GUIDELINES FOR INTERLABORATORY QUALITY CONTROL IN ACCREDITED MEDICAL LABORATORIES

Similar documents
ISO and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi

RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES

QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES

QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08

MANAGEMENT REVIEW FOR LABORATORIES AND INSPECTION BODIES

International Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection

LAB 37 Edition 3 June 2013

NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory

Medical Cannabis Laboratory Approval Program

TR SUPPLEMENTARY REQUIREMENTS FOR THE ACCREDITATION OF CONTINUOUS AMBIENT AIR QUALITY MONITORING STATIONS

Content Sheet 10-1: Overview of External Quality Assessment (EQA)

NVLAP Assessor Training. Assessor Qualification & Training Requirements

Certification Process Requirements

PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS

How do I gain confidence in an Inspection Body? Do they need ISO 9001 certification or ISO/IEC accreditation?

Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing

NIST HANDBOOK CHECKLIST CONSTRUCTION MATERIALS TESTING

ISO/IEC QUALITY MANUAL

ISO 9001:2008 Clause PR018 Internal Audit Procedure

Cargo by Cargo. Carbon and Sustainability (C&S) Assurance Guide

Validation and Calibration. Definitions and Terminology

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

Quality Assurance (QA) & Quality Control (QC)

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

REGIONAL CENTRE EUROPE OF THE INTERNATIONAL FEDERATION OF TRANSLATORS

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION

Standard Practice for Quality Control Systems for Nondestructive Testing Agencies 1

Checklist. Standard for Medical Laboratory

Steps to ISO/IEC Accreditation

Accreditation and Competence?

Procedure for Equipment Calibration and Maintenance

3. Criteria for Recognition of Certification Bodies

USE OF REFERENCE MATERIALS IN THE LABORATORY

EffiValidation 3.0 software basic training module

R221 - Specific Requirements: Forensic

Quality Management Systems for Seed Testing Laboratories: Presented to the 2010 CSAAC Meeting. Valerie Martz Senior Laboratory Accreditation Officer

EA IAF/ILAC Guidance. on the Application of ISO/IEC 17020:1998

Example of a food company quality

Monitoring the Quality and Performance of Analytical Process Testing

SP007 - Procedure for the Control of Documents & Data

Certifying Reference Materials to ISO Guide 34

R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC LABORATORIES

Asset Support Contract Model Service Information. Annex 25 Integrated Asset Management

Procedures and General Requirements

Advanced Systems, Inc. Course. Registration Form. (Please complete all information)

rtech laboratories began working toward ISO 9002:1994 in ISO 17025: Practical Benefits of Implementing a Quality System STATISTICAL ANALYSIS

American Association for Laboratory Accreditation

Memorandum of Understanding

GLOBAL FUND QUALITY ASSURANCE POLICY FOR DIAGNOSTICS PRODUCTS. (Issued on 14 December 2010, amended on 5 February 2014)

IMPACTS OF BLUE DROP CERTIFICATION ON WATER QUALITY TESTING LABORATORIES

Procedure for Conducting Audits and Management Reviews

Good Clinical Laboratory Practice (GCLP) An international quality system for laboratories which undertake the analysis of samples from clinical trials

MODULE 7 REFERENCE TO ACCREDITATION AND ADVERTISING

NABL NATIONAL ACCREDITATION

OH&S MANAGEMENT SYSTEM CHECKLIST - AS 4801:2001 (STATUS A = Acceptable; N = Not Acceptable; N/A = Not Applicable)

Quality Management Standard BS EN ISO 9001:

List of EA Publications. And International. Documents

High Level Importance: ISO Microbiology compliance

P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014

QSS 0: Products and Services without Bespoke Contracts.

OPERATIONAL STANDARD

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009

EPA NATIONAL LEAD LABORATORY ACCREDITATION PROGRAM. LABORATORY QUALITY SYSTEM REQUIREMENTS (LQSR) REVISION 3.0 (November 05, 2007)

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009

Everything you want to know about ISO 15189:2012 Medical Laboratories Requirements for Quality and Competence

Training Programme. Laboratory Quality Manager and Laboratory Assessor November Vienna, Austria

Quality Management System Policy Manual

cc quality management

EA-7/01. EA Guidelines. on the application. Of EN Publication Reference PURPOSE

Presented by Rosemarie Bell 24 April 2014

1 For more information T: / E: DAP@ul.com / W: ul.com/dap

FSSC Certification scheme for food safety systems in compliance with ISO 22000: 2005 and technical specifications for sector PRPs PART I

