Opaque Bubble Layer Risk Factors in Femtosecond Laser-assisted LASIK



Similar documents
Alain Saad, MD, Alice Grise-Dulac, MD, Damien Gatinel, MD, PhD

Femtosecond Lasers in LASIK Surgery

Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps

Simultaneous Correction of Unilateral Rainbow Glare and Residual Astigmatism by Undersurface Flap Photoablation After Femtosecond Laser-Assisted LASIK

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US

Uniquely Safe. predictably better for our patients. enhancement, may be significantly reduced.

Total Corneal Astigmatism Correction With Limbal Relaxing Incisions Made by Femtosecond Laser System

Retreatment by Lifting the Original Laser in Situ Keratomileusis Flap after Eleven Years

How To Implant A Keraring

Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking

Consumer s Guide to LASIK

Initial clinical experience with the FS200 Femto and EX500 excimer

Refractive Surgery Issue. Inlays and Presbyopia: On the Horizon P. 24. Crack a SMILE or Raise a Flap? P. 30. LASIK Xtra: Who Should Get It? P.

Overview of Refractive Surgery

Femto-LASIK. Pulsewidth: Ultrashort-pulse micro- machining can make sub- wavelength holes. micromachining

OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment

Diego Fernando Suárez Sierra, MD Fellow Cornea and Refractive Surgery Fellow Lens and Ocular Surface Vejarano Laser Vision Center

How do you adjust for IOP in a LASIK patient?

Validation of a New Scoring System for the Detection of Early Forme of Keratoconus

Richard S. Hoffman, MD. Clinical Associate Professor of Ophthalmology Oregon Health & Science University

Conductive keratoplasty (CK) utilizes radiofrequency energy. Original Article

Common Co-management Questions

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

When refractive surgeons began to understand

Minimally Invasive Surgery: Femtosecond Lasers and Other Innovative Microsurgical Techniques

LASIK Surgery. One Use-Plus. The Moria Option for SBK

ALL-LASER LASIK USING THE ZEISS VisuMax MEL 80 REFRACTIVE WORKSTATION

Case Report Laser Vision Correction on Patients with Sick Optic Nerve: A Case Report

Ectasia after laser in-situ keratomileusis (LASIK)

Accelerated Refractive Performance

Crosslinking and Long-Term Hyperopic LASIK Stability Initial Clinical Findings in Contralateral Eye Study

NEW HORIZONS IN CORNEAL SURGERY VERSATILE FEMTOSECOND LASER WORKSTATION WE FOCUS ON PERFECTION

Long-term stability of the posterior cornea after laser in situ keratomileusis

Ophthalmology Department,Tanta University, Egypt

Lasik Xtra in: Hyperopia AK Xtra Clear cornea cataract surgery

Effect of Preoperative Keratometric Power on Intraoperative Complications in LASIK in 34,099 Eyes

Comparing Femtosecond Lenticule Extraction (FLEx) and Femtosecond Laser In-situ Keratomileusis (LASIK) for Myopia and Astigmatism

One Use-PlusSBK versus LDV Femto Laser Clinical evaluation

The concept of a proactive intervention involving in situ

Collagen Cross-linking combined with PRK and LASIK

Predictability of Corneal Flap Thickness in LASIK Using the WaveLight FS200. Femtosecond Laser

Predictability of corneal flap thickness in laser in situ keratomileusis using a 200 khz femtosecond laser

ANGLE KAPPA. Copyrighted material. Not for distribution. chapter. Anastasios John Kanellopoulos, MD

Refractive Surgery. Evolution of Refractive Error Correction

Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday?

