Method Validation of HPLC

Size: px
Start display at page:

Download "Method Validation of HPLC"

Transcription

1 Method Validation of HPLC Customer Support Centre Shimadzu (Asia Pacific) Pte Ltd 1 Materials required to establish an analytical method Analytical Methods Analytical Procedure Controlling procedure of standard Controlling procedure of test sample Explanations of the instruments and apparatus System suitability test Calculating formula and measurement number Specification of standards Materials of validation Test methods and evaluation methods to get results Analytical results Calculating results 2

2 Consideration in method development Separation Mobile phase Select of buffer solution ph Buffer concentration Select of organic solvent polarity Detection does not interrupt ETC Ionpair methods Derivatized methods Column Small subsidiary work High theoretical plate number Small dispersion for the lot Detector Sensitivity Selectivity Applied with SPDM10A VP 3 DOCUMENTS AVAILABLE ON VALIDATION OF ANALYTICAL PROCEDURES ICH Text on Validation of Analytical Procedures (October 1994) Extension of the ICH Text "Validation of Analytical Procedures (November 1994) FDA(CDER) Validation of Chromatographic Methods (November 1994) USP23 <1225> Validation of Compendial Methods Others Pharmaceutical Technology, LCGC, etc.

3 Evaluation of Parameters for Analytical Capability Parameters for analytical capability are evaluated by analytical results using statistical method. Estimation and test Regression analysis and Variance analysis 5 VALIDATION CHARACTERISTICS Validation of an analytical procedure is the process by which it is established that the validation characteristics of the analytical procedure meet proper standards of validation characteristics. [Typical Validation Characteristics] Accuracy Precision Repeatability Intermediate Precision Specificity Detection Limit Quantitation Limit Linearity Range (ICH Q2A,October 1994)

4 VALIDATION CHARACTERISTICS The table lists the validation characteristics regarded as the most important for the validation of different types of analytical procedures. IDENTIFICATION TESTING FOR IMPURITIES ASSAY characteristics quantitat. limit Accuracy Precision Repeatability Interm.Precision Specificity Detection Limit Quantitation Limit Linearity Range signifies that this characteristic is not normally evaluated signifies that this characteristic is normally evaluated Validation example(1) Test Confirmation test for acetylsalicylic acid in acetylsalicylic acid raw drug using HPLC. Analytical capability parameter to evaluate Specificity Data for evaluation Acetylsalicylic acid drug and standard of their impurity compound are analyzed. And, acetylsalicylic acid drug is analyzed on hard conditions. Get the chromatograms and spectrum of each peaks 8

5 SPECIFICITY Specificity is the ability to assess unequivocally the analyte in the presence of components which may be expected to be present. Typically these might include impurities, degradant, matrix, etc. Lack of specificity of an individual analytical procedure may be compensated by other supporting analytical procedure(s). [Determination] Specificity can be determined for instance by spiking pure substances with excipients and/or impurities and/or degradation products and to compare the test results with those of pure substances. Specificity test for acetylsalicylic acid(1) Acetylsalicylic acid phydroxybenzoic acid Salicylic acid A B (min) Resolution between target and impurity is more than 1.5 and each impurity resolution is more than 1.2. Spectrum of drug substance(b) is similar to one of standard acetylsalicylic acid(a). 10

6 Specificity test for acetylsalicylic acid(2) Acetylsalicylic Acid Salicylic Acid A B C D min Chromatogaram of raw drug A: 10mMHCl 25 darkness 1hr B: 10mMNaOH 25 darkness 1hr C: White light5000 Neutral buffer 25 1hr D: 80 Neutral buffer darkness 1hr 11 Specificity test for acetylsalicylic acid(3) Method(peak purity) The spectrum of upslope, peaktop, downslope correspond each other. And theoretical plate number and symmetry factor are calculated by the chromatograms of different wavelength. Acetylsalicylic Acid 200 nm N=12804, Tf = nm N=12970, Tf = nm N=12924, Tf = nm N=12887, Tf = nm N=12924, Tf = min Chromatogarams of raw drug 12

