1 MR. TELLIARD: Our next speaker is Paula Hogg. Paula is currently a lab manager at Hampton Roads Sanitation District s Central Environmental Laboratory. We would like to also thank Hampton Roads for sending most of their lab people. APPLICATION OF ICP-MS TECHNOLOGY FOR TRACE METALS ANALYSIS MS. HOGG: Today, I will be talking about the use of ICP-MS technology for trace metals analysis. HRSD's Central Environmental Laboratory performs a variety of trace metals analysis of environmental samples, primarily in support of wastewater treatment plant operations. The current total number of metals results is approaching 70,000 annually. Sample matrices include treatment process, industrial pretreatment program samples, treatment plant effluent for NPDES monitoring applications, groundwater, surface water, and biosolids. HRSD purchased an ICP-MS over four years, and the laboratory uses this instrumentation for trace metals analysis along with a Graphite Furnace AA(GFAA), a flame AA, and an ICP. A starting point in evaluating ICP-MS is to compare this technique with more traditional methods. Today, I will discuss and compare ICP-MS to the Stabilized Temperature Graphite Furnace AA(STGFAA) method, including methodologies, applications, method performance, and cost effectiveness of these methods. There are several considerations in choosing an appropriate method for trace metal analysis. These include sample concentration levels, data quality objectives, the application of the analytical results for decision-making criteria and costs. There are various levels of analysis, clean techniques, sample concentration levels, methodologies, and applications for trace metals monitoring as shown on this slide. ICP-MS can be widely applied over concentration ranges, but it is most appropriate and suited for low to sub part per billion concentration ranges and this s the primary application of ICP-MS at HRSD. Moving from the bottom level or at high concentrations to the top level of the triangle or at low concentration ranges, the amount of effort required for contamination control and the use of clean techniques, becomes more important, and with that, costs increase as well. HRSD's approach is to use the most cost effective and practical method in meeting data quality objectives. The purpose of clean techniques is to produce data free from bias due to contamination to achieve an adequate level of reliability and accuracy as defined by data quality objectives. So, as sample concentrations and regulatory levels decrease, clean techniques become necessary. For application of ISP-MS, clean techniques become a given for monitoring samples at water quality-based permit levels. In evaluating the distribution of the metals analysis workload at HRSD, nearly half of the work is dedicated to analysis of effluent for NPDES monitoring applications in Virginia and for other water quality monitoring projects. Clean techniques are used for this level of monitoring, coupled with STGFAA and ICP-MS methods. June

2 For NPDES monitoring and other regulatory program requirements, STGFAA is used, because this method is approved under 40 CFR Part 136. The ICP-MS technique is used for nonregulatory water quality studies, process characterization studies, and is an integral part of our clean sampling and analytical program at HRSD. All of the sampling equipment used for clean level monitoring is prepared using clean techniques and equipment blanks are screened using the ICP-MS prior to taking equipment out into the field. All sampling equipment is certified as free from contamination. Currently, the STGFAA workload is at maximum capacity, due to an increase in clean level monitoring. Pending the method approval, this workload can be shifted to ICP-MS. These are typical analytical concentration range capabilities of STGFAA and ICP-MS. These may vary depending on operating and analytical conditions as well as instrument model and manufacturer. ICP-MS can be applied over a wide concentration range, as seen on the next slide. This is due to the wide linear dynamic range offered by the ICP-MS technology, up to ten orders of magnitude for some elements. The next two slides summarize EPA published methods. The EPA 200 series methods, published in 1983, are currently approved in 40 CFR Part 136, and these are used for NPDES monitoring programs. The elements listed are typically required for effluent monitoring. The Ys indicate which elements are included in each method. The Ns indicate elements not listed in the methods. The more recent EPA methods, and 1639 for Graphite Furnace AA are not approved in 40 CFR, Part 136. However, the inter-laboratory validation studies for 1600 series are currently underway. The ICP-MS methods, and 1638, are also not currently approved. For Method 1638, as with 1639, there are some elements not listed as part of the 1600 series, which may be required for effluent or NPDES monitoring. What has changed and what is different about the more recent EPA methods, and and 1600 series methods? First, all of the digestion protocols include sample concentration steps or an alternative for in-bottle digestion, which helps with contamination control. Also, digestions require the addition of HCl, and all elements can be determined using one digestion method, instead of using separate digestions analysis such as arsenic, selenium and silver. In methods and 1639, a hydrogen argon purge is used. The 1600 series methods include both sampling and analytical methods to specifically address the needs for measuring toxic metals at water quality criteria levels. These methods are designed to preclude contamination in nearly all situations. The 1600 series are also performance based methods or, The laboratory is permitted to omit any steps or modify any procedure, provided all performance requirements set forth in the method are met. QC analysis, however, cannot be omitted. In the 1600 series methods, the terms must, may, and shall are used to indicate the relative importance of steps and indicate which procedures are necessary for successful analysis. These methods also include clean sampling techniques which are incorporated into Method 1669, as mentioned earlier. There are particular clean techniques, which HRSD has found important in controlling contamination for the NPDES monitoring level. For both sampling and analysis, sample handling by personnel, and exposure to air is minimized, and these are the two key components of successful clean level monitoring. Equipment and glassware preparation and analysis are performed in HEPA filtered clean areas, not necessarily Class 100 clean rooms in Class 100 clean benches. Traffic is June 1999

