INSTRUCTION MANUAL MODEL 157 AC/DC LEAKAGE CURRENT TESTER

Size: px
Start display at page:

Download "INSTRUCTION MANUAL MODEL 157 AC/DC LEAKAGE CURRENT TESTER"

Transcription

1 INSTRUCTION MANUAL MODEL 157 AC/DC LEAKAGE CURRENT TESTER a world leader in test & measurement FINE INSTRUMENTS CORPORATION Room # 601, DAEWOO TECHNOPIA, 768-1, WONSI-DONG, DANWON-GU, ANSAN-SI, GYEONGGI-DO, SOUTH KOREA - TEL: (82-31) FAX: (82-31) fine@finest.co.kr - Internet: co.kr Copyright 2000 Fine Instruments Corp. All right reserved. Specifications subject to change without notice. Litho in Korea. FINE INSTRUMENTS CORPORATION

2 Table of Contents 1. Introduction 2. Safety 3. Technical Specifications 4. Operating Instructions 5. How to Measure 6. Maintenance and Calibration SOURCES LIKE SMALL HAND-HELD RADIO TRANSCEIVERS, FIXED STATION RADIO AND TELEVISION TRANSMITTERS, VEHICLE RADIO TRANSMITTERS AND CELLULAR PHONES GENERATE ELECTROMAGNETIC RADIATION THAT MAY INDUCE VOLTAGES IN THE TEST LEADS OF THE MULTIMETER. IN SUCH CASES THE ACCURACY OF THE MULTIMETER CANNOT BE GUARANTEED DUE TO PHYSICAL REASONS. 1. Introduction This Meter is a handheld battery operated AC/DC Leakage Current Tester that is designed to measure the AC/DC leakage current and high current up to 40A. This Meter is designed and tested according to IEC Publication ( )(Overvoltage Category lll) and the EMC Directive (EN and EN ) and other safety standards (see Technical Specifications ). This Meter has a frequency filtering function for measuring currents of fundamental frequency. This filter function can be used when we check whether the leakage breaker was tripped by leakage currents or by effects of harmonics. This Meter provides : Shielded transformer jaws to minimize the effect of external stray magnetic field 2A AC range with the 1mA best resolution Can measure from 1 ma to 40A AC Auto Ranging for Volts and Ohms Manual Ranging for AC/DC Amperes Can accurately measure leakage current from high frequency electric appliances such as inverters 1

3 DC Amps Zero Adjustment function Data hold function Law battery indicator Automatic Power off after 10 minutes inactivity in any measurement function Frequency Response Range : 50Hz to 400Hz 2. Safety Read the following safety information carefully before attempting to operate or service the Meter. Although the Meter is designed to be as safe as possible, safety in use is the responsibility of the operator who must be suitably qualified and/or authorized. Maximum voltage between current carrying conductors, which are uninsulated, and ground must not exceed 600V RMS. Always choose an insulated conductor if available. Do not attempt to measure any current or voltage higher than the maximum limits of the input values shown in the electrical specifications. Never measure any current while the test leads are inserted into the input jacks. Never use the Meter or test leads whose insulating protection has been impaired. Be extremely cautious when clamping around uninsulated conductors or bus bars. Never leave the Meter to contact with any surface water. Ingress of water could damage the electronics as well as cause a potential shock hazard. Accidental contact with the conductor could result in electric shock. Frequently inspect the Meter and the test leads for damage. If the Meter is physically damaged or does not function correctly, don t use it. Avoid working alone with high voltage circuits. Use the meter only as specified in this manual. Otherwise, the protections provided by this Meter may be impaired. When taking measurements in high voltage circuits, connections should be made to the circuits while the power is off. Safety Symbols Warning There is a risk of electrical shock. Caution Refer to this manual before using this meter. Double Insulation The meter is protected throughout by double insulation or reinforced insulation. When servicing, use only specified replacement parts. CAT III - Installation category (overvoltage category) lll according to EN and EN Alternating Current (AC) Only use the FTL-500V1 test lead supplied by Fine Instruments Corporation. Disconnect the live test lead before disconnecting the common test lead. 2 Direct Current (DC) Both DC and AC Earth Ground 3

4 4 3. Technical Specifications 3.1 General Specifications Digital Display (LCD) : Counts 4000 Updates 3 times/sec. Maximum Volts to Ground : 600V RMS Overflow Indication : LCD will show an OFL Low Battery Indication : The is displayed when the battery voltage drops below the operating voltage. Battery Life : 60 hours typical (alkaline) Battery Type : NEDA 1604A 9V or 6LF 2 9V Maximum Jaw Opening : 36mm Maximum Conductor Size : ø 33mm Maximum Conductor Voltage: 600V RMS Indoor use Storage Temperature : -10 C to 50 C Operating Temperature : 0 C to 45 C at and Humidity R.H < 80% non-condensing Aititude : 2,000m Relative Humidity : 0% to 80% at 32 F to 95 F (0 C to 35 C ) 0% to 70% at 95 F to 115 F (35 C to 45 C ) Pollution Degree : 2 Safety Conformance : Conforms to the requirements of UL , CSA C22.2 No and ANSI/ISA-S Weight : Approx. 320g Size : 190 (L) x 76.8 (W) x 42.5 (D)mm Standard Accessories : Battery (9V), Instruction manual, test leads and soft carry case 3.2 Electrical Specifications Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18 C to 28 C with a relative humidity of up to 80%, for a period of one year after calibration. AC/DC Current FUNCTION RANGE RESOLUTION ACCURACY DC A AC A (50Hz to 400Hz) 2 A * 40 A 2 A * 40 A 1 ma 10 ma 1 ma 10 ma 2.0% % % % + 5 OVERLOAD PROTECTION 600V RMS Filtering Function for Measuring Currents of Fundamental Frequency The Filtering function can make the Meter measure currents of fundamental frequency range when the Filter switch is at the ON position. The fundamental frequency range (Filter ON position) has the frequency characteristics for the limited frequency range of 50Hz to 100Hz and the wide frequency range (Filter OFF position) has the frequency characteristics for the frequency range of 50Hz to 400Hz. This Filtering function is very useful for checking whether the currents contain harmonics or not. Recently there has been increased usage of power through inverters, switching regulators, etc. When the high frequency noise from such appliances leaks or flows into the ground through capacitors not filtering completely, the earth leakage breaker may not trip. In such a case the Meter may not give correct current readings with the Filter switch at the ON position. Therefore, it is necessary to make current measurements with the Filter switch at the OFF position. 5

