A planning study for palliative spine treatment using StatRT and megavoltage CT simulation *
|
|
|
- Gervase Elliott
- 9 years ago
- Views:
Transcription
1 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2011 A planning study for palliative spine treatment using StatRT and megavoltage CT simulation * Yi Rong 1,3a Poonam Yadav 1,4 Bhudatt Paliwal 1,2 Lu Shang, 5 James S. Welsh 1,2 Department of Human Oncology, 1 University of Wisconsin - Madison, Madison, WI, USA; Department of Medical Physics, 2 University of Wisconsin - Madison, Madison, WI, USA; University of Wisconsin Riverview Cancer Centre, 3 Wisconsin Rapids, WI; Vellore Institute of Technology University, 4 Vellore, Tamil Nadu, India; Guangxi Polytechnic of Construction and Technology, 5 Guangxi, China [email protected] Received 22 April, 2010; accepted 23 September, 2010 Megavoltage CT (MVCT) simulation on the TomoTherapy Hi Art system is an alternative to conventional CT for treatment planning in the presence of severe metal artifact. StatRT is a new feature that was implemented on the TomoTherapy operator station for performing online MVCT scanning, treatment planning and treatment delivery in one session. The clinical feasibility of using the StatRT technique and MVCT simulation to palliative treatment for a patient with substantial spinal metallic hardware is described. A patient with metastatic non-small-cell lung cancer involving the thoracic spine underwent conventional kilovoltage CT simulation. The metal artifact due to stainless steel spine-stabilizing rods was too severe for treatment planning, despite attempts to correct using density override. The patient was then re-scanned using MVCT on a tomotherapy unit. Plans were generated using both StatRT and conventional tomotherapy planning (Tomo plan) with different settings for comparison. StatRT planning ran a total of five iterations in a short planning window (10 15 min). Two Tomo plans were generated using: (1) five iterations in the full scatter mode, and (2) 300 iterations in the beamlet mode. It was noted that the DVH of the StatRT plan was almost identical to the Tomo plan optimized by the full scatter mode and the same number of iterations. Dose distribution analysis reveals that these three planning methods yielded comparable doses to heart, lungs and targets. This work also demonstrated that undermodulation can result in a high degree of thread effects. The overall time for the treatment process (including 7 minutes for simulation, 15 minutes for contouring, 10 minutes for planning and 5 minutes for delivery) decreases from hours to around 40 minutes using the StatRT procedure. StatRT is a feasible treatment-planning tool for physicians to scan, contour and treat patients within one hour. This can be particularly beneficial in urgent palliative treatments. PACS numbers: D-, C- Key words: helical tomotherapy, StatRT planning, megavoltage CT imaging, palliative treatment Conflict of Interest Statement: James S. Welsh has received honoraria for speaking for TomoTherapy, Inc.; Yi Rong has received travel sponsorship for TomoTherapy, Inc. a Corresponding author: Yi Rong, 410 Dewey St. Wisconsin Rapids, WI 54494; phone: 715/ ; fax 715/ ; [email protected] * This work was presented in part at the 19th annual meeting of the American College of Radiation Oncology (ACRO) in Feb
2 98 Rong et al.: StatRT for palliative spine treatment using MVCT 98 I. Introduction The use of computed tomography (CT) in radiation therapy has been largely increased due to the essential role of the accurate 3D anatomical information and the attenuation coefficient data that allow tissue heterogeneity corrections in the treatment planning process. Though the vastly improved kilovoltage CT (kvct) scanners have 3D imaging capabilities and have reduced the scanning time, metallic artifacts persist in the present generation of CT. CT artifacts have many causes including patient movement, the limitation of the detector, the image reconstruction method, and the presence of high-density materials such as metal in or on the patient. (1) Streak artifact is one of the most significant challenges in the application of diagnostic kvct simulation. It can be caused by one or more effects such as beam hardening, scattered radiation, poor signal-to-noise ratio, aliasing and object motion, which may result in significant bright and dark streaks on the reconstructed image. (2) Metals can also cause streak artifacts, the main cause of which is the high attenuation and scattering properties of metal in diagnostic X-ray energy range. A direct solution is to use less-attenuating materials (e.g., titanium) or devices with smaller scatter cross-section. (3-5) However such case-by-case solutions are not ideal for clinical practice. Promising mathematical algorithms were also proposed for artifact reduction on diagnostic CT images. (6,7) But manufacturers have not yet implemented them for clinical use. In radiotherapy application, metal artifacts greatly limit the usefulness of kvct images for contouring the target as well as organs at risk. This can be overcome by megavoltage CT (MVCT) imaging. (8) It has been observed that compared to kvct, MVCT images exhibit