MALDI-TOF Mass Spectrometry

Similar documents
ProteinChip Energy Absorbing Molecules (EAM)

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100)

Electrospray Ion Trap Mass Spectrometry. Introduction

An In-Gel Digestion Protocol

Plant Genomic DNA Extraction using CTAB

PrepTip. Reverse Phase PrepTip User Guide

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C

Mass Spectrometry. Overview

[ Care and Use Manual ]

Pesticide Analysis by Mass Spectrometry

m/z

Benchtop Mitochondria Isolation Protocol

Covalent Conjugation to Cytodiagnostics Carboxylated Gold Nanoparticles Tech Note #105

SUCRALOSE. White to off-white, practically odourless crystalline powder

Wizard DNA Clean-Up System INSTRUCTIONS FOR USE OF PRODUCT A7280.

Molecular Cell Biology. Prof. D. Karunagaran. Department of Biotechnology. Indian Institute of Technology Madras

HiPer Ion Exchange Chromatography Teaching Kit

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Functional Data Analysis of MALDI TOF Protein Spectra

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

F321 THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important are... in the nucleus of an atom

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

Classic Immunoprecipitation

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Genomic DNA Extraction Kit INSTRUCTION MANUAL

Chapter 13 Solution Dynamics. An Introduction to Chemistry by Mark Bishop

Introduction to Proteomics 1.0

50 g 650 L. *Average yields will vary depending upon a number of factors including type of phage, growth conditions used and developmental stage.

CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING

Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass

PCR and Sequencing Reaction Clean-Up Kit (Magnetic Bead System) 50 preps Product #60200

6 Characterization of Casein and Bovine Serum Albumin

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

13C NMR Spectroscopy

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

AMD Analysis & Technology AG

A Beer s Law Experiment

Supplementary Materials and Methods (Metabolomics analysis)

TIANquick Mini Purification Kit

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Laboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik

PROTEINS (LOWRY) PROTOCOL

Optimal Conditions for F(ab ) 2 Antibody Fragment Production from Mouse IgG2a

Honors Chemistry: Unit 6 Test Stoichiometry PRACTICE TEST ANSWER KEY Page 1. A chemical equation. (C-4.4)

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

ab Protein Sumoylation Assay Ultra Kit

Background Information

Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)

The Theory of HPLC. Gradient HPLC

Element of same atomic number, but different atomic mass o Example: Hydrogen

Paper 1 (7404/1): Inorganic and Physical Chemistry Mark scheme

Experiment 8 Synthesis of Aspirin

Chapter 1: Moles and equations. Learning outcomes. you should be able to:

Procedure for RNA isolation from human muscle or fat

How to Grow Single Crystals for X-ray Analysis by Solution Crystallisation

LC-MS/MS for Chromatographers

Oasis HLB Cartridges and 96-Well Plates

Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?

Hands-On Labs SM-1 Lab Manual

DNA SPOOLING 1 ISOLATION OF DNA FROM ONION

How To Use An Acquity Qda Detector

Human serum albumin (HSA) nanoparticles stabilized with. intermolecular disulfide bonds. Supporting Information

ISOLATION OF CAFFEINE FROM TEA

Recrystallization II 23

LAB 14 ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA)

Affi-Prep Protein A Matrix Instruction Manual

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry

Agencourt RNAdvance Blood Kit for Free Circulating DNA and mirna/rna Isolation from μL of Plasma and Serum

Separation by Solvent Extraction

SYNTHESIS AND ANALYSIS OF A COORDINATION COMPOUND OF COPPER

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

UltraClean Soil DNA Isolation Kit

Peptide Synthesis Zheng Miao* and Zhen Cheng

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

The Empirical Formula of a Compound

General Chemistry I (FC, 09-10) Lab #3: The Empirical Formula of a Compound. Introduction

INSTRUCTIONS Edition AC

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis

GRS Plasmid Purification Kit Transfection Grade GK (2 MaxiPreps)

CHE334 Identification of an Unknown Compound By NMR/IR/MS

Marmara Üniversitesi Fen-Edebiyat Fakültesi Kimya Bölümü / Biyokimya Anabilim Dalı PURIFICATION AND CHARACTERIZATION OF PROTEINS

Protease Peptide Microarrays Ready-to-use microarrays for protease profiling

Amount of Substance.

