Results of Proficiency Test Bisphenol A in Plastic May 2014
|
|
|
- Shauna Garrison
- 9 years ago
- Views:
Transcription
1 Results of Proficiency Test Bisphenol A in Plastic May 214 Organised by: Author: Corrector: Report: Spijkenisse, the Netherlands ing. R.J. Starink dr. R.G. Visser & ing. N. Boelhouwer iis14p4 July 214
2 CONTENTS 1 INTRODUCTION SET UP QUALITY SYSTEM PROTOCOL CONFIDENTIALITY STATEMENT SAMPLES ANALYSIS RESULTS STATISTICS GRAPHICS Z-SCORES EVALUATION EVALUATION PER SAMPLE PERFORMANCE EVALUATION FOR THE GROUP OF LABORATORIES DISCUSSION... 9 Appendices: 1. Data, statistical results and graphic results Details test test methods Number of participating laboratories per country Abbreviations and literature Bisphenol A in Plastic: iis14p4 page 2 of 18
3 1 INTRODUCTION Bisphenol A (BPA) is a chemical that is mainly used in combination with other chemicals to manufacture plastics and resins. For example, BPA is used in polycarbonate, a high performance transparent, rigid plastic. Polycarbonate is used to make food containers, such as returnable beverage bottles, infant feeding (baby) bottles, tableware (plates and mugs) and storage containers. Residues of BPA are also present in epoxy resins used to make protective coatings and linings for food and beverage cans and vats. BPA can migrate in small amounts into food and beverages stored in materials containing the substance. Bisphenol A is classified in Directive 29/48/EC under Regulation (EC) No 1272/28 as toxic. In the absence of any specific requirements, bisphenol A can be contained in toys in concentrations equal to or smaller than the relevant concentration established for the classification of mixtures containing it as CMRs, namely 5 % as from 2 July 213 and 3 % as from 1 June 215 respectively. It cannot be excluded that that concentration may lead to increased exposure to bisphenol A, compared to the migration limit of,1 mg/l for bisphenol A set by European standards EN 71-9:25+A1:27, EN 71-1:25 and EN 71-11:25. The determination of Bisphenol A in plastics is known to give problems with the comparability of laboratory results. However, no appropriate Bisphenol A reference materials are yet available. As an alternative, participation in a proficiency test may enable laboratories to check their performance. Therefore, a proficiency test (laboratory-evaluating interlaboratory study) for the determination of Bisphenol A in plastics was organized by the Institute for Interlaboratory Studies in April 214. In this new proficiency test iis14p4, 66 laboratories in 19 different countries did participate. See appendix 3 for the number of participating laboratories per country. In this report the results of the 214 proficiency test are presented and discussed. 2 SET UP The in Spijkenisse was the organiser of this proficiency test. It was decided to send two different plastic samples. The first sample, a polypropylene (PP) granulate, was especially prepared by a Chinese factory by addition of Bisphenol A to PP. The second sample, a PVC granulate, was especially prepared by a Chinese factory by addition of Bisphenol A to PVC and subsequent homogenization. Analyses for fit-for-use and homogeneity were subcontracted. The participants were asked to report the analytical results with one extra figure using the indicated units on the report form. These results with an extra figure are preferably used for statistical evaluation. Bisphenol A in Plastic: iis14p4 page 3 of 18
4 2.1 QUALITY SYSTEM The in Spijkenisse, the Netherlands, has implemented a quality system based on ISO/IEC 1743:21. This ensures strict adherence to protocols for sample preparation and statistical evaluation and 1% confidentiality of participant s data. Also customer s satisfaction is measured on a regular basis by sending out questionnaires. 2.2 PROTOCOL The protocol followed in the organisation was the one as described for proficiency testing in the report iis Interlaboratory Studies: Protocol for the Organisation, Statistics and Evaluation of April 214 (iis-protocol, version 3.3). This protocol can be downloaded via the FAQ page of the iis website CONFIDENTIALITY STATEMENT All data presented in this report must be regarded as confidential and for use by the participating companies only. Disclosure of the information in this report is only allowed by means of the entire report. Use of the contents of this report for third parties is only allowed by written permission of the. Disclosure of the identity of one or more of the participating companies will be done only after receipt of a written agreement of the companies involved. 2.4 SAMPLES Two different samples, one polypropylene and one PVC batch both artificially fortified to be positive on Bisphenol A (with respective approx..21 %M/M and.39%m/m), were selected. Both materials were divided over plastic bags, approx. 3 grams for each sample. The homogeneity of the subsamples was checked by determination of Bisphenol A (BPA) content on 7 stratified randomly selected subsamples. BPA in %M/M BPA in %M/M Sample # Sample # Sample # Sample # Sample # Sample # Sample # Sample # Sample # Sample # Sample # Sample # Sample # Sample # Table 1: homogeneity test results of the subsamples #1467 and #1468 From the above test results the repeatabilities were calculated and compared with.3 times the estimated reproducibility of EN14372:4 in agreement with the procedure of ISO 13528, Annex B2 in the next table; Bisphenol A in Plastic: iis14p4 page 4 of 18
