Tonometry after Laser in Situ Keratomileusis Treatment: A Preliminary Study in Thai Patients



Similar documents
Underestimation of Intraocular Pressure in Eyes After Laser In Situ Keratomileusis

How do you adjust for IOP in a LASIK patient?

Goldmann applanation tonometry (GAT) is the standard for

T he assessment of precise intraocular pressure (IOP) is of

Evaluation of the Orssengo-Pye IOP corrective algorithm in LASIK patients with thick corneas

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US

Contact Lens & Anterior Eye

Retreatment by Lifting the Original Laser in Situ Keratomileusis Flap after Eleven Years

INTRODUCTION. Ş. BAYRAKTAR 1 ' 2, Z. BAYRAKTAR 2 Istanbul Surgery Hospital, Istanbul 2. BeyogIu Eye Education and Research Hospital, Istanbul - Turkey

Tonometric Changes of Latanoprost-Induced Intraocular Pressure Reduction after Photorefractive Keratectomy MATERIALS AND METHODS.

Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure

Clinical Characteristics of Glaucomatous Subjects Treated with Refractive Corneal Ablation Surgery

Case Report Laser Vision Correction on Patients with Sick Optic Nerve: A Case Report

Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps

Long-term stability of the posterior cornea after laser in situ keratomileusis

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.

OriginalArticle. Quality of Life Assessment Before and After Laser in situ Keratomileusis

Anterior Elevation Maps as the Screening Test for the Ablation Power of Previous Myopic Refractive Surgery

Overview of Refractive Surgery

LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY

Comparison of Two Procedures: Photorefractive Keratectomy Versus Laser In Situ Keratomileusis for Low to Moderate Myopia

VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment)

Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking

Original Articles. Laser in situ Keratomileusis to Correct Residual Myopia After Cataract Surgery

LASIK: Clinical Results and Their Relationship to Patient Satisfaction

Validation of a New Scoring System for the Detection of Early Forme of Keratoconus

The Efficacy of Multi-Zone Cross-Cylinder Method for Astigmatism Correction

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM

Wavefront technology has been used in our

Uniquely Safe. predictably better for our patients. enhancement, may be significantly reduced.

Rosario G. Anera, PhD, Jose R. Jiménez, PhD, Luis Jiménez del Barco, PhD, Javier Bermúdez, PhD, Enrique Hita, PhD

Archive of SID. Primary Open Angle Glaucoma and Post-LASIK Keratectasia. Challenging Case

Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy

LASIK in the Presbyopic Age Group

Initial Supervised Refractive Surgical Experience: Outcome of PRK and LASIK

LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth?

How To See With An Cl

The pinnacle of refractive performance.

Dr. Booth received his medical degree from the University of California: San Diego and his bachelor of science from Stanford University.

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay

Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser

KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,

Intraoperative Complications of Laser in Situ Keratomileusis in Yazd

LASIK SURGERY OUTCOMES, VOLUME AND RESOURCES

Straylight values 1 month after laser in situ keratomileusis and photorefractive keratectomy

Wavefront-guided versus standard laser in situ keratomileusis in low to moderate myopia

Changes in higher order aberrations after wavefront-guided PRK for correction of low to moderate myopia and myopic astigmatism: Two-year follow-up

Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery

What is Refractive Error?

Conductive keratoplasty (CK) utilizes radiofrequency energy. Original Article

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation

Factors Affecting Long-term Myopic Regression after Laser In Situ Keratomileusis and Laser-assisted Subepithelial Keratectomy for Moderate Myopia

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

Cornea and Refractive Surgery Update

Diego Fernando Suárez Sierra, MD Fellow Cornea and Refractive Surgery Fellow Lens and Ocular Surface Vejarano Laser Vision Center

final corrected draft

EFFECT OF MYOPIC LASIK ON RETINAL NERVE FIBER LAYER THICKNESS- IS IT SAFE OR UNSAFE?

