Ion-Selective Electrode Determination of Fluoride Ion

Similar documents
A Potentiometric Analysis of Fluoride Ion in Toothpaste

9. Analysis of an Acid-Base Titration Curve: The Gran Plot

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

Experiment 6 Titration II Acid Dissociation Constant

Experiment 9 Electrochemistry I Galvanic Cell

Experiment 4 (Future - Lab needs an unknown)

Acid Dissociation Constants and the Titration of a Weak Acid

ph: Measurement and Uses

Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid

Solubility Product Constants

Experiment 17: Potentiometric Titration

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

18 Conductometric Titration

USEPA Electrode Method Method 8156

Determination of the Mass Percentage of Copper in a Penny. Introduction

Determining the Identity of an Unknown Weak Acid

Determiniation of Flouride Ion in Well Waters of Eastwood School District, Wood, County, Ohio

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets

Upon completion of this lab, the student will be able to: H3O + (aq) + A (aq) BH + (aq) + OH (aq) O + ] [H 3. ph = log[h 3.

Phenolphthalein-NaOH Kinetics

Calibration of Glassware

SOLUBILITY, IONIC STRENGTH AND ACTIVITY COEFFICIENTS

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

III. Chemical Kinetics

GA/7 Potentiometric Titration

Measurement by Ion Selective Electrodes

EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy

Chem 131A: Absorbance of Riboflavin

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

Electrochemistry Revised 04/29/15

Acid Base Titration: ph Titration Curve

Acid Base Titrations

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

SOLID STATE CHEMISTRY - SURFACE ADSORPTION

Color, True and Apparent

Electrical Conductivity of Aqueous Solutions

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

Color, True and Apparent

3 The Preparation of Buffers at Desired ph

A Beer s Law Experiment

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Determination of sodium with the ion-selective

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

The Determination of an Equilibrium Constant

EXPERIMENT INTRODUCTION TO INDICATORS AND ACID-BASE TITRATIONS

ph units constitute a scale which allows scientists to determine the acid or base content of a substance or solution. The ph 0

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Appendix C. Vernier Tutorial

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.

Experiment 4 The Relationship of Density and Molarity of an Aqueous Salt Solution

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

AIRMoN, ph and Conductivity Field Measurement

Ion Exchange Determination of Na+ by Displacement and Zn 2+ Using Preconcentration. Reading: Harris pp , 699

ph Measurements of Common Substances

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

TITRATION CURVES, INDICATORS, AND ACID DISSOCIATION CONSTANTS

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

EXPERIMENT 2 THE HYDROLYSIS OF t-butyl CHLORIDE. PURPOSE: To verify a proposed mechanism for the hydrolysis of t-butyl Chloride.

EXPERIMENT # 3 ELECTROLYTES AND NON-ELECTROLYTES

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Measurement and Calibration

EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.

Experiment 6 Coffee-cup Calorimetry

Experiment 8 - Double Displacement Reactions

CHEM 161: Beer s Law and Analysis of a Sports Drink

EXPERIMENT 12 A SOLUBILITY PRODUCT CONSTANT

Analysis of Vitamin C Using Iodine. Introduction

TITRIMETRIC ANALYSIS OF CHLORIDE

Determination of calcium by Standardized EDTA Solution

POTENTIOMETRIC TITRATION OF A WEAK ACID

Colorimetric Determination of Iron in Vitamin Tablets

This value, called the ionic product of water, Kw, is related to the equilibrium constant of water

Galvanic Cells and the Nernst Equation

COMMON LABORATORY APPARATUS

Ion Selective Electrodes

Lab #11: Determination of a Chemical Equilibrium Constant

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Determining Equivalent Weight by Copper Electrolysis

Net ionic equation: 2I (aq) + 2H (aq) + H O (aq) I (s) + 2H O(l)

Calcium Analysis by EDTA Titration

Acid-Base Titrations Using ph Measurements

SOLVOLYSIS OF tert-butyl CHLORIDE: TESTING A MECHANISM

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

Concentrations and Dilutions of Food Dyes

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

SenTix 950 SenTix 980

Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer. NAME: Lab Section: Date: Sign-Off:

Factors Affecting Enzyme Activity

The introduction of your report should be written on the on the topic of the role of indicators on acid base titrations.

ph and Acidity in Wine and Fruit Juice

6 H2O + 6 CO 2 (g) + energy

Experiment 7: Titration of an Antacid

Transcription:

Chemistry 321L Manual Page 23 Ion-Selective Electrode Determination of Fluoride Ion I. Introduction A. Ion-Selective Electrodes The amount of a specific ion contained in an aqueous solution can be determined by direct potentiometric measurement of the voltage of a galvanic cell such as shown below. reference electrode analyte solution ion selective electrode The most common example of this is a ph measurement using a glass membrane sensitive to H +. The ion-selective electrode (ISE) typically consists of an inner reference electrode plus a membrane that provides the interface between the sample solution and the ISE. A potential develops across the membrane that depends on the difference in the activity of a specific ion on each side of the membrane. An internal solution with a fixed concentration (activity) of the analyte ion means that the potential developed across the membrane is related to the analyte activity in the sample solution. The overall measured cell potential can be expressed as where E outer ref is the potential of the outer reference electrode; E inner ref is the potential of the inner reference electrode; E junc represents the various junction potentials that develop at liquid junction in the cell; E ISE is the potential developed across the ion-selective membrane. If the measurements are made with very little current flowing in the cell, the reference electrode potentials are fixed, and if the sample solution is essentially the same matrix for all measurements the junction potentials are also unchanged. Then the measured cell potential can be expressed as where R is the gas constant, T is the temperature (K), F is the Faraday constant, n represents the charge on the analyte ion and a is the activity of the analyte ion.

