English Rev 0-07/11. IFU 9419 Final Document
|
|
|
- Reginald Wilkerson
- 9 years ago
- Views:
Transcription
1 INSTRUCTIONS FOR USE ZIMMER ANGLED TAPERED ABUTMENTS FOR TAPERED SCREW-VENT AND SCREW-VENT IMPLANT SYSTEMS Before using a product, the operating surgeon/practitioner in charge should carefully study the indications, contraindications, recommendations, warnings and instructions, as well as all other product-specific information (technical product description, description of the surgical and restorative technique, catalogue sheet, etc.) and fully comply with them. Detailed instructions over and above those contained in this instruction for use concerning the possible combinations, productspecific risks, preparatory steps, indications and contraindications, etc. can be found in the description of the surgical technique, in the technical description of the product or on the appropriate catalogue sheet. Zimmer also recommends attending the appropriate user-training courses. The aforementioned documents and details of the training courses may be obtained from the appropriate representatives in the various countries. The manufacturer, the importer and the suppliers of Zimmer products are not liable for complications, other negative effects or damages that might occur for reasons such as incorrect indications or surgical technique, unsuitable choice of material or handling thereof, unsuitable use or handling of the instruments, asepsis and so on. The operating surgeon is responsible for any such complications or other consequences. It is also the operating surgeon's responsibility to properly instruct and inform the patient on the functions, handling and necessary care of the product and on all known product risks. DESCRIPTION The Angled Tapered abutment has a hex which engages the internal hex of the Tapered Screw-Vent and Screw-Vent implant. The abutments are available in multiple cuff heights and in both 15 and 30 configurations to provide angulation correction for off-angle implant placement. The abutment is secured to the implant with an abutment retaining screw which is preassembled in the abutment. The abutment screw is not removable from the abutment. The abutment has a tapered cone with an internal screw access for the attachment of various restorative components using a separate coping screw. The abutment cone has a 15 taper and is 1.2mm tall. Abutments are packaged with a metal delivery tool to assist with the placement and orientation of the abutment into the implant. The abutment and abutment retaining screw are fabricated from titanium alloy. INDICATIONS The Angled Tapered Abutment is used for a terminal or intermediate abutment for screw-retained multiple-unit restorations. The 30 Angled Tapered Abutment must be used within 45 of parallelism for a splinted restoration. The 15 Angled Tapered Abutment must be used within 30 of parallelism for a splinted restoration. CONTRAINDICATIONS Not for use as a single tooth restoration, with a cemented prosthesis, or with limited interocclusal space less than 7.5mm. Not for use when implants are divergent greater than 45 with the 30 Angled Tapered Abutment or when implants are divergent greater than 30 with the 15 Angled Tapered Abutment. The Angled Tapered Abutment may not be prepped. The Angled Tapered Abutment is not for use with the 3.3mm Screw-Vent Implant. implants should not be placed if there is an insufficient volume of alveolar bone to minimally support the implant (minimum 1mm circumferential and 2mm apical). Implants placed in the maxilla should not perforate the sinus floor membrane. Poor bone quality, poor patient oral hygiene, heavy tobacco use, uncontrolled systematic diseases (diabetes, etc.), reduced immunity, alcoholism, drug addiction, and psychological instability may contribute to lack of integration and/or subsequent implant failure. Severe bruxism, clenching, and overloading, may cause bone loss, screw loosening, component fracture, and/or implant failure. Exposure to radiation and chemotherapy may impact health of the implant. Dental implant patients should be instructed to consult with their physician prior to undergoing such treatment options. WARNINGS Surgical and restorative techniques required to place dental implants are highly specialized and complex procedures. Practitioners should attend courses of study to familiarize themselves with implantology techniques. Improper technique can cause bone loss and implant failure. implant systems are intended to be used only with specially designed bone drills and prosthetics. Implants placed at severe angles relative to existing dentition or multiple implants placed at convergent/divergent manner can result in complex restorations that may overload implants, potentially leading to implant failure. A thorough diagnostic work-up and use of surgical templates are recommended to help ensure proper angulation. Other relative contraindications include steroid and anticoagulant treatment which may affect the surgical site, surrounding tissue, or patient s healing function.exposure to long-term use of bisphosphonate drugs especially with chemotherapy may impact implant survival. Careful patient selection including consultation with the attending physician is strongly recommended prior to implant treatment.excessive mobility, bone loss, or infection may indicate the implant is failing. Any implant which appears to be failing should be treated or removed as soon as possible. If removal is necessary, curette any soft tissue from the implant site and allow site to heal as though it were an atraumatic extraction.due to the metal conductivity, electrosurgery around the implants and intraoral abutment preparations without irrigation could result in tissue damage and implant failure. Patients should consult with their physician and imaging technician prior to undergoing an MRI procedure. The Angled Tapered Abutment has not been evaluated for safety and compatibility in the MR environment. The Angled Tapered Abutment has not been tested for heating or migration in the MR environment. PRECAUTIONS Proper case planning is essential to the long-term success of both the prosthesis and the implant. Overload is one of the key contributors to implant failure. Ensure the implant size and abutment angulation are appropriate for the occlusal load. Highly angulated abutments (>30 ) should be avoided. Off Axis Loading if implants are placed up to 45 off axis then it may be necessary to splint(join the restoration) the divergent implants.
