A Guide to Laboratory Fume Hoods
|
|
|
- Annabelle Lucas
- 10 years ago
- Views:
Transcription
1 A Guide to Laboratory Fume Hoods
2 INTRODUCTION Laboratory fume hoods are a type of ventilation system with the primary function is to exhaust chemical fumes, vapors, gasses, dust, mist and aerosol. Fume hoods also serve as physical barriers between reactions and the laboratory, offering a measure of protection against inhalation exposure, chemical spills, run-away reactions and fires. A typical fume hood has a box like structure with a moveable sash window. Experimental procedures are performed within the hood which is consistently and safely ventilated, usually by means of an extract blower and ductwork. Chemical fumes exhausted and diluted many times over in the atmosphere and has a negligible effect to human health. When environment concerns are of importance, an extract treatment system, often referred to as a scrubber is installed to remove most of the vapors from the exhaust air stream. 2 The hood functions by maintaining a relatively negative pressure in the interior of the hood to prevent any contaminant from escaping while drawing air in through the hood opening at a consistent rate. A suitable hood face velocity (the speed at which air is drawn into the hood) is of importance to the safe and effective operation of a fume hood. While excessive face velocities can often result in turbulence and reduce containment, insufficient velocities can also compromise hood performance. In general, a hood s face velocity is recommended to be between 0.3 m/s (60 fpm) and 0.5 m/s (100 fpm), however it is important to check with local safety regulations on the face velocity recommendation before using the fume hoods. Most hoods commonly sized for a minimum face velocity at full sash opening; but as means to conserve energy some hoods size the minimum face velocity of the hood at half-sash opening creating new low flow fume hoods, which are now present in the market. Types of Fume Hoods General Purpose, Bench Top Fume Hood Distillation Fume Hood A distillation hood is characterized by a low worktop height which results in a large working height for the operator. This allows tall distillation equipment to be installed and mounted in the work chamber. Otherwise, it has similar features to that of a standard fume hood. Perchloric Acid Fume Hood Perchloric acid reacts violently with organic materials. Dried perchloric acid is also highly explosive. Therefore, perchloric fume hoods require built-in water wash down systems in order to prevent perchlorate salt deposits. Interior liners are made of acid resistant materials like stainless steel. Interior corners are coved to aid in cleaning. All procedures that use perchloric acid must be confined to a perchloric fume hood, to prevent dangerous reactions with other chemicals. A Guide to Laboratory Fume Hoods Introduction
3 Radioisotope Fume Hood Radioisotope fume hoods are constructed specifically to protect users from radioactive materials. They have specially constructed worktops to withstand the weight of lead shielding plates, and may also have lead laced sashes. Interiors are made of stainless steel with coved corners to aid in decontamination. Acid Digestion Fume Hood Acid digestion fume hoods have special liners manufactured of acid resistant materials such as unplasticized PVC. For acid digestion applications involving high service temperatures, other materials such as PVDF may be used. Sashes may be made of polycarbonate to resist hydrofluoric acid etching. Floor Mounted Fume Hood Floor mounted fume hoods are used for applications which require large apparatus. As the name implies, these hoods are floor mounted without any work surface. This facilitates the transfer of equipment and materials into, and out from the hood. Floor mounted hoods are sometimes referred to, although wrongly, as walk-in fume hoods. Demonstration Fume Hood A demonstration hood has all 4 sides made of safety glass, and this hood is commonly used in educational institutions to allow students to easily view the teachers demonstrations inside the fume hood from all angles, enhancing efficiency in teaching laboratories. Ductless Fume Hood Ductless fume hoods utilize activated carbon filtration to adsorb chemical vapors and fumes. These hoods recirculate air to the laboratory, and are growing in popularity because of energy savings and the green movement. 3 Not Fume Hoods Biological safety cabinets, laminar flow clean benches, pharmacy isolators and glove boxes are sometimes mistaken for fume hoods. These devices utilize particulate filters which do not remove chemical vapors. With specific rare exceptions, they should not be used for work involving chemicals.
4 Important components of a Fume Hood Hood Body The visible part of the fume hood that serves to contain hazardous gases and vapors. 2. Baffles Moveable partitions used to create slotted openings along the back of the hood body. Baffles keep the airflow uniform across the hood opening, thus eliminating dead spots and optimizing capture efficiency. 3. Sash The sliding door to the hood. By using the sash to adjust the front opening, airflow across the hood can be adjusted to the point where capture of contaminants is maximized. Each hood has its optimum sash configuration. The sash should be held in this position when working in the hood and closed completely when the hood is not in use. The sash may be opened completely to set up equipment, but must be returned to the optimum setting prior to generating contaminants inside the hood. Hoods are equipped with one of three types of sashes: Vertical Vertical sashes provide the greatest overall access and are the most common. 5 A Guide to Laboratory Fume Hoods Important Components of a Fume Hood
5 Horizontal Horizontal sashes provide the best access to the top interior of the hood, but do not provide access to the entire width at the same time. However, they can be positioned as a body shield to protect the operator, while enabling access through the sides. Combination vertical and horizontal Combination sashes combine horizontal sash panes, in a vertical rising frame. As such, they offer the advantages of both vertical and horizontal sashes. 4. Airfoil Located along the bottom and side edges the airfoil streamlines airflow into the hood, preventing the creation of turbulent eddies that can carry vapors out of the hood. The space below the bottom airfoil provides source of room air for the hood to exhaust when the sash is fully closed. Removing the airfoil can cause turbulence and loss of containment. 5. Work surface Generally a laboratory bench top, or the floor in the case of a floor-mounted hood, this is the area where the work is conducted. 6. Exhaust plenum The exhaust plenum helps distribute airflow evenly across the hood face. Materials such as paper towels drawn into the plenum can create turbulence in this part of the hood, resulting in areas of poor airflow and uneven performance Face The imaginary plane between the bottom of the sash and the work surface. Hood face velocity is measured across this plane.
