Polyethylene Hip Resurfacing for Women
|
|
|
- Martina Knight
- 9 years ago
- Views:
Transcription
1 1 ORIGINAL ARTICLE Polyethylene Hip Resurfacing for Women James W. Pritchett MD Abstract Background: Since 1995, most hip resurfacing procedures have been performed using a metal-on-metal prosthesis with excellent functional results. However, there have been concerns about metallosis, particularly for women. Also, there may be a higher early revision rate compared to total hip replacement. These concerns suggest there may be a role for polyethylene as the acetabular bearing surface for hip resurfacing. Currently available cross-linked polyethylene has superior wear characteristics and a lower failure rate compared to the polyethylene used in the past for both resurfacing and total hip replacement. Methods: We performed 200 resurfacing procedures using a metal or ceramic femoral prosthesis and a polyethylene acetabular prosthesis. The procedures were performed as primary procedures or as acetabular only revisions for metal-on-metal resurfacing procedures that had failed due to metallosis. Either a one-piece cemented or two-piece acetabular component with a titanium shell and polyethylene insert was used. The patients averaged 51 years of age and 69% of the patients were women. The average follow-up was 4 years (range, 2 to 11 years). No patients were lost to follow-up. The author certifies that he has no commercial associations (consultancies, stock ownership, equity interest, patent/ licensing arrangements, etc.) that might pose a conflict of interest with this article. The author certifies he has institutional approval for this investigation and the investigation was conducted in conformity with ethical principles of research and that informed consent was obtained. Address for correspondence or questions: James W. Pritchett MD, Orthopedics International 901 Boren Ave. #800, Seattle, WA [email protected] Results: There were two infections. There were no dislocations. 95% of patients considered their procedure completely successful. Two patients underwent successful revision surgery for acetabular loosening. Four patients underwent successful revision to a total hip replacement for femoral neck fracture, femoral loosening, pain, or infection. There were no instances of osteolysis and there were no revisions for acetabular wear. Two patients had radiographic signs of polyethylene wear. None of the patients reported squeaking or clunking from their resurfaced hip. The mean Harris hip score was 93. Discussion: Hip resurfacing with a polyethylene acetabular component is a reliable procedure at midterm follow-up. Some of the concerns that exist namely metallosis with metal-on-metal prostheses can be avoided. The functional results are comparable to metal-on-metal resurfacing but long-term follow up is needed to determine if implant survivorship with polyethylene acetabular components will equal metal-on-metal prostheses. Polyethylene can be a useful option in acetabular revision situations or for women fearing metallosis. Introduction The current generation of metal-on-metal hip resurfacing arthroplasty is the fourth attempt at trying to preserve the femoral head and eliminate a femoral component inserted into the shaft of the femur. The first-generation implants were done on a limited basis using metal-on-metal, acrylic, or crude polymers. 10,23 The second generation used a cemented polyethylene acetabular component and usually a stemless femoral component. After initial enthusiasm, the high rate of failure from femoral component loosening, femoral neck fracture, and late acetabular loosening, led to abandonment of this technique. 1,6,8,22,30,32
2 2 With a necessary large-diameter femoral head and thin polyethylene, wear debris was substantial, primarily because resurfacing patients were young and active. 22,32 The third generation of resurfacing prostheses used cross-linked polyethylene and a stemmed ceramic or metal femoral component. 27 These prostheses were never widely used. Due to concerns about polyethylene, when improved metallurgy was developed, a fourth generation of resurfacing prostheses was born. These implants are metal-on-metal and employ a so-called hybrid concept: a cementless, porous coated, non-modular (monobloc) acetabular component and a stemmed femoral component implanted with bone cement. 2,28 The results of fourth generation prostheses have been better than early-generation prostheses except for smaller size patients (women), who have a heightened risk of an adverse reaction to wear debris (metallosis). 17 We asked three questions: (1) What are the results of hip resurfacing using a cross-linked polyethylene acetabular component? (2) What are the complications of using polyethylene for hip resurfacing? and (3) What is the survivorship of hip resurfacing prostheses using polyethylene? Development of Polyethylene Polyethylene was not the initial choice of a polymer for hip arthroplasty. Sir John Charnley originally used polytetrafluorethylene (Teflon). The initial results were very positive but all of his implants failed over a few years. Charnley s technician, Harry Craven, was introduced to polyethylene by a bearing salesman and Charnley began using it in November, Charnley was opposed to the use of metal-on-metal, stating Nevertheless, the conditions for film lubrication in a metal-to-metal joint must inevitably become less favorable as the diameter of the femoral head is reduced. 5 In 1960, Dr. Charles O. Townley used polyurethane for hip resurfacing but over a few years the polyurethane wore away and he also moved to polyethylene. 23,24 Polyester and polyformaldehyde were also used but never became popular, as the results compared unfavorably with polyethylene in long-term follow-up. 16,31 Nylon was used unsuccessfully in a limited number of early procedures. 18 Recently, poly-ether-ether-ketone (PEEK) has been used successfully on an investigational basis. 