LASIK Flap Characteristics Using the Moria M2 Microkeratome With the 90-µm Single Use Head
|
|
|
- Amice Garrett
- 10 years ago
- Views:
Transcription
1 ORIGINAL ARTICLE LASIK Flap Characteristics Using the Moria M2 Microkeratome With the 90-µm Single Use Head Ioannis M. Aslanides, MD, PhD; Nikolaos S. Tsiklis, MD; Nikolaos I. Astyrakakis, OD; Ioannis G. Pallikaris, MD, PhD; Mirko R. Jankov, MD ABSTRACT PURPOSE: To evaluate the accuracy and consistency of corneal fl ap thickness, horizontal diameter, and hinge size with the Moria M2 90-µm single use head (Moria, Antony, France). METHODS: Fifty-two myopic patients (104 eyes), mean age 32.6 years, underwent bilateral LASIK with a superior hinged fl ap using the Moria M2 microkeratome (90- µm single use head). Prospective evaluation included fl ap thickness (subtraction method), diameter, hinge size, interface particles, intraoperative complications, and visual recovery. RESULTS: The mean preoperative spherical equivalent refraction was diopters (D) (range: 2.88 to D) and D (range: 3.13 to 9.38 D) for right and left eyes, respectively. The mean preoperative central pachymetry was µm and µm for right and left eyes, respectively. The mean preoperative steepest K was D and D for right and left eyes, respectively. Corneal diameter (white to white) was mm and mm for right and left eyes, respectively. The mean postoperative fl ap thickness was µm (range: 67 to 152 µm) and µm (range: 65 to 151 µm) for right and left eyes, respectively. The mean postoperative fl ap diameter was mm (expected mean according to the nomogram given by the company was 9.5 mm). The mean postoperative hinge chord was mm (expected mean 4.2 mm). No interface particles were detected on slit-lamp examination. CONCLUSIONS: The Moria M2 90-µm single use head is safe with good predictability for LASIK fl ap creation. [J Refract Surg. 2006;xx:xxx-xxx.] L aser in situ keratomileusis (LASIK) is a common refractive surgery for the correction of myopia, hyperopia, and astigmatism. 1 The advantages of the procedure 2 (fast, painless recovery of vision, lack of subepithelial haze) are mainly due to the creation of a corneal flap. At the same time, the use of a microkeratome to create a corneal flap increases the risks of intraoperative complications such as free cap, irregular buttonhole, or incomplete flap There are many different microkeratomes and corresponding heads that can be used to create the corneal flap. 1 The most popular microkeratomes for LASIK are based on the oscillating blade principle. The Moria M2 (Antony, France) is a popular compact automated microkeratome with mechanical stop designed for maximum safety. This prospective study evaluates the accuracy and consistency of corneal flap thickness, horizontal diameter, and hinge size produced by the Moria M2 90-µm single use head, as well as visual acuity on postoperative days 1 and 3, which is related to both the microkeratome cut and the excimer laser ablation. PATIENTS AND METHODS This prospective study included 52 myopic patients (24 men and 28 women, 104 eyes) with a mean age of years (range: 19 to 53 years). Preoperative measurements included uncorrected visual acuity (UCVA), best spectaclecorrected visual acuity, manifest and cycloplegic refraction, slit-lamp examination with fundus evaluation, corneal topography (C-Scan; Technomed Arnold-Sommerfeld-Ring1, Baesweiler, Germany), ultrasonic pachymetry (Pac Scan 300P; From VEIC - Vardinoyiannion Eye Institute of Crete, School of Medicine, University of Crete, Greece. The authors have no financial or proprietary interest in the materials presented herein. Correspondence: Mirko R. Jankov, MD, VEIC - Vardinoyiannion Eye Institute of Crete, University of Crete, PO Box 1352, Voutes, Heraklion Crete, Greece. Tel: ; Fax: ; visioncare@mac. com Received: June 10, 2005 Accepted: April 11, 2006 Journal of Refractive Surgery Volume XX Month/Month 200X 1
2 Sonomed, Lake Success, NY), and corneal diameter (white-to-white distance; Canon autorefractometer; Canon Medical Systems, Irvine, Calif). All pachymetry measurements were done by the same individual (A.N.I.). The mean preoperative spherical equivalent refraction was diopters (D) (range: 2.88 to D) in the right eye and D (range: 3.13 to 9.38 D) in the left eye. The mean preoperative steepest K was D in the right eye and D in the left eye. Mean pachymetry was µm (range: 491 to 595 µm) in the right eye and µm (range: 490 to 603 µm) in the left eye, whereas mean white-to-white corneal diameter in the right eye was mm (range: 11 to 12.5 mm) and mm (range: 11 to 12.5 mm) in the left eye. All patients underwent primary LASIK in both eyes by the same surgeon (A.I.M.) using the Allegretto Wave excimer laser (WaveLight Technologies, Erlangen, Germany). Inclusion criteria were myopia between 2.00 and D, 3.0 D of corneal astigmatism, (manufacturer s guideline), and age 18 years. Exclusion criteria included history of corneal dystrophy or herpetic eye disease, topographic evidence of keratoconus or warpage from contact lenses, corneal scaring, glaucoma, severe