Quality Control in Refractive Surgery
|
|
|
- Carmella Lewis
- 10 years ago
- Views:
Transcription
1 Quality Control in Refractive Surgery Stefan Pieger*, M.Sc. Ingenieurbüro Pieger GmbH, Wendelstein, Germany Introduction Personal experience as application specialist for refractive excimer lasers since (Meditec, Schwind, Nidek) PRK & LASIK Nomogram development based on systematic outcomes analysis. Excimer Laser Surgery and refractive surgery in general well suited for a systematic approach on quality control. 1
2 Why Quality Control? Verify current nomogram settings and make adjustments if necessary. Reduce enhancement rate. Use for marketing and advertisement. Discover trends and technical problems in order to react more rapidly. Fulfill requirements of ophthalmic societies. Increase confidence level about refractive procedures offered in your center. How to collect your data? Patients files Excel Spread Sheet Database Software (Access; Filemaker; etc.) Outcomes Analysis Software (; ASSORT; Refr. Consultant; etc.) 2
3 How to analyze refractive data? Standard Refractive Outcomes (JRS) Stability / Safety / Predictability / Efficacy Additional Outcome Parameters Astigmatism Outcomes: SIRC - Surgically Induced Refractive Change (= Vector Analysis ); Double Angle Scatter Plot Defocus Equivalent / Contrast Sensitivity in mesopic conditions / pre OP BSCVA vs. post OP UCVA Wavefront Based Outcomes Defocus + Cyl ( aberrometer refraction ) Higher Order RMS / Spherical Aberration (Z12) / Coma / Trefoil Standard Outcomes 3
4 vol164/war.pdf Standard Refractive Outcomes: Safety 100% 90% 80% 2 or more lines lost 3 % at 1 m post OP 70% 60% 50% 40% 30% 20% 10% 0% 1% 0% 2% 0% 22% 13% 54% 57% 25% 19% 2% 3% 0% 0% lost > 2 lost 2 lost 1 unchanged gained 1 gained 2 gained > 2 month (eyes) 1 (770) 3 (653) Number of eyes per Follow up visit. Follow up time interval 2. SAFETY: Change in BSCVA - Percent 4
5 Standard Refractive Outcomes: Efficacy 100% 90% 80% 72% 20/20 or better 3 month post Op 70% 60% 50% 40% 30% 20% 10% 0% 4% 6% 20/12 or better 18% 26% 46% 40% 14%14% 10%7% 3%4% 5% 3% 20/15 20/20 20/25 20/30 20/40 20/50 or worse month (eyes) 1 m (691) 3 m (562) Number of eyes per Follow up visit. Follow up time interval 4. EFFICACY: UCVA - Percent Standard Refractive Outcomes: Cumulative UCVA vs. preop SCVA 100% 90% 80% 70% 60% 50% 90% 95% 95% 100% 100% 88% 84% 76% 48% Number of eyes at Follow up visit. month (eyes) 40% 30% 20% 10% 8% 32% 12% PreOP SCVA 0% 3 m (25) prescva (39) '20/10' 20/12,5 or better 20/16 or better 20/20 or better 20/25 or better 20/32 or better 20/40 or better preop SCVA vs. postop UCVA - Percentage 20/50 or better 20/63 or worse 5
6 Standard Refractive Outcomes: Stability (SEQ) 1,00-0,15-0,20-0,18-0,22 0,00 pre OP 1 m 3 m 6 m 12 m -1,00 ± 1 StDev -2,00-3,00-4,00-5,00-6,00-4,70 Follow up time interval Mean value of SEQ -7,00-8, STABILITY: Achieved Change in Refr. over Time Number of eyes per Follow up visit. Standard Refractive Outcomes: Predictability (SEQ) Achieved [D] PREDICTABILITY: Attempted vs Achieved (Scatter) 744 eyes overcorrected y = -0.00x x undercorrected Attempted delta SR equiv. [D] Number of eyes at follow up ±1 D happiness Zone Trend line y = -0.00x x Regression Formula: Achieved = 0.94*Attempted (~6% undercorrection) 6