Certification Process Requirements

CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

AS/NZS 4801:2001. Safety Management Systems (SMS) Self-Assessment Checklist. Revision 1 (January 2014)

2008 review 2013 review Targeted measures Targeted deadline. ENQA Criterion / ESG. February 2016/ Measures already taken. Level of compliance

IAF Informative Document. Transition Planning Guidance for ISO 9001:2015. Issue 1 (IAF ID 9:2015)

Continuing registration as a national VET regulator (NVR) registered training organisation (RTO)

American Association for Laboratory Accreditation

ONTARIO'S DRINKING WATER QUALITY MANAGEMENT STANDARD

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Derbyshire Trading Standards Service Quality Manual

Beamex. Calibration White Paper. How often should instruments be calibrated?

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

Guidelines on the Reporting of Compliance with Specification

DOCUMENTED PROCEDURE MANUAL

Quality Manual ISO 9001:2015 Quality Management System

COPY. Revision History

ISO/IEC Inspection Standard Application Document

JOHN HART GENERATING STATION REPLACEMENT PROJECT. Schedule 9. Quality Management

ASCLD/LAB-International

How to implement a Quality Management System

Application for Approved (Overseas/Local) Third Party Inspection Agency for Lifting Equipment

Transcription:

GUIDELINES FOR INTERLABORATORY QUALITY CONTROL IN ACCREDITED MEDICAL LABORATORIES Approved By: Chief Executive Office: Mike Peet Senior Manager: Christinah Leballo Date of Approval: 2008-02-15 Date of Implementation: 2008-02-15 SANAS Page 1 of 6

CONTENTS: 1. Purpose and Scope 2. Definitions and References 3. Requirements for laboratories participating in Interlaboratory Quality Control 3.1 Types of Interlaboratory Proficiency Testing 3.2 Supplier Requirements 3.3 Material Transport 3.4 Organisation and Management 3.5 General Requirements ADDENDUM 1: Amendment Record SANAS Page 2 of 6

1. Purpose and Scope This document is to be used as a guideline for medical laboratories implementing interlaboratory proficiency testing (IPT), also referred to as Internal Quality Assessment (IQA), or Interlab Quality Control. All assessments are to be conducted according to the applicable ISO Standard, i.e. either ISO/IEC 17025:2005 or ISO 15189. 2. Definitions and References SANAS PM SANAS A01 SANAS F29 SANAS F88 SANAS Policy Manual References, Acronyms and Definitions Generic Checklist for Accreditation of Medical Laboratories Accreditation Requirements for Medical Laboratories 3. Requirements for Laboratories Participating in IPT Any laboratory implementing an Interlab QC should conform to the relevant sections of ISO Guide 43, as well as the relevant sections of the ILAC document Guidelines for the Requirements for the Competence of Providers of Proficiency Testing Schemes ILAC-G13:2000. IPT is not intended to replace conventional EQA, which is considered an essential component of a laboratory s quality system. Rather, an IPT such as described in this guideline should only be used for tests where an appropriate EQA is not available. In addition, there may be instances where certain tests are analysed in such low volumes in laboratory that a formal EQA is not cost effective, and an IPT is more appropriate. Where it is not possible to perform an IPT, perhaps due to the lack of another laboratory performing the particular tests, then the laboratory should demonstrate traceability to a reference preparation or reference measurement procedure where possible; or if that is not possible, then acceptable repeatability should be well documented. 3.1 Types of Interlaboratory Proficiency Testing (IPT) The following types of IPT are defined by this guide : 3.1.1 Secondary IQA (2-IQA). This refers to interlaboratory testing of randomly selected fresh or prepared in-house samples, or alternatively, assayed or unassayed manufactured samples or reference samples between two or more laboratories. The preparation methods for the 2-IQA samples should have been thoroughly validated and documented to demonstrate that stability is maintained throughout preparation, transport and storage. 3.1.2 Interlaboratory random sample exchange between two or more laboratories, of which one may be a reference laboratory, where no EQA program is available or otherwise affordable due to low-volume testing of specific tests. 3.2 Supplier Requirements 3.2.1 If the material used is supplied by a manufacturer, the manufacturer should follow or conform to : i) ISO 9000; ii) DIN standards; iii) ANSI standards; iv) GMP; or v) be of proven international repute. 3.2.2 If the material used is manufactured or processed in-house, the laboratory should follow or conform to; i) standards as per ISO guide 43; or SANAS Page 3 of 6

ii) ILAC-G13:2000. 3.2.3 The material should be deemed suitable for testing with regard to: 3.3 Material Transport i) Each IPT sample should be uniquely identifiable; ii) There should be no patient identification on the sample container or in accompanying documentation; iii) Each sample should be stored and handled according to test-specific prescribed conditions; iv) Possible matrix effects should be documented for the participating laboratory's information and should be taken into account; v) Stability of the material is a pre-requisite for conducting IPT, and it should not be assumed that material sufficiently stable for internal QC will be suitable for use in an IPT scheme. 3.3.1 When the material is transported to the laboratory/ies the following should apply: i) The material should reach the laboratory within the stated time-frame; ii) The material should be transported within the specific temperature range; iii) The material should be transported in a manner which is safe to handlers and the general public; iv) There should be documented instructions for storage of the IPT material. 3.3.2 Specifications should be made on accompanying documentation of the analytical investigation/s to be undertaken on the IPT sample/s. 3.3.3 The accompanying documentation should specify whether a diagnostic opinion is required based on the specified investigation/s (where applicable). 3.4 Organisation and Management 3.4.1 The laboratory undertaking the management of an IPT program should have demonstrable competence in the operation, management and control of Proficiency Tests. 3.4.2 All aspects of the operation of the IPT must be documented, including but not restricted to achievable standards of performance and suitable statistical analysis, taking cognizance of the different instruments and methods that may be used for the tests covered by the IPT. 3.4.3 The IPT program should be undertaken at a frequency that allows poor performance to be identified timeously so that such poor performance does not affect patient sample analysis. 3.4.4 The stated frequency should be validated and sustained throughout the period of running the IPT. 3.4.5 There should be a documented procedure in each participating laboratory for material processing and analysis of the IPT material. 3.4.6 There should be a secure procedure for capturing the IPT analyte results, whether manual or electronic. 3.4.7 The results should be archived and retrievable. 3.4.8 There should be an independent site available in or outside the laboratory for the IPT data analysis. 3.4.9 The individuals responsible for the data analysis should be competent to undertake the task, and be experienced in handling statistical data. SANAS Page 4 of 6

3.4.10 There should be a secure procedure for relaying these results to the independent site for data analysis. 3.4.11 Timelines should be set for the receipt of results by the independent site prior to data analysis. 3.4.12 The laboratory independent site should have a documented system detailing the process for data analysis. 3.4.13 The data analysis system should discriminate between laboratory instruments and methodologies (where applicable). 3.4.14 The data analyses should be based on appropriate statistical methods and assumptions. 3.4.15 The data analysis system should report on levels of accuracy and precision based on consensus data or specific target values supplied by the manufacturer. 3.4.16 Access to the data analysis system should be secure. 3.4.17 The data should be archived for a specified time period. 3.4.18 Access to the data should be controlled. 3.4.19 Documented procedures should be available for the following: i) Report generation. ii) Appropriate action for poor performance in the IPT program 3.4.20 There should be a documented instruction detailing the individual/s designated responsible for receiving and acting on the report. 3.5 General Requirements 3.5.1 IPTs should conform to guidelines contained in ISO guide 43 and ILAC-G13:2000 for operating Proficiency Tests (PTs) (also known as External Quality Assessment (EQA) schemes). 3.5.2 These requirements include,but are not restricted to the following: i) Sample matrix and stability requirements, especially with regard to stability during transportation ii) Turn-around times of reports should be stipulated and adhered to. iii) Documentation and management of the IPT system iv) Selection of an appropriate model for statistical evaluation of the laboratory results, taking the different instruments and methods into consideration. v) Stipulated and validated standards of performance for laboratories taking part in the IPT. vi) Competence and training in the set-up and operation of IPTs. 3.5.3 All materials used for IPTs should be deemed potentially infectious and treated accordingly. SANAS Page 5 of 6

ADDENDUM 1. AMENDMENT RECORD Proposed By: Section Change CEO Section 2 Changed QM to PM CEO Page 1 Changed to new logo & Front page QM Changed from R 74 (Requirement) to TG 74 (Technical Guidance) document. SANAS Page 6 of 6