1 User Manual Luminescent Dissolved Oxygen Probe: Model LDO10101, LDO10103, LDO10105, LDO10110, LDO10115 or LDO10130 DOC Safety information Precautionary labels Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement. Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user. Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal. Specifications Note: Specifications are subject to change without notice. Specifications Probe type Dissolved oxygen range Dissolved oxygen accuracy Details Luminescent dissolved oxygen (LDO) probe 0.1 to 20.0 mg/l (ppm) 1 to 200% saturation ±0.1 mg/l for 0 to 8 mg/l ±0.2 mg/l for greater than 8 mg/l % saturation resolution 0.1% Stabilization time Temperature resolution Temperature accuracy Pressure resolution T90% at 10 seconds (when stirred) 0.1 C (0.18 F) ±0.3 C (±0.54 F) 1 hpa Pressure accuracy ±0.8% Operating temperature range Storage temperature range Minimum sample depth Dimensions (standard) Dimensions (rugged) Cable connection 0 to 50 C (32 to 122 F) 0 to 40 C (32 to 104 F) 25 mm (0.984 in.) Diameter: 29 mm (1.14 in.) Length: 191 mm (7.52 in.) Cable length: 1 or 3 m (3.28 or 9.84 ft) Diameter: 46 mm (1.81 in.) Length: 223 mm (8.73 in.) Cable length: 5, 10, 15 or 30 meter (16.40, 32.81, or ft) M12 digital output and connector compatible with HQd meters Product overview The LDO101 series probe is a luminescent dissolved oxygen probe (Figure 1). The LDO10105, LDO10110, LDO10115 or LDO10130 rugged probe is available with a 5, 10, 15 or 30 m (16.40, 32.81, or ft) cable. The LDO10101 or LDO10103 standard 1

2 Figure 1 Probe overview probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe measures the dissolved oxygen concentration in wastewater, drinking water and general applications. Note: For BOD applications, use the LBOD10101 probe which has LDO technology and a stirrer. 1 LDO probe cap 5 Pressure sensor module 2 Rugged probe (5, 10, 15 or 30 meter cable) 6 ibutton compartment 1 3 Standard probe (1 or 3 meter cable) 7 Pressure sensor module cap 4 Probe cable 1 ibutton is a registered trademark of Maxim Integrated Products, Inc. Preparation for use Note: Do not touch the probe cap with a hand, fingers or any surface that can scratch the cap. Prepare the probe for use before calibration or sample measurement. 1. Make sure that the probe cap and ibutton are installed correctly. The ibutton label should be up. 2. Make sure that the probe cap and ibutton have the same lot code. 3. If a rugged probe, make sure that the shroud is installed before field use (refer to Install the shroud on page 9). Note: Damage to the sensing elements can occur if the shroud is not installed during field use. Damage under these conditions is not covered by the product warranty. 4. Rinse the probe cap with deionized water. Blot dry with a lint-free cloth. 5. If dissolved oxygen monitoring periods are longer than 6 hours, condition the probe cap for 72 hours. Calibrate the probe once every 8 hours. Note: After 72 hours, the probe cap will reach a fully hydrated state. 2

3 Calibration Before calibration: The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached. It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options. To view the current calibration, push, select View Probe Data, then select View Current Calibration. If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option. Prepare the probe for use (refer to Preparation for use on page 2. If a rugged probe, remove the shroud from the probe (refer to Remove the shroud on page 9). Calibration notes: % saturation or mg/l calibration methods are available in the Modify Current Settings menu. The slope value is the comparison between the latest calibration and the factory calibration shown as a percentage. An additional zero point calibration can be added to the calibration routine. Refer to Change calibration options on page 8. The calibration is recorded in the probe and the data log. The calibration is also sent to a PC, printer or flash memory stick if connected. Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed. If a calibration error occurs, refer to Troubleshooting on page 10. Water-saturated air (100%) calibration procedure: 1. Connect the probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn on the meter. 2. Push Calibrate. 3. Push Methods. Select User Cal - 100%. Push OK. 4. Rinse the probe cap with deionized water. Blot dry with a lintfree cloth. 5. Add approximately ¼ inch (6.4 mm) of reagent water to a narrow-neck bottle, such as a BOD bottle. 3

