EVALUATION OF LINEAR ACCELERATOR GATING WITH REAL-TIME ELECTROMAGNETIC TRACKING
|
|
|
- Dylan White
- 10 years ago
- Views:
Transcription
1 doi: /j.ijrobp Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 3, pp , 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved /09/$ see front matter PHYSICS CONTRIBUTION EVALUATION OF LINEAR ACCELERATOR GATING WITH REAL-TIME ELECTROMAGNETIC TRACKING RYAN L. SMITH, B.S.,* KRISTEN LECHLEITER, M.S.,* KATHLEEN MALINOWSKI, M.S.,* D. M. SHEPARD, PH.D., y D. J. HOUSLEY, B.S., y M. AFGHAN, PH.D., y JEFF NEWELL, M.S., z JAY PETERSEN, B.S., z BRIAN SARGENT, B.S., z AND PARAG PARIKH, M.D.* *Washington University School of Medicine, St. Louis, MO; y Swedish Cancer Institute, Seattle, WA; and z Calypso Medical Technologies, Seattle, WA Purpose: Intrafraction organ motion can produce dosimetric errors in radiotherapy. Commonly, the linear accelerator is gated using real-time breathing phase obtained by way of external sensors. However, the external anatomy does not always correlate well with the internal position. We examined a beam gating technique using signals from implanted wireless transponders that provided real-time feedback on the tumor location without an imaging dose to the patient. Methods and Materials: An interface was developed between Calypso Medical s four-dimensional electromagnetic tracking system and a Varian Trilogy linear accelerator. A film phantom was mounted on a motion platform programmed with lung motion trajectories. Deliveries were performed when the beam was gated according to the signal from the wireless transponders. The dosimetric advantages of beam gating and the system latencies were quantified. Results: Beam gating using on internal position monitoring provided up to a twofold increase in the dose gradients. The percentage of points failing to be within ±10 cgy of the planned dose (maximal dose, 200 cgy) was 3.4% for gating and 32.1% for no intervention in the presence of motion. The mean latencies between the transponder position and linear accelerator modulation were 75.0 ±12.7 ms for beam on and 65.1 ± 12.9 ms for beam off. Conclusion: We have presented the results from a novel method for gating the linear accelerator using trackable wireless internal fiducial markers without the use of ionizing radiation for imaging. The latencies observed were suitable for gating using electromagnetic fiducial markers, which results in dosimetric improvements for irradiation in the presence of motion. Ó 2009 Elsevier Inc. Respiratory gating, Radiotherapy, Organ motion, Treatment delivery, Real-time tracking. INTRODUCTION The goal of radiotherapy is to maximize the absorbed dose to the target volume while minimizing the dose to the surrounding healthy tissue. Intrafraction motion resulting from respiration can cause the tumor to move considerably throughout treatment. The displacements associated with respiration can be up to 3 cm within the thorax (1). To account for this motion, radiation oncologists must incorporate substantial margins (typically 1 cm in the superoinferior direction and 0.5 cm for both the anteroposterior and the lateral directions) in the design of each planning target volume. This leads to large volumes of irradiated healthy tissue that can limit the total dose the patient can safely receive. This has led researchers to explore beam gating techniques with the goal of more accurate radiation delivery to tumors affected by respiratory motion. Respiratory correlation has been used extensively in computed tomography and magnetic resonance imaging in an effort to reduce breathing-related image artifacts (2, 3). More recently, similar techniques have been used to localize the tumor and gate the linear accelerator (4 8). Conventional gating setups use a variety of techniques to measure breathing motion, including optically tracked external marker blocks, thermocouples, thermistors, strain gauges, and pneumotachographs (9). Current techniques rely on the use of external markers or sensors to determine the internal position of the target. Although a correlation exists between external markers and internal tumor position, for some patients, external marker trajectories do not serve as an adequate surrogate for internal tumor position (10). Respiration induces considerable deformation within the thoracic cavity. As the diaphragm contracts, the internal Reprint requests to: Parag J. Parikh, M.D., Department of Radiation Oncology, Washington University School of Medicine, 4921 Parkview Pl., Lower Level, St. Louis, MO Tel: (314) ; Fax: (314) ; [email protected] Supported in part by Calypso Medical. 920 Conflict of interest: J. Newell, B. Sargent, and J. Petersen are employees of Calypso Medical Technologies, and P. Parikh receives research funding from Calypso Medical Technologies. Received April 25, 2008, and in revised form Jan 20, Accepted for publication Jan 23, 2009.
