Qualification, calibration and maintenance of stability chambers. Dr. Mathias Kramer

Size: px
Start display at page:

Download "Qualification, calibration and maintenance of stability chambers. Dr. Mathias Kramer"

Transcription

1 Qualification, calibration and maintenance of stability chambers Dr. Mathias Kramer

2 Agenda 1. Qualification of stability chambers 2. Calibration of equipment of stability chambers 3. Maintenance of stability chambers

3 Qualification 1. Qualification of stability chambers 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification

4 Qualification general aspects Qualification = Project management define good Team-Members, experts where needed QA, Engineer, technician, system owner (teamleader) the team determines your success approach: planning doing reporting using professional project management Tool (MS-Project) define milestones arrange periodical status- meetings

5 Qualification general aspects MS-Project-Plan:

6 Qualification 1. Qualification of stability chambers 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification

7 Qualification User Requirement Specification URS: Document to collect all requirements define usage very precisely (which kind of samples will be stored?) define climatic condition -nominal value, according ICH Q1A, WHO (e.g. accelerated conditions: 40 C ± 2 C / 75 % r.h. ± 5 % r.h.) - limits, according ICH Q1A, - additional conditions (e.g. Humidity in 5 C-Cells is not recommended, but very important)

8 Qualification User Requirement Specification URS: Document to collect all requirements define location, shape and area needed define equipment (shelves, drawer needed?) define safety and enviromental aspects (fire detector) define security requirements (on-line monitoring, locking system) define budget define basis and scale of qualification

9 Qualification analysis of local conditions Find the best location (building) for stability chambers consider flow line (Short way, minimal horizontal moving) infrastructure (water, air condition, additional deconstruction) safety and enviromental aspects (energy saving) security requirements (available access control-systems)

10 Qualification 1. Qualification of stability chambers: 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification

11 Qualification overview Qualification of a (new) climatic chamber: 4 Qualification Phases 1.Phase 2.Phase 3.Phase 4.Phase MQP DQ IQ OQ PQ MQR MQP: Master-Qualification Plan MQR: Master-Qualification Report DQ: Design Qualification IQ: Installation Qualification OQ: Operational Qualification PQ: Performance Qualification

12 Qualification 1. Phase: DQ-Design Qualification Start need of a climatic chamber write and approve DQ-Plan getting different tenders Check technical manuals URS available write and approve DQ-Report (with reported open issues) before starting IQ End Order of the designed chamber

13 Qualification 2. Phase: IQ-Installaltion Qualification Start delivery of the climatic chamber write and approve IQ-Plan compare ordered and delivered material Installation of the climatic chamber according local requirements / standards first function check prepare a (technical) logbook Installation of additional equipment (monitoring sensor or flash-light) write and approve IQ-Report (with reported open issues) before starting OQ End installed climatic chamber

14 Qualification 3. Phase: OQ-Operational Qualification Start installed climatic chamber write and approve OQ-Plan OQ-tests with empty chamber: running of the climatic cell: does it work as expected perform extensive Temperature- and Humidity-Room-Mapping testing any influence of the climatic conditions (light, open doors) simulate a power failure and protocol the behavior of the chamber to make conclusions find a representative position for the monitoring sensor write and approve OQ-Report (with reported open issues) before starting PQ End Climatic chamber ready for operation

15 Qualification 4. Phase: PQ-Performance Qualification Start running / working climatic chamber (without samples) write and approve PQ-Plan repeat the OQ-tests with filled chamber: running of the climatic cell: does it work as expected perform extensive Temperature- and Humidity-Room-Mapping testing any influence of the climatic conditions (light, open doors) simulate a power failure and protocol the behavior of the chamber to make conclusions Go live with the on-line monitoring, if available Calibrate the sensors write + approve PQ-Report (without any open issues) End completely qualified climatic chamber

16 Qualification 4. Phase: PQ-Performance Qualification perform extensive Temperature- and Humidity-Room-Mapping Possible data logger positions to map temperature and humidity

17 Qualification data logger to collect temperature and humidity data

18 Qualification 1. Qualification of stability chambers: 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification 2. Calibration of equipment: 3. Maintenance of stability chambers:

19 Qualification Periodic Re-Qualification Start running climatic chamber Strategy: Review of qualification status after a pre-defined time period write and approve Re-Qualification-Plan check all relevant parameters (i.e. MSR) is the change documentation complete? Repeat the Temperatur- and Humidity- Room-Mapping write and approve PQ-Report before writing the Final Report End re-qualified climatic chamber

20 Calibration 1. Qualification of stability chambers 2. Calibration of equipment of stability chambers 3. Maintenance of stability chambers

21 Calibration concepts Calibration of equipment (e.g. sensors for on-line monitoring) in place Calibration including whole measured distance one point calibration at set point Calibration of equipment (e.g. data-logger for temperature mapping) with a certified, adjusted master tool (Certified Institution) Calibration without measured distance Multi point calibration all set points must lie within calibration range Data logger must be suitable for intended conditions (e.g. 40 C / 75%r.H.)