Wavefront-guided versus standard laser in situ keratomileusis in low to moderate myopia

Anterior Elevation Maps as the Screening Test for the Ablation Power of Previous Myopic Refractive Surgery

Intacs for keratoconus Yaron S. Rabinowitz

Ectasia Risk Factors: Lessons Learned

Alexandria s Guide to LASIK

LASIK. Complications. Customized Ablations. Photorefractive Keratectomy. Femtosecond Keratome for LASIK. Cornea Resculpted

Anastasios John Kanellopoulos, M.D., Clinical Professor of Ophthalmology in NYU Medical School, Director of the Laser Vision Institute, Athens, Greece

Wavefront technology has been used in our

Intracorneal Ring Segments Implantation Followed by Same-day Topography-guided PRK and Corneal Collagen CXL in Low to Moderate Keratoconus

Detection of static cyclotorsion and compensation for dynamic cyclotorsion in laser in situ keratomileusis

Comparison of Two Procedures: Photorefractive Keratectomy Versus Laser In Situ Keratomileusis for Low to Moderate Myopia

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

SMART TEC NANOTECHNOLOGY LASER

IntraLase Corp. Clinical Studies Fact Sheet

Factors Affecting Long-term Myopic Regression after Laser In Situ Keratomileusis and Laser-assisted Subepithelial Keratectomy for Moderate Myopia

INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the

The pinnacle of refractive performance.

Wavefront-guided Enhancements Using the WaveLight Excimer Laser in Symptomatic Eyes Previously Treated With LASIK

Laser in situ keratomileusis (LASIK) has been. Retreatment of Hyperopia After Primary Hyperopic LASIK REPORTS

ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography

LASIK Eye Surgery Report

Comparison of higher-order aberrations after wavefront-guided laser in situ keratomileusis and laser-assisted subepithelial keratectomy

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay

Collagen cross-linking should be done separately. Canan Asli Utine, MD, MSc, FICO Yeditepe University, Istanbul, Turkey

Wavefront Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography

PATIENT CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)

Bascom Palmer Eye Institute. LASIK and Your Vision Correction

By Dr Waleed Al-Tuwairqi, MD Dr Omnia Sherif, MD Ophthalmology Consultants, Elite Medical & Surgical Center Riyadh -KSA.

Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY

RELEX SMILE AND SMILE EXTRA.. OUR 1 YEAR RESULTS AND PATIENTS SURVEY

Femtosecond Laser versus Mechanical Keratome LASIK for Myopia

LASIK for Hyperopia With the WaveLight Excimer Laser

KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,

Cornea and Refractive Surgery Update

A Prospective, Contralateral Eye Study Comparing Thin-Flap LASIK (Sub-Bowman Keratomileusis) with Photorefractive Keratectomy

Advanced personalized nomogram for myopic laser surgery: First 100 eyes

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM

Femtosecond LASIK: Practical Pearls From Five Years of Experience. Copyrighted material. Not for distribution. chapter

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation

Refractive Surgery. Common Refractive Errors

Long-Term Outcomes of Flap Amputation After LASIK

ICRS implantation with the Femto LDV laser in stabilized KC patients: 6 months results

Laser Vision Correction

LASIK & Refractive Surgery

Refractive Errors & Correction

Corneal flap thickness with the Moria M2 single-use head 90 microkeratome

Intraoperative Complications of Laser in Situ Keratomileusis in Yazd

Correction of Myopic Astigmatism With Small Incision Lenticule Extraction

Excimer Laser Refractive Surgery

Original Article. Outcomes of Retreatment after Aborted Laser In Situ Keratomileusis due to Flap Complications

Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System

XXXII nd Congress of the ESCRS, London, September 13, 2014 Instructional Course # 7. LASIK: basic steps for safety and great results

Therapeutic Flap Amputation for Atypical LASIK Flap and Interface Abnormalities

Refractive Change Induced by the LASIK Flap in a Biomechanical Finite Element Model

Epi-LASIK Surgery. The procedure of choice for surface ablation.

Transcription:

ORIGINAL ARTICLE Opaque Bubble Layer Risk Factors in Femtosecond Laser-assisted LASIK Romain Courtin, MD; Alain Saad, MD; Emmanuel Guilbert, MD; Alice Grise-Dulac, MD; Damien Gatinel, MD ABSTRACT PURPOSE: To determine the characteristics and risk factors for occurrence of opaque bubble layer (OBL) during femtosecond laser-assisted flap creation for LASIK. METHODS: One hundred ninety-eight eyes of 102 consecutive patients who underwent LASIK flap creation performed with the Alcon WaveLight FS200 laser (Alcon Laboratories, Inc., Fort Worth, TX) were retrospectively analyzed in a cohort study. Preoperative manifest refraction, corneal keratometry, central corneal thickness, white-to-white corneal diameter, corneal hysteresis, corneal resistance factor, and programmed flaps parameters were collected. Digital images automatically recorded after flap creation were analyzed to measure OBL areas. Correlation tests were performed between preoperative corneal parameters and OBL areas. RESULTS: The incidence rate of OBL was 48% (103 eyes). The mean OBL area as a percentage of the corneal flap area in the OBL group was 4.25% ± 7.16% (range: 0% to 32.9%). The central corneal thickness, corneal resistance factor, and corneal hysteresis were significantly positively correlated with the OBL area (r = 0.242, P =.001; r = 0.254, P =.028; and r = 0.351, P <.0001, respectively). Corneal hysteresis and OBL area were positively correlated, independently of the central corneal thickness and other confounder factors with standardized coefficient (r = 0.353 ± 0.227, P =.002). CONCLUSIONS: This study confirms the already known OBL risk factors with a larger cohort and suggests for the first time that an elevated corneal hysteresis is an independent predictive risk for OBL occurrence. [J Refract Surg. 2015;31(9):608-612.] L ASIK flap creation is the most used femtosecond laser application in refractive surgery. Compared to the microkeratome, the femtosecond laser technique has equivalent visual and refractive outcomes. 1-4 However, this technology increases predictability, 5 accuracy, and precision 6 of LASIK flap creation. Also, it allows for flap customization (thickness, diameter, side-cut angle) and is better suited for thin corneas, high spherical ametropia, and high astigmatisms with a fast learning curve. 7 Despite the improved safety gained by using femtosecond laser in flap creation, specific complications exist including the occurrence of opaque bubble layer (OBL). Femtosecond pulses are focused in the corneal tissue where it causes covalent bonds to rupture between the atomic nucleus and electrons (ionization). Plasma expansion in the tissue creates a cavitation bubble causing a separation between stromal lamellae. OBL is the accumulation of gas bubbles temporarily detained in the intrastromal interface, creating transient opacity. The mechanism of OBL is not clearly understood. However, it is suggested that the gas bubbles infiltrate the stroma because they cannot escape and due to corneal compression with a high level of vacuum created by applanation. Histopathology of OBL has not been described because of the transient nature of the phenomenon. In vivo structural characteristics with ultra-high-resolution optical coherence tomography has allowed a more detailed structural analysis without providing further explanation of the mechanism. 8 Although OBL formation is rarely dangerous, it represents a frequent and sometimes inconvenient complication. The purpose of this study was to determine the incidence From the Department of Anterior Segment and Refractive Surgery, Rothschild Foundation, Paris, France (RC, AS, EG, AG-D, DG); Laser Vision Institute Noémie de Rothschild, Paris, France (RC, AS, EG, AG-D, DG); and the Department of Ophthalmology, University Hospital, Clermont-Ferrand, France (RC). The authors have no financial or proprietary interest in the materials presented herein. Correspondence: Damien Gatinel, MD, Department of Anterior Segment and Refractive Surgery, Fondation Ophtalmologique A. de Rothschild, 25-29 rue Manin, 75019 Paris, France. E-mail: gatinel@gmail.com doi:10.3928/1081597x-20150820-06 608 Copyright SLACK Incorporated

