CC-65/CC-100 CFC-Free Immersion Coolers. Thermo NESLAB Manual P/N 002782 Rev. 04/05/99. Instruction and Operation Manual



Similar documents
HX-500 Outdoor Industrial Recirculating Chiller

Advantium 2 Plus Alarm

Mold Alert... Operation Manual P0270

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model #

Single Station Remote Alarm

, User s Manual

Dehumidifier Users manual. For Models: DH45S DH65S

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

Installation Guide Smart-UPS X External Battery Pack SMX120BP

Ceiling Mount Air Handler Manual

Model: Model: Description: Voltage: Amps: Ice Type: Unit Weight:

Leak Detection System

user s manual Battery Case model #: SPB3200 Battery Case Charger for Samsung Galaxy S 4

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

Air Conditioning Systems Warranty Manual. Version 1.8

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W

Weatherproofing Coaxial Cable Connections

SAFETY PRECAUTIONS Read all of the instructions before using this appliance. When using this appliance, always exercise basic safely precautions,

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

Bottom Loading Water Dispenser

DC REFRIGERATORS 12/24 VOLTS INSTALLATION AND OWNER S MANUAL

2-3 SAS/SATA II HDD Canister USER S MANUAL XC-23D1-SA10-0-R. Document number: MAN A

Single Phase Soft Starter

6 & 12 Volt Battery and Systems Tester with 100 Amp Load

Operators Manual. Circulating Bath / Immersion Circulator With Standard Controller - LED Display /99

SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / Instruction Manual. Page 1 of 9

User Manual. CFG253 3 MHz Function Generator

GAS SAFETY PRECAUTIONS

Kramer Electronics, Ltd. USER MANUAL. Rack Adapter Models: RK-1T2PT RK-2T1PT RK-3T RK-6T RK-9T RK-3TR RK-4PT RK-T2B RK-T2SB

Panel Mounted Air Conditioners by Dantherm, Inc.

User and Installation Guide

CABLE MODEM QUICK START

Support. Phone Number. Technical. Security. On-Site Repair. Peace of Mind. EPSON PreferredSM. Priority. Toll-Free. and

Foodservice Equipment Specialists P.O. Box 880 Saco, ME. / U.S.A * FAX (207)

TIG INVERTER INSTRUCTION MANUAL

Electric Fryers Instruction Manual Models: 8047D, 8048D, 8049D, 8050D, 8051D 8066, 8068, 8068FL, 8073, 8073BF and 8075

AMU PROFESSIONAL VIDEO AND AUDIO MONITORING UNIT OPERATOR'S HANDBOOK ISSUE A3

UNIVERSAL LUMBAR INSTALLATION INSTRUCTIONS

DUAL%CHANNEL BROADBAND%LINEAR%AMPLIFIER Model&A800D

Whynter 62 QT DUAL ZONE PORTABLE FRIDGE / FREEZER

INSTRUCTION MANUAL MODEL PCD-88

BC-5000 OPERATIONS MANUAL BATTERY CAPACITY TESTER COFKO LLC.

BroadBand PowerShield. User Manual

Use & Care Guide AIR CONDITIONERS ACM052 ACM062 ACM072

Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan

SunMaxx Solar Filling Station Operating Instructions

SOLAR ELECTRIC MODULE ES-124 & ES-62T Owners Manual and Installation Guide

Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR İ-COLD Rev. 00

Perseus. Owner s Manual. Vacuum Tube Preamplifier. Rogue Audio, Inc. 3 Marian Lane Brodheadsville, PA Issue date: 02/01/06

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual

Model SPDT Relays

Name of Equipment Silver King Model SKMCD1P/C1. This equipment chapter is to be inserted in the appropriate section of the Equipment Manual.

Installation and Operation Guide for PD5100 Automatic Transfer Switch

How To Install A Power Supply (Uplast) With A Battery Pack

NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

EPSON Preferred. Priority Technical Support. Toll-Free Phone Number. Security and Peace of Mind. On-Site Repair

el-720 Carbon Dioxide Gas Leak Detector User s Guide

Smart-UPS RT External Battery Pack Stack/Rack-Mount 6U

User Guide. HDMI Active Cable Extender. DVI-7370c

MCR1900 Media Converter 19-Slot Chassis

BUILT-IN WINE & BEVERAGE COOLER WITH DUAL TEMPERATURE ZONES INSTRUCTION MANUAL. Model No.: MCWBC77DZC

EPSON PreferredSM. Toll-Free Phone Number. Priority Technical Support. Security and Peace of Mind. On-Site Repair. Epson Stylus Pro 7900/9900

Portable Air Conditioner

NFC Communication Tray Model: HHX-IT3-Z

ROUSH PERFORMANCE - Limited Warranty

Owner s Manual Phoenix Solution. LGR Performance Amplifier

FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

Automatic taper of charge rate for superior battery life through good equalization of cells and low water use rate.

ATS Overhead Table Shelf System INSTRUCTION MANUAL

IR Repeater. Operation Manual P8620

Instruction Manual. Fun Pop 8oz. Popper

Thermistor Thermometer MODEL NO

DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

Owners Manual 2009 PRICE LIST DRYMAXTM. Compressed Air Treatment Products. Refrigerated Air Dryers

SWIMMING POOL HEAT PUMP

ASP-DTH. DVI & Audio to HDMI Converter. User Manual. Manual Number:

INSTALLATION, OPERATING AND SERVICE INSTRUCTIONS

American Fireglass Outdoor Fire pit kit information. General instructions/warnings for outdoor gas pan and burner systems

Mini-HDMI Series MINIATURE MULTIMODE FIBER OPTIC HDMI TRANSMISSION SYSTEM

MODEL G300 BRAKE BLEEDER

i410/i1010 AC/DC Current Clamp

INSTANT HOT WATER RECIRCULATING SYSTEM

How To Use A Cdm250 Digital Multimeter

About this Manual. Support for Your Product

Instructions & Safety Information Models A220-20D and A220-20L Version 2

4 in 1 Air Compressor

Please keep these operating instructions in a safe place.

Instructions A Amp AC/DC Current Probe

Service & Parts Manual

MICA HEATER INSTRUCTION MANUAL Model No: UHM V 50Hz 2200W

Laptop Power Adapter for home/office User Manual

Indirect-Fired Storage Water Heater Models WH-30 through WH-80 INSTALLATION AND OPERATING INSTRUCTIONS

PROAIR Water-Cooled Air Conditioner. CR43WC Model INSTRUCTION MANUAL. Rev. E 2013 Pentair Technical Products P/N

ADP Autoline 3A0986A ENG

Battery Charger. Instruction Manual G

CARING FOR YOUR WATER HEATER

Transcription:

CC-65/CC-100 CFC-Free Immersion Coolers Thermo NESLAB Manual P/N 002782 Rev. 04/05/99 Instruction and Operation Manual

CryoCool Series Immersion Coolers Instruction and Operation Manual PREFACE Tab able of Contents Compliance... 2 Unpacking... 2 Warranty... 2 NES-care... 2 After-sale Support... 2 SECTION I Safety SECTION II General Information SECTION III Installation SECTION IV Operation SECTION V Maintenance SECTION VI Troubleshooting Warnings... 3 Description... 4 Specifications... 4 Site... 6 Electrical Requirements... 6 Fluids... 7 Cooling Probes... 7 Operation... 8 Temperature Adjustment... 8 Service Contracts... 9 Cleaning... 9 Cooling Fluid... 9 Checklist... 10 Service Assistance... 10 Technical Support... 10 SECTION VII Warranty... 11-1 -

Preface Compliance Products tested and found to be in compliance with the requirements defined in the EMC standards defined by 89/336/EEC as well as Low Voltage Directive (LVD) 73/23/EEC can be identified by the CE label on the rear of the unit. The testing has demonstrated compliance with the following directives: LVD, 73/23/EEC Complies with UL 3101-1:93 EMC, 89/336/EEC EN 55011, Class A Verification EN 50082-1:1992 IEC 1000-4-2:1995 IEC 1000-4-3:1994 IEC 1000-4-4:1995 For any additional information refer to the Letter of Compliance that shipped with the unit (Declaration of Conformity). Unpacking Warranty Retain all cartons and packing material until the unit is operated and found to be in good condition. If the unit shows external or internal damage, or does not operate properly, contact the transportation company and file a damage claim. Under ICC regulations, this is your responsibility. Units have a warranty against defective parts and workmanship for one full year from date of shipment. See back page for more details. NES-care Extended Warranty Contract Extend parts and labor coverage for an additional year. Worry-free operation. Control service costs. Eliminate the need to generate repair orders. No unexpected repair costs. Other contract options are available. Please contact Thermo NESLAB for more information. After-sale Support Thermo NESLAB is committed to customer service both during and after the sale. If you have questions concerning the operation of your unit or the information in this manual, contact our Sales Department. If your unit fails to operate properly or if you have questions concerning spare parts or Service Contracts, contact our Service Department. Before calling, please refer to the serial number label on the rear of the case top to obtain the following information: - BOM number - Serial number - 2 -

Warnings Section I Safety Make sure you read and understand all instructions and safety precautions listed in this manual before installing or operating your unit. If you have any questions concerning the operation of your unit or the information in this manual, contact our Sales Department (see After-sale Support). Performance of installation, operation, or maintenance procedures other than those described in this manual may result in a hazardous situation and may void the manufacturer's warranty. Transport the unit with care. Sudden jolts or drops can damage the refrigeration lines. Observe all warning labels. Never remove warning labels. Never operate damaged equipment. Always turn off the unit and disconnect the line cord from the power source before performing any service or maintenance procedures, or before moving the unit. Never operate equipment with damaged line cords. Refer service and repairs to a qualified technician. In addition to the safety warnings listed above, warnings are posted throughout the manual. These warnings are designated by an exclamation mark inside an equilateral triangle with text highlighted in bold print. Read and follow these important instructions. Failure to observe these instructions can result in permanent damage to the unit, significant property damage, or personal injury or death. - 3 -

Description Section II General Information The CryoCool Series of immersion coolers is designed as a refrigeration source for sub-ambient work in liquid baths. CryoCools employ mechanical refrigeration systems constructed with fullhermetic refrigeration compressors. The CC-65 employs a single stage refrigeration system with one compressor. The CC-100 has a dual stage refrigeration system using two compressors in cascade configuration. An insulated coaxial hose assembly carries refrigerant to the cooling probe. Specifications Temperature Range 1 Temperature Stability Evaporator Head Dimensions 2 (Length x Diameter) F-head FV-head R-head Minimum Bend Radius 3 F-head FV-head R-head CC-65-20 C to -55 C ±0.5 C 25 x 1 /2 63.5 x 1.3 7 ¼ x 1 1 /4 18.4 x 3.2 1 2.5 CC-100-25 C to -90 C ±0.5 C 18 x 5 /8 45.7 x 1.6 25 x 1 /2 63.5 x 1.3 7 ¼ x 1 1 /4 18.4 x 3.2 1 1 /2 3.8 1 2.5 1. Low end specifications listed for "no load" conditions. 2. See Section III, Cooling Probes for a description of the evaporator heads. 3. The minimum bend radius is the smallest radius that the cooling probe can be bent without suffering damage at room temperature. Never bend the probe when it is cold. - 4 -

Evaporator Hose Dimensions (Length x Diameter) Unit Dimensions (H x W x D) Weight Pounds Kilograms Cooling Capacity 4 CC-65 44 x 1 1 /4 111.8 x 3.2 15 1 /4 x 7 1 /2 x 10 1 /4 38.7 x 19.1 x 26.0 41.0 18.6 CC-100 65 x 1 ½ 165.1 x 3.8 20 1 /2 x 14 1 /2 x 17 3 /4 52.1 x 36.8 x 45.1 130.0 59.0 Heat Removal (Watts) 200 180 160 140 120 100 80 60 A=CC-100, 60Hz B=CC100, 50Hz C=CC65, 60Hz R-Probe D=CC65, 50Hz R-Probe A B C D 40 20-90 -80-70 -60-50 -40-30 -20 Temperature ( C) 4. Specifications obtained in a 2 liter container at +20 C ambient with cooling fluid specific heat of 0.5. - 5 -

Site Section III Installation The unit should be located in a laboratory or clean industrial environment where ambient temperatures are inside the range of +55 F to +95 F (+13 C to +35 C). Never place the unit in a location where excessive heat, moisture, or corrosive materials are present. The unit has an air-cooled refrigeration system. On the CC-65 air is drawn through the top of the unit and discharged through the sides and bottom of the unit. On the CC-100 air is drawn through the front of the unit and discharged through the rear. The unit must be positioned so the intake and discharge are not impeded. A minimum clearance of 12 inches (30 centimeters) on all vented sides is necessary for adequate ventilation. Inadequate ventilation will cause a reduction in cooling capacity and, in extreme cases, compressor failure. Excessively dusty areas should be avoided and a periodic cleaning schedule should be instituted (see Section V, Cleaning). The unit will retain its full rated capacity in ambient temperatures up to approximately +75 F (+24 C). Above +75 F, reduce the cooling capacity 1% for every 1 F above +75 F, up to a maximum ambient temperature of +95 F. In terms of Celsius, reduce the cooling capacity 1% for every 0.5 C above +24 C, up to a maximum ambient temperature of +35 C. Electrical Requirements Refer to the serial number label on the rear of the unit for the specific electrical requirements of your unit. Make sure the voltage of the power source meets the specified voltage, ±10%. Make sure an adequate ground connection is provided. - 6 -

Fluids The selected fluid must have a viscosity of 50 centistokes or less at the lowest operating temperature. Never use flammable or corrosive fluids with this unit. Cooling Probes Three cooling probes (or evaporator heads) are available with CryoCool units: F (flexible), FV (flexible, very), and R (rigid). Refer to Section II, Specifications to determine the specific cooling probe on your unit. Place the cooling probe in the work area. For maximum cooling efficiency, the probe should be fully immersed in the fluid and should be located as close as possible to the agitator or stirrer motor (if so equipped). Care should be taken not to stretch or bend the probe beyond it limits or exceed its minimum bend radius. Bend the probe only at room temperatures, never bend it when it is cold. - 7 -

Operation Section IV Operation Before starting the unit, double check all electrical connections and make sure the cooling probe is properly immersed in the work area. All CryoCools are operated by a single power switch. The CC-100 has three lights: one indicates the unit is operating, a second indicates the high stage compressor is operating, and a third indicates the low stage compressor is operating. Once the unit is turned on, the refrigeration system will reduce the temperature of the cooling fluid in the work area to the lowest achievable temperature under the existing heat load conditions. For best results, stirring or agitation in the work area is recommended. When not stirred, cooling capacities will be reduced and the work area will show temperature layering, with the coldest, most dense zone at the bottom. When the unit is shut off, wait approximately five minutes before restarting. This allows time for the refrigeration pressures to equalize. If the pressures are not allowed to equalize, the compressor will short-cycle (clicking sound) and no cooling will occur. Temperature Adjustment If temperature control is desired, a Thermo NESLAB Cryotrol Temperature Controller is required. Refer to the Cryotrol Temperature Controller Instruction and Operation Manual for installation and operation instructions. Never operate any CryoCool unit above -25 C. Older Cryotrols are equipped with a round 8 pin connector. If you wish to connect your older Cryotrol unit to the CryoCool, an adapter cable is available from Thermo NESLAB. Contact our Service Department for more information (see Preface, After-Sale Support). A Thermo NESLAB Exatrol Digital Temperature controller may also be used to control temperature. The Exatrol is designed for more precise temperature control and offers tighter temperature stability than the Cryotrol. Refer to the Exatrol manual for additional information. - 8 -

Service Contracts Cleaning Section V Maintenance Thermo NESLAB offers on-site Service Contracts that are designed to provide extended life and minimal down-time for your unit. For more information, contact our Service Department (see Preface, After-sale Support). For proper operation, the unit needs to pull substantial amounts of air through the condenser. A build up of dust or debris on the fins of the condenser will lead to a loss of cooling capacity. Periodic vacuuming of the condenser fins is necessary. The frequency of cleaning depends on the operating environment. We recommend making a monthly visual inspection of the condenser after initial installation. After several months, the cleaning frequency will be established. Cooling Fluid The cooling fluid in the work area should be periodically replaced when operating at low temperatures. At low temperatures, the cooling fluid may collect water vapor from the air. As the concentration of water in the cooling fluid increases, performance is adversely affected. - 9 -

Checklist Section VI Troub oubleshooting Unit will not start. Make sure the voltage of the power source meets the specified voltage, ±10%. Refer to the serial number label on the rear of the unit for the specific electrical requirements of your unit. Loss of cooling capacity. Make sure the heat load in the work area is not greater than the cooling capacity of the unit. Refer to Section II, Specifications to review the cooling capacity specifications. Check for ice build up on the cooling probe. A layer of ice will act as insulation and reduce the cooling capacity. Ice build up is often an indication that the cooling fluid needs replacing. Defrost the cooling probe and change the cooling fluid. When the unit is shut off, wait approximately five minutes before restarting. This allows time for the refrigeration pressures to equalize. If the pressures are not allowed to equalize, the compressor will short-cycle. Service Assistance If, after following these troubleshooting steps, your unit fails to operate properly, contact our Service Department for assistance (see Preface, After-sale Support). Before calling, please obtain the following information: Part number Serial number Voltage of unit Voltage of power source Technical Support Our Service Department can provide you with a complete list of spare parts for your unit (see Preface, After-sale Support). Before calling, please obtain the following information: Part number Serial number - 10 -

WARRANTY Thermo NESLAB Instruments, Inc. warrants for 12 months from date of shipment any Thermo NESLAB unit according to the following terms. Any part of the unit manufactured or supplied by Thermo NESLAB and found in the reasonable judgment of Thermo NESLAB to be defective in material or workmanship will be repaired at an authorized Thermo NESLAB Repair Depot without charge for parts or labor. The unit, including any defective part must be returned to an authorized Thermo NESLAB Repair Depot within the warranty period. The expense of returning the unit to the authorized Thermo NESLAB Repair Depot for warranty service will be paid for by the buyer. Thermo NESLAB s responsibility in respect to warranty claims is limited to performing the required repairs or replacements, and no claim of breach of warranty shall be cause for cancellation or recision of the contract of sales of any unit.with respect to units that qualify for field service repairs, Thermo NESLAB s responsibility is limited to the component parts necessary for the repair and the labor that is required on site to perform the repair. Any travel labor or mileage charges are the financial responsibility of the buyer. The buyer shall be responsible for any evaluation or warranty service call (including labor charges) if no defects are found with the Thermo NESLAB product. This warranty does not cover any unit that has been subject to misuse, neglect, or accident. This warranty does not apply to any damage to the unit that is the result of improper installation or maintenance, or to any unit that has been operated or maintained in any way contrary to the operating or maintenance instructions specified in Thermo NESLAB s Instruction and Operation Manual. This warranty does not cover any unit that has been altered or modified so as to change its intended use. In addition, this warranty does not extend to repairs made by the use of parts, accessories, or fluids which are either incompatible with the unit or adversely affect its operation, performance, or durability. Thermo NESLAB reserves the right to change or improve the design of any unit without assuming any obligation to modify any unit previously manufactured. THE FOREGOING EXPRESS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Thermo NESLAB S OBLIGATION UNDER THIS WARRANTY IS STRICTLY AND EXCLUSIVELY LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE COMPONENT PARTS AND Thermo NESLAB DOES NOT ASSUME OR AUTHORIZE ANYONE TO ASSUME FOR IT ANY OTHER OBLIGATION. Thermo NESLAB ASSUMES NO RESPONSIBILITY FOR INCIDENTAL, CONSEQUENTIAL, OR OTHER DAMAGES INCLUDING, BUT NOT LIMITED TO LOSS OR DAMAGE TO PROPERTY, LOSS OF PROFITS OR REVENUE, LOSS OF THE UNIT, LOSS OF TIME, OR INCONVENIENCE. This warranty applies to units sold in the United States. Any units sold elsewhere are warranted by the affiliated marketing company of Thermo NESLAB Instruments, Inc. This warranty and all matters arising pursuant to it shall be governed by the law of the State of New Hampshire, United States. All legal actions brought in relation hereto shall be filed in the appropriate state or federal courts in New Hampshire, unless waived by Thermo NESLAB. - 11 -