7 Astigmatism Outcomes: must be based on Vector Analysis! SIRC (Surgically Induced Refractive Change) PreOP: -5/-2 x month: Plano/-1 x 130 SIRC: -4.31/-2.39 x 12 'The cylinder was actually overcorrectd and the axis has been misaligned by 12!' Astigmatism Outcomes: SIA (based on Vector Analysis) Achieved [D] Attempted Cyl vs SIA (Scatter) overcorrected 3 month postop 490 eyes 0/-0.5 x 180-2/-1@ D Cyl Undercorrection! 0/-0.5 x 90 (-0.5/+0.5x180 ) 0.5 D Cyl Overcorrection! 3 2 y = -0.00x x undercorrected Attempted Cyl [D] y = 0.90 * x (~10% undercorrection) 7
8 Astigmatism Outcomes: Double Angle Scatter Plot PreOP Cyl & Axis PostOP Cyl & Axis Double Angle Minus Cyl Scatter Plot (160 eyes) Double Angle Minus Cyl Scatter Plot (79 eyes) Mean Cyl 6,9-5 Mean Cyl 162, Additional Outcomes: Defocus Equivalent 100% 90% 80% 86% 88% 99%99% 100%100% 100%100% 100%100% SEQ = SPH + ½ CYL 70% 60% 50% 40% 60% 64% DEQ = SEQ + ½ CYL 30% 20% 10% month (eyes) 1 m (615) 3 m (554) 0% <=0,5D <=1D <=2D <=3D <=4D <=5D DEFOCUS EQUIVALENT - Percent 8
9 Defocus Equivalent vs. Refractive Outcome 100% 90% 80% 70% 76% 72% 92% 92% 99% 99% 100%100% -5.0/+10.0 x 90 SEQ Plano DEQ +5 60% 50% 40% 100% 90% 86% 88% 99% 99% 100%100% 1 m (615) 3 m (554) 30% 80% 20% 10% 70% 60% 60% 64% 0% 50% +- 0,5 +- 1,0 +- 2,0 +- 3,0 40% Refractive outcome - % within Attempted 1 m (744) 3 m (662) 30% 20% DEQ ~ Blur Circle 10% 0% <=0,5D <=1D <=2D <=3D DEFOCUS EQUIVALENT - Percent Additional Outcomes: Mesopic Contrast Sensitivity Contrast Sensitivity Mesopic 2,20 2,00 1,80 1,60 1,68 1,63 1,92 1,88 1,61 1,57 month (eyes) pre op (112) 1 m (72) Normal Population Range (Vector Vision CSV 1000) 1,40 1,20 1,22 1,21 1,00 0,80 0,60 0,40 A (3cpd) B (6cpd) C (12cpd) D (18cpd) Spatial Frequency [cycles/degree] VA in LogMar Scale 9
10 Wavefront based Outcomes Work in progress! Important parameters: HO RMS, spherical Aberration and Coma Report on same (6mm) pupil diameter or convert to diopters Presenting horizontal and vertical Coma individually? Vector calculation to present magnitude and axis in [D]! (0.5 D 230 ) Wavefront based Outcomes Higher Order RMS [µm ±StDev] 0,80 0,70 0,60 0,53 Pupil Diameter: 6.0 mm 0,54 0,54 0,51 0,50 0,40 0,31 0,30 0,20 0,10 0, pre OP 1 m 3 m 6 m 9 m RMS higher order over time 10
11 Wavefront based Outcomes Spherical Aberration [µm] or [D] 2,50 0,10 0,05 Pupil Diameter: 6.0 mm 2,00 0,00-0,03 pre OP 1 m 3 m 1,52 9 m -0,05 1,31 1,36 1,42 1,50-0,10-0,15 1,00-0,20-0,25 0,50-0,30 0,41-0,23-0,25-0,24-0, , ,40 pre OP 1 m 3 m 6 m 9 m Spherical Spherical Aberration Aberration (Z12) (Z12) DEQ over over time time [µm] [D] Coma as Vector- Scatter Plot Coma Scatter Plot (18 eyes) 90-2,5 Coma [D] 202,4-2,0-1,5-1,0-0, ,0-2,5-2,0-1,5-1,0-0,5-0,0-0,5-1,0-1,5-2,0-2,5-0,5 0-1,0-1,5-2,0-2,
12 Making Outcome-based Nomogram Adjustments Comparison of Laser Settings vs. Achieved change in refraction (and not Attempted vs. Achieved). Reduce random errors as far as possible as nomograms can only compensate systematic errors! Must be specific for major laser parameters like OZ, TZ, ablation profile type as well as for refraction types. Nomograms for Individual Patient Groups Data must be filtered on certain parameters: Refraction Type (Myp/MyoAsti/Hyp/HypAsti ) Surgery Type (PRK; LASIK; LASEK; Custom ) Optical Zone Diameter Others (age, laser software version; humidity ) 12