5 BPA in %M/M BPA in %M/M r (observed) # r (observed) # reference method EN14372:4 EN14372:4.3 x R (ref. method) Table 2: evaluation of repeatabilities of BPA contents of the subsamples #1467 and #1468 As the observed repeatabilities of the results of the homogeneity tests were all in agreement with the target precision data, the homogeneity of subsamples #1467 and #1468 was assumed. To each of the participating laboratories, one sample of approx. 3 grams granulate, labelled #1467 and one sample of approx. 3 grams granulate (labelled #1468) were sent on April 3, ANALYSIS The participants were requested to determine and report the total Bisphenol A content on both samples #1467 and #1468. The participants were explicitly asked to treat the samples as if they were routine samples and to report the analytical results using the indicated units on the report form and not to round the test results, but report as much significant figures as possible. The participants were also asked not to report less than results which are above the detection limit, because such results can not be used for meaningful statistical calculations. To get comparable results a detailed report form, on which the units were prescribed and a report form on which some analytical details were requested, was sent together with each set of samples. Also a letter of instructions was added to the package. The laboratories were asked to complete the report form with the requested details of the methods used. 3 RESULTS During four weeks after sample despatch, the results of the individual laboratories were gathered. The original data are tabulated in the appendices of this report. The laboratories are presented by their code numbers. Directly after the deadline, a reminder fax was sent to those laboratories that had not yet reported. Shortly after the deadline, the available results were screened for suspect data. A result was called suspect in case the Huber Elimination Rule (a robust outlier test, see lit.5) found it to be an outlier. The laboratories that produced these suspect data were asked to check the results. Additional or corrected data are placed under 'Remarks' in the result tables in appendix 1. A list of abbreviations used in the tables can be found in appendix 3. Bisphenol A in Plastic: iis14p4 page 5 of 18
6 3.1 STATISTICS Statistical calculations were performed as described in the report iis Interlaboratory Studies: Protocol for the Organisation, Statistics and Evaluation of April 214 (iis-protocol, version 3.3) For the statistical evaluation the unrounded (when available) figures were used instead of the rounded results. Results reported as < or > were not used in the statistical evaluation. First, the normality of the distribution of the various data sets per determination was checked by means of the Lilliefors-test a variant of the Kolmogorov-Smirnov test and by the calculation of skewness and kurtosis. Evaluation of the three normality indicators in combination with the visual evaluation of the graphic Kernel density plot, lead to judgement of the normality being either unknown, OK, suspect or not OK. After removal of outliers, this check was repeated. Not all data sets proved to have a normal distribution, in which cases the statistical evaluation of the results should be used with due care. According to ISO 5725 (1986 and 1994, lit.8 and 9) the original results per determination were submitted subsequently to Dixon s, Grubbs' and Rosner outlier tests. Outliers are marked by D(.1) for the Dixon s test, by G(.1) or DG(.1) for the Grubbs test and by R(.1) for the Rosner General ESD test (ref. 15). Stragglers are marked by D(.5) for the Dixon s test, by G(.5) or DG(.5) for the Grubbs test and by R(.5) for the Rosner General ESD test (ref. 15). Both outliers and stragglers were not included in the calculations of averages and standard deviations. For each assigned value the uncertainty was determined in accordance with ISO Subsequently the calculated uncertainty was evaluated against the respective requirement based on the target reproducibility in accordance with ISO When the uncertainty passed the evaluation no remarks are made in the report. However, when the uncertainty failed the evaluation it is mentioned in the report and it will have consequences for the evaluation of the test results. Finally, the reproducibilities were calculated from the standard deviations by multiplying them with a factor of 2.8. Bisphenol A in Plastic: iis14p4 page 6 of 18