dry eye, and collagen vascular diseases. LASIK procedures were performed in a standardized manner. One drop of proparacaine 0.5% (Alcaine; Alcon, Ft Worth, Tex) was instilled in each eye 5 minutes and just before the procedure. This was followed by a povidone-iodine (Betadine) preparation of the eyelids. Eyelashes were isolated by a drape, and a speculum with suction was placed into the operative eye. The cornea was marked with a corneal marker using gentian violet staining. The microkeratome settings (suction ring, flap stop) were chosen according to the steepest K (manufacturer s nomogram), aiming for the maximum flap diameter. The Moria M2 90-µm single use head was used for a desired cut depth of 120 µm and a superior hinge. The standard speed of pass ( speed 2 : 15,000 rpm, 2 seconds of cutting time) was used in all cases. One single use head was used in both eyes of each patient (the right eye was always done first), and the head was discarded upon completion of the procedure. After the microkeratome pass, the flap was lifted and central ultrasound pachymetry of the residual stromal bed was performed. Three measurements were taken and the mean value was subtracted from the preoperative corneal thickness. This difference was considered the flap thickness (subtraction pachymetry). After performing the laser ablation, the flap was floated back into position, and the stomal bed was irrigated with balanced 2 salt solution. Flap alignment was checked using gentian violet premarkings on the cornea, and a striae test was performed to ensure proper flap adherence. Using one drop of prednisolone acetate 1%, which clearly identified the outline of the flap furrow in its whole circumference, as well as the exact dimensions of the hinge chord, the measurements were taken with the use of a surgical caliber. All patients were examined 60 minutes after surgery to check flap adherence. Postoperatively, they were given flurbiprofen sodium 0.03% (Ocuflur; Allergan, Irvine, Calif) drops 4 times a day for 2 days, dexamethasone 0.1% tobramycin 0.3% (Tobradex, Alcon) drops 4 times a day for 2 weeks, and sodium hyaluronate 0.18% (Vismed; TRB Chemedica, Newcastle under Lyme, United Kingdom) drops hourly for 1 month. Patients were instructed to wear protective eye shields at night and to return the following day and on postoperative day 3. Statistical analysis was done using two-tailed Student t test for independent events and linear correlation. RESULTS The mean flap thickness for all eyes was µm, being µm (range: 67 to 152 µm) in the right eye and µm (range: 65 to 151 µm) in the left eye (Figs 1 and 2). The difference was not statistically significant (P=.08). The mean flap diameter was mm (range: 8.5 to 9.9 mm) and the mean hinge chord size was mm (range: 3.8 to 6 mm) (Figs 3 and 4). The relationship between flap thickness and preoperative corneal pachymetry is demonstrated in Figure 5. A statistically significant correlation (R 2 =0.49, P.0001) showed that thicker corneas produce thicker flaps. None of the other tested parameters, such as age, flat K, steep K, or preoperative spherical equivalent refraction, showed any correlation with flap thickness. No intraoperative complications occurred in this series. No interface particles were observed on slit-lamp examination. Mean UCVA (logmar) on postoperative day 1 was (range: 0.30 to 0.08) and (range: 0.28 to 0.08) on postoperative day 3. DISCUSSION A crucial step during LASIK is the creation of a hinged corneal flap 12 due to the potential complication that may occur and cause the postponement of the procedure. The precision in reproducing flap thickness is significant when planning and performing LASIK surgery. A thin flap is generally desirable because it reveals a thicker stromal bed for the ablation, although it is more difficult to manipulate and more prone to journalofrefractivesurgery.com
3 Figure 1. Distribution of flap thickness in 104 myopic eyes that underwent LASIK with the Moria M2 90-µm single use head. Figure 2. Comparison between flap thickness in first (right eyes) and second (left eyes) cuts in 104 myopic eyes that underwent LASIK with the Moria M2 90-µm single use head. Figure 3. Distribution of flap diameter and hinge chord size. The solid line represents the mean flap diameter for all eyes (9.4 mm) and the dashed line represents the expected diameter size (9.5 mm) according to the manufacturer s nomogram. Figure 4. Distribution of hinge chord size (all eyes). The solid line represents the mean hinge chord size for all eyes (4.4 mm) and the dashed line represents the expected hinge chord size (4.2 mm) according to the manufacturer s nomogram. complications such as folds or striae. Given the current recommendation of leaving at least 250 µm 4,13 of residual corneal tissue, a thinner flap allows the correction of a larger amount of myopia. Therefore, a thin LASIK flap theoretically decreases the risk of corneal ectasia. On the other hand, the risk of flap irregularities, buttonholes, and epithelial defects may increase when thinner flaps are created. 14 The concept of a disposable microkeratome has been investigated and introduced in clinical practice by a Journal of Refractive Surgery Volume XX Month/Month 200X number of manufacturers, but it was not until the first encouraging results with the Moria M2 single use 130 head 15 had it been shown that with an adequate design or plastic material choice, the increased flap-related complications with other disposable microkeratomes 16 can be overcome. This study evaluates the safety and efficiency of the Moria M2 90-µm single use head. The M2 90-µm single use head is a disposable unit made of plastic with several advantages over a conventional reusable 3