20 New formulations of polyurethane have been developed but they are not approved for use. 21 Neither PEEK nor polyurethane is available at this time. Polyethylene was originally rejected as a candidate material for both resurfacing and total hip replacement. It failed completely when used on the femoral side 6,19,31 but proved useful on the acetabular side. 14 Cross-linked polyethylene when used as a dual mobility (unconstrained tripolar) prosthesis, however, works well. 9 We now routinely use dual mobility prostheses if a femoral failure (fracture, loosening, or osteonecrosis) occurs following a successful resurfacing procedure. We also offer dual mobility prostheses as an alternative to treat metallosis occurring after resurfacing or total hip procedures. The dual mobility option permits a single component revision while preserving the natural femoral head geometry. Polyethylene wears over time and its wear debris may cause osteolysis Cross-linking has reduced both wear and osteolysis considerably. All conventional hip prostheses today employ crosslinked polyethylene. Because of reduced wear, larger diameter femoral head prostheses are now used routinely. 7,12,13 Acetabular prostheses using cross-linked polyethylene are now manufactured with sufficient internal diameters to accommodate the natural femoral head preserved during hip resurfacing surgery (at least for smaller size individuals). 11
3 3 Materials and Methods For hip resurfacing we offered a polyethylene prosthesis to patients who had the following indications: (1) small femoral geometry (women), defined as a femoral head diameter of less than 46 mm, (2) prior adverse reaction to metal wear debris, and (3) concern for metal sensitivity. All patients were also offered the option of total hip replacement procedures. All patients were aware of the availability of metal-on-metal resurfacing prostheses. All femoral prostheses were stemmed and either a modular magnesia-stabilized zirconium or cobaltchromium femoral prosthesis was used (Figure 1). Femoral prostheses were used with or without cement. The acetabular prostheses were either cemented in place or implanted without cement. The cementless prostheses were two-piece with a titanium backing and cross-linked polyethylene of a composite thickness of 10 mm. Patients were allowed full weight bearing immediately and were evaluated annually. No limitations were placed on patients following their initial recovery. No blood transfusions were given. Results The follow-up ranged from 2 to 11 years. Fortyfour patients had 2 to 3 years of follow-up, 51 had 8 to 11 years of follow-up, and 105 had 3 to 8 years of follow-up. No patients were lost to follow-up. There were 200 resurfacing procedures using a polyethylene acetabular prosthesis and a metal or ceramic femoral prosthesis performed and prospectively followed. The average patient age was 51 years and 69% of the patients were women. Of the 200 procedures, 158 were performed as primary procedures (Figures 2A, 2B) and 42 were acetabular revisions for metal-on-metal resurfacing procedures (Figures 3A, 3B, 3C) that had failed due to metallosis. Figure 1. This is a photograph of the two-piece acetabular component consisting of a porous titanium shell and a 44 mm polyethylene liner. The femoral component is zirconium with a curved modular stem. For 179 procedures, a two-piece acetabular component with a titanium shell and a polyethylene insert was used. For 21 procedures the polyethylene was cemented to the acetabular bone. There were two wound infections and three patients developed substantial heterotopic ossification. There were no dislocations or nerve palsies. Five patients continued to report pain: two had mild pain, two had moderate pain and one had substantial pain. Two patients, one with a cemented and one with a cementless acetabulum, underwent successful revision for acetabular loosening. Four
4 4 Figure 2A. This is an anteroposterior radiograph of a 49-year-old woman with severe osteoarthritis. Figure 3A. This is a lateral radiograph of a 51-yearold woman who developed metallosis and acetabular loosening following a metal-on-metal resurfacing procedure. Figure 2B. The postoperative radiograph shows a hybrid resurfacing using a two-piece acetabular component a cemented cobalt-chromium femoral component. Figure 3B. This is an anteroposterior radiograph after successful revision of the acetabular prosthesis to a two-piece polyethylene bearing prosthesis.
5 5 patients underwent successful revision to total hip replacement for femoral neck fracture, loosening, persistent pain, or infection. In three of the four revisions to total hip replacement procedures, the metal backing of the acetabular component was preserved and the acetabular liner was exchanged. In the fourth revision procedure, the cemented one-piece acetabular component was revised to a two-piece component. There was no appreciable wear at 2, 3, 5, and 6 years seen on the polyethylene. There were no instances of osteolysis but two patients had radiographic signs of polyethylene wear at 7 and 8 years. (Figure 4A-B) No patients reported squeaking or clunking from their resurfaced hip. The mean Harris hip score was 93 and 95% of the patients claimed no functional limitations. Figure 3C. This is a lateral radiograph after the acetabular revision Discussion Polyethylene has been an orthopedic bearing material since the 1960s. 5,14 It is chemically and conceptually simple; it is produced by the polymerization of ethylene gas into a macromolecular carbon chain with pendant hydrogen atoms. Crosslinks, bonds that interconnect polyethylene molecules, can be produced by gamma or electron beam radiation. They are then annealed or re-melted by thermal treatments. 7 In 1998, highly cross-linked polyethylenes were introduced for clinical use. Clinical studies to date show a 50% to 87% reduction in wear. 7,12,13 Cross-linked polyethylene has been produced and approved for use for femoral head diameters Figure 4A-B. This is an anteroposterior radiograph of a 44-year-old woman showing a cemented polyethylene acetabular component and cobalt-chromium femoral resurfacing prosthesis. On the left, thinning of the polyethylene is seen 8 years following implantation. On the right, the original thickness of the polyethylene is seen.
