Metallurgical analysis of five failed cast cobalt-chromium-molybdenum alloy hip prostheses

Size: px
Start display at page:

Download "Metallurgical analysis of five failed cast cobalt-chromium-molybdenum alloy hip prostheses"

Transcription

1 Veterans Administration Journal of Rehabilitation Research and Development Vol. 23 No. 4 Pages Metallurgical analysis of five failed cast cobalt-chromium-molybdenum alloy hip prostheses STEPHEN D. COOK, Ph.D.; MARCUS A. KESTER, Ph.D.; AMANDA F. HARDING, B.S. ; T. DESMOND BROWN, M.D.; PATRICIA M. SANDBORN, B.S. Rehabilitation Research and Development, Veterans Administration Medical Center, New Orleans, Louisiana and Department of Orthopaedic Surgery, Tulane University School of Medicine, New Orleans, Louisiana Abstract The clinical and metallurgical characteristics of five cast cobalt-chromium-molybdenum alloy femoral hip prostheses which failed in vivo were evaluated. The devices included: two of the Howmedica Muller-Charnley design, two of the Howmedica Charnley design, and one of the Zimmer Aufranc-Turner design. Fractographic analyses demonstrated that the five devices had failed by fatigue which originated on the lateral aspect. Failure occurred after an average in vivo time of 80.4 months (approximately 6.7 years). Only two of the devices had Rockwell hardness values that were within the ASTM specifications for the alloy. Upon metallurgical examination, moderate to severe levels of gas porosity, interdendritic shrinkage, and nonmetallic inclusions were found in all of the devices. As expected, extremely large grain sizes also were observed in the devices examined. These results indicate that the metallurgical flaws and defects associated with the cast cobalt-chromium-molybdenum alloys used in these devices may preclude successful longterm performance and warrant manufacturer's attention. INTRODUCTION Total hip arthroplasty has proved to be a successful treatment modality for the restoration of daily activities and relief of pain in the disabled joints of the elderly. However, because of continuing problems, the procedure has not enjoyed the same degree of success in young or middle-aged patients. Device loosening and pain are the principle reasons for revision surgery. A late problem with total hip prostheses is fatigue failure of the femoral component. Charnley has reported a fracture incidence Send reprint requests to : Stephen D. Cook, Ph.D.; Department of Orthopaedic Surgery ; Tulane University School of Medicine ; 1430 Tulane Avenue; New Orleans, Louisiana of 0.23 percent in his initial 6500 hip replacement procedures (1). Martens et al reported an incidence of 11 percent in a smaller group of 56 patients who had received Charnley-Muller devices of an earlier design (9). Iatrogenic factors, fixation methods, implant design, and metallurgic characteristics have all been implicated in causing fatigue failures in the femoral stems (1 9, 1 I ). Failure to obtain or maintain proper valgus positioning of the device is a primary factor in device mechanical failure. Varus positioning or migration can lead to increased stress on the lateral aspect of the device (2,3,5,7,9). Likewise, failure to remove enough of the calcar trabecular bone may lead to resorption and a loss of calcar support (1 3,5 8,11).-If the device is properly seated distally, this situation can generate a cantilever effect on the device during daily activities (5,7,8). Such an alteration of stresses can produce catastrophic device failure through the fatigue mechanism. Design flaws, which can lead to mechanical fatigue, are uncommon. Most manufacturers have abandoned such features as sharp corners and mechanically etched serial numbers on the devices. The use of better metal alloys has also significantly reduced the incidence of crack initiation. However, the use of cast cobalt-chromium-molybdenum in early hip designs has led to a number of problems. These problems arose from metallurgical flaws such as gas porosity, nonmetallic inclusion content, grain size variability and interdendritic shrinkage (2 5, 9,11,12). Such characteristics predispose the implant to crack formation and propagation. Normal daily activities such as gait cause fluctuating stresses on the implant (5). These fluctuations can produce small cracks that originate 27

2 28 Journal of Rehabilitation Research and Development Vol. 23 No. 4 October 1986 principally along the lateral and anterior borders of the device which experience the highest tensile stresses (5,7,8,11). Once the crack is formed, loads producing mean tensile stresses significantly less than the material's nominal strength can lead to fatigue failure, due to the stress concentration at the leading edge of the crack. As part of the ongoing Orthopaedic Implant Retrieval and Analysis Program sponsored by the Veterans Administration, all total joint prostheses removed from patients at Tulane-affiliated hospitals were evaluated for in vivo performance. It was the purpose of this study to review the clinical histories of five patients with fractured cast cobalt-chromium-molybdenum alloy femoral hip prostheses and correlate findings with the metallurgical properties and failure characteristics of the devices. The following is a typical case of a device failure : the patient, a 63-year-old white male, had a Muller-Charnley (32 mm head) type total hip arthroplasty inserted in 1972 for severe left hip pain due to avascular necrosis (Figure la). He had a history of chronic alcohol abuse. Following surgery, the patient had good relief of pain and returned to office work until 1982, when he began to have some left hip discomfort. The pain became more severe after December 1984 (Figure lb) and by August 1985, the patient was only able to ambulate one-block distances with the aid of a cane. At that time, he had a painful range of motion of 0-90 Deg. flexion, 30 Deg. external rotation, 0 Deg. internal rotation, and 15 Deg. abduction. He also had a slight leg-length discrepancy of 1 cm and slight quadriceps atrophy. He had no flexion contracture and no drainage. MATERIAL AND METHODS Upon surgical removal, the five fractured femoral hip prostheses were forwarded to the Biomaterials Laboratory for mechanical failure and metallurgical evaluations. The hip components were removed between the years 1976 and 1985 from patients at Tulane-affiliated hospitals. All femoral prosthetic components were fabricated from cast cobalt-chromium-molybdenum alloy. After removal, the devices were ultrasonically cleaned in a hydrogen peroxide bath, brushed with a mild detergent, and photographed. Proper care was taken to insure that the fracture surfaces of the broken femoral stems remained unharmed in order to facilitate determination of the mode of device failure. The fracture surfaces were viewed using light and scanning electron microscopes. Using an abrasion wheel cutoff saw, a transverse section was obtained from each component just below the distal fracture site. This section was mounted in epoxy and polished to a mirror image for metallurgical analysis. Using ASTM Standard techniques, each specimen was evaluated for Rockwell hardness, inclusion content, porosity, and grain size. Gas porosity, inclusion content, and interdendritic shrinkage were evaluated on a grading scale where 0 represented no significant levels, 1 represented moderate levels, and 2 represented severe levels. The controls incorporated in the grading scheme were similar cast cobalt-chromiummolybdenum prostheses, but were of current designs and materials. Clinical histories also were obtained from each patient's chart. Parameters evaluated included patient age at implant insertion, insertion diagnosis, implant time in situ, and reason for implant removal. Figure la. Postoperative radiograph of a Muller-Charnley total hip arthroplasty (prosthesis #1). 4

3 COOK ET AL: Metallurgical analysis of failed hip prostheses 29 Figure lb. Follow-up radiograph of Muller-Charnley hip, December Figure Id. Radiograph of Muller-Charnley arthroplasty, May Figure lc. Radiograph of Muller-Charnley arthroplasty showing the fractured femoral stem component, August Radiographic review in August 1985 showed a transverse fracture of the femoral component at the junction of the distal and middle third of the stem (Figure lc). There was solid fixation of the stem tip in cement, yet there was visable radiolucency around the proximal stem and resorption of bone in the calcar region. The acetabular component showed evidence of proximal migration, early protrusio and radiolucency at the cement-bone interface. Radiographs taken in May 1985 (Figure 1d) revealed evidence of the femoral stem fracture, although this remained undiagnosed until August An earlier film from December 1984 did not show the stem fracture or acetabular loosening. In August 1985, the patient was taken to surgery for removal of the loose acetabular and femoral components. The fractured piece of the distal femoral stem was firmly fixed, necessitating a window cut in the bone for removal. The prosthesis was replaced with a non-cemented Dupuy type AML bipolar endoprosthesis (Figure le).

4 30 Journal of Rehabilitation Research and Development Vol. 23 No. 4 October 1986 RESULTS Femoral prostheses were removed from five patients (four males and one female) with an average age at insertion of 54.7 years (range: years). The cases included three right hip arthroplasties and two left hip arthroplasties. Insertion diagnoses included avascular necrosis (three cases), joint dislocation (one case) and Paget's disease (one case). All prostheses remained in situ for an average of 80.4 months (range: months) and all were removed due to fracture of the femoral component with associated loosening. The clinical data are presented in Table 1. All devices had metallurgical defects which could have contributed to their mechanical failure (Table 2). With the grading scheme employed, three (60 percent) of the devices had moderate levels of gas porosity, while the two (40 percent) remaining devices had severe levels (Figure 2). Three (60 percent) devices also exhibited moderate levels of nonmetallic inclusions, while two (40 percent) had severe levels (Figure 3). Interdendritic shrinkage voids were present in all prostheses examined and were judged to be severe in three (60 percent) devices and moderate in two (40 percent) devices (Figure 4). ASTM guidelines suggest that cast cobalt-chromiummolybdenum alloy for implantation should have Rockwell "C" scale hardness values between 25 and 35 (11). Figure le. (above) Radiograph at revision surgery showing the AML bipolar endoprosthesis, August Figure 2. (below) Photomicrograph showing severe levels of gas porosity in a fractured femoral stem component (prosthesis #1).

5 31 COOK ET AL: Metallurgical analysis of failed hip prostheses Figure 3. Photomicrograph showing severe levels of nonmetallic inclusions in a fractured femoral stem component (prosthesis #3). Table 1 Clinical Data for Patients with Fractured Hip Prostheses Prosthesis Number Sex Age (years) Mfg./ Head Diameter Insertion Diagnosis Removal Reason Months in situ Limb M 50 H* (32mm) Avas.Necr. Breakage 158 L 2 M 50 H* (32mm) Avas.Necr. Breakage 43 R 3 F 67 H** (22mm) Dislocation Breakage 108 L 4 M H** (22mm) Avas.Necr. Breakage 56 R 5 M 52 Z*** (32mm) Pagets Dis. Breakage 37 R H = Howmedica Z = Zimmer * Muller-Charnley : Model # ** Charnley : Model # *** Aufranc-Turner : Model # Table 2 Metallurgical Parameters for Fractured Hip Prostheses Prosthesis Number Level of Fracture* 1 Distal 1 / 3 Vlfile 1) 3 3 Proximal 1/3 4 Proximal I / 3 5 Middle 1/3 Rockwell C Hardness** 15.0 (1.8) 22.0 (2.6) 20.0 (2.2) 26.0 (1.5) 28.0 (4.7) Gas Porosity*** Inclusions*** Interdendritic Shrinkage*** * See Figure 5 ** Mean (Standard Deviation) ASTM guidelines suggest Rockwell *** 0 = None I = Moderate 2 = Severe hardness in the range of for RC values.

6 32 Journal of Rehabilitation Research and Development Vol. 23 No. 4 October 1986 Figure 4. Photomicrograph showing severe levels of interdendritic shrinkage voids in a fractured femoral stem component (prosthesis #1). Figure 5. Photograph showing the location of fracture sites along the femoral stems : A. Howmedica Charnley device, B. Howmedica Muller-Charnley device, C. Aufranc-Turner device.

7 33 COOK ET AL: Metallurgical analysis of failed hip prostheses Three of the five (60 percent) femoral components had mean hardness values less than 25 (Table 2). There were no consistent findings regarding the variation of hardness values across the sections, although in three (60 percent) devices, the hardness values increased in the center of the device. The fractures of the femoral stems occurred in a variety of locations (Figure 5) (Table 2). Two (40 percent) of the devices broke in the area of the proximal third of the stem. The central third of the stem was the location of two (40 percent) device failures, while the remaining device fractured in the distal third of the stem. Both of the devices that fractured in the proximal third were the Howmedica Charnley total hip (Model ). Additionally, one device (prosthesis #2) had a secondary crack forming slightly distal (0.5 cm) to the level of the fracture site (Figure 6). Due to in vivo fretting abrasion subsequent to fracture, it was extremely difficult to determine the precise origination site of the fatigue. The fracture surfaces normally showed a somewhat smooth or slow fatigue which encompassed the lateral one-third to one-half of the transverse cross-sectional area (Figure 7). This portion of the fracture surface had stress striations typical of fatigue failure (Figure 8). The remaining portion of the fracture surface had a rough appearance, which is typical of a brittle or fast fracture (Figures 7,9). Fast fracture ultimately occurs when the effective cross-sectional area, already reduced due to slow fatigue, can no longer support the demands placed upon it. DISCUSSION The results of this study concerning the metallurgical defects present in many cast cobalt-chromiummolybdenum prostheses are in agreement with the findings of other researchers (34,9 11). These flaws Figure 6. Scanning electron microscope photograph of a secondary fatigue crack formation distal to the fracture surface (prosthesis #2).

