1. Overview... A - 2. Intended Use... A - 2. Warnings... A - 2. AIRVO 2 and Accessories... A Setting up AIRVO 2... A - 4
|
|
|
- Barbara Wilcox
- 10 years ago
- Views:
Transcription
1 User Manual AIRVO 2
2 BEFORE YOU START This User Manual is intended for healthcare professionals. This User Manual applies to AIRVO 2 units with LOT numbers and above. Read this User Manual including all warnings. Failure to do so may result in injury. In addition, watch the AIRVO 2 Video Guide. Keep them both in a safe place for future reference. Before the AIRVO 2 is used for the first time, it must be set up according to the instructions in the AIRVO 2 Technical Manual. The AIRVO 2 must be cleaned and disinfected between patients according to the instructions in the Disinfection Kit Manual (900PT600). For further assistance, please contact your Fisher & Paykel Healthcare representative. English TABLE OF CONTENTS 1. Overview... A - 2 Intended Use... A - 2 Warnings... A - 2 AIRVO 2 and Accessories... A Setting up AIRVO 2... A Using AIRVO 2... A - 6 Target dew-point temperature... A - 7 Target flow... A - 7 Oxygen... A - 8 Alarms... A Reprocessing... A - 12 Schedule for changing accessories... A - 12 Filter replacement... A - 12 Servicing... A Technical Information... A - 13 A 1
3 1. OVERVIEW The AIRVO 2 is a humidifier with integrated flow generator that delivers high flow warmed and humidified respiratory gases to spontaneously breathing patients through a variety of patient interfaces. INTENDED USE The AIRVO 2 is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases. This includes patients who have had upper airways bypassed. The flow may be from 2-60L/min depending on the patient interface. The AIRVO 2 is for patients in hospitals and long-term care facilities. USA Federal Law restricts this unit for sale by or on the order of a physician.! Nasal delivery of respiratory gases generates flow-dependent positive airway pressure (PAP). This must be taken into account where PAP could have adverse effects on a patient. The unit is not intended for life support. To avoid burns: The unit should only be used with interfaces, water chambers and breathing tubes specified in this user manual. Using the breathing tube or interface for longer than the specified time can result in serious injury including infection. Before using oxygen with the unit, read all warnings in the Oxygen section of this manual. Never operate the unit if: the heated breathing tube has been damaged with holes, tears or kinks, it is not working properly, the case screws have ever been loosened. Do not block the flow of the air through the unit and breathing tube. The unit should be located in a position where ventilation around the unit is not restricted. Never block the air openings of the unit or place it on a soft surface such as a bed or couch/sofa, where the filter area may be blocked. Keep the air openings free of lint, hair etc. To avoid electric shock: Do not store or use the unit where it can fall or be pulled into water. If water has entered the unit enclosure, disconnect the power cord and discontinue use. Never operate the unit if: it has been dropped or damaged, it has a damaged power cord or plug, it has been dropped into water. Avoid unnecessary removal of the power cord from the rear of the device. If removal is necessary, hold the connector during removal. Avoid pulling on the power cord. Return the unit to an authorized service center for examination and repair, except as outlined in this manual. To avoid choking, or inhalation of a foreign object: Ensure an air filter is fitted when operating your unit. Never drop or insert any object into any opening or tube. Miscellaneous: Do not use the unit when the room temperature exceeds 30 C (86 F) or is below 10 C (50 F) as the unit may switch off. Humidity output will be compromised below 18 C (64 F) and above 28 C (82 F). The unit is not suitable for use in the presence of a flammable, anesthetic mixture with air or oxygen or nitrous oxide. A 2
4 AIRVO 2 AND ACCESSORIES AUDIO PAUSE UP DOWN ON/OFF (STANDBY) DISPLAY Heated breathing tube Patient interface English MODE HOSPITAL STAND POLE MOUNTING TRAY OXYGEN INLET PORT HEATED BREATHING TUBE CONNECTION PORT MEASUREMENT POINT OF DISPLAYED DEW POINT TEMPERATURE CHAMBER PORTS Water chamber POWER CORD and CONNECTOR SERIAL PORT FILTER COVER AIRVO 2 (PT101AZ/ PT101UK) HEATER PLATE FINGER GUARD AIR FILTER AUTO-FILL WATER CHAMBER (MR290) (with adapter fitted) Tube & chamber kits and patient interfaces Tube & chamber kit Interfaces 900PT531 Heated breathing tube, MR290 auto-fill chamber and adapter (10-Pack) à OPT316 Nasal Cannula - Infant (20-Pack) OPT318 Nasal Cannula - Pediatric (20-Pack) 900PT501 Heated breathing tube, MR290 auto-fill chamber and adapter (10-Pack) à OPT842 OPT844 OPT846 OPT870 RT013 Nasal Cannula - Small (20-Pack) Nasal Cannula - Medium (20-Pack) Nasal Cannula - Large (20-Pack) Tracheostomy Direct Connection (20-Pack) Mask Interface Adapter - 22mm (20-Pack) Cleaning and Disinfection Miscellaneous 900PT PT PT PT603 Disinfection Kit Disinfection Filter (2-Pack) Cleaning Sponge-Stick (20-Pack) Clean Storage Cover (20-Pack) 900PT PT PT PT PT913 OPT012 OPT014 Pole mounting tray Hospital stand Oxygen inlet extension kit Filter holder Air filter (2-Pack) Wigglepads (OPT316/OPT318) (20-pack) Oxygen Tubing (Optiflow Junior) A 3
5 2. SETTING UP AIRVO 2 1. BEFORE YOU BEGIN The AIRVO 2 should be fixed on a pole mounting tray (900PT405) below patient head height. Open the packaging of the tube & chamber kit (heated breathing tube, MR290 auto-fill chamber and adapter). 2. INSTALL WATER CHAMBER Remove the blue port caps from the chamber by pulling the tear tab upwards then remove the bracket holding the water supply tube. Fit the supplied adapter over the two vertical ports on the chamber and push on fully then clip the water supply tube into position. Fit the water chamber to the unit by pressing down the finger guard and sliding the chamber on, carefully aligning with the blue chamber port ends. Push the chamber on firmly until the finger guard clicks into place. To avoid burns: Do not start the unit without the water chamber in place. The water in the chamber becomes hot during use. Exercise caution when removing and emptying the chamber. Do not touch the heater plate, water chamber or chamber base during use. To avoid electric shock: When handling the unit with the water chamber in place, avoid tilting the machine to prevent any chance of water entering the unit enclosure. Empty all the water from the water chamber before transporting the unit. CAUTIONS To ensure optimal therapy (MR290 only): Do not use the auto-fill MR290 chamber if it has been dropped, or been run dry and the water out alarm has been activated. H 2 O 3. CONNECT WATER BAG Attach the sterile water bag to the hanging bracket 20cm (8 ) above the unit, and push the bag spike into the fitting at the bottom of the bag. Open the vent cap on the side of the bag spike. The chamber will now automatically fill to the required level and maintain that level until the water bag is empty. To ensure continual humidification, always ensure that the water chamber and/or water bag are not allowed to run out of water. Check that water flows into the chamber and is maintained below the fill line. If the water level rises above the fill line, replace the chamber immediately. MR290: Flow setting vs usage time (2-litre sterile water bag) L/min hrs A 4
