Biomechanical Testing of Suture-Based Meniscal Repair Devices Containing Ultrahigh-Molecular-Weight Polyethylene Suture: Update 2011
|
|
|
- Rosamund Thornton
- 10 years ago
- Views:
Transcription
1 Biomechanical Testing of Suture-Based Meniscal Repair s Containing Ultrahigh-Molecular-Weight Polyethylene Suture: Update 2011 F. Alan Barber, M.D., Morley A. Herbert, Ph.D., Eric D. Bava, M.D., and Otis R. Drew, M.D. Purpose: To evaluate the biomechanical characteristics of recently introduced ultrahigh-molecularweight polyethylene suture based, self-adjusting meniscal repair devices. Methods: Updating a prior study published in 2009, we made vertical longitudinal cuts 3 mm from the periphery in fresh-frozen adult human menisci to simulate a bucket-handle meniscus tear. Each tear was then repaired by a single repair technique in 10 meniscus specimens. Group 1 menisci were repaired with a vertical mattress suture of No. 2-0 Ethibond (Ethicon, Somerville, NJ). Group 2 menisci were repaired with a vertical mattress suture of No. 2-0 OrthoCord (DePuy Mitek, Raynham, MA). Group 3 menisci were repaired with a single OmniSpan device with No. 2-0 OrthoCord suture (DePuy Mitek). Group 4 menisci were repaired with a single Meniscal Cinch device with No. 2-0 FiberWire suture (Arthrex, Naples, FL). Group 5 menisci were repaired with a single MaxFire device inserted with the MarXmen gun (Biomet Sports Medicine, Warsaw, IN). Group 6 menisci were repaired with a Sequent device with No. 0 Hi-Fi suture (ConMed Linvatec, Largo, FL) in a V suture configuration. Group 7 menisci were repaired with a single FasT-Fix 360 device (Smith & Nephew Endoscopy, Andover, MA). By use of a mechanical testing machine, all samples were preloaded at 5 N and cycled 200 times between 5 and 50 N. Those specimens that survived were destructively tested at 5 mm/min. Endpoints included maximum load, displacement, stiffness, and failure mode. Results: Mean failure loads were as follows: Ethibond suture, 73 N; OrthoCord suture, 88 N; OmniSpan, 88 N; Cinch, 71 N; MarXmen/MaxFire, 54 N; Sequent, 66 N; and FasT-Fix 360, 60 N. Ethibond was stronger than MarXmen/MaxFire. The mean displacement after 100 cycles was as follows: Ethibond, 2.58 mm; OrthoCord, 2.75 mm; OmniSpan, 2.51 mm; Cinch, 2.65 mm; MarXmen/MaxFire, 3.67 mm; Sequent, 3.35 mm; and FasT-Fix 360, 1.13 mm. The MarXmen/MaxFire showed greater 100-cycle displacement than Ethibond and FasT-Fix 360. No difference in stiffness existed for these devices, and failure mode varied without specific trends. Conclusions: The biomechanical properties of meniscal repairs using the OmniSpan, Cinch, Sequent, and FasT-Fix 360 devices are equivalent to suture repair techniques. However, the MarXmen/MaxFire meniscal repair device showed significantly lower failure loads and survived less cyclic loading in the human cadaveric meniscus than other tested repairs. Clinical Relevance: Most commercially available devices for all-inside meniscal repair using ultrahigh-molecular-weight polyethylene suture provide fixation comparable to the classic vertical mattress suture repair technique in human cadaveric meniscus. From the Plano Orthopedic Sports Medicine and Spine Center (F.A.B., E.D.B., O.R.D.), Plano, Texas; and Advanced Surgical Institutes, Medical City Dallas Hospital (M.A.H.), Dallas, Texas, U.S.A. The authors report the following potential conflict of interest or source of funding in relation to this article: The meniscus repair devices tested were provided at no cost by the various companies making them. In addition, a $500 fee to underwrite the costs of the Instron machine was paid. Received April 26, 2011; accepted November 17, Address correspondence to F. Alan Barber, M.D., Plano Orthopedic Sports Medicine and Spine Center, 5228 W Plano Pkwy, Plano, TX 75093, U.S.A by the Arthroscopy Association of North America /11263/$36.00 doi: /j.arthro Arthroscopy: The Journal of Arthroscopic and Related Surgery, Vol 28, No 6 (June), 2012: pp
2 828 F. A. BARBER ET AL. Meniscus repair can be accomplished by a variety of surgical techniques. The inside-out suture repair using a braided polyester suture has become a well-established and accepted technique for the treatment of reparable meniscal tears. 1-6 However, technical difficulty exists with suture repairs of posterior horn meniscal tears, and the strength of No. 2-0 braided polyester suture makes it subject to breaking. Newer meniscal repair devices now allow for repair of the meniscus by all-inside techniques. The initial generation of all-inside rigid fixation devices (i.e., tacks, staples, and screws) has been shown to provide lower load-to-failure strength and carry the risk of chondral abrasion because a portion of these devices remain exposed on the meniscal surface The development of all-inside meniscal repair devices with self-adjusting sutures provided greater strength and safety, but the braided polyester suture material was still subject to breaking. 7,10,11 The current generation of self-adjusting suture devices made with ultrahigh-molecular-weight polyethylene (UHMWPE) material may provide a stronger allarthroscopic technique that is less likely to break, avoids the need for additional incisions about the knee, decreases the potential for injury to the neurovascular structures about the knee, and avoids many of the complications linked to the prior meniscal repair techniques Previous testing of these self-adjusting UHMWPE suture devices in a porcine meniscus model has shown promising biomechanical data. 19 However, it is unclear how these devices perform in human meniscus tissue. In addition, prior studies using a single destructive pull for testing do not replicate the clinical condition as well as cyclic loading followed by destructive testing. The purpose of this study was to evaluate the biomechanical characteristics of recently introduced UHMWPE suture based, selfadjusting meniscal repair devices in a human cadaveric model. The hypothesis of this study was that the different UHMWPE suture containing devices would show better structural properties when subjected to cyclic loading in the human meniscus model. METHODS vertical longitudinal cut 3 mm from the periphery of each meniscus was created with a knife. This cut was created after the meniscus was completely removed from the tibia. The cut was not initially extended into the anterior and posterior meniscal horns to allow better control of the test meniscus during repair device insertion. A single suture or repair device was placed in each meniscus to approximate the 2 meniscal edges. All meniscal repair sutures or devices were placed in the posterior horn, and both medial and lateral menisci from the same knee were used with each device. Once the repair construct was placed and the repair complete, we divided the 2 remaining tissue bridges, completely separating the meniscus into 2 segments connected only by the single repair. Testing was conducted on a 3345 Instron mechanical testing machine (Instron, Canton, MA) and cyclically loaded at a displacement rate of 5 mm/min with a distraction stress always parallel to the axis of the repair device being tested (Fig 1). Each segment of the repaired meniscus was held with 2 metal clamps that were in turn attached to the mechanical testing We used 70 fresh-frozen adult human menisci (from 35 knees) provided by the Musculoskeletal Transplant Foundation (Edison, NJ) for research purposes. The meniscus specimens used in this bench test were all free meniscus specimens suitable for use as meniscal allografts. All menisci were intact and unblemished. A FIGURE 1. Testing was conducted on a mechanical testing machine with each meniscal segment held with 2 metal clamps. F. Alan Barber.