6 Important Facts about Laboratory Fume Hoods Fume hoods should be suitably sited to ensure maximum containment. Locate hoods to avoid crosscurrents at the hood face due to pedestrian traffic, doors and supply air diffusers. Sufficient laboratory supply air must be available to allow hoods to operate at their specified face velocities. A fume hood is not designed to contain high velocity releases of particulate contaminants unless the sash is fully closed. The hood is not capable of containing explosions, even when the sash is fully closed. If an explosion hazard exists, the user should provide anchored barriers, shields or enclosures of sufficient strength to deflect or contain it. Such barriers can significantly affect the airflow in the hood. 6 A conventional fume hood must not be used for perchloric acid. Perchloric acid vapors can settle on ductwork, resulting in the deposition of perchlorate crystals. Perchlorates can accumulate on surfaces and have been known to detonate on contact, causing serious injury to researchers and maintenance personnel. Specialized perchloric acid hoods, made of stainless steel and equipped with a wash down system must be used for such work. Many fume hoods are equipped with flat or rounded sills or air foils which direct the flow of air smoothly across the work surface. Sills should not be removed or modified by the hood user. Objects should never be placed on these sills. Materials released from containers placed on the sills may not be adequately captured. In addition, an object on the sill may prevent the quick and complete closure of the sash in an emergency. Tubing is frequently used to channel exhaust to the hood from equipment located some distance away. This is not an effective control method. Occasionally, a researcher may need local exhaust ventilation other than that provided by an existing fume hood. A new device may not be connected to an existing fume hood without the approval of the manufacturer. Adding devices to even the simplest exhaust system without adequate evaluation and adjustment will usually result in decreased performance of the existing hood and/or inadequate performance of the additional device. Work involving harmful microorganisms should be done in a biosafety cabinet, rather than a chemical fume hood. A fume hood should not be used for waste disposal. It is a violation of environmental regulations to intentionally send waste up the hood stack. A Guide to Laboratory Fume Hoods Important Facts About Laboratory Fume Hood
7 Guidelines for Using a Fume Hood Effectively Adequate planning and preparation is the key. The hood user should know the Standard Operating Procedure (SOP) of the hood and should design experiments so that the SOP can be maintained whenever hazardous materials might be released. Look for process changes that improve safety and reduce losses to the environment (e.g. more accurate chemical delivery systems vs. pouring volatile chemicals from bottles). Develop a process to evaluate research proposals ahead of time for potential emissions and look for opportunities to reduce them. The level of protection provided by a fume hood is affected by the manner in which the fume hood is used. No fume hood, however well designed, can provide adequate containment unless good laboratory practices are used, as follow: Ensure the exhaust is operating before beginning work. Check the baffles for obstructions. If the hood is fitted with an airflow monitor, check the monitor s status. Even while working, be alert to changes in airflow. When using the fume hood, keep your face outside the plane of the hood. Use the sash for partial protection during hazardous work. Work at least 6 inches back from the face of the hood. A stripe on the bench surface is a good reminder. Use appropriate personal protective equipment such as splash goggles and gloves. This enhances safety in case of catastrophic spills, run-away reactions or fire. Wear a full face shield if there is possibility of an explosion or eruption. When using large apparatus inside the hood, place the equipment on blocks, when safe and practical, to allow air flow beneath it. Do not make quick motions into or out of the hood, use fans, or walk quickly by the hood opening. These will cause airflow disturbances which reduces the effectiveness of the hood. Substitute with less hazardous or less volatile chemicals where possible. The hood should be kept closed, except during apparatus set-up or when working within the hood is necessary. Keep the hood sash closed as much as possible to maximize the hood s performance. Keep the sash closed when not in use to maximize energy conservation. If performance is suspected, or an airflow alarm is triggered (if installed), terminate usage, and close the sash completely. Items contaminated with odorous or hazardous materials should be removed from the hood only after decontamination or if placed in a closed outer container to avoid releasing contaminants into the laboratory air. 7
8 Risks Involved with the Improper Use of Fume Hoods A fume hood is a piece of safety equipment that can be misused to the extent that they can be less effective than expected. Injury from misuse can arise from two causes: From the fume hood not providing adequate flow rates for the work required (i.e. sash being left open or from excess clutter that reduces the containment of noxious substances). From the hood itself (i.e. if the fan belts are slipping, the exhaust duct has blockage due to paper towels being sucked into the duct, the duct damper is restricted). Always realize the most likely person to be injured is the hood user. Escaping noxious material into the laboratory can also affect all laboratory occupants. A. Power Outages In case of a power outage, the hood sash should be lowered within an inch to maintain a chimney effect to keep some air flowing into the hood. 8 B. Exhaust Care should be taken with use of paper products, aluminum foil, and other lightweight materials within a hood. For example, a single paper towel or chemical wipe can potentially decrease the airflow into the hood if it restricts exhaust flow. Good House Keeping Practices in a Laboratory using Fume Hoods Do not use the hood as a waste disposal mechanism. Apparatus used in a hood should be fitted with condensers, traps, or scrubbers to contain and collect waste solvents, toxic vapors or dust. Proper disposal of waste and code compliance should be practiced. Limit chemical storage in fume hoods. Keep the smallest amount of chemicals in the hood needed to conduct the procedure at hand. Do not use the hood as a storage area. Items can block airflow and interfere with containment. Store hazardous chemicals such as flammable liquids in an approved safety cabinet. Do not leave uncapped bottles of reagents in a hood. Although a hood is used to exhaust fumes out of the laboratory, minimizing the amount of evaporation and fuming is desirable for the environment and operator safety. Always use good housekeeping techniques to maintain the hood at optimal performance levels. Excessive storage of materials or equipment can cause eddy currents or reverse flow resulting in contaminants escaping from the hood. A Guide to Laboratory Fume Hoods Risks Involved With The Improper Use of Fume Hoods