3 minimized. Analytically, glassware is washed immediately prior to use. Non-colored labware, powder-free gloves, and high purity reagents are also used. Analysts are dedicated to clean protcols. The use of cosmetics, jewelry, and lotions which sometimes contain low levels of the trace metals are avoided. For dissolved metals determination, capsule filtration is used. In summary, adequate contamination control can be achieved using practical, preventive approaches. This is a picture of the automated enclosed sampling system for unattended collection of composite samples. Tubings, containers, fittings, and all components are metal-free. The sampling equipment is prepared using clean protocols, and is certified as clean using ICP/MS screening prior to each sampling event. The system can be wired to the plant flow measurement system for the unattended collection of composite samples and simultaneous collection of field blanks. Blank data including laboratory reagent blank, field blank, and equipment blank, are used to derive the level of effort needed to control contamination. Another consideration in selection and application of a method is detection and quantitation capabilities. Overall, detection levels for trace metals using ICP-MS technology are significantly lower than that of STGFAA, often by orders of magnitude. For quantification, the minimum level is defined as the lowest level at which the entire analytical system gives a recognizable and reliable signal, an acceptable calibration point, or approximately 3.18 times the method detection limit. Generally, the minimum levels of both STGFAA nd ICP-MS methods are able to measure at water quality criteria levels required by water quality based permitting programs. However, there are exceptions for Graphite Furnace AA. Note that the low ICP-MS quantification levels for arsenic and selenium, are the result of coupling ICP- MS technology with the hydride generation technique. Lead, silver, and thallium are the exceptions for water quality based monitoring for STGFAA, because the minimum levels tend to be slightly above water quality criteria. Based on the experience of HRSD over the last four years, overall method performance, is typically measured by both productivity and quality. Analytically, method performance can be measured by sensitivity, accuracy, and precision. Productivity factors, such as sample through-put or output, are important, because there is a direct effect on laboratory operating costs and data turnaround. With regard to method detection limits for the two methods, the next two slides compare the MDLs listed in Method 1638 and 1639 with HRSD s current MDLs for ICP-MS and STGFAA. These have been determined in reagent water. The MDL for selenium is very low, as a result of coupling ICP-MS with hydride generation technique. For ICP-MS, MDLs are either below or very close to the Method 1638 MDLs. These MDLs have been determined numerous times and these levels can consistently be achieved. For STGFAA, the majority of HRSD current MDLs are slightly higher than those listed in Method In stressing the importance of eliminating unacceptable levels of contamination, laboratory reagent blank data for the two methods can be compared. For most elements, ICP-MS LRBs are an order of magnitude lower than those of STGFAA. These are based on direct instrument readings and therefore may be a result of noise. However, there is a significant difference in the two measurement systems. June

4 For accuracy, a summary of recent laboratory-fortified blank data for both methods is shown. In a pure reagent water matrix, accuracy is excellent for both methods. However, for laboratoryfortified effluent samples, recoveries change. Typically, method acceptance limits can be met. However, for silver and selenium analysis by STGFAA, often the method of standard additions must be employed to overcome sample matrix interferences. As a result, productivity decreases and costs increase due to increased analytical time requirements. For trace metals analysis by ICP-MS, there are fewer matrix interferences for effluent samples and spike recoveries are improved, with the exception of copper. Low copper recoveries may be a result of sulfate levels in effluent. In comparing precision of the two methods, both methods are very good. ICP-MS has slightly improved precision over GFAA. One exception is with selenium, which is 8% for GFAA and 11% for ICP-MS. When productivity of the two methods is compared, there is a significant difference. This is due to the capability of simultaneous, multi-element analysis by the ICP-MS. Although the sample injection time is the same for both, up to 21 elements can be determined simultaneously with ICP- MS. In addition, elemental isotopes can be monitored. On a good day, using overnight operation, up to three elements per day can be determined using STGFAA, compared to less than one day for ICP- MS. Currently, there are STGFAA instruments available on the market that can determine up to six elements simultaneously. This has the potential to improve productivity but not to the level of ICP- MS technology. Higher productivity leads to improved data turnaround and cost effectiveness. Costs can be broken those down into three areas- initial investment costs, ongoing operating costs, and a resulting cost per test. Initial cost for instrumentation, training, and method development are outlined on this slide. These costs are current and may vary, depending on the manufacturer and model. For initial costs, ICP-MS instrumentation is typically twice as costly as that of STGFAA instrumentation. Usually, training and installation are provided with purchase of a new instrument, so these training and development costs are for laboratory hours or labor required to implement the methods. For operating costs, there are several components, including all expenses associated with the ongoing analysis. Operating costs include all labor costs calculated from time requirements, salaries and fringe benefits, material and supply costs, and operational overhead calculated from utility and other facility operating costs. The evaluation for annual operating costs is based on 10,000 results per year. This will vary, depending on the workload. Based on 10,000 results per year, the operating costs of the ICP-MS method is approximately one-third less than that of GFAA. The higher service contract, maintenance and material and supply costs are outweighed by the productivity of ICP-MS. When these two components are combined and evaluated over the long term, approximately $50,000 per year can be saved with a workload of 10,000 results per year, after the first year, using ICP-MS for trace metal analysis. Therefore, initial investment costs can be recovered in a three to five-year period, depending on workload. In summary, application of ICP-MS for trace metal analysis has many advantages, including a lower cost per test, improved data turnaround, and increased reliability, all which are important to the data user and customer. For the laboratory, the enhanced method performance of ICP-MS, low June 1999