7 Validation example(2) Test Quantitative testing for acetylsalicylic acid in their drug using HPLC Evaluation of Parameters Specificity Accuracy Repeatability Intermediate Precision Linearity Range Data 5 concentration level sample are analyzed 5 times. It is analyzed on different date, operator, instrument, column and reagent. Conc. 80 % 90 % 100 % 110 % 120 % 1st run 79.2 % 90.8 % 99.3 % % % 2nd run 79.7 % 90.2 % % % % 3rd run 80.6 % 89.7 % % % % 4th run 80.0 % 89.3 % 99.1 % % % 5th run 80.4 % 90.5 % 99.9 % % % Test 1 Test 2 1st run 99.3 % 99.6 % 2nd run % 98.9 % 3rd run % % 4th run 99.1 % 99.7 % 5th run 99.9 % 99.2 % 13 LINEARITY The linearity of an analytical procedure is its ability (within a given range) to obtain test results which are directly proportional to the concentration (amount) of analyte in the sample. For those analytical procedures which are not linear, another mathematical relationship (proportionality) must be demonstrated. [Determination] The linearity is determined by calculating a regression line by the method of least squares of test results versus analyte concentrations. Five points in the range of standard value±20% for an assay method and in the range from 20% of target concentration down to the limit of quantitation of the drug substance or impurity for an assay/impurities combination method based on area % (for impurities) are recommended. It is often combined with accuracy, precision and range, and carried out as a single study. Under most circumstances, regression coefficient (r) is Intercept and slope should be indicated.

8 Linearity test for acetylsalicylic acid Conc. 80% 90% 100% 110% 120% 1st run nd run rd run th run th run % 20% 40% 60% 80% 100% 120% 140% Evaluation of linearity for acetylsalicylic acid Regression coefficient :r = Slope : b = Σ{(x i Σx i /n) (y i Σy i /n)} {[Σ(x i Σx i /n)][σ(y i Σy i /n) 2 ]} 1/2 Σ{(x i Σx i /n) (y i Σy i /n)} Σ(x i Σx i /n) 2 Intercept : a = Σy i /n bσx i /n r = , a = 0.124, b = More than of regression coefficient is obtained. 16

9 ACCURACY The accuracy of an analytical procedure expresses the closeness of agreement between the value which is accepted either as a conventional true value or an accepted reference value and the value found. [Determination] Accuracy studies are performed, at least in triplicate, by the addition of known amounts of drug (80, 100 and 120% levels of label claim) to the placebo formulation working in the linear range of detection of the analyte. The mean of the replicates, expressed as % label claim, is an estimate of accuracy. It is often combined with precision, linearity and range, and carried out as a single study. Accuracy Method After the results of 5 level concentration, 5 numbers repeat was normalized, it is calculated with the accuracy and 95% confidence interval of their variance by 1Dvariance analysis. Variance analysis Factor Square sum Freedom Square mean Fvalue Sample S I = I ν= 4 V I = F= Erroer S e = e ν= 20 V e = Total S T = T ν = 24 S I = Σ(Σy i) 2 j /m (ΣΣy ij) 2 /n I ν= l 1 V I = S I/ν I F= V I/V e S e = S T S I e ν= T ν ν I V e = S e/ν e S T = ΣΣy 2 ij (ΣΣy ij) 2 /n T ν = lm 1 Trueness(each step) dj = Σ{yji(100/xj)}/m 100 d80 = d90 = d100 = d110 = d120 = Trueness( Total) d = ΣΣ{yij(100/xj)}/n 100 d = Confidence interval d 1.96(VI /n) 1/2 D d1.96(vi /n) 1/ D ] or d t0.025(l 1)(VI /n) 1/2 D dt0.025(l 1)(VI /n) 1/ D ( l = 5: step number m = 5: repeat number n = lm = 25: total number of analysis) It does not deviate as 0 is contained in confidence interval of accuracy. 18

10 PRECISION The precision of an analytical procedure expresses the closeness of agreement (degree of scatter) between a series of measurements obtained from multiple sampling of the same homogeneous sample under the prescribed conditions. Precision may be considered at three levels: Repeatability Intermediate Precision Reproducibility Reproducibility is not normally expected if intermediate precision is accomplished. Repeatability Repeatability expresses the precision under the same operating conditions over a short interval of time. Repeatability is also termed intraassay precision. Repeatability is considered at two levels: Injection Repeatability and Analysis Repeatability. Injection Repeatability Injection repeatability, expressed as the RSD, is determined by multiple injections of a homogeneous sample (prepared solution) under the analytical conditions. A minimum of 10 injections with an RSD of 1% is recommended. For low level impurities, higher variations may be acceptable. Analysis Repeatability Analysis repeatability, expressed as the RSD, is determined by multiple measurements of a sample by the same analyst under the same analytical conditions. It is often combined with accuracy, linearity and range, and carried out as a single study.

