Universal Endo System Directions for Use. ProTaper



Similar documents
Removal of a fractured instrument with a new extractor: clinical cases

Mani Silk: A new and novel means of predictable canal shaping

Feel the Safety and Control

SuperEndo B&L-alpha II Cordless Heat Carrier Operation & Maintenance Instruction Manual

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference

Pet hair clipper. Model Diagrams within this manual may not be drawn proportionally.

Precision Endodontic System. Precision Obturation System. So sophisticated, it s simple

KNOW YOUR FOOD SLICER

IMPORTANT SAFETY RULES TO FOLLOW

Fixing Dental Implant Component Failure and Fracture: Retrieving the Fractured Implant Abutment Screw. Charles A. Mastrovich, DDS

EP ; WO2012/080751; GB

Waterpik Water Flosser

Achieving Independence

2-SLICE COOL TOUCH TOASTER

CanalPro TM Apex Locator

Step-by-Step Guide. ENHANCEMENTS Brisa UV Gel Enhancements Rebalancing. STEP 1: Cleanse

VACUUM CLEANER INSTRUCTION MANUAL

ARCO Electric Products Installation and Maintenance Manual Low Voltage Automatic Power Factor Correction Capacitor Systems 2013

Tassimo The Art of the Perfect Cup

Instruction Manual. Image of SP-3015 & SP Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

Step-by-step instructions:

Instruction manual for Firstline FCS12000CH

LUCCI AIRFUSION QUEST II CEILING FAN

Construction of Quality Hot Mix Asphalt Pavements

OPERATOR S MANUAL 18 VOLT, 1 HOUR CHARGER

Lunette 2 Series. Curved Fixed Frame Projection Screen. User s Guide

WARNING To reduce the risk of. READ ALL INSTRUCTIONS BEFORE USING. KEEP AWAY FROM WATER DANGER Any appliance is

Content Ⅰ Safety caution Ⅱ Introduction and technical parameter Ⅲ Standard configuration Ⅳ Structure and theory Ⅴ Assemble Ⅵ Function and settings

100 Series Maxi Internal Filters. Instruction Manual

We make a point of ensuring safety.

VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised

OPERATION MANUAL. Total Heat Exchanger HRV (Heat Reclaim Ventilation) (Ceiling mounted duct type)

National- Spencer Inc.

CIRCULAR SAW SAFETY Reviewed 9/24/2007

How to Prepare Formula for Bottle-Feeding at Home

digital fever alert ear thermometer suitable from birth instruction manual

Ultrasonic support removal tank

#12 Professional Meat Grinders

Electric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL

MBSAW. Meat Cutting Band Saw With Meat Grinder Assembly & Operating Instructions

HAIR CLIPPER. Model No. HC5150 HC

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102

Brisa UV Gel Enhancements Sculpted on a Tip

Congratulations on your purchase of the Great Planes SLOT MACHINE, the first truly easy way to cut hinge slots in your model airplanes.

Electric Egg Boiler. Instruction Manual. Model Number: AEB-917

RICE COOKER USER MANUAL. Model: TRC-6STW

PART 1 - INTRODUCTION...

19 LED Tube Controller ORDERCODE 41003

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral

16/32 Channel 1U Rack Mount CCTV Power Supply

Endo Easy Efficient Sonic Powered Endo Irrigation

Espresso/Cappuccino Maker MODELS MEXKITB, MEX1B

Selection, Use, and Care of Brushes

MICA HEATER INSTRUCTION MANUAL Model No: UHM V 50Hz 2200W

Quick Start Guide See Inside for Use and Safety Information

BUT PRECAUTIONS MUST BE TAKEN OR SERIOUS BURNS CAN RESULT.

75th anniversary blender

CURL. Have a Question? Call Us First!