INTRODUCTION. Trans Am Ophthalmol Soc 2006;104:

Laser in situ keratomileusis in patients with corneal guttata and family history of Fuchs endothelial dystrophy

Outcome of Laser in Situ Keratomeliusis (Lasik) in Low to High Myopia: Review of 200 Cases

S canning laser polarimetry (SLP) of the retinal nerve fibre

Comparison of higher-order aberrations after wavefront-guided laser in situ keratomileusis and laser-assisted subepithelial keratectomy

Refractive Surgery. Common Refractive Errors

Changes in central corneal thickness and refractive error after thin-flap laser in situ keratomileusis in Chinese eyes

Ectasia after laser in-situ keratomileusis (LASIK)

LASIK, Epi LASIK and PRK Past present and future

Tamer O. Gamaly, FRCS; Alaa El Danasoury, MD, FRCS; Akef El Maghraby, MD

All surgical procedures have the potential for

Accelerated Refractive Performance

Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD

SUB-PROTOCOL Title: Healthcare Performance Measurement & Reporting for LASIK surgery services provided by OPTIMAX Malaysia

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

Corneal intrastromal implantation surgery for the treatment of moderate and high myopia

Effect of Preoperative Keratometric Power on Intraoperative Complications in LASIK in 34,099 Eyes

Excimer Laser Refractive Surgery

Central Islands After LASIK Detected by Corneal Topography

Laser Vision Correction: A Tutorial for Medical Students

Laser in situ keratomileusis (LASIK) has been. Retreatment of Hyperopia After Primary Hyperopic LASIK REPORTS

Vision Glossary of Terms

Laser refractive surgery is becoming increasingly

EVEN THOUGH EARLY CUSTOMIZED ABLATION RESULTS

Topographically-guided Laser In Situ Keratomileusis to Treat Corneal Irregularities

5/24/2013 ESOIRS Moderator: Alaa Ghaith, MD. Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD

Treatment of Myopia and Myopic Astigmatism by Customized Laser In Situ Keratomileusis Based on Corneal Topography

refractive surgery a closer look

Life Science Journal 2014;11(9) Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000)

Vision Correction Surgery Patient Information

Intraocular Lens Power Calculation after Myopic Refractive Surgery

significantly different from that of sequential treatments. Methods: Data were obtained from 254 consecutive patients that were

INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the

For approximately two decades photorefractive keratectomy. Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK.

Alain Saad, MD, Alice Grise-Dulac, MD, Damien Gatinel, MD, PhD

Prophylactic Effects of Mitomycin-C on Regression and Haze Formation in Photorefractive Keratectomy

LASER VISION CORRECTION

Corneal flap thickness with the Moria M2 single-use head 90 microkeratome

Dynamic Contour Tonometry

Keratoconus Detection Using Corneal Topography

LASER VISION C ORRECTION REFRACTIVE SURGERY CENTER

Transcription:

Tonometry after Laser in Situ Keratomileusis Treatment: A Preliminary Study in Thai Patients Ngamkae Ruangvaravate, MD*, Atiporn Thuangtong, MD*, Panida Kosrirukvongs, MD*, Pinnita Prabhasawat, MD*, Wipawee Booranapong, MD*, Sabong Srivannaboon, MD* * Department of Ophthalmology, Faculty of Medicine, Siriraj Hospital, Mahidol University Objective: To evaluate the change in intraocular pressure (IOP) measurement by Goldmann applanation tonometer after Laser in Situ Keratomileusis (LASIK) for myopia and myopic astigmatism, and to assess the correlation between the changes of IOP reading and the reduction of central corneal thickness (CCT) after LASIK in Thai patients. Study design: Prospective correlational study. Material and Method: LASIK was performed on 65 eyes of 33 patients for correction of myopia and myopic astigmatism. IOP was measured by Goldmann applanation tonometer before and 3 months after LASIK. The correlation between the change in IOP reading and the change in central corneal thickness were evaluated. Results: IOP reading was significantly reduced by mean of 2.9 + 2.5 mmhg (p = 0.0001). The authors used Pearson analysis to study the correlation between the change in IOP and the reduction of CCT. In subgroup analysis the patients were divided by degree of myopia: group 1, myopia less than -3 diopters (D) (n = 14); group 2, myopia -3 to -6 D (n = 31); group 3, myopia greater than -6 D (n = 20)). The result showed more correlation in higher myopia group (Pearson; r = 0.158 in group 3, r = -0.098 in group 2 and r = -0.102 in group 1). Conclusion: Goldmann applanation tonometry underestimates the IOP in thin cornea. Variability in CCT is a potent confounder of this tonometry technique. Therefore, it has important implications for considering CCT measurement incorporated with Goldmann applanation tonometry for glaucoma diagnosis especially in myopic patients who undergo LASIK surgery. Keywords: Goldmann applanation tonometry, Intraocular pressure, Laser in Situ Keratomileusis, Central corneal thickness J Med Assoc Thai 2005; 88(3): 340-4 Full text. e-journal: http://www.medassocthai.org/journal Refractive procedures have become a popular surgical option for the treatment of myopia. Laser in Situ Keratomileusis (LASIK) has become the most accepted surgical modality for the correction of a wide range of myopia. LASIK is performed routinely in more and more patients. LASIK corrects myopia by altering the thickness and curvature of the central cornea. Myopia has been implicated as one risk factor in the development of glaucoma (1). The procedures for diagnosis and treatment of glaucoma are visualization of Correspondence to : Ruangvaravate N, Department of Ophthalmology, Faculty of Medicine, Siriraj Hospital, Mahidol University, Bangkok 10700, Thailand. Phone: 0-2419-8033-4, 01-820-8610, E-mail: sinrv@mahidol.ac.th the optic nerve head, visual field examination and measurement of the intraocular pressure (IOP). However, visual field and optic nerve head examination are difficult in some myopic eyes. In these cases, accurate evaluation of IOP is of particular relevance. Goldmann applanation tonometry is considered the gold standard for measuring IOP. It is based on the modified Imbert-Fick law, which takes into account the corneal rigidity force, affected by central corneal thickness. Variations in central corneal thickness (CCT) should yield variable corresponding IOP. The relationship between IOP and CCT has been confirmed (2,3). Hansen (4) reported with normotensive eyes a positive linear correlation existed between 340 J Med Assoc Thai Vol. 88 No.3 2005