4 Figure 5. Relationship between flap thickness and preoperative corneal pachymetry. A statistically significant correlation shows thicker corneas to produce thicker flaps. The horizontal solid line shows mean flap thickness (106 µm) for all 104 eyes and the two horizontal dotted lines represent one standard deviation (SD) (17 µm). Flaps thicker and thinner than mean 1 SD have been observed in corneas thicker and thinner than 550 µm prior to the surgery, respectively. microkeratome head. The most important is that a new sterilized head is provided for every patient, which eliminates the need for sterilization by the surgical facility, in turn avoiding possible contamination by microbial pathogens and bacterial endotoxins. Moreover, it requires minimal technical manipulation as the microkeratome blade is pre-assembled with the disposable head, and being in a special package, it can be mounted without being touched. In our study, the M2 single use head tended to cut less ( µm) than the manufacturer expected (120 µm), although with a large range (65 to 152µm). The M2 90-µm single use head seems to cut thinner flaps with smaller variability, comparable not only to other Moria heads 14,15,17-21 but also to other microkeratomes with different heads. 17,19,20,22-26 The mean flap thickness in the second (left) eye was approximately 5.5% thinner than in the fist (right) eye ( µm compared to µm), which is in agreement with other studies. 17,19,20 Although no statistically significant difference (P=.08) was noted between the first (right eye) and second (left eye) cut, this may change with a larger number of eyes. There was a trend showing that thicker corneas produce thicker flaps, which is comparable to other studies. 20,22,23,25,27-30 Corneas as thin as 500 µm and as thick as 589 µm produced flaps with thicknesses between 90 and 133 µm. However, all flaps thicker than 133 µm were been observed in corneas thicker than 550 µm 4 whereas all flaps thinner than 89 µm occurred in corneas that were thinner than 550 µm prior to surgery. None of the other tested parameters, such as age, flat K, steep K, or preoperative spherical equivalent refraction, showed any correlation with flap thickness. The mean flap diameter was mm and the mean hinge chord size was mm, which were close to the expected values according to the manufacturer s nomogram (9.5 mm and 4.2 mm, respectively). A larger flap diameter with a small hinge allows a larger ablation zone, reduces the risk of ablating the hinge, and can be significant in completing the procedure in cases with eccentric flaps. On the other hand, it increases the risk for intraoperative bleeding in cases of neovascularization of the peripheral cornea, as in chronic contact lens users. Although several interspersed interface particles were observed in a previous confocal microscope study after LASIK with the Moria microkeratome, 31,32 no interface particles were noted on slit lamp examination in the present study. The Moria M2 90-µm single use head seems to be safe and accurate. However, intraoperative pachymetry should be done routinely, considering high values of flap thickness, not only to prevent ectasia after LASIK but also for possible enhancement reoperation. REFERENCES 1. Pallikaris IG, Papatzanaki ME, Stathi EZ, Frenschock O, Georgiadis A. Laser in situ keratomileusis. Lasers Surg Med. 1990;10: El Danasoury MA, El Maghraby A, Klyce SD, Mehrez K. Comparison of photorefractive keratectomy with excimer laser in situ keratomileusis in correcting low myopia (from to 5.50 D): a randomized study. Ophthalmology. 1999;106: Jacobs JM, Taravella MJ. Incidence of intraoperative flap complications in laser in situ keratomileusis. J Cataract Refract Surg. 2002;28: Tham VM, Maloney RK. Microkeratome complications of laser in keratomileusis. Ophthalmology. 2000;107: Lin RT, Maloney RK. Flap complications associated with lamellar refractive surgery. Am J Ophthalmol. 1999;127: Stulting RD, Carr JD, Thompson KP, Ferensowicz M, Furlong MT. Complications of laser in situ keratomileusis for the correction of myopia. Ophthalmology. 1999;106: Gimbel HV, Penno EE, van Westenbrugge JA, Ferensowicz M, Furlong MT. Incidence and management of intraoperative and early postoperative complications in 1000 consecutive laser in situ keratomileusis cases. Ophthalmology. 1998;105: Gimbel HV, van Westenbrugge JA, Penno EE, Ferensowicz M, Feinerman GA, Chen R. Simultaneous bilateral laser in situ keratomileusis; safety and efficacy. Ophthalmology. 1999;106: Knorz MC, Jendritza B, Hugger P, Liermann A. Complications after laser in situ keratomileusis (LASIK). Ophthalmologe. 1999;96: Davidorf JM, Zaldivar R, Oscherow S. Results and complica- journalofrefractivesurgery.com