3 MENISCAL REPAIR DEVICES 829 FIGURE 2. A single vertical suture of No. 2-0 Ethibond was placed 3 mm inside the meniscal cut and hand tied on the capsular side of the meniscus to approximate the 2 meniscal fragments. F. Alan Barber. machine. This allowed the consistent application of force to the repair system without the plastic deformation associated with suture or tape materials. After an initial preload to take the slack out of the system, a cyclic loading force between 5 and 50 N was applied at 1 Hz for 200 cycles. Those specimens that survived the cycling were destructively tested (loaded to failure) at a rate of 5 mm/min Load and displacement were sampled continuously at 10 Hz. The displacement was measured by the travel of the actuator of the materials testing machine. This loading rate simulates the stresses on the meniscus during early postoperative rehabilitation exercises and activities of daily living, such as slow-speed walking, stair climbing or descending, and squatting. This testing protocol was consistent with previously published testing protocols. 7,10,19,20,22 A total of 10 samples for each repair device tested were used. The sampling rate for force and position data was 50 per second. This was downloaded into a spreadsheet and analyzed with statistical software. Endpoints for this testing were as follows: 1. Maximum load to failure with destructive testing after 200 cycles 2. Cyclic displacement at both 100 cycles and 200 cycles 3. Stiffness, determined from the maximal endpoints of the linear region of the load-displacement plot 4. Mode of failure, determined by visual inspection of the failure result The repair devices tested included the following: In group 1 a single vertical suture of No. 2-0 Ethibond (Ethicon, Somerville, NJ) was placed 3 mm inside the meniscal cut, extending from the superior surface to the inferior surface of the meniscus, and tied by hand with 6 square knots on the capsular side of the meniscus, approximating the 2 fragments of meniscus (inside-out vertical stitch). This served as the classic control device repair to provide comparison to the historical record (Fig 2). In group 2 a single vertical suture of No. 2-0 OrthoCord (DePuy Mitek, Raynham, MA) (55% polydioxanone and 45% UHMWPE) was placed 3 mm inside the meniscal cut, extending from the superior surface to the inferior surface of the meniscus, and tied by hand with 6 square knots on the capsular side of the meniscus, approximating the 2 fragments of meniscus (high-strength suture inside-out vertical stitch control). In group 3 a repair using the OmniSpan device (DePuy Mitek) (Fig 3) was placed in a vertical mattress configuration on the superior surface of the meniscus, 3 mm inside the meniscal cut and angled to orient 1 arm toward the superior peripheral capsule and the second toward the inferior meniscal capsule. The OmniSpan has a strand of No. 2-0 OrthoCord that is doubled between 2 poly ether ether ketone (PEEK) anchors, inserted by a gun with a curved needle. The sliding locking knot is located on the outside of the first PEEK anchor inserted and creates a repair with 2 sutures crossing between the 2 anchors without a knot on the meniscal surface. FIGURE 3. The OmniSpan has a strand of No. 2-0 OrthoCord that is doubled between 2 PEEK anchors. The sliding locking knot is located on the outside of the first PEEK anchor and creates a repair with 2 sutures crossing between the 2 anchors without a knot on the meniscal surface. F. Alan Barber.
4 830 F. A. BARBER ET AL. FIGURE 4. The Meniscal Cinch has 2 PEEK anchors connected by No. 0 FiberWire with a pre-tied sliding locking knot. F. Alan Barber. FIGURE 6. The Sequent meniscal repair device has up to 7 PEEK anchors connected with No. 0 Hi-Fi suture and allows a continuous stitching technique. The stitch is secured at each PEEK anchor by making 2 full clockwise rotations of the insertion device and tensioning the suture to lock it into the PEEK anchor s slots. F. Alan Barber. In group 4 a repair using the Meniscal Cinch (Arthrex, Naples, FL) (Fig 4) was placed in a vertical mattress configuration by use of a dual-needle insertion gun on the superior surface of the meniscus, 3 mm inside the meniscal cut and angled to orient 1 arm toward the superior peripheral capsule and the second toward the inferior meniscal capsule. This anchor has 2 PEEK anchors connected by No. 0 FiberWire (braided polyester and UHMWPE) (Arthrex) with a pre-tied sliding locking knot. In group 5 a repair using the MarXmen gun to insert the MaxFire repair device (Biomet Sports Medicine, Warsaw, IN) (Fig 5) was placed 3 mm inside the meniscal cut, creating a vertical repair. The MaxFire FIGURE 5. The MaxFire device (inserted by the MarXmen gun) is made from MaxBraid suture with 2 soft anchors consisting of polyethylene sleeves that bunch up to create anchors for stitch security. F. Alan Barber. device is made from MaxBraid suture (Biomet Sports Medicine) with 2 anchors consisting of polyethylene sleeves. After deploying, these sleeves bunch up to create anchors for stitch security. A pre-tied sliding locking knot is part of the repair. In group 6 a repair using the Linvatec Sequent meniscal repair device (ConMed Linvatec, Largo, FL) (Fig 6) was inserted, by use of the associated insertion device, starting at the periphery of the cut and then moving 3 mm inside the meniscal cut and back to the periphery again to create a V -shaped vertical repair. This repair device allows for a continuous stitching technique and comes with up to 7 PEEK anchors. Three of these anchors were used to create this V- shaped repair. These anchors are connected by No. 0 Hi-Fi (UHMWPE) suture (ConMed Linvatec), and the knots are created at each subsequent anchor by making 2 full clockwise rotations of the insertion device and tensioning the suture to lock it into the slots of the PEEK anchor. In group 7 a repair using the FasT-Fix 360 device (Smith & Nephew Endoscopy, Andover, MA) (Fig 7) was placed by use of a curved-needle insertion device in a vertical mattress configuration on the superior surface of the meniscus, 3 mm inside the meniscal cut and angled to orient 1 arm toward the superior peripheral capsule and the second toward the inferior meniscal capsule. This anchor also has 2 reconfigured PEEK anchors connected by No. 2-0 UltraBraid (UHMWPE) (Smith & Nephew Endoscopy) with a sliding locking knot. In contrast, the prior version of this device (Ultra FasT-Fix) contained a larger No. 0 UHMWPE suture. In addition, the PEEK anchors of this version are arrow shaped.