6 6 up to 46 mm. Some, but not all, studies have shown increased wear with femoral head diameters greater than 32 mm. There is substantial and favorable experience with femoral head diameters of 36 and 40 mm. 11,13,27 There is favorable wear simulator data from polyethylene diameters of 44 and 46 mm but no long-term clinical data are available. 11 Limiting oxidation has been an additional concern and polyethylene containing Vitamin E is now available. 4 Early polyethylene prostheses were secured to the pelvis with polymethylmethacrylate during hip replacement or resurfacing procedures. 1,6,10,30,32 This was very successful but late loosening is common and, therefore, the use of porous-coated metal backing had become a very popular and successful alternative. Because cross-linked polyethylene can fracture, its thickness and the thickness of the metal backing are subject to engineering limitations. Most engineers recommend using a polyethylene thickness of 3.8 mm or greater and a composite thickness including the metal backing of 10 mm or more if a two-piece component is selected. Cobalt-chromium alloys are used widely as bearing surfaces against polyethylene for hip and knee implants. Cobalt-chromium is harder and more resistant to corrosion than previous metals used in joint replacement, such as stainless steel. Titanium is much too soft to use as a bearing surface. Cobalt-chromium surfaces can be damaged and exhibit low wetability. Newer cobalt-chromium surfaces are superior to older implants with respect to smoothness. Ceramic materials generally offer harder and more hydrophilic surfaces compared to cobalt chromium and can be polished to a very low degree of roughness. 14 Alumina-based ceramics have very favorable wear characteristics but there have been rare reports of implant fracture. 29 Reports of yttria-stabilized zirconia showed no reduction in wear when used with cross-linked polyethylene. 25 Oxidized zirconia has favorable wear results in hip simulator studies but has not been manufactured for use in resurfacing. 3 Magnesia-stabilized zirconia was chosen for use in our patients because of its superior wear characteristics in a hip simulator. 25 Also, there was no difficulty in preparing thin-walled stemmed prostheses appropriate for resurfacing applications. 21,27 There are no long-term data available for using cross-linked polyethylene for resurfacing applications either with cobalt-chromium or ceramic prostheses. A nitrated (ceramized) resurfacing prosthesis has been used on a limited basis articulating with non-cross-linked polyethylene. The durability has been up to 11 years. 15 Polyethylene should be reconsidered for resurfacing because of the superior wear characteristics of cross-linked polyethylene. Also, newer cobaltchromium prostheses have reduced roughness. It will take many years to confirm the wisdom of this approach. When polyethylene wear occurs, it is anticipated that revision to another polyethylene bearing without disturbing the well-fixed metal shell will be possible. Women need not be denied hip resurfacing surgery.
7 7 References 1. Amstutz HC, Grigoris P, Dorey FJ. Evolution and future of surface replacement of the hip. J Orthop Sci 1998;3: Amstutz H, Le Duff M, Campbell PA, Gruen TA, Wisk LE. Clinical and radiographic results of metal-on-metal hip resurfacing with a minimum ten-year follow-up. J Bone Joint Surg Am 2010;92: Bourne RB, Barrack R, Rorabeck CH, Salehi A, Good V. Arthroplasty options for the young patient: Oxinium on crosslinked polyethylene. Clin Orthop Relat Res 2005;441: Bracco P, Oral E. Vitamin E-stabilized UHMWPE for total joint implants: A review. Clin Orthop Relat Res 2011;469: Charnley J. Total hip replacement by low-friction arthroplasty. Clin Orthop Relat Res 1970;72: Freeman MA, Cameron HU, Brown GC. Cemented double cup arthroplasty of the hip: a 5 year experience with the ICLH prosthesis. Clin Orthop Relat Res 1978;134: Gordon AC, D Lima DD, Colwell CW Jr. Highly cross-linked polyethylene in total hip arthroplasty. J Am Acad Orthop Surg 2006;14: Grigoris P, Roberts P, Panousis K, Bosch H. The evolution of hip resurfacing arthroplasty. Orthop Clin North Am 2005;36: Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res 2009;467: Haboush EJ. A new operation for arthroplasty of the hip based on biomechanics, photoelasticity, fast setting dental acrylic, and other considerations. Bull Hosp Joint Dis 1953;14: Kelly NH, Rajadhyaksha AD, Wright TM, Maher SA, Westeich GH. High stress conditions do not increase wear of thin highly crosslinked UHMWPE. Clin Orthop Relat Res 2010;468: Kuzyk PR, Saccone M, Sprague S, Simunovic N. Bhandari M, Schemitsch EH. Cross-linked versus conventional polyethylene for total hip replacement: a meta-analysis of randomized clinical trials. J Bone Joint Surg Br 2011;93: Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res 2011;469: Learmonth ID, Young C, Rorabeck C. The operation of the century: total hip replacement. Lancet 2007;370: Malviya A, Lobaz S, Holland J. Mechanism of failure eleven years following a Buechel Pappas hip resurfacing. Acta Orthop Belg 2007;73: Mathiesen EB, Lindgren U, Reinholdt FP, Sudmann E. Wear of the acetabular socket. Comparison of polyacetal and polyethylene. Acta Orthop Scan 1986;57: McBryde CW, Thievendran K, Thomas AM, Treacy RB, Pynsent PB. The influence of head size and sex on the outcome of Birmingham hip resurfacing. J Bone Joint Surg Am 2010;92: McKeever DC. The choice of prosthetic materials and evaluation of results. Clin Orthop Relat Res 1955;6: Newman PH, Scales JT. The unsuitability of polyethylene for movable weight-bearing prostheses; report of a case of cup arthroplasty of the hip. J Bone Joint Surg Br 1951;33B: Pace N, Marinelli M, Spurio S. Technical and histologic analysis of a retrieved carbon fiber-reinforced poly-etherether-ketone composite alumina-bearing liner 28 months after implantation. J Arthroplasty 2008;23: Pritchett JW. Heat generated by hip resurfacing prostheses: an in vivo pilot study. J Long Term Effects Med Implants 2011;21: Pritchett JW. Success rates of the TARA hip. Am J Orthop 1998;27: Pritchett JW. Curved-stem hip resurfacing: minimum 20-year followup. Clin Orthop Relat Res. 2008;466: Pritchett JW. Conservative total articular replacement arthroplasty: minimum 20 year follow-up. In: McMinn DJW, ed. Modern Hip Resurfacing. London: Springer;2009: Roy ME, Whiteside LA, Magill ME, Katerberg BJ. Reduced wear of cross-linked UHMWPE using magnesia-stabilized zirconia femoral heads in a hip simulator. Clin Orthop Relat Res 2011;469: Shen FW, Lu Z, McKellop HA. Wear versus thickness and other features of 5-Mrad crosslinked UHMWPE acetabular liners. Clin Orthop Relat Res 2011;469: Townley CO, inventor,biopro Inc, assignee. Modular ball and socket joint preferably with a ceramic head ball. United States Patent US 6,096, August Treacy RB, McBryde CW, Pynsent PB. Birmingham hip resurfacing arthroplasty. A minimum follow-up of five years. J Bone Joint Surg Br 2005:87: Urban JA, Garvin KL, Boese CK, Bryson L, Pederson DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty. Seventeen to twenty-oneyear results. J Bone Joint Surg Am 2001;83: Wagner H. Surface replacement arthroplasty of the hip. Clin Orthop Relat Res 1978;134: Weber BG. Total hip replacement with rotationendoprosthesis. (Trunion-bearing prosthesis). Clin Orthop Relat Res 1970;72: Yue EJ, Cabanella M, Duffy GP, Heckman MG, O Connor MI. Hip resurfacing arthroplasty;risk factors for failure of 25 years. Clin Orthop Relat Res 2009;467:
Hip Resurfacing 2011 ORIGINAL ARTICLE. James W. Pritchett MD. Introduction. Abstract
1 ORIGINAL ARTICLE Hip Resurfacing 2011 James W. Pritchett MD Abstract In 1938 Marion Smith-Peterson placed a cobalt chromium cup on a reshaped femoral head to perform the first hip resurfacing. 15 Also,
The Journal of Arthroplasty
The Journal of Arthroplasty 29 (2014) 219 224 Contents lists available at ScienceDirect The Journal of Arthroplasty journal homepage: www.arthroplastyjournal.org One-Component Revision of Failed Hip Resurfacing
TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION
Richard A. Sweet, M.D. Louisville Orthopaedic Clinic Louisville, KY TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION INTRODUCTION Total hip replacement surgery (THR) has
Why an Exactech Hip is Right for You
Why an Exactech Hip is Right for You Why do I need a total hip replacement? Which surgical approach is best for me? How long will it last? Which implant is right for me? Founded in 1985 by an orthopaedic
VERILAST Technology for Hip Replacement Implants
VERILAST Technology for Hip Replacement Implants Surgeon Name Clinic Name Clinic Address Clinic i Address Phone Number Web Address Total Hip Replacement What Is VERILAST Hip Technology? OXINIUM Oxidized
Total Hip Joint Replacement. A Patient s Guide
Total Hip Joint Replacement A Patient s Guide Don t Let Hip Pain Slow You Down What is a Hip Joint? Your joints are involved in almost every activity you do. Simple movements such as walking, bending,
NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal
NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Health Technology Appraisal Total hip replacement and resurfacing arthroplasty for the treatment of pain or disability resulting from end stage arthritis
Answers to commonly asked questions from patients with metal-on-metal hip replacements / resurfacings. Contents
Answers to commonly asked questions from patients with metal-on-metal hip replacements / resurfacings John Skinner 1 and Alister Hart 2, Consultant Orthopaedic Surgeons and Directors of the London Implant
on-metal Hips: Device Mechanics and Failure Modes
1 Metal-on on-metal Hips: Device Mechanics and Failure Modes Steven M. Kurtz, Ph.D., and Richard Underwood, Ph.D. Exponent, Inc., and Drexel University NIH R01 AR47904 2 Contracts: DePuy Orthopaedics,
Metal-on-Metal Hip Systems
Meeting of the Orthopaedic and Rehabilitation Devices Advisory Panel June 27-28, 2012 Metal-on-Metal Hip Systems Center for Devices and Radiological Health U.S. Food and Drug Administration Meeting of
Patient Labeling Information System Description
Patient Labeling Information System Description The Trident Ceramic Acetabular System is an artificial hip replacement device that features a new, state-of-the-art ceramic-on-ceramic bearing couple. The
Clinically Proven Fiber Metal Material. Power to choose advanced bearing technologies to match patient demands
Trilogy IT Acetabular System Design Rationale 1 For orthopedic surgeons treating a wide range of patients, the Trilogy IT Acetabular System provides a clinically proven1-5 Fiber Metal Material and the
ReCap Product Rationale
FLH 155 01/09 Biomet UK Ltd Waterton Industrial Estate Bridgend, South Wales CF31 3XA, United Kingdom Tel. +44 (0)1656 655221 Fax. +44 (0)1656 645454 ReCap Product Rationale A complete resurfacing system