9 Testing and maintenance of Fume hood Hoods should be evaluated by the user before each use to ensure adequate face velocities and the absence of excessive turbulence. In case of exhaust system failure while using a hood, shut off all services and accessories and lower the sash completely. Leave the area immediately. Fume hoods should be certified, at least annually, to ensure they are operating safely. Typical tests include face velocity measurements, smoke tests and tracer gas containment. Tracer gas containment tests are especially crucial, as studies have shown that face velocity is not a good predictor of fume hood leakage. In fact, these same studies have shown that approximately a third of fume hoods in use in the US, fail to meet the safety criteria specified in ANSI Z9.5. Laboratory fume hoods are one of the most important used and abused hazard control devices. We should understand that the combined use of safety glasses, protective gloves, laboratory smocks, good safety practices, and laboratory fume hoods are very important elements in protecting us from a potentially hazardous exposure. Laboratory fume hoods only protect users when they are used properly and are working correctly. A fume hood is designed to protect the user and room occupants from exposure to vapors, aerosols, toxic materials, odorous, and other harmful substances. A secondary purpose is to serve as a protective shield when working with potentially explosive or highly reactive materials. This is accomplished by lowering the hood sash. 9 Misconceptions Associated with Fume Hoods When working with highly hazardous materials, the higher the face velocity the better. While it is important to have a face velocity between 0.3 m/s (60 fpm) and 0.5 m/s (100 fpm), velocities higher than this are actually harmful. When face velocity exceeds 0.6 m/s (125 fpm) eddy currents are created which allow contaminants to be drawn out of the hood, increasing worker exposures. Check with local safety regulations on the maximum face velocity before using the hood. A chemical hood can be used for storage of volatile, flammable, or odiferous materials when an appropriate storage cabinet is not available. While it is appropriate to keep chemicals that are being used during a particular experiment inside the chemical hood, hoods are not designed for permanent chemical storage. Each item placed on the work surface interferes with the directional airflow, causing turbulence and eddy currents that allow contaminants to be drawn out of the hood. Even with highly volatile materials, as long as a container is properly capped evaporation will
10 not add significantly to worker exposures. Unlike a chemical hood, flammable materials storage cabinets provide additional protection in the event of a fire. The airfoil on the front of a hood is of minor importance. It can safely be removed if it interferes with my experimental apparatus. Airfoils are critical to efficient operation of a chemical hood. With the sash open an airfoil smoothes flow over the hood edges. Without an airfoil eddy currents form, causing contaminates to be drawn out of the hood. With the sash closed, the opening beneath the bottom airfoil provides for a source of exhaust air. Salient Feature of Esco s Fume Hoods 10 (Left to Right) Frontier Acela Fume Hood, Frontier PVC Acid Digestion Fume Hood, Frontier R a d i o i s o t o p e F u m e H o o d, Frontier Floor Mounted Fume Hood and Frontier Perchloric Fume Hood Industrial-grade support structure constructed of electro-galvanized steel and aluminum with corrosion and abrasion resistant oven-baked powder-coated finish. Internal liner is constructed of phenolic resin laminates; more corrosion-resistant and durable than other common internal liner materials. Single wall design ensures the internal work zone width is maximized. Instant-start electronically ballasted 5000k lighting system is flicker-free and the warm glow of the lighting system is comfortable to the operator s eyes during extended hours of usage. Integral fluorescent lighting is isolated from the air stream in order to increase airflow uniformity. A Guide to Laboratory Fume Hoods Salient Features of Esco s Fume Hood
11 Standard vertical rising fail-safe counterbalanced sash. The sash will tilt and lock if a cable snaps; all sashes are constructed of tempered glass for maximum operator safety. State-of-the-art baffle system constructed of phenolic resin laminates delivers maximum containment by ensuring airflow uniformity throughout the main chamber of hood. Aerodynamic design eliminates air turbulence within the hood which can otherwise compromise containment; noise levels and static pressure losses are also minimized. Aerodynamic curved side profile improves containment. Corrosion-resistant stainless steel airfoil provides spill retention and safely ventilates fumes generated towards the front of the hood for superior operator protection; the curved front edge also facilitates an ergonomic working position. Generously-sized work room due to improved positioning of the service fixtures. Black colour work surface resists staining. No exposed metal screws are present in the work zone construction, which improves corrosion resistance of the body. Electrical system is designed to meet the latest international regulations and standards for safety. Independently certified to the latest safety and performance requirements of the ANSI/ ASHRAE Standard The ANSI/ASHRAE 110 is the American standard test method for fume hoods. Esco possesses test laboratories in Asia capable of performing the full set of ASHRAE tests using an electron capture and SF6 as a tracer gas. All test results are captured via a sophisticated computer logger. Laboratory conditions can also be controlled to simulate real world operating conditions and airflow disturbances. Esco s laboratory is also equipped with instruments to perform air velocity and smoke visualisation. Esco uses the ANSI / ASHRAE 110 protocol to evaluate the performance of new fume hood designs and as a quality control measure. 11 The KI-Discus test is an alternative method for evaluating fume hood containment. The advantage of this test is that it can be readily performed in the field (unlike the ANSI / ASHRAE 110 test method which requires an expensive and usually non-portable source of SF6 gas) with a mobile test kit. This test has already gained wide acceptance in Europe, especially the UK, and was originally pioneered for use on safety cabinets.