considerably reduced artifacts in metal implants such as hip implants, surgical clips and dental fillings, (9) and provide good agreement to measurements in dose prediction. (10,11) Although MVCT scans are inferior in soft tissue contrast and spatial resolution compared to those of kvct scans, (12) they can be used for patient alignment and treatment planning as an alternative for the situations when kvct scans are not available or are compromised by artifacts. (13-17) StatRT is a new software package for the helical TomoTherapy Hi Art (TomoTherapy Inc., Madison, WI) treatment planning system and is particularly important in time-sensitive cases. The software is designed to manage scanning, planning and delivery in a timesaving manner. Conventionally, tomotherapy offers three options for plan optimization on the planning station: TERMA, Full Scatter, and Beamlet. The TERMA mode only accounts for the total energy released per unit mass (TERMA) of each beamlet during the process of optimization, without taking into account the transportation of charged particles and scattered particles. Thus, it is the fastest approach of the three, but not the most accurate. The Full Scatter mode calculates both TERMA as well as electron transportation in each iteration of optimization. It is the slowest but most accurate mode for dose optimization. The Beamlet mode allows users to batch beamlets for calculating dose contributions from each beamlet and optimize beamlet weights in each iteration with the real-time display of the dose volume histograms. Thus, it is the most common option for conventional tomotherapy optimization. With StatRT, plans are still calculated using the same photon dose calculation engine, which is the convolution/ superposition algorithm, and delivered in a helical IMRT manner to achieve highly conformal dose distributions. The differences from the conventional procedures are that the Full Scatter mode is the only option for plan optimization and all steps including image acquisition, target and organ delineation, optimization, dose calculation and plan delivery are performed on the operator station. It should eliminate the time for the physical transfer of patients from simulation to treatment and the data transfer from imaging to planning. An emergency palliation can be efficiently initiated with the StatRT approach on the operator station. Also, it offers operating ease as the entire process can be managed on a single computer, the Hi Art treatment system s operator station. It is found that the dose distributions from StatRT are much more conformal than the traditional palliative techniques. (18,19) This allows better tissue sparing and possibly better tolerance of the treatment.
3 99 Rong et al.: StatRT for palliative spine treatment using MVCT 99 However, there are several aspects yet to be studied, such as the shortcomings of this approach, the appropriate parameters for scanning and planning, and the timelines for the first fraction. The present planning study aimed to establish a clinical workflow and feasibility for performing the StatRT palliative treatment. The reported case is also a typical example where MVCT images are the superior option due to the presence of metal artifacts in kvct. The treatment quality of StatRT is also compared with the conventional tomotherapy planning technique on a planning station (Tomo plan) in order to determine the advantages and disadvantages of both approaches. II. Materials and Methods A. Patient history and diagnosis A 72-year-old man was initially diagnosed and treated for stage one bronchoalveolar adenocarcinoma of left lung in Two years later he presented with back pains. CT images from a Philips Brilliance 40-channel CT scanner (Philips Healthcare, Andover, MA) revealed a large lytic lesion at T10, consistent with metastatic disease. A bone scan showed radiotracer activity at T10. He underwent surgical intervention for what appeared to be impending spinal cord compression, specifically undergoing decompression of the spinal cord with laminectomy of T6 and T10 and a component of T11 with posterior instrumentation from T7 through L1 using stainless steel rods. The patient was prescribed for a course of palliative RT for the metastatic non-small-cell lung cancer, with the prescription regimen of 30 Gy in 10 fractions directed to T8 through T12. B. Image acquisition and treatment planning Significant metallic artifacts due to the stainless steel spine stabilizing rods were observed on the initial CT images (Fig. 1(a)). These artifacts were determined too severe to allow treatment planning even with the density override. Alternatively, a MVCT image set (Fig. 1(b)) was obtained using the 3.5 MV beam on the tomotherapy unit for treatment planning. Tomotherapy has two laser systems installed in the treatment room, with a fixed green laser system to indicate the virtual isocenter outside the bore and a movable red laser system to indicate the treatment position. When performing the image acquisition during the StatRT approach, the red lasers Fig. 1. Two scans: diagnostic kvct scan (a) and MVCT scan (b) on the TomoTherapy Hi Art system for a patient with metal implants in the spine.
4 100 Rong et al.: StatRT for palliative spine treatment using MVCT 100 are in their home position (coinciding with the green) and projecting to the virtual isocenter. The virtual isocenter is defined outside the bore, 70 cm away from the actual isocenter in the longitudinal direction. Patients are usually aligned to the green lasers for anterior-posterior and left-right directions, but to the red laser on both sides for leveling. MVCT interslice distance can be selected as coarse (6 mm), normal (4 mm), and fine (2 mm). (20) In this case, the MVCT scan took about nine minutes (around 520 seconds) using a normal grid and slice selection from -20 cm to +20 cm, with the central slice being the virtual setup isocenter, giving a total scan length of 40 cm. As we gained experience, we found that a shorter (around 7 minutes) scan, from -15 cm to +15 cm, is often sufficient to cover most of the tumor sites. The artifacts could not be completely eliminated in the MVCT image but were substantially reduced in comparison to kvct. The patient was sent home after MVCT acquisition and came back for a regular treatment using an approved and verified tomotherapy plan that was done in a conventional way. That is to say, the MVCT image set was exported to the Pinnacle planning system for contouring of the planning target volume (PTV), heart and lungs. It was then sent back to tomotherapy for conventional tomo planning using the Beamlet mode and 300 iterations (Tomo Plan 1). In our clinic, the image-value-to-density table (IVDT) for MVCT scans was periodically verified and also acquired every time prior to a MVCT planning. To test the feasibility of the StatRT planning on the operator station, the PTV, lungs and heart were also contoured on the MVCT images, with the auto-contour tool used for lungs, and the StatRT plan was generated. It took approximately 15 minutes in contouring, mostly as a result of the complex PTV contoured by the physician. StatRT optimization with five iterations in the Full Scatter mode took around 10 minutes. For comparison, Tomo Plan 2 was also generated in the Full Scatter mode and five iterations to mimic the StatRT approach, but was completed on the planning station using the first set of contours as used for the Tomo Plan 1. Table 1 shows complete information on the parameters setup for StatRT and two Tomo plans. The field width and pitch remained the same for all three plans. Ta b l e 1. Treatment planning parameter setup for StatRT and two Tomo plans. StatRT Plan Tomo Plan 1 Tomo Plan 2 Optimization Mode Full Scatter Full Scatter Beamlet Iteration Pitch Field Width 5 cm 5 cm 5 cm Modulatin Factor (actual) 2.7 (1.33) 2.7 (1.30) 2.7 (2.50) Treatment Duration 203 sec 203 sec 376 sec Gantry Period 15 sec 15 sec 27 sec C. Pretreatment verification and patient DQA following the first fraction The tomotherapy cheese phantom with two high-density and two low-density plugs were MVCT scanned and planned using the StatRT software (Fig. 2). A hypothetical PTV was contoured on the high-density plugs and given the same prescription dose as the patient treatment. Avoidance structures were contoured on the low-density plugs. The delivery quality assurance (DQA) plan was generated on the same MVCT image and measured using the ion chamber/ film system. This verification was created to test the dose delivery accuracy with density heterogeneity using MVCT scans prior to the patient treatment. For a genuine clinical emergency with the patient scheduled to be treated immediately using the StatRT, the DQA would be conducted after the first fraction. DQA is fully integrated into the planning software, and it can be delivered from the operator station to compare with the planar chamber dose measurements using the OmniPro-I mrt MatriXX system (IBA Dosimetry, Schwarzenbruck, Germany) with the 2D ionization chamber array. (21,22)
5 101 Rong et al.: StatRT for palliative spine treatment using MVCT 101 Fig. 2. Transverse, sagittal and coronal views of the MVCT scan of the cheese phantom with two high-density inserts (1.335 g/cm 3 and g/cm 3 ) and two low-density inserts (0.46 g/cm 3 and 0.29 g/cm 3 ). The two high-density inserts are delineated in red as the PTV (prescribed to 30 Gy). The two low-density inserts are delineated in blue as avoidance structures. Isdodose curves (100%, 90%, 70% and 60%) are also shown. III. Results Figure 3(a) shows that the absolute dose difference is 2.175% at the high-density region and 0.037% near the low-density region. The gamma map on Fig. 3(b) showed a dose agreement within ± 3%/3 mm between ion chamber/film measurements and dose calculation on the MVCT image set of the cheese phantom with density heterogeneity. The choice of the Beamlet mode and the high level of the intensity modulation resulted in longer gantry period and thus a longer treatment time (by 46%) for Tomo Plan 2 compared to the other two plans (Table 1). The dose parameters along with the homogeneity index for all three plans are calculated and tabulated in Table 2. The homogeneity index describes the target dose uniformity and is calculated by: Homogeneity Indexes (D 2 - D D p 98 ) 100% (1) where D 2 and D 98 represent the doses to 2% and 98% of the PTV, and D p represents the prescription dose (23) Dose analysis in Table 2 reveals that these three planning methods yield comparable results for max/min/mean doses to heart, lungs, and PTVs. The max/min/mean Fig. 3. (a) Dose profile comparison between measured dose by ion chamber matrix and calculated dose; (b) gamma map for 3% / 3 mm criteria between the film measured dose and calculated dose.
6 102 Rong et al.: StatRT for palliative spine treatment using MVCT 102 Ta b l e 2. Dose index comparisons for: (i) the StatRT plan, (ii) the Tomo Plan 1 and (iii) the Tomo Plan 2. Max Dose Min Dose D 40% Mean Dose Homogeneity Physical Volume (Gy) (Gy) (Gy) (Gy) Index (cc) (i) PTV Heart Lungs (ii) PTV Heart Lungs (iii) PTV Heart Lungs doses to PTV in the StatRT plan are comparable to Tomo plan 1 and 2. However, the homogeneity index comparison shows a significant improvement in dose homogeneity in Tomo Plan 2 compared to StatRT and Tomo Plan 1, due to the reduced thread effects with higher number of iterations. Doses to the heart and lungs are within clinically acceptable dose limits in the StatRT and the two Tomo plans. Dose volume histograms (DVHs) of the three plans are shown in Fig. 4. The DVH of the StatRT plan is almost identical to Tomo Plan 2. Highly conformal dose coverage to the tumor site was achieved in all three plans. Six isodose levels 105%, 100%, 90%, 80%, 70% and 60% were compared (Fig. 5). The isodose lines are identical except for the 31.5 Gy dose level (i.e., 105% of the prescription), represented as a distinct thread effect, which resulted from the undermodulation in the StatRT plan and Tomo Plan 1 where only five iterations were run in optimization. The thread effects in this case present as a series of dose ripples near the outer edge of dose distribution along the longitudinal direction. The results show that the radiation dose distributions in the StatRT planning are equivalent to a Tomo plan with the same optimization parameters. Also, we observed that treatment time is significantly shorter when less iteration is used in the planning process. The overall time for the treatment process (including 7 minutes for simulation, 15 minutes for contouring, 10 minutes for planning and 5 minutes for delivery) decreases from hours to around 40 minutes using the StatRT procedure. DQA performed for the StatRT plan is shown in Fig. 6. The calculated dose from the StatRT plan was verified to be within 3%/3 mm agreement with the measured dose.
7 103 Rong et al.: StatRT for palliative spine treatment using MVCT 103 Fig. 4. DVH comparisons for: (i) StatRT, (ii) Tomo Plan 1 and (iii) Tomo Plan 2. PTV, heart and lungs are illustrated by red, pink and blue, respectively.
8 104 Rong et al.: StatRT for palliative spine treatment using MVCT 104 Fig. 5. Comparison of six isodose levels (105%, 100%, 90%, 80%, 70% and 60%) for three plans with different plan settings: (i) StatRT plan, (ii) Tomo Plan 1 and (iii) Tomo Plan 2. Overall, all isodose lines are almost identical in the three plans, except for the 105% level where severe thread effects occurred in (i) and (ii). Fig. 6. Following first fraction DQA for the patient treatment using the MatriXX system: (a) isodose comparison for levels 100%, 90%, 70%, 60%, and 50%; (b) gamma map with 3%/3 mm criteria for planer dose comparison; (c) and (d) lateral and longitudinal profile comparisons between measured (red) and calculated (blue) doses, respectively.
9 105 Rong et al.: StatRT for palliative spine treatment using MVCT 105 IV. DISCUSSION Palliative radiotherapy is generally appropriate when there are slim chances of cure and the clinical aim is for pain control or other symptomatic relief. During palliative radiotherapy it becomes important to efficiently synchronize the scanning, treatment planning, contouring, calculations or simulation-associated statistical data, and treatment delivery in an efficient manner. StatRT is an extended utility on tomotherapy for urgent situations when time is a significant constraint. This efficiency is invaluable for facilities that have only one tomotherapy unit or circumstances in which urgent palliation is considered desirable. With the current toolset, it is possible to scan, plan and deliver radiotherapy treatment within one hour in comparison to the ordinary situation that takes hours of waiting time between simulation and treatment. A shorter planning window used in the StatRT plan still allows at least five iterations of optimization, which generates plans with clinically acceptable quality. Yet thread effects are distinct without sufficient iterations of optimization. Using recommended pitch factors (0.86/n) helps in minimizing the them. (24) Nevertheless plans with a D 1% value smaller than 110% and homogeneity index smaller than 17 are still classified good or outstanding, based on the RTOG criteria. (25) Further study is needed to understand the clinical significance of the heterogeneous dose distribution with the thread effects. Though higher number of iterations might improve plan quality, it is at the cost of increased planning and delivery time. The simulation time can be restricted to 10 minutes using the normal mode and the contouring time can be reduced also as more experience accumulated. Total time from scanning to treatment can be reduced to 40 minutes, facilitated by the use of a fast auto-contouring tool and fewer (<10) iterations in optimization. Comparable DVHs and dose distributions were observed in the Tomo plan using the Full Scatter optimization setting. The StatRT technique is beneficial to the patients in need of palliative treatment by reducing the pain from being moved around. Typically, AP-PA sets of beams are used for such a palliative treatment since the intended doses can be achieved with a homogenous dose distribution. (26) Generating a palliative plan takes much less time on a conventional linear accelerator than the procedure described above. The portal imager or a conventional kilovoltage X-ray simulator is normally used where the radiation oncologist specifies field borders and designs beam blocks to isolate the tumor from critical sensitive organs. However, in our facility only one tomotherapy machine is available and the volumetric CT image is required for treatment planning. This increases difficulties in palliative treatment especially for a case as described in this study. It is noted that the absorbed dose in the soft tissue regions affected by the high-density artifacts is higher in the kvct image than the MVCT-based recalculation. The dose differential can be as large as 15%. (27) Due to the mild appearance of the metallic streak artifacts, MVCT provides better image quality in the region surrounding the metal hardware. The MVCT image reveals sufficient information of the tumor and normal anatomy for patient repositioning and treatment planning. (10,11,28) It is worth noting that recent studies show output instabilities over time of the tomotherapy imaging beam and the consequent dose effect can be as large as 3%. (29,30) Therefore, our protocol requires acquisition of a new IVDT prior to every MVCT scans for dose calculation purpose. It might be difficult for patients with significant pain to remain in the same position for more than 30 minutes; thus we recommend a second MVCT scan prior to the treatment in order to verify the target position. The time for this additional MVCT scan was not included in the total time mentioned above. We would expect another 5 10 minutes, depending on the length of the target, the choice of the scan mode and the difficulty in registration. The entire procedure takes 40 minutes to one hour, which limits the usage of the tomotherapy unit for other patients. It is also important to coordinate the schedule with all personnel involved, including physician and medical physicist.
10 106 Rong et al.: StatRT for palliative spine treatment using MVCT 106 V. Conclusions In situations where conventional kvct is unable to provide accurate anatomical information due to interference from high-density materials such as metal, MVCT-based planning can be advantageous. With the successful integration of MVCT on the workstation, tomotherapy offers two options for treatment planning using either Tomo plan or StatRT plan. Satisfying essentially all of the treatment planning objectives, the StatRT technique has proved to be an excellent solution for combining image acquisition, treatment planning and dose delivery in one session when time is the limiting factor. The dose distribution from StatRT appears equivalent to a Tomo plan with a limited number of iterations. It also appears comparable to a Tomo plan with large number of iteration except for the presence of a noticeable thread effect at high isodose levels. In conjunction with MVCT imaging, StatRT is a feasible tool for physicians to scan, contour and treat patients within one hour, and can be beneficial when there are severe imaging artifacts in the target area. References 1. Barrett JF, Keat N. Artifacts in CT: Recognition and Avoidance. Radiographics. 2004;24(6): De Man B, Nuyts J, Dupont P, Marchal G, Suetens P. Metal streak artifacts in X-ray computed tomography: a simulation study. IEEE Trans Nucl Sci. 1999;46(3): Veiga-Pires JA, Kaiser M. Artefacts in CT scanning. Br J Radiol. 1979;52(614): Robertson DD, Weiss PJ, Fishman EK, Magid D, Walker PS. Evaluation of CT techniques for reducing artifacts in the presence of metallic orthopedic implants. J Comput Assist Tomogr. 1988;12(2): Henrich G. A simple computational method for reducing streak artifacts in CT images. Compt Tomogr. 1980;4(1): Haramati N, Staron RB, Mazel-Sperling K, et al. CT scans through metal scanning technique versus hardware composition. Comput Med Imaging Graph. 1994;18(6): De Man B, Nuyts J, Dupont P, Marchal G, Suetens P. Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm. IEEE Trans Nucl Sci. 2000;47(3): Ruchala KJ, Olivera G, Schloesser E, Mackie T. Megavoltage CT on a tomotherapy system. Phys Med Biol. 1999;44(10): Schreiner LJ, Rogers M, Salomons G, Kerr A. Metal artifacts supression megavoltage computed tomography [proceedings paper].in: Flynn MJ, editor. Medical Imaging 2005: Physics of medical imaging. Bellingham (WA): SPIE; p Langen KM, Meeks SL, Poole DO, et al. The use of megavoltage CT (MVCT) images for dose recomputations. Phys Med Biol. 2005;50(18): Yang C, Liu T, Jennelle RL, et al. Utility of megavoltage fan-beam CT for treatment planning in a head-and-neck cancer patient with extensive dental fillings undergoing helical tomotherapy. Med Dosim. 2010;35(2): Stützel J, Oelfke U, Nill SA. Quantitative image quality comparison of four different image guided radiotherapy devices. Radiother Oncol. 2008;86(1): Zhao S, Robeltson D, Wang G, Whiting B, Bae K. X-ray CT metal artifact reduction using wavelets: an application for imaging total hip prostheses. IEEE Trans Med Imaging. 2000;19(12): Welsh JS, Bradley K, Manon R, et al. Megavoltage CT imaging for adaptive helical tomotherapy of lung cancer. Clin Lung Cancer. 2004;5: Hong TS, Welsh JS, Ritter MA, et al. Megavoltage computer tomography: an emerging tool for image-guided radiotherapy. Am J Clin Oncol. 2007;30(6): Rong Y, Fahner T, Welsh JS. Hypofractionated breast and chest wall irradiation using simultaneous in-field boost IMRT delivered via helical tomotherapy. Technol Cancer Res Treat. 2008;7(6): Rong Y, Paliwal BR, Howard S, Welsh JS. Pulsed reduced dose-rate radiotherapy using helical tomotherapy. Int J Radiat Oncol Biol Phys Epub online. 18. Mclntosh A, Dunlap N, Sheng K, et al. Helical tomotherapy-based statrt: dosimetry evaluation for clinical implementation of a rapid radiation palliation program. Med Dosim In press. 19. MacPherson M, Montgomery L, Fox G, et al. On-line rapid palliation using helical tomotherapy: a prospective feasibility study. Radiother Oncol. 2008;87(1): Lee IJ, Seong J, Lee CG, et al. Early clinical experience and outcome of helical tomotherapy for multiple metastatic lesions. Int J Radiat Oncol Biol Phys. 2009;73(5): Oinam AS, Singh L, Sharma SC, Goswami P. Improvement of I mrt MatriXX in terms of spatial resolution and large area acquisition for patient-specific intensity-modulated radiotherapy verification. J Med Phys. 2009;34(3): Dobler B, Streck N, Klein E, Loeschel R, Haertl P, Koelbl O. Hybrid plan verification for intensity modulated radiation therapy (IMRT) using the 2D ionization chamber array I mrt MatriXX a feasibility study. Phys Med Biol. 2010;55(2):N39 N55.
11 107 Rong et al.: StatRT for palliative spine treatment using MVCT Wu Q, Mohan R, Morris M, Lauve A, Schmidt Ullrich R. Simultaneous integrated boost intensity-modulated radiotherapy for locally advanced head-and-neck squamous cell carcinomas. I: dosimetric results. Int J Radiat Oncol Biol Phys. 2003;56(2): Kissick MW, Fenwick J, James JA, et al. The helical tomotherapy thread effect. Med Phys. 2005;32(5): Radiation Therapy Oncology Group. RTOG 0022: Phase I/II study of conformal and intensity modulated irradiation for oropharyngeal cancer. Philadelphia (PA): Radiation Therapy Oncology Group; Retrieved September 14, 2006, from Andic F, Baz Cifci S, Ors Y, Niang U, Dirier A, Adli MA. A dosimetric comparison of different treatment plans of palliative spinal bone irradiation: analysis of dose coverage with respect to ICRU 50 Report. J Exp Clin Cancer Res. 2009;28(1): Meeks S, Langen K, Wagner T, Willoughby, Kupelian P. The use of MVCT images for dose calculation in the presence of metaliic objects [abstract]. Med Phys. 2005;32(6): Forrest LJ, Mackie R, Ruchala K, et al. The utility of megavoltage computed tomotherapy images from a helical tomotherapy system for setup verification purposes. Int J Radiat Oncol Biol Phys. 2004;60(5): Duchateau M, Tournel K, Verellen D, et al. The effect of tomotherapy imaging beam output instabilities on dose calculation. Phys Med Biol. 2010;55(11):N329 N Yadav P, Tolakanahalli R, Rong Y, Paliwal BR. The effect and stability of MVCT images on adaptive tomotherapy. J App Clin Med Phys. 2010; 11(4):4-14.
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients
대한방사선종양학회지 2005;23(3):169~175 Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients Department of Radiation Oncology, College of Medicine, Ewha Womans University
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT
Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Wen Li, Andrew Vassil, Lama Mossolly, Qingyang Shang, Ping Xia Department of Radiation Oncology Why
Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com
Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider
Dosimetric impact of the 160 MLC on head and neck IMRT treatments
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Dosimetric impact of the 160 MLC on head and neck IMRT treatments Prema Rassiah-Szegedi, a Martin Szegedi, Vikren Sarkar, Seth Streitmatter,
THE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning Karin M. Andersson,
NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Introduction Prostate is one of the treatment sites that is well suited for IMRT. For
Intensity-Modulated Radiation Therapy (IMRT)
Scan for mobile link. Intensity-Modulated Radiation Therapy (IMRT) Intensity-modulated radiotherapy (IMRT) uses linear accelerators to safely and painlessly deliver precise radiation doses to a tumor while
QA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
TITLE: Comparison of the dosimetric planning of partial breast irradiation with and without the aid of 3D virtual reality simulation (VRS) software.
SAMPLE CLINICAL RESEARCH APPLICATION ABSTRACT: TITLE: Comparison of the dosimetric planning of partial breast irradiation with and without the aid of 3D virtual reality simulation (VRS) software. Hypothesis:
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of the Chest
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of the Chest File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_the_chest
Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD
Good Morning Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD Barbara Agrimson, BS RT(T)(R), CMD Steve Rhodes, BS RT(T), CMD Disclaimer This presentation
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_the_prostate
Key words: treatment planning, quality assurance, 3D treatment planning
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical
Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.
Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan
Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation
Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation Deborah Schofield, Laurence Court, Aaron Allen, Fred Hacker, Maria Czerminska Department of Radiation Oncology Dana
A Revolution in the Fight Against Cancer. What TomoTherapy Technology Means to You
A Revolution in the Fight Against Cancer What TomoTherapy Technology Means to You Cancer Treatment that Revolves Around You When it comes to choosing the right treatment for your cancer, your team of healthcare
Clinical Education A comprehensive and specific training program. carry out effective treatments from day one
Proton Therapy Clinical Education A comprehensive and specific training program carry out effective treatments from day one Forewarned is forearmed Although over 100,000 patients have been treated in proton
Helical TomoTherapy for Lung Cancer Radiotherapy: Good Science Pays Clinical Dividends Peter Hoban, Ph.D., TomoTherapy Inc.