M 2 a-magnum. Design Rationale. Large Metal Articulation. Knees Hips Extremities Cement and Accessories PMI. Technology
M 2 a-magnum Large Metal Articulation Design Rationale Knees Hips Extremities Cement and Accessories PMI Technology M 2 a-magnum Large Metal Articulation History of Metal-on-Metal Early metal-on-metal
NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE
NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Overview Total hip replacement and resurfacing arthroplasty for treating pain or disability resulting from end stage arthritis of the hip (review of technology
P REPLACEMENT SURGERY
P REPLACEMENT SURGERY DIRECT ANTERIOR APPROACH M I N I M I Z I N G R E C O V E R Y. M A X I M I Z I N G R E S U L T S. CENTER FOR MINIMAL INVASIVE JOINT SURGERY 2301 25TH STREET SOUTH FARGO ND 58103 701-241-9300
Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology
Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology Hips designed to fit the unique anatomies of men and women Independent control for a natural fit Simple, practical solutions for optimal restoration
REVISION HIP RESURFACING
1 ORIGINL RTICLE REVISION HIP RESURFCING James W. Pritchett MD bstract Most hip resurfacing patients assume that when a resurfacing procedure fails, revision to a total hip replacement will be their only
Y O U R S U R G E O N S. choice of. implants F O R Y O U R S U R G E R Y
Y O U R S U R G E O N S choice of implants F O R Y O U R S U R G E R Y Y O U R S U R G E O N S choice of implants F O R Y O U R S U R G E R Y Your Surgeon Has Chosen the C 2 a-taper Acetabular System The
Wear-contact analysis of hip implant coated with functionally graded material
2014; 1(7): 439-443 IJMRD 2014; 1(7): 439-443 www.allsubjectjournal.com Received: 13-12-2014 Accepted: 30-12-2014 e-issn: 2349-4182 p-issn: 2349-5979 Impact Factor: 3.762 Vinod Pallapu M. Tech Student
Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability
Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability Robert Wetzel, MD, Lalit Puri, MD, S. David Stulberg, MD Disclosure In accordance with ACCME
Bringing together proven technologies. Zimmer Continuum Acetabular System
Bringing together proven technologies. Zimmer Continuum Acetabular System The power to meet individual patient needs. Zimmer Continuum Acetabular System provides highly flexible solutions for orthopaedic
Hip Resurfacing (Re-Review)
20, 2012 Health Technology Assessment Hip Resurfacing (Re-Review) Final Evidence Report October 14, 2013 Health Technology Assessment Program (HTA) Washington State Health Care Authority PO Box 42712 Olympia,
High-Flex Solutions for the MIS Era. Zimmer Unicompartmental High Flex Knee System
High-Flex Solutions for the MIS Era Zimmer Unicompartmental High Flex Knee System Zimmer Unicompartmental High Flex Knee Built On Success In today s health care environment, meeting patient demands means
Zimmer Longevity Highly Cross-linked Polyethylene
Zimmer Longevity Highly Cross-linked Polyethylene Clinical Value Dossier April 9, 2012 TOC Clinical Value Dossier for Zimmer Longevity Highly Cross-linked Polyethylene 1 Executive Summary 7 2 Burden of
BIRMINGHAM HIP Resurfacing (BHR ) System PATIENT INFORMATION
BIRMINGHAM HIP Resurfacing (BHR ) System PATIENT INFORMATION Table of Contents 1.0 What is the BHR Device? 2.0 What is the Purpose of the BHR Device? 3.0 When Should the BHR Device Not Be Used? (Contraindications)
Appendix B: FDA Safety Communication (Website) for MoM Hip Systems
Appendix B: FDA Safety Communication (Website) for MoM Hip Systems. 1 Metal on Metal Hip Implant Systems Home Page The Hip Joint Hip Implant Systems Metal-on-Metal Hip Implant Systems Total Hip Replacement
frequently asked questions Knee and Hip Joint Replacement Technology
frequently asked questions Knee and Hip Joint Replacement Technology frequently asked questions Knee and Hip Joint Replacement Technology Recently, you may have seen advertisements from legal companies
BONE PRESERVATION STEM
TRI-LOCK BONE PRESERVATION STEM Featuring GRIPTION Technology SURGICAL TECHNIQUE IMPLANT GEOMETRY Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was
Imagine being an orthopedic surgeon and knowing that your patient s
http://www.surgicalproductsmag.com/product-releases/2009/05/safer-methods-internal-bone-fixation Advances in Internal Bone Fixation Sharps safety for orthopedic surgeons by Ron Stoker Imagine being an
Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner
The Journal of Arthroplasty Vol. 24 No. 2 2009 Case Report Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner William J. Robertson, MD, Christopher J. Mattern, MD/MBA, John
Total Hip Replacement Hip replacement surgery, or arthroplasty, uses implants to resurface and replace the bones in the joint, re-creating the smooth gliding surfaces that were once intact. Hip replacement
Introduction to Biomaterials in Orthopaedic Surgery
CHAPTER 1 Introduction to Biomaterials in Orthopaedic Surgery 1.1 Definition of Biomaterial A biomaterial is a material that interacts with human tissue and body fluids to treat, improve, or replace anatomical
Disclosures. ! Consultant. ! Stryker. ! Smith and Nephew. Why Resurface? ! Save bone the day of surgery. ! No bone loss from stressshielding
Disclosures irmingham Hip Resurfacing Hip Resurfacing in 2015 Peter rooks MD, FRCS(C) Cleveland Clinic! Consultant! Stryker! Smith and Nephew HR Goals Why Resurface? Why Resurface?! Summarize benefits!
