DRG Giardia Ag (stool) (RAP-5407: 20 cassettes) / (RAP-5408 : 20 strips) As of 22 Dec rm (Vers. 1.1)

Similar documents
OneStep Fecal Occult Blood RapiDip InstaTest. Cat #

Anti-Zona Pellucida Antibody Latex Agglutination Test

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA

Classic Immunoprecipitation

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

HBV Quantitative Real Time PCR Kit

Epinephrine/Norepinephrine ELISA Kit

Revised 3 Oct rm (Vers. 16.1)

Dengue IgM ELISA. For the quantitative determination of IgM-class antibodies to Dengue Virus in serum.

STEP-BY-STEP INSTRUCTIONS FOR INVESTIGATIONAL USE. Rapid HCV Antibody Test FOR ORAQUICK RAPID HCV ANTIBODY TEST

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

LAB 14 ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA)

Toxoplasma gondii IgM ELISA Kit Protocol

Immunoglobulin E (IgE) concentrations in Human. Immunoglobulin E (IgE) Human ELISA Kit

Dot Blot Analysis. Teacher s Guidebook. (Cat. # BE 502) think proteins! think G-Biosciences

Cold Agglutination Titer detecting Cold Reacting Antibodies

IgE (Human) ELISA Kit

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

UltraClean Soil DNA Isolation Kit

Malondialdehyde (MDA) ELISA

ELISA-VIDITEST-MICROCYSTIN LR Lot: XX

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

Mouse krebs von den lungen 6 (KL-6) ELISA

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

How Does a Doctor Test for AIDS?

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit

RealLine HCV PCR Qualitative - Uni-Format

Aptima Specimen Transfer Kit

Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit

ab Protein Sumoylation Assay Ultra Kit

Albumin ELISA Kit. Albumin ELISA Kit. Zur in vitro Bestimmung des Albumin in Urin und Stuhl

ABORhCard. ABORhCard Package Insert ABO and Rh Blood Grouping Device

Bovine Vitamin B12 (VB12) ELISA Kit

The kit is a sandwich enzyme immunoassay for in vitro quantitative measurement of PST in mouse tissue homogenates and other biological fluids.

Human Free Testosterone(F-TESTO) ELISA Kit

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit

Aseptic Technique. A GMP/GTP Training Module

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit

POC.PC.17 Point of Care Testing AMNISURE PROCEDURE FOR RUPTURED FETAL MEMBRANES TABLE OF CONTENTS. 1.0 Principle of Procedure.. 2

FBS06 P30 Antigen Test for the Presence of Seminal Fluid

Canine Creatine Kinase MM isoenzyme(ck-mm) ELISA. kit

Human IP-10 ELISA Kit, pink-one

empcr Amplification Method Manual - Lib-A

Rat Creatine Kinase MB Isoenzyme (CKMB) ELISA

Human TNF- Product #

PCR and Sequencing Reaction Clean-Up Kit (Magnetic Bead System) 50 preps Product #60200

NimbleGen DNA Methylation Microarrays and Services

Cancer Antigen CA125 Human ELISA Kit

Recommended Procedures for the Extraction of RNA. Jan Pedersen USDA, APHIS, VS, National Veterinary Services Laboratories, Ames, IA 50010

UltraClean PCR Clean-Up Kit

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Rat Fibronectin ELISA Kit

Agencourt RNAdvance Blood Kit for Free Circulating DNA and mirna/rna Isolation from μL of Plasma and Serum

ab Albumin Human ELISA Kit

Blood Collection and Processing SOP

PowerFecal DNA Isolation Kit

QuickTiter Hepatitis B Surface Antigen (HBsAg) ELISA Kit

AxyPrep TM Mag PCR Clean-up Protocol

Uscn Life Science Inc. Wuhan

50 g 650 L. *Average yields will vary depending upon a number of factors including type of phage, growth conditions used and developmental stage.