5 detection capabilities, and a wide linear concentration range, this technology offers reliability, flexibility, increased control over matrix interferences, and a long-term cost savings. The disadvantages of ICP-MS are few. Initial investment and ongoing maintenance costs are higher. These disadvantages are quickly outweighed by high productivity. Because method approval is pending, currently ICP-MS cannot be utilized to its full potential. In addition, coupling ICP-MS with the hydride technique for arsenic and selenium analysis can provide additional benefits. Based on HRSD s experience, ICP-MS is overall superior, and we look forward to method promulgation in the near future. Thank you. Finally, I would like to acknowledge HRSD s dedicated analysts. QUESTION AND ANSWER SESSION MR. TELLIARD: Any questions? MR. CIARLEGLIO: Peter Ciarleglio from Dames & Moore. I was wondering if...i don't know how variable the waters that you are measuring in there are, but if you did have samples that were extremely high in metals or anything like that, what experience do you have? Can it shut the ICP-MS down for a while because of carryover and you have to restart analyses after cleaning up or any experiences in that regard? MS. HOGG: For contamination control, high concentration level samples are separated from low level sample analysis. At this point, high level samples are not analyzed using ICP-MS due to the large workload associated with NPDES monitoring. The ICP method is used for analysis of higher concentration samples. MR. CIARLEGLIO: So, you pre-screen your samples? MS. HOGG: No, higher level concentration samples are primarily associated with pretreatment program monitoring. ICP method will meet regulatory limits for this program and is more appropriate based on the historical database. MR. CIARLEGLIO: Okay. In a commercial lab, you don't always know which samples are going to be high ahead of time. MS. HOGG: Yes.. MR. CIARLEGLIO: Then, I did have a second question. In studies that I had read before in papers dealing with, well, particularly, lead and zinc, even when they claimed to use clean room techniques and stuff like that, they gave average zinc blank levels that were like 2 and 3 ug/l, and I notice there you had them down around 0.1 or 0.2. Do you think you have got your zinc contamination under control that well, or does it flare up sometimes? June

6 MS. HOGG: Zinc will flare up from time to time. Over the past five years of experience with clean techniques, both zinc and lead can be a problem. The exposure to air and the handling by personnel are the two key components in controlling this problem. When sample handling by personnel is fully minimized the problem can be controlled. MR. CIARLEGLIO: Yeah, you mentioned cosmetics. I think zinc is kind of a component of a lot of cosmetics. MS. HOGG: It is. use? MR. JAGESSAR: Patrick Jagessar, New York City DEP. May I ask what ICP/MS do you MS. HOGG: Fisons Plasmaquad. MR. JAGESSAR: Do you use internal standards in your analysis? MS. HOGG: Yes, we do. That is required by the method. MR. JAGESSAR: Do you do mercury by the ICP/MS? MS. HOGG: No, we don't. We have a mercury analyzer with an atomic fluorescence capability. MR. JAGESSAR: What about marine water? You do marine water, ambient water, marine? MS. HOGG: No, we do not have extensive experience with marine water. We have anlayzed surface water, groundwater, lake water, some river water but this is a very minor part of the workload. MR. JAGESSAR: Thank you. MR. TELLIARD: Thank you, Paula June 1999

7 1999 EPA 22nd Annual Conference on Analysis of Pollutants in the Environment APPLICATION OF ICP-MS TECHNOLOGY FOR TRACE METALS ANALYSIS Paula Hogg, Patty Lee Hampton Roads Sanitation District Virginia Beach, Virginia June

8 TOPICS COMPARISON OF ICP-MS & STGFAA METHODS & APPLICATIONS PERFORMANCE COST June 1999

9 METHODS & APPLICATIONS APPLICATIONS Ambient Monitoring In-Stream Studies WQ Criteria NPDES Regulatory Groundwater ULTRA- CLEAN < 1 ppb CLEAN 1-20 ppb METHODS Pre-Concentration ICP-MS GFAA ICP-MS POTW Process Industrial TRACE METALS > 20 ppb ICP-MS FAA June

10 METALS ANALYSIS WORKLOAD ICP STGFAA FAA PRETREATMENT PRETREATMENT 40% 20% NPDES ICP STGFAA ICP-MS 40% SPECIAL PROGRAMS WQ MONITORING ICP STGFAA ICP-MS FAA June 1999

11 ANALYTICAL CONCENTRATION RANGES Element STGFAA ICP-MS Cd ppb 0.02 ppb ppm Cu ppb 0.08 ppb ppm Cr ppb 0.02 ppb ppm Pb ppb ppb -100 ppm Ni ppb 0.02 ppb ppm Ag ppb 0.02 ppb ppm Zn ppb 0.2 ppb ppm June

12 METHOD DYNAMIC RANGE STGFAA ICP-MS ppt ppb ppm Concentration June 1999

13 EPA METHODS -STGFAA Elements Series Sb Y Y Y Cd Y Y Y Cu Y Y N Cr Y Y Cr 3+ Pb Y Y N Ni Y Y N Ag Y Y N Tl Y Y N Zn Y Y N As Y Y N Se Y Y Y June

