Ettan IPGphor Cup Loading Manifold



Similar documents
designed WORK IPGphor II First-Dimension Isoelectric Focusing System AA

2D gel Protocol. 2. Determining Protein Concentration of cell lysates

ReadyKart Mobile Processing Station Instructions for Use

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

Amersham High Molecular Weight Calibration Kit for native electrophoresis

GE Healthcare. UNICORN 6.x. SQL Server 2008 configuration for UNICORN installations

UniFlux 10: hollow fiber conversion kit ( ) installation instruction

HiTrap Heparin HP, 1 ml and 5 ml

How to Download a GE Healthcare License File

2D gel electrophoresis

Page finder. 1. Legal Handling Safety warnings and precautions Storage Expiry 4

Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use

GE Healthcare Life Sciences. Validation Services. Compliance support through life cycle management

Instrument services for WAVE Bioreactor systems

GE Healthcare. UNICORN 5.2 vs Software Change Description

CU-950 Installation Guide

UNICORN. Hardening Guide

soft ware change description

Immobiline DryStrip Reswelling Tray

Western Blotting For Protein Analysis

EZ-PAGE Electrophoresis System USER MANUAL

PART 1 - INTRODUCTION...

Please read and keep these instructions. Run water through your machine before first use

Blue Sepharose 6 Fast Flow

Process-scale purification of monoclonal antibodies polishing using Capto Q

HAIR CLIPPER. Model No. HC5150 HC

Troubleshooting Polyacrylamide Gel Electrophoresis (PAGE)

Summary of GE Healthcare's Quality Management System (QMS) Covering BioProcess chromatography media, equipment, software, and spare parts

Convection TCO600. For your safety and continued enjoyment of this product, always read the instruction book carefully before using.

Two-Dimensional Gel Electrophoresis

INSTRUCTIONS Edition AC

Troubleshooting Guide for DNA Electrophoresis

illustra MicroSpin G-25 Columns

Electric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL

Trace Dissolved Oxygen Sensor

NFC Communication Tray Model: HHX-IT3-Z

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference

UltraClean Soil DNA Isolation Kit

VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised

San josé OWNER S MANUAL

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

PCR and Sequencing Reaction Clean-Up Kit (Magnetic Bead System) 50 preps Product #60200

Convection TCO650. For your safety and continued enjoyment of this product, always read the instruction book carefully before using.

OPERATOR S MANUAL 18 VOLT, 1 HOUR CHARGER

Two-Dimensional Gel Electrophoresis (2-DGE)

SDS-PAGE Protocol Mutated from the SDS-PAGE protocol written by the Lord of the Flies

Milwaukee USER MANUAL

His GraviTrap. GE Healthcare. Operation

SenTix 950 SenTix 980

SCION tc DASH APPLIQUE Preparation

Care and use guide. Contents. a handy solution made easy. page

Gain efficiency in your process development with ÄKTA avant

CORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

15GAL STEEL OIL DRAIN WITH 110V PUMP

PROFESSIONAL POPCORN MAKER

Installation Instructions S

Pet hair clipper. Model Diagrams within this manual may not be drawn proportionally.

For additional information, please visit our website at

POOL ROVER SERIES 2 ROBOTIC POOL CLEANER PRODUCT SETTINGS AND TROUBLESHOOTING GUIDE

SBC90. Abrasive Blast Cabinet Assembly & Operating Instructions

STANDARD OPERATING PROCEDURE

Dissolved Oxygen Sensors

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

DNA Electrophoresis Lesson Plan

MAINTENANCE & TROUBLESHOOTING

Beautifully Traditional

Trillium 63 Instruction Manual

Electronics and Soldering Notes

Record Cleaning Machine

Objectives: Vocabulary:

1000-LB. TRAILER JACK OWNER S MANUAL

75th anniversary blender

Tassimo The Art of the Perfect Cup

ReadyPrep Protein Extraction Kit (Soluble/Insoluble) Instruction Manual. Catalog #

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS

CHARISMA SHOWER/TUB DOOR INSTALLATION INSTRUCTIONS

SAFETY & OPERATING INSTRUCTIONS

3 WATT LED SPOTLIGHT Model No. SLM

PLEASE READ ALL INSTRUCTIONS BEFORE USE AND SAVE A COPY FOR FUTURE REFERENCE!

