Picture archiving and communication systems (PACS) and quality assurance

Size: px
Start display at page:

Download "Picture archiving and communication systems (PACS) and quality assurance"

Transcription

1 IT Guidance Documents Picture archiving and communication systems (PACS) and quality assurance Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists

2 This guidance forms part of a series on the developments in information technology in radiology. This is a fast-moving field and developments are occurring rapidly. Consequently, this guidance will be updated regularly and readers should check regularly that they are using the most up-to-date guidance available. 2

3 This document gives radiologists essential guidance on the areas of quality assurance which need to be monitored throughout the life of picture archiving and communication systems (PACS). It includes the need to ensure consistency in image appearance, the testing of image display devices, and the issue of rejecting and analysing inadequate digital images. This document replaces BFCR(08)8 Picture archiving and communication systems (PACS) and quality assurance, which has now been withdrawn. 1. Overview 1.1 PACS quality assurance (QA) is an overarching topic that covers all aspects of system design and implementation, embracing everything from information governance to network performance. PACS QA is a core component of PACS acceptance testing, and quality of service should be maintained thereafter through regular performance monitoring. 1.2 From a clinical perspective, image quality is the major consideration for anyone using PACS for primary (diagnostic) image interpretation. PACS should support image QA through system design, and PACS training should ensure staff can acquire and manipulate images to optimise image display. Medical physicists have an important role in the acceptance testing and monitoring of acquisition modalities and medical display devices. 2. Consistent presentation of images 2.1 QA of PACS image display relies on consistent presentation of images across the enterprise, irrespective of whether images are viewed on a soft copy display device or on printed hard copy. 2.2 PACS and acquisition modality workstations can optimise image presentation by enabling users to adjust the contrast and brightness of images, and to apply spatial and graphical operations, such as user annotations, shutters, flip/rotate, display area selection and zoom. These changes can be saved with the images in the form of a Digital Imaging and Communications in Medicine (DICOM) Greyscale Softcopy Presentation State (GSPS). To ensure safe and consistent presentation of images, all PACS display devices must display these images with their saved presentation state applied. Where more than one presentation state is saved with an image, the system should ensure that one of these is the default QA presentation state for image display. 2.3 For the consistency of the perceived pixel intensity, a standard contrast curve has been defined as the DICOM Greyscale Standard Display Function (GSDF). This provides a standard against which different types of display and hard copy output devices can be calibrated. The Integrating the Healthcare Enterprise (IHE) consistent presentation of images profile requires support of both the DICOM GSPS and the DICOM GSDF. 3. QA of image display devices 3.1 All PACS display devices used for primary diagnostic image viewing should undergo formal QA acceptance testing and regular performance monitoring. In the UK, guidance on these procedures is provided by Institute of Physics and Engineering in Medicine (IPEM), 1 referencing guidance from the Medicines and Healthcare products Regulatory Agency, the American Association of Physicists in Medicine and the International Electrotechnical Commission. A summary of the 2005 IPEM guidance is given in Table In addition to the 2005 IPEM guidance, 1 it is recommended that all primary diagnostic displays meet acceptance criteria matching the minimum specification guidelines for display devices used for clinical image interpretation summarised in Table If a primary display workstation does not meet all the recommendations outlined above, a local risk assessment should be undertaken as to whether the workstation can be used for primary diagnostic clinical interpretation. PACS workstations not suitable for diagnostic use should be clearly labelled and the rationale for imposing restrictions on clinical use should be covered in PACS training. 3.4 These guidelines do not require colour image display capability for primary diagnostic displays. Colour display is desirable to support the display of a number of advanced imaging modalities and annotations. Examples include nuclear medicine, functional magnetic resonance imaging (MRI), physiological MRI/CT, dual-energy CT, computer-aided diagnosis (CAD). Historically, colour display devices have been unable to meet the recommended standards for primary diagnostic use, hence in high workflow areas, the use of monochrome displays has been preferred. Technological advances mean that colour devices which achieve the recommended greyscale performance are now available. 3.5 The calibration and display of colour is a complex topic, and beyond the scope of this document. The guidance documents cited above are themselves limited in scope to monochrome performance, and are 3