7 3.2 GRAPHICS In order to visualise the data against the reproducibilities from literature, Gauss plots were made, using the sorted data for one determination (see appendix 1). On the Y-axis the reported analysis results are plotted. The corresponding laboratory numbers are under the X-axis. The straight horizontal line presents the consensus value (a trimmed mean). The four striped lines, parallel to the consensus value line, are the +3s, +2s, -2s and -3s target reproducibility limits of the selected standard. Outliers and other data, which were excluded from the calculations, are represented as a cross. Accepted data are represented as a triangle. Furthermore, Kernel Density Graphs were made. This is a method for producing a smooth density approximation to a set of data that avoids some problems associated with histograms (see appendix 3, nos.13-14). Also a normal Gauss curve was projected over the Kernel Density Graph for reference. 3.3 Z-SCORES To evaluate the performance of the participating laboratories the z-scores were calculated. As it was decided to evaluate the performance of the participants in this proficiency test (PT) against the literature requirements, e.g. ASTM reproducibilities, the z-scores were calculated using a target standard deviation. These results in an evaluation independent of the spread of this interlaboratory study. The target standard deviation was calculated from the literature reproducibility by division with 2.8. When a laboratory did use a test method with a reproducibility that is significantly different from the reproducibility of the reference test method used in this report, it is strongly advised to recalculate the z-score, while using the reproducibility of the actual test method used this in order to evaluate whether the reported test result is fit-for-use. In case no literature reproducibility was available, other target values were used. In some cases literature repeatability is available; in other cases a reproducibility of a former iis proficiency test could be used and also the Horwitz equation can be used to estimate target reproducibility. The z-scores were calculated according to: z (target) = (result - average of PT) / target standard deviation Absolute values for z<2 are very common and absolute values for z>3 are very rare. The usual interpretation of z-scores is as follows: z < 1 good 1 < z < 2 satisfactory 2 < z < 3 questionable 3 < z unsatisfactory Bisphenol A in Plastic: iis14p4 page 7 of 18
8 4 EVALUATION In this interlaboratory study no problems were encountered during the execution. Five participants did not report any test results due to several unknown reasons. Finally, 6 laboratories reported 12 numerical results. Observed were 6 statistically outlying test results, which is 4.8% of all results. In proficiency studies outlier percentages of 3% - 7.5% are quite normal. 4.1 EVALUATION PER SAMPLE In this section the results are discussed per sample. Almost all participants reported to have used an in house test method. However, when evaluating the test method details provided by the participants, differences are not large. The majority of the participants reported to have used an ultrasonic bad as technique for release of the BPA. Due to the lack of a suitable test method with precision data, it was decided to use the requirements from the standardised method EN14372:4, Child use and care articles, Cutlery and feeding utensils, Safety requirements and tests for evaluation of the results of this interlaboratory study. Regretfully, only a relative within-laboratory standard deviation RSDr is given in EN14372:4. Multiplication of RSDr by 2.8 gives the repeatability. Multiplication of the repeatability by 3 gives a good estimate of the target reproducibility. For comparison the estimated reproducibility calculated using the Horwitz equation is also given. Sample #1467 Bisphenol A: The determination of BPA was very problematic at the level of.145 %M/M. Three statistical outliers were detected. The calculated reproducibility after rejection of the statistical outliers is not at all in agreement with the estimated reproducibility of EN14372:4. See also discussion (chapter 5) Sample #1468 Bisphenol A: The determination of BPA was problematic at the level of.35 %M/M. Three statistical outliers were detected. The calculated reproducibility after rejection of the statistical outliers is not in agreement with the estimated reproducibility of EN14372:4. See also discussion (chapter 5) Bisphenol A in Plastic: iis14p4 page 8 of 18
9 4.2 PERFORMANCE EVALUATION FOR THE GROUP OF LABORATORIES A comparison has been made between the reproducibilities as found for the group of participating laboratories and the estimated reproducibilities of EN14372:24 (R target ) in the next tables: Parameter Unit n Average 2.8 * sd R (target) Bisphenol A %M/M Table 4: overview of results for sample #1467 Parameter Unit n Average 2.8 * sd R (target) Bisphenol A %M/M Table 5: overview of results for sample # DISCUSSION A number of different test methods were reported to have been used. Most often in house (49 laboratories) was mentioned as test method used, followed by JETRO29 (5 laboratories) and EPA355C (3 laboratories). From the analytical details in appendix 2, it can be noticed that several different extraction techniques and solvents were used. It is remarkable to see that when only the reported results were evaluated of the laboratories that used Ultrasonic as release technique the calculated reproducibility was smaller (and the consensus value was higher) for sample #1467, while this had no consequences for the spread and consensus value of sample #1468. When for both samples the data set of the group, that used Ultrasonic as release technique, was evaluated separately for the type of solvent used. It is surprisingly to see that only the calculated reproducibility of sample #1468 for the group that used THF as solvent meets the estimated reproducibility of EN14372:4. The three other data sets did not improve or even got worse. Not surprisingly, the choice of solvent used is of utmost importance. Parameter Unit n Average 2.8 * sd R (target) Bisphenol A (only ultrasonic) %M/M Bisphenol A (DCM-ultrasonic) %M/M Bisphenol A (THF-ultrasonic) %M/M Table 6: overview of separate evaluation for sample #1467 Parameter Unit n Average 2.8 * sd R (target) Bisphenol A (only ultrasonic) %M/M Bisphenol A (DCM-ultrasonic) %M/M Bisphenol A (THF-ultrasonic) %M/M Table 7: overview of separate evaluation for sample #1468 It can be concluded that the observed spread in this interlaboratory study may not be caused by just one critical point in the analysis. Each participating laboratory will have to evaluate its performance in this study and decide about any corrective actions if necessary. Bisphenol A in Plastic: iis14p4 page 9 of 18
10 APPENDIX 1 Determination of Total Bisphenol A (BPA) on sample #1467; results in %M/M lab method value mark z(targ) remarks 11 INH C -.8 Reported (unit error?) 33 in house in house in house in house in house in house.6 C,R(.5) -7.1 First reported in house in house.7649 C -3.5 Reported (unit error?) 2139 INH in house in house in house in house GB/T in house in house in house in house EN in house in house in house in house in house in house in house.1286 C -.85 First reported in house JETRO JETRO in house JETRO JETRO EPA355C in house in house in house JETRO in house INH in house.76 C First reported in house in house in house in house in house in house in house EPA355C EPA355C W Results withdrawn, reported first 43 mg/kg 3172 in house.28 R(.5) in house.2576 R(.5) CPSD-AN in house.1158 C Reported (unit error?) 321 in house in house in house in house in house.1636 C.94 First reported.196 normality OK n 57 outliers 3 mean (n) st.dev. (n).4916 R(calc.).1375 R(EN14372:4).5488 Compare R(Horwitz) =.2174 Bisphenol A in Plastic: iis14p4 page 1 of 18
11 Kernel Density Bisphenol A in Plastic: iis14p4 page 11 of 18
12 Determination of Total Bisphenol A (BPA) on sample #1467; results in %M/M (evaluated per technique and solvent) Kernel Density After exclusion of non ultrasonic techniques normality OK n 49 outliers excl. mean (n).1511 st.dev. (n).4127 R(calc.).1156 R(EN14372:4).571 Compare R(Horwitz) = Kernel Density Only Ultrasonic - DCM Normality OK N 31 Outliers 2 mean (n).1513 st.dev. (n).4239 R(calc.).1187 R(EN14372:4).572 Compare R(Horwitz) = Kernel Density Only Ultrasonic - THF normality OK n 18 outliers mean (n).1493 st.dev. (n).399 R(calc.).194 R(EN14372:4).564 Compare R(Horwitz) =.223 Bisphenol A in Plastic: iis14p4 page 12 of 18
13 Determination of Total Bisphenol A (BPA) on sample #1468; results in %M/M lab method value mark z(targ) remarks 11 INH C Reported (unit error?) 33 in house in house in house in house in house in house.359 C,R(.5) First reported in house in house.6877 C,R(.5) Reported (unit error?) 2139 INH in house in house in house in house GB/T in house in house in house in house EN in house in house in house in house in house in house R(.5) in house.2734 C -.77 First reported in house JETRO JETRO in house JETRO JETRO EPA355C in house in house in house JETRO in house INH in house.319 C.34 First reported in house in house in house in house in house in house in house EPA355C EPA355C W Results withdrawn, first reported 497 mg/kg 3172 in house in house CPSD-AN in house C.39 Reported (unit error?) 321 in house in house in house in house in house normality OK n 57 outliers 3 mean (n).353 st.dev. (n) R(calc.) R(EN14372:4).1153 Compare R(Horwitz) =.485 Bisphenol A in Plastic: iis14p4 page 13 of 18
14 Kernel Density Bisphenol A in Plastic: iis14p4 page 14 of 18
15 Determination of Total Bisphenol A (BPA) on sample #1468; results in %M/M (evaluated per technique and solvent) Kernel Density After exclusion of non ultrasonic techniques normality suspect n 51 outliers excl. mean (n).354 st.dev. (n).6148 R(calc.).1721 R(EN14372:4).1154 Compare R(Horwitz) = Kernel Density Only Ultrasonic - DCM normality OK n 34 outliers mean (n).2952 st.dev. (n).6819 R(calc.).199 R(EN14372:4).1116 Compare R(Horwitz) = Kernel Density Only Ultrasonic - THF normality suspect n 18 outliers mean (n).3256 st.dev. (n).3837 R(calc.).174 R(EN14372:4).1231 Compare R(Horwitz) = Bisphenol A in Plastic: iis14p4 page 15 of 18