5 When in doubt as to the presence of high frequencies and harmonics that affect AC current measurements, take current readings with the switch at the ON and OFF positions respectively and then compare the results obtained. Effect of External Stray Magnetic Field 50mA AC max. in proximity to 100A AC carrying, ø10mm conductor 4. Operating Instructions 4.1 Controls and Indicators (1) CLAMP. Opens 36mm to enclose conductors. (2) LEVER. Open the close clamp jaws. AC/DC Volts FUNCTION RANGE RESOLUTION ACCURACY INPUT IMPEDANCE 4V 1mV DC V 40V 10mV 400V 100mV 0.9% V 1V 10MΩ 4V 1mV AC V 40V 10mV (50Hz to 1.5% V 100mV 400Hz) 600V 1V Ohms FUNCTION RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION 400Ω 0.1Ω 4kΩ 1Ω 0.9% kΩ 10Ω Ohms 600V RMS 400kΩ 100Ω 4MΩ 1kΩ 40MΩ 10kΩ 1.2% + 5 (1) (2) (4) (5) (6) (3) (9) (10) Continuity Beeps less than approximate 50Ω (7) (8) 6 7

6 (3) POWER ON / FUNCTION SELECTION SWITCH. Turns power on and selects functions. Always set the switch to the OFF position after use. (4) MODE. Toggles between AC mode and DC mode in the either Amperes or Volts measurement range. Toggles between mode and Ω mode in the / Ω range. (5) DATA HOLD (DC A ZERO). Freezes reading in digital display. Press this button for 2 seconds in the DC amperes measurement mode to initiate the Zero Adjustment procedure. (6) LCD. 3 3/4 digit, liquid crystal display with maximum indication of Function symbols, decimal points, (negative) indicator or low battery indicator appears automatically. OFL is displayed for overrange indication. (7) COM. Common Terminal. (8) VΩ. Volts or Ohms Input Terminal. (9) CURRENT RANGE SWITCH. Selects between 4A and 40A range in the AC/DC amperes measurement mode. (10) FILTER ON/OFF SWITCH. Selects between FILTER ON and FILTER OFF in the AC amperes measurement mode. 5. How to Measure 5.1 Before using the Meter, familiarize yourself with the instrument and this manual. READ AND REVIEW THIS MANUAL FREQUENTLY. 5.2 Visual Inspection of the Meter Before use, inspect the Meter for any conditions which would make it unsafe. Check for missing or loose parts. Check also for damage. Make certain that the Meter is free from oils, grease, dirt and excessive moisture. IF ANY OF THESE CONDITIONS ARE FOUND, DO NOT USE THE METER. 5.3 Visual Inspection of the Test Leads Before use, check the test leads for deterioration of the insulation. IF THE TEST LEADS ARE IN ANY WAY DAMAGED OR DEGRADED, DO NOT USE THEM. THEY MUST BE REPLACED. Replace the battery before using the Meter when the low battery indicator is displayed on the LCD. If the Meter does not turn ON, the battery is missing or worn out. To replace the battery, see Battery Replacement of section 6. The Meter shuts OFF after 10 minutes if no pushbutton is pressed, even though it is making a measurement in any measurement function. 8 9

7 5.4 Measuring AC Current THE METER IS NOT DESIGNED TO MAKE CURRENT MEASUREMENTS ON CIRCUITS WITH VOLTAGE ABOVE 600V AC. MEASUREMENTS ON CIRCUITS ABOVE THIS VOLTAGE MAY CAUSE SHOCK HAZARD AND DAMAGE TO THE METER OR EQUIPMENT UNDER TEST. For measurements of earth leakage current and small current that flow through a grounded wire, refer to Fig.2. It is recommended that 4A range is used for these measurements. 1. Put the Meter in the desired AC amperes range. 2. For normal measurements, refer to the Fig.1. or (CORRECT METHOD) Fig.2 Leakage Current Measurements on the Single Phase Systems For measurements of unbalanced leakage current, clamp the jaws around all the current carrying conductors except a grounded wire. Refer to the Fig. 3. (WRONG METHOD) Fig.1 Normal AC Current Measurements 10 11

8 THE TRANSFORMER JAWS ARE MADE OF METAL AND THEIR TIPS ARE NOT INSULATED. THEREFORE, IT IS VERY DANGEROUS TO TOUCH THE METAL SECTION OF EQUIPMENT UNDER TEST WITH THE JAW TIPS. BE EXTREMELY CAREFUL WHEN CLAMPING AROUND A BARE CONDUCTOR. (3 Wires System without Neutral) 5.5 Measuring DC Current For normal measurements, 1. Put the Meter in the desired DC amperes range. 2. Set the Meter display zero value by pressing the Data Hold button for 2 seconds. 3. Press the lever to open the clamp jaws and clamp around a conductor. (If necessary, use the arrow direction mark to identify the direction of DC current flow.) 4. The measured current will appear on the display of the meter. (4 Wires System With Neutral) For measurements of DC leakage current, repeat the above 1 to 3 referring to the Fig.2 and Fig.3 of 5.4 above. Fig.3 Leakage current Measurements on the Three Phase Systems It is recommended that the conductors are placed at the center of the closed jaws as much as possible. 3. If you want to measure currents in the fundamental frequency range only, set the Filter switch to the ON position 4. Set the Power/Function switch to the OFF position after using the Meter