4 not intended to ensure the reproducibility of colour information. Agreed guidelines for the quality assurance of colour image reproduction on medical displays have not yet been published. Advice on the calibration and any QA required, in excess of that given in the monochrome guidance, should be sought from the manufacturer. 3.6 Due to the lack of standardised methodology for verifying medical image colour reproduction precision, radiologists reporting colour images should be aware that colour fidelity may be altered if viewed for primary diagnosis on equipment maintained to different standards to the device that created the image (for example, equipment supplied by different vendors). A risk assessment should be performed if colour shifts could be diagnostically relevant, although in the case of false-colour volume rendered images, and pseudocolour parametric maps, this is unlikely to be relevant. 3.7 Handheld portable devices (such as mobile phones, ipads and so on), which can be used for image viewing, are not included in the guidance. While they can offer sufficient resolution (if viewed under appropriate low lighting conditions) for gross image review/aide memoire purposes, they are clearly not suitable for primary diagnosis. Table 1. Summary of image display monitor guidance from IPEM Physical parameter Frequency Remedial level Image display monitor condition Daily to weekly Image display monitors should be clean, and the perceived contrast of the test pattern should be consistent between monitors connected to the same workstation. Ensure that the 5% and 95% details superimposed on the 0% and 100% squares respectively are visible Greyscale contrast ratio 3 monthly Ratio white to black <250 Distance and angle calibration 3 monthly ± 5mm ± 3 Resolution 3 monthly Grade AAPM TG18-QC resolution patterns according to the reference score (CX >4) Greyscale drift 6 to 12 monthly Black baseline ±25% White baseline ±20% DICOM greyscale calibration 6 to 12 monthly GSDF ±10% Uniformity 6 to 12 monthly U% >30% Variation between monitors 6 to 12 monthly Black baseline >30% White baseline >30% Room illumination 6 to 12 monthly >15 lux 4

5 Table 2. Summary of the PACS and Imaging Informatics Group minimum and recommended specification for primary diagnostic display devices used for clinical image interpretation. This guidance applies to all workstations where CR, DR, fluoroscopy, ultrasound, CT, MR, nuclear medicine and PET images are viewed (excluding mammography). Screen resolution b (Native pixel array) Minimum a 1280 x 1024 (~1.3 megapixels) Recommended 1500 x 2000 c (~3 megapixels) Screen size (viewable diagonal) 42 cm (~17 ) 50 cm (~ 20 ) Maximum luminance 170 cd/m2 d 500 cd/m2 Luminance ratio (maximum/minimum) 250:1 e 500:1 Greyscale calibration Within 10% GSDF Calibrated to GSDF Greyscale bit depth Video display interface 8-bit greyscale (24-bit colour) Digital (where possible) Analogue f 10-bit greyscale Digital Notes a. The minimum specification for diagnostic display devices is only appropriate if clinical image viewing is performed according to image viewing guidelines, making use of the application software zoom, pan, magnification and windowing tools. b. LCD devices should be run at their native resolution to ensure there is a 1:1 match between screen pixels and screen resolution, and therefore no loss of image quality due to screen interpolation. CRT displays can be run at a variety of resolutions with no loss of display quality; however, care should be taken that the correct aspect ratio is maintained to avoid distortion of the image. c. Displays with a variety of aspect ratios are in common usage, in both landscape and portrait orientations. The absolute values of the horizontal and vertical screen matrix dimensions are the limiting factors in the resolution of image display. The total pixel count is therefore of secondary importance. d. High fidelity displays ( 3 megapixels) are recommended in radiology and other areas where large numbers of images are reported to reduce image interpretation and reporting times, and thereby assist department workflow. e. AAPM TG18 recommendation. f. AAPM TG18 & IPEM recommendation. g. New installations should not be analogue, except where a workstation/display package is provided as a single unit by a medical device manufacturer. 4. Image rejects and reject analysis 4.1 PACS must be able to hide images that are not considered adequate for clinical image viewing. These images should not be deleted from the system, but should be retained so that they can be accessed by authorised personnel for reject analysis. It is helpful if PACS enables these images to be categorised at the time of rejection, and can collate them in a PACS reject folder to facilitate audit analysis. 4.2 PACS can support QA workflow by optionally hiding images received from modalities from general clinical viewing until the images have been quality assured on a PACS workstation. By combining PACS support for QA workflow with PACS reject analysis functionality, it should be possible to store all images acquired on PACS, without the need for a separate image archive and reject analysis package. Approved by the Board of the Faculty of Clinical Radiology: 15 June 2012 Reference 1. Institute of Physics and Engineering in Medicine. Report 91. Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Imaging Systems. London: IPEM, TestingofDiagnosticX-RayImagingSystems.aspx (last accessed 27/11/12) 5