6 4. INSTALL HEATED BREATHING TUBE One end of the heated breathing tube has a blue plastic sleeve. Lift the sleeve and slide the connector onto the unit. Push the sleeve down to lock. To avoid burns: Do not modify the breathing tube or interface in any way. Do not allow the breathing tube to remain in direct contact with skin for prolonged periods of time. Adding heat, above ambient levels, to any part of the breathing tube or interface e.g. covering with a blanket, or heating it in an incubator or overhead heater for a neonate, could result in serious injury. Do not use an insulating sleeve or any similar accessories which are not recommended by Fisher & Paykel Healthcare. CAUTIONS Position the heated breathing tube away from any electrical monitoring leads (EEG, ECG/EKG, EMG, etc), to minimize any possible interference with the monitored signal. English 5. SELECT PATIENT INTERFACE The AIRVO 2 can be used with a variety of patient interfaces. Read the separate user instructions for the patient interface that will be used, including all warnings. Nasal cannula Tracheostomy interface Mask interface adapter OPT842 OPT844 OPT846 AIRVO TM 2 OPT316, OPT318 (Refer to Using AIRVO 2 - Junior Mode ) OPT870 RT013 (with mask) Note that the RT013 Mask Interface Adapter is designed to be used with vented masks only. Do not use sealed masks. The following table shows the target dew-point temperature settings and target flow settings able to be used with these interfaces Low temperature ambient conditions may prevent the unit from reaching a 37 C target temperature setting at high target flow settings. In these cases, consider decreasing the target flow setting. To avoid burns: Do not modify the breathing tube or interface in any way. Do not use any patient interfaces not listed here. A 5
7 3. USING AIRVO 2 1. SWITCH ON UNIT Plug the unit s power cord into the mains power supply. The connector at the other end of the power cord should be well secured to the rear of the unit. To avoid electric shock: Ensure that the unit is dry before plugging into the power socket. Switch on the unit by pressing the On/Off button. 2. CHECK DISINFECTION STATUS The unit will show you whether it is safe for use on a new patient. Last Disinfection: #16 This AIRVO 2 is safe for use on a new patient. This AIRVO 2 has not been cleaned and disinfected since last use. This AIRVO 2 is NOT safe for use on a new patient. 3. WARM-UP The unit will begin to warm up. You will see numbers showing the current output dew-point temperature, flow and oxygen values. These numbers will pulse until they approach their target settings. This screen is called the Summary screen. 4. JUNIOR MODE If the patient will be using an Optiflow Junior nasal cannula (OPT316/ OPT318), you must activate Junior Mode. Junior Mode limits the target settings to: 34 C and 2-25 L/min, in increments of 1 L/min. To activate Junior Mode: Hold the Mode button for 5 seconds. New target settings New target settings The target settings for dew-point temperature and flow will be changed automatically. The colorful icons in the corners of the screen indicate that this unit is in Junior Mode. To deactivate Junior Mode, follow the same procedure: hold the Mode button for 5 seconds. A 6
8 N 5. CONFIGURE TARGET SETTINGS Press the Mode button to view target settings. These settings are locked by default. TARGET DEW-POINT TEMPERATURE You can set the AIRVO 2 to three target dew-point temperature settings: 37 C (98.6 F) 34 C (93 F) [if compliance at 37 C is a problem] 31 C (88 F) [for face masks only]. You may not have access to all settings, if: the unit is in Junior Mode (limited to 34 C), the unit was initially set up with tighter limits. The AIRVO 2 will return to its default setting (37 C) after every disinfection cycle. English To change the target dew-point temperature setting: Hold the Up and Down buttons for 3 seconds to unlock the setting. The lock will disappear and be replaced by an arrow showing the minimum and maximum accessible settings. Press the Up and Down buttons to choose the new setting. When you have finished, press the Mode button to 'lock' the setting again. The lock will reappear. Transport Mode: If Transport Mode has been enabled, you can activate it on this screen by holding the Audio pause button for 5 seconds. The unit will enter a low-power, low-humidity mode for 20 minutes, designed for use when transporting patients. For more information, refer to REF To deactivate Transport Mode, follow the same procedure: hold the Audio pause button for 5 seconds. A Press the Mode button to move on to the next screen. TARGET FLOW You can set the AIRVO 2 to flows between 10 L/min and 60 L/min, in increments of 1 L/min (10-25 L/min) and 5 L/min (25-60 L/min). You may not have access to all settings, if: the unit is in Junior Mode (limited to 2-25 L/min, in increments of 1 L/min), the unit was initially set up with tighter limits. The AIRVO 2 will remember its target flow setting when you switch it off. To change the target flow setting: Follow the same sequence of steps as above in To change the target dew-point temperature setting. A 7
9 Press the Mode button to move on to the next screen. OXYGEN You can connect supplementary oxygen to the AIRVO 2 (up to 60 L/min). The AIRVO 2 contains an oxygen analyzer to help you determine the oxygen fraction you are delivering to the patient. Your unit may have been initially set up with tighter limits. Use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply. Before using the AIRVO 2 with oxygen, read all of the following warnings: The use of oxygen requires that special care be taken to reduce the risk of fire. Accordingly, for safety it is necessary that all sources of ignition be kept away from the unit and preferably out of the room in which it is being used. Oxygen should not be used while smoking or in the presence of an open flame. The unit should be located in a position where ventilation around the unit is not restricted. A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances must be kept away from all oxygen equipment. Ensure that the AIRVO 2 is switched on before connecting oxygen. Oxygen must only be added through the special oxygen inlet port on the back of the unit. To ensure that oxygen enters the unit correctly, the oxygen inlet port must be fitted properly to the filter holder and the filter holder must be fitted properly to the unit. The power cord connector should also be well secured. Do not connect more than 60L/min O2 to the oxygen inlet port on the back of the unit. The oxygen concentration delivered to the patient can be affected by changes to the flow setting, oxygen setting, patient interface or if the airpath is obstructed. When finished, turn off the oxygen source. Remove the output of the oxygen source from the oxygen inlet port on the back of the unit. The oxygen flow must be turned off when the unit is not operating, so that oxygen does not build up inside the device. The oxygen analyzer within the AIRVO 2 uses ultrasonic measurement technology. It does not require in-field calibration. It is designed for use with pure oxygen - connecting any other gases or mixtures of gases will cause it to function incorrectly. CONNECT OXYGEN Connect the output from the oxygen source to the oxygen inlet port on the back of the unit. Make sure you push the oxygen tube firmly onto this connection port. ADJUST OXYGEN Adjust the level of oxygen from the oxygen source, until the desired oxygen fraction is displayed onscreen. It may take the reading several minutes to settle. You can set the oxygen fraction between the maximum and minimum values displayed above and below the arrow. If the oxygen fraction exceeds 95%, the oxygen reading will pulse red and the device will beep. Note that if the patient s peak inspiratory demand exceeds the flow delivered by the unit, the fraction of oxygen inspired by the patient will be lower than the value shown onscreen, due to the additional entrainment of ambient air. Check that suitable blood saturation levels are achieved at the prescribed flow. Press the Mode button to return to the Summary screen. A 8