5 MENISCAL REPAIR DEVICES 831 TABLE 2. Failure Loads of s Completing 200 Cycles n Force (N) Mean SD Minimum Maximum Ethibond OrthoCord OmniSpan Cinch MarXmen/MaxFire Sequent FasT-Fix NOTE. The MarXmen/MaxFire completed fewer cycles than the other devices except for FasT-Fix 360 (P.05). FIGURE 7. The FasT-Fix 360 device has 2 arrow-shaped PEEK anchors connected by No. 2-0 UltraBraid with a sliding locking knot. F. Alan Barber. Statistical Analysis A general linear model was used to perform an analysis of variance for the data. If the data indicated statistical significance, Duncan multiple range tests were performed. Statistical significance was placed at P.05. RESULTS Mean load-to-failure data of the tested all-inside meniscal repairs (devices and sutures) are shown in Table 1. When compared with the Ethibond suture repair, the OrthoCord suture, OmniSpan, Cinch, Sequent, and FasT-Fix 360 repairs did not exhibit a TABLE 1. Load-to-Failure Data statistically significant difference. However, the Ethibond suture repair was superior to MarXmen/MaxFire with regard to load to failure (P.03). Because not all devices tested completed all 200 cycles, the mean number of cycles completed varied for the different devices: Ethibond suture, 140 cycles; OrthoCord suture, 160 cycles; OmniSpan, 121 cycles; Cinch, 140 cycles; MarXmen/MaxFire, 37 cycles; Sequent, 110 cycles, and FasT-Fix 360, 66 cycles. The MarXmen/MaxFire device completed statistically fewer cycles than the other devices except for FasT-Fix 360 (P.05). Considering only those tests that completed the full 200 cycles of testing, we generated load-to-failure data (Table 2). The MarXmen/MaxFire data should not be compared with the other devices because only 1 sample completed the cyclic loading. The amount of cyclic displacement at 100 cycles and 200 cycles was the second endpoint studied (Table 3). TABLE 3. Cyclic Displacement After 100 Cycles and 200 Cycles Cyclic Displacement (mm) Force (N) 100 Cycles 200 Cycles n Mean SD Minimum Maximum Ethibond OrthoCord OmniSpan Cinch MarXmen/MaxFire Sequent FasT-Fix NOTE. The mean failure load of the Ethibond suture repair was superior to that of the MarXmen/MaxFire repair (P.03). Ethibond OrthoCord OmniSpan Cinch MarXmen/MaxFire Sequent FasT-Fix NOTE. The MarXmen/MaxFire device showed more displacement at 100 cycles than the Ethibond suture and FasT-Fix 360 (P.05). The FasT-Fix 360 device showed less displacement at 200 cycles than the other repairs (P.05).
6 832 F. A. BARBER ET AL. *P.58. TABLE 4. Stiffness The MarXmen/MaxFire device showed statistically more displacement at 100 cycles than Ethibond suture and FasT-Fix 360 (P.05). The increase in displacement was less between 100 cycles and 200 cycles than for the first 100 cycles. The FasT-Fix 360 device showed statistically less displacement at 200 cycles than the other repairs (P.05). The third endpoint studied was stiffness. There was no statistical difference among the repair devices for stiffness (Table 4). The mode of failure during the cyclic testing was recorded. The specific modes of failure are listed in Table 5. DISCUSSION Stiffness (N/mm)* Ethibond 26.5 OrthoCord 27.0 OmniSpan Cinch 28.6 MarXmen/MaxFire 27.0 Sequent 31.7 FasT-Fix The development of all-inside meniscal repair devices with UHMWPE suture is a significant advance in the repair of meniscus tears. UHMWPE suture is significantly stronger than braided polyester suture. 19,25 The mechanical properties of these self-adjusting UHMWPE suture repair devices were reported in a porcine meniscus model. 19 The porcine meniscus model has been observed to show results similar to that of a young adult human meniscus with more consistent mechanical properties than those found in an older human cadaveric meniscus. 2 Our hypothesis was not supported by these data. The different UHMWPE suture containing devices did not show better structural properties with cyclic loading in the human meniscus model than previously reported. Several factors contributed to this. A comparison of this human cadaveric study to the porcine model data illustrates the inconsistency of the human cadaveric model. The strength and consistency of the young porcine tissue resulted in the suture (usually UHMWPE) breaking in 82% of the tests, 19 whereas in this study only 41% of the tests resulted in suture breaking. In contrast, the loads at failure are very similar when the porcine model data are compared with the human cadaveric model data. The bovine meniscus model also shows better consistency than the human model; however, its greater density results in higher failure loads than in the human model. 26 Performing the current study in human cadaveric meniscus allows for this comparison to the historical data for other biologic materials. The human cadaveric meniscus tissue was often the weakest point of the repair, as evidenced by the numerous specimens that failed by the suture cutting through the meniscus, the clamps pulling through the meniscus, or even the device anchors pulling through the meniscus. A comparison of the performance of the human meniscus in this test to that of porcine and bovine tests supports the conclusion that the human meniscus is not the best model for comparison testing of these devices. The younger porcine model is more consistent and provides a more reliable testing environment. The testing mechanism in this study was cyclic distraction followed by destructive loading. During the postoperative rehabilitation and healing phase, a meniscal repair is subjected to compression and shear loads but probably less distraction In addition, these loads can be under 10 N. 29 Previous studies have shown that distraction loads are not significant across the repair site. 27,28,30 Whereas 1 biomechanical study reported significantly higher shear failure loads than distraction failure loads for a suture-based device, in that study the reverse was true for braided polyester suture repairs. 31 No statistical difference between distraction and shear failure loads was shown for an earlier-generation rigid meniscal fixation device. 31 Other studies comparing distraction load testing and shear load testing have failed to show any significant difference between shear and distraction test modes for the meniscal repair devices tested. 32,33 It is there- TABLE 5. Failure Modes for s Tested Pulled Out Suture Broke Suture Pull Through Clamp Failure Ethibond 10 OrthoCord OmniSpan Cinch 6 4 MarXmen/MaxFire 4 6 Sequent 4 6 FasT-Fix NOTE. The Ethibond sutures consistently broke away from the knot. On the basis of prior studies, 23,24 slippage seldom occurs with Ethibond. OrthoCord pulled out of the meniscus with the suture and knot intact. This is an example of why the human meniscus has some limitations as a study model.