Robotic-Arm Assisted Surgery
Mako TM Robotic-Arm Assisted Surgery for Total Hip Replacement A Patient s Guide Causes of Your Hip Pain Your joints are involved in almost every activity you do. Movements such as walking, bending and
HEADER TOTAL HIP REPLACEMENT SURGERY FROM PREPARATION TO RECOVERY
HEADER TOTAL HIP REPLACEMENT SURGERY FROM PREPARATION TO RECOVERY ABOUT THE HIP JOINT The hip joint is a ball and socket joint that connects the body to the legs. The leg bone is called the femur. The
Randal S. Ford. ATTORNEY AT LAW (205) 752-4440 [email protected]. DePuy ASR Hip Implant Recall Fact Sheet
DePuy ASR Hip Implant Recall Fact Sheet Recalled Devices: ASR XL Acetabular System on the market since 2004, and sold worldwide. Dates of Implant: July 2003 to 2010 ASR Hip Resurfacing System on the market
Cormet Hip Resurfacing System
Cormet Hip Resurfacing System Patient Product Information 325 Corporate Drive Mahwah, NJ 07430 t: 1-888-STRYKER www.aboutstryker.com The information presented in this brochure is for educational purposes
Total Hip Replacement
Please contactmethroughthegoldcoasthospitaswityouhaveanyproblemsafteryoursurgery. Dr. Benjamin Hewitt Orthopaedic Surgeon Total Hip Replacement The hip joint is a ball and socket joint that connects the
Mid-term results for hip resurfacing in patients under 30 years old with childhood hip disorders
Acta Orthop. Belg., 2015, 81, 264-273 ORIGINAL STUDY Mid-term results for hip resurfacing in patients under 30 years old with childhood hip disorders Simon MacLean, Scott Evans, Paul Pynsent, John O hara
BONE CEMENT. History, Performance, and Choice TECHNICAL MONOGRAPH
BONE CEMENT History, Performance, and Choice TECHNICAL MONOGRAPH HIStory PERFORMANCE CHOICE HISTORY PERFORMANCE CHOICE Started with DePuy Synthes Companies The Bone Cementing Story CMW Laboratories, now
A Magazine Dedicated to Mass Torts Law. The Dangers of Metal-on-Metal Hips. Details of DePuy ASR Settlement Announced. www.langdonemison.
d3 Dangerous Drugs and Devices A Magazine Dedicated to Mass Torts Law The Dangers of Metal-on-Metal Hips Details of DePuy ASR Settlement Announced www.langdonemison.com Johnson & Johnson Agrees to Pay
The Right Choice. Exeter. total hip system
The Right Choice Exeter total hip system Exeter The Right Choice Anatomic Reconstruction Offset The objectives of total hip replacement are to: relieve pain increase mobility and function Achieving a correct
Acetabular Wedge Augment System
Orthopaedics Acetabular Wedge Augment System Stryker Hips. Implant with confidence. Restoration Acetabular Wedge Augment System Restoration Acetabular Wedge Augments, united with Tritanium, Trident, MDM
Research Letter Biomimetic Composite-Metal Hip Resurfacing Implant
Research Letters in Materials Science Volume 2008, Article ID 368985, 4 pages doi:10.1155/2008/368985 Research Letter Biomimetic Composite-Metal Hip Resurfacing Implant Habiba Bougherara and Martin N.
YOUR GUIDE TO TOTAL HIP REPLACEMENT
A Partnership for Better Healthcare A Partnership for Better Healthcare YOUR GUIDE TO TOTAL HIP REPLACEMENT PEI Limited M50 Business Park Ballymount Road Upper Ballymount Dublin 12 Tel: 01-419 6900 Fax:
Clinical Testing for Metal-on-Metal Prosthetic Wear and Tear
Clinical Testing for Metal-on-Metal Prosthetic Wear and Tear Frederick G. Strathmann, PhD, DABCC (CC, TC) February 8 th, 2013 Disclosures None Objectives Compare heavy metal analysis in synovial fluid
Photograph showing the components of an artificial total hip
Case Study 4 (CS4) Artificial Total Hip Replacement Learning Objectives After studying this case study, you should be able to do the following: 1. List and briefly explain six biocompatibility considerations
The Crosslinked Ultra-high Molecular Weight Polyethylene
The Crosslinked Ultra-high Molecular Weight Polyethylene Fu-Wen Shen Ph.D. and Harry McKellop, Ph.D. The J Vernon Luck Orthop Research Center, Orthopaedic Hospital/University of California Los Angeles,
Anterior Hip Replacement
Disclaimer This movie is an educational resource only and should not be used to manage Orthopaedic health. All decisions about the management of hip replacement and arthritis management must be made in
Your Practice Online
P R E S E N T S Your Practice Online Disclaimer This information is an educational resource only and should not be used to make a decision on Revision Hip Replacement or arthritis management. All decisions
Disappointing Short-Term Results With the DePuy ASR XL Metal-on-Metal Total Hip Arthroplasty
The Journal of Arthroplasty Vol. 27 No. 4 2012 Disappointing Short-Term Results With the DePuy ASR XL Metal-on-Metal Total Hip Arthroplasty Nicholas M. Bernthal, MD,* Paul C. Celestre, MD,* Alexandra I.