9 5.6 Measuring AC (or DC) Volts TO AVOID THE RISK OF ELECTRICAL SHOCK, INSTRUMENT DAMAGE AND/OR EQUIPMENT DAMAGE, INPUT VOLTAGES MUST NOT EXCEED 600 VOLTS AC (or DC). DO NOT ATTEMPT TO TAKE ANY UNKNOWN VOLTAGE MEASUREMENTS THAT MAY BE IN EXCESS OF 600 VOLTS AC (or DC). 1. Put the Meter in the Volts AC (or DC) range. 2. Plug the red test lead into the VΩ input terminal and the black test lead into the COM input terminal of the Meter. 3. Disconnect the power from the circuit to be tested. 4. Connect the test leads to the circuit to be tested. 5. Reapply power to the circuit. The measured voltage will appear on the display of the Meter. 6. When measuring DC volts only, if the red test lead is connected to the negative (or lower voltage) side of the circuit, a minus indicator will appear on the left hand side of the LCD. 5.7 Measuring Resistance (Ohms, Continuity) 1. Set the Function switch to the /Ω position. 2. Insert the black test lead into the COM input terminal and the red test lead into the VΩ input terminal. 3. Connect the test leads to the circuit to be measured. 4. The measured resistance will be displayed. Audible Continuity Buzzer 1. Set the Function switch to the /Ω position and select the continuity test mode by pressing the MODE button. 2. Insert the black test lead into the COM input terminal and the red test lead into the VΩ input terminal. 3. Connect the test leads to the circuit to be tested. 4. The Meter will emit a continuous tone for resistances of less than 50 ohms. Data Hold 1. Obtain a stable reading on any function or range of the Meter. 2. Press the Data Hold button. (HOLD will be displayed.) 3. The reading will be frozen on the display until the Data Hold button is pressed again. Auto-Power-Off The Meter automatically shuts off after 10 minutes of inactivity. CAUTION TURN OFF POWER AND DISCHARGE ALL CAPACITORS ON CIRCUIT TO BE TESTED BEFORE ATTEMPTING IN-CIRCUIT RESISTANCE MEASUREMENTS. FAILURE TO DO SO MAY END UP IN EQUIPMENT AND/OR INSTRUMENT DAMAGE

10 6. Maintenance and Calibration Battery Replacement IF THIS METER IS USED IN A MANNER NOT SPECIFIED BY THIS MANUAL, THE PROTECTION PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED. REPAIRS OR SERVICING NOT COVERED IN THIS MANUAL SHOULD ONLY BE PERFORMED BY QUALIFIED PERSONNEL. TO AVOID ELECTRICAL SHOCK, DO NOT SERVICE UNLESS YOU ARE QUALIFIED TO DO SO. Cleaning and Decontamination Periodically clean your meter s case using clamp cloth. Do not use abrasives, cleaning solvents or strong detergents, as they may damage the finish or affect the reliability of the structural components. TO PREVENT ELECTRICAL SHOCK HAZARD, TURN OFF THE METER. DISCONNECT ALL TEST LEADS AND UNCLAMP THE METER FROM ANY CONDUCTOR BEFORE REMOVING THE BATTERY COVER. 1. Unplug the test leads from the instrument. 2. Remove the screw from the battery compartment cover and lift the cover. 3. Replace the battery with the same type and size as the one removed. 4. After a battery change, this battery cover must be reinstalled and that it must not be discarded. 5. Reattach the battery compartment cover to the meter and reinstall the screw. Calibration Calibrate this Meter once a year to maintain its performance specifications

UT202A Operating Manual. Contents

UT202A Operating Manual. Contents Title Contents Page Overview Unpacking Inspection Safety Information Rules for Safe Operation International Electrical Symbols The Meter Structure Functional Buttons and auto power off Display Symbols

More information

Model UT33A: OPERATING MANUAL Table of Contents (1)

Model UT33A: OPERATING MANUAL Table of Contents (1) Table of Contents (1) Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter structure Functional Buttons LCD Display Measurement Operation

More information

Model UT201/202: OPERATING MANUAL. Table of Contents

Model UT201/202: OPERATING MANUAL. Table of Contents Table of Contents Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons The Effectiveness

More information

Manual Ranging MultiMeter

Manual Ranging MultiMeter Owner s Manual Manual Ranging MultiMeter Model 82345 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product.! Safety! Operation! Maintenance!