12 Since 1978, Esco has emerged as a leader in the development of controlled environment, laboratory and cleanroom equipment solutions. Products sold in more than 100 countries include biological safety cabinets, fume hoods, ductless fume hoods, laminar flow clean benches, animal containment workstations, cytotoxic cabinets, hospital pharmacy isolators, and PCR cabinets and instrumentation. With the most extensive product line in the industry, Esco has passed more tests, in more languages, for more certifications, throughout more countries than any biosafety cabinet manufacturer in the world. Esco remains dedicated to delivering innovative solutions for the clinical, life science, research and industrial laboratory community. Airflow Alarms and Monitors Biological Safety Cabinets Exhaust Blowers Benchtop Laboratory Fume Hoods High Performance Low Flow Fume Hoods Esco Technologies, Inc Turnpike Drive, Units Hatboro, PA 19040, USA Toll-Free USA and Canada ESCO Tel Fax us.escoglobal.com [email protected] Esco Micro Pte. Ltd. 21 Changi South Street 1 Singapore Tel Fax [email protected] Esco Global Offices Kuala Lumpur, Malaysia Leiden, The Netherlands Manama, Bahrain Mumbai, India Philadelphia, USA Salisbury, UK Shanghai, China Singapore 2009 Esco Micro Pte. Ltd. Specifications Subject to Change. ES1316_V2_3K_10/09 escoglobal.com LIVE SUPPORT Esco Micro Pte Ltd Cert. No: PT Esco Bintan Indonesia Cert. No: /1
Chemical Fume Hoods. Environmental Health and Safety 974-5084
Chemical Fume Hoods Environmental Health and Safety 974-5084 Regulatory References Prudent Practices in the Laboratory 2011 (National Research Council of The National Academies): Laboratory chemical hoods
Fume Hood Questions and Answers
Fume Hood Questions and Answers 1. When should I use a fume hood? Use a chemical fume hood anytime your work involves: Chemicals with a National Fire Protection Association (NFPA) Health rating of 3 or
CelCulture. CelCulture CO2 Incubator Model CCL-170_-_. CO2 Incubator with Integrated Cooling System Solution for Highly Specialized Applications
1 CO2 Incubator Model CCL-170_-_ CO2 Incubator with Integrated Cooling System Solution for Highly Specialized Applications Welcome to Esco Esco s Vision is to provide enabling technologies for scientific
Procedures for the Effective Use of Biological Safety Cabinets
Procedures for the Effective Use of Biological Safety Cabinets APPROVAL DATE BY BIOHAZARDS SUBCOMMITTEE: August 10, 2012 SUPERSEDES: March 1, 2012 This corporate guideline/procedure is intended as a minimum
CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010
CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010 1. Background Class II Type B Biosafety Cabinets (BSCs) are unique laboratory
Laminar Flow Clean Benches, Horizontal and Vertical The Leading Solution for Research Laboratories
Esco Labculture Laminar Flow Clean Bench Model LHC-4A_ AS 1386.5, EN12469 (Microbiological Testing Method) Laminar Flow Clean Benches, Horizontal and Vertical The Leading Solution for Research Laboratories
Risk Assessment: Biosafety Cabinet
Risk Assessment: Biosafety Cabinet Biosafety cabinets (BSCs) serve as primary containment devices for operations involving potential splashes, spills or aerosolization of hazardous biological materials.
Department of Environmental Health and Safety & Emergency Management
Date: February 2014 Revision: 02 Page: 1 of 13 This Guideline is issued jointly by the Department of Environmental Health and Safety & (EHSEM) and Facilities Management to provide guidance and consistency
LABORATORY GUIDELINES AND STANDARDS
LABORATORY GUIDELINES AND STANDARDS APPLICATION NOTE LC-125 Introduction This publication provides excerpts from some of the many guidelines and standards that pertain to the construction and operation
Laminar Flow Cabinets
Laminar Flow Cabinets The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. 24 36 48 Simple, Effective Protection for Samples and Work Processes, Uncompromised Performance, Economical
Laboratory Ventilation Safety. J. Scott Ward
Laboratory Ventilation Safety J. Scott Ward In 1925, Laboratory Construction Company was born. The first product was a Kjeldahl Nitrogen Determination Apparatus. We may have shortened our name, but we
SPILLS & SPILL KITS. Spills -General Guidelines:
Spills -General Guidelines: If handled properly, a spill may be nothing more than a nuisance. If handled improperly, a spill can seriously disrupt your activities and the work of your colleagues. At worst,
In Vitro Fertilization Workstation. Esco IVF Workstation, Model IVF-4A_. Shown with optional support stand.
In Vitro Fertilization Workstation Esco IVF Workstation, Model IVF-4A_. Shown with optional support stand. Introduction The Labculture IVF workstation is specially designed for use in fertility laboratories
How To Select The Right Laboratory Hood System
How To Select The Right Laboratory Hood System An Industry Service Publication Foreword This booklet has been developed to serve as an aid in selecting a laboratory fume hood ventilation system. The information
Subpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715,
4658.4500 PLUMBING SYSTEMS; NEW CONSTRUCTION. Subpart 1. Installation. All plumbing systems must be installed and tested according to this chapter and chapter 4715, the Minnesota Plumbing Code. Subp. 2.