UltraClean Forensic DNA Isolation Kit (Single Prep Format)

HiPer Immunoelectrophoresis Teaching Kit

QuickTiter FeLV Core Antigen ELISA Kit (FeLV p27)

Measles (Rubeola) IgM ELISA Catalog No. CB (96 Tests)

Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit

SSI STREPTOCOCCUS LATEX GROUP KIT

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

Genomic DNA Extraction Kit INSTRUCTION MANUAL

Human Peripheral Blood Mononuclear Cell (PBMC) Manual

ProteoMiner Protein Enrichment Kits

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: Genscreen ULTRA HIV Ag-Ab Number: PQDx Abstract

CatCombi Research ELISA Kit

MYCOLOGY PARASITOLOGY RHEUMATOID FACTORS VIROLOGY SAMPLE COLLECTION

Frozen-EZ Yeast Transformation II Catalog No. T2001

Arcturus PicoPure RNA Isolation Kit. User Guide

ab Complement C1q Human ELISA Kit

Agilent High Sensitivity DNA Kit Guide

HBV PCR detection Kit USER MANUAL

Thrombin Generation Assay

Kevin Bogart and Justen Andrews. Extraction of Total RNA from Drosophila. CGB Technical Report doi: /cgbtr

IP10 Human ELISA Kit

ReadyPrep Protein Extraction Kit (Soluble/Insoluble) Instruction Manual. Catalog #

LAB 11 PLASMID DNA MINIPREP

QuickZyme Soluble Collagen Assay

竞 争 性 分 析 Epitope Mapping 实 验 方 法

HUMULIN 70/30 KwikPen

Register your instrument! MagSep Viral DNA/RNA Kit. Instructions for use

TECHNICAL BULLETIN. Angiotensin II EIA Kit for serum, plasma, culture supernatant, and cell lysates. Catalog Number RAB0010 Storage Temperature 20 C

HiPer Ion Exchange Chromatography Teaching Kit

How To Use An Enzymatics Spark Dna Sample Prep Kit For Ion Torrent

Direct Testing Systems and Serology

VACUETTE RNAgard Blood Tubes BioMaxi Blood RNA Purification Kit. For the collection, preservation and purification of RNA from whole blood

LIVE/DEAD Fixable Dead Cell Stain Kits

Carcinoembryonic Antigen (CEA)

Parallel Circuits Charles Lang

Mix-N-Go Protein A Assays

HighPure Maxi Plasmid Kit

STANDARD OPERATING PROCEDURE

How to safely collect blood samples from persons suspected to be infected with highly infectious blood-borne pathogens (e.g.

Transcription:

This kit is intended for Research Use Only. Not for use in diagnostic procedures. Please use only the valid version of the package insert provided with the kit. One-step immunochromatographic test for the detection of Giardia antigens in stool samples 1 INTENDED USE Giardia Ag Rapid Test (RAP-5407 and RAP-5408) chromatographic immunoassay is a procedure for detection of Giardia antigens in stool samples. The test is based on the immunological capture of colored microparticles as they pass through a membrane on which the monoclonal antibody has been immobilized. 2 PRINCIPLE Giardia Ag Rapid Tests (RAP-5407 and RAP-5408) use monoclonal Giardia parvum-specific antibodies that detect all forms of the parasite's life cycle. The test is based on the use of red microspheres linked covalently to an anti-giardia lamblia monoclonal antibody, plus blue microspheres as test control. The parasite present in stool samples reacts with the latex particles which are coated with monoclonal antibodies specific against the antigen. This latex particles/antibodies/parasite complex migrates through a chromatographic process towards the reaction area. In this area, there are anti-giardia antibodies that react with the latex particles/antibodies/parasite complex. This reaction leads to the appearance of a red line. These lines are used to interpret the result, at five minutes' room-temperature incubation. 3 KIT CONTENTS Giardia Ag Rapid Test (Stool) (Cassettes) RAP-5407 Giardia Ag Rapid Test (Stool) (Strips) RAP-5408 20 Reaction devices (cassettes) 20 Reaction devices (strips) 20 vials containing dilution buffer The number of vials provided should be the same as the number of cassettes. 1 vial containing dilution buffer. The volume of the dilution buffer provided is proportional to the number of strips included and is indicated on the label of the vial containing it. The total volume of dilution buffer for all the strips is provided in a single vial. DRG International, Inc., USA Fax: (908) 233 0758 e-mail: corp@drg-international.com 1