More information

Model UT33B/C/D: OPERATING MANUAL Table of Contents (1)

Model UT33B/C/D: OPERATING MANUAL Table of Contents (1) Table of Contents (1) Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter structure Functional Buttons Measurement Operation A. DC

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763 61-765

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763 61-765 IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763 61-765 The Service Information provides the following information: Precautions and safety information Specifications Performance test procedure Calibration

More information

AutoRanging Digital MultiMeter

AutoRanging Digital MultiMeter Owner's Manual AutoRanging Digital MultiMeter Model No. 82139 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance

More information

User Guide. Model 380260 Insulation Tester / Megohmmeter

User Guide. Model 380260 Insulation Tester / Megohmmeter User Guide Model 380260 Insulation Tester / Megohmmeter Introduction Congratulations on your purchase of Extech s Insulation Tester/Megohmmeter. The Model 380260 provides three test ranges plus continuity

More information

Ground Resistance Clamp On Tester

Ground Resistance Clamp On Tester USER MANUAL Ground Resistance Clamp On Tester MODEL 382357 Introduction Congratulations on your purchase of Extech s 382357 Ground Resistance Tester. This Clamp on device allows the user to measure ground

More information

Model UT58A/B/C: OPERATING MANUAL

Model UT58A/B/C: OPERATING MANUAL Table of Contents Title Page Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons Display Symbols

More information

Auto-ranging Digital Multimeter 52-0052-2 INSTRUCTION MANUAL

Auto-ranging Digital Multimeter 52-0052-2 INSTRUCTION MANUAL Auto-ranging Digital Multimeter 52-0052-2 INSTRUCTION MANUAL WARNING: READ AND UNDERSTAND THIS MANUAL BEFORE USING YOUR MULTIMETER. FAILURE TO UNDERSTAND AND COMPLY WITH WARNINGS AND OPERATING INSTRUCTIONS

More information

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual. i800 AC Current Clamp Instruction Sheet Introduction The i800 AC Current Clamp, the Clamp, has been designed for use with multimeters, recorders, power analyzers, safety testers, etc., for accurate non-intrusive

More information

10 Multimeter. Users Manual

10 Multimeter. Users Manual 10 Multimeter Users Manual 1991-2001 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies. 10 MULTIMETER m m Mk SELECT RANGE OFF VDC VAC

More information

User's Guide. True RMS Industrial Multimeter

User's Guide. True RMS Industrial Multimeter User's Guide 97650 True RMS Industrial Multimeter Ω C ã F ã 10A V µ 10A V ã ã ma A Introduction This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency (electrical & electronic),

More information

User Guide. www.burntec.com. 4 wire Earth Resistance Tester Model GRT300

User Guide. www.burntec.com. 4 wire Earth Resistance Tester Model GRT300 User Guide 4 wire Earth Resistance Tester Model GRT300 Introduction Congratulations on your purchase of Extech s 4 Wire Earth Resistance Tester. The Model GRT300 has been designed and tested according

More information

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual EPM3 Phase Sequence and Motor Rotation Tester Users Manual 1 L1 L2 L3 2 3 A B C CAT 600V 3-PHASE TESTER 5 TEST M1 M2 BATT M3 EPM3 MOTOR ROTATION TESTER EPM3 MOTOR ROTATION DETERMINED WHILE FACING MOTOR

More information

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12 Table of Contents The Basics of Electricity 2 Using a Digital Multimeter 4 IDEAL Digital Multimeters An Introduction The Basics of Digital Multimeters is designed to give you a fundamental knowledge of

More information

HVAC Clamp Meter. Users Manual

HVAC Clamp Meter. Users Manual 902 HVAC Clamp Meter Users Manual PN 2547887 May 2006 Rev. 1, 3/07 2006-2007 Fluke Corporation. All rights reserved. Printed in China. All product names are trademarks of their respective companies. LIMITED

More information

How To Use A Cdm250 Digital Multimeter

How To Use A Cdm250 Digital Multimeter User Manual CDM250 Digital Multimeter 070-6736-03 Copyright Tektronix, Inc. 1987. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this

More information

Table of Contents. Page 3 4 5 6 8 9 10 11 13 13 14 16 17 19

Table of Contents. Page 3 4 5 6 8 9 10 11 13 13 14 16 17 19 Table of Contents Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Multimeter Structure Functional Buttons Display Symbols Measurement

More information

Model UT713 OPERATING MANUAL

Model UT713 OPERATING MANUAL Model UT713 OPERATING MANUAL TABLE OF CONTENTS TITLE PAGE Introduction Unpacking Inspection Safety Information Turning the Calibrator On Simulating a Thermocouple Measuring a Thermocouple Simulating TC

More information

7 FUNCTION DIGITAL MULTIMETER

7 FUNCTION DIGITAL MULTIMETER 7 FUNCTION DIGITAL MULTIMETER 90899 OPERATING INSTRUCTIONS 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our Web site at http://www.harborfreight.com Copyright 2004 by Harbor Freight Tools. All rights

More information

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011 Instruction Manual 2in1 LAN Tester & Multimeter Model: LA-1011 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.

More information

i410/i1010 AC/DC Current Clamp

i410/i1010 AC/DC Current Clamp /i1010 AC/DC Current Clamp Instruction Sheet Safety Information XW Read First: Safety Information To ensure safe operation and service of the current clamp, follow these instructions: Read all operating

More information

T+ and T+ PRO Electrical Tester

T+ and T+ PRO Electrical Tester T+ and T+ PRO Electrical Tester Instruction Sheet Introduction The Fluke T+ and T+ PRO Electrical Testers (the Tester ) have the following features: AC and dc voltage measurement, 12 V to 600 V, with or

More information

Ground Resistance Clamp-On Tester

Ground Resistance Clamp-On Tester User s Manual Ground Resistance Clamp-On Tester MODEL 382357 Warranty EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants the basic instrument to be free of defects in parts and workmanship for one

More information

Seven function digital multimeter

Seven function digital multimeter Seven function digital multimeter 98025 Set up And Operating Instructions Distributed exclusively by Harbor Freight Tools. 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our website at: http://www.harborfreight.com

More information

Instructions A622 100 Amp AC/DC Current Probe 070-8883-03

Instructions A622 100 Amp AC/DC Current Probe 070-8883-03 Instructions A622 100 Amp AC/DC Current Probe 070-8883-03 Revision A www.tektronix.com 070888303 Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents,

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

1587/1577. Insulation Multimeters. Technical Data. Two powerful tools in one.