Introduction to the Bertram Hip Spacer
Introduction to the Bertram Hip Spacer Approximately 200,000 hip replacement surgeries are done every year in United States. Fortunately the incidence of infection is routinely 0.5 percent-1 percent. However
Comparative study between patients with osteonecrosis and osteoarthritis after hip resurfacing arthroplasty
Acta Orthop. Belg., 2012, 78, 735-744 ORIGINAL STUDY Comparative study between patients with osteonecrosis and osteoarthritis after hip resurfacing arthroplasty Thomas P. GROSS, Fei LIU From Midlands Orthopaedics,
Clinical performance of endoprosthetic and total hip replacement systems
Veterans Administration Journal of Rehabilitation Research and Development Vol. 24 No. 3 Pages 49 56 Clinical performance of endoprosthetic and total hip replacement systems P. M. SANDBORN, B.S. ; S. D.
Metallurgical analysis of five failed cast cobalt-chromium-molybdenum alloy hip prostheses
Veterans Administration Journal of Rehabilitation Research and Development Vol. 23 No. 4 Pages 27-36 Metallurgical analysis of five failed cast cobalt-chromium-molybdenum alloy hip prostheses STEPHEN D.
Taperloc Complete Hip System
Taperloc Complete Hip System One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art of medical care is to provide
Metasul LDH Large Diameter Head
Metasul LDH Large Diameter Head Surgical Technique Metasul LDH Large Diameter Head Surgical Technique Enhancing Stability and Increasing Range of Motion Metasul LDH Large Diameter Head Surgical Technique
Metal on Metal Hips. Total Hip Replacement. Total Hip Replacement
Metal on metal hips: Questions Metal on Metal Hips Jim Sullivan Why were metal on metal hips popular? Are all metal on metal hips the same? What are the problems with the ASR hip? What are the issues with
Graphic courtesy of DePuy Orthopaedics, Inc. HealthEast Joint Replacement Registry: 20 Year Report
Graphic courtesy of DePuy Orthopaedics, Inc. HealthEast Joint Replacement Registry: 20 Year Report HealthEast Joint Replacement Registry: 20 Year Report Foreword HealthEast Care System began the first
TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP
TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP John T. Dearborn, M.D. and Alexander P. Sah, M.D. The Center for Joint Replacement Please read this pamphlet before you see me so that
it s time for rubber to meet the road
your total knee replacement surgery Steps to returning to a Lifestyle You Deserve it s time for rubber to meet the road AGAIN The knee is the largest joint in the body. The knee is made up of the lower
Rim Cracking of the Cross-Linked Longevity Polyethylene Acetabular Liner After Total Hip Arthroplasty
This is an enhanced PDF from The Journal of Bone and Joint Surgery The PDF of the article you requested follows this cover page. Rim Cracking of the Cross-Linked Longevity Polyethylene Acetabular Liner
Hip joint replacement surgery has been the most common
Radiographic Imaging of Hip Replacement Hardware Catherine C. Roberts, MD,* and Felix S. Chew, MD, MBA Hip joint replacement surgery has been the most common type of elective and semi-elective orthopedic
HIP RESURFACING ARTHROPLASTY
HIP RESURFACING ARTHROPLASTY MEDICAL POLICY Policy Number: 2015T0503K Effective Date: December 1, 2015 Table of Contents BENEFIT CONSIDERATIONS COVERAGE RATIONALE APPLICABLE CODES.. DESCRIPTION OF SERVICES...
.URGENT DEVICE CORRECTION
zimmer.urgent DEVICE CORRECTION Zimmer, Inc. P.O. Box 706 Warsaw, IN 46561-0706 574.267.6131 www.zimmer.com July 22, 2008 Dear Surgeon: Since we last wrote to you in May 2008 regarding the Durom Acetabular
Medical Device Alert. Device All metal-on-metal (MoM) hip replacements. Action by. CAS deadlines. Ref: MDA/2012/036 Issued: 25 June 2012 at 11:00
Medical Device Alert Issued: 25 June 2012 at 11:00 Device All metal-on-metal (MoM) hip replacements Problem The MHRA is issuing updated information and advice about the follow-up of implanted with metal-on-metal
Metal-on-metal (MoM) total hip arthroplasty
S77 Metal-on-Metal Total Hip Arthroplasty Five- to 11-Year Follow-Up Vassilios S. Nikolaou, M.D., Ph.D., Alain Petit, Ph.D., Kevin Debiparshad, M.D., Olga L. Huk, M.D., M.Sc., David J. Zukor, M.D., and
i-polymers for i-arthroplasty New Polymer Technologies for HIP Arthroplasty
i-polymers for i-arthroplasty New Polymer Technologies for HIP Arthroplasty Leo Smit Business Director Implant Solutions Materials that belong to the body Total Hip Arthroplasty One of the largest successes
PRESENTATIONS/PAPERS INTERNATIONAL. RAISING STANDARDS IN JOINT ARTHROPLASTY Course Co-Chairman, The Great Debate - London, June 2013
PRESENTATIONS/PAPERS INTERNATIONAL 1998 2013 RAISING STANDARDS IN JOINT ARTHROPLASTY Course Co-Chairman, The Great Debate - London, June 2013 HOW TO CHOOSE THE RIGHT STEM FOR THE RIGHT PATIENT Corin Symposium
Analysis of Functional and Radiological Outcome of Total Hip Replacements in Rheumatoid and Osteoarthritis Patients *
Open Journal of Rheumatology and Autoimmune Diseases, 2013, 3, 246-250 Published Online November 2013 (http://www.scirp.org/journal/ojra) http://dx.doi.org/10.4236/ojra.2013.34038 Hip Replacements in Rheumatoid