Class II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health
Background Class II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health This document provides guidelines for the design and installation of Class
Guide on the Segregation, Packaging, Labelling and Storage of Laboratory Chemical Wastes for Schools
APPENDIX II Guide on the Segregation, Packaging, Labelling and Storage of Laboratory Chemical Wastes for Schools (Prepared by Environmental Protection Department) 1. Introduction The Waste Disposal (Chemical
How To Test A Fume Cupboard
Jürgen Liebsch WALDNER Labor EN 14175 Requirements for Fume Cupboards DIN 12924 EN 14175 The contents of the DIN 12924 part 1 standard is divided into 5 parts in the EN 14175 standard: 14175-1 Terms and
Particularly Hazardous Substances (PHS) Standard Operating Procedure (SOP)
Particularly Hazardous Substances (PHS) Standard Operating Procedure (SOP) Formaldehyde, formalin, paraformaldehyde solutions, and paraformaldehyde solids Principal Investigator: Room & Building #: Department:
Laminar Flow Clean Benches, Horizontal and Vertical The Leading Solution for Research Laboratories
Laminar Flow Clean Benches, Horizontal and Vertical The Leading Solution for Research Laboratories Esco Airstream Laminar Flow Clean Bench Model AHC-4D_ Esco Experience Esco is the world leader in premium
Spray Booth Guideline
Department of Development Service 135 North D Street, Perris CA. 92570 Phone: (951) 443-1029 Fax: (951) 943-3293 PURPOSE The intent of this guideline is to provide the information necessary to ensure that
Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:
Autoclave Safety Purpose: Sterilization refers to the complete killing of all living organisms, including spores. Common sterilization techniques include the application of wet heat, dry heat, chemicals,
North Carolina State University Emergency Facilities Closure Checklist- Part I
North Carolina State University Emergency Facilities Closure Checklist- Part I Unplanned or spontaneous events often disrupt daily operations on campus. In the event that an incident may interrupt your
Biological Safety Program
Risk Management & Safety Main Office, Wyoming Hall Phone: (307) 766-3277 Fax: (307)766-6116 Regulated Materials Management Center Phone: (307)766-3696 Fax: (307)766-3699 Web: www.uwyo.edu/ehs Email: [email protected]
Level I Inspection Guidance Document 2014
Level I Inspection Guidance Document 2014 LEVEL I Required Weekly Checklist Note: Items listed in the Level I Required Weekly Checklist are to be inspected weekly. For Satellite Accumulation Areas, it
Best Practices for Handling Nanomaterials in Laboratories
Best Practices for Handling Nanomaterials in Laboratories Much of the toxicology of nanomaterials is currently uncertain or unknown. Some nanosized particles are more toxic than their respective micron
Fume Cupboard Guidance
Health and Safety Office Fume Cupboard Guidance Document control information Published document name: fume-cupboards-gn.pdf Date issued: Version: 1.0 Previous review dates: December 2013 Next review date:
Series 18 Pass Thru Air Lock Cleanroom Entry System
OPERATION & MAINTENANCE INSTRUCTIONS FOR Series 18 Pass Thru Air Lock Cleanroom Entry System SOLUTIONS BUILT TO YOUR SPECIFICATIONS 8605 Wyoming Ave. N. Minneapolis MN 55445 fax: 763.425.2004 e-mail: [email protected]
Biosafety Level 2 (BSL-2) Safety Guidelines
BLS-4 Biosafety Level 2 (BSL-2) Safety Guidelines BSL-3 BSL-2 BSL-1 BSL-2 builds upon BSL-1. If you work in a lab that is designated a BSL-2, the microbes used pose moderate hazards to laboratory staff
TYPICAL FIRE SAFETY INSPECTION VIOLATIONS
TYPICAL FIRE SAFETY INSPECTION VIOLATIONS The following is a list of typical violations often found by inspectors and a generic solution. You can use this list to improve the safety of your facility, to
Laboratory Hazardous Waste Disposal Guidelines
Satellite Accumulation Area (SAA) For safety and environmental reasons, hazardous waste must be stored in a designated "Satellite Accumulation Area." These areas must be inspected weekly for container
Laboratory Biosafety Level 3 Criteria
Laboratory Biosafety Level 3 Criteria Biosafety in Microbiological and Biomedical Laboratories (BMBL) 5th Edition Section IV Biosafety Level 3 (BSL-3): Biosafety Level 3 is applicable to clinical, diagnostic,
PART-II (COMMERCIAL / PRICE BID) DESIGN, SUPPLY AND INSTALLATION OF CUSTOMIZED STEEL LAB FURNITURE.AT N.C.C.S., PUNE
(An Autonomous Institution of the Department of Biotechnology, Govt. of India) N.C.C.S. Complex, Ganeshkhind, Pune University Campus, Pune 411007. PART-II (COMMERCIAL / PRICE BID) DESIGN, AND INSTALLATION
CHEMICAL SPILL MANAGEMENT GUIDELINES
WHS Unit CHEMICAL SPILL MANAGEMENT GUIDELINES Contents 1 Purpose... 2 2 Scope... 2 3 Definitions... 2 4 Roles and Responsibilities... 2 5 General... 2 6 Defining Major or Minor Spill... 3 7 Major Spill
SPACE TYPE: LABORATORY (WET) FOR CHEMISTRY AND BIOLOGY
SPACE TYPE: LABORATORY (WET) FOR CHEMISTRY AND BIOLOGY Construction Criteria The unit costs for (Wet) space types are based on the construction quality and design features in the following table. This
BIOLOGICAL SAFETY CABINET PROGRAM
BIOLOGICAL SAFETY CABINET PROGRAM TABLE OF CONTENTS: 1. INTRODUCTION 2. PURPOSE AND SCOPE 3. SELECTION AND INSTALLATION 4. USE OF BIOLOGICAL SAFETY CABINETS 5. CERTIFICATION OF BIOLOGICAL SAFETY CABINETS
