Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material



Similar documents
Coating and Extraction of Honeycomb Denuders

DRAFT Standard Operating Procedure for Long-Term Archiving of PM 2.5 Filters and Extracts

Private Water Supplies Sampling Manual. A Field Guide

SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES

Solids, Volatile Dissolved and Fixed Dissolved

Extraction: Separation of Acidic Substances

The Analytical Balance

Aptima Specimen Transfer Kit

Experiment 8 - Double Displacement Reactions

Paper Chromatography: Separation and Identification of Five Metal Cations

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2

PLEASE READ ALL INSTRUCTIONS BEFORE USE AND SAVE A COPY FOR FUTURE REFERENCE!

EFFECT OF SALT ON CELL MEMBRANES

The Determination of an Equilibrium Constant

Dissolved and precipitated oxalate

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

Relative Humidity Calibration Kit

Protocol Micro Lab. K:\Microlab_protocols\Protocols_active\03 Instrument manuals\ml03_001_001 DGGE.doc. Preparation of gels

RAPID FIXER FACT SHEET A LIQUID RAPID FIXER FOR BLACK AND WHITE FILMS AND PAPERS. August 2002

Bluelab Soil ph Meter Instruction Manual

ESS Method 130.1: General Auto Analyzer Procedures

EXPERIMENT FIVE. Preparation of Cyclohexene from Cyclohexanol: an Elimination Reaction DISCUSSION

NO TARN 15. 5) Can be used on Silver AND Gold. Protects gold from blackening during shipping due to reaction with box chemicals.

Digital Photo Picture Frame. Operation Manual

EFFECT OF ALCOHOL ON CELL MEMBRANES

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

DNA Analyst Training Laboratory Training Manual Protocol 2.02 Clean Technique

Revised: June 6, 2007

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

105 Adopted:

Blood Collection and Processing SOP

UltraClean Soil DNA Isolation Kit

Recovery of Elemental Copper from Copper (II) Nitrate

Physical and Chemical Properties and Changes

Determination of Insoluble Solids in Pretreated Biomass March 2008 Material

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

Making Biodiesel from Virgin Vegetable Oil: Teacher Manual

USEPA Electrode Method Method 8156

Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:

Guidelines for Ethidium Bromide Waste Management & Disposal. University of Tennessee-Knoxville

How to Prepare Powdered Infant Formula in Care Settings

Household Acids and Bases

Enzyme Pre-Lab. Using the Enzyme worksheet and Enzyme lab handout answer the Pre-Lab questions the pre-lab must be complete before beginning the lab.

Please read and keep these instructions. Run water through your machine before first use

Determination of a Chemical Formula

Introduction. ph = log [H + ]

Solubility Product Constants

experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.

WESTERN BLOT DETECTION KIT Buffers and detection reagents for up to ten 10 x 10 cm 2 blots. Fluorescent detection via: Goat anti-mouse SureLight P3

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference

BLENDER. Safety Guidelines. Meet Your Rapid Boiler. Before First Use. Using Your Blender. Blender Setting Guide. Cleaning Your Rapid Boiler.

Specific volume and absorption properties

Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid

SEPARATION OF A MIXTURE OF SUBSTANCES LAB

COMMON LABORATORY APPARATUS

THE ACTIVITY OF LACTASE

PREPARATION AND PROPERTIES OF A SOAP

Beware that ordinary prescription glasses do not provide adequate protection. This is also true with poorly fitting safety glasses.

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

Color, True and Apparent

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

Color, True and Apparent

ABC CIRCUITS 10 Circuit Drive Circuit Town, IN SLUG CONTROL PLAN

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Cidex OPA Solution Training Checklist and Competency for Semi-Critical Medical Devices

ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST

PM 2.5 Speciation Network Newsletter

Southern Blot Analysis (from Baker lab, university of Florida)

Neutralizing an Acid and a Base

Luminol Test PROCESS SKILLS SCIENCE TOPICS VOCABULARY

Safety, Cleaning, and Chemical Disposal Procedures

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES

TECHNICAL MEMORANDUM TO: FROM: AUTHOR: DATE: SUBJECT: Introduction Experimental Design