CLINICAL FEATURE Helical TomoTherapy for Lung Cancer Radiotherapy: Good Science Pays Clinical Dividends Peter Hoban, Ph.D., TomoTherapy Inc. The Challenge Lung cancer kills more people each year in the
Radiation Protection in Radiotherapy
Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France [email protected] Radiation therapy The lecture is oriented to
Management of spinal cord compression
Management of spinal cord compression (SUMMARY) Main points a) On diagnosis, all patients should receive dexamethasone 10mg IV one dose, then 4mg every 6h. then switched to oral dose and tapered as tolerated
Corporate Medical Policy
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Head and Neck File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_head_and_neck
Volumetric modulated arc therapy (VMAT) for prostate cancer
Planning study Volumetric modulated arc therapy (VMAT) for prostate cancer Institution: Purpose: Swedish Cancer Institute, Seattle, WA, USA In a community cancer center, treating prostate cancer comprises
Purchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage
Proton Therapy for Head & Neck Cancers
Proton Therapy for Head & Neck Cancers Robert S Malyapa, MD, PhD and William M Mendenhall, MD University of Florida Proton Therapy Institute, Jacksonville, USA Carmen Ares, MD and Ralf Schneider, MD Paul
Goals and Objectives: Breast Cancer Service Department of Radiation Oncology
Goals and Objectives: Breast Cancer Service Department of Radiation Oncology The breast cancer service provides training in the diagnosis, management, treatment, and follow-up of breast malignancies, including
Radiation Therapy. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements
PRIMARY CERTIFICATION DIDACTIC AND CLINICAL COMPETENCY REQUIREMENTS EFFECTIVE JANUARY 2014 Radiation Therapy 1. Introduction Candidates for certification and registration are required to meet the Professional
IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect [email protected]. Protect, Enhance and Save Lives
Vincent Bossier System Architect [email protected] IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT
intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis 11/2009 5/2016 5/2017 5/2016
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Abdomen File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review
What is the CyberKnife System?
CYBERKNIFE Robotic Radiosurgery System PATIENT BROChURE 1 What is the CyberKnife System? have stabilizing frames bolted to their head or limit their breathing during treatment to minimize movement of the
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota
Creation of 3D printed phantoms for clinical radiation therapy Eric Ehler, PhD Assistant Professor University of Minnesota ehler [email protected] Overview Background of 3D Printing Practical Information Current
Our Department: structure and organization
EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled
Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers
Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers H. Helen Liu, PhD Department of Radiation Physics, U.T. MD Anderson Cancer Center, Houston, TX AAPM, Seattle, 2005 Acknowledgement
INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS
INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS HOW DOES RADIATION THERAPY WORK? Cancer is the unrestrained growth of abnormal cells within the body. Radiation therapy kills both
Accreditation a tool to help reduce medical errors. Professor Arthur T Porter PC MD FACR FRCPC FACRO
Accreditation a tool to help reduce medical errors Professor Arthur T Porter PC MD FACR FRCPC FACRO Errors in Radiotherapy Radiation therapy is a highly regulated medical practice with historically low
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
Implementation Date: April 2015 Clinical Operations
National Imaging Associates, Inc. Clinical guideline PROSTATE CANCER Original Date: March 2011 Page 1 of 5 Radiation Oncology Last Review Date: March 2015 Guideline Number: NIA_CG_124 Last Revised Date:
Clinical Trials and Radiation Treatment. Gerard Morton Odette Cancer Centre Sunnybrook Research Institute University of Toronto
Clinical Trials and Radiation Treatment Gerard Morton Odette Cancer Centre Sunnybrook Research Institute University of Toronto What I will cover.. A little about radiation treatment The clinical trials
Multi-slice Helical CT Scanning of the Chest
Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.
Independent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
First Three Years After Project Proton Therapy Facility:
39. Project Related Utilization and Financial Information. Proton Therapy Facility: Unit of Service Delivery Projected Utilization FY2017 FY2018 FY2019 Estimated Patients 448 670 808 Average Revenue Per
Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
Development and current status of proton therapy for lung cancer in Korea
Thoracic Cancer ISSN 1759-7706 INVITED REVIEW Development and current status of proton therapy for lung cancer in Korea Myonggeun Yoon Department of Radiological Science, College of Health Science, Yonsei
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
Baylor Radiosurgery Center
Radiosurgery Center Baylor Radiosurgery Center Sophisticated Radiosurgery for both Brain and Body University Medical Center at Dallas Radiosurgery Center 3500 Gaston Avenue Hoblitzelle Hospital, First
The Canadian National System for Incident Reporting in Radiation Treatment (NSIR-RT) Taxonomy March 2, 2015 V2
The Canadian National System for Incident Reporting in Radiation Treatment (NSIR-RT) Taxonomy March 2, 2015 V2 Taxonomy Data Category Number Description Data Fields and Menu Choices 1. Impact 1.1 Incident
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
External dosimetry Dosimetry in new radiotherapeutic techniques
External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau 44805 Saint-Herblain France [email protected] Objective : To describe
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY
Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY W. PARKER, H. PATROCINIO Department of Medical Physics, McGill University Health Centre, Montreal, Quebec, Canada 7.1. INTRODUCTION
Basic Radiation Therapy Terms
Basic Radiation Therapy Terms accelerated radiation: radiation schedule in which the total dose is given over a shorter period of time. (Compare to hyperfractionated radiation.) adjuvant therapy (add-joo-vunt):
02 CyberKnife: Treatment Delivery
TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography
Today s agenda Computed Tomography Presented by: Dina Hernandez, BSRT, RT (R), CT, QM Krista Bush, RT, MBA Leonard Lucey, JD ACR Quality & Safety MIPPA overview and CMS requirements CT accreditation How
A new score predicting the survival of patients with spinal cord compression from myeloma
A new score predicting the survival of patients with spinal cord compression from myeloma (1) Sarah Douglas, Department of Radiation Oncology, University of Lubeck, Germany; [email protected] (2) Steven
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
DIFFERENT FROM THE REST FOCUSED ON THE FUTURE. A guide to understanding TomoTherapy treatments
DIFFERENT FROM THE REST FOCUSED ON THE FUTURE A guide to understanding TomoTherapy treatments Advanced treatment planning Integrated, daily CTrue imaging 360º treatment delivery A platform for truly personalized
Page: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
Failure Modes and Effects Analysis (FMEA)
Failure Modes and Effects Analysis (FMEA) Sasa Mutic Washington University School of Medicine St. Louis Missouri Failure Modes and Effects Analysis Objectives: To motivate the use of FMEA and to provide
Proton Therapy. What is proton therapy and how is it used?