11 SCL10A::::::: SIL10A:::::: SAMPLE COOLER LC10AD ::::::: SPD10A ::::::: CTO10A:::::: LC10AD ::::::: SCL10A::::::: SIL10A:::::: LC10AD ::::::: SPD10A ::::::: CTO10A:::::: Repeatability Method After the results of 5 level concentration, 5 numbers repeat was normalized, it is calculated with the accuracy and 95% confidence interval of their variance by 1Dvariance analysis. Standard deviation( each level) sj = {Σ(yji Σyji/m) 2 /(m 1)} 1/2 Variance analysis Factor Square sum Freedom Square mean F value Sample SI = νi = 4 VI = F = Error Se = νe = 20 Ve = Total ST = νt = 24 SI = Σ(Σyi)j 2 /m (ΣΣyij) 2 /n νi = l 1 VI = SI/νI F = VI/Ve Se = ST SI νe = νt νi Ve = Se/νe ST = ΣΣyij 2 (ΣΣyij) 2 /n νt = lm 1 s80 = s90 = s100 = s110 = s120 = Relative standard deviation( each level) RSDj = sj/(σyji/m) RSD80 = 0.70 (%) RSD90 = 0.67 (%) RSD100 = 0.72 (%) RSD110 = 0.69 (%) RSD120 = 0.63 (%) Standrd deviation( total) s = Ve 1/2 s = Confidence interval Se /χ (n l) σ 2 Se /χ (n l) σ ( l = 5: level, m = 5: repeat n = lm = 25: total number of analysis) 21 Intermediate precision Intermediate precision expresses withinlaboratories variations: different days, different analysis, different equipment. [Determination] Intermediate precision, expressed as the RSD, is determined by multiple measurements, as a minimum, for two separate occasions. It is often combined with accuracy, linearity and range, and carried out as a single study.

12 Intermediate precision Method To use two analytical results, the difference of population mean and 95% confidence interval of the ratio of population variance. mean( each test) Avej = Σyji/mj Ave1 = Ave2 = standard deviation( each test) sj = {Σ(yji Σyji/mj) 2 /(mj 1)} 1/2 s1 = s2 = relative standard deviation( each test) RSDj = sj/(σyji/mj) RSD1 = 0.72 (%) RSD2 = 0.61 (%) ( m = 5: Repeat number of Test1 n = 5: Repeat number of Test2) Degree of freedom ν = (s1 2 /m1s2 2 /m2) 2 /{(s1 2 /m1) 2 /(m1 1)(s2 2 /m2) 2 /(m2 1)} ν = 4.0 ν* = 4 Difference of population mean d = Σy1i/m1 Σy2i/m2 d = 0.32 Confidence of difference of population mean d t0.025(ν*)(s1 2 /m1s2 2 /m2) 1/2 D d t0.025(ν*)(s1 2 /m1s2 2 /m2) 1/ D Variance ratio F = s1 2 /s2 2 F = Confidence interval of variance ratio F0.975(m1 1, m2 1)s2 2 /s1 2 F F0.025(m1 1, m2 1)s2 2 /s F It does not signify as 0 is contained in the confidence interval of difference of population mean and 1 is contained in the confidence in the ratio of population variance. 23 Validation example(3) Test Limitation test of impurities in acetylsalicylic acid drug using HPLC Evaluation of Parameters Specificity Limit of detection Data 5 level (lower) concentration sample, 5 number of repeat Conc % 0.06 % 0.08 % 0.10 % 1.20 % 1st run nd run rd run th run th run

13 LIMIT OF DETECTION The limit of detection of an individual analytical procedure is the lowest amount of analyte in a sample which can be detected but not necessarily quantitated as an exact value. [Determination] The limit of detection may be determined from the signaltonoise ratio (S/N) by comparing test results from samples with known concentrations of analyte with those of blank samples and by establishing the minimum concentration at which the analyte can be reliably detected. S/N 3:1 is generally accepted. Standard deviation and slope of calibration curve method Limit of detection Method The slope for regression curve and standard deviation were calculated by regression analysis using the results of 5 concentration,5times repeats. Then the limit of quantitation was calculated from these results. SD of redisual sy/x = {Σ{yi (bxia)} 2 /(n 2)} 1/2 sy/x = SD of y intercept sy = sy/x{(σxi 2 /nσ(xi Σxi/n) 2 } 1/2 sy = or SD of y intercept sy = sy/x{11/n(σxi/n) 2 /Σ(xi Σxi/n) 2 } 1/2 sy = Limit of detection DL = 3.3s y/x/a DL = (%) ( From SD of redisual) or DL = (%) ( From SD of y intercept) or DL = (%) ( From SD of y intercept) ( a = 76182: Slope of regression curve b = : y intercept n = 25: total number of analysis) 26