22 Maintenance 3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations

23 Maintenance periodical Maintenance, technical support Defined maintenance according service-protocol: checking pressure, air flow, valves on-going maintenance including auxiliary media (steam, cooling water) support in any case of emergency, 24h a day

24 Maintenance service-protocol to document pressure

25 Maintenance 3. Maintenance of stability chambers: 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations

26 Maintenance microbiological monitoring and cleaning periodical microbiological monitoring according written procedure periodical cleaning of stability chambers including according written procedure, e.g. every 6 months define acceptance criteria

27 Maintenance 3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations

28 Maintenance on-line monitoring (Temperature and Humidity) confirm the defined position for any monitoring sensor which is used to visualize the climatic conditions, generating GMP-Data to document any deviations Date could be requested by inspectors monitoring sensors are calibrated annually

29 Maintenance on-line monitoring (calibrated sensor)

30 Maintenance 3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations

31 Maintenance Alarm Systems double- control of the climatic limits using another, different, independent alarm system

32 Maintenance 3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations

33 Maintenance Change Control Approved changes Planned change has to be assessed Relevant changes have to be approved by Quality Assurance before realization Emergency changes Approved by Quality Assurance after implementation of the change

34 Maintenance 3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations

35 Maintenance Temperature- or Humidity Deviations Short time deviations are allowed by ICH ( short term spikes due to opening of doors are accepted as unavoidable ) important deviations have to be documented (i.e. in a Deviation Report ) according ICH ( Excursions that exceed the defined tolerances for more than 24 hours should be describes in the study report )

36 Stability Chambers Overview F. Hoffmann-La Roche (post approval) all ICH conform conditions are available Chamber KTZ 1, 8, 10 KTZ 4 Conditions T ( C) r. H. (%) KTZ 3, KPS E +I KTZ 5 KTZ

37 Stability Chambers Example: 25 C/60% r.h. Chamber

38 Stability Chambers Example: 25 C/60% r.h. Chamber drawer

39 Stability Chambers Example: 5 C climatic chamber Calibrated online monitoring sensor (Pt100)

40 Stability Chambers Example: 25 C/60% r.h. climatic chamber Flash light (emergency case) Calibrated on-line monitoring sensor (Temperature and Humidity)

41 Stability Chambers Example: 25 C/60% r.h. climatic chamber access controlsystems escape route ramp

42 Dr. Mathias Kramer Thanks for your attention!

43 We Innovate Healthcare

Environmental Chamber. Stability/ GMP 2000 Series. Stability Testing Chambers

Environmental Chamber. Stability/ GMP 2000 Series. Stability Testing Chambers GMP 2000 Series ISO 9001:2000 21 CFR PART 11 COMPLIANCE SOFTWARE Stability/ Environmental Chamber Stability Testing Chambers Walk-in Test Chambers for Stability Tests on Pharmaceutical Products according

More information

Installation, Operation and Performance Qualification of a Thermoguard Installation

Installation, Operation and Performance Qualification of a Thermoguard Installation Installation, Operation and Performance Qualification of a Thermoguard Installation Thermoguard Qualification in Accordance with GMP To operate Thermoguard hardware and software within a regulated environment

More information

Qualification of temperature-controlled storage areas

Qualification of temperature-controlled storage areas WHO Vaccine Qualification of temperature-controlled storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport of time and temperature

More information

OPERATIONAL STANDARD

OPERATIONAL STANDARD 1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.

More information

(WHO Technical Report Series, No. 961, 2011), Annex 9. Temperature and humidity monitoring systems for fixed storage areas 107

(WHO Technical Report Series, No. 961, 2011), Annex 9. Temperature and humidity monitoring systems for fixed storage areas 107 Annex 5 Technical supplements to Model guidance for the storage and transport of timeand temperature-sensitive pharmaceutical products (WHO Technical Report Series, No. 961, 2011), Annex 9 1. The technical

More information

Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers

Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers Laboratory incubators (CL) and drying ovens (SL)

More information

FAT Factory Acceptance Testing SAT Site Acceptance Testing

FAT Factory Acceptance Testing SAT Site Acceptance Testing FAT Factory Acceptance Testing SAT Site Acceptance Testing Dr. Ingrid Walther May 2015, Page 1 Factory Acceptance Testing - FAT Site Acceptance Testing - SAT FAT and SAT - Important Milestones in the Project