10 6. CONNECT YOUR PATIENT Wait until the Ready for use symbol is displayed on the Summary screen. Ready for use symbol English Connect the patient interface to the heated breathing tube. Monitor the flow and oxygen values displayed on the Summary screen. Adjust the level of oxygen from the oxygen source as necessary. When the patient first uses the unit, the air will feel warm. This is normal. The patient should continue to breathe normally through the nose and/or mouth, or tracheostomy. 7. DURING USE If the Ready for use symbol has been displayed for 1 minute and no button has been pushed in this time, a screensaver will be launched. If excess condensate accumulates in the heated breathing tube, drain by lifting the patient end of the tube, allowing the condensate to run into the water chamber. 8. AFTER USE Switch off the unit by pressing the On/Off button. A 9
11 ALARMS The AIRVO 2 has visual and auditory alarms to warn you about interruptions to your patient s treatment. These alarms are generated by an intelligent alarm system, which processes information from the sensors and target settings of the unit and compares this information to pre-programmed limits. ALARM SIGNALS Visual alarm signal Symbols Meaning Alarm condition. (message) Audio paused. Auditory alarm signal 3 beeps in 3 seconds. Repeated every 5 seconds. Press this button to mute the auditory alarm for 115 seconds. The auditory alarm can be reactivated by pressing this button again. ALARM CONDITIONS All of the alarms listed below have been assessed as Medium Priority. These priorities have been allocated for an operator s position within 1 meter of the device. The unit also uses an internal priorityranking system. If multiple alarm conditions occur simultaneously, the unit will display the highest-priority alarm. The following table lists all of the alarm conditions from highest-priority to lowest priority, their causes, possible solutions and delays. Alarm conditions that affect oxygen delivery require an immediate response to assess the patient s saturation levels. Alarm conditions that affect humidity delivery require a prompt response to assess potential drying of mucus and associated blockages. Message Meaning Affects delivery of: Fault (E###) Check tube Check for leaks Check for blockages O 2 too low O 2 too high The unit has detected an internal fault and has shut itself down. Switch the unit off and then restart. If the problem persists, note the fault code and contact your Fisher & Paykel Healthcare representative. The unit cannot detect the heated breathing tube. Check that the heated breathing tube is not damaged and that it is plugged in correctly. If the problem persists, then change the heated breathing tube. The unit has detected a leak in the system. The most likely cause is that the water chamber has been removed or has not been pushed into place correctly. Check that the heated breathing tube is not damaged and that it is plugged in correctly. Check that the nasal interface is fitted. Check that the filter is fitted. The unit has detected a blockage in the system. Check the heated breathing tube or patient interface for blockage. Check the air filter and filter holder for blockage. Check whether the unit should be in Junior Mode. If the patient will be using an Optiflow Junior nasal cannula (OPT316/OPT318), you must activate Junior Mode. The measured oxygen level has fallen below the allowed limit. Check that the oxygen source is still correctly connected. Adjust the level of oxygen from the oxygen source as necessary. The measured oxygen level has exceeded the allowed limit. Adjust the level of oxygen from the oxygen source as necessary. Oxygen, humidity. Oxygen, humidity. Oxygen, humidity. Oxygen, humidity. Oxygen Oxygen Delays < 5 seconds < 5 seconds < 5 seconds < 10 seconds < 20 seconds < 20 seconds A 10
12 (continued) Message Meaning Affects delivery of: Cannot reach target flow Check water Cannot reach target temperature Check operating conditions [Power out] The unit cannot reach the target flow setting. Check the heated breathing tube or patient interface for blockage. Check whether the target flow setting is too high for the patient interface being used (refer to Setting up AIRVO 2 - Select Patient Interface ). The unit will choose appropriate new target settings. You will be prompted for acknowledgement. The oxygen concentration delivered to the patient can be affected by changes to the flow setting. Adjust the level of oxygen from the oxygen source as necessary. The chamber has run out of water. When a chamber runs dry, the chamber float may be damaged. Replace the chamber and water bag. [Twenty seconds after the chamber is removed, the Check for leaks alarm is activated (see above). When the chamber is replaced, the unit enters Warm-up Mode and resumes normal operation.] To ensure continual humidification, always ensure that the water chamber and/or water bag are not allowed to run out of water. The unit cannot reach the target temperature setting. You will be prompted for acknowledgement. The most likely cause for this is that the unit is operating at a high flow rate in low ambient conditions. Consider decreasing the target flow setting. The oxygen concentration delivered to the patient can be affected by changes to the flow setting. Adjust the level of oxygen from the oxygen source as necessary. The unit has detected that it is operating in unsuitable ambient conditions. Do not use the device when the ambient temperature is less than 10 C. Do not use the device when the ambient temperature is greater than 30 C. This alarm may be caused by a sudden change in ambient conditions (eg. storing the unit in a cold place then using it in a warm place). Leave the unit running for 30 minutes. Switch the unit off and then restart. The unit has been disconnected from the mains power supply. No visual alarm. The auditory alarm will sound for 120 seconds. Oxygen Humidity Humidity Humidity Oxygen, humidity. Delays 10 +/- 1 minutes Flows above 20 L/min: < 20 minutes Flows of and below 20 L/min: < 40 minutes 30 +/- 3 minutes 60 +/- 6 seconds < 5 seconds English ALARM LIMITS Most alarm limits are pre-programmed. The exceptions are listed below. These alarm limits may be changed to other values by authorized personnel. Changes will be preserved during or after any power loss. Alarm condition Factory-set alarm limit Possible preset values O 2 too low 21% O % O 2 O 2 too high 95% O % O 2 A hazard can exist if different alarm presets are used on different units within any single area, eg. an intensive care unit. Alarm limits set to extreme values can render the alarm system useless. CHECKING ALARM SYSTEM FUNCTIONALITY The functionality of the alarm system can be checked at any time when the unit is turned on. Remove the heated breathing tube. You should see the Check tube visual alarm signal and hear the auditory alarm signal. If either alarm signal is absent, do not use the unit. Contact your Fisher & Paykel Healthcare representative. AUDITORY INFORMATION SIGNALS In addition to auditory alarm signals, auditory information signals are provided. These are described below. Melody Ascending sequence of 5 tones Ascending sequence of 3 tones Single tone every 5 seconds A 11 Meaning The Ready for use symbol has appeared Activation/deactivation of Junior Mode Measured oxygen level > 95%, OR, Measured oxygen level > 32% at turn-off