14 Signal to Noise ratio Limit of detection : S/N= 3 Limit of quantitation : S/N=10 Signal to Noise ratio 50ppm : S/N=200mm/1mm=200? ppm : S/N=2 Concentration=50 ppm (N = 1mm) (S = 200mm)? =0.5ppm

15 Standard deviation and slope of calibration curve method Limit of detection : 3.3 σ/s Limit of quantitation : 10 σ/s σ : standard deviation of response signal S : slope value of calibration curve σ slope [resp.] [conc.] Limit of Detection test for acetylsalicylic acid Conc. 0.04% 0.06% 0.08% 0.10% 1.20% 1st run nd run rd run th run th run s y/x = {Σ(y i (bx i a)} 2 /(n2)} 1/2 = a = : slop of regression line b = : y intercept n = 25 : total run number DL = 3.3s y/x /a = 0.011(%) 30

16 Validation example(4) Test Quantitative test of salicylic acid which is impurity, in acetylsalicylic acid raw drug using HPLC Evaluation of Parameters Data Specificity Accuracy Repeatability Intermediate precision Limit of quantitation Linearity Range 5 concentration level(50%~120%) sample, 5 times repeat. Conc 0.04 % 0.06 % 0.08 % 0.10 % 1.20 % 1st run nd run rd run th run th run LIMIT OF QUANTITATION The limit of quantitation of an individual analytical procedure is the lowest amount of analyte in a sample which can be quantitatively determined with suitable precision and accuracy. [Determination] The limit of quantitation may be determined from the signaltonoise ratio (S/N) by comparing test results from samples with known concentrations of analyte with those of blank samples and by establishing the minimum concentration at which the analyte can be detected with suitable precision and accuracy. S/N 10:1 is generally accepted.

17 Limit of quantitation Method The slope for regression curve and standard deviation were calculated by regression analysis using the results of 5 concentration,5times repeats. Then the limit of quantitation was calculated from these results. Standard deviation sy/x = {Σ{yi (bxia)} 2 /(n 2)} 1/2 Quantitative limit QL = 10 sy/x/a sy/x = DL = (%) ( From SD of redisual) or Standard deviation of y intercept sy = sy/x{(σxi 2 /nσ(xi Σxi/n) 2 } 1/2 DL = (%) ( From SD of y intercept) sy = or Standard deviation of y intercept sy = sy/x{11/n(σxi/n) 2 /Σ(xi Σxi/n) 2 } 1/2 or DL = (%) ( From SD of y intercept) sy = ( a = 53161: Slope of regression line b = : y intercept n = 25: Total number of analysis) 33 Quantitation Limit test for salicylic acid as a impurity Conc. 0.04% 0.06% 0.08% 0.10% 1.20% 1st run nd run rd run th run th run s y/x = {Σ(y i (bx i a)} 2 /(n2)} 1/2 = a = : slop of regression line b = : y intercept n = 25 : total run number DL = 10s y/x /a = 0.041(%) 34

18 RANGE The range of an analytical procedure is the interval between the upper and lower concentration (amounts) of analyte in the sample (including these concentrations) for which it has been demonstrated that the analytical procedure has a suitable level of precision, accuracy and linearity. [Determination] The range is validated by verifying that the analytical procedure provides acceptable precision, accuracy and linearity when applied to samples containing analyte at the extremes of the range as well as within the range. The recommended range is NLT±20% for an assay method and is20% of target concentration down to the limit of quantitation of the drug substance or impurity for an assay/impurities combination method based on area % (for imputies). It is often combined with accuracy, precision and linearity, and carried out as a single study. REFERRENCE STANDARDS A reference standard is a highly purified compound that is well characterized. USP/NF reference standard that does not need characterization. Noncompendial standard that should be of the highest purity that can be obtained by reasonable effort and should be thoroughly characterized to assure its identity, strength, quality and purity. The working concentration is the target concentration of the compound of interest. Keeping the concentrations of the sample and the standard close to each other for the external standard method improves the accuracy of the method. Include the purity correction factor, if known, of the reference standard in the calculation. State the working concentrations of the standard and sample in the method.

