Quantification of Polysorbate Using the ACQUITY UPLC H-Class Bio System with ELS Detection
|
|
|
- Jemima Bradley
- 9 years ago
- Views:
Transcription
1 Quantification of Polysorbate Using the ACQUITY UPLC H-Class Bio System with ELS Detection Sean M. McCarthy Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY UPLC H-Class Bio System, configured with an ACQUITY UPLC Column Manager and ACQUITY UPLC ELS Detector, is an ideal system for the quantification of detergents commonly used in biopharmaceutical formulations. INTRODUCTION Biotherapeutics formulations are complex and their composition is tightly controlled to ensure product stability, and efficacy. Formulations of biotherapeutics contain detergents that are used to prevent denaturation and aggregation of the active pharmaceutical ingredient. Of the detergents used, polysorbates 20 and 80 are very common due to their low toxicity and biocompatibility. By their nature, polysorbates are structurally and chemically heterogeneous mixtures of species, making their quantification challenging. There are recent reports that demonstrate the use of a trap-and-elute strategy for detergent analysis in formulated biopharmaceuticals. 1 Often, analysis of detergent in formulated samples is necessary to ensure stability of the detergents. These strategies provide a rapid and robust process for separating the therapeutic and other excipients from the detergent in the formulation. WATERS SOLUTIONS ACQUITY UPLC H-Class Bio System with Column Manager ACQUITY UPLC ELS Detector In this work, we present the use of the ACQUITY UPLC H-Class Bio System with a Column Manager and an evaporative light scattering detector (ELSD) for determining the concentration of polysorbate in formulated samples. Using the mixed-mode Oasis MAX online cartridge, we effectively trap polysorbate at low ph while the therapeutic and other excipients are unretained. Using the ACQUITY UPLC Column Manager equipped with a switching valve, unretained components are diverted to waste, then the valve position is changed so flow is directed to the ELSD prior to elution of polysorbate from the column. The proposed workflow represents a straightforward and efficient method for analyzing polysorbate content in biotherapeutic formulations. Empower 3 CDS Software Oasis MAX Solid-Phase Extraction KEY WORDS Detergents, polysorbate, evaporative light scattering detection, biopharmaceutical 1
2 E X P E R IM E N TA L LC conditions QSM Mobile phase A: 2% formic acid in water Mobile phase B: 2% formic acid in isopropanol Seal wash: 10% acetonitrile in water Seal wash period: 5 min. Gradient: Time Flow %A %B %C %D Curve (ml/min) Initial Initial Polysorbate 20, Polysorbate 80, and bovine serum albumin were purchased from Sigma Aldrich. Stock solutions of each were prepared in Milli-Q water at 1 mg/ml. Stock solutions were prepared for each polysorbate as follows: 0.4, 0.3, 0.2, 0.1, and 0.05 mg/ml for polysorbate 20 and 0.3, 0.2, 0.1, 0.05, and mg/ml for polysorbate 80. Each sample was spiked with bovine serum albumin to a concentration to 30 mg/ml. Sample manager Wash solvent: Purge solvent: 50/50 water/isopropanol 50/50 water/isopropanol Column manager Advanced mode: For plumbing diagram, see Figure 1 Column: Oasis MAX, 30 µm, 2.1 x 20 mm (p/n ) Column temp.: 30 C Events: Time Event Action (min) 2.8 Right valve Position Right valve Position 1 ELSD Gain: 100 Data rate: 2 pps Time constant: Slow Mode: Heating Power level: 75% Temp.: 80 C Gas pressure: 20 psi (1.64 slpm) 2
3 RESULTS AND DISCUSSION Instrument configuration and plumbing To prevent fouling of the ELS detector, we configured our system to efficiently divert unretained protein and other excipient material to waste following injection. This was accomplished using 6-port, 2-position valves in the column manager while operating in advanced mode. As shown in Figure 1, the right valve is configured so that the LC flow can be directed as needed during the separation. In position 1 (top) LC flow is directed to waste. In position 2 (bottom) LC flow is directed to the ELS detector. At the end of the separation, the valve is returned to position 1 for the following injection. Programming the valve switch is accomplished within the inlet method editor without the need for additional components or connections. Evaluation of chromatographic reproducibility For routine use, methods used for quantitation need to be reproducible in retention time and area. Since our configuration utilized both a trap-and-elute method as well as a diversion valve, we evaluated the reproducibility of the separation at each of the mass load levels for polysorbate 20 and 80. Shown in Figure 2 is an overlay of triplicate injections for each polysorbate 80 standard prepared as described above, containing 30 mg/ml protein with varying levels of polysorbate. As shown, the separations are reproducible both within and between polysorbate concentrations. From this data we determined that both the instrument and the plumbing configuration used were acceptable for routine polysorbate analysis. Figure 1. Fluidic diagram of column manager. Figure 2. Overlay of triplicate injections of polysorbate 80 at five concentration levels. 3