13 4. REPROCESSING The AIRVO 2 must be cleaned and disinfected between patients according to the instructions in the Disinfection Kit Manual (900PT600). This should take place as soon as possible after use. The unit utilizes warmed water and can pose a risk of bacterial colonization and patient infection if cleaning, disinfection and replacement procedures are not followed. Standard aseptic techniques to minimize contamination should be followed when handling the unit and accessories. This includes proper hand-washing, avoiding hand contact with connection ports, safe disposal of the used consumables and suitable storage of the unit after cleaning and disinfection. SCHEDULE FOR CHANGING ACCESSORIES The accessories for the unit must be changed frequently to avoid the risk of infection. Parts should be replaced immediately if they are damaged or discolored; otherwise they must be replaced within the periods shown in the following table. Maximum period of use 1 week (single-patient use) 2 weeks (single-patient use) 3 months or 1000 hours All patient interfaces OPT316 Nasal Cannula - Infant OPT318 Nasal Cannula - Pediatric OPT842 Nasal Cannula - Small OPT844 Nasal Cannula - Medium OPT846 Nasal Cannula - Large OPT870 Tracheostomy Interface RT013 Mask Interface Adapter - 22mm Part number and description All tube & chamber kits 900PT501 Heated breathing tube, MR290 auto-fill chamber and adapter 900PT531 Heated breathing tube, MR290 auto-fill chamber and adapter (for use with OPT316/318 only) 900PT913 Air filter (or more often if significantly discolored) FILTER REPLACEMENT Air filter change due If the unit tells you that a filter change is due: 1. Take the filter holder from the back of the unit and remove the filter. 2. Replace the old filter with a new one. 3. Reattach the filter holder to the unit (clip the bottom of the filter holder in first, then rotate it upwards until the top clips into place). 4. Press the Mode button to move on to the next screen. SERVICING This device contains no serviceable parts. A 12
14 5. TECHNICAL INFORMATION SYMBOL DEFINITIONS Caution Hot Surfaces Type BF Applied Part ATTENTION Consult accompanying documents Do not throw away Drip Proof Alternating Current Class ll Double Insulated Power On/Off (Standby) 93/42/EEC Class IIa English PRODUCT SPECIFICATIONS Dimensions Weight 295 mm x 170 mm x 175 mm (11.6 x 6.7 x 6.9 ) 2.2 kg (4.8 lb) unit only, 3.4 kg (7.5 lb) packaged in bag incl. accessories Humidity Maximum temperature of delivered gas >33 mg/l at 37 C target >10 mg/l at 34 C target >10 mg/l at 31 C target 43 C (109 F) Supply frequency Hz Supply voltage/current V 2.2 A (2.4 A max) V 1.8 A (2.0 A max) Sound pressure level Auditory alarm pause Serial port Alarms exceed 1 m 115 seconds The serial port is used for downloading product data, using F&P Infosmart software. Maximum flow range (default) Maximum flow range (Junior Mode) L/min 2-25 L/min Maximum oxygen input 60 L/min Warm-up time 10 minutes to 31 C (88 F), 30 minutes to 37 C (98.6 F) using a MR290 chamber with flow rate of 35 L/min and starting temperature 23 ± 2 C (73 ± 3 F) Oxygen analyzer accuracy < ± (2.5% + 2.5% of gas level) (within the range 25-95% O 2 ) Operating conditions: C (64-82 F), 30-70% RH Designed to conform to the requirements of: IEC UL CSA C22.2/No AS EN The unit complies with the electromagnetic compatibility requirements of IEC In certain circumstances, the unit may affect or be affected by nearby equipment due to the effects of electromagnetic interference. If this should happen, try moving the unit or the location of the unit causing interference, or alternatively consult your healthcare provider. Accessory equipment connected to the serial port of the device must be certified to either IEC or IEC Furthermore all configurations shall comply with the system standard IEC Anyone who connects additional equipment to the signal input part or signal output part configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of the system standard IEC If in doubt, consult the technical services department or your local representative. OPERATING CONDITIONS Minimum/maximum ambient temperature Humidity Altitude Mode of operation C (64-82 F) 10-95% RH m (6000 ft) Continuous operation STORAGE AND TRANSPORT CONDITIONS The unit should be stored and transported in environmental conditions of -10 C to 60 C (14 F to 140 F), 10 to 95% RH, non-condensing. DISPOSAL INSTRUCTIONS Unit Disposal Instructions This unit contains electronics. Please do not discard with regular waste. Return to Fisher & Paykel Healthcare or dispose according to local guidelines for disposing of electronics. Dispose according to Waste Electrical and Electronic Equipment (WEEE) directive in European Union. Consumables Disposal Instructions Place the interface, breathing tube and chamber in a waste bag at the end of use. Hospitals should discard according to their standard method for disposing of contaminated product. A 13
15 For more information please contact your local Fisher & Paykel Healthcare representative REF REV A 2013-June 2013 Fisher & Paykel Healthcare Limited Manufacturer Fisher & Paykel Healthcare Ltd 15 Maurice Paykel Place East Tamaki, Auckland 2013 PO Box , Panmure Auckland 1741 New Zealand Tel: Fax: [email protected] Web: Australia Fisher & Paykel Healthcare Pty Limited New Street, PO Box 167 Ringwood, Melbourne Victoria 3134, Australia Tel: Fax: Austria Tel: Fax: Benelux Tel: Fax: China Tel: Fax: France Tel: Fax: Germany Tel: Fax: India Tel: Fax: Irish Republic Tel: Italy Tel: Fax: Spain Tel: Fax: Sweden Tel: Fax: Switzerland Tel: Fax: Taiwan Tel: Fax: Turkey Fisher Paykel Sağlık Ürünleri Ticaret Limited Şirketi, Alinteri Bulvari 1161/1 Sokak No , P.O. Box Ostim, Ankara, Turkey Tel: Fax: UK Fisher & Paykel Healthcare Ltd Unit 16, Cordwallis Park Clivemont Road, Maidenhead Berkshire SL6 7BU, UK Tel: Fax: USA/Canada Tel: or Fax:
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
PORTABLE AIR CONDITIONER
VERY IMPORTANT! MODEL: GDC-AC9RW / GDC-AC9RCS GDC-AC9RCW / GDC-AC12RW GDC-AC12RB / GDC-AC12RCB GDC-AC12RCW INSTRUCTIONS FOR USE PORTABLE AIR CONDITIONER Do not install and use your portable air conditioner
USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR
USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W
MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries
Vertical Display and Storage B1350-2. SKOPE Gen2: Three Door Chiller
Vertical Display and Storage User Manual MAN1227 Rev. 3.0 March 2008 edition CONTACT ADDRESSES Designed and Manufactured by New Zealand SKOPE INDUSTRIES LIMITED PO Box 1091, Christchurch New Zealand Freephone:
Online UPS. PowerWalker VFI 1000R/1U
Online UPS PowerWalker VFI 1000R/1U Manual Uninterruptible Power Supply System Table of Contents 1. Important Safety Warning... 1 1-1. Transportation... 1 1-2. Preparation... 1 1-3. Installation... 1 1-4.
SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / 8353. Instruction Manual. Page 1 of 9
SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / 8353 Instruction Manual Page 1 of 9 TABLE OF CONTENTS Introduction...3 Important Safeguards..3 System Components...5 Setup and Operation...5 Setup...5 Operation....6
NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS ELECTRICAL SAFETY This appliance is for indoor use only. Always turn off the unit and unplug
Use and Care Manual. Model CPA12KH AIR CONDITIONER
Use and Care Manual Model CPA12KH AIR CONDITIONER Introduction Thank you for choosing this air conditioner to provide you and your family with all of the "Home Comfort" requirements for your home, cottage
Operating Instructions
PLEASE READ BEFORE USE AND SAVE. Integrated Telephone System Model No. KX-T7FXW Pulse-or-tone dialing capability Operating Instructions Preparation Basic Operation Advanced Operation Useful Information
Operating instructions in the back. www.blackanddecker.co.uk PD1020L
Operating instructions in the back www.blackanddecker.co.uk PD1020L 2 ENGLISH (Original instructions) Intended use Your Black & Decker Dustbuster handheld vacuum cleaner has been designed for vacuum cleaning
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W
Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CCS Customer Support : 1-800-474-2147 Voltage rating: 115V~60Hz Power rating : 1400W For product inquiries or support
Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF
Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator
USER S MANUAL HSC-24A
AIRREX AIR CONDITIONER USER S MANUAL HSC-24A Thank you for purchasing an AIRREX AIR CONDITIONER. BEFORE operation please read this user s manual carefully. Keep this manual readily available. It is ESSENTIAL
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW
Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MN12CES / MN10CESWW Voltage rating: 120V~60Hz Power rating : 1100W (MN12CES) Power rating : 900W (MN10CESWW) Customer
USER S MANUAL FH052EAV1 FH070EAV1. System Air Conditioner (Cooling and Heating) ENGLISH ESPAÑOL FRANÇAIS ITALIANO PORTUGUÊS DEUTSCH E HNIKA
USER S MANUAL FH052EAV1 FH070EAV1 E HNIKA PORTUGUÊS ENGLISH ESPAÑOL ITALIANO DEUTSCH FRANÇAIS System Air Conditioner (Cooling and Heating) E S F I P D G DB98-29263A(1) Safety Precautions Register your
UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS
UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS: Carefully read these instructions before installing your new air-conditioner. AUSTRALIAN AUTOMOTIVE AIR AL00500054E 1 Table
Introduction. Intended use. Contraindications. General warnings and cautions. Patient Instructions WARNING CAUTION. English
Patient Instructions English Introduction ApneaLinkAir ApneaLink Plus Your physician or healthcare provider has given you the ApneaLink Air device to test if you have sleep-disordered breathing. The device
COOLING THERAPY SYSTEM
COOLING THERAPY SYSTEM Q U I C K R E F E R E N C E G U I D E Garment Models Description and Operation Controller Troubleshooting The Quick Reference Guide is a guideline only and does not replace the User's
IMPORTANT INSTRUCTIONS & OPERATING MANUAL. Houston 50 Inch Electric Wall Mounted Fireplace Black / White
IMPORTANT INSTRUCTIONS & OPERATING MANUAL Houston 50 Inch Electric Wall Mounted Fireplace Black / White Model Number:MFE5050BK Model Number:MFE5050WH Read these instructions carefully before attempting
SWIMMING POOL HEAT PUMP Owners Manual
SWIMMING POOL HEAT PUMP Owners Manual This manual refers to the 17.0kw and 21.0kw models only. The heat pump unit is sold with a 1 year warranty. In addition there is a 2 year parts warranty on the compressor
Meaco 30L and Meaco 40L dehumidifier instruction manual
Meaco 30L and Meaco 40L dehumidifier instruction manual Please read this instruction manual before using the dehumidifier and keep safe for future reference SAFETY INSTRUCTIONS PLEASE READ ALL INSTRUCTIONS
System One Heated Humidifier USER MANUAL
System One Heated Humidifier USER MANUAL 2011 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Warnings... 2 Cautions... 2 Contraindications... 3 Symbol Key...