8 34 Journal of Rehabilitation Research and Development Vol. 23 No. 4 October 1986 Figure 7. Fractograph of the surface of a failed femoral stem component (prosthesis #2). Note the smooth appearance of the lateral (L) aspect and the rough appearance of the medial aspect (M). include gas porosity, nonmetallic inclusions, and interdendritic shrinkage, which can all function as both crack initiators and intensifiers in the crack propagation process. The lateral bias of the fatigue initiation site is also in accord with other researchers (3,8). Gas pores or a string of pores as are present in interdendritic shrinkage are stress raisers and their presence near a surface warrants concern (11). Further, the frequency of nonmetallic inclusions has been demonstrated to have a detrimental effect on the endurance limit of cast materials ( 11). The effect of porosity on the hardness of a device can he appreciated by noting that device #1 (Figure 2, Table 2) had the greatest porosity and the lowest Rockwell hardness values of the five hips examined. Manufacturing processes can enhance the chances of prosthetic survival by including steps to reduce the chance of microstructural flaws being present. Such steps include hot isostatic pressing, forging, and remelting processes (3 5,11), as well as the use of other alloys such as titanium. These metallurgic flaws are particularly important when considering the fluctuating stresses that occur during the working life of a hip prosthesis. The extreme proximal location of the fracture of the two femoral components of identical models illustrates the importance of design considerations in the performance of prosthetic devices (Table 2) (1,2,8 10). Device features such as sharp corners, tapers, small crosssectional areas, and overly curved stems should be avoided. As shown in Figure 9, the stem on the Charnley (22 mm head) device is curved in such a manner that large bending stresses are generated at the shoulder of the device (1). The area of fatigue failure was in this shoulder region in both hip stems of this type. Prosthetic parameters affecting mechanical and clinical failure rates include: iatrogenic factors such as cement techniques, device placement and selection, patient selection and education, as well as the surgical handling of the device. Proper calcar support of the device is extremely critical in obviating device fatigue. Although the proximal end of the device migrates medially with the

9 35 COOK ET AL : Metallurgical analysis of failed hip prostheses Figure S. Scanning electron microscope photograph showing slow fatigue striations on the lateral 1/3 to 1/2 of the transverse cross-sectional area (prosthesis #2). Figure 9. Scanning electron microscope photograph showing fast fracture striations on the fracture surface of a femoral stem (prosthesis #3).

10 36 Journal of Rehabilitation Research and Development Vol. 23 No. 4 October 1986 loss of calcar support, the distal end remains fixed in the cement. Once the proximal load transfer is lost, the system resembles a bending cantilever and the device must resist enhanced cyclic bending stresses (7). Factors such as patient weight (3,5,6,8) and high levels of activities (5) can also increase these bending stresses and have been demonstrated to have a positive correlation with fatigue failure. It is hoped that with the development of the newer alloys such as titanium and the use of new manufacturing techniques, the frequency of mechanical failures can be reduced. Education regarding the data generated through implant retrieval and analysis, along with these new alloys and techniques, should result in an enhanced survival rate for current hip devices. REFERENCES 1. CHARNLEY J. Fracture of femoral prostheses in total hip replacement. Clin Orthop Rel Res 106: , cemented stem-type femoral components. 141 :17-27, Clin Orthop Rel Res 2. CHAO E AND COVENTRY M. Fracture of the femoral component after total hip replacement. J Bone Joint Surg 63A(7): , COLLIS D. Femoral stem failure in total hip replacement. J Bone Joint Surg 59A(8) : , DUCHEYNE P, DEMEESTER P AND AERNOULDT E. Fatigue fractures of the femoral component of Charnley and Charnley- Muller type total hip prostheses. J Biomech Mater Res 6: , GALANTE J. Causes of fractures of the femoral component in total hip replacement. J Bone Joint Surg 62A(4): , GALANTE J, ROSTOKER W AND DOYLE J. Failed femoral stems in total hip prosthesis. J Bone Joint Surg 57A(2) : , GRUEN T, MCNEICE G AND AMSTUTZ H. `Modes of failure' of 8. MARKOLF K AND AMSTUTZ H. A comparative experimental study of stresses in femoral total hip replacement components: The effects of prosthesis orientation and acrylic fixation. J Biomech 9:73-79, MARTENS M, AERNOULDT E, DEMEESTER P, ET AL. Factors in the mechanical failure of the femoral component in total hip prosthesis. Acta Orthop Scand 45 : , REUBEN J, EISMUNT F, BURSTEIN A AND WRIGHT T. Comparative mechanical properties of forty-five total hip stems. Clin Orthop Rel Res 141:55-65, ROSTOKER W, CHAO E AND GALANTE J. Defects in failed stems of hip prostheses. J Biomech Mater Res 12: , Standard specification for cast cobalt-chromium-molybdenum alloy for surgical implantation applications. ASTM Designation: F American National Standards Institute, pp 13-14, 1985.

Clinical performance of endoprosthetic and total hip replacement systems

Clinical performance of endoprosthetic and total hip replacement systems Veterans Administration Journal of Rehabilitation Research and Development Vol. 24 No. 3 Pages 49 56 Clinical performance of endoprosthetic and total hip replacement systems P. M. SANDBORN, B.S. ; S. D.

More information

Why an Exactech Hip is Right for You

Why an Exactech Hip is Right for You Why an Exactech Hip is Right for You Why do I need a total hip replacement? Which surgical approach is best for me? How long will it last? Which implant is right for me? Founded in 1985 by an orthopaedic

More information

BONE PRESERVATION STEM

BONE PRESERVATION STEM TRI-LOCK BONE PRESERVATION STEM Featuring GRIPTION Technology SURGICAL TECHNIQUE IMPLANT GEOMETRY Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was

More information

Metal-on-Metal Hip Systems

Metal-on-Metal Hip Systems Meeting of the Orthopaedic and Rehabilitation Devices Advisory Panel June 27-28, 2012 Metal-on-Metal Hip Systems Center for Devices and Radiological Health U.S. Food and Drug Administration Meeting of

More information

on-metal Hips: Device Mechanics and Failure Modes

on-metal Hips: Device Mechanics and Failure Modes 1 Metal-on on-metal Hips: Device Mechanics and Failure Modes Steven M. Kurtz, Ph.D., and Richard Underwood, Ph.D. Exponent, Inc., and Drexel University NIH R01 AR47904 2 Contracts: DePuy Orthopaedics,

More information

Total Hip Replacement

Total Hip Replacement Please contactmethroughthegoldcoasthospitaswityouhaveanyproblemsafteryoursurgery. Dr. Benjamin Hewitt Orthopaedic Surgeon Total Hip Replacement The hip joint is a ball and socket joint that connects the

More information

Introduction to the Bertram Hip Spacer

Introduction to the Bertram Hip Spacer Introduction to the Bertram Hip Spacer Approximately 200,000 hip replacement surgeries are done every year in United States. Fortunately the incidence of infection is routinely 0.5 percent-1 percent. However

More information

The Right Choice. Exeter. total hip system

The Right Choice. Exeter. total hip system The Right Choice Exeter total hip system Exeter The Right Choice Anatomic Reconstruction Offset The objectives of total hip replacement are to: relieve pain increase mobility and function Achieving a correct

More information

frequently asked questions Knee and Hip Joint Replacement Technology

frequently asked questions Knee and Hip Joint Replacement Technology frequently asked questions Knee and Hip Joint Replacement Technology frequently asked questions Knee and Hip Joint Replacement Technology Recently, you may have seen advertisements from legal companies

More information

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION Richard A. Sweet, M.D. Louisville Orthopaedic Clinic Louisville, KY TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION INTRODUCTION Total hip replacement surgery (THR) has

More information

Patient Labeling Information System Description

Patient Labeling Information System Description Patient Labeling Information System Description The Trident Ceramic Acetabular System is an artificial hip replacement device that features a new, state-of-the-art ceramic-on-ceramic bearing couple. The

More information

Exeter. Surgical Technique. V40 Stem Cement-in-Cement. Orthopaedics

Exeter. Surgical Technique. V40 Stem Cement-in-Cement. Orthopaedics Exeter Orthopaedics V40 Stem Cement-in-Cement Surgical Technique Exeter V40 Stem Cement-in-Cement Surgical Technique Table of Contents Indications and Contraindications...2 Warnings and Precautions...2

More information

Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology

Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology Hips designed to fit the unique anatomies of men and women Independent control for a natural fit Simple, practical solutions for optimal restoration

More information

Zimmer M/L Taper Hip Prosthesis. Surgical Technique

Zimmer M/L Taper Hip Prosthesis. Surgical Technique Zimmer M/L Taper Hip Prosthesis Surgical Technique Zimmer M/L Taper Hip Prosthesis 1 Zimmer M/L Taper Hip Prosthesis Surgical Technique Table of Contents Preoperative Planning 2 Determination of Leg Length

More information

Minimally Invasive Hip Replacement through the Direct Lateral Approach

Minimally Invasive Hip Replacement through the Direct Lateral Approach Surgical Technique INNOVATIONS IN MINIMALLY INVASIVE JOINT SURGERY Minimally Invasive Hip Replacement through the Direct Lateral Approach *smith&nephew Introduction Prosthetic replacement of the hip joint

More information

PUBLISHED VERSION PERMISSIONS. This document has been archived with permission from the editor of the Medical Journal of Australia, 26 April 2007.

PUBLISHED VERSION PERMISSIONS. This document has been archived with permission from the editor of the Medical Journal of Australia, 26 April 2007. PUBLISHED VERSION Graves, Stephen Ellis; Davidson, David Charles; Ingerson, Lisa; Ryan, Philip; Griffith, Elizabeth Catherine; McDermott, Brian Francis John; Pratt, Nicole Leanne The Australian Orthopaedic

More information

Y O U R S U R G E O N S. choice of. implants F O R Y O U R S U R G E R Y

Y O U R S U R G E O N S. choice of. implants F O R Y O U R S U R G E R Y Y O U R S U R G E O N S choice of implants F O R Y O U R S U R G E R Y Y O U R S U R G E O N S choice of implants F O R Y O U R S U R G E R Y Your Surgeon Has Chosen the C 2 a-taper Acetabular System The

More information

FEMORAL NECK FRACTURE FOLLOWING TOTAL KNEE REPLACEMENT

FEMORAL NECK FRACTURE FOLLOWING TOTAL KNEE REPLACEMENT 1 FEMORAL NECK FRACTURE FOLLOWING TOTAL KNEE REPLACEMENT László Sólyom ( ), András Vajda & József Lakatos Orthopaedic Department, Semmelweis University, Medical Faculty, Budapest, Hungary Correspondence:

More information

Hip Resurfacing 2011 ORIGINAL ARTICLE. James W. Pritchett MD. Introduction. Abstract

Hip Resurfacing 2011 ORIGINAL ARTICLE. James W. Pritchett MD. Introduction. Abstract 1 ORIGINAL ARTICLE Hip Resurfacing 2011 James W. Pritchett MD Abstract In 1938 Marion Smith-Peterson placed a cobalt chromium cup on a reshaped femoral head to perform the first hip resurfacing. 15 Also,

More information

FAILURE ANALYSIS OF A FEMORAL HIP STEM MADE OF STAINLESS STEEL AFTER SHORT TIME OF EXPOSURE

FAILURE ANALYSIS OF A FEMORAL HIP STEM MADE OF STAINLESS STEEL AFTER SHORT TIME OF EXPOSURE FAILURE ANALYSIS OF A FEMORAL HIP STEM MADE OF STAINLESS STEEL AFTER SHORT TIME OF EXPOSURE Żaneta Anna MIERZEJEWSKA *, Zbigniew OKSIUTA * * Faculty of Mechanical Engineering, Division of Materials and

More information

Total Hip Joint Replacement. A Patient s Guide

Total Hip Joint Replacement. A Patient s Guide Total Hip Joint Replacement A Patient s Guide Don t Let Hip Pain Slow You Down What is a Hip Joint? Your joints are involved in almost every activity you do. Simple movements such as walking, bending,

More information

Rehabilitation Protocol: Total Hip Arthroplasty (THA)

Rehabilitation Protocol: Total Hip Arthroplasty (THA) Rehabilitation Protocol: Total Hip Arthroplasty (THA) Department of Orthopaedic Surgery Lahey Hospital & Medical Center, Burlington 781-744-8650 Lahey Outpatient Center, Lexington 781-372-7020 Lahey Medical

More information

Accolade. System. Brochure. Achieving Perfect Balance. Cemented and Cementless Femoral Hip System

Accolade. System. Brochure. Achieving Perfect Balance. Cemented and Cementless Femoral Hip System Accolade System Brochure Achieving Perfect Balance Cemented and Cementless Femoral Hip System Accolade Cemented Hip System Potential for Perfect Balance The Accolade C Femoral Component: A perfect balance

More information

How To Fix A Radial Head Plate

How To Fix A Radial Head Plate Mayo Clinic CoNGRUENT RADIAL HEAD PLATE Since 1988 Acumed has been designing solutions to the demanding situations facing orthopedic surgeons, hospitals and their patients. Our strategy has been to know

More information

VERSYS HERITAGE CDH HIP PROSTHESIS. Surgical Technique for CDH Hip Arthroplasty

VERSYS HERITAGE CDH HIP PROSTHESIS. Surgical Technique for CDH Hip Arthroplasty VERSYS HERITAGE CDH HIP PROSTHESIS Surgical Technique for CDH Hip Arthroplasty SURGICAL TECHNIQUE FOR VERSYS HERITAGE CDH HIP PROSTHESIS CONTENTS ANATOMICAL CONSIDERATIONS....... 2 PREOPERATIVE PLANNING............