Ultrasonic support removal tank

Care and use guide. soil ph meter. Contents. page

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

MOSQUITO AND INSECT REPELLENT PRODUCTS

CONTROLLING CROSS INFECTION IN THE DENTAL LABORATORY. Best Practice Guide

DryWeight BulkVolume

KODAK Developer System Cleaner and Neutralizer

18 Conductometric Titration

Acid Base Titrations

Instruction manual. VWR unstirred water baths

Standard Operating Procedure (SOP) Work Package 8. Sample Collection and Storage

SOLAR PV SYSTEM MAINTENANCE GUIDE

Milwaukee USER MANUAL

Biosafety Spill Response Guide

Loss-on-Ignition Standard Operating Procedure

ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST

Standard Operating Procedure for a Chemical Spill

Dry Ice Color Show Dry Ice Demonstrations

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Separation by Solvent Extraction

Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance

PART I: PREPARATION OF SOLUTIONS AND STANDARDIZATION OF A BASE

Quantifying Bacterial Concentration using a Calibrated Growth Curve

Transcription:

PM 2.5 Nylon Filter Cleaning Date: August 25,2009 Page I of8 Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material Environmental and Industrial Sciences Division RTI Intemational* Research Triangle Park, North Carolina Prepared by: &v fj.~()-.j Date: ~(J.Jo (() q a l/a- sh CJt; :7 Reviewed by: ~~ Q ~ Date: J /. ~~ Approved by: ~ Date: f1y~ " ' / ~RTI INTERNATIONAL * RTI International is a trade name of Research Triangle Institute.

Page 2 of 8 Contents Section Page 1.0 Procedural Section...3 1.1 Purpose and Applicability...3 1.2 Summary of Method...3 1.3 Health and Safety Warnings...3 1.4 Cautions...3 2.0 Apparatus and Reagents...4 3.0 Filter Cleaning...4 3.1 Cleaning Procedure...4 3.2 Filter Acceptance Testing...6 4.0 Quality Control...7 5.0 References...7 Attachment A...8

Page 3 of 8 Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2.5 Material 1.0 Procedural Section 1.1 Purpose and Applicability Nylon filters are used for the collection of PM 2.5 material in the chemical speciation particulate samplers. These filters are analyzed for the following ions: nitrate, sulfate, ammonium, sodium, and potassium. The filters, as purchased and received from different manufacturers, show unacceptable levels of these ions, often exceeding the maximum level of 1 Fg per filter for a particular ion. This has prompted the development of a procedure for cleaning the nylon filters prior to their use for field sampling; this procedure is described in this standard operating procedure (SOP). 1.2 Summary of Method Fifty nylon filters are placed in a 2 L polypropylene jar with approximately 1000 ml of polished deionized water (18.2 MΩ-cm; water that has been passed through a secondary deionization system). The filters are shaken in the water for approximately 2 minutes, and the water is decanted and discarded. This process is repeated. The jar is then filled with polished deionized water and placed on a Toxicity Characteristic Leaching Procedure (TCLP) apparatus (TLCP, EPA SW-846 Method 1311). The jar is rotated for 7 to 8 hours, and the water is replaced with fresh polished deionized water. The jar is then rotated overnight for 14 to 16 hours before the water is replaced again. After another 24 hours of washing, the water is drained from the filters and the filters are dried. (The order of extended washing may vary; that is the sequence may be 24 hours, 7 to 8 hours, and then 12 to 14 hours rather than 7 to 8 hours, 12 to 14 hours, and 24 hours.) The filters are dried on glass racks in a convection oven set at 45EC. One filter out of 50 is desorbed and analyzed by IC to test for residual contamination, prior to being approved for later use. 1.3 Health and Safety Warnings The PM 2.5 filter-preparation operations do not involve unusual risks from electrical equipment or chemical exposures. Standard RTI laboratory health and safety precautions will be followed. 1.4 Cautions Laboratory personnel should always wear clean clothes and wash hands thoroughly before performing filter handling and analysis procedures. The use of gloves rinsed with deionized water is required for all steps of the filter cleaning process because this will minimize the potential for laboratory contamination.

