ph and Acidity in Wine and Fruit Juice

Similar documents
ph and Acidity in Tomato Sauce

ph Alkalinity of Water

DETERMINATION OF PHOSPHORIC ACID CONTENT IN SOFT DRINKS

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets

ph units constitute a scale which allows scientists to determine the acid or base content of a substance or solution. The ph 0

Experiment 4 (Future - Lab needs an unknown)

Acid Base Titrations

GA/7 Potentiometric Titration

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

Potentiometric determination. Application Bulletin 125/3 e. Branch General analytical laboratories; water analysis; beverages

18 Conductometric Titration

Experiment 6 Titration II Acid Dissociation Constant

TITRATION CURVES, INDICATORS, AND ACID DISSOCIATION CONSTANTS

1. To perform a potentiometric titration of a sample of Liquid Drano with hydrochloric acid.

Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Electrical Conductivity of Aqueous Solutions

Determining the Identity of an Unknown Weak Acid

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration

TITRATION OF AN ACID; USING A ph METER. The ph meter is an instrument that measures the ph of a solution and affords a

Acid-Base Titrations Using ph Measurements

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

POTENTIOMETRIC TITRATION OF A WEAK ACID

ph: Measurement and Uses

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

EXPERIMENT 12 A SOLUBILITY PRODUCT CONSTANT

EXPERIMENT INTRODUCTION TO INDICATORS AND ACID-BASE TITRATIONS

Acid Dissociation Constants and the Titration of a Weak Acid

Additional Lecture: TITRATION BASICS

MOISTURE (Karl Fischer, Buffered)

9. Analysis of an Acid-Base Titration Curve: The Gran Plot

Experiment 17: Potentiometric Titration

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

USEPA Electrode Method Method 8156

Determination of Citric Acid in Powdered Drink Mixes

Analysis of Vitamin C Using Iodine. Introduction

Lab 25. Acid-Base Titration and Neutralization Reactions: What Is the Concentration of Acetic Acid in Each Sample of Vinegar?

Acid Base Titration: ph Titration Curve

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration

The introduction of your report should be written on the on the topic of the role of indicators on acid base titrations.

Neutralizing an Acid and a Base

Acetic Acid Content of Vinegar: An Acid-Base Titration E10-1

CHM 130LL: ph, Buffers, and Indicators

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

Juice Titration. Background. Acid/Base Titration

Determination of the total acid number in petroleum products

The Determination of Acid Content in Vinegar

ph Measurement and Control

Ascorbic Acid Titration of Vitamin C Tablets This lab will be completed individually! Make sure you come prepared!

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)

Drop Counter Sensor Product Number: ENDRP-AD100

Properties of Acids and Bases

Determination of calcium by Standardized EDTA Solution

Customer application report

SenTix 950 SenTix 980

ABC of Titration. School experiments. Titration Theory. Natural science laws experience ABC of Titration live

Standard Methods for the Examination of Water and Wastewater

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

6 H2O + 6 CO 2 (g) + energy

(1) Hydrochloric acid reacts with sodium hypochlorite to form hypochlorous acid: NaOCl(aq) + HCl(aq) HOCl(aq) + NaCl(aq) hypochlorous acid

Quick Guide. DL50/DL53/DL55/DL58 Titrators

Experiment 6 Coffee-cup Calorimetry

Experiment 9: Acids and Bases Adapted from: Chemistry, Experimental Foundations, 4th Ed. Laboratory Manual, by Merrill, Parry & Bassow.

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

Analyzing the Acid in Vinegar

Winemaker s Guide to Chemistry Analysis

Titration. Lecture # 8. Titrations in Analytical Chemistry. Other Forms of Titration. End Point vs. Equivalence Point. Minimizing Titration Error

Precipitation Titration: Determination of Chloride by the Mohr Method by Dr. Deniz Korkmaz

LESSON ASSIGNMENT. After completing this lesson, you should be able to: 7-1. Solve basic titration problems.

Experiment 7: Titration of an Antacid

Color, True and Apparent

Milwaukee USER MANUAL

Formulas, Equations and Moles

Volumetric Analysis. Lecture 5 Experiment 9 in Beran page 109 Prelab = Page 115

Color, True and Apparent

Phenolphthalein-NaOH Kinetics

TITRATION OF VITAMIN C

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

Simulation of the determination of lead azide content in waste water from explosives manufacture

Chem 321 Lecture 13 - Acid-Base Titrations 10/10/13

The Determination of an Equilibrium Constant

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Preparation of frequently used solutions

Syllabus OC18 Use litmus or a universal indicator to test a variety of solutions, and classify these as acidic, basic or neutral

Acid-Base Titrations. Setup for a Typical Titration. Titration 1

Solutions and Dilutions

Acid Base Titrations in Aqueous Solvents

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 10: TITRATION AND STANDARDIZATION

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

A Potentiometric Analysis of Fluoride Ion in Toothpaste

EXPERIMENT 10 Chemistry 110. Solutions Part 2 ACIDS, BASES, AND ELECTROLYTES

Experiment 3 Limiting Reactants

ph Measurements of Common Substances

Tutorial 4 SOLUTION STOICHIOMETRY. Solution stoichiometry calculations involve chemical reactions taking place in solution.

Transcription:

ph and Acidity in Wine and Fruit Juice DOC316.52.93089 Endpoint potentiometric titration Scope and application: Wine and fruit juice 1. Introduction This working procedure refers to the titration of total acidity in wines and fruit juices. The main acid in wine is tartaric acid, in juices it is usually citric or malic acid. These beverages also contain several other acids in different proportions: total acidity corresponds to the sum of titratable acidities. 2. Principle The protocol rests on a weak acid titration by a strong base. Despite the presence of several acids in the sample, the titration curves show only one ph jump. The equivalent volume is determined from this ph jump in endpoint mode. By default, the endpoint ph value is set to 8.2, according to the AOAC (Association of Analytical Communities) recommendation. It is also possible to set the endpoint to 7.0, as recommended by the OIV (International Organization of Vine and Wine). Total acidity in eq/l can be directly calculated from the equivalent volume. Total acidity can also be expressed in g/l of a chosen acid, considering its molar weight and the stoichiometry of the acid-base reaction: 3. Electrode and reagents Acid Molar weight (g/mol) Exchanged electrons Tartaric 150.1 2 Malic 134.1 2 Citric 192.1 3 Sulfuric 98.1 2 Acetic 60.1 1 Electrodes: Titrant: Intellical combined ph electrode with integrated temperature sensor, PHC725 NaOH 0.25 eq/l. Use a commercial solution or dissolve 10.00 g of sodium hydroxide in 1 L of CO 2 -free boiled water. Standard for titrant calibration: Dihydrate oxalic acid, molar weight = 126.07 g/mol ph standards: Colored 4.01, 7.00, 10.01 (part numbers 2283449, 2283549, 2283649) Deionized water 4. Ranges and settings 4.1. Default parameters The working procedure is described using the following parameters: V sample = 30 ml Syringe volume = 10 ml 4.2. Working range For most samples, 1 syringe (10 ml) of titrant should be sufficient to reach the equivalent point. It provides the following range: Veq (ml) 1.5 9.5 Acidity in g/l of tartaric acid 0.9 5.9 Acidity in g/l of citric acid 0.8 5.0 Page 1 of 6