Appendix J IBC Biohazard Spill Management Plan
OVERVIEW Prevention is the most important part of any spill management plan. Be sure to read and understand standard operating procedures (SOP s) and protocols for safe manipulation of biohazards before
UniMax Large Floor Mount (Walk-In) Fume Hoods
UniMax Large Floor Mount (Walk-In) Fume Hoods UniMax Fume Hoods HazMax Fume Hoods Access Door/Sash Options Plumbing/Electrical Services Exhaust/Ventilation Services Filter/Scrubber Services Laboratory
Mechanical Systems Competency 1.20
Competency 1.20 Mechanical systems personnel shall demonstrate a working level knowledge of the safety and health fundamentals of mechanical systems and/or components. 1. Supporting Knowledge and Skills
Geismar Site Reviewed: 07/05/16 Effective: 07/05/16 Supersedes: 08/04/14 Preparer: Owner: Approver: S/IH Team Leader S/IH Team Leader Manager LA Hub
BASF Corporation Title: LOCAL EXHAUST VENTILATION Individual Unit Function: Industrial Hygiene Procedure No.: IH-3 Page: 1 of 12 Geismar Site Reviewed: 07/05/16 Effective: 07/05/16 Supersedes: 08/04/14
STANDARD OPERATING PROCEDURES SPILL RESPONSE AND CLEAN-UP OUTSIDE BIOSAFETY CABINET
BIOLOGICAL SPILL KIT IN A 5 GALLON BUCKET WITH LID Spill response and cleanup procedures (SOP) 1 Notepad 1 Pen 6 Business cards 1 Permanent marker 1 trash bag 6 Biohazard stickers 1 roll duct tape 1 roll
The following standard practices, safety equipment, and facility requirements apply to BSL-1:
Section IV Laboratory Biosafety Level Criteria The essential elements of the four biosafety levels for activities involving infectious microorganisms and laboratory animals are summarized in Table 1 of
Biosafety Level 2 Criteria
Biosafety Level 2 Criteria Biosafety in Microbiological and Biomedical Laboratories (BMBL) 5th Edition Biosafety Level 2 (BSL-2): Biosafety Level 2 builds upon BSL-1. BSL-2 is suitable for work involving
THE UNIVERSITY OF NEWCASTLE- SCHOOL of BIOMEDICAL SCIENCES
Page: 1 of 7 1. Purpose: 1.1. To describe the procedures to be used when dealing with chemical or microbiological spills. 2. Equipment: 2.1. Spill Kit 2.2. Miscellaneous items as listed 3. Materials: 3.1.
GEORGIA INSTITUTE OF TECHNOLOGY ENVIRONMENTAL HEALTH AND SAFETY PERSONAL PROTECTIVE EQUIPMENT
PURPOSE To protect the health and welfare of GEORGIA TECH employees in areas where there may be a risk of injury or exposure to hazardous substances or conditions employees who work in areas where physical
PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3)
PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3) Yes No 1. Is access to the laboratory limited or restricted at the discretion of the laboratory director when experiments are in progress? Yes No
Fertilisafe TM Multi-Zone ART Workstation. Designed in Denmark
Fertilisafe TM Multi-Zone ART Workstation Designed in Denmark Fertilisafe TM Workstation Multi-zone is at the top of its class with a complete set of advanced features: Next-generation Multi-Zone (9+1)
Spetec. Laminar Flow Box. Portable Clean Room Technology
Spetec Laminar Flow Box Portable Clean Room Technology 1 Fundamentals Portable Clean Room Technology Laminar Flow The basis for the development of the clean room technology was first established in the
Notes. Material 1. Appropriate Flammable Liquids
29 CFR 1910.106 Flammable Materials Flammable Liquids Preparation 1. Read Applicable Background information and related Company Policy Chapter. 2. Make Copies of this Lesson Plan for Personnel 3. Make
WIBObarrier OCS - Open Containment Systems WIBObarrier CCS - Closed Containment Systems
WIBOjekt and WIBObarrier Air Curtain Systems. The perfect protection for person, product and environment. Filling Weighing Sampling Research Safe handling with high potent products / API WIBObarrier OCS
A SELF-EVALUATION CHECKLIST of BEST PRACTICES for AUTO REFINISH SHOPS and SCHOOLS
A of BEST PRACTICES for AUTO REFINISH SHOPS and SCHOOLS www.epa.gov/dfe/pubs/projects/auto/ WHAT IS THE DESIGN FOR THE ENVIRONMENT (DFE) AUTO REFINISH PROJECT? The DfE Program has been working with the
General Purpose Laminar Flow Cabinets, Horizontal and Vertical
24 36 48 60 72 96 The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. VLF-48 Provides reliable protection for samples and work processes for a multitude of applications. p:2 Product
Installation Instructions
Installation Instructions READ BEFORE INSTALLING UNIT For Low Profile Window Air Conditioner INSTALLATION WARNINGS AND CAUTION Carefully read the installation manual before beginning. Follow each step
Laboratory Design & Construction Guidelines
Laboratory Design & Construction Guidelines Department of Environmental Health & Safety EHS-M-35 Page 1 Issue Date: 03/01/10 General Health and Safety The USC Environmental Health and Safety Department
If you have any questions about how to safely undertake a task or project, ask your supervisor before you begin.
Laboratory Health and Safety Manual Section 1: Introduction and Objectives This manual provides a set of minimum standards and practices for the safe and healthy operation of a laboratory. Following the
Hazard/Risk Identification and Control Procedure
Hazard/Risk Identification and Control Procedure Introduction Hazard identification and the steps taken to minimize the risks associated with identified hazards are a critical component of working safely.