Page 4 of 8 2.0 Apparatus and Reagents The nylon filters used are Whatman 47 mm nylon membrane filters, 1.0 μm pore size (Whatman catalog number 7410-004). The only reagent needed is polished deionized water (18.2 MΩ-cm; water that has been passed through a secondary deionization system). Several pieces of equipment are used for cleaning the nylon filters. Included are: 1. 2-L polypropylene wide-mouth Mason jars (VWR Catalog Number 16128-660 or equivalent) 2. TCLP apparatus (TLCP, EPA SW-846 Method 1311) that holds six 2-L jars. 3. Programmable timer (VWR Lab Controller or equivalent) 4. Convection drying oven (VWR Model 1320 or equivalent) 5. 11'' x 11'' glass drying rack (custom made from 1/4'' glass rods in parallel rows attached to 3/8'' glass rods serving as a frame; center-to-center distance for the 1/4'' parallel glass rods is ½'') 6. Plastic colander approximately 8'' in diameter from a kitchen appliance store. 3.0 Filter Cleaning 3.1 Cleaning Procedure The nylon filters are cleaned using the following procedure, which should be started at the beginning of a work day. The date when the cleaning is started is entered into the log book and the batch is identified by this date. 1. Fifty 47-mm nylon filters are carefully removed from the manufacturer s filter container using either gloves or forceps. Each filter is separated with a blue tissue which is removed prior to placing the 50 filters into a 2-L polypropylene jar that contains approximately 1000 ml of polished deionized water. The lid is attached, and the jar is shaken gently for approximately 2 minutes. The water is then carefully poured out of the jar without losing any filters. This rinse step is repeated. The two-step rinse procedure is then duplicated with five additional 2-L jars each loaded with 50 filters. Each jar is labeled with a letter (i.e., A, B, C), using a marker. 2. Each jar is carefully filled with polished deionized water until it is overflowing; it is then capped tightly and placed on the TCLP apparatus. The apparatus is turned on and mixes the filters and polished water by rotating the jars end-over-end. It cycles until the end of the day (i.e., 7 to 8 hours). The water is carefully poured out of each jar, and the jars are again filled to overflowing. The jars are placed on the TCLP apparatus, turned on, and allowed to run overnight, or for 14 to 16

Page 5 of 8 hours. At the beginning of the next work day, the water is poured out again and replaced, and the jar is placed back on the apparatus for approximately 24 hours, or until the beginning of the next work day. Depending on a person s work schedule, the order of the extended washing may be varied; that is, the sequence may be 24 hours, 7 to 8 hours, and then 12 to 14 hours rather than 7 to 8 hours, 12 to 14 hours, and 24 hours. Note: Because the filters tend to stick to the sides of the jars during rotation, the TCLP apparatus is connected to a timer that is programmed to rotate the jars for 15 minutes, and then allow them to sit at rest for 2 minutes. During this rest period, the filters that were stuck to the sides of the jar slip away and fall to the lower part of the jar of water. This timed cycle of rotation, followed by resting, continues through each cleaning period. The procedure for programming the time is given in Attachment A. 3. The jars are removed from the TCLP apparatus after the final wash and are taken to a Class 100 clean room for drying of the filters. The lid of a jar is removed and the water and filters are gently poured into a pre-rinsed plastic colander placed in a sink in the clean room. It may be necessary to add polished deionized water to the jar several times to remove all of the filters. The excess water is allowed to drain from the filters and the colander for several minutes. Any filters that fall into the sink during this process will be discarded. 4. With gloves and a clean forceps, the filters are removed from the colander one by one and placed separately on the drying rack, which has been thoroughly prerinsed with polished deionized water shortly before use. The loaded rack is carefully placed in the oven, which is set at 45EC. The filters are allowed to dry completely. The filters sometimes curl slightly during the drying process. A large amount of curling indicates that the oven temperature is too high. If so, slightly reduce the temperature so the filters dry without curling. Note: The drying oven must be kept free of any dust or particulate material and should only be operated in a clean environment. The oven should be visually inspected for any contamination prior to each use. A dedicated oven used only for drying PM 2.5 nylon filters is used. 5. The dried filters are removed from the drying rack using clean forceps and are placed back into the original manufacturer s plastic containers. These containers are washed with deionized water and are dried before reuse. Filters will be inspected for pinholes and tears; any damaged filter will be discarded. Twentyfive filters are placed in each container. Each container is labeled with the batch number (i.e., start date for cleaning) and the jar identifier (i.e., A, B, C).