7 MENISCAL REPAIR DEVICES 833 fore reasonable to conclude at this point, in the absence of a better model, that it is valid to test meniscal repair device performance by destructive distraction after cyclic loading. Early designs using rigid meniscal repair devices did not provide repair strength comparable to that of a single braided polyester suture. 7,9,10,25 These early designs were typically rigid devices, which often left part of the device on the meniscal surface and were associated with reports of articular cartilage damage and device migration Testing comparing these earlier designs with suture-based devices showed, in addition to statistically lower failure loads, device breakage and migration as part of the failure mode. 32 Our study shows that the OmniSpan, Cinch, Sequent, and FasT-Fix 360 meniscal repair devices, all of which use UHMWPE-containing suture, provide repair strengths that are comparable to an inside-out repair performed with a braided polyester suture (Ethibond) or a UHMWPE-containing suture (OrthoCord). However, the MarXmen/MaxFire meniscal repair showed significantly lower mean failure loads, completed fewer load cycles, and had significantly greater displacement during the first 100 cycles of loading when compared with the inside-out repair using braided polyester suture. This lower failure strength and increased failure occurrence during cyclic loading of the MarXmen/ MaxFire device have been previously observed. 37 One possible reason for these results may be the device design. All of the other device designs (OmniSpan, Cinch, Sequent, and FasT-Fix 360) incorporate rigid PEEK capsular anchors. In contrast, the MaxFire is an all-suture device that does not use a rigid capsular anchor, which may affect its biomechanical properties in the human cadaveric meniscus. Even though the current data do not necessarily show an ultimate strength benefit with the UHMWPE suture containing devices, there are clear benefits with these devices that are responsible for their overwhelming clinical popularity. These self-adjusting suture devices allow an all-inside technique and are arguably easier to use than inside-out sutures. Secondary incisions associated with the inside-out technique are not required, and perhaps most importantly, the No. 2-0 braided polyester inside-out sutures were subject to frequent breaking during handling and knot tying whereas the UHMWPE sutures of comparable size are not. Various modes of failure were found for the selfadjusting repair devices, whereas Ethibond failed exclusively by suture breaking. Because of the distribution of failure modes and the limited number of specimens tested, no conclusion can be drawn from these data. The human cadaveric meniscus tissue used performed with greater variability than porcine meniscus used previously. Consequently, no device had all of its specimens survive 200 cycles. One weakness of this study is that the test protocol does not exactly re-create the mechanism by which a meniscal repair fails. The repairs were subjected to distractive forces, not compressive axially applied rotational loads or shear forces, during testing. Testing was carried out at room temperature and not in an isotonic aqueous environment. This was a time 0 test, which also decreases any impact of hydrolysis on the material over time. Precisely how much force is required to hold a repaired meniscus together during loaded knee flexion and extension is currently not established, and the data generated are only 1 aspect of repair device performance, which may not directly correlate with clinical healing. Consequently, any of these repair devices may be clinically appropriate. The values generated in this study cannot provide a direct application to performance in the clinical model, and there are no data generated on the potential for clinical healing or the potential for complications. The number of samples tested was limited by the availability of human menisci. Consequently, the finding that no difference existed for some comparisons does not mean that a difference did not exist ( error). Finally, a comparison of these data to the historical biologic alternatives suggests that a human cadaveric meniscus may not be the best biologic model for this testing. CONCLUSIONS The biomechanical properties of meniscal repairs using the OmniSpan, Cinch, Sequent, and FasT-Fix 360 devices are equivalent to suture repair techniques. However, the MarXmen/MaxFire meniscal repair device showed significantly lower failure loads and survived less cyclic loading in the human cadaveric meniscus than other tested repairs. Acknowledgment: The authors appreciate the assistance of Jennifer Heldreth in this research. REFERENCES 1. Morgan CD. The all-inside meniscus repair. Arthroscopy 1991;7: Post WR, Akers SR, Kish V. Load to failure of common meniscal repair techniques: Effects of suture technique and suture material. Arthroscopy 1997;13:
8 834 F. A. BARBER ET AL. 3. Rimmer MG, Nawana NS, Keene GC, Pearcy MJ. Failure strengths of different meniscal suturing techniques. Arthroscopy 1995;11: Maruyama M. The all-inside meniscal suture technique using new instruments. Arthroscopy 1996;12: Feng H, Hong L, Geng XS, Zhang H, Wang XS, Jiang XY. Second-look arthroscopic evaluation of bucket-handle meniscus tear repairs with anterior cruciate ligament reconstruction: 67 consecutive cases. Arthroscopy 2008;24: Espejo-Baena A, Figueroa-Mata A, Serrano-Fernandez J, et al. All-inside suture technique using anterior portals in posterior horn tears of lateral meniscus. Arthroscopy 2008;24:369.e1-369.e4. Available online at 7. Barber FA, Herbert MA. Meniscal repair devices. Arthroscopy 2000;16: Naqui SZ, Thiryayi WA, Hopgood P, Ryan WG. A biomechanical comparison of the Mitek RapidLoc, Mitek Meniscal repair system, Clearfix screws and vertical PDS and Ti-Cron sutures. Knee 2006;13: Becker R, Starke C, Heymann M, et al. Biomechanical properties under cyclic loading of seven meniscus repair techniques. Clin Orthop Relat Res 2002: Barber FA, Herbert MA, Richards DP. Load to failure testing of new meniscal repair devices. Arthroscopy 2004;20: Zantop T, Eggers AK, Weimann A, Hassenpflug J, Petersen W. Initial fixation strength of flexible all-inside meniscus suture anchors in comparison to conventional suture technique and rigid anchors: Biomechanical evaluation of new meniscus refixation systems. Am J Sports Med 2004;32: Albrecht-Olsen P, Lind T, Kristensen G, Falkenberg B. Failure strength of a new meniscus arrow repair technique: Biomechanical comparison with horizontal suture. Arthroscopy 1997; 13: Barber FA, Coons DA. Midterm results of meniscal repair using the BioStinger meniscal repair device. Arthroscopy 2006;22: Barber FA, Coons DA, Ruiz-Suarez M. Meniscal repair with the RapidLoc meniscal repair device. Arthroscopy 2006;22: Haas AL, Schepsis AA, Hornstein J, Edgar CM. Meniscal repair using the FasT-Fix all-inside meniscal repair device. Arthroscopy 2005;21: Siebold R, Dehler C, Boes L, Ellermann A. Arthroscopic all-inside repair using the Meniscus Arrow: Long-term clinical follow-up of 113 patients. Arthroscopy 2007;23: Barber FA, Schroeder FA, Oro FB, Beavis RC. FasT-Fix meniscal repair: Mid-term results. Arthroscopy 2008;24: Kalliakmanis A, Zourntos S, Bousgas D, Nikolaou P. Comparison of arthroscopic meniscal repair results using 3 different meniscal repair devices in anterior cruciate ligament reconstruction patients. Arthroscopy 2008;24: Barber FA, Herbert MA, Schroeder FA, Aziz-Jacobo J, Sutker MJ. Biomechanical testing of new meniscal repair techniques containing ultra high-molecular weight polyethylene suture. Arthroscopy 2009;25: Chang HC, Nyland J, Caborn DN, Burden R. Biomechanical evaluation of meniscal repair systems: A comparison of the Meniscal Viper Repair System, the vertical mattress FasT-Fix, and vertical mattress Ethibond sutures. Am J Sports Med 2005;33: Kohn D, Siebert W. Meniscus suture techniques: A comparative biomechanical cadaver study. Arthroscopy 1989;5: Kocabey Y, Chang HC, Brand JC Jr, Nawab A, Nyland J, Caborn DN. A biomechanical comparison of the FasT-Fix meniscal repair suture system and the RapidLoc device in cadaver meniscus. Arthroscopy 2006;22: Abbi G, Espinoza L, Odell T, Mahar A, Pedowitz R. Evaluation of 5 knots and 2 suture materials