Scan for mobile link. Proton Therapy Proton therapy delivers radiation to tumor tissue in a much more confined way than conventional photon therapy thus allowing the radiation oncologist to use a greater
Radiation Therapy in Prostate Cancer Current Status and New Advances
Radiation Therapy in Prostate Cancer Current Status and New Advances Arno J. Mundt MD Professor and Chairman Dept Radiation Oncology Moores Cancer Center UCSD Radiation Therapy Wilhelm Roentgen (1845-1923)
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
Scripps Proton Therapy Center: Configuration and Implementation
Scripps Proton Therapy Center: Configuration and Implementation Anthony Mascia Proton Symposium 2013 Annual AAPM Meeting Indianapolis, Indiana USA Facility Configuration Scripps Proton Therapy Center ***All
Intensity-modulated radiotherapy and volumetric-modulated arc therapy for malignant pleural mesothelioma after extrapleural pleuropneumonectomy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 4, 2013 Intensity-modulated radiotherapy and volumetric-modulated arc therapy for malignant pleural mesothelioma after extrapleural pleuropneumonectomy
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
METASTASES TO THE BONE
RADIATION THERAPY FOR METASTASES TO THE BONE Facts to Help Patients Make an Informed Decision TARGETING CANCER CARE AMERICAN SOCIETY FOR RADIATION ONCOLOGY WHAT ARE BONE METASTASES? Cancer that starts
Case Number: RT2009-124(M) Potential Audiences: Intent Doctor, Oncology Special Nurse, Resident Doctor
Renal Cell Carcinoma of the Left Kidney Post Radical Surgery with pt4 Classification with Multiple Lung and Single Brain Metastases: the Role and Treatment Consideration of Radiotherapy Case Number: RT2009-124(M)
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Accreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
GE Healthcare. pet/ct for simulation. Precision in motion.
GE Healthcare pet/ct for simulation. Precision in motion. Patient profile H: 5 10 W: 370 lbs BMI: 53 Each cancer patient is just as unique as his or her cancer. Approaches to cancer treatment are changing
4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology [email protected] http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
Quality Assurance of accelerators; the technologists responsibility
Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section
Principles of radiation therapy
CHAPTER 2 Principles of radiation therapy Michael J. Gazda, MS, and Lawrence R. Coia, MD This chapter provides a brief overview of the principles of radiation therapy. The topics to be discussed include
P R E S E N T S Dr. Mufa T. Ghadiali is skilled in all aspects of General Surgery. His General Surgery Services include: General Surgery Advanced Laparoscopic Surgery Surgical Oncology Gastrointestinal
Follow this and additional works at: http://digitalcommons.wustl.edu/open_access_pubs
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2012 Evaluation of the efficiency and effectiveness of independent dose calculation followed by machine log file
Department of Radiation Oncology
Department of Radiation Oncology Welcome to Radiation Oncology at Emory Clinic Every member of Emory Clinic Department of Radiation Oncology strives to provide the highest quality of care for you as our
Target Volumes for Anal Carcinoma For RTOG 0529
Target Volumes for Anal Carcinoma For RTOG 0529 Robert Myerson, M.D. Ph.D.*, Lisa Kachnic, M.D.**, Jacqueline Esthappan, Ph.D.*,Parag Parikh M.D.*William Straube M.S.*, John Willins, Ph.D.** *Washington
The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning.
The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning. KAROLINA KOKUREWICZ S U P E R V I S O R S : D I N O J A R O S Z Y N S K I, G I U S E P P E S C
Proton Therapy for Prostate Cancer
Proton Therapy for Prostate Cancer Andrew K. Lee, MD, MPH Director, Proton Therapy Center Associate Professor Department of Radiation Oncology M.D. Anderson Cancer Center Randomized studies showing benefit
1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
Innovative RT SBRT. The variables with REQ in superscript are required.
The variables with REQ in superscript are required. The variables with a are single-select variables; only one answer can be selected. The variables with a are multi-select variables; multiple answers
Table of Contents. Data Supplement 1: Summary of ASTRO Guideline Statements. Data Supplement 2: Definition of Terms
Definitive and Adjuvant Radiotherapy in Locally Advanced Non-Small-Cell Lung Cancer: American Society of Clinical Oncology Clinical Practice Guideline Endorsement of the American Society for Radiation
A Patient s Guide to the Calypso System for Breast Cancer Treatment
A Patient s Guide to the Calypso System for Breast Cancer Treatment Contents Introduction How Radiation Works To Treat Cancer Radiation Treatment for Breast Cancer Special Concerns for Treating the Left