EEG Cap Instructions

Contents Contents Contents Overview Charging your headset Wearing the headset Powering on the headset Connecting your headset for dictation

ENGLISH INSTRUCTION & INSTALLATION MANUAL DUCTLESS MINI SPLIT AIR CONDITIONING SYSTEMS

Recommended with NESCAFÉ GOLD Barista Style. Experience the full range today.* NESCAFÉ GOLD BLEND

Odyssey of the Mind Technology Fair. Simple Electronics

English. Symbols used to mark instructions...3. Congratulations...5 Getting the best results...5. Warnings...6 Operating Procedure...

Ambiente Quick Reference Manual

Nucleus Freedom BTE Rechargeable Battery (A25) and Nucleus Freedom BTE Recharger

PLEASE READ ALL INSTRUCTIONS BEFORE USE AND SAVE A COPY FOR FUTURE REFERENCE!

EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board.

Wiper Motor Marinco 2.5. Installation Instructions

PICCs and Midline Catheters

INFRARED QUARTZ WALL HEATER

Manual Ranging MultiMeter

DASH KIT INSTALLATION

GeniJet Pipeless Motor Spa Installation

FLOOR WASHING ROBOT. 300 Series

RomiApex A-15. APEX LOCATOR User Manual

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR


rainbowsystem.com Rexair LLC Troy, Michigan 2011 Rexair LLC R14388A-0 Printed in U.S.A.

Bottom Loading Water Dispenser

Pediatric Multi-Venous IV Arm Kit. Directions for Use

Active and Passive Sash Replacement

2.4 GHz Dual Handset Cordless Telephone Answering System 2255 with Caller ID/Call Waiting

Instruction Manual. AWK-115S

PROFESSIONAL POPCORN MAKER

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL

BELLMAN CXE-25/27 INSTRUCTIONS

IMPORTANT SAFETY INSTRUCTIONS WARNING READ AND SAVE THESE OPERATING AND SAFETY INSTRUCTIONS BEFORE USING THIS HEATER.

CYGNAL Application Note

ENDO REMOVAL SYSTEM Set of endodontic micro instruments for removing broken canal instruments

ClearView 100. Instruction Manual

Fish Grader User Manual

WARNING: To prevent fire or electric shock, do not expose this appliance to rain or moisture.

Multi-Pitch Pitching Machine USER MANUAL

FOR THE FOLLOWING MODELS: EE-8075W EE-8075O EE-8075R EE-8075BK

StorTrends 3400 Hardware Guide for Onsite Support

Convection TCO600. For your safety and continued enjoyment of this product, always read the instruction book carefully before using.

12-Volt 10-Amp Regulated Power Supply

Female Multi-Venous IV Arm Kit. Directions for Use

HOW TO MAKE A MOTOR BRACKET

Transcription:

ProTaper Universal Endo System Directions for Use An instructional DVD is included with your system purchase. Please view this DVD, in addition to reading this booklet, to gain a complete understanding of the system and the technique. If you would like to speak with one of our staff clinicians, please call 1-800-662-1202, and we ll provide one-on-one assistance. Indications for Use Contraindications For the removal of dentin, shaping and obturating of the root canal. None known. Warnings This product contains dry natural rubber. Do not use near standing water. Dropping the oven and/or motor into water may cause electric shock. Do not immerse the oven and/or motor into water. Do not expose the oven to moisture. Turn off and unplug when not in use. Do not remove the cover of the oven. This oven is designed for use in a dental office. Do not use in any other environment. Precautions As with all new products, you must exercise caution until you become proficient in its use. While we have implemented safeguards against possible misuse, there are several important points to remember: A slow-speed handpiece is required for rotary file use. Operate the handpiece at 300 RPM (revolutions per minute). Straight-line access is a prerequisite for proper endodontic treatment. ProTaper Universal files are no exception. Always utilize minimal apical pressure. Don t force the files down the canal. Clean file flutes frequently during instrumentation (after the file is removed from the canal). Frequently irrigate and lubricate the canal throughout the procedure. To promote the mechanical preparation objectives, take any

ProTaper Universal rotary finishing instrument to length only one time and for no more than one second. Refer to your Endo DTC instruction manual for more operational and maintenance procedures. Adverse Reactions Step-By-Step Instructions Exercise caution in the apical area and around significant curvatures. Files are single use devices. They can become less efficient after multiple uses causing undue stress to the file. This can lead to file separation. When instrumenting the canal, select appropriately sized instruments, as choosing an overly large file can lead to dangerous over-enlargement of the coronal portion of narrow root forms. Additionally, too large a file taken to length increases the risk of separation. To reduce the risk of electric shock, do not remove oven cover; there are no user-serviceable parts inside. This oven is intended for use in heating gutta-percha coated obturators only. Do not insert anything other than an obturator into the oven. Patients with latex sensitivity may experience allergic reactions to gutta-percha, which contains dry natural rubber. Turning On Safe and Efficient Power ProTaper Universal rotary instruments are used in a controlled, slow-speed rotary handpiece, thus providing the reaming action that produces a continuous tapered preparation from orifice to apex. A torque control motor, such as the Aseptico Endo DTC (model #AEU-25) included in your system, should be used. Procedures for Using the Aseptico Endo DTC Motor* Preset torque settings are recommended for each file series. They are given without guarantee of risk or fracture. To avoid file separation, remember torque control motors are not a substitute for proper technique. ProTaper Universal System Technique Steps File Sterilization Files must be sterilized before use. ANSI/ADA Specification 28 recommends: Scrub the instruments with soap and warm water. Rinse thoroughly with distilled or deionized water. Allow to air dry. Place the instruments, unwrapped, in the autoclave tray. Use fresh distilled or deionized water. Steam Autoclave under pressure of 0.22 MNm -2 at 136 C (± 2 C ) for 20 minutes. The crown down technique is the technique of choice for rotary instruments. This technique initiates in the coronal area and progresses in an apical direction. The progression of instrumentation and examination continues until you take an appropriate finishing file to working length. The basic ProTaper Universal System sequence will be the same, regardless of the canal size. Create straight-line access to canal orifice. In the presence of a viscous chelator (such as ProLube root canal conditioner), passively scout the coronal 2/3 with 10 and 15 hand files. Gently work these instruments until a smooth, reproducible glide path is confirmed. In the presence of NaOCl, float the S1 into the canal and passively follow the glide path. Before light resistance is encountered, laterally brush and cut dentin on the outstroke to improve straight line access and apical progression. Always brush away from the furcation. When you remove the Endo DTC from the box, you must set the motor to accommodate the reduction contra-angle you will use. To do this, turn on the power, and the motor automatically will lead you through the process. On the screen, you will see 1/8 handpiece TUL. Enter 2 for yes or 1 to go to the next handpiece selection. Simply select the handpiece you will use. You will notice the ProTaper Universal system technique is set as a preset. Press the appropriate preset button repeatedly to cycle through the ProTaper Universal file series. Continue shaping with the S1 as described until the depth of the 15 hand file is reached. Use the S2, exactly as described for the S1, until the depth of the 15 hand file is reached. In the presence of NaOCl or a viscous chelator, scout the apical 1/3 with 10 and 15 hand files and gently work them until they are loose at length. Establish working length, confirm patency and verify the presence of a smooth, reproducible glide path in the apical 1/3. * Refer to manufacturers directions for use for all other motors.

ProTaper Universal Obturator Selection Use the S1, as described, until working length is reached. Use the S2, as described, until working length is reached. Reconfirm working length, especially in more curved canals. Use the F1 in a non-brushing action until working length is reached. Gauge the foramen with a 20 hand file. If this instrument is snug at length, the canal is shaped and ready to obturate. If the 20 hand file is loose at length, proceed to the F2 and, when necessary, the F3, F4 and F5 gauging after each finishing file with corresponding hand file. Important Tips Always establish a reproducible glide path in each canal with a 15 hand file prior to rotary instrumentation. Lubricate files generously with a file prep such as ProLube root canal conditioner. Irrigate, recapitulate and re-irrigate after using each rotary file. Always confirm working length. Take files to resistance never force files. Use all ProTaper Universal rotary files in constant rotation at a speed of 300 RPM. The file will appropriately follow the canal. Inspect files for distortion or wear, such as uneven flutes, dull spots or bending. Use shaping files with a brushing action on the withdrawal stroke in order to create straight-line radicular access and to passively progress apically. Finishing files should follow canal passively to working length, then be withdrawn. Use the SX after the S1 and S2 to optimally shape canals in shorter roots, relocate the coronal aspect of canals away from furcal concavities, or to expand the shape as desired. In more curved canals, take any ProTaper Universal rotary finishing instrument to length only one time and for no more than one second or you may risk transportation. ProTaper Universal rotary files are single patient use devices. Select the ProTaper Universal size verifier that matches the finishing file taken to working length. To set the length in this technique, set the rubber stop at your working length. Take the size verifier to working length. The size verifier should fit passively (can easily be rotated by hand at working length up to 1/2 rotation). If it does not fit passively, the verifier may be used as a finishing file to gently enlarge the canal terminus for a passive fit. Select the obturator that matches the chosen verifier. Always use the length calibration marks on the obturator for length determination. These calibration marks are at 18 mm, 19 mm, 20 mm, 22 mm, and 24 mm. Because gutta-percha distribution on the carrier can vary, do not attempt to set this length based on a ruler measurement of the obturator. Preparation of the Canal Irrigate with BioPure MTAD Antibacterial Root Canal cleanser according to product directions. BioPure MTAD cleanser should be the last liquid introduced into the canal before sealer is placed. Thoroughly dry the canal. ProTaper Universal obturators should be placed only in a canal that has been thoroughly dried. ProTaper Universal absorbent points are appropriate because they conform to these instruments predefined canal shapes. After the canal has been dried, pick up sealer with a ProTaper Universal absorbent point measured to working length. Lightly coat the walls of the canal with the sealer to the working length. Take a dry, appropriately sized ProTaper Universal absorbent point, also measured to working length, and wick up any excess sealer. It is recommended that noneugenol sealers like ThermaSeal Plus sealer be used with the ProTaper Universal obturators alpha phase guttapercha. The sealer, in addition to penetrating minute canal spaces, acts as a lubricant, enhancing the flow characteristics of the centrally condensed gutta-percha. Heating the Obturator Heat the obturator only in the ProTaper Universal obturator oven. This oven is specifically designed to guarantee full and uniform gutta-percha heating. Using the Obturator Oven Plug the oven into a 110 VAC outlet. Then turn the oven on with the power on/off switch located on the back. The yellow stand-by indicator will light up. Put both obturator holders in the upper position by pressing the back of the holder (arrow up) and guiding the holder up with your finger.

After disinfecting the obturator and setting the rubber stop, place the ProTaper Universal obturator in the left obturator holder. Regardless of what working length you have determined, the rubber stop must be under the holder. Make sure the obturator is hanging straight, not at an angle. The obturator must clear all sides of the heating chamber. Push the holder down (arrow down) until you hear it click. Push the button which corresponds to the size of the ProTaper Universal obturator you wish to heat. Then push the start left button. The obturator is now being heated. The heating time varies depending on obturator size and is regulated automatically. After the first signal beep, the obturator is ready for use. Push the obturator holder (arrow up) and guide the holder with your finger. Take the obturator carefully out of the holder by pulling it toward you, making sure not to scrape the obturator on any part of the holder. You may leave the obturator in the ProTaper Universal obturator oven for up to 90 seconds after the first signal beep. It will keep the obturator at the appropriate temperature and ready for use. The oven will beep every 15 seconds to remind you that the obturator is still in the oven. After 90 seconds, the heating element will switch off automatically. As an extra safety precaution, after you ve taken the obturator out of the holder, switch off the heating element by pushing the start left button again. Keep in mind that you won t be able to switch the heating element off during the heating cycle. If you want to heat more obturators, you may alternate using the left and right holders to continue your work efficiently. Wait until after the signal beeps for the first holder, then you may immediately start to heat the other one. You will be unable to heat both sides at the same time. When you re finished using the oven, use the power on/off switch located at the back of the oven to turn it off. Then unplug the oven until you are ready to use it again. Placement of the Obturator In teeth with multiple canals, placement of ProTaper Universal absorbent points in unfilled canals will prevent excess gutta-percha material from going into those canals. After removing the obturator from the oven, place the obturator into the canal using a slow, smooth, continuous motion up to the stopper. The placement motion should not require any serious force. To remove the handle at the orifice level, use one finger to steady the handle while severing the shaft with a Prepi bur, high-speed bur, or ProUltra endo tip. Creating Post Space A Prepi bur or high-speed round bur can be helpful in initiating post space. Using the selected bur at 150,000 RPM to 200,000 RPM, create a dimple 2 mm 3 mm into the obturated canal. Center the dimple on the plastic carrier in cases obturated with solid core techniques. Then select an appropriately sized Pro-Post drill. Set the motor speed to 2,000 RPM. The eccentric (angle) cutting tip will remove the plastic core and gutta-percha. Use intermittent apical pressure to create the post space. Keep the preparation parallel. Retreatment of the ProTaper Universal Obturators ProTaper Universal retreatment files (D1, D2 and D3) are designed to remove filling materials from the root canal prior to reshaping procedures. The technique for removing either a carrier surrounded by gutta-percha or gutta-percha alone is the same. General Tips Always begin with a well-angulated set of radiographs. A good understanding of the anatomy is critical to success. Obtain good straight-line access to the canals. Clinically assess the obturation material. Heat and friction will help to remove obturation material. Throughout the procedure, remove the instruments, inspect and clean the flutes. Then re-insert and continue to auger material out of canal. Recommended speed for the ProTaper Universal retreatment instruments is 500 700 RPM for gutta-percha and carrier based obtruators. Use 300 RPM to remove paste filers. Retreatment Technique Establish a pilot hole using small sized stainless steel hand files with an appropriate solvent, heat carrier or ultrasonic instrument. Without engaging dentin, gently press the spinning D1 into the obturation material. Use the D1 to remove the obturation material from the coronal 1/3. Next, use the D2 to progressively remove material from the middle 1/3. When appropriate, use the D3 to remove obturation material from the apical 1/3. Remove the files frequently and inspect flutes. Continue as long as obturation material is visualized between the cutting blades.

Use hand files with a solvent to remove obturation materials from the apical 1/3 when encountering intricate anatomy. Oven Cleaning Instructions Clean oven after each use. Any gutta-percha remaining on the holder or in the heating chamber may be effectively removed with Orange Solvent (generic dental solvent) or any cotton pledget. Oven Technical Specifications Voltage rating: 110 VAC, 50/60 Hertz, range of the supply voltage +/- 10%, Fuse 2 Amps Slow. Explanation of Marking Symbols Alternate current Power On I Power Off O Caution, Hot Surface PRESS HERE TO RAISE OBTURATOR HOLDER POWER ON/OFF SWITCH PRESS HERE TO LOWER OBTURATOR HOLDER PLACE OBTURATOR HERE PLACE OBTURATOR HERE HEATING CHAMBER HEATING CHAMBER AFTER SELECTING PROPER SIZE BUTTON, PRESS START LEFT FOR LEFT OBTURATOR HEATER AFTER SELECTING PROPER SIZE BUTTON, PRESS START RIGHT FOR RIGHT OBTURATOR HEATER YELLOW LIGHT TO INDICATE OVEN HAS BEEN TURNED ON PRESS HERE FOR OBTURATOR SIZES F1 & F2 PRESS HERE FOR OBTURATOR SIZES F3, F4 & F5

Manufactured For: DENTSPLY Tulsa Dental Specialties DENTSPLY International, Inc. 608 Rolling Hills Drive Johnson City, TN 37604 1-800-662-1202 1-800-597-2779 (fax) www.tulsadentalspecialties.com 2009 DENTSPLY International, Inc. IMPTE03/09E