Advantium 2 Plus Alarm
ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia
MOUNTING AND OPERATING INSTRUCTIONS
MOUNTING AND OPERATING INSTRUCTIONS CF 51/55 1. Delivery CF 51/55-1 calculator with battery (optionally with mains power supply) - 1 wall mounting bracket - package with material for sealing, screws, wall
TS1200 Intelligent Lead Free Soldering Station. Operation Manual
TS1200 Intelligent Lead Free Soldering Station Operation Manual Thank you for purchasing an Intelligent Lead Free Soldering Station. It is designed for lead free soldering. Please read this manual before
AN500T, AN1000, AN1000T, AN1500, AN1500T AN2000, AN2000T
Product Instruction Manual Accona AN500T, AN1000, AN1000T, AN1500, AN1500T AN2000, AN2000T Panel heater v16.5/5 Version 3.2 Jan 2015 Contents 1. Important safety points 2. Installation 2.1. Wall mounting
Carpet Washer. vax.co.uk VRS5W. Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308. Vax model number: Version 1.0
VRS5W Powermax User Guide V1.0.qxd:V1.0 23/7/10 15:35 Page 1 Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308 Carpet Washer Vax model number: VRS5W instruction manual Version 1.0 Please read carefully
SUB Universal water baths SBB Boiling baths JB, PB Unstirred water baths
SUB Universal water baths SBB Boiling baths JB, PB Unstirred water baths Operating instructions Universal water baths SUB6: 6 litres SUB14: 14 litres SUB28: 28 litres SUB36: 36 litres Boiling baths SBB6:
B150 Series Commercial Brewer. Direct Plumb Kit
B150 Series Commercial Brewer Welcome Keurig Authorized Dealer! Direct Plumb Kit This manual provides: Installation of the Direct Plumb Kit (DPK) that allows you to connect your customer s B150 brewer
VACUUM CLEANER INSTRUCTION MANUAL
VACUUM CLEANER INSTRUCTION MANUAL MODEL: SVC9702 PLEASE READ THIS MANUAL CAREFULLY BEFORE USING AND KEEP IT PROPERLY FOR FUTURE USE 1 Safety Cautions: WARNING: To reduce the risk of fire, electrical shock,
Owner s Manual. Not suitable for mobile applications. Important Safety Instructions 2. Quick Installation 3. Basic Operation 4. Storage & Service 7
Register online today for a chance to win a FREE Tripp Lite product! www.tripplite.com/warranty Owner s Manual Internet Office & BC Personal UPS Systems Not suitable for mobile applications. Important
RomiApex A-15. APEX LOCATOR User Manual
RomiApex A-15 APEX LOCATOR User Manual Table of Contents Introduction..... 2 1. Indications for use... 3 2. Precautions..... 3 3. Adverse Reactions...... 3 4. Package Contents........ 3 5. Getting Started.....
OASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE
OASIS-PLUS 120V INFORMATION & OPERATING INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE 56041852 WARNING:
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE
16/32 Channel 1U Rack Mount CCTV Power Supply
16/32 Channel 1U Rack Mount CCTV Power Supply Manual PH-A3224-GUQ Shown 16-Channel 32-Channel PTC PH-A1612-PUQ PH-A3224-PUQ Glass Fuse PH-A1612-GUQ PH-A3224-GUQ Industrial design 12 Amp 3 Amps per channel
1. Introduction...1 1.1 Revision History...1. 2. Warnings, Cautions, Contraindictions...2 2.1 Contraindications...2 2.2 Warnings...2 2.3 Cautions...
Technical Manual TABLE OF CONTENTS 1. Introduction...1 1.1 Revision History...1 2. Warnings, Cautions, Contraindictions...2 2.1 Contraindications...2 2.2 Warnings...2 2.3 Cautions...3 3. Specifications...4
CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN
BeoLab 4000 Guide CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN CAUTION: To reduce the risk of electric shock, do not remove cover (or back). No User-serviceable parts inside. Refer servicing to qualified
SuperEndo B&L-alpha II Cordless Heat Carrier Operation & Maintenance Instruction Manual
SuperEndo B&L-alpha II Cordless Heat Carrier Operation & Maintenance Instruction Manual PLEASE READ ALL INSTRUCTIONS PROVIDED FOR THIS DEVICE BEFORE USING IT. Caution: This product should only be used
Ultrasonic support removal tank
Ultrasonic support removal tank User guide and safety manual See inside for use and safety information. The Ultrasonic support removal tank automatically removes PLA support material from a printed ABS
EnergySmart monitor user instructions
EnergySmart monitor user instructions britishgas.co.uk/energysmart Your new EnergySmart monitor has arrived It s a great way to learn how to use less electricity and save money, and just one of the ways
PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: AW 120 / AW 122 / AW 125 AWD 120 / AWD 121 / AWD 129
PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: The information included in this Splendide Repair Manual may change without notice. Please see our web site www.splendide.com/service/docs.html for updates,
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CHCSCS
Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CHCSCS Voltage rating: 120V~60Hz Power rating : 1400W(Cooling) Power rating : 1350W(Heating) Customer Support : 1-800-474-21477
Get Cleaning... User Guide Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308. Carpet Washer. vax.co.uk. Rapide Ultra series
W90-RU & W89-RU Series User Guide v1.4.qxd:user guide 2/3/11 10:03 Page 1 Carpet Washer User Guide Vax Careline: (UK) 0844 412 8455 Get Cleaning... What s your Vax model number? (Located on the top flap
MICA HEATER INSTRUCTION MANUAL Model No: UHM-786 230V 50Hz 2200W
MICA HEATER INSTRUCTION MANUAL Model No: UHM-786 230V 50Hz 2200W Safety Precautions To reduce the risk of personal injury or damage to property, basic safety precautions must be observed including the
car + boat Box Contents x2 x2 22350_DC16_CAR_BOAT_OPSMAN.indd 1 29/7/08 15:27:01
car + boat Box Contents x x 350_DC6_CAR_BOAT_OPSMAN.indd 9/7/08 5:7:0 SAVE THESE INSTRUCTIONS THIS APPLIANCE IS INTENDED FOR HOUSEHOLD USE ONLY IMPORTANT SAFETY INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE
Drayton Digistat +2RF/+3RF
/+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer
D7050/D7050TH. Installation Instructions. Multiplex Photoelectric Smoke Detectors
D7050/D7050TH EN Installation Instructions Multiplex Photoelectric Smoke Detectors D7050/D7050TH Installation Instructions.0 Description Notices These instructions cover the installation of the D7050/D7050TH
User Manual 1K/2K/3K Online UPS
User Manual 1K/2K/3K Online UPS Uninterruptible Power Supply System Table of Contents 1. Important Safety Warning... 1 1-1. Transportation... 1 1-2. Preparation... 1 1-3. Installation... 1 1-4. Operation...