More information

Radial Head Fracture Repair and Rehabilitation

Radial Head Fracture Repair and Rehabilitation 1 Radial Head Fracture Repair and Rehabilitation Surgical Indications and Considerations Anatomical Considerations: The elbow is a complex joint due to its intricate functional anatomy. The ulna, radius

More information

Computer Aided Engineering (CAE) Techniques Applied To Hip Implant

Computer Aided Engineering (CAE) Techniques Applied To Hip Implant International Journal Of Computational Engineering Research (ijceronline.com) Vol. 3 Issue. 3 Computer Aided Engineering (CAE) Techniques Applied To Hip Implant 1, M. S. Abo_Elkhair, 2, M. E. Abo-Elnor,

More information

Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner

Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner The Journal of Arthroplasty Vol. 24 No. 2 2009 Case Report Failure Mechanisms and Closed Reduction of a Constrained Tripolar Acetabular Liner William J. Robertson, MD, Christopher J. Mattern, MD/MBA, John

More information

An Alternative Conservative Approach to Hip Reconstruction

An Alternative Conservative Approach to Hip Reconstruction Joint Implant Surgery & Research Foundation Chagrin Falls, Ohio, USA An Alternative Conservative Approach to Hip Reconstruction Craig S. Waller, MD* and Timothy McTighe, Dr. H.S. (hc)** Acknow ledgement:

More information

High-Flex Solutions for the MIS Era. Zimmer Unicompartmental High Flex Knee System

High-Flex Solutions for the MIS Era. Zimmer Unicompartmental High Flex Knee System High-Flex Solutions for the MIS Era Zimmer Unicompartmental High Flex Knee System Zimmer Unicompartmental High Flex Knee Built On Success In today s health care environment, meeting patient demands means

More information

FATIGUE CONSIDERATION IN DESIGN

FATIGUE CONSIDERATION IN DESIGN FATIGUE CONSIDERATION IN DESIGN OBJECTIVES AND SCOPE In this module we will be discussing on design aspects related to fatigue failure, an important mode of failure in engineering components. Fatigue failure

More information

Treatment of hip dysplasia in children with. Terje Terjesen Department of Orthopaedic Surgery Oslo University Hospital, Rikshospitalet

Treatment of hip dysplasia in children with. Terje Terjesen Department of Orthopaedic Surgery Oslo University Hospital, Rikshospitalet Treatment of hip dysplasia in children with cerebral palsy Terje Terjesen Department of Orthopaedic Surgery Oslo University Hospital, Rikshospitalet Oslo, Norway Why do CP children have problems in the

More information

P REPLACEMENT SURGERY

P REPLACEMENT SURGERY P REPLACEMENT SURGERY DIRECT ANTERIOR APPROACH M I N I M I Z I N G R E C O V E R Y. M A X I M I Z I N G R E S U L T S. CENTER FOR MINIMAL INVASIVE JOINT SURGERY 2301 25TH STREET SOUTH FARGO ND 58103 701-241-9300

More information

Zimmer Persona Adverse Events Reported to FDA Through 3/27/2015

Zimmer Persona Adverse Events Reported to FDA Through 3/27/2015 Event Description: IT IS REPORTED THAT THE PATIENT WAS REVISED DUE TO GROSS ASEPTIC LOOSENING. Manufacturer 1822565-2015- 00224 PERSONA STEMMED CEMENTED TIBIAL COMPONENT 2/23/15 1/21/15 Injury Narrative:

More information

Small stem Exeter total hip replacement: clinical and radiological follow-up over a minimum of 2.5 years

Small stem Exeter total hip replacement: clinical and radiological follow-up over a minimum of 2.5 years Journal of Orthopaedic Surgery 2003: 11(2): 148 153 Small stem Exeter total hip replacement: clinical and radiological follow-up over a minimum of 2.5 years S Sivananthan, M Arif, DSK Choon Department

More information

SCIENTIFIC PAPER Anterior Dislocation Following Primary Total Hip Replacement by the Posterior Approach Aetiology and Treatment

SCIENTIFIC PAPER Anterior Dislocation Following Primary Total Hip Replacement by the Posterior Approach Aetiology and Treatment Hong HKJOS Kong Journal of Orthopaedic Surgery 2003;7(1):14-18. SCIENTIFIC PAPER Anterior Dislocation Following Primary Total Hip Replacement by the Posterior Approach Aetiology and Treatment Ng TP, Yau

More information

Metal Injection Molding (MIM) of components made of Titanium and its alloys

Metal Injection Molding (MIM) of components made of Titanium and its alloys Metal Injection Molding (MIM) of components made of Titanium and its alloys 1 Presentation content Introduction to Metal Injection Molding (MIM) Technology - explaination Products - examples Company brief

More information

Imagine being an orthopedic surgeon and knowing that your patient s

Imagine being an orthopedic surgeon and knowing that your patient s http://www.surgicalproductsmag.com/product-releases/2009/05/safer-methods-internal-bone-fixation Advances in Internal Bone Fixation Sharps safety for orthopedic surgeons by Ron Stoker Imagine being an

More information

Photograph showing the components of an artificial total hip

Photograph showing the components of an artificial total hip Case Study 4 (CS4) Artificial Total Hip Replacement Learning Objectives After studying this case study, you should be able to do the following: 1. List and briefly explain six biocompatibility considerations

More information

EMPLOYMENT: PEACHTREE ORTHOPAEDIC CLINIC ATLANTA- GA 2013- PRESENT Orthopaedic Surgeon Northside Hospital, Atlanta, GA

EMPLOYMENT: PEACHTREE ORTHOPAEDIC CLINIC ATLANTA- GA 2013- PRESENT Orthopaedic Surgeon Northside Hospital, Atlanta, GA GEORGE N. GUILD III, M.D. ORTHOPAEDIC SURGEON: ADULT HIP AND KNEE RECONSTRUCTION PEACHTREE ORTHOPAEDIC CLINIC 5505 PEACHTREE DUNWOODY ROAD SUITE 600 ATLANTA, GA 30342 TEL. (404)- 355-0743 FAX: 404-355-