14 EPA METHODS -ICP-MS Elements Sb Y Y Cd Y Y Cu Y Y Cr Y N Pb Y Y Ni Y Y Ag Y Y Tl Y Y Zn Y Y As Y N Se Y Y June 1999

15 200.8, AND 1600 METHODS Sample Concentration HCl/HNO 3 Digestion and Hydrogen/Argon Purge 1600 Series Performance Based Clean Techniques-Sampling & Analysis WQ Criteria Levels Additional Quality Control June

16 CLEAN TECHNIQUES Minimize Exposure to Air Enclosed Sampling Equipment Clean Area/ Clean Benches Limit Traffic Clean Techniques Equipment Preparation Glassware Preparation Containers Clean Hands/Dirty Hands Capsule Filtration Reagent Purity June 1999

17 June

18 Se Cr 22 nd Annual Conference METHOD DETECTION LIMITS ug/l Sb Cd Cu Pb Ni Ag Tl Zn As ICP-MS STGFAA June 1999

19 MINIMUM LEVELS (ug/l) Element STGFAA WQ Criteria ICP-MS Sb Cd Cr Cu Pb Ni Ag Tl Zn As-Hydride (As III) 0.05 Se-Hydride June

20 MINIMUM LEVELS ug/l HRSD 1638 WQ Criteria HRSD 1639 Cd Pb Ag Tl June 1999

21 METHOD PERFORMANCE MDL STUDIES ACCURACY Lab Fortified Blanks Lab Fortified Matrix PRECISION PRODUCTIVITY June

22 METHOD 1638 DETECTION LEVELS ug/l EPA Method 1638 HRSD1638 Hydride Sb Cd Cu Pb Ni Se Ag Tl Zn June 1999

23 METHOD 1639 DETECTION LEVELS ug/l EPA 1639 HRSD Sb Cd Cr III Ni Se Zn June

24 LABORATORY REAGENT BLANK DATA ug/l 1 ICP-MS STGFAA Sb Cd Cu Pb Ni Se Ag Zn June 1999

25 ACCURACY- Laboratory Fortified Blanks ELEMENT STGFAA ICP-MS Mean Range Mean Range Ag Cd Cu Pb Se 102% % 107% % 99% % 100% % 97% % 100% % 98% % 106% % 102% % 96% 91-99% (Hydride) June

26 ACCURACY- Laboratory Fortified Matrix ELEMENT STGFAA ICP-MS Mean Range Mean Range Ag Cd Cu Pb Se 81% % 109% % 91% % 101% % 80% % 92% 87-96% 94% 90-97% 81% 73-92% 93% 87-99% 88% % (Hydride) June 1999

27 PRECISION - EFFLUENT RPD - % 15 STGFAA ICP/MS 10 8 % 5 0 Cd Cu Pb Ag Se June

28 PRODUCTIVITY METHOD SAMPLE RESULTS STGFAA 6 Minutes 1 Element ICP-MS 6 Minutes 21 Elements June 1999

29 COSTS INITIAL COSTS ANNUAL OPERATING COSTS COST/TEST June

30 INITIAL COSTS ITEM INSTRUMENT TRAINING DEVELOPMENT MDL IPR TOTAL FURNACE AA $ 70,000 $ 5,000 $ 1,600 $ 1,600 $ 78,200 ICP/MS $ 150,000 $ 5,000 $ 500 $ 500 $ 156, June 1999

31 OPERATING COSTS - COMPONENTS LABOR Analytical Administrative Data Management Fringe MATERIALS & SUPPLIES Service Contract OVERHEAD June

32 ANNUAL OPERATING COSTS 10,000 Results/Year ITEM FURNACE AA ICP-MS SERVICE CONTRACT MATERIALS & SUPPLIES LABOR OVERHEAD TOTAL COST/RESULT $ 10,000 $ 5,000 $ 104,000 $ 28,000 $ 147,000 $ $ 15,000 $ 9,000 $ 56,000 $ 19,000 $ 99,000 $ June 1999

33 LONG TERM OPERATING COSTS $250,000 $200,000 $220,000 GFAA ICP-MS $150,000 $100,000 $50,000 $ YEAR June

34 ICP/MS ADVANTAGES Low Multi-Element Analytical Cost Sensitivity Data Reliability Instrument Stability Data Turnaround Productivity Flexibility June 1999

35 ICP/MS DISADVANTAGES High Initial Cost Increased Instrument Maintenance High Service Contract Costs Method Approval Pending (40 CFR Part 136) ICP/MS-Hydride Methods Needed for As & Se June

36 ACKNOWLEDGEMENTS Cynthia Bato, Chemist Rolando Castro, Specialist HRSD Central Environmental Laboratory P.O. Box 5911 Virginia Beach, VA June 1999

How To Use The Nexion 300

How To Use The Nexion 300 APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of

More information

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure

More information

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,

More information

Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist

Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist Application of ICP-MS as a Tool for Environmental Monitoring Bill Spence Environmental Market Specialist Introduction Knowledge of the concentration of metallic species in the environment has been important

More information

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy

More information

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development

More information

Elemental Scientific. Stability. Elemental Scientific

Elemental Scientific. Stability. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast inline

More information

Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT

Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.