CCT and IOP measured by Goldmann applanation tonometry. The other study demonstrated a reduced corneal thickness in a group of patients with lowtension glaucoma (2). The present study correlated with the study in ocular hypertensive patients, the corneal thickness is greater than normal leads to artificially high estimations of IOP (5). Therefore, the corneal thickness has a major influence and should be taken into consideration when evaluating Goldmann applanation tonometry reading (6,7). The meta-analysis confirms that eyes with low CCT values can result in low tonometry readings and high CCT values can result in elevated tonometry readings by Goldmann applanation tonometry (10). Therefore, LASIK can change the accuracy of intraocular pressure measurements by Goldmann applanation tonometry. This prospective study evaluates the changes that occur in IOP measurements by Goldmann applanation tonometry after LASIK for myopia and myopic astigmatism in a Thai population. Material and Method Sixty-five eyes of 33 myopic or myopic astigmatism patients who had LASIK in one or both eyes were enrolled in the present study. Patients selected for the study met the inclusion criteria including myopic or myopic astigmatism patients without any active ocular disease, received LASIK at Siriraj Hospital and accepted for joining the study. All patients signed their individual informed consent forms. Exclusion criteria included patients who had contraindication for LASIK and who used topical steroid for more than 4 weeks - to avoid steroidinduced ocular hypertension. Complete eye examinations including visual acuity measurement, manifest refraction, CCT measurements by pachymetry (OBSCAN), and IOP measurements by Goldmann applanation tonometer were performed before and 3 months after LASIK. LASIK was performed in all patients by the staff in the refractive surgery unit of Siriraj Hospital. A corneal flap was created with microkeratome (Hansatome or Automated Corneal Shaper). Photoablative keratectomy was applied to the bed of the cornea with Technolas excimer laser. All of the patients received antibiotic eye drops and steroid eye drops (fluorometholone, 1% methylprednisolone) or combine antibiotic and steroid eyedrops (Spersadexolin, Dexamytrex) after LASIK range from 1-4 weeks (mean 1.7 weeks). Data were analyzed with descriptive statistics (mean + SD and range) the paired t test. The Pearson product-moment correlation coefficient was used to access the relationship between the reduction of IOP readings and the mean reduction of the central corneal thickness. Results Thirty-three patients including 17 females (52.3%) and 16 males (47.7%), with a mean age of 30.9 + 7.1 years (range 16-42 years) were studied. The uncorrected visual acuity of almost all patients before LASIK procedure was 20/200 or worse. The mean preoperative spherical equivalent was -5.1 + 2.2 D (range -1.5 to -11.5 D). After LASIK procedure, almost all patients had uncorrected visual acuity of 20/25 or better. The mean postoperative spherical equivalent was -0.2 + 0.5 D (range -1.25 to 1.75 D). The mean preoperative CCT was 523.7 + 24.0 microns (range 483 to 583 microns). The mean postoperative CCT was 450.9 + 43.3 microns (range 362 to 537 microns). The mean reduction of CCT after LASIK was 73.6 + 36.2 microns (range 13 to 147 microns). The mean preoperative IOP was 13.6 + 2.6 mmhg (range 8 to 21 mmhg) and the mean postoperative IOP was 10.7 + 2.2 mmhg (range 6 to 18 mmhg) (Table 1). The mean reduction of IOP was 2.9 + 2.5 mmhg (range 5 to 8 mmhg). Mean IOP pre- and post LASIK showed a significant difference (p = 0.0001). Then, the patients were classified into 3 subgroups depending on the degree of myopia (group 1: myopia less than -3 D, group 2: Table 1. The preoperative and postoperative data (n = 65 eyes) Preoperative Postoperative Age (years) 30.9+7.1 (16-42) SE (diopters) -5.1+2.2 (-1.5 to -11.5) -0.2+0.5 (-1.25 to 1.75) CCT (microns) 523.7+24.0(483-583) 450.9+43.3(362-537) IOP (mmhg) 13.6+2.6 (8-21) 10.7+2.2 (6-18)** SE = spherical equivalent, CCT = central corneal thickness, IOP = intraocular pressure ** p = 0.0001 J Med Assoc Thai Vol. 88 No.3 2005 341