2 Internal wireless gating d R. L. SMITH et al. 921 anatomy compresses and distends. Often the external anatomy exhibits a good correlation with the motion of the internal structures such as the diaphragm and/or lung tumors (5, 10). The external anatomy moves from respiration; however, studies have shown considerable differences between the external anatomy and internal motion. These differences can come in the form of correlated motion with a phase lag between the external and internal motion, or, less frequently, the motion might not exhibit any correlation. Margeras and Yorke (11) have reported up to a 0.5-s lag between the Varian RPM marker block position and the diaphragm position measured fluoroscopically. Koch et al. (12) found that the correlation was poor and unstable unless the external surrogate measuring the skin surface position was near the tumor. In a study by Berbeco et al. (13), lung tumor motion was measured using continuous fluoroscopy concurrently with measurement of external abdominal surface positions. The amount of residual tumor motion, defined as the amount of tumor motion during a respiratory gate determined from the movement of the external surrogate, showed large fluctuations (>300%) for both intra- and interfraction motion. The residual motion was found to be up to 8 mm in magnitude, which strongly suggested that external position monitoring cannot accurately reflect the internal position of a tumor in all cases. The periods in which the external and internal motion exhibit poor correlation are often transient; however, these transient periods could have dosimetric implications (14). The lack of correlation between the internal and external positions has led investigators to examine alternative techniques for accurately tracking the position of targets inside the thoracic cavity. Shimizu et al. (15) developed a realtime target tracking system that uses four integrated kilovoltage imaging systems. The fluoroscopic imaging system used in this technique provides accurate information on the location of discrete points inside the abdomen. However, accurate tracking of the target comes at the expense of an increased imaging dose. For a single fluoroscope, the estimated skin surface dose rate can up be to 118 cgy/h (16). In addition, for three-dimensional (3D) target tracking, stereoscopic fluoroscopes are necessary, resulting in accumulated dose because of imaging. The Synchrony Respiratory Tracking System treatment option of the CyberKnife robotic radiotherapy system provides another image-based system for tracking internal fiducial markers (17). With that technique, gold fiducial markers are placed inside the thoracic cavity near the tumor, and the patient wears a vest with light-emitting diodes that indicate the position of the chest or abdomen. Before the treatment begins, a series of orthogonal X-ray images are acquired that are used to correlate the position of the external markers to the internal fiducial markers. A correspondence model is developed, and periodic images are obtained during the course of delivery to ensure that the correspondence model remains valid. Although the Synchrony Respiratory Tracking System delivers a lower radiation dose to the patient compared with continuous fluoroscopic imaging, this is achieved at the expense of intermittent absolute knowledge of the internal positions. The Task Group 75 guidelines state that the entrance dose per image can be as great as 0.2 cgy (18). For a 2-h session with imaging performed every 30 s, the patient would receive 48 cgy during the treatment course. Alternative image-based solutions have been investigated that use the on-board imaging functionality of many modern linear accelerators (19). Similar to the fluoroscopy-based solutions, on-board imaging solutions deliver a dose to the patient to image and track the internal markers or tumors. Another factor limiting this technique is that, if used simultaneously, high-energy megavoltage scatter from the treatment beam can degrade the image quality of the kilovoltage images used for tracking (20). Imaging-based methods have the advantage of providing information about the surrounding tissue, something a pure electromagnetic position monitoring solution cannot provide. Continuous electromagnetic position monitoring is now available without an additional dose to the patient (Calypso Medical, Seattle WA). The system uses one or more wireless transponders, which are subject to performance testing as a part of the manufacturing operation to ensure they can stand up to high radiation levels throughout the treatment process. The transponders are currently implanted into the prostate using a 14-gauge needle in a procedure similar to gold fiducial implants currently in use clinically. During treatment planning, the transponder locations, as indicated by volumetric imaging, are recorded with respect to the isocenter and a plan is developed. During delivery, an array is placed above the patient. Four source coils in the array excite the transponders by magnetic induction. After excitation, 32 receiver coils in the array detect the resulting response signal. Each transponder has a unique resonant frequency, and they are sequentially exited to independently query the 3D position information. The array is registered to the room using stereoscopic infrared cameras, and the transponder position is known with respect to isocenter. Balter et al. (21) have reported submillimeter accuracy when tracking the transponders moving at 3 cm/s in a volume cm in width and #27 cm away from the array. An additional study by Santanam et al. (22) confirmed submillimeter accuracy using concurrent onboard kilovoltage imaging. In a clinical prostate cancer treatment study, Willoughby et al. (23) have shown the system to be functional in a linear accelerator environment, even when the linear accelerator is treating directly through the array. No failures of the transponders from the radiation dose have occurred. The system has been cleared by the Food and Drug Administration for use in the prostate and prostatic bed. Potential applications in the lung and abdomen (where motion is substantial) are promising. In the present study, we investigated the feasibility of using real-time electromagnetic tracking for linear accelerator gating. The system uses a spatial gating technique that gates the beam by way of the absolute 3D position of the internal fiducial markers (Fig. 1) instead of using phase or amplitude like conventional external surrogate systems currently in use.
