STAGO PT / aptt / Fibrinogen General Procedure

Similar documents
Collection Guidelines for Routine & Special Coagulation Testing

Abnormal Basic Coagulation Testing Laboratory Testing Algorithms

Bleeding disorders or haemorrhagic diatheses are a group of disorders characterised by defective haemostasis with abnormal bleeding.

Contents. Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope Introduction Standard Precautions...

INR = (patient PT/mean normal PT) ISI.

Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA

Lupus anticoagulant aptt & drvvt screening panel W Reflex

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

Lupus anticoagulant Pocket card

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Laboratory Detection of Newer Anticoagulant Drugs

Thrombin Generation Assay

Vitamin B12. ADVIA Centaur System

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

MCHENRY WESTERN LAKE COUNTY EMS SYSTEM OPTIONAL CE ADVANCED LEVEL (EMTP, PHRN, ECRN) August Anticoagulants

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

The Golden Rule of Specimen Collection: The Patient Test Result is Only as Good as the Sample We Get

APPENDIX N BULK MILK TANKER SCREENING TEST FORM

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

Carcinoembryonic Antigen (CEA)

New Oral Anticoagulants

Cold Agglutination Titer detecting Cold Reacting Antibodies

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

Cancer Antigen CA 125

Anticoagulation Essentials! Parenteral and Oral!

RealLine HCV PCR Qualitative - Uni-Format

Standard Operating Procedure (SOP) Work Package 8. Sample Collection and Storage

Consultative Coagulation How to Effectively Answer Common Questions About Hemostasis Testing Session #5020

4/24/2015. Urgent, STAT, Super STAT, ASAP! Achieving timely lab testing for the Emergency Department. Learning Objectives.

Biochemistry Validation Form

THE PREPARATION OF SINGLE DONOR CRYOPRECIPITATE

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

USE AND INTERPRETATION OF LABORATORY COAGULATION TESTS IN PATIENTS WHO ARE RECEIVING A NEW ORAL ANTICOAGULANT (DABIGATRAN, RIVAROXABAN, APIXABAN)

Malondialdehyde (MDA) ELISA

Collect and label sample according to standard protocols. Gently invert tube 8-10 times immediately after draw. DO NOT SHAKE. Do not centrifuge.

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #:

HBV Quantitative Real Time PCR Kit

TOTAL PROTEIN FIBRINOGEN

Disclosure. New Agents for Treatment of DVT. Prevalence of DVT VTE. Normal Hemostasis 7/17/2015. Mark Oliver, MD, RVT, RPVI,FSVU

DNA Isolation Kit for Cells and Tissues

NnEeWw DdEeVvEeLlOoPpMmEeNnTtSs IiıNn OoRrAaLl AaNnTtIiıCcOoAaGgUuLlAaTtIiıOoNn AaNnDd RrEeVvEeRrSsAaLl

Lab 2. Spectrophotometric Measurement of Glucose

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Coatron A4. Fully automated Hemostasis Analyzer. Made in Germany REF: Innovation in Coagulation

Note: Page numbers in italics indicate figures. Page numbers followed by a t indicate tables.

Critical Bleeding Reversal Protocol

STEP-BY-STEP INSTRUCTIONS FOR INVESTIGATIONAL USE. Rapid HCV Antibody Test FOR ORAQUICK RAPID HCV ANTIBODY TEST

QuantiFERON TB Gold In Tube Method (QGIT Test) Manufacturer: Cellestis Inc. (USA) a QIAGEN Company

ab83369 Alkaline Phosphatase Assay kit (Colorimetric)

ab Protein Sumoylation Assay Ultra Kit

Hemostasis analyzer system

Department of Pathology and Laboratory Medicine Capital District Health Authority Nova Scotia

Mouse glycated hemoglobin A1c (GHbA1c) ELISA Kit

Session 3 Topics. Argatroban. Argatroban. Drug Use and Adverse Effects. Laboratory Monitoring of Anticoagulant Therapy

