Determination of Carbonyl Compounds by Reversed-Phase High-Performance Liquid Chromatography



Similar documents
METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

METHOD 8315A DETERMINATION OF CARBONYL COMPOUNDS BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

METHOD 554 DETERMINATION OF CARBONYL COMPOUNDS IN DRINKING WATER BY DINITROPHENYLHYDRAZINE DERIVATIZATION AND HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Determination of Calcium and Magnesium in Brine

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Tamsulosin Hydrochloride Capsules

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

ISOLATION OF CAFFEINE FROM TEA

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.

Separation by Solvent Extraction

Extraction: Separation of Acidic Substances

Experiment 8 Preparation of Cyclohexanone by Hypochlorite Oxidation

Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance

105 Adopted:

METHOD 550. July J.W. Hodgeson. W.J. Bashe (Technology Applications Inc.) T.V. Baker (Technology Applications Inc.)

experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.

INSTRUCTIONS Edition AC

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

A Complete Solution for Method Linearity in HPLC and UHPLC

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

SUCRALOSE. White to off-white, practically odourless crystalline powder

NWTPH-HCID. Hydrocarbon Identification Method for Soil and Water

Analytical Test Report

α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase

MOISTURE (Karl Fischer, Buffered)

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

POLYVINYL ALCOHOL. SYNONYMS Vinyl alcohol polymer, PVOH, INS No DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS

Experiment 5 Preparation of Cyclohexene

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

PURIFICATION TECHNIQUES

METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Paper Chromatography: Separation and Identification of Five Metal Cations

DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included.

Innovative vs Traditional

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Purification of reaction mixtures using flash chromatography.

Experiment 8 Synthesis of Aspirin

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

An In-Gel Digestion Protocol

Isolation of Caffeine from Tea

SODIUM CARBOXYMETHYL CELLULOSE

COMMON LABORATORY APPARATUS

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

NITRIC OXIDE and NITROGEN DIOXIDE 6014

Dissolved and precipitated oxalate

Recovery of Elemental Copper from Copper (II) Nitrate

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

CHEM 2423 Recrystallization of Benzoic Acid EXPERIMENT 4 - Purification - Recrystallization of Benzoic acid

Experiment #7: Esterification

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Apparatus error for each piece of equipment = 100 x margin of error quantity measured

United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science

Hot water extractable acidity or alkalinity of paper (Reaffirmation of T 428 om-08) (No changes from Draft 1)

Color, True and Apparent

Protocol: HPLC (amino acids)

OXIDATION-REDUCTION TITRATIONS-Permanganometry

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Organic Lab 1 Make-up Experiment. Extraction of Caffeine from Beverages. Introduction

Determination of Caffeine in Beverages HPLC-1

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Color, True and Apparent

This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA

ORGANIC SAMPLE PREPARATION

PECTINS. SYNONYMS INS No. 440 DEFINITION DESCRIPTION. FUNCTIONAL USES Gelling agent, thickener, stabilizer, emulsifier CHARACTERISTICS

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

[ Care and Use Manual ]

Mixtures and Pure Substances

Experiment 12- Classification of Matter Experiment

SYNTHESIS AND ANALYSIS OF A COORDINATION COMPOUND OF COPPER

Experiment 7: Titration of an Antacid

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Aliphatic/Aromatic Separation of Extractable Petroleum Hydrocarbons in Solids or Water by Silica gel Fractionation

CHEMICAL REACTIONS OF COPPER AND PERCENT YIELD KEY

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

HS 1003 Part 2 HS 1003 Heavy Metals Test

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

Physical and Chemical Properties and Changes

Project SMT-CT

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

Total Cyanide by Segmented Flow Injection Analysis, On-Line UV Digestion, and Amperometric Detection

Determination of calcium by Standardized EDTA Solution

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup

PART I: PREPARATION OF SOLUTIONS AND STANDARDIZATION OF A BASE

Making Biodiesel from Virgin Vegetable Oil: Teacher Manual

Solids, Volatile Dissolved and Fixed Dissolved

Estimation of Alcohol Content in Wine by Dichromate Oxidation followed by Redox Titration

Austin Peay State University Department of Chemistry CHEM 1021 TESTING FOR ORGANIC FUNCTIONAL GROUPS

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

Transcription:

Application Note 97 Determination of Carbonyl Compounds by Reversed-Phase High-Performance Liquid Chromatography INTRODUCTION Large quantities of carbonyl compounds are used worldwide as raw materials and intermediates in the chemical and plastics manufacturing industries. These compounds are eye, skin, and respiratory system irritants and many have also been identified as animal teratogens, mutagens, and carcinogens. Because of their widespread, high-volume industrial usage they are present in workplace air spaces and industrial waste materials. They are also present in dump site leachates, which can contaminate ground water that may eventually be used for drinking water. This application note describes a method wherein carbonyl compounds in aqueous samples or extracts are first derivatized with dinitrophenylhydrazine (DNPH). After derivatization, the carbonyl compounds are extracted and injected onto a C18, reversed-phase highperformance liquid chromatography (RP-HPLC) column and separated using gradient elution. The separated compounds are then quantified by UV detection at 360 nm. These analytical procedures are patterned after and consistent with U.S. EPA Method 554, which specifically addresses drinking water samples. It is also consistent with U.S. EPA Method 8315, which addresses aqueous and solid waste, soil, stack gas, and indoor air samples. EQUIPMENT Dionex DX-500 HPLC system consisting of: GP40 Gradient Pump (SST) LC5 High-Pressure Injection Valve (SST) AD20 Absorbance Detector Eluent Organizer PeakNet Chromatography Workstation REAGENTS AND STANDARDS Reagents Reagent grade chemicals should be used for all tests. Water should be free from substances that interfere with the test. Such substances include the analytes of interest that should be below the limits of detection as well as residual chlorine and phenolic compounds. Because formaldehyde is ubiquitous, care must be taken to prevent it from contaminating the water. Acetonitrile (HPLC grade or equivalent) Acetic acid Sodium acetate Citric acid Sodium citrate (trisodium dihydrate) 2,4-dinitrophenylhydrazine (DNPH), recrystallized with acetonitrile Hydrochloric acid Methylene chloride (HPLC grade or equivalent) Perchloric acid Sodium chloride Sodium hydroxide (pellets) Sodium sulfate (granular, anhydrous) Sodium sulfite (anhydrous) Standards Analytical grade aldehydes and ketones should be purchased for the preparation of the DNPH derivative standards. Alternatively, prederivatized DNPH stock solutions can be obtained as 100 µg/ml concentrates from AccuStandard (New Haven, Connecticut, USA).

CONDITIONS Option 1 Column: Zorbax ODS column, 4.6 mm x 250 mm i.d., 5-µm particle size Eluent 1: Acetonitrile Eluent 2: Water Gradient: Time A B (min) (%) (%) 0 60 40 20 60 40 35 100 0 50 100 0 Flow Rate: 1.2 ml/min Inj. Volume: 20 µl Detection: UV, 360 nm Option 2 Column: Eluent 1: Eluent 2: Two Zorbax ODS columns, 4.6 mm x 250 mm i.d., 5-µm particle size Acetonitrile Water Gradient: Time A B (min) (%) (%) 0 60 40 30 75 25 50 100 0 55 100 0 56 60 40 Flow Rate: 1.0 ml/min Inj. Volume: 25 µl Detection: UV, 360 nm PREPARATION OF SOLUTIONS Stock Solutions Acetate Buffer (for formaldehyde analysis only) Add 40 ml of 5 M acetic acid to 60 ml of 5 M sodium acetate. Mix thoroughly and adjust to ph 5 with NaOH (or HCl) as needed. 1 M Citric Acid Place 192.1 g of citric acid in a 1-L volumetric flask. Add 500 ml of water and mix until dissolved. Mix and dilute to volume with water. 1 M Sodium Citrate Place 258.1 g of sodium citrate in a 1-L volumetric flask. Add 500 ml of water and mix until dissolved. Mix and dilute to volume with water. 1 M Citrate Buffer, ph 3 Add 80 ml of 1 M citric acid to 20 ml of 1 M sodium citrate. Mix thoroughly and adjust the ph with NaOH (or HCl) as needed. 2,4-Dinitrophenylhydrazine, 3.00 mg/ml Dissolve 428.7 mg of DNPH (solids containing 30% water) into 100 ml of acetonitrile. (See Appendix A for DNPH purification procedure.) Sodium Chloride, Saturated Solution Dissolve an excess of sodium chloride (greater than 38 g/100 ml) in water. 5.0 N Sodium Hydroxide Place 100 ml water in a 200-mL volumetric flask. Add 40.0 g of sodium hydroxide, and mix and cool the solution to 25 C. Bring to volume with water. 0.1 M Sodium Sulfite Solution (freshly prepared) Place 150 ml of water in a 250-mL volumetric flask. Add 3.15 g of anhydrous sodium sulfite, dissolve, and dilute to volume with water. Stock Standard Solutions Prederivatized DNPH stock solutions can be obtained as 100 µg/ml concentrates, in acetonitrile, from AccuStandard (New Haven, Connecticut, USA). Alternatively, underivatized stock standard solutions may be made from the pure standards and derivatized, as needed, according to the directions listed below. 1 2 Determination of Carbonyl Compounds by RP-HPLC

Preparation of Stock Standard Solutions (1000 mg/l) Add 100 mg of the pure material other than formaldehyde to 90 ml of water and dilute to 100 ml to give a final volume concentration of 1000 mg/l. Store all standard solutions in the dark at 4 C in a glass vial with a Teflon -lined cap with minimum headspace. They should be stable for about six weeks. Preparation of Stock Formaldehyde Solution (1000 mg/ml) Prepare by diluting an appropriate amount of the certified formaldehyde solution. If a certified formaldehyde solution is not available, dilute approximately 2.67 ml of commercially available 37% formaldehyde solution to 1 L with water. Standardize by using the procedure discussed in Appendix B. The solution should be stored at room temperature in the dark and it should be standardized every 6 months. The solution appears to be stable indefinitely. Preparation of Calibration Standards Calibration Standards Prepare calibration standards at a minimum of five concentrations for each analyte of interest in water from the stock standard solution. The lowest concentration of each standard should meet or slightly exceed the MDLs listed in Table 1. The other concentrations of the calibration curve should correspond to the expected range of concentrations found in the samples of interest. Table 1 Method Detection Limits Analyte MDL a MDL a Option 1 Option 2 Acetaldehyde 12 (µg/l) 0.6 (µg/l) Acetone N/A 2 Acrolein N/A 0.4 Benzaldehyde N/A 2 Butanal 9 9 Crotonaldehyde 1 2 Cyclohexanone 0.5 N/A Decanal 6 N/A 2,5-Dimethylbenzaldehyde N/A 10 Formaldehyde 0.8 0.4 Heptanal 3 N/A Hexanal 3 3 Isovaleraldehyde N/A 5 Nonanal 9 N/A Octanal 3 N/A Pentanal 2 5 Propanal 8 2 m-tolualdehyde N/A 5 o-tolualdehyde N/A 4 p-tolualdehyde N/A 5 Formaldehyde Solution (10.0 mg/l) Dilute 1.00 ml of the formaldehyde stock solution to 100 ml with water. Prepare fresh daily. SAMPLE PREPARATION AND PRESERVATION Aqueous Samples Aqueous samples should be stored at 4 C and derivatized within three days of being taken. Derivatized compounds are extracted using methylene chloride (liquid/liquid) extraction or C18 cartridge (liquid/solid) adsorption, followed by ethanol extraction. The methylene chloride extracts are exchanged into either acetonitrile or ethanol and analyzed by HPLC. The ethanol extracts are analyzed directly by HPLC. a M.D.L. = (s.d.) x (t s ) 99.5%, where (t s ) is for a 99.5% single-sided student s test distribution. N/A = Not applicable Solid Samples Solid samples should be stored at 4 C and extracted within three days. Samples are extracted with an acetate buffer solution and this buffered extract is derivatized, extracted and analyzed by HPLC as described for aqueous samples. Stack Gas Samples Stack gas samples are drawn through a sampling train where the carbonyl compounds are derivatized in situ. These samples should be stored at 4 C and analyzed within 30 days of being taken. The derivatized samples are extracted with methylene chloride and the extracted compounds exchanged into acetonitrile, followed by HPLC analysis. Application Note 97 3