6 The Royal College of Radiologists 38 Portland Place London W1B 1JQ Tel +44 (0) Fax +44 (0) A Charity registered with the Charity Commission No Citation details: The Royal College of Radiologists. Picture archiving and communication systems (PACS) and quality assurance, Second edition. London: The Royal College of Radiologists, Ref No. BFCR(12)15 The Royal College of Radiologists, November 2012 For permission to reproduce any of the content contained herein, please permissions@rcr.ac.uk This material has been produced by The Royal College of Radiologists (RCR) for use internally within the specialties of clinical oncology and clinical radiology in the United Kingdom. It is provided for use by appropriately qualified professionals, and the making of any decision regarding the applicability and suitability of the material in any particular circumstance is subject to the user s professional judgement. While every reasonable care has been taken to ensure the accuracy of the material, RCR cannot accept any responsibility for any action taken, or not taken, on the basis of it. As publisher, RCR shall not be liable to any person for any loss or damage, which may arise from the use of any of the material. The RCR does not exclude or limit liability for death or personal injury to the extent only that the same arises as a result of the negligence of RCR, its employees, Officers, members and Fellows, or any other person contributing to the formulation of the material.

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists

More information

Medical imaging monitors specification guidelines

Medical imaging monitors specification guidelines Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,

More information

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for

More information

The Royal College of Radiologists

The Royal College of Radiologists The Royal College of Radiologists Board of the Faculty of Clinical Radiology Electronic remote requesting This guidance is only available electronically from www.rcr.ac.uk This guidance forms part of a

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

MQSA Quality Control Manual for Monochrome Displays for Mammography

MQSA Quality Control Manual for Monochrome Displays for Mammography MQSA Quality Control Manual for Monochrome Displays for Mammography Ver1.0 Applicable Models LMD-DM50, LMD-DM30 Sony Corporation - 1 - MQSA Quality Control Manual for Monochrome Displays for Mammography

More information

Nio. Industry-standard diagnostic display systems

Nio. Industry-standard diagnostic display systems Nio Industry-standard diagnostic display systems Diagnostic confidence in grayscale With the Nio diagnostic display system, Barco brings dependable diagnostic imaging to its true potential. Presenting

More information

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors Health NYC Quality Assurance Guidelines for Primary Diagnostic Monitors June 2015 Content I. Overview... 1 II. Definitions... 1 III. Acceptance Testing Requirements... 4 IV. Bi-weekly PDM Testing... 4

More information

CAR Standards for Teleradiology

CAR Standards for Teleradiology CAR Standards for Teleradiology The standards of the Canadian Association of Radiologists (CAR) are not rules, but are guidelines that attempt to define principles of practice that should generally produce

More information

Paediatric radiology experts

Paediatric radiology experts Radiology Paediatric radiology experts Board of the Faculty of Clinical Radiology The Royal College of Radiologists Foreword This statement looks at issues faced by paediatric radiologists both as general

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

Highmark Professional Provider Privileging Application Teleradiology Supplement INSTRUCTIONS

Highmark Professional Provider Privileging Application Teleradiology Supplement INSTRUCTIONS Highmark Professional Provider Privileging Application Teleradiology Supplement INSTRUCTIONS 1. Please complete a separate Teleradiology Application Supplement for each physical location where imaging

More information

Supports screening, diagnostic, and multimodality workflows

Supports screening, diagnostic, and multimodality workflows GE Healthcare Universal Viewer Breast Imaging Product Data Sheet Supports screening, diagnostic, and multimodality workflows Introduction Centricity PACS with Universal Viewer 1 puts clinical insight within

More information

TTK DIAGNOSTIC MONITORS

TTK DIAGNOSTIC MONITORS TTK DIAGNOSTIC MONITORS Greyscale and colour diagnostic displays Resolution of 2 to 15 megapixels with ISD Five-year diagnostic guarantee Calibrated automatically in the network TOTOKU DIAGNOSTIC MONITORS

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Monitor Quality Control Solutions

Monitor Quality Control Solutions Monitor Quality Control Solutions Managing the quality of monitors used every day leads to the improvement of the quality of medical care itself. Detecting changes in a monitor s display quality Controlling

More information

Technical Paper DENTAL MONITOR CALIBRATION

Technical Paper DENTAL MONITOR CALIBRATION Technical Paper DENTAL MONITOR CALIBRATION A REPORT ON DENTAL IMAGE PRESENTATION By Tom Schulte Dental radiographs and oral photographs are often both viewed on the same dental workstation monitor. The

More information

Guidelines On Teleradiology In Malaysia

Guidelines On Teleradiology In Malaysia 1 The following are Teleradiology Standards prepared by our College (MRS) and submitted to the Ministry of Health. The Ministry of Health has not endorsed these standards yet. However, we hope they will

More information

Barco medical display systems. Images you can trust. Every time. Everywhere.