for arthroscopic rotator cuff repair: Very strong sutures can still slip. Arthroscopy 2006;22: Barber FA, Herbert MA, Beavis RC. Cyclic load and failure behavior of arthroscopic knots and high strength sutures. Arthroscopy 2009;25: Barber FA, Herbert MA, Coons DA, et al. Sutures and suture anchors Update Arthroscopy 2006;22:1063.e e9. Available online at Aros BC, Pedroza A, Vasileff WK, Litsky AS, Flanigan DC. Mechanical comparison of meniscal repair devices with mattress suture devices in vitro. Knee Surg Sports Traumatol Arthrosc 2010;18: Richards DP, Barber FA, Herbert MA. Compressive loads in longitudinal lateral meniscus tears: A biomechanical study in porcine knees. Arthroscopy 2005;21: Richards DP, Barber FA, Herbert MA. Meniscal tear biomechanics: Loads across meniscal tears in human cadaveric knees. Orthopedics 2008;31: Staerke C, Brettschneider O, Gröbel KH, Becker R. Tensile forces on sutures in the human lateral knee meniscus. Knee Surg Sports Traumatol Arthrosc 2009;17: Becker R, Brettschneider O, Gröbel KH, von Versen R, Stärke C. Distraction forces on repaired bucket-handle lesions in the medial meniscus. Am J Sports Med 2006;34: Fisher SR, Markel DC, Koman JD, Atkinson TS. Pull-out and shear failure strengths of arthroscopic meniscal repair systems. Knee Surg Sports Traumatol Arthrosc 2002;10: Brucker PU, Favre P, Puskas GJ, von Campe A, Meyer DC, Koch PP. Tensile and shear loading stability of all-inside meniscal repairs: An in vitro biomechanical evaluation. Am J Sports Med 2010;38: Zantop T, Temmig K, Weimann A, Eggers AK, Raschke MJ, Petersen W. Elongation and structural properties of meniscal repair using suture techniques in distraction and shear force scenarios: Biomechanical evaluation using a cyclic loading protocol. Am J Sports Med 2006;34: Barber FA. Chondral injury after meniscal repair with Rapid- Loc. J Knee Surg 2005;18: Kelly JD IV, Ebrahimpour P. Chondral injury and synovitis after arthroscopic meniscal repair using an outside-in mulberry knot suture technique. Arthroscopy 2004;20:e49-e Seil R, Rupp S, Dienst M, Mueller B, Bonkhoff H, Kohn DM. Chondral lesions after arthroscopic meniscus repair using meniscus arrows. Arthroscopy 2000;16:E Mehta VM, Terry MA. Cyclic testing of 3 all-inside meniscal repair devices: A biomechanical analysis. Am J Sports Med 2009;37:
FAST-FIX 360 Meniscal Repair System
FAST-FIX 360 Meniscal Repair System FAST-FIX 360 Meniscal Repair System All-Inside Meniscal Repair Knee Series Technique Guide as described by: Charles H. Brown, Jr., MD Nicholas Sgaglione, MD All-Inside
the revo / mini-revo shoulder fixation system s u r g i c a l t e c h n i q u e
the revo / mini-revo shoulder fixation system s u r g i c a l t e c h n i q u e REVO The following techniques are described by Stephen J. Snyder, M.D., Van Nuys, CA. Arthroscopic repair of the rotator
Bankart Repair using the Smith & Nephew BIORAPTOR 2.9 Suture Anchor
Shoulder Series Technique Guide *smith&nephew BIORAPTOR 2.9 Suture Anchor Bankart Repair using the Smith & Nephew BIORAPTOR 2.9 Suture Anchor Gary M. Gartsman, M.D. Introduction Arthroscopic studies of
Double Bone Plug Meniscus Reconstruction Surgical Technique
Double Bone Plug Meniscus Reconstruction Surgical Technique Double Bone Plug Meniscus Low Profile Reamer 2-0 FiberWire Meniscus Repair Needles Collared Pin & Coring Reamer Set RetroConstruction Drill Guide
The Material Difference. Options for Rotator Cuff Repair, Labral Repair and Suture Management
The Material Difference Options for Rotator Cuff Repair, Labral Repair and Suture Management The Material Difference Biomet Sports Medicine recognizes the benefit of material options. Many times surgeons
The rapid evolution of arthroscopic shoulder surgery
Technical Note Arthroscopic Repair of SLAP Lesions With a Bioknotless Suture Anchor Edward Yian, M.D., Conrad Wang, M.D., Peter J. Millett, M.D., and Jon J. P. Warner, M.D. Abstract: The diagnosis and
Shoulder Restoration System
Shoulder Restoration System PopLok Knotless Suture Anchor Simple, Secure, Versatile all-peek knotless anchor system for rotator cuff and instability repairs CO M M I T T ED TO I N N OVATI O N SURGICAL
Arthroscopic Shoulder Procedures. David C. Neuschwander MD. Shoulder Instability. Allegheny Health Network Orthopedic Associates of Pittsburgh
Arthroscopic Shoulder Procedures David C. Neuschwander MD Allegheny Health Network Orthopedic Associates of Pittsburgh Shoulder Instability Anterior Instability Posterior Instability Glenohumeral Joint
Biomechanics of Joints, Ligaments and Tendons.
Hippocrates (460-377 B.C.) Biomechanics of Joints, s and Tendons. Course Text: Hamill & Knutzen (some in chapter 2 and 3, but ligament and tendon mechanics is not well covered in the text) Nordin & Frankel
Achilles Tendon Repair, Operative Technique
*smith&nephew ANKLE TECHNIQUE GUIDE Achilles Tendon Repair, Operative Technique Prepared in Consultation with: C. Niek van Dijk, MD, PhD KNEE HIP SHOULDER EXTREMITIES Achilles Tendon Repair, Operative
Arthroscopic Shoulder Instability Repair Using the SUTUREFIX ULTRA Suture Anchor and SUTUREFIX ULTRA Instrumentation System
*smith&nephew SHOULDER TECHNIQUE GUIDE Arthroscopic Shoulder Instability Repair Using the SUTUREFIX ULTRA Suture Anchor and SUTUREFIX ULTRA Instrumentation System KNEE HIP SHOULDER EXTREMITIES Arthroscopic
Anatomic Percutaneous Ankle Reconstruction of Lateral Ligaments (A Percutaneous Anti ROLL)
Anatomic Percutaneous Ankle Reconstruction of Lateral Ligaments (A Percutaneous Anti ROLL) Mark Glazebrook James Stone Masato Takao Stephane Guillo Introduction Ankle stabilization is required when a patient
Knee Kinematics and Kinetics
Knee Kinematics and Kinetics Definitions: Kinematics is the study of movement without reference to forces http://www.cogsci.princeton.edu/cgi-bin/webwn2.0?stage=1&word=kinematics Kinetics is the study
Title of Presentation Arial Regular 38pt
Engineered Fascia ECM for Soft Tissue Repair Title of Presentation Arial Regular 38pt Date 15pts Author Name 15pts Author Title 15pts Cleveland Clinic Kathleen Derwin, PhD October 30, 2008 Rotator Cuff
SHOULDER INSTABILITY IN PATIENTS WITH EDS
EDNF 2012 CONFERENCE LIVING WITH EDS SHOULDER INSTABILITY IN PATIENTS WITH EDS Keith Kenter, MD Associate Professor Sports Medicine & Shoulder Reconstruction Director, Orthopaedic Residency Program Department
Stabilization of Acute Acromioclavicular Joint Dislocations using Dog Bone Button Technology Surgical Technique
Stabilization of Acute Acromioclavicular Joint Dislocations using Dog Bone Button Technology Surgical Technique AC Repair - Dog Bone Button Stabilization of Acute Acromioclavicular Joint Dislocations using
Mary LaBarre, PT, DPT,ATRIC
Aquatic Therapy and the ACL Current Concepts on Prevention and Rehab Mary LaBarre, PT, DPT,ATRIC Anterior Cruciate Ligament (ACL) tears are a common knee injury in athletic rehab. Each year, approximately
SHOULDER INSTABILITY. E. Edward Khalfayan, MD
SHOULDER INSTABILITY E. Edward Khalfayan, MD Instability of the shoulder can occur from a single injury or as the result of repetitive activity such as overhead sports. Dislocations of the shoulder are
How To Fix A Radial Head Plate
Mayo Clinic CoNGRUENT RADIAL HEAD PLATE Since 1988 Acumed has been designing solutions to the demanding situations facing orthopedic surgeons, hospitals and their patients. Our strategy has been to know
Meniscus. Loading of Meniscus. Meniscus Biomechanics. Rehabilitation Following Meniscus Repair Alone & Combined with ACL Reconstruction
Rehabilitation Following Meniscus Repair Alone & Combined with ACL Reconstruction Russ Paine, PT Director Memorial Hermann Sportsmedicine Rehabilitation Houston, Texas Meniscus Knee bends 2-4 million x
How To Know If You Can Recover From A Knee Injury
David R. Cooper, M.D. www.thekneecenter.com Wilkes-Barre, Pa. Knee Joint- Anatomy Is not a pure hinge Ligaments are balanced Mechanism of injury determines what structures get damaged Medial meniscus tears
The Trans-Rotator Cuff Approach to SLAP Lesions: Technical Aspects for Repair and a Clinical Follow-up of 31 Patients at a Minimum of 2 Years
The Trans-Rotator Cuff Approach to SLAP Lesions: Technical Aspects for Repair and a Clinical Follow-up of 31 Patients at a Minimum of 2 Years Stephen J. O Brien, M.D., Answorth A. Allen, M.D., Struan H.