More information

September 13-16, 2009

September 13-16, 2009 September 13-16, 2009 Waikoloa, HI September 13-16, 2009, Waokoloa, Waikoloa, HI Biomimetic i Tooth Implants by Electron Beam Melting Mari Koike, DDS, PhD 1 ; Gilbert Chahine 2 ; Radovan Kovacevic, PhD

More information

Posterior Referencing. Surgical Technique

Posterior Referencing. Surgical Technique Posterior Referencing Surgical Technique Posterior Referencing Surgical Technique INTRO Introduction Instrumentation Successful total knee arthroplasty depends in part on re-establishment of normal lower

More information

Accelerated Patient Rehabilitation

Accelerated Patient Rehabilitation Accelerated Patient Rehabilitation Providing The Tools Necessary For An Immediate Solution: NanoTite Implants QuickBridge Provisional Components Navigator System For CT Guided Surgery Initial Patient Presentation

More information

Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability

Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability Modular Dual Mobility Acetabular Components: An Important Extension of a Proven Approach to Hip Instability Robert Wetzel, MD, Lalit Puri, MD, S. David Stulberg, MD Disclosure In accordance with ACCME

More information

Malrotation Causing Patellofemoral Complications After Total Knee Arthroplasty

Malrotation Causing Patellofemoral Complications After Total Knee Arthroplasty CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 356, pp 144-153 1998 Lippincott Williams 8 Wilkins Malrotation Causing Patellofemoral Complications After Total Knee Arthroplasty R.A. Berger, MD *; L.S.

More information

Fatigue. 3. Final fracture (rough zone) 1. Fatigue origin. 2. Beach marks (velvety zone)

Fatigue. 3. Final fracture (rough zone) 1. Fatigue origin. 2. Beach marks (velvety zone) Fatigue Term fatigue introduced by Poncelet (France) 1839 progressive fracture is more descriptive 1. Minute crack at critical area of high local stress (geometric stress raiser, flaws, preexisting cracks)

More information

Zimmer Gender Solutions NexGen High-Flex Implants

Zimmer Gender Solutions NexGen High-Flex Implants Zimmer Gender Solutions NexGen High-Flex Implants Because Women and Men are Different Something new is taking shape It s all about shape. Women and men are different. That s not news to the medical establishment.

More information

Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology. Surgical Technique

Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology. Surgical Technique Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology Surgical Technique Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology 1 Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology Surgical

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. Dall-Miles Cabling System Dall-Miles Recon and Trauma Cable System Trochanteric Reattachment Using the Trochanteric

More information

Restoration Modular Cone Body/Conical Distal Stem Femoral Components Using the Restoration Modular Instrument System Surgical Technique

Restoration Modular Cone Body/Conical Distal Stem Femoral Components Using the Restoration Modular Instrument System Surgical Technique Restoration Modular Cone Body/Conical Distal Stem Femoral Components Using the Restoration Modular Instrument System Surgical Technique CONE CONICAL BROACHED CALCAR MILLED CONE MT3 FLUTED PLASMA CONICAL

More information

In this brochure you will: Keep in mind:

In this brochure you will: Keep in mind: Hip Implant Options Bearing Basics Today s hip implant patients are fortunate. Doctors can now choose from advanced implant materials. One choice doctors make when selecting the implant for you is called

More information

Taking the Mystique out of Implant Dentistry. Dr. Michael Weinberg B.Sc., DDS, FICOI

Taking the Mystique out of Implant Dentistry. Dr. Michael Weinberg B.Sc., DDS, FICOI Taking the Mystique out of Implant Dentistry Dr. Michael Weinberg B.Sc., DDS, FICOI What is Restorative Implant Dentistry? Restorative implant dentistry involves taking a few simple mechanical principles

More information

INTRODUCTION Cerclage Cable Vs. Monofilament Wire

INTRODUCTION Cerclage Cable Vs. Monofilament Wire B M P CABLE S Y S T E M S U R G I C A L TECHNIQUE INTRODUCTION The Biomet Medical Products (BMP ) Cable System provides a cost effective and simple method of trochanteric reattachment, fracture fixation,

More information

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare.

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare. Use of CP Titanium and Titanium alloys for Dental Implants Steve Hurson, Nobel Biocare U.S.A. Yorba Linda CA Abstract Professor Per Ingvar Branemark of Sweden published his results on 15 years of research

More information

Radiologic Assessment of Total Hip Arthroplasty: Loosening

Radiologic Assessment of Total Hip Arthroplasty: Loosening March 2004 Radiologic Assessment of Total Hip Arthroplasty: Loosening Grant Garrigues, Harvard Medical School Anatomy Liner Acetabular Component Calcar Stem Femoral Component 2 www.yoursurgery.com/proceduredetails.cfm?br=5&proc=27

More information

What Dental Implants Can Do For You!

What Dental Implants Can Do For You! What Dental Implants Can Do For You! Putting Smiles into Motion About Implants 01. What if a Tooth is Lost and the Area is Left Untreated? 02. Do You Want to Restore Confidence in Your Appearance? 03.

More information

Total Hip Replacement Hip replacement surgery, or arthroplasty, uses implants to resurface and replace the bones in the joint, re-creating the smooth gliding surfaces that were once intact. Hip replacement

More information

Total Elbow Arthroplasty and Rehabilitation

Total Elbow Arthroplasty and Rehabilitation Total Elbow Arthroplasty and Rehabilitation Surgical Indications and Considerations Anatomical Considerations: There are three bones and four joint articulations that have a high degree of congruence in

More information

Foot and Ankle Technique Guide Proximal Inter-Phalangeal (PIP) Fusion

Foot and Ankle Technique Guide Proximal Inter-Phalangeal (PIP) Fusion Surgical Technique Foot and Ankle Technique Guide Proximal Inter-Phalangeal (PIP) Fusion Prepared in consultation with: Phinit Phisitkul, MD Department of Orthopedics and Rehabilitation University of Iowa

More information

QUESTION I HAVE BEEN ASKED TO REHAB GRADE II AND III MCL INJURIES DIFFERENTLY BY DIFFERENT SURGEONS IN THE FIRST 6WEEKS FOLLOWING INJURY.

