Technical Spotlight: Thermal Expansion Measurements of Insulative Materials and the Effect of the TMA Furnace Thermal Gradient

Similar documents
4 Thermomechanical Analysis (TMA)

New Methods of Testing PCB Traces Capacity and Fusing

CHARACTERIZATION OF POLYMERS BY TMA. W.J. Sichina, National Marketing Manager

Module 1 : Conduction. Lecture 5 : 1D conduction example problems. 2D conduction

Thermal cover:layout 1 1/18/11 3:56 PM Page 2 TA Instruments

Characterization of Electronic Materials Using Thermal Analysis

CMMs and PROFICIENCY TESTING

Analytical Testing Services Commercial Price List ManTech International Corporation January 2016

Thermal diffusivity and conductivity - an introduction to theory and practice

Using NTC Temperature Sensors Integrated into Power Modules

Making Sense of Laminate Dielectric Properties By Michael J. Gay and Richard Pangier Isola Group December 1, 2008

FORCE ON A CURRENT IN A MAGNETIC FIELD

Expansion and shrinkage of fibers

Temperatures of Bundled Electrical Cables Installed in Bored Holes in Residential Wood Framing

INTERNATIONAL ASSOCIATION OF CLASSIFICATION SOCIETIES. Interpretations of the FTP

TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here. Operating Instructions

Determination of Thermal Conductivity of Coarse and Fine Sand Soils

Physics 3 Summer 1989 Lab 7 - Elasticity

Exergy Analysis of a Water Heat Storage Tank

IC Temperature Sensor Provides Thermocouple Cold-Junction Compensation

How To Calculate Thermal Resistance On A Pb (Plastipo)

Practical Application of Industrial Fiber Optic Sensing Systems

Chapter 3 Non-parametric Models for Magneto-Rheological Dampers

Basic RTD Measurements. Basics of Resistance Temperature Detectors

/12/$ IEEE 1488

Lead-Free Product Transition: Impact on Printed Circuit Board Design and Material Selection

Dimensional Change Characteristics for Printed Circuit Board Films

Energy and Buildings

THERMAL CONDUCTIVITY AND THERMAL EXPANSION COEFFICIENT OF GFRP COMPOSITE LAMINATES WITH FILLERS

NTC Thermistors, Radial Leaded and Coated

Lecture 9, Thermal Notes, 3.054

Silicone Rubber Thermal Interface Materials: Applications and Performance Considerations

Characterization of Polymers Using TGA

Theoretical and Practical Aspects of Thermo Mechanical Reliability in Printed Circuit Boards with Copper Plated Through Holes

INVESTIGATION OF VISCOELASTICITY AND CURE SHRINKAGE IN AN EPOXY RESIN DURING PROCESSING

EFFECT OF OBSTRUCTION NEAR FAN INLET ON FAN HEAT SINK PERFORMANCE

Effects of Tg and CTE on Semiconductor Encapsulants

(Issued 1 Dec. 1965) CRD-C METHOD OF TEST FOR THERMAL CONDUCTIVITY OF LIGHTWEIGHT INSULATING CONCRETE 1

Evaluation of Laminated Flexible Printed Circuit Boards Under Wide Temperature Cycling

Measurement of Residual Stress in Plastics

Developments in Low Temperature Testing of Rubber Materials

The T-lam System

FEATURES BODY DIA. MAX. Type M AWG 30: [0.254] Type C AWG 28: [0.320] Type T AWG 30: [0.254]

Integration of a fin experiment into the undergraduate heat transfer laboratory

Phys222 W11 Quiz 1: Chapters Keys. Name:

Effect of Pressure and UV Intensity on the Rheological Cure Profile and Dimensional Change of Developing Networks

T H A N K S F O R A T T E N D I N G OUR. FLEX-RIGID PCBs. Presented by: Nechan Naicker

Unit 7: Electrical devices LO2: Understand electrical sensors and actuators Sensors temperature the thermistor

BIASING OF CONSTANT CURRENT MMIC AMPLIFIERS (e.g., ERA SERIES) (AN )

J.Instrum.Soc.India 30(1)29-34 PROGRAMMABLE CONTROL OF TEMPERATURE: A SIMPLE AND VERSATILE METHOD. N. Asha Bhat and K. S. Sangunni.

Experiment #3, Ohm s Law

Materials. Thermally Conductive Substrates & Thermal Management. by Ian Mayoh Page 12. March 2014

MOLAR AND PARTIAL MOLAR VOLUMES IN AQUEOUS SOLUTIONS (10/21/03)

Vapor Chambers. Figure 1: Example of vapor chamber. Benefits of Using Vapor Chambers

Table of content. L81/RITA high speed Thermo Balance. Quattro Dilatometer. L75/1250/B/S Macro Dilatometer. New air cooled furnace program

IDEAL AND NON-IDEAL GASES

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

Fox-2 MPPT Unit Testing

Compliant Terminal Technology Summary Test Report

Thermocline Management of Stratified Tanks for Heat Storage

HEAVY OIL FLOW MEASUREMENT CHALLENGES

Simple linear regression

16 2. Glass and heat. 2. Glass and heat. 2.1 Viscosity. Viscosity is the resistance against flowing.

Thermal Adhesives Ther-O-Bond 1500

Influence of material data on injection moulding simulation Application examples Ass.Prof. Dr. Thomas Lucyshyn

Corrosion of Copper in Water

Effect of design parameters on temperature rise of windings of dry type electrical transformer

ELECTRON SPIN RESONANCE Last Revised: July 2007

Optical modeling of finite element surface displacements using commercial software

DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE. E. Günther, S. Hiebler, H. Mehling

HOW ACCURATE ARE THOSE THERMOCOUPLES?

Engineering Feasibility Study: Vehicle Shock Absorption System

Determining Measurement Uncertainty for Dimensional Measurements

A simple scaling approach to produce climate scenarios of local precipitation extremes for the Netherlands

Fire Preventive and Fireproof Performance Test and Evaluation Procedure Manual

Chapter 10. Key Ideas Correlation, Correlation Coefficient (r),

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

75 Ω Transmission System

Design of Experiments (DOE) Tutorial

RESPONSE TIME INDEX OF SPRINKLERS

Check-list for auditors. Audit visit at candidate registered laboratories

KINETICS RISER SUPPORT SYSTEMS

TSL Professional Services

Testing and Reliability Improvement of High Reliability Consumer Electronics Products Manufactured on Printed Circuit Boards

HFM Heat Flow Meter Thermal Conductivity Analyzer

Evaluation of Soft Soldering on Aluminium Nitride (AlN) ESTEC Contract No /05/NL/PA. CTB Hybrids WG ESTEC-22nd May 2007

3.3 kv IGBT Modules. Takeharu Koga Yasuhiko Arita Takatoshi Kobayashi. 1. Introduction. 2. Specifications of 3.3 kv IGBT Module

STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H)

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

THE IMPACT OF YIELD STRENGTH OF THE INTERCONNECTOR ON THE INTERNAL STRESS OF THE SOLAR CELL WITHIN A MODULE

3 Technical Bulletin. Characteristics of Thermal Interface Materials. January, 2001

R&D Engineer. equipment. the power

TECH BULLETIN. Roofing No. Impact of Temperature on Polyiso R-value. Subject: Impact of Temperature on the R-value for Polyisocyanurate Insulation

CS4525 Power Calculator

VISCOSITY OF A LIQUID. To determine the viscosity of a lubricating oil. Time permitting, the temperature variation of viscosity can also be studied.

P O W E R C A P A C I T O R A S S E M B L I E S

Everline Module Application Note: Round LED Module Thermal Management

Understanding Boiling Water Heat Transfer in Metallurgical Operations

Prelab Exercises: Hooke's Law and the Behavior of Springs

A SURVEY AND TUTORIAL OF DIELECTRIC MATERIALS USED IN THE MANUFACTURE OF PRINTED CIRCUIT BOARDS.