myopia -3 to -6 D and group 3: myopia greater than -6 D) for subgroup analysis. There were 14, 31, 20 eyes in each group respectively. The more myopia the patients had, the more the reduction in CCT was. The reduction of IOP correlated with the reduction of the CCT. So the patients in group 3 had the lowest reduction of IOP (3.3 +/- 1.9 mmhg). However, the IOP was significantly reduced in all 3 groups (p = 0.018, p < 0.001, p < 0.001 respectively) (Table 2). The Pearson product-moment correlation coefficient was used to assess the relationship between the reduction of IOP readings and the reduction of central corneal thickness. There was a low correlation between the reduction in IOP values measured by Goldmann applanation tonometry and the reduction of CCT (Pearson r = 0.09, p = 0.481). When this correlation was done in subgroup analysis, the result was the same (Table 3), but there was a tendency of correlation in group 3 that had more myopia than others. Discussion IOP measurement is an important part of ophthalmic practice; it is essential for screening, early diagnosis and management of many eye diseases especially glaucoma. The most common way to assess IOP is by Goldmann applanation tonometer, which gained widespread and rapid acceptance after its introduction in the 1950s. Goldmann applanation tonometry s status as a gold standard went largely unchallenged for 50 years, even though Professor Goldmann himself recognized various potential sources of error for the device in his first description Table 2. The change of IOP and CCT in each subgroup No. CCT change IOP change of yrs (microns) (mmhg) < -3 D 14 44.3+27.4 2.0+2.8 p = 0.018-3 to -6 D 31 65.1+26.7 3.2+2.6 p < 0.001 > -6 D 20 110.1+25.2 3.3+1.9 p < 0.001 CCT = central corneal thickness, IOP = intraocular pressure, D = diopters Table 3. The Pearson correlation coefficient in each subgroup Group r p < -3 D -0.102 0.730-3 to -6 D -0.098 0.611 > -6 D 0.158 0.532 D = diopters of his tonometer. In particular, Goldmann and Schmidt acknowledged that their design assumptions were based on a CCT of 500 microns and that the accuracy of their device would vary if CCT deviated from this value (8). The mean central corneal thickness in healthy subjects are 520 microns with a Gussian distribution (4). Many investigators have shown that CCT may exert a large influence on applanation measurements; thin cornea produces a false low reading, as does a thick cornea swollen with edema. In contrast, thick cornea secondary to increased collagen content results in false-high reading (9). Ehlers and co-workers (9) calculated from their linear regression equation that at a true IOP of 20 mmhg, applanation tonometry would produce a mean underestimation of 5.2 mmhg in eyes with a corneal thickness of 450 microns and on overestimation of 4.7 mmhg if the corneal thickness were 590 microns. A mean error of approximately 5 mmhg in the Goldmann applanation tonometer reading at a true IOP of 20 mmhg for each 70 microns change in the corneal thickness. Whitacre and coworkers (3) predicted a smaller mean error of 3.5 mmhg for a decrease in corneal thickness of 70 microns. Doughty and Zaman (10) found from linear regression model predicted that a 10% difference in CCT would result in 1.1 + 0.6 mmhg difference in IOP measures by Goldmann applanation tonometer (p < 0.05). Singh and co-workers (11). assessed the correlations between CCT and tonometry in eyes with normal-pressure glaucoma and primary open angle glaucoma. They found that CCT was lowest in normalpressure glaucoma, highest in eyes with ocular hypertension with a 0.2 mmhg change per 10 microns variation in central corneal thickness. Refractive surgery procedures (Photorefractive keratectomy - PRK, LASIK) leading to the decrease of CCT. In all eyes after laser surgery procedures, IOP was lower because of the decrease of CCT (12). Many studies have shown a decrease in IOP reading from Goldmann applanation tonometer. Mardelli and co-workers (7) showed in their retrospective study of a correlation between the reduction of CCT after PRK and IOP measurements; 23 + 23 microns reduction in the CCT associated with a 0.5 + 2.4 mmhg mean reduction in tonometry reading. Thinning of the cornea after LASIK affect tonometry readings. IOP difference before and after LASIK was statistically significant (13). In 85 eyes, Emara and co-workers (14) measured a mean decrease of 2.5 + 2.7 mmhg after LASIK. They found a significant correlation between IOP after LASIK and postoperative central corneal thickness 342 J Med Assoc Thai Vol. 88 No.3 2005