MANUAL PLATELET COUNT

Xabans Good for What Ails Ya? Brian Tiffany, MD, PhD, FACEP Dept of Emergency Medicine Chandler Regional Medical Center Mercy Gilbert Medical Center

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

SARASOTA MEMORIAL HOSPITAL BLOOD COMPONENT CRITERIA AND INDICATIONS SCREENING GUIDELINES

Ward 29 guide to the safe preparation and administration of intravenous (IV) antibiotics at home

Immunoglobulin E (IgE) concentrations in Human. Immunoglobulin E (IgE) Human ELISA Kit

LABORATORY DIAGNOSIS OF BLEEDING DISORDERS

New anticoagulants: Monitoring or not Monitoring? Not Monitoring

Automation is Here. Now How Do You Handle All the Data? Swedish Covenant Hospital Lab. Life IS Good

DVT/PE Management with Rivaroxaban (Xarelto)

MANUAL WBC LAB OBJECTIVE

Intact PTH. system. Intact PTH /R5. Key to symbols used 8K25

Reversal of Antiplatelet and Anticoagulant Therapy: What You Need To Know. Ronald Walsh, MD Chief Medical Officer Community Blood Services

Medical Necessity and Advanced Beneficiary Notice (ABN) Policy and Form

Simoa HD-1 Analyzer Troubleshooting Guide. USER Sep 2014 Software Version 1.3 Beta

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

Agencourt RNAdvance Blood Kit for Free Circulating DNA and mirna/rna Isolation from μL of Plasma and Serum

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE

ph units constitute a scale which allows scientists to determine the acid or base content of a substance or solution. The ph 0

INTRODUCTION. Viral shedding is crucial for Gene Therapy Products Safety linked with shedding data

LAMC Reversal Agent Guideline for Anticoagulants Time to resolution of hemostasis (hrs) Therapeutic Options

Mouse krebs von den lungen 6 (KL-6) ELISA

Anti-Zona Pellucida Antibody Latex Agglutination Test

Human Parathyroid Hormone (PTH) ELISA Kit

Dr Gordon Royle Haematologist, Middlemore Hospital

Rat Fibronectin ELISA Kit

Human Free Testosterone(F-TESTO) ELISA Kit

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit

UltraClean Soil DNA Isolation Kit

Canine Creatine Kinase MM isoenzyme(ck-mm) ELISA. kit

Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit

Data Sheet. PD-L1:B7-1[Biotinylated] Inhibitor Screening Assay Kit Catalog # Size: 96 reactions

Hemoglobin Testing using the Data Management HemoCue Analyzer

VACUETTE RNAgard Blood Tubes BioMaxi Blood RNA Purification Kit. For the collection, preservation and purification of RNA from whole blood

Bovine Vitamin B12 (VB12) ELISA Kit

Quantifying Bacterial Concentration using a Calibrated Growth Curve

Blood Collection and Processing SOP

Division of Laboratory Medicine Department of Pathology and Laboratory Medicine Hospital of the University of Pennsylvania

PACKAGE LEAFLET: INFORMATION FOR THE USER. PARACETAMOL MACOPHARMA 10 mg/ml, solution for infusion. Paracetamol

UltraClean Forensic DNA Isolation Kit (Single Prep Format)

EXERCISE 5: ERYTHROCYTES SEDIMENTATION RATE - ESR, SED RATE

18 Conductometric Titration

Transcription:

STAGO PT / aptt / Fibrinogen General Procedure Principle: The Stago Coagulation analyzer is an automated system used for in vitro testing of the coagulation system. For PT/aPTT/Fibrinogen, the Stago analyzer measures the time of clot formation, using an electro-magnetic mechanical clot detection system. The Prothrombin time (PT) is a basic coagulation screening test for the assessment of congenital and acquired deficiencies of the extrinsic pathway (factors II, V, VII, X). It is also used in monitoring warfarin therapy because of its sensitivity to variations in the concentration of vitamin-k dependent factors. The activated partial thromboplastin time (APTT) is a basic screening test for the intrinsic coagulation pathway (factors XII, XI, IX, VIII, X, V, II, and I). It is used to detect congenital and acquired deficiencies of these factors and to monitor heparin therapy. Low functional fibrinogen levels may be seen in DIC, during massive bleeding, liver disease, in the rare congenitally deficient patient (afibrinogenemia), or patients with nonfunctional fibrinogen protein. Elevated fibrinogen levels can be observed in diabetes or inflammatory syndromes, and may be associated with an increased risk of cardiovascular disease and pre-thrombolytic states. The concentration of fibrinogen is inversely proportional to the clotting time when a known amount of thrombin (in excess) is added to dilute patient plasma. The measured time is read from a stored curve on the analyzer, and is the plasma fibrinogen level. Safety: The required personal protective equipment for this procedure - Gloves - Impermeable lab coats, worn closed - Shield - Approved Protective eyewear Gloves and lab coats should be worn at all times during analysis of the samples. Samples that must be run in the secondary mode must be opened behind a safety shield. Gloves, lab coat, and a face shield or goggles/safety glasses must be worn when performing maintenance where there is a chance of splashing. Page 1 of 10

Specimen: 1. Specimen Collection: Specimens are collected in 3.2% sodium citrate tubes with a collection ratio of 9:1 blood to anti-coagulant. No other anticoagulant is acceptable. The sodium citrate tube should fill completely. The specimen should be gently mixed after collection. 2. Centrifugation: Specimens are centrifuged capped, in non-refrigerated centrifuges to achieve plateletpoor plasma. Refer to chart for centrifuge model, time and speed used at each site. 3. Unacceptable Specimens: Specimens processed in a centrifuge that can not achieve platelet-poor plasma (<10.000 k/ul) are unacceptable. Samples that are short draws, clotted, over-filled or hemolyzed may yield incorrect results and should be recollected. Samples with a HCT >55% should be collected in tubes that have an adjusted amount of anticoagulant to insure accurate results. Use the Criteria Nomogram for Acceptablity of Citrated Plasma (0544F) to visually detect a HCT >55%. Once it has been determined that citrate adjustment should occur, use the Citrate Adjustment Chart under Special Preparation for anticoagulant adjustments. 4. Special Preparation: Citrate Adjustment Chart for HCT>55%: Small Plastic Citrate Tubes, 3 ml (Normal draw 2.7 ml blood/ 0.3 ml citrate) Hematocrit Volume citrate to REMOVE from tube 55.0 57.0 Remove 0.08 ml 57.1 59.0 Remove 0.09 ml 59.1 61.0 Remove 0.10 ml 61.1 63.0 Remove 0.11 ml 63.1 65.0 Remove 0.12 ml 65.1 67.0 Remove 0.13 ml 67.1 69.0 Remove 0.14 ml 69.1 71.0 Remove 0.15 ml 71.1 73.0 Remove 0.16 ml 73.1 75.0 Remove 0.17 ml 75.1 77.0 Remove 0.18 ml 77.1 79.0 Remove 0.19 ml 79.1 80.0 Remove 0.20 ml Adjust the amount of anticoagulant in a sodium citrate tube according to the chart. Since the tube vacuum has been lost after removing the cap to do the adjustment, Page 2 of 10