Indoor Air Samples Indoor air samples are drawn through a sampling train where the carbonyl compounds are derivatized in situ. These samples should be stored at 4 C and nalyzed within 30 days of being taken. The derivatized compounds are extracted with acetonitrile and analyzed by HPLC. DERIVATIZATION Measure an aliquot of the sample which is appropriate to the anticipated concentration range (nominally 100 ml). Quantitatively transfer the sample aliquot to a 250- ml Florence flask. In cases where the selected sample or extract volume is less than 100 ml, the total volume of the aqueous layer should be adjusted to 100 ml with water. Record the original sample volume prior to dilution. For analytes other than formaldehyde, add 4 ml of citrate buffer and adjust the ph to 3.0 + 0.1 with 6 N HCl or 6 N NaOH. If formaldehyde is the only analyte of interest, add 4 ml of acetate buffer and adjust the ph to 5.0 + 0.1 with 6 N HCl or 6 N NaOH. In either case, following ph adjustment add 6 ml of DNPH reagent, seal the container, and place in a heated orbital shaker at 40 C for 1 hour. Adjust the agitation to produce a gentle swirling of the reaction solution. EXTRACTION PROCEDURE Liquid-Solid Extraction This procedure requires the use of a vacuum manifold, which is capable of simultaneous extraction of up to 12 samples. In addition, solid sorbent cartridges packed with 2 g of C18 resin will be required. Attach the assembled vacuum manifold to a water aspirator or vacuum pump. Attach the 2 g sorbent cartridges to the vacuum manifold and condition them by passing 10 ml of 40 mm citrate buffer through each one. Remove the Florence flask from the shaker at the end of the 1-hour period and add to it 10 ml of saturated sodium chloride solution. Pour the reaction solution into the sorbent cartridges and apply a vacuum so that the solution is drawn through at a rate of 3 to 5 ml/min. Continue applying the vacuum for about 1 minute after the liquid sample has passed through the cartridge. While maintaining the vacuum, elute each cartridge into a 10-mL volumetric flask, with approximately 9 ml of acetonitrile. Dilute the solution to volume with acetonitrile, mix thoroughly, and place in a tightly sealed vial until analyzed. Due to the excess of DNPH used in the derivatization process, the sorbent cartridges will retain the yellow color after elution. This discoloration does not indicate non-recovery of the analyte derivatives. 1, 2 Liquid-Liquid Extraction After the 1-hour derivatization period, serially extract the solution with three 20-mL portions of methylene chloride using a 250-mL separatory funnel. If an emulsion forms upon extraction, remove the entire emulsion and centrifuge at 2000 rpm for 10 minutes. Separate the layers and proceed with the next extraction. Combine the methylene chloride layers in a 125-mL Erlenmeyer flask containing 5.0 g of anhydrous sodium sulfate. Swirl the contents to complete the drying process. Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10-mL concentrator tube to a 500-mL evaporator flask. Pour the methylene chloride extract into the evaporator flask, taking care to minimize transfer of the sodium sulfate granules. Wash the Erlenmeyer flask with 30 ml of methylene chloride and add the wash to the evaporator flask. Add one or two boiling chips to the flask and attach a three-ball Snyder column. Pre-wet the Snyder column by adding about 1 ml of methylene chloride to the top. Place the K-D apparatus in a hot water bath (80 90 C) so that the concentrator tube is partially immersed in the hot water and the entire lower rounded surface of the flask is bathed with hot vapor. Adjust the vertical position of the apparatus and the water temperature as required to complete the concentration in 10 15 minutes. At the proper rate of distillation the balls of the column will actively chatter, but the chambers will not flood with the condensed solvent. When the apparent volume of the liquid reaches 5 ml, remove the K-D apparatus and allow it to drain and cool for at least 10 minutes. Prior to liquid chromatographic analysis, the extract solvent must be exchanged to acetonitrile. Remove the three-ball Snyder column and evaporator flask. Add 5 ml of acetonitrile and a new boiling chip and attach a micro-snyder column to the concentrator tube. Concentrate the extract using 1 ml of acetonitrile to pre-wet the Snyder column. Place the K-D apparatus in the water bath so that the concentrator tube is partially immersed in the hot water and proceed as described above. When 4 Determination of Carbonyl Compounds by RP-HPLC