3 922 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 3, 2009 is to be gated off. This causes the electron gun to be asynchronous to the RF pulse of the linear accelerator; thus, suppressing the beam. When the target is inside the 3D volume, the logic on the timer interface sets the gun triggers to be synchronous with the RF pulse of the Clinac, allowing a beam pulse to be delivered. Fig. 1. Spatial gating setup. Four-dimensional phantom moves attached film phantom in realistic breathing trajectories. Real-time position information of transponders implanted in film phantom are acquired using array and sent to decision-making computer. Each position measurement is analyzed to determine whether it is inside a predefined three-dimensional volume. If so, the beam is turned on and delivery proceeds. If position is outside volume, beam hold is enacted and delivery halts until target returns. This approach has two primary advantages. First, the beam is gated using the internal position of the tumor, thereby providing highly accurate positional information. Second, it does not require an additional imaging dose. METHODS AND MATERIALS The Calypso four-dimensional (4D) electromagnetic localization system was modified to export real-time position information to a spatial gating prototype. The prototype consisted of a personal computer that accepted the real-time information via Ethernet and a software application that controlled a transistor-transistor logic signal through the parallel port. The software compared the real-time 3D position information with a predetermined 3D volume to determine whether to enact a beam hold (Fig. 1). Issues surrounding the clinical determination of this 3D volume are addressed in the discussion section. The gating prototype would open and close an electrically isolated relay attached to the beam hold interface of a Varian linear accelerator (Varian Medical, Palo Alto, CA) in the same manner described by others (9). The transistor-transistor logic signal from the decision gating prototype was routed directly to the gating board, with plans to later incorporate a Varian-designed gating interface, which will provide additional buffering, isolation, redundancy, and safety checks. This transistor-transistor logic signal operates by opening a +12-volt signal, which goes to the timer interface board in the Clinac. This signal combines with other logic on the timer interface board and sets delayed electron gun triggers when the Clinac Gating: arc trajectories Feasibility tests were performed to demonstrate linear accelerator gating using positional information from the Calypso system. A standard Calypso quality assurance phantom preloaded with three wireless transponders was placed on a rotating circular motion platform 6 cm from the center of rotation. The Calypso system was set to track one transponder with an update frequency of 30 Hz. The individual transponders are excited sequentially. To increase the frequency of spatial position information, a single transponder was used for this gating application. Single transponders do not provide rotational information. For other applications, multiple transponders could be used, and the target could be gated by way of the centroid (and/or rotation of the transponder plane); however, this would come at the expense of increased latency. In addition, Parikh et al. (24) have shown comparable RMS error when measuring a single transponder position vs. multiple transponders (0.2 mm vs mm), with superior latencies for a single transponder (100-ms integration time per transponder). Typically, the Calypso system operates with an integration time of approximately 80% of the update period. The clinical system therefore has a 100-ms update period (10 Hz) with an 80-ms integration time. For this phase of the study at 30 Hz, the integration time was approximately 26 ms. A2 2-cm field was used to irradiate the phantom on a Varian 23EX accelerator. A gating window was established, and when the beacon was outside this window, the treatment was halted. Although the motion was circular, it was confined to a relatively short arc (2.8 cm). Over the width of the gating window (2 mm), the motion can be considered linear, given that the radius of the arc was 6 cm. The gating window was defined only along the direction of motion. Three film measurements were performed to establish proof of concept. First, a beam was delivered to the static phantom. Second, a nongated delivery was performed, where the disk was programmed to rotate back and forth at 1.2 cm/s over an arc covering 2.8 cm. Finally, a gated beam delivery was performed using the same motion parameters with a gating window of 2 mm. The gating was performed over the center of the arc, with the phantom moving at peak constant velocity. The gating window was positioned such that the phantom moved into and out of the window in both directions of motion along the arc trajectory, allowing for a beam hold on exiting the volume and re-establishment of the beam when the phantom re-entered the volume. Latency estimates To determine the latency of the system, the signal directly from the dynamic phantom was compared with the target current test-point signal from the linear accelerator using a logic analyzer. The target current is the current measured at the metal target of the linear accelerator s electron beam; hence, this signal is analogous to the presence or absence of the treatment beam. Using the target current instead of radiographic methods permits a more precise measurement of the latency using standard test equipment. It also facilitates acquisition of large numbers of beam transitions to accumulate a histogram of latencies. The Calypso system was set to monitor the positions at 30 Hz, with an update period of approximately 26 ms. Using this method, a histogram of latencies was generated for the motion from 200 circular motion cycles (Fig. 2). The latencies
4 Internal wireless gating d R. L. SMITH et al. 923 Fig. 2. Latency histograms for gating system. were separated into two categories: beam-on latency and beam-off latency. Beam-on latency was defined as the duration from the target entering the gating volume (as indicated by a signal from the motion phantom) to the first observed target current pulse on the linear accelerator. Beam-off latency was defined as the duration from the target leaving the gating volume (as indicated by a signal from the motion phantom) to the last observed target current pulse on the linear accelerator. Gating: clinical dosimetry Clinically relevant dosimetric analyses were performed. A fourfield, 6-MV, 200-cGy, 3D conformal radiotherapy plan for a random lung cancer patient was selected for this study. The treatment plan was developed using Pinnacle, version 7.1 (Philips, Madison, WI) and delivered using a Varian Trilogy linear accelerator. The phantom in the present study was composed of a standard solid water phantom with one sheet of the solid water replaced with an equivalently sized acrylic slab containing three electromagnetic transponders. The film phantom was attached to the Washington University 4D motion platform (25), which is a custom stage capable of reproducing physiologic tumor motion in all three axes to submillimeter accuracy. For each exposure, the film was placed in the coronal plane. The platform was programmed using respiratory motion data measured for a lung cancer patient using 4D-computed tomography and spirometry (26) (Fig. 3). The gating system was operated using one transponder and a refresh rate of 25 Hz. The decision was made to use 25 Hz in this clinical dosimetry experiment instead of 30 Hz in the feasibility test because the greater sampling frequency is a difficult computation load for processing on a range of current-generation computing platforms. This should have a negligible effect on the latency, because the integration time only increased by 6 ms compared with 30 Hz. A gating window corresponding to exhalation (3-mm superoinferior margin) was used for lung delivery. The dose was delivered according to the following three scenarios: (1) to the static phantom, (2) to the phantom with the programmed motion patterns with no beam gating, and (3) to the moving phantom gated with signals from the Calypso system. The phantom motion was started simultaneously with the linear accelerator for each run. Analysis was performed to evaluate the dosimetric difference between the gated and nongated films compared with the reference static film irradiated in the absence of motion. Radiographic film was used to record the dose (Kodak EDR2 Readypack, Rochester, NY) and was later digitized using a Vidar film scanner. During the experiment, an additional film was irradiated with a multileaf collimator plan that delivered a series of known doses. This film was digitized and used for Hurter and Driffield (H&D) curve calibration. This H&D curve was then used to calibrate the dose for the remaining films in the study. RESULTS Gating: arc trajectories The film results from the initial phantom study showed a dramatic decrease in motion artifacts when comparing the dynamic gated and nongated films (Fig. 4). Because of the reduced motion artifacts, the dose for the dynamic gated delivery closely matched the dose delivered to the static phantom. Fig. 3. Lung trajectories reconstructed from four-dimensional computed tomography and spirometry from a lung cancer patient (26).