QUESTION I HAVE BEEN ASKED TO REHAB GRADE II AND III MCL INJURIES DIFFERENTLY BY DIFFERENT SURGEONS IN THE FIRST 6WEEKS FOLLOWING INJURY. QUESTION I HAVE BEEN ASKED TO REHAB GRADE II AND III MCL INJURIES DIFFERENTLY BY DIFFERENT SURGEONS IN THE FIRST 6WEEKS FOLLOWING INJURY. SOME ARE HINGE BRACED 0-90 DEGREES AND ASKED TO REHAB INCLUDING

More information

Performance evaluation of CLS Brevius Stem with Kinectiv Technology

Performance evaluation of CLS Brevius Stem with Kinectiv Technology Performance evaluation of CLS Brevius Stem with Kinectiv Technology Dirk Weidmann, Ronja Bruhn, Susanne Frei, Daniel Hertig, Xavier Langlois, Paul Hulme, PhD Zimmer GmbH Background A primary goal of total

More information

Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling

Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling Qasem Alyazji Near East University, Department of Biomedical Engineering Nicosia, North Cyprus, Turkey

More information

Patellofemoral Chondrosis

Patellofemoral Chondrosis Patellofemoral Chondrosis What is PF chondrosis? PF chondrosis (cartilage deterioration) is the softening or loss of smooth cartilage, most frequently that which covers the back of the kneecap, but the

More information

Understanding Total Hip Replacement

Understanding Total Hip Replacement Understanding Total Hip Replacement Brian J. White MD Orthopaedic Specialist in Disorders of the Hip Assistant Team Physician Denver Nuggets Western Orthopaedics Denver, Colorado Introduction This is designed

More information

CH 6: Fatigue Failure Resulting from Variable Loading

CH 6: Fatigue Failure Resulting from Variable Loading CH 6: Fatigue Failure Resulting from Variable Loading Some machine elements are subjected to static loads and for such elements static failure theories are used to predict failure (yielding or fracture).

More information

Metal-On-Metal Hip Resurfacing for Young Active Adults with Degenerative Hip Disease

Metal-On-Metal Hip Resurfacing for Young Active Adults with Degenerative Hip Disease Metal-On-Metal Hip Resurfacing for Young Active Adults with Degenerative Hip Disease November 1, 2006 Life is Movement TABLE OF CONTENTS The Alberta Bone & Joint Health Institute.. 2 Request. 2 Distribution

More information

ABGII Cement Free Total Hip Replacement System

ABGII Cement Free Total Hip Replacement System Cement Free Total Hip Replacement System Efficacy Innovation Cement Free Hip System The ABG Cement Free Hip System is one of the foremost ranges of cementless total hip replacement prosthesis. Used extensively

More information

Figure 1: Typical S-N Curves

Figure 1: Typical S-N Curves Stress-Life Diagram (S-N Diagram) The basis of the Stress-Life method is the Wohler S-N diagram, shown schematically for two materials in Figure 1. The S-N diagram plots nominal stress amplitude S versus

More information

THE ARTHROPLASTICAL POSSIBILITIES OF ELBOW JOINT IN RHEUMATOID ARTHRITIS. Miklós Farkasházi M.D.

THE ARTHROPLASTICAL POSSIBILITIES OF ELBOW JOINT IN RHEUMATOID ARTHRITIS. Miklós Farkasházi M.D. Abstract of the Ph.D. Thesis THE ARTHROPLASTICAL POSSIBILITIES OF ELBOW JOINT IN RHEUMATOID ARTHRITIS by Miklós Farkasházi M.D. Doctoral School of Semmelweis University Faculty of Physical Education and

More information

Patellofemoral Joint: Superior Glide of the Patella

Patellofemoral Joint: Superior Glide of the Patella Patellofemoral Joint: Superior Glide of the Patella Purpose: To increase knee extension. Precautions: Do not compress the patella against the femoral condyles. Do not force the knee into hyperextension

More information

Metasul LDH Large Diameter Head

Metasul LDH Large Diameter Head Metasul LDH Large Diameter Head Surgical Technique Metasul LDH Large Diameter Head Surgical Technique Enhancing Stability and Increasing Range of Motion Metasul LDH Large Diameter Head Surgical Technique

More information

INSTRUCTIONS FOR USING THE KNEE SOCIETY RADIOGRAPHIC EVALUATION FORM

INSTRUCTIONS FOR USING THE KNEE SOCIETY RADIOGRAPHIC EVALUATION FORM INSTRUCTIONS FOR USING THE KNEE SOCIETY RADIOGRAPHIC EVALUATION FORM Routine standing and recumbent anterior-posterior, lateral, and skyline patellar roentgenograms are made pre-operative, at time of hospital

More information

Acetabular Wedge Augment System

Acetabular Wedge Augment System Orthopaedics Acetabular Wedge Augment System Stryker Hips. Implant with confidence. Restoration Acetabular Wedge Augment System Restoration Acetabular Wedge Augments, united with Tritanium, Trident, MDM

More information

Exeter X3 RimFit. Acetabular Cup Surgical Protocol

Exeter X3 RimFit. Acetabular Cup Surgical Protocol Exeter X3 RimFit Acetabular Cup Surgical Protocol 2 Exeter X3 RimFit Surgical Protocol Table of Contents Indications and Contraindications... 4 Introduction... 5 Surgical Protocol... 6 Step 1 - Pre-Operative

More information

Integrated Computational Materials Engineering (ICME) for Steel Industry

Integrated Computational Materials Engineering (ICME) for Steel Industry Integrated Computational Materials Engineering (ICME) for Steel Industry Dr G Balachandran Head ( R&D) Kalyani Carpenter Special Steels Ltd., Pune 411 036. Indo-US Workshop on ICME for Integrated Realization

More information

The Mechanical Properties of Glass

The Mechanical Properties of Glass The Mechanical Properties of Glass Theoretical strength, practical strength, fatigue, flaws, toughness, chemical processes Glass Engineering 150:312 Professor Richard Lehman Department of Ceramics and

More information

DIAGNOSING SCAPHOID FRACTURES. Anthony Hewitt

DIAGNOSING SCAPHOID FRACTURES. Anthony Hewitt DIAGNOSING SCAPHOID FRACTURES Anthony Hewitt Introduction Anatomy of the scaphoid Resembles a deformed peanut Articular cartilage covers 80% of the surface It rests in a plane 45 degrees to the longitudinal

More information

ATI 2205 ATI 2205. Technical Data Sheet. Duplex Stainless Steel GENERAL PROPERTIES. (UNS S31803 and S32205)

ATI 2205 ATI 2205. Technical Data Sheet. Duplex Stainless Steel GENERAL PROPERTIES. (UNS S31803 and S32205) ATI 2205 Duplex Stainless Steel (UNS S31803 and S32205) GENERAL PROPERTIES ATI 2205 alloy (UNS S31803 and/or S32205) is a nitrogen-enhanced duplex stainless steel alloy. The nitrogen serves to significantly

More information

TOTAL KNEE REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE KNEE

TOTAL KNEE REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE KNEE TOTAL KNEE REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE KNEE John T. Dearborn, M.D. and Alexander P. Sah, M.D. The Center for Joint Replacement Please read this pamphlet before you see me so

More information

From Department of Orthopaedics, Conquest Hospital, The Ridge, East Sussex. n N.A. Sandiford, FRCS (Tr/Orth), MFSEM, Specialist. Registrar.