13 Nomogram Improvements Laser Settings vs. Achieved Achieved [D] Laser Settings vs Achieved 1166 eyes overcorrected y = -0,01x 2 + 1,22x - 0, undercorrected Laser setting [D] y = -0,01x 2 + 1,22x - 0,08 reduce attempted SEQ by 22%! Identify and exclude outliers Achieved [D] Laser Settings vs Achieved 1166 eyes overcorrected y = -0,01x 2 + 1,22x - 0, undercorrected Laser setting [D] 13
14 Laser Setting CYL vs. Surgical Induced change in Astigmatism Achieved [D] 7 Laser Setting (cyl) vs SIA 1480 eyes 6 overcorrected y = 0.03x x High Scatter! (further analysis necessary!) % systematic undercorrection 1 undercorrected LaserSetCyl [D] Nomograms: General Comments 1. Reduce Scatter by Standardized Surgery and OR Environment 2. Exclude Outliers from Data Analysis 3. Exclude Enhancements (separate nomogram) 4. Choose appropriate follow up interval ( 3 m) 5. Create Formula ( -10% ) / Lookup Table or use Nomogram Software 14
15 Summary Improving the results of refractive surgery procedures must be based on an individual quality control system. Nomograms can compensate for systematic errors, but not for random errors. Modern outcomes analysis software allows constant monitoring of your results. Conventional Outcomes will be extended by HO-RMS, Spherical Aberration and Coma. Thank You! 15
16 16
Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia
Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia Alaa El Danasoury, FRCS Magrabi Hospitals & Centers Surgical options for the Correction of Presbyopia Monovision Reversal of Presbyopia:
VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment)
CustomVue Advantage Patient Information Sheet VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment) Statements
Early results at 1 and 3 months after Trans-PRK with AMARIS. a no-touch, one-step treatment
Early results at 1 and 3 months after Trans-PRK with AMARIS a no-touch, one-step treatment Trans-PRK advantages - No-Touch treatment in one step (much faster than in the past with 2 steps) - Significantly
Standardized Analyses of Correction of Astigmatism With the Visian Toric Phakic Implantable Collamer Lens
Standardized Analyses of Correction of Astigmatism With the Visian Toric Phakic Implantable Collamer Lens Donald R. Sanders, MD, PhD; Edwin J. Sarver, PhD ABSTRACT PURPOSE: To demonstrate the methodology
Aberrations caused by decentration in customized laser refractive surgery
Aberrations caused by decentration in customized laser refractive surgery Diana C. Lozano 1,2 Advisors: Jason Porter 2,3a, Geunyoung Yoon 2,3b, and David R. Williams 2,3a San Diego State University, San
Wavefront technology has been used in our
Wavefront Customized Ablations With the WASCA Asclepion Workstation Sophia I. Panagopoulou, BSc; Ioannis G. Pallikaris, MD ABSTRACT PURPOSE: WASCA (Wavefront Aberration Supported Cornea Ablation) is a
Comparison of Retinal Image Quality between SBK and PRK. Allen Boghossian, D.O. Durrie Vision Overland Park, KS
Comparison of Retinal Image Quality between and Allen Boghossian, D.O. Durrie Vision Overland Park, KS Study Design Purpose: To compare optical quality and intraocular scatter between and using double-pass
Richard S. Hoffman, MD. Clinical Associate Professor of Ophthalmology Oregon Health & Science University
Zeiss Mel 80 and Visumax Refractive Laser Systems Richard S. Hoffman, MD Clinical Associate Professor of Ophthalmology Oregon Health & Science University No Financial Interest ZEISS Workstation CRS-Master
Wavefront Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography
Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography Maria Regina Chalita, MD, 1 Sai Chavala, MD, 1 Meng Xu, MS, 2 Ronald R. Krueger, MD, MSE 1 Purpose: To evaluate
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System Jan Venter, MD ABSTRACT PURPOSE: To determine the predictability, effi cacy, safety, and stability of LASIK using custom ablation
FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US
FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US JON DISHLER, MD DENVER, COLORADO, USA INTRODUCTION AND STUDY OBJECTIVES This article summarizes the first US
Laser in situ keratomileusis for mixed astigmatism using a modified formula for bitoric ablation
European Journal of Ophthalmology / Vol. 18 no. 6, 2008 / pp. 869-876 Laser in situ keratomileusis for mixed astigmatism using a modified formula for bitoric ablation D. DE ORTUETA, C. HAECKER Augenlaserzentrum
Wavefront Analysis of Flap & Laser-induced aberrations in 2-step LASIK. Ann Laurenzi, O.D. Cole Eye Institute Cleveland Clinic Foundation
Wavefront Analysis of Flap & Laser-induced aberrations in 2-step LASIK Ann Laurenzi, O.D. Cole Eye Institute Cleveland Clinic Foundation Co-Authors Maria Regina Chalita, M.D. Samra Waheed, M.D. Meng Xu,
The pinnacle of refractive performance.
Introducing! The pinnacle of refractive performance. REFRACTIVE SURGERY sets a new standard in LASIK outcomes More than 98% of patients would choose it again. 1 It even outperformed glasses and contacts
Common visual problems in older LASIK patients
丘 子 宏 LASIK 手 術 後 的 視 覺 Visual acuity:the measurement of high contrast Snellen acuity but not other functions under different condition Quality of vision: measure the visual functions in variable condition
Wavefront higher order aberrations in Dohlman/Boston keratoprosthesis and normal eyes
Wavefront higher order aberrations in Dohlman/Boston keratoprosthesis and normal eyes Kalliopi Stasi, Seth Pantanelli, Geunyoung Yoon, Greg McCormick, James Aquavella Rochester Eye Institute and Center
LASIK for Hyperopia With the WaveLight Excimer Laser
LASIK for Hyperopia With the WaveLight Excimer Laser A. John Kanellopoulos, MD; Joseph Conway, MD; Lawrence H. Pe, MD ABSTRACT PURPOSE: To evaluate the safety and effi cacy of the ALLEGRETTO WAVE excimer
Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser
Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser Sudhank Bharti, MD; Harkaran S. Bains ABSTRACT PURPOSE: To investigate the predictability
Changes in higher order aberrations after wavefront-guided PRK for correction of low to moderate myopia and myopic astigmatism: Two-year follow-up
European Journal of Ophthalmology / Vol. 17 no. 4, 2007 / pp. 507-514 Changes in higher order aberrations after wavefront-guided PRK for correction of low to moderate myopia and myopic astigmatism: Two-year
RELEX SMILE AND SMILE EXTRA.. OUR 1 YEAR RESULTS AND PATIENTS SURVEY
RELEX SMILE AND SMILE EXTRA.. OUR 1 YEAR RESULTS AND PATIENTS SURVEY DR SANDIP MITRA MD FRCS CORNEA AND REFRACTIVE FELLOW (ROYAL VICTORIA EYE AND EAR HOSPITAL, AUSTRALIA) RELEX SMILE UNIT AT THE ALZAHRA
Faster recovery of visual acuity at all distances
news Special issue PresbyMAX Faster recovery of visual acuity at all distances PresbyMAX μ-monovision: The latest technique for treating presbyopia SCHWIND now offers PresbyMAX µ-monovision as a new option
The Evolution of the Optical Zone in Corneal Refractive Surgery. Bruce Drum, Ph.D.