More information

STD - STANDARD VERSION extended possibilities reasonable price wide range of optional equipment

STD - STANDARD VERSION extended possibilities reasonable price wide range of optional equipment ooled Incubators ooled incubators ooled incubators () are available in two versions: STD and TOP. STD STANDARD VERSION extended possibilities reasonable price wide range of optional equipment stainless

More information

Heating oven with mechanical control

Heating oven with mechanical control ED series 53 Drying ovens / Heating ovens Heating oven with mechanical control The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 C (572 F). Thanks

More information

Incubator with mechanical convection

Incubator with mechanical convection BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,

More information

Thermostatic Equipment

Thermostatic Equipment Thermostatic Equipment comming soon (06/2015): Climatic Chambers with LED lighting Climatic Chambers with steam humidifier Peltier - Cooled Incubators Refrigerators with back - up system ultra - Low Freezers

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice

Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single market, regulatory environment, industries under vertical legislation Pharmaceuticals and cosmetics Brussels, July 2001 Working Party on Control

More information

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known

More information

SECTION 15950 - TESTING, ADJUSTING, AND BALANCING

SECTION 15950 - TESTING, ADJUSTING, AND BALANCING SECTION 15950 - TESTING, ADJUSTING, AND BALANCING PART 1 - GENERAL 1.1 RELATED DOCUMENTS Drawings and general provisions of the Contract, including General and Supplementary Conditions and Division 1 Specification

More information

Heating oven with mechanical convection

Heating oven with mechanical convection FD series 53 Drying ovens / Heating ovens Heating oven with mechanical convection A BINDER FD series heating oven is always used when fast drying and sterilization is required. Thanks to its fully homogeneous

More information

Process Validation: Practical Aspects of the New FDA Guidance

Process Validation: Practical Aspects of the New FDA Guidance Process Validation: Practical Aspects of the New FDA Guidance ISPE Boston Chapter Meeting April 18, 2013 Rusty Morrison Commissioning Agents, Inc. Objectives / Summary What is Process Validation? Regulatory

More information

Temperature mapping of storage areas

Temperature mapping of storage areas QAS/14.598 Supplement 8 WHO Vaccine Temperature mapping of storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport of time

More information

PTB 311E Manual 3-in-1 Tablet Testing Instrument

PTB 311E Manual 3-in-1 Tablet Testing Instrument PTB 311E Manual 3-in-1 Tablet Testing Instrument The manual tablet testing instrument PTB 311E is a dual force mode hardness test apparatus as it can be used for either linear force or linear speed increase

More information

Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46. 1 Index... 2

Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46. 1 Index... 2 0932 Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46 1 Index 1 Index... 2 2 Description of the Inputs / Outputs... 4 2.1 Inputs / Module OMRON ID201 / ID211...4 2.2 Outputs / Module OMRON OC211...5

More information

To purchase a full bound copy of this document contact

To purchase a full bound copy of this document contact 29570 SUB Aqua Plus IQOQ document Unique Product Serial number label placed here To purchase a full bound copy of this document contact Grant Instruments (Cambridge) Ltd Shepreth Cambridgeshire SG8 6GB

More information

Temperature mapping of storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011

Temperature mapping of storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 WHO Vaccine Temperature mapping of storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport of time and temperature sensitive

More information

PT-MT3 Magnetic Tablet for dynamic calibration and perodical qualification of Tablet Harndness Testing Instruments

PT-MT3 Magnetic Tablet for dynamic calibration and perodical qualification of Tablet Harndness Testing Instruments PT-MT3 Magnetic Tablet for dynamic calibration and perodical qualification of Tablet Harndness Testing Instruments What is a PT-MT3 and why to use it? This instrument represents a system for calibrating

More information

Material test chamber with mechanical convection

Material test chamber with mechanical convection FP series 400 Material test chambers Material test chamber with mechanical convection A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and

More information

GUIDE TO GOOD MANUFACTURING PRACTICE FOR MEDICINAL PRODUCTS ANNEX 15 *

GUIDE TO GOOD MANUFACTURING PRACTICE FOR MEDICINAL PRODUCTS ANNEX 15 * PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PS/INF 11/2015 1 April 2015 GUIDE TO GOOD MANUFACTURING PRACTICE FOR MEDICINAL PRODUCTS ANNEX 15 * * Entry into force:

More information

Guidance on Qualification of existing facilities, systems, equipment and utilities

Guidance on Qualification of existing facilities, systems, equipment and utilities QUALIFICATION_EXISTING_EQUIPMENT_FINAL page 1 / 16 1. Acknowledgement...3 2. Introduction...3 3. Scope...4 4. Regulatory requirements...4 5. Guidance...4 5.1 Risk Assessment... 4 5.2 Procedure... 7 5.3

More information

A Stability Program for the Distribution of Drug Products

A Stability Program for the Distribution of Drug Products A Stability Program for the Distribution of Drug Products Teresa I. Lucas*, Rafik H. Bishara, and Robert H. Seevers Drug products must be transported in a manner that ensures products will be maintained

More information

Data Quality Assurance and Control Methods for Weather Observing Networks. Cindy Luttrell University of Oklahoma Oklahoma Mesonet

Data Quality Assurance and Control Methods for Weather Observing Networks. Cindy Luttrell University of Oklahoma Oklahoma Mesonet Data Quality Assurance and Control Methods for Weather Observing Networks Cindy Luttrell University of Oklahoma Oklahoma Mesonet Quality data are Trustworthy Reliable Accurate Precise Accessible Data life

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

Process Validation for Medical Devices

Process Validation for Medical Devices Process Validation for Medical Devices Dan O Leary CBA, CQE, CRE, CSSBB, CIRM, LLC 603-209-0600 [email protected] Copyright 2010 by, LLC Process Validation for Medical Devices 1 Instructor Introduction

More information

TECHNICAL SPECIFICATION. StirLIN-1 Economy Liquid Nitrogen System

TECHNICAL SPECIFICATION. StirLIN-1 Economy Liquid Nitrogen System TECHNICAL SPECIFICATION StirLIN-1 Economy Liquid Nitrogen System Reference 80 8120_01 Issue Date August 1, 2015 1. INTRODUCTION Since more than sixty years Stirling Cryogenics has designed and manufactured

More information

Freezers - Ultrafreezers. Platilab & Irilab Series

Freezers - Ultrafreezers. Platilab & Irilab Series Freezers - Ultrafreezers Platilab & Irilab Series PLATILAB: top security and performances with reliable temperature control. IRILAB: combining the high reliability of PLATILAB with the possibility to store

More information

GOOD LABORATORY PRACTICE (GLP) GUIDELINES FOR THE VALIDATION OF COMPUTERISED SYSTEMS. Working Group on Information Technology (AGIT)

GOOD LABORATORY PRACTICE (GLP) GUIDELINES FOR THE VALIDATION OF COMPUTERISED SYSTEMS. Working Group on Information Technology (AGIT) GOOD LABORATORY PRACTICE (GLP) GUIDELINES FOR THE VALIDATION OF COMPUTERISED SYSTEMS Working Group on Information Technology (AGIT) Release Date: 14 December 2007 Version: 02 AGIT - Validation of Computerised

More information

NSW Air Quality Monitoring Network: Quality Assurance Procedures

NSW Air Quality Monitoring Network: Quality Assurance Procedures NSW Air Quality Monitoring Network: Quality Assurance Procedures Copyright State of NSW and Office of Environment and Heritage With the exception of photographs, the State of NSW and Office of Environment

More information

Hygro-Thermometer + InfraRed Thermometer Model RH101

Hygro-Thermometer + InfraRed Thermometer Model RH101 User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative

More information

LABORATORY / MEDICINE / PHARMACEUTICAL REFRIGERATORS

LABORATORY / MEDICINE / PHARMACEUTICAL REFRIGERATORS TECHNOLOGY FOR LIFE LABORATORY / MEDICINE / PHARMACEUTICAL REFRIGERATORS Biomedical Refrigeration I LR / ML (with solid door) Safe storage of temperature-sensitive preparations at +4 C/ +5 C According

More information

Safety Storage Cabinets

Safety Storage Cabinets Safety Storage Cabinets Safe storage of flammables? The workplace without a fire resistant Safety Storage Cabinet Dangers, risks and your disadvantages... STORAGE ROOM Transport/working time - Increased

More information

Smart. Scalable. Simple.

Smart. Scalable. Simple. Thermo Scientific Wireless Monitoring Solution Smart. Scalable. Simple. providing the protection you need with the simplicity you deserve Thermo Scientific Wireless Monitoring Solution Optimized Sample

More information

DataView DVTH Data Logger for Temperature and Humidity

DataView DVTH Data Logger for Temperature and Humidity DataView DVTH Data Logger for Temperature and Humidity Logs temperature, humidity, and dew point Built-in real time text and graphical display Stand-alone or PCconnected operation Deep sample memory Battery-powered

More information

Training Course Computerized System Validation in the Pharmaceutical Industry Istanbul, 16-17 January 2003. Change Control