More information

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu

More information

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth

More information

This example of a completed sampling plan worksheet has been included to illustrate the information

This example of a completed sampling plan worksheet has been included to illustrate the information APPENDIX B EXAMPLE OF A COMPLETED SAMPLING PLAN WORKSHEET This example of a completed sampling plan worksheet has been included to illustrate the information necessary to document a sampling program for

More information

VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol

VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol Page 1 van 36 pages VALIDATION REPORT Investigation of the major interferences for proposed Horizontal ICP AES method Identification code Validation in accordance with Reference method Scope Reference

More information

HS 1003 Part 2 HS 1003 Heavy Metals Test

HS 1003 Part 2 HS 1003 Heavy Metals Test HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: Method No.: Lead Dust Wipes Modification of the Automated AAII-25 Colorimetric method Nitric and Hydrochloric Acid Digestion followed by Atomic Absorption

More information

Universal Data Acquisition (UDA)

Universal Data Acquisition (UDA) Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published

More information

Removing Thallium from Industrial FGD Scrubber Water with Sorbster Adsorbent Media

Removing Thallium from Industrial FGD Scrubber Water with Sorbster Adsorbent Media Case History MAR Systems Inc. Removing Thallium from Industrial FGD Scrubber Water with Sorbster Adsorbent Media Trace thallium levels in process and wastewater streams pose a human toxicity threat. Tidwell

More information

MILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE

MILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE MILESTONE CLEAN CHEMISTRY LINE A Breakthrough in Reduction and Control of the Analytical Blank DUOPUR TRACECLEAN ULTRA-TRACE INSERTS MILESTONE MILESTONE CLEAN CHEMISTRY LINE An innovative and complete

More information

Sample Analysis Design Isotope Dilution

Sample Analysis Design Isotope Dilution Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS Attention: Angela Mazza / Kathleen Wilson Catholic Education Centre 80 Sheppard Ave E North York, ON M2N 6E8 Your Project #: ANNUNCIATION Your C.O.C. #: na Report #: R3046860 Version: 1 MAXXAM JOB #: B488250

More information

Elemental Analyses by ICP-AES

Elemental Analyses by ICP-AES Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to

More information

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially

More information

Flow Injection Analysis

Flow Injection Analysis Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since

More information

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters 1. SCOPE and APPLICATION 1.1 Gravimetric analysis is used to determine total suspended

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

Clean Laboratory Techniques

Clean Laboratory Techniques Clean Laboratory Techniques Thursday November 18 th, 2010 Vanaja Sivakumar, Ph.D. Vice President, Inorganic Manufacturing Housekeeping Everyone in attendance will receive a copy of the slide deck The webinar

More information

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE Department of Bioresources Engineering Rm 111 Worrilow Hall 531 S. College Ave. Newark, DE 19716-2140

More information

Stream Monitoring at Tumacácori NHP

Stream Monitoring at Tumacácori NHP Stream Monitoring at Tumacácori NHP Santa Cruz River Researcher s Day Tucson Community Service Center March 27 2009 Presented by E.L.Gwilliam Overview why, what, where, when and how Stream Protocol Importance

More information

DATA QUALITY OBJECTIVES

DATA QUALITY OBJECTIVES The Wastewater Treatment Plant Operators Guide to Biosolids Sampling Plans CHAPTER 5 DATA QUALITY OBJECTIVES Goals of the Sampling Plan ESSENTIAL ELEMENTS OF A SAMPLING PLAN Description of the Facility

More information

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

ANALYTICAL DETECTION LIMIT GUIDANCE. & Laboratory Guide for Determining Method Detection Limits

ANALYTICAL DETECTION LIMIT GUIDANCE. & Laboratory Guide for Determining Method Detection Limits ANALYTICAL DETECTION LIMIT GUIDANCE & Laboratory Guide for Determining Method Detection Limits Wisconsin Department of Natural Resources Laboratory Certification Program April 1996 PUBL-TS-056-96 introduction

More information

Standard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry

Standard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Standard Operation Procedure Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu

More information

QA/QC. Standard methods in field. QA/QC in the field. QA/QC in analysis

QA/QC. Standard methods in field. QA/QC in the field. QA/QC in analysis QA/QC To identify, measure and keep systematic and random errors under control in planning, sampling, analysis and reporting of data. Quality assurance = how samples and data are collected and managed.

More information

Ecology Quality Assurance Glossary

Ecology Quality Assurance Glossary Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability

More information

Mercury Speciation in Flue Gas from Coal-fired Power Stations

Mercury Speciation in Flue Gas from Coal-fired Power Stations Mercury Speciation in Flue Gas from Coal-fired Power Stations Matthew A Dexter, Warren T Corns and Peter B Stockwell P S Analytical Arthur House Crayfields Industrial Estate Main Road Orpington BR5 3HP

More information

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS 1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years

More information

ANALYTICAL REPORT LABORATORY DETAILS. Telephone Facsimile Email. SGS Reference Report Number Date Reported Date Received

ANALYTICAL REPORT LABORATORY DETAILS. Telephone Facsimile Email. SGS Reference Report Number Date Reported Date Received ANALYTICAL REPORT CLIENT DETAILS Contact Client Address Matthew Behre SGS Metallurgical Services 431 Victoria Street Malaga MALAGA WA 6090 LABORATORY DETAILS Manager Laboratory Address Said Hirad SGS Newburn