19 OTHER VALIDATION CHARACTERISTICS Recovery Recovery is expressed as the amount/weight of the compound of interest analyzed as a percentage to the theoretical amount present in the medium. Accuracy may be expressed as percent recovery by the assay of known, added amounts of analyte. Sample Solution Stability Solution stability of the drug substance or drug product after preparation according to the test method should be evaluated. Data to support the sample solution stability under normal laboratory conditions for the duration of the test procedure, e.g., 24 hours, should be generated. Robustness The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage. Testing varying some or all conditions, e.g., age of columns, column type, column temperature, ph of buffer in mobile phase, reagents, is normally performed. SYSTEM SUITABILITY TESTS System suitability is an integral part of many analytical procedures. The tests are based on the concept, that the equipment, electronics, analytical operations and samples to be analyzed constitute an integral system that can be evaluated as such. [ System Suitability Parameters ] [ System Suitability Parameters ] Capacity Factor ( k' ) Precision/Injection Repeatability ( RSD ) Relative Retention (α) Resolution ( Rs ) Tailing Factor ( T ) Theoretical Plate Number ( N )

20 SYSTEM SUITABILITY PARAMETERS [Determination and Standards of Parameters] Capacity Factors ( k' ) k' = (t R t 0 ) /t 0 k'>2 Precision/Injection Repeatability ( RSD ) n 5 RSD 1% Relative Retention (α ) t α = k' 1 / k' 0 2 There is not an essential parameter as long as the resolution( Rs ) is stated. Resolution ( Rs ) Rs = (t R2 t R1 ) / (1/2)(t W1 t W2 ) Rs 2 Tailing Factor ( T ) T = Wx / 2f T 2 Theoretical Plate Number ( N ) N =16(t R /t W ) 2 = L/H N>2000 Injection Air peak t R1 Solvent peak f Solvent tail t R2 Wx t W1 t W2 0.05h h Thank you 40

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in

More information

Applying Statistics Recommended by Regulatory Documents

Applying Statistics Recommended by Regulatory Documents Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

Design of Experiments for Analytical Method Development and Validation

Design of Experiments for Analytical Method Development and Validation Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories

More information

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND

More information

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this

More information

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.

More information

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 Simultaneous determination of Telmisartan

More information

Validating Methods using Waters Empower TM 2 Method. Validation. Manager

Validating Methods using Waters Empower TM 2 Method. Validation. Manager Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,

More information

To avoid potential inspection

To avoid potential inspection Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

Assay Development and Method Validation Essentials

Assay Development and Method Validation Essentials Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

ERDOSTEINE - MONOGRAPH.

ERDOSTEINE - MONOGRAPH. STRUCTURAL FORMULA (±1S-(2-[N-3-(2-oxotetrahydro thienyl)]acetamido)-thioglycolic acid) C 8 H 11 NO 4 S 2 M.W. = 249.307 DESCRIPTION Color : White to ivory white Appearance : Microcrystalline powder SOLUBILITY

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number

More information

Implementing New USP Chapters for Analytical Method Validation

Implementing New USP Chapters for Analytical Method Validation Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: Ludwig_Huber@labcompliance.com Today s Agenda Handling Method Changes vs. Adjustments

More information

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction

More information

Regression Analysis: A Complete Example

Regression Analysis: A Complete Example Regression Analysis: A Complete Example This section works out an example that includes all the topics we have discussed so far in this chapter. A complete example of regression analysis. PhotoDisc, Inc./Getty

More information

Method Validation. Introduction. Página Web 1 de 16

Method Validation. Introduction. Página Web 1 de 16 Página Web 1 de 16 Method Validation Introduction Strategy for Validation of Methods Validation of Standard Methods Revalidation Parameters for Method Validation Selectivity/specificity Precision and Reproducibility

More information

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,

More information

Center for Drug Evaluation and Research (CDER) Reviewer Guidance' Validation of Chromatographic Methods. November 1994 CMC 3

Center for Drug Evaluation and Research (CDER) Reviewer Guidance' Validation of Chromatographic Methods. November 1994 CMC 3 Center for Drug Evaluation and Research (CDER) Reviewer Guidance' Validation of Chromatographic Methods November 1994 CMC 3 TABLE OF CONTENTS I. INTRODUCTION... 1 II. TYPES OF CHROMATOGRAPHY... 2 A. High

More information

Qualification Study CHO 360-HCP ELISA (Type A to D)

Qualification Study CHO 360-HCP ELISA (Type A to D) Short Report Qualification Study CHO 360-HCP ELISA (Type A to D) Presented by: http://www.biogenes.de Version 01 Issue date: 27.11.2013 Version 01 Page 1 of 8 Table of Content TABLE OF CONTENT... 2 1 INTRODUCTION...