5 tions of laser in situ keratomileusis by experienced surgeons. J Refract Surg. 1998;14: Asano-Kato N, Toda I, Hori-Komai Y, Takano Y, Tsubota K. Risk factors for insufficient fixation of microkeratome during laser in situ keratomileusis. J Refract Surg. 2002;18: Salah T, Waring GO III, El-Maghraby A, Moadel K, Grimm SB. Excimer laser in situ keratomileusis under a corneal flap for myopia of 2 to 20 diopters. Am J Ophthalmol. 1996;121: Walker MB, Wilson SE. Lower intraoperative flap complications rate with the Hansatome microkeratome compared to the Automated Corneal Shaper. J Refract Surg. 2000;16: Pietila J, Makinen P. Getting thinner flaps with Moria M2 microkeratome. Presented at: XXII Congress of the European Society of Cataract & Refractive Surgeons; September 18-22, 2004; Paris, France. 15. Kanellopoulos AJ, Pe LH, Kleiman L. Moria M2 single use microkeratome head in 100 consecutive LASIK procedures. J Refract Surg. 2005;21: Pallikaris IG, Katsanevaki VJ, Panagopoulou SI. Laser in situ keratomileusis intraoperative complications using one type of microkeratome. Ophthalmology. 2002;109: Muallem MS, Yoo SY, Romano AC, Schiffman JC, Culbertson WW. Corneal flap thickness in laser in situ keratomileusis using the Moria M2 microkeratome. J Cataract Refract Surg. 2004;30: Jacobs BJ, Deutsch TA, Rubenstein JB. Reproducibility of corneal flap thickness in LASIK. Ophthalmic Surg Lasers. 1999;30: Solomon KD, Donnenfeld E, Sandoval HP, Al Sarraf O, Kasper TJ, Holzer MP, Slate EH, Vroman DT. Flap thickness accuracy: comparison of 6 microkeratome models. J Cataract Refract Surg. 2004;30: Miranda D, Smith SD, Krueger RR. Comparison of flap thickness reproducibility using microkeratomes with a second motor for advancement. Ophthalmology. 2003;110: Srivannaboon S. Flap analysis: critical point in laser in situ keratomileusis. J Med Assoc Thai. 2001;84: Shemesh G, Dotan G, Lipshitz I. Predictability of corneal flap thickness in laser in situ keratomileusis using three different microkeratomes. J Refract Surg. 2002;18:S347-S Giledi O, Mulhern MG, Espinosa M, Kerr A, Daya SM. Reproducibility of LASIK flap thickness using the Hansatome microkeratome. J Cataract Refract Surg. 2004;30: Kezirian GM, Stonecipher KG. Comparison of the IntraLase femtosecond laser and mechanical keratomes for laser in situ keratomileusis. J Cataract Refract Surg. 2004;30: Flanagan GW, Binder PS. Precision of flap measurements for laser in situ keratomileusis in 4428 eyes. J Refract Surg. 2003;19: Gailitis RP, Lagzdins M. Factors that affect corneal flap thickness with the Hansatome microkeratome. J Refract Surg. 2002;18: Choi YI, Park SJ, Song BJ. Corneal flap dimensions in laser in situ keratomileusis using the Innovatome automatic microkeratome. Korean J Ophthalmol. 2000;14: Yildirim R, Aras C, Ozdamar A, Bahcecioglu H, Ozkan S. Reproducibility of corneal flap thickness in laser in situ keratomileusis using the Hansatome microkeratome. J Cataract Refract Surg. 2000;26: Ucakhan OO. Corneal flap thickness in laser in situ keratomileusis using the Summit Krumeich-Barraquer microkeratome. J Cataract Refract Surg. 2002;28: Yi WM, Joo CK. Corneal flap thickness in laser in situ keratomileusis using an SCMD manual microkeratome. J Cataract Refract Surg. 1999;25: Javaloy Estan J, Vidal MT, Quinto A, Rojas V, Alio JL. Quality assessment model of 3 different microkeratomes through confocal microscopy. J Cataract Refract Surg. 2004;30: Perez-Gomez I, Efron N. Confocal microscopic evaluation of particles at the corneal flap interface after myopic laser in situ keratomileusis. J Cataract Refract Surg. 2003;29: Journal of Refractive Surgery Volume XX Month/Month 200X 5
6 AUTHOR QUERIES Please clarify the manufacturer of the C-Scan. Is the full company name Technomed Arnold-Sommerfeld- Ring1? Please provide the manufacturer (and its location) of Betadine. What is meant by can be significant in the following statement? A larger flap diameter with a small hinge allows a larger ablation zone, reduces the risk for ablating the hinge, and can be significant in completing the procedure in cases with eccentric flaps. AUTHOR QUERIES PER DR WARING Please revise your conclusion in the Abstract and Discussion, which currently states that the instrument gives good predictability for flap thickness, but you have a standard deviation of approximately 17 µm with a range of approximately 70 µm. That does not qualify for good predictability. In addition, the microkeratome cuts approximately 15 µm thicker than the stated 90 µm, although your desired cut depth was 120 µm, which acknowledges this disparity. You begin your Discussion by mentioning flap folds and striae and irregularities. You do not discuss these topics in your Results. Please comment on the number of eyes that had flap folds or striae or irregularities, including those that needed some type of postoperative adjustment. In your Discussion, you note that there is a large range (65 to 152 µm), and yet your Abstract and Conclusion state that the unit is accurate. Please remove allusions to accuracy, and in your Abstract Conclusion and Discussion conclusion please indicate the actual range. 6 journalofrefractivesurgery.com