Transcription:

Technical Spotlight: Measurements of Insulative Materials and the Effect of the TMA Furnace Thermal Gradient www.nts.com John Radman Senior Applications Engineer NTS Baltimore 5 North Park Dr. Hunt Valley, MD 213 (41) 584-999 john.radman@nts.com Abstract The Thermomechanical Analyzer (TMA) has long been the standard means of measuring the Coefficient of (CTE) of materials. In previous work, we have demonstrated that a significant vertical thermal gradient exists within the TMA furnace when testing materials of low thermal conductivity. The thermal gradient was measured to be approximately 2.5 to 3 o C per millimeter. (1) This paper studies the impact of this thermal gradient on thermal expansion measurements of materialsals with varying thermal conductivity characteristicss and thickness. The TMA instrument thermocouple location versus sample thickness was varied for each material type and the variation in the coefficient of thermal expansion was determined. Introduction This study focuses on the determination of the Coefficient of measurementsments by TMA methodology. The typical sample measuring.25 by.25 inches by the thickness (.75 inches maximum) sits on a quartz sample stage. A quartz probe exerting a light force contacts the top of the sample and records the sample movement along the thickness direction. The sample and probe are enclosed in a furnace chamber, Figure 1, where the temperature is controlled to ramp at a specified rate over a temperature range. As the temperature ramp progresses the expansion is plotted versus temperature. The thermal expansionn can be calculated from the curve. Figure 1 Arrangement of the sample on the TMA stage under the probe and the position of the thermocouple. It has been observed that differences exist in the coefficient of thermal expansion of insulative materials relative to sample thickness. The suspected cause was a temperature gradient across the thickness of the sample within the TMA furnace chamber coupled with the low thermal conductivity of the material. With the TMA thermocouple located at the level of the sample stage the temperature at the top of the sample, depending on thickness, may reach a higher temperature than at the bottom of the sample. The temperature gradient results in an approximate o C difference between the top and bottom of a 7 mm thick specimen. Page 1 of 5

Experiments and Results Three materials of low thermal conductivity were prepared at varying thickness. The materials were tested over a temperature range below the glass transition temperature. The TMA thermocouple was positioned at but not touching the sample stage. The instrument was calibrated with the thermocouple in this position. Assuming a sample thickness of 6 mm (approx..25 inch) a temperature difference between the top and the bottom of the sample is 3 o C/mm * 6 mm = 18 o C. Therefore during the temperature ramp, for example to 1 o C, the bottom of the sample is at 1 o C but the top of the sample is at 118 o C. Because of the low thermal conductivity of the material, temperature equilibrium is never reached across the sample thickness. The result of this equilibrium imbalance is the top of the specimen is expanding at a greater rate than the bottom of the specimen thus skewing the results toward higher thermal expansion values., an Acrylic polymer, was tested at thickness values of nominally 1.3, 6.8, and 12.2 mm. Calculation of the CTE values indicated a clearly increasing CTE versus thickness trend. The results are presented in Figure 2 as a comparative representation of the increasing trends of thickness, CTE, and percent increase in CTE. The CTE increase from the thinnest, 1.3 mm, to the thickest, 12.2 mm, sample was 23% or 2.1% increase in CTE/mm increase in thickness. Figure 3, a plot of the thickness versus CTE values, indicates the increasing CTE values are linear over the thickness range. CTE, ppm/deg c 9 7 5 3 1 Figure 2 Comparative test results from. 1 9 7 5 5 1 15 Thicknes s, m m Figure 3 CTE versus sample thickness for. Page 2 of 5