The Evolution of the Optical Zone in Corneal Refractive Surgery. Bruce Drum, Ph.D. FDA, Division of Ophthalmic and ENT Devices, Rockville, MD Disclaimer This presentation represents the professional opinion
Advances in the measurement of the eye's
Visual Benefit of Correcting Higher Order Aberrations of the Eye David Williams, PhD; Geun-Young Yoon, PhD; Jason Porter, MS; Antonio Guirao, PhD; Heidi Hofer, BS; Ian Cox, PhD ABSTRACT There is currently
Refractive Surgery. Evolution of Refractive Error Correction
Refractive Surgery Techniques that correct for refractive error in the eye have undergone dramatic evolution. The cornea is the easiest place to place a correction, so most techniques have focused on modifying
INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the
INTRACOR An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the 1 Dr Luis Ruiz Presbyopia treatment with INTRACOR Luis
MicroScan. Excimer laser system for all types of vision corrections OPTOSYSTEMS LTD.
MicroScan Excimer laser system for all types of vision corrections OPTOSYSTEMS LTD. МicroScan is a new generation excimer laser system for all types of vision corrections: myopia, hypermetropia, astigmatism.
Insert to October 2013 THE COMPLETE PICTURE. Experiences with the ALADDIN system. Sponsored by Topcon
Insert to October 2013 THE COMPLETE PICTURE Experiences with the ALADDIN system Sponsored by Topcon THE COMPLETE PICTURE Advantages of The Aladdin How this all-in-one optical biometer edged out the IOLMaster
Subjective Image Quality Metrics from The Wave Aberration
Subjective Image Quality Metrics from The Wave Aberration David R. Williams William G. Allyn Professor of Medical Optics Center For Visual Science University of Rochester Commercial Relationship: Bausch
IBRA Professional. Leading Technology for Refractive Outcome Analysis. Brochure 2015
IBRA Professional Leading Technology for Refractive Outcome Analysis Brochure 2015 Introduction IBRA is the leading all-in-one system for refractive laser and lens surgery outcomes, allowing eye surgeons
Customized corneal ablation and super vision. Customized Corneal Ablation and Super Vision
Customized Corneal Ablation and Super Vision Scott M. MacRae, MD; James Schwiegerling, PhD; Robert Snyder, MD, PhD ABSTRACT PURPOSE: To review the early development of new technologies that are becoming
SCHWIND CAM Perfect Planning wide range of applications
SCHWIND CAM Perfect Planning wide range of applications ORK-CAM PresbyMAX PALK-CAM PTK-CAM 2 SCHWIND CAM the system solution The latest version of the modular SCHWIND CAM represents an even more efficient
UPDATE ON AVEDRO CROSSLINKING STUDIES
UPDATE ON AVEDRO CROSSLINKING STUDIES Peter S. Hersh, M.D. Director, Cornea & Laser Eye Institute CLEI Center for Keratoconus Clinical Professor of Ophthalmology, Rutgers Medical School Visiting Researcher,
Comparison of Higher Order Aberrations and Contrast Sensitivity After LASIK, Verisyse Phakic IOL, and Array Multifocal IOL
ORIGINAL ARTICLES Comparison of Higher Order Aberrations and Contrast Sensitivity After LASIK, Verisyse Phakic IOL, and Array Multifocal IOL Sasivimon Chandhrasri, MD; Michael C. Knorz, MD ABSTRACT PURPOSE:
CustomVue Treatments for Monovision in Presbyopic Patients with Low to Moderate Myopia and Myopic Astigmatism
CustomVue Treatments for Monovision in Presbyopic Patients with Low to Moderate and Myopic Introduction Pre-Operative Examination Surgical Technique 1 2 IMPORTANT INFORMATION CustomVue Monovision treatments
Accelerated Refractive Performance
Accelerated Refractive Performance Get There at the Speed of WaveLight Designed to accommodate your refractive technology goals now and into the future, the WaveLight Workstation is a faster way to get
New topographic custom ablation procedure for treating irregular astigmatism post keratoplasty with high frequency (1 KHz) excimer laser.