(p < 0.002), but there was no significant correlation between the decrease in IOP and the decrease in central corneal thickness. Fournier et al (15) reported a mean decrease of 1.9 + 2.9 mmhg in a retrospective study of 145 patients who received LASIK. Rashad et al (16) found a mean decrease in Goldmann applanation tonometry of 3.69 + 1.63 mmhg after LASIK for myopia, although the relationship between IOP and CCT was not assessed. Montes-Mico and Charman (17) found the reduction in measured IOP following refractive surgery (PRK,LASIK), by about 0.5 mmhg/ D of myopic correction. No statistically significant differences were found between the results obtained following the two types of surgery. Duch and coworkers (18) showed the correlation between CCT and IOP measured with Goldmann applanation tonometer by regression analysis was a decreased 2.9 mmhg per 70 microns reduction in CCT. In the present study, the mean IOP measured by Goldmann applanation tonometer decreased 2.9 + 2.5 mmhg after 3 months post-lasik, this was a statistically significant difference (p = 0.0001) (Table 1). However, there was no correlation between the mean IOP reduction by Goldmann applanation tonometry and the reduction of CCT by Pearson analysis, this may be attributed to small sample size in the present study (Table 2). But after the authors analyzed in subgroups depending on the degree of myopia, it was found that there was a tendency of correlation in the higher myopic group (Table 3). In conclusion, the Goldmann applanation tonometer simply measures the force needed to deform the cornea in a standardized manner. IOP is derived from the force measurement indirectly, based on a number of assumptions about corneal deformability. Corneal deformability in turn represents a summation of the corneal curvature, elastic properties, surface tension, and the IOP. Central corneal thickness is probably the major component of corneal elasticity. Therefore, IOP is overestimated for thicker corneas and underestimated for thinner values. Currently, there is no consensus on the magnitude of IOP measurement error with thin or thick corneas. It is not known whether this relation is linear or nonlinear across the range of corneal thickness in the general population. It is also not known whether the iatrogenically thinned cornea after LASIK has the same implications in terms of tonometric artifact as do naturally thin corneas. However, there is an increasing need for awareness of underestimation of IOP measured by Goldmann applanation tonometer after LASIK. This has important implications for myopic patients who undergo LASIK surgery and may be at risk to develop glaucoma. Their pressures will be read as normal or low because the cornea is thin as a result of LASIK. The authors suggest measuring the CCT and IOP by Goldmann applanation tonometer pre- and post- LASIK in all cases to avoid a misleading diagnosis of early glaucoma in this group of patients in the future. References 1. Thomas JV. Primary open angle glaucoma. In: Albert DM, Jakobiec FA, eds. Principles and Practice of Ophthalmology. Philadelphia: Saunders, 1994: 1342-5. 2. Ehlers N, Hansen FK. Central corneal thickness in lowtension glaucoma. Acta Ophthalmol 1974; 52: 740-6. 3. Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993; 115: 592-6. 4. Hansen FK. A clinical study of the normal human central thickness. Acta Ophthalmol 1971; 49: 82-9. 5. Argus WA. Ocular hypertension and central corneal thickness. Ophthalmology 1995; 102: 1810-2. 6. Chatterjee A, Shah S, Bessant DA, et al. Reduction in intraocular pressure after Excimer Laser Photorefractive Keratectomy. Ophthalmology 1997; 104: 355-9. 7. Mardelli PG, Piebenga LW, Whitacre MM, et al. The effect of Excimer Laser Photorefractive Keratectomy on Intraocular Pressure Measurements Using the Goldmann applanation tonometer. Ophthalmology 1997; 104: 945-8. 8. Brandt JD. Corneal thickness in glaucoma screening, diagnosis, and management. Current Opinion in Ophthalmology 2004; 15: 85-9. 9. Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol 1975; 53: 34-43. 10. Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach. Surv Ophthalmol 2000; 44: 367-408. 11. Singh RP, Goldberg I, Graham SL, et al. Central corneal thickness,tonometry and Ocular Dimensions in Glaucoma and Ocular Hypertension. J Glaucoma 2001; 10: 206-10. 12. Filipecka I,Gierek-Ciaciura S.Influence of corneal thickness changes after refractive surgery on intraocular pressure measurements. Klin Oczna 2001; 103: 25-7. 13. Arimoto A, Shimizu K, Shoji N, et al. Underestimation of intraocular pressure after LASIK. Nippon Ganka Gakkai Zasshi 2001; 105: 771-5. 14. Emara B, Probst LE, Tingey DP, Kennedy DW, Willms LJ, Machat J. Correlation of Intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ kera-tomileusis. J Cataract Refract J Med Assoc Thai Vol. 88 No.3 2005 343