1587/1577. Insulation Multimeters. Technical Data. Two powerful tools in one. 1587/1577 Insulation Multimeters Technical Data Two powerful tools in one. The Fluke 1587 and 1577 Insulation Multimeters combine a digital insulation tester with a full-featured, true-rms digital multimeter

More information

162 CB CABLE TRACER. Filter Probe & Tone Generator INSTRUCTION MANUAL

162 CB CABLE TRACER. Filter Probe & Tone Generator INSTRUCTION MANUAL 162 CB CABLE TRACER Filter Probe & Tone Generator INSTRUCTION MANUAL INDEX PAGE 1. INTRODUCTION... 1 2. FILTER PROBE... 1-3 3. TONE GENERATOR... 3-6 4. SPECIFICATION... 7-8 5. MAINTENANCE... 8 1. INTRODUCTION

More information

Essential Electrical Concepts

Essential Electrical Concepts Essential Electrical Concepts Introduction Modern vehicles incorporate many electrical and electronic components and systems: Audio Lights Navigation Engine control Transmission control Braking and traction

More information

Model UT60A: OPERATING MANUAL. Table of Contents. Page 3 4 5 6 8 9 10 11 14 17 17 17 19 19

Model UT60A: OPERATING MANUAL. Table of Contents. Page 3 4 5 6 8 9 10 11 14 17 17 17 19 19 Table of Contents Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons Display Symbols Measurement

More information

323/324/325. Calibration Manual. Clamp Meter

323/324/325. Calibration Manual. Clamp Meter 323/324/325 Clamp Meter Calibration Manual February 2013 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective

More information

Model UT61A/61B/61C/61D/61E: OPERATING MANUAL. Table of Contents

Model UT61A/61B/61C/61D/61E: OPERATING MANUAL. Table of Contents P/N:110401103671 Table of Contents Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons

More information

Sense Components. Sense monitor. Antenna assembly. External mounting kit Current sensors. Download the Sense app. Go to sense.com/app.

Sense Components. Sense monitor. Antenna assembly. External mounting kit Current sensors. Download the Sense app. Go to sense.com/app. Sense Components Sense monitor Antenna assembly External mounting kit Current sensors Download the Sense app. Go to sense.com/app Power cable Technical Specifications Sense is a home energy monitoring

More information

DC POWER SUPPLY ALIMENTATION C.C.

DC POWER SUPPLY ALIMENTATION C.C. DC POWER SUPPLY ALIMENTATION C.C. ISO-TECH IPS 303A 201-3424 ISO-TECH IPS 601A 201-3446 82IP-303A0ME E1 SAFETY TERMS AND SYMBOLS These terms may appear in this manual or on the product: WARNING. Warning

More information

Microwave Meter. Instruction Manual

Microwave Meter. Instruction Manual Microwave Meter 840046 Instruction Manual Microwave Meter 840046 Copyright 2009 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted

More information

72-7730 72-7732 Intelligent Digital Multimeters

72-7730 72-7732 Intelligent Digital Multimeters 72-7730 72-7732 Intelligent Digital Multimeters Model 72-7730/72-7732 OPERATING MANUAL TABLE OF CONTENTS CHAPTER TITLE PAGE 1. Before You Start Overview Inspection Safety Information Rules For Safe Operation

More information

Model 77 Series IV Digital Multimeter

Model 77 Series IV Digital Multimeter Model 77 Series IV Digital Multimeter Introduction XWWarning To avoid electric shock or injury, do not perform the performance tests or calibration adjustment procedures unless qualified to do so. The

More information

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 General Specifications...1 2-2 Electrical Specifications... 2 3. FRONT PANEL DESCRIPTION... 4 3-1

More information

Users Manual. ProcessMeter

Users Manual. ProcessMeter 787 ProcessMeter Users Manual April 1997, Rev.3, 12/01 1997, 1898, 2000, 2001 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. LIMITED

More information

What is a multimeter?

What is a multimeter? What is a multimeter? A multimeter is a devise used to measure voltage, resistance and current in electronics & electrical equipment It is also used to test continuity between to 2 points to verify if

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

Current Monitoring Kit QUICK START GUIDE

Current Monitoring Kit QUICK START GUIDE Current Monitoring Kit QUICK START GUIDE Pico Technology EL040 Current Monitoring Kit Quick Start Guide CONTENTS 1 Introduction... 1 2 Kit contents... 3 3 Installation... 3 3.1 Connecting the system together...

More information

1. SAFETY INFORMATION

1. SAFETY INFORMATION RS-232 Sound Level Meter 72-860A INSTRUCTION MANUAL www.tenma.com 1. SAFETY INFORMATION Read the following safety information carefully before attempting to operate or service the meter. Use the meter

More information

Table of Contents. Page No.

Table of Contents. Page No. Table of Contents Title Page No. INTRODUCTION... 2 SAFETY PRECAUTIONS/S... 2 SPECIFICATIONS... 3 CONTROLS AND INDICATORS... 8 TESTING PROCEDURES... 10 AUXILIARY FUNCTIONS... 17 BATTERY AND FUSE REPLACEMENT...

More information

Getting started with

Getting started with PART NO. CMA113 MADE IN CHINA 1. Measuring CAT II 2. Max. voltage 250V ~ 3. Max. current 71 Amp Getting started with Electricity consumption monitoring single phase for homes and some smaller light commercial

More information

Model UT71A/B OPERATING MANUAL

Model UT71A/B OPERATING MANUAL Model UT71A/B OPERATING MANUAL CHAPTER TITLE PAGE 1. 2. Before You Start Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols Getting Acquainted Turning

More information

INSTRUCTION MANUAL IPM 3600 3 Phase Power Analyser EN FR IT DE ES

INSTRUCTION MANUAL IPM 3600 3 Phase Power Analyser EN FR IT DE ES INSTRUCTION MANUAL IPM 3600 3 Phase Power Analyser EN FR IT DE ES CONTENTS / EN CONTENTS Title Page I. SAFETY INFORMATION...1 II. INTRODUCTION...3 III. SPECIFICATIONS...4 3-1 Environmental Conditions...4

More information

INTRODUCTION SAFETY PRECAUTIONS/

INTRODUCTION SAFETY PRECAUTIONS/ Title Page No. INTRODUCTION..................................1 SAFTY PRCAUTIONS/WARNINGS.................1 SPCIFICATIONS................................3 CONTROLS AND INDICATORS......................6

More information

User s Manual Before using the inverter, you need to read and save the safety instructions.

User s Manual Before using the inverter, you need to read and save the safety instructions. User s Manual Before using the inverter, you need to read and save the safety instructions. STI SERIES (STI200, STI300, STI500, STI700, STI1000) Power Frequency Pure Sine Wave Inverter The information

More information

User s manual FLIR VP50/VP52 Non-contact AC voltage detector

User s manual FLIR VP50/VP52 Non-contact AC voltage detector User s manual FLIR VP50/VP52 Non-contact AC voltage detector User s manual FLIR VP50/VP52 #T559851; r. AD/ 9134/9134; en-us Table of contents 1 Disclaimers... 1 1.1 Copyright... 1 1.2 Quality assurance...

More information

The Basics of Digital Multimeters

The Basics of Digital Multimeters IDEAL INDUSTRIES INC. The Basics of Digital Multimeters A guide to help you understand the basic Features and Functions of a Digital Multimeter. Author: Patrick C Elliott Field Sales Engineer IDEAL Industries,

More information

SOUND LEVEL METER 94405

SOUND LEVEL METER 94405 SOUND LEVEL METER 94405 INSTRUCTION MANUAL REC FAST A/C 1. SAFETY INFORMATION Read the following safety information carefully before attempting to operate or service the meter. Use the meter only as specified

More information

DIGITAL MULTIMETER OPERATION MANUAL ITEM 98674 DIGITAL MULTIMETER CONTENTS

DIGITAL MULTIMETER OPERATION MANUAL ITEM 98674 DIGITAL MULTIMETER CONTENTS CONTENTS ITEM 98674 OPERATION MANUAL CONTENTS 1. SAFETY INFORMATION 1 1.1 PRELIMINARY 1 1.2 DOS AND DON TS 2 1.3 SYMBOLS 3 1.4 MAINTENANCE 4 2. DESCRIPTION 6 2.1 NAMES OF PARTS 7 2.2 SWITCH, BUTTONS AND

More information

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and

More information

EKOS Cart. Instructions for Use

EKOS Cart. Instructions for Use EKOS Cart Instructions for Use EKOS Corporation 11911 North Creek Parkway South Bothell, WA 98011 USA (425) 415-3100 (tel) (425) 415-3102 (fax) info@ekoscorp.com (e-mail) - 1-4913-002 REV E Intended Use

More information

Drayton Digistat +2RF/+3RF

Drayton Digistat +2RF/+3RF /+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer

More information

Achat 115MA full-range speaker. user manual

Achat 115MA full-range speaker. user manual Achat 115MA full-range speaker user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de

More information

Thermistor Thermometer MODEL NO. 93210-00

Thermistor Thermometer MODEL NO. 93210-00 Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:

More information

Wiser Panel Meter, Model Number WISERCTPM200 Installer s Guide

Wiser Panel Meter, Model Number WISERCTPM200 Installer s Guide Instruction Bulletin EAV85226 08/2014 Wiser Panel Meter, Model Number WISERCTPM200 Installer s Guide Retain for future use. Product Description Kit Contents The Wiser Panel Meter is for use in energy management

More information

Thermocouple Thermometer Models

Thermocouple Thermometer Models Technology Made Easy... Thermocouple Thermometer Models 60010-00 (Type J) 60010-10 (Type K) 60010-20 (Type T) 68X309905 Rev 1 04/04 TABLE OF CONTENTS 1. INTRODUCTION 4 2. SAFETY PRECAUTIONS 6 3. SPECIFICATIONS

More information

TIG INVERTER INSTRUCTION MANUAL

TIG INVERTER INSTRUCTION MANUAL TIG INVERTER INSTRUCTION MANUAL Contents Warning General Description Block Diagram Main Parameters Circuit Diagram Installation and Operation Caution Maintenance Spare Parts List Troubleshooting 3 4 4

More information

ABCs of DMMs Multimeter features and functions explained Application Note

ABCs of DMMs Multimeter features and functions explained Application Note ABCs of DMMs Multimeter features and functions explained Application Note Digital multimeters offer a wide selection of features. Choosing the right meter for the job can be challenging unless you know

More information

Fluke 3000 FC Series Test Tools Technical Data

Fluke 3000 FC Series Test Tools Technical Data Fluke 3000 FC Series Test Tools Technical Data Now compatible with Fluke Connect Mobile App Start building tomorrow s test tool system today with the Fluke 3000 FC Digital Multimeter. The new Fluke Connect

More information

Digital Clamp-on DMMs with uncompromising performance

Digital Clamp-on DMMs with uncompromising performance Digital Clamp-on DMMs with uncompromising performance With this new range of clamp-on meters, Metrix introduces practical and inexpensive instruments with uncompromising performance. Designed to satisfy

More information

Current Monitoring Kit

Current Monitoring Kit Current Monitoring Kit QUICK START GUIDE DO090-6 CONTENTS Issues: 1) 2.10.02 WP A4 format 2) 2.10.02 Added safety warning 3) 17.3.06 Word A5 format. S1: Removed relay modules. S2: Added MI010. S4.1: added

More information

INSTRUCTION MANUAL PLEASE READ ALL THE INSTRUCTIONS COMPLETELY BEFORE USE AND SAVE THIS MANUAL FOR FUTURE REFERENCE

INSTRUCTION MANUAL PLEASE READ ALL THE INSTRUCTIONS COMPLETELY BEFORE USE AND SAVE THIS MANUAL FOR FUTURE REFERENCE INSTRUCTION MANUAL PLEASE READ ALL THE INSTRUCTIONS COMPLETELY BEFORE USE Ver. 2.0 AND SAVE THIS MANUAL FOR FUTURE REFERENCE Table of Contents Unpacking... 3 About the CCRadio-EP... 4 Quick Start Guide...

More information

Model 1756 Test Lead Kit

Model 1756 Test Lead Kit Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Model 1756 Test Lead Kit Gerneral Purpose Test Lead Information Description These test leads allow you

More information

12-Volt 10-Amp Regulated Power Supply

12-Volt 10-Amp Regulated Power Supply 22-506.fm Page 1 Friday, August 6, 1999 12:55 PM Cat. No. 22-506 OWNER S MANUAL Please read before using this equipment. 12-Volt 10-Amp Regulated Power Supply 22-506.fm Page 2 Friday, August 6, 1999 12:55

More information

2XWSXWFKDQQHOV 3KDVH &+ CH 2 &+ CH 4 &+ CH 6 L1 X X L2 X X L3 X X

2XWSXWFKDQQHOV 3KDVH &+ CH 2 &+ CH 4 &+ CH 6 L1 X X L2 X X L3 X X INSTALLATION INSTRUCTIONS ENGLISH Version 1 1 January 2004 EUROLIGHT LD6230 IMPORTANT SAFETY INSTRUCTIONS DETAILED SAFETY INSTRUCTIONS: 1) Read these instructions 2) Keep these instructions 3) Heed all

More information

Multi Media Cable Tester TM-901N/TM-902/TM-903

Multi Media Cable Tester TM-901N/TM-902/TM-903 Multi Media Cable Tester TM-901N/TM-902/TM-903 User Manual HB2TM9020010 TENMARS ELECTRONICS CO., LTD Table of Contents 1 Introduction... 1 2 Safety Precaution... 1 3 The verifier does the following:...

More information

Multimeter measurements on variable frequency drives using the new Fluke 289 DMM

Multimeter measurements on variable frequency drives using the new Fluke 289 DMM Multimeter measurements on variable frequency drives using the new Fluke 289 DMM Application Note Editor s note: For similar instructions using the Fluke 87V DMM, reference Fluke article 12345. In the

More information

Operating instructions in the back. www.blackanddecker.co.uk PD1020L

Operating instructions in the back. www.blackanddecker.co.uk PD1020L Operating instructions in the back www.blackanddecker.co.uk PD1020L 2 ENGLISH (Original instructions) Intended use Your Black & Decker Dustbuster handheld vacuum cleaner has been designed for vacuum cleaning

More information

Product and functional description

Product and functional description Product and functional description The KNX / DALI gateway N 141/02 is a 4 MU wide, DINrail mounted KNX device with one DALI interface to which up to 64 DALI actuators (e.g. DALI ballasts) can be connected

More information

MODEL 2202IQ (1991-MSRP $549.00)

MODEL 2202IQ (1991-MSRP $549.00) F O R T H E L O V E O F M U S I C F O R T H E L O V E O F M U S I C MODEL 2202IQ (1991-MSRP $549.00) OWNER'S MANUAL AND INSTALLATION GUIDE INTRODUCTION Congratulations on your decision to purchase a LINEAR

More information

MEGGER 210170 Insulation & Low Resistance Tester

MEGGER 210170 Insulation & Low Resistance Tester MEGGER 210170 Insulation & Low Resistance Tester User Guide SAFETY WARNINGS Safety Warnings and Precautions must be read and understood before the instrument is used. They must be observed during use.

More information

Table of Contents. Title

Table of Contents. Title Table of Contents Title Page No. INTRODUCTION... 2 SAFETY PRECAUTIONS/WARNINGS... 2 SPECIFICATIONS... 3 CONTROLS AND INDICATORS... 8 TESTING PROCEDURES... 10 AUXILIARY FUNCTIONS... 17 BATTERY AND FUSE

More information

Digital Photo Picture Frame. Operation Manual

Digital Photo Picture Frame. Operation Manual Digital Photo Picture Frame Operation Manual 20070309 CONGRATULATIONS on your purchase of a Polaroid 7 LCD Digital Photo Picture Frame. Please read carefully and follow all warnings and instructions in

More information

How To Use A Power Supply Unit (Upu)

How To Use A Power Supply Unit (Upu) BRAVER UPS (Uninterruptible Power System) User s Manual Safety CAUTION! This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user replaceable parts.

More information

3000W Power Inverter

3000W Power Inverter 3000W Power Inverter OWNER S MANUAL Model number-4573000 TO REDUCE THE RISK OF INJURY, USER MUST READ AND UNDERSTAND THIS INSTRUCTIONAL MANUAL. THIS MANUAL CONTAINS IMPORTANT INFORMATION REGARDING THE

More information

Integrating Sound Level Meter and Datalogger

Integrating Sound Level Meter and Datalogger USER GUIDE Integrating Sound Level Meter and Datalogger Model 407780A Introduction Thank you for selecting the Extech Instruments Model 407780A. This device is shipped fully tested and calibrated and,

More information

STEREO PREAMPLIFIER INSTRUCTIONS FOR USE

STEREO PREAMPLIFIER INSTRUCTIONS FOR USE XX STEREO PREAMPLIFIER INSTRUCTIONS FOR USE Thank you for purchasing the Musical Fidelity A5 CR Preamplifier. Used properly and carefully, it should give many years of outstanding musical reproduction.