4 Evaluation of assay linearity After determining that our system configuration was suitable for polysorbate analysis, we generated calibration curves for polysorbate 20 and 80. The calibration curves were constructed in Empower 3 Software using Quickset and defining concentration levels. The data was then processed using a basic processing method and the calibration curve was generated within Empower. The data was plotted using a log vs. log scale, yielding a linear relationship between concentration and response. As shown in Figures 3 and 4, excellent linearity was achieved for polysorbate 20 and 80, respectively. Figure 3. Polysorbate 20 calibration curve. Data points were collected in triplicate. Figure 4. Polysorbate 80 calibration curve. Data points were collected in triplicate. 4
5 CONCLUSION The analysis of detergents in biopharmaceutical formulations is a required but often challenging task in the laboratory. The ACQUITY UPLC H-Class System with ELSD and column manager in a trap-and-elute configuration provides a comprehensive system for performing these analyses. As shown, the integrated valves on the ACQUITY UPLC Column Manager provides reproducible results without the need for additional external switching valves. In addition, the ACQUITY UPLC ELSD yields a linear response for quantification. References 1. Hewitt D, Zhang T, Kao YH. Quantitation of polysorbate 20 in protein solutions using mixed-mode chromatography and evaporative light scattering detection. J Chromatogr A Dec 26;1215(1 2): doi: /j.chroma Megoulas NC, Koupparis MA. Twenty years of evaporative light scattering detection. Crit. Rev. Anal. Chem. 2005;35: Waters, The Science of What s Possible, ACQUITY UPLC, Oasis, and Empower are registered trademarks of Waters Corporation. All other trademarks are the property of their respective owners Waters Corporation. Produced in the U.S.A. March EN AG-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A. T: F:
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS
Automating the Determination of Drug-to-Antibody Ratio (DAR) of Antibody Drug Conjugates (ADCs) Based on Separation by Hydrophobic Interaction Chromatography (HIC) Robert Birdsall, Eoin Cosgrave, Henry
Analysis of Vegetable Oils by High Performance Liquid Chromatography
Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
High-Throughput 3-D Chromatography Through Ion Exchange SPE
High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample
A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples
A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples Mary Lame, Hua Yang, Sherri Naughton, and Erin Chambers Waters Corporation, Milford,
Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS
Development of Integrated Informatics Workflows for the Automated Assessment of Comparability for Antibody Drug Conjugates (ADCs) Using LC-UV and LC-UV/MS Henry Shion, Robert Birdsall, Steve Cubbedge,
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
How To Use An Acquity Qda Detector
Mass-Directed Isolation of a Synthetic Peptide Using the ACQUITY QDa Detector Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY QDa Detector
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )
Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 ) Andrew Aubin Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S Fast analysis of
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
Protein Separation Technology
[ ] Protein Separation Technology Protein Separation Technology The analysis and characterization of protein samples requires the detection of small chemical differences between large molecules. Most often
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna
Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna Paula Hong and Pat McConville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics
The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.
The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen. Mark Crawford Single Presenter Name Abstract With the recent decline in hive bee populations this year the investigation of acaricides
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Introduction to the ACQUITY UPLC H-Class Amino Acid Analysis System
Introduction to the ACQUITY UPLC H-Class Amino Acid Analysis System 0.22 0.20 0.18 0.16 0.14 0.12 Cys 0.10 0.08 0.06 0.04 0.02 1.5 0 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 2012
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
Discovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
ORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3
Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3 Hongwei Xie and Weibin Chen Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B
EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS
EKSIGENT EKSPERT NANOLC 4 Unmatched flexibility for low flow LC/MS EKSIGENT EKSPERT NANOLC 4 Focused ekspertise In 23, Eksigent changed the way scientists looked at nanoscale chromatography with the introduction
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Chapter 28: High-Performance Liquid Chromatography (HPLC)
Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion
A Novel Approach to Low Volume Sample Preparation
A Novel Approach to Low Volume Sample Preparation Tony Edge, Jon Bardsley, Ken Meadows, Luisa Pereira International Symposium on Chromatography Salzburg, Austria September 2014 1 The world leader in serving
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control
Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control Dr. Frank Steiner, 1 Andreas Brunner, 1 Fraser McLeod, 1 Dr. Sergey Galushko 2 1 Dionex Corporation, Germering,
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System
An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System Mary E. Lame, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA,