Anterior Approach. to Hip Replacement Surgery
Anterior Approach to Hip Replacement Surgery Introduction When debilitating pain and stiffness in your hip limits your daily activities, you may need a total hip replacement. The development of total hip
Engine Bearing Materials
Engine Bearing Materials Dr. Dmitri Kopeliovich (Research & Development Manager) The durable operation of an engine bearing is achieved if its materials combine high strength (load capacity, wear resistance,
Small stem Exeter total hip replacement: clinical and radiological follow-up over a minimum of 2.5 years
Journal of Orthopaedic Surgery 2003: 11(2): 148 153 Small stem Exeter total hip replacement: clinical and radiological follow-up over a minimum of 2.5 years S Sivananthan, M Arif, DSK Choon Department
The Total Ankle Replacement
The Total Ankle Replacement Patient Information Patient Information This patient education brochure is presented by Small Bone Innovations, Inc. Patient results may vary. Please consult your physician
Total Knee Replacement Specifications 2014 (01/01/2012 to 12/31/2012 Dates of Procedure)
Summary of Changes Removed following ICD-9 Procedure s: 81.54 Total Knee Replacement (Bicompartmental, Partial Knee Replacement, Tricompartmental, Unicompartmental (hemijoint)). 81.55 Revision of Knee
Hip Replacement. Department of Orthopaedic Surgery Tel: 01473 702107
Information for Patients Hip Replacement Department of Orthopaedic Surgery Tel: 01473 702107 DMI ref: 0134-08.indd(RP) Issue 3: February 2008 The Ipswich Hospital NHS Trust, 2005-2008. All rights reserved.
Biomaterials in tissue engineering
Biomaterials in tissue engineering S. Swaminathan Director Centre for Nanotechnology & Advanced Biomaterials School of Chemical & Biotechnology SASTRA University Thanjavur 613 401 Tamil Nadu Page 1 of
THE REVERSE SHOULDER REPLACEMENT
THE REVERSE SHOULDER REPLACEMENT The Reverse Shoulder Replacement is a newly approved implant that has been used successfully for over ten years in Europe. It was approved by the FDA for use in the U.S.A.
Metal on Metal Vs Other Bearing Surfaces in Total Hip Replacement
Article ID: WMC001843 ISSN 2046-1690 Metal on Metal Vs Other Bearing Surfaces in Total Hip Replacement Corresponding Author: Dr. Atif Shafique, Doctor, Liverpool - United Kingdom Submitting Author: Mr.
Understanding Total Hip Replacement
Understanding Total Hip Replacement Brian J. White MD Orthopaedic Specialist in Disorders of the Hip Assistant Team Physician Denver Nuggets Western Orthopaedics Denver, Colorado Introduction This is designed
An Alternative Conservative Approach to Hip Reconstruction
Joint Implant Surgery & Research Foundation Chagrin Falls, Ohio, USA An Alternative Conservative Approach to Hip Reconstruction Craig S. Waller, MD* and Timothy McTighe, Dr. H.S. (hc)** Acknow ledgement:
From Department of Orthopaedics, Conquest Hospital, The Ridge, East Sussex. n N.A. Sandiford, FRCS (Tr/Orth), MFSEM, Specialist. Registrar.
Acta Orthop. Belg., 2014, 80, 372-379 ORIGINAL STUDY Primary total hip arthroplasty with hydroxyapatite coated titanium femoral stems. Does design philosophy influence long term outcome? : results of a
IN THE FEDERAL COURT OF AUSTRALIA (FCA) NEW SOUTH WALES REGISTRY - FEDERAL COURT OF AUSTRALIA
IN THE FEDERAL COURT OF AUSTRALIA (FCA) NEW SOUTH WALES REGISTRY - FEDERAL COURT OF AUSTRALIA GENERAL DIVISION No: NSD213/2011 NOTICE OF FILING This document was filed electronically in the FEDERAL COURT
Working principle of the dual mobility (total hip replacement): wear mechanisms and design optimization
Working principle of the dual mobility (total hip replacement): wear mechanisms and design optimization by Diploma work No. 64/2011 at Department of Materials and Manufacturing Technology CHALMERS UNIVERSITY
Hip Transplant Detailed Study
Hip Transplant Detailed Study INTRODUCTION This case involves four patients who underwent hip replacements with a "metal-on-metal" device designed and manufactured by "Replacement Manufacturer." The patients
Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling
Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling Qasem Alyazji Near East University, Department of Biomedical Engineering Nicosia, North Cyprus, Turkey
After Hip Arthroscopy
After Hip Arthroscopy On your road to recovery... Rehabilitation is essential to help you return to an active life and reach your personal goals. This booklet provides goals, activities and milestones
Materials Sciences. Dr.-Ing. Norbert Hort [email protected]. International Masters Programme in Biomedical Engineering
Materials Sciences International Masters Programme in Biomedical Engineering Magnesium Innovations Center (MagIC) GKSS Forschungszentrum Geesthacht GmbH Dr.-Ing. Norbert Hort [email protected] Contents
Controversies around modern bearing surfaces in total joint replacement surgery
Page 44 SA Orthopaedic Journal Autumn 2013 Vol 12 No 1 Controversies around modern bearing surfaces in total joint replacement surgery PB Oosthuizen MBChB Senior Registrar CH Snyckers MBChB, Dip PEC(SA),