Wall-Mounting your HP computer. User Guide

HYDRAULIC LIFT TABLE CART 2200-LB.

DC REFRIGERATORS 12/24 VOLTS INSTALLATION AND OWNER S MANUAL

illustra puretaq Ready-To-Go PCR Beads

How To Test For Cleaning Efficiency With A Predictor 96 Well Filter Plate

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102

ATS Overhead Table Shelf System INSTRUCTION MANUAL

BedRug Bedliner Installation Instructions. Part Number(s): Multiple. Preparation

Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Chemicals Needed:

UV100A Ultraviolet Air Treatment System

Instruction Manual. AWK-115S

HiPer Ion Exchange Chromatography Teaching Kit

12 Volt 30 Amp Digital Solar Charge Controller

Antenna Splitter ASA 1. Instruction manual

GAMP5 - a lifecycle management framework for customized bioprocess solutions

SunMaxx Solar Filling Station Operating Instructions

Hegel H1 High End Integrated Amplifier

Cleaning Instructions, Pilot Replacement and Valve Change. Model No.: Natural Gas

Transcription:

GE Healthcare Life Sciences Ettan IPGphor Cup Loading Manifold User Manual

Table of Contents Table of Contents 1 Introduction... 1.1 Important user information... 2 Description... 2.1 Overview... 2.2 Accessories... 3 Operation... 3.1 Instructions... 3.2 Run conditions... 4 Maintenance... 5 Troubleshooting... 6 Reference information... 5 5 7 7 13 15 15 23 25 27 31 Ettan IPGphor User Manual 11-0034-60 AD 3

1 Introduction 1 Introduction Purpose of the User Manual The User Manual provides you with the instructions needed to use Ettan IPGphor Cup Loading Manifold together with an Ettan IPGphor unit. In this chapter Section 1.1 Important user information See page 5 1.1 Important user information Read this before operating Ettan IPGphor unit All users must read the entire Ettan IPGphor Operating Instructions before installing, operating or maintaining the instrument. Always keep the Operating Instructions at hand when using Ettan IPGphor. Do not operate Ettan IPGphor in any other way than described in the user documentation. If you do, you may be exposed to hazards that can lead to personal injury and you may cause damage to the equipment. Intended use The Ettan IPGphor Cup Loading Manifold in combination with GE Healthcare's Ettan IPGphor, Ettan IPGphor II and Ettan IPGphor 3 Isoelectric Focusing Units, is suitable for first-dimension isoelectric focusing (IEF) of proteins on Immobiline DryStrip immobilized ph gradient (IPG) gel strips 3 mm wide, from 7 cm to 24 cm long. Ettan IPGphor is intended to be used as a laboratory device for research purposes. Ettan IPGphor shall not be used in any clinical procedures, or for diagnostic purposes. Ettan IPGphor User Manual 11-0034-60 AD 5

1 Introduction 1.1 Important user information Safety notices This user documentation contains WARNINGS, CAUTIONS and NOTICES concerning the safe use of the product. See definitions below. Warnings WARNING WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury. It is important not to proceed until all stated conditions are met and clearly understood. Cautions CAUTION CAUTION indicates a hazardous situation which, if not avoided, could result in minor or moderate injury. It is important not to proceed until all stated conditions are met and clearly understood. Notices NOTICE NOTICE indicates instructions that must be followed to avoid damage to the product or other equipment. Notes and tips Note: Tip: A Note is used to indicate information that is important for trouble-free and optimal use of the product. A tip contains useful information that can improve or optimize your procedures. Typographical conventions Software items are identified in the text by bold italic text. A colon separates menu levels, thus File:Open refers to the Open command in the File menu. Hardware items are identified in the text by bold text (e.g., Power switch). 6 Ettan IPGphor User Manual 11-0034-60 AD