Intraoperative Management of Partial Flap during LASIK
Intraoperative Management of Partial Flap during LASIK A Small Case Series Report Vikentia J. Katsanevaki, MD, PhD, Nikolaos S. Tsiklis, MD, Nikolaos I. Astyrakakis, OD, Ioannis G. Pallikaris, MD, PhD
Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps
Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps Lingo Y. Lai, MD William G. Zeh, MD Clark L. Springs, MD The authors
Moria M2 Single Use Microkeratome Head in 100 Consecutive LASIK Procedures
Moria M2 Single Use Microkeratome Head in 100 Consecutive LASIK Procedures A. John Kanellopoulos, MD; Lawrence H. Pe, MD; Lynda Kleiman, MD ABSTRACT PURPOSE: To evaluate the safety and effi cacy of the
FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US
FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US JON DISHLER, MD DENVER, COLORADO, USA INTRODUCTION AND STUDY OBJECTIVES This article summarizes the first US
Corneal flap thickness with the Moria M2 single-use head 90 microkeratome
Corneal flap thickness with the Moria M2 single-use head microkeratome Anne Huhtala, 1,2 Juhani Pietila, 1,2 Petri Ma kinen, 1,2 Sakari Suominen, 1,2 Matti Seppa nen 1,2 and Hannu Uusitalo 2,3 1 Mehila
Intraoperative flap complications in laser in situ keratomileusis with two types of microkeratomes q
Saudi Journal of Ophthalmology (2011) 25, 239 243 King Saud University Saudi Journal of Ophthalmology www.saudiophthaljournal.com www.ksu.edu.sa www.sciencedirect.com ORIGINAL ARTICLE Intraoperative flap
Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking
Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking Kay Lam, MD, Dan B. Rootman, MSc, Alejandro Lichtinger, and David S. Rootman, MD, FRCSC Author affiliations:
Overview of Refractive Surgery
Overview of Refractive Surgery Michael N. Wiggins, MD Assistant Professor, College of Health Related Professions and College of Medicine, Department of Ophthalmology Jones Eye Institute University of Arkansas
Thin-flap (sub-bowman keratomileusis) versus thick-flap laser in situ keratomileusis for moderate to high myopia: Case-control analysis
ARTICLE Thin-flap (sub-bowman keratomileusis) versus thick-flap laser in situ keratomileusis for moderate to high myopia: Case-control analysis Dimitri T. Azar, MD, Ramon C. Ghanem, MD, Jose de la Cruz,
Long-Term Outcomes of Flap Amputation After LASIK
Long-Term Outcomes of Flap Amputation After LASIK Priyanka Chhadva BS, Florence Cabot MD, Anat Galor MD, Sonia H. Yoo MD Bascom Palmer Eye Institute, University of Miami Miller School of Medicine Miami
Laser in situ keratomileusis (LASIK) has been. Retreatment of Hyperopia After Primary Hyperopic LASIK REPORTS
REPORTS Retreatment of Hyperopia After Primary Hyperopic LASIK Julio Ortega-Usobiaga, MD, PhD; Rosario Cobo-Soriano, MD, PhD; Fernando Llovet, MD; Francisco Ramos, MD; Jaime Beltrán, MD; Julio Baviera-Sabater,
Initial Supervised Refractive Surgical Experience: Outcome of PRK and LASIK
Initial Supervised Refractive Surgical Experience: Outcome of PRK and LASIK Urmil Shah 1, Mahesh Shah 2, Bharati Shah 3 1 DO, DNB, FICO Resident, Aravind eye hospital, Tirunelveli 627001, Tamilnadu, India
LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY
Thi-Qar Medical Journal (TQMJ): Vol(4) No(4):1(14-21) SUMMARY: LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY Dr. Ali Jawad AL- Gidis (M.B.Ch.B., D.O., F.I.C.O.)* Background: LASIK which
significantly different from that of sequential treatments. Methods: Data were obtained from 254 consecutive patients that were
PROSPECTIVE, RANDOMIZED COMPARISON OF SIMULTANEOUS AND SEQUENTIAL BILATERAL LASIK FOR THE CORRECTION OF MYOPIA* BY G. 0. Waring III, MD, FACS, FRCOPHTH, J. D. Carr, MD, MA, FRCOPHTH, R. D. Stulting, MD,
LASIK for Hyperopia With the WaveLight Excimer Laser
LASIK for Hyperopia With the WaveLight Excimer Laser A. John Kanellopoulos, MD; Joseph Conway, MD; Lawrence H. Pe, MD ABSTRACT PURPOSE: To evaluate the safety and effi cacy of the ALLEGRETTO WAVE excimer
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System Jan Venter, MD ABSTRACT PURPOSE: To determine the predictability, effi cacy, safety, and stability of LASIK using custom ablation
XXXII nd Congress of the ESCRS, London, September 13, 2014 Instructional Course # 7. LASIK: basic steps for safety and great results
XXXII nd Congress of the ESCRS, London, September 13, 2014 Instructional Course # 7 LASIK: basic steps for safety and great results Microkeratomes Jérôme C. VRYGHEM, M.D. Brussels Eye Doctors Brussels
Tamer O. Gamaly, FRCS; Alaa El Danasoury, MD, FRCS; Akef El Maghraby, MD