Chemistry 321L Manual Page 24 The ISE filling solution contains a large concentration (activity) of the analyte ion and is essentially unchanged during operation of the electrode (a ion inner is fixed). Thus, at 25EC, For fluoride ion solutions at 25EC and constant ionic strength, Thus, for an ideal fluoride ISE, the cell potential is linearly related to the logarithm of the fluoride ion concentration and should increase 59.16 mv for every 10-fold decrease in the [F - ]. When the ionic strength of all standards and samples is constant, the response of a real fluoride ISE is described by a similar relationship where β is the electromotive efficiency and typically has a value very close to unity (> 0.98). B. Direct Potentiometric Measurement To check if the electrode is working properly, you will measure the the cell potential of three fluoride standards prepared in a total ionic strength adjustment buffer (TISAB). The TISAB contains an acetic acid/acetate buffer that fixes the ph of the solution at about 5. At this ph the formation of HF is negligible and the concentration of OH -, the only other anion that the electrode responds to, is insignificant. It also contains NaCl to establish a high and constant ionic strength, and a complexing agent that removes cations that could interfere by forming complexes with fluoride. From a linear least-squares fit to a plot of E meas versus log[f - ] you can obtain the slope [S = β(0.05916)]. Typically S equals 56 ± 2 mv. C. Method of Standard Addition The method of variable volume standard addition will be used to determine the fluoride content of an unknown solution. In this approach, a solution containing fluoride will be mixed with the TISAB and the potential will be measured. Then successive amounts of a fluoride standard solution will be added and the potential will be measured after each addition. The following describes how the unknown fluoride concentration can be obtained from these measurements.

Chemistry 321L Manual Page 25 The measured potential (E) can be represented by where K is a constant; S is the slope of the calibration curve and equals β(0.05916); C is the analyte ion (F - ) concentration. This equation can be rearranged to give The analyte ion concentration after any addition of the standard is given by where C 0 is the analyte concentration before any standard is added; V 0 is the volume of the solution before any standard is added; C std is the concentration of the standard solution; V std is the volume of standard solution that is added. Substituting this expression for C in the previous equation gives This equation can be rearranged to give

Chemistry 321L Manual Page 26 A plot of 10 E/S (V 0 + V std ) versus C std V std will give a linear plot with an x-intercept (y = 0) equal to the negative of the amount (μg) of analyte in the solution before addition of the standard. The analyte concentration (μg/ml) in the original unknown solution (C unk ) can then be determined by dividing by the volume of the unknown fluoride solution (V unk ). II. Procedure A. Preparation of Fluoride Standard Solutions 1. By serial dilution of the 1000 μg/ml fluoride standard solution, prepare 50 ml each of 200, 20 and 2 μg/ml fluoride standards in 50-mL volumetric flasks. After thorough mixing, transfer each diluted standard solution to a labeled plastic reagent bottle for storage. Calculate the concentration of each diluted standard using the exact concentration of the stock solution. B. Calibration Check of Electrode 1. Carefully pipet 25.00 ml of the most dilute fluoride standard into a 50-mL volumetric flask and dilute to the mark with the TISAB. Stopper the flask and thoroughly mix the solution. 2. Transfer this solution to a 100-mL plastic beaker. Place the beaker on a stirring plate, add a magnetic stirring bar and begin stirring at a constant rate. 3. Connect the fluoride ISE to a ph meter and set the meter to the mv mode. Rinse the electrode with dionized water and blot dry. 4. Lower the electrode into the standard solution. When the reading is stable record the mv value. 5. Repeat steps II.B.1-4 for each of the remaining fluoride standards. 6. Estimate of the slope (S) from the difference in the mv readings for each factor of ten increase in the fluoride ion concentration. If your value is outside the expected range, consult with your lab instructor.

Chemistry 321L Manual Page 27 C. Analysis of Unknown 1. Prepare a 500 μg/ml fluoride standard by pipeting 5.000 ml of the 1000 μg/ml fluoride standard solution into a 10-mL volumetric flask and diluting to the mark with the TISAB. 2. Carefully pipet 50.00 ml of your fluoride unknown, which contains the TISAB at the same concentration as used for the standard calibrations, into a 100-mL plastic beaker. Place the beaker on a stirring plate, add a magnetic stirring bar and begin stirring at a constant rate. 3. Rinse the ISE with deionized water and blot dry. 4. Lower the electrode into the unknown solution. When the reading is stable record the mv value. 5. Pipet 1.000 ml of the 500 μg/ml F - standard solution into the unknown solution. When the reading is stable record the mv value. 6. Make three additional 1.000 ml additions of the standard solution and record the mv reading after each addition as before. 7. When finished, rinse the ISE with deionized water and place it in the storage container. III. Calculations A. Determination of Calibration Slope 1. Using EXCEL, plot the mv reading for the diluted fluoride standards versus the log of the actual fluoride ion concentration. 2. Fit the data points with a linear least-squares line and from the equation for the line obtain the slope (S). B. Determination of Unknown Concentration by Standard Addition IV. Results 1. Using the slope determined in III.A, plot 10 E/S (V 0 + V std ) versus C std V std. Remember to include the initial reading with no added standard. 2. Fit the data points with a linear least-squares line and obtain the equation for the line. 3. Use the equation for the line to determine the x-intercept and from this calculate the fluoride ion concentration in the unknown solution. Report the fluoride ion concentration (μg/ml) in the unknown solution.