the patient s blood must be drawn using a syringe (draw 2.7 ml of blood), and then carefully delivered to the adjusted tube, capped, and mix thoroughly. Report results with the comment specimen collected in anticoagulant adjusted tube. 5. Stability: Centrifugation and testing should be performed as soon as possible, to maintain the integrity of the specimen. While performing these tests outside the limits of their stability is unacceptable, specimens are capped and retained for at least 24 hours for purposes of specimen identity/integrity investigation. If testing is not performed within 24 hours for PT and four hours for aptt, plasma should double-spun and platelet-poor plasma should be frozen (-20 C up to two weeks, or -70 C for long-term storage). Frozen plasma should be rapidly thawed at 37 C, then gently mixed and tested immediately. Mixing is critical before testing, in order to resuspend proteins that may have been precipitated by freezing. Analyte PT aptt Fibrinogen Stability 8 hrs @ 15-25 C 4 hrs @ 15-25 C 8 hrs @ 15-25 C 6. Specimen Storage While performing these tests outside the limits of their stability is unacceptable, specimens are capped and retained for at least 24 hours for purposes of specimen identity/integrity investigation. Equipment and Materials: Centrifuge Pipettes Pipette tips ABC Stago analyzer Cuvette roll (contains 1000 cuvettes) Magnetic stir bars Reducer Wash Solution Clinical Laboratory Reagent Water or Commercially Bottled Purified Water Page 3 of 10

Reagents and QC Material: Product Reconstitution Stabilization time before use Neoplastine Cl Plus* PTT-Automate Fibrinogen Pour contents of Rgt. 2 vial into Rgt. 1 vial 5 ml Rgt. Grade Water 5 ml Rgt. Grade Water Stand at RT 30 min. Swirl gently. Stand at RT 30 min. Vortex on low for 5 sec before loading. Stand at RT 30 min. Swirl gently. Stability on STAGO (15-19 C) Reconstituted reagent stability in refrigerator (2-8 C) 48 hours 8 days in original capped vial (stand at RT 30 min before use) 24 hours 7 days in original capped vial (stand at RT 30 min before use) 120 hours (5 days) 14 days in original capped vial (RT 30 min before use) Special Notes Use with stir bar & reducer Calcium Chloride Coag N & ABN Owren-Koller Load without white cap Reconstitute each vial with 1 ml Rgt. Grade Water Load without white cap Stand at RT 30 min. Swirl gently. 72 hours N/A (3 days) 8 hours 8 hours 72 hours (3 days) N/A *Caution: Neoplastine Reagent 2 contains sodium azide as a preservative and should be discarded with care. When dumping waste in the sink, follow with copious amount water to clear the plumbing. Before loading any product, check surface for foam/bubbles and if present remove. RT = room temperature Quality Control: Coag Control N and Abn: After the reconstitution period: 1. Load QC material onto analyzer: Request product drawer to open through Main Menu under Loading Reagents. Scan bar-coded controls, confirm volume, press <enter> on keyboard, and place in designated position on the analyzer. 2. To test QC: Select Main Menu, Calib/Control, Quality Control, <enter>. Place the cursor on the tests, and select each test by pressing F1. When all the tests have been selected, press F10 to run the QC. Page 4 of 10

When prompted, enter QC or CQ (depending on site) as the access code and press <enter>. Escape from QC screen to begin testing of controls. Both levels of QC are tested every 8 hours of patient testing. QC also is tested if reagents are added, or if maintenance is performed during a shift. Control ranges are monitored automatically by the analyzer. If any controls are outside the range stated on the manufacturer package insert, the analyzer will audibly and/or visually alarm the operator. All results for a 24-hour period will be held in the Stago QC file, and reduced to a mean value at midnight. This mean is plotted on the Levy- Jennings chart as a daily mean. New lots of QC are sequestered annually for SLN and ranges are established during annual reagent lot switchover. System established ranges are then set up in the LIS and cannot exceed the ranges on the STAGO package insert. QC results entered into the LIS must meet the Westgard rules in the SLN QC policy. To investigate an unacceptable control value: First check control for time of reconstitution. Repeat the QC run. If results are still out of range, reconstitute a fresh control sample and re-test. If results are still out of range, prepare new reagents and run QC. If results are still out of range, examine the analyzer set-up for any visible problems (i.e. bent needles, leaks). If no resolution, call technical service. Patient results cannot be reported until control value(s) are within acceptable limits. Blind Duplicates: A previously tested patient sample, with results in the normal range should be tested between analyzers at sites who alternate the use of two analyzers. This sample, along with QC samples, should be run on the newly set-up analyzer before running patient samples. The sample should be within the specimen age requirements for patients. If a result is out of range, troubleshoot and repeat the blind duplicate. If commercial QC is acceptable, choose another sample to use for a blind duplicate. Calibration: No calibration is required for PT or aptt. The pre-calibrated fibrinogen values are identical for all the vials of each lot. Entering the data for the Fibrinogen calibration curve: When the operator scans a new lot of fibrinogen reagent, the STA / STA Compact will request the operator to scan the bar code printed on the bar code insert across the analyzer bar code reader. The curve is automatically read from this bar code label. Page 5 of 10