A Prospective, Randomized, Contralateral Eye Comparison of Epithelial Laser in situ Keratomileusis and Photorefractive Keratectomy in Eyes Prone to Haze Tamer O. Gamaly, FRCS; Alaa El Danasoury, MD, FRCS;
LASIK, Epi LASIK and PRK Past present and future
LASIK, Epi LASIK and PRK Past present and future Ioannis G. Pallikaris MD, PhD Institute of Vision and Optics University of Crete Medical School Heraklion Crete Greece Photorefractive Keratectomy Kerr-Muir
Wavefront technology has been used in our
Wavefront Customized Ablations With the WASCA Asclepion Workstation Sophia I. Panagopoulou, BSc; Ioannis G. Pallikaris, MD ABSTRACT PURPOSE: WASCA (Wavefront Aberration Supported Cornea Ablation) is a
Laser Vision Correction: A Tutorial for Medical Students
Laser Vision Correction: A Tutorial for Medical Students Written by: Reid Turner, M4 Reviewed by: Anna Kitzmann, MD Illustrations by: Steve McGaughey, M4 November 29, 2011 1. Introduction Laser vision
INTRODUCTION MATERIAL AND METHODS
J Med Sci 2013;33(4):191-197 http://jms.ndmctsgh.edu.tw/3304191.pdf DOI:10.6136/JMS.2013.33(4).191 Copyright 2013 JMS Yen-Shou Chou, et al. Accuracy of LASIK Flap Thickness Obtained Using 2 Types of Moria
Ophthalmology Department,Tanta University, Egypt
Moataz M. Sabry, MD, PhD. Ophthalmology Department,Tanta University, Egypt 1 SHOULD WE ABONDONE MECHANICAL MICROKERATOMES? 2 With Mechanical, LASIK is a one-laser procedure, there s no moving of patient
Comparison of Two Procedures: Photorefractive Keratectomy Versus Laser In Situ Keratomileusis for Low to Moderate Myopia
Comparison of Two Procedures: Photorefractive Keratectomy Versus Laser In Situ Keratomileusis for Low to Moderate Myopia Jae Bum Lee, Jae Sung Kim, Chul-Myong Choe, Gong Je Seong and Eung Kweon Kim Institute
LASIK SURGERY OUTCOMES, VOLUME AND RESOURCES
MOH Information Paper: 2006/17 LASIK SURGERY OUTCOMES, VOLUME AND RESOURCES By Dr. Ganga Ganesan 1 I INTRODUCTION LASIK stands for Laser-Assisted In Situ Keratomileusis and is a surgical procedure that
Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD
Post LASIK Ectasia Gina M. Rogers, MD and Kenneth M. Goins, MD October 6, 2012 Chief Complaint: Decreasing vision after laser- assisted in- situ keratomileusis (LASIK) History of Present Illness: This
Consumer s Guide to LASIK
Consumer s Guide to LASIK A Community Service Project brought to you by Price Vision Group Your Guide To A Successful LASIK Procedure The purpose of this educational guide is to help prospective patients
Conductive Keratoplasty to Treat Complications of LASIK and Photorefractive Keratectomy
Conductive Keratoplasty to Treat Complications of LASIK and Photorefractive Keratectomy Peter S. Hersh, MD, 1,2 Kristen L. Fry, OD, 1,2 Ravindra Chandrashekhar, MD, 1 Desipina S. Fikaris, BS 1 Purpose:
Topography-guided Treatment of Irregular Astigmatism With the WaveLight Excimer Laser
ORIGINAL ARTICLES Topography-guided Treatment of Irregular Astigmatism With the WaveLight Excimer Laser Mirko R. Jankov II, MD; Sophia I. Panagopoulou, MS; Nikolaos S. Tsiklis, MD; Georgos C. Hajitanasis,
LASIK. Complications. Customized Ablations. Photorefractive Keratectomy. Femtosecond Keratome for LASIK. Cornea Resculpted
Refractive Surgery: Which Procedure for Which Patient? David R. Hardten, M.D. Minneapolis, Minnesota Have done research, consulting, or speaking for: Alcon, Allergan, AMO, Bausch & Lomb, Inspire, Medtronic,
Topographically-guided Laser In Situ Keratomileusis to Treat Corneal Irregularities
Topographically-guided Laser In Situ Keratomileusis to Treat Corneal Irregularities Michael C. Knorz, MD, Bettina Jendritza, MD Objective: To evaluate the predictability and safety of topographically guided
Alexandria s Guide to LASIK
Alexandria s Guide to LASIK A Community Service Project sponsored by: Wallace Laser Center Your Guide To A Successful LASIK Procedure The word LASIK is actually an acronym for Laser Assisted In-Situ Keratomileusis.
Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy
Keratorefractive Surgery for Post-Cataract Refractive Surprise Moataz El Sawy Departmentof Ophthalmology, Faculty of Medicine,MenoufiyaUniversity, Egypt [email protected] Abstract: Purpose: To evaluate
Comparison of Epi-LASIK and Off-Flap Epi-LASIK for the Treatment of Low and Moderate Myopia
Comparison of Epi-LASIK and Off-Flap Epi-LASIK for the Treatment of Low and Moderate Myopia Maria I. Kalyvianaki, MD, PhD, 1,2 George D. Kymionis, MD, PhD, 1,2 George A. Kounis, PhD, 1 Sophia I. Panagopoulou,
One-year results of photorefractive keratectomy and laser in situ keratomileusis for myopia using a 213 nm wavelength solid-state laser
ARTICLE One-year results of photorefractive keratectomy and laser in situ keratomileusis for myopia using a 213 nm wavelength solid-state laser Nikolaos S. Tsiklis, MD, MSc, George D. Kymionis, MD, PhD,
Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY
Sponsored by Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Results of the T-CAT Phase III Clinical Trial TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Topography-Guided Custom Ablation Treatments (T-CAT) with
The pinnacle of refractive performance.