Articular Cartilage Injury to the Knee: Current Concepts in Surgical Techniques and Rehabilitation Management
Articular Cartilage Injury to the Knee: Current Concepts in Surgical Techniques and Rehabilitation Management Combined Sections Meeting 2014 Las Vegas, Nevada, February 3 6, 2014 James L. Carey, MD, MPH
Combined lesions of the glenoid labrum include labral
9(1):10 14, 2008 Ó 2008 Lippincott Williams & Wilkins, Philadelphia T E C H N I Q U E Arthroscopic Repair of Combined Labral Lesions MAJ Brett D. Owens, MD, Bradley J. Nelson, MD, and COL Thomas M. DeBerardino,
Pre - Operative Rehabilitation Program for Anterior Cruciate Ligament Reconstruction
Pre - Operative Rehabilitation Program for Anterior Cruciate Ligament Reconstruction This protocol is designed to assist you with your preparation for surgery and should be followed under the direction
Flying Swan Arthroscopic Labral Repair using a Tensioned Suture Bridge Construct
SHOULDER TECHNIQUE GUIDE Flying Swan Arthroscopic Labral Repair using a Tensioned Suture Bridge Construct Andrew L. Wallace, MFSEM PhD FRCS FRACS Susan Alexander, MSc PhD FRCS KNEE HIP SHOULDER EXTREMITIES
www.ghadialisurgery.com
P R E S E N T S Dr. Mufa T. Ghadiali is skilled in all aspects of General Surgery. His General Surgery Services include: General Surgery Advanced Laparoscopic Surgery Surgical Oncology Gastrointestinal
Arthroscopy of the Hand and Wrist
Arthroscopy of the Hand and Wrist Arthroscopy is a minimally invasive procedure whereby a small camera is inserted through small incisions of a few millimeters each around a joint to view the joint directly.
Labral Repair. Surgical Protocol by Ronald Glousman, M.D. and Nicholas Sgaglione, M.D.
Labral Repair Surgical Protocol by Ronald Glousman, M.D. and Nicholas Sgaglione, M.D. It s small. It s strong. And it's all suture. The JuggerKnot Soft Anchor represents the next generation of suture anchor
QUESTION I HAVE BEEN ASKED TO REHAB GRADE II AND III MCL INJURIES DIFFERENTLY BY DIFFERENT SURGEONS IN THE FIRST 6WEEKS FOLLOWING INJURY.
QUESTION I HAVE BEEN ASKED TO REHAB GRADE II AND III MCL INJURIES DIFFERENTLY BY DIFFERENT SURGEONS IN THE FIRST 6WEEKS FOLLOWING INJURY. SOME ARE HINGE BRACED 0-90 DEGREES AND ASKED TO REHAB INCLUDING
Howell ACL System. Howell 65 Tibial Guide EZLoc Femoral Fixation Device WasherLoc Tibial Fixation Device Graft Choice
Howell ACL System Howell 65 Tibial Guide EZLoc Femoral Fixation Device WasherLoc Tibial Fixation Device Graft Choice Howell ACL System Howell 65 Tibial Guide EZLoc Femoral Fixation Device WasherLoc Tibial
Rehabilitation Guidelines for Meniscal Repair
UW Health Sports Rehabilitation Rehabilitation Guidelines for Meniscal Repair There are two types of cartilage in the knee, articular cartilage and cartilage. Articular cartilage is made up of collagen,
Labral Repair. Surgical Protocol by Ronald Glousman, M.D. and Nicholas Sgaglione, M.D.
Labral Repair Surgical Protocol by Ronald Glousman, M.D. and Nicholas Sgaglione, M.D. It s small. It s strong. And it's all suture. The JuggerKnot Soft Anchor represents the next generation of suture anchor
it s time for rubber to meet the road
your total knee replacement surgery Steps to returning to a Lifestyle You Deserve it s time for rubber to meet the road AGAIN The knee is the largest joint in the body. The knee is made up of the lower
PRINCIPLES AND TECHNIQUES OF TENDON AND LIGAMENT REPAIR
C H A P T E R 5 2 PRINCIPLES AND TECHNIQUES OF TENDON AND LIGAMENT REPAIR Jay D. Ryan, DPM INTRODUCTION Surgical repair of soft tissue structures, including tendon, ligament, or capsule can be quite challenging.
Portal Placement for Shoulder Arthroscopy: Basic to Advanced William B. Stetson, MD
Portal Placement for Shoulder Arthroscopy: Basic to Advanced William B. Stetson, MD 1. Cannulas Smooth Ribbed Lipped Partial Threaded Fully Threaded Flexible 5.75 mm, 6 mm, 7 mm & 8.25 mm x 7cm or 9 cm
.org. Rotator Cuff Tears: Surgical Treatment Options. When Rotator Cuff Surgery is Recommended. Surgical Repair Options
Rotator Cuff Tears: Surgical Treatment Options Page ( 1 ) The following article provides in-depth information about surgical treatment for rotator cuff injuries, and is a continuation of the article Rotator
Rehabilitation Guidelines for Knee Arthroscopy
Rehabilitation Guidelines for Knee Arthroscopy Arthroscopy is a common surgical procedure in which a joint is viewed using a small camera. This technique allows the surgeon to have a clear view of the
Technique Guide. VersiTomic. Michael A. Rauh, MD. Anterior Cruciate Ligament Reconstruction
Technique Guide VersiTomic Anterior Cruciate Ligament Reconstruction Michael A. Rauh, MD The opinions expressed are those of Dr. Rauh and are not necessarily those of Stryker VersiTomic Anterior Cruciate
Double Endobutton Technique for Repair of Complete Acromioclavicular Joint Dislocations. Volume 8, Issue 4 175
Techniques in Shoulder & Elbow Surgery 8(4):175 179, 2007 T E C H N I Q U E Double Endobutton Technique for Repair of Complete Acromioclavicular Joint Dislocations Steven Struhl, MD New York UniversityYHospital
Rehabilitation Guidelines for Anterior Shoulder Reconstruction with Arthroscopic Bankart Repair
Rehabilitation Guidelines for Anterior Shoulder Reconstruction with Arthroscopic Bankart Repair The anatomic configuration of the shoulder joint (glenohumeral joint) is often compared to a golf ball on
Brian P. McKeon MD Jason D. Rand, PA-C, PT Patient Information Sheet: Anterior Cruciate Ligament
Brian P. McKeon MD Jason D. Rand, PA-C, PT Patient Information Sheet: Anterior Cruciate Ligament The anterior cruciate ligament or ACL is one of the major ligaments located in the knee joint. This ligament
KNEE LIGAMENT REPAIR AND RECONSTRUCTION INFORMED CONSENT INFORMATION
KNEE LIGAMENT REPAIR AND RECONSTRUCTION INFORMED CONSENT INFORMATION The purpose of this document is to provide written information regarding the risks, benefits and alternatives of the procedure named
Rehabilitation Guidelines For SLAP Lesion Repair