, Rexolite 1422, was tested at thickness values of nominally 1.6, 12.3, and 16.5 mm. The CTE values increased with each thickness increment. The results are presented in Figure 4 as a comparative representation of the increasing trends of thickness, CTE, and percent increase in CTE. The results in Figure 4 indicate an increase of 18% from the 1.6 mm sample to the 16.5 mm sample. A 14% increase in the CTE was calculated between the 1.6 and 12.3 mm specimens. The increase in CTE/mm increase in thickness = 1.7%. Figure 5, a plot of the thickness versus CTE values, indicates the increasing CTE values are linear over the thickness range., a glass/epoxy laminate, was tested at thickness values of nominally 1.1, 5.3, and 14.5 mm. The CTE values increased with each thickness increment. The results in Figure 6, a comparative representation of the increasing trends of thickness, CTE, and percent increase in CTE, illustrates a CTE increase of 1% from the 1.1 mm sample to the 5.3 mm sample and a 32% increase in the CTE between the 1.1 and 14.5 mm specimens. The increase in CTE/mm increase in thickness = 2.4%. Figure 7, a plot of the thickness versus CTE values, indicates the increasing CTE values are linear over the thickness range. 9 1 7 1 5 3 1 Figure 4 Comparative test results from. Figure 6 Comparative test results from. 9 1 85 11 1 75 9 7 65 7 5 1 15 5 1 15 Figure 5 CTE versus sample thickness for. Figure 7 CTE versus sample thickness for. Page 3 of 5

Copper control samples were tested at similar conditions to determine if CTE differences occur with thermally conductive material. Testing over the thickness range of 1.5 mm to 1.4 mm resulted in a CTE increase of approximately 4%, or an increase in CTE/mm increase in thickness =.4%. Figure 8. These results indicate thermally conductive materials exhibit significantly less CTE variation versus thickness than thermal insulators. Figure 9 suggests the rate of CTE increase is dependent upon the thermal conductivity of the material. The slope of the lines for Materials A and C are nearly parallel. These materials have more closely matched thermal conductivity values than. The rate of CTE increase versus thickness is lower for the more thermally conductive. 17 16.8 16.6 16.4 16.2 16 15.8 15.6 15.4 15.2 15 Copper 2 4 6 8 1 12 Figure 8 CTE versus sample thickness for Copper. Conclusion From previous work it is known that a thermal gradient exists within samples in the TMA furnace while testing materials of low thermal conductivity. The thermal gradient is considerably less pronounced for high thermal conductivity materials such as copper (TC = 398W/mK). The materials selected for the experimental procedures were of low thermal conductivity, =.14 W/mK and =.56 W/mK represent the lower and upper ends of the thermal conductivity scale for polymers, and =. W/mK is typical for circuit board laminate. 1 1 CTE vs Thickness 5 1 15 Figure 9 CTE versus sample thickness for Materials A, B, and C. The test results indicate that higher CTE values are obtained as a function of increasing sample thickness when tested in a TMA at the typical thermal ramp rate of 1 o C/min. The CTE increase versus material thickness was linear for all three materials although the rate of increase was greater for the less thermally conductive samples. The less thermally conductive samples, Materials A and C, exhibited a CTE increase/mm increase in thickness of 2.1 and 2.4%, whereas more thermally conductive exhibited an increase in CTE/mm increase in thickness of 1.7%. High thermal conductivity materials such as copper the increase in CTE/mm increase in thickness becomes nearly negligible at.4%. The varying CTE values could have serious implications in tight tolerance or highly critical applications. When practical, the material should be tested as close to the application thickness as possible, within the confines of the analytical instrument, at conditions simulating actual operating thermal environments. Similar improvement may be possible by adjusting the temperature ramp rate of the instrument to allow more time for thermal equilibrium to occur within the test specimen. Page 4 of 5

References 1. Z. Mei, M. Hu, K. Ogle, J. Radman, R. Michalkiewicz, An Issue in Time to Delamination (T 2 ) Testing for PCBs, Electronic Circuit World Convention 1 Technical Conference Proceedings, February 27, 5 For more information concerning the testing of your products, please contact us at () 27-2516, or email us at RFQ@nts.com. Page 5 of 5