From Department of Orthopaedics, Conquest Hospital, The Ridge, East Sussex. n N.A. Sandiford, FRCS (Tr/Orth), MFSEM, Specialist. Registrar. Acta Orthop. Belg., 2014, 80, 372-379 ORIGINAL STUDY Primary total hip arthroplasty with hydroxyapatite coated titanium femoral stems. Does design philosophy influence long term outcome? : results of a

More information

Dental Implant Options in Atrophic Jaws

Dental Implant Options in Atrophic Jaws Dental Implant Options in Atrophic Jaws Orthopedic Application Jay B. Reznick, D.M.D., M.D. Diplomate, American Board of Oral and Maxillofacial Surgery Tarzana, CA Endopore Dental Implant System Screw-Type

More information

Clinically Proven Fiber Metal Material. Power to choose advanced bearing technologies to match patient demands

Clinically Proven Fiber Metal Material. Power to choose advanced bearing technologies to match patient demands Trilogy IT Acetabular System Design Rationale 1 For orthopedic surgeons treating a wide range of patients, the Trilogy IT Acetabular System provides a clinically proven1-5 Fiber Metal Material and the

More information

Lentur Cable System. Surgical Technique

Lentur Cable System. Surgical Technique Lentur Cable System Surgical Technique Contents Introduction... Page 1 System Design Features And Benefits... Page 2 Implants... Page 3 Instrumentation... Page 4 Surgical Technique... Page 5 Single Cable...

More information

Adult Forearm Fractures

Adult Forearm Fractures Adult Forearm Fractures Your forearm is made up of two bones, the radius and ulna. In most cases of adult forearm fractures, both bones are broken. Fractures of the forearm can occur near the wrist at

More information

PERIPROSTHETIC IMPLANTS

PERIPROSTHETIC IMPLANTS PERIPROSTHETIC IMPLANTS PRODUCT OVERVIEW CLINICAL SOLUTIONS Periprosthetic fractures present unique challenges, such as how to gain fixation when the medullary canal is occupied. Special techniques and

More information

The LOCATOR concept. Simplicity and versatility for prosthesis fixation

The LOCATOR concept. Simplicity and versatility for prosthesis fixation The concept Simplicity and versatility for prosthesis fixation The concept Experience the freedom in prosthesis fixation Simple and secure fixation of implant-supported prostheses is essential for successful

More information

THE MEASUREMENT AND ANALYSIS OF AXIAL DEFORMITY AT THE KNEE

THE MEASUREMENT AND ANALYSIS OF AXIAL DEFORMITY AT THE KNEE THE MEASUREMENT AND ANALYSIS OF AXIAL DEFORMITY AT THE KNEE Kenneth A. Krackow, M.D. Clinical Director, Department of Orthopaedic Surgery Kaleida Health System Buffalo General Hospital Professor and Full

More information

.URGENT DEVICE CORRECTION

.URGENT DEVICE CORRECTION zimmer.urgent DEVICE CORRECTION Zimmer, Inc. P.O. Box 706 Warsaw, IN 46561-0706 574.267.6131 www.zimmer.com July 22, 2008 Dear Surgeon: Since we last wrote to you in May 2008 regarding the Durom Acetabular

More information

Anterior Hip Replacement

Anterior Hip Replacement Disclaimer This movie is an educational resource only and should not be used to manage Orthopaedic health. All decisions about the management of hip replacement and arthritis management must be made in

More information

A 15-year follow-up study of 4606 primary total knee replacements

A 15-year follow-up study of 4606 primary total knee replacements Knee A 15-year follow-up study of 466 primary total knee replacements V. I. Roberts, C. N. A. Esler, W. M. Harper From Glenfield General Hospital, Leicester, England This is a 15-year follow-up observational

More information

TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP

TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP John T. Dearborn, M.D. and Alexander P. Sah, M.D. The Center for Joint Replacement Please read this pamphlet before you see me so that

More information

M 2 a-magnum. Design Rationale. Large Metal Articulation. Knees Hips Extremities Cement and Accessories PMI. Technology

M 2 a-magnum. Design Rationale. Large Metal Articulation. Knees Hips Extremities Cement and Accessories PMI. Technology M 2 a-magnum Large Metal Articulation Design Rationale Knees Hips Extremities Cement and Accessories PMI Technology M 2 a-magnum Large Metal Articulation History of Metal-on-Metal Early metal-on-metal

More information

The Journal of Arthroplasty

The Journal of Arthroplasty The Journal of Arthroplasty 29 (2014) 219 224 Contents lists available at ScienceDirect The Journal of Arthroplasty journal homepage: www.arthroplastyjournal.org One-Component Revision of Failed Hip Resurfacing

More information

Mini Medical School _ Focus on Orthopaedics

Mini Medical School _ Focus on Orthopaedics from The Cleveland Clinic Mini Medical School _ Focus on Orthopaedics Arthritis of the Shoulder: Treatment Options Joseph P. Iannotti MD, PhD Professor and Chairman, Department of Orthopaedic Surgery The

More information

Praxis Technology: Enhancements to Ti-MIM processing bring medical implants a step closer

Praxis Technology: Enhancements to Ti-MIM processing bring medical implants a step closer Praxis Technology: Enhancements to Ti-MIM processing bring medical implants a step closer Praxis Technology, a US-based manufacturer focused exclusively on titanium products has made substantial progress

More information

Corporate Medical Policy Computer Assisted Surgical Navigational Orthopedic Procedures

Corporate Medical Policy Computer Assisted Surgical Navigational Orthopedic Procedures Corporate Medical Policy Computer Assisted Surgical Navigational Orthopedic File Name: Origination: Last CAP Review: Next CAP Review: Last Review: computer_assisted_surgical_navigational_orthopedic_procedures

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Health Technology Appraisal Total hip replacement and resurfacing arthroplasty for the treatment of pain or disability resulting from end stage arthritis

More information

.org. Arthritis of the Hand. Description

.org. Arthritis of the Hand. Description Arthritis of the Hand Page ( 1 ) The hand and wrist have multiple small joints that work together to produce motion, including the fine motion needed to thread a needle or tie a shoelace. When the joints

More information

.org. Shoulder Joint Replacement. Anatomy

.org. Shoulder Joint Replacement. Anatomy Shoulder Joint Replacement Page ( 1 ) Although shoulder joint replacement is less common than knee or hip replacement, it is just as successful in relieving joint pain. Shoulder replacement surgery was

More information