More information

Secondary Lead Smelting

Secondary Lead Smelting Secondary Lead Smelting Secondary Lead Smelting Objectives Describe the basic smelting process terms: smelting, refining, and alloying List key chemicals associated with secondary lead smelting Define

More information

Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems

Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES FIMS 100 and 400 Flow Injection Mercury Systems Delivering the Performance You Need The determination and monitoring of sub-ppb concentrations

More information

Importance of Proper Sampling and Preservative Techniques Analysis dependent upon quality of sampling procedure Eliminate contamination Poor sampling

Importance of Proper Sampling and Preservative Techniques Analysis dependent upon quality of sampling procedure Eliminate contamination Poor sampling Proper Sampling and Preservative Techniques for Wastewater Analysis By: Tom Horn Industrial Fluid Management Laboratory/Project Director LICENSES, STATE OF OHIO Class II Wastewater Analyst #2-00-0202 Class

More information

9. Quality Assurance 9.1 INTRODUCTION 9.2 FIELD SAMPLING QUALITY ASSURANCE 9.3 ANALYTICAL QUALITY ASSURANCE. 9.3.1 Internal Quality Control

9. Quality Assurance 9.1 INTRODUCTION 9.2 FIELD SAMPLING QUALITY ASSURANCE 9.3 ANALYTICAL QUALITY ASSURANCE. 9.3.1 Internal Quality Control 9. Quality Assurance Setting The overall goal of a well-designed and well-implemented sampling and analysis program is to measure accurately what is really there. Environmental decisions are made on the

More information

METHOD 200.8 DETERMINATION OF TRACE ELEMENTS IN WATERS AND WASTES BY INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY. Revision 5.

METHOD 200.8 DETERMINATION OF TRACE ELEMENTS IN WATERS AND WASTES BY INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY. Revision 5. METHOD 200.8 DETERMINATION OF TRACE ELEMENTS IN WATERS AND WASTES BY INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY Revision 5.4 EMMC Version S.E. Long (Technology Applications Inc.), T.D. Martin, and

More information

Gold Refining and Coin Manufacturing at the Royal Canadian Mint

Gold Refining and Coin Manufacturing at the Royal Canadian Mint INTERNET Sub title will go here The Application of XRF to Gold Refining and Coin Manufacturing at the Royal Canadian Mint Michael W. Hinds, Ph.D. The Business of the Mint Circulation Coins (Canada & Other

More information

Appendix B. Eugene/Springfield Environmental Services Laboratory Quality Assurance Plan

Appendix B. Eugene/Springfield Environmental Services Laboratory Quality Assurance Plan Appendix B Eugene/Springfield Environmental Services Laboratory Quality Assurance Plan City of Eugene NPDES Stormwater Monitoring Plan May 2011 Users of this document are responsible to ensure it is the

More information

Rapid Thermophilic Digestion Technology

Rapid Thermophilic Digestion Technology Rapid Thermophilic Digestion Technology Biomax Technologies Pte Ltd Mr Sim Eng Tong, CEO Block 4 Kaki Bukit Avenue 1, #05-07/08, Singapore 417939 Phone: + 65 6274 8606 Fax: + 65 6274 8607 Website: www.biomaxtech.com

More information

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,

More information

EPA NATIONAL LEAD LABORATORY ACCREDITATION PROGRAM. LABORATORY QUALITY SYSTEM REQUIREMENTS (LQSR) REVISION 3.0 (November 05, 2007)

EPA NATIONAL LEAD LABORATORY ACCREDITATION PROGRAM. LABORATORY QUALITY SYSTEM REQUIREMENTS (LQSR) REVISION 3.0 (November 05, 2007) EPA NATIONAL LEAD LABORATORY ACCREDITATION PROGRAM LABORATORY QUALITY SYSTEM REQUIREMENTS (LQSR) REVISION 3.0 (November 05, 2007) The requirements in this document apply to all lead testing laboratories

More information

OLIVÉR BÁNHIDI 1. Introduction

OLIVÉR BÁNHIDI 1. Introduction Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY

More information

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014 SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014 This experiment explores quality assurance practices which are commonly

More information

FMA-80 Fluorescence Mercury Analyzer

FMA-80 Fluorescence Mercury Analyzer FMA-8 Fluorescence Mercury Analyzer www.milestonesrl.com MILESTONE H E L P I N G C H E M I S T S The new Milestone FMA-8 The mercury problem A wide range of mercury analyzers Milestone offers a wide range

More information

Analyses on copper samples from Micans

Analyses on copper samples from Micans PDF rendering: DokumentID 1473479, Version 1., Status Godkänt, Sekretessklass Öppen Analyses on copper samples from Micans P. Berastegui, M. Hahlin, M. Ottosson, M. Korvela, Y. Andersson, R. Berger and

More information

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly

More information

Innovative Technology Verification Report

Innovative Technology Verification Report United States Office of Research and EPA/600-R-03/149 Environmental Protection Development May 2004 Agency Washington DC 20460 Innovative Technology Verification Report Field Measurement Technology for

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic Absorption Spectrometry

Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic Absorption Spectrometry application Note Atomic Absorption Authors Jijun Yao Renkang Yang Jianmin Chen PerkinElmer, Inc. Shelton, CT 06484 USA Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic

More information

METHOD 200.7 DETERMINATION OF METALS AND TRACE ELEMENTS IN WATER AND WASTES BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY

METHOD 200.7 DETERMINATION OF METALS AND TRACE ELEMENTS IN WATER AND WASTES BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY METHOD 200.7 DETERMINATION OF METALS AND TRACE ELEMENTS IN WATER AND WASTES BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY Revision 4.4 EMMC Version USEPA-ICP Users Group (Edited by T.D. Martin

More information

Introduction. The following definitions may help you better understand the components of the data report.