the apparent volume of liquid reaches less than 5 ml, remove the K-D apparatus and allow it to drain and cool for at least 10 minutes. Remove the Snyder column and rinse the flask and its lower joint with 1 2 ml of acetonitrile and add to the concentrator tube. Quantitatively transfer the sample to a 10-mL volumetric flask using a 5-mL syringe with an attached Acrodisk 0.45-mm filter cassette. Adjust the extract volume to 10 ml, stopper the flask, and store refrigerated at 4 C if further processing will not be performed immediately. If the extract will be stored for longer than two days, it should be transferred to a vial with a Teflon-lined screw cap or crimp top. Proceed with HPLC chromatographic analysis if further cleanup is not required. 1 PREPARATION FOR HPLC ANALYSIS Filter the mobile phase through a 0.22-mm polyester membrane filter in an all glass and Teflon suction apparatus. Degas each filtered solution by purging with helium for 10 to 15 minutes (100 ml/ min) or by heating to 60 C for 5 to 10 minutes in an Erlenmeyer flask covered with a watch glass. A constant backpressure restrictor 500 kpa (75 psi) or 15 30 cm of 0.25-mm i.d. Teflon tubing should be placed after the detector to eliminate further mobile phase outgassing. Allow the system to pump the initial solvent mixture ratio (60% CH 3 CN/40% H 2 O) for 20 to 30 minutes at a flow rate of 1 ml/min and monitor the detector output to establish a stable baseline. DISCUSSION AND RESULTS This application note describes two options for this HPLC analysis of carbonyl compounds. Option 1 uses a single C18 column and separates a limited set of compounds. This option has been found suitable for the analysis of drinking water, waste water, soil and stack gas samples. Option 2 separates a somewhat different and broader range of compounds. This option has been found suitable for the analysis of indoor air samples. Table 2 lists the compounds determined by each option. See Figures 1 and 2 on page 6 for example chromatograms. Method detection limits for both options were obtained using fortified, organic-free reagent water. MDLs are based upon analyses of eight replicate spiked blank injections. Method performance is illustrated in U.S. EPA Method 8315. Table 2 Carbonyl Compounds that Can Be Determined by This Method Compound Name Option Acetaldehyde 1, 2 Acetone 2 Acrolein 2 Benzaldehyde 2 Butanal 1, 2 Crotonaldehyde 1, 2 Cyclohexanone 1 Decanal 1 2,5-Dimethylbenzaldehyde 2 Formaldehyde 1 Heptanal 1 Hexanal 1, 2 Isovaleraldehyde 2 Nonanal 1 Octanal 1 Pentanal 1, 2 Propanal 1, 2 m-tolualdehyde 2 o-tolualdehyde 2 PRECAUTIONS Hazards The toxicity or carcinogenicity of each reagent in this method has not been precisely defined; however, each compound should be treated as a potential health hazard. For example, glacial acetic acid can produce skin burns and severely injure the eyes. Formaldehyde is a severe skin, eye, and mucous membrane irritant and can cause acute hypersensitivity. Formaldehyde may also be a potential carcinogen. DNPH is flammable. The laboratory is responsible for maintaining a current awareness of OSHA regulations regarding the safe handling of the chemicals specified in this method. Application Note 97 5

0.007 AU 1 2 3 4 5 Peaks: 1. Formaldehyde 250 µg/l 2. Acetaldehyde 3. Propanal 4. Crotonaldehyde 5. Butanal 6. Cyclohexanone 7. Pentanal 8. Hexanal 9. Heptanal 10. Octanal 11. Nonanal 12. Decanal 8 9 12 10 11 0.004 Peaks: 1. Formaldehyde 100 µg/l 2. Acetaldehye 3. Acrolein 4. Acetone 5. Propanal 6. Crotonaldehyde 7. Butanal 8. Benzaldehyde 9. Isovaleraldehyde 10. Pentanal 11. o-tolualdehyde 12. m-tolualdehyde 13. p-tolualdehyde 14. Hexanal 15. 2,4-Dimethylbenzaldehyde 6 7 AU 1 2 3 4 5 6 7 8 1113 14 9 10 12 15 0.000 0 5 10 15 20 25 30 35 40 Minutes 0.000 Figure 1. Separation of carbonyl compounds using Option 1. Interferences Method interferences may be caused by contaminants in solvents, reagents, glassware, and other sample processing hardware that lead to discrete artifacts and/ or elevated baselines in the chromatogram. All of these materials must be routinely demonstrated to be free from interferences by running blanks. Formaldehyde contamination of the DNPH is a frequently encountered problem because formaldehyde is prevalent in the environment; thus, DNPH should be recrystallized to prevent this type of contamination using the procedure Appendix A. Matrix interference may be caused by contaminants that are co-extracted from the sample. If this is a problem, additional sample cleanup may be necessary. Acetaldehyde is generated during derivatization if ethanol is present in the sample. This background will impair the measurement of acetaldehyde at levels below 500 µg/l. Phenols, residual chlorines, sulfites, and other reactive substances combine with some free carbonyls to form other compounds. Such bound carbonyls are not meant to be measured by this test method. 0 5 10 15 20 25 30 35 40 Minutes Figure 2. Separation of carbonyl compounds using Option 2. REFERENCES 1. U.S. Environmental Protection Agency. U.S. EPA Method 8315, Determination of Carbonyl Compounds by High-Performance Liquid Chromatography, November 1992. 2. U.S. Environmental Protection Agency. U.S. EPA Method 554 Determination of Carbonyl Compounds in Drinking Water by Dinitrophenylhydrazine Derivatization and High-Performance Liquid Chromatography (HPLC), August 1992. 6 Determination of Carbonyl Compounds by RP-HPLC

APPENDIX A: 2,4-DINITROPHENYLHYDRAZINE PURIFICATION Formaldehyde contamination of DNPH reagent is a frequently encountered problem due to its widespread occurrence in the environment. The DNPH reagent must be purified using the following procedure. Prepare a saturated solution of DNPH by boiling excess DNPH in 200 ml of acetonitrile for approximately 1 hour, transfer the supernatant to a covered beaker on a hot plate and allow gradual cooling to 40 60 C. Maintain this temperature range until 95% of the solvent has evaporated, leaving crystals. Decant the solution to waste and rinse the remaining crystals twice with acetonitrile. Repeat the recrystallization process. Take an aliquot of the second rinse, dilute 10 times with acetonitrile, acidify with 1 ml of 3.8 M perchloric acid per 100 ml of DNPH solution, and analyze with HPLC conditions for Option 2. The impurity level for formaldehyde should be less than 0.025 ng/µl. If the impurity level is not satisfactory, pipette off the solution to waste, repeat the recrystallization procedure, and rinse with two 25-mL portions of acetonitrile. Prepare the second rinse as before and analyze. When the impurity level is satisfactory place the crystals in an all-glass reagent bottle, add another 25 ml of acetonitrile, cap, and shake the bottle. Always use clean pipettes when removing the saturated DNPH stock solution to reduce contamination of the solution. 1 Note: This procedure should be performed under a ventilated hood. APPENDIX B: STANDARDIZATION OF FORMALDEHYDE STOCK Calibrate a ph meter with the standard 7.0 and 10.0 ph buffers. Transfer a 25.0-mL aliquot of the 0.1 M sodium sulfite solution to an Erlenmeyer flask. Measure and record the ph while stirring. Add a 25.0-mL aliquot of the formaldehyde stock solution, mix, and allow to stand for 5 minutes. While stirring, back titrate this mixture to the original ph using 0.1 N HCl. Perform three duplications and calculate the mean concentration. Duplicates should agree within 0.3%. Calculate the concentration of the formaldehyde stock solution as follows: C F = V A x C A x MW F V F where: C F = Concentration of formaldehyde (mg/l) C A =Concentration of acid used (N) V F = Volume of formaldehyde (ml) V A = Volume of acid used (ml) MW F = Moleccular weight of formaldehyde (30.03) Application Note 97 7

PeakNet is a registered trademark of Dionex Corporation. Zorbax and Teflon are registered trademarks of E.I du Pont de Nemours & Co. Acrodisk is a registered trademark of Gelman Sciences, Inc. Printed on recycled and recyclable paper. Dionex Corporation Dionex Corporation Dionex U.S. Regional Offices Dionex International Subsidiaries 1228 Titan Way Salt Lake City Technical Center Sunnyvale, CA (408) 737-8522 Austria (01) 616 51 25 Belgium (015) 203800 Canada (905) 844-9650 China (852) 2428 3282 Denmark (45) 36 36 90 90 P.O. Box 3603 1515 West 2200 South, Suite A Westmont, IL (630) 789-3660 France 01 39 30 01 10 Germany 06126-991-0 Italy (06) 66 51 50 52 Japan (06) 6885-1213 The Netherlands (0161) 43 43 03 Sunnyvale, CA Salt Lake City, UT Houston, TX (281) 847-5652 Switzerland (062) 205 99 66 United Kingdom (01276) 691722 8 Determination of Carbonyl Compounds (408) 737-0700 by RP-HPLC (801) 972-9292 Marlton, NJ (856) 596-0600 94088-3603 84119-1484 Atlanta, GA (770) 432-8100 * Designed, developed, and manufactured under an NSAI registered ISO 9001 Quality System. LPN 0578 500 3/01 2001 Dionex Corporation