Principles of Ventilation. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this
FIRE DAMPER APPLICATION GUIDE
FIRE AND CEILING RADIATION DAMPERS FIRE DAMPER APPLICATION GUIDE Use this application guide to help determine what type of fire damper best suits the application. Below are listed the ten Metal-Fab fire
FLAMMABLE LIQUIDS COMPLIANCE NOTE
Issued by: Heather Rice Page: 1 of 7 COMPLIANCE NOTE INTRODUCTION This document provides additional guidance to support the Arrangement on Flammable Liquids (Ref No. CYC/HS/A27) CONTENTS Section Topic
plenum slot diffusers
light powered Redefine your comfort zone. www.titus-hvac.com dual-function smart logic light powered energy solutions Table of Contents PLENUM SLOT IFFUSERS products Plenum Slot iffusers Products... 4
Emergency Showers and Eyewashes Mike Adkins Wednesday, October 31, 10:00-11:1511:15 a.m.
Emergency Showers and Eyewashes Mike Adkins Wednesday, October 31, 10:00-11:1511:15 a.m. Safety First! How to Specify Emergency Showers and Eyewashes Presented by Mike Adkins, Guardian Equipment. ASPE
Emergency Showers and Eyewash Stations
Emergency Showers and Eyewash Stations Here is a guide regarding what types and specifications are required when selecting an emergency Shower and/or eyewash station. Why are emergency showers or eyewash
Lab PPE Guide Examples of Personal Protective Equipment
Lab PPE Guide Examples of Personal Protective Equipment 1140 South Paulina St. 245 (PSB) Chicago IL 60612 Phone: 312-413-3387 Fax: 312-413-3700 www.uic.edu/depts/envh This document is to be used as a supplement
Laboratory Work Risk Assessment
1. Introduction Laboratory Work Risk Assessment The legislation governing Hazardous Substances and Dangerous Goods requires that all operations and experiments involving hazardous materials must have documentation
CEILING SYSTEMS. CI/SfB (35) Xy April 2006. [Between us, ideas become reality.] ULTIMA CANOPY. Data sheet. Acoustic performance meets High Design
CEILING SYSTEMS [Between us, ideas become reality.] CI/SfB (35) Xy April 2006 ULTIMA CANOPY Data sheet Acoustic performance meets High Design ULTIMA CANOPY Installation Instructions I. GENERAL 1. Product
SPECIAL APPLICATIONS
Page 1 SPECIAL APPLICATIONS The protection of microelectronics clean rooms using aspirating smoke detection systems requires special consideration. The physical configuration of the room(s) and the direction
Walking and Working Surface Checklist:
Walking and Working Surfaces Purpose: This information is presented to improve the level of safety in our operations, and to inform you that we will comply with the OSHA Standard 29 CFR 1910.23. Responsibility:
Material Safety Data Sheet Ampicillin, sodium salt MSDS
Material Safety Data Sheet Ampicillin, sodium salt MSDS Section 1: Chemical Product and Company Identification Product Name: Ampicillin, sodium salt Catalog Codes: SLA1666 CAS#: 69-52-3 RTECS: XH8400000
Element D Services Heating, Ventilating, and Air Conditioning
PART 1 - GENERAL 1.01 OVERVIEW A. This section supplements Design Guideline Element D3041 on air handling distribution with specific criteria for projects involving design of a Data Center spaces B. Refer
DNA Analyst Training Laboratory Training Manual Protocol 2.02 Clean Technique
DNA Analyst Training Laboratory Training Manual Protocol 2.02 Clean Technique This laboratory protocol (or part thereof) has been provided as an example of a laboratory SOP, courtesy of the Illinois State
Laboratory Safety Inspection Checklist (Biology and Biochemistry)
EOSMS 201-1A Date: 11/15/14 Page 1 of 9 Building Room Inspection Date Department Principal Investigator(s) Inspector(s) Instructions: This form is to be completed for laboratory self-inspection. Complete
LABORATORY SAFETY INSPECTION CHECKLIST
LABORATORY SAFETY INSPECTION CHECKLIST As part of EH&S's laboratory safety survey program, we have compiled a list of the most common safety problems. This is not a list of every possible hazard, but can
COMMERCIAL COOKING HOODS, VENTILATION & FIRE SUPPRESSION SYSTEM GUIDELINES AND PROCEDURES
COMMERCIAL COOKING HOODS, VENTILATION & FIRE SUPPRESSION SYSTEM GUIDELINES AND PROCEDURES These guidelines are to be used for ALL commercial cooking hoods, ventilation systems and related fire suppression
Laminar Flow Cabinets and Workstations
VERTICAL AND HORIZONTAL HEPA FILTERED SYSTEMS FOR PROCESS OR OPERATOR PROTECTION Laminar Flow Cabinets and Workstations For procedures requiring protection from particulate contamination and for aseptic
COMPREHESIVE LAB SAFETY CHECKLIST
COMPREHESIVE LAB SAFETY CHECKLIST The SAFETY & ENVIRONMENTAL COMPLIANCE DEPARTMENT has developed this form to assist and inform USA faculty and staff in identifying and correcting potential laboratory
F-900 HOPEC Bench Hood Framed Vertical Sash
F-900 Series HOPEC Bench Hoods Product Features F-900 HOPEC Bench Hood Framed Vertical Sash The ISOLATOR HOPEC (Hand Operated Positive Energy Control) bench fume hood is based on the design of Jerry Koenigsberg*.