More information

A practical and novel standard addition strategy to screen. pharmacodynamic components in traditional Chinese medicine using

A practical and novel standard addition strategy to screen. pharmacodynamic components in traditional Chinese medicine using Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2015 A practical and novel standard addition strategy to screen pharmacodynamic components in traditional

More information

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

PVV Satyanaryana et al., IJSID, 2012, 2 (1), 226-231. International Journal of Science Innovations and Discoveries

PVV Satyanaryana et al., IJSID, 2012, 2 (1), 226-231. International Journal of Science Innovations and Discoveries ISSN:2249-5347 IJSID International Journal of Science Innovations and Discoveries An International peer Review Journal for Science Research Article Available online through www.ijsidonline.info RP-HPLC

More information

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified

More information

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA

More information

QbD Approach to Assay Development and Method Validation

QbD Approach to Assay Development and Method Validation QbD Approach to Assay Development and Method Validation Thomas A. Little Ph.D. 11/105/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

Analytical Methods for Cleaning Validation

Analytical Methods for Cleaning Validation Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2011, 3 (6): 232-239 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4

More information

High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form

High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form ISSN: 0973-4945; CODEN ECJHAO E- Chemistry http://www.e-journals.net Vol. 5, No.3, pp. 427-430, July 2008 High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage

More information

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for

More information

A commitment to quality and continuous improvement

A commitment to quality and continuous improvement Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous

More information

Revision of The Dissolution Procedure: Development and Validation 1092

Revision of The Dissolution Procedure: Development and Validation 1092 Page 1 of 5 STIMULI TO THE REVISION PROCESS Stimuli articles do not necessarily reflect the policies of the USPC or the USP Council of Experts Revision of The Dissolution Procedure: Development and Validation

More information

VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI

VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI Fakultas Farmasi Universitas Pancasila Jln. Srengseng Sawah,

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 204 3R HANDLING AND USE OF REFERENCE STANDARDS IN THE OMCL NETWORK Full document title and reference Document type Legislative

More information

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs Agilent Application Solution Transfer of a USP method for prednisolone from normal phase PLC to SFC using the Agilent 126 Infinity ybrid SFC/UPLC System Saving time and costs Application Note Pharmaceutical

More information

How To Test For Contamination In Large Volume Water

How To Test For Contamination In Large Volume Water Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

Introduction. Methods. Sample Processing

Introduction. Methods. Sample Processing PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko

More information

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

Alignment and Preprocessing for Data Analysis

Alignment and Preprocessing for Data Analysis Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Performance and advantages of qnmr measurements

Performance and advantages of qnmr measurements Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended

More information

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

Analytical Procedures and Methods Validation for Drugs and Biologics

Analytical Procedures and Methods Validation for Drugs and Biologics Analytical Procedures and Methods Validation for Drugs and Biologics Guidance for Industry U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research

More information

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

Uses of Derivative Spectroscopy

Uses of Derivative Spectroscopy Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Analysis of Various Vitamins in Multivitamin Tablets

Analysis of Various Vitamins in Multivitamin Tablets High Performance Liquid Chromatography Analysis of Various Vitamins in Multivitamin Tablets Effective August 24, 2007, the US Food and Drug Administration (FDA) issued a regulation (21 CFR Part 111) that

More information

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

IMPURITIES IN NEW DRUG PRODUCTS

IMPURITIES IN NEW DRUG PRODUCTS INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE IMPURITIES IN NEW DRUG PRODUCTS Q3B(R2) Current

More information

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection 52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Analytical Procedures and Methods Validation Chemistry, Manufacturing, and Controls Documentation DRAFT GUIDANCE This guidance document is being distributed for comment purposes only.