2 BREAKAGE Implant and abutment fractures can occur when applied loads exceed the normal functional design tolerances of the implant components. Potential overloading conditions may result from deficiencies in implant numbers, lengths and/or diameters to adequately support a restoration, excessive cantilever length, incomplete abutment seating, abutment angles greater than 30 degrees, occlusal interferences causing excessive lateral forces, patient parafunction (e.g., bruxing clenching), improper casting procedures, inadequate prosthesis fit, and physical trauma. CHANGES IN PERFORMANCE It is the responsibility of the clinician to instruct the patient on all appropriate contraindications, side effects, and precautions as well as the need to seek the services of a trained dental professional if there are any changes in the performance of the implant (e.g., looseness of the prosthesis, infection or exudate around the implant, pain, or any other unusual symptoms that the patient has not been told to expect). HYGIENE & MAINTENANCE Long-term implant health is directly related to the maintenance of oral hygiene. Potential implant candidates should establish an adequate oral hygiene regimen prior to implant therapy. Following implant placement, the clinician should instruct the patient on proper tools and techniques to ensure long-term maintenance of the implant(s). The patient should also be instructed to maintain routinely scheduled prophylaxis and evaluation appointments. TREATMENT PLANNING Appropriate imaging techniques should be used to determine if adequate bone is available, and to determine the location of important anatomical landmarks, such as the mandibular canal, maxillary sinuses and adjacent teeth. Thorough clinical evaluation is imperative prior to all implant surgeries. GENERAL CONSIDERATIONS Control of biomechanical stresses is the key factor to long-term success of the prosthesis. Even after implant integration, imbalances in occlusal forces can lead to implant failure. Implant patients should be monitored for signs of screw loosening, periimplant bone loss and tooth wear as signs of occlusal overloading. ADVERSE EFFECTS The following complications may occur relative to implant placement: pain, discomfort, dehiscence, delayed healing, paresthesia, hyperesthesia, edema, hemorrhage, hematoma, infection, inflammation, local and generalized allergic reaction, lack of integration, loss of bone, and loss of implant. Other adverse effects may also occur as a result of iatrogenic factors and host responses. STERILITY All implants have been gamma radiation sterilized and are for single use only. Do not resterilize implants. prosthetic and ancillary components are sold sterile or non-sterile. Refer to the specific packaging for verification of sterility. Sterilize non-sterile product prior to use in patients. SINGLE USE Reuse of a single use device that has come in contact with blood, bone, tissue or other body fluids may lead to patient or user injury. Possible risks associated with reuse of a single use device include, but are not limited to, mechanical failure and transmission of infectious agents. SHELF LIFE The product expiration date is indicated by the hourglass symbol on the product label, followed by the year and month of expiration. Caution: Do not use sterile devices if the packaging providing the sterile barrier, including the outer cap, vial, or tray has been damaged or compromised in any manner (i.e. cracked or crushed). PRODUCT PACKAGING All implants have been cleaned, packaged in double vials within an environmentally controlled room, and sterilized for convenience and immediate use. The implants are suspended on a carrier for transfer to the prepared surgical site without risk of contact contamination. Both the implant and the inner vial packaging are sterile. The label on the outer vial packaging for each implant contains a lot number that should be recorded in the patient s file to ensure complete traceability of the product. Prosthetic components provided in sealed plastic vials or Tyvek * sealed blister packages are also pre-cleaned and sterilized for your convenience. CLEANING/STERILIZATION INFORMATION Disinfection and sterilization procedures should conform to OSHA or local guidelines for blood borne pathogens. Clinically contaminated implants should not be cleaned and resterilized under any circumstances. CLEANING Use the following guidelines for cleaning products: Drills, Instruments and Components - Disassemble multi-piece components, if applicable. Rinse with cool-to-lukewarm water for two-and-one-half minutes. For Drills, use the Zimmer cleaning wire to remove any debris from the irrigation channel. Using a 25 gauge needle, flush the drill lumen with water to remove any remaining debris. For all parts place in an ultrasonic cleaner with an enzymatic detergent diluted with tap water per the manufacture s guidelines. Sonicate for 10 minutes. Rinse with tap water for three minutes. Kits, Trays, and Blocks - Remove all parts and insert from the tray. Clean parts per the above instructions. Thoroughly rinse the kits under running tap water to remove all visible soil. Use a soft bristle brush to clean the kits until all visible soil is removed. A syringe or pipe cleaner may be used to aid in the rinsing. Assure that all hard to reach areas are accessed. After the rinsing, prepare the enzymatic detergent following the manufacturer s specifications. Fully immerse the kit in the prepared
3 detergent and allow the kit to soak in the detergent for a minimum of one minute. Following the soak use a damp cloth and/or a soft bristle brush to wipe and remove any excess debris/soil from each component. A syringe or a pipe cleaner may be used to aid in the cleaning. Rinse the kits with lukewarm tap water to eliminate all residual enzymes and detergent, thoroughly for a minimum of three minutes. Dry the components. Reassemble the contents of the kit and follow the guidelines for sterilization. Note: This procedure should be performed after an instrument used during a surgery comes into contact with the surgical tray or prosthetic tray. STERILIZATION Individual parts should be placed in appropriate autoclave or dry heat pouch prior to sterilization. When sterilizing parts within a kit, parts should be placed in appropriate locations and kit should be wrapped in sterilization wrap. The following sterilization parameters (method, time and temperature) are required to achieve a 10-6 sterility assurance level (SAL). Local or national specifications should be followed where steam sterilization requirements are stricter or more conservative than those listed in the table. Exceeding these sterilization parameters may result in damage to plastic components. Verify the calibration of your unit to ensure recommended temperatures are not being exceeded. To ensure autoclave is performing effectively, periodic use of biologic indicators should be considered. Chemclave sterilization is NOT recommended for any products. Parts individually pouched Cycle Type Temperature Exposure Time Gravity 121 C 250 F Dry Time (only for kits) 40 minutes n/a Gravity 121 C 250 F 80 minutes 30 minutes Parts individually pouched or in kit Pre-vacuum Pre-vacuum Dry Heat 132 C 270 F 134 C 273 F 160 C 320 F 3 minutes 30 minutes 18 minutes 30 minutes 120 minutes n/a TECHNIQUE INFORMATION Procedure for Angled Tapered Abutments Note: During implant placement, it is recommended to orient the flat of the internal hex of the implant to be opposite the angle correction. The pre-attached multi-purpose fixture mount can be used to index the internal hex of the implant. The flat side on the wall of the fixture mount will line up with the flat side of the internal hex. Note: Prior to delivering the abutment to the mouth with the abutment delivery tool, the delivery tool should be hand tightened to the abutment to confirm adequate attachment of the tool to the abutment. Note: Use appropriate abutments and ancillary components that correspond to the implant system being restored. 1. Remove the Angled Tapered Abutment from the abutment packaging over a sterile field. Hand tighten the abutment delivery tool to confirm attachment to the cone of the abutment. 2. Thread dental floss thru the floss hole in the delivery tool and tie. Utilizing the abutment delivery tool, deliver the abutment to the mouth, aligning the angled abutment in the appropriate orientation for desired angulation correction. 3. Use a 1.25mm (0.50 ) Hex Driver to hand tighten the abutment retaining screw. A contra angle hand piece with a 1.25mmD latch-lock driver may also be used for initial delivery. The long latch-lock driver (part# HXL1.25D) must be used if the abutment delivery tool is attached to the abutment. The standard latch-lock driver (part# HX1.25D) can be used if the abutment delivery tool is removed from the abutment. 4. Verify with periapical radiograph that the abutment is seated completely into the implant and has engaged the internal hexagon. 5. Tighten the abutment retaining screw to 30 Ncm with a calibrated torque wrench. If using the Zimmer Restorative Torque Wrench (part# TWR), the long Torque Wrench Insert Hex Driver, (part# TW1.25L) must be used if the abutment delivery tool is attached to the abutment. The short Torque Wrench Insert Hex Driver (part# TW1.25) can be used if the abutment delivery tool is removed from the abutment.
4 6. If the abutments will not be immediately restored with a provisional or final restoration, it t is recommended to place the Tapered Abutment Titanium Healing Cap (part# TATHC) to prevent irritation of the soft tissue and to prevent the ingress of material in the screw access of the abutment cone. Labeling Symbols Symbols may be used on some international package labeling for easy identification. Symbols Use of Symbols Symbol for European Conformity 0197 Symbol for Angled Abutment Cuff Height Symbol for Do Not Reuse Symbol for Attention see instructions for use Symbol for Sterile by gamma irradiation Symbol for Use By Symbol for Hexagon Symbol for Use by Prescription only Authorized Representative in the European Community CONTACT AND ORDERING INFORMATION USA: Inc Aston Avenue Carlsbad, CA France SAS 2,Place Gustave Eiffel Rungis Cedex Spain: Ibérica S.L.U. C/Verge del Pilar 49,local Cardedeu Tel: , France Tel: +33-(0)1-45- (Barcelona) Spain Tel: Fax: Fax: +33-(0)1-45- Fax: Canada: Corp. Germany: 2323 Argentia Road Mississauga, Ontario, L5N 5N3 Tel: GmbH Italy: Wentzinger Strasse 23 Viale Matteotti Vittorio Veneto D Freiburg (TV) Tel: / Italy Fax: / 15 Tel:
5 Fax: Fax: English Rev 0-07/11 Australia: Israel: Ltd. Units 1-2, 1 Skyline Place 143 Bialik Street Frenchs Forest NSW 2086 Ramat Gan Tel: +61(0) or Tel: +972-(0) (NZ) Fax: +972-(0) MAH-TSV & SP: Fax: Implatex Co. Ltd YDM Nippori Bldg MAH- Spline & AdVent : Hakuho Corporation Kojimachi 1, 2 chome Nishi Nippori Chome Sanbanchi Bldg. 202 Arakawa-ku, Tokyo Kojimachi , Japan Chiyoda-ku, Tokyo Tel: Fax: Japan Tel: Fax: ** Tyvek Pouch is a registered trademark of E.I. du Pont de Nemours and Company. All other names and logos, and trademarks referred to within this package insert are the property of Zimmer Inc. or its affiliates. Rev. 0-07/11 (and date of printing) Printed in U.S.A. (if applicable) 2011 Zimmer, Inc. All rights reserved. (Reflects date of current revision and first release of insert/ifu) X-XXX-XX or D xxx 0197 The CE mark is valid only if it is also printed on the product label.
THE LOCATOR OVERDENTURE IMPLANT SYSTEM.