16 APPENDIX 2 Method information Lab maximum particle size technique to release BPA extraction solvent used quantification technique recovery checked 11 3x3 mm Ultrasonic Dichloromethane LC-MS Yes, 16% 33 As received Ultrasonic Dichloromethane HPLC/MS/MS Yes 339 As received Ultrasonic Methanol/Dichloromethane HPLC/MS/MS No 623 5x5 mm Ultrasonic Dichloromethane LC-MS Yes, <2 ul Ultrasonic Acetonitrile -- No mm Ultrasonic THF/methanol LC/MS/MS No Soxhlet Dichlormethane LC/MS/MS No 2132 As received Ultrasonic THF LC/MS/MS No 2135 As received Ultrasonic Methanol GC/MS No µm Ultrasonic Dichloromethane LC-FLD No x3x3 mm Ultrasonic THF LC/MS/MS No 2172 <2 mm Ultrasonic THF LC-MS Yes, x3x3 mm Ultrasonic THF LC/MS/MS No 219 As received ASE Acetone/Hexane GC/MS (Sim) No mm Ultrasonic THF/methanol HPLC mm Shaking Dichloromethane HPLC-MS Yes, 1.6% x3 mm Ultrasonic THF HPLC-MS No 2216 ground -- Dichloromethane HPLC-MS Yes, 11% Ultrasonic Dichloromethane HPLC-DAD Yes, 17% Ultrasonic THF -- Yes, 13% 2236 As received Ultrasonic Chloroform/methanol HPLC-FLD Yes, 11.6% 2237 >1mm Ultrasonic Toluene GC/MSD Yes, 94% x2x2 mm Ultrasonic Dichloromethane LC No 2247 As received Ultrasonic Chloroform/methanol LC/MS/MS No x5 mm Ultrasonic Dichloromethane HPLC-MS Yes, 85% 2266 As received Ultrasonic Methanol GC-MS No x5 mm Ultrasonic Methanol/Dichloromethane HPLC-FLD No 229 3x3 mm Ultrasonic Chloroform/methanol LC/MS/MS Yes, 95% x5 mm Ultrasonic Dichloromethane LC/DAD/MS Yes, 111% x2 mm Ultrasonic Dichloromethane/acetone -- Yes, 98.9% x5 mm Ultrasonic Dichloromethane LC/MS/MS No x2x2 mm Ultrasonic Dichloromethane HPLC/DAD/MS Yes, 91.3% 2363 <2x2 mm Ultrasonic Dichloromethane HPLC-FLD x2 mm Ultrasonic Dichloromethane HPLC/MS/FLD Yes, 1.6% x2x2 mm Ultrasonic Dichloromethane HPLC-FLD/LC/MS/MS Yes, 98.2 % 237 5x5 mm Ultrasonic Dichloromethane LC-MS Yes, 97.4 % mm Ultrasonic Dichloromethane LC/MS/MS Yes, 18% x2 mm Ultrasonic Dichloromethane HPLC/DAD/MS Yes, 95% 2386 As received Ultrasonic Dichloromethane LC/MS/MS No 239 As received Ultrasonic Dichloromethane HPLC-DAD/MS Yes, 133% 2413 As received Ultrasonic THF GC-MS No 2448 As received Reflux Cyclohexane HPLC-DAD Yes, 119% <.1 mm Ultrasonic Dichloromethane GC-MS No 251 1x1x1 mm Ultrasonic Dichloromethane HPLC-FLD No x5 mm Ultrasonic THF LC-MS Yes, 95% Ultrasonic THF/methanol -- No 267 As received Ultrasonic THF HPLC-FLD Yes, 9% 269 As received (#1467) Dissolving Toluene HPLC-FLD No 1 mm (#1468) Ultrasonic Dichloromethane HPLC-FLD No x.1 mm Ultrasonic THF HPLC-FLD No x3 mm Ultrasonic Chloroform/methanol HPLC-FLD/MS/MS Yes, 15% 3163 As received Thermaldesorption -- GC/MS Ultrasonic Dichloromethane LC-FLD/MS No 3182 No Ultrasonic Chloroform/methanol HPLC No 3199 As received Ultrasonic THF HPLC/MS Yes, 94.5% 329 5x5 mm Ultrasonic THF HPLC/MS/MS Yes, 95% 321 As received Ultrasonic THF HPLC-Fluorescence No 3218 As received Ultrasonic Chloroform/methanol HPLC-FLD Yes, 98% x3x3 mm Ultrasonic THF LC/MS/MS No 3233 As received Ultrasonic THF LC/MS Yes, 95.6% 3243 <2 mm Ultrasonic Dichloromethane GC-MS No Bisphenol A in Plastic: iis14p4 page 16 of 18
17 APPENDIX 3 Number of participating laboratories per country 6 labs in FRANCE 5 labs in GERMANY 5 labs in HONG KONG 2 labs in HUNGARY 3 labs in INDIA 1 lab in INDONESIA 1 lab in IRELAND 3 labs in ITALY 3 labs in KOREA 2 labs in P.R. of CHINA 1 lab in PAKISTAN 1 lab in SAUDI ARABIA 1 lab in SWITZERLAND 3 labs in TAIWAN R.O.C. 2 labs in THAILAND 1 lab in THE NETHERLANDS 2 labs in TURKEY 5 labs in U.S.A. 1 lab in VIETNAM Bisphenol A in Plastic: iis14p4 page 17 of 18
18 APPENDIX 4 Abbreviations: C D(.1) D(.5) G(.1) G(.5) DG(.1) DG(.5) R(.1) R(.5) n.a. n.d. fr = final result after checking of first reported suspect result = outlier in Dixon s outlier test = straggler in Dixon s outlier test = outlier in Grubbs outlier test = straggler in Grubbs outlier test = outlier in Double Grubbs outlier test = straggler in Double Grubbs outlier test = outlier in Rosner outlier test = straggler in Rosner outlier test = not applicable = not detected = first reported result Literature: 1 iis Interlaboratory Studies, Protocol for the Organisation, Statistics & Evaluation, January 21 2 ASTM E ASTM E ISO 5725: ISO 5725, parts 1-6, /84/EC, Official Journal of the European Communities, L34/26, (21) 7 98/485/EC, Official Journal of the European Comminities, L217/35, (1998) 8 M. Thompson and R. Wood, J. AOAC Int, 76, 926, (1993) 9 W.J. Youden and E.H. Steiner, Statistical Manual of the AOAC, (1975) 1 IP 367/84 11 DIN 3842 T41/42 12 P.L. Davies, Fr. Z. Anal. Chem, 331, 513, (1988) 13 J.N. Miller, Analyst, 118, 455, (1993) 14 ASTM F963: standard consumer safety specification on toy safety 15 Analytical Methods Committee Technical brief, No4 January The Royal Society of Chemistry 22, Analyst 22, 127 pages , P.J. Lowthian and M. Thompson 17 ISO/FDIS 13528, 25e, Statistical methods for use in proficiency testing by interlaboratory comparisons 18 R.G. Visser, Reliability of proficiency test results for metals and phthalates in plastics, Accred Qual Assur, 14:29-34 (29) 19 Annex XVII to REACH Regulation 197/26 2 Guideline on the interpretation of the concept which can be placed in the mouth as laid down in the entry 52 of Annex XVII to REACH Regulation 197/26, published July 18, 212 on C.F.R. Part 151 and 16 C.F.R C.F.R. 1199, Federal Register /Vol. 78, No. 31 /Thursday, February 14, Bernard Rosner, Percentage Points for a Generalized ESD Many-Outlier Procedure, Technometrics, 25(2), pp , (1983) Bisphenol A in Plastic: iis14p4 page 18 of 18
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.
Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description
Interlaboratory studies
Interlaboratory studies Department of Food Chemistry and Analysis, ICT Prague Vladimír Kocourek Prague, 2012 Interlaboratory studies Various titles: Interlaboratory proficiency test or studies Interlaboratory
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
European Union Reference Laboratory for Dioxins and PCBs in Feed and Food
for Dioxins and s in Feed and Food Chemisches und Veterinäruntersuchungsamt Freiburg PO Box 46 D-793 Freiburg Germany CEN Ring Trial - Animal Feed Ring trial on determination of PCDD/Fs and s in animal
NABL NATIONAL ACCREDITATION
NABL 162 NABL NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES GUIDELINES FOR PROFICIENCY TESTING PROGRAM for TESTING AND CALIBRATION LABORATORIES ISSUE NO : 04 AMENDMENT NO : 00 ISSUE
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Magruder Statistics & Data Analysis
Magruder Statistics & Data Analysis Caution: There will be Equations! Based Closely On: Program Model The International Harmonized Protocol for the Proficiency Testing of Analytical Laboratories, 2006
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
Benzo[a]pyrene Method Development
Benzo[a]pyrene Method Development Dr Michael Intorp CORESTA Cooperation Center for Scientific Research Relative to Tobacco FDA Center for Tobacco Products Scientific Workshop Tobacco Product Analysis 30
PTA proficiency testing for metal testing laboratories
PTA proficiency testing for metal testing laboratories BRIGGS Philip (Proficiency Testing Australia PO Box 7507, Silverwater, Rhodes NSW 2128 Australia ) Abstract: Proficiency testing is used internationally
Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)
Application Note 340 Determination of Fat in Dried Milk Products INTRODUCTION Many extraction techniques for the determination of fat in food are labor-intensive or require long extraction times. The Roese-Gottlieb
American Association for Laboratory Accreditation
Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Descriptive Statistics
Descriptive Statistics Suppose following data have been collected (heights of 99 five-year-old boys) 117.9 11.2 112.9 115.9 18. 14.6 17.1 117.9 111.8 16.3 111. 1.4 112.1 19.2 11. 15.4 99.4 11.1 13.3 16.9
The Basic Principles of Sieve Analysis
The Basic Principles of Sieve Analysis Introduction Many natural and manufactured materials occur in a disperse form, which means that they consist of differently shaped and sized particles. The particle
Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007
Proficiency Testing In A Global Analytical Environment Benefits & Challenges AOAC September 2007 1 The Coca-Cola Company The Coca-Cola Company is the largest manufacturer, distributor and marketer of nonalcoholic
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail [email protected] Web
Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team
Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for
Implementing New USP Chapters for Analytical Method Validation
Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: [email protected] Today s Agenda Handling Method Changes vs. Adjustments
Performance and advantages of qnmr measurements
Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure
CETANZ Technical Report TR 5. Grant Bosma Fulton Hogan. Report Date First Issue August 2011. Report Revision Date - Revision Number -
TECHNICAL REPORT ON PENETRATION OF BITUMINOUS MATERIALS CETANZ Technical Report TR 5 Author(s) Grant Bosma Fulton Hogan Report Date First Issue August 2011 Report Revision Date - Revision Number - Associated
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING. The Expression of Uncertainty in Quantitative Testing
European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING Accreditation of Laboratories Publication Reference EAL-G23 The Expression of Uncertainty in Quantitative Testing
Certificate of Analysis
The Japan Society for Analytical Chemistry Certificate of Analysis Certified Reference Material JSAC 0602-3 Plastics (tip form) for Chemical Analysis of Hazardous Metals This certified reference material
How To Comply With The New Toy Directive
An Expert s View of the Implications of the Intertek, Centre Court, Meridian Business Park, Leicester LE19 1WD UK [email protected] +44 (0)116 263 0330 www.intertek.com Implications of the
BT Premium Event Call and Web Rate Card
BT Managed Event and BT Self-Managed Event (also referred to as Express, Plus and Premium) Conference Bridge and Call for Booked Audio Conferencing Services will comprise the following for each phone-conference:
MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM AN ONLINE TOOL DESIGNED TO TAKE THE WORK OUT OF YOUR COMPENSATION REVIEW PROCESS
MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM AN ONLINE TOOL DESIGNED TO TAKE THE WORK OUT OF YOUR COMPENSATION REVIEW PROCESS MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM www.imercer.com/cars Mercer
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
A ROUND-ROBIN TO MEASURE THE DIRECT PIEZOELECTRIC COEFFICIENT USING THE BERLINCOURT METHOD
Technical Working Area 24 Performance Related Properties for Electroceramics A ROUND-ROBIN TO MEASURE THE DIRECT PIEZOELECTRIC COEFFICIENT USING THE BERLINCOURT METHOD Report No. 47 ISSN 1016-2186 June
Plastics Converting in Europe
Understanding the plastics industry Plastics Converting in Europe The European Plastics Converters (EuPC) is the leading EU-level Trade Association, based in Brussels, representing European Plastics Converters.
Author General Management Quality Assurance
Lab Ofichem B.V. Pag.: 1 / 14 Date: 20-01- Version: 02 Status: geldig Contents 1. General... 2 2. Shared Third Party audit program... 3 3. Reassurance absence of Conflict of Interest... 5 4. Contract structure...
