U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Similar documents
U S E R M A N U A L. (CGA 870 Connection)

Single- and Two- Stage Regulators

INSTALLATION AND OPERATION INSTRUCTIONS NOT FOR OXYGEN OR OTHER OXIDIZING GASES

MODEL G300 BRAKE BLEEDER

NITROUS TRANSFER PUMP INSTRUCTIONS

American Fireglass Outdoor Fire pit kit information. General instructions/warnings for outdoor gas pan and burner systems

Compressed Gas Cylinder Program

INSPECTION, MAINTENANCE AND RECHARGE SERVICE MANUAL P/N 16303

TWO BURNER PROPANE CAMPING STOVE

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

San josé OWNER S MANUAL

PNEUMATIC PLANISHING HAMMER

TWO BURNER PROPANE CAMPING STOVE

Operator s Manual. Pressure Injection Cell Model PC8500. Congratulations! Contents

Hydraulic Transmission Jacks Operating Instructions & Parts Manual

PALLET JACK TON

tire inflator with pressure gauge

Your Portable Oxygen System. Use & Care Guide

STC 3S A Series Solenoid Valves

3M Water Filtration Products. CFS01 Series, CFS02 Series, CFS11 Series, and CFS12 Series Housings. Installation and Instruction Manual For:

Your safety and the safety of others are very important.

Propane Conversion Kit PROPANE CONVERSION KIT INSTALLATION MANUAL

GAS SAFETY PRECAUTIONS

PAINT SPRAY GUN WASHER

Compressed Gas Cylinder Safety

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

Fluid Regulator. Instructions Parts List L LOW PRESSURE

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen. Series: LP Gas

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

Table of Contents WARNING SYMBOLS AND DEFINITIONS

FASCINATION 700 HVLP TANNING PRO SYSTEM USER MANUAL

Cylinder Leak- Down Tester

GAS CYLINDER SAFETY. 2. GRADES OF GASES BOC have divided the gases that they supply into 3 grades.

NFC Communication Tray Model: HHX-IT3-Z

Instructions For SF 6 Refill Kit

VW TDI (Golf, Jetta, New Beetle) Upgrade Turbocharger

SKILL ATTACK PITCHING MACHINE PATENTS APPLIED FOR INSTRUCTION MANUAL. Includes: OPERATION SETUP USE & CARE SERVICE

Tech. Services: (800) Fax: (800) Order Entry: (800) Fax: (800) Internet Address:

Air vacuum pump with r134a & r12/22

Customer Support

4 in 1 Air Compressor

VANGUARD A OPERATOR S MANUAL PORTER. FM-952 Rev. E 7/07

Oxy-Fuel Gas Welding. Given a functional oxy-fuel gas unit, instruction and demonstration of use, each student will be able to:

ATS Overhead Table Shelf System INSTRUCTION MANUAL

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

el-720 Carbon Dioxide Gas Leak Detector User s Guide

Bottom Loading Water Dispenser

COMPRESSED GASES. 1.2 The contents of each cylinder and container must be clearly identified (by tag or stamp) on the cylinder.

orlando OWNER S MANUAL

!"" #$% "!&' ( ( ) *

Service Manual. Model PM4150 Series. Precision Medical, Inc. 300 Held Drive Northampton, PA USA

Threaded Inline Air Heaters

OXYGEN CONCENTRATOR, OXYGEN CYLINDER

STORED PRESSURE STAINLESS STEEL

Instructions and precautions. Fork Height. Visit our website at:

1 Ton Telescoping Gantry Crane

QuickAir - 5, - 10 Minute Emergency Escape Breathing Apparatus. Operating Manual

Recharge Procedure for Telemark TVP-2000 and TVP-3500

Operating Instructions & Maintenance Guidelines for Precision Vaporizers

Ozone Generator Manual

Cleaning Instructions, Pilot Replacement and Valve Change. Model No.: Natural Gas

ROUTER TABLE INSERT ASSEMBLY AND OPERATING INSTRUCTIONS

OWNERS MANUAL FWT SERIES TRANSOM MOTOR

USE &CARE GUIDE. Remote Faucet Pump System. See Important Safeguards on page 2

SE-100-1, SE-200-1, SE-500-1, and SE AIR CHAMP PRODUCTS. User Manual SE BRAKE MODELS: (i) MTY (81)

OAK WORKBENCH WITH 2 DRAWERS

EzFlex Combustible Gas Leak Detector Model EZ40

Mini-led spotlight with magnetic base

FJ2. 2 Ton Trolley Floor Jack Assembly & Operating Instructions

Extrusion Flo-Valve INSTRUCTIONS-PARTS LIST psi (210 bar) Maximum Working Pressure. Model , Series K. Rev.