THE LOCATOR OVERDENTURE IMPLANT SYSTEM. FOUR DECADES OF ATTACHMENT KNOWLEDGE INCORPORATED INTO NARROW DIAMETER OVERDENTURE IMPLANTS. The LOCATOR Overdenture Implant System (LODI) is comprised of 2.4mm
One Abutment - One Time
One Abutment - One Time A new treatment concept simple, innovative, easy to implement. Something so simple that does so much. What makes One Abutment - One Time a favorite among clinicians and patients
Product Catalog. ERA Mini TM Dental Implant System
Product Catalog ERA Mini TM Dental System How to Order Return Procedure To order, call Zimmer Dental Customer Service between 6:00 a.m. and 4:30 p.m. PST, Monday through Friday, or fax your order to 1
screw-retained crown using a PEEK temporary abutment
screw-retained crown using a PEEK temporary abutment This prosthetic technique module may contain references to the complete Prosthetic Manual (L02015). To download the full Prosthetic Manual, please visit
Guidance for implant removal. Straumann Dental Implant System
Guidance for implant removal Straumann Dental Implant System The ITI (International Team for Implantology) is academic partner of Institut Straumann AG in the areas of research and education. ContentS
IMPLANT CONSENT FORM WHAT ARE DENTAL IMPLANTS?
IMPLANT CONSENT FORM WHAT ARE DENTAL IMPLANTS? Dental implants are a very successful and accepted treatment option to replace lost or missing teeth. A dental implant is essentially an artificial tooth
Taking the Mystique out of Implant Dentistry. Dr. Michael Weinberg B.Sc., DDS, FICOI
Taking the Mystique out of Implant Dentistry Dr. Michael Weinberg B.Sc., DDS, FICOI What is Restorative Implant Dentistry? Restorative implant dentistry involves taking a few simple mechanical principles
Dental Updates. Excerpted Article e-mail: [email protected]. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS
¼ ½ ¾ µ mw/cm 2 Volume 17; 2007 Dental Updates "CUTTING EDGE INFORMATION FOR THE DENTAL PROFESSIONAL " 200 SEMINARS AND 30 JOURNALS REVIEWED YEARLY FOR THE LATEST, CUTTING EDGE INFORMATION Excerpted Article
Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery.
Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction Scout Vessel Guard 2
SAMBA Screw System INSTRUCTIONS FOR USE TABLE OF CONTENTS. Description. Indications. Contraindications. Warnings. Potential Adverse Events
SAMBA Screw System INSTRUCTIONS FOR USE SAMBA Screw System TABLE OF CONTENTS Description Indications Contraindications Warnings Potential Adverse Events How Supplied and Handled Inspection Cleaning Sterilization
Outlines: Original Artwork: LeMaitre. Light Post Adapter. Light Post Adapter. Fits Olympus, Storz, and Stryker Equipment* English Instructions for Use
Outlines: Original Artwork: LeMaitre VASCULAR Fits Olympus, Storz, and Stryker Equipment* English Instructions for Use Fits Olympus, Storz, and Stryker Equipment* Instructions for Use - English Model Numbers
Multi-unit abutment bar overdenture
This prosthetic technique module may contain references to the complete Prosthetic Manual (L02015). To download the full Prosthetic Manual, please visit www.biohorizons.com/prostheticmanual.aspx Use this
screw-retained single crowns using custom-cast abutments
screw-retained single crowns using custom-cast abutments This prosthetic technique module may contain references to the complete Prosthetic Manual (L02015). To download the full Prosthetic Manual, please
NARROW DIAMETER implant
ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball
páêçå~=aéåí~ä=`^al`^j=póëíéã på~åmçëí
kéï=~ë=çñw= MRKOMNP páêçå~=aéåí~ä=`^al`^j=póëíéã på~åmçëí lééê~íáåö=fåëíêìåíáçåë båöäáëü=erpf Operating Instructions = Sirona Dental Systems GmbH Table of contents 1 Symbols used... 3 2 Product description...
BASIC INFORMATION ON THE STRAUMANN VARIOBASE ABUTMENT. Straumann Variobase Abutment
BASIC INFORMATION ON THE STRAUMANN VARIOBASE ABUTMENT Straumann Variobase Abutment 1 The ITI (International Team for Implantology) is academic partner of Institut Straumann AG in the areas of research
Contents. Cement retained restoration. Screw retained restoration. Overdenture retained restoration. TS Implant System. 70 ComOcta Gold Abutment
Contents TS Implant System Cement retained restoration Screw retained restoration 06 Cement-retained bridges with the Solid abutment system (non- 72 Screw retained crown with the ComOcta Gold abutment
RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS
RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education. CONTENTS Product
Standard Internal Hex
Standard Internal Hex Touareg TM -OS Touareg TM -S Swell TM Touareg -S Touareg -OS Swell About ADIN Adin Dental Implant Systems Ltd., designs, manufactures and markets state of the art, technologically
FUSEFORCE. Hand Fixation System SURGIC A L T ECHNIQUE
FUSEFORCE Hand Fixation System SURGIC A L T ECHNIQUE Contents Chapter 1 3 Chapter 2 4 Appendix A 7 Intended Use Surgical Technique Ordering Information Indication Sizing Reference Proper surgical procedures
What Dental Implants Can Do For You!
What Dental Implants Can Do For You! Putting Smiles into Motion About Implants 01. What if a Tooth is Lost and the Area is Left Untreated? 02. Do You Want to Restore Confidence in Your Appearance? 03.