Surg 1998; 24: 1320-5. 15. Fournier AV, Podtetenev M, Lemier J, et al. Intraocular pressure change measured by Goldmann tonometry after laser in situ keratomileusis. J Cataract Refract Surg 1998; 24: 905-10. 16. Rashad KM, Bahnassy AA. Changes in Intraocular Pressure after Laser in Situ Keratomileusis. J Cataract Refract Surg 2001; 17: 420-7. 17. Montes-Mico R, Charman WN. Intraocular pressure after excimer laser myopic refractive surgery. Ophthalmic Physiol Opt 2001; 21: 228-35. 18. Duch S, Serra A, Castanera J, et al. Tonometry after Laser in Situ Keratomileusis Treatment. J Glaucoma 2001; 10: 261-5. 19. Whitarce MM, Stein R. Sources of error with the use of Goldmann type tonometers. Surv Ophthalmol 1993; 38: 1-30. 20. Goldmann H. Applanation tonometry. In: Newell FE.ed. Glaucoma: Transactions of the second conference. New York: Josiah Macy, Jr Foundation, 1957: 167-220. 21. El Danasoury MA, El Maghraby A, Coorpender SJ. Change in intraocular pressure in myopic eyes measured with contact and non-contact tonometers after laser in situ keratomileusis. J Cataract Refract Surg 2001; 17: 97-104. 22. Faucher A, Gregoire J, Blondeau P. Accuracy of Goldmann tonometry after refractive surgery. J Cataract Refract Surg 1997; 23: 832-8. 23. Park HJ, Uhm KB, Hong C. Reduction in intraocular pressure after laser in situ keratomileusis. J Cataract Refract Surg 2001; 27: 303-9. การว ดความด นตาภายหล งการทำเลส ค: การศ กษาในคนไทย งามแข เร องวรเวทย, อต พร ตวงทอง, พน ดา โกส ยร กษ วงศ, ภ ญน ตา ประภาสะว ต, ว ภาว บ รณพงศ, สบง ศร วรรณบ รณ ว ตถ ประสงค : เพ อศ กษาการเปล ยนแปลงของค าความด นตา โดยการว ดด วยเคร องว ด Applanation ในผ ป วย ท ได ร บการทำเลส คเพ อร กษาสายตาส นและสายตาเอ ยง และศ กษาความส มพ นธ ระหว างการเปล ยนแปลงของ ค าความด นตา และการลดลงของความหนาของกระจกตาในผ ป วยไทย ว ธ การศ กษา: การศ กษาแบบไปข างหน า ว สด และว ธ การ: ผ ป วย 33 ราย (65 ตา) ท ได ร บการทำเลส ค เพ อแก ไขสายตาส นและสายตาเอ ยง ได ร บการตรวจ ตาอย างละเอ ยด และว ดความด นตาก อนและหล งการทำเลส ค เป นเวลา 3 เด อน โดยด ความส มพ นธ ก บค าความหนา ของกระจกตาก อนและหล งทำเลส ค ผลการศ กษา: ค าความด นตาท ว ดได หล งทำเลส คลดลงเฉล ย 2.9 + 2.5 มม. ปรอท (p = 0.0001) และเม อศ กษาถ ง ความส มพ นธ ระหว างการลดลงของค าความด นตา ก บการลดลงของความหนากระจกตา โดยแบ งผ ป วยเป น 3 กล ม กล ม 1 สายตาส นน อยกว า -3 ไดออฟเตอร กล ม 2 สายตาส น -3 ถ ง -6 ไดออฟเตอร และกล ม 3 สายตาส นมากกว า -6 ไดออฟเตอร พบว าม แนวโน มท จะม ความส มพ นธ ก นในกล มท ม สายตาส นมากกว า -6 ไดออฟเตอร ข นไป สร ป: การว ดความด นตาด วยเคร อง Applanation จะให ค าท ต ำลงในผ ท ม กระจกตาบาง ด งน น การว ดความด นตา ด วยเคร อง Applanation เพ อใช ในการว น จฉ ยโรคต อห น ควรคำน งถ งค าความหนาของกระจกตาร วมด วย โดยเฉพาะผ ป วยท ได ร บการทำเลส ค 344 J Med Assoc Thai Vol. 88 No.3 2005