STANDARD OPERATING PROCEDURES Safe Autoclave Operations
Department of Environmental Health & Safety Biosafety & Biosecurity STANDARD OPERATING PROCEDURES Safe Autoclave Operations The purpose of this document is to provide standard operating procedures for
Grey Nuns Community Hospital. Chemical Spill / Hazardous Material
Grey Nuns Community Hospital Chemical Spill / Hazardous Material Introduction This plan addresses the UgeneralU procedures to be followed for a spill involving a hazardous substance. Some departments will
Job Hazard/Safety Analysis For the Lab Worker
Job Hazard/Safety Analysis For the Lab Worker 1 INTRODUCTION Job Hazard/Safety Analysis for the Lab Worker is a partial list of known task performed by the professional laboratory worker. We have placed
Cleanroom Solutions Ltd T: 01273 760 060 F: 01273 760 061 E: [email protected] W: www.cleanroom-solutions.co.uk
Cleanroom Solutions Ltd T: 01273 760 060 F: 01273 760 061 E: [email protected] W: Introduction Cleanroom Solutions Ltd has a wealth of experience together with a solid track record in the
Mechanical Rooftop Custom Capabilities
Mechanical Rooftop Custom Capabilities Mechanical Construction Capabilities Price manufactures a wide range of custom gas heat and DX cooled units to meet your new construction and retrofit needs. In addition
Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested
DOP Solutions Technical Document: Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested Issue 1: Prepared by Neil Stephenson and Tim Triggs and edited by
Hazardous materials can be silent killers. Almost every household and workplace has varying amounts of chemicals that, if spilled or combined, will
HAZARDOUS MATERIALS Hazardous materials can be silent killers. Almost every household and workplace has varying amounts of chemicals that, if spilled or combined, will cause great harm and even death.
Laboratory Safety and Environmental Health Assessment Program
Laboratory Safety and Environmental Health Assessment Program Principal Investigators and Faculty are vital to the success of safety and environmental compliance programs in Colorado School of Mines research
Cooktop Low-Profile Ventilation Hoods
INSTALLATION GUIDE Cooktop Low-Profile Ventilation Hoods Contents Wolf Cooktop Low-Profile Ventilation Hoods........ 3 Cooktop Low-Profile Hood Specifications.......... 4 Cooktop Low-Profile Hood Installation............
NFPA Combustible Dust Compliance and the Industrial Ventilation System
NFPA Combustible Dust Compliance and the Industrial Ventilation System Presented By: Tom Kroeger PE [email protected] Kirk & Blum, a CECO Environmental Company Marty Schloss PE [email protected]
Stainless Steel Worksurface 3 Feet Hood A10-WS-PR3 A10-WS-SR3 4 Feet Hood A10-WS-PR4 A10-WS-SR4 6 Feet Hood A10-WS-PR6 A10-WS-SR6.
Product Sampler #1 Laminar Flow Hoods / Laminar Flow Cabinets: 1- Benchtop Vertical Laminar Flow Hoods- Air Recirculation Available s: Clean Bench Class 100 Bench-top Vertical Laminar Flow Hoods with Air-recirculation
Radford University. Indoor Air Quality Management Plan
Radford University Indoor Air Quality Management Plan Spring 2015 Introduction Concerns with Indoor Air Quality (IAQ) have increased since energy conservation measures were instituted in office buildings
Carnegie Mellon University s Policy and Procedures for Recombinant and Synthetic Nucleic Acid Materials Spills
Carnegie Mellon University s Policy and Procedures for Recombinant and Synthetic Nucleic Acid Materials Spills Background In accordance with Section IV-B-2-b-(6) of the NIH Guidelines for Research Involving
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
IX. EMERGENCY PLANNING
IX. EMERGENCY PLANNING EMERGENCY PLANNING, SPILL RESPONSE & REPORTING Any unexpected event or accident situation that occurs during the transport, storage or use of hazardous materials constitutes a potential
Standard Operating Procedure for a Chemical Spill
Standard Operating Procedure for a Chemical Spill Facility: Lab Director: Scope: ERC Clean Room C17 Engineering Research Complex Electrical and Computer Engineering Brian Wright 3234 Engineering 355-5233
How To Install A Perfect 16 Id Series Filter System
IQAir Perfect 16 ID-Series MERV16 Micro-Particle Filtration System Perfect 16 Installation Manual Product Data High-Performance Medical-Grade Residential/Commercial Air Filtration System Application The
Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.
Requirements for Decontamination by Autoclaving Summary: Steam sterilization (autoclaving) is one of the most effective methods for decontaminating biohazardous material. Autoclaves use saturated steam
Material Safety Data Sheet
Date of Issue: 26/5/2009 MSDS No.: Q54-1300B Material Safety Data Sheet 1. PRODUCT AND COMPANY IDENTIFICATION Product Name : Daily Maintenance Kit(Wipe Cleaning) Product Code : IP6-108 -Application model
Recommendations for the Safe Use of Handling of Cytotoxic Drugs
Recommendations for the Safe Use of Handling of Cytotoxic Drugs Introduction Cytotoxic drugs are toxic compounds and are known to have carcinogenic, mutagenic and/or teratogenic potential. With direct
Conservation of Momentum Greg Kifer
SCIENCE EXPERIMENTS ON FILE Revised Edition 6.7-1 Conservation of Momentum Greg Kifer Topic Conservation of momentum Time 1 hour! Safety Please click on the safety icon to view the safety precautions.
LAMINAR FLOW FOR ALL YOUR NEEDS
LAMINAR FLOW FOR ALL YOUR NEEDS 2 LAMINAR FLOW FOR ALL YOUR NEEDS EXPLORE TECNIPLAST COMPLETE RANGE OF LAMINAR FLOW SOLUTIONS: OUR CUTTING-EDGE TECNOLOGY IS AT YOUR SERVICE, TO ASSURE TOP ENVIRONMENTAL,
Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection
Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection A wide range of products to meet the needs of transferring animals, operators,
FlowFAST V. Vertical Laminar Airflow Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.
Vertical Laminar Airflow Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE. Vertical Laminar Flow Cabinets BEYOND MINIMUM SAFETY REQUIREMENT Cabinets are ISO Class 3 vertical laminar flow cabinets which