New topographic custom ablation procedure for treating irregular astigmatism post keratoplasty with high frequency (1 KHz) excimer laser. G. COLONNA M.D., G. Lorusso M.D., S. Santoro M.D. ESCRS Berlin
Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation
Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation Bettina B. Jendritza, MD; Michael C. Knorz, MD; Steve Morton ABSTRACT PURPOSE: To investigate the use of wavefront-guided
Initial clinical experience with the FS200 Femto and EX500 excimer
Initial clinical experience with the FS200 Femto and EX500 excimer lasers for LASIK ASCRS, San Diego 2011 A. John Kanellopoulos, MD Professor NYU Medical School, NY Director, Laservision.gr Institute,
Patient-Reported Outcomes with LASIK (PROWL-1) Results
Patient-Reported Outcomes with LASIK (PROWL-1) Results Elizabeth M. Hofmeister, MD CAPT, MC, USN Naval Medical Center San Diego Refractive Surgery Advisor for Navy Ophthalmology Assistant Professor of
Multi-Centre Evaluation of TransPRK outcomes with SCHWIND AMARIS using SmartPulse Technology
study 2015 Multi-Centre Evaluation of TransPRK outcomes with SCHWIND AMARIS using SmartPulse Technology David T. Lin, MD, Canada, Paolo Vinciguerra, MD, Italy, Maria C. Arbelaez, MD, Oman, Shady T. Awwad,
Irregular astigmatism:
Irregular astigmatism: definition, classification, topographic and clinical presentation Ming X. Wang, MD, PhD Clinical Associate Professor of Ophthalmology of University of Tennessee Director, Wang Vision
Is LASIK Good Enough? 75% of flights have a CL wearer 53% of all EVAs have either a spectacle or CL wearer Microbial keratitis has occurred
Optical Correction of Astronauts 21 Sept 27: NASA Approves LASIK 75% of flights have a CL wearer 53% of all EVAs have either a spectacle or CL wearer Microbial keratitis has occurred Staph epi Presbyopia
final corrected draft
Archived at the Flinders Academic Commons http://dspace.flinders.edu.au/dspace/ This is the author s final corrected draft of this article. It has undergone peer review. Citation for the publisher s version:
Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday?
Financial Disclosure Arturo Chayet, MD Tijuana, BC Mexico Perry S. Binder, MS, MD San Diego CA USA I have the following financial interests or relationships to disclose: AMO/IntraLase Corporation - C Acufocus
Assessment of Contrast Sensitivity and Aberrations After Photorefractive Keratectomy in Patients with Myopia Greater than 5 Diopters
ORIGINAL REPORT Assessment of Contrast Sensitivity and Aberrations After Photorefractive Keratectomy in Patients with Myopia Greater than 5 Diopters Alireza Fahim 1, Bijan Rezvan 1, and Hassan Hashemi
Challenging Refractive Surgery Cases. Vance Thompson, MD, FACS Refractive and Cataract Surgery Vance Thompson Vision Sioux Falls, South Dakota
Challenging Refractive Surgery Cases Vance Thompson, MD, FACS Refractive and Cataract Surgery Vance Thompson Vision Sioux Falls, South Dakota Financial Disclosures Research/consulting: Alcon AMO B & L
Wavefront-Guided LASIK for the Correction of Primary Myopia and Astigmatism
Ophthalmic Technology Assessment Wavefront-Guided LASIK for the Correction of Primary Myopia and Astigmatism A Report by the American Academy of Ophthalmology Steven C. Schallhorn, MD, Ayad A. Farjo, MD,
Life Science Journal 2014;11(9) http://www.lifesciencesite.com. Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000)
Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000) Gamal Mostafa Abo El Maaty, Mohamed Elmoddather, Mahmoud Ibrahem Ghazy, Mohamed Al-Taher Ophthalmology Department, Faculty of
The effect of corneal wavefront aberrations on corneal pseudo-accommodation
The effect of corneal wavefront aberrations on corneal pseudo-accommodation Li Wang, MD, PhD Elizabeth Yeu, MD, Douglas D. Koch, MD Cullen Eye Institute Baylor College of Medicine, Houston, TX Background
When To Laser, When To Implant, When To Do Both
When To Laser, When To Implant, When To Do Both Scott MacRae, MD Professor of Ophthalmology Professor of Visual Sciences StrongVision Refractive Surgery Center University of Rochester Eye Institute Refractive
How To Treat Eye Sight Problems With Eye Care
Dealing With The Refractive Surprise Richard S. Hoffman, MD Clinical Associate Professor of Ophthalmology Casey Eye Institute Oregon Health and Science University No Financial Interest Why The Surprise?