Barco medical display systems. Images you can trust. Every time. Everywhere. Barco medical display systems Images you can trust. Every time. Everywhere. In healthcare imaging, accuracy is of vital importance. As a medical professional, you count on display technology that delivers

More information

THE ROYAL COLLEGE OF RADIOLOGISTS

THE ROYAL COLLEGE OF RADIOLOGISTS THE ROYAL COLLEGE OF RADIOLOGISTS TEL: 020-7636-4432 FAX: 020-7323-3100 38 PORTLAND PLACE LONDON W1B 1JQ BFCR(06)4 (updated February 2008) Retention and Storage of Images and Radiological Patient Data

More information

Recommendations for cross-sectional imaging in cancer management, Second edition

Recommendations for cross-sectional imaging in cancer management, Second edition www.rcr.ac.uk Recommendations for cross-sectional imaging in cancer management, Second edition Breast cancer Faculty of Clinical Radiology www.rcr.ac.uk Contents Breast cancer 2 Clinical background 2 Who

More information

Sustainable future for diagnostic radiology: working for alternative and/or multiple providers

Sustainable future for diagnostic radiology: working for alternative and/or multiple providers Sustainable future for diagnostic radiology: working for alternative and/or multiple providers www.rcr.ac.uk 2 Contents Key points 3 Introduction 4 Setting up as a service provider either independently

More information

510(K) Summary K102148

510(K) Summary K102148 510(K) Summary K102148 Moksha Digital Software PVT Limited #161 1, Janardhana Towers, 7th Cross, 19th Main, Sector 1, HSR Layout, Bangalore 560034, India AG1721 Phone:±+91 (80) 4110 1208AU 1720 Fax: +91

More information

DICOM Digital Imaging and Communications in Medicine

DICOM Digital Imaging and Communications in Medicine DICOM Digital Imaging and Communications in Medicine INF5081 / Vårsemester 2007 Wolfgang Leister Reidar Øksnevad Pedro Cruz G. April 2007 PACS PACS = Picture Archive and Communication System Archive for

More information

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen

More information

What s New in DICOM. DICOM Workshop @ SPIE MEDICAL IMAGING February 2010. Bas Revet, Philips Healthcare (Chair WG 6)

What s New in DICOM. DICOM Workshop @ SPIE MEDICAL IMAGING February 2010. Bas Revet, Philips Healthcare (Chair WG 6) What s New in DICOM DICOM Workshop @ SPIE MEDICAL IMAGING February 2010 Bas Revet, Philips Healthcare (Chair WG 6) DICOM is the International Standard for Medical Imaging and related information: images,

More information

A PACS-Aware DICOM Image Object

A PACS-Aware DICOM Image Object Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry

More information

Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display

Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display Marco Eichelberg a, Jörg Riesmeier a, Klaus Kleber b, Jörg Holstein b, Herman Oosterwijk c,

More information

Measuring Size and Distance in Digital Mammography

Measuring Size and Distance in Digital Mammography On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval

More information

Monitor Qualit y Control Solutions

Monitor Qualit y Control Solutions Monitor Qualit y Control Solutions Controlling the Quality of Monitors Managing the quality of monitors used every day leads to the improvement of the quality of medical care itself. Q1 A1 Do all monitors

More information

Workstation, Server Hosted

Workstation, Server Hosted Workstation, Server Hosted With so many PACS out there, Why just make another? So, we didn't. Distribute, Burn, Print, Import, Archive The Internet DICOM Eye Web Web Viewer PACStoGo DICOM Distribution

More information

ACR AAPM SIIM TECHNICAL STANDARD FOR ELECTRONIC PRACTICE OF MEDICAL IMAGING

ACR AAPM SIIM TECHNICAL STANDARD FOR ELECTRONIC PRACTICE OF MEDICAL IMAGING The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

Digital Mammography Monitors

Digital Mammography Monitors Digital Mammography Monitors Film to Filmless Mammography Diagnosis As the incidence of breast cancer rises, regular mammograms are recommended for early detection. It is vital in the process of early

More information

Supporting information for appraisal and revalidation: guidance for clinical oncology and clinical radiology, Second edition

Supporting information for appraisal and revalidation: guidance for clinical oncology and clinical radiology, Second edition www.rcr.ac.uk Supporting information for appraisal and revalidation: guidance for clinical oncology and clinical radiology, Second edition Based on the Academy of Medical Royal Colleges and Faculties Core

More information

Fast mobile reading on the go November 2013. syngo.via WebViewer. Unrestricted Siemens AG 2013 All rights reserved.