Introducing! The pinnacle of refractive performance. REFRACTIVE SURGERY sets a new standard in LASIK outcomes More than 98% of patients would choose it again. 1 It even outperformed glasses and contacts
Flap striae after LASIK can be treated successfully
Flap striae after LASIK can be treated successfully Following a few key rules leads to positive outcomes, surgeons say. by Insun Lee, Miten Vasa, and Emil W. Chynn, MD Special to OCULAR SURGERY NEWS LASIK
Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Advances in refractive surgery: microkeratome and femtosecond laser flap creation in relation to safety, efficacy, predictability, and biomechanical stability Karl Stonecipher a, Teresa S. Ignacio b and
PATIENT CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)
INTRODUCTION: You have been diagnosed with myopia (nearsightedness) or hyperopia (farsightedness) with or without astigmatism, or astigmatism alone. Myopia is a result of light entering the eye and focusing
Cornea and Refractive Surgery Update
Cornea and Refractive Surgery Update Fall 2015 Optometric Education Dinner Sebastian Lesniak MD Matossian Eye Associates Disclosures: None Bio: Anterior Segment and Cornea Surgery Fellowship Wills Eye
Conductive keratoplasty (CK) utilizes radiofrequency energy. Original Article
Original Article Comparing the Rate of Regression after Conductive Keratoplasty with or without Prior Laser-Assisted in situ Keratomileusis or Photorefractive Keratectomy Majid Moshirfar, Erik Anderson
One-Year Clinical Results after Epi-LASIK for Myopia
One-Year Clinical Results after Epi-LASIK for Myopia Vikentia J. Katsanevaki, MD, PhD, 1,2 Maria I. Kalyvianaki, MD, 1,2 Dimitra S. Kavroulaki, MD, 1 Ioannis G. Pallikaris, MD, PhD 1,2 Purpose: To evaluate
KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,
Comparison of and Intacs for Keratoconus and Post-LASIK Ectasia MUNISH SHARMA, MD, AND BRIAN S. BOXER WACHLER, MD PURPOSE: To evaluate the efficacy of single-segment Intacs and compare with double-segment
Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery
Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery Farhad Hafezi, MD; Michael Mrochen, PhD; Franz Fankhauser II, MD; Theo Seiler, MD, PhD
Mostafa A. El-Husseiny, Olfat A. Hassanin, Iman M. Eissa and Mohamed Yasser Farag
COMPARISON OF LASER IN SITU KERATOMILEUSIS (LASIK) AND PHOTOREFRACTIVE KERATECTOMY (PRK) IN CORRECTING LOW TO MODERATE MYOPIA WITH OR WITHOUT ASTIGMATISM Mostafa A. El-Husseiny, Olfat A. Hassanin, Iman
Central Islands After LASIK Detected by Corneal Topography
Korean J Ophthalmol Vol. 15:8-14, 2001 Central Islands After LASIK Detected by Corneal Topography Jin Seok Lee, MD, Choun-Ki Joo, MD Department of Ophthalmology, Kangnam St. Mary s Hospital, College of
Laser in situ keratomileusis in patients with corneal guttata and family history of Fuchs endothelial dystrophy
J CATARACT REFRACT SURG - VOL 31, DECEMBER 2005 Laser in situ keratomileusis in patients with corneal guttata and family history of Fuchs endothelial dystrophy Majid Moshirfar, MD, Vahid Feiz, MD, Michael
Simultaneous Topography-guided PRK Followed by Corneal Collagen Cross-linking for Keratoconus
Simultaneous Topography-guided PRK Followed by Corneal Collagen Cross-linking for Keratoconus George D. Kymionis, MD, PhD; Georgios A. Kontadakis, MD, MSc; George A. Kounis, PhD; Dimitra M. Portaliou,
Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser
Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser Sudhank Bharti, MD; Harkaran S. Bains ABSTRACT PURPOSE: To investigate the predictability
Common Co-management Questions
Issue 037 efocus Innovation. Leadership. Passion for Perfection 415.922.9500 --- www.pacificvision.org Common Co-management Questions Top questions recently asked by optometrists co-managing refractive
IntraLase Corp. Clinical Studies Fact Sheet
Page 1 of 12 Contact: Liana Miller (949) 595-4320 [email protected] IntraLase Corp. Clinical Studies Fact Sheet Clinical data validates IntraLase as a superior technology for creating corneal flaps
Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation
Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation Mohammad Naser Hashemian, MD 1 Mahdi AliZadeh, MD 2 Hassan Hashemi, MD 1,3 Firoozeh Rahimi, MD 4 Abstract Purpose: To present
Ectasia after laser in-situ keratomileusis (LASIK)
Ectasia after laser in-situ keratomileusis (LASIK) 長 庚 紀 念 醫 院 眼 科 蕭 靜 熹 Post-LASIK ectasia A rare complication of LASIK Manhattan jury awarded a former investment banker a record $7.25 million for post-lasik
LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth?
JKAU: Med. Sci., Vol. 18 No. 4, pp: 29-36 (2011 A.D. / 1432 A.H.) DOI: 10.4197/Med. 18-4.3 LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth? Ali M. El-Ghatit, MD,
Factors Affecting Long-term Myopic Regression after Laser In Situ Keratomileusis and Laser-assisted Subepithelial Keratectomy for Moderate Myopia
pissn: 111-8942 eissn: 292-9382 Korean J Ophthalmol 216;3(2):92-1 http://dx.doi.org/1.3341/kjo.216.3.2.92 Original Article Factors Affecting Long-term Myopic Regression after Laser In Situ Keratomileusis
Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State
Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State University Columbus, Ohio Refractive Errors Myopia (nearsightedness)
LASIK Eye Surgery Report
LASIK Eye Surgery Report LASIK eye surgery can be a liberating experience for people hoping to reduce or eliminate their dependence on glasses and contact lenses. Most patients do not realize how evolved
Alain Saad, MD, Alice Grise-Dulac, MD, Damien Gatinel, MD, PhD
CASE REPORT Bilateral loss in the quality of vision associated with anterior corneal protrusion after hyperopic LASIK followed by intrastromal femtolaser-assisted incisions Alain Saad, MD, Alice Grise-Dulac,
LASIK Surgery. One Use-Plus. The Moria Option for SBK
LASIK Surgery One Use-Plus The Moria Option for SBK One Use-Plus Accuracy and predictability equivalent to Femto-SBK One Use-Plus SBK Flap thickness using ultrasound pachymetry 1 Flap thickness with Femto-SBK
Refractive Surgery Issue. Inlays and Presbyopia: On the Horizon P. 24. Crack a SMILE or Raise a Flap? P. 30. LASIK Xtra: Who Should Get It? P.
MULTIMODAL IMAGING OF PLACOID DISORDERS P. 42 WILLS RESIDENT CASE SERIES P. 67 NTG: THE NOCTURNAL BLOOD PRESSURE FACTOR P. 54 WISE CHOICES FOR OCULAR DIAGNOSES P. 50 NEW WAYS TO DETECT KERATOCONUS P. 58
Femto-LASIK. Pulsewidth: Ultrashort-pulse micro- machining can make sub- wavelength holes. micromachining
All-laser laser LASIK (Femto( Femto-LASIK) Femto-LASIK 台 大 眼 科 王 一 中 IntraLase 2/1 Perfect Vision Ziemer (DaVinci) Carl Zeiss Meditec Pulsewidth: Femtosecond laser (Nd:Glass)) 153 nm (near infrared) Each
Intracorneal Ring Segments Implantation Followed by Same-day Topography-guided PRK and Corneal Collagen CXL in Low to Moderate Keratoconus
SURGICAL TECHNIQUE Intracorneal Ring Segments Implantation Followed by Same-day Topography-guided PRK and Corneal Collagen CXL in Low to Moderate Keratoconus Waleed Al-Tuwairqi, MD; Mazen M. Sinjab, MD,
All surgical procedures have the potential for
Laser in situ Keratomileusis Retreatment for Residual Myopia and Astigmatism Khaled M. Rashad, MD ABSTRACT PURPOSE: To evaluate the visual and refractive results of laser in situ keratomileusis (LASIK)
Effect of Lasik on Endothelial Cell Count in Patients Treated for Myopia
Original Article Effect of Lasik on Endothelial Cell Count in Patients Treated for Myopia Mirza Jamil Ud din Baig, Khalid Mahmood, Tariq Khan, Zaheer Uddin Aqil Qazi Pak J Ophthalmol 2010, Vol. 26 No.1.....................................................................................................
EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com
ejpmr, 2015,2(3), 436-440 EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com Tumram et al. SJIF Impact Factor 2.026 Research Article ISSN 3294-3211 EJPMR CLINICAL OUTCOME OF TORIC IOL
Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday?
Financial Disclosure Arturo Chayet, MD Tijuana, BC Mexico Perry S. Binder, MS, MD San Diego CA USA I have the following financial interests or relationships to disclose: AMO/IntraLase Corporation - C Acufocus
Comparison of higher-order aberrations after wavefront-guided laser in situ keratomileusis and laser-assisted subepithelial keratectomy
J CATARACT REFRACT SURG - VOL 32, MAY 2006 Comparison of higher-order aberrations after wavefront-guided laser in situ keratomileusis and laser-assisted subepithelial keratectomy So-Hyang Chung, MD, In
OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment
Shinagawa LASIK Center OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment Minoru Tomita, MD, Ph.D 1) Executive Medical Director at Shinagawa LASIK Center, Tokyo, Japan 2) Clinical Professor
Complications of Combined Topography-Guided Photorefractive Keratectomy and Corneal Collagen Crosslinking in Keratoconus
Complications of Combined Topography-Guided Photorefractive Keratectomy and Corneal Collagen Crosslinking in Keratoconus Michelle Cho, M.D. 1 Anastasios John Kanellopoulos, M.D 1,2 New York University
Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure
ORIGINAL ARTICLE Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure Joseph Frucht-Pery, MD; David Landau, MD; Frederik Raiskup, MD; Faik Orucov, MD;