The syringe is allowed to refill once during titration, so the range is widened: Veq (ml) 1.5 19 Acidity in g/l of tartaric acid 0.9 11.8 Acidity in g/l of citric acid 0.8 10.1 4.3. Settings 4.3.1. ph Probe Recommended probe PHC725 Stirring Stirring speed 15 [%] Time 15 [s] ph measurement Stirring speed 0 [%] Measured parameter [ph] Stability criteria 0.05 [ph/min] Stability max time 120 [s] 4.3.2. Acidity Sample Wine or juice Amount 30 [ml] Amount min 0 [ml] Amount max 35 [ml] Titrant NaOH Titrant concentration 0.2500 [eq/l] Syringe Syringe 1 Probe Recommended probe PHC725 Leveling Active no Time 30 [s] EP titration Stirring speed 25 [%] Measured parameter [ph] Predose 0 [ml] Max volume stop point 20 [ml] Stop on last EQP True Delay 0 [s] Min increment size 0.15 [ml] Max increment size 1.5 [ml] EP Ordinates 8.2 [ph] Result 1 name Acidity (g/l tartaric acid) R1 resolution R1 min 0.5 [g/l] R1 max 12 [g/l] R1 QC min 0.5 [g/l] R1 QC max 12 [g/l] Result 2 name Acidity (g/l citric acid) R2 resolution R2 min 0.5 [g/l] R2 max 10 [g/l] R2 QC min 0.5 [g/l] Page 2 of 6

R2 QC max 10 [g/l] Result 3 name Acidity (g/l malic acid) R3 resolution R3 min 0.5 [g/l] R3 max 11 [g/l] R3 QC min 0.5 [g/l] R3 QC max 11 [g/l] Result 4 name Acidity (g/l acetic acid) R4 resolution R4 min 0.5 [g/l] R4 max 10 [g/l] R4 QC min 0.5 [g/l] R4 QC max 10 [g/l] Result 5 name Acidity (g/l sulfuric acid) R5 resolution R5 min 0.5 [g/l] R5 max 8 [g/l] R5 QC min 0.5 [g/l] R5 QC max 8 [g/l] 4.4. Modification of the settings The parameters are defined in order to have the best compromise between accuracy and titration time. For higher concentration with a high titrant volume, titration time can be reduced with an addition of titrant (predose) at the beginning of the titration. Enter the predose volume (in ml) and the stirring time after the addition in the application edit window. 5. Titration procedure 5.1. Leveling To use this method, an external pump is required. All elements (probes, tubes from the titrator and the tube from the external pump) have to be well installed on the probe holder. The beaker has to contain a level of sample higher than the position of the tube of the external pump. When the beaker is attached to the probe holder, this method allows the system to automatically remove the excess sample by a defined pump working time, and always keep the same sample volume before launching the analysis. In order to define this volume, autoleveling calibration sequence has to be previously executed (see section 8.3 Autoleveling calibration). When this option is active, the working time of the external pump must be set (default 30 s). The minimum working time must allow the pump to be removing air during the last few seconds of the external pump activation. Note: Do not forget to re-edit the sample amount with the expected value when deactivating the leveling method. 5.2. Titration Rinse the ph probe with deionized water. If leveling is disabled, use a pipette to collect precisely 30 ml of sample. Pour the sample into the 50 ml polypropylene beaker. Put in a magnetic stir bar, dip the probe and the delivery tip in the solution. Start the application. At the end of the titration, a first window displays the result. A second window displays the titration curve and the equivalent point coordinates. After the titration, there are two possibilities: Replicate the sample. This is used to study the repeatability by analyzing several samples successively. At the end of each titration, a window displays the average value, the standard deviation (SD) and the relative standard deviation (RSD in %). Analyze a new sample. Another titration can be started but no Standard Deviation and RSD value will be made. Page 3 of 6