Rehabilitation Guidelines For SLAP Lesion Repair The anatomic configuration of the shoulder joint (glenohumeral joint) is often compared to a golf ball on a tee. This is because the articular surface of
Biomechanical Evaluation of a Cable-Crimp System Designed for Repair of Tendons and Ligaments in the Hand
iomechanical Evaluation of a Cable-Crimp System Designed for Repair of Tendons and Ligaments in the Hand Joshua A. Gordon, MD 1 ;ErikMcDonald,S 2 ; Jenni M. uckley, PhD 3 ; and Leonard Gordon, MD 4 The
The Petrylaw Lawsuits Settlements and Injury Settlement Report
The Petrylaw Lawsuits Settlements and Injury Settlement Report KNEE INJURIES How Minnesota Juries Decide the Value of Pain and Suffering in Knee Injury Cases The Petrylaw Lawsuits Settlements and Injury
ACL Reconstruction with Bone-Tendon-Bone Transplants Using the ENDOBUTTON CL BTB Fixation System
Knee Series Technique Guide ACL Reconstruction with Bone-Tendon-Bone Transplants Using the ENDOBUTTON CL BTB Fixation System Stuart Fromm, M.D. As described by: Stuart Fromm, M.D. Black Hills Orthopedic
Shoulder Arthroscopy
Copyright 2011 American Academy of Orthopaedic Surgeons Shoulder Arthroscopy Arthroscopy is a procedure that orthopaedic surgeons use to inspect, diagnose, and repair problems inside a joint. The word
Dr. Benjamin Hewitt. Shoulder Stabilisation
Please contactmethroughthegoldcoasthospitaswityouhaveanyproblemsafteryoursurgery. Dr. Benjamin Hewitt Orthopaedic Surgeon Shoulder Stabilisation The shoulder is the most flexible joint in the body, allowing
Anteromedial ACL Reconstruction Surgical Technique for Soft Tissue Grafts
Anteromedial ACL Reconstruction Surgical Technique for Soft Tissue Grafts Advancing ACL Reconstruction with implant absorption and ossification of the Implant site 2,3,5 The MILAGRO BR Interference Screw
Frequently Asked Questions following Anterior Cruciate Ligament Reconstruction Surgery 1
Frequently Asked Questions following Anterior Cruciate Ligament Reconstruction Surgery 1 Will my knee be normal after surgery and recovery? Unfortunately, even with an ACL reconstructive procedure, it
MATERIALS AND METHODS
REDISTRIBUTION OF LOAD OF INJURED LOWER CERVICAL SPINE UNDER AXIAL COMPRESSION USING FEM E. C. Teo 1 and H. W. Ng 1 1 School of Mechanical and Production Engineering, Nanyang Technological University,
Mini TightRope CMC Surgical Technique
Mini TightRope CMC Surgical Technique Mini TightRope CMC Mini TightRope CMC Fixation The Mini TightRope provides a unique means to suspend the thumb metacarpal after partial or complete trapezial resection
Rotator Cuff Tears in Football
Disclosures Rotator Cuff Tears in Football Roger Ostrander, MD Consultant: Mitek Consultant: On-Q Research Support: Arthrex Research Support: Breg Research Support: Arthrosurface 2 Anatomy 4 major muscles:
Anterior Cruciate Ligament Reconstruction. ACL Rehab Protocol
Anterior Cruciate Ligament Reconstruction Rehab Protocol This rehabilitation protocol has been designed for patients following ACL reconstruction who anticipate returning to a high level of activity as
Field Evaluation and Management of Non-Battle Related Knee and Ankle Injuries by the ATP in the
Field Evaluation and Management of Non-Battle Related Knee and Ankle Injuries by the ATP in the JF Rick Hammesfahr, MD Editor s Note: Part Three consists of ankle injury evaluation and taping. Part Two
Michael K. McAdam, M.D. Orthopedic Surgeon Specializing in Arthroscopy and Sports Medicine
Michael K. McAdam, M.D. Orthopedic Surgeon Specializing in Arthroscopy and Sports Medicine Anterior Cruciate Ligament Injury Injury to the anterior cruciate ligament (ACL) is common, especially in athletic
TRIGEN INTERTAN vs Stryker Gamma3 summary
TRIGEN INTERTAN vs Stryker Gamma3 summary Initial stiffness of INTERTAN construct almost 4% greater than Gamma3 INTERTAN femoral head rotation and varus collapse initially up to 84% lower than Gamma3 After
Foot and Ankle Technique Guide Proximal Inter-Phalangeal (PIP) Fusion
Surgical Technique Foot and Ankle Technique Guide Proximal Inter-Phalangeal (PIP) Fusion Prepared in consultation with: Phinit Phisitkul, MD Department of Orthopedics and Rehabilitation University of Iowa
Lateral Ankle Instability Repair using TWINFIX Ti 3.5 mm Suture Anchors
*smith&nephew ANKLE TECHNIQUE GUIDE Lateral Ankle Instability Repair using TWINFIX Ti 3.5 mm Suture Anchors Prepared in Consultation with: James Calder, MD KNEE HIP SHOULDER EXTREMITIES Lateral Ankle Instability
Arthroscopic Subscapularis Repair
CHPTER 18 rthroscopic Subscapularis Repair mmar nbari, MD nthony. Romeo, MD n all-arthroscopic repair of the subscapularis tendon has seen significant interest in the past 10 years. s we refine our knowledge
The information contained in this document is intended for healthcare professionals only.
The information contained in this document is intended for healthcare professionals only. X3 Knee Technology The Next Generation in Highly Crosslinked Knee Polyethylene Innovation Matters As a pioneer
Ankle Syndesmosis Surgical Protocol by Timothy Charlton, M.D.
Featuring Ankle Syndesmosis Surgical Protocol by Timothy Charlton, M.D. ZipLoop Technology is a unique weave in which a single strand of braided polyethylene is woven through itself twice in opposite directions.
Business Plan XYZ Medical Device Company. Current Date
Business Plan XYZ Medical Device Company Current Date Strategic Objectives Three Year Sales Goal: Increase current 2007 sales of $450,000 by a minimum of 50% each year in order to attain and exceed sales
METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING
METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING Met E 206 MATERIALS LABORATORY EXPERIMENT 1 Prof. Dr. Rıza GÜRBÜZ Res. Assist. Gül ÇEVİK (Room: B-306) INTRODUCTION TENSION TEST Mechanical testing
Why knees hurt, and what you can do about it.
Why knees hurt, and what you can do about it. Table of Contents How Knees Work Common Conditions and Treatment Options Surgical Procedures START Start the Journey to Pain-Free Knees START What Sets Us
INJURIES OF THE HAND AND WRIST By Derya Dincer, M.D.