The calibration curve will be validated for the lot in use once the two fibrinogen control levels have been tested. If the controls are out of range, the analyzer will not test patient samples or validate the curve. To examine the curve on the analyzer screen: Through Main Menu, under Calib/Control, select Calibration. Move the cursor to FIB and <enter>. Curve will display on the screen. To print the calibration curve: With the curve displayed on the screen, press ESC for options; select PRINT. A curve can not be printed while the analyzer is running. Refer to STAGO Maintenance Procedure for instructions on Fibrinogen linearity verification. Procedure: Changing Sampling Modes 1. The Compact Stago can operate in two different patient loading modes, auto mode or manual mode. To change loading modes, select Loading from the main menu. 2. Select Samples. The mode will be displayed near the top of the screen. If you wish to change modes press ESC for options. Choose the desired mode. Loading Patient Samples in Manual Mode 1. Choose Loading from the main menu. 2. Select Samples (the sample drawer will open) 3. Ensure that the sample mode is set to manual 4. Scan or type the accession and press enter on the keyboard 5. Load the tube into the sample drawer. 6. Select tests from the test menu. 7. Press F10 to save. 8. Continue loading samples until all samples have been loaded, pressing F10 after each sample. 9. After loading the last sample, press ESC. The drawer will close and processing will begin. Loading Patient Samples in Auto Mode. 1. Choose Loading from the main menu. 2. Select Samples (the sample drawer will open) 3. Ensure that the sample mode is set to auto. 4. Scan or type the accession and press enter on the keyboard 5. Load the tube into the sample drawer. 6. After loading the last sample, press ESC. The drawer will close and processing will begin. Page 6 of 10

Microsample pour-offs Select F8 to indicate volume once drawer has opened. Be sure to turn off micro volume before running normal volume samples again. Instrument Processing As soon as the sample drawer closes, the Test Status screen will appear. If there are insufficient reagents to run the test, the Blocked Sample Pipetting will appear in red with the amount of deficiency. This deficiency will BLOCK the sample pipettor. When this occurs, add the necessary reagents and run QC. Dilutions are automatically prepared by the analyzer, according to the established parameters that are entered in each TEST SETUP at time of installation. If the patients results fall outside the assay range, the analyzer will automatically retest that sample at an appropriate dilution, if the test has been previously defined for this response. Results: All patient results display on the TEST PANEL screen of analyzer s monitor. Results will automatically transmit to the LIS. Results may automatically print or be batch printed at the end of each day. For results in question that need operator intervention, cursor to the identification number in the Test Panel screen; <enter>. This will display the File Processing screen. Follow the options in the left hand corner of the screen, (i.e. re-run test). Reporting Results: See Stago Information Sheet, #0716C and Stago Compact Error Codes, #0544C, for result reporting and error code information. Reference Range: See Stago Information Sheet, #0716C for current reference ranges. Therapeutic Ranges INR Comments <1.5 1.5-2.0 Low-intensity Therapeutic Range 2.0-3.0 Mod-intensity Therapeutic Range 2.5-3.5 High-intensity Therapeutic Range 3.0-4.0 High-intensity Therapeutic Range 5 Common Critical/Alarm Value 10 Common Upper Limit reported Page 7 of 10