5 924 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 3, 2009 Fig. 4. Film demonstration of linear accelerator beam gating. (A) Static exposure, (B) dynamic exposure without gating, and (C) dynamic exposure with beam gating. Latency estimates The measurements showed that the latency was sufficiently small compared with the tumor velocity, thus providing high-quality dose localization. The mean latencies between the transponder position and linear accelerator modulation were ms for beam-on and ms for beam-off, given as the mean standard deviation (Fig. 2). The difference between the beam-on and beam-off times could be attributed to asymmetry in the linear accelerator turn-on and turn-off times or partially to imperfect alignment of the phantom with respect to the Calypso gating volume. The range in the latencies can be attributed to the Fig. 5. Dose difference maps. Films irradiated in presence of clinically relevant motion subtracted from the static reference case. (A) Entire dose profile. (B) High-gradient region of interest as denoted by box in Fig. a. (C,D) Normalized difference maps calculated to show over- and underdosing as percentage of maximal dose. Red and blue regions indicate over- and underdosing, respectively. Gating reduced spread and magnitude of dose mismatch occurring in presence of motion.
6 Internal wireless gating d R. L. SMITH et al. 925 software implementation of the gating decision unit, as well as to the finite integration times of the transponders (26 ms). The latency associated with enacting a beam hold or re-establishing treatment by way of the linear accelerator was relatively small, approximately 17 ms (27). Given the experimental setup, this value was incorporated into the total latency values reported for the spatial gating system. The update rates and latencies of the system were comparable to the optical-based (28) and fluoroscopy-based (15) gating systems reported previously. Gating: clinical dosimetry Dosimetric films were used to determine the dose profile from one fraction of treatment. One baseline run with no motion was used to generate a static film. This film was used as the ideal dose distribution in the absence of patient motion. The static film was compared with the films irradiated using the same treatment plan delivered both with and without gating in the presence of motion. Using beam gating, better dose localization was observed, and the film results showed better correlation with the static dose distribution (Fig. 5). The effects of gating were most evident in the regions of a highdose gradient, because the nongated case effectively blurs the dose over the region that passes through the isocenter during respiratory motion. Difference maps have shown that the dose blurring found in the nongated dynamic case is significantly reduced when the gating solution is implemented (Fig. 5). Dosimetric analysis was performed to quantify the level of over- and underdosing. For the no-intervention case, 32.1% of the points failed to be within 10 cgy from the ideal dose and 8.6% failed for 20 cgy. For gating, 3.4% failed for 10 cgy and 0.0% failed to be within 20 cgy. Gamma analysis was performed on the nongated and gated films. Although no points failed a 3-mm/3% test, 8.3% of the points in the nongated film failed at 1.5 mm/1.5% compared with 0% of the points in the gated film (Fig. 6). It is evident from multiple line profiles that gating produces an increase in the achievable dose gradients (Fig. 7). This increase in dose gradients has clinical implications, which have been addressed in the Discussion section. Even when considering the small gating window used for the present study, the duty cycle was 47% and 49%, respectively, for each of the two lung trajectories. This shows that for most cases the increase in treatment time was small compared with the time spent initially aligning the patient and moving the gantry to the various beam angles. This will not be the case for instances of drastic motion or when the target leaves the gating volume for an extended period. DISCUSSION Gating is a widely used technique for dose localization. One of the limiting factors in the effectiveness of gating is that most implementations use external markers to predict the internal movement of tumors. Although studies have shown a correlation between external and internal motion, variations of about 1 cm have been found between internal fiducial motion and external markers (10, 13). Thus, it is important to implement a solution for determining the precise location of the internal anatomy without exposing the patient to an additional imaging dose throughout the treatment course. It has been shown that large latencies can produce a phase mismatch between beam gating and the tumor position (28). For the initial studies shown in the present report, a software-based decision-making setup was implemented. For clinical implementation, a hardware-based solution would offer lower latencies. The latencies associated with our system are as good as or better than alternative options. For instance, fluoroscopic and optical gating systems have claimed latencies of 90 ms (15) and 170 ms (cite 28) ms, respectively. Note, the low latencies associated with our setup demonstrated a measurable dosimetric difference without the use of predictive algorithms (29). This internal tracking implementation Fig. 6. Gamma (3-mm/3%) maps for gated and nongated cases for region of interest denoted in Fig. 5.