4 MATERIAL NOT INCLUDED IN THE KIT Test tubes Pipettes that can be graduated Pipette tips 5 PRECAUTIONS 1. Specimen samples (stool) may contain infectious agents and should be treated and discarded as potentially dangerous biological materials. 2. The buffer contains sodium azide as an antimicrobial agent. Avoid direct contact with skin and mucosa. Discard appropriately. Do not use the buffer if there are indications of contamination or precipitation. 3. Do not store or prepare food, eat, drink or smoke in the area where the reagents and samples are handled. 4. Wear disposable gloves when handling the samples. Wash your hands thoroughly when you have finished working. 5. Do not exchange components from kits with different lot numbers. 6. Before using, let all kit components and samples reach room temperature, because cold reagents and/or samples can reduce test functionality. About 20-30 minutes are usually sufficient for reaching room temperature. 7. Do not use kit components after the expiration dates. 8. If the package is broken, the product can still be used as long as none of the components have been damaged. 9. It is important to add the correct sample amount (4-5 drops of suspension). If less than indicated is used, the chromatography may not occur because the sample may not reach the reaction area; if there is too much, brown lines may appear instead of red and blue ones. 10. The used product should be discarded as indicated by current legislation. 11. Do not use the test if a colored line appears in the result area before you start to use it. 12. It is important to take the appropriate amount of stool sample: 30-50 mg of solid stool or <100 µl of liquid stool. An excessive sample amount prevents correct chromatography. 13. In the case of the STRIP product (RAP-5408) packaged in a tube, it is important to close the tube again immediately after taking the reactive strip out; otherwise, high room humidity might damage the rest of the strips inside the tube. 6 LIMITATIONS 1. The test should be used only for detecting Giardia lamblia antigens in stool. 2. This is a qualitative test; no quantitative interpretation of the result should be made with respect to the intensity of the positive line. 3. A large number of samples were evaluated to ensure proper test functioning. The correlation of results with other techniques (ELISA) was excellent. However, interferences in the test performance cannot be excluded. DRG International, Inc., USA Fax: (908) 233 0758 e-mail: corp@drg-international.com 2

4. If there is too much sample, brown lines may appear instead of red and blue ones, or there may be no test and control lines because chromatography is not produced correctly. The brown lines have no value. If this happens, repeat the test with a smaller amount of stool sample or dilute the previously-prepared extract. 5. No crossed reaction with other substances has been observed during test evaluation. A negative result does not totally exclude the possibility of Giardia lamblia infection. 6. The analysis of some samples can produce lines of imprecise color, caused by negative samples in the majority of the cases. The test should be repeated if lines of indeterminate color appear. If the same result is obtained again, the analysis should be performed with another analytic method. 7 STORAGE The kit can be stored at temperatures between 2 C and 30 C (it can be stored in a refrigerator). The expiration date is stamped on the package. 8 SAMPLES Stool samples should be collected in an appropriate sterile container as soon as possible after symptom onset. The sample should be representative, if possible. The samples can be stored in the refrigerator (at approximately 4 C) for 1-2 days before being analyzed. To conserve samples for a longer period, store them in a freezer at -20 C, with no further handling. In this case, the sample must be completely defrosted, taken to room temperature and homogenized before analysis. 9 PROCEDURE 9.1 Giardia Ag Rapid Test (RAP-5407, cassette version) 1. Unscrew the vial cap carefully, so as not to spill the extraction buffer. Collect a sufficient amount of fecal matter (30-50 mg) with the end of the applicator. If the stool is liquid, collect 100 µl with the help of a pipette and transfer to the vial. 2. Introduce the applicator with the specimen into the vial. 3. Screw the cap on tightly and shake vigorously to ensure a homogenous mixture. 4. Take the reaction device out of the aluminum pouch. 5. Discard the small bag of desiccant; it is not used during the test, as its only purpose is to protect the test against moisture. 6. Break the upper end of the vial. 7. Add 4 or 5 drops to the sample area of the reaction device (round window marked with an arrow). Do not add solid particles along with the liquid. DRG International, Inc., USA Fax: (908) 233 0758 e-mail: corp@drg-international.com 3

8. Wait between no less than 5 minutes and no more than 10 minutes, then read and interpret the result. 9.2 Giardia Ag Rapid Test (RAP-5408, strip version) 1. Take one test tube per sample. Add approximately 1-1.5 ml of buffer. 2. Next, take a small amount of stool (30-50 mg) and resuspend it in the buffer. If a cotton swab is used, submerge it in the buffer and press the cotton end against the tube sides, rotating the swab. If the stool is liquid, collect 100 µl with the help of a pipette and transfer to the vial. 3. Shake vigorously to suspend the sample in the diluent (or vortex for 15 seconds). 4. Wait at least 3 minutes, until the solid particles have settled on the bottom, or centrifuge for 1 minute at 700 x g. Then, with the help of a pipette, transfer 150 µl of supernatant to another test tube. 5. Take the reaction device out of the aluminum pouch. Discard the small bag of desiccant; it is not used during the test, as its only purpose is to protect the test against moisture. 6. Insert the reactive strip into the 2 nd test tube, with the arrows pointing towards the bottom of the tube. IMPORTANT: The liquid must never reach the area with the arrows. If necessary, use a longer test tube or reduce the sample amount. 7. Incubate; read the result in the white area on the center of the strip at 5 minutes. Note: The strip can also be inserted into the first test tube or vial, as long as it is long enough for the liquid not to reach the area with the arrows. However, on rare occasions, absorption can be hindered due to a very concentrated sample. Alternatively, the strip can be submerged in the first test tube for 10 seconds (without going past the tips of the arrows) and then left to react on a horizontal surface. 10 READING RESULTS Interpretation code for strip images. G: Giardia lamblia N: Negative NEGATIVE: There is only a single BLUE line in the results area (in the cassette test, aligned with the letter "C" (Control) marked on the device frame). This line should always appear. POSITIVE Giardia: A BLUE line and a separate RED/PINK line appear in the results area (in the device test, aligned with the letter "T" (Test) marked on the device frame). The intensity can vary according to the antigen concentration present. DRG International, Inc., USA Fax: (908) 233 0758 e-mail: corp@drg-international.com 4

INVALID: The test is INVALID if no blue line appears because the procedure was not followed correctly, the reagents have deteriorated or because an incorrect amount of sample was added. Repeat the test with a new reaction device. Any line that may appear after 10 minutes due to the nature of the sample has no value. G Positive Negative Invalid Invalid 11 QUALITY CONTROL The test is invalid if no blue line appears, whether because the test was not performed properly or the reagents have deteriorated. If this happens, repeat the analysis, following the work protocol indicated in these instruction sheets closely. WARNING: Including a control with an established result is recommended to ensure that the data obtained are correct. 12 PROCEDURE LIMITATIONS 1. The test should be used only for detecting Giardia lamblia antigens in stool. DRG International, Inc., USA Fax: (908) 233 0758 e-mail: corp@drg-international.com 5

2. This is a qualitative test; no quantitative interpretation of the result should be made with respect to the intensity of the positive line. 3. Over 200 samples were evaluated to ensure proper test functioning. The correlation of results with other techniques (ELISA) was excellent. However, interferences in the test performance cannot be excluded. 4. If there is too much sample, brown lines may appear instead of red and blue ones, or there may be no test and control lines because chromatography is not produced correctly. The brown lines have no value. If this happens, repeat the test with a smaller amount of stool sample or dilute the previously-prepared extract. 5. No crossed reaction with other substances has been observed during test evaluation. A negative result does not totally exclude the possibility of Giardia lamblia infection. 6. The analysis of some samples can produce lines of imprecise color, caused by negative samples in the majority of the cases. The test should be repeated if lines of indeterminate color appear. If the same result is obtained again, the analysis should be performed with another analytic method. Rev. 12/13/11 cc DRG International, Inc., USA Fax: (908) 233 0758 e-mail: corp@drg-international.com 6