for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns
for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns Product Manual for SolEx TM HRP Cartridges and SolEx TM HRP RSLC Columns Page 1 of 15 Product Manual for SolEx TM HRP Cartridges SolEx HRP (P/N 074400)
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
CORESTA RECOMMENDED METHOD N 72
CORESTA RECOMMENDED METHOD N 72 DETERMINATION OF TOBACCO-SPECIFIC NITROSAMINES IN SMOKELESS TOBACCO PRODUCTS BY LC-MS/MS (February 216). INTRODUCTION The CORESTA Smokeless Tobacco Sub-Group studied various
Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis
Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis Marcia Santos, Staff Applications Scientist, Beckman Coulter Life Sciences, Brea, CA USA Mark Lies, Marketing
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns Jonathan P. Danaceau, Kenneth J. Fountain, and Erin E. Chambers Waters
α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase
α-cyclodextrin New specifications prepared at the 57th JECFA (2001) and published in FNP 52 Add 9 (2001). An ADI not specified was established at the 57th JECFA (2001). SYNONYMS α-schardinger dextrin,
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
Anion chromatography using on-line recycled eluents
Journal of Chromatography A, 1089 (2005) 82 86 Anion chromatography using on-line recycled eluents Takashi Yokoyama, Hitoe Maekubo, Asami Sakai, Michio Zenki Department of Chemistry, Faculty of Science,
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
Purification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS
Therapeutic Drug Monitoring of Antiretroviral Drugs with PLC-M Ursula Gutteck-Amsler, Katharina M. Rentsch Abstract Prospective and retrospective studies have provided some evidence of the clinical and
Automation of Solid Phase Extraction and Column Chromatographic Cleanup
Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for
of Biologics Using LC/MS in Compliant Laboratories
Demonstrating Comparability and Consistency of Biologics Using LC/MS in Compliant Laboratories Patrick Boyce Biopharmaceutical Marketing Manager Europe & India Waters 2011 Waters Corporation Overview Challenges
MEDICINAL CHEMISTRY APPLICATIONS BOOK
MEDICINAL CHEMISTRY APPLICATIONS BOOK Introduction...3 The Role of LC and MS in Medicinal Chemistry...5 SCREENING System Management for a High Throughput Open Access UPLC/MS System Used During the Analysis
ß-CYCLODEXTRIN SYNONYMS
ß-CYCLODEXTRIN Prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995) superseding specifications prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications
Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
PrepTip. Reverse Phase PrepTip User Guide
PrepTip Reverse Phase PrepTip User Guide All text, photographs and illustrations are copyrighted by Harvard Apparatus, Inc. 2004. PrepTip is a trademark of Harvard Apparatus, Inc. Harvard Apparatus 84
Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC
Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC Abstract: High performance liquid chromatography (HPLC) was used to determine the identity and concentrations of food dyes
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH Cristiana C. Leandro 1, Peter Hancock 1, Christian Soulier 2, Françoise Aime 2 1 Waters Corporation, Manchester,
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.
Salman Azimi, 1 Nayan S. Mistry, 2 and Michelle Wood 2. Drug Quality Control Laboratory, Supreme Council of Health, Doha, Qatar 2
UPLC/MS/MS for the Screening, Confirmation, and Quantification of Drugs Illegally Added to Herbal/Dietary Supplements for the Enhancement of Male Sexual Performance Salman Azimi, 1 Nayan S. Mistry, 2 and
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report
of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley
SUPPLEMENTARY FIGURES
SUPPLEMENTARY FIGURES Fig. S1: Effect of ISO- and TAC-treatments on the biosynthesis of FAS-II elongation products in M. tb H37Ra. LC/MS chromatograms showing a decrease in products with elemental compositions
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
INSTRUCTIONS 56-1190-98. Edition AC
Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography
Protocol: HPLC (amino acids)
Page 1 of 6 1. Chemicals Composition M MW gram volume (ml) product code 0-pthaldialdehyde C 8H 6O 2 134.13 Sigma P0657 Perchloric acid HClO 4 3.5 100.46 Dilute 70% stock solution 1:1 with demi to obtain
RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml
GE Healthcare Life Sciences Instructions 71-7146-00 AI Ion Exchange Columns RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml Introduction RESOURCE Q and S are pre-packed columns for separating biomolecules
INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE
Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND
IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities
ICMS and LCMS Determination of Ionic Liquids, Counterions, and Impurities Leo Wang, Detlef Jensen, and William Schnute Dionex Corporation, Sunnyvale, CA, USA; Dionex Corporation, Olten, Switzerland INTRODUCTION
Lean Process Improvements for the GMP Laboratory
Lean Process Improvements for the GMP Laboratory ITRODUCTIO Pharmaceutical manufacturing costs the industry approximately $90 billion each year and represents twice the expenditure of research and development.
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
Introduction. Methods. Sample Processing
PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko
Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction