANCE PROGRAM UL Client Test Data Program Investigations of products by UL typically involve the testing of product samples. While UL maintains extensive facilities for these purposes, customers
More informationISO 9001: 2008 Boosting quality to differentiate yourself from the competition. xxxx November 2008
ISO 9001: 2008 Boosting quality to differentiate yourself from the competition xxxx November 2008 ISO 9001 - Periodic Review ISO 9001:2008 Periodic Review ISO 9001, like all standards is subject to periodic
More informationOMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (06) 72 2R (FORM/158-Rev.02) STANDARD AIDE-MEMOIRE FOR THE MJA/MJV OF OMCLs Full document title and reference Document type
More informationSurgi Manufacturing Quality Manual
Surgi Manufacturing Page 1 of 18 Approvals: QA: Eng. Mgt. : A Date: 18Aug98 1. Introduction... 4 1.1 Scope... 4 1.2 Purpose... 4 1.3 Authority... 4 1.4 Issue of the Manual... 4 1.5 Amendments... 4 1.6
More informationCompany Quality Manual Document No. QM Rev 0. 0 John Rickey Initial Release. Controlled Copy Stamp. authorized signature
Far West Technology, Inc. ISO 9001 Quality Manual Document No.: QM Revision: 0 Issue Date: 27 August 1997 Approval Signatures President/CEO Executive Vice President Vice President/CFO Change Record Rev
More informationISO 9001:2000 AUDIT CHECKLIST
ISO 9001:2000 AUDIT CHECKLIST No. Question Proc. Ref. Comments 4 Quality Management System 4.1 General Requirements 1 Has the organization established, documented, implemented and maintained a quality
More informationComparison ISO/TS 16949 (1999) to VDA 6.1 (1998)
1 APPLICABILITY VDA 6.1: Section: 3.1; 7 new: In addition to the applicability for supplier sites for production, services and their subcontractors for: products and production materials, or services like
More informationDEC STD 017-2 - ISO 9002 - Quality Systems - Model for Quality Assurance in Production and Installation
Digital Internal Use Only DEC STD 017-2 - ISO 9002 - Quality Systems - Model for Quality Assurance in Production and Installation DOCUMENT IDENTIFIER: A-DS-EL00017-02-0000 Rev A, ABSTRACT: This standard
More informationCartel Electronics. AS 9100 Quality Systems Manual
Cartel Electronics AS 9100 Quality Systems Manual 1900 C Petra Lane Placentia, California 92870 Introduction Cartel Electronics, as a global supplier to the aviation, space, and space industries, has developed
More informationQuality Management Plan Procedure
Quality Management Plan Procedure Check the Capital Project Delivery website to ensure this is the current version. Table of Contents Quality Management Plan Overview... 1 Introduction... 1 Objectives...
More informationEA IAF/ILAC Guidance. on the Application of ISO/IEC 17020:1998
Publication Reference EA IAF/ILAC-A4: 2004 EA IAF/ILAC Guidance on the Application of ISO/IEC 17020:1998 PURPOSE This guidance document is for ISO/IEC 17020: General Criteria for the operation of various
More informationcalibrate confidence with Risk Management: The Critical Role of Measurement and Calibration WHITE PAPER SERVICES
calibrate with confidence The Critical Role of Measurement and Calibration The ultimate goal of your company is to be profitable in providing a superior, effective, and safe product or service. Your use
More informationCalibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
More informationISO 9000 Introduction and Support Package: Guidance on the Documentation Requirements of ISO 9001:2008
Document: ISO/TC 176/SC 2/N 525R2 ISO 9000 Introduction and Support Package: 1 Introduction Two of the most important objectives in the revision of the ISO 9000 series of standards have been a) to develop
More informationACCEPTANCE CRITERIA FOR QUALITY DOCUMENTATION (AC10) June 2014
ACCEPTANCE CRITERIA FOR QUALITY DOCUMENTATION (AC10) 1.0 INTRODUCTION June 2014 1.1 Purpose: The purpose of this acceptance criteria is to establish requirements for information that is provided to ICC-ES
More informationApplication of ISO/IEC 17011 for the Accreditation of Food Safety Management Systems (FSMS) Certification Bodies