Multi Tube Rotator. Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11
Multi Tube Rotator Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11 Table of Contents Safety Information...3 Alert Signals...3 Description...4 Specifications...5 Power Requirements...5 Shaker Motion...5
floater imports DF169 ECO AIR COOLER EVAPORATIVE AIR COOLER WITH IONIZER USER S MANUAL
floater imports DF169 ECO AIR COOLER EVAPORATIVE AIR COOLER WITH IONIZER USER S MANUAL WHAT IS AN ECO AIR COOLER? The ECO AIR COOLER uses evaporative cooling to provide an airflow that is cooler than the
MR850 RESPIRATORY HUMIDIFIER
MR850 RESPIRATORY HUMIDIFIER Technical Manual REVISION J Copyright 2005 Fisher & Paykel Healthcare Ltd. Auckland, New Zealand Fisher & Paykel Healthcare Head Office: PO Box 14-348, Panmure, Auckland 1134
Stove Guard Kit User Manual
Stove Guard Kit User Manual Innohome improves the Safety of your Home. inno home www.innohome.com Stove Guard Kit User Manual Stove Guard Kit User Manual Congratulations! You now own one of the most intelligent
ENGLISH INSTRUCTION & INSTALLATION MANUAL DUCTLESS MINI SPLIT AIR CONDITIONING SYSTEMS
ENGLISH INSTRUCTION & INSTALLATION MANUAL DUCTLESS MINI SPLIT AIR CONDITIONING SYSTEMS Céliera Corporation. All rights reserved. Unauthorized duplication, reproduction prohibited. CONTENTS SAFETY PRECAUTIONS...
Quick Start Guide See Inside for Use and Safety Information
3 rd Generation Personal 3D Printer Quick Start Guide See Inside for Use and Safety Information The USB Mass Storage Device Contains the User Guide and Quick Start Guide in other Languages Congratulations
aseries A13B Mini Bullet Camera User Manual
aseries A13B Mini Bullet Camera User Manual Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer. This manual applies to the MicroView
Instructions Manual AmazonBasics 12-Sheet Cross-Cut Paper / CD / Credit Card Shredder with Basket ASIN# B005QAQFFS. Customer Service 1-866-216-1072
Instructions Manual AmazonBasics 12-Sheet Cross-Cut Paper / CD / Credit Card Shredder with Basket ASIN# B005QAQFFS Customer Service 1-866-216-1072 Paper Shredder Instructions Manual Sheet Capacity 12 Sheets
CFMRI Physiological Monitoring System. Operator s Manual
CFMRI Physiological Monitoring System Operator s Manual This manual provides a quick guidance to using the physiological monitoring system during an MRI session. It is not intended as a substitute for
COLOR VIDEO DOOR PHONE CDV-71BE/D
COLOR VIDEO DOOR PHONE CDV-71BE/D 513-11, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Int l Business Dept. : Tel.; +82-31-7393-540~550 Fax.; +82-31-745-2133 Web site : www.commax.com Printed
Infrarot-Bewegungsmelder IP44
Infrarot-Bewegungsmelder IP44 infrared motion sensors IP44 ODA (weiß) slim ODA (schwarz) slim 95174 96000 ODA (weiß) ODA (schwarz) 95175 96001 Betriebsanleitung User s Manual User s Manual infrared motion
Uninterruptible Power Supply ERA LED 1.5 ERA LED 2.0 ERA LED 2.6. User s manual
Uninterruptible Power Supply ERA LED 1.5 ERA LED 2.0 ERA LED 2.6 User s manual Index Safety Warnings... 2 1 Introduction... 3 2 General Characteristics... 4 3 Receipt and site selection... 4 4 EXTERNAL
2 YEAR WARRANTY S7200PAU USE & CARE MANUAL ADVANCED CERAMIC TECHNOLOGY. To register your product go to www.remington-products.com.
USE & CARE MANUAL PLEASE READ PRIOR TO USE To register your product go to www.remington-products.com.au ADVANCED CERAMIC TECHNOLOGY 2 YEAR WARRANTY S7200PAU 1 IMPORTANT SAFEGUARDS Warning: When the straightener
Daily use. Never use alcohol or other solvents to clean any part of the loudspeakers!
BeoLab 6002 Guide WARNING: To reduce the risk of fire or electric shock, do not expose this appliance to rain or moisture. Do not expose this equip ment to dripping or splashing and ensure that no objects
Operating Instructions
Integrated Telephone System Model No. KX-TS05HGW PLEASE READ BEFORE USE AND SAVE. Pulse-or-tone dialing capability Operating Instructions Preparation Basic Operation Advanced Operation Useful Information
UK s best selling phone brand. User Guide. BT Big Button 200 Corded Phone
UK s best selling phone brand User Guide BT Big Button 200 Corded Phone 2 Welcome. to your BT Big Button 200 telephone Large buttons for easy, accurate and more comfortable dialling Large display to assist
User Manual Color video door phone
User Manual Color video door phone CDV-70KM Thank you for purchasing COMMAX products. Please carefully read this User s Guide (in particular, precautions for safety) before using a product and follow instructions
OPTIONAL SLENDER REMOTE CONTROL
DAIKIN ROOM AIR CONDITIONER Operation Manual OPTIONAL SLENDER REMOTE CONTROL BRC944A2B READ BEFORE OPERATION Safety Precautions...2 Names of Functions of Parts...4 Preparation before Operation...5 OPERATION
STAINLESS STEEL TANK TABLE OF CONTENTS TROUBLESHOOTING... 1 TESTING WATER TEMPERATURE... 3 TANK & JACKET ASSEMBLY... 4.
TABLE OF CONTENTS STAINLESS TROUBLESHOOTING... 1 TESTING WATER TEMPERATURE... 3 & JACKET ASSEMBLY... 4 SST Section 052506 1 Troubleshooting performed by untrained personnel could result in electrical
Operating Instructions Bedienungsanleitung Mode d emploi
Operating Instructions Bedienungsanleitung Mode d emploi DW 400 www.bron-kobold.com Operating instructions DW 400 Before use Please read all the information contained in these operating instructions carefully.
NEWTECH PRO 1-2-3 KVA ONLINE UNINTERRUPTIBLE POWER SUPPLY USER MANUAL
NEWTECH PRO 1-2-3 KVA ONLINE UNINTERRUPTIBLE POWER SUPPLY USER MANUAL [email protected] / www.tuncmatik.com Table of Contents 1. Important Safety Warning 2 1-1. Transportation 2 1-2. Preparation 2 1-3.