Implants in your Laboratory: Abutment Design
1/2 point CDT documented scientific credit. See Page 41. Implants in your Laboratory: Abutment Design By Leon Hermanides, CDT A patient s anatomical limitations have the greatest predictive value for successful
TRI-CAM Implant System
TRI-CAM Implant System infinite opportunities In Implantology For almost 50 years ACE Surgical has been dedicated to dental surgical advancements. We continue to develop and manufacture the highest quality,
Standard laryngoscope blade assemblies Directions for use
1 2014 Welch Allyn, Inc. MM 721358 Ver. C Standard laryngoscope blade assemblies Directions for use Intended use About this document The rigid laryngoscope is a device used to examine and visualize a patient
Restorative Manual Impression Transfer System
Restorative Manual Impression Transfer System Indirect (closed-tray) transfer technique Implant-level indirect transfers for closed-tray, transfer impression technique Designed to transfer the soft tissue
SCD Case Study. Treatment Considerations for Implant Rehabilitation
SCD Case Study Treatment Considerations for Implant Rehabilitation Multiple surgical and restorative factors play a role in the treatment planning of implant restorations for the edentulous patient (Ali
TABLE OF CONTENTS. Surgical Technique 2. Indications 4. Product Information 5. 1. Patient Positioning and Approach 2
Surgical Technique TABLE OF CONTENTS Surgical Technique 2 1. Patient Positioning and Approach 2 2. Intervertebral Device Implanted 2 3. Buttress Plate Selection 2 4. Awl Insertion 2 5. Screw Insertion
Understanding Dental Implants
Understanding Dental Implants Comfort and Confidence Again A new smile It s no fun when you re missing teeth. You may not feel comfortable eating or speaking. You might even avoid smiling in public. Fortunately,
Straumann Bone Level Tapered Implant Peer-to-peer communication
Straumann Bone Level Tapered Implant Peer-to-peer communication Clinical cases April, 2015 Clinical Cases Case No. Site 1 Single unit; Anterior Maxilla 2 Multi-unit; Anterior Maxilla Implant placement
RomiApex A-15. APEX LOCATOR User Manual
RomiApex A-15 APEX LOCATOR User Manual Table of Contents Introduction..... 2 1. Indications for use... 3 2. Precautions..... 3 3. Adverse Reactions...... 3 4. Package Contents........ 3 5. Getting Started.....
DENT IMPLANT restoring qualit S: of LIfE
DENTAL IMPLANTS: restoring quality of life Dental Implants: A Better Treatment Option. What are dental implants? Dental implants are a safe, esthetic alternative to traditional crowns, bridgework, and
Spedding Dental Clinic. 73 Warwick Road Carlisle CA1 1EB T: 01228 521889 www.speddingdental.co.uk
DENTAL IMPLANTS Spedding Dental Clinic 73 Warwick Road Carlisle CA1 1EB T: 01228 521889 www.speddingdental.co.uk SPEDDING DENTAL CLINIC Jack Spedding is a partner in Spedding dental clinic. He is a highly
Don t Let Life Pass You By Because Of Oral Bone Loss
Don t Let Life Pass You By Because Of Oral Bone Loss Ask For Dental Implant Solutions From BIOMET 3i Scan With Your Smartphone! In order to scan QR codes, your mobile device must have a QR code reader
Ideal treatment of the impaired
RESEARCH IMPLANTS AS ANCHORAGE IN ORTHODONTICS: ACLINICAL CASE REPORT Dale B. Herrero, DDS KEY WORDS External anchorage Pneumatized Often, in dental reconstruction, orthodontics is required for either
GUIDELINES. Educational Requirements & Professional Responsibilities for Implant Dentistry CONTENTS. The Guidelines of the Royal College of
Educational Requirements & Professional GUIDELINES Approved by Council May 2013 This is replacing the document last published in August 2002. Educational Requirements & Professional The Guidelines of the
More than a fixed rehabilitation.
More than a fixed rehabilitation. A reason to smile. In combination with: Patient expectations drive dental treatments for fixed edentulous immediate restorations. Patients today have increasingly high
While the prosthetic rehabilitation of
Restoring Mandibular Single Teeth with the Inclusive Tooth Replacement Solution Go online for in-depth content by Bradley C. Bockhorst, DMD While the prosthetic rehabilitation of full-arch cases provides
What is a dental implant?