Introduction. The following definitions may help you better understand the components of the data report. Introduction This document is designed to help our clients understand the quality control requirements and limitations of data reporting. There are three sections to this document. The first section will

More information

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy In support of the toy industry, many laboratories are being called upon to analyze a wide variety of

More information

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications

More information

LED Life-Cycle Assessment & Environmental Testing

LED Life-Cycle Assessment & Environmental Testing LED Life-Cycle Assessment & Environmental Testing DOE Solid State Lighting Booth Presentation LightFair 2013 Jason Tuenge Pacific Northwest National Laboratory Program Name or Ancillary Text eere.energy.gov

More information

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009 Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production

More information

Medical Cannabis Laboratory Approval Program

Medical Cannabis Laboratory Approval Program Medical Cannabis Laboratory Approval Program Application Process and Required Documentation After the publication of the Medical Cannabis Laboratory Application, currently expected by February 16, 2015,

More information

PERFORMANCE REPORT. Quality Assurance Program - January to June 2011

PERFORMANCE REPORT. Quality Assurance Program - January to June 2011 PERFORMANCE REPORT Quality Assurance Program - January to June 2011 The following discussion is prepared in accordance with the requirements for a Semiannual Performance Report on the Quality Assurance

More information

Triad Characterization of Soil Contamination at Former Small Arms Training Ranges

Triad Characterization of Soil Contamination at Former Small Arms Training Ranges Triad Characterization of Soil Contamination at Former Small Arms Training Ranges Kira P Lynch, MS USACE, Seattle District National Site Assessment Symposium San Diego, CA, June 28, 2004 1 Site Background

More information

How To Test For Contamination In Large Volume Water

How To Test For Contamination In Large Volume Water Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed

More information

4.2 Bias, Standards and Standardization

4.2 Bias, Standards and Standardization 4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

Storage and treatment of cathode ray tubes (CRTs) and display equipment containing CRTs

Storage and treatment of cathode ray tubes (CRTs) and display equipment containing CRTs Storage and treatment of cathode ray tubes (CRTs) and display equipment containing CRTs Quick guide 874_11 Issued 27/10/2011 What s this document about? Who does this apply to? This document covers the

More information

Advancing the State of Analytical Science. for air, water and soil analysis. and beyond

Advancing the State of Analytical Science. for air, water and soil analysis. and beyond Advancing the State of Analytical Science for air, water and soil analysis and beyond Full Spectrum of Services SEMICONDUCTOR SECTOR ENVIRONMENTAL INDUSTRY HIGH TECHNOLOGY R&D Process Analysis Substrate,

More information

DATA VALIDATION STANDARD OPERATING PROCEDURES FOR CONTRACT LABORATORY PROGRAM ROUTINE ANALYTICAL SERVICES

DATA VALIDATION STANDARD OPERATING PROCEDURES FOR CONTRACT LABORATORY PROGRAM ROUTINE ANALYTICAL SERVICES DATA VALIDATION STANDARD OPERATING PROCEDURES FOR CONTRACT LABORATORY PROGRAM ROUTINE ANALYTICAL SERVICES UNITED STATES ENVIRONMENTAL PROTECTION AGENCY REGION IV SCIENCE AND ECOSYSTEM SUPPORT DIVISION

More information

Determination of elements by ICP-AES and ICP-MS

Determination of elements by ICP-AES and ICP-MS September 2003 HORIZONTAL - 19 Determination of elements by ICP-AES and ICP-MS Henk J van de Wiel National Institute of Public Health and the Environment (RIVM) Bilthoven, The Netherlands 1 HORIZONTAL

More information

Laboratory water purification systems

Laboratory water purification systems Laboratory water purification systems Choosing a water purification system Application ISO 3696 classification Suggested model and configuration Page Feed water for labware washers Sterifeed 20 Feed water

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007

Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007 Proficiency Testing In A Global Analytical Environment Benefits & Challenges AOAC September 2007 1 The Coca-Cola Company The Coca-Cola Company is the largest manufacturer, distributor and marketer of nonalcoholic

More information

LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e

LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e 1 SIDE BY SIDE COMPARISON TABLE FOR THE DETERMINATION OF CYANIDE BY UV DIGESTION AND AMPEROMETRIC DETECTION FIA METHOD - 10-204-00-5-D (ASTM D7511-09 e ) TOPIC Method ASTM D7511-09e LACHAT METHOD NUMBER