Fast mobile reading on the go November 2013. syngo.via WebViewer. Unrestricted Siemens AG 2013 All rights reserved. Fast mobile reading on the go November 2013 syngo.via WebViewer Answers for life. syngo.via WebViewer The mobile extension of syngo.via syngo.via WebViewer 1, enables mobile reading of images with brilliant

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0. DICOM Conformance Statement

CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0. DICOM Conformance Statement CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0 DICOM Conformance Statement Publication # 8G8913 22 April 2009 Table of Contents Table of Contents...2 1 Introduction...5 1.1 Abbreviations

More information

DR Retrofit Kit 1417 WL Configuration Integrated PACS

DR Retrofit Kit 1417 WL Configuration Integrated PACS DR Retrofit Kit 1417 WL Configuration Integrated PACS Description: The DR Retrofit Kit 1417WL - Configuration Integrated PACS makes the use of CR Cassettes/Imaging Plates or the use of analogue x-ray films,

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

GE Healthcare. Centricity PACS and PACS-IW with Universal Viewer* Where it all comes together

GE Healthcare. Centricity PACS and PACS-IW with Universal Viewer* Where it all comes together GE Healthcare Centricity PACS and PACS-IW with Universal Viewer* Where it all comes together The healthcare industry is going through an unprecedented period of change with providers being called upon

More information

DryView CHROMA Imaging System

DryView CHROMA Imaging System DryView CHROMA Imaging System VERSATILE PRINTING ON FILM OR PAPER One of the most trusted names in medical printing now offers a highly versatile printer that empowers physicians and other healthcare professionals

More information

Twelve. Figure 12.1: 3D Curved MPR Viewer Window

Twelve. Figure 12.1: 3D Curved MPR Viewer Window Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from

More information

Quality Assurance Guidance

Quality Assurance Guidance Quality Assurance Guidance Digitisation and Disposal of Patient Records Policy Toolkit December 2014 Effective: December 2014 TITLE: QUALITY ASSURANCE GUIDANCE 1. BACKGROUND WA Health is reliant on access

More information

Teleradiology: The Present Perspective

Teleradiology: The Present Perspective Teleradiology: The Present Perspective P. K. Bhuyan Department of Radio-Diagnosis, MIMER Medical College, Talegaon Dabhade, Pune 410507. Introduction Telemedicine is a broad term encompassing all methods,

More information

Department of Veterans Affairs VHA DIRECTIVE 2011-005 Veterans Health Administration Washington, DC 20420 February 8, 2011

Department of Veterans Affairs VHA DIRECTIVE 2011-005 Veterans Health Administration Washington, DC 20420 February 8, 2011 Department of Veterans Affairs VHA DIRECTIVE 2011-005 Veterans Health Administration Washington, DC 20420 RADIOLOGY PICTURE ARCHIVING AND COMMUNICATION SYSTEMS (PACS) 1. PURPOSE: This Veterans Health Administration

More information

Version 8 DICOM Conformance Statement. Version 3.04, September 2014

Version 8 DICOM Conformance Statement. Version 3.04, September 2014 Version 8 DICOM Conformance Statement Version 3.04, September 2014 1 Conformance Statement Overview The application described in this Conformance Statement VEPRO EMR Manager PACS is a collection of processes

More information

REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) TERMINOLOGY NEEDED TO UNDERSTAND MONITOR PERFORMANCE

REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) TERMINOLOGY NEEDED TO UNDERSTAND MONITOR PERFORMANCE REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) By Ken Compton and Herman Oosterwijk The question of what monitor to use for diagnostic purposes with digital medical images has been a recurring theme

More information

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals

More information

Mobile Healthcare Security Whitepaper

Mobile Healthcare Security Whitepaper Mobile Healthcare Security Whitepaper aycan OsiriX PRO, aycan mobile and Apple's ipad as DICOM image distribution enhancement for PACS environments. Copyright aycan Digitalsysteme GmbH 2013 ayc_mobile_healthcare_security_whitepaper_130529_sp_ds.

More information

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Guidelines on education and training for digital radiology Author:

More information

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

ZEBRAFISH SYSTEMS SYSTEMS ZEBRAFISH. Fast, automatic analyses

ZEBRAFISH SYSTEMS SYSTEMS ZEBRAFISH. Fast, automatic analyses ZEBRAFISH SYSTEMS ZEBRAFISH SYSTEMS Fast, automatic analyses Wide-scale evaluation possibilities ( averaged ECG, QT, QTv, HRV) Distortion-free ECG signal transfer Social interaction examinations Measurements

More information

Health Management Information Systems: Medical Imaging Systems. Slide 1 Welcome to Health Management Information Systems, Medical Imaging Systems.