An overview of using Hong Kong as a platform for trade and investment with China. Daniel Booth Director Vistra (Hong Kong)
An overview of using Hong Kong as a platform for trade and investment with China Daniel Booth Director Vistra (Hong Kong) Breda May, 2012 The role of Hong Kong A recognized and respected jurisdiction for
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
rtech laboratories began working toward ISO 9002:1994 in ISO 17025: Practical Benefits of Implementing a Quality System STATISTICAL ANALYSIS
1038 HONSA & MCINTYRE : JOURNAL OF AOAC INTERNATIONAL VOL. 86, NO. 5, 2003 STATISTICAL ANALYSIS ISO 17025: Practical Benefits of Implementing a Quality System JULIE D. HONSA and DEBORAH A. MCINTYRE rtech
European standardization of Solid Recovered Fuels
European standardization of Solid Recovered Fuels Chairman CEN/TC 343 Workshop: Processing routes for Solid Recovered Fuels 20 October, 2011 Dublin Outline Objectives and Challenges Background Mandate
LAB 37 Edition 3 June 2013
LAB 37 Edition 3 June 2013 Accreditation Requirements for Sampling and Testing in Accordance with the Drinking Water Testing Specification (DWTS) CONTENTS SECTION PAGE 1 Introduction 2 2 Regulatory requirements
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
NimkarTek Technical Services Pvt Ltd www.nimkartek.com
(A) s on RSLs RSLs A1 A2 A3 A4 A5 Name Understanding a Brand RSL Document Restricted Substances in Textiles Best Practices for RSL compliance Testing requirements for RSL compliance Fundamentals of RSL
European Union Regulations for packaging materials
European Union Regulations for packaging materials From March 01, 2005 onwards, all import shipments for the EU territory (25 member states as per Annex) have to comply with the wood packaging standard
Report on Government Information Requests
Report on Government Information July 1 - December 31, 2015 Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software and
R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 LABORATORIES
Page 1 of 7 R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 November 2009 2009 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
Digital vs Traditional Media Consumption
Digital vs Traditional Media Consumption Summary Comparing time spent on traditional and digital media at a global level as well as analyzing behaviors between countries and across the age groups GWI Q3
Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK
Schedule of United Kingdom Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK ISO/IEC 17021:2011 to provide environmental management systems certification Kitemark Court Davy Avenue Knowlhill
What Does the Normal Distribution Sound Like?
What Does the Normal Distribution Sound Like? Ananda Jayawardhana Pittsburg State University [email protected] Published: June 2013 Overview of Lesson In this activity, students conduct an investigation
Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
INTERNATIONAL COMPARISONS OF PART-TIME WORK
OECD Economic Studies No. 29, 1997/II INTERNATIONAL COMPARISONS OF PART-TIME WORK Georges Lemaitre, Pascal Marianna and Alois van Bastelaer TABLE OF CONTENTS Introduction... 140 International definitions
Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
Part 1: Background - Graphing
Department of Physics and Geology Graphing Astronomy 1401 Equipment Needed Qty Computer with Data Studio Software 1 1.1 Graphing Part 1: Background - Graphing In science it is very important to find and
Project Management Salary Survey Ninth Edition Project Management Institute Newtown Square, Pennsylvania, USA
Project Management Salary Survey Ninth Edition Project Management Institute Newtown Square, Pennsylvania, USA 2015 Project Management Institute, Inc. (PMI). All rights reserved. No part of this work may
Hybrid Wide-Area Network Application-centric, agile and end-to-end
Hybrid Wide-Area Network Application-centric, agile and end-to-end How do you close the gap between the demands on your network and your capabilities? Wide-area networks, by their nature, connect geographically
Density Curve. A density curve is the graph of a continuous probability distribution. It must satisfy the following properties:
Density Curve A density curve is the graph of a continuous probability distribution. It must satisfy the following properties: 1. The total area under the curve must equal 1. 2. Every point on the curve
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
Application of Welding Standards in Hong Kong. S. K. Babu Laboratory Manager - Foundation Techniques Ltd
Application of Welding Standards in Hong Kong S. K. Babu Laboratory Manager - Foundation Techniques Ltd Introduction: This paper summarizes the current welding standards practiced in Hong Kong & also outlines
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
Image Lab Software for the GS-900 Densitometer
Image Lab Software for the GS-900 Densitometer Quick Start Guide Catalog # 170-9690 Bio-Rad Technical Support For help and technical advice, please contact the Bio-Rad Technical Support department. In
Wine Proficiency test Organization 2009
DIRECTION NATIONALE 21-23, rue de Croulebarbe 75013 PARIS Tél.: 01 58 52 20 20 Fax: 01 58 52 20 30 e-mail : [email protected] www.oenologuesdefrance.fr Wine Proficiency test Organization
Governance, Risk and Compliance Assessment
Governance, Risk and Compliance Assessment Information security is a pervasive business requirement and one that no organisation can afford to get wrong. If it s not handled properly, your business could
Quality Assurance of test results using the techniques of statistical quality control According to ISO/IEC 17025:2005
Global Advanced Research Journal of Engineering, Technology and Innovation (ISSN: 2315-5124) Vol. 2(10) pp. 284-294, November, 2013 Available online http://garj.org/garjeti/index.htm Copyright 2013 Global
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
CLUB SURVEY 2015 BANKING ARGENTINA
CLUB SURVEY 2015 BANKING ARGENTINA BANKING CLUB SURVEY 2015 THE KEY TO DEVELOPING COMPETITIVE BANKING INDUSTRY-SPECIFIC COMPENSATION & BENEFITS PACKAGES With the objective of satisfying the needs of HR
In terms of food quality, you want to be absolutely confident that the results you get are reliable.