What is a dental implant? Today, the preferred method of tooth replacement is a dental implant. They replace missing tooth roots and form a stable foundation for replacement teeth that look, feel and function
ball abutment overdenture: chairside pick-up using existing denture
ball abutment overdenture: chairside pick-up using existing denture This prosthetic technique module may contain references to the complete Prosthetic Manual (L02015). To download the full Prosthetic Manual,
Perkins Statewide Articulation Agreement. Documentation item: Secondary Competency Task List Coversheet
Perkins Statewide Articulation Agreement Documentation item: Secondary Task List Coversheet The Secondary School agrees to: A. Implement the approved PDE Program(s) of Study. B. Provide assessment of student
Implant Bar Overdenture Utilizing Locator Attachments
Utilizing Locator Attachments Step-by-Step Restorative Protocol Implant Bar Overdentures offer a removable implant solution for edentulous patients desiring a stable and esthetic prosthesis that improves
Surgical Manual OSSEOTITE Certain Implant Systems And OSSEOTITE External Hex Implant Systems
Surgical Manual OSSEOTITE Certain Implant Systems And OSSEOTITE External Hex Implant Systems No-Touch Mountless OSSEOTITE Certain Implant Delivery System Surgical Protocol Implant Placement Surgical Protocol
(English) NEXUS SPINE SPACER SYSTEM
(English) NEXUS SPINE SPACER SYSTEM INDICATIONS FOR USE NEXUS Spine Spacer System, a GEO Structure is indicated for use in the thoraco-lumbar spine (i.e., T1 to L5) to replace a diseased vertebral body
4052 Slimplicity Tech final_layout 1 6/29/15 3:29 PM Page 2 Surgical Technique
Surgical Technique TABLE OF CONTENTS Slimplicity Anterior Cervical Plate System Overview 2 Indications 2 Implants 3 Instruments 4 Surgical Technique 6 1. Patient Positioning and Approach 6 2. Plate Selection
páêçå~=aéåí~ä=`^al`^j=póëíéã qá_~ëé
kéï=~ë=çñw MSKOMNS páêçå~=aéåí~ä=`^al`^j=póëíéã qá_~ëé lééê~íáåö=fåëíêìåíáçåë båöäáëü=erpf Operating Instructions = Table of contents Sirona Dental Systems GmbH Table of contents 1 Symbols used... 3 2
PROSTHETIC PROCEDURE. for HG IMPLANT SYSTEM
PROSTHETIC PROCEDURE for HG IMPLANT SYSTEM PROSTHETIC PROCEDURE for HG IMPLANT SYSTEM HG Implant System Contents Cement retained restoration Rigid abutment When abutment reduction is unnecessary When abutment
Prosthodontist s Perspective
Unless otherwise noted, the content of this course material is licensed under a Creative Commons Attribution - Non-Commercial - Share Alike 3.0 License. Copyright 2008, Dr. Jeff Shotwell. The following
Surgical Manual. Preservation By Design. Implant
Surgical Manual Preservation By Design Implant Table Of Contents Important Product Information... 1-2 Preoperative Planning.... 3 Introduction And Treatment Planning... 4 Top-Down Treatment Planning....
Best Practices for Loaner Surgical Instruments
Best Practices for Loaner Surgical Instruments Copyright 2015, Sterile Processing University, LLC, All Rights Reserved. This material may not be copied or used without permission of the author. Nancy Chobin,
INSTALLATION MANUAL. Contents JAPANESE ENGLISH GERMAN FRENCH ITALIAN SPANISH. To the installing technician: CEILING BRACKET
Be sure to read this installation manual thoroughly prior to installation. If installation methods or non-standard parts not specified in this installation manual are used, accidents or injury may result.
prosthetic technique manual
prosthetic technique manual TABLE OF CONTENTS Introduction 1 Treatment Planning 2-5 Restorative Options Implant-level Cement-retained Restorations Implant-level Screw-retained Restorations Abutment-level
procedures & products NOBELESTHETICS including Procera
procedures & products NOBELESTHETICS including Procera First from Nobel Biocare. NOBELPERFECT, (NP, RP, WP), NOBELDIRECT (NP, RP, WP), Brånemark System, NOBELREPLACE and NOBELSPEEDY Implants. A complete
DENTAL IMPLANTS DR JEBIN,MDS.,D.ICOI
Good Morning DENTAL IMPLANTS DR JEBIN,MDS.,D.ICOI What is implant? A dental implant is an artificial root that replaces the natural tooth root. Crown Gum Implant Tooth Root Jawbone Parts of implant Cover
NobelActive. procedures and products
NobelActive procedures and products precautions and warnings Manufacturer: Nobel Biocare AB, Box 5190, SE-402 26 Göteborg, Sweden. Phone: +46 31 81 88 00. Fax: +46 31 16 31 52 www.nobelbiocare.com Important!
BioHorizons Education Programme 2015
BioHorizons Education Programme 2015 SPMP14328GB Rev A November 2014 Contents The Role of Implants in Restorative Dentistry An Introduction to Contemporary Implant Prosthodontics Sinus Elevation Socket
Gl GinGival level implant
GL Gingival Level implant TABLE OF CONTENTS GL - Gingival Level implant Implant characteristics page 04 Dental implant page 05 Prosthetic components page 06 Ball attachment system page 08 Temporary peek