4 6. Put a stopper in the bottle and shake the bottle vigorously for approximately 30 seconds to saturate the entrapped air with water. Allow up to 30 minutes for contents to equilibrate to room temperature. 7. Remove the stopper. Carefully dry the probe cap with a nonabrasive cloth. Put the probe in the bottle. 8. Push Read. The display shows "Stabilizing" and a progress bar as the probe stabilizes. The display shows the standard value when the reading is stable. 9. Push Done to view the calibration summary. 10. Push Store to accept the calibration and return to the measurement mode. If a rugged probe, install the shroud on the probe (refer to Install the shroud on page 9). Sample measurement Before measurement: The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached. If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual for more information. Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3). Calibrate the probe if accuracy better than ±0.50 mg/l is necessary for the application. Prepare the probe for use (refer to Preparation for use on page 2). To deploy a rugged probe at a distance, toss the probe body with a gentle underhand throw. Do not swing the probe by the cable as this may cause injury to the user, will cause severe strain on the cable and will shorten the service life of the cell. Damage under these conditions is not covered by the product warranty. Do not submerge the pressure sensor module. Measurement notes: Stabilization times with smaller concentration changes generally will be longer and can be minimized by correct stirring and conditioning. Experiment to determine the correct stir rate if necessary. Salinity affects the concentration of dissolved oxygen in the sample. To correct for salinity effects, refer to Advanced operation on page 6 or do the Auto salinity correction on page 5. Data is automatically stored in the data log when Press to Read or Interval is selected in the Measurement Mode. When Continuous is selected, data will only be stored when Store is selected. Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed. If a measurement error occurs, refer to Troubleshooting on page 10. 4

5 Measurement procedure: Note: Procedures also apply for rugged model probes. 1. Connect the probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn on the meter. 2. Rinse the probe cap with deionized water. Blot dry with a lintfree cloth. 3. Put the probe in the sample and stir gently or add a stir bar. Do not put the probe on the bottom or sides of the container. Stir the sample at a moderate rate or put the probe in flowing conditions. 4. Put the probe in the sample at least 25 mm (0.984 in.) deep. Push Read. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display will show the lock icon when the reading stabilizes. 5. Repeat steps 2-4 for additional measurements. When measurements are done, store the probe Storage on page 10. Auto salinity correction Dissolved substances affect the amount of oxygen water can hold. Manually enter salinity settings for the most accurate dissolved oxygen measurements or use the optional Auto Salinity Correction feature. The manufacturer recommends using the Auto Salinity Correction feature when measuring dissolved oxygen in samples where salinity varies. Auto salinity correction measures dissolved oxygen through the connection of one LDO101 probe and one CDC401 conductivity cell (set to Salinity parameter). The value obtained by the CDC401 conductivity cell automatically adjusts the salinity value for the LDO101 series probe. Salinity units are represented as parts per thousand (ppt) or (º/ºº). 1. Connect the LDO101 probe and CDC401 conductivity cell to the HQd meter. Turn the meter on. Note: Security options must be turned off to use auto salinity correction. 2. Push and select the CDC401 Settings. 3. Select Current Method, Hach Salinity. Push OK. Note: To change more measurement options, choose Modify Current Settings, change the parameter to Salinity and any other desired settings. 4. Push EXIT until the Full Access Options menu is shown. 5. Select LDO101 Settings, Modify Current Settings, Measurement Options, Salinity Correction: Off. Then select Sal Correction Mode: Off. Push the DOWN arrow to select Auto (*) - Use CDC401. Push OK. 6. Push EXIT until the meter returns to the measurement mode. The HQ40d meter is now set up to automatically use the salinity values obtained by the CDC401 with the LDO101 probe. If the salinity value is out of range, the display will show *S= ---- above the dissolved oxygen reading as shown in Figure 2. Note: The asterisk (*) indicates that salinity is automatically correcting the dissolved oxygen value. No asterisk indicates that salinity is being manually corrected. Warning messages will override the asterisk (*) priority. 5

6 Figure 2 Salinity value out of range Advanced operation Parameter-specific settings can be changed through the Full Access Options menu. Details about menu navigation, available options and how to change them are given in the screens, tables and procedures throughout this section. The settings that can be changed are shown in Table 1. Setting Options Description Table 1 Parameter-specific settings Measurement Options Resolution Measurement Limits Salinity Correction Pressure Units Averaging Interval Defines measurement resolution Upper and lower measurement limits Value for salinity correction Atmospheric pressure units How often the meter calculates an average readings Units Calibration Options mg/l % Calibration Primary unit of measurement User 100% User 100% with 0 User mg/l User mg/l with 0 Factory Calibration Reminder Reminder Repeat Off, 8 h, 12 h, 1 d, 2 d, 5 d or 7 d Expires Immediately, Reminder + 30 m, Reminder + 1 h, Reminder + 2 h or Continue Reading Change measurement options 6 Methods are groups of default or user-defined settings relevant to specific applications. If the meter is set to the default method and the Modify Current Settings option is chosen, a prompt for a new name is shown after the changes are entered. The settings are saved with this name to distinguish them from the default method settings, which cannot be changed. A saved method can be used instead of multiple adjustments to the individual settings. Changes made to a user defined method are automatically saved with the existing name. Multiple methods can be saved for the same probe on each meter. There are three default methods available for the LDO101 probe: Factory Cal (Calibration with default LDO calibration) User Calibration 100% (allows user calibration)

7 Default 1. Make sure a probe is connected to the meter. 2. Push and select LDO101 Settings. 3. Select Modify Current Settings. 4. Select Units. Select mg/l (default) or %. 5. Select Measurement Options and update the settings: Option Resolution Description Sets the resolution: 0.1 Fast (0.35 mg/l)/min 0.01 Fast (0.35 mg/l)/min 0.01 Medium (0.15 mg/l)/min (default), or 0.01 Slow (0.05 mg/l)/min The resolution affects the number of decimal places and the stabilization time. Higher resolution measurements take more time to stabilize. Measurement Limits Salinity Correction Set the measurement limits Lower limit (default: 0.0 mg/l; 0%) or Upper limit (default: 20.0 mg/l; 200%). The measurement limits can be set to match the acceptable values for the sample. When the measurement is above the upper limit setting or below the lower limit setting, the meter shows an "Out of limits" message. This message is an alert to a potential problem with the process conditions. Sets the salinity correction Off (default), Manual or Auto (connect conductivity probe). Salinity lowers the solubility of dissolved oxygen in water. To correct for salinity in the sample, set salinity correction to manual and then enter the salinity value. Note: When the HQ40d meter is used, a conductivity probe can also be connected for automatic salinity measurement and correction. The parameter setting for the conductivity probe must show salinity. Salinity Value Sets the salinity value (default: 35.0 ). When salinity correction is set to manual, sets the salinity value of the sample. Salinity can be measured with a conductivity probe. Pressure Units Averaging Interval Sets the pressure units hpa, mbar, inhg or mmhg. The meter shows the atmospheric pressure at the current elevation, which is necessary for accurate measurements. Note: This pressure reading will not agree with readings from sources such as weather stations, which report atmospheric pressure at sea level. Sets the averaging interval Off, 30, 60, 90 seconds, 3, 5 minutes. The averaging interval is useful for samples that contain a lot of air bubbles, for example in an aeration basin. The air bubbles cause the dissolved oxygen readings to vary greatly from one reading to the next. To make the readings more consistent, increase the averaging interval. The meter will take measurements at the same frequency but show only the average over a longer interval. Note: Labels and options may vary depending on the units selected. 6. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatically saved with the same method name. 7. Push EXIT until the meter returns to the measurement mode. 7

8 Change calibration options 1. Make sure that a probe is connected to the meter. 2. Push and select LDO101 Settings. 3. Select Modify Current Settings. 4. Select Calibration Options and update the settings: Option Calibration Standard Value Description User 100% (water-saturated air (100%) calibration) User 100% with 0 (water-saturated air (100%) calibration with 0 point) User mg/l (calibration with a specified dissolved oxygen concentration (mg/l) solution) User mg/l with 0 (calibration with a specified dissolved oxygen concentration (mg/l) solution with 0 point ) Factory (calibration with the default LDO calibration) When Calibration is set to mg/l or mg/l with 0, sets the concentration of the solution that will be used for calibration 2.00 to mg/l (default=7.00 mg/l) 5. Select Calibration Reminder and update the settings: Option Reminder Repeat Description Meter will make an audible sound when calibration is due and repeat the sound at selected interval Off, 8 h, 12 h, 1 d, 2 d, 5 d or 7 d. Expires Calibration expires after the selected time Immediately, Reminder + 30 min, Reminder + 1 h, Reminder + 2 h or Continue Reading. Note: The meter cannot be used to read samples after calibration has expired unless Continue Reading is selected. 6. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatically saved with the same method name. 7. Push EXIT until the meter goes back to the measurement mode. Maintenance Clean the probe Keep the probe cap free of deposits for the best measurements. Note: Do not touch the black colored substrate of the probe cap. Do not use alcohol or other organic solvents to clean the black colored substrate of the probe cap. These solvents cause damage to the probe cap. 1. Remove the shroud (refer to Remove the shroud on page 9). 2. Gently clean the probe cap with a mild detergent, water and a soft cloth or cotton swab. Do not remove the black colored substrate from the probe cap. Do not scrub the probe cap or lens. 3. If water is present between the probe cap and lens: a. Remove the probe cap. b. Blot dry the probe cap and lens with a soft dry cloth. c. Install the probe cap. 4. Install the shroud (refer to Install the shroud on page 9). 8

9 Replace probe cap Remove the shroud Figure 3 Probe exploded view The probe cap must be replaced every 365 days or more often if the cap becomes damaged or fouled. The meter will show a reminder message when 30 days of probe service life remains on the probe cap. For LDO probe cap replacement instructions, refer to the instructions provided with the LDO probe cap replacement kit. 1. Loosen and remove the locking ring (Figure 3). 2. Slide the shroud and locking ring off the probe. 1 Shroud 7 Probe lens 2 Locking ridges (8x) 8 Cap seal 3 Locking rib 9 Temperature sensor 4 Locking ring 10 Locking groove 5 Probe cap 11 Locking ribs (4x) 6 O-ring Install the shroud 1. Put the locking ring on the probe with the threads toward the probe cap (Figure 3 on page 9). 2. Slide the shroud on the probe until it is against the locking groove (rugged) or ribs (standard). Slide the standard probe shroud on the standard probe until the inside locking ridges align halfway between the ribs on the probe. Turn the shroud slightly until it is seated. 3. Hand-tighten the locking ring on the shroud. 9

10 Storage Troubleshooting Dry storage the manufacturer recommends that the probe is stored dry when the probe is used for measurements of short duration (less than 6 hours). Wet storage the probe must be stored wet when the probe is used for monitoring periods longer than 6 hours. Dry storage Note: Rugged probes may be stored dry with the shroud installed if the storage container is sufficiently large. 1. Rinse the probe with deionized water. Blot dry with a lint-free cloth. Note: The probe must be conditioned again in tap water for at least 30 minutes prior to use. 2. If a rugged probe, remove the shroud (optional). Refer to Remove the shroud on page 9. Wet storage Note: The need for recalibration is minimized if the probe cap is kept wet. 1. Put the probe in tap water. 2. During the initial 72 hours in tap water, calibrate the probe once every 8 hours. Note: After 72 hours of storage in tap water, the probe cap will reach a fully hydrated state. Message or symptom Possible cause Action Probe not supported Software not updated To download the most current version of the software, refer to the applicable product page on the manufacturer's website. Refer to the HQd Series meter manual for specific instructions for the meter model. Connect a probe or probe requires service HQd meter does not support IntelliCAL probe Probe not connected correctly Software not updated Large number of methods stored on the probe Damaged probe Contact a Technical Support Representative. Disconnect, then connect the probe. Tighten the locking nut. To download the most current version of the software, refer to the applicable product page on the manufacturer's website. Refer to the HQd Series meter manual. Continue to let the probe connect. Do not disconnect the probe. Make sure there is connectivity with another probe or meter to confirm isolated issue with probe. Contact a Technical Support Representative. 10

11 Message or symptom Possible cause Action Out of range Out of range Slope out of range LDO calibration not supported (factory calibration) O2 Sensor 0 days remaining O2 Sensor ## of days remaining Probe cap loose, scratched or damaged Temperature and/or pressure sensor error Damaged probe Sample outside of specifications ibutton number does not match probe cap lot number Bubbles trapped under probe tip Probe cap exposed to direct sunlight Probe not prepared for sample Calibration method settings Probe cap loose, scratched or damaged Temperature and pressure errors LED lights do not function Bubbles trapped under probe tip LDO method calibration option is set to Factory. There are 0 days remaining in the life of the probe cap Meter set to incorrect date and time Software not updated There are 30 days or fewer remaining in the life of the probe cap. Reposition or replace the probe cap. Make sure that the temperature and pressure sensors are both reading correctly. 1 Make sure that the blue and red LEDs are both illuminated on the probe. If not, replace the probe or contact a Technical Support Representative. Make sure that the sample concentration, temperature and pressure are within the range of the probe. Replace the ibutton or probe cap or do a user calibration. Gently shake the probe until bubbles are removed. Install the protective shroud. Let the probe reach equilibrium in a water-saturated air environment and do the calibration again. Make sure that the calibration standards in the method are correct. Locate and install the ibutton that matches the probe cap and replace the probe cap. Make sure that the temperature and pressure sensors are both reading accurately. Contact a Technical Support Representative. 1 Contact a Technical Support Representative. Gently shake the probe until bubbles are removed. If user calibration is necessary, change the settings in Calibration Options. Refer to Change calibration options on page 8. Replace the probe cap. Calibration will be allowed, however the calibration icon and question mark will be shown on the measurement screen even if the calibration has passed. Disconnect the probe from the meter. Set the correct date and time in the Meter Options menu. Connect the probe and make sure that the message has been removed. Update the HQd software to the latest version and test again. Replace the probe cap soon. 11

12 Message or symptom Possible cause Action Calibration failed: outside of acceptance criteria/temperature out of range/offset out of limits Water Saturated air equilibration not reached Probe cap loose, scratched, or damaged Temperature and/or pressure sensor error Damaged probe Allow longer equilibration time. Change the location of the probe cap or replace the probe cap. Make sure that the temperature and pressure sensors are both reading correctly and within range. 1 Make sure that the blue and red LEDs are both illuminated on the probe. If not, replace the probe or contact a Technical Support Representative. 1 The pressure as measured by the probe is what is referred to as atmospheric pressure and is not corrected to sea level. Weather station pressures are reported at sea level and commonly referred to as mean sea level pressure. As a result the probe will not read the same as most household or professional barometers or weather station reports (which are compensated) unless reported at sea level. In order to compare the pressure results obtained from the probe barometer and these compensated barometers, it is necessary to first compensate the pressure reported by the probes mathematically. HACH COMPANY World Headquarters P.O. Box 389, Loveland, CO U.S.A. Tel. (970) (800) (U.S.A. only) Fax (970) orders@hach.com HACH LANGE GMBH Willstätterstraße 11 D Düsseldorf, Germany Tel. +49 (0) Fax +49 (0) info@hach-lange.de HACH LANGE Sàrl 6, route de Compois 1222 Vésenaz SWITZERLAND Tel Fax Hach Company/Hach Lange GmbH, 2010, All rights reserved. Printed in U.S.A. 05/2013, Edition 2

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process

More information

DISSOLVED OXYGEN PROBE

DISSOLVED OXYGEN PROBE 51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe

More information

USEPA Electrode Method Method 8156

USEPA Electrode Method Method 8156 ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method

More information

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01 Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery

More information

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination

More information

Milwaukee USER MANUAL

Milwaukee USER MANUAL Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the

More information

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure

More information

ExStik TM ph Waterproof Meters

ExStik TM ph Waterproof Meters User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL

More information

Pocket Colorimeter II

Pocket Colorimeter II DOC022.53.80451 Pocket Colorimeter II User Manual 04/2014, Edition 1 Table of Contents Section 1 Specifications...3 Section 2 General information...5 2.1 Safety information...5 2.1.1 Use of hazard information...5

More information

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600 User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously

More information

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn

More information

RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer

RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer EN FR IT DE ES CONTENTS / EN Title CONTENTS Page 1. SAFETY INFORMATION...1 2. INTRODUCTION...2 3. SPECIFICATIONS...3 4. PARTS & CONTROLS...4

More information

6 H2O + 6 CO 2 (g) + energy

6 H2O + 6 CO 2 (g) + energy AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic

More information

SenTix 950 SenTix 980

SenTix 950 SenTix 980 SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the

More information

Instruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter

Instruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter Instruction Manual HI 9146 Portable Waterproof Microprocessor Dissolved Oxygen Meter w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction

More information

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Trace Dissolved Oxygen Sensor

Trace Dissolved Oxygen Sensor Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION

More information

OxyGuard. Dissolved Oxygen Probe

OxyGuard. Dissolved Oxygen Probe OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard

More information

HOBO Dissolved Oxygen Logger (U26-001) Manual

HOBO Dissolved Oxygen Logger (U26-001) Manual HOBO Dissolved Oxygen Logger (U26-001) Manual The HOBO Dissolved Oxygen logger is a standalone logger that uses RDO Basic Technology to measure dissolved oxygen (DO). The logger has an optical sensor that

More information

Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph

Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or

More information

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510 User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance

STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance Page 1 of 9 KEY WORDS- Dissolved Oxygen, calibration, probe preparation, meter preparation APPROVALS APPROVED BY: Kean S. Go, DATE: +/?? APPROVED BY: A Pement DATE: a// A/? f Lisa Ross, EHAP Senior Scientist

More information

Dissolved Oxygen Sensors

Dissolved Oxygen Sensors Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit

More information

Relative Humidity Calibration Kit

Relative Humidity Calibration Kit Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration

More information

Appendix C. Vernier Tutorial

Appendix C. Vernier Tutorial C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer

More information

YSI Model 5000 YSI Model 5100. Dissolved Oxygen Temperature. Operations Manual

YSI Model 5000 YSI Model 5100. Dissolved Oxygen Temperature. Operations Manual YSI Model 5000 YSI Model 5100 Dissolved Oxygen Temperature Operations Manual Table of Contents 1. General Description...1 2. Specifications...3 3. Quick Start...5 4. Controls...6 4.1 Front Panel... 6 4.2

More information

PART 1 - INTRODUCTION...

PART 1 - INTRODUCTION... Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART

More information

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference

More information

User Guide. 9708 Dissolved Oxygen Probes

User Guide. 9708 Dissolved Oxygen Probes User Guide 9708 Dissolved Oxygen Probes Introduction The Thermo Scientific Orion 9708 dissolved oxygen probe simplifies measurements of dissolved oxygen, particularly Biochemical Oxygen Demand (BOD).

More information

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation

More information

24 Wide Screen LED Monitor with HDMI

24 Wide Screen LED Monitor with HDMI 24 Wide Screen LED Monitor with HDMI USER MANUAL L24LE11B_MANUAL_EN_R1 Please read this manual carefully before using, and keep it for future reference. TABLE OF CONTENTS GETTING STARTED Monitor Stand

More information

DISSOLVED OXYGEN METER Model : DO-5509

DISSOLVED OXYGEN METER Model : DO-5509 DISSOLVED OXYGEN METER Model : DO-5509 Your purchase of this DISSOLVED OXYGEN METER marks a step forward for you into the field of precision measurement. Although this DISSOLVED OXYGEN METER is a complex

More information

Care and use guide. Contents. page

Care and use guide. Contents. page Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement

More information

MAINTENANCE & TROUBLESHOOTING

MAINTENANCE & TROUBLESHOOTING MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens replace the projection lamp replace the batteries in the remote use the security lock feature troubleshoot the projector Cleaning

More information

BODY FAT & HYDRATION % MONITOR SCALE

BODY FAT & HYDRATION % MONITOR SCALE EVALUATION BODY FAT & HYDRATION % MONITOR SCALE INSTRUCTION MANUAL MODEL 12-1190 ver 7/08 BODY FAT & HYDRATION % MONITOR SCALE Introduction and Precaution: MODEL: 12-1190 1. Congratulations for purchasing

More information

User Guide. Temperature and Humidity Datalogger. Model 42280

User Guide. Temperature and Humidity Datalogger. Model 42280 User Guide Temperature and Humidity Datalogger Model 42280 Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger. The 42280 is a wall-mount, tripod

More information

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506 Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking

More information

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407780 Product Page Integrating Sound Level Datalogger

More information

Portable Waterproof ph/dissolved Oxygen Meter

Portable Waterproof ph/dissolved Oxygen Meter Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35632-Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ /

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510

User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510 User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater

More information

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required

More information

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from

More information

User s manual FLIR VP50/VP52 Non-contact AC voltage detector

User s manual FLIR VP50/VP52 Non-contact AC voltage detector User s manual FLIR VP50/VP52 Non-contact AC voltage detector User s manual FLIR VP50/VP52 #T559851; r. AD/ 9134/9134; en-us Table of contents 1 Disclaimers... 1 1.1 Copyright... 1 1.2 Quality assurance...

More information

SINGLE RANGE DS METERS Operation Instructions

SINGLE RANGE DS METERS Operation Instructions SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax

More information

Digital Photo Picture Frame. Operation Manual

Digital Photo Picture Frame. Operation Manual Digital Photo Picture Frame Operation Manual 20070309 CONGRATULATIONS on your purchase of a Polaroid 7 LCD Digital Photo Picture Frame. Please read carefully and follow all warnings and instructions in

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method

More information

User Manual. Thermo Scientific Orion VERSA STAR

User Manual. Thermo Scientific Orion VERSA STAR User Manual Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 68X006501 Revision A April 2014 Contents Chapter 1 Meter Introduction... 6 Meter Overview... 6 Packing List... 7

More information

User's Guide. Pinless Moisture Meter. Model MO257

User's Guide. Pinless Moisture Meter. Model MO257 User's Guide Pinless Moisture Meter Model MO257 Introduction Congratulations on your purchase of the Extech MO257 Pinless Moisture Meter. The pinless moisture sensor monitors the moisture in wood and other

More information

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003)

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003) (Part # S-LIB-M003) The smart sensor is designed to work with the HOBO Weather Station logger. The smart sensor has a plug-in modular connector that allows it to be added easily to a HOBO Weather Station.

More information

site monitoring Kit Site Monitoring Kit User Manual we prove it.

site monitoring Kit Site Monitoring Kit User Manual we prove it. site monitoring Kit 1 User Manual Site Monitoring Kit User Manual we prove it. 2 site monitoring Kit Content 1. Introduction 3 2. Content of the Site Monitoring Kit 4 3. Preparation 6 1. Positioning of

More information

Instruction Manual Temp 4/5/6 Temperature Meter Series

Instruction Manual Temp 4/5/6 Temperature Meter Series Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.

More information

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide

More information

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately: BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data

More information

Mini Digital Altimeter & Climb rate & Barometer & Thermometer & Compass & Weather forecast & Time

Mini Digital Altimeter & Climb rate & Barometer & Thermometer & Compass & Weather forecast & Time Mini Digital Altimeter & Climb rate & Barometer & Thermometer & Compass & Weather forecast & Time User Manual 1.Introduction Thank you for purchase of Mini Digital Altimeter. This mini digital atimeter

More information

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for

More information

Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements

Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

PA500II Enterprise Mobile Computer

PA500II Enterprise Mobile Computer PA500II Enterprise Mobile Computer - PA500II - Quick Reference Guide 400891G Version 1.0 PA500II Product Introduction & Accessory Kit After opening the box, ensure the following accessories for the PA500II

More information

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

MAINTENANCE & TROUBLESHOOTING

MAINTENANCE & TROUBLESHOOTING MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens clean the fan intake filter replace the projection lamp replace the batteries in the remote control use the Kensington lock feature

More information

Nokia Bluetooth Headset BH-800 User Guide. 9246974 Issue 2

Nokia Bluetooth Headset BH-800 User Guide. 9246974 Issue 2 Nokia Bluetooth Headset BH-800 User Guide 9246974 Issue 2 DECLARATION OF CONFORMITY We, NOKIA CORPORATION declare under our sole responsibility that the product HS-24W is in conformity with the provisions

More information

UltraBasic Benchtop ph Meter Operation Manual

UltraBasic Benchtop ph Meter Operation Manual UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.

More information

User Guide. RDO Optical Dissolved Oxygen Sensor

User Guide. RDO Optical Dissolved Oxygen Sensor User Guide RDO Optical Dissolved Oxygen Sensor ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

Quaternary Ammonium Compounds DOC316.53.01128

Quaternary Ammonium Compounds DOC316.53.01128 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower

More information

Windows 10.1 Tablet (UB-15MS10 and UB-15MS10SA) FAQ December 2014

Windows 10.1 Tablet (UB-15MS10 and UB-15MS10SA) FAQ December 2014 Windows 10.1 Tablet (UB-15MS10 and UB-15MS10SA) FAQ December 2014 Navigation... 2 Question 1: What are charms used for?... 2 Question 2: How do I navigate while using a mouse or trackpad?... 3 Question

More information

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

SCREAMIN' EAGLE HEAVY BREATHER PERFORMANCE AIR CLEANER KIT REV. 05-04- SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT GENERAL Kit Numbers 95-08B, 9006-09B, 876-0A Models For model fitment information, see the P&A Retail Catalog or the Parts and Accessories

More information

2 (. 3 4 5 . -,.. -,. 3. : 1 2 3 > 4, :!! 1,, 17 Lt MAX 2 3 4,, > 5.,,, 6 :, :1)! 2)! 7,, 1, :,. 2,, Bottom of steam generator 7 8 9 10 11 12 automatically and this phenomenon will continue for

More information

Instruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X361305 Rev. 0-11/05. Technology Made Easy...

Instruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X361305 Rev. 0-11/05. Technology Made Easy... Instruction Manual DO 110 Hand-held Dissolved Oxygen Meter with RS 232C Technology Made Easy... 68X361305 Rev. 0-11/05 Preface This manual serves to explain the use of the hand-held series meters. The

More information

QFP910. Water detector. Building Technologies. CE1N2732en 2016-03-01. Synco living

QFP910. Water detector. Building Technologies. CE1N2732en 2016-03-01. Synco living 2 732 Synco living Water detector QFP910 Wireless radio sensor to record water puddles RF communication based on KNX standard (868 MHz unidirectional) Battery operation with standard 1.5 V batteries Use

More information

IN-OUT Thermometer with Cable Free Sensor and Clock

IN-OUT Thermometer with Cable Free Sensor and Clock IN-OUT Thermometer with Cable Free Sensor and Clock MODEL: RAR232 USER'S MANUAL INTRODUCTION Congratulations on your purchase of the RAR232 In-Out Thermometer with 433MHz cable free sensor and calendar

More information

Bluelab Soil ph Meter Instruction Manual

Bluelab Soil ph Meter Instruction Manual Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for

More information

Pro1030 USER MANUAL. English

Pro1030 USER MANUAL. English Pro1030 USER MANUAL English Contents Warranty...i Introduction...1 Item #605182 Rev A, January 2013 For the latest version of this manual, visit ysi.com 2013 YSI Incorporated. The YSI logo is a registered

More information

VISM CQB Scope Series

VISM CQB Scope Series 1 VISM CQB Scope Series Congratulations on the purchase of your new VISM CQB Scope! The CQB Series of Scopes give you many great options so you can choose the scope that best fits your needs. Backed by

More information

For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/.

For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. Instruction Sheet PN 51A-499AOZ/rev.H September 2010 Model 499A OZ Ozone Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS APPLICATION

More information

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of the blood 7 Procedure to test glucose 7 Procedure to

More information

Operator Quick Guide EC SENSOR

Operator Quick Guide EC SENSOR Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra

More information

GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G WI-FI

GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G WI-FI GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G WI-FI LET S GET THIS SHOW ON THE ROAD You must be excited about your brand new Telstra Pre-Paid 3G Wi-Fi. This guide will help you get connected as quickly and

More information

Thermistor Thermometer MODEL NO. 93210-00

Thermistor Thermometer MODEL NO. 93210-00 Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:

More information

125 8880 telstra.com/ppmbb visit a telstra store or partner GETTING TO KNOW YOUR

125 8880 telstra.com/ppmbb visit a telstra store or partner GETTING TO KNOW YOUR FOR MORE INFORMATIoN: 125 8880 telstra.com/ppmbb visit a telstra store or partner GETTING TO KNOW YOUR Telstra PRE-paid 4G wi-fi LET S GET THIS SHOW ON THE ROad WHAT S INSIDE Your Telstra Pre-Paid 4G Wi-Fi

More information

Integrating Digiquartz MET3 and MET3A Broadband Meteorological Systems with Leica GX1200 Series GPS Receivers

Integrating Digiquartz MET3 and MET3A Broadband Meteorological Systems with Leica GX1200 Series GPS Receivers Integrating Digiquartz MET3 and MET3A Broadband Meteorological Systems with Leica GX1200 Series GPS Receivers The standard by which other standards are measured Paroscientific, Inc. 4500 148 th Ave. N.E.