TABLE 1 Femtosecond Laser Flap Creation Settings of the Alcon WaveLight FS200 Setting Flap Diameter (range) Measurement 7.70 to 9.50 mm Thickness (range) 120.00 to 135.00 µm Canal Length (range) Width Bed cut 0.30 to 1.20 mm 1.70 mm Spot separations 8.0 µm Line separations 8.0 µm Pulse energy (range) 0.79 to 0.85 µj Side cut Spot separations 5.0 µm Line separations 3.0 µm Angle 70 Pulse energy (range) 0.79 to 0.85 µj Hinge Position 90 Length 3.8 mm Angle 50 Width 0.4 mm The Alcon Wavelight FS200 is manufactured by Alcon Laboratories, Inc., Fort Worth, TX. and risk factors of OBL occurrence using the Alcon WaveLight FS200 laser (Alcon Laboratories, Inc., Fort Worth, TX) for current femtosecond laser-assisted LASIK procedures. PATIENTS AND METHODS Patients and Femtosecond Laser Settings The retrospective study included 102 successive patients (198 eyes) who had femtosecond laser-assisted LASIK surgery for myopia, astigmatism, or hyperopia. Interventions were performed by two surgeons (DG and AS) using the same operating methods from January to December 2013. All procedures were performed with the WaveLight FS200 laser. The study protocol was approved by the local research ethics committee. Preoperative data analyzed were manifest refraction (sphere, cylinder, and spherical equivalent). The corneal keratometry, central corneal thickness (CCT), and white-to-white corneal diameter were measured with the ORBSCAN II Corneal Topographer (Bausch & Lomb, Orbtek, Inc., Salt Lake City, UT). The corneal Journal of Refractive Surgery Vol. 31, No. 9, 2015 Figure 1. Screenshot of delimitation and extraction using Preview software (Apple, Inc., Cupertino, CA). biomechanical parameters, corneal hysteresis (CH), and corneal resistance factor (CRF) were measured with the Ocular Response Analyzer (Reichert Ophthalmic Instruments, Buffalo, NY). For femtosecond laser flap creation, the only set parameters were flap diameter, flap thickness, and canal length. Canal width and flap bed cut, side cut, and hinge were WaveLight FS200 laser default settings and the same for all procedures (Table 1). Patients with missing or incomplete data were excluded from the study. Standardized OBL Measurement Procedure The OBL measurement procedure was highly reproducible to ensure the best repeatability of OBL area quantification. Digital images automatically recorded and stored in the WaveLight FS200 storage system during the flap creation procedure were collected. Different software were used to analyze the images and evaluate the OBL areas. To obtain a reproducible method of measurement, each step was standardized. Step 1: Image extracting and editing (Preview software; Apple, Inc., Cupertino, CA). 1. Laser Alcon WaveLight FS200 flap creation: treatment parameters and procedure screenshot. 2. Screenshot image extraction. 3. Flap side-cut delimitation using the circular selection tool (Figure 1). Step 2: Image processing and analysis (ImageJ software; National Institutes of Health, Bethesda, MA). 1. Image color conversion to grayscale (8-bit) (Figure A1, available in the online version of this article). 609