2 Description 2 Description About this chapter This section provides an overview of the Ettan IPGphor Cup Loading Manifold unit. In this chapter This chapter contains the following sections: Section 2.1 Overview 2.2 Accessories See page 7 13 2.1 Overview Introduction The Ettan IPGphor Cup Loading Manifold, in combination with GE Healthcare's Ettan IPGphor, Ettan IPGphor II and Ettan IPGphor 3 Isoelectric Focusing Units, is suitable for first-dimension isoelectric focusing (IEF) of proteins on Immobiline DryStrip immobilized ph gradient (IPG) gel strips 3 mm wide, from 7 cm to 24 cm long. IEF is performed with the gel side facing up. Samples up to 150 μl are applied in an open-bottom loading cup. Cup loading has been found to improve protein focusing patterns, particularly on basic IPG strips (ph 6 9 and 6 11). The manifold can accommodate anodic and cathodic cup loading. Larger volumes of the sample, 375-500 μl, are applied with paper bridge application. Sample volumes up to the volume of the strip (18 cm strip = 340 μl) can be applied by in gel rehydration loading. Ettan IPGphor User Manual 11-0034-60 AD 7

2 Description 2.1 Overview Illustration of the Ettan IPGphor Cup Loading Manifold The illustration below shows the main components of the Ettan IPGphor Cup Loading Manifold: 2 4 7 1 3 6 9 5 8 10 Part Function Part Function 1 Manifold 6 Pre-cut paper bridges 2 Pre-cut electrode wicks 7 Sample cup insertion tool 3 Electrode assembly 8 Cleaning solution 4 Sample cups 9 Forceps 5 Cleaning brush 10 Spirit level 8 Ettan IPGphor User Manual 11-0034-60 AD

2 Description 2.1 Overview Manifold Ceramic Manifold The manifold (see figure below) is made of aluminum oxide ceramic for efficient heat transfer and temperature control during IEF. Protrusions along the channel inside the manifold align the rehydrated IPG strip, keeping it straight and centered when placed inside the manifold. The manifold has a coating to minimize protein adsorption. Because some cleaning agents can remove this coating, clean the manifold only with the Ettan IPGphor Strip Holder Cleaner. Note: Manifold Light The manifold is fragile and should be handled with care. The manifold light is made of the polymers PC (polycarbonate) and ABS (acrylonitrilebutadiene-styrene). The manifold light's flat design is important for the heat transfer. If exposed to very high temperatures, the flatness could be affected. Refer to Manifold Light, on page26 regarding cleaning temperature. General information IPG strips from 7 cm to 24 cm long may be used. The unit may also be used for IPG strip equilibration prior to second dimension electrophoresis. Forcep wells are located at the ends of the channels to aid strip removal after electrophoresis. forcep wells channel numbering centering protrusion Ettan IPGphor User Manual 11-0034-60 AD 9

2 Description 2.1 Overview Electrode assembly The movable electrodes, as shown below, can be placed along the manifold to fit the various length of Immobiline Dry Strip. Electrode pins make electrical contact with the power supply pads on the bed of the Ettan IPGphor. The electrodes have a platinum wire that is strung across the bottom of the electrode teeth to provide electrical contact to the IPG strips. The hold down teeth are found along the opposite length of the electrode. They are used to apply pressure on the paper bridges when the manifold is used for bridge loading. Cams on the sides of the electrode secure the electrodes in place and must be in the closed position during IEF. cam in open position hold down teeth electrode teeth cam in closed position 10 Ettan IPGphor User Manual 11-0034-60 AD

2 Description 2.1 Overview Sample cups Sample cups, as shown in the figure below, can be placed along the length of the Ettan IPGphor manifold that is not blocked by a centering protrusion. The sample cups are supplied in strips of 6 for easy handling and placement. If fewer than 6 cups are required, they may be easily separated by cutting the thin plastic bridging the strip of cups together. The sample cups can accommodate sample volumes of up to 150 μl. For proper sealing of the cup to the gel, all of the feet of the sample cup must rest on the bottom of the channel and all cup arms must be fully pressed down into the channel. cup arms cup feet Ettan IPGphor User Manual 11-0034-60 AD 11

2 Description 2.1 Overview Sample cup insertion tool The sample cup insertion tool, shown below, is supplied with the manifold kit and refill packages of sample cups. It is used for proper placement and sealing of the cup over the IPG strips. IEF electrode paper wicks The small, precut rectangular wicks, shown below, must be placed at both the anodal and cathodal ends of the rehydrated IPG strips just under the electrodes. The wicks absorb excess water, salts, and proteins with pis that lie outside the ph range of the IPG strip. 12 Ettan IPGphor User Manual 11-0034-60 AD

2 Description 2.1 Overview Paper bridge loading pads Paper bridge loading pads, as shown below, are used to load large volumes of sample (375 500 μl). The paper bridges are 0.8 cm 5.0 cm in size with pointed ends. One end of the pad contacts the IPG strip and is held in place by the hold down teeth of the electrode assembly. The other end of the paper bridge loading pad makes contact with a paper wick placed under the platinum wire of the electrode assembly. 2.2 Accessories Introduction The following accessory items are not supplied with the unit and need to be ordered separately. Immobiline DryStrip Cover Fluid Immobiline DryStrip Cover Fluid is required to ensure that the rehydrated IPG strip gels do not dry out during electrophoresis. Without cover fluid, the strips will dry out, urea will crystallize, and the sample will not focus properly. Ettan IPGphor User Manual 11-0034-60 AD 13

2 Description 2.2 Accessories Immobiline DryStrip Reswelling Tray The Immobiline DryStrip Reswelling Tray is required for proper strip rehydration. The channel in the Cup Loading Manifold is too wide to ensure proper absorption of the required volumes of rehydration solution. Two trays are available - one for 7 18 cm strip lengths and a second for 7 24 cm strip lengths. 14 Ettan IPGphor User Manual 11-0034-60 AD

3 Operation 3 Operation About this chapter This section provides instructions on how to operate the Ettan IPGphor Cup Loading Manifold unit. In this chapter This chapter contains the following sections: Section 3.1 Instructions 3.2 Run conditions See page 15 23 3.1 Instructions Rehydrating the IPG strips Follow the instructions below to rehydrate the IPG strips. Step 1 Action Rehydrate the IPG strips with the gel side down in the appropriate volume of rehydration solution using the Immobiline DryStrip Reswelling Tray. Rehydration in the Cup Loading Manifold is not recommended: the channel is too wide to ensure proper rehydration. Ettan IPGphor User Manual 11-0034-60 AD 15

3 Operation 3.1 Instructions Step 2 Action Follow the instructions included with the Immobiline DryStrips. The table below is provided as a reference. IPG strip length (cm) 7 11 13 18 24 Rehydration volume (μl) 125 200 250 340 450 3 Cover the IPG strips with Immobiline DryStrip Cover Fluid and allow the strips to rehydrate overnight (10 20 hours). Placing the Manifold on the IEF unit Follow the instructions below to place the manifold on the IEF unit. Step 1 Action Clean and dry the IPGphor bed before placing the manifold on the unit. Position the manifold on the IPGphor platform. 16 Ettan IPGphor User Manual 11-0034-60 AD

3 Operation 3.1 Instructions Step 2 Action The small T-shaped protrusion fits into a cutout section of the IPGphor bed near the lid hinge, see figure below. Ensure that the manifold is level by placing the round spirit level on the center of the manifold after it is placed on the Ettan IPGphor unit. Transferring the IPG strips to the Manifold NOTICE Make sure that the Ettan IPGphor unit is placed on a stable flat surface. Follow the steps below to transfer the IPG strips to the manifold. Step 1 Action Measure out 108 ml of Immobiline DryStrip Cover Fluid (even if fewer than 12 strips will be loaded into the manifold). Add the cover fluid evenly in the 12 manifold channels. Ettan IPGphor User Manual 11-0034-60 AD 17

3 Operation 3.1 Instructions Step 2 3 Action Transfer the strips to the Ettan IPGphor Cup Loading Manifold. Place the strips under the cover fluid with the gel surface face up and the anodic (+, pointed) end to the anode of the Manifold. Place the strip so that the electrode assembly accommodates at each end of the strip without hiding the strip, see figure below, to be able to see the brome phenol blue moving away at start. Center the strip down the length of the manifold channel. Protrusions along the sides guide the strip approximately straight, although some manual adjustment of the strip may be necessary, as shown below. 24 cm 18 cm 13 cm 11 cm 7 cm Strip length (cm) 7 11 13 18 24 Note: Using 24 cm long Immobiline Dry Strip and cathodic cup loading is going to be used, the strips should be placed 3 4 cm further up compared to the figure above to make the cathode electrode assembly accommodated at the cathode end, not hiding the sample application place. 18 Ettan IPGphor User Manual 11-0034-60 AD

3 Operation 3.1 Instructions Placing sample cups Place the sample cups following the instructions below. Step 1 2 3 Action Place a strip of cups in the appropriate position as shown in the figure below, for example ~1 cm from the end of the gel portion of the IPG strip. Do NOT place the cup with the feet over a center protrusion. Push the cups into the channels with gloved fingers, starting at one end of the strip and working towards the other. Align the insertion tool over the cups and push down to ensure that the feet of the cups are properly seated at the bottom of the channel (wiggle the tool gently while pushing down in order to ensure that the cups are seated as far down as they will go). Take care not to move the cups while removing the insertion tool. Fill the cups with cover fluid to test for proper seating of the cups. Remove the cover fluid after 10 minutes. incorrect cup position correct cup position Note: Cups must not straddle the centering protrusions on the bottom of the channels. Ettan IPGphor User Manual 11-0034-60 AD 19

3 Operation 3.1 Instructions Placing paper wicks Follow the instructions below to place the paper wicks. Step 1 2 Action Count out the appropriate number of precut paper wicks. Two wicks per strip are required. Separate the wicks from each other. Add 150 μl distilled water to each wick. Place the wicks on the IPG strips such that one end of the wick overlaps the end of the gel on the IPG strip, as shown in figure below. The electrode must contact the wick. 20 Ettan IPGphor User Manual 11-0034-60 AD

3 Operation 3.1 Instructions Step 3 Action With the electrode cams in the open position, place the electrode assembly on top of all the wicks. Swivel the cams into the closed position under the external lip of the tray. The electrodes should not be moved while the cams are in the closed position, see figure below. Electrode Cup Paper wick IPG Strip Sample application (loading) Follow the instructions below to load the sample. Step 1 Action Briefly centrifuge the protein sample prior to loading to remove insoluble material and particulate matter. These materials will impede sample entry and result in vertical streaks in the second-dimension gel. Ettan IPGphor User Manual 11-0034-60 AD 21

3 Operation 3.1 Instructions Step 2 Action Load samples into the sample cups. A maximum of 150 μl of sample may be placed in these cups. Check to make sure that there is cover fluid over the samples. When the cups are initially placed on the manifold, cover fluid will flow into the cups as they are seated. When sample is introduced into the cups, the sample will sink to the bottom of the cup and contact the IPG strip. Note: For basic IPG strips, superior focusing patterns are generally obtained when the sample cup is placed at the anodic (+) end of the Immobiline Dry Strip. 3 Close the Ettan IPGphor lid. Paper-bridge loading Large sample volumes and large protein amounts can be applied using paper bridge loading. A paper pad (paper bridge) is soaked with sample and placed between the anodic end of the Immobiline DryStrip gel and the electrode 375 500 μl sample can be applied using the paper bridge pads. Solutions containing up to 5 mg protein have been loaded on a 18 cm-long narrow ph range Immobiline DryStrip gel. The rehydrated Immobiline DryStrip gel is first positioned in the bottom of the manifold channel. Then the paper bridge with sample is positioned, followed by a paper wick (see figure below). With anodic application the anode electrode is positioned as far out as possible in the electrode assembly, while the cathode electrode is positioned close to the end of the Immobiline DryStrip gel to ensure good contact between the paper wick and Immobiline DryStrip gel. Note: The anodic application point is an important factor for obtaining good results. electrode assembly paper wick IPG strip anode paper bridge hold down teeth 22 Ettan IPGphor User Manual 11-0034-60 AD

3 Operation 3.2 Run conditions 3.2 Run conditions Introduction Program the Ettan IPGphor with the recommended run parameters. Ramping the voltage slowly while the sample is entering the IPG strip will improve results. Instructions for optimal ramp, voltages and times, or Vhrs (volt-hours) totals are attached to each package of Immobiline Dry Strip. WARNING Ettan IPGphor is a high voltage instrument that can cause fatal electrical shock if the safety features are disabled. The safety lid must be securely closed before starting a protocol. WARNING The Ettan IPGphor unit is capable of producing thousands of volts. Before operating the unit, read and fully understand the Ettan IPGphor operating instructions and warnings. Ceramic manifold Follow the recommended run conditions found in the Instructions attached to each Immobiline DryStrip package. Manifold Light Follow the recommended run conditions found in the Instructions attached to each Immobiline DryStrip package except for focusing time. CAUTION Do not use any organic solvents during cleaning. Focusing time As a guideline, use the lower number of recommended volt-hours to avoid overfocusing. Normally the focusing results does not improve after this time. Ettan IPGphor User Manual 11-0034-60 AD 23

3 Operation 3.2 Run conditions WARNING Check that the high voltage is turned off before opening the safety lid. The HV ON indicator on the front panel should not light. 24 Ettan IPGphor User Manual 11-0034-60 AD

4 Maintenance 4 Maintenance Precautions WARNING Disconnect power. Always disconnect power from the instrument before performing any maintenance task. NOTICE The ceramic strip holders and manifold are very brittle and fragile. Take care in handling and do not subject to impacts. NOTICE Avoid scratching the electrode surfaces. NOTICE Keep the instrument dry and clean. Wipe regularly with a soft damp tissue and, if necessary, a mild cleaning agent. Let the instrument dry completely before use. Ceramic Manifold NOTICE The ceramic strip holders and manifold trays are very brittle and fragile. Handle carefully: Do not bump or drop. Ettan IPGphor User Manual 11-0034-60 AD 25

4 Maintenance Gently clean the manifold after each use. Wash the manifold channels with the Ettan IPGphor Strip Holder Cleaner. Rinse thoroughly with water and then deionized water and allow the manifold to air dry. Do not use abrasive cleaning agents or pads. Take care to avoid chipping the manifold. Manifold Light Clean the Ettan IPGphor Manifold Light with water, no warmer than 40 C, and with IPG Strip Holder cleaning solution. NOTICE Do not use any organic solvents. Electrodes Wash the electrode assemblies with the Ettan IPGphor Strip Holder Cleaner. Rinse thoroughly with water then deionized water and allow to air dry. NOTICE Do not use strong acids, bases, ketones, alcohols, or other reagents to clean the covers, sample cups, or electrodes or the parts may be damaged. Cups may be briefly rinsed with ethanol if desired. 26 Ettan IPGphor User Manual 11-0034-60 AD

5 Troubleshooting 5 Troubleshooting Troubleshooting first-dimension IEF The following table specifies the troubleshooting situations that you might encounter while using the Ettan IPGphor Isoelectric Focusing System employing the Cup Loading Manifold. Symptom Current is too low or zero Possible cause Electrical continuity is impeded. Remedy Check the external electrode contacts: The electrodes at the bottom of the strip holder (one at each end) must make metal to-metal contact with the appropriate electrode contact area. Check the internal electrode contacts: The gel (which becomes visible because of the dye in the rehydration solution) must contact both electrodes in the manifold through the paper wicks and/or paper bridge parts. Check that the IPG strip is fully rehydrated along its entire length. Electrical contact at the electrodes is reduced by incomplete rehydration. Check that the paper wicks are present, wet and properly positioned. Ettan IPGphor User Manual 11-0034-60 AD 27

5 Troubleshooting Symptom Voltage too low or does not reach maximum set value Possible cause Ettan IPGphor protocol settings are incorrect for the experiment. Conductivity/ionic strength is too high. Remedy Check that the current limit is properly set. Check that the actual number of Immobiline DryStrip gels on the Ettan IPGphor platform is the same as the number of gels entered in the protocol. Prepare the sample to yield a salt concentration less than 10 mm. The recommended IPG Buffer concentration is 0.5%. A maximum of 2% is advisable only if sample solubility is a problem. High conductivity can also arise from the use of poor quality urea or other denaturants. Urea is also prone to decompose to charged breakdown products. Salts and ionic impurities in the sample can raise the conductivity of the strip. Shorter length IPG strips (e.g. 7 cm strips) will not reach 8000 V. The distance between the electrodes is shorter so that the voltage gradient (V/cm) required to reach the 50 μa current limit is reached at a lower overall voltage. 28 Ettan IPGphor User Manual 11-0034-60 AD

5 Troubleshooting Symptom Sample leaks from cup Sparking or burning in the Immobiline DryStrip gels Immobiline DryStrips turn white and opaque after focusing Possible cause Incorrect cup placement. Incorrect strip placement. Current limit setting is too high. Immobiline DryStrip gel is not fully rehydrated. Immobiline DryStrip gels dried out during IEF. Immobiline DryStrip gels dried out during IEF. Remedy Check that the feet of the cups are resting on the bottom of the manifold channel. Check for correct positioning of sample cup arms. Check that the feet of the cups are not resting on a centering protrusion in the channel. Check that the strip is centered inside of the channel. Do not exceed the maximum recommended setting of 75 μa to 18 and 24 cm strips and 50 μa to 7, 11 and 13 cm strips per Immobiline DryStrip gel. Ensure that the Immobiline DryStrip gels are rehydrated with a sufficient volume of rehydration solution. Remove any large bubbles trapped under the Immobiline DryStrip gel after placing it on rehydration solution. Always apply Immobiline DryStrip Cover Fluid to prevent dehydration of rehydrated Immobiline DryStrip gels. Always apply recommended amount of Immobiline DryStrip Cover Fluid to prevent dehydration of rehydrated Immobiline DryStrip gels. Ettan IPGphor User Manual 11-0034-60 AD 29

5 Troubleshooting Symptom Immobiline DryStrip Cover Fluid overflows from Cup Loading Manifold Possible cause Excess cover fluid added. Remedy Do not add more than the recommended volume. Ensure that the outside rim of the tray does not have any oil on it. 30 Ettan IPGphor User Manual 11-0034-60 AD

6 Reference information 6 Reference information Technical service and repair GE Healthcare offers complete technical support for all our products. If you have any questions about how to use this product, please contact your local GE Healthcare representative. Ordering information The following table provides you with information about replacement parts and their code numbers. Replacement part Ettan IPGphor Cup Loading Manifold, Ceramic Ettan IPGphor Cup Loading Manifold Light Sample cups, pack of 20 (6x) Paper electrode wicks, pack of 40 (6x) Paper bridge pads, pack of 20 (6x) Electrode set Cleaning brush Spirit level Forceps SS Code number 80-6498-57 11-0025-80 80-6498-95 80-6499-14 80-6499-33 80-6498-76 80-6505-98 80-6194-19 80-6506-17 The following table provides you information about accessory parts and code number. Accessory part Immobiline DryStrip Reswelling Tray, 7 18 cm Immobiline DryStrip Reswelling Tray, 7 24 cm Immobiline DryStrip Cover Fluid IPGphor Strip Holder Cleaning Solution Quantity 1 1 1 liter 1 liter Code number 80-6371-84 80-6465-32 17-1335-01 29-0119-27 Ettan IPGphor User Manual 11-0034-60 AD 31

For local office contact information, visit www.gelifesciences.com GE Healthcare Bio-Sciences AB Björkgatan 30 751 84 Uppsala Sweden www.gelifesciences.com/2de GE, imagination at work and GE monogram are trademarks of General Electric Company. Ettan, Immobiline, IPGphor, and PlusOne are trademarks of GE Healthcare companies. 2012 General Electric Company All rights reserved. First published 2005 All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information. GE Healthcare Europe GmbH Munzinger Strasse 5, D-79111 Freiburg, Germany GE Healthcare UK Limited Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK GE Healthcare Bio-Sciences Corp. 800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327, USA GE Healthcare Japan Corporation Sanken Bldg. 3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073, Japan imagination at work 11-0034-60 AD 04/2012