Encode Restorative System Procedure & Laboratory Manual
Procedure & Laboratory Manual Encode Virtual Abutment Encode Abutment Blank Encode Abutment 1/2 Milled Encode Abutment Completed ENCODE RESTORATIVE SYSTEM Simply Impressive With the introduction of the
Navigator System For CT Guided Surgery Manual
Navigator System For CT Guided Surgery Manual Instructions For Use OSSEOTITE, OSSEOTITE XP, OSSEOTITE Certain, Certain PREVAIL, OSSEOTITE NT, Encode, IOL, Miniplant, Microminiplant, GingiHue, Gold-Tite,
IMPLANT DENTISTRY EXAM BANK
IMPLANT DENTISTRY EXAM BANK 1. Define osseointegration. (4 points, 1/4 2. What are the critical components of an acceptable clinical trial? (10 points) 3. Compare the masticatory performance of individuals
platform shifting Nobel Biocare is shifting gingival beauty to the next level
platform shifting Nobel Biocare is shifting gingival beauty to the next level Nobel Biocare is shifting gingival beauty to the next level Platform Shifting is the latest in Nobel Biocare s offering of
Regular C/X Prosthetics. Prosthetics
Regular C/X Prosthetics /X C/ Prosthetics ANKYLOS C/X Prosthetics For more than 20 years, the ANKYLOS system developed by Prof. Dr. G.-H. Nentwig and Dr. Dipl.-Ing. Walter Moser with its TissueCare Connection
Improving Esthetics with Sequential Treatment Planning and Implant-Retained Dentures
Improving Esthetics with Sequential Treatment Planning and Implant-Retained Dentures by Timothy F. Kosinski, DDS, MAGD While oral function is the primary concern for most patients, the importance of esthetics
Zimmer PSI Knee System For Use with Persona The Personalized Knee System Surgical Technique
Zimmer PSI Knee System For Use with Persona The Personalized Knee System Surgical Technique TOC Zimmer PSI Knee Surgical Technique Zimmer PSI Knee Surgical Technique Table of Contents Intra-Operative Guide
Job Ready Assessment Blueprint. Dental Assisting. Test Code: 4026 / Version: 01
Job Ready Assessment Blueprint Dental Assisting Test Code: 4026 / Version: 01 Measuring What Matters Specific Competencies and Skills Tested in this Assessment: Introduction to the Dental Assisting Profession
Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material
Healing Abutment Selection Perio Implant Part I Credit and Thanks for Lecture Material Implant Surface Characteristics!CAPT Robert Taft!CAPT Greg Waskewicz!Periodontal Residents NPDS and UMN!Machined Titanium!Tiunite!Osseotite
Straumann Dental Implant System. Fax Order Form Collection.*
Straumann Dental Implant System. Fax Order Form Collection.* * The NNC Fax Order Form is still a stand alone document due to the unique restorative portfolio Straumann Roxolid SLActive Tissue Level Implants
ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning. Orthodontics and Implantology
ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning Orthodontics and Implantology 2 All rights are reserved. Any reproduction of the present publication is prohibited in whole or in part and by
Long-term success of osseointegrated implants
Against All Odds A No Bone Solution Long-term success of osseointegrated implants depends on the length of the implants used and the quality and quantity of bone surrounding these implants. As surgical
STEP-BY-STEP INSTRUCTIONS ON THE PROSTHETIC PROCEDURES. Straumann Anatomic IPS e.max Abutment
STEP-BY-STEP INSTRUCTIONS ON THE PROSTHETIC PROCEDURES Straumann Anatomic IPS e.max Abutment The ITI (International Team for Implantology) is academic partner of Institut Straumann in the areas of research
A Comprehensive Explanation
Dental Implants A Comprehensive Explanation Overview Since the 1980s, dental implants have become more popular among dentists and patients. 1 In some clinical situations, implants may be the best treatment
DENTAL IMPLANT THERAPY
DENTAL IMPLANT THERAPY PATIENT WELCOME PACK Dr. Syed Abdullah BDS, MSc (Dental Implants) What are dental implants? In the early 1950s, a Swedish Scientist, Per-Ingvar Branemark observed that titanium metal
The LOCATOR concept. Simplicity and versatility for prosthesis fixation
The concept Simplicity and versatility for prosthesis fixation The concept Experience the freedom in prosthesis fixation Simple and secure fixation of implant-supported prostheses is essential for successful
SURGICAL MANUAL. Step By Step Techniques
SURGICAL MANUAL Step By Step Techniques TABLE OF CONTENTS PRE-SURGICAL 1 8 MEASUREMENT OF BONE.......................... 2 BONE CLASSIFICATION........................... 3 IMPLANT SIZE SELECTION.........................
GINGIVAL LEVEL implant
GL GINGIVAL LEVEL implant English version TABLE OF CONTENTS GL - GINGIVAL LEVEL IMPLANT Implant characteristics page 04 Dental implant page 05 Prosthetic components page 06 Ball attachment system page
Technique Guide. Screw Removal Set. Instruments for removing Synthes screws.