TABLE 2 Preoperative Corneal Parameters Analyzed Parameter Minimum Maximum Mean SD CCT (µm) 492.00 632.00 570.32 28.39 K1 (D) 40.30 47.20 44.27 1.37 K2 (D) 40.00 46.80 43.20 1.42 Sphere (D) -11.50 5.50-3.53 3.48 Cylinder (D) -5.00 0.00-0.92 0.93 SE (D) -12.12 4.62-3.99 3.45 CRF 6.70 15.00 11.28 1.60 CH 6.70 14.80 11.84 1.39 WW (mm) 10.70 13.10 11.84 0.39 SD = standard deviation; CCT = central corneal thickness; K1 = flat keratometry; K2 = steep keratometry; D = diopters; SE = spherical equivalent; CRF = corneal resistance factor; CH = corneal hysteresis; WW = white-towhite distance 2. Standardized contrast adjusting. The optimal adjustment was previously subjectively defined (Figure A2). 3. Image binarization (Figure A3). 4. The calculation tool allowed defining the percentage of black area in the flap area (Figure A4). The flap area was defined by the WaveLight FS200 settings because several studies showed an excellent accuracy and precision of flap size. 6,9 Then, knowing the flap area in mm 2 and the OBL percentage of the flap, OBL areas were calculated using a simple rule of three. Statistical Analysis All statistical analyses were performed with XL- STAT software (Addinsoft, Paris, France). Data were expressed as the mean ± standard deviation. The Mann Whitney test was used for preoperative comparisons, and logistic regression was used for the subsidiary analysis. The Pearson correlation was used to express the relationship between the size of the OBL and the measured data. A P value less than.05 was considered statistically significant. RESULTS In our study of 198 eyes, 101 were right eyes and 97 were left eyes. Of the 102 patients, 62.7% (64 patients) were female. The mean age was 34.20 ± 9.83 years (range: 20 to 69 years). The incidence rate of OBL was 48% (95 eyes). The mean OBL area in the OBL group was 8.86% ± 8.09% (range: 0% to 32.90%) of the flap area. The preoperative corneal parameters are shown in Table 2. TABLE 3 Comparison of the Main Parameters Between the Two Groups Parameter No OBL (n = 103) OBL (n = 95) P OBL area (%) 0 8.86 ± 8.09 Age 35.33 ± 11.19 32.80 ± 7.81.527 Corneal parameters CCT (µm) 563.87 ± 28.75 577.32 ± 26.40.001 K1 (D) 44.20 ± 1.46 44.34 ± 1.26.577 K2 (D) 43.17 ± 1.51 43.24 ± 1.31.739 Sphere (D) -3.18 ± 3.46-3.91 ± 3.48.088 Cylinder (D) -0.88 ± 0.83-0.97 ± 1.02.945 SE (D) -3.62 ± 3.52-4.40 ± 3.34.092 CRF 11.07 ± 1.64 11.50 ± 1.52.028 CH 10.99 ± 1.28 11.85 ± 1.37 <.0001 WW (mm) 11.89 ± 0.42 11.78 ± 0.36.093 Flap parameters Flap area (mm 2 ) 82.22 ± 4.54 82.11 ± 4.37.761 Flap thickness (mm) 121.84 ± 4.08 122.00 ± 4.02.702 Canal width (mm) 1.70 ± 0.00 1.70 ± 0.00 1.00 Canal length (mm) 0.80 ± 0.19 0.77 ± 0.18.183 OBL = opaque bubble layer; CCT = central corneal thickness; K1 = flat keratometry; K2 = steep keratometry; D = diopters; SE = spherical equivalent; CRF = corneal resistance factor; CH = corneal hysteresis; WW = whiteto-white distance For the preoperative corneal biomechanical parameters, the mean CRF and CH were 11.28 ± 1.60 mm Hg (range: 6.70 to 15.00 mm Hg) and 11.40 ± 1.39 mm Hg (range: 6.70 to 14.80 mm Hg), respectively. The mean flap diameter, flap thickness, and canal were 9.06 ± 0.32 mm (range: 7.80 to 9.50 mm), 121.92 ± 4.04 µm (range: 120 to 135 µm), and 0.78 ± 0.18 mm (range: 0.30 to 1.20 mm), respectively. The mean flap area was 82.17 ± 4.45 mm 2 (range: 69.60 to 90.25 mm 2 ). The CCT, CRF, and CH were statistically different between the no OBL and the OBL group. There was no significant difference between the two groups for flap creation settings (Table 3). Pearson correlation tests were performed between preoperative corneal parameters and the OBL area. The CCT, CRF, and CH were positively correlated with the OBL area (r = 0.242, P <.001; r = 0.254, P <.001; and r = 0.351, P <.001, respectively). Thicker corneas and higher corneal biomechanical values have a significantly higher risk of developing OBL. None of the flap parameters were correlated with a lower or higher risk of OBL. A linear regression analysis was performed and showed a positive correlation between CH and OBL 610 Copyright SLACK Incorporated