Training Course Computerized System Validation in the Pharmaceutical Industry Istanbul, 16-17 January 2003. Change Control Training Course Computerized System Validation in the Pharmaceutical Industry Istanbul, 16-17 January 2003 Change Control Wolfgang Schumacher Roche Pharmaceuticals, Basel Agenda Change Control Definitions

More information

US Environmental Protection Agency Office of Pesticide Programs

US Environmental Protection Agency Office of Pesticide Programs US Environmental Protection Agency Office of Pesticide Programs Office of Pesticide Programs Microbiology Laboratory Environmental Science Center, Ft. Meade, MD Standard Operating Procedure for Monitoring

More information

Example Retrocommissioning Measure: Opening Throttled Discharge Valves

Example Retrocommissioning Measure: Opening Throttled Discharge Valves Opening Throttled Discharge Valves This 21-story building, constructed in 1997, is located in Oregon and contains 589,000 gross square feet of mostly office occupancy. The HVAC system includes six large

More information

Contamination Control and Environmental Monitoring Systems

Contamination Control and Environmental Monitoring Systems Life Science Contamination Control and Environmental Monitoring Systems Hospital Electronics Brookhaven Instruments deliver, install and validate particle counters, Microbial Air Samplers and Monitoring

More information

OVERVIEW OF STANDARD 7E & STANDARD 20

OVERVIEW OF STANDARD 7E & STANDARD 20 Standard for Thermal Transport Packaging Used in Parcel Delivery System Shipment 7E 2 010 ISTA 7E Usage Modes VERSION DATE OCTOBER 2010 Initial Release ISTA, Distributing Confidence, Worldwide ISTA 7-Series

More information

Operating Principle of the Sample Positioning Flaps

Operating Principle of the Sample Positioning Flaps Test WEIGHT, THICKNESS, DIAMETER (or LENGTH), and HARDNESS of regular tablets, odd shaped tablets (caplets, oblongs etc.) and other solid samples fully automatically. Use the WHT 3ME for the fully automated

More information

Solar energy... ...Solar technology in endurance testing

Solar energy... ...Solar technology in endurance testing Solar energy......solar technology in endurance testing Solar & Photovoltaic systems - Environmental testing...... we are your partner Energy generated by the sun (solar energy) is considered forward looking

More information

Horizontal Steam Sterilizers. Healthcare Line. Since

Horizontal Steam Sterilizers. Healthcare Line. Since Horizontal Steam Sterilizers Healthcare Line Since 1925 Horizontal Steam Sterilizers for Healthcare Applications The Tuttnauer line of healthcare sterilizers are designed for hospital operating rooms,

More information

Climate chamber with illumination

Climate chamber with illumination KBWF series (E5.2) Climate chambers with illumination Climate chamber with illumination The BINDER climate chamber with illumination of the KBW series achieves homogeneous light distribution with its natural

More information

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification White paper System Suitability Tests and AIQ R.D.McDowall PhD Principal, McDowall Consulting Paul Smith European Validation Program Manager, PerkinElmer Nicola Vosloo PhD European Market Development Leader,

More information

Systec Laboratory Autoclaves Systec H-Series. Horizontal floor-standing autoclaves. Systec H-Series 2D. Pass-through autoclaves.

Systec Laboratory Autoclaves Systec H-Series. Horizontal floor-standing autoclaves. Systec H-Series 2D. Pass-through autoclaves. Fotolia Layout Systec Laboratory Autoclaves Systec H-Series. Horizontal floor-standing autoclaves. Systec H-Series 2D. Pass-through autoclaves. Performance and competence. Experience counts We focus on

More information

Cold Chain Shipping: Protecting Temperature Sensitive Products. Diane McLean Pfizer Global Manufacturing

Cold Chain Shipping: Protecting Temperature Sensitive Products. Diane McLean Pfizer Global Manufacturing Cold Chain Shipping: Protecting Temperature Sensitive Products Diane McLean Pfizer Global Manufacturing January 8, 2009 1 What is Cold Chain? The term cold chain or cold chain management refers to controlled

More information

BERMAD Fire Protection

BERMAD Fire Protection Model: 400E-2M BERMAD Electrically Controlled Deluge Valve with Easy Lock Manual Reset Installation Operation Maintenance Safety First BERMAD believes that the safety of personnel working with and around

More information

FIRE ALARM APPLICATION

FIRE ALARM APPLICATION FIRE ALARM APPLICATION ENUMCLAW FIRE DEPARTMENT 1330 Wells St, Enumclaw, WA 98022 Phone: 360-825-5544 Fax: 360-825-9442 COMMUNITY DEVELOPMENT, BUILDING DEPARTMENT 1309 Myrtle Ave Enumclaw WA 98022 Phone:

More information

Large Steam Sterilizers for the Life Sciences. Laboratory Line

Large Steam Sterilizers for the Life Sciences. Laboratory Line Large Steam Sterilizers for the Life Sciences Laboratory Line Large Steam Sterilizers for the Life Sciences Sterilization in a laboratory environment has its unique requirements. The types of loads that

More information

GAMP 5 and the Supplier Leveraging supplier advantage out of compliance

GAMP 5 and the Supplier Leveraging supplier advantage out of compliance GAMP 5 and the Supplier Leveraging supplier advantage out of compliance Paul Osborne Performance PharmaTech Ltd. Overview This document is designed to assist suppliers who wish to sell computer based equipment

More information

Medical Waste Autoclaves. Medical Waste Autoclaves

Medical Waste Autoclaves. Medical Waste Autoclaves Medical Waste Autoclaves Medical Waste Autoclaves Safe, Cost-Effective, Medical Waste Sterilization Treat Medical Waste with high performance Tuttnauer Medical Waste autoclave systems designed for small,

More information

BSL3 and BSL4 Autoclaves. Laboratory Line

BSL3 and BSL4 Autoclaves. Laboratory Line BSL3 and BSL4 Autoclaves Laboratory Line BSL3 and BSL4 Autoclaves In today s society the need for safe bio-containment systems is increasingly important. This is especially true in the sterilization process

More information

The Central Monitoring System we prove it.

The Central Monitoring System we prove it. - Reliable - Flexible - Compliant The Central Monitoring System we prove it. 2 intro intro 3 The Central Monitoring System A networked LAN wired or wireless sensor system Reliable There are several attributes

More information

USER S GUIDE 94150 Humidity and Temperature USB Datalogger

USER S GUIDE 94150 Humidity and Temperature USB Datalogger USER S GUIDE 94150 Humidity and Temperature USB Datalogger FEATURES Memory for 32,000 readings (16000 temperature and 16,000 humidity readings) Dew point indication Status Indication USB Interface User-Selectable

More information

Vacuum drying oven for non-flammable solvents

Vacuum drying oven for non-flammable solvents VD series 53 Vacuum drying ovens Vacuum drying oven for non-flammable solvents A BINDER vacuum drying oven of the VD series is impressive while gently drying with its homogeneous temperature distribution.

More information

Packaging Validation according to ISO 11607

Packaging Validation according to ISO 11607 Synergy makes sence Packaging Validation according to ISO 11607 MG-FSI72-104 Last revision: March 2011 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel. 33 (0)4 72 81 22 62 - Fax : 33 (0)4

More information

Microwave Meter. Instruction Manual

Microwave Meter. Instruction Manual Microwave Meter 840046 Instruction Manual Microwave Meter 840046 Copyright 2009 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted

More information

Technology: Heating, Ventilation and Air Condition (HVAC) Zone Controls

Technology: Heating, Ventilation and Air Condition (HVAC) Zone Controls Accelerated Capital Allowances Eligibility Criteria Ref No.: HZC C2.1 Category: Process and Heating, Ventilation and Air conditioning (HVAC) Control Systems Technology: Heating, Ventilation and Air Condition

More information

Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi

Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi Instrument/Method Validation Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi Instrument/Method Validation QUESTION AND ANSWER PRESENTATION Instrument Validation What is instrument validation?

More information

Model: 400E-2M. Bermad Electrically Controlled Deluge Valve with Easy Lock Manual Reset. Installation Operation Maintenance. Application Engineering

Model: 400E-2M. Bermad Electrically Controlled Deluge Valve with Easy Lock Manual Reset. Installation Operation Maintenance. Application Engineering Model: 400E-2M Bermad Electrically Controlled Deluge Valve with Easy Lock Manual Reset Installation Operation Maintenance Application Engineering BERMAD 1. Safety First BERMAD believes that the safety

More information

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam

More information

Temperature and humidity monitoring systems for fixed storage areas

Temperature and humidity monitoring systems for fixed storage areas WHO Vaccine Temperature and humidity monitoring systems for fixed storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport

More information

CATALOGUE 2015. products of POL-EKO-APARATURA. Your partner in laboratory analysis and technological processes

CATALOGUE 2015. products of POL-EKO-APARATURA. Your partner in laboratory analysis and technological processes CATALOGUE 2015 products of POLEKOAPARATURA Your partner in laboratory analysis and technological processes POLEKOAPARATURA sp.j. POLEKOAPARATURA has been active in the polish market since 1990. In 2006

More information

VALIDATION AND QUALIFICATION

VALIDATION AND QUALIFICATION VALIDATION AND QUALIFICATION Lizette Caballero, B.S., M.T. (ASCP) Laboratory Manager Cellular Therapy Laboratory Florida Hospital Cancer Institute Orlando, FL 1 Learning Objectives Validation of Equipment