05/05/2007 INJURIES OF THE HAND AND WRIST By Derya Dincer, M.D. Hand injuries, especially the fractures of metacarpals and phalanges, are the most common fractures in the skeletal system. Hand injuries
With advances made in arthroscopic surgery, more
Suture Capsulorrhaphy Versus Capsulolabral Advancement for Shoulder Instability Andrew D. Kersten, M.D., Meredith Fabing, D.O., Scott Ensminger, B.S., Constantine K. Demetropoulos, Ph.D., Ross Cooper,
Basic Wound Closure & Knot Tying. Joslyn Albright, MD General Surgery PGY-4 Research Resident
Basic Wound Closure & Knot Tying Joslyn Albright, MD General Surgery PGY-4 Research Resident Objectives Provide basic information on commonly used suture materials Review general principles of wound closure
Preventing Knee Injuries in Women s Soccer
Preventing Knee Injuries in Women s Soccer By Wayne Nelson, DC, CCRS The United States has recently seen a rapid increase in participation of young athletes with organized youth soccer leagues. As parents
A New Weight-Bearing Meniscal Test and a Comparison With McMurray s Test and Joint Line Tenderness
A New Weight-Bearing Meniscal Test and a Comparison With McMurray s Test and Joint Line Tenderness Devrim Akseki, M.D., Özal Özcan, M.D., Hakan Boya, M.D., and Halit Pınar, M.D. Purpose: The purpose of
Hip Arthroscopy Product Guide
Hip Arthroscopy Product Guide Tissue Preserving Hip Solutions Recent advances in the knowledge of and treatment of femoroacetabular impingement have led to the development of minimally invasive techniques
ACL Injuries in Women Webcast December 17, 2007 Christina Allen, M.D. Introduction
ACL Injuries in Women Webcast December 17, 2007 Christina Allen, M.D. Please remember the opinions expressed on Patient Power are not necessarily the views of UCSF Medical Center, its medical staff or
Shoulder Instability. Fig 1: Intact labrum and biceps tendon
Shoulder Instability What is it? The shoulder joint is a ball and socket joint, with the humeral head (upper arm bone) as the ball and the glenoid as the socket. The glenoid (socket) is a shallow bone
with Medial Patellofemoral Ligament (MPFL) Reconstruction Surgical Protocol by Ronald Navarro, M.D.
with Medial Patellofemoral Ligament (MPFL) Reconstruction Surgical Protocol by Ronald Navarro, M.D. Features A unique weave in which a single strand of braided polyethylene is woven through itself twice
Posttraumatic medial ankle instability
Posttraumatic medial ankle instability Alexej Barg, Markus Knupp, Beat Hintermann Orthopaedic Department University Hospital of Basel, Switzerland Clinic of Orthopaedic Surgery, Kantonsspital Baselland
ACCELERATED REHABILITATION PROTOCOL FOR POST OPERATIVE POSTERIOR CRUCIATE LIGAMENT RECONSTRUCTION DR LEO PINCZEWSKI DR JUSTIN ROE
ACCELERATED REHABILITATION PROTOCOL FOR POST OPERATIVE POSTERIOR CRUCIATE LIGAMENT RECONSTRUCTION DR LEO PINCZEWSKI DR JUSTIN ROE January 2005 Rationale of Accelerated Rehabilitation Rehabilitation after
Tissue Reinforcement with Strattice Reconstructive Tissue Matrix following Correction of Severe Breast Deformity
Tissue Reinforcement with Strattice Reconstructive Tissue Matrix following Correction of Severe Breast Deformity Robert Cohen, MD, FACS* Paradise Valley, AZ Case summary A 41-year old woman with a history
The Treatment of Traumatic Recurrent Anterior Shoulder Instability with Arthroscopic Bankart Repair
J Med Sci 22;22(2):63-68 http://jms.ndmctsgh.edu.tw/22263.pdf Copyright 22 JMS Hsing-Ning Yu, et al. The Treatment of Traumatic Recurrent Anterior Shoulder Instability with Arthroscopic Bankart Repair
Experimental determination of cervical spine mechanical properties
Acta of Bioengineering and Biomechanics Vol. 10, No. 4, 2008 Experimental determination of cervical spine mechanical properties MAREK GZIK*, WOJCIECH WOLAŃSKI, DAGMARA TEJSZERSKA Division of General Mechanics
Sports Radiology : Making a diagnosis or solving a problem
Sports Radiology : Making a diagnosis or solving a problem Professor Dr John George University of Malaya Research Imaging Centre University of Malaya Medical Centre Kuala Lumpur, Malaysia. Abstract: Aim:
Malleolar fractures Anna Ekman, Lena Brauer
Malleolar fractures Anna Ekman, Lena Brauer How to use this handout? The left column is the information as given during the lecture. The column at the right gives you space to make personal notes. Learning
Football Medicine Strategies for Knee Injuries FIRST ANNOUNCEMENT. and call for papers
International Conference on Sports Rehabilitation and Traumatology FIRST ANNOUNCEMENT and call for papers Football Medicine Strategies for Knee Injuries In partnership with: 21 st -22 nd April, 2012 Stamford
.org. Ankle Fractures (Broken Ankle) Anatomy
Ankle Fractures (Broken Ankle) Page ( 1 ) A broken ankle is also known as an ankle fracture. This means that one or more of the bones that make up the ankle joint are broken. A fractured ankle can range
Anterior Cruciate Ligament Reconstruction
1 Anterior Cruciate Ligament Reconstruction Surgical Indications and Considerations Anatomical Considerations: The anterior cruciate ligament (ACL) lies in the middle of the knee. It arises from the anterior
Anterior Capsular Repair Rehabilitation Program Methodist Sports Medicine Center, Indianapolis, IN Department of Physical Therapy
Anterior Capsular Repair Rehabilitation Program Methodist Sports Medicine Center, Indianapolis, IN Department of Physical Therapy Anterior Capsule reconstruction is a surgical procedure utilized for anterior
Structural Axial, Shear and Bending Moments
Structural Axial, Shear and Bending Moments Positive Internal Forces Acting Recall from mechanics of materials that the internal forces P (generic axial), V (shear) and M (moment) represent resultants
All Suture Anchor. with IntelliBraid Technology
All Suture Anchor with IntelliBraid Technology Introduction Searching For a Better Option The ICONIX family of anchors symbolize the next generation of suture anchor technology. The all suture based system
Surgical Technique. Struan H Coleman MD, PhD
Surgical Technique Guide Struan H Coleman MD, PhD the Author Struan H. Coleman MD, PhD, specializes in Sports Medicine at Hospital for Special Surgery where he treats orthopedic conditions of the shoulder,
1 of 6 1/22/2015 10:06 AM
1 of 6 1/22/2015 10:06 AM 2 of 6 1/22/2015 10:06 AM This cross-section view of the shoulder socket shows a typical SLAP tear. Injuries to the superior labrum can be caused by acute trauma or by repetitive
Knotilus TM. Anchor Instability Repair. Technique Guide
Knotilus TM Anchor Instability Repair Technique Guide Instability Repair Using the Knotilus TM Anchor Introduction While the shoulder has more mobility than any other joint in the body, it is also the
Linvatec Has Taken Absorbable Technology to a New Level
2003 JUNE EDITION LINVATEC L EAST INVASIVE TECHNOLOGIES LINVATEC SPECIALISES IN HALL SURGICAL POWER EQUIPMENT Small and Large bone orthopaedics Otolaryngology ARTHROSCOPY Shaver systems Manual instruments
Meniscal Lesions in the Anterior Cruciate Insufficient Knee: the Accuracy of Clinical Evaluation
Meniscal Lesions in the Anterior Cruciate Insufficient Knee: the Accuracy of Clinical Evaluation Chathchai Pookarnjanamorakot MD*, Thongchai Korsantirat MD**, Patarawan Woratanarat MD* * Department of