VW TDI (Golf, Jetta, New Beetle) Upgrade Turbocharger

Hand Held Metal Detector

In-Line Filters and Coalescers with Automatic Drains

MODEL J SERIES CONVERTER

System Saver 318 Air Compressor for Mack E-Tech and ASET Engines

SBC90. Abrasive Blast Cabinet Assembly & Operating Instructions

DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual

Advantium 2 Plus Alarm

SERIES VOLT ELECTRIC INSTANTANEOUS WATER HEATER FOR EMERGENCY SAFTEY EQUIPMENT INSTALLATION AND OWNERS MANUAL REQUIRED MANTENANCE

STC 2P050 Series Solenoid Valves NSF Compliant Materials, 2 Way, Direct Acting

6 & 12 Volt Battery and Systems Tester with 100 Amp Load

CUNO TM Series NB Bags for Liquid Filtration

Owner s Manual & Safety Instructions

1 / 4 IN. FFL x 4 FT. 9 IN. 1 / 2 IN. ACME x 4 FT. 9 IN. Gauge Accuracy +/-2% Set includes Quick Disconnect Valves

Whole House Water Treatment System. Installation, Operation and Maintenance Manual. Multi-Pure Model TO15000

Owner s Manual & Safety Instructions

PAPER SHREDDER PRS-F-01 USER MANUAL

Electric Water Heater

RGC-IR Remote Gas Calibrator for IR400

National- Spencer Inc.

Operators Manual SOIL COMPACTION TESTER

DiscovR R134a Refrigerant Identifier

3 WATT LED SPOTLIGHT Model No. SLM

Hydraulic Punch Driver Kit

WARRANTY VINYL SIDING CERTAINTEED CORPORATION LIFETIME LIMITED

Owner s Manual & Safety Instructions

Oxygen. Do not use flammable products, such as aerosol sprays, rubbing alcohol, paint thinners, and other oil-based lubricants near oxygen.

THC 85 INDUSTRIAL / COMMERCIAL SPACE HEATER

55-Gallon Dispenser Package

CONGRATULATIONS! TABLE OF CONTENTS. 1 Safety Warnings 2 About Your Wheels 3 Setting Up Your Wheels 7 Care and Cleaning 8 Warranty

Transcription:

U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 ISO 13485 Certified www.precisionmedical.com

RECEIVING / INSPECTION Remove the Precision Medical Gauge Regulator from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. INTENDED USE To regulate flow of oxygen from a high pressure oxygen cylinder for medical use only. To be used by or after instruction from a trained, healthcare professional. READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the Gauge Regulator. This is provided for your safety and to prevent damage to the Gauge Regulator. If you do not understand this manual, DO NOT USE the Gauge Regulator and contact your Provider. EXPLANATION OF ABBREVIATIONS psi kpa lpm Pounds Per Square Inch Kilopascal Liters Per Minute WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury. Used without the safety alert symbol, indicates a potentially hazardous situation which, if not avoided, could result in property damage. CONSULT ACCOMPANYING DOCUMENTS Symbol for USE NO OIL Symbol for NO SMOKING

WARNING To prevent risk of burns, fire or injury to person(s): Handle the Gauge Regulator carefully at all times and use the Regulator according to the instructions in this manual. Oxygen cylinder should be in an upright position when in use. ALWAYS confirm prescribed flow before administering to patient and monitor flow on a frequent basis. Always follow ANSI and CGA standards for Medical Gas Products and Flowmeters and Oxygen Handling. DO NOT use tape, oils, greases, petroleum jellies, organic lubricants, paint or any other hydrocarbon based or combustible materials on or near this Gauge Regulator. DO NOT use if oil, grease or contaminants are present on or around Regulator, cylinder valve or connecting devices. DO NOT use near any type of flame or flammable/explosive substances, vapors or atmosphere. DO NOT smoke in an area where oxygen is being administered. Only personnel instructed and trained in its use should operate this Gauge Regulator. Regulator contains magnetic, ferrous material that may affect the results of an MRI. Be sure all connections are tight and leak free. Only use oxygen safe leak detector to test for leaks. Store Regulator in a clean area when not in use. DO NOT direct flow of oxygen at any person, or flammable material when cracking cylinder. DO NOT autoclave. DO NOT gas sterilize with EtO (Ethylene Oxide). DO NOT clean with aromatic hydrocarbons. DO NOT attempt to repair Regulator. DO NOT immerse the Regulator in any kind of liquid. This will void the warranty.

SPECIFICATIONS Accuracy: ±10% of indicated setting @ 70 F (21 C) Transport / Storage Temperature Range: -4 F (-20 C) to 140 F (60 C) Inlet Pressure Range: 300-3000 psi (2068-20684 kpa) Ignition and Fault Tolerance: meets ASTM G175-03 Specifications are subject to change without prior notice. PRODUCT DIAGRAM Missing or illegible labels must be replaced, contact Precision Medical, Inc. Flow Gauge Gauge Guard T-Handle Outlet Connection Regulator Knob Contents Gauge

OPERATING INSTRUCTIONS WARNING Read this User Manual before installing or operating the Gauge Regulator. 1. Check cylinder valve to be sure it is not bent or damaged. A damaged valve could leak or fail, or it may not make a tight connection when cylinder is placed into use. 2. Make sure the Regulator and cylinder valve are free of oil, grease or other contaminants. 3. Before installing Regulator, crack cylinder valve by opening the valve slightly and closing to remove any foreign particles. 4. Turn Regulator Knob clockwise to the OFF position. 5. a) For CGA 870 connection (Yoke): Be sure the high pressure inlet port has a Yoke Seal. Attach Regulator to the oxygen cylinder post valve and align index pins with mating holes. Hand tighten T-handle. b) For CGA 540 connection (Nut & Nipple): Attach the Regulator to the oxygen cylinder valve and tighten with wrench. 6. Position the Regulator so the cylinder valve opening is pointing away from the user and any other person(s). 7. S-L-O-W-L-Y open cylinder valve counterclockwise, until completely open. 8. Adjust Flow: To INCREASE flow - Turn Knob counterclockwise To DECREASE flow - Turn Knob clockwise 9. Set flow by aligning indicator needle with marks on flow gauge face. 10. Adjust flow beyond the last calibrated indicator line will result in an undetermined oxygen flow. 11. Continue procedure as prescribed by your healthcare professional. NOTE: Precision Medical, Inc. strongly recommends the use of kink proof patient tubing.

12. To remove Regulator from cylinder: Completely close oxygen cylinder valve. Turn the Regulator to the OFF position. S-L-O-W-L-Y loosen T-handle or Nut and remove Regulator. WARNING To prevent risk of burns, fire or injury to person(s): Prior to handling regulator and cylinder valve, be sure hands are free of oil, grease and other contaminates. DO NOT use Regulator if oil, grease or dirt is found. Particles propelled in a high-velocity oxygen stream could create a spark, causing a fire. Oil and grease can react with oxygen and other oxidizers, causing an explosion. DO NOT use tools to tighten T-handle. This can lead to over tightening and will cause damage to the Regulator. Inspect the Regulator for visual damage before use,do NOT USE if damaged. For optimum safety, use only a Precision Medical, Inc. Yoke Seal part number 1221. The Yoke Seal supplied with the Regulator is reusable. DO NOT use plastic yoke seals.

MAINTENANCE / CLEANING 1. Disconnect all connections before cleaning. 2. Clean exterior surfaces of the Gauge Regulator with a cloth dampened with disinfectant SporicidinR or equivalent. 3. Wipe dry with a clean cloth. 4. Store the Regulator in a clean area free from grease oil and other sources of contamination. DO NOT use cleaning solutions. DO NOT attempt to repair the Regulator. All repairs must be performed by Precision Medical, Inc. RETURNS Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com. TROUBLESHOOTING If the Gauge Regulator fails to function, consult the Troubleshooting Guide. If problem cannot be solved, consult your Provider. Problem Probable Cause Remedy No flow Cylinder valve closed Regulator in OFF position Cylinder empty Turn ON cylinder Set to prescribed flow Replace cylinder

LIMITED WARRANTY AND LIMITATION OF LIABILITY Precision Medical, Inc. warrants that the Medical Gas Gauge Regulator (the Product) will be free of defects in workmanship and/or material for the following period: Five (5) years from date of shipment Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc. s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense. ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES. The representative of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract. THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED. Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice. Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty. 503369 Rev5 (E) 03/30/10 (?M) Printed in USA