Page 6 of 8 3.2 Filter Acceptance Testing One filter from each set of dried filters is selected at random for analysis. Blank filters are analyzed according to the analytical procedure described elsewhere in the SOPs for Anion (Hardison, 2008) and Cation (Hardison, 2008) analysis contained in the laboratory Quality Assurance Project Plan. For lot acceptance, the filter loadings of the ions of interest (i.e., sodium, potassium, ammonium, nitrate, and sulfate) must each be less than 1.0 Fg per filter. If any ion exceeds the limit, the entire lot must be rejected. Rejected lots may be re-cleaned using the same procedure. Each accepted batch of filters is assigned a unique number. Each filter s batch number is recorded in the PM 2.5 database when it is loaded into a sample module in the Sample Handling and Archiving Laboratory. The lot number can be used to trace the acceptance test results in case there is a question about any filter. Note: Several different cleaning procedures were used during the course of the PM 2.5 Speciation Trends Network contract, which began in early 2000. This note summarizes the procedures used for cleaning nylon filters prior to finalization of the method described in this SOP. Prior to March 28, 2000, filters were soaked three times for 30 minutes in deionized water without shaking or ultrasonication. Drying and acceptance procedures were identical to those previously described. Prior to December 1, 2001, filters were cleaned using a shaker for the final 24-hour wash in deionized water. In fall 2001, some batches of filters received from the supplier were noted to be partially disintegrating in the shaker. It was concluded that the filter s durability was somewhat variable, and that shaking for 24 hours was too forceful for the less durable filters; therefore, the more gently rolling method was adopted. Prior to December 1, 2002, filters were placed in a polypropylene jar of sodium carbonate/sodium bicarbonate solution (the eluent used for anion analysis). The jar containing the filters was placed in an ultrasonic bath for 1 hour. The filters were then rinsed three times with deionized water, rinsed gently using a jar roller mill in deionized water for about 1 hour, rinsed again manually three or four times, and then rinsed gently in fresh deionized water for 24 hours using the jar roller mill. This procedure was abandoned for the following reasons: the ultrasonic bath sometimes caused partial disintegration of the filters, sodium from the eluent solution was sometimes still present on the filters, and the TCLP apparatus was better than a roller because it provides endover-end mixing. The method described in this SOP was subsequently adopted.

Page 7 of 8 4.0 Quality Control The quality control activities include the following: 1. Perform ion analyses of the polished deionized water whenever the deionizer beds are changed to determine that the ions of interest are below their maximum allowable concentration, as presented in Table 1. Replace the ion exchange beds in the water deionization system if these limits are exceeded. Table 1. Maximum Allowable Concentration (MAC) for Ions of Interest Ion of Interest MAC, Fg/mL Nitrate 0.01 Sulfate 0.01 Ammonium 0.01 Sodium 0.01 Potassium 0.02 2. Keep all jars closed and stored in a clean environment when not in use. 3. Periodically wipe down the inside of the drying oven with wet, lint-free tissues. 5.0 References Hardison, E. 2008. Standard Operating Procedure for PM 2.5 Anion Analysis. Quality Assurance Project Plan Chemical Speciation of Particulate Matter, Volume II, Appendix A-5.1, revision 6. Hardison, E. 2008. Standard Operating Procedure for PM 2.5 Cation Analysis. Quality Assurance Project Plan Chemical Speciation of Particulate Matter, Volume II, Appendix A-5.2, revision 6.

Page 8 of 8 Attachment A Method for Programming the VWR Lab Controller The device is programmed for the repeat mode, which permits repeatedly turning equipment on or off at one or two unique time intervals. 1. Press the CHANNEL SELECT key until the OUTLET channel is selected. 2. Delete all time-of-day program times by pressing the C key, and then the REPEAT key. 3. Press the OUTLET ON/OFF key to ON. 4. Press the 1, 5, 0, and 0 keys to program 15 minutes (15.00) power on. 5. Press the REPEAT key. 6. Press the 2, 0, and 0 keys to program 2 minutes power off. 7. Turn on the toggle switch on the TCLP apparatus (if it is not already on) and press the START/STOP key to begin counting down. 8. At zero, the outlet switches to OFF, the alarm sounds for 2 seconds, the display automatically returns to the programmed 2 minutes, and the timer begins counting down. At the next zero, it switches, alarms, displays 15.00, and begins counting down. This process will repeat until the C key is pressed, or the toggle switch is turned off.