Technique Guide Screw Removal Set. Instruments for removing Synthes screws. Table of Contents Introduction Screw Removal Set 2 Surgical Technique Preoperative Planning and Preparation 6 Removal of Intact
PATIENT INFORM CONSENT for IMPLANT RESTORATION Rev 04.2012
PATIENT INFORM CONSENT for IMPLANT RESTORATION Rev 04.2012 Implant placement and restoration involves two major stages: surgical placement of the implant(s) followed by the restoration of the implant after
Zimmer Small Fragment Universal Locking System. Surgical Technique
Zimmer Small Fragment Universal Locking System Surgical Technique Zimmer Small Fragment Universal Locking System 1 Zimmer Small Fragment Universal Locking System Surgical Technique Table of Contents Introduction
Restorative Guidelines
Restorative Guidelines Contents Restorative Guidelines 4.1 Neoss Implant System 4.2 4.2 Esthetiline Solution 4.3 4.3 Provisional Abutments 4.8 4.4 Impression Techniques Implant Level 4.12 4.5 NeoLink
Ridge Reconstruction for Implant Placement
Volume 1, No. 5 July/August 2009 The Journal of Implant & Advanced Clinical Dentistry Ridge Reconstruction for Implant Placement 2 Hours of CE Credit Oral Implications of Cancer Chemotherapy Immediate
Procedures & Products Powered by Procera. All-on-4
Procedures & Products Powered by Procera All-on-4 First from Nobel Biocare. NobelPerfect, (NP, RP, WP), NobelDirect (NP, RP, WP), Brånemark System, NobelReplace and NobelSpeedy Implants. A complete assortment
Implant Replacement of the Maxillary Central Incisor Utilizing a Modified Ceramic Abutment (Thommen SPI ART) and Ceramic Restoration
Implant Replacement of the Maxillary Central Incisor Utilizing a Modified Ceramic Abutment (Thommen SPI ART) and Ceramic Restoration ROBERT SCHNEIDER, DDS, MS* ABSTRACT The prosthetic restoration of a
Eastman Dental Hospital. Dental implants - general information for patients. Department of Restorative Dentistry
Eastman Dental Hospital Dental implants - general information for patients Department of Restorative Dentistry First published: January 2004 Last review date: March 2014 Next review date: March 2016 Leafl
STEP-BY-STEP INSTRUCTIONS ON TEMPORARY ABUTMENTS. Straumann Temporary Abutments, VITA CAD-Temp
STEP-BY-STEP INSTRUCTIONS ON TEMPORARY ABUTMENTS Straumann Temporary Abutments, VITA CAD-Temp The ITI (International Team for Implantology) is academic partner of Institut Straumann AG in the areas of
Implant Systems MAX LOGIC MONO. Today s name in tomorrow s implant technology
Implant Systems MAX LOGIC MONO Today s name in tomorrow s implant technology Implant Systems Etgar Medical Implant Systems brings to the market an entire spectrum of dental implant products based on its
INTRODUCTION TO OSTEO-TI IMPLANTS; A NURSES MANUAL
INTRODUCTION TO OSTEO-TI IMPLANTS; A NURSES MANUAL Osteo-Ti INTRODUCTION This handbook has been designed to give you a basic understanding of dental implants and related surgery procedures. As with all
RESTORE. Implant System Surgical Manual
RESTORE Implant System Surgical Manual table of contents Pre-Surgical Considerations Introduction 2 Indications 2 Contraindications 2-3 Warnings 3 Adverse Reactions 3 Sterilization 4 Instrumentation 5
STEP BY STEP INSTRUCTIONS ON TEMPORARY ABUTMENTS. Straumann Temporary Abutments, VITA CAD-Temp
STEP BY STEP INSTRUCTIONS ON TEMPORARY ABUTMENTS Straumann Temporary Abutments, VITA CAD-Temp The ITI (International Team for Implantology) is academic partner of Institut Straumann AG in the areas of
A New Beginning with Dental Implants. A Guide to Understanding Your Treatment Options
A New Beginning with Dental Implants A Guide to Understanding Your Treatment Options Why Should I Replace My Missing Teeth? Usually, when you lose a tooth, it is best for your oral health to have it replaced.
Abutment fracture in a bridge supported by natural teeth and implants
Abutment fracture in a bridge supported by natural teeth and implants Authors_Dr Gregory-George Zafiropoulos, Dr Giorgio Deli & Dr Rainer Valentin, Germany/Italy _Introduction Implant treatment has evolved
Prosthetic Manual. Implant System. Contact Us: Smarter thinking. Simpler design. Phone: 866-902-9272 781-328-3490. Fax: 866-903-9272 781-328-3400
Contact Us: Implant System Phone: 866-902-9272 781-328-3490 Fax: 866-903-9272 781-328-3400 Prosthetic Manual Mail: Keystone Dental, Inc. 144 Middlesex Turnpike Burlington, MA 01803 USA www.keystonedental.com
LOCATOR Implant Attachment System for XiVE, FRIALIT and ANKYLOS Implants
Technique Manual LOCATOR Implant Attachment System for XiVE, FRIALIT and ANKYLOS Implants Status of information: L8002-TM REV. F - DF4515 Rev. 000 09/2010 1 CLASSIFICATION Universal hinge, resilient attachment
BICON DENTAL IMPLANTS
BICON DENTAL IMPLANTS The Bicon Dental Implant System, since 1985, has offered discerning dentists the ability to provide secure implant restorations that look, feel, and function like natural teeth. With
SURGICAL PROCEDURES GUIDE
SURGICAL PROCEDURES GUIDE Endopore Contents Contents 3 Introduction 4 Product Descriptions Endopore implants with internal connection for one-stage surgery Endopore implants with internal hex for one-stage
Dental Implant Restoration
Dental Implant Restoration Principles and Procedures Stuart H. Jacobs Brian C. O Connell London, Berlin, Chicago, Tokyo, Barcelona, Beijing, Istanbul, Milan, Moscow, New Delhi, Paris, Prague, São Paulo,
IMPLANTS IN FOCUS. Endosseous dental implant restorations PLANNING FOR IMPLANT RESTORATIONS
IMPLANTS IN FOCUS PLANNING FOR IMPLANT RESTORATIONS Replacing a missing maxillary central incisor with a dental implant can be the most demanding restoration in dentistry, so it s important to consider
March 20, 2015. Dear Mr. Chen:
DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002 March 20, 2015 Biomate
Lentur Cable System. Surgical Technique
Lentur Cable System Surgical Technique Contents Introduction... Page 1 System Design Features And Benefits... Page 2 Implants... Page 3 Instrumentation... Page 4 Surgical Technique... Page 5 Single Cable...
How To Use A Cresco
Cresco Prosthetic and laboratory procedures 2 Discover the freedom of precision When it comes to screw-retained implant bridges, Cresco is proving to be the preferred choice of both dentists and dental