Getting the most from this book...4 About this book...5

Expectations for GC-MS Lab

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Chapter 5, Lesson 3 Why Does Water Dissolve Salt?

Experimental procedures. Solid phase peptide synthesis (SPPS)

THE ACTIVITY OF LACTASE

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

Novel Method for Solid Phase Peptide Synthesis Using Microwave Energy

Transcription:

MALDI-TOF Mass Spectrometry by Reena Zutshi I. Introduction Mass spectrometry is a powerful tool used for studying the es of atoms, molecular fragments, and molecules. In general, molecules in the gas phase (or species desorbed from a condensed phase) are ionized, and the ions are then accelerated by an electric field and separated on the basis of their -to-charge ratio (m/z). For an ion with a charge of +1, m/z will be numerically equal to the. The electron ionization that converts molecules to ions can not only remove electrons, but may impart so much energy that the molecule fragments. The molecular ions are deflected by a magnet as they travel through the analyzer tube toward the detector. A spectrum is a chromatogram presenting the signal intensity (y-axis) versus m/z (x-axis). See Figure 1 below. The peak intensities are expressed as a percentage of the most intense signal (the base beak). A time-of-flight (TOF) spectrometer separates ions with identical kinetic energy but different m/z, since lighter ions travel faster than heavier ones. Interpretation of the fragmentation patterns and isotopic peaks can provide valuable clues for the structure determination of organic molecules. Mass spectrometry of proteins is primarily used for determination of molecular ; therefore, high-energy ionization that leads to fragmentation is not desirable. Matrix-assisted laser desorption/ionization spectrometry (MALDI-MS) emerged as an effective analytical tool about twelve years ago. MALDI provides a 'soft' ionization source that prevents decomposition of fragile biomolecules, producing primarily singly charged, intact protein ions. In MALDI, the sample (such as an aliquot of purified peptide) is embedded in an excess of matrix, which is a solution of an ultraviolet absorbing compound. When the liquid evaporates, the sample consists of dry crystals of sample mixed with matrix. When this mixture is irradiated with a laser, the matrix assists in the volatilization and ionization of the analyte. The ions produced with the MALDI technique are analyzed using a time-of-flight (TOF) spectrometer, which is characterized by a high ion throughput and therefore high sensitivity. By knowing the molecular of your peptide or protein, you can determine whether it is present in a given sample by examining the spectrum for peaks with the right.the Chemistry department has a MALDI-TOF spectrometer located in the instrument center. Chemical Biology students will be trained by the instrument center staff and their TA. As always, your TA will provide specific guidance and suggest modifications to the protocol below as necessary. Figure 1. Mass Spectrum of bovine serum albumin obtained using MALDI-TOF. Figure adapted from the American Society for Mass Spectrometry website (http://www.asms.org/whatisms/p12.html). II. Sample preparation for MALDI-TOF The general protocol for preparing samples for MALDI is as follows:

Select matrix Prepare matrix Prepare sample Mix sample and matrix Load sample/matrix on a clean sample plate Dry A. Selecting a matrix Selection of proper matrix is important for getting a good spec, since the matrix plays a key role in ionization. The chart below suggests guidelines to consider. Matrix Sinapinic Acid (3,5-dimethoxy-4-hydroxy cinnamic acid) CHCA (a-cyano-4-hydroxycinnamic acid) THAP (2,4,6-Trihydroxyacetophenone) HPA (3-hydroxypicolinic acid) in diammonium citrate Application Peptides and proteins greater than 10 kda in Peptides and proteins less than 10 kda in Small oligonucleotides less than 3.5 kda in Large nucleotides greater than 3.5 kda in I have found that app derived molecules and other hydrophobic peptides work well in sinapinic acid. For smaller peptides (18-24 mer), CHCA can be used. B. Preparing stock solutions of matrix 1. Sinapinic Acid: In an eppendorf tube, weigh out 10 mg of sinapinic acid. Add 600 µl of deionized water, 100 µl of 3% TFA and 300 µl of acetonitrile to the matrix. Vortex for 1 minute to disslove, then centrifuge for 1 minute to precipitate any undissolved sinapinic acid. Use only the supernatant for applications. Note: Buffers, salts and detergent retard the ionization of the matrix. As far as possible, avoid using the last two (salts and detergents). If you must use a sample prepared in a buffer, increase the concentration of TFA in the matrix stock to enhance sample ionization. 2. CHCA: In an eppendorf tube, weigh out 10 mg of CHCA. Add 400 µl of deionized water, 100 µl of 3% TFA and 500 µl of acetonitrile to the matrix. Vortex for 1 minute to disslove, then centrifuge for 1 minute to precipitate any undissolved sinapinic acid. Use only the supernatant for applications. Note 1: Buffers, salts and detergent retard the ionization of the matrix. As far as possible, avoid using the last two (salts and detergents). If you must use a sample prepared in a buffer, increase the concentration of TFA in the matrix stock to enhance sample ionization. Note 2: If the dry matrix is a mustard-yellow color instead of bright yellow, it may contain impurities. To purify, dissolve CHCA in warm ethanol. Filter and add 2 volumes of deionized water. Let the solution stand in the refrigerator for 2 hrs. Filter and wash the precipitate with cold water. 3. THAP:

Make a 50 mg/ml solution of diammonium citrate in deionized water. Dissolve 10 mg of THAP in 50% acetonitrile/deionized water. Combine 8:1:: THAP solution:diammonium citrate solution. Note: For oligonucleotide applications, do not use HPLC grade water for sample preparation. Use deionized water only. 4. HPA: Make a 50 mg/ml solution of diammonium citrate in deionized water. Dissolve 50 mg of THAP in 50% acetonitrile/deionized water. Combine 8:1:: THAP solution:diammonium citrate solution. Note: For oligonucleotide applications, do not use HPLC grade water for sample preparation. Use deionized water only. C. Preparing Sample Stock for MALDI Stock solutions of samples should be at the following concentrations before adding to the matrix solution: Peptides and proteins: 1-100 pmol/µl (Lower concentration for smaller peptides and higher concentration for proteins and larger peptides, i.e. > 5 kda) Oligonucleotides: 10-100 pmol/l Samples should be preferably dissolved in water. If insoluble in water, add acetonitrile to the solution (up to 50%) and then 0.1% TFA to increase solubility. Consult your TA for specific guidelines on sample stock solutions. 1. Mixing Sample and Matrix Mix 1 µl of sample and 9 µl of matrix in a microcentrifuge tube for a final concentration of 0.1-10 pmol/µl. Your particular sample may require a different ratio of sample to matrix; consult your TA for guidance. D. Loading sample/matrix on the plate Load 1-2 µl of the sample/matrix solution onto the plate. Make sure that the sample is placed in the center of the cell you are loading onto and note the cell # for future reference (a spec sample record table is provided at the end of this section for your convenience). Allow the sample plate to dry. E. Cleaning the sample plate Wipe the sample plate with a Kimwipe soaked in ethanol, then with soap/water, then with water and finally with ethanol again (for each of these I suggest using a Kimwipe soaked in the appropriate solution to wipe the top of the plate only - a wet sample plate can damage the instrument). III. Troubleshooting If you do not see any signal on the spec, it could be due to the following reasons: Possible Cause Possible Solution a) Use a 0.1-10 pmol/µl (final Sample concentration too low conc) of peptides and proteins b) Use at least 1:1 ratio of sample :

Sample concentration too high - sample signal may be suppressed Sample made in buffer - poor crystallization on sample plate Sample contains salt or detergent - poor crystallization on sample plate Matrix is old matrix 1. Dilute the sample 2. Use at least 1:1 ratio of sample : matrix If you must use buffer, use a low ionic strength nitrogen containing buffer - avoid PBS or sulfate buffers Use a higher conc of TFA (upto 1% final conc) to enhance ionization Get rid of salt/detergent on desalting column or by dialysis Make fresh matrix Be sure to avoid: Using organic solvents to dissolve samples - causes the sample to spread out and not crystallize properly. Touching the surface of the sample plate with the pipet tip - causes uneven crystallization.

Mass Spec Sample Record Your Name: Date: Peptide ID Sample/Cell # Matrix Expected Mass Spec Sample Record Your Name: Date: Peptide ID Sample/Cell # Matrix Expected