Health Management Information Systems: Medical Imaging Systems. Slide 1 Welcome to Health Management Information Systems, Medical Imaging Systems. Health Management Information Systems: Medical Imaging Systems Lecture 8 Audio Transcript Slide 1 Welcome to Health Management Information Systems, Medical Imaging Systems. The component, Health Management

More information

CORPORATE SCANNING POLICY

CORPORATE SCANNING POLICY STAFFORDSHIRE COUNTY COUNCIL CORPORATE SCANNING POLICY Document Management Document Disclaimer This document is issued only for the purpose for which it is supplied. Document Owner This document is produced

More information

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

STAMFORD HOSPITAL DEPARTMENT OF RADIOLOGY RULES AND REGULATIONS. Preamble

STAMFORD HOSPITAL DEPARTMENT OF RADIOLOGY RULES AND REGULATIONS. Preamble STAMFORD HOSPITAL DEPARTMENT OF RADIOLOGY RULES AND REGULATIONS Preamble Stamford Hospital and its radiology staff shall maintain radiological facilities and services sufficient to meet the needs of the

More information

Workshop: Defining the Medical Imaging Requirements for a Health Center. Teleradiology and Networking

Workshop: Defining the Medical Imaging Requirements for a Health Center. Teleradiology and Networking Workshop: Defining the Medical Imaging Requirements for a Health Center April 17 2011 E-health and Telemedicine ehealth is the use, in the health sector, of digital data - transmitted, stored and retrieved

More information

White Paper. Study to Demonstrate Efficacy of 8MP Color Display - Mammography 1.INTRODUCTION... 2 2.PURPOSE... 2 3.METHODS... 2 4.RESULTS...

White Paper. Study to Demonstrate Efficacy of 8MP Color Display - Mammography 1.INTRODUCTION... 2 2.PURPOSE... 2 3.METHODS... 2 4.RESULTS... White Paper Study to Demonstrate Efficacy of 8MP Color Display - Mammography CONTENTS 1.INTRODUCTION... 2 2.PURPOSE... 2 3.METHODS... 2 4.RESULTS... 3 5.CONCLUSIONS... 5 6.REPORT FROM DR. KRUPINSKI, UNIVERSITY

More information

GENERAL INFORMATION BROCHURE FOR ACCREDITATION OF MEDICAL IMAGING SERVICES

GENERAL INFORMATION BROCHURE FOR ACCREDITATION OF MEDICAL IMAGING SERVICES GENERAL INFORMATION BROCHURE FOR ACCREDITATION OF MEDICAL IMAGING SERVICES 2010 Page 1 Introduction to Accreditation Program for Medical Imaging Services Definition of Medical Imaging Services (MIS) Medical

More information

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950 HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple

More information

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Routine Ultrasound Equipment Tests for Quality Assurance. Contents Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI,

More information

Medical Imaging Display Colour Space (mrgb) Teleconference 17 July 2014 15:00 (UK) / 10:00 (EST)

Medical Imaging Display Colour Space (mrgb) Teleconference 17 July 2014 15:00 (UK) / 10:00 (EST) Medical Imaging Display Colour Space (mrgb) Teleconference 17 July 2014 15:00 (UK) / 10:00 (EST) The meeting was called to order at 10:00 am (EST) by Craig Revie, acting chair, with the following attendees:

More information

DICOM Conformance Statement For Diagnostic Review Workstation Software Version 5-x MAN-00546

DICOM Conformance Statement For Diagnostic Review Workstation Software Version 5-x MAN-00546 DICOM Conformance Statement For Diagnostic Review Workstation Software Version 5-x MAN-00546 SecurView DX DICOM Conformance Statement Part Number MAN-00546 Revision 001 For Software version 5-X Service

More information

A technical overview of the Fuel3D system.

A technical overview of the Fuel3D system. A technical overview of the Fuel3D system. Contents Introduction 3 How does Fuel3D actually work? 4 Photometric imaging for high-resolution surface detail 4 Optical localization to track movement during

More information

anatomage table Interactive anatomy study table

anatomage table Interactive anatomy study table anatomage table Interactive anatomy study table Anatomage offers a unique, life-size interactive anatomy visualization table for the medical community. Anatomage Table offers an unprecedented realistic

More information

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl Cirrus 0.2T MRI for Everyone North America, Asia, Europe contact: kturek@mri-tech.pl MRI-TECH inc. Cirrus MRI system for all your needs: Low costs Low maintenance High quality Open geometry Imaging of

More information

Gear View Basic. User Help. Version 2.2.0. Written by: Product Documentation and R&D Date: August 2015 LX-DOC-GVB-2.2.0-UH-EN-REVA

Gear View Basic. User Help. Version 2.2.0. Written by: Product Documentation and R&D Date: August 2015 LX-DOC-GVB-2.2.0-UH-EN-REVA Gear View Basic User Help Version 2.2.0 Written by: Product Documentation and R&D Date: August 2015 2015 Lexmark International Technology, S.A. All rights reserved. No part of this publication may be reproduced,

More information

PARCA Certified PACS Interface Analyst (CPIA) Requirements

PARCA Certified PACS Interface Analyst (CPIA) Requirements PARCA Certified PACS Interface Analyst (CPIA) Requirements Copyright notice: Copyright 2005 PACS Administrators in Radiology Certification Association (PARCA). All rights reserved. All rights reserved.