7 926 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 3, 2009 Fig. 7. (A) Static dose profile. Gating reduced dose blurring and improved dose gradients compared with no intervention. (B D) Dose gradients of line profiles analyzed for lines at y = 3 mm, 8 mm, and 13 mm. Raw data plotted with polynomial fit overlaid. Gating improved dose gradients to match delivery in absence of motion. can be incorporated with any linear accelerator in a standardsize vault. In clinical implementation, the exact dimensions of the 3D gating volume will likely vary from patient to patient. The 3D volume would be chosen according to a number of factors: the relationship and level of correlation between the transponder and the tumor as evidenced by respiratory-correlated imaging, the proximity to normal structures, the amount of target motion, and the desired efficiency of the treatment. The number of implanted transponders will have an adverse affect on the update rate of the system. The use of a single transponder increases the acquisition frequency for the spatial position information, but at the expense of the rotational information obtained by using multiple transponders. Studies are needed to determine the cost/benefit ratio from acquiring spatial information from multiple transponders compared with the additional latency associated with multiple transponder readings. For instance, in one potential clinical implementation, multiple transponders could be used during the patient setup process, but a single transponder could be localized for gating throughout the treatment course. Work is needed to ensure that implantation in the lung is safe. Pneumothorax is a typical complication with percutaneous implantation of a fiducial marker in the lung. Although bronchoscopic implants have lower pneumothorax rates (1.8%) (30) than implants done percutaneously (33%) (31), additional work is needed to ensure the system is safe for patient use. Research to develop a bronchoscopic implantation technique for electromagnetic transponders is promising (32). Additionally, the implanted electromagnetic transponders have been shown to be stable in the prostate. Targeting of a lung tumor might be more challenging, because the transponders will not likely have a fixed relationship to the lung tumor. The incorporation of the uncertainty will affect the size of a gating window. Work on a modified transponder design with stability features has shown good fixation to lung tissue (32).
8 Internal wireless gating d R. L. SMITH et al. 927 If left unchecked, breathing motion prevents high-dose gradient regions in which the delivered dose to the surrounding healthy tissue decays rapidly. High-dose gradients are necessary for dose escalation to tumor sites while ensuring that the surrounding critical structures do not receive a substantial dose. As noted in Fig. 7, the dose gradients achieved using the gating solution were larger than those achieved with no intervention in the presence of motion. CONCLUSION An electromagnetic tracking system has been successfully interfaced with a linear accelerator gating system. The latencies measured were comparable to those of other real-time radiotherapy systems, and film experiments using realistic lung trajectories showed that gating provides significant dosimetric improvements. REFERENCES 1. Langen KM, Jones DT. Organ motion and its management. Int J Radiat Oncol Biol Phys 2001;50: Ramsey CR, Cordrey IL, Oliver AL. A comparison of beam characteristics for gated and nongated clinical x-ray beams. Med Phys 1999;26: Pan T, Lee TY, Rietzel E, et al. 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT. Med Phys 2004;31: Shirato H, Shimizu S, Kunieda T, et al. Physical aspects of a real-time tumor-tracking system for gated radiotherapy. Int J Radiat Oncol Biol Phys 2000;48: Margeras G, Yorke E, Rosenzweig K. Initial clinical evaluation of a respiratory gating radiotherapy system. Proceedings of the 22nd annual EMBS International Conference, July 23 28, 2000, Chicago, IL. 6. Kini VR, Vedam SS, Keall PJ, et al. Patient training in respiratory-gated radiotherapy. Med Dosim 2003;28: Kubo HD, Len PM, Minohara S, et al. Breathing-synchronized radiotherapy program at the University of California Davis Cancer Center. Med Phys 2000;27: Berson AM, Emery R, Rodriguez L, et al. Clinical experience using respiratory gated radiation therapy: Comparison of freebreathing and breath-hold techniques. Int J Radiat Oncol Biol Phys 2004;60: Kubo HD, Hill BC. Respiration gated radiotherapy treatment: A technical study. Phys Med Biol 1996;41: Gierga DP, Brewer J, Sharp GC, et al. The correlation between internal and external markers for abdominal tumors: Implications for respiratory gating. Int J Radiat Oncol Biol Phys 2005;62: Margeras GS, Yorke E. Deep inspiration breath hold and respiratory gating strategies for reducing organ motion in radiation treatment. Semin Radiat Oncol 2004;14: Koch N, Liu HH, Starkschall G, et al. Evaluation of internal lung motion for respiratory-gated radiotherapy using MRI: Part I Correlating internal lung motion with skin fiducial motion. Int J Radiat Oncol Biol Phys 2004;60: Berbeco RI, Nishioka S, Shirato H, et al. Residual motion of lung tumours in gated radiotherapy with external respiratory surrogates. Phys Med Biol 2005;50: Ozhasoglu C, Murphy MJ. Issues in respiratory motion compensation during external-beam radiotherapy. Int J Radiat Oncol Biol Phys 2002;52: Shimizu S, Shirato H, Ogura S, et al. Detection of lung tumor movement in real-time tumor-tracking radiotherapy. Int J Radiat Oncol Biol Phys 2001;51: Shirato H, Oita M, Fujita K, et al. Feasibility of synchronization of real-time tumor-tracking radiotherapy and intensity-modulated radiotherapy from viewpoint of excessive dose from fluoroscopy. Int J Radiat Oncol Biol Phys 2004;60: Seppenwoolde Y, Berbeco RI, Nishioka S, et al. Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: A simulation study. Med Phys 2007;34: Murphy MJ, Balter J, Balter S, et al. The management of imaging dose during image-guided radiotherapy: Report of the AAPM Task Group 75. Med Phys 2007;34: Berbeco RI, Jiang SB, Sharp GC, et al. Integrated radiotherapy imaging system (IRIS): Design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors. Phys Med Biol 2004;49: Luo W, Yoo S, Wu J, et al. Effect of MV scatter on kv image quality during simultaneous kv-mv imaging. Int J Radiat Oncol Biol Phys 2007;69:S Balter JM, Wright JN, Newell LJ, et al. Accuracy of a wireless localization system for radiotherapy. Int J Radiat Oncol Biol Phys 2005;61: Santanam L, Malinowski K, Hubenshmidt J, et al. Fiducialbased translational localization accuracy of electromagnetic tracking system and on-board kilovoltage imaging system. Int J Radiat Oncol Biol Phys 2008;70: Willoughby TR, Kupelian PA, Pouliot J, et al. Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer. Int J Radiat Oncol Biol Phys 2006;65: Parikh P, Hubenschmidt J, Dimmer S, et al. 4Dverificationof real-time accuracy of the Calypso system with lung cancer patient trajectory data. Int J Radiat Oncol Biol Phys 2005;63:S26 S Malinowski K, Lechleiter K, Hubenschmidt J, et al. Use of the 4D phantom to test real-time targeted radiation therapy device accuracy. Med Phys 2007;34: Low DA, Parikh PJ, Lu W, et al. Novel breathing motion model for radiotherapy. Int J Radiat Oncol Biol Phys 2005;63: Hanna G, Wang Z, Yin F. Time-delay measurement of a Varian real-time position management system. Med Phys 2006;33: Jin J, Yin F. Time delay measurement for linac based treatment delivery in synchronized respiratory gating radiotherapy. Med Phys 2005;32: Sharp GC, Jiang SB, Shimizu S, et al. Prediction of respiratory tumour motion for real-time image-guided radiotherapy. Phys Med Biol 2004;49: Imura M, Yamazaki K, Shirato H, et al. Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy. Int J Radiat Oncol Biol Phys 2005;63: Whyte RI, Crownover R, Murphy MJ, et al. Stereotactic radiosurgery for lung tumors: Preliminary report of a phase I trial. Ann Thorac Surg 2003;75: Mayse ML, Smith RL, Park M. Development of a non-migrating electromagnetic transponder system for lung tumor tracking. Int J Radiat Oncol Biol Phys 2008;72(Suppl.):S430.
Gated Radiotherapy for Lung Cancer
Gated Radiotherapy for Lung Cancer Steve B. Jiang, Ph.D. Depart Of Radiation Oncology University of California San Diego [email protected] radonc.ucsd.edu/research/cart Two Types of Gating Internal gating
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
A Patient s Guide to the Calypso System for Breast Cancer Treatment
A Patient s Guide to the Calypso System for Breast Cancer Treatment Contents Introduction How Radiation Works To Treat Cancer Radiation Treatment for Breast Cancer Special Concerns for Treating the Left
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology [email protected] http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
Radiation Protection in Radiotherapy
Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France [email protected] Radiation therapy The lecture is oriented to
Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients
대한방사선종양학회지 2005;23(3):169~175 Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients Department of Radiation Oncology, College of Medicine, Ewha Womans University
World-first Proton Pencil Beam Scanning System with FDA Clearance
Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi
02 CyberKnife: Treatment Delivery
TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options
Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan
An Brief Introduction to 4D CT Scanning Steve B. Jiang, Ph.D. Dept of Radiation Oncology Univ of California San Diego Outline Problems with free breathing 3D scanning What is 4D CT? How does it work? Acknowledgements
QA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
What is the CyberKnife System?
CYBERKNIFE Robotic Radiosurgery System PATIENT BROChURE 1 What is the CyberKnife System? have stabilizing frames bolted to their head or limit their breathing during treatment to minimize movement of the
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
CBCT for Prone Breast. Todd Jenkins, MS, DABR Nash Cancer Treatment Center
CBCT for Prone Breast Todd Jenkins, MS, DABR Nash Cancer Treatment Center Disclosures No outside funding or support Disclosures Techniques likely apply across vendors Prone Breast Technique Rationale
Deformable Registration for Image-Guided Radiation Therapy
Symposium Multimodality Imaging for Radiotherapy : State of the art, needs and perspectives 1/33 Deformable Registration for Image-Guided Radiation Therapy David Sarrut «Radiation, images, oncology» team
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_the_prostate
NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
Horizon Blue Cross Blue Shield of New Jersey 2012 Radiation Therapy Payment Rules
55875 55876 55876 Description Transperineal placement of needles or catheters into prostate for interstitial radioelement application, with or wout cytosocopy 55920 Placement of interstitial device(s)