More information

Ethernet Networking Installation & Operations Manual Humminbird ETHERNET

Ethernet Networking Installation & Operations Manual Humminbird ETHERNET Ethernet Networking Installation & Operations Manual 531906-2EN_A GPS Unit 1 Humminbird ETHERNET Unit 2 Thank You! Thank you for choosing Humminbird, the #1 name in marine electronics. Humminbird has built

More information

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer User Manual Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer MODEL 42515 Introduction Congratulations on your purchase of the Model 42515 IR Thermometer. The Model 42515 IR thermometer

More information

INSTRUCTION MANUAL. NC-500 NETcat Pro Wiring Tester

INSTRUCTION MANUAL. NC-500 NETcat Pro Wiring Tester INSTRUCTION MANUAL English...1 16 Français...17 32 Español...33 48 Deutsch...49 64 NC-500 NETcat Pro Wiring Tester Read and understand all of the instructions and safety information in this manual before

More information

SYMBOL MC9060 HAND HELD TERMINAL USER MANUAL

SYMBOL MC9060 HAND HELD TERMINAL USER MANUAL SYMBOL MC9060 HAND HELD TERMINAL USER MANUAL CORPORATE VERSION FEBRUARY 2007 TABLE OF CONTENTS Welcome... 3 Getting Started... 3 Equipment Setup... 3 Battery Installation... 4 About the Stylus... 5 Aligning

More information

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

SCREAMIN' EAGLE HEAVY BREATHER PERFORMANCE AIR CLEANER KIT REV. 2015-04-23 SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT GENERAL Kit Numbers 29299-08, 29098-09 Models For model fitment information, see the P&A Retail Catalog or the Parts and Accessories

More information

4-Channel Thermometer / Datalogger

4-Channel Thermometer / Datalogger USER GUIDE 4-Channel Thermometer / Datalogger RTD and Thermocouple Inputs Model SDL200 Introduction Congratulations on your purchase of the Extech SDL200 Thermometer, an SD Logger Series meter. This meter

More information

PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques

PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques Introduction The amount of dissolved oxygen in process water is continually gaining importance in many industries as a critical parameter

More information

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and

More information

Submersion/Insertion Toroidal Sensor

Submersion/Insertion Toroidal Sensor Instruction Sheet PN 51A-228/rev.P December 2010 Model 228 Submersion/Insertion Toroidal Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS

More information

Nokia Bluetooth Headset BH-607. Issue 2. 1

Nokia Bluetooth Headset BH-607. Issue 2. 1 Nokia Bluetooth Headset BH-607 2 3 1 4 5 6 7 8 9 11 10 Issue 2. 1 12 13 14 DECLARATION OF CONFORMITY Hereby, NOKIA CORPORATION declares that this BH-607 product is in compliance with the essential requirements

More information

RS DIGITAL THERMOMETER

RS DIGITAL THERMOMETER RS DIGITAL THERMOMETER 1319A INSTRUCTION MANUAL C F MX MN AVG TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION... 1 2. SPECIFICATIONS... 1 3. NAME OF PARTS AND POSITIONS... 3 4. OPERATION INSTRUCTIONS... 5

More information

Xbox LIVE Vision. english

Xbox LIVE Vision. english Xbox LIVE Vision Thanks for choosing the Xbox LIVE Vision camera. Xbox LIVE Vision adds video to your Xbox 360 system and games, including: Video chat and picture messages. Photo snapshots. Family Settings

More information

W850. Maintenance Guide

W850. Maintenance Guide W850 Maintenance Guide August 2011 www.lexmark.com Contents...3 Cleaning the exterior of the printer...3 Storing supplies...3 Conserving supplies...3 Checking the status of supplies...4 Checking the status

More information

The Vortex Spitfire 1x Prism Scope

The Vortex Spitfire 1x Prism Scope REPLACE image The Vortex Spitfire 1x Prism Scope Perfect for the AR platform, the Vortex Spitfire 1x prism scope combines a compact, prism-based design with the intuitive, glass-etched DRT (Dual Ring Tactical)

More information

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...

More information

Hygro-Thermometer + InfraRed Thermometer Model RH101

Hygro-Thermometer + InfraRed Thermometer Model RH101 User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative

More information

Duct Humidity Transmitter

Duct Humidity Transmitter SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers

More information

T0118 T2118 T3118. Instruction Manual

T0118 T2118 T3118. Instruction Manual Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity

More information

Chloride and Salinity

Chloride and Salinity INTRODUCTION Chloride Chloride and Chloride, in the form of the Cl ion, is one of the major inorganic anions, or negative ions, in saltwater and freshwater. It originates from the dissociation of salts,

More information