6. Results 6.1. Result calculation The calculation used is: 6.2. Experimental results C titrant (eq L) V titrant (ml) Acidity (g L) = n e- chosen acid n e- titrant V sample (ml) M chosen acid (g/mol) = 0.25 (eq L ) V titrant (ml) n e- chosen acid 1 30 (ml) M chosen acid (g/mol) These results are indicative and have been obtained for a given sample for five successive determinations. On red wine: On orange juice: Measurement Parameter Unit Mean concentration 5.059 [g/l of tartaric acid] Standard deviation 0.025 [g/l of tartaric acid] Relative standard deviation 0.5 [%] Mean titration duration 98 [s] Measurement Parameter Unit Mean concentration 7.363 [g/l of citric acid] Standard deviation 0.023 [g/l of citric acid] Relative standard deviation 0.3 [%] Mean titration duration 132 [s] 6.3. Example of a titration curve This curve has been obtained during the analysis of one of the five wine samples. 7. Recommendations Always rinse the ph probe and the delivery tip between measurements. Refill the electrode regularly with KCl saturated solution to maintain the level of internal solution around 1 cm (0.4 inches) below the refill hole. Page 4 of 6

8. Appendices 8.1. Electrode calibration For the precision of the titration, it is recommended to calibrate the probe with ph standards at the same temperature as the samples. For calibration, the ph is compensated at 25 C. Pour a sufficient amount of the first buffer into a 50 ml beaker and add a magnetic stir bar. Place it on the titrator under the probe holder and dip the probe into the beaker. Start the electrode calibration sequence. Rinse the probe between ach buffers. Repeat this operation for each buffer (at least two buffers are recommended). At the end of the calibration, the results of the slope and offset are displayed and the user can accept or reject this result. Default settings for electrode calibration 8.2. Titrant calibration Setting Unit Stirring speed 25 [%] Stability criteria 0.05 [ph/min] Stability max time 300 [s] Slope limit min 97 [%] Slope limit max 102 [%] Offset limit min 44 [mv] Offset limit max 74 [mv] The sodium hydroxide solution can be calibrated. The exact concentration can be determined from an acidbase titration using oxalic acid. Weigh 75 mg of dihydrate oxalic acid in a 50 ml beaker and use a graduated cylinder to add 30 ml of deionized water. Put in a stir bar, dip the probe and the delivery tip into the solution and launch the titrant calibration sequence. When prompted, type in the exact weighed amount (three digits). At the end of the titrant calibration, titer (eq/l) is displayed and the user can reject, replicate, or save the result. Default settings for titrant calibration Titrant NaOH Titrant concentration 0.2500 [eq/l] Syringe Syringe 1 Standard Oxalic acid Amount 75 [mg] Amount min 60 [mg] Amount max 100 [mg] Molar weight 126.07 [g/mol] EP titration Stirring speed 25 [%] Measured parameter [ph] Predose 2 [ml] Max volume stop point 8 [ml] Stop on last EQP True Delay 0 [s] Min increment size 0.08 [ml] Max increment size 0.8 [ml] EP Ordinates 8.55 [ph] Result name Titer Result resolution 4 decimals Result min 0.2250 [eq/l] Result max 0.2750 [eq/l] Page 5 of 6

8.3. Autoleveling calibration The aim of this method is to calibrate the volume of sample by leveling. The result of this calibration will be used as sample volume for the following titrations. This option is ONLY available from the calibration menu if Method Leveling is set to Active (Yes). Refer to the documentation delivered with the external pump for a correct installation, paying particular attention to the suction tube from the pump. Prepare a 0.02 mol/l oxalic acid solution: dissolve 2.521 g of dihydrate oxalic acid in 1 L of deionized water. Pour a sufficient amount of the solution in a beaker allowing the external pump tube to be immerged in the liquid. In the calibration menu select Autoleveling calibration and then ph acidity wine/juice. The result in ml is compared to minimum and maximum amounts defined for the sample volume. The calculation used is: V sample = V titrant (ml) C titrant (eq/l) n e- oxalic acid C oxalic acid solution (mol/l) = V titrant(ml) 0.25 (eq/l) 2 0.02 (mol/l) Autoleveling calibration uses the same settings as for a titrant calibration (see section 8.2 Titrant calibration). Default settings for autoleveling calibration Sample Amount min 0 [ml] Amount max 35 [ml] Autoleveling calibration Solution name Oxalic Acid Concentration 0.02 [mol/l] Resolution Page 6 of 6