Optimized Profiles for Astigmatic Refractive Surgery
14 Optimized Profiles for Astigmatic Refractive Surgery Samuel Arba-Mosquera 1,, Sara Padroni 3, Sai Kolli 4 and Ioannis M. Aslanides 3 1 Grupo de Investigación de Cirugía Refractiva y Calidad de Visión,
Despite being a relatively safe and
APACRS Asia-Pacific The news magazine of the Asia-Pacific Association of Cataract & Refractive Surgeons Supplement to EyeWorld Asia-Pacific Winter 2015 Shaping the refractive laser landscape Sponsored
Surface Ablation After Corneal
Surface Ablation After Corneal Surgery: Management of Haze Helen K. Wu, MD New England Eye Center Tufts University School of Medicine Boston, MA Financial Disclosures Travel Stipend/Honoraries: IOP Ophthalmics
Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information
Rediscover quality of life thanks to vision correction with technology from Carl Zeiss Patient Information 5 2 It was really w Vision defects: Light that goes astray For clear vision the eyes, cornea and
Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY
Sponsored by Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Results of the T-CAT Phase III Clinical Trial TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Topography-Guided Custom Ablation Treatments (T-CAT) with
Uncorrected astigmatism can cause visual
Spatial Vision and Astigmatism Correction What is the minimum amount of astigmatism that should be corrected? By Eloy A. Villegas, PhD; and Pablo Artal, PhD Uncorrected astigmatism can cause visual quality
Femtolaser and ocular surgery
Femtolaser and ocular surgery Safety Quick visual recovery Excellent vision without glasses It is an ultrafast laser using a near infrared light to create cuts in ocular tissues First used in corneal refractive
Bitoric Laser In Situ Keratomileusis for the Correction of Simple Myopic and Mixed Astigmatism
Bitoric Laser In Situ Keratomileusis for the Correction of Simple Myopic and Mixed Astigmatism Arturo S. Chayet, MD, 1 Miguel Montes, MD, 1 Laura Gómez, MD, 1 Xavier Rodríguez, MD, 1 Nora Robledo, OD,
Techniques for Enhancing Cataract Surgery Patients with Residual Refractive Error. Director of Cornea Center For Excellence In Eye Care Miami, FL
Techniques for Enhancing Cataract Surgery Patients with Residual Refractive Error William Trattler, MD Director of Cornea Center For Excellence In Eye Care Miami, FL 1 Financial Disclosures Consulting
Surgical Advances in Keratoconus. Keratoconus. Innovations in Ophthalmology. New Surgical Advances. Diagnosis of Keratoconus. Scheimpflug imaging
Surgical Advances in Keratoconus Keratoconus Ectatic disorder 1 in 1,000 individuals Starts in adolescence & early adulthood Uncertain cause 20% require corneal transplant Innovations in Ophthalmology
OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment
Shinagawa LASIK Center OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment Minoru Tomita, MD, Ph.D 1) Executive Medical Director at Shinagawa LASIK Center, Tokyo, Japan 2) Clinical Professor
What now, when Presbyopia sets in? Minoru Tomita, MD, PhD Executive Director Shinagawa LASIK Center, Tokyo, JAPAN
What now, when Presbyopia sets in? Minoru Tomita, MD, PhD Executive Director, Tokyo, JAPAN Financial disclosure Travel expenses for the ASCRS meeting were paid by Ziemer and AcuFocus, Inc. Presbyopia:
IOL Calculation After LASIK. Chapter (3)
the cornea (P) may be considered as the sum of the power of the anterior (Pa) and posterior (Pp) corneal surfaces, as shown in the formula: P = Pa + Pp = (n2 n1)/r1 + (n3 n2)/r2, where n1 is the refractive
Kerry D. Solomon, MD, is Director of the Carolina Eyecare Research Institute at Carolina Eyecare Physicians in Charleston, S.C.