More information

WATER QUALITY DEPARTMENT JOB DESCRIPTION: LAB TECHNICIAN HAMPTON ROADS SANITATION DISTRICT I. FILE GUIDE REFERENCE

WATER QUALITY DEPARTMENT JOB DESCRIPTION: LAB TECHNICIAN HAMPTON ROADS SANITATION DISTRICT I. FILE GUIDE REFERENCE I. FILE GUIDE REFERENCE WATER QUALITY DEPARTMENT JOB DESCRIPTION: LAB TECHNICIAN HAMPTON ROADS SANITATION DISTRICT A. DIVISION : CEL B. GRADE : 5d-5 C. IMMEDIATE SUPERVISOR : Chemist Manager/Laboratory

More information

Developing Quality Assurance Project Plans using Data Quality Objectives and other planning tools. Developing QAPPs 5/10/2012 1

Developing Quality Assurance Project Plans using Data Quality Objectives and other planning tools. Developing QAPPs 5/10/2012 1 Developing Quality Assurance Project Plans using Data Quality Objectives and other planning tools 1 Introductions 2 Agenda I. Developing a QAPP II. Systematic Planning using Data Quality Objectives III.

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY

METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY 1.0 SCOPE AND APPLICATION 1.1 Inductively coupled plasma-atomic emission spectrometry (ICP-AES) determines trace elements, including

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

Sludge Treatment Facility Stack Gas Monitoring Report February 2016

Sludge Treatment Facility Stack Gas Monitoring Report February 2016 I. INTRODUCTION This Monthly Report aims to provide a summary of environmental performance of the Sludge Treatment Facility (STF) over the monitoring period, which includes the air emission data collected

More information

CNSolution TM 9310 On-line Cyanide Analyzer. Product Release Synopsis. I Product Description/Function. 1 3 Cyanide Recycle

CNSolution TM 9310 On-line Cyanide Analyzer. Product Release Synopsis. I Product Description/Function. 1 3 Cyanide Recycle CNSolution TM 9310 Product Release Synopsis I Product Description/Function The CNSolution 9310 is designed to measure and control cyanide used in hydrometallurgical leaching of gold and silver from ores.

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so

More information

DMA-80 Direct Mercury Analyzer

DMA-80 Direct Mercury Analyzer DMA-80 Direct Mercury Analyzer Milestone Srl Milestone Srl 1 Introduction Mercury in the atmosphere can be from manmade sources (coal-fired power plants, municipal incinerators, industrial boilers) or

More information

Potassium Aluminium Silicate (Mica) Chemical and Technical Assessment (CTA) Prepared by Daniel E. Folmer, Ph.D., and reviewed by Madduri V. Rao, Ph.D.

Potassium Aluminium Silicate (Mica) Chemical and Technical Assessment (CTA) Prepared by Daniel E. Folmer, Ph.D., and reviewed by Madduri V. Rao, Ph.D. Potassium Aluminium Silicate (Mica) Chemical and Technical Assessment (CTA) Prepared by Daniel E. Folmer, Ph.D., and reviewed by Madduri V. Rao, Ph.D. 1. Summary Potassium aluminium silicate (PAS) was

More information

Study of Arsenic Levels in Private Wells in the Great Miami River Watershed

Study of Arsenic Levels in Private Wells in the Great Miami River Watershed 2011-06 Study of Arsenic Levels in Private Wells in the Great Miami River Watershed Introduction Arsenic is one of the most common elements in the earth s crust and is commonly found in many locations

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS

aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS aurora M90 A new era in ICP-MS Innovation with Integrity ICP-MS Your Partner in ICP-MS Solutions Bruker continues to find new and novel ways to meet your changing needs. As a leader in elemental analysis

More information

Soil Toxicity in post-katrina New Orleans

Soil Toxicity in post-katrina New Orleans Soil Toxicity in post-katrina New Orleans William Kern 1, Scott Bell 2, Amy Kracker Selzer 3 1 Center for Environmental Studies, Brown University 2 Population Studies and S4, Brown & University of Saskatchewan

More information

limit to percentage levels, a limited working range translates

limit to percentage levels, a limited working range translates Figure 1. The AutoPrep 50 Automatic Dilution System. Figure 2. Schematic diagram of the AutoPrep 50. Flame atomic absorption spectroscopy (flame AA) is still the most frequently used instrumental technique

More information

Analytical Data Package Reports

Analytical Data Package Reports Analytical Data Package Reports TriMatrix offers a variety of analytical data packages formatted to meet the reporting needs and varying data quality objectives of its clients. The data packages offered

More information

RoHS Test Methods. Presented by: Bruce Peterson

RoHS Test Methods. Presented by: Bruce Peterson RoHS Test Methods Presented by: Bruce Peterson Motivation for Presentation The EU RoHS directive takes effect July 1, 2006 Customer confusion about - Scope of directive How much testing needs to be performed

More information

ANALYTICAL SUMMARY REPORT July 06, 2011 303 Irene St Helena, MT 59601 Workorder No.: H11060340 Project Name: Beal Mtn Quote ID: H634 - Beal 2011 Site Wide Monitoring Energy Laboratories Inc Helena MT received

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

How to read your Oil Analysis Report

How to read your Oil Analysis Report How to read your Oil Analysis Report A better understanding Petroleum Technologies Group, LLC. 4665 Broadmoor S.E. Ste#15 Grand Rapids, MI 49512 Phone # 616/698 9399 Fax # 616/698 9395 www.oil lab.com

More information