IAF Mandatory Document Application of ISO/IEC 17011 for the Accreditation of Food Safety Management Systems (FSMS) Certification Bodies (IAF MD 16:2015) Version 2 Food Safety Management Systems (FSMS)
More informationPRODUCT CERTIFICATION REQUIREMENTS
PRODUCT CERTIFICATION REQUIREMENTS SIRIM QAS International Product Certification Page 1 of 14 QUALITY SYSTEM 1. GENERAL 1.1 This document defines the responsibilities and obligations of the Licensee with
More informationTesting Facility Laboratory Manual
Quality Synthetic Rubber's Testing Facility Laboratory Manual Quality Synthetic Rubber, Inc. 1700 Highland Road P.O. Box 1030 Twinsburg, Ohio 44087-9030 Phone: 330-425-8472 Duns Code: 01-753-2458 Z-Code:
More informationGUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E)
GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E) Guidelines for the determination of calibration intervals of measuring instruments Guide pour la détermination
More informationSUPPLIER QUALITY PROGRAM, V (U)
FEDERAL MANUFACTURING & TECHNOLOGIES PQR 1060 PAGE 1 OF 17 SUPPLIER QUALITY PROGRAM, V (U) CHANGE HISTORY ISSUE DESCRIPTION OF CHANGE AUTHOR DATE E CORRECTED CHANGE HISTORY S. L. HALTER 10/18/2000 F UPDATED
More informationOPERATIONAL STANDARD
1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 9000 Third edition 2005-09-15 Quality management systems Fundamentals and vocabulary Systèmes de management de la qualité Principes essentiels et vocabulaire Reference number
More informationISO 9001:2015 Overview of the Revised International Standard
ISO 9001:2015 Overview of the Revised International Standard Introduction This document provides: a summary of the new ISO 9001:2015 structure. an overview of the new and revised ISO 9001:2015 requirements
More informationCriminal Justice Offender Tracking System Certification Program Requirements
U.S. Department of Justice Office of Justice Programs National Institute of Justice Criminal Justice Offender Tracking System Certification Program Requirements NIJ CR-1004.00.00 June 2012 NCJ xxxxxx National
More informationwww.itilhelp.com ISO 9000 Quality Standard Background Information Quality Systems Implementation
ISO 9000 Quality Standard Background Information www.itilhelp.com There is a worldwide trend towards more stringent customer expectations with regard to quality. Accompanying this trend has been a growing
More informationThis Business Management System Manual is based on ISO 9001:2008 requirements
Page 1 of 19 (Formerly: Quality Management System Manual) When the ISO 9001:2008 standard is identifying the Quality Manual or Quality Management System, Turbonetics is representing this requirement with
More informationSection i POLICY AND AUTHORITY STATEMENT
Section i POLICY AND AUTHORITY STATEMENT It is the policy of Laboratory Testing, Incorporated (LTI)*, as a Material Organization, to furnish and perform testing services in nondestructive, mechanical,
More informationAEROSPACE STANDARD. Quality Management Systems - Requirements for Aviation, Space and Defense Organizations RATIONALE
AEROSPACE STANDARD AS9100C Issued 1999-11 Revised 2009-01 Superseding AS9100B Quality Management Systems - Requirements for Aviation, Space and Defense Organizations RATIONALE This standard has been revised
More informationISO-9001:2000 Quality Management Systems
ISO-9001:2000 Quality Management Systems REQUIREMENTS 10/10/2003 ISO-9001:2000 Requirements 1 Process Based Approach C U S MANAGEMENT RESPONSIBILITY RESOURCE MANAGEMENT C U S T O M Requirements PRODUCT
More informationEA Document on. Accreditation. For Notification Purposes
Publication Reference EA-2/17 INF: 2014 EA Document on Accreditation For Notification Purposes PURPOSE The document presents the policy agreed by EA Members for accreditation of Conformity Assessment Bodies
More informationALACC Frequently Asked Questions (FAQs)
Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC
More informationCOPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.