More information

Intelligent cloud solution for telemedicine and archiving

Intelligent cloud solution for telemedicine and archiving Intelligent cloud solution for telemedicine and archiving Cloud solution for images and diagnostics - The cost efficient and safe service for images and diagnostic findings Image generation Automatic archiving

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement CARESTREAM Vue PACS v11.3.2.4 Lesion Management and Bookmark Applications Part # 6K4188 2012-07-02 PAGE 1 of 142 Table of Contents 1 INTRODUCTION... 4 1.1 ABBREVIATIONS AND

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

Final Report. Quality Use of Diagnostic Imaging

Final Report. Quality Use of Diagnostic Imaging Final Report Quality Use of Diagnostic Imaging QS7(ii) Establish technical standards for accreditation requirements for clinical teleradiology Matthew Swain and Paul O Keefe COPYRIGHT NOTICE: This report

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Changing The World One Hospital At a Time

Changing The World One Hospital At a Time Changing The World One Hospital At a Time miplatform Integrated Medical Image and Information Management Platform PACS/RIS System Mobile Imaging System Image-based Conferencing System TeleMed/TeleRad System

More information

George W. Seeley, Ph.D.*t. Kevin MCNeiII, M.S.t. William J. Dallas, ph.d.*t. Past

George W. Seeley, Ph.D.*t. Kevin MCNeiII, M.S.t. William J. Dallas, ph.d.*t. Past RathoGphlcsdexten Displays in radiology: Curnutafiveindoxterms: Past, present, a nd future George W. Seeley, Ph.D.*t Kevin MCNeiII, M.S.t William J. Dallas, ph.d.*t Past. Almost immediately after Wilhelm

More information

MERGE PACS : PRODUCT ROADMAP. Dennis Roscoe, Ph.D., Director PACS Solution Management, Merge Healthcare Tuesday, August 28, 2012

MERGE PACS : PRODUCT ROADMAP. Dennis Roscoe, Ph.D., Director PACS Solution Management, Merge Healthcare Tuesday, August 28, 2012 MERGE PACS : PRODUCT ROADMAP Dennis Roscoe, Ph.D., Director PACS Solution Management, Merge Healthcare Tuesday, August 28, 2012 agenda 2 Merge PACS Roadmap End-to-End Total Solution Questions MERGE PACSROADMAP

More information

Telemedicine, EMR & Remote Connectivity

Telemedicine, EMR & Remote Connectivity Telemedicine, EMR & Remote Connectivity Saroj Mishra, MS, FACS Prof & Head, Dept. of Endocrine Surgery & Faculty I/C, SGPGI Telemedicine Program Sanjay Gandhi Postgraduate Institute of Medical Sciences

More information

KIP 3000 Series MULTIFUNCTION SIMPLICITY

KIP 3000 Series MULTIFUNCTION SIMPLICITY KIP 3000 Series MULTIFUNCTION SIMPLICITY KIP 3000 SERIES SYSTEMS The KIP 3000 is a highly hl advanced d wide format digital copy system that provides the ability to easily add network printing and scan-to-file

More information

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Performance evaluation and quality assurance of Varian enhanced dynamic wedges JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department

More information

Clarity the clear solution for managing digital medical imaging data

Clarity the clear solution for managing digital medical imaging data Clarity the clear solution for managing digital medical imaging data Daniel J. Valentino, Ph.D. Chief Scientist and Chief Technology Officer, icrco, Inc. dvalentino at icrcompany.com August 7, 2009 Abstract

More information

Standard-Compliant Streaming of Images in Electronic Health Records

Standard-Compliant Streaming of Images in Electronic Health Records WHITE PAPER Standard-Compliant Streaming of Images in Electronic Health Records Combining JPIP streaming and WADO within the XDS-I framework 03.09 Copyright 2010 Aware, Inc. All Rights Reserved. No part

More information

NHS Imaging and Radiodiagnostic activity in England. 2012/13 Release. August 2013

NHS Imaging and Radiodiagnostic activity in England. 2012/13 Release. August 2013 NHS Imaging and Radiodiagnostic activity in England 2012/13 Release August 2013 Commentary This National Statistics release covers Imaging and Radiodiagnostic examinations or tests carried out in the NHS

More information

The Convenient and Compact CR System for Veterinary Practices

The Convenient and Compact CR System for Veterinary Practices The Convenient and Compact CR System for Veterinary Practices Fire CR+ 20 / 40 / 60 / 80 vet with a throughput of 20 to 70 cassettes per hour and the option of a space - saving wall mount The small and