OWNER S MANUAL DRAINVAC ATOMIK PRINTED APRIL 4-2011
OWNER S MANUAL th DRAINVAC ATOMIK PRINTED APRIL 4-2011 INTRODUCTION We take this opportunity to express our gratitude and extend our congratulations for your decision to purchase a Drainvac product. A
Hydraulic control unit series 0086-372-01
Technical Product Information No. 1290 EN Hydraulic control unit series 0086-372-01 Contents Page About this Technical Product Information (TPI) 2 The ORTLINGHAUS numbering system 2 About the product 3
GENANO 4500 MEDICAL AIR PURIFICATION DEVICE
GENANO AIR PURIFICATION TECHNOLOGY GENANO 4500 MEDICAL AIR PURIFICATION DEVICE OPERATING INSTRUCTION 1 GENANO LTD Kimmeltie 3 FI- 02110 Espoo, Finland Tel + 358 9 7743 870 Fax + 358 9 7533 130 www.genano.fi
FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION
FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION Caution! UV-radiation of this device is in the range of UV-A (320-400 nm). Direct exposure to eyes shall therefore be avoided. UV protection glasses shall
GE Healthcare. Lullaby Warmer. Because every baby deserves a high standard of care
GE Healthcare Lullaby Warmer Because every baby deserves a high standard of care A confident choice for infant warming At GE Healthcare, we share your commitment to providing high quality care for all
This unit is a pump that will clear the waste from a domestic shower and washbasin, bath, washing machine, glass washer and a sink.
1. Description This unit is a pump that will clear the waste from a domestic shower and washbasin, bath, washing machine, glass washer and a sink. The unit will start automatically. Please pay particular
=============================== WARNING
=============================== WARNING EXPLANATION OF GRAPHICAL SYMBOLS This symbol is intended to alert the user to the presence of unprotected dangerous voltage" within the product's enclosure that
Getting started with
PART NO. CMA113 MADE IN CHINA 1. Measuring CAT II 2. Max. voltage 250V ~ 3. Max. current 71 Amp Getting started with Electricity consumption monitoring single phase for homes and some smaller light commercial
FCC COMPLIANCE STATEMENT FOR AMERICAN USERS
FCC COMPLIANCE STATEMENT FOR AMERICAN USERS This equipment has been tested and found to comply with the limits for a CLASS A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
OPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
User Manual. Video Doorphone CDV-1020AQ
User Manual Video Doorphone CDV-1020AQ 513-11, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Int l Business Dept. Tel. : +82-31-7393-540~550 Fax. : +82-31-745-2133 Web site : www.commax.com
PBX Series Quick Fit Connector Bimetallic Steam Traps
6262100/6 IM-P626-01 ST Issue 6 PBX Series Quick Fit Connector Bimetallic Steam Traps Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning
Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS
Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Technical Information sheet 0001/07/Rev01 ENG FUNCTION Series S001 and S002 circulation units are applied to the primary circuit of solar
Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan
Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan TABLE OF CONTENTS INTRODUCTION 2 IMPORTANT SAFEGUARDS...2 PACKAGE CONTAINS..2 NAMES OF PARTS.3
TMS TANK MANAGEMENT SYSTEM
TMS TANK MANAGEMENT SYSTEM Page 1 of 9 Operating Instructions GENERAL The Tank Management System is a bespoke design to control, monitor and accommodate efficient storage and dispensing of TMS. FUNCTIONS
Wireless Alarm system s manual
MOUNTVIEW TECH AUSTRALIA PTY LTD Wireless Alarm system s manual ADS A180 Series ADS Security 1/11/2011 1. Before You Begin For your safety and the safety of others, and to ensure that you get the most
MCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
GNOME PELLET E.I. Pellet Heater Owner's Manual Installation and Operating Instructions. Please read this entire manual before installation.
Pellet Heater Owner's Manual Installation and Operating Instructions Please read this entire manual before installation. Save these instructions. SAFETY NOTICE HEATER MUST BE PROPERLY INSTALLED AND MAINTAINED
Important Safeguards
Table of Contents Important Safeguards...2 Product Layout...3 Preparing for Use...4 Air-conditioning without installation...4 Air-conditioning with installation...5 Control Panel...6 Operating from the
When you switch off your system, or mute the sound, the red indicator light appears immediately, indicating that the subwoofer is not in use.
BeoLab 11 Guide WARNING: To reduce the risk of fire or electric shock, do not expose this appliance to rain or moisture. Do not expose this equip ment to dripping or splashing and ensure that no objects
Register your product and get support at. www.philips.com/welcome AS111. User manual
Register your product and get support at www.philips.com/welcome AS111 User manual Contents 1 Important 4 Safety 4 English 2 Your docking speaker for AndroidTM 5 Introduction 5 What's in the box 5 Overview
EverFlo. EverFlo Q EN, DE DSF 4/16/08
1053621 EN, DE DSF 4/16/08 EverFlo EverFlo Q English Table of Contents Symbol Key... 2 Abbreviations... 2 Chapter 1: Introduction... 3 Intended Use... 3 About Your EverFlo / EverFlo Q...3 Parts of Your
CAUTION OPC-LM1-IL. Option Card for Encoder of Line Driver Output. Instruction Manual
Instruction Manual OPC-LM1-IL Option Card for Encoder of Line Driver Output CAUTION Deliver this instruction manual without fail to those who actually operate the equipment. Read this operation manual
PG_IM_A.indd 1 23/06/2014 15:22
Max Min Linen Lin Cotton Coton Wool Laine Silk Soie PG_IM_A.indd 1 23/06/2014 15:22 PG_IM_A.indd 2 23/06/2014 15:22 Max Cotton Coton Silk Soie Wool Laine Linen Lin Min Thank you for purchasing Phoenix
esata External Storage
esata External Storage Operation Manual DA-ES110 Before reading this manual This operation manual contains basic instruction on installing and using esata External Storage, an IDIS product. Users who are
Installation Instructions for Alarm Module Kit A043F059
Instruction Sheet 07-2013 Installation Instructions for Alarm Module Kit A043F059 1 Introduction The information contained within is based on information available at the time of going to print. In line
USER S MANUAL. Duct Type Series. Free Joint Multi Air Conditioner (Cooling and Heating) FUEA Series E S F I P D G DB98-29565A(1) ENGLISH ESPAÑOL
USER S MANUAL Duct Type Series MH MH FEEA Series FUEA Series ENGLISH ESPAÑOL PORTUGUÊS E HNIKA DEUTSCH ITALIANO FRANÇAIS Free Joint Multi Air Conditioner (Cooling and Heating) E S F I P D G DB98-29565A(1)
WALL MOUNTED SPLIT TYPE AIR CONDITIONER
WALL MOUNTED SPLIT TYPE AIR CONDITIONER MANUAL 8 OM-G12-ACSON CONTENTS - Operating Guide page 1 - G12 Remote Controller Indication page 2 - Indicator Lights page 4 - Installation of Aroma page 5 - Auto
BT Big Button 100. User Guide
BT Big Button 100 User Guide Welcome to your BT Big Button 100 phone Large buttons for easy dialling. Handsfree make and receive calls using the loudspeaker. One-touch 1571 button dials your network answering
SDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual January 2011 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