Revision History COPY The top row of this table shows the most recent changes to this controlled document. For previous revision history information, archived versions of this document are maintained by
More informationREQUIREMENTS FOR CERTIFICATION BODIES TO DETERMINE COMPLIANCE OF APPLICANT ORGANIZATIONS TO THE MAGEN TZEDEK SERVICE MARK STANDARD
REQUIREMENTS FOR CERTIFICATION BODIES TO DETERMINE COMPLIANCE OF APPLICANT ORGANIZATIONS TO THE MAGEN TZEDEK SERVICE MARK STANDARD Foreword The Magen Tzedek Commission has established a standards and certification
More informationISO 9001 Quality Systems Manual
ISO 9001 Quality Systems Manual Revision: D Issue Date: March 10, 2004 Introduction Micro Memory Bank, Inc. developed and implemented a Quality Management System in order to document the company s best
More informationNLECTC-National Ballistic-resistant Body Armor Inspection Scheme
Subject: 24 June 2015 1 Scope The National Law Enforcement and Corrections Technology Center-National (NLECTC-National) is a program of the U.S. Department of Justice (DOJ), Office of Justice Programs
More informationChecklist for Operational Risk Management
Checklist for Operational Risk Management I. Development and Establishment of Comprehensive Operational Risk Management System by Management Checkpoints - Operational risk is the risk of loss resulting
More informationRow Manufacturing Inc. Quality Manual ISO 9001:2008
Row Manufacturing Inc. Quality Manual ISO 9001:2008 Row Manufacturing 210 Durham Drive Athens, Alabama 35611 Phone:256.232.4151 Fax:256.232.4133 Page 2 of 33 This Page intentionally left Blank Page 3 of
More informationInternal Audit Checklist
Internal Audit Checklist 4.2 Policy Verify required elements Verify management commitment Verify available to the public Verify implementation by tracing links back to policy statement Check review/revisions
More informationNational Institute of Justice
U.S. Department of Justice Office of Justice Programs National Institute of Justice National Institute of Justice Law Enforcement and Corrections Standards and Testing Program NIJ Duty Holster Certification
More informationR103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 LABORATORIES
Page 1 of 7 R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 November 2009 2009 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
More informationCopies of this document shall be classified either as Controlled or Uncontrolled, and shall be identified as such on the cover sheet.
INTRODUCTION This document describes an inspection system broken down into four separate levels. When Inspection Level IV is invoked, this system meets the requirements of former MIL-I-45208A. The primary
More informationISO/IEC 17020 ACCREDITATION REQUIREMENTS FOR FORENSIC INSPECTION BODIES
ISO/IEC 17020 ACCREDITATION REQUIREMENTS FOR FORENSIC INSPECTION BODIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/02/02 DOCUMENT NUMBER: MA 3012 TABLE OF CONTENTS Foreword... 3 1. Application... 4
More informationSupplier Quality Assurance Requirements Manual
12716 EAST PINE STREET TULSA, OK 74116 Supplier Quality Assurance Requirements Manual Approved: Mike Collins Quality Assurance Manager Precise Machining & Manufacturing 1 Record of Revisions Revision D
More informationThe following paragraphs, identified to coincide with the OHSAS 18001:2007 numbering system, provide a clause-by-clause summary of the standard.
Summary of OHSAS 18001:2007 Requirements With this article, the 18000 store provides a brief and clear summary of the OHSAS 18001:2007 requirements. First of all, OHSAS 18001 is an international standard
More informationGuidance from the Group of Notified Bodies for the Construction Products Directive 89/106/EEC. GNB-CPD position paper from SG17 - EN 1090-1:2009
GNB-CPD SG17 Guidance from the Group of Notified Bodies for the Construction Products Directive 89/106/EEC NB-CPD/SG17/09/069 Issued: 7 September 2009 APPROVED GUIDANCE GNB-CPD position paper from SG17
More informationMeasuring equipment in a laboratory.
Measuring equipment in a laboratory. Laboratory and metrology are two words that practically do not function separately. They are integrally connected to measuring equipment for tests and measurements.
More informationIAF Informative Document. IAF Informative Document for the Transition of Management System Accreditation to ISO/IEC 17021:2011 from ISO/IEC 17021:2006
IAF ID 2:2011 International Accreditation Forum, Inc. IAF Informative Document IAF Informative Document for the of Management System Accreditation to ISO/IEC 17021:2011 from (IAF ID 2:2011) The International
More informationDNV GL Assessment Checklist ISO 9001:2015
DNV GL Assessment Checklist ISO 9001:2015 Rev 0 - December 2015 4 Context of the Organization No. Question Proc. Ref. Comments 4.1 Understanding the Organization and its context 1 Has the organization
More information