I think the ideal diagnostic technology for all of us would be a device where we could take a measurement, make an adjustment based on the patient s history, including past surgery, and come up with an
KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,
Comparison of and Intacs for Keratoconus and Post-LASIK Ectasia MUNISH SHARMA, MD, AND BRIAN S. BOXER WACHLER, MD PURPOSE: To evaluate the efficacy of single-segment Intacs and compare with double-segment
LASIK Eye Surgery Report
LASIK Eye Surgery Report LASIK eye surgery can be a liberating experience for people hoping to reduce or eliminate their dependence on glasses and contact lenses. Most patients do not realize how evolved
EVALUATING ASTIGMATISM
EVALUATING ASTIGMATISM Dr. Saurabh Sawhney Dr. Aashima Aggarwal Insight Eye Clinic, New Delhi Introduction: Astigmatism, the so-called final frontier of cataract surgery, is an area of research where few
Our Commitment To You
SYSTEM SUPPORT Quality-crafted, the system boasts dependability with high efficiency and low gas usage. We provide responsive service and maintenance contract options, supported by our nationwide direct
By Dr Waleed Al-Tuwairqi, MD Dr Omnia Sherif, MD Ophthalmology Consultants, Elite Medical & Surgical Center Riyadh -KSA.
By Dr Waleed Al-Tuwairqi, MD Dr Omnia Sherif, MD Ophthalmology Consultants, Elite Medical & Surgical Center Riyadh -KSA Rome, Italy 2013 بسم الرحمن الرحيم In the name of Allah, Most Gracious, Most Merciful
Incision along Steep Axis
Toric IOL An option or a must? ~ 15% cataract surgical patients >1.5 D Options: spectacles, CLs, Incision along steep axis, LRI, AK, toric IOL, Excimer Laser or a combination Walter J. Stark, MD Professor
Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda
Slit Skiascopic-guided Ablation Using the Nidek Laser Scott MacRae, MD; Masanao Fujieda ABSTRACT PURPOSE: To present the approach of using a scanning slit refractometer (the ARK 10000) in conjunction with
ABrand New Refraction Meas urement Instrument
PW ABrand New Refraction Meas urement Instrument The TOPCON KR-9000PW WAVEFRONT ANALYZER incorporates comprehensive measurement of the eye's total optical system which has not been achieved until now,
Comparison of Epi-LASIK and Off-Flap Epi-LASIK for the Treatment of Low and Moderate Myopia
Comparison of Epi-LASIK and Off-Flap Epi-LASIK for the Treatment of Low and Moderate Myopia Maria I. Kalyvianaki, MD, PhD, 1,2 George D. Kymionis, MD, PhD, 1,2 George A. Kounis, PhD, 1 Sophia I. Panagopoulou,
LASIK/PRK following previous eye Surgery
AAO Chicago 2010 LASIK/PRK following previous eye Surgery A. John Kanellopoulos, MD Associate Clinical Professor, NYU Medical School Director: Laservision.gr Eye Institute, Athens, Greece www.brilliantvision.com
Improved glare (halos and scattered light) measurement for post-lasik surgery
Improved glare (halos and scattered light) measurement for post-lasik surgery Stanley Klein, Sam Hoffmann, Adam Hickenbotham, Sabrina Chan, and many more optometry students School of Optometry, Univ. of
LASIK complications and their management ESCRS Munich 9.2003
LASIK complications and their management ESCRS Munich 9.2003 A. John Kanellopoulos, M.D. Associate Professor NYU Medical School Director, LaserVision.gr Eye Institute Case report 1 2 years s/p PK LASIK
Lasik Xtra in: Hyperopia AK Xtra Clear cornea cataract surgery
Lasik Xtra in: Hyperopia AK Xtra Clear cornea cataract surgery A. John Kanellopoulos, M.D. Clinical Profesor of Ophthalmology New York University School of Medicine, New York, NY, USA Laservision.gr Institute,
Looking for Keratoconus
07.31.06 Screening for LASIK: Tips and Techniques Surgeons share helpful strategies for identifying which patients are good candidates and which could be trouble. Christopher Kent, Senior Editor Screening