More information

The Theory of HPLC. Gradient HPLC

The Theory of HPLC. Gradient HPLC The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims

More information

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

Shimadzu Simulated Distillation Gas Chromatograph System C184-E030

Shimadzu Simulated Distillation Gas Chromatograph System C184-E030 Shimadzu Simulated Distillation Gas Chromatograph System C184-E030 Meets All Simulated Distillation Gas Chromatograph Standards - Provides Highly Accurate Analysis Results with Excellent Reproducibility-

More information

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification White paper System Suitability Tests and AIQ R.D.McDowall PhD Principal, McDowall Consulting Paul Smith European Validation Program Manager, PerkinElmer Nicola Vosloo PhD European Market Development Leader,

More information

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth

More information

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93)

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93) INTRODUCTION HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93) Headache, sore muscles, arthritis pain... How do you spell relief?

More information

Articaine & Epinephrine Injection according to USP method

Articaine & Epinephrine Injection according to USP method Application Note Biotech & pharmaceutical The most reliable LC-EC applications for Pharmaceutical & Biotech analysis ever formulated Aminoglycosides Amikacin Framycetin Sulphate Gentamicin Sulphate Kanamycin

More information

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number 1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number A. 3(x - x) B. x 3 x C. 3x - x D. x - 3x 2) Write the following as an algebraic expression

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

Evaluating Current Practices in Shelf Life Estimation

Evaluating Current Practices in Shelf Life Estimation Definition of Evaluating Current Practices in Estimation PQRI Stability Working Group Pat Forenzo Novartis James Schwenke Applied Research Consultants, LLC From ICH Q1E An appropriate approach to retest

More information

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting

More information

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone

More information

Optimizing the Post-implementation Monitoring of a LC-MS/MS Method

Optimizing the Post-implementation Monitoring of a LC-MS/MS Method Optimizing the Post-implementation Monitoring of a LC-MS/MS Method Julianne Cook Botelho, PhD Centers for Disease Control and Prevention Atlanta, GA AACC- Mass Spectrometry in the Clinical Lab: Best Practice

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research J. Chem. Pharm. Res., 2010, 2(1): 91-99 ISSN No: 0975-7384 A Stability-Indicating RP-LC method for the Determination of Related

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

Quality by Design Approach for the Separation of Naproxcinod and its Related Substances by Fused Core Particle Technology Column

Quality by Design Approach for the Separation of Naproxcinod and its Related Substances by Fused Core Particle Technology Column Journal of Chromatographic Science Advance Access published October 11, 2012 Journal of Chromatographic Science 2012;00:1 7 doi:10.1093/chromsci/bms162 Article Quality by Design Approach for the Separation

More information

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT

More information

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC Abstract: High performance liquid chromatography (HPLC) was used to determine the identity and concentrations of food dyes

More information

IN VITRO BINDING BIOEQUIVALENCE STUDY SUMMARY TABLES AND SAS TRANSPORT FORMATTED TABLES FOR DATASET SUBMISSION

IN VITRO BINDING BIOEQUIVALENCE STUDY SUMMARY TABLES AND SAS TRANSPORT FORMATTED TABLES FOR DATASET SUBMISSION IN VITRO BINDING BIOEQUIVALENCE STUDY SUMMARY TABLES AND SAS TRANSPORT FORMATTED TABLES FOR DATASET SUBMISSION I. For Calcium Acetate Drug Products Table I.1 Submission Summary * Drug Product Name Strength(s)

More information

Drospirenone and Ethinyl Estradiol Tablets. Type of Posting. Revision Bulletin Posting Date. 31 July 2015 Official Date

Drospirenone and Ethinyl Estradiol Tablets. Type of Posting. Revision Bulletin Posting Date. 31 July 2015 Official Date Drospirenone and Ethinyl Estradiol Tablets Type of Posting Revision Bulletin Posting Date 31 July 2015 Official Date 01 August 2015 Expert Committee MonographsSmall Molecules 4 Reason for Revision Compliance

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLCDAD and HPLCMS/MS Jongheon Shin College of Pharmacy Seoul National University Summary The standard analytical

More information

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal

More information

Research Article. Development and validation of a HPLC analytical assay method for dapoxetine tablets: A medicine for premature ejaculation

Research Article. Development and validation of a HPLC analytical assay method for dapoxetine tablets: A medicine for premature ejaculation Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2014, 6(7):1613-1618 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Development and validation of a HPLC analytical

More information

Assay Qualification Template for Host Cell Protein ELISA

Assay Qualification Template for Host Cell Protein ELISA Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to

More information