More information

FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION

FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION Caution! UV-radiation of this device is in the range of UV-A (320-400 nm). Direct exposure to eyes shall therefore be avoided. UV protection glasses shall

More information

Hygro-Thermometer + InfraRed Thermometer Model RH101

Hygro-Thermometer + InfraRed Thermometer Model RH101 User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative

More information

DX-395. FM/AM/SW1/SW2 PLL Synthesized Receiver OWNER S MANUAL. Cat. No. 20-225. Please read before using this equipment.

DX-395. FM/AM/SW1/SW2 PLL Synthesized Receiver OWNER S MANUAL. Cat. No. 20-225. Please read before using this equipment. 20-225.fm Page 1 Wednesday, August 4, 1999 10:07 AM Cat. No. 20-225 OWNER S MANUAL Please read before using this equipment. DX-395 FM/AM/SW1/SW2 PLL Synthesized Receiver 20-225.fm Page 2 Wednesday, August

More information

Pocket Tach 99 (PT99) Non-Contact Tachometer

Pocket Tach 99 (PT99) Non-Contact Tachometer CE DECLARATION OF CONFORMITY As Manufacturer: Monarch Instrument Division of Monarch International Inc. 15 Columbia Drive, Amherst NH 03031 USA declares under Monarch s sole responsibility that the product:

More information

HOW TO USE MULTIMETER. COMPILE BY: Dzulautotech

HOW TO USE MULTIMETER. COMPILE BY: Dzulautotech HOW TO USE MULTIMETER COMPILE BY: Dzulautotech 1. GENERAL Electricity is absolutely necessary for an automobile. It is indispensable when the engine is started, the air fuel mixture is ignited and exploded,

More information

************* OWNER'S MANUAL BAMF800/2 BAMF1250/2 BAMF1800/2 BAMF2200/2 BAMF2600/2 BAMF1200/4 BAMF1600/4 BAMF2000/1D BAMF4000/1D BAMF5500/1D

************* OWNER'S MANUAL BAMF800/2 BAMF1250/2 BAMF1800/2 BAMF2200/2 BAMF2600/2 BAMF1200/4 BAMF1600/4 BAMF2000/1D BAMF4000/1D BAMF5500/1D ************* OWNER'S MANUAL BAMF800/2 BAMF1250/2 BAMF1800/2 BAMF2200/2 BAMF2600/2 BAMF1200/4 BAMF1600/4 BAMF2000/1D BAMF4000/1D BAMF5500/1D INTRODUCTION Power Acoustik amplifiers provide high-performance

More information

EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM

EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM INSTALLATION MANUAL Protec Fire Detection PLC, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT. Telephone: +44 (0) 1282 717171 Fax: +44 (0) 1282 717273

More information

115C and 117C. True-rms Multimeters. Users Manual

115C and 117C. True-rms Multimeters. Users Manual 115C and 117C True-rms Multimeters Users Manual PN 2538695 July 2006, Rev. 1, 2/07 2006, 2007 Fluke Corporation, All rights reserved. Printed in China All product names are trademarks of their respective

More information

Power supply unit Model: PAC-SC50KUA

Power supply unit Model: PAC-SC50KUA Building Air Conditioning Control System Power supply unit Model: PAC-SC50KUA Installation Manual Contents 1. Safety precautions... 1 2. Product feature... 2 1. Specification... 2 2. Appearance... 2 3.

More information

User's Guide. Pinless Moisture Meter. Model MO257

User's Guide. Pinless Moisture Meter. Model MO257 User's Guide Pinless Moisture Meter Model MO257 Introduction Congratulations on your purchase of the Extech MO257 Pinless Moisture Meter. The pinless moisture sensor monitors the moisture in wood and other

More information

User s Manual AURORA 1.2K/2.2K

User s Manual AURORA 1.2K/2.2K User s Manual AURORA 1.2K/2.2K Uninterruptible Power System Safety CAUTION This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user serviceable

More information

DLP-PU/E Instruction Manual

DLP-PU/E Instruction Manual Instruction Manual BEFORE USING THE POWER SUPPLY UNIT Pay attention to all warnings and cautions before using the unit. Incorrect usage could lead to an electrical shock, damage to the unit or a fire hazard.

More information

12 Volt 30 Amp Digital Solar Charge Controller

12 Volt 30 Amp Digital Solar Charge Controller 12 Volt 30 Amp Digital Solar Charge Controller User s Manual WARNING Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions

More information

Digital Ground Resistance Tester. Models 4620 & 4630. Digital Ground Resistance Tester

Digital Ground Resistance Tester. Models 4620 & 4630. Digital Ground Resistance Tester Digital Ground Resistance Tester Models 4620 & 4630 Digital Ground Resistance Tester Models 4620 and 4630 perform both ground resistance and soil resistivity tests. These testers measure from 0 to 2000Ω,

More information

CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN

CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN BeoLab 7-6 Guide CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN CAUTION: To reduce the risk of electric shock, do not remove cover (or back). No User-serviceable parts inside. Refer servicing to qualified

More information

RS DIGITAL THERMOMETER

RS DIGITAL THERMOMETER RS DIGITAL THERMOMETER 1319A INSTRUCTION MANUAL C F MX MN AVG TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION... 1 2. SPECIFICATIONS... 1 3. NAME OF PARTS AND POSITIONS... 3 4. OPERATION INSTRUCTIONS... 5

More information