More information

REGIUS History. Touch REGIUS110. REGIUS170 ImagePilot REGIUS150 REGIUS330 CS-1

REGIUS History. Touch REGIUS110. REGIUS170 ImagePilot REGIUS150 REGIUS330 CS-1 Konica Minolta has been a CR technology leader since 1996, consistently designing, developing, and delivering innovative solutions. The REGIUS Σ represents the sum of this experience. REGIUS Σ delivers

More information

Head Office #5 Sugar Creek Road Piedmont, Missouri 63957. Tel: 877-838-9273, option 2, sub-option 8 Fax: 573-223-3551 Dragonfly@avimark.net.

Head Office #5 Sugar Creek Road Piedmont, Missouri 63957. Tel: 877-838-9273, option 2, sub-option 8 Fax: 573-223-3551 Dragonfly@avimark.net. Head Office #5 Sugar Creek Road Piedmont, Missouri 63957 Tel: 877-838-9273, option 2, sub-option 8 Fax: 573-223-3551 Dragonfly@avimark.net. DragonflyPACS Contact Information Dragonfly Specialized Support

More information

Presentation Objective: Benefits to New Employee. Goal for Participants. Benefits to the Administration

Presentation Objective: Benefits to New Employee. Goal for Participants. Benefits to the Administration Imaging Department Quick Step How to Create and Implement a Rapid Integration Orientation New Employee Orientation builds a bridge between the old job and the new job for better quality in health care

More information

Parkway College of Nursing and Allied Health School of Nursing and Allied Health. BSc (Hons) Diagnostic Radiography and Imaging Module Synopses

Parkway College of Nursing and Allied Health School of Nursing and Allied Health. BSc (Hons) Diagnostic Radiography and Imaging Module Synopses Parkway College of Nursing and Allied Health School of Nursing and Allied Health 4FHH1157 Anatomy, Physiology and Pathology for Imaging 1 Develop a broad understanding of anatomy, physiology, pathology

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

Dear Ms. Lucas: This letter corrects our substantially equivalent letter of February 18, 2016.

Dear Ms. Lucas: This letter corrects our substantially equivalent letter of February 18, 2016. DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002 Duerr Dental AG February

More information

Highmark Provider Privileging Requirements

Highmark Provider Privileging Requirements Highmark Provider Privileging Requirements Copyright 2008 by Highmark Inc. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or

More information

Intelligent Tools For A Productive Radiologist Workflow: How Machine Learning Enriches Hanging Protocols

Intelligent Tools For A Productive Radiologist Workflow: How Machine Learning Enriches Hanging Protocols GE Healthcare Intelligent Tools For A Productive Radiologist Workflow: How Machine Learning Enriches Hanging Protocols Authors: Tianyi Wang Information Scientist Machine Learning Lab Software Science &

More information

Universal Viewer Zero Footprint Client 1

Universal Viewer Zero Footprint Client 1 GE Healthcare Universal Viewer Zero Footprint Client 1 Product Data Sheet Helping to improve communication and collaboration Introduction Centricity PACS and Centricity PACS IW with Universal Viewer puts

More information

IPEM guidance on coding Medical Physics and Clinical Engineering staff within ESR for Healthcare Science Leads

IPEM guidance on coding Medical Physics and Clinical Engineering staff within ESR for Healthcare Science Leads IPEM guidance on coding Medical Physics and Clinical Engineering staff within ESR for Healthcare Science Leads Background Workforce Standards at the Health & Social Care Information Centre have recently

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

3D Viewer. user's manual 10017352_2

3D Viewer. user's manual 10017352_2 EN 3D Viewer user's manual 10017352_2 TABLE OF CONTENTS 1 SYSTEM REQUIREMENTS...1 2 STARTING PLANMECA 3D VIEWER...2 3 PLANMECA 3D VIEWER INTRODUCTION...3 3.1 Menu Toolbar... 4 4 EXPLORER...6 4.1 3D Volume

More information

Network Management for Picture Archiving and Communication Systems (PACS)

Network Management for Picture Archiving and Communication Systems (PACS) Network Management for Picture Archiving and Communication Systems (PACS) Edwood Yiu M.Eng. project presentation November 21, 2006 School of Engineering Science Simon Fraser University Road map Purpose

More information

Cedars-Sinai Medical Center:

Cedars-Sinai Medical Center: Cedars-Sinai Medical Center: Delivering Better Patient Care with Infoimaging Technology An IDC Infoimaging Case Study THE SUBJECT Cedars-Sinai Medical Center in Los Angeles, California, is the largest

More information