Radiation Therapy in Prostate Cancer Current Status and New Advances
Radiation Therapy in Prostate Cancer Current Status and New Advances Arno J. Mundt MD Professor and Chairman Dept Radiation Oncology Moores Cancer Center UCSD Radiation Therapy Wilhelm Roentgen (1845-1923)
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard
Department of Radiation Oncology
Department of Radiation Oncology Welcome to Radiation Oncology at Emory Clinic Every member of Emory Clinic Department of Radiation Oncology strives to provide the highest quality of care for you as our
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
The Management of Respiratory Motion in Radiation Oncology
AAPM REPORT NO. 91 The Management of Respiratory Motion in Radiation Oncology Report of AAPM Task Group 76 July 2006 DISCLAIMER: This publication is based on sources and information believed to be reliable,
Gamma Knife and Axesse Radiosurgery
Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery
RADIOTHERAPY Giovanna Mantello
F The impact on RADIOTHERAPY Giovanna Mantello AOU Osp. Riuniti Ancona [email protected] RADIOTHERAPY OF PULMONARY LESIONS INTENSITY MODULATION HIGH GRADIENT DOSE DOSE ESCALATION IPOFRACTIONATION STEREO BODY
REPORT ORGANIZATION ABSTRACT I. INTRODUCTION CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE
E S T I M A T I O N O F T H E I M A G I N G D O S E F O R T H E C Y B E R K N I F E R O B O T I C R A D I O S U R G E R Y S Y S T E M REPORT ORGANIZATION ABSTRACT I. INTRODUCTION II. III. IV. CALCULATION
Radiation Therapy in Prostate Cancer Current Status and New Advances
Radiation Therapy in Prostate Cancer Current Status and New Advances Arno J. Mundt MD Professor and Chairman Dept Radiation Oncology Moores Cancer Center UCSD Presentation Welcome Overview of UCSD, Moores
IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect [email protected]. Protect, Enhance and Save Lives
Vincent Bossier System Architect [email protected] IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT
Real-time Position Management System
Real-time Position Management System COMPREHENSIVE SYSTEM FOR TOTAL MOTION MANAGEMENT RPM gates the imaging and treatment beams on and off at predetermined intervals based upon the patient s breathing
MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015)
MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015) CODE CODE TYPE Description 0330 REVCODE RADIOLOGY/THERAPEUTIC - GENERAL CLASSIFICATION 0333 REVCODE RADIOLOGY/THERAPEUTIC
Cyberknife Information Guide. Prostate Cancer Treatment
Cyberknife Information Guide Prostate Cancer Treatment CYBERKNIFE INFORMATION GUIDE PROSTATE CANCER TREATMENT As a patient recently diagnosed with localized prostate cancer, it is important that you familiarize
AK IMRT Würzburg 26./27.03.2009. Roland Kramer, Conmedica GmbH, Schriesheim
AK IMRT Würzburg 26./27.03.2009 Roland Kramer, Conmedica GmbH, Schriesheim 2 AK IMRT Würzburg 26./27.03.2009 Roland Kramer, Conmedica GmbH, Schriesheim OVERVIEW 3 Company Background AlignRT GateCT and
Protons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D.
Protons vs. CyberKnife UC SF Protons vs. CyberKnife UC SF Alexander R. Gottschalk, M.D., Ph.D. Associate Professor and Director of the CyberKnife Radiosurgery Program Department of Radiation Oncology University
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth
Executive Office of Health and Human s Office of Medicaid www.mass.gov/masshealth May 2015 TO: FROM: RE: Radiation Oncology Centers Participating in Daniel Tsai, Assistant Secretary and Director of (2015
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
First Three Years After Project Proton Therapy Facility:
39. Project Related Utilization and Financial Information. Proton Therapy Facility: Unit of Service Delivery Projected Utilization FY2017 FY2018 FY2019 Estimated Patients 448 670 808 Average Revenue Per
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Introduction Prostate is one of the treatment sites that is well suited for IMRT. For
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
METHODS OF BOLUSING THE TRACHEOSTOMY STOMA
PII S0360-3016(00)01550-9 Int. J. Radiation Oncology Biol. Phys., Vol. 50, No. 1, pp. 69 74, 2001 Copyright 2001 Elsevier Science Inc. Printed in the USA. All rights reserved 0360-3016/01/$ see front matter
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis 11/2009 5/2016 5/2017 5/2016
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Abdomen File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis
FOR IMMEDIATE RELEASE
FOR IMMEDIATE RELEASE The PROBEAT-RT Proton Beam Therapy System Incorporating Real-time Tumor-tracking is Approved for Commercial Manufacturing and Sales under Japan s Pharmaceutical Affairs Law Treatments
Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015
INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design
Implementation Date: April 2015 Clinical Operations
National Imaging Associates, Inc. Clinical guideline PROSTATE CANCER Original Date: March 2011 Page 1 of 5 Radiation Oncology Last Review Date: March 2015 Guideline Number: NIA_CG_124 Last Revised Date:
Comparison of spirometry and abdominal height as four-dimensional computed tomography metrics in lung
Comparison of spirometry and abdominal height as four-dimensional computed tomography metrics in lung Wei Lu, Daniel A. Low, a Parag J. Parikh, Michelle M. Nystrom, Issam M. El Naqa, Sasha H. Wahab, Maureen
Proton Therapy. What is proton therapy and how is it used?