TABLE 4 Standardized Correlation Coefficients of Corneal and Flap Parameters With the OBL Area Variable Corneal parameters Standardized Coefficients Standard Error t P Lower 95% CI Upper 95% CI CCT 0.175 0.084 2.094.038 0.010 0.340 K1-0.066 0.243-0.273.785-0.546 0.413 K2 0.009 0.253 0.038.970-0.490 0.509 Sphere 0.061 0.098 0.622.535-0.133 0.255 Cylinder -0.069 0.126-0.547.585-0.317 0.179 Spherical equivalent -0.020 0.070-0.291.771-0.158 0.117 CRF -0.081 0.115-0.705.482-0.309 0.146 CH 0.353 0.115 3.069.002 0.126 0.580 WW 0.030 0.082 0.364.716-0.131 0.191 Flap parameters Flap diameter 0.052 0.092 0.563.574-0.130 0.233 Flap area 0.011 0.100 0.114.910-0.186 0.208 Flap thickness -0.130 0.094-1.383.168-0.314 0.055 Canal length -0.019 0.073-0.254.800-0.163 0.126 95% CI = 95% confidence interval; CCT = central corneal thickness; K1 = flat keratometry; K2 = steep keratometry; CRF = corneal resistance factor; CH = corneal hysteresis; WW = white-to-white distance area independent of the CCT and other confounder factors with standardized coefficient (r = 0.353 ± 0.227, P =.002) (Table 4). Journal of Refractive Surgery Vol. 31, No. 9, 2015 DISCUSSION In our study, the occurrence of OBL was frequent with an incidence of 48.0%. Excessive OBL may interfere with eye tracker pupil recognition. It also causes difficulties in flap lifting and intraoperative pachymetry measurements. These problems can result in longer interventions and possibly induce an undercorrection during the photoablation procedure due to a reduced fluence. In our experience, we have not encountered serious problems; however, after flap creation complicated with OBL we found subjectively that flaps were more difficult to lift. 10 Our study is concordant with the literature, because no serious complications have been reported after OBL occurrence. 10,11 Femtosecond lasers use near infrared (wavelength of 1,053 nm) with a femtosecond pulse duration (range: 10 to 15 seconds) producing photodisruption. Multiple pulses of the laser, with a pulse frequency of 200 Hz with the Alcon WaveLight FS200, generate an ionization resulting in formation of microscopic gas bubbles next to each other, dissipating and expanding into corneal tissue. This cutting process causes a separation between stromal lamellae with minimal collateral tissue damage, creating the LASIK flap. 12 Previous studies have described two different types of OBLs. 11,12,13 Soft, diffuse, or delayed OBLs have a more transparent appearance and occur later, after completion of the laser dissection in a particular area. Hard, advancing, or early OBLs appear earlier and have a denser appearance. 6 The preoperative and intraoperative characteristics comparison of groups with hard and diffuse OBL showed no statistically significant difference. 12 However, it is sometimes difficult to determine whether it is diffuse or hard OBL. In our study, we did not differentiate between the two types of OBLs. Our results confirm the correlations between CCT and OBL shown in previous studies 12,14 with a larger population of 198 eyes: thicker corneas were associated with more frequent and larger OBL. In the statistical methods, we decided to use fellow eyes so that our study can be compared to those already published in the literature, despite possible changes in relationships. To our knowledge, this is the first time in the literature that corneal biomechanical parameters are correlated with the occurrence of OBL. Incidence and importance of OBL have been positively correlated with high CH and high CRF. Furthermore, a linear correla- 611