Critical Values: Critical values must be verified, called immediately, and documented in the LIS. See Stago Information Sheet #0716C for current critical values. Procedural Notes: 1. New lot numbers of Neoplastine, PTT-A, and Fibrinogen reagent are changed annually and are done as a System. No changes are made by individual sites. Refer to Annual Lot Conversion procedure for instructions. 2. Patients receiving thrombolytic therapy will have a rapid drop in the plasma Fibrinogen level. These samples must be collected with an anticoagulant containing a plasmin inhibitor to determine accurate Fibrinogen result. DIC Panel Reporting: The panel for testing suspected cases of Disseminated Intravascular Coagulation (DIC) includes the following tests: PT, APTT, Fibrinogen, D-dimer, Platelet count, and smear review (Tech Review) for the presence of shistocytes. The panel tests appear collectively on the patient report when ordered as a panel. This panel is available for testing only in the hospital setting. NOTE: when reporting the Smear Review portion of the panel, report either schistocytes noted on the smear or no schistocytes on the smear using site specific entry codes. Limitations of Procedure: 1. STA Neoplastine CI and CI+, both contain a specific inhibitor of heparin. Therefore, only levels of heparin outside of the therapeutic range will affect the PT results 2. Many commonly administered drugs may affect the results obtained in APTT testing. (Example: heparin and coumadin). 3. High levels of paraproteins may interfere with the fibrinogen assay. 4. The clinical use of topical bovine thrombin can cause the generation of antibodies which can lead to artificial prolongation of fibrinogen results. 5. The STA-Fibrinogen reagent contains a specific inhibitor of heparin. However, heparin levels above 2 IU/mL may interfere with this assay. 8. The STA Fibrinogen procedure is insensitive to hirudin levels up to 3 ug/ml. 9. Fibrinogen levels will be falsely low in the presence of thrombolytic agents (tpa, streptokinase). Plasma from patients receiving these agents must be treated with an inhibitor (ie, aprotonin) before testing for fibrinogen in order to obtain accurate values. Page 8 of 10

References: 1. STA Compact Operator s Manual, Diagnostica Stago, v 1.0, January 1996. 2. CLSI, Collection, Transport and Processing of Blood Specimens for Coagulation Testing, approved guidelines, H21-A5, 2008. 3. Coagulation: The Good, the Bad and the Unacceptable, Advance, May 31, 1999. 4. Criteria Nomogram for Acceptability of Citrated Plasma, UNC Hospital 2004. 5. STA- Fibrinogen package insert, revised March 2007. 6. STA - Neoplastine CI, Determination of Prothrombin Time (PT). Package insert for use in PT determinations. Revised February 2007. 7. STA - Coag Control N + Abn, Control Plasmas for Coagulation Tests on STA, Package insert (Cat. No. 00676), revised July 2002. 8. Hirsh, J. Ed. DL, Oral Anti-Coagulants; Mechanism of Action; Clinical Effectiveness and Optimal Therapeutic Range, Chest. 119 (Suppl): 85-215, 2001. 9. Williams, Hematology, Chapter 21, page 285, McGraw Hill Med., 7 th Edition, 2006. 10. Ansell, J. et al. Pharmacology and Management of Vitamin K Antagonists: Conference on Antiarrythmic and Thrombolytic Therapy, Chest: 126: 2045, 2004. 11. STA PTT-A. Determination of Activated Partial Thromboplastin Time, Package insert for use in APTT determinations, revised February 2007. 12. Triplett, Douglas A., Laboratory Evaluation of Coagulation, ASCP Press, Chicago, 1982, pps 29, 30, 40. Page 9 of 10