Scan for mobile link. Proton Therapy Proton therapy delivers radiation to tumor tissue in a much more confined way than conventional photon therapy thus allowing the radiation oncologist to use a greater
Current and Future Trends in Proton Treatment of Prostate Cancer
Current and Future Trends in Proton Treatment of Prostate Cancer Reinhard W. Schulte Assistant Professor Department of Radiation Medicine Loma Linda University Medical Center Loma Linda, CA, USA Outline
Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA
Cancer Research and Oncology: Open Access Received: Nov 16, 2015, Accepted: Jan 12, 2016, Published: Jan 15, 2016 CancerResOncol, Volume 1, Issue 1 http://crescopublications.org/pdf/crooa/crooa-1-001.pdf
HITACHI Proton Beam Therapy System
HITACHI Proton Beam Therapy System Masumi Umezawa, M.S. Hitachi, Ltd., Hitachi Research Laboratory HITACHI Proton Beam Therapy System Contents 1. Overview of HITACHI 2. HITACHI Proton Beam Therapy and
Intensity-Modulated Radiation Therapy (IMRT)
Scan for mobile link. Intensity-Modulated Radiation Therapy (IMRT) Intensity-modulated radiotherapy (IMRT) uses linear accelerators to safely and painlessly deliver precise radiation doses to a tumor while
4DCT radiotherapy for NSCLC: a review of planning methods
Journal of Radiotherapy in Practice Journal of Radiotherapy in Practice (2014) Page 1 of 10 & Cambridge University Press 2014 doi:10.1017/s1460396914000041 Literature Review 4DCT radiotherapy for NSCLC:
Basic Radiation Therapy Terms
Basic Radiation Therapy Terms accelerated radiation: radiation schedule in which the total dose is given over a shorter period of time. (Compare to hyperfractionated radiation.) adjuvant therapy (add-joo-vunt):
Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies
Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies Presented by Susan Granucci, Healthcare Reimbursement Specialist Topics 1. Intensity Modulated Radiation Therapy (IMRT) 2. Tomotherapy
Volumetric modulated arc therapy (VMAT) for prostate cancer
Planning study Volumetric modulated arc therapy (VMAT) for prostate cancer Institution: Purpose: Swedish Cancer Institute, Seattle, WA, USA In a community cancer center, treating prostate cancer comprises
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
Two years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions
Two years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions Laurent Desbat TIMC-IMAG, UMR 5525, Grenoble University, France [email protected] Rolf
Corporate Medical Policy
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Head and Neck File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_head_and_neck
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
Intensity-modulated radiotherapy and volumetric-modulated arc therapy for malignant pleural mesothelioma after extrapleural pleuropneumonectomy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 4, 2013 Intensity-modulated radiotherapy and volumetric-modulated arc therapy for malignant pleural mesothelioma after extrapleural pleuropneumonectomy
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Quality Assurance in Stereotactic. Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
Modelling physiological features of Human body behavior in car crash simulations
Modelling physiological features of Human body behavior in car crash simulations Michel BEHR, Yves GODIO, Maxime LLARI, Christian BRUNET Laboratoire de Biomécanique Appliquée (Marseille, France) Presented
LINAC-based stereotactic radiosurgery for treatment of trigeminal neuralgia
LINAC-based stereotactic radiosurgery for treatment of trigeminal neuralgia Bruce J. Gerbi, Patrick D. Higgins, 1 Kwan H. Cho, 2 and Walter A. Hall 3 Department of Therapeutic Radiology-Radiation Oncology,
TITLE: Comparison of the dosimetric planning of partial breast irradiation with and without the aid of 3D virtual reality simulation (VRS) software.
SAMPLE CLINICAL RESEARCH APPLICATION ABSTRACT: TITLE: Comparison of the dosimetric planning of partial breast irradiation with and without the aid of 3D virtual reality simulation (VRS) software. Hypothesis:
Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com
Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider
Stereotactic Radiotherapy for Prostate Cancer using CyberKnife
Stereotactic Radiotherapy for Prostate Cancer using CyberKnife Delivering the best in care UHB is a no smoking Trust To see all of our current patient information leaflets please visit www.uhb.nhs.uk/patient-information-leaflets.htm
International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters
www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in
Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT
Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Wen Li, Andrew Vassil, Lama Mossolly, Qingyang Shang, Ping Xia Department of Radiation Oncology Why
Dose Tracking System:
Dose Tracking System: A Paradigm Shift in Patient Dose Monitoring Andrew Kuhls-Gilcrist, Ph.D. XRVL Clinical Science Manager Toshiba America Medical Systems, Inc. 2 Dose Tracking System: A Paradigm Shift
Proton Therapy: Cutting Edge Treatment for Cancerous Tumors. By: Cherilyn G. Murer, JD, CRA
Proton Therapy: Cutting Edge Treatment for Cancerous Tumors By: Cherilyn G. Murer, JD, CRA Introduction Put simply, proton therapy is a new cutting edge cancer treatment that promises better outcomes for
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
Quantitative Imaging In Clinical Trials Using PET/CT: Update
Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative
Radiation Therapy for Prostate Cancer: Treatment options and future directions
Radiation Therapy for Prostate Cancer: Treatment options and future directions David Weksberg, M.D., Ph.D. PinnacleHealth Cancer Institute September 12, 2015 Radiation Therapy for Prostate Cancer: Treatment
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
Brain Tumor Treatment
Scan for mobile link. Brain Tumor Treatment Brain Tumors Overview A brain tumor is a group of abnormal cells that grows in or around the brain. Tumors can directly destroy healthy brain cells. They can
Patient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University
Key words: treatment planning, quality assurance, 3D treatment planning
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical
Clinical Education A comprehensive and specific training program. carry out effective treatments from day one
Proton Therapy Clinical Education A comprehensive and specific training program carry out effective treatments from day one Forewarned is forearmed Although over 100,000 patients have been treated in proton