tion between CH and the occurrence and importance of OBL was identified independently of the CCT and other confounder factors. However, these correlations are low, constituting the main limitation of this study. We know that the CH value is proportional to the degree of the corneal viscosity and inversely proportional to its degree of elasticity. Thus, we believe that the lower corneal elasticity and higher viscosity might increase the OBL occurrence due to the lower capacity for reversible deformation of the cornea with a greater gas bubble infiltration between stromal lamellae. Gas bubbles have been described in other locations, and may escape in corneal sub-epithelial space, subconjunctiva, or anterior chamber. 15-17 The WaveLight FS200 femtosecond laser can also create a gas bubble evacuation canal to minimize OBL. However, this canal may be associated with blood extending from the canal into the flap interface and may require a flap lift. 18 Based on our study, the occurrence of OBL with the Alcon WaveLight FS200 laser is frequent (42%) and usually innocuous. Thicker corneas and higher corneal biomechanical parameters are associated with an increased risk of OBL. In the future, it could be interesting to compare refractive and visual outcomes (halos, glare, and higherorder aberrations) between laser flap creation complicated with OBL and those without OBL. AUTHOR CONTRIBUTIONS Study concept and design (AS, DG); data collection (RC, AS, DG); analysis and interpretation of data (RC, AS, EG, AG-D, DG); writing the manuscript (RC); critical revision of the manuscript (AS, EG, AG-D, DG); statistical expertise (AS) REFERENCES 1. Patel SV, Maguire LJ, McLaren JW, Hodge DO, Bourne WM. Femtosecond laser versus mechanical microkeratome for LASIK: a randomized controlled study. Ophthalmology. 2007;114:1482-1490. 2. Chen S, Feng Y, Stojanovic A, Jankov MR 2nd, Wang Q. IntraLase femtosecond laser vs mechanical microkeratomes in LASIK for myopia: a systematic review and meta-analysis. J Refract Surg. 2012;28:15-24. 3. Gil-Cazorla R, Teus MA, de Benito-Llopis L, Mikropoulos DG. Femtosecond laser vs mechanical microkeratome for hyperopic laser in situ keratomileusis. Am J Ophthalmol. 2011;152:16-21. 4. Montés-Micó R, Rodríguez-Galietero A, Alió JL. Femtosecond laser versus mechanical keratome LASIK for myopia. Ophthalmology. 2007;114:62-68. 5. Ahn H, Kim JK, Kim CK, et al. Comparison of laser in situ keratomileusis flaps created by 3 femtosecond lasers and a microkeratome. J Cataract Refract Surg. 2011;37:349-357. 6. Sutton G, Hodge C. Accuracy and precision of LASIK flap thickness using the IntraLase femtosecond laser in 1000 consecutive cases. J Refract Surg. 2008;24:802-806. 7. Reinstein DZ, Carp GI, Archer TJ, Gobbe M. Transitioning from mechanical microkeratome to femtosecond laser flap creation: visual outcomes of an experienced and a novice LASIK surgeon. J Cataract Refract Surg. 2012;38:1788-1795. 8. Hurmeric V, Yoo SH, Fishler J, Chang VS, Wang J, Culbertson WW. In vivo structural characteristics of the femtosecond LASIK-induced opaque bubble layers with ultrahigh-resolution SD-OCT. Ophthalmic Surg Lasers Imaging. 2010;41:S109-S113. 9. Santhiago MR, Kara-Junior N, Waring GO. Microkeratome versus femtosecond flaps: accuracy and complications. Curr Opin Ophthalmol. 2014;25:270-274. 10. Kaiserman I, Maresky HS, Bahar I, Rootman DS. Incidence, possible risk factors, and potential effects of an opaque bubble layer created by a femtosecond laser. J Cataract Refract Surg. 2008;34:417-423. 11. Kanellopoulos AJ, Asimellis G. Essential opaque bubble layer elimination with novel LASIK flap settings in the FS200 Femtosecond Laser. Clin Ophthalmol. 2013;7:765-770. 12. Soong HK, Malta JB. Femtosecond lasers in ophthalmology. Am J Ophthalmol. 2009;147:189-197. 13. Kanellopoulos AJ, Asimellis G. Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique. Clin Ophthalmol. 2013;7:343-51. 14. Liu CH, Sun CC, Hui-Kang Ma D, et al. Opaque bubble layer: incidence, risk factors, and clinical relevance. J Cataract Refract Surg. 2014;40:435-440. 15. Ide T, Kymionis GD, Goldman DA, Yoo SH, O Brien TP. Subconjunctival gas bubble Formation during LASIK flap creation using femtosecond laser. J Refract Surg. 2008;24:850-851. 16. Srinivasan S, Herzig S. Sub-epithelial gas breakthrough during femtosecond laser flap creation for LASIK. Br J Ophthalmol. 2007;91:1373. 17. Srinivasan S, Rootman DS. Anterior chamber gas bubble formation during femtosecond laser flap creation for LASIK. J Refract Surg. 2007;23:828-830. 18. Au J, Krueger RR. Interface blood as a new indication for flap lift after LASIK using the WaveLight FS200 femtosecond laser. J Refract Surg. 2014;30:858-860. 612 Copyright SLACK Incorporated

1 2 3 4 Figure A. Steps of image processing and analysis using ImageJ software (National Institutes of Health, Bethesda, MD) after flap image extraction. (1) Conversion in 8-bit (grayscale). (2) Standardized contrast adjusting. (3) Binarization (black and white). (4) Measure of black area percentage.