More information

EL-USB-2-LCD FEATURES WINDOWS CONTROL SOFTWARE DISPLAY FUNCTIONS. www.lascarelectronics.com Page 1 of 5

EL-USB-2-LCD FEATURES WINDOWS CONTROL SOFTWARE DISPLAY FUNCTIONS. www.lascarelectronics.com Page 1 of 5 This standalone data logger measures and stores up to 16,379 relative humidity and 16,379 temperature readings over 0 to 100%RH and -35 to +80 C (-31 to +176 F) measurement ranges. The user can easily

More information

Scope of Work. See www.sei.ie for resources available

Scope of Work. See www.sei.ie for resources available HVAC SWG Summary Background Pool energy-efficiency knowledge Increase cost-competitiveness Develop resources aimed at assisting in reducing HVAC energy consumption Address barriers to energy-saving HVAC

More information

LABOKLAV. Autoclaves Made in Germany. www.shp-steriltechnik.de

LABOKLAV. Autoclaves Made in Germany. www.shp-steriltechnik.de LABOKLAV Autoclaves Made in Germany www.shp-steriltechnik.de 2 LABOKLAV STEAM STERILIZERS Steam sterilization is the sterilization method with the widest range of applications efficient, safe, reproducible

More information

User Guide. Temperature and Humidity Datalogger. Model 42280

User Guide. Temperature and Humidity Datalogger. Model 42280 User Guide Temperature and Humidity Datalogger Model 42280 Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger. The 42280 is a wall-mount, tripod

More information

Vertical and Benchtop Autoclaves for Life Sciences. Laboratory Line

Vertical and Benchtop Autoclaves for Life Sciences. Laboratory Line Vertical and Benchtop Autoclaves for Life Sciences Laboratory Line Designed for Your Laboratory Applications Advanced Laboratory Autoclaves Tuttnauer laboratory autoclaves have been designed to provide

More information

ICH Topic Q 1 A Stability Testing Guidelines: Stability Testing of New Drug Substances and Products

ICH Topic Q 1 A Stability Testing Guidelines: Stability Testing of New Drug Substances and Products The European Agency for the Evaluation of Medicinal Products Human Medicines Evaluation Unit CPMP/ICH/380/95 ICH Topic Q 1 A Stability Testing Guidelines: Stability Testing of New Drug Substances and Products

More information

Wireless Data Logger Monitoring Systems

Wireless Data Logger Monitoring Systems Wireless Data Logger Monitoring Systems Table of Contents About MadgeTech, Inc. Wireless Systems & Components Wireless Data Loggers 2000A Wireless Data Logger Models MadgeTech Cloud Services MadgeTech

More information

Training Material for Balances (1) Prepared by R&D Division 1. Marketing Department A&D Company, Limited

Training Material for Balances (1) Prepared by R&D Division 1. Marketing Department A&D Company, Limited Training Material for Balances (1) Prepared by R&D Division 1 Marketing Department A&D Company, Limited 1 Contents Things you should know before using a balance --------------------------- P.3 1. Convection

More information

STERILIZERS, LABORATORY DRYING OVENS

STERILIZERS, LABORATORY DRYING OVENS TS9026 TS9053 TS9135 TS9430 TS 9000 SERIES: STERILIZERS, LABORATORY DRYING OVENS Series TS9000 consists of four different cabinets available in sizes from 26 litres to 430 litres. Interior housing and

More information

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM Purpose: To provide licensees and registrants with a

More information

GMP Pre-Qualification of Separators Ex Works Flexible and Safe on the GEA Pharmaceutical Test Bay

GMP Pre-Qualification of Separators Ex Works Flexible and Safe on the GEA Pharmaceutical Test Bay GMP Pre-Qualification of Separators Ex Works Flexible and Safe on the GEA Pharmaceutical Test Bay engineering for a better world GEA Mechanical Equipment Customer Specific, Economical and Safe For manufacturers

More information

Guide Specification for Laboratory Fume Hood Commissioning

Guide Specification for Laboratory Fume Hood Commissioning Guide Specification for Laboratory Fume Hood Commissioning 3 February, 1997 Application notes: This specification is intended to be used by qualified, experienced engineers and industrial hygienists as

More information

Dual Laser InfraRed (IR) Thermometer

Dual Laser InfraRed (IR) Thermometer User Manual Dual Laser InfraRed (IR) Thermometer MODEL 42570 Introduction Congratulations on your purchase of the Model 42570 IR Thermometer. This Infrared thermometer measures and displays non-contact

More information

A WHO guide to good manufacturing practice (GMP) requirements

A WHO guide to good manufacturing practice (GMP) requirements WHO/VSQ/97.02 ENGLISH ONLY DISTR.: LIMITED A WHO guide to good manufacturing practice (GMP) requirements Part 2: Validation Written by: Gillian Chaloner-Larsson, Ph.D, GCL Bioconsult, Ottawa Roger Anderson,

More information

LIBERO: Innovation Leader in Cold Chain Monitoring

LIBERO: Innovation Leader in Cold Chain Monitoring LIBERO: Innovation Leader in Cold Chain Monitoring we prove it. Efficient cold chain monitoring Cold chain monitoring is more than just a monitor logging temperature. It is about simplifying your entire

More information

Audit Report in the framework of the APIC Audit Programme

Audit Report in the framework of the APIC Audit Programme Audit Report in the framework of the APIC Audit Programme Company Location Address Country Subject of Audit Audit Date Auditor (lead) Co-Auditor(s) Participants Signature of Auditor(s) Lead Auditor Co-Auditor

More information

CANNON T4 MINI / MICRO DATA CENTRE SYSTEMS

CANNON T4 MINI / MICRO DATA CENTRE SYSTEMS CANNON T4 MINI / MICRO DATA CENTRE SYSTEMS Air / Water / DX Cooled Cabinet Solutions Mini Data Centre All in One Solution: Where there is a requirement for standalone computing and communications, or highly

More information

QFP910. Water detector. Building Technologies. CE1N2732en 2016-03-01. Synco living

QFP910. Water detector. Building Technologies. CE1N2732en 2016-03-01. Synco living 2 732 Synco living Water detector QFP910 Wireless radio sensor to record water puddles RF communication based on KNX standard (868 MHz unidirectional) Battery operation with standard 1.5 V batteries Use

More information

The leading voice in Nuclear Healthcare

The leading voice in Nuclear Healthcare Calder Group Nuclear Healthcare The leading voice in Nuclear Healthcare Introduction Shaping the future Together, the Calder Group delivers the only world-class vertical integration in the Nuclear Healthcare

More information

SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES (GMP): VALIDATION

SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES (GMP): VALIDATION WORLD HEALTH ORGANIZATION ORGANISATION MONDIALE DE LA SANTE Working document QAS/03.055/Rev.2 RESTRICTED SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES (GMP): VALIDATION This document has followed

More information

Hospital Sterilizers. T-Max Line

Hospital Sterilizers. T-Max Line Hospital Sterilizers T-Max Line Hospital Sterilizers Sterilization is a complex process. Like any complex process, many things can go wrong. A Hospital depends on its CSSD: Central Sterile Services Department.

More information

REAL TIME LABORATORY MONITORING SYSTEM

REAL TIME LABORATORY MONITORING SYSTEM REAL TIME LABORATORY MONITORING SYSTEM MONITORING LET XILTRIX WATCH OVER YOUR LABORATORY! 20 years experience and IKS expertise makes XiltriX the total monitoring solution for any laboratory. All day,

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

PROCEDURE FOR FRESH APPLE AND PEAR EXPORTS FROM SOUTH AFRICA TO MEXICO. Responsibility: Service Manager: Special Shipments Name: Initial:

PROCEDURE FOR FRESH APPLE AND PEAR EXPORTS FROM SOUTH AFRICA TO MEXICO. Responsibility: Service Manager: Special Shipments Name: Initial: Page 1 of 10 Rev: 0 1. Reference 1.1 This procedure refers to the cold treatment of fruit flies and Eastern fruit moth in apples and pears, as prescribed by the Secretariat of Agriculture, Cattle and Rural

More information

Integrating Good Manufacturing Practices During the Transition from Clinical Trials to Commercial Manufacturing

Integrating Good Manufacturing Practices During the Transition from Clinical Trials to Commercial Manufacturing Integrating Good Manufacturing Practices During the Transition from Clinical Trials to Commercial Manufacturing By Mark Zemler W H I T E P A P E R Executive Summary In support of our customers, this whitepaper

More information

Fire. The Fire Installers Mate. A guide to fire alarm systems design

Fire. The Fire Installers Mate. A guide to fire alarm systems design Fire The Fire Installers Mate A guide to fire alarm systems design BS 5839 Part 1:2002 A guide to BS 5839 Part 1:2002 Disclaimer This booklet is not intended to be a comprehensive guide to all aspects

More information

Validation of process gas systems

Validation of process gas systems Validation of Process Gas Systems By Jeff Hargroves Alza Corporation Validation of process gas systems involve documenting the expected system behavior, and verifying that the system performs as expected.

More information