In terms of food quality, you want to be absolutely confident that the results you get are reliable. Welcome to the centre of excellence for food analysis Our strong core competences in food chemistry
RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES
RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES Reviewed by: Natalija Jovičić Zarić MSc, Quality Manager Approved by: Milica Lukešević, Acting Director Any
ApplicationNOTE Introduction
Introduction Honey, like other foods, is subject to strict quality control measures before it can be sold commercially. It is a natural product, produced solely by bees from flower and tree pollen. Human
Certifying Reference Materials to ISO Guide 34
Certifying Reference Materials to ISO Guide 34 Gill Holcombe Head of Reference Materials Production Environmental Laboratory Quality Day 17 November 2011 4005 Outline LGC background Quality systems - ISO
L 70/12 Official Journal of the European Union 9.3.2006
L 70/12 Official Journal of the European Union 9.3.2006 COMMISSION REGULATION (EC) No 401/2006 of 23 February 2006 laying down the methods of sampling and analysis for the official control of the levels
Chart 1: Zambia's Major Trading Partners (Exports + Imports) Q4 2008 - Q4 2009. Switzernd RSA Congo DR China UAE Kuwait UK Zimbabwe India Egypt Other
Bank of Zambia us $ Million 1. INTRODUCTION This report shows Zambia s direction of merchandise trade for the fourth quarter of 2009 compared with the corresponding quarter in 2008. Revised 1 statistics,
Reporting practices for domestic and total debt securities
Last updated: 4 September 2015 Reporting practices for domestic and total debt securities While the BIS debt securities statistics are in principle harmonised with the recommendations in the Handbook on
Dow Jones Titans Indices Methodology
Dow Jones Titans Indices Methodology S&P Dow Jones Indices: Index Methodology May 2016 Table of Contents Introduction 3 Highlights and Index Family 3 Eligibility Criteria and Index Construction 5 Dow Jones
CONSIDERATIONS WHEN CONSTRUCTING A FOREIGN PORTFOLIO: AN ANALYSIS OF ADRs VS ORDINARIES
THE APERIO DIFFERENCE. Authors Michael Branch, CFA Ran Leshem CONSIDERATIONS WHEN CONSTRUCTING A FOREIGN PORTFOLIO: AN ANALYSIS OF ADRs VS ORDINARIES U.S. investors can capture international equity exposure
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES October 2008 Revision 08 Great Lakes Institute for Environmental Research University of Windsor Windsor, Ontario
ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:
How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated
Chapter Test B. Chapter: Measurements and Calculations
Assessment Chapter Test B Chapter: Measurements and Calculations PART I In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1.
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
German proposal for the restriction of PAHs in consumer products
Federal Institute for Occupational Safety and Health Dortmund, 04/06/2010 German proposal for the restriction of PAHs in consumer products 1. Introduction The government of the Federal Republic of Germany
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
FOLPET (041) Folpet was first evaluated in 1969, and most recently re-evaluated in 1993.
619 FOLPET (041) EXPLANATION Folpet was first evaluated in 1969, and most recently re-evaluated in 1993. Residue data were received in 1993, but no recommendations could be made for citrus fruits, apples,
Format for Experiment Preparation and Write-Up
Format for Experiment Preparation and Write-Up Scientists try to answer questions by applying consistent, logical reasoning to describe, explain, and predict observations; and by performing experiments
Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
Report on Government Information Requests
Report on Government Information January 1 - June 30, 2015 Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software and
Normal and Binomial. Distributions
Normal and Binomial Distributions Library, Teaching and Learning 14 By now, you know about averages means in particular and are familiar with words like data, standard deviation, variance, probability,
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
BEIPH Final Report. QCMD 2010 Hepatitis B Virus DNA (HBVDNA10A) EQA Programme. William G MacKay on behalf of QCMD and its Scientific Council July 2010
QUALITY CONTROL for MOLECULAR DIAGNOSTICS The Altum Building, Todd Campus, West of Scotland Science Park, Glasgow, G20 0XA Scotland Tel: +44 (0) 141 945 6474 Fax: +44 (0) 141 945 5795 www.qcmd.org [email protected]
Appendix 1: Full Country Rankings
Appendix 1: Full Country Rankings Below please find the complete rankings of all 75 markets considered in the analysis. Rankings are broken into overall rankings and subsector rankings. Overall Renewable
CETANZ Technical Report TR 6. Report Date First Issue 13 November 2015. Report Revision Date - Revision Number 1
TECHNICAL REPORT ON CONE PENETRATION, PLASTIC LIMIT AND PLASTICITY INDEX TEST PROFICIENCY 2015 CETANZ Technical Report TR 6 Author(s) J. Ellis Report Date First Issue 13 November 2015 Report Revision Date
5 Correlation and Data Exploration
5 Correlation and Data Exploration Correlation In Unit 3, we did some correlation analyses of data from studies related to the acquisition order and acquisition difficulty of English morphemes by both
SEFAR PHARMA GMP solutions Life Science
SEFAR PHARMA GMP solutions Life Science Sefar Life Science Solutions Sefar responded to the stringent requirements of the pharmaceutical industry by creating the highest quality standards available. Developed
